You are on page 1of 400

FUJIFILM DR

FUJIFILM DR-ID 300


Console : DR-ID 300CL
Service Manual

Document No. 010-305-19A


First Edition : Aug. 20, 2009
19th Edition : Jun. 29, 2018

Printed in Japan
Export Restrictions Applying to this Equipment
Japanese export restrictions and foreign exchange regulations apply to
this equipment. Therefore, government approval obtained through due
procedure is required in order to export or otherwise remove such
equipment from Japan.
Note that the same restrictions and regulations apply not only to the
main unit but also to after-sale parts, service manuals and service
training (for exporting or transferring technology) when such items are
dispatched or shipped to foreign countries and when service training is
given for trainees from foreign countries.

Caution : RX only in the United States. (Federal law restricts this device to sale by
or on the order of a physician.)

Precautions in Using the Service Manual


This manual should not be used by anyone other than the servicing staff
members qualified through the training specified by our company. Since
this manual contains confidential information including engineering know-
how related to product development, it is prohibited to distribute the manual
to users or to allow them free access to the manual without obtaining prior
permission from FUJIFILM Corporation.

This manual is intended to assist the service person in the safe and efficient
servicing of the equipment described.
The equipment must be used in accordance with the procedures contained in this
manual and must not be used for purposes other than those which are described
herein.

The equipment should only be used by persons having recognized qualifications


and, if relevant, having adequate training on this equipment, especially regarding
protective measure such as laser radiation protection.

It is the responsibility of the user to ensure that existing legal regulations regarding
installation of the building with respect to the equipment are observed. Moreover
the legal regulations regarding the operation of the equipment are to be observed.

Incorrect operation, or failure of the user to maintain the equipment in accordance


with the schedule of maintenance, relieves the manufacturer or his agent from all
responsibility for consequent non-compliance, damage, injury, defects and/or other
malfunction.

FDR and FCR are trademarks or registered trademarks of FUJIFILM Corporation.

Copyright © 2009-2018 by FUJIFILM Corporation. All rights reserved.

No part of this publication may be reproduced, stored in a retrieval system, or transmitted


in any form or by means, electronic, mechanical, photocopying, recording or otherwise,
without the prior written permission of FUJIFILM Corporation.
010-305-19E
DR-ID 300CL Service Manual 06.29.2018 FM9522
WARNING
This manual is written for International distribution (including
distribution in the U.S.) and therefore may contain references to

Important Notice Concerning this Manual


mammography related options.
Mammography functionality for this product is not
commercially available in the US, pending FDA Approval.
This functionality must not be activated on US models, the only
breast related functionality available in the U.S. is the Specimen
Imaging Application.

