You are on page 1of 8

The International Journal of Artificial Organs / Vol. 23 / no. 10, 2000 / pp.

47-51

Artificial Heart and Cardiac Assist Devices

Modeling and control of a magnetic bearing system for


the magnetically suspended centrifugal blood pump
H. KIM 1 AND H.C. KIM 2
1
Interdisciplinary Program, Biomedical Engineering Major, Graduate School
2
Department of Biomedical Engineering, College of Medicine, and Institute of Medical and Biological Engineering, MCR,
Seoul National University, Seoul - Korea

ABSTRACT: We have developed a centrifugal blood pump with a magnetically suspended impeller to
reduce the hemolysis level for long-term use. The main advance we have made was simplifying the
traditional 5 axes controlled magnetic bearing system, since the total device should ultimately be small
enough to implant into a human body. A simplified magnetic bearing system with only 3 actively
controlled axes was designed to levitate the impeller. The state equations of the magnetic bearing
plant were established using Lagrange’s methods, and using the results, a linear quadratic optimal
controller was designed to regulate the position of the impeller. Flow rate and differential pressure of
5L/min and 300mmHg, respectively at a motor speed of 2750 rpm was achieved from a mock
circulation setup. These results show satisfactory performance as a cardiopulmonary bypass device.
The lowering of pumping efficiency as the clearance between the impeller and pump housing
increased was negligible. Thus for power efficiency, the smallest gap not leading to a hemolysis level
rise can be determined as a feasible operation clearance.

KEY WORDS: Centrifugal blood pump, Magnetic bearing system, Linear quadratic optimal controller

INTRODUCTION As a class of rotary pumps, the centrifugal blood pump


has now become popular and widely used in clinical
Mechanical circulatory support system are widely used application. This is because, in addition to the advantages
as cardiopulmonary bypass devices and implantable of the rotary pumps, they can operagte at an even lower
ventricular assist devices. Recently, research on long-term rotational speed than the axial pump. However, conventional
use of mechanical circulatory support systems, focusing centrifugal blood pumps still have a comparatively high
on implantable circulatory assist applications are in hemolysis level not suitable for long-term use due to heat
progress by several groups. Rotary blood pumps are and friction generated from the mechanical bearings or
attracting more and more interest by researchers as a next pivots. This detriment to the blood cells limits centrifugal
generation implantable circulatory device. In the early blood pumps to short-term use only. Recent studies show
stages of development, arguments against their long-term that active magnetic bearings can reduce the hemolysis
usage were wide spread due to the non-pulsatile output level in rotary blood pumps by eliminating this heat and
flow and high risk of hemolysis. However, through friction (2-5). This provides a significant improvement
continuous research, rotar y pumps have proven to towards long-term usage of centrifugal blood pumps.
overcome these drawbacks and have the advantages of We have developed a prototype magnetically suspended
compact size, less hemolysis, less thrombosis, low power centrifugal blood pump as a cardiopulmonary bypass
consumption, easy control and low cost compared to the device or a ventricular assist device. Since the ultimate
pulsatile blood pump. A series of studies verifying the goal is to make a small implantable device, we set our
suitability of nonpulsatile blood pumps have also been sights on simplifying the active magnetic bearing system. In
performed by Jett et al (1). traditional active magnetic bearing systems, shown in

©
by Wichtig Editore, 2000 0391-3988/000-09 $04.50/0
Centrifugal blood pump with a magnetically suspended impeller to reduce hemolysis level

a) b)

Fig. 1 - (a) Axis identification of the traditional 5 axes controlled magnetic bearing. The 5 axes of x1-x2, x3-x4, y1-y2, y3-y4, z1-z2 must have
their own dedicated sensors and actuators that make the total system bulky. (b) In our design we use the permanent magnet bearing to
regulate the axes x5-x6 and y5-y6. This strucutre has the advantage over (a) since only 3 axes (z, Θx, Θy) require the active magnetic bearing,
making the system small enough to implant into a human body.