010-305-07E
DR-ID 300CL Service Manual 09.12.2012 FM6055
BLANK PAGE

010-305-07E
DR-ID 300CL Service Manual 09.12.2012 FM6055
How to Use This Manual ■ Warning and Cautionary Information
Precautions during the service work are indicated by the following terms
and symbols.
■ Notes on the Use of the Service Manual
1. This document is intended to be read by the service WARNING
personnel of FUJIFILM Corporation or authorized agent who
is commissioned by FUJIFILM Corporation to perform Serious injuries or death may result if this precaution is not
servicing tasks. observed.
Damage of the unit and customer database may also occur.
Under these circumstances, no one is allowed to perform the
following actions without the prior written permission from
FUJIFILM Corporation CAUTION
Duplicating or reproducing this document in whole or in part Mild or moderate injuries may result if this precaution is not
Disclosing this document, in whole or in part, to a person other than observed.
the authorized service personnel Serious malfunction (which may be unrepairable or difficult to
Lending or transferring this document, in whole or in part, to a third
repair) may also occur.
party
➤ IMPORTANT
2. Due care must be exercised when handling this document
because it contains product internal structure descriptions Abnormal operations (easy to repair) may result if this precaution is not
and other confidential information. observed.
* When the contents of this document need be changed, the
information about such changes will be conveyed using NOTE
“Engineering Change Notice (ECN)”.
Indicates items which require particular attention.
The “Engineering Change Notice (ECN)” must also be handled
using the same care as for this document.
➥ REFERENCE
3. Due to a product development schedule, the product
specifications and available options/variations prevailing at Explanation of related items and terms.
the time of the release of this document may differ from the
descriptions given in this document. In such an instance, a
Service Engineering Information will be issued as needed to
furnish latest information about inapplicable specifications
and components.
4. If you have any doubt about the contents of this document
or find errors, contact your local FUJIFILM representative.
010-305-01E
08.20.2009 FM5581
DR-ID 300CL Service Manual 1
■ Servicing Instruments and Tools That Require ■ Software version notation
Inspection/Calibration Software versions are shown according to the following version
The machine should be installed and serviced by use of servicing notation.
instruments and tools that have been regularly inspected and calibrated V1. 0. 0003
as appropriate. X Y Z
If the machine were serviced using servicing instruments and tools that
have not been inspected and calibrated, proper performance of the X. Major version
machine could not be guaranteed. The number increasing at every function addition. 1 to 99.
Servicing instruments and tools that require inspection/calibration are However, this number does not always be a continuous number at
as listed below. every release.
The inspection/calibration should be planned and performed in
Y. Minor version (sometimes called serviced package)
accordance with the specifications and instruction manuals of the
The number increasing at every error corrections and function
applicable servicing instruments and tools.
addition. 0 to 99.
● Instruments and tools that require inspection/calibration However, this number does not always be a continuous number at
every release.
Name Inspection Calibration
Z. Hot-fix status number
Digital multimeter
The number indicating the status of Hot Fix (hereinafter referred to as
HF). 0001 to 7777.
■ Screen Displays According to Different OSs and Versions The sum of all of the HF numbers is displayed as this number. If no
This manual uses screen displays of Windows Vista or Windows 7 or HF included, this Z part is not written.
Windows 10. Each digit of HF is formed by only 3 elements, 1, 2 and 4, and a
When setting procedures to be performed and displayed windows are maximum of 12 combinations of elements can be indicated by these
4 digits. You can learn which elements are included from the HF
different between Windows Vista, Windows 7 and Windows 10, follow
status number.
the procedure specific to the used OS to perform necessary settings.
Numbers represent unique combinations as follows: 7 = 1+2+4,
6 = 2+4, 5 = 1+4, 4 = 4 only, 3 = 1+2, 2 = 2 only, 1 = 1 only, 0 = none.
For example, the Z part “1234” indicates that 5 HF, HF1000, HF0200,
HF0010, HF0020, and HF0004 among 12 HF have been installed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 2
Contents (DR-ID 300CL Service Manual)
Contents (DR-ID 300CL Service Manual)
Safety Precautions
1. Safety Precautions ...................................................... 1
2. Electromagnetic Compatibility (EMC) ....................... 5
2.1 Electromagnetic Compatibility (EMC) Related Standard ... 5
2.2 Further information for IEC 60601-1-2: 2001+A1:2004... 6
3. Caution for Handling Personal Information .............. 10
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual 3
Contents (DR-ID 300CL Service Manual)
Product Specifications
1. Equipment Specifications .......................................... 1
1.1 Product Code ......................................................... 1
1.2 Optional Components ............................................. 2
1.3 Dimensions/Weight ................................................. 5
1.3.1 PC ................................................................ 5
1.3.2 Monitor .......................................................... 5
1.4 Environmental Conditions ....................................... 6
1.4.1 PC ................................................................ 6
1.4.2 Monitor .......................................................... 6
1.5 Electrical Specifications .......................................... 8
1.5.1 PC ................................................................ 8
1.5.2 Monitor .......................................................... 9
1.6 Applicable Standards .............................................. 10
1.7 Network Cable ........................................................ 10
1.8 Equipment Disposal ................................................ 10
010-305-16E
07.31.2016 FM9434
DR-ID 300CL Service Manual 4
Contents (DR-ID 300CL Service Manual)
MT: Machine Troubleshooting
Appendix A DR-ID 300CL Error
Message Table (V4.0 or Earlier) .... Appx MT A-1
Appendix B DR-ID 300CL Error
Message Table (V5.0 or Later) ...... Appx MT B-1
Appendix C Cooper Application
Error Message Table...................... Appx MT C-1
Appendix D Precise Enlargement Function
Error Message Table...................... Appx MT D-1
Appendix E Free Layout Print
Error Message Table...................... Appx MT E-1
Appendix F Referral Viewing Function
Error Message Table...................... Appx MT F-1
Appendix G Network Image Search Function
Error Message Table...................... Appx MT G-1
Appendix H QA ROI Measurement
Error Message Table...................... Appx MT H-1
Appendix I QA Measurement Function
Error Message Table...................... Appx MT I-1
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 5
Contents (DR-ID 300CL Service Manual)
SP: Service Parts
1. Service Parts List ........................................................ SP-1
1.1 How to Use Service Parts List ................................. SP-1
1.2 Service Parts List .................................................... SP-3
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 6
Contents (DR-ID 300CL Service Manual)
PM: Preventive Maintenance
1. Preventive Maintenance Items ................................... PM-1
3. Safety ........................................................................... PM-2
3.1 Service Voltage ...................................................... PM-2
3.2 Power Plug ............................................................. PM-2
4. Hardware Cleaning...................................................... PM-3
4.1 PC Internal Cleaning ............................................... PM-3
4.2 Keyboard and Mouse Cleaning................................ PM-3
4.3 LCD Cleaning ......................................................... PM-4
4.4 DVD/CD-RW Combo Drive Cleaning ....................... PM-4
6. Functional Checks ...................................................... PM-5
6.1 Confirmation of Image Input to Storage onto HDD .... PM-5
6.2 Image Output to Film .............................................. PM-5
6.3 Image Transfer ....................................................... PM-5
7. Backup of Setup Files and
Verification of Date and Time ..................................... PM-6
7.1 Backup of Setup Files and Log Files ........................ PM-6
7.2 Verification of Date and Time .................................. PM-6
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 7
Contents (DR-ID 300CL Service Manual)
IN: Installation
1. DR-ID 300CL Installation Conditions ......................... IN-1 4. DR-ID 300CL Software Setup ..................................... IN-13
1.1 Dimensions and Weight .......................................... IN-1 4.1 DR-ID 300CL-AP Startup ........................................ IN-14
1.1.1 PC ................................................................ IN-1 4.1.1 OS Startup ..................................................... IN-15
1.1.2 Monitor .......................................................... IN-1 4.1.2 OS Setups ..................................................... IN-16
1.2 Temperature and Humidity ...................................... IN-2 4.1.3 Installing and Setting the Touch Panel Driver
(Only on Desktop PC)
1.2.1 PC ................................................................ IN-2 [Skip this procedure if the 17" monitor with
1.2.2 Monitor .......................................................... IN-2 a touch panel is provided.]................................ IN-17
4.1.4 Touch Panel Monitor Setups/
1.3 Places of Installation ............................................... IN-3 Operational Verification (Only on Laptop PC) ...... IN-28
1.4 Power Requirements .............................................. IN-3 4.1.5 Setup for Addition of Descriptive Characters to Icons
[Applicable only in Japan] ................................. IN-29
2. Installation Procedure Overview ............................... IN-4 4.1.6 Installing the DR-ID 300CL Standard Kit
(V5.0 or Later) ................................................ IN-30
3. DR-ID 300CL Hardware Setup .................................... IN-5 4.1.7 Option Key Installation ..................................... IN-32
3.1 Unloading and Unpacking ....................................... IN-5 4.1.8 AP Key Installation .......................................... IN-35
3.2 Checking the Component Items ............................... IN-6 4.1.9 DR-ID 300CL Startup Verification ...................... IN-36
3.3 Installing the Accessories, 4.2 Settings on the DR-ID 300CL for Connection with
Optional Items, and Available Items ......................... IN-10 the FPD/Cassette-type Image Readers .................... IN-37
3.3.1 Installing the Accessories and Optional Items 4.2.1 Starting the Service Utility ................................ IN-37
(for Desktop PC) ............................................. IN-10 4.2.2 DR-ID 300CL System Information Setup ............ IN-38
3.3.2 Installing the Accessories and Available Items
4.2.3 Registering the Image Detector (V5.0 or Later).... IN-79
(for Laptop PC) ............................................... IN-11
4.2.4 Setting the Selectors ....................................... IN-80
3.4 Making a Cable Connection to
4.2.5 Optimizing the Monitor Display .......................... IN-84
the FPD/Cassette-type Image Reader ..................... IN-12
4.2.6 Changing the Menu Database........................... IN-85
3.5 Connecting to the Other Equipment ......................... IN-12
4.2.7 Exiting the Service Utility .................................. IN-87
4.2.8 Installing the MC Software ................................ IN-87
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 8
Contents (DR-ID 300CL Service Manual)
4.3 Setting Network Connection in Laptop PC ................ IN-88 4.14 Enabling the New GUI for
4.3.1 Setting the IP Address in Laptop PCs (V7.4 or Later)...................................... IN-105
Each Network Configuration ............................. IN-88 4.15 Changing the Password of
4.3.2 Enabling the Built-in Wireless Network Adaptor of the Service Utility (V8.0 or Later) ............................. IN-107
a Laptop PC ................................................... IN-92
4.16 Batch setting of VG parameter (V9.0 or Later) .......... IN-107
4.3.3 Setting the Connection Between
DR-ID 300CL (Laptop PC) and
the Image Local Network AP ............................ IN-93 5. Checkout Procedures ................................................. IN-108
4.3.4 Registration of Connection Settings of 5.1 Checking the Connection between
the In-Hospital Network (V7.3 or Later) .............. IN-94 the DR-ID 300CL and FPD/Cassette-type
Image Reader......................................................... IN-108
4.4 Setting Network Connection for
the Desktop PC with Additional NIC ......................... IN-95 5.2 Checking the Connection between
the DR-ID 300CL and Other Equipment ................... IN-108
4.5 Setting the Priority Order of the Network Binding ...... IN-96
4.5.1 For the Laptop PC (Windows 7 only)/DR-XD 1000 ... IN-96
5.3 Checking the Output Image ..................................... IN-108
4.5.2 For the Desktop PC (Windows Vista/7 only) with 5.4 Operation Check of the Network Switching Function ... IN-109
Additional NIC ................................................ IN-97
4.5.3 In Windows 10 (Laptop PC, Desktop PC shared) ... IN-98 6. Setup File Backup, Image Database Initialization and
Error Log Verification/Deletion .................................. IN-110
4.6 Procedures for Changing the DR-ID 300CL and
RU (FPD and Cassette-type Image Reader) 6.1 Backing Up the DR-ID 300CL Setup Files ................ IN-111
IP Addresses .......................................................... IN-101 6.1.1 Backing Up the System Setup Files ................... IN-111
4.7 Procedures for Changing 6.1.2 Backing Up the Network Adapter Setup Files
the DR-ID 300CL Host Name .................................. IN-101 (Only Laptop PC) ............................................ IN-111
4.8 DR-ID 300CL Setup for 6.2 Initializing Various Image Databases ....................... IN-112
Connection to Other Equipment ............................... IN-101 6.2.1 Initializing Image Databases ............................. IN-112
4.9 Barcode Reader/Magnetic Card Reader Connection ... IN-102 6.2.2 Clearing the Image Output Queue ..................... IN-113
4.10 Setting the DX Parameter ....................................... IN-102 6.2.3 Initializing the Patient Database ........................ IN-113
6.2.4 Initializing the Calibration Database
4.11 Setting the VIP Account Password........................... IN-103 (V4.0 or Earlier) .............................................. IN-114
4.12 Setting the Connection Device 6.3 Checking and Deleting the DR-ID 300CL Error Log .. IN-115
Status Icon (V7.3 or Later) ...................................... IN-104
6.4 Turning OFF the DR-ID 300CL ................................ IN-115
4.13 Setting the Images Folder Size (V7.3 or Later) ......... IN-104
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 9
Contents (DR-ID 300CL Service Manual)
Appx IN: Installation – Appendix
Appendix 1 Connecting the Barcode Reader 2. Settings Required When Using
[Option] ........................................... Appx IN1-1 the Study List Search Function Based on
the Hospital Barcode ................................... Appx IN1-22
1. Connecting
the Barcode Reader to the DR-ID 300CL .... Appx IN1-1 2.1 Setting the Search Key .......................... Appx IN1-22
1.1 For the Barcode Reader 2.2 Setting the Barcode Type ...................... Appx IN1-23
Using USB Connection .......................... Appx IN1-1
3. Operation Setting of
1.1.1 Connecting wired Barcode Reader .. Appx IN1-1 Wireless Barcode Reader ............................ Appx IN1-24
1.1.2 Connecting
Wireless Barcode Reader .............. Appx IN1-2 Appendix 2 Setting Up the Magnetic Card
1.1.3 Installing the Driver ....................... Appx IN1-3 [Option] ........................................... Appx IN2-1
1.2 For the Barcode Reader Using Bluetooth Connection 1. Installing the Bracket ................................... Appx IN2-1
(Mobile PC / DR-XD 1000 Built-in PC) .... Appx IN1-7
2. Connecting the Magnetic Card Reader to
1.2.1 Enabling the Bluetooth................... Appx IN1-7 the DR-ID 300CL ......................................... Appx IN2-2
1.2.2 Installing the Bluetooth Driver ......... Appx IN1-8
3. Setting the Magnetic Card Reader and
1.2.3 Disabling
Magnetic Card ............................................. Appx IN2-5
the Bluetooth Network Connection .. Appx IN1-9
1.2.4 Setting the OPN-3002i 4. Hospital Card Data Format Setup Example
(Factory Setting) ........................... Appx IN1-13 [Reference] .................................................. Appx IN2-9
1.2.5 Pairing Setting for PC 5. ID Card Data Format [Reference] ................ Appx IN2-13
(Including DR-XD 1000 built-in PC) and
the Barcode Reader ...................... Appx IN1-17 6. Updating the Driver for
1.2.6 Setting the Configuration Items ....... Appx IN1-18 the Magnetic Card Reader........................... Appx IN2-19
1.2.7 Verification ................................... Appx IN1-19 Appendix 3 Procedures for
1.3 Switching the Barcode Reader ............... Appx IN1-21 Changing the DR-ID 300CL and
1.3.1 Switching from RU IP Addresses ............................ Appx IN3-1
USB (Wired/Wireless) Barcode Reader to
Bluetooth Barcode Reader .............
1. Incorporating into the Network ..................... Appx IN3-1
Appx IN1-21
1.3.2 Switching from Bluetooth Barcode Reader to 2. Changing the Software Settings .................. Appx IN3-2
USB (Wired/Wireless) Barcode Reader ... Appx IN1-21
Appendix 4 Changing the RU Host Name ........ Appx IN4-1
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 10
Contents (DR-ID 300CL Service Manual)
Appendix 5 Changing Appendix 8 Upgrading the AP .......................... Appx IN8-1
the DR-ID 300CL Host Name ......... Appx IN5-1
1. Upgrading the DR-ID 300CL-AP.................. Appx IN8-1
Appendix 6 Changing 1.1 Precautions against
the DR-ID 300CL IP Address ......... Appx IN6-1 DR-ID 300CL upgrading ........................ Appx IN8-1
1. IP Version 4 ................................................. Appx IN6-1 1.2 Upgrading the DR-ID 300CL .................. Appx IN8-8
2. IP Version 6 ................................................. Appx IN6-3 2. System Migration ......................................... Appx IN8-63
2.1 System Migration to
Appendix 7 Uninstalling the AP ........................ Appx IN7-1 DR-ID 300CL V5.0 - V8.1 ...................... Appx IN8-63
1. DR-ID 300CL-AP ......................................... Appx IN7-2 2.2 System Migration to
2. SQL Server .................................................. Appx IN7-3 the DR-ID 300CL V9.0 or Later .............. Appx IN8-77
2.1 SQL Server 2005 .................................. Appx IN7-3 2.3 System Migration to
the DR-ID 300CL V12.0 or Later ............ Appx IN8-88
2.2 SQL Server 2008 .................................. Appx IN7-5
3. Upgrading the PPU-AP ................................ Appx IN8-98
2.3 SQL Server 2012 .................................. Appx IN7-8
4. Upgrade of the DR-ID 300CL-AP
3. Option Keys ................................................. Appx IN7-11 Which Accompanies
4. Console Software (Built_inConsole) Creation of Image drive ............................... Appx IN8-110
(V4.0 or Earlier) ........................................... Appx IN7-12
Appendix 9 Setup for Security Enhancement ... Appx IN9-1
5. FRIS Modules [Applicable only in Japan] .... Appx IN7-13
6. Synapse Client (Synapse Workstation) ....... Appx IN7-14
7. Precise Enlargement Software .................... Appx IN7-15
8. Free Layout Print Software .......................... Appx IN7-17
9. Referral Viewing Software ........................... Appx IN7-20
10. PPS Cache Server....................................... Appx IN7-22
11. Network Image Search Software ................. Appx IN7-23
12. QA ROI Measurement Software .................. Appx IN7-25
13. QA Measurement Software ......................... Appx IN7-27
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 11
Contents (DR-ID 300CL Service Manual)
PC: Performance Check
DR-ID 300CL Performance Checklist ............................... PC-1
DR-ID 300CL System Configuration List ......................... PC-3
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual 12
Safety Precautions-I
DR-ID 300CL Service Manual
Safety Precautions
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Safety Precautions-I
Safety Precautions-II
Control Sheet
Issue date Revision number Reason Pages affected
08/20/2009 01 New release (FM5581) All pages
04/15/2010 02 Revision for V4.0 (FM5668) 1–10
02/28/2011 03 Revision for V5.0 (FM5818) 2
12/15/2011 05 Revision for V6.1 (FM5944) 2
06/29/2018 19 Revision for V12.0 (FM9522) 3-4
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Safety Precautions-II
Safety Precautions-1
1. Safety Precautions CAUTIONS
The grounding wire must be connected.
■ Precautions to be observed during operations To avoid damages to the parts/units due to static discharge, use
a wrist strap putting one end on your wrist and the other end on
To prevent danger to humans and serious accidents, observe the the grounded section of the equipment, as illustrated below.
following.
When doing so, make sure to unplug the power connector from
the socket to avoid electrical hazards.
WARNING
Unless indicated otherwise, be sure to turn off the power switch
before maintenance work, and disconnect the power plug from Wrist strap
the outlet.
Even if the power switch is OFF, 100 to 240VAC (differs according
to machine specifications) will still be supplied to the primary
input terminal of the unit if the power cable is connected, and this
may result in possible electric hazards.
As neckties, scarves, and necklaces may get caught in the unit
during work, remove these items or take precautions against
such accidents.
Never modify the unit unless otherwise specified.
Modification of the safety unit (interlock, breaker, temperature
fuse, etc.) will cause serious problems as it is not only dangerous
00010003.EPS
but violates the related laws and regulations.
When performing settings for several software programs, ensure
that applications other than those of specified procedures must
have been shut down appropriately.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-1
Safety Precautions-2
This equipment is affixed with a label. Follow the instructions provided ■ Shock Hazard Precautions
on the label when performing necessary works.
300CL-PC 300CL-PC
● Control Unit
(DT) (MT) When installing, using, or servicing the Control Unit (hereinafter referred
to as the PC), observe the following precautions. For details, refer to
the User’s Guide, which is supplied with the PC.
Prevention of electric shock
For the PC main body and monitor, use a power cord provided with a
3-prong plug. Connect the PC main body and monitor to a power
outlet that is properly grounded. If the use of an extension cord is
required, use a three-wire type having a grounding pin.
● LCD
Identification Identification WARNING
label label
A high voltage is applied to the inside of the LCD monitor.
DXL00010002.ai
To avoid safety hazards such as electric shock hazard, never
access the interior of the LCD. The LCD monitor has no parts that
can be serviced by a service engineer.
Identification ● Power supply
label
DXCL0500Sa001.ai
WARNING
● Identification label When supplying power to system components, do not use a multi-
outlet power strip or extension cable because the capacity of such
a power strip or extension cable may be exceeded to create a fire
hazard or shock hazard.
Sample year of manufacture
MADE IN JAPAN
010-305-05E
12.15.2011 FM5944
DR-ID 300CL Service Manual Safety Precautions-2
Safety Precautions-3
■ Installation Precautions ■ Bluetooth Barcode Reader Charging Precautions
When installing, using, or servicing the PC, observe the following Do not keep the bluetooth barcode reader in a condition that the
precautions. For details, refer to the User’s Guide, which is supplied battery has been completely burn out for a long time.
with the PC. Stop charging the bluetooth barcode reader when it has already been
fully charged.
● Installation site
Do not keep the bluetooth barcode reader under a high-temperature
Exercise care so that no dust or foreign matter enters an opening in the
environment.
PC or the interior of the PC. Do not install the PC near a heating device
or other heat source. When mounting the PC on a floor, ensure that it
is positioned at least 5 cm above the surface of the floor. ■ Connection Precautions
● Precautions to observe when servicing the interior of the PC WARNING
(1) Turn off the PC, monitor, and other currently used peripheral devices. Be sure that only specified external devices are connected to the
(2) Disconnect the power cord from the power outlet (to avoid electric apparatus.
shock hazard).
(3) Open the I/O panel on the rear of the PC main body, and touch a
metal surface to discharge any static buildup on your body before ■ Isolation Precautions
accessing the interior of the PC.
(4) If it is necessary to remove, for instance, a circuit board or a WARNING
memory module, handle it without directly touching its connector or
IC pin and place it in an antistatic bag. To assure complete system isolation, do not install unspecified
accessories or options.
● Environmental conditions
Avoid installing the machine at the following places:
Places where the temperature changes drastically
Places near heaters or other heat sources
Places where the machine is likely to be wet or submerged
Places where the machine is likely to come into contact with any
corrosive gas
Dusty or dirty places. Places where the machine is subjected to
constant vibration or shock.
Places where the machine is subjected to excessive vibration or
shock. Places exposed to direct sunlight
Also, be sure that the monitor is not installed at the following places:
Places where the monitor is surrounded, such as in a rack case
Places whether the display surface is exposed to direct sunlight
Places near a transformer, another monitor, a power line, or a 0.02
Gauss or more of AC magnetic field
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Safety Precautions-3
Safety Precautions-4
■ Disinfection Precautions
WARNING
Make sure that a legal limit is not exceeded by the respiratory
concentration of an employed disinfectant containing a solvent.
Some disinfectants may pose a hazard to human health. Before
using any disinfectant, be sure to thoroughly read its
documentation.
WARNING
Do not use the following disinfectants. If you use them, the
apparatus becomes damaged so that the quality, performance, and
safety of the apparatus cannot be guaranteed.
● Chlorine disinfectants that readily corrode metals and rubber
members.
● Disinfectants that are described in their documentation as being
inapplicable to metals, plastics, rubber members, and paints.
● Formalin gases or spray-type disinfectants that may enter the
apparatus.
It is recommended that ethanol for disinfection be used as a
disinfectant. Before using a disinfectant, thoroughly read its
documentation.
■ Cleaning Precautions
CAUTION
For exterior cleaning, use a cloth moistened with neutral detergent
or ethanol. Do not use thinner, benzine, or other similar solvent
because the exterior will corrode.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Safety Precautions-4
Safety Precautions-5
2. Electromagnetic Compatibility (EMC) This Service Manual contains the names of standards applied in
IEC 60601-1-2.
The names of standards applied in EN 60601-1-2 are replaced as
follows.
2.1 Electromagnetic Compatibility (EMC)
Related Standard IEC 60601-1-2
(descriptions in this Service Manual)
Standards applied in EN 60601-1-2
As required by IEC 60601-1-2:2001+A1:2004/EN 60601-1-2:2001+A1: CISPR 11 EN 55011
2006, this equipment complies with the following internationally IEC 61000-3-2 EN 61000-3-2
recognized standard regarding Electromagnetic Compatibility (EMC). IEC 61000-3-3 EN 61000-3-3
EMC standard : IEC 60601-1-2:2001+A1:2004 IEC 61000-4-2 EN 61000-4-2
This does not guarantee that there is no harmful electromagnetic IEC 61000-4-3 EN 61000-4-3
interference under any installation environment. IEC 61000-4-4 EN 61000-4-4
This equipment can generate, use and radiate radio frequency energy.
IEC 61000-4-5 EN 61000-4-5
If the equipment is not installed and used in accordance with the
instructions, or if peripheral devices that are not complied with the EMC IEC 61000-4-6 EN 61000-4-6
standard, harmful interference may be generated under a particular IEC 61000-4-8 EN 61000-4-8
environment causing malfunction of the equipment and other devices. IEC 61000-4-11 EN 61000-4-11
If this equipment causes harmful interference to other devices, or if this
equipment is affected by interference from other devices, the user is
encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to
which the other device(s) are connected.
If the problem cannot be solved with the above measures, stop using
this equipment and consult the manufacturer or our official dealer for
help.
WARNING
Do not place devices generating electromagnetic wave near this
equipment.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-5
Safety Precautions-6
2.2 Further information for IEC 60601-1-2: Table 1
2001+A1:2004 Guidance and manufacturer’s declaration - electromagnetic emissions
The Console is intended for use in the electromagnetic environment specified
below. The customer or the user of the Console should assure that it is used in
(1) Medical electrical equipment is subject to special precautions such an environment.
regarding EMC. The equipment must be installed and put into Emissions test Compliance Electromagnetic environment - guidance
service according to the EMC information provided in Tables 1 to 4. The Console uses RF energy only for its
(2) Portable and mobile RF communications equipment can affect RF emissions internal function. Therefore, its RF emissions
Group 1
CISPR 11 are very low and are not likely to cause any
medical electrical equipment. interference in nearby electronic equipment.
(3) Information regarding the cable affecting EMC is as follows. RF emissions
Class A
CISPR 11 The Console is suitable for use in all
Name Maximum Length General Specification
Harmonic emissions establishments other than domestic and
Cat5e or more, UTP type and straight Class D
Network Cable 100m IEC 61000-3-2 those directly connected to the public low-
cable Voltage fluctuations/ voltage power supply network that supplies
Does not buildings used for domestic purposes.
Power Cable Depends on the cable length of a personal computer. flicker emissions
comply
IEC 61000-3-3
(4) The use of accessories, transducers and cables other than those
specified, with the exception of transducers and cables sold by
FUJIFILM Corporation as replacement parts for internal
components, may result in increased emissions or decreased
immunity of the equipment.
(5) The equipment should not be used adjacent to or stacked with other
equipment.
If adjacent or stacked use is necessary, the equipment should be
observed to verify normal operation in the configuration in which it
will be used.
(6) Test items (Tables 1 to 4)
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-6
Safety Precautions-7
Table 2
Guidance and manufacturer’s declaration - electromagnetic immunity
The Console is intended for use in the electromagnetic environment specified below.
The customer or the user of the Console should assure that it is used in such an environment.
Immunity test EN/IEC test level Compliance level Electromagnetic environment - guidance
±6kV contact ±6kV contact
Electrostatic discharge (ESD) Floors should be wood, concrete or ceramic tile. If floors are covered with
IEC61000-4-2 synthetic material, the relative humidity should be at least 30%.
±8kV air ±8kV air
±2kV for power supply lines ±2kV for power supply lines
Electrical fast transient/burst Mains power quality should be that of a typical commercial or hospital
IEC61000-4-4 environment.
±1kV for input/output lines ±1kV for input/output lines
±1kV between lines ±1kV between lines
Surge Mains power quality should be that of a typical commercial or hospital
IEC61000-4-5 environment.
±2kV between line and contact ±2kV between line and contact
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 0.5 cycles for 0.5 cycles
40% UT 40% UT
(>60% dip in UT) (>60% dip in UT)
Voltage dips, short interruptions and Mains power quality should be that of a typical commercial or hospital
for 5 cycles for 5 cycles
voltage variations of power supply environment. If the user of the Console requires continued operation during
input lines power mains interruptions, it is recommended that the Console be powered
70% UT 70% UT
IEC61000-4-11 from an uninterruptible power supply or a battery.
(30% dip in UT) (30% dip in UT)
for 25 cycles for 25 cycles
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 cycles for 5 cycles
Power frequency (50/60Hz)
Power frequency magnetic fields should be at levels characteristic of a typical
magnetic field 3 A/m 3 A/m
location in a typical commercial or hospital environment.
IEC61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-7
Safety Precautions-8
Table 3
Guidance and manufacturer’s declaration - electromagnetic immunity
The Console is intended for use in the electromagnetic environment specified below.
The customer or the user of the Console should assure that it is used in such an environment.
Immunity test EN/IEC test level Compliance level Electromagnetic environment - guidance
Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment should be used no closer to any
IEC61000-4-6 150 kHz to 80 MHz part of the Console, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80 MHz to 800 MHz
Radiated RF 3 V/m 3 V/m
IEC61000-4-3 80 MHz to 2.5 GHz d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in
which the Console is used exceeds the applicable RF compliance, the Console should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Console.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-8
Safety Precautions-9
Table 4
Recommended separation distances between
Portable and mobile RF communications equipment and the Console
The Console is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled.
The customer or the user of the Console can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Console as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
m
output power of
transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
W
d = 1.2 P d = 1.2 P d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-9
Safety Precautions-10
3. Caution for Handling Personal
Information
● Discarding the Equipment
When disposing of the DR-ID 300CL, be sure to completely discard any
private information stored on it.
● When to Obtain and to Bring Out Personal Information
Follow the operating procedures under the Private Information
Protection Law and prescribed by your local FUJIFILM representative.
When you collect logs, use such as a security-ensured jig provided by
your local FUJIFILM representative, and handle the medium with
adequate care.
010-305-02E
04.15.2010 FM5668
DR-ID 300CL Service Manual Safety Precautions-10
Product Specifications-I
DR-ID 300CL Service Manual
Product Specifications
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-I
Product Specifications-II
Control Sheet
Issue date Revision number Reason Pages affected
08/20/2009 01 New release (FM5581) All pages
04/15/2010 02 Revision for V4.0 (FM5668) 1–6
02/28/2011 03 Revision for V5.0 (FM5818) 1–3
08/31/2011 04 Revision for V6.0 (FM5897) 1
08/31/2011 04 Changes in pagination (FM5897) 2
12/15/2011 05 Revision for V6.1 (FM5944) 1–4
03/31/2012 06 Revision for V6.2 (FM6000) 2
03/31/2012 06 Changes in pagination (FM6000) 1, 3–5
09/12/2012 07 Revision for V7.0 (FM6055) 2–6
02/15/2013 08 Revision for V7.1 (FM6142) 2–5
05/25/2013 09 Revision for V7.2 (FM6179) 1, 2
05/25/2013 09 Changes in pagination (FM6179) 3
03/31/2014 11 Revision for V7.4 (FM6225) 3–7, 9, 10
03/31/2014 11 Changes in pagination (FM6225) 1, 2, 8
08/31/2014 12 Revision for V8.0 (FM6242) 2, 3, 5–9
08/31/2014 12 Changes in pagination (FM6242) 10
12/31/2015 14 Revision for V9.0 (FM9380) 1–3
12/31/2015 14 Changes in pagination (FM9380) 4
07/31/2016 16 Revision for V10.0 (FM9434) 2-6, 8
10/31/2016 17 Revision (FM9444) 3-9
010-305-17E
10.31.2016 FM9444
DR-ID 300CL Service Manual Product Specifications-II
Product Specifications-III
Control Sheet
Issue date Revision number Reason Pages affected
03/31/2017 18 Revision for V11.0 (FM9475) 2-4, 11-12
06/29/2018 19 Revision for V12.0 (FM9522) 4-8
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-III
Product Specifications-IV
BLANK PAGE
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Product Specifications-IV
Product Specifications-1
1. Equipment Specifications
1.1 Product Code
Product Code Description Q’ty Remarks
DR300CL A*-V* Control Unit (Non-RAID type) 1
DR300CL B*-V* Control Unit (RAID type) 1
DR300CL LITE A*-V* Control Unit (LITE specification Non-RAID type) 1 V5.0 or later
DR300CL LITE B*-V* Control Unit (LITE specification RAID type) 1 V5.0 or later
DR300CL A*-V* MOBILE 3 Control Unit (Mobile specification) 1 V7.3 or later
DX CL SOFT V* Software 1 V6.1 or later
DX CL SOFT LITE V* Software (LITE specification) 1 V7.1 or later
DX CL SOFT MOBILE V* Software (Laptop type) 1 V6.1 or later
010-305-14E
12.31.2015 FM9380
DR-ID 300CL Service Manual Product Specifications-1
Product Specifications-2
1.2 Optional Components
Product Code Description Q’ty Remarks
DX CL DICOM PRINT E / DX CL DICOM PRINT LIC. E DICOM print key (CD/License) 1
DX CL MFP E / DX CL MFP LIC. E MFP Processing key (CD/License) 1
DX CL DICOM ORDER MWM E / DX CL DICOM ORDER MWM LIC. E DICOM Worklist (MWM) key (CD/License) 1
DX CL DICOM ID MWM E / DX CL DICOM ID MWM LIC. E DICOM worklist (MWM) key (CD/License) (only for patient information) 1
DX CL DICOM MPPS E / DX CL DICOM MPPS LIC. E DICOM PPS key (CD/License) 1
DX CL DICOM COMMITMENT E / DX CL DICOM COMMITMENT LIC. E DICOM Commitment Push key (CD/License) 1
DX CL SHUTTER-PROC E / DX CL SHUTTER-PROC LIC. E Shuttering Processing key (CD/License) 1
DX CL TILE-IM QA E / DX CL TILE-IM OA LIC. E QA Tiling Confirmation key CD 1 V4.0 or earlier
V4.0 or earlier, and V7.2
DX CL DICOM QR / DX CL DICOM QR LIC. E DICOM Image Query/Retrieve key (CD/License) 1
or later
DX CL GPR E / DX CL GPR LIC. E Grid pattern removal processing key (CD/License) 1
DX CL IM COMPOSITION E / DX CL IM COMPOSITION LIC. E Automatic image composition key (CD/License) 1
DX CL RETAKE ANALYSIS E / DX CL RETAKE ANALYSIS LIC. E Retake Analysis key (CD/License) 1
DX CL QC TEST SOFT E / DX CL QC TEST SOFT LIC. E Image Reader QC key CD 1 V4.0 or earlier
DX CL RUQC E / DX CL RUQC LIC. E Image Reader QC key (CD/License) 1 V6.1 or later
DX CL DICOM MEDIA STORAGE E /
DICOM Media Storage key (CD/License) 1
DX CL DICOM MEDIA STORAGE LIC. E
DX CL PRECISE ENLARGEMENT E Precise Enlargement function software CD 1 For V4.0
DX CL PRECISE ENLARGEMENT A E /
Precise Enlargement function key (CD/License) 1 V5.0 or later
DX CL PRECISE ENLARGEMENT A LIC. E
DX CL DISPLAY OF PATIENT INF E /
Patient Information Display Linkage function key (CD/License) 1 V4.0 or later
DX CL DISP OF PATIENT INF LIC. E
DX CL SIMPLE ORDERING E / DX CL SIMPLE ORDERING LIC. E Simple ordering function key (CD/License) 1 V5.0 or later
DX CL DELUX-QA FOR LITE E / DX CL DELUX-QA FOR LITE LIC. E Detail QA key (CD/License) (for Lite type) 1 V5.0 or later
DX CL IM-MAGNIFY FOR LITE E / DX CL IM-MAGNIFY FOR LITE LIC. E Image magnification key (CD/License) (for Lite type) 1 V5.0 or later
DX CL LUT-ADJ FOR LITE E / DX CL LUT-ADJ FOR LITE LIC. E LUT adjustment key for competitor’s PACS (CD/License) (for Lite type) 1 V5.0 or later
DX CL ANNOTATION FOR LITE E / DX CL ANNOTATION FOR LITE LIC. E Annotation key (CD/License) (for Lite type) 1 V5.0 or later
DX CL X ONLINE FOR LITE E / DX CL X ONLINE FOR LITE LIC. E X-CON control function key (CD/License) (for Lite type) 1 V5.0 or later
DX CL FPD FOR LITE E / DX CL FPD FOR LITE LIC. E DR exposure device connection key (CD/License) (for Lite type) 1 V5.0 or later
DX CL DICOM PDI STORAGE E / DX CL DICOM PDI STORAGE LIC. E PDI key (CD/License) 1 V6.0 or later
DX CL F-LAYOUT PRINT E / DX CL F-LAYOUT PRINT LIC. E Free Layout Print key (CD/License) 1 V6.0 or later
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Product Specifications-2
Product Specifications-3
Product Code Description Q’ty Remarks
DX CL REMOTE IMAGE DISPLAY E /
Remote Display Image Display function key (CD/License) 1 V6.1 or later
DX CL REMOTE IMG DISPLAY LIC. E
DX CL ES FOR DR-ID200 E / DX CL ES FOR DR-ID200 LIC. E Energy Subtraction key (CD/License) 1 V6.1 or later
DX CL TOMO FOR DR-ID200 A E Tomosynthesis key CD 1 V6.1 to V7.4
DX CL TOMO FOR DR-ID200 B E / DX CL TOMO FOR DR-ID200 B LIC. E Tomosynthesis key (CD/License) 1 V8.0 or later
DX CL REMOTE DESKTOP E / DX CL REMOTE DESKTOP LIC. E Remote Desktop function key (CD/License) 1 V6.2 or later
IR CL PEM E / IR CL PEM LIC. E PEM Processing key (CD/License) 1 V7.0 or later
DX CL MAMMOGRAPHY QC E / DX CL MANMO QC LIC. E Mammography QC key (CD/License) 1 V7.1 or later
DX CL REFERRAL VIEWING E / DX CL REFERRAL VIEWING LIC. E Referral Viewing function key (CD/License) 1 V7.1 or later
DX CL SMART CONTROLLER E / DX CL SMART CONTROLLER LIC. E Smart-Controller key (CD/License) 1 V7.2 or later
DX CL MOTION DETECTION E / DX CL MOTION DETECTION LIC. E General exposure motion detection function key (CD/License) 1 V7.2 or later
DX CL VIRTUAL GRID E / DX CL VIRTUAL GRID LIC. E Virtual Grid function key (CD/License) 1 V8.0 or later
DX CL GENERAL RIS BROWSER E /
General RIS Browser key (CD/License) 1 V7.4 or later
DX CL GEN RIS BROWSER LIC. E
DX CL DYN VISUAL II LIC. E Dynamic Visualization II function key (Licence) 1 V10.0 or later
- V10.0 or later
DX CL DYN VISUAL II UPG LIC. E Dynamic Visualization II function option upgrading (Licence) 1 - For upgrading MFP to
DYN.
DX CL QA ROI MEASUREMENT E / DX CL QA ROI MEASURE LIC. E ROI measurement function key (Licence) 1
IR 346 MCR ISO B Magnetic Card Reader (ISO) 1
IR 346 BARCODE NB Barcode Reader 1
MCR ISO E Magnetic Card Reader (ISO) 1 From 2016.4
IR CL BARCODE KIT DL 3 E Barcode reader (Datalogic Gryphon GD4130) 1
IR CL BARCODE STAND DL E Barcode reader stand Datalogic Gryphon GD4130) 1
IR CL BARCODE OA E Barcode Reader (OmniScan)/stand 1
DR300 CL DISPLAY 17IN LCD E 17" Touch Panel LCD Display 1
LCD 17IN TP CLR EIZ E 17" Touch Panel Color Display 1
DR300 CL 2M DISPLAY CLR E 2M Color LCD (EIZO Radiforce RX211/RX220/RX240) 1
LCD 2M CLR EIZ E 2M Color LCD (EIZO Radiforce RX250) 1
DR300 2M DISPLAY DUAL GRAY A E DUAL Monitor LCD/2M Monochrome (EIZO Radiforce GS220-CL) 1
2M DISPLAY DUAL GRAY A E DUAL Monitor LCD/2M Monochrome (NANAO Radiforce GX240-CL) 1
DR300 3M DISPLAY DUAL CLR B E DUAL Monitor LCD/3M Color (EIZO Radiforce RX320/RX340) 1
LCD 3M CLR EIZ E 3M Color LCD (EIZO Radiforce RX350) 1
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Product Specifications-3
Product Specifications-4
Product Code Description Q’ty Remarks
DUAL Monitor LCD/3M Monochrome
DR300 3M DISPLAY DUAL GRAY E 1
(EIZO Radiforce GS310-CL/GS310-CL ECN)
3M DISPLAY GRAY AG E DUAL Monitor LCD/3M Monochrome (NANAO Radiforce GX340-CL) 1
VIDEO BD V3600 DUAL Monitor Video Board (ATI FireGL V3600) 1 Windows Vista only
VIDEO BD V3800 DUAL Monitor Video Board (ATI FireGL V3800) 1 For Windows Vista/7
VIDEO BD X3900 DUAL Monitor Video Board (EIZO MED-X3900) 1 Windows 7 only
VIDEO BD K620 E DUAL Monitor Video Board (NVIDIA Quadro K620) 1 For Windows 7/10
RS232C cable (9 pin) for connection to other manufacturer’s RIS terminal
IDT3/4 ONLINE 9S-CBL nM* E 1 *n: Cable length
or X-ray apparatus
RS232C cable (25 pin) for connection to other manufacturer’s RIS
IDT3/4 ONLINE 25P-CBL nM* E 1 *n: Cable length
terminal or X-ray apparatus
SERIAL PORT OPTI TYPE 1 E Additional serial port (for Optiplex GX620/745/755/760/780) 1
SERIAL PORT OPTI TYPE 10 E Additional serial port (for Optiplex 790DT/790SFF/990DT/990SFF) 1 V6.1 or later
- V7.0 or later
SERIAL PORT OPTI TYPE 20 E Additional serial cable (for HP 6200/6300/600G1/600G2/8200) 1
- For PCI
SERIAL PORT OPTI TYPE 21 Additional serial cable (for HP EliteDesk 800 G3 SFF) 1 V11.0 or later
XCON CONNECT KIT E XCON connection kit for RS-232C 1
SERIAL PORT BS E Board for serial capacity enhancement (MOXA CP-112UL-I-DB9) 1 V5.0 or later
Board for serial capacity enhancement - V7.4 or later
SERIAL PORT BD I PCIEX1 E 1
(MOXA CP-114EL-I-DB9M/RS-232C or RS-422) - For PCI Express x1
MARKER KIT E Positioning marker 1 V6.1 or later
GIGABIT NIC BD PCIEX1 E Additional LAN board (Gigabit CT Desktop Adapter EXPI9301CT) 1 V7.0 or later
GIGABIT NIC BD PCI E Additional LAN board (for OptiPlex 780DT) 1 V7.0 or later
OPTIPLEX MEM 4GB TYPE 3 E Memory extension 2GB×2 (for OptiPlex 760/790) 2 V7.0 or later
Memory extension 2GB×2
OPTIPLEX MEM 4GB TYPE 4 E 2 V7.0 or later
(for OptiPlex GX280/745/755/760/960/WS370/WS380/WS390)
OPTIPLEX MEM 4GB TYPE 11 Memory extension 2GB×2 (for OptiPlex 780/790/990) 2 V7.0 or later
DR300CL PPU A*-V* E Control Unit (PPU) 1 V6.1 or later
DR-XD 1000 BT BCR KIT E Bluetooth reader (opto-electronics OPN-3002i) 1 V10.0 or later
DISPLAYPORT DVI SINGLE 2 Displayport - DVI conversion adapter (single) (for HP PC) 1
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-4
Product Specifications-5
1.3 Dimensions/Weight
1.3.1 PC
Model Dimensions (height width depth) Weight Remarks
Optiplex 760MT 408.9 186.9 445 (mm) Approx. 14.3 kg
Optiplex 780MT 408 187 433 (mm) Approx. 11.7 kg
Optiplex 780DT 399 114 353 (mm) Approx. 8.26 kg V5.0 or later
Optiplex 790DT 360 102 410 (mm) Approx. 7.56 kg V6.1 or later
Compaq 6200 Pro SF 338 100 378 (mm) Approx. 7.6 kg V7.0 or later
Compaq 6300 Pro SF 338 100 378 (mm) Approx. 7.6 kg V7.0 or later
Compaq 8200 Elite MT 448 178 445 (mm) Approx. 11.2 kg V7.0 or later
ProDesk 600 G1 SF 338 100 379 (mm) Approx. 7.6 kg V7.4 or later
ProDesk 600 G2 SF 338 100 379 (mm) Approx. 7.6 kg V9.0 or later
EliteDesk 800 G3 SFF 338 100 308 (mm) Approx. 5.3 kg V11.0 or later
BFC-1208 292.9 225.4 50.14 (mm) Approx. 3.0 kg V10.0 or later
1.3.2 Monitor
Model Dimensions (height width depth) Weight Remarks
L560T-C 432.5 - 514.5 380 208.5 (mm) Approx. 8.2kg
T1781 411 392 157 (mm) Approx. 6.7kg
RadiForce RX211 522.5 - 604.5 376 208.5 (mm) Approx. 9.4kg
RadiForce RX220 522.5 - 604.5 376 208.5 (mm) Approx. 10.4kg
RadiForce RX240 520.5 - 599.0 376 245.5 (mm) Approx. 10.2kg V7.0 or later
RadiForce RX250 428.5 - 572.5 361 200 (mm) Approx. 8.2kg V10.0 or later
RadiForce GS220-CL 522.5 - 604.5 376 208.5 (mm) Approx. 9.9kg
RadiForce GX240-CL 520.5 - 599.0 376 245.5 (mm) Approx. 10.2kg V7.0 or later
RadiForce RX320 522.5 - 604.5 376 208.5 (mm) Approx. 10.4kg
RadiForce RX340 520.5 - 599.0 376 245.5 (mm) Approx. 10.7kg V7.0 or later
RadiForce RX350 481 - 571 354 200 (mm) Approx. 8.1kg V10.0 or later
RadiForce GS310-CL 515.5 - 597.5 368 208.5 (mm) Approx. 9.2kg
RadiForce GS310-CL ECN 515.5 - 597.5 368 208.5 (mm) Approx. 9.2kg
RadiForce GX340-CL 520.5 - 599.0 376 245.5 (mm) Approx. 10.2kg V7.0 or later
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-5
Product Specifications-6
1.4 Environmental Conditions
1.4.1 PC
➥ REFERENCE
Increase humidity appropriately if it is too dry to operate the equipment.
Model Condition Temperature Humidity * Atmospheric Pressure
Optiplex 760MT Operating 10ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
Optiplex 780MT
Optiplex 780DT
Optiplex 790DT Non-operating -40ºC - 65ºC 20% - 80%RH 200hpa - 1060hpa
Compaq 6200 Pro SF Operating 10ºC - 35ºC 10% - 90%RH 710hpa or more
Compaq 6300 Pro SF
Compaq 8200 Elite MT Non-operating -30ºC - 60ºC 10% - 90%RH 320hpa or more
ProDesk 600 G1 SF Operating 10ºC - 35ºC 10% - 90%RH 710hpa or more
ProDesk 600 G2 SF Non-operating -30ºC - 60ºC 5% - 95%RH 320hpa or more
Operating 10ºC - 35ºC 10% - 90%RH
EliteDesk 800 G3 SFF
Non-operating -30ºC - 60ºC 5% - 95%RH
Operating 0ºC - 40ºC 0% - 90%RH
BFC-1208
Non-operating -10ºC - 50ºC 0% - 90%RH
* No dew condensation
1.4.2 Monitor
➥ REFERENCE
Increase humidity appropriately if it is too dry to operate the equipment.
Model Condition Temperature Humidity * Atmospheric Pressure
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
L560T-C
Non-operating -20ºC - 60ºC 10% - 85%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 540hpa - 1060hpa
T1781
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
* No dew condensation
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-6
Product Specifications-7
Model Condition Temperature Humidity * Atmospheric Pressure
Operating 0ºC - 35ºC 30% - 80%RH 700hpa - 1060hpa
RadiForce RX211
Non-operating -20ºC - 60ºC 30% - 80%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce RX220
Non-operating -20ºC - 60ºC 10% - 85%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce RX240
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce RX250
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 30% - 80%RH 700hpa - 1060hpa
RadiForce GS220-CL
Non-operating -20ºC - 60ºC 10% - 85%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce GX240-CL
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 30% - 80%RH 700hpa - 1060hpa
RadiForce RX320
Non-operating -20ºC - 60ºC 10% - 85%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce RX340
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce RX350
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 30% - 80%RH 700hpa - 1060hpa
RadiForce GS310-CL
Non-operating -20ºC - 60ºC 30% - 80%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 30% - 80%RH 700hpa - 1060hpa
RadiForce GS310-CL ECN
Non-operating -20ºC - 60ºC 30% - 80%RH 200hpa - 1060hpa
Operating 0ºC - 35ºC 20% - 80%RH 700hpa - 1060hpa
RadiForce GX340-CL
Non-operating -20ºC - 60ºC 10% - 90%RH 200hpa - 1060hpa
* No dew condensation
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-7
Product Specifications-8
1.5 Electrical Specifications
1.5.1 PC
Model Frequency Input Voltage Max. Input Current Remarks
115V 3.6A
Optiplex 760MT 50/60Hz
230V 1.8A
90V - 135V 3.6A
Optiplex 780MT 50/60Hz
180V - 265V 1.8A
90V - 135V 3.6A
Optiplex 780DT 50/60Hz V5.0 or later
180V - 265V 1.8A
90V - 135V 3.6A
Optiplex 790DT 50/60Hz V6.1 or later
180V - 265V 1.8A
Compaq 6200 Pro SF 50/60Hz 90V - 264V 2.6A V7.0 or later
Compaq 6300 Pro SF 50/60Hz 90V - 264V 2.6A V7.0 or later
Compaq 8200 Elite MT 50/60Hz 90V - 264V 3.5A V7.0 or later
ProDesk 600 G1 SF 50/60Hz 90V - 264V 3.6A V7.4 or later
ProDesk 600 G2 SF 50/60Hz 90V - 264V 3.5A V9.0 or later
EliteDesk 800 G3 SFF 50/60Hz 90V - 264V - V11.0 or later
BFC-1208 50/60Hz 100V - 240V 5A V10.0 or later
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Product Specifications-8
Product Specifications-9
1.5.2 Monitor
Model Frequency Input Voltage Max. Input Current Remarks
100V - 120V 0.55A - 0.5A
L560T-C 50/60Hz
200V - 240V 0.35A - 0.3A
100V - 120V 0.7A - 0.6A
T1781 50/60Hz
200V - 240V 0.4A - 0.35A
100V - 120V 1.1A
RadiForce RX211 50/60Hz
200V - 240V 0.6A
100V - 120V 1.2A
RadiForce RX220 50/60Hz
200V - 240V 0.6A
100V - 120V 1.1A - 0.9A
RadiForce RX240 50/60Hz
200V - 240V 0.6A - 0.5A
100V - 120V 0.8A - 0.7A
RadiForce RX250 50/60Hz
200V - 240V 0.4A - 0.3A
100V - 120V 1.1A
RadiForce GS220-CL 50/60Hz
200V - 240V 0.5A
100V - 120V 0.85A - 0.65A
RadiForce GX240-CL 50/60Hz
200V - 240V 0.4A - 0.35A
100V - 120V 1.2A
RadiForce RX320 50/60Hz
200V - 240V 0.6A
100V - 120V 1.3A - 1.0A
RadiForce RX340 50/60Hz
200V - 240V 0.6A - 0.5A
100V - 120V 0.9A - 0.8A
RadiForce RX350 50/60Hz
200V - 240V 0.5A - 0.4A
100V - 120V 1.0A - 0.8A
RadiForce GS310-CL 50/60Hz
200V - 240V 0.5A - 0.4A
100V - 120V 1.0A - 0.8A
RadiForce GS310-CL ECN 50/60Hz
200V - 240V 0.5A - 0.4A
100V - 120V 0.9A - 0.75A
RadiForce GX340-CL 50/60Hz
200V - 240V 0.5A - 0.45A
010-305-17E
10.31.2016 FM9444
DR-ID 300CL Service Manual Product Specifications-9
Product Specifications-10
1.6 Applicable Standards
PC IEC 60950-1
Monitor IEC 60950-1
1.7 Network Cable
Use an IEC950/UL1950-compliant UTP cable of enhanced category 5
or higher.
1.8 Equipment Disposal
When disposing of the DR-ID 300CL, be sure to completely discard any
private information stored on it.
010-305-12E
08.31.2014 FM6242
DR-ID 300CL Service Manual Product Specifications-10
Product Specifications-11
BLANK PAGE
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Product Specifications-11
Product Specifications-12
BLANK PAGE
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Product Specifications-12
DR-ID 300CL Service Manual
MT: Machine Troubleshooting
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual MT-I
Control Sheet
Issue date Revision number Reason Pages affected
08/20/2009 01 New release (FM5581) All pages
04/15/2010 02 Revision for V4.0 (FM5668) 3, 4, 8, 10, 22 Appx MT A-1–136 Appx MT B-1–12
Appx MT C-1–4
02/28/2011 03 Revision for V5.0 (FM5818) 4, 5, 8, 9, 18, 19, 21–26 Appx MT A-1–136 Appx MT B-1–108
Appx MT C-1–12 Appx MT D-1–4
02/28/2011 03 Changes in pagination (FM5818) 20
08/31/2011 04 Revision for V6.0 (FM5897) Appx MT B-1–108 Appx MT C-1–16 Appx MT D-1–4
Appx MT E-1–8
12/15/2011 05 Revision for V6.1 (FM5944) 11, 25 Appx MT B-1–136 Appx MT E-1–8
03/31/2012 06 Revision for V6.2 (FM6000) Appx MT B-1–148 Appx MT C-1–16
09/12/2012 07 Revision for V7.0 (FM6055) 11, 26–34 Appx MT B-1–152
02/15/2013 08 Revision for V7.1 (FM6142) 10–12, 27, 36–38 Appx MT B-1–184 Appx MT F-1–4
02/15/2013 08 Changes in pagination (FM6142) 13–26, 28–35
05/25/2013 09 Revision for V7.2 (FM6179) 22, 24 Appx MT B-1–188 Appx MT G-1–4
05/25/2013 09 Changes in pagination (FM6179) 25
08/30/2013 10 Revision for V7.3 (FM6200) Appx MT B-1–196
03/31/2014 11 Revision for V7.4 (FM6225) 27, 28, 32–35, 39–42 Appx MT B-1–208 Appx MT H-1–4
03/31/2014 11 Changes in pagination (FM6225) 36–38
08/31/2014 12 Revision for V8.0 (FM6242) Appx MT B-1–212
01/31/2015 13 Revision for V8.1 (FM6269) Appx MT B-1–216 Appx MT I-1–4
12/31/2015 14 Revision for V9.0 (FM9380) 25, 42–46 Appx MT B-1–220
12/31/2015 14 Changes in pagination (FM9380) 24
010-305-14E
12.31.2015 FM9380
DR-ID 300CL Service Manual MT-II
Control Sheet
Issue date Revision number Reason Pages affected
07/31/2016 16 Revision for V10.0 (FM9434) 1, 25, 27, 34-36, 40, 45-50 Appx MT B-5, 6, 151
07/31/2016 16 Changes in pagination (FM9434) 37-39, 41-44, 51, 52 Appx MT B-7-150, 152-219
10/31/2016 17 Revision (FM9444) 45, 47-50, 53-56
10/31/2016 17 Changes in pagination (FM9444) 51, 52
03/31/2017 18 Revision for V11.0 (FM9475) Appx MT A-56, 79
Appx MT B-6, 11, 14, 22, 24, 45, 59, 61-63, 65, 69, 71, 131, 153-155, 162, 166, 193, 203, 212,
214-215, 218, 220-224
03/31/2017 18 Changes in pagination (FM9475) Appx MT B-7-10, 12-13, 15-21, 23, 25-44, 46-58, 60, 64, 66-68, 70, 72-130, 132-152, 156-161,
163-165, 167-192, 194-202, 204-211, 213, 216-217, 219
06/29/2018 19 Revision for V12.0 (FM9522) 1, 26, 32, 35-38, 42, 47, 53-60, 62
Appx MT B-1, 5-6, 67, 70, 153
Appx MT C-4
06/29/2018 19 Changes in pagination (FM9522) 39-41, 43-46, 48-52, 61, 63-64
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual MT-III
BLANK PAGE
010-305-16E
07.31.2016 FM9434
DR-ID 300CL Service Manual MT-IV
Appendix A DR-ID 300CL Error Message Table (V4.0 or Earlier)
Event Event Source Description Causes Remedies Remarks
1 VBRunTime The VB Application No application key is registered to the Install the DR-ID 300CL application key.
identified by the event registry.
source logged this
ApplicationPDFPassWordInput:
Thread ID: 2556 ,Logged:
PasswordInputTool can’t
start.
29 FFIIPQACtrl clsCombineDialog.cls, Failed in automatic image combining Check image orientation and rotation and
cmdAutoCombine_Click, 1. processing. then retry.
3001 FFQAMsgBoxCtrl ctlImageLaterality.ctl, The ImageLaterality tag ([0x00200062] Set up image laterality from the User Utility.
GetLateralityParam, 1, tag) does not exist in DDO.
clsImageInfoFCR.
GetBSTRValue
[0x00200062] Element Not
Found.
3001 FFQAMsgBoxCtrl ctlImageLaterality.ctl, No value has been set up for the Set up image laterality from the User Utility.
GetLateralityParam, 2, ImageLaterality tag ([0x00200062] tag) in
clsImageInfoFCR. DDO.
GetBSTRValue
[0x00200062] Length = 0.
10000 MenuManager Property data is invalid. A program internal error or inconsistency in Restart. Use the User Utility to check “%s”: Detailed
[%s] exposure menu data. parameter of the exposure menu at error explanation
Invalid Data(%s%s) occurrence.
10001 MenuManager Method Error. A program internal error. Restart. If the same symptom recurs, “%s”: Detailed
[%s] reinstall the system. explanation
Error Message: %s
10001 CooperMsgNotify Property data is invalid. The parameter to be conveyed to the I/F is Restart.
%n%1%n%2 invalid.
10002 CooperMsgNotify Function Call Error. A program error. Restart.
%n%1%n%2
10003 CooperMsgNotify Exception Error. A program error. Restart.
%n%1%n%2
10900 IIPInputEz %1 : Registry Value The acquisition/setup key or registration Check the registry key.
Get/Set Error%n information is abnormal.
[Key] %2%n
[Exp] %3%n
%4%n
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-1
Event Event Source Description Causes Remedies Remarks
10901 IIPInputEz %1 : Database Connection This error is attributable to an abnormal Check Directory.ini or database file name.
Error%n character string in Directory.ini, an incorrect
[DB] %2%n database file name, or a damaged
[Exp] %3%n database.
%4%n
%5
10902 IIPInputEz %1 : Database Data This error is attributable to an abnormal Check the database table/field name.
Get/Set Error%n database acquisition/setup key or
[DB] %2%n database.
[Table] %3%n
[Field] %4%n
[Exp] %5
10903 IIPInputEz %1 : Dicom Data Object The DDO acquisition/setup tag is Check the DDO tag.
Get/Set Error%n abnormal.
[Tag] %2%n
[Exp] %3%n
%4%n
%5
10904 IIPInputEz %1 : Other Error%n Unknown error.
[Procedure] %2%n
[ErrorCode] %3%n
[Exp] %4%n
%5
11900 IIPInputEz %1 : Registry Value The acquisition/setup key or registration Check the registry key.
Get/Set Error%n information is abnormal.
[Key] %2%n
[Exp] %3%n
%4%n
%5
11900 IIPPortable %1 : Registry Value Get/ The registry was damaged. Restart. If the same symptom recurs,
Set Error reinstall the system.
[Key] %2
[Exp] %3
%4
%5
11901 IIPInputEz %1 : Database Connection This error is attributable to an abnormal Check Directory.ini or database file name.
Error%n character string in Directory.ini, an incorrect
[DB] %2%n database file name, or a damaged
[Exp] %3%n database.
%4%n
%5
11901 IIPPortable %1 : Database Connection Failed in accessing the image database. Restart. If the same symptom recurs,
Error reinstall the system and perform
[DB] %2 “Initialization of image database”. (Image
[Exp] %3 data will be fully deleted.)
%4
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-2
Event Event Source Description Causes Remedies Remarks
11902 IIPInputEz %1 : Database Data Get/ This error is attributable to an abnormal Check the database table/field name.
Set Error%n acquisition/setup key or database.
[DB] %2%n
[Table] %3%n
[Field] %4%n
[Exp] %5
11902 IIPPortable %1 : Database Data Get/ Failed in accessing the image database. Restart. If the same symptom recurs,
Set Error reinstall the system and perform
[DB] %2 “Initialization of image database”. (Image
[Table]%3 data will be fully deleted.)
[Field]%4
[Exp]%5
11903 IIPInputEz %1 : Dicom Data Object The DDO acquisition/setup tag is Check the DDO tag.
Get/Set Error%n abnormal.
[Tag] %2%n
[Exp] %3%n
%4%n
%5
11903 IIPPortable %1 : DDO Data Get/Set The DDO file was damaged. Restart. If the same symptom recurs,
Error reinstall the system and perform
[Tag] %2 “Initialization of image database”. (Image
%3 data will be fully deleted.)
%4
%5
11904 IIPInputEz %1 : Other Error%n Unknown error.
[Procedure] %2%n
[ErrorCode] %3%n
[Exp] %4%n
%5
11904 IIPPortable %1 : Other Error Program error or unstable system status. Check POCKET id Console connection
[Procedure] %2 status and restart. If the same symptom
[Error Code]%3 recurs, reinstall the system and perform
[Exp]%4 “Initialization of image database”. (Image
%5 data will be fully deleted.)
12001 FF32 [%1] Failed in reading a bitmap file. Check to see if the filename is correct or “%1”: Instance name
It failed in CreateFile(%2). It is conceivable that there was no such available hard disk space is sufficient. “%2”: Error code
%3 file, free disk space was not sufficient or Proceed with processing. “%3”: Bitmap file name/
the system was unstable. Icon file name
12002 FF32 [%1] Failed in creating a file map object. Proceed with processing. “%1”: Instance name
It failed in It is conceivable that available resources “%2”: Error code
CreateFileMapping(%2). were insufficient or the system was “%3”: Bitmap file name/
%3 unstable. Icon file name
12003 FF32 [%1] Failed in mapping a file. It is conceivable Proceed with processing. “%1”: Instance name
It failed in that available resources were insufficient “%2”: Error code
MapViewOfFile(%2). or the system was unstable. “%3”: Bitmap file name/
%3 Icon file name
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-3
Event Event Source Description Causes Remedies Remarks
12004 FF32 [%1] Failed in creating a device context. Proceed with processing. “%1”: Instance name
It failed in It is conceivable that available resources “%2”: Error code
CreateCompatibleDC. were insufficient or the system was
[%2] unstable.
12005 FF32 [%1] Failed in creating a DIB section. Proceed with processing. “%1”: Instance name
It failed in It is conceivable that available resources “%2”: Error code
CreateDIBSection(%2). were insufficient or the system was “%3”: Bitmap file name/
%3 unstable. Icon file name
12006 FF32 [%1] Failed in creating a bitmap picture object. Proceed with processing. “%1”: Instance name
It failed in It is conceivable that available resources “%2”: Error code
OleCreatePictureIndirect(%2). were insufficient or the system was “%3”: Bitmap file name/
%3 unstable. Icon file name
12008 FF32 [%1] Failed in reading an icon file. Check to see if the filename is correct or “%1”: Instance name
It failed in ExtractIon. It is conceivable that there was no such available hard disk space is sufficient. “%2”: Error code
%2 file, free disk space was not sufficient or Proceed with processing.
the system was unstable.
12009 FF32 [%1] Failed in creating an picture object for an Proceed with processing. “%1”: Instance name
It failed in icon. It is conceivable that available “%2”: Error code
OleCreatePictureIndirect(%2). resources were insufficient or the system “%3”: Bitmap file name/
%3 was unstable. Icon file name
12011 FF32 [%1] The file path was illegal. Check to see if the filename is correct. “%1”: Instance name
It is a filepath error. %2 Proceed with processing. “%2”: Error code
12012 FF32 [%1] Failed in acquiring memory. Check to see if the filename is correct. “%1”: Instance name
Memory allocation failed(%2). It is conceivable that available resources Proceed with processing. “%2”: Error code
%3 were insufficient or the system was “%3”: Bitmap file name/
unstable. Icon file name
12014 FF32 [%1] An invalid input value was specified. Proceed with processing. Common software
“%1”: Instance name
(function name),
occurred error,
error code
12015 FF32 [%1] An invalid input value was specified. Proceed with processing. Common software
%2 “%1”: Instance name
(function name),
occurred error,
error code
“%2”: Error details
12016 FF32 [%1] An invalid input value was specified. Proceed with processing. Common software
%2 “%1”: Instance name
%3 (function name),
occurred error,
error code
“%2”: Error details (1/2)
“%3”: Error details (2/2)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-4
Event Event Source Description Causes Remedies Remarks
12900 IIPInputEz %1 : Registry Value The acquisition/setup key or registration Check the registry key.
Get/Set Error%n information is abnormal.
[Key] %2%n
[Exp] %3%n
%4%n
%5
12900 IIPPortable %1 : Registry Value Get/ A registry was damaged. Restart. If the same symptom recurs,
Set Error reinstall the system.
[Key] %2
[Exp] %3
%4
%5
12901 IIPInputEz %1 : Database Connection This error is attributable to an abnormal Check Directory.ini or database file name.
Error%n character string in Directory.ini, an incorrect
[DB] %2%n database file name, or a damaged
[Exp] %3%n database.
%4%n
%5
12901 IIPPortable %1 : Database Connection Failed in accessing the image database. Restart. If the same symptom recurs,
Error reinstall the system and perform
[DB] %2 “Initialization of image database”. (Image
[Exp] %3 data will be fully deleted.)
%4
%5
12902 IIPInputEz %1 : Database Data This error is attributable to an abnormal Check the database table/field name.
Get/Set Error%n acquisition/setup key or database.
[DB] %2%n
[Table] %3%n
[Field] %4%n
[Exp] %5
12902 IIPPortable %1 : Database Data Get/ Failed in accessing the image database. Restart. If the same symptom recurs,
Set Error reinstall the system and perform
[DB] %2 “Initialization of image database”. (Image
[Table]%3 data will be fully deleted.)
[Field]%4
[Exp]%5
12903 IIPInputEz %1 : Dicom Data Object The DDO acquisition/setup tag is Check the DDO tag.
Get/Set Error%n abnormal.
[Tag] %2%n
[Exp] %3%n
%4%n
%5
12903 IIPPortable %1 : DDO Data Get/Set The DDO file was damaged. Restart. If the same symptom recurs,
Error reinstall the system and perform
[Tag] %2 “Initialization of image database”. (Image
%3 data will be fully deleted.)
%4
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-5
Event Event Source Description Causes Remedies Remarks
12904 IIPInputEz %1 : Unknown Error%n Unknown error. Check the POCKET id Console connection
[Procedure] %2%n status.
[ErrorCode] %3%n
[Exp] %4%n
%5
12904 IIPPortable %1 : Other Error A general-purpose information event log. Check the POCKET id Console connection
[Procedure] %2 status.
[Error Code]%3
[Exp]%4
%5
13000 FFIdMenuChange It failed that [%1] gets the The “Font” key was not found in the Reinstall the IIP because the displayed
FontName registry. characters can be corrupted.
13000 FFIdMenuItem It failed that [%1] gets the The “Font” key was not found in the Reinstall the IIP because the displayed
FontName registry. characters can be corrupted.
13001 FFCustomMsgBox MessageBox DataBase The message database was not Reinstall or check whether the message
Connection Error. successfully connected. It is conceivable database exists and the contents of
that the database did not exist or that the Directory.ini are correct.
specified data source name was invalid.
13001 FFIdMenuChange It failed that [%1] gets the The “FontSizeSS”/“FontSizeS”/“FontSizeM”/ Reinstall the IIP because the displayed
FontSize “FontSizeL”/“FontSizeLL” key was not found characters can be corrupted.
in the registry.
13001 FFIdMenuItem It failed that [%1] gets the The “FontSizeSS”/“FontSizeS”/“FontSizeM”/ Reinstall the IIP because the displayed
FontSize “FontSizeL”/“FontSizeLL” key was not found characters can be corrupted.
in the registry.
13001 FFIipInput It failed that [%1] gets the The “Font” key was not found in the Reinstall the IIP because the displayed
FontName registry. characters can be corrupted.
13001 FFIpcInf [%1] Initialization was unsuccessful. Restart. “%1”: Instance name
It failed in CoInitialize(%2). It is conceivable that the system was “%2”: Error code
unstable.
13001 FFLogFile [%1] Log file creation was unsuccessful. Check for a free disk space or restart. “%1”: Instance name
It failed in CreateFile(%2). The problem might have been caused, for “%2”: Error code
instance, by an invalid filename selection, “%3”: File name/mutex
inadequate disk space, or system name
instability.
13001 FFImageDBCTL Image Database Access The image database was not successfully Restart.
Failure. accessed (ADO error).
ErrNo:%1 NativeErrNo:%2
SQLState:%3 Source:%4
Description:%5
13001 FF32LIB Image Database Access The image database was not successfully Restart.
Failure. accessed (ADO error).
ErrNo:%1 NativeErrNo:%2
SQLState:%3 Source:%4
Description:%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-6
Event Event Source Description Causes Remedies Remarks
13001 FFImageDBCTL Image Database Access The image database was not successfully Restart.
Error. accessed (ADO error).
ErroNo:%1
Source:%2
Description:%3
13001 FF32LIB FF32lib Database Access Failed in accessing the database. Restart.
Failure.%nFunction:%1
Command:%2%nErrNo:%3
NativeErrNo:%4
SQLState:%5%n
Source:%6%n
Description:%7
13002 FFIdMenuChange It failed that [%1] gets the The “Multibyte” key was not found in the Reinstall.
multibyte flag registry.
13002 FFIdMenuItem It failed that [%1] gets the The “Multibyte” key was not found in the Reinstall.
multibyte flag registry.
13002 FFIipInput It failed that [%1] gets the The “FontSizeSS”/“FontSizeS”/“FontSizeM”/ Reinstall the IIP because the displayed
FontSize “FontSizeL”/“FontSizeLL” key was not found characters can be corrupted.
in the registry.
13002 FFIpcInf [%1] A log save control was not successfully Restart. If the same symptom recurs, “%1”: Instance name
It failed in created. The problem is attributable to perform a reinstallation. “%2”: Error code
CoCreateInstance(%2). system instability or unregistered control.
LogFileCom create failure.
13002 FFLogFile [%1] File map object creation was unsuccessful. Restart. “%1”: Instance name
It failed in The problem is attributable to resource “%2”: Error code
CreateFileMapping(%2). inadequacy or system instability.
13002 FFImageDBCTL Image Database Access The image database was not successfully Restart.
Failure. accessed.
InstanceName[%1]%2
13002 FFIdMcrControl Driver initialize error[%1] The magnetic card was not connected. Connect the magnetic card reader.
13002 FFCustomMsgBox MessageBox Data Get The specified message was not Perform a reinstallation or replace the
Error. databased. message database with the latest one.
It is conceivable that the message was not
registered in the message database or that
the specified ID was incorrect (program
defect).
13002 FF32LIB FF32lib Database Access Failed in accessing the database. Restart.
Failure. InstanceName[%1]
%n%2%nFunction:%3
Command:%4
13002 FF32LIB Image Database Access Failed in accessing the image database. Restart.
Failure.
InstanceName[%1]
%2
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-7
Event Event Source Description Causes Remedies Remarks
13003 FFIdMenuChange [%1] failed in the The “DisplayDataDB” key was not found in Reinstall.
initialization for the display Directory.ini or the connection character
group data acquisition string was not properly registered in the
display group table.
13003 FFIdMenuItem Image Folder Path Get The “Bitmap” key was not found in Reinstall.
Error Directory.ini.
13003 FFIipInput It failed that [%1] gets the The “Multibyte” key was not found in the Reinstall.
multibyte flag registry.
13003 FFIpcInf [%1] Memory allocation was not successfully Restart. “%1”: Instance name
Memory allocation failed. achieved for main thread creation.
The problem is attributable to resource
inadequacy or system instability.
13003 FFLogFile [%1] File mapping was unsuccessful. Restart. “%1”: Instance name
It failed in The problem is attributable to resource “%2”: Error code
MapViewOfFile(%2). inadequacy or system instability.
13003 FFImageDBCTL Image Database Access The image database was not successfully Restart.
Failure. accessed.
InstanceName[%1]%2%3
13003 FF32LIB Image Database Access The image database was not successfully Restart.
Failure. accessed.
InstanceName[%1]%2
%3
13003 FFCustomMsgBox Same MessageBox The specified message box is already Close the message box.
Existed. open.
[%1]
13003 FF32LIB FF32lib Database Access Failed in accessing the database. Restart. “%1”: Instance name
Failure. “%2”: Error code
InstanceName[%1]%n%2
%n%3
13004 FFIdMenuChange [%1] failed in the display The “DisplayDataDB” key was not found in Reinstall.
group data acquisition Directory.ini or the connection character
string was not properly registered in the
display group table.
13004 FFIdMenuItem [Miss]Image Get Error Id_MissExposure_I.bmp was not found in Reinstall the IIP because the button cannot
the path to an image folder or the image be displayed normally.
format was illegal.
13004 FFIipInput It failed that [%1] gets the The “LangStrDB” key was not found in Reinstall.
connection string of STR Directory.ini.
DB
13004 FFIpcInf [%1] A pipe has already An attempt was made to duplicate a pipe. There are no operational problems. “%1”: Instance name
been made(%2). “%2”: Error code
13004 FFLogFile [%1] It failed in Mutex creation was unsuccessful. Restart. “%1”: Instance name
CreateMutex(%2). The problem is attributable to resource “%2”: Error code
%3 inadequacy or system instability. “%3”: file name/mutex
name
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-8
Event Event Source Description Causes Remedies Remarks
13004 FFIdMcrControl Magnetic card data The magnetic card was found damaged. Have a normal magnetic card be read.
acquisition error[%1]
13004 FFCustomMsgBox MessageBox Dialog Create The message box could not be created. Restart.
Error. This error occurred due to insufficient
[%1] resources or system instability.
13005 FFIdMenuChange [%1] failed in the liberating The connection to the display group table Restart. If the same symptom recurs,
processing of the display was not successfully closed. reinstall.
group data acquisition
13005 FFIipInput It failed that [%1] connects The connection character string was not Reinstall.
to STR DB properly registered for the “LangStrDB” key
in Directory.ini.
13005 FFIpcInf [%1] Main thread creation was not successful. Restart. “%1”: Instance name
It failed in _ The problem is attributable to resource “%2”: Error code
beginthread(%2). inadequacy or system instability.
13005 FFldMcrControl Magnetic card variation The magnetic card was not of the specified Have the specified type of magnetic card
error[%1] type. be read.
13005 FFCustomMsgBox MessageBox Dialog Show The message box could not be opened. Restart.
Error. This error occurred due to insufficient
[%1] resources or system instability.
13006 FFCustomMsgBox List For Modeless The modeless message box management Restart.
MessageBox Management table was not successfully created.
New Error. The problem is attributable to resource
inadequacy or system instability.
13006 FFIdMenuChange [%1] Failed in the The “DisplayDataDB” key was not found in Reinstall.
initialization for the displayDirectory.ini.
menu data acquisition
13006 FFIipInput It failed that [%1] gets the The “SysConfigDB“ key was not found in Reinstall.
connection string of SYS Directory.ini.
DB
13006 FFIpcInf [%1] It failed in The specified pipe was not successfully Make a change so as to let the program “%1”: Instance name
WaitNamedPipe(%2). connected. It is conceivable that the pipe initiate a retransmission, or restart. “%2”: Error code
was not created or was unconnectable or
that the system was unstable.
13006 FFIdMcrControl Magnetic card variation The card format was incorrect or an Have the specified hospital card be read.
check error[%1] attempt was made to read an unspecified
hospital card.
13006 FFCustomMsgBox List For Modeless The modeless message box management Restart.
MessageBox Management table was not successfully created.
New Error. This error occurred due to insufficient
[%1] resources or system instability.
13007 FFIdMenuChange [%1] failed in the acquisition The connection character string was not Reinstall.
of number of records of properly registered in the display group
display menu data table.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-9
Event Event Source Description Causes Remedies Remarks
13007 FFIipInput It failed that [%1] connects The connection character string was not Reinstall.
to SYS DB properly registered for the “SysConfigDB”
key in Directory.ini.
13007 FFIpcInf [%1] The specified pipe was not successfully The problem is attributable to the program. “%1”: Instance name
It failed in CreateFile(%2). connected. It is conceivable that the pipe Check the program or restart. “%2”: Error code
was closed or that the system was
unstable.
13008 FFIdMenuChange [%1] failed in the display Not properly registered to the display group Reinstall.
menu data acquisition table.
13008 FFIipInput It failed that [%1] calls theThe message database did not exist or the Reinstall.
CUSTOM MESSAGE BOX program was faulty.
(%2)
13008 FFIpcInf [%1] A message was not successfully The problem is attributable to the program. “%1”: Instance name
It failed in WriteFile(%2). transmitted. It is conceivable that the pipe Check the program or restart. “%2”: Error code
was closed or that the system was
unstable.
13008 FFCustomMsgBox An exception error An exception occurred (the system was Restart.
occurred. unstable).
[%1]
%2
13009 FFIdConnectionWatch LogFile Name Make The “Log” key was not found in Directory. Reinstall.
error(%1).%n[%2] ini.
13009 FFIdMenuChange [%1] Failed in the liberating The connection to the display menu table Restart. If the problem frequently occurs,
processing of the display was not successfully closed. initialize the image database.
menu data acquisition
13009 FFIpcInf [%1] Memory allocation was not successfully Restart. “%1”: Instance name
Memory allocation failed. achieved for reception thread creation. “%2”: Error code
aaaaa The problem is attributable to resource
inadequacy or system instability.
13009 FFIdConnectionWatchEx Log File Name Make The “Log” key was not found in Directory. Register the “Log” key in Directory.ini.
Error[%1] ini.
Error Code [%2]
13009 FFIdMcrControl DataFormatDB close The database was faulty. Reinstall.
error[%1]
13009 FFCustomMsgBox Common function error The common function returned an error. Restart.
occurred.
[%1]
%2
13010 FFIdConnectionWatch DB Connect The “SysConfigDB” key was not found in Reinstall.
error(%1).%n[%2] Directory.ini or SysConfig.mdb did not exist
in the path.
13010 FFIdMenuChange [%1] Failed in the The connection to the display menu table Restart. If the problem frequently occurs,
termination of the display was not successfully closed. initialize the image database.
menu data acquisition
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-10
Event Event Source Description Causes Remedies Remarks
13010 FFIdConnectionWatchEx DB Connect Error [%1] “SysConfigDB” key was not found in Properly register the “SysConfigDB” key in
Error Code [%2] Directory.ini or SysConfig.mdb did not exist Directory.ini.
in the specified path.
13010 FFIdMcrControl Patient info data error[%1] The patient information contained in the Check the data contained in the magnetic
%n(%2) magnetic card was incorrect. card.
13010 FFIpcInf [%1] It failed in The pipe was not successfully created. Restart.
CreateNamedPipe(%2). The system was unstable.
[Pipe:%3]
13010 FFCustomMsgBox ActiveX Data Objects(ADO) An ADO error was detected. Restart.
Error.
ErroNo:%1
Source:%2
Description:%3
13011 FFIipInput It failed that [%1] gets the The “ScallHight”/“ScallWidth” key was not Reinstall.
window size registered in the registry.
13011 FFIpcInf [%1] Pipe connection was not successfully Restart. “%1”: Instance name
It failed in awaited. It is conceivable that the system “%2”: Error code
ConnectNamedPipe(%2). was unstable.
13012 FFIpcInf [%1] The reception thread was not successfully Restart.
It failed in _ created. This error occurred due to
beginthread(%2). [Pipe:%3] insufficient resources or system instability.
13013 FFIdMenuChange It failed that [%1] get the The “Bitmap” key was not found in Reinstall.
folder path to Bitmap Directory.ini.
13013 FFIpcInf [%1] The message reception was not Restart.
It failed in ReadFile(%2). successfully completed. The system was
[Pipe:%3] unstable.
13014 FFIdMenuChange It failed that [%1] get the Id_PrevPage_U.bmp/“Id_PrevPage_D.bmp”/ Reinstall the IIP because the button cannot
bitmap to Previous Page “Id_PrevPage_X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
13014 FFIpcInf [%1] An illegal parameter was encountered “%1”: Instance name
It is a parameter error. (program defect).
13015 FFIdMenuChange It failed that [%1] get the Id_NextPage_U.bmp/“Id_NextPage_D.bmp”/ Reinstall the IIP because the button cannot
bitmap to Next Page “Id_NextPage_X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
13015 FFIpcInf [%1] An exception occurred (the system was Restart.
Exception Information. unstable).
%2
13023 FFIdPatInfoInput %1 (%2) : Patient The patient information contained illegal Correct the patient information setup.
Information Check Error%n information.
Item [%3]%n
Error [%4]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-11
Event Event Source Description Causes Remedies Remarks
13024 FFIdPatInfoInput %1 (%2) : Bitmap Image “Id_LocalSearch_U.bmp”, Place “Id_LocalSearch_U.bmp”,
Get Error%n “Id_LocalSearch_D.bmp”, or “Id_LocalSearch_D.bmp”, and
BMP File [%3] “Id_LocalSearch_X.bmp” was not found in “Id_LocalSearch_X.bmp” in the path to the
the path to the folder in which images were folder in which images are positioned.
positioned.
13025 FFIdPatInfoInput %1 (%2) : Patient Info Get The patient information database was Repair the patient information database.
Error%n damaged.
Error [%3]%n
RetCode [%4]
13026 FFIdPatInfoInput %1 (%2) : Patient Info Set The patient information to be registered Enter correct patient information or repair
Error%n contained illegal information or the patient the patient information database.
Error [%3]%n information database was damaged.
RetCode [%4]
13027 FFIdPatInfoInput %1 (%2) : Patient Info Not No matching patient information was
Correspond%n found.
RetCode [%3]%n
13027 FFIipInput %1%n%2%n%3%n%4%n Control information No particular remedy needs be taken.
%5%n%6%n%7%0
13030 FFIipInput %1%n%2%n%3%n%4%n The connection character string was not Reinstall.
%5%n%6%n%7%0 properly registered for the
“RuStatusDataDB” key in Directory.ini.
13031 FFIipInput [%s] %s (%s): Log Output The system setup database was damaged. Reinstall.
Directory Get Error
13032 FFIipInput [%s] %s (%s): MCR Control The card reader unit was not connected. There are no particular problems except
Initialize Error that the magnetic card reader is not
available.
13033 FFIipInput [%s] %s (%s): LR button The range was changed by the Service No particular remedy needs be taken.
cannot be concave set Utility.
13034 FFIipInput [%s] %s (%s): TB button The range was changed by the Service No particular remedy needs be taken.
cannot be concave set Utility.
13035 FFIipInput [%s] %s (%s): TBLR button The range was changed by the Service No particular remedy needs be taken.
cannot be concave set Utility.
13036 FFIipInput [%s] %s (%s): Rotation The range was changed by the Service No particular remedy needs be taken.
button cannot be concave Utility.
set
13037 FFIipInput [%s] %s (%s): Message The “MsgTableName” key was not found in Reinstall.
Table Name Get Error the system setup database.
13038 FFIipInput [%s] %s (%s): StrDB The “LangStrDB” key was not found in Reinstall.
Connection String Get Directory.ini.
Error
13039 FFIipInput [%s] %s (%s): StrDB The connection character string was not Reinstall.
Connect Error properly registered for the “LangStrDB” key
in Directory.ini.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-12
Event Event Source Description Causes Remedies Remarks
13040 FFIipInput [%s] %s (%s): SysDB The “SysConfigDB” key was not found in Reinstall.
Connection String Get Directory.ini.
Error
13041 FFIipInput [%s] %s (%s): SysDB The connection character string was not Reinstall.
Connect Error properly registered for the “SysConfigDB”
key in Directory.ini.
13042 FFIipInput [%s] %s (%s): ImageDB The “ImageDB“ key was not found in Reinstall.
Connection Character Directory.ini.
String Get Error
13043 FFIipInput [%s] %s (%s): Failed The connection character string was not Reinstall.
ImageDB Connection properly registered for the “ImageDB” key
in Directory.ini.
13044 FFIipInput [%s] %s (%s): Failed The image database was not successfully Reinstall.
ImageDB Open Recordset accessed.
13045 FFIipInput [%s] %s (%s): Failed The image database was not successfully Check whether the SQL service manager
ImageDB Close Recordset accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13046 FFIipInput [%s] %s (%s): ImageDB The image database was not successfully Check whether the SQL service manager
Data Get Error accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13047 FFIipInput [%s] %s (%s): ImageDB The image database was not successfully Check whether the SQL service manager
Data Set Error accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13048 FFIipInput [%s] %s (%s): ImageDB The image database was not successfully Check whether the SQL service manager
Data Delete Error accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13049 FFIipInput [%s] %s (%s): Failed The image database was not successfully Check whether the SQL service manager
ImageDBFlag Connection accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-13
Event Event Source Description Causes Remedies Remarks
13050 FFIipInput [%s] %s (%s): Failed The image database was not successfully Check whether the SQL service manager
ImageDBFlag Close accessed. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13051 FFIipInput [%s] %s (%s): The image database was not successfully Check whether the SQL service manager
ImageDBFlag Data Set accessed. is normally operating. If the same symptom
Error occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13052 FFIipInput [%s] %s (%s): Failed The initialization process for menu Reinstall.
MenuDB Initialize database access was not successfully
performed.
13053 FFIipInput [%s] %s (%s): Menu Data Menu information was not successfully Reinstall.
in ImageDB Get Error acquired from the menu database.
13054 FFIipInput [%s] %s (%s): Menu Menu parameter information was not Reinstall.
Parameter Data in successfully acquired from the menu
ImageDB Get Error database.
13055 FFIipInput [%s] %s (%s): Failed The termination process for menu Reinstall.
MenuDB Termination database access was not successfully
performed.
13056 FFIipInput [%s] %s (%s): NetDB The “NetConfigDB” key was not found in Reinstall.
Connection Character Directory.ini.
String Get Error
13057 FFIipInput [%s] %s (%s): Failed The connection character string was not Reinstall.
NetDB Connection properly registered for the “NetConfigDB”
key in Directory.ini.
13058 FFIipInput [%s] %s (%s): NetDB Data The network setup database was Reinstall.
Get Error damaged.
13059 FFIipInput [%s] %s (%s): Failed New DICOM data was not successfully Restart. If the problem frequently occurs,
Create New DICOM Data created. initialize the image database.
13060 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully Restart. If the problem frequently occurs,
Get Error acquired. initialize the image database.
13061 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully set up. Restart. If the problem frequently occurs,
Set Error initialize the image database.
13062 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully deleted. Restart. If the problem frequently occurs,
Delete Error initialize the image database.
13063 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully loaded. Restart. If the problem frequently occurs,
Load Error initialize the image database.
13064 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully saved. Restart. If the problem frequently occurs,
Save Error initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-14
Event Event Source Description Causes Remedies Remarks
13065 FFIipInput [%s] %s (%s): DICOM Data DICOM data was not successfully freed. Restart. If the problem frequently occurs,
Free Error initialize the image database.
13066 FFIipInput [%s] %s (%s): Failed in The “IdPadding” key was not found in the Reinstall.
Padding Acquisition[Id#] system setup database.
13067 FFIipInput [%s] %s (%s): Failed in The “ExaminationNoPadding” key was not Reinstall.
Padding found in the system setup database.
Acquisition[Exam#]
13068 FFIipInput [%s] %s (%s): Failed in The “DateFormat” key was not found in the Reinstall.
Format Acquisition[Date Of system setup database.
Birth]
13069 FFIipInput [%s] %s (%s): Cannot The requesting department database was Reinstall.
Access Data not found in the path indicated by the
Base[Departmdb] connection character string. The requesting
department table did not exist.
13070 FFIipInput [%s] %s (%s): Displayed The “FieldDepartList” key was not found in Reinstall.
Item Is Not the system setup database.
Understood[Depart List]
13072 FFIipInput [%s] %s (%s): Failed in The “DepartDB” key was not found in Reinstall.
Connected String Directory.ini.
Acquisition[Depart]
13073 FFIipInput [%s] %s (%s): Cannot The technician’s database was not found Reinstall.
Access Data in the path indicated by the connection
Base[Paramedicmdb] character string. The technician’s table did
not exist.
13074 FFIipInput [%s] %s (%s): Displayed The “FieldParamedicList” key was not Reinstall.
Item Is Not found in the system setup database.
Understood[Paramedic List]
13076 FFIipInput [%s] %s (%s): Failed in The “ParamedicDB” key was not found in Reinstall.
Connected String Directory.ini.
Acquisition[Paramedic]
13077 FFIipInput [%s] %s (%s): Failed in The registry was damaged. Reinstall.
Window Size Acquisition
13078 FFIipInput [%s] %s (%s): Failed in The “ExposureListErase” key was not Reinstall.
ExposureList Erase found in the system setup database.
13079 FFIipInput [%s] %s (%s): Config The “Config” key was not found in Reinstall.
Directory Character String Directory.ini.
Get Error
13080 FFIipInput [%s] %s (%s): The image database was damaged. Check whether the SQL service manager
ImageDBFlag Data Get is normally operating. If the same symptom
Error occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
13081 FFIipInput [%s] %s (%s): StrDB The character string database was Reinstall.
Message String Get Error damaged.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-15
Event Event Source Description Causes Remedies Remarks
13083 FFIipInput [%s] %s (%s): Initialize The system was unstable. Restart.
Timeout <InImg>
13084 FFIipInput [%s] %s (%s): Terminate The system was unstable. Restart.
Error <InImg>
13085 FFIipInput [%s] %s (%s): Terminate The system was unstable. Restart.
Timeout <InImg>
13086 FFIipInput [%s] %s (%s): Input The system was unstable. Restart.
Method Error <InImg>
13087 FFIipInput [%s] %s (%s): Original The system was unstable. Restart.
Image Input Failure
<InImg>
13088 FFIipInput [%s] %s (%s): Image The system was unstable. Restart.
Processing Trouble
<InImg>
13089 FFIipInput [%s] %s (%s): Menu Read The image database/information file was Reinstall.
Error <IdExposureList> damaged.
13090 FFIdDepartSelect %1 (%2) : Invalid Depart Directory.ini was damaged. Reinstall.
Code%n
Code [%3]
13090 FFIdParamedicSelect %1 (%2) : Invalid Directory.ini was damaged. Reinstall.
Paramedic Code%n
Code [%3]
13090 FFIdIndividualSelect %1 (%2) : Invalid Directory.ini was damaged. Repair Directory.ini.
Depart/Paramedic Code%n
Code [%3]
13091 FFIdDepartSelect %1 (%2) : Invalid Depart The database was damaged. Reinstall.
Name (SBCS)%n
Name (SBCS) [%3]
13091 FFIdParamedicSelect %1 (%2) : Invalid The database was damaged. Reinstall.
Paramedic Name
(SBCS)%n
Name (SBCS) [%3]
13091 FFIdIndividualSelect %1 (%2) : Invalid The database was damaged. Repair the database.
Depart/Paramedic
Name (SBCS)%n
Name (SBCS) [%3]
13094 FFIipInput [%s] %s (%s): Message The system was unstable. Restart.
Send Error <IpcInfCom>
13096 FFNetDB %1 An invalid input value was specified. Restart. Common software
%2 “%1”: Instance name
%3 (function name),
occurred error,
error code
“%2”: Error details (1/2)
“%3”: Error details (2/2)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-16
Event Event Source Description Causes Remedies Remarks
13097 FFNetDB %1 Invalid network configuration. Check network configuration using the Common software
%2 Service Utility. “%1”: Instance name
%3 (function name),
occurred error,
error code
“%2”: Error details (1/2)
“%3”: Error details (2/2)
13098 FF32LIB FF32lib A WIN32 system An error occurred when calling up the Restart.
library error occurred. WIN32 API.
[%1]%n%2
13098 FFIdPatInfoInput %1 (%2) : %n A general-use message for the input of If it is referring to an error, follow the
[%3]%n patient information (serves as an error custom message box that opens.
%4%n message and also as information).
%5
13099 FF32LIB FF32lib An unknown error An exception error occurred. Restart.
occurred. [%1]%n%2
13100 FFIipInput [%s] %s (%s): Failed in The “DisplayDataDB” key was not found in Reinstall.
Display Group Directory.ini or the connection character
Acquisition[Initialize] string was not properly registered in the
display group table.
13101 FFIipInput [%s] %s (%s): Failed in The “DisplayDataDB” key was not found in Reinstall.
Display Group Acquisition Directory.ini or the connection character
string was not properly registered in the
display group table.
13103 FFIipInput [%s] %s (%s): Failed in The “DisplayDataDB” key was not found in Reinstall.
Display Menu Directory.ini.
Acquisition[Initialize]
13104 FFIipInput [%s] %s (%s): Failed in The menu database was damaged. Reinstall.
Display Menu
Acquisition[Record]
13108 FFIipInput [%s] %s (%s): Failed in The menu database was damaged. Reinstall.
Exposure Menu Acquisition
13109 FFIipInput [%s] %s (%s): Next Set The registry was damaged. Reinstall.
Number Data Get Error
13110 FFIipInput [%s] %s (%s): Next Set The registry was damaged. Reinstall.
Number Data Set Error
13111 FFIipInput [%s] %s (%s): System ID The system setup database was damaged. Reinstall.
Data Get Error
13112 FFIipInput [%s] %s (%s): Auto The system setup database was damaged. Reinstall.
Distribution Data Get Error
13113 FFIipInput [%s] %s (%s): Enable The system setup database was damaged. Reinstall.
Selector Data Get Error
13114 FFIipInput [%s] %s (%s): Auto Menu The system setup database was damaged. Reinstall.
Select Data Get Error
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-17
Event Event Source Description Causes Remedies Remarks
13115 FFIipInput [%s] %s (%s): Next Set The system setup database was damaged. Reinstall.
Number Min Data Get Error
13116 FFIipInput [%s] %s (%s): Next Set The system setup database was damaged. Reinstall.
Number Max Data Get
Error
13120 FFIipInput [%s] %s (%s): Failed in The database was not found or the Reinstall.
Getting ID-Online Port# associated key did not exist.
13121 FFIipInput [%s] %s (%s): Failed in The database was not found or the Reinstall.
ID- Online Device Name associated key did not exist.
13123 FFIipInput [%s] %s (%s): Failed in The “SelectableDirection” key was not Reinstall.
selectable direction found in the system setup database or its
value was incorrect.
13124 FFIipInput [%s] %s (%s): Failed in The “SelectableFormat” key was not found Reinstall.
Selectable Image Format in the system setup database or its value
Acquisition was incorrect.
13125 FFIipInput [%s] %s (%s): FilmMarkDB The “FilmMarkDB“ key was not found in Reinstall.
Connection String Get Directory.ini.
Error
13126 FFIipInput [%s] %s (%s): FilmMarkDB The connection character string was not Reinstall.
Connect Error properly registered for the “FilmMarkDB“
key in Directory.ini.
13128 FFIipInput [%s] %s (%s): Failed in The “LimitFilmCopy” key was not found in Reinstall.
Limit Film Copy Acquisition the system setup database or its value was
incorrect.
13131 FFIipInput [%s] %s (%s): Failed in The “SelectableEdrMode” key was not Reinstall.
Selectable EDR Mode found in the system setup database or its
Acquisition value was incorrect.
13132 FFIipInput [%s] %s (%s): Failed in The “SelectableFunction” key was not Reinstall.
Selectable Function found in the system setup database or its
Acquisition value was incorrect.
13134 FFIipInput [%s] %s (%s): Failed in The “SelectableFilingMode” key was not Reinstall.
Selectable EDR Mode found in the system setup database or its
Acquisition value was incorrect.
13135 FFIipInput [%s] %s (%s): NetDB Data The connection character string for the Reinstall.
Get Error network setup database was not
successfully acquired or the network setup
database did not exist.
13136 FFIipInput [%s] %s (%s): Failed in The “EnableDistributionCode” key was not Reinstall.
Enable DistributionCode found in the system setup database or its
Acquisition value was incorrect.
13137 FFIipInput [%s] %s (%s): Failed in The “EnableFileString” key was not found Reinstall.
Enable FileString in the system setup database or its value
Acquisition was incorrect.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-18
Event Event Source Description Causes Remedies Remarks
13138 FFIipInput [%s] %s (%s): Failed in The “EnableFunction” key was not found in Reinstall.