Figure 1-(a), with five actively controlled axes the pump of an active magnetic bearing system, a magnetically
structure is too bulky because each control axis must have suspended impeller and a magnetic coupling with a
its own devoted sensors, electromagnets, amplifiers, and brushless direct current motor. The implemented motor
feedback controllers. Seeking a compromise between (FBLM5120A-A, Oriental Motors, Tokyo, Japan) has a
simple structure and precise positioning, we were able to maximum rotating speed and rated torque of 3000 rpm
realize a simplified magnetic bearing system as illustrated and 0.4 N.m, respectively. On the magnetic bearing side,
in Figure 1-(b), applicable to areas like implantable the three proximity sensors and three in house fabricated
centrifugal blood pumps, where the size of the system is electromagnets with the geometry shown in Figure 3 face
critical. With this simplified magnetic bearing system, which the impeller.
consists of an active magnetic bearing at 3 axes of motion A ferrite stainless steel plate on the impeller, face to
and permanent magnetic bearing at 3 axes of motion, the face with the magnetic bearing head, becomes the target
impeller could hover at a fixed position with a variation of for the magnetic bear ing system. This magnetic
less than 10 µm. Since a low rotating speed is required to suspension cancels the attracting force of the magnetic
reduce hemolysis, a pump showing high efficiency in coupling and regulates the impeller position. Twenty four
output flow and pressure head should be another design neodymium-iron-boron (NdFeB) permanent magnets are
target. We adopted the impeller structure from our former attached, one set on the motor shaft and the other set on
studies to develop the prototype centrifugal pump (6,7). A the impeller. The polarity of the driving permanent
digital signal processor based controller was installed to magnets must be alternated in the manner fo N-S-N-S... to
implement a linear quadratic control algorithm to suspend transfer the rotating force from the motor to the impeller.
the impeller. The digital controller enables sophisticated This magnetic coupling between the impeller and the
modern control algorithms with flexibility in upgrading. motor shaft also acts as a passive radial bearing,
regulating 3 degrees of freedom of the impeller.
The intended design point was a pressure head and
METHODS flow rate of 300mmHg and 5L/min, respectively at a
motor speed of 3000 rpm. With this target operating
Mechanical Structure point, calculation of the blade angles and volute specs
was done based on former design studies (6,7). The
The schematic diagram of the pump head with the weight of the pump head excluding the motor was 380 g
magnetic bearing system is shown in Figure 2. It consists and the dimensions of the pump housing were 125 mm

2
H. Kim and H.C. Kim

Fig. 2 - Schematic diagram of


the prototype magnetically
suspended centrifugal blood
pump.

and 41.5 mm in diameter and height, respectively. The


inlet and outlet nipple dimensions were designed to fit
the standard 3/8” connection tubes used in cardiac
surger y. A data acquisition board (EZAD-512, Dail
Information, Seoul, Korea) in the pump controller was
connected to a laptop computer via an RS-232
communication link. A monitoring program coded with
Labview (National Instruments, Austin, TX) on the laptop
displays the displacement, power consumption and
additional data from the pump system. A mock circulation
setup in Figure 4 was established to evaluate the pump
performances.

Fig. 3 - Geometry of the 3 sensors and 3 electromagnets of the


Magnetic Bearing System Implementation magnetic bearing system.

The active magnetic bearing system consists of three


proximity sensors, a digital signal processor, power
amplifiers and three electromagnets. Eddy current type
proximity sensors (E2CA-X1R5A, Omron, Tokyo, Japan) and
signal amplifiers (E2CA-AL4C, Omron, Tokyo, Japan) were
used to detect the distance from the pump housing to the
impeller. An additional sensor signal conditioning circuit was
designed to reduce the high frequency noise and adjust the
gain and offset, providing the signal with a reliable resolution
and a desirable dynamic range for data acquisition. This
amplification circuit also acts as an anti-aliasing filter before
the analog-to-digital conversion. The overall resolution,
sensitivity and cut off frequency were 1.8µm, 20mV/µm and
3kHz, respectively. For high-speed sampling and
calculations, a digital signal processor (TMS320C31, Texas Fig. 4 - Mock circulation test setup of the developed pump system.

3
Centrifugal blood pump with a magnetically suspended impeller to reduce hemolysis level

Fig. 5 - Block diagram of the


magnetic bearing plant and
controller.

Instruments, Dallas, TX) based control unit was installed (8). digital controller.
The digital signal processor reads the displacement from the Assuming the impeller as a rigid body, the total
proximity sensors through a 12 bit resolution analog-to- magnetic bearing system requires that all six degrees of
digital converter (MAX122, Maxim, Sunnyvale, CA) at a freedom (x, y, z, Θx, Θy, Θz) in Figure 6 to be controlled.
sampling rate of 10 kHz. The digitized displacement data Since the motor, which is magnetically coupled, controls
are lowpass filtered by an 80-tap finite impulse response the rotation Θz and this permanent magnetic bearing also
(FIR) digital filter. For maximum calculation speed, all regulates the radial coordinates x, y, the active magnetic
filtering and linear quadratic control algorithms were coded bearing system’s duty is to control the remaining three
with assembly language using application specific degrees of freedom z, Θx, Θy. Forces acting on the impeller
instructions such as MPYF,ADDF or STF,STF. Using these besides these three axes (i.e. shear stress) have not been
parallel instructions we could i) multiply and add or ii) store 2 included for simplicity.
floating point operands in a single cycle, taking full As a first step in deriving the equation of motion, we
advantage of the digital signal processor’s micro-parallelism define the vector notations of the generalized coordiantes
(9). The electromagnet coil current calculated by the digital in Figure 6 as:
control algorithm is sent to the current amplifiers via a 12-bit
resolution digital-to-analog converter (AD664, Analog [1]
Device, Norwood, MA). A current amplifier was designed
using a power operational amplifier (OPA548, Burr-Brown, A preliminary equation of
Tucson, AZ) to convert the voltage from the digital-to-analog
conver ter to a corresponding driving current. Each
electromagnet was organized with 287 turns of enameled [2]
copper wire around a ferrite El lamination core.