Enable Function Acquisition the system setup database or its value was
incorrect.
13157 FFIipInput [%s] %s (%s): Finish Mode The system setup database was damaged. Reinstall.
Get Error
13161 FFIipInput [%s] %s (%s): Failed The program was faulty or the image file Restart. If the problem frequently occurs,
Display Thumbnail Image was damaged. initialize the image database.
13165 FFIipInput [%s] %s (%s): Image Size PU-dependent symptom. Check the PU status.
Code Specify Error
13166 FFIipInput [%s] %s (%s): Pantomo The system setup database was damaged. Repair the system setup database.
Setting Get Error
13177 FFIipInput [%s] %s (%s): Default The system setup database was damaged. Reinstall.
Menu Code Get Error
13178 FFIipInput [%s] %s (%s): Next Multi The registry was damaged. Reinstall.
Frame Pair Code Data Get
Error
13179 FFIipInput [%s] %s (%s): Next Multi The registry was damaged. Reinstall.
Frame Pair Code Data Set
Error
13180 FFIipInput [%s] %s (%s): Series The system setup database was damaged. Reinstall.
Generation Type Data Get
Error
13181 FFIipInput [%s] %s (%s): Multi Frame The system setup database was damaged. Reinstall.
UID Information Get Error
13184 FFIipInput [%s] %s (%s): MCR Type The system setup database was damaged. Reinstall.
Get Error
13185 FFIipInput [%s] %s (%s): MCR The system setup database was damaged. Reinstall.
Variation Get Error
13186 FFIipInput [%s] %s (%s): Custom The system setup database was damaged. Reinstall.
Card Use Get Error
13187 FFIipInput [%s] %s (%s): Font Name The “Font” key was not found in the Reinstall the IIP because the displayed
Get Error registry. characters can be corrupted.
13188 FFIipInput [%s] %s (%s): Font Size The “FontSizeSS”/“FontSizeS”/“FontSizeM”/ Displayed characters can be corrupted.
Get Error “FontSizeL”/“FontSizeLL” key was not Reinstall the IIP as necessary.
found in the registry.
13189 FFIipInput [%s] %s (%s): Multibyte The “Multibyte” key was not found in the Reinstall.
Flag Get Error registry.
13190 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Id#] Directory.ini or the ID “30700” was not
registered in the character string database.
13191 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Exam#] Directory.ini or the ID “30701” was not
registered in the character string database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-19
Event Event Source Description Causes Remedies Remarks
13192 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[SBCS Directory.ini or the ID “30702” was not
Patient Name] registered in the character string database.
13193 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[DBCS Directory.ini or the ID “30703” was not
Patient Name] registered in the character string database.
13194 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Sex] Directory.ini or the ID “30704” was not
registered in the character string database.
13195 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Sex M] Directory.ini or the ID “30705” was not
registered in the character string database.
13196 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Sex F] Directory.ini or the ID “30706” was not
registered in the character string database.
13197 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Sex O] Directory.ini or the ID “30707” was not
registered in the character string database.
13198 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Date Of Directory.ini or the ID “30708” was not
Birth] registered in the character string database.
13199 FFIipInput [%s] %s (%s): Failed in The “IdLength” key was not found in the Reinstall.
Limitation No system setup database.
Acquisition[Id#]
13200 FFIipInput [%s] %s (%s): Failed in The “ExaminationNoLength” key was not Reinstall.
Limitation No found in the system setup database.
Acquisition[Exam#]
13201 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Depart] Directory.ini or the ID “30800” was not
registered in the character string database.
13202 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30800” was not
Acquisition[Paramedic] registered in the character string database.
13203 FFIipInput [%s] %s (%s): Image The “Bitmap” key was not found in Reinstall.
Folder Path Get Error Directory.ini.
13204 FFIipInput [%s] %s (%s): Id_Clear_U.bmp/“Id_Clear_D.bmp”/ Reinstall the IIP because the button cannot
[Clear]Button Image Get “Id_Clear_X.bmp” was not found in the be displayed normally.
Error path to an image folder or the image format
was illegal.
13205 FFIipInput [%s] %s (%s): Id_Ok_U.bmp/“Id_Ok_D.bmp”/“Id_Ok_ Reinstall the IIP because the button cannot
[Ok]Button Image Get Error X.bmp” was not found in the path to an be displayed normally.
image folder or the image format was
illegal.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-20
Event Event Source Description Causes Remedies Remarks
13206 FFIipInput [%s] %s (%s): [Cancel] Id_Cancel_U.bmp/“Id_Cancel_D.bmp”/ Reinstall the IIP because the button cannot
Button Image Get Error “Id_Cancel_X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
13207 FFIipInput [%s] %s (%s): Id_Plus_U.bmp/“Id_Plus_D.bmp”/“Id_Plus_ Reinstall the IIP because the button cannot
[Plus]Button Image Get X.bmp” was not found in the path to an be displayed normally.
Error image folder or the image format was
illegal.
13208 FFIipInput [%s] %s (%s): [Minus] Id_Minus_U.bmp/“Id_Minus_D.bmp”/ Reinstall the IIP because the button cannot
Button Image Get Error “Id_Minus_X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
13209 FFIipInput [%s] %s (%s): Id_NextPage_U.bmp/“Id_NextPage_ Reinstall the IIP because the button cannot
[Next Page]Button Image D.bmp”/“Id_NextPage_X.bmp” was not be displayed normally.
Get Error found in the path to an image folder or the
image format was illegal.
13210 FFIipInput [%s] %s (%s): Id_PrevPage_U.bmp/“Id_PrevPage_ Reinstall the IIP because the button cannot
[Previous Page]Button D.bmp”/“Id_PrevPage_X.bmp” was not be displayed normally.
Image Get Error found in the path to an image folder or the
image format was illegal.
13217 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30202” was not
Acquisition[Complete] registered in the character string database.
13218 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[QA Directory.ini or the ID “30201” was not
Unfinished] registered in the character string database.
13219 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30200” was not
Acquisition[Interrupt] registered in the character string database.
13220 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Cancel] Directory.ini or the ID “30203” was not
registered in the character string database.
13221 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[New Directory.ini or the ID “30500” was not
Study] registered in the character string database.
13222 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30501” was not
Acquisition[Registered] registered in the character string database.
13223 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[QA Directory.ini or the ID “30502” was not
Waiting] registered in the character string database.
13224 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Output Directory.ini or the ID “30503” was not
Waiting] registered in the character string database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-21
Event Event Source Description Causes Remedies Remarks
13225 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30504” was not
Acquisition[Outputted] registered in the character string database.
13226 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[All] Directory.ini or the ID “30505” was not
registered in the character string database.
13227 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Today] Directory.ini or the ID “30506” was not
registered in the character string database.
13228 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Image Acquisition[System] Directory.ini or the ID “30507” was not
registered in the character string database.
13229 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30600” was not
Acquisition[StudyReseve] registered in the character string database.
13230 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “30601” was not
Acquisition[StudyStart] registered in the character string database.
13231 FFIipInput [%s] %s (%s): Failed in The bmp file path/bmp file was abnormal. Reinstall.
Image Acquisition[Priority]
13232 FFIipInput [%s] %s (%s): Failed in The bmp file path/bmp file was abnormal. Reinstall.
Image Acquisition[End]
13233 FFIipInput [%s] %s (%s): Failed in The bmp file path/bmp file was abnormal. Reinstall.
Image Acquisition[Change]
13234 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[New]Button Image Get
Error
13235 FFIipInput [%s] %s (%s): [Repeat] The bmp file path/bmp file was abnormal. Reinstall.
Button Image Get Error
13236 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Again]Button Image Get
Error
13237 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Add]Button Image Get
Error
13238 FFIipInput [%s] %s (%s): [Delete] The bmp file path/bmp file was abnormal. Reinstall.
Button Image Get Error
13239 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Prev]Button Image Get
Error
13240 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Next]Button Image Get
Error
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-22
Event Event Source Description Causes Remedies Remarks
13241 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Monitor Mode]Button
Image Get Error
13242 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[QA]Button Image Get Error
13243 FFIipInput [%s] %s (%s): [Change] The bmp file path/bmp file was abnormal. Reinstall.
Button Image Get Error
13244 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[AP/PA]Button Image Get
Error
13245 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[EDR Mode]Button Image
Get Error
13246 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Film Format]Button Image
Get Error
13247 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Detail]Button Image Get
Error
13248 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[New]Button ToolTipText damaged.
Get Error
13249 FFIipInput [%s] %s (%s): [Repeat] The character string database was Reinstall.
Button ToolTipText Get damaged.
Error
13250 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Again]Button ToolTipText damaged.
Get Error
13251 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Add]Button ToolTipText damaged.
Get Error
13252 FFIipInput [%s] %s (%s): [Delete] The character string database was Reinstall.
Button ToolTipText Get damaged.
Error
13253 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Prev]Button ToolTipText damaged.
Get Error
13254 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Next]Button ToolTipText damaged.
Get Error
13255 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Monitor Mode]Button damaged.
ToolTipText Get Error
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-23
Event Event Source Description Causes Remedies Remarks
13256 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[QA]Button ToolTipText Get damaged.
Error
13257 FFIipInput [%s] %s (%s): [Change] The character string database was Reinstall.
Button ToolTipText Get damaged.
Error
13258 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[AP/PA]Button ToolTipText damaged.
Get Error
13259 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[EDR Mode]Button damaged.
ToolTipText Get Error
13260 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Film Format]Button damaged.
ToolTipText Get Error
13261 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Detail]Button ToolTipText damaged.
Get Error
13262 FFIipInput [%s] %s (%s): Failed in The bmp file path/bmp file was abnormal. Reinstall.
Image Acquisition[New]
13263 FFIipInput [%s] %s (%s): Failed in Id_NormalImage_U.bmp/ Reinstall the IIP because the button cannot
Image Acquisition[Normal] “Id_NormalImage_ D.bmp”/ be displayed normally.
“Id_NormalImage_X.bmp” was not found in
the path to an image folder or the image
format was illegal.
13264 FFIipInput [%s] %s (%s): Failed in Id_LR_Reverse_U.bmp/“Id_LR_Reverse_ Reinstall the IIP because the button cannot
Image Acquisition[LR] D.bmp”/“Id_LR_Reverse_X.bmp” was not be displayed normally.
found in the path to an image folder or the
image format was illegal.
13265 FFIipInput [%s] %s (%s): Failed in Id_TB_Reverse_U.bmp/“Id_TB_Reverse_ Reinstall the IIP because the button cannot
Image Acquisition[TB] D.bmp”/“Id_TB_Reverse_X.bmp” was not be displayed normally.
found in the path to an image folder or the
image format was illegal.
13266 FFIipInput [%s] %s (%s): Failed in Id_Rotation180_U.bmp/ Reinstall the IIP because the button cannot
Image Acquisition[TB LR] “Id_Rotation180_D.bmp”/ be displayed normally.
“Id_Rotation180_X.bmp” was not found in
the path to an image folder or the image
format was illegal.
13267 FFIipInput [%s] %s (%s): Failed in Id_Rotation90_U.bmp/“Id_Rotation90_ Reinstall the IIP because the button cannot
Image Acquisition[Rotation] D.bmp”/“Id_Rotation90_X.bmp” was not be displayed normally.
found in the path to an image folder or the
image format was illegal.
13268 FFIipInput [%s] %s (%s): Failed in Id_Single_U.bmp/“Id_Single_D.bmp”/ Reinstall the IIP because the button cannot
Image Acquisition[Single] “Id_Single_X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-24
Event Event Source Description Causes Remedies Remarks
13269 FFIipInput [%s] %s (%s): Failed in Id_Twin_U.bmp/“Id_Twin_D.bmp”/ Reinstall the IIP because the button cannot
Image Acquisition[Twin] “Id_Twin_X.bmp” was not found in the path be displayed normally.
to an image folder or the image format was
illegal.
13270 FFIipInput [%s] %s (%s): An Illegal The film mark database was damaged. Reinstall.
Film Mark Character Was
Detected
13271 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[FIX] Directory.ini or the ID “33502” was not
registered in the character string database.
13272 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Skfix] Directory.ini or the ID “33503” was not
registered in the character string database.
13273 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[SEMI- Directory.ini or the ID “33504” was not
X] registered in the character string database.
13274 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition Directory.ini or the ID “33505” was not
[DensityAverage] registered in the character string database.
13275 FFIipInput [%s] %s (%s): Id_Utility_U.bmp/“Id_Utility_D.bmp”/ Reinstall the IIP because the button cannot
[Utility]Button Image Get “Id_Utility_X.bmp” was not found in the be displayed normally.
Error path to an image folder or the image format
was illegal.
13276 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[Alarm Directory.ini or the ID “32700” was not
Stop] registered in the character string database.
13277 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption Acquisition[Exit] damaged.
13278 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption Acquisition[User damaged.
Utility]
13279 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption Acquisition[List damaged.
Setting]
13280 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption damaged.
Acquisition[Version]
13281 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption damaged.
Acquisition[CANCEL]
13282 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Clear]Button Image Get
Error
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-25
Event Event Source Description Causes Remedies Remarks
13283 FFIipInput [%s] %s (%s): The character string database was Reinstall.
[Clear]Button ToolTipText damaged.
Get Error
13284 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Directory.ini or the ID “35000” was not
Acquisition[CANCEL] registered in the character string database.
13285 FFIipInput [%s] %s (%s): Failed in The “LangStrDB” key was not found in Reinstall.
Caption Acquisition[DUST] Directory.ini or the ID “35001” was not
registered in the character string database.
13286 FFIipInput [%s] %s (%s): Tooltip Text The character string database was Reinstall.
Get Error damaged.
13287 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption damaged.
Acquisition[TitleBar]
13288 FFIipInput [%s] %s (%s): Failed in The character string database was Reinstall.
Caption Acquisition[Option] damaged.
13289 FFIipInput [%s] %s (%s): Version Info The registry was damaged. Reinstall.
Get Error
13290 FFIipInput [%s] %s (%s): Option List The registry was damaged. Reinstall.
Get Error
13293 FFIipInput [%s] %s (%s): The bmp file path/bmp file was abnormal. Reinstall.
[Miss]Image Get Error
13294 FFIipInput [%s] %s (%s): Id_ApPa_U.bmp/“Id_ApPa_D.bmp”/ Reinstall the IIP because the button cannot
[Image Style]Button Image “Id_ ApPa_ X.bmp” was not found in the be displayed normally.
Get Error path to an image folder or the image format
was illegal.
13295 FFIipInput [%s] %s (%s): Id_Edr_U.bmp/“Id_Edr_D.bmp”/ Reinstall the IIP because the button cannot
[EDR]Button Image Get “Id_Edr_ X.bmp” was not found in the path be displayed normally.
Error to an image folder or the image format was
illegal.
13296 FFIipInput [%s] %s (%s): [Function] Id_Edr_U.bmp/“Id_Edr_D.bmp”/“Id_Edr_ Reinstall the IIP because the button cannot
Button Image Get Error X.bmp” was not found in the path to an be displayed normally.
image folder or the image format was
illegal.
13297 FFIipInput [%s] %s (%s): Id_FilmCount_U.bmp/“Id_FilmCount_ Reinstall the IIP because the button cannot
[Film Count]Button Image D.bmp”/“Id_FilmCount_X.bmp” was not be displayed normally.
Get Error found in the path to an image folder or the
image format was illegal.
13299 FFIipInput [%s] %s (%s): Id_FilmFormat_U.bmp/“Id_FilmFormat_ Reinstall the IIP because the button cannot
[Image Format]Button D.bmp”/“Id_FilmFormat_X.bmp” was not be displayed normally.
Image Get Error found in the path to an image folder or the
image format was illegal.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-26
Event Event Source Description Causes Remedies Remarks
13300 FFIipInput [%s] %s (%s): Id_FilingMode_U.bmp/“Id_FilingMode_ Reinstall the IIP because the button cannot
[Filing]Button Image Get D.bmp”/“Id_FilingMode_X.bmp” was not be displayed normally.
Error found in the path to an image folder or the
image format was illegal.
13301 FFIipInput [%s] %s (%s): Id_FilmMark_U.bmp/“Id_FilmMark_D. Reinstall the IIP because the button cannot
[Film Mark]Button Image bmp”/“Id_ FilmMark_X.bmp” was not found be displayed normally.
Get Error in the path to an image folder or the image
format was illegal.
13302 FFIipInput [%s] %s (%s): Id_DistributionCode_U.bmp/ Reinstall the IIP because the button cannot
[Distribution Code]Button “Id_DistributionCode_D.bmp”/ be displayed normally.
Image Get Error “Id_DistributionCode_ X.bmp” was not
found in the path to an image folder or the
image format was illegal.
13303 FFIipInput [%s] %s (%s): [CSSS] Id_CsCs_U.bmp/“Id_CsCs_D.bmp”/ Reinstall the IIP because the button cannot
Button Image Get Error “Id_CsCs_ X.bmp” was not found in the be displayed normally.
path to an image folder or the image format
was illegal.
13304 FFIipInput [%s] %s (%s): Image Data The line was abnormal. Check the line.
Connection Disconnected
13309 FFIipInput [%s] %s (%s): Failed in The patient information contained an illegal Ensure that the patient information does
Patient Information character string. not contain any illegal character string.
13345 FFIdDepartSelect %1 (%2) : Database Data The database was damaged. Reinstall.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
13345 FFIdFilmMark %1 (%2) : Database Data The database was damaged. Reinstall.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
13345 FFIdMcrControl IdMcrControl The database was faulty. Reinstall.
(threadMcrRead) :
Database Data Get Error
13345 FFIdParamedicSelect %1 (%2) : Database Data The database was damaged. Reinstall.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
13345 FFIdPatInfoInput %1 (%2) : Database Data The database was damaged. Reinstall.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-27
Event Event Source Description Causes Remedies Remarks
13345 FFIipInput %1 (%2) : Database Data The database was damaged. Reinstall.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
13345 FFIdIndividualSelect %1 (%2) : Database Data The database was damaged. Repair the database.
Get Error%n
Database [%3]%n
Table [%4]%n
Key [%5]
13346 FFIipInput %1 (%2) : Patient The patient information contained illegal Correct the patient information setup.
Information Check Error%n information.
Item [%3]%n
Error [%4]
13347 FFIipInput %1 (%2) : Caption Get The character string database was Reinstall.
Error%n damaged.
CaptionId [%3]
13348 FFIdDepartSelect %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIdFilmMark %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIdMenuChange %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIdParamedicSelect %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIdPatInfoInput %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIipInput %1 (%2) : Registry Data The registry was damaged. Reinstall.
Get Error%n
Value [%3]
13348 FFIdIndividualSelect %1 (%2) : Registry Data The registry was damaged. Repair the registry.
Get Error%n
13349 FFIdDepartSelect %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
13349 FFIdFilmMark %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-28
Event Event Source Description Causes Remedies Remarks
13349 FFIdMenuChange %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
13349 FFIdParamedicSelect %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
13349 FFIdPatInfoInput %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
13349 FFIipInput %1 (%2) : INI File Data Get Directory.ini was damaged. Reinstall.
Error%n
Key [%3]
13349 FFIdIndividualSelect %1 (%2) : INI File Data Get Registry.ini was damaged. Repair Registry.ini.
Error%n
Key [%3]
13350 FFIdFilmMark %1 (%2) : Bitmap Image The bmp file path or bmp file was Reinstall.
Get Error%n abnormal.
BMP File [%3]
13350 FFIdMenuChange %1 (%2) : Bitmap Image The bmp file path or bmp file was Reinstall.
Get Error%n abnormal.
BMP File [%3]
13350 FFIipInput %1 (%2) : Bitmap Image The bmp file path or bmp file was Reinstall.
Get Error%n abnormal.
BMP File [%3]
13352 FFIipInput %1 (%2) : Dicom MWM & The DICOM MWM & PPS option was not Install the DICOM MWM & PPS option.
PPS Option Not Install installed.
13353 FFIipInput %1 (%2) : Dicom MWM The DICOM MWM option was not installed. Install the DICOM MWM option.
Option Not Install
13362 FFIipInput %1 (%2) : Not Entry Host The host name for the local enclosure was With the service utility, register the host
Name not successfully acquired. information about the PPS local enclosure.
13370 FFIipInput %1 (%2) : PPS Request The equipment connection was faulty or Retry.
Failure [IN PROGRESS] the remote equipment power was OFF.
13371 FFIipInput %1 (%2) : PPS Request The equipment connection was faulty or Retry.
Failure [COMPLETED] the remote equipment power was OFF.
13372 FFIipInput %1 (%2) : PPS Request The equipment connection was faulty or Retry.
Failure [DISCONTINUED] the remote equipment power was OFF.
13382 FFIipInput %1 (%2) : Study List Retry.
Refresh Exception%nError
[%3]
13386 FFIipInput %1 (%2) : Failed in data The serial cable was broken or the Reinstall.
reception%n[Online Time connected equipment did not respond.
Out]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-29
Event Event Source Description Causes Remedies Remarks
13389 FFIdDepartSelect %1 (%2) : Database The database was damaged. Reinstall.
Connection Error%n
Database [%3]%n
13389 FFIdFilmMark %1 (%2) : Database The database was damaged. Reinstall.
Connection Error%n
Database [%3]%n
13389 FFIdParamedicSelect %1 (%2) : Database The database was damaged. Reinstall.
Connection Error%n
Database [%3]%n
13389 FFIdPatInfoInput %1 (%2) : Database The database was damaged. Reinstall.
Connection Error%n
Database [%3]%n
13389 FFIipInput %1 (%2) : Database The database was damaged. Reinstall.
Connection Error%n
Database [%3]%
13389 FFIdIndividualSelect %1 (%2) : Database The database was damaged. Repair the database.
Connection Error%n
Database [%3]%n
13392 FFIipInput %1 (%2) : BCR Control The barcode reader was not connected. Connect the barcode reader.
Initialize Error
13443 FFIipInput %1 (%2) : Cannot Change An attempt was made to change the Change the information after completion of
Patient Info patient information during an image input image input.
EditMode [%3] sequence (patient information change
button: inactive).
13444 FFIipInput %1 (%2) : Cannot End An attempt was made to terminate a study Terminate the study after completion of
Examination during an image input sequence (study image input.
ExamEndMode [%3] end button: inactive).
13447 FFIipInput %1 : IipInput.exe [Starts] IipInput.exe is started (information).
13448 FFIipInput %1 : IipInput.exe [Normal IipInput.exe is terminated (information).
Termination]
13449 FFIipInput %1 : IipInput.exe [Forced IipInput.exe is terminated (information).
Termination]
13457 FFIipInput %1 (%2) : Image This error occurs if the image processing
Processing Information parameter file does not exist.
<InImg>
13462 FFIipInput %1 (%2) : Ordering The exposure menu received as the Register the exposure menu with the user
Exposure Menu Not Entry ordering information was not registered. utility.
[%3]
13470 FFIipInput %1 (%2) : Exposure Menu The exposure menu received as the Register the exposure menu with the user
Register Error. ordering information was not registered. utility.
13471 FFIipInput %1 (%2) : All Exposure The exposure menu received as the Register the exposure menu with the user
Menu Register Error. ordering information was not registered. utility.
13475 FFIipInput %1 (%2) : Failed in The patient exposure order information
Ordering Data.[RIS] sent from the RIS was abnormal.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-30
Event Event Source Description Causes Remedies Remarks
13477 FFIipInput %1 (%2) : Exposure Menu The exposure menu specified by the Register the exposure menu with the user
Register Error. received ordering information was not utility.
registered.
13478 FFIipInput %1 (%2) : All Exposure The exposure menu specified by the Register the exposure menu with the user
Menu Register Error. received ordering information was not utility.
registered.
13479 FFIipInput %1 (%2) : Failed in The order exposure study termination Check the RIS settings.
Examination End signal was received from the RIS although
Command.[Ris] an order exposure was not received.
13480 FFIipInput %1 (%2) : Failed in The problem is attributable to a Check the equipment and cables.
Communication with RIS. disconnected line or invalid parameter.
[%3]%n[%4]
13483 FFIipInput %1 (%2) : Menu is not An ID acquisition request was received Make a menu selection and then press the
Selected. from built-in type equipment although no OK button on the built-in type Console.
Failed in Getting ID exposure menu was selected.
Information
13484 FFIipInput %1 (%2) : Failed in the Inappropriate installation. Reinstall the system to secure matching
System Start. between DX option and menu database.
[Error: OpDcmStandardDx
or MenuData.mdb]
13489 FFIipInput %1 (%2) : Patient Info Set Patient information to be registered Enter correct patient information or repair
Error%n included illegal information items or patient patient information database.
Error [%3]%n information database was damaged.
RetCode [%4]
13490 FFIipInput %1 (%2) : UserName Illegal information was input as the Call upon the user to input a correct
Check Error%n username. username.
Item [%3]%n
Error [%4]
13491 FFIipInput User Login [UserName:%1] The user logged in the system.
13492 FFIipInput User Logout The user logged off the system.
[UserName:%1
13493 FFIipInput %1 (%2) : MCR-Data Send Failed in transferring magnetic card Repeat the same operation. If the symptom
Error.[RIS] information to the FRIS. remains uncorrected, check the barcode
information.
13494 FFIipInput %1 (%2) : BCR-Data Send Failed in transferring barcode information Repeat the same operation. If the symptom
Error.[RIS] to the FRIS. remains uncorrected, check the barcode
information.
13495 FFIipInput [%1] Failed in setting MCR Failed in setting magnetic card information. Repeat the same operation. If the symptom
data [%2] remains uncorrected, check the magnetic
card information.
13496 FFIipInput [%1] Failed in setting BCR Failed in setting barcode information. Repeat the same operation. If the symptom
data [%2] remains uncorrected, check the barcode
information.
13497 FFIipInput %1 (%2) : Failed in deleting Program error.
multibyte data
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-31
Event Event Source Description Causes Remedies Remarks
13901 IIPInputEz %1 : Registry Value The acquisition/setup key or registration Check the registry key.
Get/Set Error%n information was abnormal.
[Key] %2%n
[Exp] %3%n
%4%n
%5
13902 IIPInputEz %1 : Database Connection This error is attributable to an abnormal Check Directory.ini or database file name.
Error%n character string in Directory.ini, an incorrect
[DB] %2%n database file name, or a damaged
[Exp] %3%n database.
%4%n
%5
13903 IIPInputEz %1 : Database Data The acquisition/setup key or database was Check the database table/field name.
Get/Set Error%n abnormal.
[DB] %2%n
[Table] %3%n
[Field] %4%n
[Exp] %5
13903 IIPInputEz %1 : Dicom Data Object The DDO acquisition/setup tag was Check the DDO tag.
Get/Set Error%n abnormal.
[Tag] %2%n
[Exp] %3%n
%4%n
%5
13904 IIPInputEz %1 : Other Error%n Unknown error.
[Procedure] %2%n
[ErrorCode] %3%n
[Exp] %4%n
%5
13998 FFIipInput %1 (%2) : [%3] General information event log.
13998 FFIipInput SendToCommPort, COMM The data transmission port is defective. Check connection and settings of the port.
Port Send Error
13998 FFIipInput m_hQUEMutex, Create An exclusive processing error occurred. Restart.
Mutex Error
13998 FFIipInput CommOpen, COMM Port A failure resulted in an attempt to establish Check connection and settings of the port.
Open Error connection to the data transmission port.
13998 FFIipInput CommOpen, COMM Port Settings of the data transmission port are Check settings for connection to the port.
Param Set Error defective.
13998 FFIipInput tControlMain, _ A failure occurred in an attempt to start the Restart.
beginthreadex Create Error main thread.
13998 FFIipInput End_ControlProc, COMM A failure occurred in an attempt to close No particular remedial measures need to
Port Close Error the port. be taken as the DR-ID 300CL is in process
of shutdown.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-32
Event Event Source Description Causes Remedies Remarks
13998 FFIipInput MakeSendMessage, The received data is defective. Check the connected cable and device.
Message Code Convert
Error
13998 FFIipInput lFocus, The “Focus” data received is defective. Check the equipment targeted for
DicomDataFormatError connection or restart it.
13998 FFIipInput lTubeVoltage, The “Tube Voltage” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput lExposureTime, The “Exposure Time” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput lmAs, The “mAs” data received is defective. Check the equipment targeted for
DicomDataFormatError connection or restart it.
13998 FFIipInput csGrid, The “Grid” data received is defective. Check the equipment targeted for
DicomDataFormatError connection or restart it.
13998 FFIipInput lGlandulardose, The “Glandular Dose” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput csFilterMaterial, The “Filter Material” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput BodyPartThickness, The “Body Part Thickness” data received Check the equipment targeted for
DicomDataFormatError is defective. connection or restart it.
13998 FFIipInput csActualStereoAngle, The “Actual Stereo Angle” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput csTubeHeadAngle, The “Tube Head Angle” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput lmAsActualValue, The “Actual Value” data received is Check the equipment targeted for
DicomDataFormatError defective. connection or restart it.
13998 FFIipInput SaveResExpParamData, An attempt to store a DICOM file resulted Restart.
DicomFile Save Error in a failure.
13998 FFIipInput SaveResExpParamData, An attempt to obtain information on the Restart.
Not Use Dicom Directory folder targeted for storing a DICOM file
resulted in a failure.
13998 FFIipInput tComRecvMain, Sub Tread A attempt to create a data reception thread Restart.
Create Error resulted in a failure.
13998 FFIipInput MainTH_MainPrc, An attempt to obtain an area resulted in a Restart.
Received from Port [Alloc failure.
Error]
13998 FFIipInput MainTH_MainPrc, The received data is defective. Check the equipment targeted for
Received Buffer Claer connection or restart it.
13998 FFIipInput RespOperation, response The received data is defective. Check the equipment targeted for
data Format error connection or restart it.
13998 FFIipInput RespOperation, response A checksum error was found in the Check the equipment targeted for
data Checksum error received data. connection or restart it.
13998 FFIipInput RespOperation, Alloc Error An attempt to obtain an area resulted in a Restart.
for Dicom Data failure.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-33
Event Event Source Description Causes Remedies Remarks
13998 FFIipInput ACKOperation, No Target Illegal ACK data was received. Nothing particular.
[ACK] Receive
13998 FFIipInput SendToCommPort, Send Data transmission resulted in a failure. Check the equipment targeted for
Error connection or restart it.
13998 FFIipInput RetryOperation, Retry Send Data retransmission processing resulted in Check the equipment targeted for
Error a failure. connection or restart it.
13998 FFIipInput RetryOperation, Comm An attempt to retransmit data resulted in a Check the equipment targeted for
Send Retry Over failure. connection or restart it.
13998 FFIipInput RetryOperation, Response A timeout error occurred while in wait of Check the equipment targeted for
Timeout data reception. connection or restart it.
14900 IIPInputEz %1 : Registry Value Get/Set An acquisition/setup key error. A registry Check the registry key.
Error%n information error.
[Key] %2%n
[Exp] %3%n
%4%n
%5
14900 IIPPortable %1 : Registry Value Get/ A registry was damaged. There is no particular problem with
Set Error operations.
[Key] %2
[Exp] %3
%4
%5
14901 IIPInputEz %1 : Database Connection Incorrect Directory.ini file character strings. Check the Directory.ini file. Also check the
Error%n An incorrect database filename. Damaged database filename.
[DB] %2%n database.
[Exp] %3%n
%4%n
%5
14901 IIPPortable %1 : Database Connection Failed in accessing the image database. There is no particular problem with
Error operations.
[DB] %2
[Exp] %3
%4
%5
14902 IIPInputEz %1 : Database Data An acquisition/setup key error. A database Check the database table/filed name.
Get/Set Error%n error.
[DB] %2%n
[Table] %3%n
[Field] %4%n
[Exp] %5
14902 IIPPortable %1 : Database Data Get/ Failed in accessing the image database. There is no particular problem with
Set Error operations.
[DB] %2
[Table]%3
[Field]%4
[Exp]%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-34
Event Event Source Description Causes Remedies Remarks
14903 IIPInputEz %1 : DDO Data Get/Set A DDO acquisition/setup tag error. Check the DDO tag.
Error%n
[Tag] %2%n
%3%n
%4%n
%5
14903 IIPPortable %1 : DDO Data Get/Set The DDO file was damaged. There is no particular problem with
Error operations.
[Tag] %2
%3
%4
%5
14904 IIPInputEz %1 : Other Error%n Other error.
[Procedure] %2%n
[ErrorCode] %3%n
[Exp] %4%n
%5
14904 IIPPortable %1: Other Error A general-purpose information event log. There is no particular problem with
[Procedure] %2 operations.
[Error Code]%3
[Exp]%4
%5
19001 FFDICOMCommit SCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMCommit SCU Initialization failed. DPC thread generation was unsuccessful. Restart.
[DPC Invoke fail] The system was unstable.
19001 FFDICOMCommit SCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19001 FFDICOMEchoSCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMEchoSCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19001 FFDICOMMPPSSCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMMPPSSCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19001 FFDICOMMWLSCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMMWLSCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19001 FFDICOMPrintSCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMPrintSCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-35
Event Event Source Description Causes Remedies Remarks
19001 FFDICOMStoreSCP Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMStoreSCP Initialization failed. DPC thread generation was unsuccessful. Restart.
[DPC Invoke fail] The system was unstable.
19001 FFDICOMStoreSCP Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19001 FFDICOMStoreSCU Initialization failed. DPM thread generation was unsuccessful. Restart.
[DPM Invoke fail] The system was unstable.
19001 FFDICOMStoreSCU Initialization failed. Instance generation was unsuccessful. Restart.
[Insufficient memory] The system was unstable.
19002 FFDICOMCommit SCU Method Invoke failed. A message was not successfully Restart.
[PushPrc] transmitted.
The system was unstable.
19002 FFDICOMCommit SCU Method Invoke failed. A message was not successfully Restart.
[DiscontinueRequest] transmitted.
The system was unstable.
19002 FFDICOMEchoSCU Method Invoke failed. A message was not successfully Restart.
[EchoPrc] transmitted.
The system was unstable.
19002 FFDICOMMPPSSCU Method Invoke failed. A message was not successfully Restart.
[CreatePrc] transmitted.
The system was unstable.
19002 FFDICOMMPPSSCU Method Invoke failed. A message was not successfully Restart.
[SetPrc] transmitted.
The system was unstable.
19002 FFDICOMMWLSCU Method Invoke failed. A message was not successfully Restart.
[FindPrc] transmitted.
The system was unstable.
19002 FFDICOMMWLSCU Method Invoke failed. A message was not successfully Restart.
[CancelPrc] transmitted.
The system was unstable.
19002 FFDICOMQRSCU Method Invoke failed. A message was not successfully Restart.
[FindPrc] transmitted.
The system was unstable.
19002 FFDICOMQRSCU Method Invoke failed. A message was not successfully Restart.
[MovePrc] transmitted.
The system was unstable.
19002 FFDICOMQRSCU Method Invoke failed. A message was not successfully Restart.
[CancelPrc] transmitted.
The system was unstable.
19002 FFDICOMQRSCU Method Invoke failed. A message was not successfully Restart.
[FreePrc] transmitted.
The system was unstable.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-36
Event Event Source Description Causes Remedies Remarks
19002 FFDICOMPrintSCU Method Invoke failed. A message was not successfully Restart.
[PrintPrc] transmitted.
The system was unstable.
19002 FFDICOMPrintSCU Method Invoke failed. A message was not successfully Restart.
[GetPrc] transmitted.
The system was unstable.
19002 FFDICOMPrintSCU Method Invoke failed. A message was not successfully Restart.
[DiscontinueRequest] transmitted.
The system was unstable.
19002 FFDICOMStoreSCU Method Invoke failed. A message was not successfully Restart.
[StorePrc] transmitted.
The system was unstable.
19003 FFDICOMCommit SCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMEchoSCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMMPPSSCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMMWLSCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMPrintSCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMStoreSCP File I/O failed. An adequate disk space was not available. Check the available disk space.
[Communication log file
name]
19003 FFDICOMStoreSCP File I/O failed. An image-related file was not successfully Restart. If the same symptom frequently
[Communication log file accessed. The problem is attributable to a occurs, replace the disk.
name] disk.
19003 FFDICOMStoreSCP File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19003 FFDICOMStoreSCU File I/O failed. A communication log file was not Although the communication log cannot be
[Communication log file successfully opened. obtained, execution is possible.
name]
19900 IIPInputEz %1 : Registry Value The acquisition/setup key or registration Check the registry key.
Get/Set Error%n information was abnormal.
[Key] %2%n
[Exp] %3%n
%4%n
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-37
Event Event Source Description Causes Remedies Remarks
19901 IIPInputEz %1 : Database Connection This error is attributable to an abnormal Check Directory.ini or database file name.
Error%n character string in Directory.ini, an incorrect
[DB] %2%n database file name, or a damaged
[Exp] %3%n database.
%4%n
%5
19902 IIPInputEz %1 : Database Data The acquisition/setup key or database was Check the database table/field name.
Get/Set Error%n abnormal.
[DB] %2%n
[Table] %3%n
[Field] %4%n
[Exp] %5
19903 IIPInputEz %1 : Dicom Data Object The DDO acquisition/setup tag was Check the DDO tag.
Get/Set Error%n abnormal.
[Tag] %2%n
[Exp] %3%n
%4%n
%5
19904 IIPInputEz %1 : Other Error%n Unknown error.
[Procedure] %2%n
[ErrorCode] %3%n
[Exp] %4%n
%5
20001 FFCtrlProc It failed in CreateProcess. A subprocess was not successfully started Restart. If the same symptom occurs after A subsystem name is
It has the possibility that an up. The system was unstable or a a restart, reinstall. placed in the “%1”
executive file doesn’t exist. necessary executable file was not found. position.
The start treatment can’t be
continued. /
CtrlProc:RunProc()
%1
20002 FFCtrlProc Process start information is An illegal parameter was used for process Restart. If the same symptom occurs after
unusual. The start startup. It is conceivable that the contents a restart, reinstall.
treatment can’t be of the process information database were
continued. / wrong.
CtrlProc:RunProc()
20003 FFCtrlProc It failed in the completion A process end request message was not Restart.
message preparation successfully created. It is conceivable that
treatment. The treatment the available resource was not adequate
can’t be continued. or that the system was unstable.
/CtrlProc:EndProc()
20004 FFCtrlProc It failed in the completion A process end request message was not Restart.
message transmitting successfully transmitted. It is conceivable
treatment. The treatment that the transmission destination process
can’t be continued. was down or that the transmission
/CtrlProc:EndProc() destination was wrong.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-38
Event Event Source Description Causes Remedies Remarks
20005 FFCtrlProc It failed in the switching A window show/hide message was not Restart.
message preparation successfully created. It is conceivable that
treatment. The treatment the available resource was not adequate
can’t be continued. or that the system was unstable.
/CtrlProc:ChangeProc()
20006 FFCtrlProc It failed in the switching A window show/hide message was not Restart. If the same symptom occurs after
message transmitting successfully transmitted. It is conceivable a restart, reinstall.
treatment. The treatment that the transmission destination process
can’t be continued. was down or that the transmission
/CtrlProc:ChangeProc() destination was wrong.
20007 FFCtrlProc It failed in CreateProcess. A window show/hide message was not Restart. If the same symptom occurs after
It has the possibility that an successfully transmitted. It is conceivable a restart, reinstall.
executive file doesn’t exist. that the transmission destination process
The start treatment can’t be was down or that the transmission
continued. destination was wrong.
/CtrlProc:SetInitProcEnd()
20102 FFCustomMsgBox [20102] Setup not It is conceivable that the application key Install the application key. Main
completed. was not installed.
20104 FFCustomMsgBox [20104] Failed in securing It is conceivable that the file was damaged Restart. If the same symptom occurs after Main
system settings. or deleted. a restart, reinstall.
20104 FFIIPMAIN It failed in the acquisition of The configuration information was not Restart. If the same symptom recurs,
the configuration information. successfully acquired. This error occurred perform a reinstallation.
The initialization treatment due to an improper installation or system
can’t be continued. instability.
/MainProcess:Main()
20105 FFCustomMsgBox [20105] Failed in accessing It is conceivable that the file was damaged Restart. If the same symptom occurs after Main
the process control or deleted. a restart, reinstall.
database.
20105 FFIIPMAIN It failed in the acquisition of Data was not successfully acquired from Restart. If the same symptom recurs,
the process information the process information database. perform a reinstallation.
data base. The initialization This error occurred due to an improper
treatment can’t be installation or system instability.
continued. /
MainProcess:Main()
20106 FFCustomMsgBox [20106] Failed in accessing It is conceivable that the file was damaged Restart. If the same symptom occurs after Main
the message distribution or deleted. a restart, reinstall.
control database.
20106 FFIIPMAIN It failed in opening of the The message database was not Restart. If the same symptom recurs,
message data base. successfully opened. This error occurred perform a reinstallation.
The initialization treatment due to an improper installation or system
can’t be continued. instability.
/MainProcess:Main()
20108 FFCustomMsgBox [20108] Failed in process A subprocess was not successfully started View the event logs around to examine the Main
activation. up for some reason. cause.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-39
Event Event Source Description Causes Remedies Remarks
20108 FFIIPMAIN It failed in the start of the The subprocess was not successfully Restart. If the same symptom recurs,
sub-process. started. This error occurred due to an perform a reinstallation.
The initialization treatment improper installation or system instability.
can’t be continued.
/MainProcess:Main()
20110 FFCustomMsgBox [20110] Failed in The subprocess was not successfully Check for error messages and warnings Main
initialization processing. started. It is conceivable that the system issued before the occurrence of this error
was unstable. This error may also occur as and then take a remedial action.
a consequence of the occurrence of Or, restart. If the same symptom recurs,
another error. perform a reinstallation.
20110 FFIIPMAIN It failed in the start of the The subprocess was not successfully Restart. If the same symptom recurs,
sub-process. started. This error occurred due to an perform a reinstallation.
The initialization treatment improper installation or system instability.
can’t be continued.
/MainProcess:
m_ProcessControl_
OnRunProcComplete()
20118 FFCustomMsgBox [20118] Failed in process An error occurred for some reason during No particular remedy needs be taken. Main
termination. subprocess end processing.
20118 FFIIPMAIN It failed in the completion of The subprocess was not successfully Restart.
the sub-process. terminated. The system was unstable.
/MainProcess:
ProcessEndRequirement()
20124 FFCustomMsgBox [20124] Failed in process A subprocess was not successfully View the event logs around to examine the Main
switchover. controlled for some reason. cause.
20124 FFIIPMAIN It failed in switching of the A subprocess changeover was not Restart.
process. / successfully effected. The target
MainProcess:xxxxx subprocess was not active or the system
was unstable.
20147 FFIIPMAIN Study Automatic Deletion The log was deleted successfully by study This is not an error.
successfully processed. automatic deletion processing.
(Loop Count is [ %1 ]. )/
MainProcess:m_Timer_
Timer()
20148 FFIIPMAIN Study Automatic Deletion Deleting the log using study automatic This is not an error.
warning. deletion processing resulted in a failure. If a message box opens, follow what is
(Loop Count is [ %1 ]. )/ directed in the box.
MainProcess:m_Timer_
Timer()
21001 FFCustomMsgBox [21001] Failed in The problem is attributable, for instance, to Check whether the SQL service manager List
connecting to the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-40
Event Event Source Description Causes Remedies Remarks
21002 FFCustomMsgBox [21002] Failed in accessing The problem is attributable, for instance, to Check whether the SQL service manager List
the image database. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
21003 FFCustomMsgBox [21003] Failed in obtaining The problem is attributable, for instance, to Check whether the SQL service manager List
information from the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
21004 FFCustomMsgBox [21004] Failed in accessing The problem is attributable, for instance, to Check whether the SQL service manager List
the image database. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
21005 FFCustomMsgBox [21005] Failed in setting The problem is attributable, for instance, to Check whether the SQL service manager List
information to the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
21006 FFCustomMsgBox [21006] Failed in deleting The problem is attributable, for instance, to Check whether the SQL service manager List
the study information. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
21007 FFCustomMsgBox [21007] Failed in It is conceivable that the database file was Restart. If the same symptom recurs, List
connecting to the system damaged or deleted. perform a reinstallation.
setting database.
21008 FFCustomMsgBox [21008] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs, List
the system setting damaged or deleted. perform a reinstallation.
database.
21009 FFCustomMsgBox [21009] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs, List
information from the system damaged or deleted. perform a reinstallation.
setting database.
21010 FFCustomMsgBox [21010] Failed in It is conceivable that the database file was Restart. If the same symptom recurs, List
connecting to the character damaged or deleted. perform a reinstallation.
database.
21011 FFCustomMsgBox [21011] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs, List
the character database. damaged or deleted. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-41
Event Event Source Description Causes Remedies Remarks
21012 FFCustomMsgBox [21012] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs, List
information from the damaged or deleted. perform a reinstallation.
character database.
21013 FFCustomMsgBox [21013] Failed in accessing It is conceivable that the file was damaged Restart. If the same symptom recurs, List
the definition file. or deleted. perform a reinstallation.
21014 FFCustomMsgBox [21014] Failed in opening It is conceivable that the registry Restart. If the same symptom recurs, List
the registry information. information was damaged or deleted. perform a reinstallation.
21015 FFCustomMsgBox [21015] Failed in retrieving It is conceivable that the registry Restart. If the same symptom recurs, List
the registry information. information was damaged or deleted. perform a reinstallation.
21016 FFCustomMsgBox [21016] Failed in writing the It is conceivable that the registry Restart. If the same symptom recurs, List
registry information. information was damaged or deleted. perform a reinstallation.
21017 FFCustomMsgBox [21017] Failed in closing It is conceivable that the registry Restart. If the same symptom recurs, List
the registry information. information was damaged or deleted. perform a reinstallation.
21018 FFCustomMsgBox [21018] Failed in receiving The system was unstable. Restart. If the same symptom recurs, List
message. perform a reinstallation.
21019 FFCustomMsgBox [21019] Failed in canceling The selected image was being output or No particular response is required. List
the output processing. already deleted.
21020 FFCustomMsgBox [21020] Failed in output. The selected image was being output or No particular response is required. List
already deleted.
21021 FFCustomMsgBox [21021] Failed in receiving The system was unstable. No particular response is required. List
the message.
21022 FFCustomMsgBox [21022] Failed in initializing The system was unstable. Restart. If the same symptom recurs, List
process for message perform a reinstallation.
communication.
21023 FFCustomMsgBox [21023] Failed in The system was unstable. Restart. If the same symptom recurs, List
terminating process for perform a reinstallation.
message communication.
21024 FFCustomMsgBox [21024] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs, List
information from the host damaged or deleted. perform a reinstallation.
setting database.
21025 FFCustomMsgBox [21025] Failed in editing the The system was unstable. Restart. If the same symptom recurs, List
host setting database. perform a reinstallation.
21030 FFCustomMsgBox [21030] Failed in study Connection equipment was not operative Check the connected equipment or List
start. or the network was unstable. network.
21031 FFCustomMsgBox [21031] Failed in QA start. Connection equipment was not operative Check the connected equipment or List
or the network was unstable. network.
21032 FFCustomMsgBox [21032] Failed in study Connection equipment was not operative Check the connected equipment or List
re-start. or the network was unstable. network.
21033 FFCustomMsgBox [21033] Failed in obtaining Connection equipment was not operative Check the connected equipment or List
study information from host or the network was unstable. network.
%s.
21034 FFCustomMsgBox [21034] Failed in locking Connection equipment was not operative Check the connected equipment or List
the study. or the network was unstable. network.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-42
Event Event Source Description Causes Remedies Remarks
21035 FFCustomMsgBox [21035] Failed in study Connection equipment was not operative Check the connected equipment or List
completion processing. or the network was unstable. network.
21036 FFCustomMsgBox [21036] The target study Connection equipment was not operative Check the connected equipment or List
exists but is not displayed or the network was unstable. network.
on the list.
21037 FFCustomMsgBox [21037] The target study The specified study may have been No particular measures need to be taken. List
could not be found. deleted.
21038 FFCustomMsgBox [21038] The input format The input format was incorrect. No particular measures need to be taken. List
for date is wrong.
21039 FFCustomMsgBox [21039] The input format The input format was incorrect. No particular measures need to be taken. List
for time is wrong.
21040 FFCustomMsgBox [21040] No patient ID or The input word was incorrect. No particular measures need to be taken. List
Accession number exists.
21041 FFCustomMsgBox [21041] Multiple target Two or more relevant studies existed. No particular measures need to be taken. List
studies exist. The study
cannot be started.
21051 FFCustomMsgBox [21051] An unknown study Illegal data was contained in the study Check the RIS side setup information and
was received from the RIS. information received from the RIS. the like.
21052 FFMWMHelper [MWMHelper]%1 This is an information log. There is no particular problem with
operations.
21053 FFMWMHelper [MWMHelper]%1 A warning condition is encountered. There is no particular problem with
For details, refer to the contents of the operations.
event log.
21054 FFMWMHelper [MWMHelper]%1 An error occurred. Restart.
A study received by means of MWM didn’t
include StudyInstanceUID (0020,000D).
Request the MWM Server to send
StudyInstanceUID.
21129 FFCustomMsgBox [21129] The hard disk was • The free hard disk space was insufficient. Delete unnecessary images. If any images Main
full. • A forced termination procedure was are left without being output, output them.
performed during a startup or freeze.
21201 FFCustomMsgBox [21201] An error occurred. The system was unstable. Restart. If the same symptom recurs, List
perform a reinstallation.
21202 FFCustomMsgBox [21202] Initialization has The system was unstable. Restart. If the same symptom recurs, List
not been completed. perform a reinstallation.
21203 FFCustomMsgBox [21203] Radiographer It is conceivable that the database file was Restart. If the same symptom recurs, List
information database is damaged or deleted. perform a reinstallation.
abnormal.
21204 FFCustomMsgBox [21204] Failed in obtaining The system was unstable. Restart. If the same symptom recurs, List
the registry information. perform a reinstallation.
21205 FFCustomMsgBox [21205] Failed in setting The system was unstable. Restart. If the same symptom recurs, List
registry values. perform a reinstallation.
21206 FFCustomMsgBox [21206] Failed in clearing a The system was unstable. Restart. If the same symptom recurs, List
displayed image. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-43
Event Event Source Description Causes Remedies Remarks
21207 FFCustomMsgBox [21207] Failed in Connection equipment was not operative Check other connected equipment or List
connecting to the database. or the network was unstable. network.
21208 FFCustomMsgBox [21208] Failed in Connection equipment was not operative Check other connected equipment or List
connecting to the database. or the network was unstable. network.
21209 FFCustomMsgBox [21209] Failed in setting the The system was unstable. Restart. If the same symptom recurs, List
study information sharing perform a reinstallation.
host.
21210 FFCustomMsgBox [21210] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs, List
the study information from damaged or deleted. perform a reinstallation.
the database.
21211 FFCustomMsgBox [21211] An error occurred It is conceivable that the database file was Restart. If the same symptom recurs, List
in obtaining radiographer damaged or deleted. perform a reinstallation.
information.
21212 FFCustomMsgBox [21212] An error occurred. The system was unstable. Restart. If the same symptom recurs, List
perform a reinstallation.
21213 FFCustomMsgBox [21213] An error occurred. The system was unstable. Restart. If the same symptom recurs, List
Study information cannot perform a reinstallation.
be obtained.
21214 FFCustomMsgBox [21214] Failed in starting The system was unstable. Restart. If the same symptom recurs, List
remote search/acquisition perform a reinstallation.
processing.
21215 FFCustomMsgBox [21215] An error occurred It is possible that the definition file and/or Restart. If the same symptom recurs,
in checking the definition database file has been damaged or perform a reinstallation.
file for patient ID auto- deleted.
convert.
21216 FFCustomMsgBox [21216] An error occurred. The system was unstable. Restart. If the same symptom recurs,
perform a reinstallation.
21217 FFCustomMsgBox [21217] Failed in auto- It is possible that the definition file and/or Restart. If the same symptom recurs,
conversion of patient ID. database file has been damaged or perform a reinstallation.
deleted.
21218 FFCustomMsgBox [21218] The upper-limit of The limit number of the items allowed on Reduce the number of selected studies.
the number of studies the list is exceeded.
selectable by patient ID
auto-convert is exceeded.
21219 FFCustomMsgBox [21219] No setting file for The patient ID auto-convert setting file Allocate the patient ID auto-convert setting
patient ID auto-convert does not exist at the specified location. file at the specified location, and then
exists. restart.
21220 FFCustomMsgBox [21220] A format error Format of the patient ID auto-convert Correct the error so that the format of the
occurred in the setting file setting file is invalid. patient ID auto-convert setting file is
for patient ID auto-convert. appropriate, and then restart.
21221 FFCustomMsgBox [21221] No definition file of The definition file of examinee-list for Allocate the definition file of examinee-list
examinee-list for patient ID patient ID auto-convert does not exist at for patient ID auto-convert at the specified
auto-convert exists. the specified location. location, and then restart.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-44
Event Event Source Description Causes Remedies Remarks
21222 FFCustomMsgBox [21222] A format error Format of the definition file of examinee-list Correct the error so that the format of the
occurred in the definition is invalid. definition file of examinee-list is
file of examinee list for appropriate, and then restart.
patient ID auto-convert.
21223 FFCustomMsgBox [21223] No definition DB of The definition DB file of examinee-list for Allocate the definition DB file of examinee-
examinee-list for patient ID patient ID auto-convert does not exist at list for patient ID auto-convert at the
auto-convert exists. the specified location. specified location, and then restart.
22100 FFIIPListMoveImage Warning. A warning was issued. For details, see the There is no particular problem with
%1 contents of the event log. operations.
22103 FFIIPMAIN It failed in the acquisition of The version information was not Reinstall.
the configuration information. successfully acquired. It is conceivable that
The initialization treatment the Version.env file did not exist in the
can’t be continued. specified directory or that the contents of
/MainProcess:Main() the file were damaged.
22128 FFIIPMAIN An error occurred in logfile A log file backup was not successfully Reinstall. (Although there are no
backup treatment. made. It is conceivable that the specified operational problems, various
/MainProcess:xxxxx pathname was incorrect. communication logs cannot be saved.)
22200 FFIIPListMoveImage Error. An error occurred. For details, see the Since the process was aborted due to an
%1 contents of the event log. error occurrence, it is conceivable that a
data inconsistency may exist. Verify the
data consistency.
22201 FFIIPListMoveImage Initialize Error. The initialization process failed. Restart.
%1
22202 FFIIPListMoveImage FFGetString Error. The FFGetString function returned an Restart.
%1 error.
22203 FFIIPListMoveImage DB Connect Error. The database connection was not Restart.
%1 successfully established.
22204 FFIIPListMoveImage DB Get Error. Data could not be acquired from the Restart.
% database.
22205 FFIIPListMoveImage SetDB Error. Data could not be set up for the database. Restart.
%1
22206 FFIIPListMoveImage Dicom Data Object Get Data could not be acquired from the Dicom Restart.
Error. Data Object.
%1
22207 FFIIPListMoveImage Dicom Data Object Set Data could not be set up for the Dicom Restart.
Error. Data Object.
%1
22208 FFIIPListMoveImage An unknown error An exception error occurred. Restart.
occurred.%n[%1]
23000 FFMWMStudyList RegCreatekey() returned The RegCreateKey function returned an Restart.
with Error at %1. [%2] error.
23000 FFStudyList, RegCreatekey() returned The user authority may not have been Use the normal startup procedure or
FFOutputList with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-45
Event Event Source Description Causes Remedies Remarks
23000 FFIIPListCustomize RegCreatekey() returned The user authority may not have been Use the normal startup procedure or
with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
23000 FFIIPListSearch RegCreatekey() returned The user authority may not have been Use the normal startup procedure or
with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
23000 FFIIPListMoveImage Information. Information log. There are no operational problems.
%1
23000 FFMWMStudyList Study couldn’t be started Failed in starting a study automatically for There are no operational problems.
automatically. %1. [%2] any reason.
23001 FFMWMStudyList RegOpenkey() returned The RegOpenKey function returned an Restart.
with Error at %1. [%2] error.
23001 FFStudyList, RegOpenkey() returned The user authority may not have been Use the normal startup procedure or
FFOutputList with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
23001 FFIIPListCustomize RegOpenkey() returned The user authority may not have been Use the normal startup procedure or
with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
23001 FFIIPListSearch RegOpenkey() returned The user authority may not have been Use the normal startup procedure or
with Error at %1.%n[%2] granted or installation may have been perform reinstallation.
ended abnormally.
23002 FFMWMStudyList RegClosekey() returned The RegOpenKey function returned an Restart.
with Error at %1. [%2] error.
23002 FFStudyList, RegClosekey() returned The user authority may not have been Use the normal startup procedure.
FFOutputList with Error at %1.%n[%2] granted.
23002 FFIIPListCustomize RegClosekey() returned The user authority may not have been Use the normal startup procedure.
with Error at %1.%n[%2] granted.
23002 FFIIPListSearch RegClosekey() returned The user authority may not have been Use the normal startup procedure.
with Error at %1.%n[%2] granted.
23003 FFMWMStudyList RegQueryNumericValue() The RegOpenKey function returned an Restart.
returned with Error at %1. error.
[%2]
23003 FFStudyList, RegQueryNumericValue() The user authority may not have been Use the normal startup procedure.
FFOutputList returned with Error at granted.
%1.%n[%2]
23003 FFIIPListCustomize RegQueryNumericValue() The user authority may not have been Use the normal startup procedure.
returned with Error at granted.
%1.%n[%2]
23003 FFIIPListSearch RegQueryNumericValue() The user authority may not have been Use the normal startup procedure.
returned with Error at granted.
%1.%n[%2]
23004 FFMWMStudyList RegSetNumericValue() The RegOpenKey function returned an Restart.
returned with Error at %1. error.
[%2]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-46
Event Event Source Description Causes Remedies Remarks
23004 FFStudyList, RegSetNumericValue() The user authority may not have been Use the normal startup procedure.
FFOutputList returned with Error at granted.
%1.%n[%2]
23004 FFIIPListCustomize RegSetNumericValue() The user authority may not have been Use the normal startup procedure.
returned with Error at granted.
%1.%n[%2]
23004 FFIIPListSearch RegSetNumericValue() The user authority may not have been Use the normal startup procedure.
returned with Error at granted.
%1.%n[%2]
23005 FFMWMStudyList FFGetString() returned with The FFGetString function returned an Restart.
Error at %1. [%2] error.
23005 FFStudyList, FFGetString() returned with The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
FFOutputList Error at %1.%n[%2] Directory.ini file. and then restart.
23005 FFIIPListCustomize FFGetString() returned with The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
Error at %1.%n[%2] Directory.ini file. and then restart.
23005 FFIIPListSearch FFGetString() returned with The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
Error at %1.%n[%2] Directory.ini file. and then restart.
23006 FFIIPListCustomize FFStrDBConnect() returned The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
with Error at %1.%n[%2] Directory.ini file or missing LangStr.mdb and then restart. Or, check the LangStr.
file. mdb file.
23006 FFIIPListSearch FFStrDBConnect() returned The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
with Error at %1.%n[%2] Directory.ini file or missing LangStr.mdb and then restart. Or, check the LangStr.
file. mdb file.
23006 FFMWMStudyList FFStrDBConnect() returned The FFStrDBConnect function returned an Restart.
with Error at %1. [%2] error.
23006 FFStudyList, FFStrDBConnect() returned The problem is attributable to a damaged Ensure that the Directory.ini file is normal,
FFOutputList with Error at %1.%n[%2] Directory.ini file or missing LangStr.mdb and then restart. Or, check the LangStr.
file. mdb file.
23007 FFIIPListCustomize FFStrDBClose() returned The LangStr.mdb file was damaged. Check the LangStr.mdb file.
with Error at %1.%n[%2]
23007 FFIIPListSearch FFStrDBClose() returned The LangStr.mdb file was damaged. Check the LangStr.mdb file.
with Error at %1.%n[%2]
23007 FFMWMStudyList FFStrDBClose() returned The FFStrDBClose function failed. Restart.
with Error at %1. [%2]
23007 FFStudyList, FFStrDBClose() returned The LangStr.mdb file was damaged. Check the LangStr.mdb file.
FFOutputList with Error at %1.%n[%2]
23008 FFIIPListCustomize FFStrDBReadString() The LangStr.mdb file was damaged. Check the LangStr.mdb file.
returned with Error at
%1.%n[%2]
23008 FFIIPListSearch FFStrDBReadString() The LangStr.mdb file was damaged. Check the LangStr.mdb file.
returned with Error at
%1.%n[%2]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-47
Event Event Source Description Causes Remedies Remarks
23008 FFMWMStudyList FFStrDBReadString() The FFStrDBReadString function returned Restart.
returned with Error at %1. an error.
[%2]
23008 FFStudyList, FFStrDBReadString() The LangStr.mdb file was damaged. Check the LangStr.mdb file.
FFOutputList returned with Error at
%1.%n[%2]
23009 FFIIPListCustomize FFRegGetKeySZ() returned The problem is attributable to an Use the normal startup procedure.
with Error at %1.%n[%2] unprivileged user.
23009 FFIIPListSearch FFRegGetKeySZ() returned The problem is attributable to an Use the normal startup procedure.
with Error at %1.%n[%2] unprivileged user.
23009 FFMWMStudyList FFRegGetKeySZ() returned The FFRegGetKeySZ function returned an Restart.
with Error at %1. [%2] error.
23009 FFStudyList, FFRegGetKeySZ() returned The problem is attributable to an Use the normal startup procedure.
FFOutputList with Error at %1.%n[%2] unprivileged user.
23010 FFIIPListCustomize FFRegGetKeyDWORD() The problem is attributable to an Use the normal startup procedure.
returned with Error at unprivileged user.
%1.%n[%2]
23010 FFIIPListSearch FFRegGetKeyDWORD() The problem is attributable to an Use the normal startup procedure.
returned with Error at unprivileged user.
%1.%n[%2]
23010 FFMWMStudyList FFRegGetKeyDWORD() The FFRegGetKeyDWORD function Restart.
returned with Error at %1. returned an error.
[%2]
23010 FFStudyList, FFRegGetKeyDWORD() The problem is attributable to an Use the normal startup procedure.
FFOutputList returned with Error at unprivileged user.
%1.%n[%2]
23011 FFStudyList, FFSysDBConnect() It is conceivable that the Directory.ini file Ensure that the Directory.ini file is normal,
FFOutputList returned with Error at was damaged or the Hosts.mdb file did not and then restart. Or, check the SysConfig.
%1.%n[%2] exist. mdb file.
23011 FFMWMStudyList FFHostDBInitialize() FFHostDBInitialize() sent back an error. Restart the system.
returned with Error at %1.
[%2]
23012 FFStudyList, FFSysDBClose() returned This error is attributable to a damaged Check the SysConfig.mdb file.
FFOutputList with Error at %1.%n[%2] Directory.ini file.
23012 FFIIPListCustomize FFHostDBGetQueryHost() It is conceivable that the Hosts.mdb file Check the Hosts.mdb file.
returned with Error at was damaged.
%1.%n[%2]
23012 FFMWMStudyList FFHostDBGetQueryHost() FFHostDBGetQueryHost() sent back an Restart the system.
returned with Error at %1. error.
[%2]
23013 FFStudyList, FFSysDBReadszValue() This error is attributable to a damaged Check the SysConfig.mdb file.
FFOutputList returned with Error at Directory.ini file.
%1.%n[%2]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-48
Event Event Source Description Causes Remedies Remarks
23013 FFIIPListCustomize FFHostDBSetQueryHost() It is conceivable that the Hosts.mdb file Check the Hosts.mdb file.
returned with Error at was damaged.
%1.%n[%2]
23013 FFMWMStudyList FFHostDBSetQueryHost() FFHostDBSetQueryHost() sent back an Restart the system.
returned with Error at %1. error.
[%2]
23014 FFStudyList, FFSysDBReadnValue() This error is attributable to a damaged Check the SysConfig.mdb file.
FFOutputList returned with Error at Directory.ini file.
%1.%n[%2]
23015 FFStudyList, Failed ImageDB The SQL Server Service Manager starts Ensure that the SQL Server is running.
FFOutputList Connection at %1.%n[%2] running or necessary files are not installed.
23016 FFStudyList, Failed ImageDB The SQL Server Service Manager starts Ensure that the SQL Server is running.
FFOutputList Disconnection at running or necessary files are not installed.
%1.%n[%2]
23017 FFStudyList, Failed ImageDB Open The SQL Server Service Manager starts Ensure that the SQL Server is running.
FFOutputList Recordset at %1.%n[%2] running or necessary files are not installed.
23018 FFStudyList, Failed ImageDB Update The SQL Server Service Manager starts Ensure that the SQL Server is running.
FFOutputList Recordset at %1.%n[%2] running or necessary files are not installed.
23019 FFStudyList, Failed ImageDB Close The SQL Server Service Manager starts Ensure that the SQL Server is running.
FFOutputList Recordset at %1.%n[%2] running or necessary files are not installed.
23020 FFStudyList Failed Get HostName as Failed in acquiring a host name. Check the network/network settings.
%1.%n[%2] The network may have been set up
incorrectly.
23021 FFStudyList Failed HostDB Data at Failed in acquiring data from the Hosts. Check the Hosts.mdb file.
%1.%n[%2] mdb file. The database may have been
damaged.
23022 FFStudyList Failed GetPossession at Failed in acquiring property. Check the network/equipment.
%1.%n[%2] It is conceivable that any other equipment
has already acquired property, network
disconnected or power not turned ON.
23023 FFStudyList Failed A failure occurred in file change or delivery Restart the system.
DDOChangeOrDelivery at processing at film division/extension
%1.%n[%2] (12on1) output.
23024 FFStudyList Failed DicomQ/R at A failure occurred when performing DICOM Check the network/network settings, and
%1.%n[%2] Q/R processing. then restart the system as necessary.
23025 FFStudyList FFNetDBConnect() An attempt to establish connection to the Check NetConfig.mdb and restart the
returned with Error at network setup database resulted in a system.
%1.%n [%2] failure.
23026 FFStudyList Failed DicomQ/R Initialize A failure occurred in initialization of DICOM Restart the system.
at %1.%n[%2] Q/R processing.
23027 FFStudyList Failed A failure occurred in file change processing Restart the system.
MultiPrintDDOUpdate at at film division/extension (12on1) output.
%1.%n[%2]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-49
Event Event Source Description Causes Remedies Remarks
23028 FFStudyList Failed MultiPrintDelivery A failure occurred in delivery processing at Restart the system.
%1.%n[%2] film division/extension (12on1) output.
23029 FFStudyList Failed FFNetDBGet An attempt to acquire data from the Check NetConfig.mdb and restart the
%1.%n[%2] network setup database resulted in a system.
failure.
23031 FFStudyList CSIdmCheckResource() CSIdmCheckResource() returned an error Restart. If the same symptom recurs,
returned with Error at at %1. perform a reinstallation.
%1.%n[%2]
23032 FFStudyList Failed IDAutoConvertDLG Initialization of the auto-convert screen Restart. If the same symptom recurs,
Initialize at %1.%n[%2] returned an error at %1. perform a reinstallation.
23033 FFStudyList frm_AutoConvertProgDlg. frm_AutoConvertProgDIg.SetExecData() Restart. If the same symptom recurs,
SetExecData() returned returned an error at %1. perform a reinstallation.
with Error at %1.%n[%2]
23034 FFStudyList LimitOver SelectedRows The number of selected studies is Reduce the number of selected studies.
with Error at %1.%n[%2] exceeded the upper limit at %1.
23035 FFStudyList No Config File On A configuration file browse error occurred Allocate the patient ID auto-convert
AutoConvert Proc.%n[%1] during auto-convert processing. configuration file at the specified location,
and then restart.
23036 FFStudyList Config File Format Error A configuration file format error occurred Correct the error so that the format of the
On AutoConvert during auto-convert processing. patient ID auto-convert configuration file is
Proc.%n[%1] appropriate, and then restart.
23037 FFStudyList No Schema File On A definition file browse error occurred Allocate the definition file of examinee-list
AutoConvert Proc.%n[%1] during auto-convert processing. for patient ID auto-convert at the specified
location, and then restart.
23038 FFStudyList Scehma File Format Error A definition file format error occurred during Correct the error so that the format of the
On AutoConvert auto-convert processing. definition file of examinee-list is
Proc.%n[%1] appropriate, and then restart.
23039 FFStudyList No MatchingDB Error On A definition DB browse error occurred Allocate the definition DB file of examinee-
AutoConvert Proc.%n[%1] during auto-convert processing. list for patient ID auto-convert at the
specified location, and then restart.
23099 FFIIPListCustomize An unknown error An unknown error occurred. Restart.
occurred.%n[%1]
23099 FFIIPListSearch An unknown error An unknown error occurred. Restart.
occurred.%n[%1]
23099 FFMWMStudyList An unknown error occurred. An exception error occurred. Restart.
[%1]
23099 FFStudyList, An unknown error An unknown error occurred. Restart.
FFOutputList occurred.%n[%1]
24244 FFCustomMsgBox [24244] Patient ID auto- The patient ID auto-convert button was No problems will arise.
convert processing is pressed.
executed.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-50
Event Event Source Description Causes Remedies Remarks
24245 FFCustomMsgBox [24245] No examinee-list Target processing was performed Select the examinee-list file and re-perform
file has been selected. preliminarily or definitively with no the target processing preliminarily or
examinee-list file for patient ID auto- definitively.
convert selected appropriately.
24246 FFCustomMsgBox [24246] Change the study Processing to output a specified study was No problems will arise.
status from QA List to performed.
Study Completed and
output to %s.
24247 FFCustomMsgBox [24247] No output device No devices have been registered to output Register a device for output of the specified
has been registered. a specified study. study, and then redistribute the target
study.
24248 FFCustomMsgBox [24248] Failed in output a It is possible that the database file has Restart. If the same symptom recurs,
part of the study been damaged or deleted. perform a reinstallation.
24249 FFCustomMsgBox [24249] No destination has No destination has been specified for the Register a destination for distribution of the
been set. specified study. specified study, and then redistribute it.
24250 FFCustomMsgBox [24250] There is a study No patient ID or name has been input for Input the patient ID and name for the
into which a patient ID or the specified study. specified study, and then redistribute it.
patient name has not been
entered.
25029 FFCustomMsgBox [25029] Failed in obtaining Access to the definition file failed while the Restart.
the setting info of the target media for connection was being
connected media. changed.
25030 FFCustomMsgBox [25030] Failed in changing In an attempt to change the target media Restart.
the connecting destination. for connection, access to the database
failed or an unexpected error occurred.
25031 FFCustomMsgBox [25031] Failed in accessing A drive where no media has been inserted Ensure that a disk has been inserted into
the selected drive. was selected. the drive correctly.
29010 FFIIPEvtMonitoringCtrl [%1]%nThe error occurred The Win32 API function returned an error. Restart. Worklist
in Win32 API.%n%2
29011 FFIIPEvtMonitoringCtrl [%1]%nThe exception error An exception error occurred. Restart.
occurred.%n%2
29107 FFIIPMAIN It failed in the preparation The pipe for inter-process communication Restart.
of the pipe. The treatment was not successfully created.
can’t be continued. It is conceivable that the system was
/MainProcess:Main() unstable.
29112 FFIIPMAIN It failed in the indication of The routine transition instruction message Restart. If the same symptom occurs after
the routine mode transition. was not successfully issued. a restart, reinstall.
/MainProcess: It is conceivable that the subprocess was
m_ProcessControl not running or that the pipe was not
_OnRunProcComplete() created.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-51
Event Event Source Description Causes Remedies Remarks
29114 FFIIPMAIN An unusual message was The received message was unusual. Restart. If the same symptom occurs after
received. This message is It is conceivable that the message was not a restart, reinstall.
ignored. registered in the message information
/MainProcess: database or that the system was unstable.
m_ProcessComControl
_OnRecvNotification()
29115 FFIIPMAIN An unusual send cause A message was received from an unknown Restart. If the same symptom occurs after
was received. This send source. It is conceivable that the a restart, reinstall.
cause is ignored. transmission source was not registered in
/MainProcess: the message information database or that
m_ProcessComControl a wrong transmission source was selected.
_OnRecvNotification()
29116 FFIIPMAIN The form of the place of the An unknown transmission destination was Restart. If the same symptom occurs after
transmission isn’t right. specified for the received message. a restart, reinstall.
/MainProcess: It is conceivable that the transmission
SendMessage() destination was not registered in the
message information database or that a
wrong transmission destination was
selected.
29117 FFIIPMAIN It failed in the message A message delivery was not successful. Restart. If the same symptom occurs after
delivery. It is conceivable that the subprocess at the a restart, reinstall.
/MainProcess: transmission destination was not running.
SendMessage()
29119 FFIIPMAIN It failed in the process An end notice was not successfully sent to Restart.
completion notice. the subprocess. It is conceivable that the
/MainProcess:ProcessEnd- subprocess was not running or that the
AnswerMessage() system was unstable.
29122 FFIIPMAIN It failed in the unusual An error notice was not successfully sent Restart.
occurrence notice. to the subprocess. It is conceivable that
/MainProcess: the subprocess was not running or that the
ErrorNoticeMessage() system was unstable.
29123 FFIIPMAIN It failed in the pipe close. An inter-process communication pipe was Restart.
/MainProcess: not successfully closed. It is conceivable
EndMainProcess() that the system was unstable.
29125 FFIIPMAIN It failed in close of the The message database was not Restart.
message data base. successfully closed. It is conceivable that
/MainProcess:xxxxx the system was unstable.
29126 FFIIPMAIN It failed in reading of the Data was not successfully acquired from Reinstall.
registry information. the registry. It is conceivable that
/MainProcess:xxxxx installation was not properly completed.
29127 FFIIPMAIN It failed in writing of the Data was not successfully written into the Restart.
registry information. registry. It is conceivable that the system
/MainProcess:xxxxx was unstable.
30001 FFCustomMsgBox [30001] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the character database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-52
Event Event Source Description Causes Remedies Remarks
30002 FFCustomMsgBox [30002] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting to the character damaged or deleted. perform a reinstallation.
database.
30003 FFCustomMsgBox [30003] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the system setting
database.
30004 FFCustomMsgBox [30004] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting to the system damaged or deleted. perform a reinstallation.
setting database.
30005 FFCustomMsgBox [30005] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the image database.
30013 FFCustomMsgBox [30013] Failed in saving It is conceivable that the file was damaged Restart. If the same symptom recurs,
listed (worklist) display or deleted. perform a reinstallation.
settings.
30019 FFCustomMsgBox [30019] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the connection characters damaged or deleted. perform a reinstallation.
to the network setting
database.
30027 FFCustomMsgBox [30027] Failed in initializing It is conceivable that the registry Restart. If the same symptom recurs,
the list (worklist). information was damaged or deleted. perform a reinstallation.
obtained.
30045 FFCustomMsgBox [30045] IP address on the It is conceivable that the database file was Restart. If the same symptom recurs,
Image Reader not damaged or improperly set. perform a reinstallation.
obtained.
30057 FFCustomMsgBox [30057] Failed in receiving The system was unstable. Restart. If the same symptom recurs,
the message. perform a reinstallation.
30085 FFCustomMsgBox [30085] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the Image Reader status
information database.
30105 FFCustomMsgBox [30105] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (waiting for perform a reinstallation.
study).
30106 FFCustomMsgBox [30106] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (waiting for perform a reinstallation.
QA).
30107 FFCustomMsgBox [30107] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (waiting for perform a reinstallation.
output).
30108 FFCustomMsgBox [30108] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (already perform a reinstallation.
output).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-53
Event Event Source Description Causes Remedies Remarks
30109 FFCustomMsgBox [30109] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (all). perform a reinstallation.
30110 FFCustomMsgBox [30110] Failed in list The system was unstable. Restart. If the same symptom recurs,
initialization (today). perform a reinstallation.
30117 FFCustomMsgBox [30117] Failed in updating The system was unstable. Restart. If the same symptom recurs,
list (worklist) display. perform a reinstallation.
30121 FFCustomMsgBox [30121] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the Image Reader
information database.
30122 FFCustomMsgBox [30122] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the Image Reader damaged or deleted. perform a reinstallation.
information database.
30147 FFCustomMsgBox [30147] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Magnetic card
type).
30148 FFCustomMsgBox [30148] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Magnetic card
type).
30149 FFCustomMsgBox [30149] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Hospital card
used).
30150 FFCustomMsgBox [30150] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
database.
30151 FFCustomMsgBox [30151] Failed in • The connection to the network was not • Connect to the network.
initialization processing. established. • Set a dummy node for “THIS HOST (IIP)”
• Node setup was not completed for “THIS (e.g., DICOM Print).
HOST (IIP)” under “Network Config”. • Restart. If the same symptom recurs,
• The system was unstable. perform a reinstallation.
30152 FFCustomMsgBox [30152] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the pass information (image or deleted. perform a reinstallation.
data folder).
30153 FFCustomMsgBox [30153] The shutdown The system was unstable. Restart. If the same symptom recurs,
process resulted in a perform a reinstallation.
failure.
30155 FFCustomMsgBox [30155] Failed in list It is conceivable that the file was damaged Restart. If the same symptom recurs,
initialization (waiting for or deleted. perform a reinstallation.
study).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-54
Event Event Source Description Causes Remedies Remarks
30157 FFCustomMsgBox [30157] Database access The problem is attributable, for instance, to Check whether the SQL service manager
was unsuccessful. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
30200 FFCustomMsgBox [30200] Failed in RIS There may be a problem with the RIS Verify the settings and then restart.
activation. module or RIS setup. If the same symptom recurs, perform a
reinstallation.
30201 FFCustomMsgBox [30201] A communication There may be a problem with the RIS Verify the settings and then restart.
error with RIS occurred. module or RIS setup. If the same symptom recurs, perform a
reinstallation.
30202 FFCustomMsgBox [30202] A communication There may be a problem with the RIS Verify the settings and then restart.
error with RIS occurred. module or RIS setup. If the same symptom recurs, perform a
reinstallation.
30204 FFCustomMsgBox [30204] Because of Failed in installation. Reinstall.
improperly performed
installation processing, the
system cannot be started.
30205 FFCustomMsgBox [30205] Failed in search The system was unstable. Restart. If the same symptom recurs,
processing. perform a reinstallation.
30206 FFCustomMsgBox [30206] Failed in initializing The system was unstable. Restart. If the same symptom recurs,
patient-incidental perform a reinstallation.
information.
30208 FFCustomMsgBox [30208] System error A return error from the Shimadzu FPD Restart the FPD system. If the same error
occurred. controller and a program error. recurs, reinstall the system.
30209 FFCustomMsgBox [30209] System error A return error from the Shimadzu FPD Restart the FPD system. If the same error
occurred. controller and a program error. recurs, reinstall the system.
30210 FFCustomMsgBox [30210] System error An error of level: 3,4 originated from the Restart the FPD system. If the same error
occurred. FPD system. recurs, reinstall the system.
30211 FFCustomMsgBox [30211] System error An error of level: 3,4 originated from the Restart the FPD system. If the same error
occurred. FPD system. recurs, reinstall the system.
30213 FFCustomMsgBox [30213] Failed in displaying A failure occurred in preview display of a Restart the FPD system. If the same Improvement of
the preview screen. CR long-view image before it is subjected symptom recurs, perform a reinstallation. operation performance
to image combining processing. of the automatic image
combining processing.
30215 FFCustomMsgBox [30215] An error occurred An error occurred in an attempt to change Confirm connection to the FPD device. Exposure technique
when changing the an exposure technique is notified change
exposure technique. accordingly. (Not only an error occurring
while the exposure controller controlling
method is being called up, but also an
error occurring in an attempt to access
DDO and image database are notified
accordingly.)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-55
Event Event Source Description Causes Remedies Remarks
30216 FFCustomMsgBox [30216] No configurable The database does not contain trimming Restart. If the same symptom recurs,
trimming size available in sizes configurable for the FPD selector perform a reinstallation.
%s. (upright or supine position).
30507 FFCustomMsgBox [30507] An error occurred. Communication with the linking application Restart the system.
failed. If the same error recurs after restarting the
system, reinstall the system.
31001 FFCustomMsgBox [31001] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information information was damaged or deleted. perform a reinstallation.
(font name).
31002 FFCustomMsgBox [31002] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information information was damaged or deleted. perform a reinstallation.
(font size).
31003 FFCustomMsgBox [31003] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information information was damaged or deleted. perform a reinstallation.
(multi-byte language
information).
31004 FFCustomMsgBox [31004] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the used language in damaged or deleted. perform a reinstallation.
system setting.
31005 FFCustomMsgBox [31005] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the connection characters damaged or deleted. perform a reinstallation.
to the character database.
31006 FFCustomMsgBox [31006] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting to the character damaged or deleted. perform a reinstallation.
database.
31007 FFCustomMsgBox [31007] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the connection characters damaged or deleted. perform a reinstallation.
to the system setting
database.
31008 FFCustomMsgBox [31008] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting to the system damaged or deleted. perform a reinstallation.
setting database.
31010 FFCustomMsgBox [31010] Failed in The problem is attributable, for instance, to Check whether the SQL service manager
connecting to the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31011 FFCustomMsgBox [31011] Failed in acquiring The problem is attributable, for instance, to Check whether the SQL service manager
information from the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Appx MT A-56
Event Event Source Description Causes Remedies Remarks
31012 FFCustomMsgBox [31012] Failed in accessing The problem is attributable, for instance, to Check whether the SQL service manager
the image database. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31013 FFCustomMsgBox [31013] Failed in obtaining The problem is attributable, for instance, to Check whether the SQL service manager
the information from the an inactive SQL service manager or is normally operating. If the symptom
image database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31014 FFCustomMsgBox [31014] Failed in setting The problem is attributable, for instance, to Check whether the SQL service manager
information to the image an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31015 FFCustomMsgBox [31015] Failed in deleting The problem is attributable, for instance, to Check whether the SQL service manager
the information from the an inactive SQL service manager or is normally operating. If the symptom
image database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31016 FFCustomMsgBox [31016] Failed in The problem is attributable, for instance, to Check whether the SQL service manager
connecting to the flag an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31017 FFCustomMsgBox [31017] Failed in accessing The problem is attributable, for instance, to Check whether the SQL service manager
the flag database. an inactive SQL service manager or is normally operating. If the symptom
uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31018 FFCustomMsgBox [31018] Failed in setting The problem is attributable, for instance, to Check whether the SQL service manager
information to the flag an inactive SQL service manager or is normally operating. If the symptom
database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31019 FFCustomMsgBox [31019] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
initialization process for damaged or deleted. perform a reinstallation.
menu setting database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-57
Event Event Source Description Causes Remedies Remarks
31020 FFCustomMsgBox [31020] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the menu information from damaged or deleted. perform a reinstallation.
the menu setting database.
31021 FFCustomMsgBox [31021] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the menu parameter damaged or deleted. perform a reinstallation.
information from the menu
setting database.
31022 FFCustomMsgBox [31022] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
terminating the menu damaged or deleted. perform a reinstallation.
setting database.
31023 FFCustomMsgBox [31023] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the connection characters damaged or deleted. perform a reinstallation.
to the network setting
database.
31024 FFCustomMsgBox [31024] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting to the network damaged or deleted. perform a reinstallation.
setting database.
31025 FFCustomMsgBox [31025] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
network setting database.
31026 FFCustomMsgBox [31026] Failed in issuing The system was unstable. Restart. If the same symptom recurs,
the image information. perform a reinstallation.
31027 FFCustomMsgBox [31027] Failed in obtaining The system was unstable. Restart. If the same symptom recurs,
the image information. perform a reinstallation.
Use the DICOM MWM C-FIND Tag Setting
of CL Service Utility to see if Distribution
Code (ID:141) is checked or not.
If checked, uncheck it.
31028 FFCustomMsgBox [31028] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
image information. perform a reinstallation.
31029 FFCustomMsgBox [31029] Failed in deleting The system was unstable. Restart. If the same symptom recurs,
the image information. perform a reinstallation.
31030 FFCustomMsgBox [31030] Failed in reading An image-related file was not successfully Restart. If the same symptom frequently
the image information file. accessed. The problem is attributable to a occurs, replace the disk.
disk.
31031 FFCustomMsgBox [31031] Failed in writing the An image-related file was not successfully Restart. If the same symptom frequently
image information file. accessed. The problem is attributable to a occurs, replace the disk.
disk.
31032 FFCustomMsgBox [31032] Failed in releasing The system was unstable. Restart. If the same symptom recurs,
the image information. perform a reinstallation.
31033 FFCustomMsgBox [31033] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(ID#).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-58
Event Event Source Description Causes Remedies Remarks
31034 FFCustomMsgBox [31034] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(study#).
31035 FFCustomMsgBox [31035] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(patient name).