System Modeling and Controller Design

To levitate the impeller with the pump structure as


shown in Figure 5, a centralized controller would be with
desirable for sufficient stability and robustness (10). m: impeller mass 6.26x10–2 kg
Description of the pump system verifying the system I: impeller inertia 1.6172x10–5 kg·m2
parameters must be made before designing this optimal Z1, Z2, Z3: generalized force [N]

4
H. Kim and H.C. Kim

Fig. 6 - Cross-sectional view of


the magnetic bearing facing
the impeller. Coordinates are
defined for system modeling.

can be constructed applying the Lagrange equation, and coil currents ia, ib, ic (10).
ignoring the gyroscopic effects. The generalized forces,
which are the vector sum of the acting forces (including
the electromagnetic forces and load forces), can be
expressed as:

[3] [4]

Here, ks, ki are the force-displacement factor and force-


current factor, respectively. Force, displacement and
current are all regarded as incremental quantities. The
dimension of k s and k i were 1600 N/m and 21.6 N/A
respectively, and were measured by a separate setup. In
real control applications, it is convenient to express the
with equations using the sensor coordinates. After transforming
I: distance from the center to the electromagnets in Figure 6 [4] into the sensor coordinates from the geometrical
1.9x10–2 m constraints, we derive the following ultimate equation of
fa, fb, fc: force applied to the impeller at the 3 electromagnet motion from which we can design an optimal controller.
position in Figure 6 [N].
Based on the literature on handling electromagnetics or [5]
magnetic bearings, force activating within a magnetic
circuit can be expressed as a linear function of the where vectors, matrices and system parameters are
clearances za, zb, zc between the impeller and the housing, defined as:

5
Centrifugal blood pump with a magnetically suspended impeller to reduce hemolysis level

Transformation of this formula into the digital domain


with a sampling period of 100 µs was made since we are
designing a digital controller. Finally, using the difference
of the displacements instead of the actual derivative, a
simple centralized linear quadratic optimal controller, as
in Figure 7, was designed (12). The optimal state
feedback gain K minimizing the cost function J = *`0 (u’
Ru + x B + x B Q xB ) dt was found using Matlab
— — —
(Mathwor ks, Natick, MA) from the der ived state
equations in [5]. u is the system input vector and xB is

the state vector. The respective weights R and Q were
fixed as unit matrices.

[6]
RESULTS

Controller Performance

The performance of the optima linear quadratic (LQ)


controller was ver ified and compared with the
propor tional der ivative (PD) controller. The step
response without motor rotoation and the steady state
response at a motor speed of 3000 rpm are shown in
Figure 8. The rising time (<5 ms) and overshoot (<10
µm) during the step response were similar in both the LQ
and PD cases. But at high motor speed, the LQ controller
showed better perfor mance in less peak-to-peak
p: distance from the center to the sensor in Figure 6 fluctuation (45 µm) than the PD controller (60 µm). Since
2.3x10–2 m the centralized LQ controller also considers the coupling
ZBa, ZBd, ZBc: clearance between the impeller and housing gains between the channels, it would be more robust
at the 3 sensor positions in Figure 6. [m] under varying conditions.
Fig. 7 - Multiple input multiple
output linear quadratic
controller for the magnetic
bearing system. r, x B , u are
the reference input —vector,
feedback input vector and state
vector of the system, respecti-
vely.

6
H. Kim and H.C. Kim

Fig. 8 - The step response and


steady state response of the
linear quadratic (LQ) and
proportional derivative (PD)
controller are compared. The
LQ controller shows less
fluctuation in steady state
response at a motor speed of
3000 rpm.

Output Characteristics and Power Consumption

T h e f l ow ra t e a n d d i f fe r e n t i a l p r e s s u r e o f t h e
prototype pump with the magnetic suspension showed
higher hydraulic parameters than our mechanical
bearing pump developed in 1998 (7). The principal
reason for this improvement is that the clearance
between the impeller and the pump housing in the
magnetic bearing version is less than in the mechanical
b e a r i n g ve r s i o n . T h u s, t h e p r e c i s e p o s i t i o n i n g
characteristics of the magnetic bearing system can also
p r ov i d e b e t t e r p u m p i n g e f f i c i e n c y. Fo r d i f fe r e n t Fig. 9 - Pressure and flow curves for different rotating speeds.

Fig. 10 - Pressure and flow curves for different clearances between the Fig. 11 - Current consumption of the magnetic bearing system for dif-
impeller and the pump housing at the active magnetic bearing side. ferent clearances. The supply voltage for the current amplifier was 12 V.

7
Centrifugal blood pump with a magnetically suspended impeller to reduce hemolysis level

rotational speeds, the flow-pressure curves are shown ACKNOWLEDGEMENTS


i n F i g u r e 9 . T h e d e s i ra bl e o p e ra t i o n p o i n t fo r a
cardiopulmonary bypass device (5L/min, 300mmHg) The authors wish to acknowledge the financial support given
was achieved at a motor speed of 2750 rpm. Flow and by the Korea Research Foundation in the program year of 1998.
pressure relations in Figure 10 show little deviation as
the nominal clearance was changed from 200 µm to
100 µm. Since this decline is negligible, there is no Reprint requests to:
Hee Chan Kim, Ph.D.
need to consider the lowering of pumping efficiency due
Department of Biomedical Engineering
to the change of nominal clearance in determining the College of Medicine, Seoul National University
optimal clearance. 28 Yongun-Dong, Jongno-Gu
Seoul, 110-744, Korea
The current consumption of the magnetic bearing e-mail: hckim@snu.ac.kr
system almost halved as the nominal clearance changed
from 200 µm to 100 µm (Fig. 11). As the rotational speed
increased, the input current slightly decreased due to
the buoyancy. It is advantageous to operate the pump at
a smaller gap to consume less power. But if this
clearance is reduced to a value comparable to the size
of the blood cells, the hemolysis level will increase (7). REFERENCES
Conclusively, the optimal gap must be the smallest
01. Jett GK. Physiology of nonpulsatile circulation: acute versus
possible one that wont’t result in higher hemolysis. The
chronic support. ASAIO J 1999; 45: 119-22.
developed pump showed a power consumption of 5W at 02. Nojiri C, Kijima TR, Maekawa J, et al. More than 1 year
a clearance of 100 µm and rotational speed of 2700 continuous operation of a centr ifugal pump with a
rpm. magnetically suspended impeller. ASAIO J 1997; 43: M548-
52.
03. Hilton EF, Allaire PE, Baloh MJ, et al. Magnetic suspension
DISCUSSION controls for a new continuous flow ventricular assist device.
ASAIO J 1997; 43: M598-603.
We h ave d e s i g ned a prot ot ype magne ti ca l l y 04. Akamatsu T, Nakazeki T, Itoh H. Centrifugal blood pump with
suspended centr ifugal blood pump showing high a magnetically suspended impeller. Artif Organs 1992; 16:
305-8.
differential pressure and flow rate at a comparatively low
05. Malson EH, Bearnson GB, Allaire PE, et al. Feedback control
rotation speed. The magnetic bearing system with only 3
applicaitons in artificial hearts. IEEE Control Systems
actively controlled axes could provide sufficient stability
December 1998; 26-34.
and robustness during in vitro tests. A digital LQ control 06. Kim WG, Chung CH, Yang WS, et al. Development of a cen-
algorithm was implemented in a digital signal processor trifugal pump with thck blodes. Artif Organs 2000: 24: 160-4..
to regulate the position of the magnetically suspended 07. Yang WS. Design methods. In: Design and performance test
impeller. Less fluctuation was observed than with the PD of centrifugal blood pump. MS Dissertation: Seoul National
controller, when the impeller was rotating at a high University 1998; 4-13.
speed. Laboratory tests to determine the optimal gap 08. Bleuler H, Guler C, Herzog R, et al. Application of digital
between the impeller and the pump housing were signal processors for industrial magnetic bearings. IEEE
performed considering power consumption, pressure Trans Control Systems Technology 1994; 2: 280-9.
and output flow. Since the decline of pressure and flow 09. Parallel-Operations Instructions. In: TMS320C3X User’s
Guide. Dallas: Texas Instruments, 1997; 13.6-13.8.
rate was negligible as the clearance between the
10. Schweitzer R, Bleuler H, Traxler A. Magnetic force. In: Active
impeller and pump housing varied, the least clearance
magnetic bearingo. Zürich: vdf Hochschulverlag AG, 1994; 63-7.
not causing a rise in hemolysis level should become the
11. Phillips CL, Nagle HT. Discrete state equations. In: Digital
operating clearance for maximum power efficiency. control system analysis and design. New Jersey: Prentice-
Further in vitro and in vivo hemolysis test results should Hall, 1995; 156-9.
be considered. 12. Anderson BDO, Moore JB. The standard regulator problem-1.
In: Optimal Control. New Jersey: Prentice-Hall, 1989; 7-34.

You might also like