31036 FFCustomMsgBox [31036] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Kanji-patient name).
31037 FFCustomMsgBox [31037] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(sex).
31038 FFCustomMsgBox [31038] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(sex: M).
31039 FFCustomMsgBox [31039] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(sex: F).
31040 FFCustomMsgBox [31040] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(sex: O).
31041 FFCustomMsgBox [31041] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(data of birth).
31042 FFCustomMsgBox [31042] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (ID# input
character limitation).
31043 FFCustomMsgBox [31043] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Study # input
character limitation).
31044 FFCustomMsgBox [31044] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (ID # padding
format).
31045 FFCustomMsgBox [31045] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Study #
padding format).
31047 FFCustomMsgBox [31047] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (date of birth
format).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-59
Event Event Source Description Causes Remedies Remarks
31051 FFCustomMsgBox [31051] Invalid character is A 2-byte character, @, or other Make a proper entry.
input in ID #. unacceptable character was contained.
31052 FFCustomMsgBox [31052] Invalid character is A 2-byte character, @, or other Make a proper entry.
input in Study #. unacceptable character was contained.
31053 FFCustomMsgBox [31053] Invalid character is A 2-byte character, @, or other Make a proper entry.
input in patient name. unacceptable character was contained.
31054 FFCustomMsgBox [31054] Invalid character is A 1-byte character or other unacceptable Make a proper entry.
input in kanji-patient name. character was contained.
31055 FFCustomMsgBox [31055] Invalid character is A character other than “M”, “F”, and “O” Make a proper entry.
input in sex. was contained.
31056 FFCustomMsgBox [31056] Invalid format for The entry did not agree with a preselected Make a proper entry.
date of birth is input. birth date format.
31057 FFCustomMsgBox [31057] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the requesting department damaged or deleted. perform a reinstallation.
database.
31058 FFCustomMsgBox [31058] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(requesting dep.).
31059 FFCustomMsgBox [31059] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the item information for damaged or deleted. perform a reinstallation.
displaying in the requesting
dep. Selection list.
31060 FFCustomMsgBox [31060] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the requesting department damaged or deleted. perform a reinstallation.
information.
31061 FFCustomMsgBox [31061] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the requesting
department database.
31062 FFCustomMsgBox [31062] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the technician information damaged or deleted. perform a reinstallation.
database.
31063 FFCustomMsgBox [31063] Failed in caption It is conceivable that the database file was Restart. If the same symptom recurs,
information (technician). damaged or deleted. perform a reinstallation.
31064 FFCustomMsgBox [31064] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the item information for damaged or deleted. perform a reinstallation.
displaying in the technician
selection list.
31065 FFCustomMsgBox [31065] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the technician information. damaged or deleted. perform a reinstallation.
31066 FFCustomMsgBox [31066] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the connection characters damaged or deleted. perform a reinstallation.
to the technician database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-60
Event Event Source Description Causes Remedies Remarks
31067 FFCustomMsgBox [31067] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information information was damaged or deleted. perform a reinstallation.
(window size).
31068 FFCustomMsgBox [31068] No patient Nothing was entered in the patient Enter at least the patient’s sex.
information is input. information input field.
31069 FFCustomMsgBox [31069] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (succeeded
from exposure menu
reservation list).
31070 FFCustomMsgBox [31070] Multiple studies An attempt was made to start an Start an examination process with one
cannot be executed. examination process with more than one examination item selected.
examination item selected from an
examination waiting list.
31071 FFCustomMsgBox [31071] No list is selected. An attempt was made to start execution Select at least one examination item.
with no examination item selected.
31072 FFCustomMsgBox [31072] Invalid character is An unacceptable character was contained. Make a proper entry.
input for film annotation.
31074 FFCustomMsgBox [31074] Failed in obtaining The problem is attributable, for instance, to Check whether the SQL service manager
the information from the an inactive SQL service manager or is normally operating. If the symptom
flag database. uninitialized image database. occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
31076 FFCustomMsgBox [31076] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Clear damaged or deleted. perform a reinstallation.
button).
31077 FFCustomMsgBox [31077] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (OK damaged or deleted. perform a reinstallation.
button).
31078 FFCustomMsgBox [31078] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Cancel damaged or deleted. perform a reinstallation.
button).
31079 FFCustomMsgBox [31079] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (increment damaged or deleted. perform a reinstallation.
button).
31080 FFCustomMsgBox [31080] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image damaged or deleted. perform a reinstallation.
(decrement button).
31081 FFCustomMsgBox [31081] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Next damaged or deleted. perform a reinstallation.
page button).
31082 FFCustomMsgBox [31082] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (previous damaged or deleted. perform a reinstallation.
page button).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-61
Event Event Source Description Causes Remedies Remarks
31097 FFCustomMsgBox [31097] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
character database.
31099 FFCustomMsgBox [31099] An error of It is conceivable that there is an Restart. If the same symptom recurs,
unknown cause occurred. inconsistency in the image information file. perform a reinstallation.
31100 FFCustomMsgBox [31100] The image display It is conceivable that the image data or Restart. If the same symptom recurs,
process resulted in a failure image information file was damaged. perform a reinstallation.
while operating in the
image input subsystem.
31101 FFCustomMsgBox [31101] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the pass information (log or deleted. perform a reinstallation.
saving folder).
31103 FFCustomMsgBox [31103] Failed in calling up The system was unstable. Restart. If the same symptom recurs,
the patient information. perform a reinstallation.
31105 FFCustomMsgBox [31105] Failed in The system was unstable. Restart. If the same symptom recurs,
terminating the image input perform a reinstallation.
subsystem.
31106 FFCustomMsgBox [31106] Failed in The system was unstable. Restart. If the same symptom recurs,
terminating the image input perform a reinstallation.
subsystem.
31107 FFCustomMsgBox [31107] Failed in image The system was unstable. Restart. If the same symptom recurs,
inputting process in the perform a reinstallation.
image input sub-system.
31108 FFCustomMsgBox [31108] Failed in image The system was unstable. Restart. If the same symptom recurs,
inputting process in the perform a reinstallation.
image input sub-system.
31109 FFCustomMsgBox [31109] Failed in image The system was unstable. Restart. If the same symptom recurs,
inputting process in the perform a reinstallation.
image input sub-system.
31110 FFCustomMsgBox [31110] Failed in calling up The system was unstable. Restart. If the same symptom recurs,
the exposure menu. perform a reinstallation.
31111 FFCustomMsgBox [31111] Failed in study The system was unstable. Restart. If the same symptom recurs,
terminating process. perform a reinstallation.
31112 FFCustomMsgBox [31112] Failed in image The system was unstable. Restart. If the same symptom recurs,
inputting process. perform a reinstallation.
31113 FFCustomMsgBox [31113] Failed in image The system was unstable. Restart. If the same symptom recurs,
inputting process. perform a reinstallation.
31114 FFCustomMsgBox [31114] Failed in specifying The system was unstable. Restart. If the same symptom recurs,
the urgent processing. perform a reinstallation.
31127 FFCustomMsgBox [31127] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study with QA completed).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-62
Event Event Source Description Causes Remedies Remarks
31128 FFCustomMsgBox [31128] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption (only exposure damaged or deleted. perform a reinstallation.
competed).
31129 FFCustomMsgBox [31129] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study interrupt button).
31130 FFCustomMsgBox [31130] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Cancel button).
31131 FFCustomMsgBox [31131] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study reception tab).
31132 FFCustomMsgBox [31132] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(waiting for study tab).
31133 FFCustomMsgBox [31133] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(waiting for QA tab).
31134 FFCustomMsgBox [31134] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Waiting for output tab).
31135 FFCustomMsgBox [31135] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Already output tab).
31136 FFCustomMsgBox [31136] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(All tab).
31137 FFCustomMsgBox [31137] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Today tab).
31138 FFCustomMsgBox [31138] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (System damaged or deleted. perform a reinstallation.
button).
31139 FFCustomMsgBox [31139] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study reservation button).
31140 FFCustomMsgBox [31140] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study start button).
31141 FFCustomMsgBox [31141] Patient information An unacceptable character was contained. Make a proper entry.
is illegal.
31142 FFCustomMsgBox [31142] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-63
Event Event Source Description Causes Remedies Remarks
31143 FFCustomMsgBox [31143] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31144 FFCustomMsgBox [31144] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31145 FFCustomMsgBox [31145] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31146 FFCustomMsgBox [31146] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31147 FFCustomMsgBox [31147] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31148 FFCustomMsgBox [31148] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31149 FFCustomMsgBox [31149] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group data. damaged or deleted. perform a reinstallation.
31152 FFCustomMsgBox [31152] Failed in obtaining The system was unstable. Restart. If the same symptom recurs,
the exposure menu perform a reinstallation.
information.
31153 FFCustomMsgBox [31153] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Urgent damaged or deleted. perform a reinstallation.
button).
31154 FFCustomMsgBox [31154] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Study damaged or deleted. perform a reinstallation.
completed button).
31155 FFCustomMsgBox [31155] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Update damaged or deleted. perform a reinstallation.
button).
31158 FFCustomMsgBox [31158] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (New damaged or deleted. perform a reinstallation.
button).
31159 FFCustomMsgBox [31159] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Repeat damaged or deleted. perform a reinstallation.
button).
31160 FFCustomMsgBox [31160] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Re-shoot damaged or deleted. perform a reinstallation.
button).
31161 FFCustomMsgBox [31161] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Add damaged or deleted. perform a reinstallation.
button).
31162 FFCustomMsgBox [31162] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Delete damaged or deleted. perform a reinstallation.
button).
31163 FFCustomMsgBox [31163] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Before damaged or deleted. perform a reinstallation.
scroll button).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-64
Event Event Source Description Causes Remedies Remarks
31164 FFCustomMsgBox [31164] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (After damaged or deleted. perform a reinstallation.
Scroll button).
31165 FFCustomMsgBox [31165] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Monitor damaged or deleted. perform a reinstallation.
mode button).
31166 FFCustomMsgBox [31166] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (QA damaged or deleted. perform a reinstallation.
button).
31167 FFCustomMsgBox [31167] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Change damaged or deleted. perform a reinstallation.
button).
31168 FFCustomMsgBox [31168] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Rotation/ damaged or deleted. perform a reinstallation.
Inversion button).
31169 FFCustomMsgBox [31169] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (EDR damaged or deleted. perform a reinstallation.
mode button).
31170 FFCustomMsgBox [31170] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Image damaged or deleted. perform a reinstallation.
format button).
31171 FFCustomMsgBox [31171] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Change damaged or deleted. perform a reinstallation.
menu parameter button).
31172 FFCustomMsgBox [31172] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (New button). damaged or deleted. perform a reinstallation.
31173 FFCustomMsgBox [31173] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Repeat button). damaged or deleted. perform a reinstallation.
31174 FFCustomMsgBox [31174] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Re-shoot damaged or deleted. perform a reinstallation.
button).
31175 FFCustomMsgBox [31175] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Add button). damaged or deleted. perform a reinstallation.
31176 FFCustomMsgBox [31176] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Delete button). damaged or deleted. perform a reinstallation.
31177 FFCustomMsgBox [31177] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Before scroll damaged or deleted. perform a reinstallation.
button).
31178 FFCustomMsgBox [31178] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (After scroll damaged or deleted. perform a reinstallation.
button).
31179 FFCustomMsgBox [31179] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Monitor mode damaged or deleted. perform a reinstallation.
button).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-65
Event Event Source Description Causes Remedies Remarks
31180 FFCustomMsgBox [31180] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (QA button ). damaged or deleted. perform a reinstallation.
31181 FFCustomMsgBox [31181] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Change damaged or deleted. perform a reinstallation.
button).
31182 FFCustomMsgBox [31182] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Rotation/ damaged or deleted. perform a reinstallation.
Inversion button).
31183 FFCustomMsgBox [31183] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (EDR mode damaged or deleted. perform a reinstallation.
button).
31184 FFCustomMsgBox [31184] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Image format damaged or deleted. perform a reinstallation.
button).
31185 FFCustomMsgBox [31185] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip (Change menu damaged or deleted. perform a reinstallation.
parameter button).
31186 FFCustomMsgBox [31186] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information (Set information was damaged or deleted. perform a reinstallation.
No.).
31187 FFCustomMsgBox [31187] Failed in setting the It is conceivable that the registry Restart. If the same symptom recurs,
tool tip (Before scroll information was damaged or deleted. perform a reinstallation.
button).
31188 FFCustomMsgBox [31188] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (System ID).
31189 FFCustomMsgBox [31189] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Auto-
distribution instruction
information).
31191 FFCustomMsgBox [31191] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Selector
auto-switchover
information).
31192 FFCustomMsgBox [31192] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Menu
auto-sequence
information).
31193 FFCustomMsgBox [31193] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Set No.
minimum value
information).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-66
Event Event Source Description Causes Remedies Remarks
31194 FFCustomMsgBox [31194] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Set No.
maximum value
information).
31196 FFCustomMsgBox [31196] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Series
information issuing
method).
31197 FFCustomMsgBox [31197] No matching work The contents of the image file did not Initialize the image database.
ID exists. agree with those of the image database.
31198 FFCustomMsgBox [31198] Failed in distribution The system was unstable. Restart. If the same symptom recurs,
process. perform a reinstallation.
31201 FFInImgMonitorCtrl Data error.%n[%1] Input data was incorrect. Restart the system.
31202 FFCustomMsgBox [31202] Failed in obtaining It is conceivable that the database file was Check the FPD and the 5000 Series
the system setting damaged or deleted. equipment. Perform an image input
information (Patient process again or restart.
information online serial
port No.).
31202 FFIIPInImgCtrl Initialize error. [Detailed An abnormality occurred in the initialization Restart the system. If the same symptom
error information 31202 14 process. The common probable causes occurs after a restart, perform a
items] are as follows: reinstallation and initialize the image
database.
• Installation was not normally ended.
• The parameter file was incorrect.
31202 FFInImgMonitorCtrl Initialize error.%n[%1] Failed in initialization processing. Restart the system.
31203 FFCustomMsgBox [31203] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Patient
information online device
name.).
31203 FFInImgMonitorDispCtrl Method process error.%n An error was detected during method Restart the system.
[%1] processing.
31204 FFInImgMonitorDispCtrl Windows API access error. The system was unstable. Restart the system.
%n[%1]
31205 FFInImgMonitorDispCtrl Parameter error.%n[%1] An abnormal parameter was detected. Restart the system.
31206 FFInImgMonitorDispCtrl Memory access error. System memory was insufficient. Restart the system.
%n[%1]
31207 FFInImgMonitorDispCtrl Function error.%n[%1] An error was detected during internal Restart the system.
processing.
31207 FFCustomMsgBox [31207] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
Image Reader machine reinstall the system.
code.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-67
Event Event Source Description Causes Remedies Remarks
31208 FFCustomMsgBox [31208] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (New damaged or deleted. perform a reinstallation.
button).
31208 FFInImgMonitorDispCtrl File access error.%n[%1] The system was unstable. Restart the system.
31210 FFCustomMsgBox [31210] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Image rotation/
inversion selectable range).
31211 FFCustomMsgBox [31211] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (frontal damaged or deleted. perform a reinstallation.
image button).
31212 FFCustomMsgBox [31212] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Left-right damaged or deleted. perform a reinstallation.
invert button).
31213 FFCustomMsgBox [31213] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Up-side- damaged or deleted. perform a reinstallation.
down invert button).
31214 FFCustomMsgBox [31214] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (180 damaged or deleted. perform a reinstallation.
degree rotation button).
31215 FFCustomMsgBox [31215] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (90 damaged or deleted. perform a reinstallation.
degree rotation button).
31218 FFCustomMsgBox [31218] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Image format
selectable range).
31219 FFCustomMsgBox [31219] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (SINGLE damaged or deleted. perform a reinstallation.
button).
31220 FFCustomMsgBox [31220] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (TWIN damaged or deleted. perform a reinstallation.
button).
31225 FFCustomMsgBox [31225] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the connection characters or deleted. perform a reinstallation.
to the film annotation
database.
31226 FFCustomMsgBox [31226] Failed in It is conceivable that the database file was Restart. If the same symptom recurs,
connecting characters to damaged or deleted. perform a reinstallation.
the film annotation
database.
31227 FFCustomMsgBox [31227] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
film annotation database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-68
Event Event Source Description Causes Remedies Remarks
31232 FFCustomMsgBox [31232] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (upper limit of
No. of output film).
31235 FFCustomMsgBox [31235] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(CS).
31236 FFCustomMsgBox [31236] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(SS).
31241 FFCustomMsgBox [31241] Failed in creating The system was unstable. Restart. If the same symptom recurs,
thumbnail images with the perform a reinstallation.
image input subsystem.
31242 FFCustomMsgBox [31242] Failed in The system was unstable. Restart. If the same symptom recurs,
processing in image input perform a reinstallation.
subsystem.
31243 FFCustomMsgBox [31243] Image format A value in the image file conflicted with a Since a standard setting will be used as a
information is illegal. system setting. substitute to continue with the process, no
particular response is required.
31244 FFCustomMsgBox [31244] Annotation A value in the image file conflicted with a Since a standard setting will be used as a
information is illegal. system setting. substitute to continue with the process, no
particular response is required.
31245 FFCustomMsgBox [31245] No. of output film A value in the image file conflicted with a Since a standard setting will be used as a
information is illegal. system setting. substitute to continue with the process, no
particular response is required.
31246 FFCustomMsgBox [31246] Distribution code A value in the image file conflicted with a Since a standard setting will be used as a
information is illegal. system setting. substitute to continue with the process, no
particular response is required.
31247 FFCustomMsgBox [31247] Image reading A value in the image file conflicted with a Since a standard setting will be used as a
sensitivity is illegal. system setting. substitute to continue with the process, no
particular response is required.
31248 FFCustomMsgBox [31248] Average output A value in the image file conflicted with a Since a standard setting will be used as a
density is illegal. system setting. substitute to continue with the process, no
particular response is required.
31249 FFCustomMsgBox [31249] EDR mode is A value in the image file conflicted with a Since a standard setting will be used as a
illegal. system setting. substitute to continue with the process, no
particular response is required.
31250 FFCustomMsgBox [31250] Image filing mode A value in the image file conflicted with a Since a standard setting will be used as a
is illegal. system setting. substitute to continue with the process, no
particular response is required.
31251 FFCustomMsgBox [31251] Function mode is A value in the image file conflicted with a Since a standard setting will be used as a
illegal. system setting. substitute to continue with the process, no
particular response is required.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-69
Event Event Source Description Causes Remedies Remarks
31253 FFCustomMsgBox [31253] Contrast shift is A value in the image file conflicted with a Since a standard setting will be used as a
illegal. system setting. substitute to continue with the process, no
particular response is required.
31254 FFCustomMsgBox [31254] Sensitivity shift A value in the image file conflicted with a Since a standard setting will be used as a
information is illegal. system setting. substitute to continue with the process, no
particular response is required.
31255 FFCustomMsgBox [31255] There is a A value in the image file conflicted with a Since a standard setting will be used as a
discrepancy between system setting. substitute to continue with the process, no
system setting and image particular response is required.
information.
31256 FFCustomMsgBox [31256] There is a A value in the image file conflicted with a Since a standard setting will be used as a
discrepancy between system setting. substitute to continue with the process, no
system setting and image particular response is required.
information.
31257 FFCustomMsgBox [31257] There is a A value in the image file conflicted with a Since a standard setting will be used as a
discrepancy between system setting. substitute to continue with the process, no
system setting and image particular response is required.
information.
31258 FFCustomMsgBox [31258] There is a A value in the image file conflicted with a Since a standard setting will be used as a
discrepancy between system setting. substitute to continue with the process, no
system setting and image particular response is required.
information.
31259 FFCustomMsgBox [31259] Failed in obtaining It is conceivable that the image information Delete the associated examination.
the image invert/rotate was damaged. If the same symptom frequently recurs,
information. perform a reinstallation.
31260 FFCustomMsgBox [31260] Failed in setting the A value in the image file conflicted with a Since a standard setting will be used as a
image format information. system setting. substitute to continue with the process, no
particular response is required.
31261 FFCustomMsgBox [31261]Failed in obtaining It is conceivable that the image information Delete the associated examination.
the image format was damaged. If the same symptom frequently recurs,
information. perform a reinstallation.
31262 FFCustomMsgBox [31262] Film annotation is The registered film mark contained an Correct the film mark setup with a user
illegal. unacceptable character. utility.
31265 FFCustomMsgBox [31265] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(FIX).
31266 FFCustomMsgBox [31266] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(S value).
31267 FFCustomMsgBox [31267] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(SEMI-X).
31268 FFCustomMsgBox [31268] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Density Ave)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-70
Event Event Source Description Causes Remedies Remarks
31269 FFCustomMsgBox [31269] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Density Average).
31270 FFCustomMsgBox [31270] Input value is The entered value contained an Make a proper entry.
illegal. unacceptable character or was out of
range or otherwise abnormal.
31272 FFCustomMsgBox [31272] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Function
selectable range).
31273 FFCustomMsgBox [31273] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
function information. perform a reinstallation.
31274 FFCustomMsgBox [31274] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
No. of output film. perform a reinstallation.
31277 FFCustomMsgBox [31277] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
contrast shift. perform a reinstallation.
31278 FFCustomMsgBox [31278] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
sensitivity shift. perform a reinstallation.
31279 FFCustomMsgBox [31279] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
EDR mode information. perform a reinstallation.
31280 FFCustomMsgBox [31280] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
FIX mode S value. perform a reinstallation.
31281 FFCustomMsgBox [31281] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
SEMI-X mode average perform a reinstallation.
output density.
31282 FFCustomMsgBox [31282] Failed in QA The system was unstable. Restart. If the same symptom recurs,
starting process. perform a reinstallation.
31284 FFCustomMsgBox [31284] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Filing mode
selectable range).
31285 FFCustomMsgBox [31285] Failed in setting the The system was unstable. Restart. If the same symptom recurs,
image filing information. perform a reinstallation.
31288 FFCustomMsgBox [31288] Failed in obtaining The system was unstable. Restart. If the same symptom recurs,
the distribution destination perform a reinstallation.
information.
31291 FFCustomMsgBox [31291] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Distribution
code change).
31292 FFCustomMsgBox [31292] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Film annotation
change).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-71
Event Event Source Description Causes Remedies Remarks
31293 FFCustomMsgBox [31293] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Function
change).
31294 FFCustomMsgBox [31294] Failed in The system was unstable. Restart. If the same symptom recurs,
terminating process. perform a reinstallation.
31295 FFCustomMsgBox [31295] Failed in The system was unstable. Restart. If the same symptom recurs,
terminating process. perform a reinstallation.
31296 FFCustomMsgBox [31296] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list display. perform a reinstallation.
31303 FFCustomMsgBox [31303] No list (Waiting for It is highly probable that the program was
study) is selected. faulty.
31304 FFCustomMsgBox [31304] No list (Waiting for It is highly probable that the program was
QA) is selected. faulty.
31305 FFCustomMsgBox [31305] No list (Waiting for It is highly probable that the program was
output) is selected. faulty.
31306 FFCustomMsgBox [31306] No list (Already It is highly probable that the program was
output) is selected. faulty.
31307 FFCustomMsgBox [31307] No list (All) is It is highly probable that the program was
selected. faulty.
31308 FFCustomMsgBox [31308] No list (Today) is It is highly probable that the program was
selected. faulty.
31309 FFCustomMsgBox [31309] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (Waiting for study) perform a reinstallation.
display.
31310 FFCustomMsgBox [31310] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (Waiting for QA) perform a reinstallation.
display.
31311 FFCustomMsgBox [31311] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (Waiting for output) perform a reinstallation.
display.
31312 FFCustomMsgBox [31312] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (Already output) perform a reinstallation.
display.
31313 FFCustomMsgBox [31313] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (All) display. perform a reinstallation.
31314 FFCustomMsgBox [31314] Failed in updating The system was unstable. Restart. If the same symptom recurs,
the list (Today) display. perform a reinstallation.
31315 FFCustomMsgBox [31315] No. of selected It is highly probable that the program was
study is illegal. Study faulty.
cannot be started.
31316 FFCustomMsgBox [31316] Failed in study The system was unstable. Restart. If the same symptom recurs,
starting process. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-72
Event Event Source Description Causes Remedies Remarks
31317 FFCustomMsgBox [31317] No. of selected It is highly probable that the program was
studies is illegal. QA cannot faulty.
be started.
31318 FFCustomMsgBox [31318] Failed in QA The system was unstable. Restart. If the same symptom recurs,
starting process. perform a reinstallation.
31319 FFCustomMsgBox [31319] Failed in QA The system was unstable. Restart. If the same symptom recurs,
starting process. perform a reinstallation.
31320 FFCustomMsgBox [31320] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Operational
mode after studying).
31321 FFCustomMsgBox [31321] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Utility damaged or deleted. perform a reinstallation.
button).
31322 FFCustomMsgBox [31322] Failed in obtaining The system was unstable. Restart. If the same symptom recurs,
the patient information. perform a reinstallation.
31323 FFCustomMsgBox [31323] Date of birth could Illegal birth date information was set. Delete the associated examination. If the
not be converted to the same symptom recurs after deletion,
displaying format. perform a reinstallation.
31324 FFCustomMsgBox [31324] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Stop alarm button).
31330 FFCustomMsgBox [31330] Failed in the It is conceivable that the database file was Restart. If the same symptom recurs,
obtaining the caption damaged or deleted. perform a reinstallation.
information (System
Terminating button).
31331 FFCustomMsgBox [31331] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(User Utility button).
31332 FFCustomMsgBox [31332] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Study list setting button).
31333 FFCustomMsgBox [31333] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Version displaying button).
31334 FFCustomMsgBox [31334] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Cancel button).
31335 FFCustomMsgBox [31335] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Clear damaged or deleted. perform a reinstallation.
button).
31336 FFCustomMsgBox [31336] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip information damaged or deleted. perform a reinstallation.
(Clear button).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-73
Event Event Source Description Causes Remedies Remarks
31337 FFCustomMsgBox [31337] Failed in displaying The thumbnail image file was damaged, There are no processing problems except
the thumbnail images. deleted, or otherwise abnormal. for the inability to display thumbnail
images. If the same symptom frequently
occurs, initialize the image database.
31340 FFCustomMsgBox [31340] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Cancel button).
31341 FFCustomMsgBox [31341] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Cancel button).
31348 FFCustomMsgBox [31348] Failed in thumbnail The thumbnail image file was damaged, There are no processing problems except
image creation process. deleted, or otherwise abnormal. for the inability to display thumbnail
images. If the same symptom frequently
occurs, initialize the image database.
31349 FFCustomMsgBox [31349] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (Waiting for study). unstable or that the program was faulty. perform a reinstallation.
31350 FFCustomMsgBox [31350] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (Waiting for QA). unstable or that the program was faulty. perform a reinstallation.
31351 FFCustomMsgBox [31351] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (Waiting for output). unstable or that the program was faulty. perform a reinstallation.
31352 FFCustomMsgBox [31352] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (Already output). unstable or that the program was faulty. perform a reinstallation.
31353 FFCustomMsgBox [31353] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (All). unstable or that the program was faulty. perform a reinstallation.
31354 FFCustomMsgBox [31354] Failed in displaying It is conceivable that the system was If the same symptom recurs after a restart,
the list (Today). unstable or that the program was faulty. perform a reinstallation.
31355 FFCustomMsgBox [31355] Failed in It is conceivable that the system was If the same symptom recurs after a restart,
customizing the list display. unstable or that the program was faulty. perform a reinstallation.
31356 FFCustomMsgBox [31356] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Default menu
code).
31358 FFCustomMsgBox [31358] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the tool tip text. damaged or deleted. perform a reinstallation.
31359 FFCustomMsgBox [31359] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Title bar).
31360 FFCustomMsgBox [31360] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information damaged or deleted. perform a reinstallation.
(Option label).
31361 FFCustomMsgBox [31361] Failed in obtaining It is conceivable that the file was damaged Restart. If the same symptom recurs,
the version information. or deleted. perform a reinstallation.
31362 FFCustomMsgBox [31362] Option list could It is conceivable that the registry Restart. If the same symptom recurs,
not be obtained. information was damaged or deleted. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-74
Event Event Source Description Causes Remedies Remarks
31363 FFCustomMsgBox [31363] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information information was damaged or deleted. perform a reinstallation.
(Next multiple frame set
No.).
31364 FFCustomMsgBox [31364] Failed in setting the It is conceivable that the registry Restart. If the same symptom recurs,
registry information (Next information was damaged or deleted. perform a reinstallation.
multi-frame set No.).
31365 FFCustomMsgBox [31365] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
information (Series issuing
type).
31366 FFCustomMsgBox [31366] Failed in obtaining The system was unstable. Restart. If the same symptom recurs,
the multi-frame UID perform a reinstallation.
information.
31367 FFCustomMsgBox [31367] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the cassette insertion damaged or deleted. perform a reinstallation.
enabled image.
31368 FFCustomMsgBox [31368] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the cassette insertion damaged or deleted. perform a reinstallation.
disabled image.
31369 FFCustomMsgBox [31369] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the image icon (Mis- damaged or deleted. perform a reinstallation.
exposure).
31370 FFCustomMsgBox [31370] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Image damaged or deleted. perform a reinstallation.
style button).
31371 FFCustomMsgBox [31371] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (EDR damaged or deleted. perform a reinstallation.
button).
31372 FFCustomMsgBox [31372] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Function damaged or deleted. perform a reinstallation.
button).
31373 FFCustomMsgBox [31373] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (No. of damaged or deleted. perform a reinstallation.
output film button).
31375 FFCustomMsgBox [31375] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Film damaged or deleted. perform a reinstallation.
format button).
31376 FFCustomMsgBox [31376] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Filing damaged or deleted. perform a reinstallation.
button).
31377 FFCustomMsgBox [31377] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (Film damaged or deleted. perform a reinstallation.
annotation button).
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-75
Event Event Source Description Causes Remedies Remarks
31378 FFCustomMsgBox [31378] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image damaged or deleted. perform a reinstallation.
(Distribution code button).
31379 FFCustomMsgBox [31379] Failed in obtaining It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the button image (CS/SS damaged or deleted. perform a reinstallation.
button).
31382 FFCustomMsgBox [31382] Image data line to The line was abnormal. Check the line.
the Image Reader (%s) has
broken.
31386 FFCustomMsgBox [31386] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the caption information. damaged or deleted. perform a reinstallation.
31387 FFCustomMsgBox [31387] Failed in reading It is conceivable that the bitmap file was Restart. If the same symptom recurs,
the bit-map file. damaged or deleted. perform a reinstallation.
31388 FFCustomMsgBox [31388] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the font information. information was damaged or deleted. perform a reinstallation.
31389 FFCustomMsgBox [31389] Failed in refresh The system was unstable. Restart. If the same symptom recurs,
the list. reinstall the system.
31390 FFCustomMsgBox [31390] The information The entered information or its format was
about a requesting in error.
department was illegal.
31391 FFCustomMsgBox [31391] The information The entered information or its format was
about a paramedic was in error.
illegal.
31392 FFCustomMsgBox [31392] A patient Nothing matched the entered patient Check the patient information.
information search was information.
unsuccessful.
31393 FFCustomMsgBox [31393] Failed in setting up This error is attributable to an RIS problem Check RIS-side status or RIS setup.
connection status with the or RIS setup problem.
RIS.
31394 FFCustomMsgBox [31394] The line connected This error is attributable to an RIS problem Check the RIS status and RIS setup.
to the RIS was faulty. or RIS setup problem.
31395 FFCustomMsgBox [31395] Failed in reporting It is conceivable that the server was Check operating condition of the server
of study start (PPS). inoperative or network settings were and network settings.
incorrect.
31396 FFCustomMsgBox [31396] Failed in reporting It is conceivable that the server was Check operating condition of the server
of study termination (PPS). inoperative or network settings were and network settings.
incorrect.
31397 FFCustomMsgBox [31397] Failed in reporting It is conceivable that the server was Check operating condition of the server
of study abort (PPS). inoperative or network settings were and network settings.
incorrect.
31389 FFCustomMsgBox [31399] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the label caption. damaged or deleted. reinstall the system.
31400 FFCustomMsgBox [31400]The selected menu This error is attributable to an RIS problem Check the RIS status and RIS setup.
did not exist. or RIS setup problem.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-76
Event Event Source Description Causes Remedies Remarks
31401 FFCustomMsgBox [31401] Path information It is conceivable that the file was damaged Restart. If the same symptom recurs,
(image data folder) or deleted. perform a reinstallation.
acquisition was not
successful.
31402 FFCustomMsgBox [31402] Parameter icon It is conceivable that the bitmap file was Restart. If the same symptom recurs,
image acquisition was not damaged or deleted. perform a reinstallation.
successful.
31403 FFCustomMsgBox [31403] Caption information It is conceivable that the database file was Restart. If the same symptom recurs,
acquisition was not damaged or deleted. perform a reinstallation.
successful.
31404 FFCustomMsgBox [31404] Failed in This error is attributable to a problem with Check the connected equipment,
communicating with other the connected equipment or connection connection setup, and others.
connected device. setup.
31407 FFCustomMsgBox [31407] Failed in setting the The connected equipment may be faulty. Check the connected equipment.
exposure size.
31408 FFCustomMsgBox [31408] Failed in setting the The connected equipment may be faulty. Check the connected equipment.
reference position for
exposure.
31409 FFCustomMsgBox [31409] No response The connected equipment may be faulty. Check the connected equipment.
comes from other
connected device.
31410 FFCustomMsgBox [31410] There is no answer The connected equipment may be faulty. Check the connected equipment.
from the device that ID
information is shared.
31411 FFCustomMsgBox [31411] A Study can’t This error is attributable to an improperly Check the menu registration information.
register because the registered menu.
exposure menu registerd in
IDT isn’t registered in IIP.
31412 FFCustomMsgBox [31412] An image data line The connected equipment may be faulty. Check the connected equipment.
with the reader is cut off.
31414 FFCustomMsgBox [31414] Failed in obtaining The database file may be damaged or Restart. If the same symptom recurs,
the caption information (RIS deleted. perform a reinstallation.
tab).
31415 FFCustomMsgBox [31415] A communication The connected equipment may be faulty. Check the connected equipment.
error with RIS occurred.
31416 FFCustomMsgBox [31416] Some exposure This error is attributable to an improperly Check the settings.
menus couldn’t be set menu parameter or other setup failure.
registered.
31417 FFCustomMsgBox [31417] An exposure menu This error is attributable to an improperly Check the settings.
couldn’t be registered. set menu parameter or other setup failure.
31418 FFCustomMsgBox [31418] Failed in the The connection setup may be faulty. Check the connection setup.
reference of the reader
device.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-77
Event Event Source Description Causes Remedies Remarks
31420 FFCustomMsgBox [31420] A failure occurred It is conceivable that the connected Check the connected equipment.
in the communication with equipment may be faulty.
the RIS.
31422 FFCustomMsgBox [31422] Failed in registering The patient information database operation Restart. If the same symptom recurs,
the study to the patient was wrong. perform a reinstallation.
information database.
31423 FFCustomMsgBox [31423] Failed in image The system was unstable. Restart. If the same symptom recurs,
stitching processing. perform a reinstallation.
Ensure that only period is used as a
symbol representing a decimal point for
values to be set in “OS Settings - Regional
Options”. If any symbol other than a period
is used, correct it appropriately.
31424 FFCustomMsgBox [31424] Failed in saving the The system was unstable. Restart. If the same symptom recurs,
exposure result information. perform a reinstallation.
31425 FFCustomMsgBox [31425] Failed in setting Invalid characters were used for the No remedial action is needed because the
actual exposure result performed procedure step description, process will be continuously performed
value. protocol name, or filter type. The radiation with the invalid settings replaced with the
source-to-detector distance or radiation standard ones.
time did not consist of numerals only.
The film output count or film size ID was
out of range.
31426 FFCustomMsgBox [31426] Failed in acquiring Settings may have been corrupted. Restart the system. If the same symptom
coloring information. recurs, perform a reinstallation.
31427 FFCustomMsgBox [31427] Failed in image The system was unstable. Restart the system. If the same symptom
distribution processing. recurs, perform a reinstallation.
31428 FFCustomMsgBox [31428] Failed in displaying The system was unstable. Restart the system. If the same symptom
the study information list. recurs, perform a reinstallation.
31429 FFCustomMsgBox [31429] Failed in setting The system was unstable. Restart the system. If the same symptom
patient information. recurs, perform a reinstallation.
31430 FFCustomMsgBox [31430] Failed in acquiring The system was unstable. Restart the system. If the same symptom
patient information. recurs, perform a reinstallation.
31431 FFCustomMsgBox [31431] Failed in acquiring The bitmap file may have been damaged Restart the system. If the same symptom
image (background). or deleted. recurs, perform a reinstallation.
31441 FFCustomMsgBox [31441] Failed in acquiring It is conceivable that the file was damaged Restart. If the same symptom recurs,
parameters used for or deleted. It is also conceivable that the perform a reinstallation.
displaying mammography database file was damaged or deleted.
images.
31444 FFCustomMsgBox [31444] Illegal information Polling to the shared folder with the receipt It is conceivable that the content of the
received from the outside. computer resulted in a failure. study received from the receipt computer
does not match the DR-ID 300CL setting.
Check the connection setting and the
system setting.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-78
Event Event Source Description Causes Remedies Remarks
31445 FFCustomMsgBox [31445] Illegal information The patient cannot be registered because Check to see that characters prohibited to
received from the outside. of a format error that occurred with the text use on the DR-ID 300CL are not included
received from the Fujitsu receipt computer. in the transmitted information. Retransmit
then the study from the receipt computer,
as necessary.
31446 FFCustomMsgBox [31446] An error occurred A study transmitted from the receipt Check to see that characters prohibited to
for connection with the computer cannot be registered. use on the DR-ID 300CL are not included
equipment. in the transmitted information. Retransmit
then the study from the receipt computer,
as necessary.
31448 FFCustomMsgBox [31448] System error An error occurred in the FPD. Restart the FPD system. If the same error
occurred. recurs, reinstall the system.
31449 FFCustomMsgBox [31449] System error A disconnection error occurred in the line Check connection status to the equipment
occurred in to the FPD system. and restart the FPD system. If the same
communications. error recurs, reinstall the system.
31449 FFCustomMsgBox [31449] Connecting… A disconnection error occurred in the line Check connection status to the equipment. This is an existing
* Please wait for a moment. to the FPD system. If it is connected correctly, but the message message, but was
does not disappear after a while, restart revised so that it
the FPD system. disappears when
If the same error recurs after restarting the returning to FPD
system, reinstall the system. system line. (Dialog
can be closed via the
user operation)
31450 FFCustomMsgBox [31450] System error An error occurred on the FPD side (a error Restart the FPD system. If the same error
occurred. of level: 1,2 originated from the FPD recurs, reinstall the system.
system).
31451 FFCustomMsgBox [31451] Failed in image Due to an error occurred during image It is possible that the image can be re-
input processing. input processing, acquisition of an image acquired if it remains in the CU/MC.
If you carry input from the CU/MC or image reconstruction If the user abandoned to re-acquire the
processing out again, you based on the raw image resulted in a image, the user can delete the target
may succeed in input failure. menu.
image.
31455 FFCustomMsgBox [31455] An exposure menu This error can be attributable to the fact Check the FPD system and line status.
couldn’t be decided. that the user selected a menu on the study If such symptom does not improve, restart
window and pressed the button halfway the FPD system. If the same error recurs,
with the LED unlit. reinstall the system.
The LED may not go on by the causes
below.
FPD panel error
Line status error
Exposure disabled due to emergency
Inconsistency found in the database
31460 FFCustomMsgBox [31460] Initialization of DAP Initialization of the DAP meter resulted in a Check connection to the DAP meter and DAP
meter has failed. failure. restart the system.
010-305-18E
03.31.2017 FM9475
DR-ID 300CL Service Manual Appx MT A-79
Event Event Source Description Causes Remedies Remarks
31461 FFCustomMsgBox [31461] Initialization of DAP Initialization of the DAP meter resulted in a Retry initialization processing or suspend DAP
meter has failed. failure. the study and then restart it after checking
connection to the DAP meter.
31462 FFCustomMsgBox [31462] Dose acquisition Failed in acquiring a dose value from the Confirm the dose value with the dose-area DAP
from DAP meter has failed. DAP meter. product meter. Check connection to the
DAP meter as necessary.
31464 FFCustomMsgBox [31464] Registration of An error occurred in an attempt to register Reinstall the system. Re-scanning.
study information has study information.
failed.
31466 FFCustomMsgBox [31466] No selector is An IP is registered and the study is Review the selector setting. Cancellation of
applicable. suspended. The Console is then shut registered IP.
down. While it is being shut down, the
selector where the IP has been registered
was deleted accidentally.
31467 FFCustomMsgBox [31467] Failed in energy Failed in performing ES (energy- Reinstall the system. ES processing
subtraction processing. subtraction) processing.
31468 FFCustomMsgBox [31468] Failed in exposure A failure occurred in exposure abort Perform re-exposure as necessary. Exposure abort
abort processing. processing. processing
31469 FFCustomMsgBox [31469] A system Disconnection of the line to the Shimadzu Reconfirm FPD device settings. Line disconnected.
communication error was FPD system (while in process of ID
detected. determination).
31850 FFCustomMsgBox [31850] Failed in It is conceivable that the connected Check status of the connected equipment.
communicating with X-ray equipment may have a problem or the
equipment. network may be unstable.
31851 FFCustomMsgBox [31851] Failed in registering The system is unstable. Restart the system. If the same symptom
a reserved study. recurs, reinstall the system.
31852 FFCustomMsgBox [31852] Failed in starting a The system is unstable. Restart the system. If the same symptom
reserved study. recurs, reinstall the system.
31853 FFCustomMsgBox [31853] An error occurred The system is unstable. Restart the system. If the same symptom
with connection to the recurs, reinstall the system.
equipment.
31854 FFCustomMsgBox [31854] Failed in searching The system is unstable. Restart the system. If the same symptom
a reserved study. recurs, reinstall the system.
31855 FFCustomMsgBox [31855] Received illegal It is conceivable that the connected Check information transmitted from the
information from the equipment may have transmitted illegal connected equipment.
outside. information.
31856 FFCustomMsgBox [31856] Received illegal It is conceivable that the connected Check information transmitted from the
information from the equipment may have transmitted illegal connected equipment.
outside. information.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-80
Event Event Source Description Causes Remedies Remarks
32001 FFIIPInImgCtrl Initialize warning. [Detailed A warning-level error occurred in an image Check whether the SQL service manager
error information 32001 35 database initialization process. The is normally operating. If the same symptom
items] common probable causes are as follows: occurs while the SQL service manager is
normally operating, initialize the image
• An abnormality occurred when the image database (the entire image data is then
database was accessed. deleted).
• An abnormality occurred when an
information file was accessed.
• An abnormality occurred when an image
file was created or referenced for image
file recovery.
• The contents of the image database or
information file were illegal.
• An unrecoverable image file was
encountered.
32001 FFInImgMonitorCtrl Initialize warning.%n[%1] A warning-level error occurred during Nothing particular. (Processing was
initialization processing. continued.)
32001 FFInImgMonitorDispCtrl Initialize warning.%n[%1] A warning-level error occurred during Nothing particular. (Processing was
initialization processing. continued.)
32002 FFIIPInImgCtrl Data not found.(warning) A warning-level error was found in Nothing in particular. (The data found
[Detailed error informationinformation file data. abnormal was replaced by the default
32002 1 item] value to continue with the process.)
32002 FFInImgMonitorCtrl Data not found.(warning) A warning-level error was detected in the Nothing particular. (The error data was
%n[%1] information file data. replaced by the default value and
processing was continued.)
32002 FFInImgMonitorDispCtrl Data not found.(warning) A warning-level error was detected in the Nothing particular. (The error data was
%n[%1] information file data. replaced by the default value and
processing was continued.)
32003 FFInImgMonitorCtrl Status warning. [Detailed A process request was issued while an Nothing in particular. (The process request
error information 32003 1 initialization or termination process was was rejected.)
item] being performed.
32003 FFInImgMonitorCtrl Status warning.%n[%1] A warning-level error occurred during Nothing particular. (The error data was
method processing. replaced by the default value and
processing was continued.)
32003 FFInImgMonitorDispCtrl Status warning.%n[%1] A warning-level error occurred during Nothing particular. (The error data was
method processing. replaced by the default value and
processing was continued.)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-81
Event Event Source Description Causes Remedies Remarks
32038 FFCustomMsgBox [32038] An error was An exposure technique notified from the Reconfirm FPD device settings. Exposure ready status
detected in connection with FPD device as an error is notified. (Console target #1912)
%s. Caption of the selector
where an error occurred
is included in “%s”. (If
errors occur in both of
the exposure
techniques, standing
position and supine
position, at the same
time, the both are
included putting a
comma between them.)
Example: “Error
occurred both in the
standing position, the
supine position.”
32039 FFCustomMsgBox [32039] An error message Though an error was notified from the FPD Make sure that settings for the selector
was received from device, no selector that is for the exposure and the FPD device have been performed
unspecified technique. technique exists. properly.
32043 FFCustomMsgBox [32043] Failed in linking the Output of the examinee-list definition file Ensure that a disk has been inserted into
patient info and display failed. the drive correctly.
terminal.
34084 FFCustomMsgBox [34084] Illegal information Illegal information was received from the Correct the information properly and then
was transmitted from the RIS. restart the study from the RIS.
RIS. Information of the started study includes
the one that is not allowed for the DR-ID
300CL.
34085 FFCustomMsgBox [34085] Communication A connected device may have a problem Reconnect. If the same symptom recurs,
with the RIS was disabled. or the network can be unstable. check status of the connected device.
34091 FFCustomMsgBox [34091] An error occurred An error of unknown cause occurred with Restart.
in an attempt to connect to the connection to the RIS.
the RIS.
34092 FFCustomMsgBox [34092] A failure occurred The system was unstable. Restart.
when starting the ordered
study.
34094 FFCustomMsgBox [34094] A failure occurred An attempt to inform the RIS of the Determine the same technologist both for
when notifying the RIS of changed technologist resulted in a failure. the RIS and the DR-ID 300CL.
the changed technologist.
34095 FFCustomMsgBox [34095] A failure occurred Communication with the RIS at startup of Stop the study exposure started via the
in communication with the an ordered study resulted in a failure. RIS and then restart it.
RIS at startup of a study.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-82
Event Event Source Description Causes Remedies Remarks
34096 FFCustomMsgBox [34096] A failure occurred Communication with the RIS at the end of Confirm the actual exposure result. If
in communication with the an order study resulted in a failure. information of the actual exposure result is
RIS when finishing a study. not reflected appropriately on the data,
input it on the RIS screen.
If the same symptom remains, restart the
relevant system.
34097 FFCustomMsgBox [34097] A failure occurred The system was unstable. Match technologist registration information
when changing the RIS of this equipment with that of the
user. connected device.
34098 FFCustomMsgBox [34098] Failed in Display of RIS view screen was disabled. Restart the relevant system. If the same
communicating with other symptom remains, check settings for
connected device. connection of the connected device.
34107 FFCustomMsgBox [34107] Do you wish to When rotating a marker-specified image, When the [Yes] button is pressed, the
delete markers and execute determine whether or not to delete the markers are deleted and the image with no
the operation? markers. markers is rotated. When the [No] button is
pressed, the image is rotated without
deleting the markers.
34109 FFCustomMsgBox [34109]Images were input During preview display of the 2 CR long- After all images have fully been read, Improvement of
during the preview display. view images before subjecting them to press the CR long-view image combining operation performance
combining processing, the third image was button. of the automatic image
input. combining processing.
34113 FFCustomMsgBox [34113] Automatic linkage A failure occurred in an attempt to perform Check the settings determined on the Long-view image
of stitching images has automatic combining processing of long- Image Reader (size, reference position, combining processing
failed. view menu images, which consequently offset value).
notifies the user of the fact that the images
have been combined as determined by the
default setting.
34114 FFCustomMsgBox [34114] Conditions Inconsistency between the exposure mode Check the settings determined on the Exposure initiation
inhibiting the start of and the specified number of image Image Reader.
exposure had been sent exposures, transferred from an external
from the external device. device, is notified to the user.
34115 FFCustomMsgBox [34115] Exposure was When the number of actually exposed The hand switch may have been released Exposure abort
aborted. images did not reach the number expected too early. Retry exposure.
for image exposure, the user will be
notified that the exposure was aborted.
34116 FFCustomMsgBox [34116] Technique cannot When a pre-determined exposure Register a new menu to perform exposure Exposure technique
be changed because it is a technique was changed for a Console- of the long-view menu whether in standing change
long view menu whose IP derived reason while an IP-registered long- position or supine position.
number has been view menu was being selected, the user
registered. will be notified that the exposure technique
cannot be changed.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-83
Event Event Source Description Causes Remedies Remarks
34117 FFCustomMsgBox [34117] Exposure technique In the case of receiving from the FPD a Retry after a while. Exposure technique
cannot be changed. notification to change exposure parameters change
(such as exposure technique and exposure
size) while exposure processing is being
inactivated, with, for example, the related
dialog box left opened or when the selector
button is pressed, the user will be notified
that changing the exposure parameters is
disabled.
34118 FFCustomMsgBox [34118] Long view image When the function menu flag is set to be Perform re-exposure as necessary. Exposure initiation
exposure will be aborted. WS, the user will be notified of aborting the
exposure processing.
34119 FFCustomMsgBox [34119] Exposure for When the function menu flag is set to be Perform re-exposure as necessary. Exposure initiation
energy subtraction will be ES (energy-subtraction processing), the
aborted. user will be notified of aborting the
exposure processing.
34120 FFCustomMsgBox [34120] Reader Unit QC Confirming whether or not to perform Operate as instructed on the window. Reading Unit QC
menu is being irradiated. Reading Unit QC processing, when an
attempt was made to shut down the
system while exposure of the Reading Unit
QC menu was being performed on the
study execution screen.
34121 FFCustomMsgBox [34121] The imaging range When an attempt was made to perform Check the number of image shootings.
is exceeded. exposure of long-view images of the
number larger than that supported by the
system, the user will be notified that the
shooting is disabled.
34122 FFCustomMsgBox [34122] Reader Unit QC Confirmation required when an attempt Operate as instructed on the window. Reading Unit QC
menu is registered, but not was made to exit the Reading Unit QC
all the exposure has been menu without performing it, though
completed. properly registered on the study execution
screen.
34123 FFCustomMsgBox [34123] Exposure size In the case of receiving from the FPD a If the dialog box is left open, close it and
cannot be changed. notification to change exposure parameters then retry.
(such as exposure size) while exposure
processing is being inactivated, with, for
example, the related dialog box left opened
or when the selector button is pressed, the
user will be notified that changing the
exposure parameters is disabled.
34842 FFCustomMsgBox [34842] Incorrect password. An incorrect password was input. Input a correct password. Password
authentication
component
39001 FFIIPInImgCtrl Memory allocate error. Sufficient memory was not available to the Restart.
[E1027] system.
39001 FFInImgMonitorCtrl Memory allocate error. System memory was insufficient. Restart.
%n[%1]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-84
Event Event Source Description Causes Remedies Remarks
39002 FFIIPInImgCtrl Message post/get error. The system was unstable. Restart.
[Detailed error information
39002 5 items]
39002 FFInImgMonitorCtrl Message post/get error. The system was unstable. Restart.
%n[%1]
39003 FFIIPInImgCtrl IIPInImg control error. An abnormality was found in the IIP input Restart.
[Detailed error information image process.
39003 19 items]
39004 FFIIPInImgCtrl CR-Image control error. An abnormal termination was detected in Restart.
[Detailed error information image file creation processing or image
39004 13 items] display processing. The disk may be faulty.
39004 FFInImgMonitorCtrl CR-Image control error. An error was detected during image Restart.
%n[%1] processing.
39005 FFIIPInImgCtrl Pipe error. [Detailed error The system remains unstable. Restart.
information 39005 1 item]
39005 FFInImgMonitorCtrl Pipe error. %n[%1] The system was unstable. Restart.
39006 FFIIPInImgCtrl File error. [Detailed error The system remains unstable. Restart.
information 39006 7 items]
39006 FFInImgMonitorCtrl File error.%n[%1] The system was unstable. Restart.
39007 FFIIPInImgCtrl Data not found.(error) The data in the image database or Make an exposure again. If the error
[Detailed error information information file was abnormal. recurs, initialize the image database (the
39007 3 items] entire image data is then deleted).
39007 FFInImgMonitorCtrl Data not found.(error) There is an error in the image database/ Restart.
%n[%1] information file data.
39008 FFIIPInImgCtrl DB access error. [Detailed An abnormality occurred when a database Restart. If the same symptom frequently
error information 39008 19 was accessed. occurs, perform a reinstallation.
items]
39008 FFIIPInImgMgr Insufficient memory. The available memory was insufficient. Restart.
%n[%1]
39008 FFInImgMonitorCtrl DB access error.%n[%1] An error occurred with accessing to several Restart.
databases.
39009 FFIIPInImgMgr Create Dicom Data Object Data object creation was not successfully Restart.
Failed. %n[%1] completed.
39010 FFIIPInImgCtrl DICOMDataObj data area. An exception occurred when accessing to Restart. If the same symptom recurs,
the information file (DDO). perform a reinstallation.
39010 FFIIPInImgCtrl Exception error. Exception handling was done. Restart. If the same symptom frequently
[Detailed error information occurs, perform a reinstallation.
39010 5 items]
39010 FFIIPInImgMgr Input Image Operation is The disk was full or the connection limit This is not an operating error. Wait a while
Busy. %n[%1] was exceeded by the input request (busy and then input again.
state).
39010 FFInImgMonitorCtrl Exception error.%n[%1] An exceptional error occurred. Restart.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-85
Event Event Source Description Causes Remedies Remarks
39010 FFInImgMonitorDispCtrl Exception error.%n[%1] An exceptional error occurred. Restart.
39045 FFCustomMsgBox [39045] Image input A failure occurred during image input Restart. If the same symptom recurs, DR #1277 - A measure
processing has failed. processing at recovery of a CR image. perform a reinstallation. to prevent images from
being lost accidentally.
39901 FFInitializingStudy [%1] An exception error occurred. Restart.
An exception error
occurred(%2).
39902 FFInitializingStudy [%1] An error occurred. Restart.
%2.
40001 FFCustomMsgBox [40001] An error occurred. The system was unstable. Restart. If the same symptom recurs,
perform a reinstallation.
40002 FFCustomMsgBox [40002] An error occurred. An error occurred in communication with Restart.
Ascent View (linking application).
40200 FFQAMsgBox Process Communication The method return value for the inter- Restart.
Control Error. process communication control resulted in
[File Name] [Function Name] an error.
[Error Code]
40201 FFQAMsgBox Journal Proc Failure. The method return value for the log save Restart.
[File Name] [Function Name] control resulted in an error.
[Error Code]
40202 FFQAMsgBox CustomMessageCtrl The return value for the custom message Restart.
Access Failure. box control resulted in an error.
[File Name] [Function Name]
[Error Code]
40203 FFQAMsgBox Image DB Access Control The method return value for the image Restart.
Error. database access control resulted in an
[File Name] [Function Name] error.
[Error Code]
40205 FFQAMsgBox Process Communication The method return value for the inter- Restart.
API Error. process communication API resulted in an
[File Name] [Function Name] error.
[Error Code]
40206 FFQAMsgBox SysConfig DB Access An error occurred with connection/ Restart.
Error. disconnection/value acquisition/value
[File Name] [Function Name] setup for the system setup database
[Error Code] access API.
40207 FFQAMsgBox LangStr DB Access Error. An error occurred with connection/ Restart.
[File Name] [Function Name] disconnection of the character string
[Error Code] database access API.
40208 FFQAMsgBox LangMsgBox DB Access An error occurred with connection/ Restart.
Error. disconnection of the message box
[File Name] [Function Name] database access API.
[Error Code]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-86
Event Event Source Description Causes Remedies Remarks
40209 FFQAMsgBox Flag DB Access Error. An error occurred with connection/value Restart.
[File Name] [Function Name] acquisition/value setup/disconnection of
[Error Code] the flag database access.
40210 FFQAMsgBox Return value of a The method return value for the image Restart.
ImgDeliveryCom makes an distribution instruction component resulted
error. in an error.
[File Name] [Function Name]
[Error Code]
40211 FFQAMsgBox NetConfig DB Access Error. An error occurred with connection/ Restart.
[File Name] [Function Name] disconnection/value acquisition/value
[Error Code] setup for the network setup database
(irreplaceable).
41001 FFCustomMsgBox [41001] The device is not No floppy disk was set in position. Set media that is appropriate for the
ready. device.
41002 FFCustomMsgBox [41002] Remaining capacity The file to be copied was too large to fit Switch to larger-capacity media.
is insufficient. onto the media at the receiving end.
41003 FFCustomMsgBox [41003] Destination for file An invalid pathname was specified as the Check the system setup.
copy is illegal. file copy destination.
41004 FFCustomMsgBox [41004] E-mail program Outlook was not properly set up. Exit IIP and then complete Outlook setup.
cannot be started.
41005 FFCustomMsgBox [41005] No data cannot be The disk was write-protected or damaged. Check the disk.
written on the disk.
41106 FFCustomMsgBox [41106] Hard Disk Full. The hard disk was full. Exit QA, delete unnecessary images from
the list, and retry the distribution.
41107 FFCustomMsgBo [41107] Hard Disk Full. The hard disk was full. Exit QA, delete unnecessary images from
the list, and retry the distribution.
41108 FFCustomMsgBox [41108] Failed in reading No JPEG image gradation correction file is Check the correction file.
the LUT file for JPEG available or the format is illegal.
Image.
41109 FFCustomMsgBox [41109] A menu of a A menu of a different modality was Select a menu of the same modality.
different modality has been selected.
selected.
41110 FFCustomMsgBox [41110] Failed in access to The system was unstable. Restart. If the same symptom recurs,
the image database. perform a reinstallation.
41111 FFCustomMsgBox [41111] Failed in access to It is conceivable that the shared host has Check targets involved in the shared host.
the image database. not been started.
41112 FFCustomMsgBox [41112] Failed in access to It is conceivable that QA processing was Check other equipment or perform another
the image database. performed on other equipment. study.
41113 FFCustomMsgBox [41113] Failed in access to It is conceivable that the shared host has Check targets involved in the shared host.
the image file. not been started.
41114 FFCustomMsgBox [41114] Failed in saving the It is conceivable that the shared host has Check targets involved in the shared host.
information. not been started.
41115 FFCustomMsgBox [41115] Failed in access to It is conceivable that the shared host has Check targets involved in the shared host.
the image file. not been started.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-87
Event Event Source Description Causes Remedies Remarks
41116 FFCustomMsgBox [41116] Failed in saving the It is conceivable that the shared host has Check targets involved in the shared host.
information. not been started.
41117 FFCustomMsgBox [41117] Failed in saving the It is conceivable that the shared host has Check targets involved in the shared host.
information. not been started.
41118 FFCustomMsgBox [41118] Failed in image It is conceivable that the shared host has Check targets involved in the shared host.
output processing. not been started.
41119 FFCustomMsgBox [41119] Failed in QA startup The system was unstable. Restart. If the same symptom recurs,
processing. perform a reinstallation.
41120 FFCustomMsgBox [41120] Failed in access to It is conceivable that the shared host has Check targets involved in the shared host.
the image file. not been started.
41130 FFCustomMsgBox [41130] Failed in replacing No extension menus have been registered. Check menu registration information.
an image.
41131 FFCustomMsgBox [41131] Study information It is conceivable that system settings for Check the settings.
cannot be modified. the study # and ID # have been changed.
41132 FFCustomMessageBox [41132] Failed in energy Failed in re-performing energy subtraction Change the exposure menu again.
subtraction processing. processing in an attempt to change an ES
menu already subjected to exposure.
Abnormal resultant images (bone and soft
tissue images) were generated.
41140 FFCustomMsgBox [41140] The function cannot A function that was called up did not match Check the customized settings.
be implemented. the optional system settings.
41141 FFCustomMsgBox [41141] An error occurred. The system was unstable. Reinstall the program if the same symptom
recurs after restarting the equipment.
41150 FFCustomMsgBox [41150] Failed in file output A failure occurred in accessing the Restart. If the same symptom recurs,
processing. transmission file. The system may have perform a reinstallation.
been unstable.
41151 FFCustomMsgBox [41151] There is no free The hard disk was full. Exit QA processing, delete unnecessary
space on the hard disk. images from the list, and retry output
processing.
41152 FFCustomMsgBox [41152] Free space is The hard disk was full. Exit QA processing, delete unnecessary
insufficient on the images from the list, and retry output
destination for output. processing.
41200 FFQAMsgBox HDD FULL. The hard disk was full. Increase the free hard disk space and then
[File Name] [Function Name] retry operation.
[Error Code]
41201 FFQAMsgBox The specified path does not The specified path was not found. Check path settings.
exist.
[File Name] [Function Name]
[Error Code]
41210 FFQAMsgBox Return value of a The method return value for the image Check the destination for image distribution
ImgDeliveryCom makes an distribution instruction component resulted and then retry.
error. in an error. (Processing other than image
[File Name] [Function Name] distribution can be continued.)
[Error Code]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-88
Event Event Source Description Causes Remedies Remarks
42200 FFQAMsgBox Can’t Get Bitmap Failed in acquiring a bitmap (replaceable). Reinstall the program.
(Processing continuation).
[File Name] [Function Name]
[Error Code]
42201 FFQAMsgBox CustomMessageCtrl The return value for the custom message Reinstall the program.
Access Failure(Processing box control resulted in an error.
continuation).
[File Name] [Function Name]
[Error Code]
42202 FFQAMsgBox SysConfig DB Access An error occurred when obtaining a system Reinstall the program.
Failure(Processing setup database access API value or the
continuation). obtained value was illegal (replaceable).
[File Name] [Function Name]
[Error Code]
42203 FFQAMsgBox LangStr DB Access An error occurred when obtaining a Reinstall the program.
Failure(Processing character string database access API
continuation). value (replaceable).
[File Name] [Function Name]
[Error Code]
42204 FFQAMsgBox LangMsgBox DB Access The return value for the message box Reinstall the program.
Failure(Processing database access API resulted in an error.
continuation).
[File Name] [Function Name]
[Error Code]
42205 FFQAMsgBox FF32 Access Failure The return value for the FF32 library (FF32. Reinstall the program.
(Probably acquisition of dll) resulted in an error.
Registry goes wrong.)
[File Name] [Function Name] * When acquiring an optional key registry,
[Error Code] no error will be registered even if no key
is available.
42206 FFQAMsgBox Tag Information Access No tag information was set up though it Use default information for processing.
Failure(Processing must have been set up beforehand There is no operational problem.
continuation). (replaceable).
[File Name] [Function Name]
[Error Code]
42207 FFQAMsgBox NetConfig DB Access An error occurred when obtaining a Reinstall the program.
Failure(Processing network setup database value or obtained
continuation). value was not correct (replaceable).
[File Name] [Function Name]
[Error Code]
42209 FFQAMsgBox FF32 Access Failure Failed in obtaining/setting an FF32 library Reinstall the program.
(Processing continuation). value (replaceable).
[File Name] [Function Name]
[Error Code]
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-89
Event Event Source Description Causes Remedies Remarks
42210 FFQAMsgBox Invalid function ID Number. An illegal function ID (disregard). There is no operational problem.
(Processing continuation).
[File Name] [Function Name]
[Error Code]
42211 FFQAMsgBox Data is not found in No custom setup database registered Register custom setup database using the
CustomScreenDB. (processing can be continued). customization setting displayed on the
[File Name] [Function Name] User Utility window.
[Error Code]
42212 FFQAMsgBox LogFile Access Failure. Failed in accessing the log file. Reinstall the program.
[File Name] [Function Name]
[Error Code]
42214 FFQAMsgBox The amount of adjustments The amount of position adjustment is set to Manually adjust the positioning amount
is 0. 0 (zero). from the trimming dialog.
[File Name] [Function Name]
[Error Code]
42299 FFQAMsgBox Data resource access error. A clustering resource access error Check network connection status.
[File Name] [Function Name] (processing can be continued).
[Error Code]
43010 FFQAManagerCom Can’t Get FontSize The font size was not registered in the Reinstall the IIP because the displayed
Information. registry. characters can be corrupted.
43010 FFIIPQACtrl Can’t Get FontSize The font size was not registered in the Reinstall the IIP because the displayed
Information. registry. characters can be corrupted.
43020 FFIIPQACtrl Can’t Get Caption The character string database did not exist Reinstall the IIP because the displayed
Information. or was damaged. characters can be corrupted.
43060 FFQAManagerCom [Control: %1, The character string database did not exist Reinstall the IIP because the displayed
Function: %2 , Code: %3] or was damaged. characters can be corrupted.
43061 FFQAManagerCom [Control: %1, The character string database did not exist Reinstall the IIP because the displayed
Function: %2 , Code: %3] or was damaged. characters can be corrupted.
43062 FFQAManagerCom [Control: %1, The character string database did not exist Reinstall the IIP because the displayed
Function: %2 , Code: %3] or was damaged. characters can be corrupted.
43063 FFQAManagerCom [Control: %1, The character string database did not exist Reinstall the IIP because the displayed
Function: %2 , Code: %3] or was damaged. characters can be corrupted.
43064 FFQAManagerCom [Control: %1, The character string database did not exist Reinstall the IIP because the displayed
Function: %2 , Code: %3] or was damaged. characters can be corrupted.
43110 FFIIPQACtrl Can’t Access Local The language setting could not be Reinstall the IIP because the displayed
Configuration Data. obtained. characters can be corrupted.
43201 FFIIPQACtrl Focus Set Failure. The focus cannot be determined properly. There is no particular problem with
operations. (The focus cannot be
determined properly because the
annotation character input dialog has been
inactivated.) After shutting the
magnification function, input of annotation
characters becomes possible.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-90
Event Event Source Description Causes Remedies Remarks
43210 FFIIPQACtrl frmChageMenu.frm, Energy subtraction processing was retried. Re-perform changing of the exposure
cmdOK_Click-RunEsProc menu.
Retry
43210 FFIIPQACtrl frmChageMenu.frm, Though energy subtraction processing was Re-perform changing of the exposure
cmdOK_Click-RunEsProc retried, an abnormal image was generated. menu.
Retry Error
43301 FFQAManagerCom [Control: %1, The screen size was not obtained. Reinstall the IIP because the displayed
Function: %2 , Code: %3] characters can be corrupted.
43310 FFQAManagerCom [Control: %1, The language setup data was not obtained. Reinstall the IIP because the displayed
Function: %2 , Code: %3] characters can be corrupted.
44200 FFQAMsgBox Journal Proc Failure. The method return value for the log Restart the system.
[File Name] [Function Name] storage control resulted in an error.
[Error Code]
44201 FFQAMsgBox An unexpected Method An unexpected method call (called up at Restart the system.
call. an impossible timing, being possible
[File Name] [Function Name] though to continue processing).
[Error Code]
44262 FFCustomMsgBox [44262] The maximum Automatic marker setting was attempted Delete unnecessary markers so that there
settable number of on an image where the number of are 4 or less, and reperform the setting.
automatic markers automatic markers allowed for one
exceeded. exposure menu was exceeded.
44263 FFCustomMsgBox [44263] Do you wish to The image was rotated/flipped with Press the [Yes] button to delete the
delete markers and execute markers determined on it. markers so that the image is rotated/flipped
the operation? appropriately. Press the [No] button to
rotate/flip the image without deleting the
markers.
44266 FFCustomMsgBox [44266] Markers are not The setting of automatic markers was Press the [Yes] button to delete the
being selected. attempted on an image with an exposure already-registered automatic markers.
menu where automatic markers are Press the [No] button not to make any
already-registered, though no markers changes in the setting of the automatic
have been determined on the same image. markers.
44267 FFCustomMsgBox [44267] Automatic markers The setting of automatic markers was Press the [Yes] button to determine image
will be registered. performed for an exposure menu for which makers for the exposure menu as the
no automatic markers have been automatic markers. Press the [No] button
determined. not to perform the setting of automatic
markers.
44268 FFCustomMsgBox [44268] Automatic markers The setting of automatic markers was Press the [Yes] button to delete already-
will be registered. performed for an exposure menu for which registered automatic markers and
automatic markers have already been determine image markers as the automatic
determined. markers. Press the [No] button not to make
any changes in the setting of the automatic
markers.
49000 FFIIPQACtrl frmChageMenu.frm, An error other than generation of an Re-perform changing of the exposure
cmdOK_Click-RunEsProc abnormal image. menu.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-91
Event Event Source Description Causes Remedies Remarks
49000 FFIIPQACtrl frmChageMenu.frm, An error other than generation of an Re-perform changing of the exposure
cmdOK_Click-RunEsProc abnormal image occurred at retrial of menu.
Retry Error energy subtraction processing.
49200 FFQAMsgBox FF32 Error. The return value for the FF32 library (FF32. Restart the system.
[File Name] [Function Name] dll) resulted in an error (irreplaceable).
[Error Code]
49201 FFQAMsgBox Method Argument Error. A method argument type error (for Restart the system.
[File Name] [Function Name] instance, there were no objects).
[Error Code]
49202 FFQAMsgBox Can’t get Tag Information. No tag information was set up though it Restart the system.
[File Name] [Function Name] must have been set up beforehand
[Error Code] (irreplaceable).
49203 FFQAMsgBox Memory or Resource Failed in acquiring memory/handle Restart the system.
cannot be gained. resources.
[File Name] [Function Name]
[Error Code]
49204 FFQAMsgBox Windows API error Windows API return value resulted in an Restart the system.
occurred. error.
[File Name] [Function Name]
[Error Code]
49298 FFQAMsgBox Secondary detection code An error occurred with the return value of Restart the system.
Error. the secondary detection code.
[File Name] [Function Name]
[Error Code]
49299 FFQAMsgBox Exception Error occurred. An exceptional error (an error detected Restart the system.
[File Name] [Function Name] with Try/Catch, OnErrorGoto).
[Error Code]
50000 FFFinpSend Memory Error.%n%1 The system was abnormal (memory Restart.
allocation was not accomplished).
50000 FFOutImgPrc Initialize Error.%n%1 An error occurred during initialization Restart. If the same symptom recurs
processing. frequently, reinstall the system and then
initialize the image database.
50001 FFCustomMsgBox [50001] Failed in • Output option setup was not completed. • Perform output option setup.
initialization process. • The system was unstable. • Restart. If the same symptom recurs,
perform a reinstallation.
50001 FFFinpSend Dicom Data Object Load The DicomDataObject file for output was Perform a reinstallation.
Error.%n%1 not found.
50001 FFIIPOUTPUT Initialize Error.%n[%1] Initialization process error. Restart. If the same symptom frequently
occurs, perform a reinstallation.
• Output control was not exercised.
• The error check method was not
successfully called.
• A certain other program error occurred.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-92
Event Event Source Description Causes Remedies Remarks
50001 FFOutImgPrc DB Connect Error.%n%1 An abnormality occurred in the connection Check whether the SQL service manager
to a database. is normally operating. If the same symptom
occurs while the SQL service manager is
normally operating, initialize the image
database (the entire image data is then
deleted).
50001 Spooler Initialize Error.%n[%1] Initialization process error. Restart. If the same symptom frequently Output section
occurs, perform a reinstallation. container.
• No output control was exercised.
• An error check method call was
unsuccessful.
• Other program error.
50002 FFCustomMsgBox [50002] Failed in output The system was unstable. Restart. If the same symptom recurs after
process. a restart, perform a reinstallation.
50002 FFFinpSend File Open Error.%n%1 There is no output raw data file. Reinstall.
50002 FFIIPOUTPUT State Change State transition notification process error. Restart.
Error.%n[%1]
• The output destination was not
successfully reregistered.
• The state transition notification call was
not successful.
• An unknown program error occurred.
50002 FFOutImgPrc DB Information Error.%n%1 The database information was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
50002 Spooler State Change Error.%n[%1] State change notification process error. Restart. Output section
container.
• An output destination was not
successfully registered.
• A state change notification call was not
successful.
• Other program error.
• Line establishment message transmission
error [not used].
• Line abnormality message transmission
error [not used].
50003 FFCustomMsgBox [50003] Failed in output The system was unstable. Restart. If the same symptom recurs after
process. a restart, perform a reinstallation.
50003 FFFinpSend File Read Error.%n%1 The output raw data file is abnormal. Reinstall.
50003 FFIIPOUTPUT Routine Change Routine change process error. Restart.
Error.%n[%1]
• The resumption request was not
successfully made.
50003 FFOutImgPrc System Error.%n%1 The system was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-93
Event Event Source Description Causes Remedies Remarks
50003 Spooler Routine Change Routine change process error. Restart. Output section
Error.%n[%1] container.
• A resumption request was not
successfully made.
50004 FFCustomMsgBox [50004] Failed in output The system was unstable. Restart. If the same symptom recurs,
process. perform a reinstallation.
50004 FFFinpSend FinpSend Error.%n%1 An abnormality occurred in the output Restart. If the same symptom frequently
sequence. occurs, perform a reinstallation and then
initialize the image database.
50004 FFIIPOUTPUT Spool Error.%n[%1] Spool process error. Restart.
• The spool request call was not
successfully made.
• The reoutput request was not
successfully made.
• The study termination notification was not
successfully made.
• The spool request response status was
invalid.
• An unknown program error occurred.
50004 FFOutImgPrc I/F Timing Error.%n%1 An interface abnormality was encountered. Restart.
50004 Spooler Spool Error.%n[%1] Spool process error. Restart. Output section
container.
• A spool request call was not successfully
made.
• A reoutput request was not successfully
made.
• An examination end notification was not
successful.
• The spool request response state was
illegal.
• Other program error.
50005 FFCustomMsgBox [50005] Failed in output The system was unstable. Restart. If the same symptom recurs after
process. a restart, perform a reinstallation.
50005 FFIIPOUTPUT ExeOutput Error.%n[%1] Forced-output process error. Restart.
• The spool forced output request call was
not successfully made.
• The queue forced output response
message was not successfully
transmitted.
• An unknown program error occurred.
50005 FFOutImgPrc CreateFileMapping File mapping was unachievable. Restart.
Error.%n%1
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-94
Event Event Source Description Causes Remedies Remarks
50005 Spooler ExeOutput Error.%n[%1] Forced output process error. Restart. Output section
container.
• A spool forced output request was not
successfully called.
• A queue forced output response message
was not successfully transmitted.
• Other program error.
50006 FFCustomMsgBox [50006] Failed in output The system was unstable. Restart. If the same symptom recurs after
process. a restart, perform a reinstallation.
50006 FFIIPOUTPUT Image Proc Error.%n[%1] Image processing error. Restart.
• The image processing request call was
not successfully made.
• The image processing setup completion
call was not successfully made.
• The image processing completion
notification call was not successfully
made.
• The image processing job acquisition
request was not successfully made.
• An unknown program error occurred.
50006 FFOutImgPrc File Open Error.%n%1 The file could not be opened. Check the file.
50007 FFCustomMsgBox [50007] Failed in The system was unstable. Restart. If the same symptom recurs after
terminating process. a restart, perform a reinstallation.
50007 FFIIPOUTPUT Output Proc Error.%n[%1] Output processing error. Restart.
• The output destination information was
not successfully acquired.
• The output request call was not
successfully made.
• The output completion notification was
not successfully made.
• An unknown program error occurred.
50007 FFOutImgPrc File Read/Write A file write/read was unsuccessful. If the problem frequently occurs, replace
Error.%n%1 The problem is attributable to a disk. the disk or take some other remedial
action.
50007 Spooler Terminate Error.%n[%1] Termination process error. Restart. Output section
container.
• Other program error.
50008 FFCustomMsgBox [50008] Failed in image The system was unstable. Restart. If the same symptom recurs after
processing for output. a restart, perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-95
Event Event Source Description Causes Remedies Remarks
50008 FFIIPOUTPUT Commit Proc Error.%n[%1] Commitment processing error. Restart.
• The commitment destination information
was not successfully acquired.
• The commitment request call was not
successfully made.
• The commitment completion notification
was not successfully made.
• An unknown program error occurred.
50008 FFOutImgPrc CreateDispatch The control was not successfully initialized. Check that the control is registered.
Error.%n%1
50008 Spooler Image Proc Error.%n[%1] Image process error. Restart. Output section
container.
• An image process request was not
successfully called.
• An image process setup completion call
was not successfully made.
• An image process completion notification
was not successfully called.
• An image process job acquisition request
was not successfully issued.
• Other program error.
50009 FFCustomMsgBox [50009] Failed in output The system was unstable. Restart. If the same symptom recurs after
process. a restart, perform a reinstallation.
50009 FFIIPOUTPUT Queue Resume Queue resumption process error. Restart.
Error.%n[%1]
• The queue resumption request call was
not successfully made.
• The queue resumption response
message was not successfully
transmitted.
• An unknown program error occurred.
50009 FFOutImgPrc Memory Get Error.%n%1 The system was abnormal (memory Restart.
allocation was not achieved).
50009 Spooler Output Proc Error.%n[%1] Output process error. Restart. Output section
container.
• Output destination information was not
successfully acquired.
• An output request was not successfully
called.
• An output completion notification was not
successful.
• Other program error.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-96
Event Event Source Description Causes Remedies Remarks
50010 FFIIPOUTPUT Queue Delete Queue deletion process error. Restart.
Error.%n[%1]
• The queue deletion request was not
successfully made.
• The queue deletion response message
was not successfully transmitted.
• An unknown program error occurred.
50010 FFOutImgPrc Memory(DDO) Get The system was abnormal (memory Restart.
Error.%n%1 allocation was not achieved).
50010 Spooler Queue Resume Queue resumption process error. Restart. Output section
Error.%n[%1] container.
• A queue resumption request was not
successfully called.
• A queue resumption response message
was not successfully transmitted.
• Other program error.
50011 FFIIPOUTPUT Unit Suspend Equipment output suspension process Restart.
Error.%n[%1] error.
• The equipment output suspension
request call was not successfully made.
• The equipment output suspension
response message was not successfully
transmitted.
• An unknown program error occurred.
50011 FFOutImgPrc Memory(VR) Get The system was abnormal (memory Restart.
Error.%n%1 allocation was not achieved).
50011 Spooler Queue Delete Queue deletion process error. Restart. Output section
Error.%n[%1] container.
• A queue deletion request was not
successfully issued.
• A queue deletion response message was
not successfully transmitted.
• Other program error.
50012 FFIIPOUTPUT Unit Resume Error.%n[%1] Equipment output resumption error. Restart.
• The equipment output resumption
request call was not successfully made.
• The equipment output resumption
response message was not successfully
transmitted.
• An unknown program error occurred.
50012 FFOutImgPrc Dicom Data Object Dicom Data Object was abnormal. Restart. If the same symptom frequently
Error.%n%1 occurs, perform a reinstallation and then
initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-97
Event Event Source Description Causes Remedies Remarks
50012 Spooler Unit Suspend Error.%n[%1] Unit output suspension process error. Restart. Output section
container.
• A unit output suspension request was not
successfully called.
• A unit output suspension response
message was not successfully
transmitted.
• Other program error.
50013 FFOutImgPrc Dicom Data Object No Dicom Data Object did not contain any Restart. If the same symptom frequently
Data.%n%1 data. occurs, perform a reinstallation and then
initialize the image database.
50013 Spooler Unit Resume Error.%n[%1] Unit output resumption process error. Restart. Output section
container.
• A unit output resumption request was not
successfully called.
• A unit output resumption response
message was not successfully
transmitted.
• Other program error.
50015 FFIIPOUTPUT Terminate Error.%n[%1] Termination process error. Restart.
• An unknown program error occurred.
50020 FFOutImgPrc Data Error.%n%1 A data abnormality was encountered. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
50021 FFOutImgPrc Output Job DB Error.%n%1 The output database was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
50022 FFOutImgPrc Memory Over.%n%1 A memory area abnormality was Restart.
encountered.
50030 FFOutImgPrc OutImgPrc Error.%n%1 An abnormality occurred. Restart.
50100 FFOutImgPrc IFormatCom Error.%n%1 The format calculation object was Restart. If the same symptom frequently
abnormal. occurs, perform a reinstallation and then
initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-98
Event Event Source Description Causes Remedies Remarks
50100 Spooler Spooler Error.%n%1 Common error message. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• Queue information class related initialize the image database.
message.
An output job file was not successfully
accessed.
Output job information was not
successfully accessed.
Film information was not successfully
acquired. Information was not
successfully acquired.
The job status was not successfully
acquired (program error).
Job information was not successfully
freed.
• Unit information class related message.
Unit information was not successfully
freed.
• Spooler class related message.
The film count was not successfully
checked.
The image count/image information area
was not successfully acquired.
Unit information was not successfully
acquired.
• Other program error
50101 FFOutImgPrc IRawMakeCom The raw data creation object was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
50101 Spooler RunProc Error.%n%1 Initialization process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• A spool start process was in error.
• A database connection was not
successfully closed.
• Other program error.
50102 FFOutImgPrc FFNetDB Error.%n%1 An abnormality occurred at the time of Restart. If the same symptom frequently
network setup data acquisition. occurs, perform a reinstallation.
(It is conceivable that the network setup
database was abnormal.)
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-99
Event Event Source Description Causes Remedies Remarks
50102 Spooler OutputRegisterProc Output destination registration process Restart. If the same symptom frequently Spooler
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An illegal parameter was encountered.
• An output destination registration request
message was not successfully issued.
• An output destination was not
successfully registered.
• Other program error.
50103 FFOutImgPrc IImgJpegCtrl Error.%n%1 The decompression object was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
50103 Spooler StateChangeProc Local printer/Dicom printer state change Restart. If the same symptom frequently Spooler
Error.%n%1 notification process error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An illegal parameter was encountered.
• A printer state change notification
message was not successfully issued.
• The unit status was not successfully
acquired.
• Other program error.
50104 FFOutImgPrc IFMImgSkGpCtrl The normalization process object was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
50104 Spooler ErrorCheckProc Data integrity check process error. Restart. If the same symptom frequently Spooler
Error.%n%1 occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• A data integrity check request message
was not successfully issued.
• Other program error.
50105 FFOutImgPrc IFMImgVRSCtrl The zoom object (VRS method) was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
50105 Spooler RestoreJob Error.%n%1 Job restoration error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An unprocessed job restoration request
message was not successfully issued.
• Other program error.
50106 FFOutImgPrc IRotateCom Error.%n%1 The rotation object was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-100
Event Event Source Description Causes Remedies Remarks
50106 Spooler SpoolProc Error.%n%1 Spool process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• A parameter error occurred.
• A spool request message was not
successfully issued.
• DDO file information was not successfully
acquired.
• Other program error.
50107 FFOutImgPrc IFMImgParamCtrl The image process management Restart. If the same symptom frequently
Error.%n%1 parameter object was abnormal. occurs, perform a reinstallation and then
initialize the image database.
50107 Spooler ReoutputProc Error.%n%1 Reoutput start process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An illegal parameter was encountered.
• A reoutput start message was not
successfully issued.
• Other program error.
50108 FFOutImgPrc IFMImgHandCtrl The CR image process object was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
50108 Spooler StudyReqEndProc Study termination request processing error. Restart. If the same symptom frequently Spooler
Error.%n%1 occurs, perform a reinstallation and then
• Spooler control status error. initialize the image database.
• Illegal parameters.
• Failed in issuing study termination
request message.
• Other program errors.
50109 FFOutImgPrc IFMImgFCmpCtrl FCR compression (compression/ Restart. If the same symptom frequently
Error.%n%1 decompression (Fuji’s unique method)) occurs, perform a reinstallation and then
was abnormal. initialize the image database.
50109 Spooler ExeOutputJob Error.%n%1 Forced output request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An illegal parameter was encountered.
• A forced output request message was not
successfully issued.
• The requested output image data did not
exist.
• Other program error.
50110 FFOutImgPrc IDataToPrintCom Data conversion (for DICOM) was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-101
Event Event Source Description Causes Remedies Remarks
50110 Spooler GetImageProcJob Image processing job acquisition process Restart. If the same symptom frequently Spooler
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An image processing job acquisition
request message was not successfully
issued.
• Data was not databased.
• Other program error.
50111 FFOutImgPrc IDataToStorageCom Data conversion (for DICOM Storage) was Restart. If the same symptom frequently
Error.%n%1 abnormal. occurs, perform a reinstallation and then
initialize the image database.
50111 Spooler SetImageProcComplete Image processing completion notification Restart. If the same symptom frequently Spooler
Error.%n%1 process error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An illegal parameter was encountered.
• An image processing completion
notification message was not successfully
issued.
• Other program error.
50112 FFOutImgPrc IFMImgMoireCtrl Moire elimination was abnormal. Restart. If the same symptom frequently
Error.%n%1 occurs, perform a reinstallation and then
initialize the image database.
50112 FFQCMsgBox Custom DB Access Error. Access error to ReaderQC.mdb. Restart.
[File Name]
[Function Name]
[50112]
50112 Spooler GetOutputProcJob Output job acquisition response process Restart. If the same symptom frequently Spooler
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An output job acquisition response
message was not successfully issued.
• Data was not databased.
• Other program error.
50113 FFOutImgPrc IFilmCharMakeCom The film annotation character generation Restart. If the same symptom frequently
Error.%n%1 object was abnormal. occurs, perform a reinstallation and then
initialize the image database.
50113 Spooler SetOutputProcComplete Output completion notification process Restart. If the same symptom frequently Spooler
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An illegal parameter was encountered.
• An output completion notification
message was not successfully issued.
• Data was not databased.
• Other program error.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-102
Event Event Source Description Causes Remedies Remarks
50114 FFOutImgPrc MaskProcExec Error.%n%1 The blackening process was abnormal. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
initialize the image database.
50114 Spooler GetCommitProcJob Commitment job acquisition response Restart. If the same symptom frequently
Error.%n%1 process error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control status was abnormal.
• The commitment jog acquisition response
message was not successfully issued.
• The data was not found in the database.
• An unknown program error occurred.
50115 FFOutImgPrc RecognizeProcExec The blackening automatic recognition Restart. If the same symptom frequently
Error.%n%1 process was abnormal. occurs, perform a reinstallation and then
initialize the image database.
50115 Spooler QueueResume Queue output resumption request process Restart. If the same symptom frequently Spooler
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control state was abnormal.
• An illegal parameter was encountered.
• A queue output resumption request
message was not successfully issued.
• Other program error.
50115 Spooler SetCommitProcComplete Commitment completion notification Restart. If the same symptom frequently
Error.%n%1 process error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control status was abnormal.
• An illegal parameter was encountered.
• The commitment completion notification
message was not successfully issued.
• The queue was not found.
• The data was not found in the database.
• An unknown program error occurred.
50116 FFOutImgPrc ExecMarker Error.%n%1 The marker imbedding process was Restart. If the same symptom frequently
abnormal. occurs, perform a reinstallation and then
initialize the image database.
50116 Spooler QueueDelete Error.%n%1 Queue deletion request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An illegal parameter was encountered.
• A queue deletion request message was
not successfully issued.
• Output job file information was not
successfully accessed.
• Queue deletion was not successful.
• Database access was not successful.
• Other program error.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-103
Event Event Source Description Causes Remedies Remarks
50117 FFOutImgPrc LUTProcExec Error.%n% An error occurred in a display gradation Restart. If the same symptom frequently
correction process. occurs, perform a reinstallation and then
initialize the image database.
50117 Spooler QueueResume Queue output resumption request process Restart. If the same symptom frequently
Error.%n%1 error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control status was abnormal.
• An illegal parameter was encountered.
• The queue output resumption request
message was not successfully issued.
• An unknown program error occurred.
50117 Spooler UnitSuspend Error.%n%1 Unit suspension request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An illegal parameter was encountered.
• A unit suspension request message was
not successfully issued.
• Other program error.
50118 FFOutImgPrc MPMCode Error.%n%1 An error occurred in an output information Restart. If the same symptom frequently
conversion process. occurs, perform a reinstallation and then
initialize the image database.
50118 Spooler QueueDelete Error.%n%1 Queue deletion request process error. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
• The spooler control status was abnormal. initialize the image database.
• An illegal parameter was encountered.
• Failed in issuing a queue deletion request
message.
• Failed in accessing the output job file
information.
• Failed in deleting the queue.
• Failed in accessing the database.
• An unknown program error occurred.
50118 Spooler UnitResume Error.%n%1 Unit resumption request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An illegal parameter was encountered.
• A unit resumption request message was
not successfully issued.
• Other program error.
50119 Spooler SuspendProc Error.%n%1 Suspension request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• A suspension request message was not
successfully issued.
• Other program error.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-104
Event Event Source Description Causes Remedies Remarks
50119 Spooler UnitSuspend Error.%n%1 Equipment suspension request process Restart. If the same symptom frequently
error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control status was abnormal.
• An illegal parameter was encountered.
• The equipment suspension request
message was not successfully issued.
• An unknown program error occurred.
50119 FFOutlmgPrc Trimming Error.%n%1 An error occurred while in trimming Restart. If the same symptom frequently
processing. occurs, perform a reinstallation and then
initialize the image database.
50120 Spooler ResumeProc Error.%n%1 Resumption request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• A resumption request message was not
successfully issued.
• Other program error.
50120 Spooler UnitResume Error.%n%1 Equipment resumption request process Restart. If the same symptom frequently
error. occurs, perform a reinstallation and then
initialize the image database.
• The spooler control status was abnormal.
• An illegal parameter was encountered.
• The equipment resumption request
message was not successfully issued.
• An unknown program error occurred.
50120 FFOutlmgPrc AnnotationProc An error occurred while in annotation Restart. If the same symptom frequently
Error.%n%1 information embedding processing. occurs, perform a reinstallation and then
initialize the image database.
50121 Spooler EndProc Error.%n%1 End request process error. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• The spooler control state was abnormal. initialize the image database.
• An end request message was not
successfully issued.
• Other program error.
50121 Spooler SuspendProc Error.%n%1 Suspension request process error. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
• The spooler control status was abnormal. initialize the image database.
• The suspension request message was
not successfully issued.
• An unknown program error occurred.
50121 FFOutlmgPrc FNCProc Error.%n%1 An error occurred during FNC processing. Restart. If the same symptom recurs,
perform reinstallation and initialize the
image database.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-105
Event Event Source Description Causes Remedies Remarks
50122 Spooler ResumeProc Error.%n%1 Resumption request process error. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
• The spooler control status was abnormal. initialize the image database.
• The resumption request message was
not successfully issued.
• An unknown program error occurred.
50123 Spooler EndProc Error.%n%1 Termination request process error. Restart. If the same symptom frequently
occurs, perform a reinstallation and then
• The spooler control status was abnormal. initialize the image database.
• The termination request message was
not successfully issued.
• An unknown program error occurred.
50400 FFQCAutoCalcCom File %1, Spot %2, Fatal Error (exceptional error)
Code %3.
50500 FFQCAutoCalcCom File %1, Spot %2, Fatal Error (exceptional error)
Code %3.
50700 FFCustomMsgBox [50700] An unknown error The system was unstable. Restart. If the same symptom recurs,
occurred. perform a reinstallation.
50700 FFIIPDicomQR Unknown Error. %n[Module] Unknown error. Restart the system.
%1
%n[Function] %2
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
50701 FFCustomMsgBox [50701] Failed in The system was unstable. Restart. If the same symptom recurs,
initialization processing. perform a reinstallation.
50702 FFIIPDicomQR Logical Program Error. Not This is an anticipated error, but Restart the system.
Supported Case. unsupported.
%n[Module] %1 Hard to continue the operation as it is a
%n[Function] %2 critical error.
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
50703 FFIIPDicomQR Initialization Error. Initialization error. Restart the system.
%n[Module]
%1 %n[Function] %2
%n[SpotCode] %3
%n[Ret]%4
%n[Description] %5
50704 FFIIPDicomQR UnInitialization Error. Uninitialization error. Restart the system.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-106
Event Event Source Description Causes Remedies Remarks
50705 FFIIPDicomQR System Configuration System configuration information Shut down the system and make sure that
Information Acquisition acquisition error. the system configuration is appropriate.
Error.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50706 FFIIPDicomQR Network Configuration Network configuration information Shut down the system and make sure that
Information Acquisition acquisition error. the network configuration is appropriate.
Error.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50707 FFIIPDicomQR Resource (Message/String/ Resource (message, character string, Restart the system.
Bitmap etc.) Acquisition BMP) acquisition error.
Error.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50708 FFIIPDicomQR Registry Information Registry information acquisition error. Restart the system.
Acquisition
Error. %n[Module] %1
%n[Function] %2
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
50709 FFIIPDicomQR InterProcess Message Interprocess message communication Restart the system.
Communication Error. error.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50710 FFIIPDicomQR Database Access Error. Database access error. Restart the system.
%n[Module] %1
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-107
Event Event Source Description Causes Remedies Remarks
50711 FFIIPDicomQR DDO Access Error. DDO file access error. Restart the system.
%n[Module]
%1 %n[Function] %2
%n[SpotCode] %3
%n[Ret]%4
%n[Description] %5
50721 FFIIPDicomQR Queried Information Query result display error. Restart the system.
Display Hard to continue the operation as it is a
Error. %n[Module] %1 critical error.
%n[Function] %2
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
50741 FFIIPDicomQR Retrieved Image Display Retrieved image display error. Restart the system.
Error. Hard to continue the operation as it is a
%n[Module] %1 critical error.
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50742 FFIIPDicomQR Retrieved Image Output Image distribution error. Restart the system.
Error. Hard to continue the operation as it is a
%n[Module] %1 critical error.
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
50800 FFSFPD Logical Program Error. %n A program error occurred. Processing Restart.
%1 %n %2 cannot be continued because it is a serious
error.
50801 FFSFPD Catch Exception Error. %n A program error occurred. Processing Restart.
%1 %n %2 cannot be continued because it is an
exceptionally occurred serious error.
50802 FFSFPD Process Pipe Timeout. %n A timeout error occurred in inter-process Restart.
%1 %n %2 communication.
50803 FFSFPD Property Error. %n An illegal property. Restart.
%1 %n %2
50804 FFSFPD Arg Error. %n An illegal argument. Check the input data. Proceed with
%1 %n %2 processing.
50805 FFSFPD Data Error. %n A packet data error. Restart.
%1 %n %2
50806 FFSFPD WinAPI Error. %n A WinAPI error. Restart.
%1 %n %2
50809 FFCustomMsgBox [50809] An error occurred. The system was unstable. Restart. If the same symptom recurs,
perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-108
Event Event Source Description Causes Remedies Remarks
50900 FFSFPD Program Error. %n A program error. Processing cannot be Restart the system.
[Module]%1 %n continued because it is a serious error.
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50901 FFSFPD Resource (Message/String/ A resource (message/character string/ Check the database or BMP file.
Bitmap etc.) Acquisition BMP) acquisition error.
Error.
%n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50902 FFSFPD Registry Information A registry setting acquisition error. Check the registry key.
Acquisition
Error. %n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50903 FFSFPD System Configuration A system setting acquisition error. Check the NetConfig.mdb file.
Information Acquisition
Error.
%n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50904 FFSFPD XML Data Acquisition Error. An XML data acquisition error. Check the xml format.
%n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50905 FFSFPD Initialization Error. %n An initialization processing error. Restart the system.
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-109
Event Event Source Description Causes Remedies Remarks
50906 FFSFPD UnInitialization Error. %n A system shutdown error. Restart the system.
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50907 FFSFPD Shimazu FPD control A Shimadzu FPD control error. Restart the system.
processing
error. %n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50908 FFSFPD Calibration Execution Error. A calibration execution error. Check the error.
%n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50909 FFSFPD Database Access Error. %n A database access error. Check the actual calibration execution
[Module]%1 %n database.
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50910 FFSFPD DB Result Set Error. %n A database result setting error. Check the actual calibration execution
[Module]%1 %n database.
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50911 FFSFPD NetConfig Acquisition Error. A network setting acquisition error. Check the NetConfig.mdb file.
%n
[Module]%1 %n
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
50951 FFStdFPDCtrl %1!s! An error occurred at an external device. Restart.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-110
Event Event Source Description Causes Remedies Remarks
51001 FFIIPOUTPUT Initialize Warning.%n[%1] Initialization process warning. Check the network settings. If any setting
is incorrect, perform a reinstallation.
• The output equipment execution was not
successfully started.
• The output equipment name (AE name/
host name) was not successfully acquired
(there was an inconsistency between the
option and network settings).
• The FFNET database was not
successfully accessed.
51001 Spooler Initialize Warning.%n[%1] Initialization process warning. Restart. If the same symptom frequently Output section
occurs, perform a reinstallation. container.
• An output device execution was not
successfully started.
• An output device name (AE name/host
name) was not successfully acquired.
• FFNETDB access was unsuccessful.
51002 FFIIPOUTPUT State Change State transition notification process Restart.
Warning.%n[%1] warning.
• The response to an acquisition request
was not successfully made.
51002 Spooler State Change State change notification process warning. Restart. Output section
Warning.%n[%1] container.
• An acquisition request was not
successfully answered.
51007 Spooler Terminate Warning.%n[%1] Termination process warning. Restart. Output section
container.
• A process status flag was not successfully
set.
• A process termination response message
was not successfully transmitted.
• An abnormality occurrence notification
message was not successfully
transmitted.
• A message box was not successfully
displayed.
• A spooler termination request was not
successfully called.
• A spooler control end request was not
successfully called.
• The spooler control end response state
was abnormal.
• An image process control end request
was not successfully issued.
• An output device control end request was
not successfully issued.
• The output device control end response
state was abnormal.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-111
Event Event Source Description Causes Remedies Remarks
51010 FFCustomMsgBox [51010] Specified device The setup data was incorrect or the Reset the data or check the equipment.
does not exist. equipment not started.
51011 FFCustomMsgBox [51011] Failed in reading of The setup file format was incorrect or the Check settings.
the configuration file file could not be read.
(TagLookup.ini).
51012 FFCustomMsgBox [51012] Failed in reading of The setup file format was incorrect or the Check settings.
the gray level correction file could not be read.
file.
51013 FFCustomMsgBox [51013] Failed in output The system was unstable. Restart. If the same symptom recurs,
process. perform a reinstallation.
51015 FFIIPOUTPUT Terminate Warning.%n[%1] Termination process warning. Restart.
• The process status flag was not
successfully set.
• The process termination response
message was not successfully
transmitted.
• The abnormality occurrence notification
message was not successfully
transmitted.
• The message box was not successfully
opened.
• The spooler termination request call was
not successfully made.
• The spooler control termination request
call was not successfully made.
• The spooler control termination response
status was abnormal.
• The image processing control termination
request was not successfully made.
• The output equipment control termination
request was not successfully made.
• The output equipment control termination
response status was abnormal.
51301 FFCustomMsgBox [51301] The hard disk is The remaining free space is insufficient. Increase the free space and then perform
full. the output process again.
51302 FFCustomMsgBox [51302] Failed in the image The image output process could not be Start up the host at the output destination
output process. performed because the host at the output and then perform the output process again.
destination was not running.
51312 FFCustomMsgBox [51312] Failed in the image The selected image was not found. Select an image that can be output.
output process.
51320 FFCustomMsgBox [51320] Failed in accessing The image file was not found. Make sure that the images are entered.
the image file.
51323 FFCustomMsgBox [51323] An error occurred. Restart. If the same symptom recurs,
perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-112
Event Event Source Description Causes Remedies Remarks
51701 FFIIPDicomQR Failed in Processing. Exceptional error. Continue the operation. If the same error DICOM QR
%n[Module] The operation can be continued as it is a recurs continuously, restart the system.
%1 %n[Function] %2 minor error.
%n[SpotCode] %3
%n[Ret]%4
%n[Description] %5
51702 FFCustomMsgBox [51702] Failed in displaying The search result includes illegal data. Stop acquiring studies that include illegal
the study information list. information.
51710 FFCustomMsgBox [51710] Input information Input information includes illegal data. Input correct data.
includes illegal characters.
51711 FFCustomMsgBox [51711] Input information Input information includes illegal data. Input correct data.
exceeds the limit number of
characters.
51712 FFCustomMsgBox [51712] The hit rate of The number of hit studies is many. Narrow down the search conditions
searching for studies accordingly.
exceeded the limit value.
Search processing is
suspended.
51713 FFCustomMsgBox [51713] Failed in searching The system was unstable. Restart. If the same symptom recurs,
study information. perform a reinstallation.
51714 FFCustomMsgBox [51714] A network error The network settings were inappropriate. Check settings of network and other
occurred while in search of connected equipment.
study information.
51715 FFCustomMsgBox [51715] A timeout error The network settings were inappropriate. Check settings of network and other
occurred while in search of connected equipment.
study information. No
responses come from
equipment targeted for
search.
51717 FFCustomMsgBox [51717] Information of The network settings were inappropriate. Check settings of network and other
equipment targeted for connected equipment.
search cannot be found.
Network settings may not
be correct.
51718 FFCustomMsgBox [51718] A communication The network settings were inappropriate. Check settings of network and other
error occurred while in connected equipment.
search of study information.
51720 FFIIPDicomQR Query Processing Error. Query processing error. Continue the operation. If the same error
%n[Module] %1 It is a minor error. The operation can be recurs continuously, restart the system.
%n[Function] continued after the user issues a
%2 %n[SpotCode] %3 necessary notification.
%n[Ret]%4 %n[Description]
%5
51721 FFCustomMsgBox [51721] Failed in Not possible to obtain the selected study. Stop obtaining a study that may contain DICOM QR
processing to acquire invalid information.
studies.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-113
Event Event Source Description Causes Remedies Remarks
51721 FFIIPDicomQR Queried Information Query result display error. Continue the operation. If the same error
Display It is a minor error. The operation can be recurs continuously, restart the system.
Error. %n[Module] %1 continued after the user issues a
%n[Function] %2 necessary notification.
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
51722 FFCustomMsgBox [51722] Failed in Free space on the hard disk is insufficient. Increase free space on the hard disk
processing to acquire appropriately.
studies. Free hard disk
space is not sufficient.
51722 FFIIPDicomQR Failed in Query Processing. Query processing error. Continue the operation. If the same error DICOM QR
%n[Module] %1 It is a minor error. The operation can be recurs continuously, restart the system.
%n[Function] continued after the user issues a
%2 %n[SpotCode] %3 necessary notification.
%n[Ret]%4 %n[Description]
%5
51723 FFCustomMsgBox [51723] Image that is not An acquired study includes illegal data. Stop acquiring studies that include illegal
subject to being acquired or information.
displayed may be included.
* Image of 13.6” in size,
which was exposed by
FDR D-EVO may be
contained.
51725 FFCustomMsgBox [51725] Information of Network settings were inappropriate. Check settings of network and other
equipment targeted for connected equipment.
image acquisition cannot
be found. Network settings
may not be correct.
51726 FFCustomMsgBox [51726] A communication Network settings were inappropriate. Check settings of network and other
error occurred in an attempt connected equipment.
to acquire studies.
51727 FFCustomMsgBox [51727] A timeout error Network settings were inappropriate. Check settings of network and other
occurred while in process connected equipment.
of study acquisition. No
responses come from the
equipment targeted for
study acquisition.
51729 FFCustomMsgBox [51729] Input information is Input information includes illegal data. Input correct information.
not correct.
51730 FFCustomMsgBox [51730] An attempt is made Acquired study information includes illegal Stop acquiring studies that include illegal
to acquire a study data. information.
generated by an unknown
modality.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-114
Event Event Source Description Causes Remedies Remarks
51731 FFCustomMsgBox [51731] A study generated Acquired information includes illegal data. Stop acquiring studies that include illegal
by an unknown modality information.
was acquired.
51732 FFCustomMsgBox [51732] The hit rate of The number of hit studies is many. Narrow down the search conditions
searching for studies accordingly.
exceeded the listed limit
value.
51740 FFIIPDicomQR Retrieve Processing Error. Retrieve processing error. Continue the operation. If the same error
%n[Module] %1 It is a minor error. The operation can be recurs continuously, restart the system.
%n[Function] continued after the user issues a
%2 %n[SpotCode] %3 necessary notification.
%n[Ret]%4 %n[Description]
%5
51741 FFIIPDicomQR Retrieved Image Display Retrieved image display error. Continue the operation. If the same error
Error. It is a minor error. The operation can be recurs continuously, restart the system.
%n[Module] %1 continued after the user issues a
%n[Function] necessary notification.
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
51742 FFIIPDicomQR Failed in Retrieve Retrieve processing error. Continue the operation. If the same error DICOM QR
Processing. It is a minor error. The operation can be recurs continuously, restart the system.
%n[Module] %1 continued after the user issues a
%n[Function] necessary notification.
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
51742 FFIIPDicomQR Retrieved Image Output Image distribution error. Continue the operation. If the same error
Error. It is a minor error. The operation can be recurs continuously, restart the system.
%n[Module] %1 continued after the user issues a
%n[Function] necessary notification.
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
51805 FFCustomMsgBox [51805] Failed in image Acquired studies include illegal data. Stop acquiring studies that include illegal
display processing. Some information.
images cannot be
displayed.
51806 FFCustomMsgBox [51806] Failed in image Acquired studies include illegal data. Stop acquiring studies that include illegal
distribution processing. information.
51810 FFCustomMsgBox [51810] Information on the Network settings were inappropriate. Check settings of network and other
device targeted for image connected equipment.
printout was not found.
Network settings may
possibly be incorrect.
51950 FFStdFPDCtrl %1!s! A warning was issued from an external Check the device status.
device.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-115
Event Event Source Description Causes Remedies Remarks
52000 FFFinpSend Warning.%n%1 Information for development personnel. There is no problem with operations.
52000 FFOutImgPrc Warning.%n%1 Information for development personnel. There is no problem with operations.
52002 FFEifPrinter SMCU Driver error. An error occurred in the SMCU driver. Restart.
[%1] The problem might have been caused, for
instance, by data incompleteness or
system instability.
52003 FFEifPrinter EIFIMG Driver error. An error occurred in the EIFIMG driver. It is Restart.
[%1] conceivable that the system was unstable.
52004 FFEifPrinter Received unexpected An unknown command was received from Check the log of communications with the
command. the printer. connected printer or restart the printer/IIP.
[%1]
52006 FFEifPrinter File Error. The raw data file for output was not found. Since the problem is attributable to a
[%1] program error, try performing a reinstall.
52100 FFQCMsgBox Custom DB Access Error. The bitmap was not successfully acquired. Check that the bitmap file is installed.
[File Name]
[Function Name] [50112]
52100 Spooler Spooler Warning.%n%1 Common warning message. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• Queue information class related initialize the image database.
message. Image information
(SOPInstanceUID) was not successfully
acquired. The job status was not
successfully acquired.
• Spooler class related message.
A database connection was not
successfully closed.
• Illegal image display format data.
52101 Spooler RunProc Warning.%n%1 Initialization process warning. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• A character string database connection initialize the image database.
was not successfully opened.
52102 FFQCMsgBox SysConfig DB Access An error/invalid value was encountered Restart.
Failure (Processing when an attempt was made to acquire a
continuation). system setup database access API value.
[File Name]
[Function Name] [52102]
52103 FFQCMsgBox LangStr DB Access Failure An error occurred when an attempt was Restart.
(Processing continuation). made to acquire a character string
[File Name] database access API value.
[Function Name] [52103]
52103 Spooler StateChangeProc Local printer/Dicom printer state change Restart. If the same symptom frequently Spooler
Warning.%n%1 notification process warning. occurs, perform a reinstallation and then
initialize the image database.
• A unit state character string was not
successfully acquired.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-116
Event Event Source Description Causes Remedies Remarks
52105 FFQCMsgBox FF32 Access Failure The FF32 library (FF32dll) return value It is probable that the installation was not
(Probably acquisition of was in error. properly completed. Perform a
Registry goes wrong) reinstallation.
[File Name] * For option key registry acquisition, no
[Function Name] [52105] error occurs even if the key is missing.
52105 Spooler RestoreJob Warning.%n%1 Job restoration warning. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• An output job file was not successfully initialize the image database.
accessed.
52106 Spooler SpoolProc Warning.%n%1 Spool process warning. Restart. If the same symptom frequently Spooler
occurs, perform a reinstallation and then
• Unit information acquisition was not initialize the image database.
successful.
• File deletion was not successful.
• Queue cancellation was not successful.
• Data was not successfully written into a
log file.
52108 FFQCMsgBox Custom DB Access Failure An attempt to set/acquire a value from Restart.
(Processing continuation) ReaderQC.mdb failed.
[File Name]
[Function Name] [52108]
52108 Spooler StudyReqEndProc Examination end request process warning. Restart. If the same symptom frequently Spooler
Warning.%n%1 occurs, perform a reinstallation and then
• A database was not successfully initialize the image database.
accessed.
• DDO file information was not successfully
acquired.
52109 FFQCMsgBox FF32 DB Access Failure Registry setup/acquisition failed It is probable that the registry information
(Processing continuation) (replaceable). was not properly installed. Perform a
[File Name] reinstallation.
[Function Name] [52109]
52109 Spooler ExeOutputJob Forced output request process warning. Restart. If the same symptom frequently Spooler
Warning.%n%1 occurs, perform a reinstallation and then
• DDO file information was not successfully initialize the image database.
acquired.
52110 FFQCMsgBox FF32 DB Access Failure An image file access error occurred. Since the image is valid, make an
(Processing continuation) exposure again.
[File Name]
[Function Name] [52110]
52110 Spooler GetImageProcJob Image processing job acquisition process Restart. If the same symptom frequently Spooler
Warning.%n%1 warning. occurs, perform a reinstallation and then
initialize the image database.
• An output unit type parameter error
occurred.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-117
Event Event Source Description Causes Remedies Remarks
52112 Spooler GetOutputProcJob Output job acquisition response process Restart. If the same symptom frequently Spooler
Warning.%n%1 warning. occurs, perform a reinstallation and then
initialize the image database.
• An output unit type parameter error was
encountered.
52114 Spooler GetCommitProcJob Commitment job acquisition response Restart. If the same symptom frequently
Warning.%n%1 process warning. occurs, perform a reinstallation and then
initialize the image database.
• A commitment equipment type parameter
error occurred.
52116 Spooler QueueDelete Queue deletion request process warning. Restart. If the same symptom frequently Spooler
Warning.%n%1 occurs, perform a reinstallation and then
• An image database flag was not initialize the image database.
successfully set.
• The DDO file was not successfully
deleted.
52117 Spooler UnitSuspend Unit suspension request process warning. Restart. If the same symptom frequently Spooler
Warning.%n%1 occurs, perform a reinstallation and then
• A unit status character string was not initialize the image database.
successfully acquired.
52118 Spooler QueueDelete Queue deletion request process warning. Restart. If the same symptom frequently
Warning.%n%1 occurs, perform a reinstallation and then
• The image database flag was not initialize the image database.
successfully set.
• The DDO file was not successfully
deleted.
52118 Spooler UnitResume Unit resumption request process warning. Restart. If the same symptom frequently Spooler
Warning.%n%1 occurs, perform a reinstallation and then
• A unit status acquisition error occurred. initialize the image database.
52119 Spooler UnitSuspend Equipment suspension request process Restart. If the same symptom frequently
Warning.%n%1 warning. occurs, perform a reinstallation and then
initialize the image database.
• The equipment status character string
was not successfully acquired.
52120 Spooler UnitResume Equipment resumption request process Restart. If the same symptom frequently
Warning.%n%1 warning. occurs, perform a reinstallation and then
initialize the image database.
• An equipment status acquisition error
occurred.
52400 FFQCAAutocalccom File %1, Spot %2, Code Warning information. Perform reexposure and then
%3. measurement processing because the
image is illegal.
52500 FFQCAAutocalccom File %1, Spot %2, Code Fatal error (exceptional error).
%3.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-118
Event Event Source Description Causes Remedies Remarks
52720 FFIIPDicomQR Query Processing Error. Query processing error. Continue the operation. If the same error
%n[Module] %1 The operation can be continued as it is a recurs continuously, restart the system.
%n[Function] minor error.
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
52721 FFIIPDicomQR Queried Information Query results display error. Continue the operation. If the same error
Display The operation can be continued as it is a recurs continuously, restart the system.
Error. %n[Module] %1 minor error.
%n[Function] %2
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
52740 FFIIPDicomQR Retrieve Processing Error. Retrieve processing error. Continue the operation. If the same error
%n[Module] %1 The operation can be continued as it is a recurs continuously, restart the system.
%n[Function] minor error.
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
52741 FFIIPDicomQR Retrieved Image Display Retrieved image display error. Continue the operation. If the same error
Error. The operation can be continued as it is a recurs continuously, restart the system.
%n[Module] %1 minor error.
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
52742 FFIIPDicomQR Retrieved Image Output Image distribution error. Continue the operation. If the same error
Error. The operation can be continued as it is a recurs continuously, restart the system.
%n[Module] %1 minor error.
%n[Function]
%2 %n[SpotCode] %3
%n[Ret]%4 %n[Description]
%5
52750 FFIIPDicomQR Logical Program Error. Timeout error. Shut down the system and make sure that
%nNetwork the network configuration is appropriate.
Warning(Timeout). Check also to see that the connected
%n%1 %n %2 equipment operates correctly.
52751 FFIIPDicomQR Network Warning(Over The limit number for querying is exceeded. Suspend the query. Narrow down the
Number query condition and reperform it.
Of Result). %n%1 %n %2
52760 FFTsSearchCondition Disk Full Error. %n The hard disk is full. Restart the system.
52761 FFTsSearchCondition DB Operation Status A warning given when the database Restart the system.
Warning. %n operation result status is illegal.
52770 FFMaSearchCOM Disk Full Error. %n The hard disk is full. Restart the system.
52771 FFMaSearchCOM Network Error. %n A network error occurred. Restart the system.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-119
Event Event Source Description Causes Remedies Remarks
52772 FFMaSearchCOM Network Error. %n The number for hit results exceeded the Restart the system.
upper limit.
52780 FFMaCommon Disk Full Error. %n The HDD disk is full. Restart the system. DICOM QR
52581 FFMaCommon Debug Log OutPut. %n Event log output number for debugging. Restart the system. DICOM QR
52790 FFMaCommonImage Disk Full Error. %n The HDD disk is full. Restart the system. DICOM QR
52800 FFIIPDicomQR Network Warning(Timeout). Timeout error. Shut down the system and make sure that
%n%1 %n %2 the network configuration is appropriate.
Check also to see that the connected
equipment operates correctly.
52801 FFIIPDicomQR Network Warning(DiskFull). The HDD disk is full. Stop retrieving image information. Increase
%n%1 %n %2 free space available on the hard disk and
retry retrieval.
52802 FFIIPDicomQR Warning(Retrieve Image No images are contained in the retrieved End the study without checking images.
Nothing). %n%1 %n %2 study. Make sure that the connected equipment
retains correct study information.
52810 FFTsImgRetInd Disk Full Error. %n The hard disk is full. Restart the system.
52820 FFMaRetrieve Disk Full Error. %n The hard disk is full. Restart the system.
52821 FFMaRetrieve Storage SCP not found. %n No Storage SCP is set up. Restart the system.
52822 FFMaRetrieve Cannot Get NetWork Necessary network settings cannot be Restart the system.
Setting. %n acquired.
52823 FFMaRetrieve Integration Warning in There is a problem with consistency of the Restart the system.
DataBase. %n database.
52824 FFMaRetrieve DICOM Data Waring. %n A DICOM data-related warning. Restart the system.
52825 FFMaRetrieve Local Retrieve Warning. A warning given when images cannot be Restart the system.
%n acquired locally.
52826 FFMaRetrieve Input Image Warning A warning given when an SH image is Restart the system.
(DensityCode = SH). %n input.
52827 FFMaRetrieve File Error. %n Warning issued when an error occurs Continue the operation. If the same error DICOM QR
during file operation. recurs continuously, restart the system.
52828 FFMaRetrieve File Format Error. %n Warning issued when there is an error with There was an invalid format row in the DICOM QR
the file format. obtained ConfigQRFilter.ini filter file. Such
row was processed as an invalid condition.
Make sure that the content of the filter file
is correct.
52830 FFMaRetrieve Out Of Memory Error. %n Warning issued when a failure occurred in Continue the operation. If the same error DICOM QR
an attempt to secure memory. recurs continuously, restart the system.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-120
Event Event Source Description Causes Remedies Remarks
52831 FFMaRetrieve Input Image Infomation Warning issued when an image targeted Suspend the attempt to retrieve image DICOM QR
(Image data not match for filtering was input. information because an image that does
QRFilter). %n not match the filtering condition described
in the obtained ConfigQRFilter.ini filter file
was input. If the intended filtering condition
is not met, make sure that the content of
the filter file is correct.
52832 FFMaRetrieve Catch Exception Error.%n Warning issued when an unexpected error Continue the operation. If the same error DICOM QR
occurred exceptionally. recurs continuously, restart the system.
52840 FFTsCommon Disk Full Error. %n The HDD disk is full. Restart the system. DICOM QR
52841 FFTsCommon Catch DB exception. %n Exceptional database access error. Restart the system. DICOM QR
52850 FFMaCommon Disk Full Error. %n The HDD disk is full. Restart the system.
52851 FFMaCommon Debug Log OutPut. %n Event log output number for debugging Restart the system.
purposes.
52860 FFMaCommonImage Disk Full Error. %n The HDD disk is full Restart the system.
52870 FFTsCommon Disk Full Error. %n The HDD disk is full. Restart the system.
52871 FFTsCommon Catch DB exception. %n An exceptional access to the database. Restart the system.
53106 Spooler SpoolProc Information in a spool process. Restart. If the same symptom frequently Spooler
Infomation.%n%1 occurs, perform a reinstallation and then
• Unit type mismatch notification. initialize the image database.
53820 FFMaRetrieve Retrieve Process Start. %n Process startup. Restart the system.
53821 FFMaRetrieve Retrieve Process End. %n Process shutdown. Restart the system.
53823 FFMaRetrieve Image is Already Exists in Information given when the same image is Restart the system.
ImageDB. %n input.
52840 FFTsCommon Retry is success. %n Database retry information. Restart the system. DICOM QR
53870 FFTsCommon Retry is success. %n Database retry information. Restart the system.
54100 FFQCMsgBox Journal Proc Failure. Writing error. Check to see the media has not been
[File Name] prohibited for writing.
[Function Name] [54100]
54800 FFSFPD Network Error. %n An ArcNet network error. Check the ArcNet line. Proceed with
%1 %n %2 processing.
54801 FFSFPD Process Start. %n Started process (COM). Proceed with processing.
%1 %n %2
54802 FFSFPD Process End. %n Ended process (COM). Proceed with processing.
%1 %n %2
54803 FFSFPD Property Error. %n Invalid actual result information. Check the equipment.
%1 %n
%2
54804 FFSFPD Property Error. %n Parameter specified outside the fixed Check the exposure parameter.
%1 %n range.
%2
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-121
Event Event Source Description Causes Remedies Remarks
54807 FFSFPD Property Error. %n A filter information definition file data error. Check the filter information definition file.
%1 %n
%2
54808 FFSFPD Property Error. %n An X-ray tube code file data error. Check the X-ray tube code file
%1 %n
%2
54900 FFSFPD Impossible Error. %n A calibration execution disable error. If the system is in emergency mode, wait
[Module]%1 %n until it goes into routine mode.
[Function]%2 %n
[SpotCode]%3 %n
[Ret]%4 %n
[Description]%5
59100 FFQCMsgBox FF32 Error. An FF32 library (FF32.dll) return value It is possible that the registry information
[File Name] error (irreplaceable). has not been installed correctly. Perform a
[Function Name] [59100] reinstallation.
59102 FFQCMshBox Cant get Tag Information. No value is set up for the tag information Perform reexposure because the image is
[File Name] even though it must have been set up illegal.
[Function Name] [59102] accordingly (irreplaceable).
59103 FFQCMsgBox Memory or Resource Failed in obtaining memory/handle Increase available resources and then
cannot be gained. resource. retry.
[File Name]
[Function Name] [59103]
59198 FFQCMsgBox Secondary detection code A secondary detection code return value
Error. error.
[File Name]
[Function Name] [59198]
59701 FFIIPDicomQR Logical Program Error. Exceptional error. Continue the operation. If the same error
Exception The operation can be continued as it is a recurs continuously, restart the system.
has Occurred. %n[Module] minor error.
%1
%n[Function] %2
%n[SpotCode]
%3 %n[Ret]%4
%n[Description]
%5
59702 FFIIPDicomQR Logical Program Error. Not This is an anticipated error, but Continue the operation. If the same error
Supported Case. unsupported. recurs continuously, restart the system.
%n[Module] The operation can be continued as it is a
%1 %n[Function] %2 minor error.
%n[SpotCode] %3
%n[Ret]%4
%n[Description] %5
59751 FFIIPDicomQR Catch Exception Error. An unexpected error occurred Restart the system.
%n%1 exceptionally.
%n %2
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-122
Event Event Source Description Causes Remedies Remarks
59761 FFTsSearchCondition Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59762 FFTsSearchCondition Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59763 FFTsSearchCondition Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
59771 FFMaSearchCOM Catch Exception Error. %n An unexceptional error occurred Restart the system.
unexpectedly.
59772 FFMaSearchCOM Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59773 FFMaSearchCOM Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
59774 FFMaSearchCOM Data Base Error. %n A database error occurred. Restart the system if the relevant
processing has been completed.
If the relevant processing remains
unprocessed, there may be an illegal study
in the server targeted for search of data. In
this case, no particular remedial measures
need to be taken.
59781 FFMaCommon Catch Exception Error. %n An unexpected error occurred Restart the system. DICOM QR
exceptionally.
59782 FFMaCommon Out Of Memory Error. %n An error occurred in an attempt to secure Restart the system. DICOM QR
memory.
59783 FFMaCommon Free Memory Error. %n An error occurred in an attempt to open Restart the system. DICOM QR
memory.
59791 FFMaCommonImage Catch Exception Error. %n An unexpected error occurred Restart the system. DICOM QR
exceptionally.
59792 FFMaCommonImage Out Of Memory Error. %n An error occurred in an attempt to secure Restart the system. DICOM QR
memory.
59793 FFMaCommonImage Free Memory Error. %n An error occurred in an attempt to open Restart the system. DICOM QR
memory.
59801 FFIIPDicomQR Catch Exception Error. An unexpected error occurred Restart the system.
%n%1 exceptionally.
%n %2
59802 FFIIPDicomQR Invalid Propaty Error. Invalid property error (only for the Restart the system.
%n%1 %n AddIndicationData method).
%2
59811 FFTsImgRetInd Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59812 FFTsImgRetInd Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59813 FFTsImgRetInd Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-123
Event Event Source Description Causes Remedies Remarks
59821 FFMaRetrieve Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59822 FFMaRetrieve Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59823 FFMaRetrieve Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
59824 FFMaRetrieve Process Communication An error occurred in the communication Restart the system.
Error. %n control.
59825 FFMaRetrieve Retrieve Process Exists. A process was started double. Restart the system.
%n
59826 FFMaRetrieve Image is Already Exists in Image input is disabled because the same Restart the system.
ImageDB. %n image already exists in the image
database.
59827 FFMaRetrieve File Error. %n A file error occurred. Restart the system.
59828 FFMaRetrieve NetWork Error. %n The association was disconnected due to Restart the system.
a network error.
59829 FFMaRetrieve Cannot Get NetWork Necessary network settings cannot be Restart the system.
Setting. %n acquired.
59829 FFMaRetrieve Logical Program Error. %n Warning issued when a program error Continue the operation. If the same error DICOM QR
occurs. recurs continuously, restart the system.
59830 FFMaRetrieve DICOM Retrieve Timeout. A timeout error occurred in an attempt to Restart the system.
%n acquire a study.
59831 FFMaRetrieve DICOM Retrieve is Not DICOM acquisition processing is not Restart the system.
Able to Use Now. %n available at present.
59832 FFMaRetrieve Processing Image is Processing is disabled because the image Restart the system.
Deleted. %n was deleted.
59841 FFTsCommon Catch Exception Error. %n An unexpected error occurred Restart the system. DICOM QR
exceptionally.
59842 FFTsCommon Out Of Memory Error. %n An error occurred in an attempt to secure Restart the system. DICOM QR
memory.
59843 FFTsCommon Free Memory Error. %n An error occurred in an attempt to open Restart the system. DICOM QR
memory.
59851 FFMaCommon Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59852 FFMaCommon Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59853 FFMaCommon Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
59861 FFMaCommonImage Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59862 FFMaCommonImage Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-124
Event Event Source Description Causes Remedies Remarks
59863 FFMaCommonImage Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
59871 FFTsCommon Catch Exception Error. %n An exceptional error occurred Restart the system.
unexpectedly.
59872 FFTsCommon Out Of Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be secured appropriately.
59873 FFTsCommon Free Memory Error. %n An error occurred, which disabled the Restart the system.
memory to be released appropriately.
60000 FFCustomMsgBox [60000] Failed in accessing It is conceivable that the database file is Restart. If the same symptom recurs,
the database. damaged or deleted. perform a reinstallation.
60002 FFCustomMsgBox [60002] User Utility will be There is no alternative but to exit the Restart. If the same symptom recurs,
terminated. program due to the occurrence of an perform a reinstallation.
abnormality.
60050 FFCustomMsgBox [60050] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the character string damaged or deleted. perform a reinstallation.
database.
60051 FFCustomMsgBox [60051] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the system setting damaged or deleted. perform a reinstallation.
database.
60052 FFCustomMsgBox [60052] Failed in obtaining It is conceivable that the registry Restart. If the same symptom recurs,
the registry information. information was damaged or deleted. perform a reinstallation.
60053 FFCustomMsgBox [60053] Value of registry It is conceivable that the registry Restart. If the same symptom recurs,
information is abnormal. information was damaged or deleted. perform a reinstallation.
60054 FFCustomMsgBox [60054] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
character string database.
60055 FFCustomMsgBox [60055] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
system setting database.
60056 FFCustomMsgBox [60056] Failed in obtaining The setup file format was incorrect or the Check settings.
the information from the file could not be read.
system setting file.
60057 FFCustomMsgBox [60057] Failed in setting The setup file format was incorrect or the Check settings.
information to the system file could not be read.
setting file.
60080 FFCustomMsgBox [60080] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the display group damaged or deleted. perform a reinstallation.
information.
60081 FFCustomMsgBox [60081] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the menu information. damaged or deleted. perform a reinstallation.
60100 FFCustomMsgBox [60100] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the MPM code. damaged or deleted. perform a reinstallation.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-125
Event Event Source Description Causes Remedies Remarks
60183 FFCustomMsgBox [60183] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the technician information damaged or deleted. perform a reinstallation.
database.
60184 FFCustomMsgBox [60184] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the technician information damaged or deleted. perform a reinstallation.
database.
60185 FFCustomMsgBox [60185] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
system setting database.
60203 FFCustomMsgBox [60203] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the requesting department damaged or deleted. perform a reinstallation.
information database.
60204 FFCustomMsgBox [60204] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the requesting department damaged or deleted. perform a reinstallation.
information database.
60205 FFCustomMsgBox [60205] Failed in obtaining It is conceivable that the database file was Restart. If the same symptom recurs,
the information from the damaged or deleted. perform a reinstallation.
system setting database.
60215 FFCustomMsgBox [60215] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the film annotation damaged or deleted. perform a reinstallation.
database.
60216 FFCustomMsgBox [60216] Failed in accessing It is conceivable that the database file was Restart. If the same symptom recurs,
the film annotation damaged or deleted. perform a reinstallation.
database.
60251 FFCustomMsgBox [60251] Image processing The image processing parameter data is Check the image processing parameters.
parameter data is abnormal.
abnormal.
60260 FFCustomMsgBox [60260] System has not It is conceivable that the IIP executable file Reinstall.
been rebooted. was damaged or deleted.
60300 FFCustomMsgBox [60300] An option and a Necessary options are not installed or Reinstall.
database aren’t inappropriate menu database is installed.
conformable.
60301 FFCustomMsgBox [60301] Data storage The database file is possibly damaged or Restart. If the same symptom recurs,
resulted in a failure. deleted. perform a reinstallation.
60302 FFCustomMsgBox [60302] Data deletion The database file is possibly damaged or Restart. If the same symptom recurs,
resulted in a failure. deleted. perform a reinstallation.
60303 FFCustomMsgBox [60303] User Utility cannot Necessary options are not installed or Reinstall the system.
be started because the inappropriate menu database is installed.
necessary option has not
been installed.
60304 FFCustomMsgBox [60304] System error An error occurred on the FPD side. Check information of the connected
occurred. devices.
60305 FFCustomMsgBox [60305] System error An error occurred on the FPD side. Check information of the connected
occurred. devices.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-126
Event Event Source Description Causes Remedies Remarks
60306 FFCustomMsgBox [60306] A communication An error occurred in communication with Check information of the connected
error occurred with the the X-ray generator. devices.
X-ray generator.
61060 FFCustomMsgBox [61060] Wrong password An incorrect password was entered. The password is fixed at “****
has been input.
61082 FFCustomMsgBox [61082] No menu has been An attempt was made to start an editing or Complete a menu selection before initiating
selected. like operation without selecting any menu. an editing or like operation.
61083 FFCustomMsgBox [61083] Already registered The menu could not be deleted because it Clear the registration for display and then
for displaying and cannot was registered for display. delete the menu.
be deleted.
61084 FFCustomMsgBox [61084] Already registered The menu could not be deleted because it Remove the target menu from the
to the study menu and was registered for the examination menu. examination menu in advance.
cannot be deleted.
61085 FFCustomMsgBox [61085] No menu was A certain display group was empty because No empty display group is allowed to exist.
registered for a certain no menu was registered for it. Register one or more menus for the display
display group. group or delete the display group.
61090 FFCustomMsgBox [61090] Invalid character The entered data was abnormal due, for Make a proper entry.
has been input in the instance, to the inclusion of an
display group name. unacceptable character.
61093 FFCustomMsgBox [61093] Display group will This window opens for confirmation only.
be deleted.
61094 FFCustomMsgBox [61094] Display group The entered data was abnormal due, for Make a proper entry.
name is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61103 FFCustomMsgBox [61103] Same extension The entered data value was already Enter another value.
menu code already exists. registered or otherwise unacceptable.
61106 FFCustomMsgBox [61106] Same ANK The entered data value was already Enter another value.
exposure menu name registered or otherwise unacceptable.
already exists.
61109 FFCustomMsgBox [61109] Same kanji- The entered data value was already Enter another value.
exposure menu name registered or otherwise unacceptable.
already exists.
61110 FFCustomMsgBox [61110] Extension menu The entered data was abnormal due, for Make a proper entry.
code is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61111 FFCustomMsgBox [61111] ANK exposure The entered data was abnormal due, for Make a proper entry.
menu name is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61112 FFCustomMsgBox [61112] Kanji-exposure The entered data was abnormal due, for Make a proper entry.
menu name is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61122 FFCustomMsgBox [61122] MPM code is The entered data was abnormal due, for Make a proper entry.
illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-127
Event Event Source Description Causes Remedies Remarks
61125 FFCustomMsgBox [61125] Same film The entered data value was already Enter another value.
annotation already exists. registered or otherwise unacceptable.
61130 FFCustomMsgBox [61130] Device code is The entered data was abnormal due, for Make a proper entry.
illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61131 FFCustomMsgBox [61131] Film-related code is The entered data was abnormal due, for Make a proper entry.
illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61132 FFCustomMsgBox [61132] Auto-exposure The entered data was abnormal due, for Make a proper entry.
mechanism is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61142 FFCustomMsgBox [61142] Same study menu The entered data value was already Enter another value.
code already exists. registered or otherwise unacceptable.
61145 FFCustomMsgBox [61145] Same ANK study The entered data value was already Enter another value.
menu name already exists. registered or otherwise unacceptable.
61148 FFCustomMsgBox [61148] Same kanji-study The entered data value was already Enter another value.
menu name already exists. registered or otherwise unacceptable.
61149 FFCustomMsgBox [61149] No exposure menu An attempt was made to start an editing or Complete a menu selection before initiating
has been selected. like operation without selecting any menu. an editing or like operation.
61151 FFCustomMsgBox [61151] Exposure menu The maximum permissible number of Delete any unnecessary menu.
cannot be registered. menu registrations was exceeded.
61152 FFCustomMsgBox [61152] Same exposure The entered data value was already Enter another value.
menu has already been registered or otherwise unacceptable.
registered.
61153 FFCustomMsgBox [61153] Study menu code The entered data was abnormal due, for Make a proper entry.
is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61154 FFCustomMsgBox [61154] ANK study menu The entered data was abnormal due, for Make a proper entry.
code is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61155 FFCustomMsgBox [61155] Kanji-study menu The entered data was abnormal due, for Make a proper entry.
code is illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
61156 FFCustomMsgBox [61156] The energy An attempt was made to register an energy Energy subtraction menus are not allowed
subtraction menu cannot subtraction menu in a study menu. to be included in a study menu.
be registered to the study
menu.
61171 FFCustomMsgBox [61171] No technician code An attempt was made to register data with Enter a value.
has been input. an input field left blank.
61172 FFCustomMsgBox [61172] Technician code is The entered data was abnormal due, for Make a proper entry.
illegal. instance, to its excessive length or the
inclusion of an unacceptable character.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-128
Event Event Source Description Causes Remedies Remarks
61173 FFCustomMsgBox [61173] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the
technician code.
61174 FFCustomMsgBox [61174] Same technician The entered data value was already Enter another value.
code already exists. registered or otherwise unacceptable.
61175 FFCustomMsgBox [61175] No ANK technician An attempt was made to register data with Enter a value.
name has been input. an input field left blank.
61176 FFCustomMsgBox [61176] ANK technician The entered data was too long. Make a proper entry.
name is too long.
61177 FFCustomMsgBox [61177] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the ANK
technician name.
61178 FFCustomMsgBox [61178] Same ANK The entered data value was already Enter another value.
technician name already registered or otherwise unacceptable.
exists.
61179 FFCustomMsgBox [61179] No kanji-technician An attempt was made to register data with Enter a value.
name has been input. an input field left blank.
61180 FFCustomMsgBox [61180] Kanji-technician The entered data was too long. Make a proper entry.
name is too long.
61181 FFCustomMsgBox [61181] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the kanji-
technician name.
61182 FFCustomMsgBox [61182] Same kanji- The entered data value was already Enter another value.
technician name already registered or otherwise unacceptable.
exists.
61191 FFCustomMsgBox [61191] No requesting An attempt was made to register data with Enter a value.
department code has been an input field left blank.
input.
61192 FFCustomMsgBox [61192] Requesting The entered data was too long. Make a proper entry.
department code is too
long.
61193 FFCustomMsgBox [61193] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the
requesting department
code.
61194 FFCustomMsgBox [61194] Same requesting The entered data value was already Enter another value.
department code already registered or otherwise unacceptable.
exists.
61195 FFCustomMsgBox [61195] No ANK requesting An attempt was made to register data with Enter a value.
department name has been an input field left blank.
input.
61196 FFCustomMsgBox [61196] ANK requesting The entered data was too long. Make a proper entry.
department name is too
long.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-129
Event Event Source Description Causes Remedies Remarks
61197 FFCustomMsgBox [61197] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the ANK
requesting department
name.
61198 FFCustomMsgBox [61198] Same ANK The entered data value was already Enter another value.
requesting department registered or otherwise unacceptable.
name already exists.
61199 FFCustomMsgBox [61199] No kanji-requesting An attempt was made to register data with Enter a value.
department name has been an input field left blank.
input.
61200 FFCustomMsgBox [61200] Kanji-requesting The entered data was too long. Make a proper entry.
department name is too
long.
61201 FFCustomMsgBox [61201] Invalid character The entered data value was already Make a proper entry.
has been input in the Kanji- registered or otherwise unacceptable.
requesting department
name.
61202 FFCustomMsgBox [61202] Same kanji- The entered data value was already Enter another value.
requesting department registered or otherwise unacceptable.
name already exists.
61211 FFCustomMsgBox [61211] No film annotation An attempt was made to register data with Enter a value.
characters have been input. an input field left blank.
61212 FFCustomMsgBox [61212] Film annotation The entered data was too long. Make a proper entry.
characters are too long.
61213 FFCustomMsgBox [61213] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in the film
annotation characters.
61214 FFCustomMsgBox [61214] Same film The entered data value was already Enter another value.
annotation already exists. registered or otherwise unacceptable.
61224 FFCustomMsgBox [61224] Drive is not ready. No FD was inserted into the FDD. Insert an FD.
61234 FFCustomMsgBox [61234] Drive is not ready. No FD was inserted into the FDD. Insert an FD.
61235 FFCustomMsgBox [61235] Wrong disk A wrong FD was inserted into the FDD. Insert a correct FD.
61236 FFCustomMsgBox [61236] Wrong disk A wrong FD was inserted into the FDD. Insert a correct FD.
61280 FFCustomMsgBox [61280] No Coding Scheme No data is input. Make a proper entry.
Designator has been input.
61281 FFCustomMsgBox [61281] Coding Scheme The entered data was too long. Make a proper entry.
Designator is too long.
61282 FFCustomMsgBox [61282] Invalid character An unacceptable character was contained. Make a proper entry.
has been input in Coding
Scheme Designator.
61303 FFCustomMsgBox [61303] Failed in importing The problem is attributable, for instance, to Prepare a consumables database again.
the consumables database. a damaged file.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-130
Event Event Source Description Causes Remedies Remarks
61340 FFCustomMsgBox [61340] A file backed up An attempt was made to restore a backup To restore to the IIP, use a backup file
with another application made by the Viewer for use with the IIP or generated by the IIP. To restore to the
cannot be restored. to restore a backup made by the IIP for Viewer, use a backup file generated by the
use with the Viewer. Viewer.
61341 FFCustomMsgBox [61341] This is not a A file other than Supply.mdb was selected Select a Supply.mdb file.
contrast medium database for an import.
file.
61342 FFCustomMsgBox [61342] No backup files Specified drive or folder can be incorrect. Specify a correct destination for data
exist. backup.
61345 FFCustomMsgBox [61345] No media The DICOMDIR file does not exist. Ensure that the image file is correct.
management files exist.
64595 FFCustomMsgBox [64595] Failed in activation Access to the linking application Service Restart the system. If the same symptom
of Ascent View setting Utility failed or an unexpected error recurs, perform a reinstallation.
screen. occurred.
70000 FFCustomMsgBox [70000] Failed in It is possible that the RetakeAnalyisData. Open the RetakeAnalysisData.mdb file
conversion of mis-exposure mdb file is write-protected. properties to check to see if it is write-
image log information. protected.
70001 FFCustomMsgBox [70001] Failed in deletion of It is possible that the RetakeAnalyisData. Open the RetakeAnalysisData.mdb file
log files. mdb file is write-protected. properties to check to see if it is write-
protected.
70002 FFCustomMsgBox [70002] Failed in accessing It is possible that the file is damaged or Restart the system. If the same symptom
the definition file. deleted. recurs, perform a reinstallation.
70005 FFCustomMsgBox [70005] Connection to the It is possible that the database file is Restart the system. If the same symptom
system setting database damaged or deleted. recurs, perform a reinstallation.
resulted unsuccessful.
70006 FFCustomMsgBox [70006] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
information from the system damaged or deleted. recurs, perform a reinstallation.
setting database.
70007 FFCustomMsgBox [70007] Connection to the It is possible that the database file is Restart the system. If the same symptom
character string database damaged or deleted. recurs, perform a reinstallation.
resulted unsuccessful.
70008 FFCustomMsgBox [70008] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
information from the damaged or deleted. recurs, perform a reinstallation.
character string database.
70009 FFCustomMsgBox [70009] Failed in acquiring It is possible that the registry information is Restart the system. If the same symptom
registry information. damaged or deleted. recurs, perform a reinstallation.
70010 FFCustomMsgBox [70010] Failed in writing It is possible that the registry information is Restart the system. If the same symptom
registry information. damaged or deleted. recurs, perform a reinstallation.
70011 FFCustomMsgBox [70011] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
technologist information. damaged or deleted. recurs, perform a reinstallation.
70012 FFCustomMsgBox [70012] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
classifications of mis- damaged or deleted. recurs, perform a reinstallation.
exposure images.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-131
Event Event Source Description Causes Remedies Remarks
70013 FFCustomMsgBox [70013] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
comments on mis-exposure damaged or deleted. recurs, perform a reinstallation.
images.
70014 FFCustomMsgBox [70014] Connection to the It is possible that the database file is Restart the system. If the same symptom
log information database damaged or deleted. recurs, perform a reinstallation.
resulted unsuccessful.
70015 FFCustomMsgBox [70015] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
information for setup of damaged or deleted. recurs, perform a reinstallation.
mis-exposure image
analysis.
70016 FFCustomMsgBox [70016] Failed in reading It is possible that the image file is damaged Although the image file in question is not
image files. or deleted. displayed, there are no problems with
other operations.
70017 FFCustomMsgBox [70017] Failed in displaying An error occurred while in image Restart the system. If the same symptom
images. processing. recurs, perform a reinstallation.
70019 FFCustomMsgBox [70019] Failed in The date display format may have been Input in a correct format.
conversion of date format. input incorrectly.
70020 FFCustomMsgBox [70020] Failed in storage of This error is displayed when a failure Retry data store. If the same symptom
the mis-exposure image log occurred in data storage. recurs, change the destination for storage
list. accordingly.
70021 FFCustomMsgBox [70021] Failed in storage of This error is displayed when a failure Retry data storage. If the same symptom
the calculation result. occurred in data storage. recurs, change the destination for storage
accordingly.
70022 FFCustomMsgBox [70022] An error occurred It is possible that the image display control Restart the system. If the same symptom
while an image is being or image file has a problem. recurs, perform a reinstallation.
displayed.
70023 FFCustomMsgBox [70023] An unknown error The system was unstable. Restart the system. If the same symptom
occurred. recurs, perform a reinstallation.
70024 FFCustomMsgBox [70024] Connection to the It is possible that the database file is Restart the system. If the same symptom
database of regions to be damaged or deleted. recurs, perform a reinstallation.
exposed resulted
unsuccessful.
70025 FFCustomMsgBox [70025] Connection to the It is possible that the database file is Restart the system. If the same symptom
database of causes of damaged or deleted. recurs, perform a reinstallation.
mis-exposure images
resulted unsuccessful.
70026 FFCustomMsgBox [70026] Failed in calculation The system was unstable. Restart the system. If the same symptom
of the mis-exposure image recurs, perform a reinstallation.
log.
70027 FFCustomMsgBox [70027] Failed in displaying The system was unstable. Restart the system. If the same symptom
calculation result of the recurs, perform a reinstallation.
mis-exposure image log.
70028 FFCustomMsgBox [70028] Connection to the It is possible that the database file is Restart the system. If the same symptom
network setting database damaged or deleted. recurs, perform a reinstallation.
resulted unsuccessful.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-132
Event Event Source Description Causes Remedies Remarks
70029 FFCustomMsgBox [70029] Failed in acquiring It is possible that the database file is Restart the system. If the same symptom
necessary data from the damaged or deleted. recurs, perform a reinstallation.
network setting database.
70030 FFCustomMsgBox [70030] Failed in displaying The system was unstable. Restart the system. If the same symptom
the list. recurs, perform a reinstallation.
70032 FFCustomMsgBox [70032] Failed in deleting It is possible that the file is damaged or Restart the system. If the same symptom
the mis-exposure image deleted. recurs, perform a reinstallation.
log.
70033 FFCustomMsgBox [70033] Failed in deleting It is possible that the file is damaged or Restart the system. If the same symptom
the exposure result log. deleted. recurs, perform a reinstallation.
70034 FFCustomMsgBox [70034] Failed in deleting It is possible that the file is damaged or Restart the system. If the same symptom
an image file. deleted. recurs, perform a reinstallation.
70035 FFCustomMsgBox [70035] Failed in writing It is possible that the database file is Restart the system. If the same symptom
image data in the system damaged or deleted. recurs, perform a reinstallation.
setting database.
70036 FFCustomMsgBox [70036] Failed in writing It is possible that the database file is Restart the system. If the same symptom
image data in the display damaged or deleted. recurs, perform a reinstallation.
setting database.
70037 FFCustomMsgBox [70037] Failed in relocating The system was unstable. Restart the system. If the same symptom
the tree structure. recurs, perform a reinstallation.
70038 FFCustomMsgBox [70038] Failed in reducing The system was unstable. Restart the system. If the same symptom
the tree structure. recurs, perform a reinstallation.
70039 FFCustomMsgBox [70039] Transition to the The system was unstable. Restart the system. If the same symptom
main window resulted recurs, perform a reinstallation.
unsuccessful.
70040 FFCustomMsgBox [70040] Failed in accessing It is possible that the database file is Restart the system. If the same symptom
the menu database. damaged or deleted. recurs, perform a reinstallation.
70041 FFCustomMsgBox [70041] Failed in accessing It is possible that the database file is Restart the system. If the same symptom
the log information damaged or deleted. recurs, perform a reinstallation.
database.
70042 FFCustomMsgBox [70042] Failed in accessing It is possible that the database file is Restart the system. If the same symptom
the system setup database. damaged or deleted. recurs, perform a reinstallation.
70043 FFCustomMsgBox [70043] Failed in storage of It is possible that the file is damaged or Retry file storage. If the same symptom
image data. deleted. It is also possible that the target recurs, perform a reinstallation.
media for file storage has problems.
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-133
BLANK PAGE
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-134
BLANK PAGE
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-135
BLANK PAGE
010-305-03E
02.28.2011 FM5818
DR-ID 300CL Service Manual Appx MT A-136
Appendix B DR-ID 300CL Error Message Table (V5.0 or Later)
➥ REFERENCE
The error message of DR-ID 300CL is displayed by starting with two-character alphabet, which is an abbreviation for subsystem name.
The relationship of abbreviations for subsystem names and the content is shown below.
Abbreviated subsystem name Subsystem name Function/Role
AC ApplicationCommon Common software library
BM BMDMeasurement All kinds of bone density measurements (DXA calibration, measurement
function, report output, history fuction, DB maintenance fucntion)
BR Bridge Smart controller
DA DaemonActivity Image input on the background from CR
DE Delivery Output to Printer/PACS
DT DataEditor Study information display
IC IDAutoConvert Patient ID auto-convert
MD Modality CR/DR image input
ME MediaStorage Media output
MI ModalityInterface Communication with CR/DR/Xcon
MQ MammoQC Mammogram QC
OI OfflineImporter Image acquisition of memory exposure
OM OperationManager Login/Logout, splash window display
QA QA All QA windows
QC QC General QC
RD RDPClient Non-FUJIFILM RIS connection (remote desktop)
RE Reception Study reception to study start
SM SystemManager Overall management of DR-ID 300CL
ST Study All study windows
TD TomoDaemon Data notification to Beneo-Fx Shimadzu tomosynthesis WS
TS Tomosynthesis All tomosynthesis windows for FF
UU UserUtility UserUtility
WL WebLinker Remote image display
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-1
Event Event Source Description Causes Remedies Remarks
0 CslAdv_ ClusteringManager Error Abnormality occurred in the program. Restart the application. If the same All details of abnormality
ClusteringManager Occuerd. Refer to the phenomenon recurred, reinstall the (for example, error
Application log. system. message and error spot)
are recorded to details
log.
0 CslAdv_ 1% 2% Abnormality occurred in the program. Restart the application. If the same 1%: Error message
DaemonActivity phenomenon recurred, reinstall the 2%: Error spot
system. (Stackinformation)
0 CslAdv_DataEditor {0} Abnormality occurred in the program. Restart the application. If the same {0}: The name of the
phenomenon recurred, reinstall the function which error
system. occurred.
0 CslAdv_Delivery Can not Open Sub Key {0} Failed in opening registry key. Check the setting parameter and then {0}: Registry key
restart the application. If the same
phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Can not Send WaitOne function of signal returned an Check the setting parameter and then
SubApplication Wakeup error. restart the application. If the same
Message phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery It Failed in Configuraion Initialization function of setting returned an Check the setting parameter and then {0}: Function return value
Initialization (rc={0}) error. restart the application. If the same
phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Linker Control Init Method Initialization function of linker returned an Check the setting parameter and then {0}: Function return value
Call Error (rc={0}) error. restart the application. If the same
phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Not Found Application Key Failed in acquiring application key. Check the setting parameter and then
restart the application. If the same
phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery OutImgPrc Control An error was notified to the responding Check the setting parameter and then {0}: Status
RunProc Error (rc={0}) status of RunProcComplete event restart the application. If the same
processed in OutImagePrc. phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery OutImgPrc Control An error was notified to the responding Check the setting parameter and then {0}: Status
RunProc Method Call Error status of RunProcComplete event restart the application. If the same
(rc={0}) processed in Spooler. phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Spooler Control RestoreJob Unprocessed JOB restoration request Check the setting parameter and then {0}: Function return value
Method Call Error (rc={0}) function returned an error to Spooler. restart the application. If the same
phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-2
Event Event Source Description Causes Remedies Remarks
0 CslAdv_Delivery Spooler Control ResumeProc execution function of Spooler Check the setting parameter and then {0}: Function return value
ResumeProc Method Call returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Spooler Control RunProc An error was notified to the responding Check the setting parameter and then {0}: Status
Error (rc={0}) status of RunProcComplete event restart the application. If the same
processed in Spooler. phenomenon recurred, reinstall the
system.
0 CslAdv_Delivery Spooler Control RunProc Execution function of Spooler returned an Check the setting parameter and then {0}: Function return value
Method Call Error (rc={0}) error. restart the application. If the same
phenomenon recurred, reinstall the
system.
0 CslAdv_Re Error when initializes Mwm/ Failed in initializing Mwm/PPS Controller. Restart DR-ID 300CL.
PPS controller If the same phenomenon recurred,
initialize the image database.
0 [QA00000] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
0 [RE00000] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
0 [TS00000] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same error recurs after restarting,
reinstall DR-ID 300CL.
0 CslAdv_Qc Access OLE DB failed. Failed in accessing the database. Restart QC. If the same phenomenon
recurred, reinstall DR-ID 300CL.
0 [IC00000] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same phenomenon recurred, reinstall
DR-ID 300CL.
0 CslAdv_Delivery Spooler Control RestoreJob An error was notified to the responding Restart DR-ID 300CL. {0}: Status
Error (rc={0}) status of RestoreJobComplete event If the same error recurs after restarting,
processed in Spooler. reinstall DR-ID 300CL.
0 CslAdv_Bridge - An exception occurred in application. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ IRSLibraryException Error, Create IRSLibrary instance error. Restart DR-ID 300CL. If the same
DaemonActivity sportCode = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ Exception Error, sportCode Constructor class AnalyseCtrl error. Restart DR-ID 300CL. If the same
DaemonActivity = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ IRSLibraryException Error, Initialize component of IRS. Restart DR-ID 300CL. If the same
DaemonActivity sportCode = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-3
Event Event Source Description Causes Remedies Remarks
0 CslAdv_ IRSLibraryException Error, Terminate IRSLibrary error. Restart DR-ID 300CL. If the same
DaemonActivity sportCode = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ Exception Error, sportCode Terminate IRS FrameWork error. Restart DR-ID 300CL. If the same
DaemonActivity = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ IRSLibraryException Error, Synchronize analyzing IRS for image error. Restart DR-ID 300CL. If the same
DaemonActivity sportCode = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ Exception Error, sportCode Implement analyze IRS error. Restart DR-ID 300CL. If the same
DaemonActivity = {0} [%1] phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ 2 image is not couple Left/ Identify of two image is not couple left/ Select the left and the right menu as a pair.
DaemonActivity Right right.
0 CslAdv_ AC getter to get image Using AC getter to get image information Restart DR-ID 300CL.
DaemonActivity information fail error. If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ BC FCD to caculate adjust Using BC FCD to calculate adjust value Restart DR-ID 300CL.
DaemonActivity value fail error. If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ AC setter fail Using AC setter to set alignment error. Restart DR-ID 300CL.
DaemonActivity If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ AlignMammoImage fail Setting align mammo image error. Restart DR-ID 300CL. If the same
DaemonActivity phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ Get pair menu fail Get pair menu error. Restart DR-ID 300CL.
DaemonActivity If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ Check alignment of both Check alignment of both menu error. Restart DR-ID 300CL.
DaemonActivity menu fai If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ Image is not mammo Image is not mammo. Restart DR-ID 300CL.
DaemonActivity If the same phenomenon recurred,
initialize the image database.
0 CslAdv_ Wrong direction Two menu not same direction. Select the left and the right menu as a pair.
DaemonActivity
0 CslAdv_ Menu can not align Alignment for menu error. Restart DR-ID 300CL. If the same
DaemonActivity phenomenon recurred, reinstall DR-ID
300CL.
0 CslAdv_ FCDAppImgCreatorWS Using BC::ImageProc to create image Restart DR-ID 300CL. If the same
DaemonActivity Combine Error combine error. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-4
Event Event Source Description Causes Remedies Remarks
0 - “Main()” An error occurs when the startup of the Restart DR-ID 300CL. If the same
“Exception Notification from UserUtility.exe application failed. phenomenon recurred, reinstall DR-ID
UnhandledException 300CL.
EventHandler”
0 PatInfoDisplay CslAdv_AC_ An illegal parameter occurred inside %3. As an internal error occurred, restart DR- %1: Process ID
PatInfoDisplay: ID 300CL. %2: Process name
GeneralFailure If the error occurred again, restart the PC. %3: Detailed
processID: %1 If the error could not be recovered, reinstall information
processName: %2 the application.
%3
0 CslAdv_RDPClient LoadConfig Failed in reading the required settings Restart DR-ID 300CL. {0}: Exception message
Error.¥r¥n{0}¥r¥n{1}¥n when starting the general-purpose RIS If the same error recurs after restarting, {1}: Stack trace
Browser. reinstall DR-ID 300CL.
0 HotfixInstallerTool HotfixInstallerTool. Something is wrong with the parameter Re-register the startup script. {0}: The parameter
Exception. when executing the tool. The registration value entered when
HotfixInstallerToolException: of the startup script may not be correct. executing the tool.
Command line parameter is
invalid. Unspecified or [/a]
are supported, but your
input is [{0}]
0 HotfixInstallerTool Can not read xml file. Though the judgment of whether to apply It may be a defect on DR-ID 300CL. Ask
C:\Program Files\FujiFilm\ the Hotfix was made, the catalog file is not the developer to do the analysis.
IIP\Work\Installer\Hotfix\ existing in the local directory.
Catalogfile.xml, /update-
list/installer/installerset
0 HotfixInstallerTool Extract from zip file is Fail to decompress the zip file of Hotfix. Request for the redelivery of the {0}: The zip file name of
failed. Zip file is invalid. The zip file may be damaged. hotfixinstaller from the ActiveLine. the Hotfixinstaller.
C:\Program Files\FujiFilm\
IIP\Work\Installer
\Hotfix\{0}.zip
0 HotfixInstallerTool Can not delete C:\Program Fail to delete the directory which the zip Reinstall the Hotfix manually. {0}: The directory name
Files\FujiFilm\IIP\Work\ file of Hotfixinstall was decompressed. The which the
Installer\Hotfix\{0} files in the directory may be used by other Hotfixinstaller file
tools. was decompressed.
0 HotfixInstallerTool Xml element does not exist. There is no <release-version> tag in the It may be a defect on AcitveLine. Ask the
C:\Program Files\FujiFilm\ catalog file. developer to do the analysis.
IIP\Work\Installer\Hotfix\
Catalog.xml, /update-list/
Installer/release-version
0 HotfixInstallerTool Xml element does not exist. There is no <installerset> tag in the catalog It may be a defect on AcitveLine. Ask the
C:\Program Files\FujiFilm\ file. developer to do the analysis.
IIP\Work\Installer\Hotfix\
Catalog.xml, /update-list/
Installer/installerset
0 HotfixInstallerTool Failed to execute Setup. Setup.exe ended abnormally. Reinstall the Hotfix manually. {0}: The exit code value
exe. of Setup.exe
Return code is {0}.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-5
Event Event Source Description Causes Remedies Remarks
0 Setup Failed backup process in Fail to run the backup process in the auto Perform the backup for Hotfix manually.
auto mode. installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Failed install process in Fail to run the installation process in the Reinstall the Hotfix manually.
auto mode. auto installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Failed restore process in Fail to run the restoration process in the Restore the Hotfix manually.
auto mode. auto installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Failed backup process in Fail to run the backup process in the Perform the backup for Hotfix manually.
manual mode. manual installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Failed install process in Fail to run the installation process in the Reinstall the Hotfix manually.
manual mode. manual installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Failed restore process in Fail to run the restoration process in the Restore the Hotfix manually.
manual mode. manual installation. If it still cannot be done, ask the developer
to do the analysis.
0 Setup Free disk space is insufficient. The disk free space is insufficient. Fail to Increase the disk free space and install the {0}: Disk free space
Expand the hard disk capacity install the Hotfix. Hotfix again. {1}: Size of the Hotfix
or delete unnecessary files to installer
increase free disk space.
Free Space={0}
HotfixSize={1}
0 Setup An image being output The output queue is existing. Fail to install Reinstall the Hotfix after outputting all {0}: Number of the
exists. the Hotfix. output queue or deleting all output queue. output queue
RecordCount={0}
0 Setup PowerLineStatus={0} The PC power adapter is not connected to Connect the PC power adapter to the {0}: PC power status
power supply. Fail to install the Hotfix. power supply and reinstall the Hotfix.
0 Setup Mismatched version The applicable version information of the Confirm the applicable version of Hotfix
information. Hotfix is not matched to the DR-ID 300CL. and reinstall it.
CurrentVersion = {0}
ReleaseVersion = {1}
0 SmartCardControl Exception : {0} Message : {1} A connection error with the IC card reader, Connect the IC card reader again, or {0}: Parameter value in
Target : {2} : Source : {3} or an error during IC card authentication authenticate the IC card again. execution
may have occurred. If it does not get better, delivery the {1}: Parameter value in
analysis request to the development execution
section. {2}: Parameter value in
execution
{3}: Parameter value in
execution
0 to 2 ClsAdv_Tomo 1% An error occurred when starting Restart DR-ID 300CL. 1%: The spot and
Tomosynthesis View. If the same phenomenon recurred, reinstall cause error
DR-ID 300CL. occurred
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-6
Event Event Source Description Causes Remedies Remarks
0 to 3 MenuSel EventID (00000) AC-CGU- In %1 (Name of the file event occurred) As an internal error occurred, restart DR- %1: Fujifilm.
MenuSel:Unknown Error and %2 (Name of the method and row ID 300CL. ConsoleAdvance.
%1 %2 number event occurred), as there is flaw in If the error occurred again, restart the PC. ApplicationCommon
the parameter set in MenuSel, an error If the error could not be recovered, reinstall Library.GUI.MenuSel.
was returned. the application. MenuSel.Initialize()
%2: Z:\Source\....\
MenuSel\MenuSel.
cs:Lines 1186
0 CslAdv_QA 1% An error occurred in QA processing. Restart the application. If the same 1%: The name of the
to 2% phenomenon recurred, reinstall the module error
22103 system. occurred
2%: The spot and
cause error
occurred
1 CslAdv_AC_ImgDBC %1 An error occurred in ADO. As an internal error occurred, restart DR- %1: Method name
%2 ID 300CL. %2: Various types of
%3 If the error occurred again, restart the PC. information
%4 If the error could not be recovered, reinstall %3: Error number of ADO
%5 the application. %4: Provider-specific
%6 Error code of ADO
%7 %5: SQL status of ADO
%6: ADO object name
%7: Description of ADO
object
1 CslAdv_AC_ImgDBCEx CslAdv_AC_ImgDBCEx:Gen A general error occurred in the method of As an internal error occurred, restart DR- %1: Method name
eralFailure(Fujifilm. %1. ID 300CL. %2: Detailed information
ConsoleAdvance. If the error occurred again, restart the PC.
ApplicationCommonLibrary. If the error could not be recovered, reinstall
DataAccessor.ImgDBCEx. the application.
ImgDBCExException
Occuerd.
Code:GeneralFailure:%1,%2
1 CslAdv_ LogCtl2 Create Error. Failed in initializing AC log. Restart the application. If the same {0}: Error message of AC
ClusteringManager Msg:[{0}]\n{1} phenomenon recurred, reinstall the {1}: Error spot of ADO
system.
1 CslAdv_Delivery Spooler Control Data inconsistency checking request Check the setting parameter and then {0}: Function return value
ErrorCheckProc Method function returned an error to Spooler. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
1 CslAdv_Delivery Spooler Control Completing request function of Spooler Check the setting parameter and then {0}: Function return value
GetCommitProcJob Method returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
1 CslAdv_Modality %1 %2 Failed in initialization processing. Restart the application. If the same %1: The module error
Initialize %3 phenomenon recurred, reinstall the occurred
system. %2: Spot where error
occurred
%3: Error details
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-7
Event Event Source Description Causes Remedies Remarks
1 CslAdv_Modality Failed Init %1 Failed in initializing indicator. Restart the application. If the same %1: Indicator name.
phenomenon recurred, reinstall the
system.
1 CslAdv_Modality Failed to %1.Initialize Failed in initializing %1. Restart the application. If the same %1: Indicator name or
phenomenon recurred, reinstall the status window
system. name.
1 CslAdv_Modality Failed to Initialize%1 Failed in initializing indicator. Restart the application. If the same %1: Indicator name.
phenomenon recurred, reinstall the
system.
1 CslAdv_ Lock Error Failed in accessing the Lock database. Restart the application. If the same One of these three
OperationManager GetAllLockInfo Error phenomenon recurred, reinstall the types is set in event
GetLockType Error system. content.
1 CslAdv_ System Error Receive A system error message was received. Restart the application. If the same
SystemManager phenomenon recurred, reinstall the
system.
1 [DE00001] Failed in Failed in initialization. Restart the system.
initialization. If the same error recurs after restarting the
system, reinstall the system.
1 [DT00001] Key item setting Key item setting error Correct the item settings.
error
1 [MD00001] Failed in Failed in initialization. Restart the system.
initialization. If the same error recurs after restarting the
system, reinstall the system.
1 [OM00001] Failed in Failed in initialization. Restart the system.
initialization. If the same error recurs after restarting the
system, reinstall the system.
1 [QA00001] Failed in Failed in obtaining information from the Restart the system.
obtaining information from setting file. If the same error recurs after restarting the
the setting file. system, reinstall the system.
1 [RE00001] Failed in Failed in accessing to database. Restart the system.
accessing to database. If the same error recurs after restarting the
system, reinstall the system.
1 [SM00001] Failed in Failed in initialization processing. Restart the system.
initialization processing. If the same error recurs after restarting the
system, reinstall the system.
1 CslAdv_Modality ModalityViewException Failed in initializing the panel change Restart the system.
Msg = Failed to Initialize dialog box. If the same phenomenon recurred, reinstall
FPDPanelChangeDialog the system.
1 CslAdv_Modality ModalityViewException Failed in initializing the panel change Restart the system.
Msg = Failed to Initialize dialog box's event listener. If the same phenomenon recurred, reinstall
panelChangeViewListener the system.
Class
1 CslAdv_Modality Failed to execute Failed in initializing the panel change Restart the system.
PanelGUIInitialize dialog box. If the same phenomenon recurred, reinstall
the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-8
Event Event Source Description Causes Remedies Remarks
1 [WL00001] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1 [QC00001] Failed in Failed in accessing the database. Click the close button on the QC execution
accessing the database. window.
1 [TS00001] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1 CslAdv_IDAutoConvert catch ThreadException An exceptional event ThreadException Restart DR-ID 300CL.
occurred. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1 [DA00001] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1 [IC00001] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1 MegBoxSyncHelper CslAdv_AC_ An unexpected program error has occurred Restart DR-ID 300CL. If the same kind of %1: Detailed information
MsgBoxSyncHelper during display message synchronization phenomenon occurs again, set the smart
QueryChildWindow Error between the smart controller and DR-ID control function to OFF, or reinstall.
Illegal Message Contents: 300CL.
%1
1 MegBoxSyncHelper CslAdv_AC_ An unexpected program error has occurred Restart DR-ID 300CL. If the same kind of %1: Detailed information
MsgBoxSyncHelper during display message synchronization phenomenon occurs again, set the smart
QueryChildWindow Error between the smart controller and DR-ID control function to OFF, or reinstall.
Illegal UniqueId1: %1 300CL.
1 MegBoxSyncHelper CslAdv_AC_ An unexpected program error has occurred Restart DR-ID 300CL. If the same kind of %1: Detailed information
MsgBoxSyncHelper during display message synchronization phenomenon occurs again, set the smart
QueryChildWindow Error between the smart controller and DR-ID control function to OFF, or reinstall.
Illegal UniqueId2: %1 300CL.
1 CslAdv_Modality Exception occured Failed in rollback of the DDO file. Restart DR-ID 300CL. If the same %1: The record number
RollbackDDO. imageId : phenomenon recurred, reinstall DR-ID of the image
%1 300CL. database.
1 CslAdv_Re TreeView Filtering Release Failed in accessing the database. Restart DR-ID 300CL.
Error.
1 CslAdv_Re TreeView Filtering Error. Failed in accessing the database. Restart DR-ID 300CL.
1 CslAdv_Re TreeView Specification Failed in accessing the database. Perform the operation again.
Error. If the same phenomenon recurred, restart
DR-ID 300CL.
1 CslAdv_Re TreeView ImageMove Failed in accessing the database. Restart DR-ID 300CL.
Error. Failed in accessing the DDO file.
1 CslAdv_Re TreeView End Error. Failed in each closing process of DB, Restart DR-ID 300CL.
DDO, etc.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-9
Event Event Source Description Causes Remedies Remarks
1 CslAdv_Re TreeView DragOverEvent Failed in accessing the database. Perform the operation again.
Error. Failed in accessing the DDO file. If the same phenomenon recurred, restart
DR-ID 300CL.
1 CslAdv_Re TreeView DragEnterEvent Failed in accessing the database. Perform the operation again.
Error. Failed in accessing the DDO file. If the same phenomenon recurred, restart
DR-ID 300CL.
1 CslAdv_Re TreeView DragDropEvent Failed in accessing the database. Perform the operation again.
Error. Failed in accessing the DDO file. If the same phenomenon recurred, restart
DR-ID 300CL.
1 CslAdv_Re Initialize Exception Error. Failed in accessing the database. Restart DR-ID 300CL.
Failed in accessing the DDO file.
1 CslAdv_Re Refresh Exception Error. Failed in accessing the database. Restart DR-ID 300CL.
Failed in accessing the DDO file.
1 CslAdv_Re UpdateDBAndDDOByTo Failed in accessing the database. Restart DR-ID 300CL.
Tree Failed in accessing the DDO file.
ViewData Error.
1 CslAdv_Re UpdateDBAndDDOBy Failed in accessing the database. Restart DR-ID 300CL.
FromTreeViewData Error. Failed in accessing the DDO file.
1 CslAdv_AC_ToolMenu CslAdv_AC_ToolMenu:%1 A general error occurred in the method of Restart DR-ID 300CL. %1: Method name
Error:Exception.. %1. If the error occurred again, restart the PC. %2: Process ID
processID:%2 If the error could not be recovered, reinstall %3: Process name
processName:%3 the application. %4: Detailed
%4 information
1 CslAdv_AC_SideBar CslAdv_AC_SideBar:%1 A general error occurred in the method of Restart DR-ID 300CL. %1: Method name
Error:Exception.. %1. If the error occurred again, restart the PC. %2: Process ID
processID:%2 If the error could not be recovered, reinstall %3: Process name
processName:%3 the application. %4: Detailed information
%4
1 PatInfoDisplay CslAdv_AC_ A general error occurred inside %3. As an internal error occurred, restart DR- %1: Process ID
PatInfoDisplay: ID 300CL. %2: Process name
GeneralFailure If the error occurred again, restart the PC. %3: Detailed information
processID: %1 If the error could not be recovered, reinstall
processName: %2 the application.
%3
1 CslAdv_Modality Failed to Failed in initializing the device status Restart DR-ID 300CL.
InitializeStatusDialog dialog. If the same error recurs after restarting,
reinstall DR-ID 300CL.
1 CslAdv_Modality Failed to initialize the Failed in initializing the device status Restart DR-ID 300CL.
ModalityOutputLog dialog. If the same error recurs after restarting,
reinstall DR-ID 300CL.
1 CslAdv_Modality ScreenDesign Value Error. The system settings (ScreenDesign) are Restart DR-ID 300CL. %1: Error code
value invalid. If the same error recurs after restarting,
= %1 reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-10
Event Event Source Description Causes Remedies Remarks
1 CslAdv_Modality Initialize Failed in acquiring the value of DB. Restart DR-ID 300CL. %1: Error code
IDeviceSettingCollectionClass. If the same error recurs after restarting,
Get() ErrorCode = %1 reinstall DR-ID 300CL.
1 CslAdv_Modality ModalityException Msg = An error occurs inside the program. Restart DR-ID 300CL. %1: Error message
%1 If the same error recurs after restarting,
reinstall DR-ID 300CL.
1 CslAdv_RDPClient Save Config Failed in saving the settings for the end Restart the general-purpose RIS browser. {0}: Exception message
Error.¥r¥n{0}¥r¥n{1}¥n processing of general-purpose RIS If the error recurs restarting the general- {1}: Stack trace
Browser. purpose RIS browser, reinstall DR-ID
300CL.
1 [RD00001] An error An error occurred. Restart the system.
occured If the same symptom recurs, reinstall the
system.
1 CslAdv_AC_GuiAdp %1 The path of screen keyboard set in Check if the path of screen keyboard is %1: Error message
InitializeScreenKeyBoard() “SysConfig.mdb” file is not found. correct in “SysConfig.mdb” file. %2: Path of screen
Key:SoftKBProgPath board specified by
[%2] is not found. “SysConfig.mdb”
file
1 CslAdv_Study 1% An unrecoverable error occurred in the DR-ID 300CL restarts automatically. 1%: Error function name
to process. Restart DR-ID 300CL if it has frozen. If the
9999 same phenomenon recurred, reinstall the
system.
2 CslAdv_AC_ImgDBC FFImageDBCTL Failed in accessing the database. As an internal error occurred, restart DR- %1: Error information
%1 ID 300CL. %2: Method name
%2 If the error occurred again, restart the PC. %3: Various types of
%3 If the error could not be recovered, reinstall information
the application.
2 CslAdv_Delivery Invalid Media Status MediaStatAEName execution function of Check the setting parameter and then {0}: AE name
Parameter (ae_name={0}, Spooler returned an error. restart the application. If the same {1}: Status
media_status={1}, free_ phenomenon recurred, reinstall the {2}: Blank space
space={2}) system.
2 CslAdv_Delivery Spooler Control MediaStatAEName execution function of Check the setting parameter and then {0}: Function return value
MediaStatAEName Method Spooler returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control MediaStatErr execution function of Spooler Check the setting parameter and then {0}: Function return value
MediaStatErr Method Call returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control MediaStatFree execution function of Check the setting parameter and then {0}: Function return value
MediaStatFree Method Call Spooler returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-11
Event Event Source Description Causes Remedies Remarks
2 CslAdv_Delivery Spooler Control MediaStatMount execution function of Check the setting parameter and then {0}: Function return value
MediaStatMount Method Spooler returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control MediaStatUnMount execution function of Check the setting parameter and then {0}: Function return value
MediaStatUnMount Method Spooler returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control Image process completion notification Check the setting parameter and then {0}: Function return value
SetImageProcComplete function returned an error to Spooler. restart the application. If the same
Method Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control Output completion notification function Check the setting parameter and then {0}: Function return value
SetOutputProcComplete returned an error to Spooler. restart the application. If the same
Method Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control SpoolProc An error was notified to the responding Check the setting parameter and then {0}: Status
Error (rc={0}) status of SpoolProcComplete event restart the application. If the same
processed in Spooler. phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control SpoolProc Spool request function returned an error to Check the setting parameter and then {0}: Function return value
Method Call Error (rc={0}) Spooler. restart the application. If the same
phenomenon recurred, reinstall the
system.
2 CslAdv_Delivery Spooler Control Spool request function returned an error to Check the setting parameter and then {0}: Function return value
StudyReqEndProc Method Spooler. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
2 CslAdv_Modality %1 %2 Failed in terminating processing. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Detailed
%4 information
%3:Error outline
%4:Error details
2 CslAdv_Modality Failed to %1.Terminate Failed in completion processing of %1. Restart the application. If the same %1:Indicator name or
phenomenon recurred, reinstall the status window
system. name.
2 CslAdv_ Msg Get Error Failed in accessing the resource database. Restart the application. If the same
OperationManager phenomenon recurred, reinstall the
system.
2 CslAdv_ UPS Signal Receive UPS abnormality message was received. Restart the application. If the same
SystemManager phenomenon recurred, reinstall the
system.
2 [DE00002] Failed in output Failed in output processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-12
Event Event Source Description Causes Remedies Remarks
2 [MD00002] Failed in Failed in terminating processing. Restart the system.
terminating processing. If the same error recurs after restarting the
system, reinstall the system.
2 [OM00002] Failed in Failed in starting up the service utility. Restart the system.
starting up the service If the same error recurs after restarting the
utility. system, reinstall the system.
2 [QA00002] Failed in QA Failed in QA starting process. Restart the system.
starting process. If the same error recurs after restarting the
system, reinstall the system.
2 [RE00002] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
2 [SM00002] System error System error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
2 [WL00002] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same error recurs after restarting,
reinstall DR-ID 300CL.
2 [QC00002] Failed to Failed to initialize QC. Click the close button on the QC execution
initalize QC. window.
2 [TS00002] Failed in Failed in obtaining the information from the Restart DR-ID 300CL.
obtaining the information system setting file. If the same phenomenon recurred, reinstall
from the system setting file. DR-ID 300CL.
2 CslAdv_IDAutoConvert catch Exception An exception occurred while processing an Restart DR-ID 300CL.
exceptional event ThreadException. If the same phenomenon recurred, reinstall
DR-ID 300CL.
2 [IC00002] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
2 CslAdv_AC_ToolMenu CslAdv_AC_ %1 file does not exist. Restart DR-ID 300CL. %1: File name
ToolMenu:SysInfo If the error occurred again, restart the PC. %2: Process ID
File is not exist. If the error could not be recovered, reinstall %3: Process name
File path=%1 the application.
processID:%2
processName:%3
2 CslAdv_AC_ToolMenu CslAdv_AC_ An error occurred when accessing the Restart DR-ID 300CL. %1: Process ID
ToolMenu:SysConfig.mdb system setting file. If the error occurred again, restart the PC. %2: Process name
Access If the error could not be recovered, reinstall %3: Detailed
Error:SysDBCException. the application. information
processID:%1
processName:%2
%3
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-13
Event Event Source Description Causes Remedies Remarks
2 CslAdv_AC_ToolMenu CslAdv_AC_ToolMenu:%1 An error occurred when accessing the data Restart DR-ID 300CL. %1: Method name
Error:Exception.. in the method of %1. If the error occurred again, restart the PC. %2: Process ID
processID:%2 If the error could not be recovered, reinstall %3: Process name
processName:%3 the application. %4: Detailed
%4 information
2 PatInfoDisplay CslAdv_AC_ The access to DB is illegal inside %3. As an internal error occurred, restart DR- %1: Process ID
PatInfoDisplay: ID 300CL. %2: Process name
GeneralFailure If the error occurred again, restart the PC. %3: Detailed
processID: %1 If the error could not be recovered, reinstall information
processName: %2 the application.
%3
2 [RD00002] An error An error occurred. Restart the system.
occurred. If the same symptom recurs, reinstall the
system.
2 CslAdv_OfflineImporetr Not create thumbnail An abnormal image received from FPD Perform an exposure again. 1%: Target image file
image, target filename = when importing a image. name
%1
2 CslAdv_OfflineImporetr Not create thumbnail It is conceivable that an error occurred Confirm that the target image is imported 1%: Directory name
images. Reason : No when importing a image. by DR-ID 300CL (ImageBox). created when
images. Target directory Import the image again from the panel if imported
path = %1 not imported by DR-ID 300CL (ImageBox).
2 CslAdv_OfflineImporetr Target menu is not only Menu of image matching target could not Close the image library, and reselect the %1: Number of the
one. total menu number is be identified. menu. If the same log has been recorded, selected menu
%1 finish and restart the study.
2 CslAdv_OfflineImporetr Target selecters %1 and Selector of the acquisition of the original Close the image library, and confirm the %1: Number of selector
%2 are difference. image and the selector of the menu of the menu selector once again. that import the
image with the matching target is different. If the same log has been recorded, finish image
and restart the study. %2: Number of selector
of the menu of
image match target
2 CslAdv_OfflineImporetr Image status is invalid. %1 Menu of image matching target has been Image matching target menu is required to %1: Exposure status of
exposed. be unexposed. Confirm the status of the the image match
exposure. target menu
If the same log has been recorded, finish
and restart the study.
2 CslAdv_OfflineImporetr function type is not correct. Menu of the Function image matching The function of the image matching target %1: Function of the
%1 target is invalid. menu is required to be in RT/PS/TM. image match target
Confirm the function. menu
If the same log has been recorded, finish
and restart the study.
3 CslAdv_AC_ImgDBC Image Database %1 Error An error occurred in the method of %1. As an internal error occurred, restart DR- %1: Method name
ID 300CL.
If the error occurred again, restart the PC.
If the error could not be recovered, reinstall
the application.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-14
Event Event Source Description Causes Remedies Remarks
3 CslAdv_AC_ImgDBC Image Database %1 Unknown error. As an internal error occurred, restart DR- %1: Method name
Exception ID 300CL.
If the error occurred again, restart the PC.
If the error could not be recovered, reinstall
the application.
3 CslAdv_Delivery OutImage Control Image process completion notification Check the setting parameter and then {0}: Function return value
ExecImageProc Method function returned an error to Spooler. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
3 CslAdv_Delivery Spooler Control Acquiring request function of image Check the setting parameter and then {0}: Function return value
GetImageProcJob Method processing JOB in Spooler returned an restart the application. If the same
Call Error (rc={0}) error. phenomenon recurred, reinstall the
system.
3 CslAdv_Modality %1 %2 Failed in acquiring directory information. Restart the application. If the same %1: The module error
Failed to Read Directory.ini: phenomenon recurred, reinstall the occurred
%3. system. %2: Spot error occurred
%3: Error details
3 CslAdv_ DB Error Failed in accessing SysConfig,Paramedic. Restart the application. If the same
OperationManager mdb. phenomenon recurred, reinstall the
system.
3 CslAdv_Re Error message of A problem occurred in _console_db. Restart DR-ID 300CL.
AC::ImgDBCEx
3 CslAdv_ BCR error Failed in creating data container or xml file Restart the application. If the same
SystemManager for BCR. phenomenon recurred, reinstall the
system.
3 [DE00003] Failed in image Failed in image processing for output. Restart the system.
processing for output. If the same error recurs after restarting the
system, reinstall the system.
3 [MD00003] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
3 [OM00003] Failed in Failed in executing the lock processing. Restart the system.
executing the lock If the same error recurs after restarting the
processing. system, reinstall the system.
3 [QA00003] Failed in Failed in initialization for communication. Restart the system.
initialization for If the same error recurs after restarting the
communication. system, reinstall the system.
3 [RE00003] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
3 [SM00003] A power failure A power failure has occurred. Restart the system.
has occurred. If the same error recurs after restarting the
system, reinstall the system.
3 [QC00003] Failed to Failed to displaying thumbnail image. Click the close button on the QC execution
displaying thumbnail image. window.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-15
Event Event Source Description Causes Remedies Remarks
3 [TS00003] Recovery was Recovery was attempted from an error Restart DR-ID 300CL.
attempted from an error during the reconstruction processing but it If the same error recurs after restarting,
during the reconstruction failed. reinstall DR-ID 300CL.
processing but it failed.
3 CslAdv_IDAutoConvert catch ThreadException An exception occurred in Program.Main. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
3 [IC00003] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
3 CslAdv_AC_ToolMenu CslAdv_AC_ToolMenu:%1 An error occurred when accessing the Restart DR-ID 300CL. %1: Method name
Error:%2 other modules in the method of %1. If the error occurred again, restart the PC. %2: Spot error occurred
processID:%3 If the error could not be recovered, reinstall %3: Process ID
processName:%4 the application. %4: Process name
%5 %5: Detailed
information
3 CslAdv_AC_SideBar CslAdv_AC_SideBar:%1 An error occurred when accessing the Restart DR-ID 300CL. %1: Method name
Error:%2 other modules in the method of %1. If the error occurred again, restart the PC. %2: Spot error occurred
processID:%3 If the error could not be recovered, reinstall %3: Process ID
processName:%4 the application. %4: Process name
%5 %5: Detailed
information
3 PatInfoDisplay CslAdv_AC_ An external reference error occurred inside As an internal error occurred, restart DR- %1: Process ID
PatInfoDisplay: %3. ID 300CL. %2: Process name
GeneralFailure If the error occurred again, restart the PC. %3: Detailed
processID: %1 If the error could not be recovered, reinstall information
processName: %2 the application.
%3
3 CslAdv_Modality Failed to Read Directory.ini: Failed in finding the connection character Restart DR-ID 300CL.
Key= ModalityDB string for ModalityDB If the same error recurs after restarting,
reinstall DR-ID 300CL.
3 CslAdv_RDPClient Load Directory.ini Failed in reading the setting file (Directory. Restart DR-ID 300CL. {0}: Exception message
Error.¥r¥n{0}¥r¥n{1}¥n ini) needed to start If the same error recurs after restarting, {1}: Stack trace
reinstall DR-ID 300CL.
3 [RD00003] A connection A connection error has occurred. Please confirm the network condition and
error has occurred. restart %s.
4 CslAdv_Delivery Ansemble DataContainer Failed in registration to Ansemble. Check the setting parameter and then {0}: SANameDelivery
Create Error. (Name={0}, DataContainer. restart the application. If the same {1}: ContentType
ContentTyp={1}, Key={2}) phenomenon recurred, reinstall the {2}: Key
system.
4 CslAdv_Delivery DICOM Print Control Failed in acquiring the film number. Check the setting parameter and then {0}: Film number
PrintPrc Error restart the application. If the same
(indexNo={0}) phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-16
Event Event Source Description Causes Remedies Remarks
4 CslAdv_Delivery DICOM Print Control An error was notified to PrintPrc execution Check the setting parameter and then {0}: Execution result
PrintPrc Error (prcNo={0}) result. restart the application. If the same
phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery DICOM Print Control Dicom printer output control function Check the setting parameter and then {0}: Function return value
PrintPrc Method Call Error returned an error. restart the application. If the same
(rc={0}) phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery DICOM Store Control Output request function of storage output Check the setting parameter and then {0}: Function return value
StorePrc Method Call Error control returned an error. restart the application. If the same
(rc={0}) phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery Output ddo file is not found No DDO files exist. Check the setting parameter and then {0}: DDO file
({0}) restart the application. If the same
phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery OutputJob Get Item Error Acquired function of GetItem returned an Check the setting parameter and then {0}: function
(function={0}, id={1}, rc={2}) error. restart the application. If the same {1}: id
phenomenon recurred, reinstall the {2}: rc
system.
4 CslAdv_Delivery OutputProc Error (rc={0}, Output completion notification function Check the setting parameter and then {0}: Function return value
UnitType={1}) returned an error to Spooler. restart the application. If the same
phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery PostMessage Method Call Post function of sending message returned Check the setting parameter and then {0}: Function return value
Error (rc={0}) an error. restart the application. If the same
phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery Spooler Control Acquiring request function of output JOB in Check the setting parameter and then {0}: Function return value
GetOutputProcJob Method Spooler returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
4 CslAdv_Delivery Spooler Control NotifyUpdateImageStatusResult function Check the setting parameter and then {0}: Function return value
NotifyUpdateImage of Spooler returned an error. restart the application. If the same
StatusResult Method Call phenomenon recurred, reinstall the
Error (rc={0}) system.
4 CslAdv_Delivery Spooler Control Delivery completion notification function Check the setting parameter and then {0}: Function return value
SetCommitProcComplete returned an error to Spooler. restart the application. If the same
Method Call Error (rc={0}) phenomenon recurred, reinstall the
system.
4 CslAdv_Modality %1 %2 Value of system setting is invalid. Restart the application. If the same %1: The module error
%3 Value is invalid phenomenon recurred, reinstall the occurred
system. %2: Spot error occurred
%3: Information of
system setting key
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-17
Event Event Source Description Causes Remedies Remarks
4 CslAdv_Modality %1 %2 Failed in reading the system setting. Restart the application. If the same %1: The module error
Failed to Read SysConfig : phenomenon recurred, reinstall the occurred
%3 system. %2: Spot error occurred
[ExceptionInfomation] %3: Error outline
%4 %4: Error details
4 CslAdv_ Registory Error An error occurred when accessing registry Restart the application. If the same
OperationManager information. phenomenon recurred, reinstall the
system.
4 CslAdv_ MCR error Failed in creating data container for MCR Restart the application. If the same
SystemManager or xml file for BCR. phenomenon recurred, reinstall the
system.
4 [DE00004] Failed in output Failed in output processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
4 [MD00004] Failed in Failed in sending the message. Restart the system.
sending the message. If the same error recurs after restarting the
system, reinstall the system.
4 [OM00004] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
4 [QA00004] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
4 [RE00004] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
4 [SM00004] Failed in lock Failed in lock processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
4 [QC00004] Cannot show Cannot show magnified image. Click the close button on the QC execution
magnified image. window.
4 [TS00004] Failed in starting Failed in starting up the TomoView. Restart DR-ID 300CL.
up the TomoView. If the same error recurs after restarting,
reinstall DR-ID 300CL.
4 [IC00004] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
4 CslAdv_Modality Exception Failed in acquiring the setup value of Restart DR-ID 300CL.
[SysDBCNotFoundException] system. If the same error recurs after restarting,
reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-18
Event Event Source Description Causes Remedies Remarks
4 CslAdv_Modality %1 %2 Failed in initializing the ShotReadyIndicator Restart DR-ID 300CL. If the same %1: Error occurred
failed Init because the acquisition of phenomenon recurred, reinstall DR-ID module
ShotReadyIndicator ISelectorCollection failed. 300CL. %2: Error occurred part
[ExceptionInfomation] %3: Error code
Failed to GetResouce %4: Error outline
ISelectorCollection, Error_
code:%3
%4
5 CslAdv_Delivery Spooler Control QueueResume execution function of Check the setting parameter and then {0}: Function return value
QueueResume Method Spooler returned an error. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
5 CslAdv_Modality %1 %2 • Failed in accessing the database. Restart the application. If the same %1: The module error
%3 • Failed in acquiring information from the phenomenon recurred, reinstall the occurred
database. system. %2: Spot error occurred
• Failed in setting information on the %3: Error details
database.
5 CslAdv_Modality %1 %2 Failed in updating the database. Restart the application. If the same %1: The module error
Execute Invalid phenomenon recurred, reinstall the occurred
Operation:%3 system. %2: Spot error occurred
%3: Error details
5 CslAdv_Modality %1 %2 Failed in updating the database. Restart the application. If the same %1: The module error
Failed to Execute SQL:%3 phenomenon recurred, reinstall the occurred
system. %2: Spot error occurred
%3: Error details
5 CslAdv_Modality %1 %2 Failed in initializing the database. Restart the application. If the same %1: The module error
Failed to Initialize phenomenon recurred, reinstall the occurred
ModalityDB system. %2: Spot error occurred
%3: Error details
5 CslAdv_Modality Failed to get row number of Failed in acquiring data appricable for Restart the application. If the same %1: Search number.
DataSet from _modality_ number=%1 from ModalityDB. phenomenon recurred, reinstall the
status_tbl: number=%1 system.
5 CslAdv_Modality Failed to SqlExecutor. Failed in obtaining the FCR error Restart the application. If the same
ReferenceCommand() for information. phenomenon recurred, reinstall the
get fcr error record system.
5 CslAdv_Modality ModalityException Msg = Failed in updating the access of internal Finish the study, and perform the study %1: Selector number
Failed to SelectorChange database. again. If the same phenomenon recurred,
UpdateModalityDb - restart the application.
SelectorNumber = %1
5 CslAdv_Modality Not Found number in _ Data appricable for number=%1 does not Restart the application. If the same %1: Search number.
modality_status_tbl: exist in ModalityDB. phenomenon recurred, reinstall the
number=%1 system.
5 CslAdv_ Lock error Failed in accessing the Lock database. Restart the application. If the same
SystemManager phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-19
Event Event Source Description Causes Remedies Remarks
5 [DE00005] Failed in output Failed in output processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
5 [OM00005] Failed in Failed in registering the user name. Click the button displayed on a the screen.
registering the user name.
5 [QA00005] Failed in Failed in entering parameters to start QA Restart the system.
entering parameters to start within the study. If the same error recurs after restarting the
QA within the study. system, reinstall the system.
5 [RE00005] Failed in Failed in sending the message. Restart the system.
sending the message. If the same error recurs after restarting the
system, reinstall the system.
5 [SM00005] Failed in Failed in creating a data file. Click the button displayed on a the screen.
creating a data file.
5 CslAdv_Modality Failed to show PanelChang Failed in displaying the panel change Check for the settings of selector and %1 :The selector
eDlg:selectorNumber=%1 dialog box. device again and restart the system. If the number for
same phenomenon recurred, reinstall the changing panel
system.
5 CslAdv_Modality Nothing selector's Failed in displaying the panel change Check for the settings of selector and %1 :The selector
icon:selector num=%1 dialog box. device again and restart the system. If the number for
same phenomenon recurred, reinstall the changing panel
system.
5 CslAdv_Modality Faild to Delete _panel_ Failed in updating the ModalityDB. Restart the system.
change_request_tbl If the same phenomenon recurred, reinstall
the system.
5 CslAdv_Modality Failed to reset selector's Failed in restoring the panel to the setup Restart the system. %1 :The selector
color:SelectorNumber=%1 condition with system setting. If the same phenomenon recurred, reinstall number for
the system. changing panel
5 CslAdv_Modality Failed to show PanelChang Failed in displaying the panel change Restart DR-ID 300CL. If the same %1: Selector number
eDlg:selectorNumber=%1 dialog. phenomenon recurred, reinstall DR-ID
300CL.
5 CslAdv_Modality Faild to Delete _panel_ Failed in operating MD local database. Restart DR-ID 300CL. If the same
change_request_tbl phenomenon recurred, reinstall DR-ID
300CL.
5 CslAdv_IDAutoConvert GuiAdpException An exception occurred while retrieving Restart DR-ID 300CL.
processing result message. If the same phenomenon recurred, reinstall
DR-ID 300CL.
5 [IC00005] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6 CslAdv_Delivery Linker Control End message sending function of queue Check the setting parameter and then {0}: Function return value
PostMessage Method Call returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-20
Event Event Source Description Causes Remedies Remarks
6 CslAdv_Delivery Spooler Control QueueDelete execution function of Spooler Check the setting parameter and then {0}: Function return value
QueueDelete Method Call returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
6 CslAdv_Modality %1 Failed in browsing the image database. Finish the study, and perform the study %1: The module and
again. If the same phenomenon recurred, spot error occurred
restart the application.
6 CslAdv_ ImgDBCEx error Failed in accessing the study database. Restart the application. If the same One of these two types
SystemManager DeleteRecordByStudy phenomenon recurred, reinstall the is set in event content.
Error system.
6 [DE00006] Failed in output Failed in output processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
6 [OM00006] Failed in Failed in acquiring the initial user name. Restart the system.
acquiring the initial user If the same error recurs after restarting the
name. system, reinstall the system.
6 [RE00006] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
6 [SM00006] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
6 CslAdv_Modality Failed to ReleaseExclusion Failed in browsing the image database. Finish the study, and perform the study
again. If the same phenomenon recurred,
restart the Console.
6 CslAdv_Modality Failed to execute Failed in generating the XML to require of When the image is left in the panel, re-
CmdCRModeAcquisition the acquisition of the image. input the image.
Start.Convert If the same situations occur, restart the
DR-ID 300CL.
6 CslAdv_Modality Failed to execute Failed in generating the XML to require the When the image is left in the panel, re-
CmdCRModeDeleteStart. deletion of the image. input the image.
Convert If the same situations occur, restart the
DR-ID 300CL.
7 CslAdv_Delivery {0}\r\n{1} An exception has occurred. Check the setting parameter and then {0}: Exception message
restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery Can not Send {0} Message Message sending function of linker Check the setting parameter and then {0}: Cause of process
returned an error. restart the application. If the same abortion
phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery DICOMPrintSCU Control Request function on completing Check the setting parameter and then {0}: Function return value
EndProc Method Call Error DICOMPrintSCU returned an error. restart the application. If the same
(rc={0}) phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-21
Event Event Source Description Causes Remedies Remarks
7 CslAdv_Delivery DicomSC Control EndProc An error was notified to the responding Check the setting parameter and then {0}: Status
Error (rc={0}) status of EndProcComplete event restart the application. If the same
processed in DicomSCCtrl. phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery DicomSC Control EndProc EndProc function of SuspendComplete Check the setting parameter and then {0}: Function return value
Method Call Error (rc={0}) event processing returned an error. restart the application. If the same
phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery DICOMStoreSCU Control Complete request function to Check the setting parameter and then {0}: Function return value
EndProc Method Call Error DICOMStoreSCU returned an error. restart the application. If the same
(rc={0}) phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery EventLog.InfoEventID. Output what has completed to the event Check the setting parameter and then
Terminate log. restart the application. If the same
phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery OutImage Control EndProc An error was notified to the responding Check the setting parameter and then {0}: Status
Error (rc={0}) status of EndProcComplete event restart the application. If the same
processed in OutImagePrc. phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery OutImagePrc Control Request function on completing OutImgPrc Check the setting parameter and then {0}: Function return value
EndProc Method Call Error returned an error. restart the application. If the same
(rc={0}) phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery Spooler Control EndProc An error was notified to the responding Check the setting parameter and then {0}: Status
Error (rc={0}) status of EndProcComplete event restart the application. If the same
processed in Spooler. phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery Spooler Control EndProc Completing request function of Spooler Check the setting parameter and then {0}: Function return value
Method Call Error (rc={0}) returned an error. restart the application. If the same
phenomenon recurred, reinstall the
system.
7 CslAdv_Delivery Spooler Control SuspendProc execution function of Spooler Check the setting parameter and then {0}: Function return value
SuspendProc Method Call returned an error. restart the application. If the same
Error (rc={0}) phenomenon recurred, reinstall the
system.
7 CslAdv_Modality %1 %2 The character string database of the Image The Image Reader might not been %1: The module error
LngDBCExException Reader has not been installed. installed. Check the system. occurred
[ExceptionInfomation] %2: Spot error occurred
%3 %3: Error details
7 [DE00007] Failed in output Failed in output processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
7 [OM00007] Setup hasn't The application key has not been installed. Click the button to exit the system.
been completed yet. Install the application key.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-22
Event Event Source Description Causes Remedies Remarks
7 [QA00007] Failed in starting Failed in starting QA within the study. Restart the system.
QA within the study. If the same error recurs after restarting the
system, reinstall the system.
7 [RE00007] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
7 CslAdv_ ImgGrpDBC:Delete Failed in accessing the study database. Restart DR-ID 300CL.
SystemManager RecordByStudy Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
7 [SM00007] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same error recurs after restarting,
reinstall DR-ID 300CL.
8 CslAdv_Delivery DicomSC Control Commitment processing request function Check the setting parameter and then {0}: Function return value
CommitmentProc Method returned an error to DicomSC. restart the application. If the same
Call Error (rc={0}) phenomenon recurred, reinstall the
system.
8 CslAdv_Modality %1 %2 Failed in acquiring the selector information. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
% %3: Error code
8 CslAdv_Modality %1 %2 Failed in acquiring the selector information. Restart the application. If the same %1: The module error
Get SelectorDB Data Error. phenomenon recurred, reinstall the occurred
MenuManeger. system. %2: Spot error occurred
SelectorCollection.Get() %3: Error code
return status:%3
8 CslAdv_Re Load system config fail Failed in reading the system setting. Restart DR-ID 300CL.
8 [DE00008] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
8 [RE00008] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
8 CslAdv_ MessagePrameter error Message parameter list is abnormal. Restart DR-ID 300CL.
SystemManager (paramList.Count <= 0)
MessagePrameter error
(paramList.Count >
LockParameterCountMax)
8 [TS00008] The setting of The setting of the displayed screen is Check the screen setting, and restart
the displayed screen is invalid. DR-ID 300CL.
invalid. If the same phenomenon recurred, reinstall
DR-ID 300CL.
8 [IC00008] An error occurred An error occurred in starting auto- Restart DR-ID 300CL.
in starting auto-conversion conversion of patient ID. If the same phenomenon recurred, reinstall
of patient ID. DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-23
Event Event Source Description Causes Remedies Remarks
8 [OM00008] The modality The modality setting of the system is Check the modality settings of Service
setting of the system is contradictory. Utility. Restart DR-ID 300CL. If the same
contradictory. error recurs after restarting, reinstall DR-ID
300CL.
8 [SM00008] Failed in Failed in changing network settings. Restart the system.
changing network settings. If the same error recurs after restarting the
system, reinstall the system.
9 CslAdv_Delivery If the same error recurs Failed in connecting the image database. Check the setting parameter and then
after restarting the system, restart the application. If the same
contact your service phenomenon recurred, reinstall the
engineer." system.
9 [QA00009] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
9 [RE00009] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
9 [IC00009] Failed in auto- Failed in auto-conversion of patient ID. Restart DR-ID 300CL.
conversion of patient ID. If the same phenomenon recurred, reinstall
DR-ID 300CL.
9 [OM00009] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall DR-ID
300CL.
10 CslAdv_Modality catch DCMObjExException Failed in acquiring DICOM data. Finish the study, and perform the study
again. If the same phenomenon recurred,
restart the application.
10 CslAdv_Modality Failed to execute Failed in releasing process of DICOM file. Finish the study, and perform the study
DDOHelper.FreeDDOObj again. If the same phenomenon recurred,
restart the application.
10 CslAdv_Modality Failed to execute Failed in browsing the image database. Finish the study, and perform the study
ImageDBAccessor. again. If the same phenomenon recurred,
RenewImageStatus restart the application.
10 CslAdv_Modality SendSetPatientInformation Failed in acquiring DICOM data. Finish the study, and perform the study
DCMObjExException again. If the same phenomenon recurred,
restart the application.
10 [DE00010] Failed in Failed in updating the study information. Restart the system.
updating the study If the same error recurs after restarting the
information. system, reinstall the system.
10 [RE00010] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
10 CslAdv_Modality %1 %2 An error occurs inside the program. Restart the system %1: The module error
[ExceptionInfomation] If the same phenomenon recurred after occurred
%3 restarting the system, reinstall the system. %2: Spot error occurred
%3: Error details
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-24
Event Event Source Description Causes Remedies Remarks
10 [IC00010] Failed in auto- Failed in auto-conversion of patient ID. Restart DR-ID 300CL.
conversion of patient ID. If the same phenomenon recurred, reinstall
DR-ID 300CL.
10 CslAdv_Modality DCMObjExException DDO file operation has failed. Restart DR-ID 300CL. If the same %1: The record number
occured RollbackDDO. phenomenon recurred, reinstall DR-ID of the image
imageId : %1 300CL. database.
10 [OM00010] Setup is not Setup is not completed yet. Activate the application licence.
completed yet.
11 CslAdv_Modality %1 %2 Failed in acquiring sharing data. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
system. %2: Spot error occurred
%3: Error details
11 CslAdv_Modality %1 %2 Failed in acquiring sharing data. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
11 CslAdv_Modality Failed to execute Failed in deleting an internal processing Finish the study, and perform the study %1: The search key of
DataContainerMain. file. again. If the same phenomenon recurred, internal processing
Remove() - tempDDOKey restart the application. file.
= %1
11 CslAdv_Modality Failed to execute Failed in deleting an internal processing Finish the study, and perform the study %1: The search key of
DataContainerMain. file. again. If the same phenomenon recurred, internal processing
Remove() - XmlKey = %1 restart the application. file.
11 CslAdv_Modality OnImagingPreviewCreated Failed in creating an internal processing Finish the study, and perform the study
DataContainerMain.Create file. again. If the same phenomenon recurred,
Error restart the application.
11 [RE00011] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
12 CslAdv_Modality %1 %2 An error occurs inside the program. Finish the study, and perform the study %1: The module error
again. If the same phenomenon recurred, occurred
[ExceptionInfomation] restart the application. %2: Spot error occurred
%3 %3: Error details
12 [RE00012] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
12 CslAdv_Qc Can not execute inputted Failed in accessing the database. Restart QC. If the same phenomenon
query. recurred, reinstall DR-ID 300CL.
12 [DE00012] Failed in output Failed in output processing. Restart DR-ID 300CL.
processing. If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-25
Event Event Source Description Causes Remedies Remarks
13 CslAdv_Modality %1 %2 Failed in acquiring registry value. Restart the application. If the same %1: The module error
Get Registry Data Error. phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%3 %3: Error details
13 [RE00013] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
14 CslAdv_Modality %1 %2 Failed in displaying message box. Restart the application. If the same %1: The module error
Faild to Show MessageBox. phenomenon recurred, reinstall the occurred
MessageID = %3 system. %2: Spot error occurred
[ExceptionInfomation] %3: ID of the message
%4 which could not be
shown
%4: Error details
14 CslAdv_Modality SendSetPatientInformation Failed in displaying dialog box. Finish the study, and perform the study
MsgBoxExException again. If the same phenomenon recurred,
restart the application.
14 CslAdv_Modality SendSetPatientInformation Failed in displaying dialog box. Finish the study, and perform the study
MsgBoxExNotFound again. If the same phenomenon recurred,
Exception restart the application.
14 [RE00014] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
14 CslAdv_Delivery Spooler Control Image process completion notification Check the setting parameter and then %1: axSpooler.
SetImageProcComplete function returned an error to Spooler. restart DR-ID 300CL. If the same SetImageProc
Method Call Error (rc=%1) phenomenon recurred, reinstall DR-ID Complete() return
300CL. value
14 CslAdv_Delivery CreateSRProcComplete is Failed to register an event to notify the Restart DR-ID 300CL. %1: OutputID
null.(outputID=%1) completion of creating DDO for outputting If the same phenomenon recurred, reinstall
SR. DR-ID 300CL.
15 CslAdv_Modality %1 %2 Failed in initializing GUI. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
15 [RE00015] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
16 CslAdv_Modality %1 %2 Failed in notifying the change of exposure Check the Image Reader status. %1: The module error
ModalityException Msg = Faild conditions to Image Reader. occurred
to %2: Spot error occurred
cmdChangeExposureCondition. %3: Error details
Convert
[ExceptionInfomation]
%3
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-26
Event Event Source Description Causes Remedies Remarks
16 CslAdv_Modality Failed to execute Selector number: Failed in notifying to the Finish the study, and perform the study %1: Selector number
ClearPatientInformation device of %1. again. If the same phenomenon recurred,
ActiveSelector = %1 restart the application.
16 CslAdv_Modality Failed to execute selector. Selector number: Failed in notifying to the Restart DR-ID 300CL and the device. %1: Selector number
Devices[%1].Modality. device of %1.
DisableImaging
16 CslAdv_Modality Failed to execute selector. Selector number: Failed in notifying to the Finish the study, and perform the study %1: Selector number
Devices[%1].Modality. device of %1. again. If the same phenomenon recurred,
EnableImaging restart the application.
16 CslAdv_Modality Failed to execute Selector number: Failed in notifying to the Finish the study, and perform the study %1: Selector number
SetExposureCondition device of %1. again. If the same phenomenon recurred,
ActiveSelector = %1 restart the application.
16 CslAdv_Modality Failed to execute Selector number: Failed in notifying to the Finish the study, and perform the study %1: Selector number
SetPatientInfomation device of %1. again. If the same phenomenon recurred,
Selector = %1 restart the application.
16 CslAdv_Modality ModalityException Msg = Failed in acquiring the selector number. Finish the study, and perform the study %1: Selector number
Failed to again. If the same phenomenon recurred,
GetTubeTechniqueCode restart the application.
- SelectorNumber = %1
16 [RE00016] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
16 CslAdv_Modality %1 %2 An error occurred inside the program. Restart the system. %1: The module error
[ExceptionInfomation] If the same error recurred after restarting occurred
%3 the system, reinstall the system. %2: Spot error occurred
%3: Error details
16 CslAdv_Modality Failed to execute selector. Failed in notifying the Image Reader that Confirm the status of the Image Reader. %1 :The selector
Devices[%1].Modality. exposure has been disabled. number error
EnableImaging occurred
16 CslAdv_Modality Failed to execute selector. Failed in notifying the Image Reader that Confirm the status of the Image Reader. %1 :The selector
Devices[%1].Modality. exposure has been enabled. number error
DisableImaging occurred
17 CslAdv_Modality %1 %2 An error occurred when receiving Restart the application. If the same %1: The module error
[ExceptionInfomation] message. phenomenon recurred, reinstall the occurred
%3 system. %2: Spot error occurred
%3: Error details
17 CslAdv_Modality %1 %2 An error occurred when receiving Restart the application. If the same %1: The module error
Failed to get searchkey of message. phenomenon recurred, reinstall the occurred
DataContainer from system. %2: Spot error occurred
XmlMessage faild. %3: Error details
17 [RE00017] Failed in starting Failed in starting up the user utility. Restart the system.
up the user utility. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-27
Event Event Source Description Causes Remedies Remarks
18 CslAdv_Modality Failed to execute Linker. Failed in notifying the message. Finish the study, and perform the study
PostMessage() again. If the same phenomenon recurred,
restart the application.
18 CslAdv_Modality ModalityException Msg = Failed in notifying the message. Finish the study, and perform the study %1: The record number
Failed to again. If the same phenomenon recurred, of the image
SelectorChangedSend restart the application. database.
Message - imageId = %1
18 [RE00018] Failed in Failed in shutting down the console. Restart the system.
shutting down the console. If the same error recurs after restarting the
system, reinstall the system.
18 CslAdv_Modality %1 %2 An error occurs inside the program. Restart the system. %1: The module error
[ExceptionInfomation] If the same error recurred after restarting occurred
%3 the system, reinstall the system. %2: Spot error occurred
%3: Error details
19 CslAdv_Modality 1% An error occurs inside the program. Finish the study, and perform the study 1%: Spot error occurred
again. If the same phenomenon recurred,
restart the application.
19 [RE00019] Failed in Failed in sending a report to the simplified Restart the system.
sending a report to the ordering device. If the same error recurs after restarting the
simplified ordering device. system, reinstall the system.
19 CslAdv_Modality Failed to get value from Failed in acquiring the Tomosynthesis Finish the study, and perform the study %1: Tomosynthesis-
DDO:tag=%1 information from DDO. again. If the same phenomenon recurred, related tags
restart DR-ID 300CL. defined by DDO
19 CslAdv_Modality Failed to GetResouce Failed in acquiring the information from the Restart DR-ID 300CL. If the same
iReConstructCondition restructuring condition file. phenomenon recurred, reinstall DR-ID
300CL.
19 CslAdv_Modality ConvertXMLtoTempDDO:%1 Failed in processing in the program. Finish the study, and perform the study %1: DDO tag name
Failed to Convert:tag= %1 again. If the same phenomenon recurred,
restart DR-ID 300CL.
19 CslAdv_Modality CmdOnCRModeAcquisition The XML to notify the acquisition result of When the image is left in the panel, re-
Completed.Convert the image is abnormal. input the image.
xmlString Error. If the same situations occur, restart the
DR-ID 300CL.
19 CslAdv_Modality CmdOnCRModeAcquisition Failed in converting to notify the acquisition When the image is left in the panel, re- %1: XML of the notify
Completed.Convert Error. result of the image. input the image. the acquisition
ReceiveXml:[“ %1 “]” If the same situations occur, restart the result of the image
DR-ID 300CL.
19 CslAdv_Modality CmdOnCRModeAcquisition The XML to notify the transfer of the image When the image is left in the panel, re-
Received.Convert is abnormal. input the image.
xmlString Error. If the same situations occur, restart the
DR-ID 300CL.
19 CslAdv_Modality CmdOnCRModeAcquisition Failed in converting the notice of the image When the image is left in the panel, re- %1: XML of the notify
Received.Convert Error. transfer. input the image. the acquisition
ReceiveXml:[“ %1 “]” If the same situations occur, restart the result of the image
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-28
Event Event Source Description Causes Remedies Remarks
19 CslAdv_Modality CmdOnCRModeDelete The XML to notify the deletion result of the When the image is left in the panel, re-
Completed.Convert image is abnormal. input the image.
xmlString Error. If the same situations occur, restart the
DR-ID 300CL.
19 CslAdv_Modality CmdOnCRModeDelete Failed in converting the notice of the image When the image is left in the panel, re- %1: XML of the notify
Completed.Convert Error. deletion result. input the image. the acquisition
ReceiveXml:[“ %1 “]” If the same situations occur, restart the result of the image
DR-ID 300CL.
19 CslAdv_Modality CmdResCRMode The XML of the image acquisition response When the image is left in the panel, re-
AcquisitionStart.Convert is abnormal. input the image.
xmlString Error. If the same situations occur, restart the
DR-ID 300CL.
19 CslAdv_Modality CmdResCRMode Failed in converting the response of the When the image is left in the panel, re- %1: XML of the notify
AcquisitionStart.Convert image acquisition. input the image. the acquisition
Error. ReceiveXml:[“ %1 “]” If the same situations occur, restart the result of the image
DR-ID 300CL.
20 CslAdv_Modality Failed to execute Failed in retransmission processing. Finish the study, and perform the study %1: Device information
ModalityIF. again. If the same phenomenon recurred,
ReacquisitionImage restart the application.
(RETAKE) - imageKey =
%1
20 CslAdv_Modality Failed to execute ModalityIF. Failed in deleting retransmission menu. Finish the study, and perform the study %1: Device information
ReacquisitionImage(RETAKE_ again. If the same phenomenon recurred,
STOP) - imageKey = %1 restart the application.
20 CslAdv_Modality Failed to execute An error occurs inside the program. Finish the study, and perform the study
SelectorList. again. If the same phenomenon recurred,
GetImageData() restart the application.
20 CslAdv_Modality ImageData Create Error An error occurs inside the program. Finish the study, and perform the study
again. If the same phenomenon recurred,
restart the application.
20 [RE00020] Failed in Failed in accessing to database. Restart the system.
accessing to database. If the same error recurs after restarting the
system, reinstall the system.
21 CslAdv_Qc Unhandle exception An unclear system error occurred. Restart QC. If the same phenomenon
ocurred recurred, reinstall DR-ID 300CL.
21 CslAdv_Modality %1 %2 An error occurs inside the program. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
21 [RE00021] Failed in Failed in accessing to database. Restart the system.
accessing to database. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-29
Event Event Source Description Causes Remedies Remarks
22 CslAdv_Modality %1 %2 An error occurs inside the program. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
22 [RE00022] Failed in Failed in acquiring system setting Restart the system.
acquiring system setting information (IP size series). If the same error recurs after restarting the
information (IP size series). system, reinstall the system.
23 CslAdv_Modality %1 %2 Failed in checking menu condition. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
23 CslAdv_Modality catch Exception An error occurs inside the program. Finish the study, and perform the study
again. If the same phenomenon recurred,
restart the application.
23 CslAdv_Modality Failed to execute Failed in menu selecting process. Finish the study, and perform the study
MenuSelectStateCheck Menu selection is disabled in DR-ID 300CL again. If the same phenomenon recurred,
status. restart the application.
23 CslAdv_Modality Failed to execute Failed in menu selecting process. Finish the study, and perform the study
MenuUnSelectStateCheck Menu selection is disabled in DR-ID 300CL again. If the same phenomenon recurred,
status. restart the application.
23 CslAdv_Modality Failed to execute An error occurs inside the program. Finish the study, and perform the study
ReqExposurePreparation again. If the same phenomenon recurred,
CancelStateChecker. restart the application.
Check()
23 CslAdv_Modality ImageID incorrect Data in the image database is invalid. Shut down DR-ID 300CL, and check the %1: The record number
(imageIds.Length = %1 image database. of the image
database.
23 CslAdv_Modality ImageStatus of MixMenu Menu status is invalid. Finish the study, and perform the study
isn't ReadyUp. again. If the same phenomenon recurred,
restart the application.
23 [RE00023] Failed in Failed in accessing to database. Restart the system.
accessing to database. If the same error recurs after restarting the
system, reinstall the system.
23 CslAdv_Modality BreakOffStateChecker. An error occurs inside the program. Restart DR-ID 300CL. If the same
StateCheck is NG phenomenon recurred, reinstall DR-ID
300CL.
24 CslAdv_Modality %1 %2 Failed in initializing the log output function. Restart the application. If the same %1: The module error
failed logclass initilalize phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%3 %3: Error details
24 CslAdv_Modality %1 %2 Failed in outputting the log. Restart the application. If the same %1: The module error
failed output log phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%3 %3: Error details
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-30
Event Event Source Description Causes Remedies Remarks
24 [RE00024] Initialization has Initialization has not been completed. Restart the system.
not been completed. If the same error recurs after restarting the
system, reinstall the system.
25 CslAdv_Modality %1 %2 An error was notified from the Image Confirm the Image Reader. %1: The module error
ErrorNumber =%3, Reader. occurred
ErrorCode = %4, %2: Spot error occurred
ExtraMessage =%5, %3: Error character
DataSouceType = %6, string ID
messag_mode = %7, win_ %4: Error code
func = %8, selectorNo = %5: Detailed
%9, hostName = %10 description for
errors
%6: ID for identifying
the data source
which this error
character string is
stored
%7: Procedure of
specifying means of
error message
display
%8: WinFunc definition
(FRUP)
%9: Selector number
%10: Host name
25 [RE00025] Failed in logoff. Failed in logoff. Restart the system.
If the same error recurs after restarting the
system, reinstall the system.
26 CslAdv_Modality %1 %2 An error occurs inside the program. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
26 [RE00026] Failed in Failed in registering the patient information. Register the patient information again from
registering the patient the ordering system.
information.
26 CslAdv_Modality Exception occurred. Failed in processing in the program. Restart DR-ID 300CL. %1: Error details
%1 If the same error recurs after restarting,
reinstall DR-ID 300CL.
27 CslAdv_Modality %1 %2 Failed in executing utility of the Image Confirm the Image Reader. %1: The module error
%3 Reader. occurred
[ExceptionInfomation] %2: Spot error occurred
%4 %3: Error outline
%4: Error details
27 [RE00027] Failed in Failed in reserving a study. Reserve the study information again from
reserving a study. the ordering system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-31
Event Event Source Description Causes Remedies Remarks
28 [RE00028] Failed in starting Failed in starting a study automatically Reserve the study from the Ordering
a study automatically from from the Simplified Ordering. device again.
the Simplified Ordering.
29 [RE00029] Failed in starting Failed in starting a study automatically Register patient information from the
a study automatically from from the Simplified Ordering. Ordering device again.
the Simplified Ordering.
30 CslAdv_Modality 1% An error occurs inside the program. Finish the study, and perform the study 1%: Spot error occurred
again. If the same phenomenon recurred,
restart the application.
30 [RE00030] Invalid data has Invalid data has been entered. Register it from the Ordering device again.
been entered.
30 CslAdv_Modality Failed to execute Failed in initializing Image Reader Utility. Restart the system.
IMDIFUtilityInterface. If the same error recurred after restarting
Initialize the system, reinstall the system.
30 CslAdv_Modality Failed to GetResourceInfo : Failed in acquiring the resource of Image Restart the system. %1 :Dynamic resource
resource_info = %1 Reader Utility. If the same error recurred after restarting information
the system, reinstall the system. acquired from
ModalityIF.
30 CslAdv_Modality Failed to An error occurs inside the program. Restart the system.
CreateXmlResource If the same error recurred after restarting
the system, reinstall the system.
30 CslAdv_Modality Failed to execute Failed in setting the resource of Image Restart the system. %1 :Dynamic resource
IMDIFUtilityInterface. Reader Utility. If the same error recurred after restarting information to hand
QueryRequisiteResource : the system, reinstall the system. over to ModalityIF.
resource =
30 CslAdv_Modality Failed to Read Modality. Acquisition of the respective parameter Check Modality.exe.config. %1: Parameter name
exe.config from Modality.exe.config has failed. defined in Modality.
%1 value is invalid exe.config.
30 CslAdv_Modality It is already ID lock has been implemented more than Finish the study, and perform the study %1: MenuUID/
IDLocked:ImageKey = %1 once. again. If the same phenomenon recurred, RegistID/PlateID of
restart the DR-ID 300CL. image DB
30 CslAdv_Modality Failed to execute Failed in updating DR-ID 300CL condition. Restart DR-ID 300CL. If the same
ImageDBAccessor. phenomenon recurred, reinstall DR-ID
RenewMenuOperation 300CL.
30 CslAdv_Modality OnTerminationCompleted An error occurs inside the program. Restart DR-ID 300CL. If the same %1: Inside message
Error. ReceiveXml:[%1] phenomenon recurred, reinstall DR-ID
300CL.
30 CslAdv_Modality Failed to execute Failed in setting tags for DDO. Finish the study, and perform the study
ACCModImgPreProcCtrl. again. If the same phenomenon recurred,
ACCModImgPreProcVGP restart DR-ID 300CL.
Setter: Ret
30 CslAdv_Modality Failed to output Failed in setting tags for DDO. Finish the study, and perform the study
ImageProcParameter to again. If the same phenomenon recurred,
TempDDO restart DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-32
Event Event Source Description Causes Remedies Remarks
30 CslAdv_Modality Failed to execute Failed in unlocking of Xcon. Select the menu again. If the same
NotifyXconInterLock phenomenon recurred, restart DR-ID
300CL.
30 CslAdv_Modality Failed to execute selector. Failed in setting exposure conditions. Select the menu again. If the same %1: Error code
SendSetExposureCondition phenomenon recurred, restart DR-ID
ReturnValue = %1 300CL.
30 CslAdv_Modality ExposureResultFromXcon Failed in processing in the program. Restart DR-ID 300CL. If the same %1: Error code
LongTimeAccum() phenomenon recurred, reinstall DR-ID
NotifyExposureResult 300CL.
Receive() Error.
30 CslAdv_Modality LocalImageImportCtrl. Status of the image import destination Close the image library screen and confirm %1: Error code
LocalImageImportStart, menu is invalid. the menu.
StateCheckResult : From the image library screen, import
Error(%1) again.
If the same situations occur, restart the
DR-ID 300CL.
30 CslAdv_Modality LocalImageImportCtrl. Status of the image import destination %1: Error code
LocalImageImportFinished, menu is invalid or the image to import from
StateCheckResult : the library window is abnormal.
Error(%1)
30 CslAdv_Modality %1 %2 The URL of sound file is invalid. Restart DR-ID 300CL. If the same %1: Error occurred
File URL is Invalid. File phenomenon recurred, reinstall DR-ID module
=%3 300CL. %2: Error occurred part
[ExceptionInfomation] %3: Sound file name
%4 %4: Error outline
30 CslAdv_Modality %1 %2 The timeout occurred when playing the Restart DR-ID 300CL. If the same %1: Error occurred
File Load Timeout. File sound file. phenomenon recurred, reinstall DR-ID module
=%3 300CL. %2: Error occurred part
[ExceptionInfomation] %3: Sound file name
%4 %4: Error outline
30 CslAdv_Modality %1 %2 The sound file is not found. Restart DR-ID 300CL. If the same %1: Error occurred
File is not found. File =%3 phenomenon recurred, reinstall DR-ID module
[ExceptionInfomation] 300CL. %2: Error occurred part
%4 %3: Sound file name
%4: Error outline
30 CslAdv_Modality %1 %2 The selected file is not in PCM .wav Restart DR-ID 300CL. If the same %1: Error occurred
File is not PCM .wav File. format. phenomenon recurred, reinstall DR-ID module
File =%3 300CL. %2: Error occurred part
[ExceptionInfomation] %3: Sound file name
%4 %4: Error outline
30 CslAdv_Modality Failed to execute Failed in importing the XML to notify the When the image is left in the panel, re-
cmdOnCRModeAcquisition transfer of the image. input the image.
Completed.Convert If the same situations occur, restart the
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-33
Event Event Source Description Causes Remedies Remarks
30 CslAdv_Modality Failed to execute Failed in importing the XML to notify the When the image is left in the panel, re-
cmdOnCRModeAcquisition acquisition result of the image. input the image.
Completed.Convert If the same situations occur, restart the
DR-ID 300CL.
30 CslAdv_Modality Failed to execute Failed in importing the XML to notify the When the image is left in the panel, re-
cmdOnCRModeDelete deletion result of the image. input the image.
Completed.Convert If the same situations occur, restart the
DR-ID 300CL.
30 CslAdv_Modality Failed to execute Failed in importing the XML of the image When the image is left in the panel, re-
CmdResCRMode acquisition response. input the image.
AcquisitionStart.Convert If the same situations occur, restart the
DR-ID 300CL.
30 CslAdv_Modality Directory Not Found. path: The directory of the image saving When the image is left in the panel, re- %1: The directory of
%1 destination is not found. input the image. the image saving
If the same situations occur, restart the
DR-ID 300CL.
30 CslAdv_Modality Failed to Get hostName of Failed in acquiring a host name of the Restart DR-ID 300CL. If the same
5000(Cassette) 5000 series. phenomenon recurred, reinstall DR-ID
300CL.
31 [RE00031] Failed in starting Failed in starting the system. Restart the system.
the system. If the same error recurs after restarting the
system, reinstall the system.
32 [RE00032] Study condition Study condition cannot be changed. Restart DR-ID 300CL.
cannot be changed. If the same phenomenon recurred, reinstall
DR-ID 300CL.
40 CslAdv_Modality %1 %2 Failed in processing in the program. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
40 CslAdv_Modality Err[%1.NotifyOKButton] Failed in processing OK button on the Restart the application. If the same %1: Dialog window name
connected device status dialog window phenomenon recurred, reinstall the
%1. system.
40 CslAdv_Modality Failed to %1. Failed in displaying the connected device Restart the application. If the same %1: Dialog window name
NotifyVisibleScreen [param status dialog window %1. phenomenon recurred, reinstall the
showFlag=true] system.
40 CslAdv_Modality Failed to Failed in displaying the connected device Restart the application. If the same %1: Clicked indicator
ClickIndicator[param num = status window. phenomenon recurred, reinstall the number.
%1] system.
40 CslAdv_Modality Failed to DispScreen[param Failed in displaying the connected device Restart the application. If the same %1: Dialog window name
num=%1] status dialog window %1. phenomenon recurred, reinstall the
system.
40 CslAdv_Modality Failed to statusDlg. Failed in displaying the connected device Restart the application. If the same %1: The dialog window
PreparationShow[param status dialog window. phenomenon recurred, reinstall the number attempted
num =%1] system. to display.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-34
Event Event Source Description Causes Remedies Remarks
40 CslAdv_Modality Failed to Failed in displaying the connected device Restart the application. If the same %1: Dialog window name
SwitchScreen[param status dialog window %1. phenomenon recurred, reinstall the
num=%1] system.
40 CslAdv_Modality Failedt to Failed in requesting to enable/disable an Restart the application. If the same %1: Transition status of
ChangeShotReadyStatus: exposure of ShotReady. phenomenon recurred, reinstall the ShotReady
status=%1 system.
50 CslAdv_Re Unknown error when Unknown error was passed through during Restart DR-ID 300CL.
interact with plugin plugin command processing.
commands
64 CslAdv_Re Failed in applinker init A problem occurred in Reception.config Restart DR-ID 300CL. If the same
file. phenomenon recurred, reinstall DR-ID
300CL.
100 CslAdv_Re Can’t load plugin All plugins could not be imported. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
100 CslAdv_Study Update User_Operation for Failed to update User Operation Table. Restart DR-ID 300CL. If the same
menu ws combine failed (Update status of long-view stiched menu) phenomenon recurred, reinstall DR-ID
300CL.
101 CslAdv_Bridge - An error occurred while starting up Bridge. Check the setting of WebApplication.
101 CslAdv_Study Delete menu in User_ Failed to update User Operation Table. Restart DR-ID 300CL. If the same
Operation failed (Delete menu) phenomenon recurred, reinstall DR-ID
300CL.
101 CslAdv_Study Change select menu in Failed to update User Operation Table. Restart DR-ID 300CL. If the same
User_Operation failed (Update menu selection information) phenomenon recurred, reinstall DR-ID
300CL.
102 CslAdv_Bridge - An error occurred while terminating Bridge. Restart DR-ID 300CL.
103 CslAdv_Bridge - An error has occurred at the time of Restart DR-ID 300CL. If the same
Ansemble message reception. phenomenon recurred, reinstall DR-ID
300CL.
104 CslAdv_Bridge - An error has occurred during message Restart DR-ID 300CL. If the same
loop processing. phenomenon recurred, reinstall DR-ID
300CL.
105 CslAdv_Bridge - Acquisition of the settings from Restart DR-ID 300CL.
ConsoleGateway has failed.
107 CslAdv_Bridge - Data acquisition from the table held by the Restart DR-ID 300CL.
console has failed.
110 CslAdv_Re Unhandled error Unhandled exception has occurred. Restart DR-ID 300CL.
200 CslAdv_Re Failed in interacting with • _console_db does not exist. Restart DR-ID 300CL.
[_console_db] database • Connection with the SQL server was If the same phenomenon recurred, reinstall
disconnected. DR-ID 300CL.
• The SQL service has been stopped. If the same phenomenon recurred,
initialize the image database.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-35
Event Event Source Description Causes Remedies Remarks
200 ClsAdv_Tomo 1% An error occurred in QA processing. Restart DR-ID 300CL. 1%: Module name
2% If the same phenomenon recurred, reinstall where an error
DR-ID 300CL. occurred
2%: The spot of the
error and its cause
202 CslAdv_Qc Initialize QC error. Failed in initializing QC. Necessary Restart QC. If the same phenomenon
resource and components may not exist. recurred, reinstall DR-ID 300CL.
203 CslAdv_Qc Can't show magnified RAW image data is not in the selected The image for QC may be illegal. Expose a
image form. folder. new image and perform the operation
again.
231 CslAdv_Re Parameter values for query _recept_image_db does not exist. Restart DR-ID 300CL.
data from [_recept_image_ If the same phenomenon recurred, reinstall
db] is invalid DR-ID 300CL.
If the same phenomenon recurred,
initialize the image database.
232 CslAdv_Re Initialized adapter for [_ _recept_image_db does not exist. Restart DR-ID 300CL.
recept_image_db] DB If the same phenomenon recurred, reinstall
failed. DR-ID 300CL.
If the same phenomenon recurred,
initialize the image database.
233 CslAdv_Re SQL exception when • _recept_image_db does not exist. Restart DR-ID 300CL.
interact [_recept_image_ • Connection with the SQL server was If the same phenomenon recurred, reinstall
db] DB. disconnected. DR-ID 300CL.
• The SQL service has been stopped. If the same phenomenon recurred,
initialize the image database.
290 CslAdv_Re Failed in connecting to MS • NetConfig.mdb does not exist in C:\ Restart DR-ID 300CL. If the same
Access database Program Files\Fujifilm\IIP folder and phenomenon recurred, reinstall DR-ID
subfolder. 300CL.
• Character string of the connection to
Depart.mdb, Paramedic.mdb, NetConfig.
mdb does not exist in Directory.ini file of
C:\Program Files\Fujifilm\IIP\System
folder, or the path is invalid
290 CslAdv_Re Failed in accessing the An error occurred in the information read Restart DR-ID 300CL. If the same
IDMassHistory database during record insertion. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-36
Event Event Source Description Causes Remedies Remarks
291 CslAdv_Re Could not load information • _recept_image_db does not exist. Restart DR-ID 300CL.
from database • Connection with the SQL server was If the same phenomenon recurred, reinstall
disconnected. DR-ID 300CL.
• The SQL service has been stopped. If the same phenomenon recurred,
• Depart.mdb, Paramedic.mdb, NetConfig. initialize the image database.
mdb do not exist in "C:\Program Files\
Fujifilm\IIP" folder and subfolder.
• Character string of the connection to
Depart.mdb, Paramedic.mdb, NetConfig.
mdb does not exist in Directory.ini file of
"C:\Program Files\Fujifilm\IIP\System"
folder, or the path is invalid.
300 CslAdv_Re Failed in connecting to [_ • _recept_image_db does not exist. Restart DR-ID 300CL.
recept_image_db] • Connection with the SQL server was If the same phenomenon recurred, reinstall
database disconnected. DR-ID 300CL.
• The SQL service has been stopped. If the same phenomenon recurred,
initialize the image database.
301 CslAdv_Re Could not load information • _recept_image_db does not exist. Restart DR-ID 300CL.
from database • _console_db does not exist. If the same phenomenon recurred, reinstall
• Connection with the SQL server was DR-ID 300CL.
disconnected. If the same phenomenon recurred,
• The SQL service has been stopped. initialize the image database.
302 CslAdv_Re Could not store data in the • _recept_image_db does not exist. Restart DR-ID 300CL.
database. • _console_db does not exist. If the same phenomenon recurred, reinstall
• Connection with the SQL server was DR-ID 300CL.
disconnected. If the same phenomenon recurred,
• The SQL service has been stopped. initialize the image database.
302 CslAdv_Qc Failed in accessing image RAW image data is not in the selected Confirm the authority and the free space
files folder. for the destination folder to save.
303 CslAdv_Qc Measurement processing Illegal image data (incomplete data). Perform an exposure again. If the same
cannot be implemented. phenomenon recurred, reinstall DR-ID
300CL.
304 CslAdv_Qc An error occurred. Error Illegal image data. The data may not be supported. If the {0}: Error code
code: {0} For the cause of the failure of same phenomenon recurred, expose an
measurement, refer to the error code {0}. image again.
307 CslAdv_Qc The image is illegal Illegal image data. The image for QC may be illegal. Expose a
Its cause cannot be identified. new image and perform the operation
again.
308 CslAdv_Qc Save grid view settings QC application may have a component Restart QC. If the same phenomenon
occurs error problem. recurred, reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-37
Event Event Source Description Causes Remedies Remarks
310 CslAdv_Qc Terminate [QC Calculation] QC application may have a component Restart QC. If the same phenomenon
module occurs error problem. recurred, reinstall DR-ID 300CL.
311 CslAdv_Qc Delete image files when An error occurred when deleting the study Check that the path of the destination to
finish QC occurs error data. save study data for QC start exists.
314 CslAdv_Qc An error occurred. Error Illegal image data. The data may not be supported. If the {0}: Error code
code: {0} For the cause of the failure of same phenomenon recurred, take the
measurement, refer to the error code {0}. image again.
316 CslAdv_Qc Save QC data occurs error An error occurred when saving the QC Restart QC. If the same phenomenon
data. recurred, reinstall DR-ID 300CL.
An already existing file might not be
overwritten.
317 CslAdv_Qc Access Violation Exception No access authority. Restart QC. If the same phenomenon
recurred, reinstall DR-ID 300CL.
330 CslAdv_Re Could not delete Input.xml "C:\Program Files\Fujifilm\IIP\Work\BC_ Restart DR-ID 300CL. If the same
file return from BC::RisIf RIS\Input" folder does not exist. phenomenon recurred, reinstall DR-ID
300CL.
330 CslAdv_Re Could not read Input.xml "C:\Program Files\Fujifilm\IIP\Work\BC_ Restart DR-ID 300CL. If the same
file return from BC::RisIf RIS\Input" folder does not exist. phenomenon recurred, reinstall DR-ID
300CL.
332 CslAdv_Re Could not access MenuDB. The Menu database does not exist. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
340 CslAdv_Re Could not connect to PatientData does not exist. Restart DR-ID 300CL. If the same
PatientData database. phenomenon recurred, reinstall DR-ID
300CL.
360 CslAdv_Re Could not get file path of • The type of barcode card is invalid. Recheck the BCR-related system settings.
BCR data file • The data of barcode card is incorrect. Check that the barcode type is supported.
360 CslAdv_Re Could not read BCR data • The type of barcode card is invalid. Recheck the BCR-related system settings.
file • The data of barcode card is incorrect. Check that the barcode type is supported.
370 CslAdv_Re Could not get file path of • The type of magnetic card is invalid. Recheck the MCR-related system settings.
MCR data file • The data of magnetic card is incorrect. Check that the magnetic card type has
been set to the system settings.
Check that the magnetic card type is
supported.
370 CslAdv_Re Could not read MCR data • The type of magnetic card is invalid. Recheck the MCR-related system settings.
file • The data of magnetic card is incorrect. Check that the magnetic card type has
been set to the system settings.
Check that the magnetic card type is
supported.
402 CslAdv_Re SEHException content An error occurred in LoadDDO. Restart DR-ID 300CL.
406 CslAdv_Re Can't not convert data to Display format of the data is incorrect. Restart DR-ID 300CL.
display format
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-38
Event Event Source Description Causes Remedies Remarks
442 CslAdv_Re Format: • The XML declaration is missing. Check ECN.
[OrderInfo is invalid] • The file was not found.
<XmlTagName>:*** {0} • Opening the XML file has failed.
<XmlTagValue>:*** {1} • The XML format is not valid.
Invalid ESC existed. The
character position is *. {2}
Console has received some
invalid order data from RIS.
See the SEI about “Order
data checking”, and check
the data above.
Or
Invalid XML format
453 CslAdv_Re Failed in get registry key • Registry key value is invalid. Restart DR-ID 300CL. If the same
value. • The corresponding registry key does not phenomenon recurred, reinstall DR-ID
exist. 300CL.
500 CslAdv_Qc Get data from station table Failed in accessing the database. Restart QC. If the same phenomenon
failed recurred, reinstall DR-ID 300CL.
503 CslAdv_Qc Save CSV file failed An error occurred when saving the QC Hard disk may not have enough free
data. space, or you have no writable authority.
An already existing file might not be Confirm the destination folder to save.
overwritten.
507 CslAdv_QA QADBAccessor.dll Failed in browsing the image database. Confirm _console_db, and restart DR-ID
ImgDBCExWrapper 300CL.
GetInstance
510 CslAdv_Re Create report file error Information could not be saved in output. Restart DR-ID 300CL.
xml file.
630 CslAdv_Re Failed in connecting to the Failed in connecting to the reception image Restart DR-ID 300CL.
recept image database. database.
634 CslAdv_Re You do not have permission The data is incorrect. Restart DR-ID 300CL.
to delete study.
635 CslAdv_Re Selected items are invalid. Selected items are invalid. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-39
Event Event Source Description Causes Remedies Remarks
650 CslAdv_Re =+ No procUID specifed for The following problems occurred during Restart DR-ID 300CL. If the same Since the output of this
the request the auto start of studies requested from the phenomenon recurred, reinstall DR-ID event log is likely to be
+ Auto start for multiple outside. 300CL. caused by the defect on
studies is not supported • No study. SmartController, there is
• There are two or more studies. no way to deal with the
phenomenon as CSL
when it occurs.
The cause could be
attributed to the
appearing of the
following messages as
the study starting
message from
SmartController.
• When an ID identifying
the corresponding study
was not granted
• When multiple studies
were ordered to start
simultaneously
* Simultaneous starting
of multiple studies
from SmartController
is not supported under
the specification.
702 CslAdv_Qc Init MammoQC or RUQC Failed in initializing Mammo QC or RUQC. Check that the study data exist in the study
error. folder. Also check that the required
database is sufficient. The required data
are the files “Fujifilm\IIP\Param\
MammoQC.mdb”, “Fujifilm\IIP\Param\
QCItems.mdb”, and “Fujifilm\IIP\GUI\
String\LangStr.mdb”.
706 CslAdv_Qc Folder Input incorrect. The folder for DDO file loading was not Check that the study data save folder
found. exists.
706 CslAdv_Qc Cannot display Mammo QC cannot be displayed. Check that the study data exist in the study
IIPMammoQCControl folder. Also check that the required
database is sufficient. The required data
are the files “Fujifilm\IIP\Param\
MammoQC.mdb”, “Fujifilm\IIP\Param\
QCItems.mdb”, and “Fujifilm\IIP\GUI\
String\LangStr.mdb”.
920 CslAdv_Re Failed in setting the image MWMImage.mdb file does not exist in "C:\ Restart DR-ID 300CL. If the same
information Program Files\Fujifilm\IIP\MWMStudyList" phenomenon recurred, reinstall DR-ID
folder or is read-only file. 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-40
Event Event Source Description Causes Remedies Remarks
921 CslAdv_Re Failed in obtaining data • MWMConfig.mdb file does not exist in “C:\ Check ECN.
from the MWMConfig.mdb Program Files\Fujifilm\IIP\MWMStudyList”
database. folder.
• Invalid data are included in several fields
Or inside the database.
[OrderInfo is invalid]
<MdbFiledName>: {0}
<MdbFieldValue>: {1}
Length exceeded the
provisions of the field ( max
size = {2})
Console has received some
invalid order data from RIS.
See the SEI about “Order
data checking”, and check
the data above.
940 CslAdv_Re Could not load information MWMConfig.mdb file does not exist in "C:\ Restart DR-ID 300CL. If the same
from [MWMConfig.mdb] Program Files\Fujifilm\IIP\MWMStudyList" phenomenon recurred, reinstall DR-ID
folder. 300CL.
1000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
1000 CslAdv_Re Failed in loading SysConfig.mdb file does not exist in "C:\ Restart DR-ID 300CL. If the same
configuration information Program Files\Fujifilm\IIP\Config" folder. phenomenon recurred, reinstall DR-ID
from SysConfig DB 300CL.
1000 IIPUserUtility "Main()" Failed in launching the UserUtility.exe Restart the application. If the same
"Exception Notification from application. phenomenon recurred, reinstall the
UnhandledException system.
EventHandler"
1000 IIPUserUtility 1% 2% 3% • Out of memory Restart the application. If the same 1%: Spot error occurred
• Directory browsing error phenomenon recurred, reinstall the 2%: Message
IIP/Config directory and IIP/System system. 3%: Event ID
directory under Fujifilm do not exist. Event logs output by
• File browsing error UserUtility.exe are
Execution file (IIPUserUtility.exe) does not categorized to 5
exist. patterns.
• Null browsing error
Failed in launching the execution file
(IIPUserUtility.exe).
• Exception error
Program error.
1000 CslAdv_IDAutoConvert Initialize faild Failed in initializing GUI. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-41
Event Event Source Description Causes Remedies Remarks
1000 - SA name: Occurrence Memory is running out. Restart DR-ID 300CL. If the same The event log outputted
method name “Occurrence phenomenon recurred, reinstall DR-ID by UserUtility.exe is
location code: Error 300CL. divided into error logs
message” Event ID with five patterns.
1000 - SA name: Occurrence The following directories did not exist. Restart DR-ID 300CL. If the same
method name “Occurrence • Fujifilm\IIP\Config phenomenon recurred, reinstall DR-ID
location code: Error • Fujifilm\IIP\System 300CL.
message” Event ID
1000 - SA name: Occurrence The execution file (IIPUserUtility.exe) did Restart DR-ID 300CL. If the same
method name “Occurrence not exist. phenomenon recurred, reinstall DR-ID
location code: Error 300CL.
message” Event ID
1000 - SA name: Occurrence Failed in activation the execution file Restart DR-ID 300CL. If the same
method name “Occurrence (IIPUserUtility.exe). phenomenon recurred, reinstall DR-ID
location code: Error 300CL.
message” Event ID
1000 - SA name: Occurrence An exception error occurred. Restart DR-ID 300CL. If the same
method name “Occurrence phenomenon recurred, reinstall DR-ID
location code: Error 300CL.
message” Event ID
1000 ClsAdv_Tomo 1% An error occurred in QA processing. Restart DR-ID 300CL. If the same 1%: Module name
to 2% phenomenon recurred, reinstall DR-ID where an error
20854 300CL. occurred
2%: The spot of the
error and its cause
1001 FFQRP Another application There is a running Network Image Search Restart DR-ID 300CL.
instance is running. function process.
1005 CslAdv_MdIf Convert Tube code file Tube/technique code conversion file error. Restart the application. If the same
error. phenomenon recurred, reinstall the
system.
1005 CslAdv_Re Initialize MenuSel failed Failed in displaying the setting screen [E6] Restart DR-ID 300CL. If the same
for an active target. phenomenon recurred, reinstall DR-ID
300CL.
1006 CslAdv_MdIf Positioning file error. Positioning file error. Restart the application. If the same
phenomenon recurred, reinstall the
system.
1007 CslAdv_MdIf Filter definition file error. Error of filter information definition file data. Restart the application. If the same
phenomenon recurred, reinstall the
system.
1010 CslAdv_IDAutoConvert Linker.Init error Failed in initializing Linker. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1011 CslAdv_IDAutoConvert InitializeIDAutoConvert Failed in initialization. Restart DR-ID 300CL.
error If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-42
Event Event Source Description Causes Remedies Remarks
1012 CslAdv_IDAutoConvert Linker. Failed in processing to instruct Restart DR-ID 300CL.
SendWakeupComplete SendWakeupComplete. If the same phenomenon recurred, reinstall
error DR-ID 300CL.
1014 CslAdv_IDAutoConvert MyLinker. Failed in processing to instruct Restart DR-ID 300CL.
SendWakeupComplete SendWakeupComplete. If the same phenomenon recurred, reinstall
error DR-ID 300CL.
1015 CslAdv_IDAutoConvert DirectoryInfo.Initialize() Failed in acquiring the path information. Restart DR-ID 300CL.
error If the same phenomenon recurred, reinstall
DR-ID 300CL.
1020 CslAdv_IDAutoConvert matchingProc.Initialize Failed in initializing the matching Restart DR-ID 300CL.
failed processing section. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1030 CslAdv_IDAutoConvert AppSettingInfo() error Failed in reading the default setting Restart DR-ID 300CL.
information. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1031 CslAdv_IDAutoConvert AdjustDialogSize() error Failed in adjusting the dialog size. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1032 CslAdv_IDAutoConvert MatchingCommon() error Failed in initializing the common matching Restart DR-ID 300CL.
processing section. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1033 CslAdv_IDAutoConvert progressGUI() error Failed in generating the progress screen. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1034 CslAdv_IDAutoConvert ProgressGUI.Initialize error Failed in initializing the progress screen. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1040 CslAdv_IDAutoConvert Failed to execute Linker. Failed in the end processing of Linker. Restart DR-ID 300CL. %1: Completion state
End: Linker.End = %1 If the same phenomenon recurred, reinstall of Linker
DR-ID 300CL.
1060 CslAdv_IDAutoConvert GetProgressState error Failed in acquiring the progress Restart DR-ID 300CL.
information. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1061 CslAdv_IDAutoConvert ProgressTimeOutCount An retry error occurred in updating the Restart DR-ID 300CL.
error progress ratio. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1070 CslAdv_IDAutoConvert Matching process cancel. Failed in converting the matching process Restart DR-ID 300CL.
result. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1070 CslAdv_IDAutoConvert resultGUICtrl.SetResultInfo Failed in converting the matching process Restart DR-ID 300CL.
faild result. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1080 CslAdv_IDAutoConvert MatchingCancel Timeout Time-out occurred while waiting for cancel. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-43
Event Event Source Description Causes Remedies Remarks
1081 CslAdv_IDAutoConvert InitializeError: %1 Failed in initialization. Restart DR-ID 300CL. %1: Error code
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1082 CslAdv_IDAutoConvert CancelComplete Failed in matching. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
1082 CslAdv_IDAutoConvert MatchingComplete(NG): Failed in matching. Restart DR-ID 300CL. %1: The process status
%1 = %2 If the same phenomenon recurred, reinstall %2: Error code
DR-ID 300CL.
1109 CslAdv_Re Failed in Initialize Fris Unable to connect to the FRIS system. Check that the FRIS client has been set
Client up.
1310 CslAdv_Re Error from BCRisIf Data of starting study is invalid. Check the following configuration items
related to RIS connection.
• Item 15 “Examination No.Length” of “5.
CSL/IDT FUNCTION”
• Item 17 “Use Series Identification From
F-RIS” of “5. CSL/IDT FUNCTION”
1550 ClsAdv_Tomo 1% An error occurred in QA processing. Restart DR-ID 300CL. 1%: Module name
to 2% If the same phenomenon recurred, reinstall where an error
1555 DR-ID 300CL. occurred
2%: The spot of the
error and its cause
1601 CslAdv_Re General exception of An error occurred in the soft keyboard Restart DR-ID 300CL.
KenshinOrder screen. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1605 CslAdv_Re Failed in accessing into An error occurred when connecting to Restart DR-ID 300CL.
MassOrder.mdb MassOrderDB. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1614 CslAdv_Re Invalid Format Failed in displaying the screen while Restart DR-ID 300CL.
starting DR-ID 300CL. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1622 CslAdv_Re The site code mapping tag A siteCode mapping error occurred while Restart DR-ID 300CL.
is invalid starting DR-ID 300CL. If the same phenomenon recurred, reinstall
DR-ID 300CL.
1650 ClsAdv_Tomo 1% An error occurred in QA processing. Restart DR-ID 300CL. 1%: Module name
to 2% If the same phenomenon recurred, reinstall where an error
1653 DR-ID 300CL. occurred
2%: The spot of the
error and its cause
1700 [AC01700] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
1703 CslAdv_Re Access IDMassHistory,mdb An error occurred when connecting to Restart DR-ID 300CL. If the same
failed IDMassHistory.mdb. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-44
Event Event Source Description Causes Remedies Remarks
1705 CslAdv_Re Failed in accessing into • Error occurred at the change checking of Restart DR-ID 300CL. If the same
MassOrder.mdb the AutoIDOn, Choose examtype, and phenomenon recurred, reinstall DR-ID
btnExposurePart buttons. 300CL.
• Error occurred at the focus setting event on
the exposure part and mass screening
type.
1706 CslAdv_Re Format Failed in checking setting data while Modify the setting of
IssuanceIdMassOrder.dat starting DR-ID 300CL. IssuanceIdMassOrder.dat.
failure
1800 [AC01800] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
1801 [AC01801] Importing Setting information contains an invalid Change the corresponding item to the
issuance setting information value. specified value.
failed.
1802 [AC01802] Saving issuance The entry is invalid. Input the items with ● mark correctly.
setting information failed.
1803 [AC01803] Saving issuance Press the button below to close the dialog If the same error recurs after restarting the
setting information failed. and then retry. system, reinstall the system.
1853 CslAdv_Re Failed in accessing into the • Failed in accessing to the DB while starting Restart DR-ID 300CL. If the same
console image db DR-ID 300CL. phenomenon recurred, reinstall DR-ID
• A study other than those of IDMassHistory 300CL.
table was deleted.
• Failed in acquiring the latest study
comment when the study finished.
1884 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1884] [Code: the shutter processing. phenomenon recurred, reinstall DR-ID
%1] 300CL.
1885 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1885] [Code: the shutter processing. phenomenon recurred, reinstall DR-ID
%1] 300CL.
1886 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1886] [Code: the shutter processing. phenomenon recurred, reinstall DR-ID
%1] 300CL.
1887 FFMultiViewObjCom [File: ImageProcessFCR. Failed in preparing for the image display in Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1887] [Code: QA. (Show Image) phenomenon recurred, reinstall DR-ID
%1] 300CL.
1888 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1888] [Code: the shutter processing. (Show Image) phenomenon recurred, reinstall DR-ID
%1] 300CL.
1889 FFMultiViewObjCom [File: ImageProcessFCR. Failed in preparing for the image display in Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1889] [Code: QA. (Show Image2) phenomenon recurred, reinstall DR-ID
%1] 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-45
Event Event Source Description Causes Remedies Remarks
1890 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1890] [Code: the shutter processing. (Show Image2) phenomenon recurred, reinstall DR-ID
%1] 300CL.
1891 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1891] [Code: the shutter processing. phenomenon recurred, reinstall DR-ID
%1] 300CL.
1892 FFMultiViewObjCom [File: ImageProcessFCR. Failed in setting the Various Fill Density of Restart DR-ID 300CL. If the same %1: Error code
cpp] [Spot: 1892] [Code: the shutter processing. phenomenon recurred, reinstall DR-ID
%1] 300CL.
1999 MediaStorage MediaStorage:ctrlLinker_ An exception error occurred during the Restart DR-ID 300CL. %1: Exception message
ReceiveEvent:%1: operation in the media list at %1. If the same error recurs after restarting,
MediaStorage reinstall DR-ID 300CL.
1999 MediaStorage MediaStorage:subEndProc An exception error occurred during the Restart DR-ID 300CL. %1: Exception message
ess:%1:MediaStorage operation in the media list at %1. If the same error recurs after restarting,
reinstall DR-ID 300CL.
1999 MediaStorage MediaStorage:subChkRun An exception error occurred during the Restart DR-ID 300CL. %1: Exception message
DispFlg:%1:MediaStorage operation in the media list at %1. If the same error recurs after restarting,
reinstall DR-ID 300CL.
2000 CslAdv_ 1% 2% Abnormality occurred in acquiring setting Restart the application. If the same 1%: Error message
DaemonActivity value from SysConfig, etc. Operate with phenomenon recurred, reinstall the 2%: Error spot
the default value. system. (Stackinformation)
2000 CslAdv_ Initialization Failed Failed in initializing Ansemble linker. Restart the application. If the same
DaemonActivity phenomenon recurred, reinstall the
system.
2000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart the application. If the same %1: File name
Method=%3, %4 phenomenon recurred, reinstall the %2: Class name
system. %3: Method name
%4: Detail message
2001 [MI02001] Failed in the Failed in the reference of the reader Restart the system.
reference of the reader device. If the same error recurs after restarting the
device. system, reinstall the system.
2001 CslAdv_IDAutoConvert Faild to execute Failed in acquiring the file path at Restart DR-ID 300CL.
DataContainerMain.Peek DataContainer. If the same phenomenon recurred, reinstall
DR-ID 300CL.
2002 [MI02002] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
2002 CslAdv_IDAutoConvert Target file read error[%1] An error occurred in reading list file of the Confirm the file. If the error is caused by %1: File name
mass screening targets. the system, reinstall the system.
2003 [MI02003] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
2003 CslAdv_IDAutoConvert Target file remove error[%1] An error occurred in deleting the list file of
Confirm the file. If the error is caused by %1: File name
the selected studies. the system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-46
Event Event Source Description Causes Remedies Remarks
2010 CslAdv_IDAutoConvert Faild SendMessage Failed in sending the process completion Restart DR-ID 300CL.
notification. If the same phenomenon recurred, reinstall
DR-ID 300CL.
2020 CslAdv_IDAutoConvert Faild PostMessage Failed in sending the FatalError message. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
2036 MediaStorage SysDB SysConfig PatientID Failed in acquiring the value from Restart DR-ID 300CL.
Padding SysConfig.mdb and “IdPadding” of If the same phenomenon recurred, reinstall
ConfigCSL table. DR-ID 300CL.
2037 MediaStorage SysDB SysConfig PatientID Failed in acquiring the value from Restart DR-ID 300CL.
Padding SysConfig.mdb and If the same phenomenon recurred, reinstall
“ExaminationNoPadding” of ConfigCSL DR-ID 300CL.
table.
2600 [TS02600] Failed in Failed in accessing Image database due to Restart DR-ID 300CL.
accessing Image database an error in PPUIF. If the same phenomenon recurred, reinstall
due to an error in PPUIF. DR-ID 300CL.
2800 ClsAdv_Tomo An error occurred in TomoPPUCommCtrl. Check the connection to PPU server.
to dll. If the same phenomenon recurred, restart
2814 An error occurred in the connection with
DR-ID 300CL.
PPU.
2900 [TS02900] Failed in Failed in accessing the database. Restart DR-ID 300CL.
accessing the database. If the same phenomenon recurred, reinstall
DR-ID 300CL.
3000 CslAdv_IDAutoConvert idAutoConvert=null Failed in generating IDAutoConvert.dll Restart DR-ID 300CL.
instance. If the same phenomenon recurred, reinstall
DR-ID 300CL.
3000 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
to TomoImgProcReqQueueCtrl.dll. If the same phenomenon recurred, reinstall error and its cause
3008 An error occurred when initializing the DR-ID 300CL.
Tomosynthesis View.
3001 [MI03001] System error System error occurred. Restart the system.
occurred.
3001 CslAdv_IDAutoConvert cbProcessComplete=null Failed in acquiring the callback method Restart DR-ID 300CL.
address. If the same phenomenon recurred, reinstall
DR-ID 300CL.
3002 CslAdv_IDAutoConvert funcPtr=null Failed in registering the callback. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
3002 [MI03002] No response No response comes from X-ray controller. Check X-ray controller.
comes from X-ray
controller.
3003 CslAdv_IDAutoConvert SysConfigDBtoIdAuto Failed in writing the SysConfigDB value to Confirm the file. If the same phenomenon
ConvertIni() failed IdAutoConvert.ini. recurred, reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-47
Event Event Source Description Causes Remedies Remarks
3004 CslAdv_IDAutoConvert idAutoConvert.Initialize() Failed in initializing IdAutoConvert.dll. Restart DR-ID 300CL.
failed If the same phenomenon recurred, reinstall
DR-ID 300CL.
3010 CslAdv_IDAutoConvert resultInfo=null Failed in generating the matching result Restart DR-ID 300CL.
storage region. If the same phenomenon recurred, reinstall
DR-ID 300CL.
3115 FFQRP Unkown error An unknown error has occurred when error Restart DR-ID 300CL.
message dialog box appears.
3202 ClsAdv_Tomo Access DDO tag " + An error occurred when acquiring the Check whether the DDO files exist. If DDO
strNumberTag + " is failed values of DDO tags. files exist, check whether tags exist and
whether the data are valid.
3205 FFQRP Communication initialize The config file does not exist or is Reinstall Network Image Search function.
fail damaged.
3205 FFQRP Communication: Bind fail The config file does not exist or is Reinstall Network Image Search function.
damaged.
3205 FFQRP Unknown Error! The config file does not exist or is Reinstall Network Image Search function.
damaged.
3410 FFQRP Can't get language from The config file does not exist or is Reinstall Network Image Search function.
config file damaged.
3410 FFQRP AppName key in file QRP. The config file does not exist or is Reinstall Network Image Search function.
exe.config isn't found or damaged.
value is empty
3410 FFQRP Monitor Lut file isn't found The monitor LUT file does not exist or is Reinstall all the linking applications and
damaged. Cooper which have been installed.
3410 FFQRP Thread worker to PRINT The monitor LUT file does not exist or is Reinstall all the linking applications and
fail. damaged. Cooper which have been installed.
3410 FFQRP Unknown Error! The monitor LUT file does not exist or is Reinstall all the linking applications and
damaged. Cooper which have been installed.
3425 FFQRP Initiate QRP failed! The config file does not exist or is Reinstall Network Image Search function.
damaged.
3425 FFQRP Could not load the config The config file does not exist or is Reinstall Network Image Search function.
file damaged.
3430 FFQRP Unknown Error! The config file does not exist or is Reinstall Network Image Search function.
damaged.
3435 FFQRP Could not load the config The config file does not exist or is Reinstall Network Image Search function.
file damaged.
3435 FFQRP Unknown Error! The config file does not exist or is Reinstall Network Image Search function.
damaged.
3440 FFQRP Could not load the config The config file does not exist or is Reinstall Network Image Search function.
file damaged.
3440 FFQRP Unknown Error! The config file does not exist or is Reinstall Network Image Search function.
damaged.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-48
Event Event Source Description Causes Remedies Remarks
3450 FFQRP Can not get data from The config file does not exist or is Reinstall Network Image Search function.
config file damaged.
3490 FFQRP The string with Id={0} is not The config file does not exist or is Reinstall Network Image Search function.
found in the language file. damaged.
StackStrace: {1}
3595 FFQRP VersionInformation: The The config file does not exist or is Reinstall all the linking applications and
Application element with damaged. Cooper which have been installed.
ID={0} in the version file
can't be load. StackTrace:
{1}
3600 FFQRP VersionInformation: The The config file does not exist or is Reinstall all the linking applications and
string with Id={0} in the damaged. Cooper which have been installed.
language file can't be load.
StackTrace: {1}
3600 FFQRP Config TargetView is invalid The config file does not exist or is Reinstall Network Image Search function.
damaged.
3600 FFQRP AppName key in file QRP. The config file does not exist or is Reinstall Network Image Search function.
exe.config isn't found or damaged.
value is empty
4000 CslAdv_IDAutoConvert OnRequestEvent=NULL Failed in sending the event that the Restart DR-ID 300CL.
prematch button is clicked. If the same phenomenon recurred, reinstall
DR-ID 300CL.
4010 CslAdv_IDAutoConvert OnRequestEvent=NULL Failed in sending the event that the real Restart DR-ID 300CL.
matching button is clicked. If the same phenomenon recurred, reinstall
DR-ID 300CL.
4020 CslAdv_IDAutoConvert GuiAdpException Failed in the GuiAdp processing when the Restart DR-ID 300CL.
file selection button is clicked. If the same phenomenon recurred, reinstall
DR-ID 300CL.
5000 CslAdv_IDAutoConvert [%1] The definition file for result displaying list Confirm the file. If the error is caused by %1: File path
does not exist. the system, reinstall the system.
5001 CslAdv_IDAutoConvert mainItemDataTable=null Failed in generating the data for result Restart DR-ID 300CL.
displaying list (DataTable). If the same phenomenon recurred, reinstall
DR-ID 300CL.
5002 CslAdv_IDAutoConvert mainItemDataSet=null Failed in generating the data for result Restart DR-ID 300CL.
displaying list (DataSet). If the same phenomenon recurred, reinstall
DR-ID 300CL.
5003 CslAdv_IDAutoConvert SetupMatchingName failed Failed in initializing the screen for Restart DR-ID 300CL.
displaying the processing result. If the same phenomenon recurred, reinstall
DR-ID 300CL.
5004 CslAdv_IDAutoConvert catch GuiAdpException Failed in retrieving the strings for updating Restart DR-ID 300CL.
the result displaying list. If the same phenomenon recurred, reinstall
DR-ID 300CL.
5005 CslAdv_IDAutoConvert catch Exception An operation error of the result displaying Restart DR-ID 300CL.
list occurred. If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-49
Event Event Source Description Causes Remedies Remarks
5010 CslAdv_IDAutoConvert mainItemDataTable=null An argument error occurred while Restart DR-ID 300CL.
registering the matching result (registering If the same phenomenon recurred, reinstall
the list). DR-ID 300CL.
5020 CslAdv_IDAutoConvert GuiAdpException A GuiAdp error occurred while retrieving Restart DR-ID 300CL.
the matching type. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6000 CslAdv_IDAutoConvert ArgumentException An argument error occurred while reading Restart DR-ID 300CL.
the XML file for requesting a processing. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6001 CslAdv_IDAutoConvert XmlException Failed in processing the XML file in reading Restart DR-ID 300CL.
the XML file for requesting a processing. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6010 CslAdv_IDAutoConvert ArgumentException An argument error occurred while Restart DR-ID 300CL.
generating the matching result file. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6011 CslAdv_IDAutoConvert XmlException Failed in processing the XML file in Restart DR-ID 300CL.
generating the matching result file. If the same phenomenon recurred, reinstall
DR-ID 300CL.
6384 MediaStorRtrv FFIIPMedia 0008, 0060 Failed in acquiring the Modality (0008, Restart DR-ID 300CL.
GetString Err Return 0060). If the same error recurs after restarting,
Method (FcStorProc:: reinstall DR-ID 300CL.
ImageFilering)
6385 MediaStorRtrv FFIIPMedia DDO FILE The Modality (0008, 0060) tag does not Restart DR-ID 300CL.
INVALID Method exist. If the same error recurs after restarting,
(FcStorProc:: reinstall DR-ID 300CL.
ImageFilering)
6386 MediaStorRtrv FFIIPMedia 0018,1401 Failed in acquiring the MPM code (0018, Restart DR-ID 300CL.
GetString Err Return 1401). If the same error recurs after restarting,
Method (FcStorProc:: reinstall DR-ID 300CL.
ImageFilering)
6387 MediaStorRtrv FFIIPMedia DDO FILE The MPM code (0018, 1401) tag does not Restart DR-ID 300CL.
INVALID Method exist. If the same error recurs after restarting,
(FcStorProc:: reinstall DR-ID 300CL.
ImageFilering)
6388 MediaStorRtrv FFIIPMedia fcMenuDB. Failed in accessing the MenuData.mdb. Restart DR-ID 300CL.
Connect() Err Method If the same error recurs after restarting,
(FcStorProc:: reinstall DR-ID 300CL.
ImageFilering)
6389 MediaStorRtrv FFIIPMedia MPM Code no The MPM Code does not exist in Restart DR-ID 300CL.
Exist Method (FcStorProc:: MenuData.mdb. If the same error recurs after restarting,
ImageFilering) reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-50
Event Event Source Description Causes Remedies Remarks
7000 CslAdv_IDAutoConvert NullReferenceException Failed in converting the config information Confirm the system settings. If the problem
numerical value. is found in the system, reinstall DR-ID
300CL.
7000 CslAdv_IDAutoConvert FormatException Failed in converting the config information Confirm the system settings. If the problem
numerical value. is found in the system, reinstall DR-ID
300CL.
7001 CslAdv_IDAutoConvert GetSettingValue exception An error occurred while checking the upper Confirm the system settings. If the problem
limit of the Config value. is found in the system, reinstall DR-ID
300CL.
7001 CslAdv_IDAutoConvert ProgressRatioOfPreIn An error occurred while checking the upper Confirm the system settings. If the problem
MatchingKey NG >max limit of the Config value. is found in the system, reinstall DR-ID
300CL.
7002 CslAdv_IDAutoConvert ProgressRatioOfPreIn An error occurred while checking the lower Confirm the system settings. If the problem
MatchingKey NG <min limit of the Config value. is found in the system, reinstall DR-ID
300CL.
8000 CslAdv_IDAutoConvert GuiAdpException An error occurred in the GuiAdp conversion Restart DR-ID 300CL.
processing. If the same phenomenon recurred, reinstall
DR-ID 300CL.
8010 CslAdv_IDAutoConvert GuiAdpException An error occurred in the GuiAdp conversion Restart DR-ID 300CL.
processing. If the same phenomenon recurred, reinstall
DR-ID 300CL.
8020 CslAdv_IDAutoConvert Monitor1Resolution An error occurred while converting the Restart DR-ID 300CL.
screen resolution. If the same phenomenon recurred, reinstall
DR-ID 300CL.
8021 CslAdv_IDAutoConvert SysDBCException An error occurred while retrieving the Restart DR-ID 300CL.
resolution in system settings. If the same phenomenon recurred, reinstall
DR-ID 300CL.
9100 CslAdv_IDAutoConvert GetDirectoryString(DirKey Failed in acquiring the config directory Restart DR-ID 300CL.
Config) path. If the same phenomenon recurred, reinstall
DR-ID 300CL.
9101 CslAdv_IDAutoConvert GetDirectoryString(DirKey Failed in obtaining the log file path. Restart DR-ID 300CL.
Log) If the same phenomenon recurred, reinstall
DR-ID 300CL.
9102 CslAdv_IDAutoConvert GetDirectory(DirKeyIDAC Failed in acquiring the working directory Restart DR-ID 300CL.
Work) path. If the same phenomenon recurred, reinstall
DR-ID 300CL.
9999 CslAdv_Delivery DeviceStateListData Exception occurred when converting Check the setting parameter and then {0}: Exception message
ChangeAttrString Failed.\r\ device attribute value of the database to restart the application. If the same {1}: Stack trace
n{0}\r\n{1} the character string for display. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery DeviceStateListData An exception occurred when converting Check the setting parameter and then {0}: Exception message
ChangeStatusString device status of the database to character restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} string for display. phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-51
Event Event Source Description Causes Remedies Remarks
9999 CslAdv_Delivery DeviceStateListData An exception occurred when acquiring Check the setting parameter and then {0}: Exception message
GetListData Failed.\r\n{0}\r\ data for display of device status list. restart the application. If the same {1}: Stack trace
n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile An exception occurred during reading Check the setting parameter and then {0}: Exception message
LoadConfigFile Failed.\r\ process of XML file. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile An exception occurred during saving Check the setting parameter and then {0}: Exception message
SaveConfigFile Failed.\r\ process of XML file. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. NULL value or blank letter was notified in Check the setting parameter and then
LoadBaseConfig - Not Font name. restart the application. If the same
Defined <FontName> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of the root node Check the setting parameter and then
LoadBaseConfig - Not returned Not Found. restart the application. If the same
Found <FontSize> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of Alignment Check the setting parameter and then
LoadColumns - Not Found returned Not Found. restart the application. If the same
<Column><Alignment> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of MatchingType Check the setting parameter and then
LoadColumns - Not Found returned Not Found. restart the application. If the same
<Column><MatchingType> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of Name Check the setting parameter and then
LoadColumns - Not Found returned Not Found. restart the application. If the same
<Column><Name> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of Width Check the setting parameter and then
LoadColumns - Not Found returned Not Found. restart the application. If the same
<Column><Width> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. An exception occurred during reading Check the setting parameter and then {0}: Exception message
LoadColumns - Parse process of Columns. restart the application. If the same {1}: Stack trace
Failed phenomenon recurred, reinstall the
<Column><Alignment>\r\ system.
n{0}\r\n{1}
9999 CslAdv_Delivery ListViewConfigFile. An exception occurred during reading Check the setting parameter and then {0}: Exception message
LoadColumns - Parse process of Columns. restart the application. If the same {1}: Stack trace
Failed <Column> phenomenon recurred, reinstall the
<MatchingType>\r\n{0}\r\ system.
n{1}
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-52
Event Event Source Description Causes Remedies Remarks
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of MatchingType Check the setting parameter and then
LoadQuery - Not Defined returned Not Found. restart the application. If the same
<QueryName> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery ListViewConfigFile. SelectSingleNode method of MaxCount Check the setting parameter and then
LoadQuery - Not Found returned Not Found. restart the application. If the same
<MaxCount> phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery Not Found RootNode SelectSingleNode method of the root node Check the setting parameter and then
returned Not Found. restart the application. If the same
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputDB Close Failed.\r\ An exception occurred in database Check the setting parameter and then {0}: Exception message
n{0}\r\n{1} disconnecting process. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputDB Connect Failed.\ An exception occurred in database Check the setting parameter and then {0}: Exception message
r\n{0}\r\n{1} connecting process. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputDB Dispose Failed.\ An exception occurred in Dispose (Dispose Check the setting parameter and then {0}: Exception message
r\n{0}\r\n{1} flag) process. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputDB An exception occurred in data set acquiring Check the setting parameter and then {0}: Exception message
OpenSQLDataset Failed.\r\ process. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An exception occurred in process of Check the setting parameter and then {0}: Exception message
ChangeStatus Failed.\r\ changing to status display. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An exception occurred in database Check the setting parameter and then {0}: Exception message
dependency_OnChange changing process. restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator Click event of the indicator label is not set Check the setting parameter and then
IndicatorLabel_Click in the upper order. restart the application. If the same
ClickIndicator is null. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An exception occurred in click event Check the setting parameter and then {0}: Exception message
IndicatorLabel_Click processing. restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-53
Event Event Source Description Causes Remedies Remarks
9999 CslAdv_Delivery OutputIndicator Initialize An exception occurred in initialization. Check the setting parameter and then {0}: Exception message
Failed.\r\n{0}\r\n{1} restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator Initialize An exception occurred in GUI initialization. Check the setting parameter and then {0}: Exception message
GuiAdp Error.\r\n{0}\r\n{1} restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator NULL value or blank letter was notified to Check the setting parameter and then
InitSqlDependency SQL server connection character string. restart the application. If the same
ConnectionString is illegal. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An exception occurred in initialization of Check the setting parameter and then {0}: Exception message
InitSqlDependency Failed.\ SqlDependency. restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An operational exception occurred in Check the setting parameter and then {0}: Exception message
InitSqlDependency initialization of SqlDependency. restart the application. If the same {1}: Stack trace
InvalidOperationException.\ phenomenon recurred, reinstall the
r\n{0}\r\n{1} system.
9999 CslAdv_Delivery OutputIndicator An SQL exception occurred in initialization Check the setting parameter and then {0}: Exception message
InitSqlDependency of SqlDependency. restart the application. If the same {1}: Stack trace
SqlException.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator An exception occurred in setting process Check the setting parameter and then {0}: Exception message
SetSqlDependency Failed.\ of SqlDependency. restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator Terminate An exception occurred in completion Check the setting parameter and then {0}: Exception message
Failed.\r\n{0}\r\n{1} processing. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputIndicator Terminate An SQL exception occurred in completion Check the setting parameter and then {0}: Exception message
SqlException.\r\n{0}\r\n{1} processing. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputQueueListData An exception occurred in process of Check the setting parameter and then {0}: Exception message
ChangeDateString Failed.\ converting StudyDate value to date restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} character string for display. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputQueueListData An exception occurred in acquiring data for Check the setting parameter and then {0}: Exception message
GetListData Failed.\r\n{0}\r\ display of output queue list. restart the application. If the same {1}: Stack trace
n{1} phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-54
Event Event Source Description Causes Remedies Remarks
9999 CslAdv_Delivery OutputScreen An exception occurred in deleting process. Check the setting parameter and then {0}: Exception message
DeleteButton_Click Failed.\ restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in transition to Check the setting parameter and then {0}: Exception message
DevStsGroupButton_Click device status display. restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen DevStsList_ Device status display list control error was Check the setting parameter and then {0}: Notification
OnErrEvent.\r\n{0} notified. restart the application. If the same message
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen Initialize An exception occurred in initialization. Check the setting parameter and then {0}: Exception message
Failed.\r\n{0}\r\n{1} restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in message client Check the setting parameter and then {0}: Exception message
InitializeMsgClient Failed.\r\ initialization. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen Failed in message client initialization. Check the setting parameter and then
InitializeMsgClient restart the application. If the same
MsgClient.Init Failed. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in notification Check the setting parameter and then {0}: Exception message
NextPageButton_Click processing of page forward button restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} pressing. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in notification Check the setting parameter and then {0}: Exception message
NotifyVisibleScreen Failed.\ process of display/nondisplay switching. restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in process of Check the setting parameter and then {0}: Exception message
OutQueGroupButton_Click transition to output queue display. restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen OutQueList_ Output queue display list control error was Check the setting parameter and then {0}: Notification
OnErrEvent.\r\n{0} notified. restart the application. If the same message
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen OutQueList_ An exception occurred in notification Check the setting parameter and then {0}: Exception message
Scroll Failed.\r\n{0}\r\n{1} processing of output queue display list restart the application. If the same {1}: Stack trace
control scroll. phenomenon recurred, reinstall the
system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-55
Event Event Source Description Causes Remedies Remarks
9999 CslAdv_Delivery OutputScreen OutQueList_ An exception occurred in change Check the setting parameter and then {0}: Exception message
SelectionChanged Failed.\ notification processing of output queue restart the application. If the same {1}: Stack trace
r\n{0}\r\n{1} display list control selecting status. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen Post processing of sending message Check the setting parameter and then
PostMessage Failed. returned an error. restart the application. If the same
phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in sending process Check the setting parameter and then {0}: Exception message
PostMessage Failed.\r\n{0}\ of re-output/delete message. restart the application. If the same {1}: Stack trace
r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in notification Check the setting parameter and then {0}: Exception message
PrevPageButton_Click processing of page backward button restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} pressing. phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in message Check the setting parameter and then {0}: Exception message
ReceiveMsgFunc Error.\r\ receiving process. restart the application. If the same {1}: Stack trace
n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen An exception occurred in display updating Check the setting parameter and then {0}: Exception message
RefreshButton_Click process. restart the application. If the same {1}: Stack trace
Failed.\r\n{0}\r\n{1} phenomenon recurred, reinstall the
system.
9999 CslAdv_Delivery OutputScreen RetryButton_ An exception occurred in re-outputting Check the setting parameter and then {0}: Exception message
Click Failed.\r\n{0}\r\n{1} process. restart the application. If the same {1}: Stack trace
phenomenon recurred, reinstall the
system.
9999 CslAdv_RDPClient Form Load Filed in starting the general-purpose RIS Restart DR-ID 300CL. {0}: Exception message
Error.¥r¥n{0}¥r¥n{1}¥n browser. If the same error recurs after restarting, {1}: Stack trace
reinstall DR-ID 300CL.
9999 CslAdv_RDPClient Remote Desktop ActiveX Failed in connecting the remote. Restart the general-purpose RIS browser. {0}: Exception message
Control If the same error recurs after restarting, {1}: Stack trace
Error.¥r¥n{0}¥r¥n{1}¥n reinstall DR-ID 300CL.
10000 [MD10000] An error has An error occurred in completion processing Restart the system.
occurred in completion of the Image Reader. If the same error recurs after restarting the
processing of the Image system, reinstall the system.
Reader.
10000 [QA10000] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10000 FFQRP Another application Network Image Search function already Restart DR-ID 300CL.
instance is running. runs.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-56
Event Event Source Description Causes Remedies Remarks
10000 CslAdv_RDPClient Loading " + %1 + " is failed. Failed in reading the setting file (XML). Restart the general-purpose RIS browser. {0}: Exception message
RDPClient have set the If the error recurs restarting the general- {1}: Stack trace
default purpose RIS browser, restart DR-ID XML file name is
value.¥r¥n¥n{0}¥r¥n{1} 300CL. entered in the %1.
If the same error recurs after restarting,
reinstall DR-ID 300CL.
10001 CslAdv_Re Send message fail. Connection to firewall has been Restart DR-ID 300CL.
disconnected.
10001 [MD10001] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information image database. If the same error recurs after restarting the
from the image database. system, reinstall the system.
10001 [QA10001] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10001 [RE10001] Failed in Failed in entering information. Enter proper information in the input item
entering information. that comes with a mark.
10001 [SM10001] The hard disk is The hard disk is full. Delete unnecessary images from the list.
full.
10001 [ST10001] An error that An error that disables to continue operation Restart the system.
disables to continue occurred. If the same error recurs after restarting the
operation occurred. system, reinstall the system.
10001 FFIdAutoConvert * A method name and a An unrecoverable error occurred in the DR-ID 300CL restarts automatically. * This part is
failed process name are process. Restart DR-ID 300CL if it has frozen. If the appropriated from
output depending on the same phenomenon recurred, reinstall the V4.0.
error spot. system.
10001 [BR10001] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall DR-ID
300CL.
10001 CslAdv_RDPClient Other Reason. Failed to connect to the terminal. Press the reconnection button again, or
Check the network settings The following reasons may be considered. restart the general-purpose RIS browser. If
on the remote terminal and • The network settings on the terminal there is still no change, implement the
the console terminal. which is installed DR-ID 300CL, as well followings.
Please testing connectivity as the network settings on the remote 1. Review the power for the terminal with
with the remote terminal terminal are incorrect. DR-ID 300CL installed and the terminal
and console. • There is a problem with the status of the connected with remote, and review the
network cable and power of PC. network cables and network settings.
2. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
3. Restart DR-ID 300CL and the general-
purpose RIS browser.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-57
Event Event Source Description Causes Remedies Remarks
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Check the network settings It is possible that the network settings on restart the general-purpose RIS browser. If connection failure
on the remote terminal and the terminal which is installed DR-ID there is still no change, implement the acquired from the
the console terminal. 300CL, as well as the network settings on followings. controller is input in the
Please testing connectivity the remote terminal are incorrect. 1. Review the network settings for the [Disconnection reason]
with the remote terminal terminal with DR-ID 300CL installed and field automatically.
and console. the terminal connected with remote.
Restart DR-ID 300CL and the general-
purpose RIS browser.
2. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Check the network settings It is possible that the network settings on restart the general-purpose RIS browser. If connection failure
on the remote terminal. the remote terminal are incorrect. there is still no change, implement the acquired from the
Please testing connectivity followings. controller is input in the
with the remote terminal 1. Review the network settings for the [Disconnection reason]
and console. terminal connected with remote. Restart field automatically.
DR-ID 300CL and the general-purpose
RIS browser.
2. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Check the configuration of It is possible that the settings (Ris IP restart the general-purpose RIS browser. If connection failure
the IPAddress in Address) in Service Utility are incorrect. there is still no change, implement the acquired from the
ServiceUtility. Please followings. controller is input in the
testing connectivity with the 1. Review the settings for service utility (Ris [Disconnection reason]
remote terminal. IP Address). Restart DR-ID 300CL and field automatically.
the general-purpose RIS browser.
2. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Check the configuration of It is possible that the settings in UserUtility restart the general-purpose RIS browser. If connection failure
user account and password (user account and password of the RIS there is still no change, implement the acquired from the
in UserUtility. terminal) are incorrect. followings. controller is input in the
1. Review the settings for user utility (user [Disconnection reason]
account and password of the RIS field automatically.
terminal). Restart DR-ID 300CL and the
general-purpose RIS browser.
2. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-58
Event Event Source Description Causes Remedies Remarks
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Restart the system. and, Memory is running out and it is required to restart the general-purpose RIS browser. If connection failure
please to perform a re- restart the PC. there is still no change, restart PC. acquired from the
connection. In addition, check whether the connection controller is input in the
of Windows remote desktop is enable or [Disconnection reason]
not. field automatically.
10001 CslAdv_RDPClient [Disconnection reason] Failed to connect to the terminal. Press the reconnection button again, or The reason for remote
Check the status of the It is possible that there is a problem with restart the general-purpose RIS browser. If connection failure
network cable and power of the status of the network cable and power there is still no change, implement the acquired from the
PC, please to perform a of PC. followings. controller is input in the
re-connection. 1. Review the network cables and network [Disconnection reason]
settings. for the terminal with DR-ID field automatically.
300CL installed and the terminal
connected with remote.
2. Restart DR-ID 300CL and the general-
purpose RIS browser.
3. Make sure the remote can connect to
the terminal that the remote needs to
connect to from the Windows remote
desktop connection.
10001 [OI10001] An error An error occurred. Close the image library screen and open
occurred. again. If same errors occur, restart the
system.
10001 [TD10001] An error An error occurred. Click the button to restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10001 CslAdv_Study 1% Failed in operating data. Restart the application. If the same 1%: Error function
to phenomenon recurred, reinstall the name
19999 system. Excluding No. 1 to
44 IDs.
10001 CslAdv_Study * A method name and a Failed in operating data. Restart DR-ID 300CL.
to failed process name are If the same error recurs after restarting,
19999 output depending on the reinstall DR-ID 300CL.
error spot.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-59
Event Event Source Description Causes Remedies Remarks
10002 CslAdv_Modality %1 %2 An error was notified from the Image Confirm the Image Reader. %1: The module error
ErrorNumber =%3, Reader. occurred
ErrorCode = %4, %2: Spot error occurred
ExtraMessage =%5, %3: Error character
DataSouceType = %6, string ID
messag_mode = %7, win_ %4: Error code
func = %8, selectorNo = %5: Detailed
%9, hostName = %10 description for
errors
%6: ID for identifying
the data source
which this error
character string ID
is stored
%7: Procedure of
specifying means of
error message
display
%8: WinFunc definition
(FRUP)
%9: Selector number
%10: Host name
10002 [MD10002] Failed in setting Failed in setting the information to the Restart the system.
the information to the image database. If the same error recurs after restarting the
image database. system, reinstall the system.
10002 [QA10002] Failed in starting Failed in starting QA. Restart the system.
QA. If the same error recurs after restarting the
system, reinstall the system.
10002 [RE10002] Failed in reading Failed in reading the magnetic card. Check the card.
the magnetic card.
10002 [ST10002] Failed in Failed in receiving information from the Restart the system.
receiving information from magnetic card reader or barcode reader. If the same error recurs after restarting the
the magnetic card reader or system, reinstall the system.
barcode reader.
10002 ClsAdv_Tomo 1% An error occurred when changing image Preform the operation again. If the same 1%: The spot of the
types (exposed image or re-exposed phenomenon recurred, restart DR-ID error and its cause
image). 300CL.
10002 [SM10002] The remaining The remaining battery charge of the PC is Stop operation and recharge the battery.
battery charge of the low.
Console PC is low.
10002 [BR10002] Failed in linking Failed in linking to the patient information Check the patient information for the study
to the patient information display terminal. that is currently in process and for the
display terminal. study that is displayed on the patient
information display terminal.
10002 [OI10002] An error An error occurred. Restart the system.
occurred.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-60
Event Event Source Description Causes Remedies Remarks
10002 [TD10002] Failed in Failed in initialization. Restart the application. If the same
initialization. phenomenon recurred, reinstall the
system.
10003 CslAdv_Modality %1 %2 Failed in processing in the program. Restart the application. If the same %1: The module error
%3 phenomenon recurred, reinstall the occurred
[ExceptionInfomation] system. %2: Spot error occurred
%4 %3: Error outline
%4: Error details
10003 [MD10003] Failed in issuing Failed in issuing the image information. Restart the system.
the image information. If the same error recurs after restarting the
system, reinstall the system.
10003 [QA10003] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10003 [RE10003] Failed in reading Failed in reading the magnetic card. Swipe the card again or check the contents
the magnetic card. of the card.
10003 [ST10003] Failed in setting Failed in setting the exposure result value. Restart the system.
the exposure result value. If the same error recurs after restarting the
system, reinstall the system.
10003 [SM10003] It has passed It has passed %s hours since the Console Restart the system. %s: Time
%s hours since the Console application started. To continue the operations, click the button
application started. displayed on the screen.
10003 ClsAdv_Tomo 1% An error occurred when changing image Preform the operation again. If the same 1%: The spot of the
types (exposed image or re-exposed phenomenon recurred, restart DR-ID error and its cause
image). 300CL.
10003 [TS10003] Failed in starting Failed in starting to retrieve images from Restart DR-ID 300CL.
to retrieve images from PPU. If the same phenomenon recurred, reinstall
PPU. DR-ID 300CL.
10003 CslAdv_Re Send report pps is failed Has problem with when send report PPS Restart DR-ID 300CL.
or process update studystatus_ris_outflag
is failed.
10003 [OI10003] Failed in Failed in thumbnail creation process for the Restart the system.
thumbnail creation process mage library.
for the image library.
10003 [TD10003] Failed in Failed in terminating processing. Restart the application. If the same
terminating processing. phenomenon recurred, reinstall the
system.
10004 CslAdv_Modality Remaining battery power is Remaining battery power is low. Charge the battery.
low
10004 [MD10004] Failed in Failed in obtaining the image information. Restart the system.
obtaining the image If the same error recurs after restarting the
information. system, reinstall the system.
10004 [QA10004] Failed in Failed in acquiring study information of the Restart the system.
acquiring study information image. If the same error recurs after restarting the
of the image. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-61
Event Event Source Description Causes Remedies Remarks
10004 [RE10004] Failed in reading Failed in reading the barcode. Try reading operation again.
the barcode.
10004 [ST10004] Failed in setting Failed in setting information for the image Restart the system.
information for the image database. If the same error recurs after restarting the
database. system, reinstall the system.
10004 ClsAdv_Tomo 1% An error occurred when sending the image. Preform the operation again. If the same 1%: The spot of the
phenomenon recurred, restart DR-ID error and its cause
300CL.
10004 [SM10004] Failed in Failed in resetting the wireless LAN Restart the system.
resetting the wireless LAN adapter. If the same error recurs after restarting the
adapter. system, reinstall the system.
10004 [TD10004] Failed in Failed in transmitting the study information Click the button and continue the
transmitting the study to the workstation. operation.
information to the
workstation.
Patient Name:%s
Study Date:%s
Study Time:%s
10005 CslAdv_Modality If the same error recurs Retransmission execution has been Execute retransmission. %1: The module error
after restarting the system, selected. occurred
contact your service %2: Spot error occurred
engineer." %3: Selector number
%4: Menu UID
10005 CslAdv_Modality No remaining battery No remaining battery power. Charge the battery.
power.
10005 [MD10005] Failed in setting Failed in setting the image information. Restart the system.
the image information. If the same error recurs after restarting the
system, reinstall the system.
10005 [QA10005] Failed in image Failed in image rotation. Restart the system.
rotation. If the same error recurs after restarting the
system, reinstall the system.
10005 [RE10005] Failed in Failed in acquiring information from the Restart the system.
acquiring information from system setting. If the same error recurs after restarting the
the system setting. system, reinstall the system.
10005 [ST10005] Failed in Failed in entering information. Input information is invalid.
entering information. [*]Enter proper information in the input item
that comes with a mark.
10005 ClsAdv_Tomo 1% An error occurred when sending the image. Preform the operation again. If the same 1%: The spot of the
phenomenon recurred, restart DR-ID error and its cause
300CL.
10005 [QC10005] Failed in Failed in accessing image files. Click the close button on the QC execution
accessing image files. window.
10005 [SM10005] Failed in Failed in wireless LAN connection. Move to a place with good radio wave
wireless LAN connection. reception, and execute network recovery.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-62
Event Event Source Description Causes Remedies Remarks
10005 [TD10005] Canceling the Sending is about to be cancelled Click the button and continue the
transmission will lose the operation.
study information.
10006 [MD10006] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10006 ClsAdv_Tomo Failed in processing finish An error occurred when copying the Check the connection to PPU server. 1%: The spot of the
copy one image images from PPU to console. error and its cause
10006 [QC10006] Measurement Measurement processing cannot be Arrange the phantom position and perform
processing cannot be implemented. an exposure again.
implemented.
10006 [SM10006] The hard disk is The hard disk is full. Delete unnecessary images from the list.
full.
10007 CslAdv_Modality Remaining capacity disk Required hard disk space is insufficient. Check the free space of the hard disk.
error
10007 [QA10007] Failed in Failed in accessing the image file. Check if other device has been activated
accessing the image file. first, and then execute this function again.
10007 [RE10007] Failed in Failed in acquiring the study information. Restart DR-ID 300CL.
acquiring the study If the same phenomenon recurred, reinstall
information. DR-ID 300CL.
10007 [ST10007] Failed in Failed in communication with RIS. Restart the system.
communication with RIS. If the same error recurs after restarting the
system, reinstall the system.
10007 [QC10007] An error of An error of unknown cause occurred. Restart DR-ID 300CL.
unknown cause occurred. If the same phenomenon recurred, reinstall
DR-ID 300CL.
10007 [SM10007] The system will The system will be shut down because the Charge the battery.
be shut down because the remaining battery level of the PC is low.
remaining battery level of
Console PC is low.
10007 ClsAdv_Tomo 1% An error occurred in the process to receive Preform the operation again. If the same 1%: The spot of the
to images from PPU. phenomenon recurred, restart DR-ID error and its cause
10011 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-63
Event Event Source Description Causes Remedies Remarks
10008 CslAdv_Modality %1 %2 Information was notified from the Image Check the content displayed in the Image %1: The module error
statusId = {%3}, Reader. Reader status display dialogue. occurred
DataSourceType = {%4}, %2: Spot error occurred
selectorNo = {%5}, %3: ID of connected
hostName = {%6} character string and
status display
character string
%4: ID for identifying
the data source
which this error
character string ID
is stored
%5: Selector number
%6: Host name
10008 [QA10008] Failed in saving Failed in saving information. Check if other device has been activated
information. first, and then execute this function again.
10008 [RE10008] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10008 [ST10008] A communication A communication error with RIS occurred. Restart the system.
error with RIS occurred. If the same error recurs after restarting the
system, reinstall the system.
10008 [QC10008] Failed in Failed in accessing the database. Click the close button on the QC execution
accessing the database. window.
10008 [TS10008] An error An error occurred while obtaining Image Click the button displayed on the screen. * This function cannot
occurred while obtaining Information. be executed.
Image Information.
10008 [SM10008] The system has The system has been shut down due to the Execute again after charging the battery.
been shut down due to the low remaining battery level.
low remaining battery level.
10009 [QA10009] Failed in saving Failed in saving the information. Check if other device has been activated
the information. after completing the QA.
10009 [RE10009] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10009 [ST10009] An An communication error with RIS occurred. Restart the system.
communication error with If the same error recurs after restarting the
RIS occurred. system, reinstall the system.
10009 [QC10009] Failed to access Failed to access the registry information. Restart DR-ID 300CL.
the registry information. If the same phenomenon recurred, reinstall
DR-ID 300CL.
10009 [TS10009] An error An error occurred while obtaining the Click the button displayed on the screen. * This function cannot
occurred while obtaining exposure menu name. be executed.
the exposure menu name.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-64
Event Event Source Description Causes Remedies Remarks
10009 [SM10009] Unexpected Unexpected power shutdown has Confirm the study details and then perform
power shutdown has occurred. the exposure again.
occurred.
An inconsistent menu(s) in
the last executed study has
been deleted.
10010 [MD10010] An error An error occurred while changing an Select the menu again.
occurred while changing an exposure technique.
exposure technique.
10010 [QA10010] An error An error occurred. Restart DR-ID 300CL.
occurred. If the same error recurs after restarting,
reinstall DR-ID 300CL.
10010 [RE10010] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10010 [ST10010] Failed in Failed in obtaining information from RIS. Restart the system.
obtaining information from If the same error recurs after restarting the
RIS. system, reinstall the system.
10010 [QC10010] Free hard disk Free hard disk space is not sufficient. Confirm the free space of the hard disk,
space is not sufficient. and perform the operation again.
10010 [TS10010] An error An error occurred while obtaining the Click the button displayed on the screen. * This function cannot
occurred while obtaining mis-exposure status. be executed.
the mis-exposure status.
10010 [SM10010] Wireless LAN Wireless LAN adapter: can’t recognize %s. Please check the state of s%
adapter: can’t recognize
%s.
10011 [MD10011] Failed in setting Failed in setting the exposure size. Confirm the connected device.
the exposure size.
10011 [RE10011] Input of field Input of field name is invalid. Enter information correctly.
name is invalid.
10011 [ST10011] Failed in Failed in querying patient information. Restart the system.
querying patient If the same error recurs after restarting the
information. system, reinstall the system.
10011 [QC10011] Failed in writing Failed in writing a file. Check the output media status.
a file.
10011 [TS10011] Image Image information could not be changed. Click the button displayed on the screen. * This function cannot
information could not be be executed.
changed.
10011 [SM10011] Wireless LAN Wireless LAN adapter: abnormal in s% Please execute the restoration of network.
adapter: abnormal in s% happened If restoration fails, please restart the
happened system.
If the same error occurs even after the
system is restarted, check the status of the
Wireless LAN device.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-65
Event Event Source Description Causes Remedies Remarks
10012 [MD10012] Failed in setting Failed in setting the reference position for Check the device.
the reference position for exposure.
exposure.
10012 [QA10012] Failed in Failed in applying the changes of trimming Restart the system.
applying the changes of settings. If the same error recurs after restarting the
trimming settings. system, reinstall the system.
10012 [RE10012] Failed in data Failed in data access of magnetic card Restart the system.
access of magnetic card reader. If the same error recurs after restarting the
reader. system, reinstall the system.
10012 [TS10012] An error An error occurred while saving information. Click the button displayed on the screen. * This function cannot
occurred while saving be executed.
information.
10013 [MD10013] Image size Image size could not be detected. Confirm the connected device.
could not be detected.
10013 [QA10013] Failed in Failed in accessing the database. Check if the host has been activated.
accessing the database.
10013 [RE10013] Failed in data Failed in data access of magnetic card Restart the system.
access of magnetic card reader. If the same error recurs after restarting the
reader. system, reinstall the system.
10013 [QC10013] Folder input Folder input incorrect. Confirm the folder name.
incorrect.
10013 [TS10013] An error An error occurred while starting up the Restart DR-ID 300CL.
occurred while starting up control function of reconstruction queue. If the same error recurs after restarting,
the control function of reinstall DR-ID 300CL.
reconstruction queue.
10013 [ST10013] Failed in Failed in automatic saving for QC study. Perform QC study saving on the study list
automatic saving for QC screen.
study.
10014 [MD10014] Failed in Failed in registering the IP#. Check the device.
registering the IP#.
10014 [QA10014] An error An error occurred in image processing. Restart the system.
occurred in image If the same error recurs after restarting the
processing. system, reinstall the system.
10014 [QC10014] The image is The image is illegal. Arrange the phantom position and perform
illegal. an exposure again.
10014 [TS10014] Failed in Failed in switching exposure menu display. Check the status of PPU and the
switching exposure menu connection with PPU, and switch the
display. display of the exposure menu again.
10014 [ST10014] Failed in Failed in registering the IP#. Click the button and continue the
registering the IP#. operation.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-66
Event Event Source Description Causes Remedies Remarks
10015 [MD10015] This cassette This cassette cannot be registered Suspend the current study.
cannot be registered properly. It has been registered to the Read then the cassette’s IP# again on the
properly. following menu where image input is either Local WL list.
It has been registered to being processed or stopped halfway. If the menu is found, open the relevant
the following menu where Patient name: %s Exposure menu: %s study to check the state of such menu
image input is either being processing. Delete then the exposure
processed or stopped menu if it is judged unnecessary.
halfway.
Patient name: %s Exposure
menu: %s
10015 [QA10015] Failed in Failed in accessing the database. Click the button displayed on the screen. * This function cannot
accessing the database. be executed.
10015 [QC10015] An error An error occurred. Perform re-exposure with a default size.
occurred.
10015 [TS10015] Failed in Failed in switching exposure menu display. Click the button displayed on the screen.
switching exposure menu
display.
10016 [MD10016] Hard Disk Full. The hard disk is full. Delete unnecessary images from the list.
If the error persists, contact your service
engineer.
10016 [QA10016] Failed in Failed in accessing the database. Click the button displayed on the screen. * This function cannot
accessing the database. be executed.
10016 [QC10016] Measurement Measurement processing cannot be Check to see if an image has been input.
processing cannot be implemented.
implemented.
10016 [TS10016] Image Image information of the selected study Click the button displayed on the screen. * This function cannot
information of the selected could not be obtained. be executed.
study could not be
obtained.
10017 [MD10017] The exposure The exposure menu could not be Check the device and select appropriate
menu could not be determined. menus for the exposure again.
determined.
10017 [QA10017] Failed in Failed in calculating the S or L value. Restart the system.
calculating the S or L value.
10017 [RE10017] Forbidden Forbidden character(s) are contained. Enter information correctly.
character(s) are contained.
10017 [QC10017] Measurement Measurement processing cannot be Enter a correct image.
processing cannot be implemented.
implemented.
10017 [TS10017] Failed in loading Failed in loading the image information file. Click the button displayed on the screen. * This function cannot
the image information file. be executed.
10018 [MD10018] Failed in Failed in cancelling the exposure. Restart the system.
cancelling the exposure. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-67
Event Event Source Description Causes Remedies Remarks
10018 [QA10018] An error An error occurred in ROI calculation. Restart the system.
occurred in ROI calculation. If the same error recurs after restarting the
system, reinstall the system.
10018 [QC10018] Measurement Measurement processing cannot be Set the size of the 1shot Phantom image
processing cannot be implemented. to 24x30 or larger.
implemented.
10018 [TS10018] Failed in loading Failed in loading the exposure menu Click the button displayed on the screen.
the exposure menu information.
information.
10019 [MD10019] Failed in Failed in reading the IP#. Perform the read operation again.
reading the IP#.
10019 [QA10019] An error An error occurred in Edge Reinforcement Restart the system.
occurred in Edge processing. If the same error recurs after restarting the
Reinforcement processing. system, reinstall the system.
10019 [QC10019] Measurement Measurement processing cannot be Use only the images from the same reader
processing cannot be implemented. for one study.
implemented.
10019 [TS10019] Failed in loading Failed in loading image information files of Click the button displayed on the screen.
image information files of exposure menus.
exposure menus.
10020 [RE10020] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10020 [QC10020] Image with Image with unsupported reading density is Select an image with the designated
unsupported reading included. reading density.
density is included.
10020 [TS10020] Failed in Failed in displaying the patient information. Click the button displayed on the screen. * This function cannot
displaying the patient be executed.
information.
10021 [MD10021] Failed in image Failed in image resecuring. Restart the system.
resecuring. If the same error recurs after restarting the
system, reinstall the system.
10021 [QA10021] Failed in Failed in obtaining the login user. Restart the system.
obtaining the login user.
10021 [RE10021] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10021 [QC10021] Measurement Measurement processing cannot be Use only images with a same reading
processing cannot be implemented. density.
implemented.
10021 [TS10021] Failed in Failed in displaying the exposure menu Click the button displayed on the screen. * This function cannot
displaying the exposure name and mis-exposure status. be executed.
menu name and mis-
exposure status.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-68
Event Event Source Description Causes Remedies Remarks
10022 [MD10022] Failed in image Failed in image inputting process in the Restart the system.
inputting process in the image input sub-system. If the same error recurs after restarting the
image input sub-system. system, reinstall the system.
10022 [QA10022] Failed in Failed in updating the login user. Restart the system.
updating the login user.
10022 [RE10022] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10022 [QC10022] Data cannot be Data cannot be deleted because the basic Select data without basic value data.
deleted. value data are included.
10023 [MD10023] Image input Image input processing has failed. Restart the system.
processing has failed. If image input failed again, restart the
Reader Unit.
10023 [QA10023] Failed in Failed in creating a temporary DDO file. Restart the system.
creating a temporary DDO
file.
10023 [RE10023] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10023 [QC10023] Measurement Measurement processing cannot be Use only the images from the same reader
processing cannot be implemented. for one study.
implemented.
10023 [TS10023] Failed in saving Failed in saving the information. Confirm that the other device has been
the information. activated after completing the QA.
10024 [MD10024] An error An error notification was issued by the The image is registered to the study of
notification was issued by Imager Reader. The target image may Exposure menu and IP number.
the Imager Reader. The have been input erroneously in the study Therefore, subject the IP to image erasure
target image may have of details below. processing on the Image Reader.
been input erroneously in • Patient ID Perform re-exposure if necessary.
the study of details below: • Patient name
Patient ID: %s Patient • Study date
name: %s Study date: • Study time
%s Study time: %s
10024 [QA10024] Failed in saving Failed in saving settings. Restart the system.
settings.
10024 [QC10024] Measurement Measurement processing cannot be Confirm if the specified reader has been
processing cannot be implemented. registered.
implemented.
10024 [TS10024] Failed in saving Failed in saving the settings. Click the button displayed on the screen. * This function cannot
the settings. be executed.
10025 [QA10025] An error An error occurred in window processing. Restart the system.
occurred in window
processing.
10025 [TS10025] Failed in Failed in calculating the S or L value. Click the button displayed on the screen. * This function cannot
calculating the S or L value. be executed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-69
Event Event Source Description Causes Remedies Remarks
10025 [QC10025] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall DR-ID
300CL.
10025 [MD10025] An error An error occurred. The menu whose IP number is registered
occurred. does not exist. Select the menu again and
register IP number.
10026 [QA10026] Failed in Failed in automatic linkage of stitching Check the order, orientation and position of
automatic linkage of images. the images.
stitching images. Set the images to the same direction.
10026 CslAdv_Qc QCGridViewEx General QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10026 [TS10026] An error An error occurred in ROI calculation. Restart DR-ID 300CL.
occurred in ROI calculation.
10026 [MD10026] Hard Disk Full. The hard disk is full. Contact your service engineer.
10027 [QA10027] Failed in Failed in accessing the database. Click the button displayed on the screen,
accessing the database. and restart DR-ID 300CL.
10027 [TS10027] An error An error occurred while updating the state Restart DR-ID 300CL.
occurred while updating the of reconstruction processing.
state of reconstruction
processing.
10028 [QA10028] Failed in Failed in accessing the image file. Ensure that the other unit is activated
accessing the image file. before executing this function again.
10028 [RE10028] Failed in Failed in deleting an image file. Restart the system.
deleting an image file. If the same error recurs after restarting the
system, reinstall the system.
10028 [TS10028] The free space The free space of HDD to start the Delete unnecessary images from the study
of HDD to start the reconstruction processing is insufficient. list.
reconstruction processing
is insufficient.
10029 [QA10029] An error An error occurred in Magnification/ Restart the system.
occurred in Magnification/ Reduction function. If the same error recurs after restarting the
Reduction function. system, reinstall the system.
10030 [QA10030] An error An error occurred in Panning function. Restart the system.
occurred in Panning If the same error recurs after restarting the
function. system, reinstall the system.
10030 [RE10030] The date format The date format is invalid. Enter again.
is invalid.
10031 [QA10031] An error An error occurred in Magnification function. Restart the system.
occurred in Magnification If the same error recurs after restarting the
function. system, reinstall the system.
10031 [RE10031] Failed in query Failed in query for RIS. Confirm the RIS.
for RIS.
10031 CslAdv_Qc QCInputEX General QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-70
Event Event Source Description Causes Remedies Remarks
10032 [QA10032] An error An error occurred in Reduction function. Restart the system.
occurred in Reduction If the same error recurs after restarting the
function. system, reinstall the system.
10032 [RE10032] Failed in data Failed in data access of bar cord reader. Restart the system.
access of bar cord reader. If the same error recurs after restarting the
system, reinstall the system.
10033 [QA10033] An error An error occurred in Actual size function. Restart the system.
occurred in Actual size If the same error recurs after restarting the
function. system, reinstall the system.
10033 [RE10033] Failed in data Failed in data access of bar cord reader. Restart the system.
access of bar cord reader. If the same error recurs after restarting the
system, reinstall the system.
10034 [QA10034] An error An error occurred in 1-to-1 pixel size Restart the system.
occurred in 1-to-1 pixel size function. If the same error recurs after restarting the
function. system, reinstall the system.
10034 [RE10034] The input data The input data from the barcode reader is Enter the data again from the barcode
from the barcode reader is invalid. reader.
invalid.
10035 [RE10035] Failed in Failed in acquiring the order information. Confirm the RIS.
acquiring the order
information.
10035 [QA10035] Failed in Failed in executing the Smart-QA Restart DR-ID 300CL. If the same
executing the Smart-QA initialization. phenomenon recurred, reinstall DR-ID
initialization. 300CL.
10035 [ST10035] Failed in Failed in retrieving the order information. Check the status about the ordering
retrieving the order device.
information.
10036 [RE10036] Memory is Memory is running out. Restart the system.
running out. If the same error recurs after restarting the
system, reinstall the system.
10036 [QA10036] Failed in Failed in executing the grid shading Restart DR-ID 300CL. If the same
executing the grid shading detection function. phenomenon recurred, reinstall DR-ID
detection function. 300CL.
10036 [ST10036] Failed to Failed to perform automatic calculation of Click the button to restart the system.
perform automatic Tomosynthesis exposure conditions. If the same error recurs after restarting the
calculation of system, reinstall the system.
Tomosynthesis exposure
conditions.
10037 [RE10037] A time out A time out occurred. Delivery cannot start. Restart the system.
occurred. Delivery cannot If the same error recurs after restarting the
start. system, reinstall the system.
10037 [QA10037] Failed in Failed in executing the Smart-QA reset Restart DR-ID 300CL.
executing the Smart-QA processing. If the same phenomenon recurred, reinstall
reset processing. DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-71
Event Event Source Description Causes Remedies Remarks
10037 [ST10037] Error has Error has occurred while Free Layout Print Perform operations again.
occurred while Free Layout is in progress. If the same phenomenon recurred, restart
Print is in progress. the system.
10038 [RE10038] A time out A time out occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10038 [QA10038] Failed in Failed in executing the Smart-QA analysis. Restart DR-ID 300CL.
executing the Smart-QA If the same phenomenon recurred, reinstall
analysis. DR-ID 300CL.
10039 [RE10039] A timeout A timeout occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10040 [RE10040] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10040 [QA10040] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10041 [RE10041] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10042 [RE10042] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10043 [RE10043] The RIS unit is The RIS unit is offline. Restart the system.
offline. If the same error recurs after restarting the
system, reinstall the system.
10044 [RE10044] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10045 [RE10045] A time out A time out occurred. The order information Restart the system.
occurred. The order confirmation dialog cannot be opened. If the same error recurs after restarting the
information confirmation system, reinstall the system.
dialog cannot be opened.
10046 [RE10046] The RIS unit is The RIS unit is offline. Ensure that the device has been activated
offline. and connected to a network.
10047 [RE10047] This operation This operation is not allowed. Login again with an authorized username.
is not allowed.
10050 [RE10050] Failed in Failed in accessing the setting file. Restart the system.
accessing the setting file. If the same error recurs after restarting the
system, reinstall the system.
10051 [RE10051] Failed in Failed in acquiring the study information. Update the study list.
acquiring the study
information.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-72
Event Event Source Description Causes Remedies Remarks
10052 [RE10052] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10053 [RE10053] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10054 [RE10054] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10055 [RE10055] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10056 [RE10056] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10057 [RE10057] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10057 CslAdv_QA QACommon.dll Failed in judgment of the 2ON1 menu or Perform the operation again.
AdjustImgCtrl - the pair menu. If the same phenomenon recurred, restart
(JudgePairMenu) DR-ID 300CL.
(JudgeIs2on1Menu) -
<expcetion content>
10058 [RE10058] The RIS unit is The RIS unit is offline. Ensure that the device has been activated
offline. and connected to a network.
10059 [RE10059] Failed in reading Failed in reading the order information. Restart the system.
the order information. If the same error recurs after restarting the
system, reinstall the system.
10060 CslAdv_Re Unknow error when receive An unknown exception occurred when Restart the system.
messag from FW receiving message from firewall.
10060 [RE10060] Failed in Failed in registering the patient information. Restart the system.
registering the patient If the same error recurs after restarting the
information. system, reinstall the system.
10061 CslAdv_Re Create lock object fail. Failed in creating the Locked object. Restart DR-ID 300CL.
10061 [RE10061] Failed in Failed in reserving studies. Restart the system.
reserving studies. If the same error recurs after restarting the
system, reinstall the system.
10062 [RE10062] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10063 [RE10063] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-73
Event Event Source Description Causes Remedies Remarks
10064 [RE10064] Failed in reading Failed in reading the result data from the Restart the system.
the result data from the F-RIS system. If the same error recurs after restarting the
F-RIS system. system, reinstall the system.
10065 [RE10065] Failed in Failed in requesting to switch the current Restart the system.
requesting to switch the user. If the same error recurs after restarting the
current user. system, reinstall the system.
10065 CslAdv_Re Failed in start manual study An error occurred while starting a study. Restart DR-ID 300CL.
from KensinOrder
10065 CslAdv_QA QA.exe Failed in updating the image display when Perform the search again.
QAScreen clicking [ON] or [OFF]. If the same phenomenon recurred, restart
vwMammoAlignmentCtrl_ DR-ID 300CL.
RefreshImage - <exception
content>
10066 [RE10066] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10067 [RE10067] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10068 [RE10068] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10069 [RE10069] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10070 [RE10070] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10071 [RE10071] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10072 [RE10072] Failed in Failed in accessing the database. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
10073 [RE10073] Failed in Failed in converting the patient information. Restart the system.
converting the patient If the same error recurs after restarting the
information. system, reinstall the system.
10074 [RE10074] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
10075 [RE10075] The RIS unit is The RIS unit is offline. Restart the system.
offline.
10076 [RE10076] Failed in starting Failed in starting the study. Restart the system.
the study.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-74
Event Event Source Description Causes Remedies Remarks
10077 [RE10077] Failed in Failed in sending exposure results. Restart the system.
sending exposure results.
10078 [RE10078] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
10079 [RE10079] Failed in Failed in magnetic card reader processing. Restart the system.
magnetic card reader If the same error recurs after restarting the
processing. system, reinstall the system.
10080 [RE10080] Failed in Failed in barcode reader processing. Restart the system.
barcode reader processing. If the same error recurs after restarting the
system, reinstall the system.
10081 [RE10081] Failed in Failed in magnetic card reader processing. The card type is not acceptable.
magnetic card reader Check the card type.
processing.
10082 [RE10082] Failed to read Failed to read bar code. Check the card content.
bar code.
10083 [RE10083] Failed in Failed in magnetic card reader processing. Retry or check the card content.
magnetic card reader
processing.
10084 [RE10084] Failed in Failed in executing the study. Restart the system.
executing the study. If the same error recurs after restarting the
system, reinstall the system.
10085 [RE10085] Failed in starting Failed in starting the study. Restart the system.
the study. If the same error recurs after restarting the
system, reinstall the system.
10086 [RE10086] The beginning The beginning date is set earlier than the Set the ending date later than the
date is set earlier than the ending date. beginning date.
ending date.
10087 [RE10087] Failed in Failed in activating RIS. Restart the system.
activating RIS. If the same error recurs after restarting the
system, reinstall the system.
10088 [RE10088] Failed in data Failed in data processing from RIS. Restart the system.
processing from RIS. If the same error recurs after restarting the
system, reinstall the system.
10089 [RE10089] The latest The latest search conditions could not be Restart the system.
search conditions could not loaded. If the same error recurs after restarting the
be loaded. system, reinstall the system.
10090 [RE10090] The latest The latest search conditions could not be Restart the system.
search conditions could not saved. If the same error recurs after restarting the
be saved. system, reinstall the system.
10091 [RE10091] A temporary A temporary folder for setting a MWM Restart the system.
folder for setting a MWM worklist could not be created. If the same error recurs after restarting the
worklist could not be system, reinstall the system.
created.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-75
Event Event Source Description Causes Remedies Remarks
10092 [RE10092] A A communication error with RIS occurred. Restart the system.
communication error with If the same error recurs after restarting the
RIS occurred. system, reinstall the system.
10093 [RE10093] An error An error occurred in communication with Restart the system.
occurred in communication RIS. If the same error recurs after restarting the
with RIS. system, reinstall the system.
10094 [RE10094] Failed in Failed in communication with RIS. Restart the system.
communication with RIS. If the same error recurs after restarting the
system, reinstall the system.
10097 [RE10097] Information from Information from RIS could not be Restart the system.
RIS could not be obtained. obtained. If the same error recurs after restarting the
system, reinstall the system.
10098 [RE10098] Failed in starting Failed in starting the MWM search setting Restart the system.
the MWM search setting form. If the same error recurs after restarting the
form. system, reinstall the system.
10099 [RE10099] The custom The custom setting of MWM worklist has Restart the system.
setting of MWM worklist become disabled. If the same error recurs after restarting the
has become disabled. system, reinstall the system.
10100 [RE10100] No executable No executable study is found. Click the button displayed on the screen.
study is found.
10101 [RE10101] The DDO file of The DDO file of MWM worklist cannot be Restart the system.
MWM worklist cannot be deleted. If the same error recurs after restarting the
deleted. system, reinstall the system.
10102 [RE10102] Failed in Failed in searching patient information. Restart the system.
searching patient If the same error recurs after restarting the
information. system, reinstall the system.
10103 [RE10103] Free layout Free layout print: Failed in unlocking the Restart the system.
print: Failed in unlocking object. If the same error recurs after restarting the
the object. system, reinstall the system.
10104 [RE10104] Data from RIS Data from RIS cannot be displayed. Restart the system.
cannot be displayed. If the same error recurs after restarting the
system, reinstall the system.
10105 [RE10105] The latest The latest search conditions cannot be Restart the system.
search conditions cannot loaded. If the same error recurs after restarting the
be loaded. system, reinstall the system.
10106 [RE10106] The latest The latest search conditions cannot be Restart the system.
search conditions cannot stored. If the same error recurs after restarting the
be stored. system, reinstall the system.
10107 [RE10107] Data returned Data returned from RIS cannot be stored. Restart the system.
from RIS cannot be stored. If the same error recurs after restarting the
system, reinstall the system.
10108 [RE10108] Barcode data Barcode data cannot be obtained. Re-input from BCR.
cannot be obtained.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-76
Event Event Source Description Causes Remedies Remarks
10109 [RE10109] Data returned Data returned from RIS cannot be used to Restart the system.
from RIS cannot be used to create exposure menus or film output If the same error recurs after restarting the
create exposure menus or numbers. system, reinstall the system.
film output numbers.
10110 [RE10110] Failed in Failed in switching the user. Restart the system.
switching the user. If the same error recurs after restarting the
system, reinstall the system.
10111 [RE10111] The RIS unit is The RIS unit is offline. Ensure that the device has been activated
offline. and connected to a network.
10112 [RE10112] Failed in starting Failed in starting the MWM form search Restart the system.
the MWM form search setting. If the same error recurs after restarting the
setting. system, reinstall the system.
10113 [RE10113] Failed in Failed in restoring the image data base. Restart the system.
restoring the image data If the same error recurs after restarting the
base. system, reinstall the system.
10115 [RE10115] Failed in storing Failed in storing data returned from RIS. Restart the system.
data returned from RIS. If the same error recurs after restarting the
system, reinstall the system.
10116 [RE10116] Failed in Failed in querying patient information. Restart the system.
querying patient If the same error recurs after restarting the
information. system, reinstall the system.
10117 [RE10117] Failed in Failed in updating the setting file for Restart the system.
updating the setting file for customize study list. If the same error recurs after restarting the
customize study list. system, reinstall the system.
10120 [RE10120] Study Study information could not be collected. Ensure that the device has been activated
information could not be and connected to a network.
collected.
10121 [RE10121] The RIS unit is The RIS unit is offline. Ensure that the device has been activated
offline. and connected to a network.
10122 [RE10122] Invalid patient Invalid patient information has been Enter patient information again.
information has been entered.
entered.
10123 [RE10123] Free space to Free space to save data is insufficient. Select another destination for data storage.
save data is insufficient.
10124 [RE10124] An error has An error has occurred while saving local Restart DR-ID 300CL.
occurred while saving local data. If the same error recurs after restarting,
data. reinstall DR-ID 300CL.
10125 [RE10125] The specified The specified folder is not authorized to Select another destination for data storage.
folder is not authorized to save a study.
save a study.
10126 [RE10126] Some of the Some of the studies cannot be saved. Select studies again.
studies cannot be saved.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-77
Event Event Source Description Causes Remedies Remarks
10127 [RE10127] The study The study cannot be started while Click the button displayed on the screen to
cannot be started while processing is in progress. perform the processing in progress.
processing is in progress. Start study again after completing the
processing.
10128 [RE10128] Failed in Failed in sending exposure results. Click the button and continue other
sending exposure results. processing.
10128 CslAdv_QA QA.exe Fire event OnFinishedAnalyze(uniqueKey, Open the study and perform the operation
QAScreen result) of AC::IRS failed. again.
vwExposureListView_ If the same phenomenon recurred, restart
OnChangeImageOrder DR-ID 300CL.
- <exception content>
10129 [RE10129] The specified The specified path was not found. Enter the path again.
path was not found.
10130 [RE10130] The study The study cannot be deleted. Click the button and continue other
cannot be deleted while processing.
processing is in progress.
10131 [RE10131] The study The study cannot be locked or unlocked Click the button and continue other
cannot be locked or while processing is in progress. processing.
unlocked while processing
is in progress.
10131 CslAdv_QA QACommon.dll Call function failed. Perform the operation again.
AdjustImgCtrl - If the same phenomenon recurred, restart
(IsAutoAdjustment) DR-ID 300CL.
(JudgeAdjustment
Conditions)
(JudgeAdjustmentConditions
ForDeliverySettingView)
(ConfirmAndDeletePairProc)
(DeletePairProc)
(ConfirmAndRelease
AdjustmentProc)
(ReleaseAdjustmentProc)
(ReleaseAdjustment
PartnerProc)
(HandlePairAndAdjustment
ForImageLaterality)
(ConfirmAndRelease
AdjustmentForDelivery
SettingView)
(GetMenuAndPartner
MenuIndex)
(JudgeImageLR)
(JudgeImageSize) -
<expcetion content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-78
Event Event Source Description Causes Remedies Remarks
10131 CslAdv_QA QA.exe Call function failed. Open the study and perform the operation
IRSHelper - again.
(Analyze) If the same phenomenon recurred, restart
(Reset) DR-ID 300CL.
(GetImageInfo)
- <exception content>
QAScreen -
(ReAnalyzeIRSCurrent
Image)
(AnalyzeIRSCurrentImage)
(irsHelper_OnFinish
Analyze)
(irsHelper_OnRequest
AutoAdjust)
(vwTrimmingCtrl_On
ApplyingChange)
- <exception content>
10131 CslAdv_QA QATrimmingCtrl.dll Call function failed. Open the study and perform the operation
TrimmingCtrl again.
HandleReleaseAdjustment If the same phenomenon recurred, restart
PartnerImgInfo - <exception DR-ID 300CL.
content>
10131 CslAdv_QA QAOrderInfoChgCtrl.dll Call function failed. Perform the operation again.
OrderInfoMainView - If the same phenomenon recurred, restart
(HandlePairAndAdjustment) DR-ID 300CL.
- <exception content>
10132 [RE10132] Unable to Unable to process because a study Select a study again.
process because a study currently in progress exists in the selected
currently in progress exists studies
in the selected studies.
10133 [RE10133] Patient ID auto- Patient ID auto-convert function cannot be Click the button and continue the
convert function cannot be started. processing.
started.
10133 CslAdv_QA QACommon.dll Release Adjustment failed. Perform the operation again.
AdjustImgCtrl - If the same phenomenon recurred, restart
(ReleaseAdjustment) - DR-ID 300CL.
<expcetion content>
10133 CslAdv_QA QA.exe Release Adjustment failed. Open the study and perform the operation
QAScreen - again.
(frmOutputSetting_ If the same phenomenon recurred, restart
SavingSetting) DR-ID 300CL.
(vwTrimmingCtrl_
OnApplyingChange)
(vwImageRotationCtrl_
ChangeApplying) -
<exception content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-79
Event Event Source Description Causes Remedies Remarks
10134 [RE10134] Failed in reading Failed in reading the setting file. Click the button and confirm the setting file.
the setting file.
10135 [RE10135] Converted data Converted data cannot be obtained. Click the button and continue the
cannot be obtained. processing.
10136 [RE10136] QA cannot be QA cannot be completed. Click the button displayed on the screen.
completed.
10138 [RE10138] The selected The selected study is unable to execute Check that the image has been read by
study is unable to execute the QC. Image Reader for which QC can be
the QC. implemented.
10139 [RE10139] Study Study suspension processing has been For studies which could not be suspended,
suspension processing has completed except for some studies. reserve them again from the ordering
been completed except for system.
some studies.
10150 CslAdv_QA QA.exe Failed in showing IRS result. Open the study and perform the operation
IRSHelper - again.
(ShowContent) - If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
10151 CslAdv_QA QA.exe Failed in applying IRS auto adjustment. Open the study and perform the operation
IRSHelper - again.
(AutoAdjust) If the same phenomenon recurred, restart
(contentViewer_ DR-ID 300CL.
OnFinishedShowContents)
- <exception content>
10152 CslAdv_QA QA.exe Failed in applying IRS analyze process. Open the study and perform the operation
IRSHelper - again.
(Analyze) - <exception If the same phenomenon recurred, restart
content> DR-ID 300CL.
10153 CslAdv_QA QA.exe Failed in terminating IRS. Restart DR-ID 300CL.
IRSHelper -
(Terminate) - <exception
content>
10154 CslAdv_QA QA.exe Failed in resetting IRS. Open the study and perform the operation
IRSHelper - again.
(Reset) - <exception If the same phenomenon recurred, restart
content> DR-ID 300CL.
QAScreen -
(ReAnalyzeIRS)
(ReAnalyzeIRSCurrent
Image)- <exception
content>
10155 CslAdv_QA QA.exe Failed in enabling/disabling IRS's auto Open the study and perform the operation
IRSHelper - adjustment function. again.
(EnableAutoAdjust) - If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-80
Event Event Source Description Causes Remedies Remarks
10156 CslAdv_QA QA.exe Failed in accessing CSL database. Check that the _console_db exists and
IRSHelper - DR-ID 300CL.
(talker_
OnFinishedAnalyze) -
<exception content>
QAScreen -
(irsHelper_
OnRequestAutoAdjust)-
<exception content>
10157 CslAdv_QA QA.exe Failed in checking existence of contents of Open the study and perform the operation
IRSHelper - IRS. again.
(IsExistContents) - If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
10158 CslAdv_QA QA.exe IRS' analyzing returned result is error. Open the study and perform the operation
IRSHelper - again.
(talker_ If the same phenomenon recurred, restart
OnFinishedAnalyze) - DR-ID 300CL.
<exception content>
QAScreen -
(ReAnalyzeIRS)
(ReAnalyzeIRSCurrent
Image) - <exception
content>
10200 CslAdv_Re Log directory not exist Log directory could not be acquired from Restart DR-ID 300CL.
Directory.ini. If the same error recurs after restarting,
reinstall DR-ID 300CL.
10203 CslAdv_Qc Load resource (captions, Resource data for QC cannot be acquired. Restart QC. If the same phenomenon
icons, tooltips) error DR-ID 300CL may not have been installed recurred, reinstall DR-ID 300CL.
correctly.
10205 CslAdv_Qc Initialize QCAutoCalc QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
object fail. may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10206 CslAdv_Qc Initialize QCAutoCalc QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
object fail. Error code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10207 CslAdv_Qc MainContainer General QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10235 CslAdv_Re Could not load data from DDO file corresponding to the study has Restart DR-ID 300CL.
DDO files to memory. been deleted.
10235 CslAdv_Re Could not save DDO object DDO file corresponding to the study has Restart DR-ID 300CL.
into DDO file. been deleted, or is read-only file.
10235 CslAdv_Re Could not set study date to DDO file corresponding to the study has Restart DR-ID 300CL.
DDO object. been deleted, or is read-only file.
10235 CslAdv_Re Exception of • _console_db does not exist. Restart DR-ID 300CL.
AC::ImgDBCEx • Connection with the SQL server was If the same error recurs after restarting,
disconnected. reinstall DR-ID 300CL.
• The SQL service has been stopped.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-81
Event Event Source Description Causes Remedies Remarks
10235 CslAdv_Re Failed in accessing into the • _console_db does not exist. Restart DR-ID 300CL.
console image db • Connection with the SQL server was If the same error recurs after restarting,
disconnected. reinstall DR-ID 300CL.
• The SQL service has been stopped. If the same phenomenon recurred, initialize
the image database.
10235 CslAdv_Re Failed in accessing into the An error occurred while updating the Restart DR-ID 300CL.
recept image db information on study date and on releasing
of the ownership.
10235 CslAdv_Re Could not save DDO object An error occurred while updating the Restart DR-ID 300CL.
into DDO file information on study date and on releasing
of the ownership.
10235 CslAdv_Re The study time could not be The DDO file corresponding to the study Restart DR-ID 300CL.
set to the DDO object. was deleted or is a read-only file.
10236 FW Initialize communication Failed in Linker initialization. Check the setting of Linker library, and
failed. then restart the system.
10236 FW Send wakeup cancelled Failed in notifying the message of "wakeup Check the setting of Linker library, and
failed. cancelled" from MiniCooper to Console. then restart the system.
10236 FW Send wakeup completed Failed in notifying the message of "wakeup Check the setting of Linker library, and
failed. completed" from MiniCooper to Console. then restart the system.
10240 CslAdv_Re Access to list view config • "C:\Program File\Fujifilm\IIP\Work\ Restart DR-ID 300CL.
file failed Reception" folder is a read-only folder. If the same error recurs after restarting,
• Login user does not have the authority to reinstall DR-ID 300CL.
edit.
10257 CslAdv_Study Display SL to error It fails when display SL value. Restart DR-ID 300CL.
If the same error recurs after restarting,
reinstall DR-ID 300CL.
10260 CslAdv_Re XML file not exist • The file (PluginDefinition.xml, or Restart DR-ID 300CL.
PluginDefinition.xsd) to read the plugin If the same error recurs after restarting,
definition information does not exist. reinstall DR-ID 300CL.
• The path to the file does not exist.
10261 CslAdv_Re Read plugin definition xml • The file (PluginDefinition.xml, or Restart DR-ID 300CL.
file error PluginDefinition.xsd) to acquire the plugin If the same error recurs after restarting,
definition information does not exist. reinstall DR-ID 300CL.
• The path to the file does not exist.
10281 CslAdv_Re Failed in saving into the The path to the folder which stores Restart DR-ID 300CL.
CustomizeStudyList.xml CustomizeStudyListInfo.xml does not exist. If the same error recurs after restarting,
reinstall DR-ID 300CL.
10300 CslAdv_Qc Save QC data occurs error An error occurred when saving the QC Confirm the authority and the free space
data. for the destination folder to save.
An already existing file might not be
overwritten.
10301 CslAdv_Qc Error when read tag QC data is illegal. Some necessary Perform an exposure again. If the same {0:X8}: DDO tag ID
({0:X8}) information is not included in DDO files. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-82
Event Event Source Description Causes Remedies Remarks
10302 CslAdv_Qc Check QC conditions failed Selected data may not satisfy QC The data may not be supported. If the
measurement condition. same phenomenon recurred, expose an
the image again.
10303 CslAdv_Qc ImpView General exception QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10315 CslAdv_Qc Folder folder Browser QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Dialog exeception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10316 CslAdv_Qc Display study infor error. QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
Error code: {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10321 CslAdv_Qc Load buffer from file error. RAW images and thumbnail images are Perform an exposure again. If the same
not in the selected folder. phenomenon recurred, reinstall DR-ID
300CL.
10322 CslAdv_Qc Create xml LUT parameters QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
error. Error code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10323 CslAdv_Qc Create LUT table error. QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
Error code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10324 CslAdv_Qc Process trimming error. Errr QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10325 CslAdv_Qc Process compress JPEG QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
error. Error code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10326 CslAdv_Qc Decode JPEG Error. Error QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10327 CslAdv_Qc Generate XML parameter QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
for FCR processing error. may not have been installed correctly. recurred, reinstall DR-ID 300CL.
Error code {0}
10328 CslAdv_Qc Process FCR Error. Error QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon {0}: Error code
code {0} may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10335 CslAdv_Study Failed in constructor It fails when constructor form. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
10336 CslAdv_Qc Display thumbnail image QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
error may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10336 CslAdv_Study Failed in Set value It fails in set irradiation area, size, position, Restart DR-ID 300CL. If the same
direction, irradiation link in form. phenomenon recurred, reinstall DR-ID
300CL.
10337 CslAdv_Study Failed in get menu It fails when get information about size, Restart DR-ID 300CL. If the same
information position, direction, irradiation link in form. phenomenon recurred, reinstall DR-ID
300CL.
10338 CslAdv_Study Failed in initialize It fails when initialize form. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
10339 CslAdv_Study Failed in set icon to control It fails in set icon to control. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-83
Event Event Source Description Causes Remedies Remarks
10340 CslAdv_Study Failed in set text to control It fails in set text to control. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
10341 CslAdv_Study Failed in set font size to It fails in set font size to control. Restart DR-ID 300CL. If the same
control phenomenon recurred, reinstall DR-ID
300CL.
10342 CslAdv_Study Failed in set display control It fails in set display control in form. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
10346 CslAdv_Qc Load basic value from DB Failed in accessing the database. Restart QC. If the same phenomenon
failed recurred, reinstall DR-ID 300CL.
10347 CslAdv_Qc QCResult General QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10367 CslAdv_Qc Divide By Zero Exception QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10368 CslAdv_Qc Invalid Cast Exception QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10369 CslAdv_Qc Index Out Of Range QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10370 CslAdv_Qc Argument Out Of Range QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10372 CslAdv_QA QA.exe Failed in showing IRS result. Open the study and perform the operation
QAScreen - again.
(imageControlView_ If the same phenomenon recurred, restart
ButtonClick) - <exception DR-ID 300CL.
content>
10373 CslAdv_QA QA.exe Failed in applying IRS auto adjustment. Open the study and perform the operation
QAScreen - again.
(irsHelper_ If the same phenomenon recurred, restart
OnRequestAutoAdjust) DR-ID 300CL.
- <exception content>
10376 CslAdv_QA QA.exe Failed in starting the linking application Check the operation and the status of
QAScreen (ROI). connection for ROI, restart DR-ID 300CL.
NotifyROI
<exception content>
10377 CslAdv_QA QA.exe An error occurred to the processing in the Open the study and perform the operation
QAScreen program used in QA reset function. again.
btnReset_Click If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
10378 CslAdv_QA QA.exe An error occurred to the processing in the Open the study and perform the operation
QAScreen program used in QA reset function. again.
btnReset_Click If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-84
Event Event Source Description Causes Remedies Remarks
10380 CslAdv_QA QA.exe Failed in saving the image processing Open the study and perform the operation
QAScreen parameter used in QA reset function. again.
BackupChangeImgPrm If the same phenomenon recurred, restart
FlagForAllImage DR-ID 300CL.
<exception content>
10381 CslAdv_QA QA.exe Failed in resetting the image information. Open the study and perform the operation
QAScreen again.
ResetChangeImgInf If the same phenomenon recurred, restart
FlagForAllImage DR-ID 300CL.
<exception content>
10382 CslAdv_QA QA.exe Failed in resetting the image processing Open the study and perform the operation
QAScreen parameter. again.
ResetChangeImgPrm If the same phenomenon recurred, restart
FlagForAllImage DR-ID 300CL.
<exception content>
10383 CslAdv_QA QA.exe Failed in saving the image information Open the study and perform the operation
QAScreen used in QA reset function. again.
BackupChangeImgInf If the same phenomenon recurred, restart
FlagForAllImage DR-ID 300CL.
<exception content>
10393 CslAdv_Re A general error occurred. An error occurred in FLPDataAccessor. Restart DR-ID 300CL.
10400 CslAdv_Qc Unauthorized Access QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
Exception may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10400 [TS10400] An error An error occurred while displaying the Check the status of PPU and the
occurred while displaying representative reconstruction image. connection with PPU, and restart the
the representative system.
reconstruction image.
10401 CslAdv_Qc Argument Exception QC components crashed. DR-ID 300CL Restart QC. If the same phenomenon
may not have been installed correctly. recurred, reinstall DR-ID 300CL.
10401 [TS10401] An error An error occurred while closing the Check the status of PPU and the
occurred while closing the representative reconstruction dialog. connection with PPU, and restart the
representative system.
reconstruction dialog.
10401 CslAdv_Re No study can start Study/ No study is capable of starting study/QA. Select another study.
QA
10402 CslAdv_Qc Path Too Long Exception When saving QC data, its file path is longer Confirm the file path length.
than Windows’s.
10402 [TS10402] An error An error occurred while displaying images. Check the status of PPU and the
occurred while displaying connection with PPU.
images.
10403 CslAdv_Qc Directory Not Found The folder selected to save QC file does Restart QC. If the same phenomenon
Exception not exist. recurred, reinstall DR-ID 300CL.
10403 [TS10403] An error An error occurred while processing Check the status of PPU and the
occurred while processing images. connection with PPU.
images.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-85
Event Event Source Description Causes Remedies Remarks
10405 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. If the same 1%: The spot of the
to TomoImageDisplayMgrCtrl.dll. phenomenon recurred, reinstall DR-ID error and its cause
10415 Failed in displaying images. 300CL.
10411 CslAdv_Re Get PatientInfo false The data cannot be saved into the Patient Restart DR-ID 300CL.
database. If the same phenomenon recurred, reinstall
DR-ID 300CL.
If the same phenomenon recurred,
initialize the image database.
10413 [TS10413] Failed in Failed in switching image display. Check the status of PPU and the
switching image display. connection with PPU.
10419 CslAdv_Re Conver to JIS false The data format to be converted is invalid. Restart DR-ID 300CL.
10422 CslAdv_Qc Save standard data to Failed in accessing the database. Restart QC. If the same phenomenon
database failed. recurred, reinstall DR-ID 300CL.
10431 CslAdv_Re Failed in access • The CustomizePatientListInfo.xml file Restart DR-ID 300CL. If the same
configuration file for setting does not exist. phenomenon recurred, reinstall DR-ID
display Patient List screen • The format is invalid. 300CL.
10432 CslAdv_Study Failed update Information Update Information set mode fail. Restart DR-ID 300CL. If the same
phenomenon recurred, initialize the image
database.
10454 ClsAdv_Tomo + ArgumentException An error occurred when acquiring the path Restart DR-ID 300CL. If the same
to image processing folder. phenomenon recurred, reinstall DR-ID
300CL.
10461 CslAdv_QA QACommon.dll Failed in determining whether to implement Preform the operation again. If the same
FunctionLib - the function of image arbitrary rotation or phenomenon recurred, restart DR-ID
CheckCondition not. 300CL.
EnableFunction -
<expcetion content>
10462 CslAdv_QA QA.exe Failed in resetting the image arbitrary Open the study and perform the operation
QAScreen rotation again.
ResetFreeRotateAngle If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
10463 CslAdv_QA QATrimmingCtrl.dll Failed in saving the information after Preform the operation again. If the same
TrimmingCtrl implement the image arbitrary rotation. phenomenon recurred, restart DR-ID
objDspImg_ 300CL.
ImageEditedEvent -
<exception content>
10464 CslAdv_QA QATrimmingCtrl.dll Failed in removing the trimming frame by Preform the operation again. If the same
TrimmingCtrl dragging the mouse or rotating the image phenomenon recurred, restart DR-ID
objDspImg_ arbitrarily. 300CL.
MouseMoveEvent -
<exception content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-86
Event Event Source Description Causes Remedies Remarks
10465 CslAdv_QA QATrimmingCtrl.dll Failed in enabling the density optimizing Open the study and perform the operation
TrimmingCtrl button for the trimming function. again.
SetStatusButtonOptimize If the same phenomenon recurred, restart
Density - DR-ID 300CL.
<exception content>
10500 CslAdv_Re Warning log relate with GUI An error occurred in displaying the window. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
10500 [TS10500] Display layout of Display layout of images could not be Check the status of PPU and the
images could not be changed. connection with PPU.
changed.
10500 CslAdv_QA QADBAccessor.dll A database access error occurred. Restart the application. If the same 1%: Spot error occurred
to QADBAccessor phenomenon recurred, initialize the
10509 1% database.
10501 [TS10501] Failed in Failed in switching display of images. Check the status of PPU and the
switching display of connection with PPU, and switch the menu
images. display again.
10565 FFQRP Create file QRPFinish.xml An exception occurred while creating Check the permission folder again.
failed QRPFinish.xml file.
10600 [TS10600] An error An error occurred while loading the data Check the status of PPU and the
occurred while loading the object. connection with PPU.
data object.
10601 CslAdv_Re Does not have a normal The selected studies contain invalid Restart DR-ID 300CL. If the same
image exist in a study. images. phenomenon recurred, reinstall DR-ID
Cannot continue 300CL.
processing.
10601 [TS10601] An error An error occurred while updating the DDO Check the DDO file, and execute the
occurred while updating the file. processing again.
DDO file.
10603 CslAdv_Re Failt in get Posession Possession has been used by other Restart DR-ID 300CL.
service engineers.
10606 CslAdv_Re No neither received IDAutoConvert failed in generating the Restart DR-ID 300CL.
messages content XML file.
10609 CslAdv_Re The QC target does not • No DDO files exist. Check that the QC data of the object are a
meet the Start QC • An error occurred in opening DDO file. CR image.
conditions. • The target is DR image. Implement saving of the QC data again.
10610 CslAdv_Re The free space is not Hard disk space required for performing Check the free space of the hard disk.
enough to save QC. SaveQC is insufficient. Delete unnecessary studies from the
application.
10623 CslAdv_QA QAFCRImgParamChgCtrl. Set Param list failed. Perform the operation again.
dll If the same phenomenon recurred, restart
FCRImgParamChgCtrl DR-ID 300CL.
(SetParamList) - <exception
content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-87
Event Event Source Description Causes Remedies Remarks
10628 CslAdv_QA QAFCRImgParamChgCtrl. Get Param list failed. Perform the operation again.
dll If the same phenomenon recurred, restart
FCRImgParamChgCtrl DR-ID 300CL.
(GetParamList) -
<exception content>
10633 CslAdv_Re Failed in accessing into the • The path to the folder which stores the Restart DR-ID 300CL. If the same
DDO files for update to DDO file does not exist. phenomenon recurred, reinstall DR-ID
start study from study list • No DDO files exist. 300CL.
• DDO files are being to the read only file.
10634 CslAdv_Re Failed in copying the study An error occurred during the copy of the Restart DR-ID 300CL.
data (This event is the log data to save QC images.
at the View hierarchy)
10635 CslAdv_Re No authority to access the No authority to write the data to the saving Restart DR-ID 300CL.
saving destination folder destination folder.
(This event is the log at the
View hierarchy)
10636 CslAdv_Re File was not found (This An error occurred when accessing the file. Restart DR-ID 300CL.
event is the log in layer
view)
10637 CslAdv_Re File was not found (This An error occurred when accessing the Restart DR-ID 300CL.
event is the log at the View DDO file of the study acquiring the
hierarchy) information for saving QC images.
10638 CslAdv_Re Failed in creating a new An error occurred when creating the folder Restart DR-ID 300CL.
folder (This event is the log to save QC images.
at the View hierarchy)
10639 CslAdv_Re Failed in copying a file An error occurred during the copying of a Restart DR-ID 300CL.
(This event is the log at the file from the saving source folder to the
View hierarchy) saving destination folder.
10639 CslAdv_QA QAFCRImgParamChgCtrl. CalculateChangeValue is failed. Perform the operation again.
dll If the same phenomenon recurred, restart
FCRImgParamChgCtrl DR-ID 300CL.
(CalculateChangeValue)
- <exception content>
10640 CslAdv_Re File copying process has An error occurred during the copying of a Save QC data again.
not been completed (but file from the saving source folder to the
one or more copied files saving destination folder.
exist) (This event is the log
at the View hierarchy)
10700 [TS10700] The The reconstruction conditions cannot be Check the DDO file, and execute the
reconstruction conditions obtained from the DDO file. processing again.
cannot be obtained from
the DDO file.
10701 [TS10701] Invalid Invalid reconstruct condition exists. Check the reconstruction conditions, and
reconstruct condition exists. execute the reconstruction processing
Please input reconstruct again.
conditions again.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-88
Event Event Source Description Causes Remedies Remarks
10702 [TS10702] The parameter The parameter value to register the Check the parameters, and execute the
value to register the reconstruction by PPU is invalid. reconstruction processing again.
reconstruction by PPU is
invalid.
10740 FFQRP This image is not supported An image which is not supported exists in Confirm whether the tomosynthesis image
by system. the study. or the image exposed with long view menu
for outputting individually in the study
exists or not.
10801 CslAdv_Re Failed in update customize • The CustomizeStudyListInfo.xml file does Restart DR-ID 300CL. If the same
study list configuration file not exist. phenomenon recurred, reinstall DR-ID
• The file is being to the read only file. 300CL.
10801 [TS10801] An error An error occurred while monitoring job Check the status of PPU and the
occurred while monitoring status. connection with PPU.
job status.
10850 CslAdv_Re Update customize study list An error occurred in writing the Restart DR-ID 300CL. If the same
is failed "CustomizeStudyListInfo.xml" file. phenomenon recurred, reinstall DR-ID
300CL.
10865 CslAdv_QA QAFNCCRFParamChgCtrl. Failed in reflecting the changed VGP Open the study and perform the operation
dll processing parameter. again.
FNCCRFParamChgView If the same phenomenon recurred, restart
ApplyParamVGP- DR-ID 300CL.
10866 CslAdv_QA QAFNCCRFParamChgCtrl. Failed in determining whether to implement Open the study and perform the operation
dll the VGP processing or not. again.
FNCCRFParamChgView If the same phenomenon recurred, restart
CheckVGPCondition- DR-ID 300CL.
10867 CslAdv_QA QAFNCCRFParamChgCtrl. Failed in determining whether to implement Open the study and perform the operation
dll the VGP processing or not. again.
FNCCRFParamChgView If the same phenomenon recurred, restart
CheckVGPCondition- DR-ID 300CL.
10868 CslAdv_QA QAFNCCRFParamChgCtrl. Failed in determining whether to implement Open the study and perform the operation
dll the VGP processing or not. again.
FNCCRFParamChgView If the same phenomenon recurred, restart
CheckVGPCondition- DR-ID 300CL.
10869 CslAdv_QA QATrimmingCtrl.dll An error occurred to the VGP processing. Open the study and perform the operation
TrimmingCtrl again.
ProcessVGP - <exception If the same phenomenon recurred, restart
content> DR-ID 300CL.
10900 [TS10900] An error An error occurred while obtaining the Check the status of PPU and the
occurred while obtaining reconstruction status of the current job.
connection with PPU.
the reconstruction status of
the current job.
10901 [TS10901] Reconstruction Reconstruction processing could not be Check the status of PPU and the
processing could not be canceled. connection with PPU.
canceled.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-89
Event Event Source Description Causes Remedies Remarks
10903 [TS10903] The priority of The priority of reconstruction processing Check the status of PPU and the
reconstruction processing could not be changed. connection with PPU.
could not be changed.
10904 [TS10904] An error An error occurred while reconstructing the Check the status of PPU and the
occurred while current job again. connection with PPU.
reconstructing the current
job again.
10946 CslAdv_QA QATrimmingCtrl.dll Failed in displaying the rotation angle of Open the study and perform the operation
TrimmingCtrl the image. again.
DisplayFreeRotate - If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
10947 CslAdv_QA QATrimmingCtrl.dll Failed in starting the image arbitrary Open the study and perform the operation
TrimmingCtrl rotation function. again.
HandleWhenClickBtnFree If the same phenomenon recurred, restart
Rotate - <exception DR-ID 300CL.
content>
10997 CslAdv_QA QAESImageCtrl.dll Failed in reprocessing the ES image. Preform the operation again. If the same
ESImageView phenomenon recurred, restart DR-ID
RunESProc - <exception 300CL.
content>
11000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart the application. If the same %1: File name
Method=%3, %4 phenomenon recurred, reinstall the %2: Class name
system. %3: Method name
%4: Detail message
11000 [MD11000] System error System error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
11000 ClsAdv_Tomo CslAdv_AC_ An error occurred in TomoMenuDispCtrl. Restart DR-ID 300CL.
GuiAdp:GeneralFailure dll. If the same phenomenon recurred, reinstall
processID:5860 Cannot load mis-exposure icon. DR-ID 300CL.
MutexDBConnection() retry
out
System.Data.OleDb.
OleDbException: Could not
find file 'c:\Program files\
FujiFilm\IIP\GUI\Icon\Icon.
mdb'.
11000 CslAdv_QA QADeliverySetting.dll/ Failed in output processing. Check the output device status and 1%: Spot error occurred
to 1% perform a process again.
11099
11001 [MI11001] Failed in Failed in initialization of dose area product Check the dose area product meter.
initialization of dose area meter.
product meter.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-90
Event Event Source Description Causes Remedies Remarks
11001 FFIdAutoConvert * A method name and a Failed in manipulating the data. Restart DR-ID 300CL. * This part is
failed process name are If the same phenomenon recurred, reinstall appropriated from
output depending on the DR-ID 300CL. V4.0.
error spot.
11002 [MD11002] A system A system communication error has been Restart the system.
communication error has detected. If no image is displayed on the console
been detected. even after making x-ray exposures,
perform re-exposure as necessary.
11003 [MD11003] Failed in setting Failed in setting patient information. Check the device connection status.
patient information.
11004 [MD11004] Failed in Failed in communication with the Check the device.
communication with the connected device.
connected device.
11005 [MD11005] No response No response comes from the connected Check the device.
comes from the connected device.
device.
11006 [MD11006] The operation The operation failed. Confirm that no image is being input.
failed. Confirm the connection with device.
11010 [MD11010] The status The status displaying part is unable to Restart the system.
displaying part is unable to reflect the system status. If no image is displayed on the console
reflect the system status. even after making x-ray exposures,
perform re-exposure as necessary.
11011 [MD11011] Failed in Failed in obtaining information from the Check the Image Reader status.
obtaining information from Image Reader.
the Image Reader.
11012 [MD11012] Failed in Failed in obtaining the Image Reader Check the Image Reader status.
obtaining the Image Reader status information.
status information.
11013 [MD11013] Failed in Failed in obtaining the Image Reader error Restart the system.
obtaining the Image Reader information. If the same error recurs after restarting the
error information. system, reinstall the system.
11016 [MD11016] Remaining Remaining battery power is low Charge immediately to avoid losing
battery power is low images.
11017 [MD11017]No remaining No remaining battery power. Exposure cannot be performed any longer.
battery power. Charge the battery.
11018 [MD11018] An image An image reader with low remaining Check the Image Reader status.
reader with low remaining battery power exists.
battery power exists.
11019 [MD11019] An image An image reader without remaining battery Check the Image Reader status.
reader without remaining power exists.
battery power exists.
11020 [MD11020] Failed in Failed in controlling the Image Reader Check the Image Reader status.
controlling the Image conditions.
Reader conditions.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-91
Event Event Source Description Causes Remedies Remarks
11021 [MD11021] The command The command requested to the Image Check the Image Reader status.
requested to the Image Reader was not executed.
Reader was not executed.
11021 [MI11021] Connection line Connection line with the X-ray unit has Confirm the connected device.
with the X-ray unit has been cut off.
been cut off.
11024 [MD11024] Connection to Connection to X-ray unit has been Check the X-ray unit status.
X-ray unit has been disconnected.
disconnected.
11030 [MD11030] The specified The image could not be acquired. Check that the exposure in high-sensitivity
duration in which High mode has not exceeded the specified
Sensitivity Exposure can be possible implementation time.
performed has elapsed. Check that X-ray has been detected.
11040 [MD11040] A error occurred A error occurred while in the process of Confirm the status of the panel connection.
while in the process of obtaining images.
obtaining images.
11041 [MD11041] Images saved Images saved in the FPD may not have If the image is remained in the panel,
in %s may not have been been cleared. re-input the image and clear the image of
cleared. the panel.
11042 [MD11042] The images The images exposed by memory exposure If the image is remained in the panel,
exposed by memory mode could not be obtained from FPD. re-input the image.
exposure mode could not
be obtained from FPD.
11043 [MD11043] ImageBox is The memory of the ImageBox is not Import the image from the ImageBox of
full. enough. study screen, and increase the space.
11100 [TS11100] Failed in Failed in integrating image display. Check the status of PPU and the
integrating image display. connection with PPU.
11100 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
to TomoImgDispOperationCtrl.dll. If the same phenomenon recurred, reinstall error and its cause
11104 An error occurred in the processing of DR-ID 300CL.
displaying images.
11103 CslAdv_Re Failed in Initialize Fris Unable to connect to the FRIS system. Check that the FRIS client has been set
Client up.
11104 CslAdv_Re Failed in sending report to • Unable to connect to the FRIS system. Check the following configuration items
FRIS Client. • A report data is invalid. related to FRIS connection.
• Item 44 “Use Series Identification From
F-RIS” of “1. IMAGE MODALITY”
11108 CslAdv_Re Failed in switching current An error occurred in switching the current Restart DR-ID 300CL.
user user in the FRIS window.
11210 CslAdv_Re Failed in initializing the Failed in initializing the RisConFig xml file. Restart DR-ID 300CL. If the same
Synapse RIS screen. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-92
Event Event Source Description Causes Remedies Remarks
11211 CslAdv_Re Failed in reading Failed in reading the result data from RIS. Restart DR-ID 300CL. If the same
information is outputed phenomenon recurred, reinstall DR-ID
from RIS 300CL.
11214 CslAdv_Re Failed in accessing the Unable to connect to DR-ID 300CL Restart DR-ID 300CL.
image database database. If the same phenomenon recurred, reinstall
DR-ID 300CL.
If the same phenomenon recurred, initialize
the image database.
11215 CslAdv_Re Failed in switching current An error occurred in switching the current Restart DR-ID 300CL.
user user.
11254 CslAdv_Study Failed change patient Notify change patient information. Restart DR-ID 300CL. If the same
information phenomenon recurred, reinstall DR-ID
300CL.
11304 CslAdv_Re Failed in request start study • An error occurred when sending Restart DR-ID 300CL. If the same
to SA:STUDY messages. phenomenon recurred, reinstall DR-ID
• A time out error occurred in sending 300CL.
messages.
11312 CslAdv_Re Failed in switch current An error occurred in switching the current Restart DR-ID 300CL.
user user.
11351 CslAdv_QA QATrimmingCtrl.dll Failed in acquiring the image information Open the study and perform the operation
TrimmingCtrl used for the VGP processing. again.
ReadSTDImageBuffer- If the same phenomenon recurred, restart
<exception content> DR-ID 300CL.
11400 [TS11400] Failed in Failed in obtaining information from the Check the DDO file and execute the
obtaining information from DDO file. processing again.
the DDO file.
11401 [TS11401] Patient Patient information is invalid. Check the status of PPU and the
information is invalid. connection with PPU, and restart the
system.
11403 CslAdv_Re Failed in accessing the An error occurred when accessing the Restart DR-ID 300CL.
DDO file. DDO file of the study acquiring the
information for saving QC images.
11404 CslAdv_Re Failed in creating a new An error occurred when creating the folder Restart DR-ID 300CL.
folder. to save QC images. Recheck the setting for the QC data
storage destination.
11405 CslAdv_Re No authority to access the No authority to save QC images to a Restart DR-ID 300CL.
folder. device (USB, DVD, CD, shared network). Recheck the setting for the QC data
storage destination.
11406 CslAdv_Re File not found. An error occurred when accessing the file. Restart DR-ID 300CL.
11407 CslAdv_Re Failed in copying a file. An error occurred when copying a file from Restart DR-ID 300CL.
the saving source folder to the saving
destination folder.
11408 CslAdv_Re Failed in deleting a file. An error occurred when accessing the file Restart DR-ID 300CL.
to be deleted.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-93
Event Event Source Description Causes Remedies Remarks
11410 CslAdv_Re Failed in logging in to the An error occurred when logging in to the Restart DR-ID 300CL.
host. shared host.
11411 CslAdv_Re The result returned from An error occurred when confirming the Restart DR-ID 300CL.
the host was not received. connection.
11412 CslAdv_Re Can not register patient Failed to register patient data in local DB. Restart DR-ID 300CL.
data
11413 CslAdv_Re An error of uncertain cause An unforeseen error occurred in displaying If the general-purpose RIS browser still
occurs during RIS browser the general-purpose RIS browser. doesn't display after selecting it again,
display. restart DR-ID 300CL.
11451 CslAdv_QA QAMaskCtrl.dll Error when click apply button. Perform the operation again.
MaskView If the same phenomenon recurred, restart
(btnApplyClearPreview_ DR-ID 300CL.
Click) - exception content>
11512 CslAdv_Study Cacel align mammo image It fails when cancel align mammo image or Restart DR-ID 300CL. If the same
or update display menu when display overlay. phenomenon recurred, reinstall DR-ID
error 300CL.
11520 CslAdv_Study Failed in get menu It fails when get information about Restart DR-ID 300CL. If the same
information irradiation area, size, position, direction, phenomenon recurred, reinstall DR-ID
irradiation link in form. 300CL.
11520 CslAdv_Study Show form error It fails when show form. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
11521 CslAdv_Re Create XML file not Failed in sending the generated information Restart DR-ID 300CL.
completed (This event is to IDAutoConvert.
log in layer view)
11521 CslAdv_Re Read XML result failure Failed in sending the generated information Restart DR-ID 300CL.
(This event is log in layer to IDAutoConvert.
view)
11521 CslAdv_Study Display web page error Display web page fail. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
11522 CslAdv_Study Get current select menu It fails when get current select menu Restart DR-ID 300CL.
info error information fail. If the same phenomenon recurred, initialize
the image database.
11523 CslAdv_Study Failed auto scroll It fails when auto scroll. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
11525 CslAdv_Study Failed start next study It fails when Generate Data Key For Next Restart DR-ID 300CL. If the same
Study. phenomenon recurred, reinstall DR-ID
300CL.
11526 CslAdv_Study dxPicDeliveryStatus_ An error occurred when clicking the Restart DR-ID 300CL.
MouseDown error distribution status. If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-94
Event Event Source Description Causes Remedies Remarks
11527 CslAdv_Study dxPicDeliveryStatus_ An error occurred when moving the cursor Restart DR-ID 300CL.
MouseMove error on the distribution status. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11528 CslAdv_Study dxPicDeliveryStatus_ Failed in displaying the distribution status. Restart DR-ID 300CL.
MouseUp error If the same phenomenon recurred, reinstall
DR-ID 300CL.
11529 CslAdv_Study ReceiveDeliveryResultProc Failed in receiving the acquisition request. Restart DR-ID 300CL.
error If the same phenomenon recurred, reinstall
DR-ID 300CL.
11530 CslAdv_Study cbDeliveryStop_Click error Failed in stopping distribution. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
11531 CslAdv_Study cbDeliverySeparate_Click Failed in distribution separation. Restart DR-ID 300CL.
error If the same phenomenon recurred, reinstall
DR-ID 300CL.
11532 CslAdv_Study DeleteUserOperationEx Failed in deleting the user study operation Restart DR-ID 300CL.
error table. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11533 CslAdv_Study BuildSANYOConfigElement Failed in building the SANYO configuration Restart DR-ID 300CL.
error element. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11534 CslAdv_Study BuildManualElement error Failed in building the manual configuration Restart DR-ID 300CL.
element. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11535 CslAdv_Study Ris_OnReceiveOrder error Failed in receiving the order information Restart DR-ID 300CL.
from RIS. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11536 CslAdv_Study UpdateIpNumber Cannot find menu for registering/removing Reselect the menu. %1: Notice factor
GetMenuIndex()-"%1" IP number. If the same phenomenon recurred, restart
DR-ID 300CL.
11540 CslAdv_Study ManualAdjustMammo error Failed in manual adjustment of the mammo Restart DR-ID 300CL.
image. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11541 CslAdv_Study Error when click button An error occurred when clicking the QA Restart DR-ID 300CL.
QAshorcut shortcut button. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11542 CslAdv_Study ChangePatInforForFris Failed in changing the patient information Restart DR-ID 300CL.
of RIS order. If the same phenomenon recurred, reinstall
DR-ID 300CL.
11545 CslAdv_Study Get Tag Free Rotation error Failed in processing the initialization Restart DR-ID 300CL.
setting for image arbitrary rotation function. If the same error recurs after restarting,
reinstall DR-ID 300CL.
11546 CslAdv_Study Calculate New Image Failed in handling the click process for the Restart DR-ID 300CL.
Processing Type error image processing type switching button. If the same error recurs after restarting,
reinstall DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-95
Event Event Source Description Causes Remedies Remarks
11551 CslAdv_Study Switch Image Proc Type Failed in handling the click process for the Restart DR-ID 300CL.
error image processing type switching button. If the same error recurs after restarting,
reinstall DR-ID 300CL.
11554 CslAdv_Study GetMaskDensity Failed to get paintout density of shadow- Restart DR-ID 300CL. %1: Exception
%1 masking processing. If the same error recurs after restarting, information
reinstall DR-ID 300CL.
11555 CslAdv_Study Mouse up event of the Failed in switching between GA and GS. Restart DR-ID 300CL.
processed image If the same error recurs after restarting,
reinstall DR-ID 300CL.
11568 CslAdv_Study StMainGUI.dll The EDR process failed when performing Specify an area included exposure field as
Mediator exposure with the Long Panel. ROI. If the same error recurs, restart the
ExecuteTemplate- The ROI may not be properly specified at DR-ID 300CL.
FCDModImgPreProcEdr the ROI setting dialogue when performing
WithMaskImg’s return code exposure with the Long Panel.
: -*
[* is arbitrary value]
11568 CslAdv_Study StMainGUI.dll The exposure field recognition processing Specify an area included exposure field as
Mediator failed when performing exposure with the ROI. If the same error recurs, restart the
ExecuteTemplate- Long Panel. DR-ID 300CL.
FCDModImgPreProc The ROI may not be properly specified at
BSDetection’s return code : the ROI setting dialogue when performing
-* exposure with the Long Panel.
[* is arbitrary value]
11604 CslAdv_QA QACommon.dll Set Overlay Data fail. Perform the operation again.
AdjustImgCtrl - If the same phenomenon recurred, restart
(ReleaseAdjustment) - DR-ID 300CL.
<expcetion content>
11605 CslAdv_Re Save MassOrder Settings Failed in saving the MassOrderDB. Restart DR-ID 300CL. If the same
failed phenomenon recurred, reinstall DR-ID
300CL.
11613 CslAdv_Re Could not get file An error occurred while retrieving the study Restart DR-ID 300CL.
information by procUID comment information from DDO.
11617 CslAdv_Re Access MenuData.mdb An error occurred while mapping the study Restart DR-ID 300CL. If the same
failed data at the startup of DR-ID 300CL. phenomenon recurred, reinstall DR-ID
300CL.
11621 CslAdv_Re Delete history failed An error occurred while deleting the Restart DR-ID 300CL. If the same
IDHistoryDB table at the startup of DR-ID phenomenon recurred, reinstall DR-ID
300CL. 300CL.
11701 CslAdv_Re Reserve study failure Failed in reserving studies. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
11729 CslAdv_QA IIPQACtrl.dll/ Failed in user group information changing Check the selected user ID and perform a
QAScreen/ process. process again.
patientInfoChgLimit_
OnOk_Click
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-96
Event Event Source Description Causes Remedies Remarks
11850 CslAdv_Re Failed in accessing into the A siteCode mapping error occurred while Check the mapping tag of siteCode again.
console db starting a study or starting a next study.
11855 CslAdv_Re Reading of patient ID auto- • Failed in acquiring of the setting of the Check the format of IssuanceIdMassOrder.
issuance setting file for IssuanceIdMassOrder.dat. dat.
screen failed. • Failed in checking the appropriateness of
IssuanceIdMassOrder.dat.
11856 CslAdv_Re The format of the patientID Setting of the AUTOINC tag of Check the format of IssuanceIdMassOrder.
auto-issuance setting file IssuanceIdMassOrder.dat is inappropriate. dat.
for screening is incorrect
12000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
12000 [TS12000] An error An error occurred while loading the DDO Check the DDO file, and execute the
occurred while loading the file. processing again.
DDO file.
12000 ClsAdv_Tomo 1% An error occurred in TomoDataMgrCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
to An error occurred in the image DataEntity error and its cause
12012 control function.
12001 [MI12001] Failed in Failed in initialization of dose area product Check the dose area product meter.
initialization of dose area meter.
product meter.
12001 [TS12001] An error An error occurred while displaying images. Check the status of PPU and the
occurred while displaying connection with PPU.
images.
12003 [MI12003] No response No response comes from the device that Confirm the connected device.
comes from the device that shares ID information.
shares ID information.
12004 [MI12004] A Study can't A Study can't register because the Confirm the connected device.
register because the exposure menu registered in IDT isn't
exposure menu registered registered in IIP.
in IDT isn't registered in IIP.
12005 [MI12005] Failed in Failed in communication with the Confirm the connected device.
communication with the connected device.
connected device.
12006 [MI12006] Failed in setting Failed in setting exposure conditions. Confirm the connected device.
exposure conditions.
12007 [MI12007] Failed in setting Failed in setting exposure conditions. Confirm the connected device.
exposure conditions.
12012 [MI12012] The command The command requested to the Image Check the Image Reader status.
requested to the Image Reader was not executed.
Reader was not executed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-97
Event Event Source Description Causes Remedies Remarks
12013 [MI12013] Failed in Failed in obtaining information from the Check the Image Reader status.
obtaining information from Image Reader.
the Image Reader.
12014 [MI12014] Failed in Failed in obtaining information from the Check the Image Reader status.
obtaining information from Image Reader.
the Image Reader.
12015 [MI12015] Failed in Failed in obtaining information from the Check the Image Reader status.
obtaining information from Image Reader.
the Image Reader.
12016 [MI12016] Failed in Failed in controlling the Image Reader Check the Image Reader status.
controlling the Image conditions.
Reader conditions.
12017 [MI12017] The connected The connected Image Reader (%s) went When you are using the Image Reader,
Image Reader (%s) went into suspend mode because it had not turn the power ON again.
into suspend mode been used for a certain period.
because it had not been
used for a certain period.
12018 [MI12018] Power to the Power to the connected Image Reader When you are re-using the Image Reader,
connected Image Reader (%s) has been turned OFF. turn the power ON again.
(%s) has been turned OFF.
12020 [MI12020] The line to the The line to the device (%s) has been Confirm the connected device.
device (%s) has been disconnected.
disconnected.
12021 [MI12001] Failed in Failed in initialization of dose area product Confirm the connected device.
initialization of dose area meter.
product meter.
12022 CslAdv_QA QACommunication Failed in synchronizing the TomoView and Restart DR-ID 300CL.
Handler.dll study status.
SynchronizeStudyInfo
- <exception content>
12022 [MI12022] Failed in setting Failed in setting exposure conditions. Click the button and restart the system.
exposure conditions. If the same error recurs after restarting the
system, reinstall the system.
12025 CslAdv_QA QACommunication Failed in synchronizing the TomoView and Restart DR-ID 300CL.
Handler.dll study status.
SynchronizeStudy
Operation - <exception
content>
12027 CslAdv_QA QACommunicationHandler. Failed in sending /irs/contents/refresh. Restart DR-ID 300CL.
dll
CommunnicationHandler
(SendMsgRefreshIRS
Contents) - <exception
content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-98
Event Event Source Description Causes Remedies Remarks
12146 CslAdv_QA QACombineImageCtrl.dll Failed to acquire the image information Restart the DR-ID 300CL.
AdjustPositionView that completed with long-view composition If the same error recurs, reinstall the
GetDataObj- process. system.
Get Data Object is failed
12147 CslAdv_QA QACombineImageCtrl.dll The internal processing failed after Restart the DR-ID 300CL.
AdjustPositionView pressing the trimming screen transition If the same error recurs, reinstall the
btnTrimming_Click- button at the Long Panel 1stUP position system.
[according to the exceptions adjustment screen.
in internal processing]
12149 CslAdv_QA QACombineImageCtrl.dll The internal processing failed after Restart the DR-ID 300CL.
AdjustPositionView pressing the reset button at the Long Panel If the same error recurs, reinstall the
btnReset_Click- 1stUP position adjustment screen. system.
Reset button clicked for
Long Panel is failed
12229 CslAdv_QA QACombineImageCtrl.dll/ Failed in setup process of LUT files. Check the LUT files and perform a process
ImgViewCtrl/ again.
CrtLutFile. If the LUT file does not exist in "C:\Program
Files\FujiFilm\IIP\CrtLut\" folder, reinstall
again.
12248 CslAdv_QA QACombineImageCtrl.dll ShowFPD With Adjust Density is failed. Perform the operation again.
CombineImageCtrl - If the same phenomenon recurred, restart
(ShowFPDWithAdjust DR-ID 300CL.
Density) - <exception
content>
12249 CslAdv_QA QACombineImageCtrl.dll BackupData For Adjust S Value is failed. Perform the operation again.
CombineImageCtrl - If the same phenomenon recurred, restart
(BackupForAdjustSValue) DR-ID 300CL.
- <exception content>
12261 CslAdv_QA IIPQACtrl.dll/ Failed in acquiring the Config file Check the following Config file and perform
1% information. a process again.
"C:\Program Files\FujiFilm\IIP\Config\1%"
12270 CslAdv_QA QACombineImageCtrl.dll Failed to set the image information that Restart the DR-ID 300CL. If the same error
CombineImageCtrl completed with long-view composition recurs, reinstall the system.
SetOrgDataObjForLP- process of the initial status.
Set Original Data Object is
failed
12280 CslAdv_QA QACombineImageCtrl.dll Failed to display the image via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSelectionView dialogue when performing exposure with recurs, reinstall the system.
MultiImgVw Reset is faild the Long Panel.
12281 CslAdv_QA QACombineImageCtrl.dll Failed to display the image via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSelectionView dialogue when performing exposure with recurs, reinstall the system.
MultiImgVw SetSubMode is the Long Panel.
faild
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-99
Event Event Source Description Causes Remedies Remarks
12282 CslAdv_QA QACombineImageCtrl.dll Failed to display the image via ROI setting Restart the DR-ID 300CL.
LPAreaSelectionView dialogue when performing exposure with If the same error recurs, reinstall the
MultiImgVw the Long Panel. system.
SetDisplayMode is faild
12283 CslAdv_QA QACombineImageCtrl.dll Failed to display the image via ROI setting Restart the DR-ID 300CL.
LPAreaSelectionView dialogue when performing exposure with If the same error recurs, reinstall the
MultiImgVw SetDataObj is the Long Panel. system.
faild
12284 CslAdv_QA QACombineImageCtrl.dll The ROI setting process failed when Restart the DR-ID 300CL.
LPAreaSelectionCtrl performing exposure with the Long Panel. If the same error recurs, reinstall the
[according to the exceptions system.
in internal processing]
12468 CslAdv_QA QAOrderInfoChgCtrl.dll Failed in displaying images. Check the DDO files and perform a
StudyImageCommentView process again.
StudyImageCommentView
- <exception content>
12477 CslAdv_QA QAOrderInfoChgCtrl.dll Failed in displaying images. Check the DDO files and perform a
StudyImageCommentView process again.
StudyImageCommentView
- <exception content>
12498 CslAdv_QA QAOrderInfoChgCtrl.dll Failed in displaying images. Check the DDO files and perform a
StudyImageCommentView process again.
StudyImageCommentView
- <exception content>
12500 [TS12500] An error occurs An error occurs during image processing. Check the status of PPU and the
during image processing. connection with PPU.
12600 [TS12600] An error An error occurred while displaying images. Check the status of PPU and the
occurred while displaying connection with PPU.
images.
12601 ClsAdv_Tomo 1% An error occurred while displaying job Preform the operation again. 1%: The spot of the
setting dialog. If the same phenomenon recurred, restart error and its cause
DR-ID 300CL.
12601 ClsAdv_Tomo 1% An error occurred in TomoJobMgntCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
An error occurred in the managing error and its cause
processing of reconstruction queue.
12602 CslAdv_QA IIPQACtrl.dll/ Failed in acquiring the Config file Check the following Config file and perform
1% information. a process again.
"C:\Program Files\FujiFilm\IIP\Config\1%"
12602 ClsAdv_Tomo 1% An error occurred while changing the Preform the operation again. 1%: The spot of the
display layout. If the same phenomenon recurred, restart error and its cause
DR-ID 300CL.
12603 ClsAdv_Tomo 1% An error occurred while changing the Preform the operation again. 1%: The spot of the
display layout. If the same phenomenon recurred, restart error and its cause
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-100
Event Event Source Description Causes Remedies Remarks
12603 ClsAdv_Tomo 1% An error occurred in TomoJobMgntCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
An error occurred in the managing error and its cause
processing of reconstruction queue.
12604 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when modifying the
reconstruction conditions.
12605 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12605 [TS12605] Failed in copying Failed in copying the DDO file. Click [OK].
the DDO file.
12608 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the JPEG files. Check the path of the image files and
to RejectLogStorageCtrl/ perform a process again.
12624 ImageStorage - <exception
content>
12611 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12611 [TS12611] Failed in Failed in reconstruction processing. Check the status of PPU and the
reconstruction processing. connection with PPU.
12613 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12613 ClsAdv_Tomo 1% An error occurred in TomoJobMgntCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
to An error occurred in the managing error and its cause
12646 processing of reconstruction queue.
12615 [TS12615] Unable to Unable to execute the reconstruction Check the reconstruction conditions and
execute the reconstruction processing because the reconstruction execute the reconstruction processing
processing because the condition is invalid. again.
reconstruction condition is
invalid.
12616 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when displaying output
range.
12617 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when modifying the
reconstruction conditions.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-101
Event Event Source Description Causes Remedies Remarks
12618 CslAdv_QA QAOrderInfoChgCtrl.dll Failed in displaying images. Check the DDO files and perform a
StudyImageCommentView process again.
StudyImageCommentView
- <exception content>
12618 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when displaying output
range.
12619 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12620 [TS12620] A connection A connection error occurred with PPU. Click [Yes].
error occurred with PPU.
12621 [TS12621] Reconstruction Reconstruction processing cannot be Restart the system.
processing cannot be executed because unnecessary past data
executed because exists.
unnecessary past data
exists.
12622 [TS12622] Reconstruction Reconstruction processing cannot be Check the status of PPU and the
processing cannot be executed again. connection with PPU.
executed again. Check the reconstruction conditions and
execute the reconstruction processing
again.
12623 [TS12623] Reconstruction Reconstruction processing could not be Check the status of PPU and the
processing could not be executed. connection with PPU.
executed.
12624 ClsAdv_Tomo * String of the exception An error occurred while re-drawing the icon If the list of reconstruction jobs is not
occurred is output. of the exposed image retrieving status. displayed properly after a while, restart
DR-ID 300CL.
12624 [TS12624] Reconstruction Reconstruction processing could not be Check the status of PPU and the
processing could not be executed because the number of exposure connection with PPU.
executed because the images could not be obtained.
number of exposure images
could not be obtained.
12626 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12627 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-102
Event Event Source Description Causes Remedies Remarks
12628 CslAdv_QA QARejectLogStorage.dll/ Failed in saving the log files. Check the path of destination for log file
RejectLogStorageCtrl/ saving, and then execute the process
GetLogFolderPath - again.
<exception content>
12700 [TS12700] An error An error occurred while recovering job Check the status of PPU and the
occurred while recovering synchronization. connection with PPU.
job synchronization.
12701 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get lifesize imege pixel the Long Panel.
from ddo faild
12702 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get lifesize imege line from the Long Panel.
ddo faild
12703 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get pixelspacing from ddo the Long Panel.
faild
12704 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get StdImagePixel from the Long Panel.
ddo faild
12705 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get StdImageLine from ddo the Long Panel.
faild
12706 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get CombineImageRotation the Long Panel.
from ddo faild
12707 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get AutoCombinedInfo the Long Panel.
from ddo faild
12708 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get Image Info from the Long Panel.
DataObj faild
12709 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
LPAreaSizeCalcurator dialogue when performing exposure with recurs, reinstall the system.
Get DDO from DataObj the Long Panel.
faild
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-103
Event Event Source Description Causes Remedies Remarks
12710 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
TrimmingCtrlForLP dialogue when performing exposure with recurs, reinstall the system.
Get Number of Trimming the Long Panel.
Object from MultiImgVw
faild
12711 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
TrimmingCtrl dialogue when performing exposure with recurs, reinstall the system.
Get TrimmingObj from DDO the Long Panel.
faild
12712 CslAdv_QA QATrimmingCtrl.dll Failed to acquire the data via ROI setting Restart the DR-ID 300CL. If the same error
TrimmingCtrl dialogue when performing exposure with recurs, reinstall the system.
Add TrimmingObj to the Long Panel.
MultiImgVw faild
12801 ClsAdv_Tomo An error occurred in TomoPPUCommCtrl.dll. Check the connection to PPU server.
to An error occurred in the PPU connection. If the same phenomenon recurred, restart
12804 DR-ID 300CL.
12805 ClsAdv_Tomo An error occurred in TomoPPUCommCtrl.dll. Check the connection to PPU server.
to An error occurred in the PPU connection. Restart PPU server.
12815
12816 ClsAdv_Tomo An error occurred in TomoPPUCommCtrl.dll. Check the connection to PPU server.
An error occurred in the PPU connection. If the same phenomenon recurred, restart
DR-ID 300CL.
12817 ClsAdv_Tomo An error occurred in TomoPPUCommCtrl.dll. Check the connection to PPU server.
An error occurred in the PPU connection. If the same phenomenon recurred, restart
DR-ID 300CL.
12817 ClsAdv_Tomo Error occurs during An error occurred while updating the Restart DR-ID 300CL.
updating _reconstruction_ reconstruction status of images.
processing field in _image_
tbl
12819 ClsAdv_Tomo * String of the exception An error occurred while generating data for Restart DR-ID 300CL.
occurred is output. displaying the reconstruction jobs sent
from PPU.
12820 ClsAdv_Tomo * String of the exception An error occurred while generating data for Restart DR-ID 300CL.
occurred is output. displaying the reconstruction jobs sent
from PPU.
12821 ClsAdv_Tomo Set ReconstructionStatus An error occurred in the initialization Preform the operation again. If the same
fail processing of the re-reconstruction of the phenomenon recurred, restart DR-ID
retrieved jobs. 300CL.
12900 [TS12900] Failed in saving Failed in saving the image information. Check the status of PPU and the
the image information. connection with PPU.
12901 [TS12901] An error An error occurred while displaying images. Check the status of PPU and the
occurred while displaying connection with PPU.
images.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-104
Event Event Source Description Causes Remedies Remarks
12902 [TS12902] An error An error occurred while copying images. Check the status of PPU and the
occurred while copying connection with PPU.
images.
12903 [TS12903] Failed in saving Failed in saving information by Check the status of PPU and the
information by synchronization processing. connection with PPU, and save the
synchronization information again.
processing.
12904 [TS12904] A connection A connection error occurred with PPU. Please check PPU status and the
error occurred with PPU. connection status to PPU, and restart the
system.
13001 LngDBC Database Access Failure. An error occurred in ADO. Restart. If the same phenomenon recurred, %1: Function name.
%nFunction:%1 restart the PC. If the phenomenon still %2: The specified field
Command:%2%nErrNo:%3 recurred, reinstall it. name error
NativeErrNo:%4 occurred or the
SQLState:%5%n character string in
Source:%6%n SQL.
Description:%7 %3: The ADO number
(error number)
properties.
%4: The ADO
NativeError (unique
error code of the
provider)
properties.
%5: The ADO SQLState
(SQL status)
properties.
%6: The Source of
ADO (object name)
properties.
%7: The description of
ADO (object
description)
properties.
13001 FFIdAutoConvert * A method name and a The information is output when restoring - * This part is
failed process name are the DDO file. appropriated from
output depending on the V4.0.
error spot.
13002 LngDBC Database Access Failure. Failed in accessing the database. Restart. If the same phenomenon recurred, %1: Function name.
InstanceName restart the PC. If the phenomenon still %2: The specified field
[LngDBC]%nNo Error recurred, reinstall it. name error
Item%nFunction:%1 occurred or the
Command:%2 character string in
SQL.
13005 FFQRP ImageNumber = {0} => The DDO file does not exist or is damaged. Retrieve the image again.
invalid
13005 FFQRP Index of source item is The DDO file does not exist or is damaged. Retrieve the image again.
invalid. value = {0}
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-105
Event Event Source Description Causes Remedies Remarks
13005 FFQRP Index of destination item is The DDO file does not exist or is damaged. Retrieve the image again.
invalid. value = {0}
13005 FFQRP Number of items in The DDO file does not exist or is damaged. Retrieve the image again.
lstImageOrder not equal in
lstImageInfoForMultiViewer.
lstImageOrder.Count = {0},
lstImageInfoForMultiViewer.
Count = {1}
13005 FFQRP Can't read DDO file {0} to The DDO file does not exist or is damaged. Retrieve the image again.
get ImageOrder
13010 CslAdv_Study It failed that [%1] initializes Defective program. Restart. If the same phenomenon recurred,
[%2] reinstall the system.
13015 CslAdv_Study [%1] failed in setting Defective program. Restart. If the same phenomenon recurred,
DICOM data [%2] reinstall the system.
13016 CslAdv_Study [%1] failed in the acquisition Defective program. Restart. If the same phenomenon recurred,
of DICOM data [%2] reinstall the system.
13044 CslAdv_Study [%s] %s (%s) : Failed Failed in accessing the image database. Check that the SQL service manager is
ImageDB Open Recordset activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13045 CslAdv_Study [%s] %s (%s) : Failed Failed in accessing the image database. Check that the SQL service manager is
ImageDB Close Recordset activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13046 CslAdv_Study [%s] %s (%s) : ImageDB Failed in accessing the image database. Check that the SQL service manager is
Data Get Error activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13047 CslAdv_Study [%s] %s (%s) : ImageDB Failed in accessing the image database. Check that the SQL service manager is
Data Set Error activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13048 CslAdv_Study [%s] %s (%s) : ImageDB Failed in accessing the image database. Check that the SQL service manager is
Data Delete Error activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-106
Event Event Source Description Causes Remedies Remarks
13049 CslAdv_Study [%s] %s (%s) : Failed Failed in accessing the image database. Check that the SQL service manager is
ImageDBFlag Connection activating properly. If the same
phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13051 CslAdv_Study [%s] %s (%s) : Failed in accessing the image database. Check that the SQL service manager is
ImageDBFlag Data Set activating properly. If the same
Error phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13052 CslAdv_Study [%s] %s (%s) : Failed Failed in initializing the menu database Restart. If the same phenomenon recurred,
MenuDB Initialize access. reinstall the system.
13053 CslAdv_Study [%s] %s (%s) : Menu Data Failed in acquiring the menu information Restart. If the same phenomenon recurred,
in ImageDB Get Error from the menu database. reinstall the system.
13054 CslAdv_Study [%s] %s (%s) : Menu Failed in acquiring the menu parameter Restart. If the same phenomenon recurred,
Parameter Data in information from the menu database. reinstall the system.
ImageDB Get Error
13055 CslAdv_Study [%s] %s (%s) : Failed Failed in disconnecting the menu database Restart. If the same phenomenon recurred,
MenuDB Termination access. reinstall the system.
13059 CslAdv_Study [%s] %s (%s) : Failed Failed in creating new DICOM data. Restart. If the same phenomenon recurred,
Create New DICOM Data initialize the image database.
13060 CslAdv_Study [%s] %s (%s) : DICOM Failed in acquiring DICOM data. Restart. If the same phenomenon recurred,
Data Get Error initialize the image database.
13061 CslAdv_Study [%s] %s (%s) : DICOM Failed in setting up DICOM data. Restart. If the same phenomenon recurred,
Data Set Error initialize the image database.
13062 CslAdv_Study [%s] %s (%s) : DICOM Failed in deleting DICOM data. Restart. If the same phenomenon recurred,
Data Delete Error initialize the image database.
13063 CslAdv_Study [%s] %s (%s) : DICOM Failed in loading DICOM data. Restart. If the same phenomenon recurred,
Data Load Error initialize the image database.
13064 CslAdv_Study [%s] %s (%s) : DICOM Failed in saving DICOM data. Restart. If the same phenomenon recurred,
Data Save Error initialize the image database.
13065 CslAdv_Study [%s] %s (%s) : DICOM Failed in freeing DICOM data. Restart. If the same phenomenon recurred,
Data Free Error initialize the image database.
13066 CslAdv_Study %1 (%2) : Dicom Data Failed in copying DICOM data. Restart. If the same phenomenon recurred,
Duplicate Error[%3] initialize the image database.
13067 CslAdv_Study %1 (%2) : Dicom Data Failed in inserting DICOM data. Restart. If the same phenomenon recurred,
Push Error initialize the image database.
13068 CslAdv_Study %1 (%2) : Dicom Replace Failed in switching DICOM data. Restart. If the same phenomenon recurred,
Data Error initialize the image database.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-107
Event Event Source Description Causes Remedies Remarks
13080 CslAdv_Study [%s] %s (%s) : The image database was damaged. Check that the SQL service manager is
ImageDBFlag Data Get activating properly. If the same
Error phenomenon recurred even when the
system is activated properly, initialize the
image database.(All the image data will be
deleted)
13088 CslAdv_Study [%s] %s (%s) : Image The system was unstable. Restart. If the same phenomenon recurred,
Processing Trouble reinstall the system.
<InImg>
13093 CslAdv_Study [%s] %s (%s) : Priority Defective program. Restart. If the same phenomenon recurred,
Change Method Error reinstall the system.
<IdExposureList>
13098 LngDBC A WIN32 system library An error occurred in WINAPI. As an internal error occurred, restart DR- %1: The process ID
error occurred. [%1]%n%2 ID 300CL. and thread ID.
If the error occurred again, restart the PC. %2: The error message
If the error could not be recovered, reinstall including the error
the application. code.
13099 LngDBC An unknown error occurred. Unknown error. As an internal error occurred, restart DR- %1: The process ID
[%1]%n%2 ID 300CL. and thread ID.
If the error occurred again, restart the PC. %2: The error message
If the error could not be recovered, reinstall including the error
the application. code.
13101 ClsAdv_Tomo 1% An error occurred in TomoDBAccessor.dll. Restart DR-ID 300CL. 1%: The spot of the
to An database access error has occurred. If the same phenomenon recurred, reinstall error and its cause
13115 DR-ID 300CL.
13115 FFQRP File {0} does not exist The config file does not exist or is Reinstall Network Image Search function.
damaged.
13115 FFQRP Can't load data from {0} The config file does not exist or is Reinstall Network Image Search function.
damaged.
13115 FFQRP Load data from xml file is The config file does not exist or is Reinstall Network Image Search function.
fail damaged.
13115 FFQRP Item need to get is in wrong The config file does not exist or is Reinstall Network Image Search function.
format damaged.
13115 FFQRP Can not connect to system The config file does not exist or is Reinstall Network Image Search function.
database damaged.
13115 FFQRP Can not get data in The config file does not exist or is Reinstall Network Image Search function.
Dictionary.ini. damaged.
13115 FFQRP Passed value is null or The config file does not exist or is Reinstall Network Image Search function.
empty damaged.
13115 FFQRP DateTime input is invalid The config file does not exist or is Reinstall Network Image Search function.
damaged.
13115 FFQRP Can not get config element The config file does not exist or is Reinstall Network Image Search function.
from the xml File damaged.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-108
Event Event Source Description Causes Remedies Remarks
13115 FFQRP Can't get configuration The config file does not exist or is Reinstall Network Image Search function.
value at key '{0} damaged.
13115 FFQRP Can't PN convert patient The config file does not exist or is Reinstall Network Image Search function.
name damaged.
13117 CslAdv_Study [%s] %s (%s) : Image UID Defective program. Restart. If the same phenomenon recurred,
Issue Error reinstall the system.
13118 CslAdv_Study [%s] %s (%s) : Series Defective program. Restart. If the same phenomenon recurred,
Instance UID Issue Error reinstall the system.
13119 CslAdv_Study [%s] %s (%s) : Failed in Defective program. Restart. If the same phenomenon recurred,
Auto Distribution reinstall the system.
13201 ClsAdv_Tomo + Error occurs during An error occurred when acquiring xml Confirm DDO files and image files.
calling parameters.
ACCModImgProcCtrl.
ACCModImgProcGetter
method
13202 ClsAdv_Tomo + Error occurs during An error occurred when calling out Confirm DDO files and image files.
calling FCDModImageProc. BC:FCDModImgProc and perform FCR
FCDModImgProc method image processing.
13202 ClsAdv_Tomo + Error occurs during An error occurred when calling out Perform the operation again. If the
calling FCDAppImgPre BC:FCDAppImgPreProcGammaTbl and operation cannot be performed, restart
Proc.FCDAppImg create GammaLUT. DR-ID 300CL.
PreProcGammaTbl method
13203 ClsAdv_Tomo + COMException An error occurred at FCR image Perform the operation again. If the
processing. operation cannot be performed, restart
DR-ID 300CL.
13205 FFQRP Can not load ddo file path The DDO file does not exist. Retrieve the study again.
13205 FFQRP There is not enough HDD Not enough memory in harddisk. Delete unnecessary studies from the DR-
memory to retrive {0} ID 300CL.
images
13205 FFQRP Process not success Failed in sending the process message. Restart DR-ID 300CL.
13205 FFQRP Communication: Request Failed in requesting a queue. Restart DR-ID 300CL.
queue fail
13205 FFQRP Communication: Inform Failed in notifying the status. Restart DR-ID 300CL.
status fail
13205 FFQRP Fail to save xml file that Failed in saving the xml files. Restart DR-ID 300CL.
notify to Recept.
13205 FFQRP Get path from directoryDef. Failed in acquiring the path from Restart DR-ID 300CL.
xml failed. directoryDef.xml.
13205 FFQRP Fail to get image processor Failed in acquiring the image processor Restart DR-ID 300CL.
type. type.
13205 FFQRP Can not load temp study Can not read the study temporarily. Restart DR-ID 300CL.
13205 FFQRP Patient ID and Name are Patient ID and name are invalid. Restart DR-ID 300CL.
null
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-109
Event Event Source Description Causes Remedies Remarks
13206 ClsAdv_Tomo + Error occurs during An error occurred when acquiring xml Confirm DDO files.
calling ACCAppImgPre parameters corresponding to the
ProcCtrl.ACCAppImg processing of GammaLUT creation.
PreProcGammaTblGetter
method
13208 ClsAdv_Tomo + COMException An error occurred when performing image Perform the operation again. If the
processing and creating GammaLUT. operation cannot be performed, restart
DR-ID 300CL.
13209 ClsAdv_Tomo + Error occurs during An error occurred when acquiring Perform the operation again. If the
getting Image BufferSize ImageBufferSize. operation cannot be performed, restart
DR-ID 300CL, and perform the operation
again.
13212 ClsAdv_Tomo Error occurs during turning An error occurred when turning off Confirm DDO files and image files.
off FNC flag xmlParam’s FNC flag.
13213 ClsAdv_Tomo + ArgumentException An error occurred when acquiring HDD’s Restart DR-ID 300CL.
+ SecurityException free space.
+ PathTooLongException
+ ArgumentException
+ IOException
13214 ClsAdv_Tomo + COMException An error occurred when changing the Perform the operation again. If the
image version UID. operation cannot be performed, restart
DR-ID 300CL.
13215 ClsAdv_Tomo + COMException An error occurred when performing Perform the operation again. If the
RemoveMaskTags. operation cannot be performed, restart
DR-ID 300CL.
13305 FFQRP Error start query. • Failure in connecting to PACS server. Confirm whether or not PACS server
• Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13305 FFQRP Error cancel query • Failure in connecting to PACS server. Confirm whether or not PACS server
• Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13305 FFQRP Error start retrieve image. • Failure in connecting to PACS server. Confirm whether or not PACS server
• Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13305 FFQRP Error cancel retrieve image. • Failure in connecting to PACS server. Confirm whether or not PACS server
• Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13305 FFQRP Error in load retrieve folder • Failure in connecting to PACS server. Confirm whether or not PACS server
path. • Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13305 FFQRP Error in retrieve image • Failure in connecting to PACS server. Confirm whether or not PACS server
process. • Incorrect SOP class settings. functions normally. And, confirm the
connection settings at the Service Utility.
13307 CslAdv_Study [%s] %s (%s) : Control Defective program. Restart. If the same phenomenon recurred,
Initialize Error reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-110
Event Event Source Description Causes Remedies Remarks
13310 CslAdv_Study [%s] %s (%s) : Unmatched Defective program. Restart. If the same phenomenon recurred,
Image ID reinstall the system.
13345 CslAdv_Study %1 (%2) : Database Data The database was damaged. Restart. If the same phenomenon recurred,
Get Error%n reinstall the system.
Database [%3]%n
Table [%4]%n
Key [%5]
13348 CslAdv_Study %1 (%2) : Registry Data The registry was damaged. Restart. If the same phenomenon recurred,
Get Error%n reinstall the system.
Value [%3]
13349 CslAdv_Study %1 (%2) : INI File Data Get The Directory.ini was damaged. Restart. If the same phenomenon recurred,
Error%n reinstall the system.
Key [%3]
13356 CslAdv_Study %1 (%2) : Control Initialize Defective program. Restart. If the same phenomenon recurred,
Error [%3] reinstall the system.
13383 CslAdv_Study %1 (%2) : Failed in Defective program. Restart. If the same phenomenon recurred,
terminating process.%n reinstall the system.
Component Name [%3]
13388 CslAdv_Study %1 (%2) : Failed in event Defective program. Restart. If the same phenomenon recurred,
reception. reinstall the system.
13389 CslAdv_Study %1 (%2) : Database The database was damaged. Restart. If the same phenomenon recurred,
Connection Error%n reinstall the system.
Database [%3]%n
13410 FFQRP Can't Load ImageError file : The resource file does not exist or is Reinstall Network Image Search function.
{0} damaged.
13410 FFQRP List printer is NULL, check The displayed printer is invalid. Reinstall Network Image Search function.
setting printer
13410 FFQRP Get printer list from Failed in reading the printer setting from Reinstall Network Image Search function.
database fail. db.
13410 FFQRP Delivery process fail Failed in delivery process. Reinstall Network Image Search function.
13410 FFQRP {0} not found. Cannot find a setting. Reinstall Network Image Search function.
13410 FFQRP Can not create a screen The resource file does not exist or is Reinstall Network Image Search function.
capture bitmap image damaged.
13425 FFQRP SearchForm: Could not The config file does not exist or is Reinstall Network Image Search function.
load config of user utility damaged.
13425 FFQRP SearchForm: Load list of The connection settings of PACS server Set the connection settings for PACS
PACS server failed are incorrect, or the config file does not server in Service Utility and restart DR-ID
exist or is damaged. 300CL. If the same error recurs after
restarting, reinstall Network Image Search
function.
13435 CslAdv_Study %1 (%2) : Failed to update Failed in acquiring the DICOM date. Restart DR-ID 300CL. If the same
implant and vmcs tag phenomenon recurred,
initialize the image database.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-111
Event Event Source Description Causes Remedies Remarks
13435 FFQRP Could not save data to The config file does not exist or is Reinstall Network Image Search function.
config file damaged.
13436 CslAdv_Study %1 (%2) : Failed to update Failed in acquiring the DICOM date. Restart DR-ID 300CL. If the same
implant tag phenomenon recurred,
initialize the image database.
13440 FFQRP Date value in config file is The config file does not exist or is Reinstall Network Image Search function.
invalid damaged.
13445 CslAdv_Study %1 (%2) : Update implant Failed in acquiring the DICOM date. Restart DR-ID 300CL. If the same
present tag and VMCS tag phenomenon recurred,
failed initialize the image database.
13445 FFQRP Not found config file: QRP. The config file does not exist or is Reinstall Network Image Search function.
config damaged.
13445 FFQRP UserUtility: The string with The config file does not exist or is Reinstall Network Image Search function.
Id={0} is not found in the damaged.
language file. StackStrace:
{1}
13445 FFQRP UserUtility: The config with The config file does not exist or is Reinstall Network Image Search function.
Key={0} in the file {1} can't damaged.
be loaded. StackStrace: {2}
13445 FFQRP UserUtility: The config with The config file does not exist or is Reinstall Network Image Search function.
key={0} must has value of damaged.
attribure Enum.
StackStrace: {1}
13445 FFQRP UserUtility: The config with The config file does not exist or is Reinstall Network Image Search function.
key={0} must has value of damaged.
element Value.
StackStrace: {1}
13445 FFQRP UserUtility: The config with The config file does not exist or is Reinstall Network Image Search function.
key={0} must has value of damaged.
attribure Name.
StackStrace: {1}
13445 FFQRP UserUtility: The config with The config file does not exist or is Reinstall Network Image Search function.
Key={0} is not found in the damaged.
config file of application.
StackStrace: {1}
13450 FFQRP Save font information to The config file does not exist or is Reinstall Network Image Search function.
application config file error. damaged.
13450 FFQRP Save note information to The config file does not exist or is Reinstall Network Image Search function.
note config file error. damaged.
13455 FFQRP The config with The config file does not exist or is Reinstall Network Image Search function.
Key=OverlayFontColor is damaged.
not found in the config file
of application.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-112
Event Event Source Description Causes Remedies Remarks
13457 CslAdv_Study %1 (%2) : Image The image processing parameter files Restart. If the same phenomenon recurred,
Processing Information does not exist. reinstall the system.
<InImg>
13472 CslAdv_Study %1 (%2) : Selected Menu The exposure for the selected menu has Select a menu for which an exposure
is Done. been done. process has not been performed.
XCON Result Information is
annulled. [%3]
13473 CslAdv_Study %1 (%2) : Menu is not No exposure menu has been selected. Select a menu for which an exposure
Selected. process has not been performed.
XCON Result Information is
annulled. [%3]
13600 FFQRP Can't load path for retrieved The config file does not exist or is Reinstall Network Image Search function.
images in DirectoryDef.xml damaged.
file
13600 FFQRP AppName key in file QRP. The config file does not exist or is Reinstall Network Image Search function.
exe.config isn't found or damaged.
value is empty
13600 FFQRP Get cluster value fail in The config file does not exist or is Reinstall Network Image Search function.
cluster.ini file damaged.
13600 FFQRP Can not get cluster type The config file does not exist or is Reinstall Network Image Search function.
value or output master damaged.
value in cluster.ini file
13998 CslAdv_Study %1 (%2) : [%3] General-purpose information event log Restart. If the same phenomenon recurred,
reinstall the system.
13999 CslAdv_Study [%s] %s (%s) : <%s> Defective program. Restart. If the same phenomenon recurred,
Unknown Error reinstall the system.
18100 CslAdv_QA QAMammoAlignmentCtrl. Failed in judgment of the state of the [ON] Perform the operation again.
dll button and the [OFF] button. If the same phenomenon recurred, restart
MammoAlignmentCtrl DR-ID 300CL.
JudgeStatusOnOff
MammoAlignment - Jugde
status on/off button failed.
18101 CslAdv_QA QAMammoAlignmentCtrl. Failed in acquisition of the center position Perform the operation again.
dll of the trimming and the trimming size. If the same phenomenon recurred, restart
MammoAlignmentCtrl DR-ID 300CL.
GetInfoTrimmingImage -
Get trimming center
position and trimming size
failed
18102 CslAdv_QA QAMammoAlignmentCtrl. Failed in terminating QA. Perform the operation again.
dll If the same phenomenon recurred, restart
MammoAlignmentView DR-ID 300CL.
ExitFunction - <exception
content>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-113
Event Event Source Description Causes Remedies Remarks
18103 CslAdv_QA QAMammoAlignmentCtrl. Failed in setting the data object. Perform the operation again.
dll If the same phenomenon recurred, restart
MammoAlignmentView DR-ID 300CL.
SetDataObj - <exception
content>
18104 CslAdv_QA QAMammoAlignmentCtrl. Failed in naming the temporary DDO file. Perform the operation again.
dll If the same phenomenon recurred, restart
MammoAlignmentCtrl DR-ID 300CL.
CreateTempDDOFileName
- Create name of file temp
ddo failed
18107 CslAdv_QA QAMammoAlignmentCtrl. Failed in processing the mammo image. Perform the operation again.
dll If the same phenomenon recurred, restart
MammoAlignmentCtrl DR-ID 300CL.
ProcessMammoAlginment
- Process mammo
alignment failed
18109 CslAdv_QA QAMammoAlignmentCtrl. Failed in judgment of the horizontal Perform the operation again.
dll position of the object. If the same phenomenon recurred, restart
MammoAlignmentCtrl DR-ID 300CL.
JudgeLeftRightObj - Jugde
image left right failed
18118 CslAdv_QA QA.exe Failed in displaying the mammo image. Perform the operation again.
QAScreen If the same phenomenon recurred, restart
ProcessClickMammo DR-ID 300CL.
AlignmentButton -
<exception content>
18121 CslAdv_QA QA.exe Failed in displaying the message. Check if there is “LangMsgBox.mdb” in the
QAScreen “C:\Program Files\FujiFilm\IIP\GUI\String”
vwMammoAlignmentCtrl_ folder, and check the authority of the
CustomMsgBox - Show “LangMsgBox.mdb” file.
message failed
19000 ImgViewCtrl File %1, Spot %2, Code An error which made the process to be As an internal error occurred, restart DR- %1: Name of the file
%3. uncontinuable has occurred. ID 300CL. which output the
If the error occurred again, restart the PC. event.
If the error could not be recovered, reinstall %2: Indicates an
the application. unique ID for
identifying the error
spot.
%3: Returns error type
(fixed as -9999)
20000 CslAdv_Delivery DICOM Printer Control An error was notified to the responding Restart. If the same phenomenon recurred, {0}: Status
RunProc Error (rc={0}) status of RunProcComplete event reinstall the system.
processed in DICOMPrintSCU.
20000 CslAdv_Delivery DICOM Printer Control An error was notified to the responding Restart. If the same phenomenon recurred, {0}: Status
RunProc Method Call Error status of RunProcComplete event reinstall the system.
(rc={0}) processed in OutImagePrc.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-114
Event Event Source Description Causes Remedies Remarks
20000 CslAdv_Delivery DICOM Printer Setting was Initialization function of printer setup Restart. If the same phenomenon recurred,
not found. returned an error. reinstall the system.
20000 CslAdv_Delivery DICOM Storage Initialization function of Commitment setup Restart. If the same phenomenon recurred,
Commitmemt Setting was returned an error. reinstall the system.
not found.
20000 CslAdv_Delivery DICOM Storage Control An error was notified to the responding Restart. If the same phenomenon recurred, {0}: Status
RunProc Error (rc={0}) status of RunProcComplete event reinstall the system.
processed in DICOMStoreSCU.
20000 CslAdv_Delivery DICOM Storage Control RunProc function of storage output control Restart. If the same phenomenon recurred, {0}: Function return value
RunProc Method Call Error returned an error. reinstall the system.
(rc={0})
20000 CslAdv_Delivery DICOM Storage Setting Initialization function of storage setup Restart. If the same phenomenon recurred,
was not found. returned an error. reinstall the system.
20000 CslAdv_Delivery DICOMSC Control Registration function of output destination Restart. If the same phenomenon recurred, {0}: Function return value
OutputRegisterProc returned an error. reinstall the system.
Method Call Error (rc={0})
20000 CslAdv_Delivery DICOMSC Control RunProc An error was notified to the responding Restart. If the same phenomenon recurred, {0}: Status
Error (rc={0}) status of RunProcComplete event reinstall the system.
processed in DicomSCCtrl.
20000 CslAdv_Delivery DICOMSC Control RunProc RunProc function of Commitment control Restart. If the same phenomenon recurred, {0}: Function return value
Method Call Error (rc={0}) returned an error. reinstall the system.
20000 CslAdv_Delivery Spooler Control Registration function of output destination Restart. If the same phenomenon recurred, {0}: Function return value
OutputRegisterProc returned an error to Spooler. reinstall the system.
Method Call Error (rc={0})
20000 [QA20000] An error An error occurred in displaying the patient Restart the system.
occurred in displaying the information. If the same error recurs after restarting the
patient information. system, reinstall the system.
20000 ClsAdv_Tomo 1% An error occurred when performing data Restart DR-ID 300CL. 1%: The spot of the
synchronization. error and its cause
20000 CslAdv_Re An error occurred in FLP • The FLPFinish.xml file cannot be checked. Restart DR-ID 300CL.
processing. • An unknown exception occurred when FLP
finished.
20001 CslAdv_Delivery DICOMPrint Control GetPrc Request function on retrieving Restart. If the same phenomenon recurred, {0}: Function return value
Method Call Error (rc={0}) DICOMPrintSCU status returned an error. reinstall the system.
20001 CslAdv_Delivery Spooler Control Modification function of DICOM printer Restart. If the same phenomenon recurred, {0}: Function return value
DicomStateChangeProc status returned an error to Spooler. reinstall the system.
Method Call Error (rc={0})
20001 CslAdv_ InBcrControlStart : %1 Failed in initializing the BCR. Connect the BCR and then restart. %1: An error code.
SystemManager
20001 [DE20001] The specified The specified device does not exist. Correct the settings.
device does not exist.
20001 [QA20001] Failed in Failed in obtaining information. Close the QA or go to other study.
obtaining information.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-115
Event Event Source Description Causes Remedies Remarks
20001 [RE20001] A A communication error with RIS occurred. Check the power of the device and its
communication error with connection.
RIS occurred.
20001 [SM20001] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
20001 [ST20001] Failed in adding Failed in adding a menu. Restart the system.
a menu. If the same error recurs after restarting the
system, reinstall the system.
20001 ClsAdv_Tomo 1% An error occurred when performing data Restart DR-ID 300CL. 1%: The spot of the
synchronization. error and its cause
20001 [QC20001] Input value is Input value is incorrect. Input a correct value.
incorrect.
20001 [IC20001] The study that The study that has failed in auto- Check the cause of the failure.
has failed in auto- conversion of patient ID exists. For the failed study, correct the examinee
conversion of patient ID list file and redo the processing, or
exists. manually correct the study information
directly in the QA.
20001 CslAdv_Re Send message fail. Send message fail. Restart DR-ID 300CL.
20001 [DA20001] Failed in Failed in analyzing Smart-QA. Press the button below to continue other
analyzing Smart-QA. processing.
20001 CslAdv_Bridge - An error due to unknown cause has Restart DR-ID 300CL. If the same
occurred at the time of reception of change phenomenon recurred, reinstall DR-ID
request. 300CL.
20001 CslAdv_Study Load data overlay for 1 Failed in reading the XML file for overlay Restart DR-ID 300CL. If the same error
image screen fail display. recurs after restarting, reinstall DR-ID
300CL.
20001 [OI20001] Images cannot Images cannot be imported into the Confirm the exposure menu of import
be imported into the selected menu. destination.
selected menu.
20002 CslAdv_Modality MainControl An error occurs inside the program. Check the database of utilization number.
IpFpdUseCountUp
20002 CslAdv_ InMcrControlStart : %1 Failed in initializing the MCR. Connect the MCR and then restart. %1: An error code.
SystemManager
20002 [DE20002] Failed in Failed in delivery processing. Wait for a while and try again.
delivery processing.
20002 [QA20002] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-116
Event Event Source Description Causes Remedies Remarks
20002 [RE20002] A timeout error A timeout error occurred while in search of Restart the system.
occurred while in search of study information. If the same error recurs after restarting the
study information. No system, reinstall the system.
responses come from
equipment targeted for
search.
20002 [SM20002] It could not be It could not be defined as confirmed. Restart the system.
defined as confirmed. If the same error recurs after restarting the
system, reinstall the system.
20002 [ST20002] This operation is This operation is not allowed. Login again with an authorized username.
not allowed.
20002 [QC20002] Input value is Input value is illegal. Input a correct value.
illegal.
20002 [DA20002] Failed in image Failed in image positioning. Perform manual positioning.
positioning.
20002 CslAdv_Modality IpFpdUseCountUp An error occurs inside the program. Check the database of utilization number.
GetDDOTagValue
20002 CslAdv_Modality IpFpdUseCountUp An error occurs inside the program. Check the database of utilization number.
SetIpCountData
20002 CslAdv_Bridge - Issuance could not be done since the CSL Perform the same operation again after a
conditions at the time of change request specified time has passed. Restart if the
reception were invalid. same kind of phenomenon occurs again.
20002 CslAdv_Study Load data overlay for 6 Failed in reading the XML file for overlay Restart DR-ID 300CL. If the same error
images screen fail display. recurs after restarting, reinstall DR-ID
300CL.
20002 [OI20002] The selected The selected image cannot be imported. Confirm the image to import.
image cannot be imported.
20003 CslAdv_Delivery OutputJob Image Host Failed in acquiring images of other Restart. If the same phenomenon recurred, {0}: Function return value
Error (rc={0}) equipments. reinstall the system.
20003 CslAdv_Modality 1% Failed in browsing the image database. Finish the study, and perform the study %1: The module and
again. If the same phenomenon recurred, spot error occurred
restart the application.
20003 CslAdv_Modality DeviceType is Error. - An error occurred in analyzing device Finish the study, and perform the study For device information.
DeviceType = %1 information. again. If the same phenomenon recurred,
restart the application.
20003 CslAdv_Modality Error Notification From An error occurred in analyzing device Finish the study, and perform the study For device information.
BC::ModalityIF in information. again. If the same phenomenon recurred,
OnExposureResultCreated restart the application.
- status = %1
20003 CslAdv_Modality Error Notification From An error occurred in analyzing device Finish the study, and perform the study For device information.
BC::ModalityIF in information. again. If the same phenomenon recurred,
OnExposureResultReceive restart the application.
- status = %1
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-117
Event Event Source Description Causes Remedies Remarks
20003 CslAdv_Modality Error Notification From An error occurred in analyzing device Finish the study, and perform the study For device information.
BC::ModalityIF in information. again. If the same phenomenon recurred,
OnImagingCreated - status restart the application.
= %1
20003 CslAdv_Modality Error Notification From An error occurred in analyzing device Finish the study, and perform the study For device information.
BC::ModalityIF in information. again. If the same phenomenon recurred,
OnImagingFinished - status restart the application.
= %1
20003 CslAdv_Modality Error Notification From An error occurs inside the program. Finish the study, and perform the study %1: Internal error code.
BC::ModalityIF in again. If the same phenomenon recurred,
OnImagingPreviewCreated restart the application.
- status = %1
20003 CslAdv_Modality Error Notification From An error occurs inside the program. Finish the study, and perform the study %1: Internal error code.
BC::ModalityIF in again. If the same phenomenon recurred,
OnImagingProcessStart restart the application.
- status = %1
20003 CslAdv_Modality Error Notification From An error occurs inside the program. Finish the study, and perform the study %1: Internal error code.
BC::ModalityIF in again. If the same phenomenon recurred,
OnImagingStart - status = restart the application.
%1
20003 CslAdv_Modality Error Notification From An error occurs inside the program. Finish the study, and perform the study %1: Internal error code.
BC::ModalityIF in again. If the same phenomenon recurred,
OnReqExposure restart the application.
PreparationCancel - status
= %1
20003 CslAdv_Modality Failed to get Menu status is invalid. Finish the study, and perform the study %1: The selector menu
StudyOperation - again. If the same phenomenon recurred, status.
ImageStatus = %1 restart the application.
20003 CslAdv_Modality Failed to Menu status is invalid. Finish the study, and perform the study
mixMenuStateCheck again. If the same phenomenon recurred,
restart the application.
20003 CslAdv_Modality ImageStatus != MENU_ The menu which cannot pressed the button Select another menu and perform the %1 The selector menu
FIXED - ImageStatus = %1 halfway has been selected. operation. status.
20003 CslAdv_Modality ImageStatus != READYUP The menu which cannot press the button Finish the study, and perform the study %1: The selector menu
- ImageStatus = %1 halfway has been selected. again. If the same phenomenon recurred, status.
restart the application.
20003 CslAdv_Modality ImageStatus != READYUP Menu status is invalid. Finish the study, and perform the study %1: The selector menu
or REIMAGING - again. If the same phenomenon recurred, status.
ImageStatus = %1 restart the application.
20003 CslAdv_Modality ImageStatus != READYUP Menu status is invalid. Finish the study, and perform the study %1: The selector menu
or REIMAGING - again. If the same phenomenon recurred, status.
ImageStatus = %1 restart the application.
20003 CslAdv_Modality ImageStatus != Menu status is invalid. Finish the study, and perform the study %1: Internal error code.
READYUP_FREE or again. If the same phenomenon recurred,
REIMAGING - ImageStatus restart the application.
= %1
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-118
Event Event Source Description Causes Remedies Remarks
20003 CslAdv_Modality ImageStatus != Menu status is invalid. Finish the study, and perform the study %1: The selector menu
READYUP_FREE or again. If the same phenomenon recurred, status.
REIMAGING - ImageStatus restart the application.
= %1
20003 CslAdv_Modality ImageStatus is An error occurred in analyzing device Finish the study, and perform the study %1: The status of
Error(builtin). - ImageStatus information. again. If the same phenomenon recurred, Console.
= %1 restart the application.
20003 CslAdv_Modality ImageStatus is An error occurred in analyzing device Finish the study, and perform the study %1: The status of
Error(cassette). - information. again. If the same phenomenon recurred, Console.
ImageStatus = %1 restart the application.
20003 CslAdv_Modality MenuOperation != The menu which cannot expose has been Finish the study, and perform the study %1: The status of
SELECTED_FREE - selected. again. If the same phenomenon recurred, Console.
MenuOperation = %1 restart the application.
20003 CslAdv_Modality MenuOperation != The menu which cannot press the button Finish the study, and perform the study %1: The selector menu
SELECTED_FREE - halfway has been selected. again. If the same phenomenon recurred, status.
MenuOperation = %1 restart the application.
20003 CslAdv_Modality RegistID and MenuUID An error occurs inside the program. Finish the study, and perform the study
mismatch - RegistID = %1 again. If the same phenomenon recurred,
MenuUid = %2 restart the application.
20003 CslAdv_Modality StudyOperation != Menu status is invalid. Finish the study, and perform the study %1: The selector menu
PERFORMING - again. If the same phenomenon recurred, status.
StudyOperation = %1 restart the application.
20003 CslAdv_ Initialize Exception An error occurred in processing the In DR-ID 300CL, the indicator can activate One of these five types
SystemManager Dependency_OnChange indicator of DR-ID 300CL. properly. No remedy is required. is set in event content.
Exception
SetSqlDependency
Exception
ClickIndicator is Null
ChangeIcon Exception
20003 [DE20003] The hard disk is The hard disk is full. Delete unnecessary images from the list
full. and try again.
20003 [QA20003] Failed in saving Failed in saving information. Restart the system.
information. If the same error recurs after restarting the
system, reinstall the system.
20003 [RE20003] Failed in starting Failed in starting the study. Restart the system.
the study.
20003 [ST20003] Body movement Body movement difficult to be observed in Enlarge the image to check the details. Maximum amount of
difficult to be observed in the displayed image has been detected. movement: %s mm
the displayed image has Maximum movement: %s mm
been detected. Maximum
movement: %s mm
20003 CslAdv_Modality Failed to get Menu status is invalid. Select the menu again. %1: Menu number
MenuOperation - ImageId = If the same phenomenon recurred, restart
%1 the Console.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-119
Event Event Source Description Causes Remedies Remarks
20003 ClsAdv_Tomo 1% An error occurred in the interlocking Preform the operation again. If the 1%: The spot of the
processing of image processing operation cannot be performed, restart error and its cause
adjustment. DR-ID 300CL.
20003 [QC20003] Input value is Input value is illegal. Input a correct value.
illegal.
20003 CslAdv_Bridge - Authentication processing for user Check the technician code and the
authentication has failed. password of the respective user on the
user management screen of the User
Utility.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
20003 CslAdv_Study Create singleton instance Failed in reading the XML file for overlay Restart DR-ID 300CL. If the same error
to load overlay object error display. recurs after restarting, reinstall DR-ID
300CL.
20003 [DA20003] Failed in setting Failed in setting of auto marker.
of auto marker.
20004 CslAdv_Modality %1 %2 Information is notified from the reader. Restart. If the same phenomenon recurred, %1: The module error
LngDBCExException Failed in retrieving the translation character reinstall the system. occurred
StatusID= %3 strings. %2: Spot error occurred
[ExceptionInfomation] %3: ID of connected
%4 character string and
status display
character string
%4: Error details
20004 CslAdv_Modality %1 %2 The error information is received from the Confirm the Image Reader.
Not Found Record in %:3 reader. The received information does not Restart.
ID= %4 match with the internal information of If the same phenomenon recurred, reinstall
DR-ID 300CL. the system.
20004 CslAdv_ Initialize.Failed An error occurred in the event window of In DR-ID 300CL, the indicator can activate One of these six types
SystemManager Terminate.Faild DR-ID 300CL. properly. No remedy is required. is set in event content.
buttonDeletion_Click_
Param.Failed
buttonConfirmation_Click_
Param.Failed
PrevPageButton_Click.
Failed
NextPageButton_Click.
Failed
20004 [DE20004] Failed in reading • The display gradation correction file Check the correction file.
a display gradation (AEName.lut) corresponding to the output
correction file. destination equipment does not exist.
• The format of the display gradation
correction file (AEName.lut) corresponding
to the output destination equipment is
incorrect.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-120
Event Event Source Description Causes Remedies Remarks
20004 [QA20004] Failed in Failed in confirmation of QA. Restart the system.
confirmation of QA. If the same error recurs after restarting the
system, reinstall the system.
20004 [RE20004] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
20004 [ST20004] A non-stitched A non-stitched long-view image exists. Check the image.
long-view image exists.
20004 [QC20004] Input value is Input value is illegal. Input a correct value.
illegal.
20004 CslAdv_Bridge - An error has occurred at the time of Restart DR-ID 300CL. If the same
reception of change request (non- phenomenon recurred, reinstall DR-ID
synchronous type). 300CL.
20005 CslAdv_Modality %1 %2 The invalid value is received from the Check the Image Reader status. %1: The module error
ConvertXMLtoProperty:%3. reader. occurred
Invalid data %4 %2: Spot error occurred
%3: Invalid value
%4: Detail of invalid
value
20005 CslAdv_Modality %1 %2 Failed in reading the IP barcode Read the IP barcode again. %1: The module error
Failed to read data from information. If the same phenomenon recurred, restart occurred
%3 the application. %2: Spot error occurred
[ExceptionInfomation] %3: Error outline
%4 %4: Error details
20005 [QA20005] Failed in Failed in suspending confirmation of QA. Restart the system.
suspending confirmation of If the same error recurs after restarting the
QA. system, reinstall the system.
20005 ClsAdv_Tomo Event argument is null The message that system received is valid. Restart DR-ID 300CL.
20005 [DE20005] Failed in output Failed in output process. Confirm the device status.
process.
20005 [ST20005] A long-view A long-view menu having images not Check the image.
menu having images not aligned exists.
aligned exists.
20005 [QC20005] Option key Option key software is not installed. Exit the program, and install option key
software is not installed. software.
20006 CslAdv_Modality %1 %2 Failed in notifying the inexposable status Check the Image Reader status. %1: The module error
Failed to execute selector. to Image Reader. occurred
Devices[%3].Modality. %2: Spot error occurred
DisableImaging %3: Error occurred
selector number
20006 CslAdv_Modality %1 %2 Failed in notifying the exposable status to Check the Image Reader status. %1: The module error
Failed to execute selector. Image Reader. occurred
Devices[%3].Modality. %2: Spot error occurred
EnableImaging %3: Error occurred
selector number
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-121
Event Event Source Description Causes Remedies Remarks
20006 CslAdv_Modality %1 %2 Failed in notifying the clearing exposure Check the Image Reader status. %1: The module error
Failed to selector. conditions to Image Reader. occurred
SendClearExposureCondition %2: Spot error occurred
- SelectorNumber = %3 %3: Error occurred
selector number
20006 CslAdv_Modality %1 %2 An error occurred in requesting to delete Restart. If the same phenomenon recurred, %1: The module error
OnResetImageKey the IP information from the Image Reader. reinstall the system. occurred
Recieved from ModalityIF %2: Spot error occurred
at Error Status.%3 %3: Error details
20006 CslAdv_Modality Failed to execute selector. Selector number: Failed in notifying to the Check the device status and connection.
%1: Selector number
Devices[%1].Modality. device of %1.
DisableImaging
20006 [QA20006] Failed in Failed in resuming the study. Restart the system.
resuming the study. If the same error recurs after restarting the
system, reinstall the system.
20006 ClsAdv_Tomo Event argument is null The message that system received is valid. Restart DR-ID 300CL.
20006 [QC20006] Image Image sharpness is illegal. Arrange the phantom position and perform
sharpness is illegal. an exposure again.
20006 [ST20006] Reading of the Reading of the patient ID auto-issuance Confirm the setting file.
patient ID auto-issuance setting file for screening failed.
setting file for screening
failed.
20007 [QA20007] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
20007 ClsAdv_Tomo 1% An error occurred in the interlocking Preform the operation again. If the same 1%: The spot of the
processing of image processing phenomenon recurred, restart DR-ID error and its cause
adjustment. 300CL.
20007 [QC20007] Failed to delete Failed to delete this folder. Restart DR-ID 300CL.
this folder.
%s
20007 [TS20007] An error An error occurred. Check the status of PPU and the
occurred. connection with PPU, and restart the
system.
20007 [ST20007] A value outside A value outside the allowable range has Restore the auto-issuance ID to the
the allowable range has been set for auto-issuance ID. minimum value.
been set for auto-issuance
ID.
20008 [QA20008] Failed in Failed in applying the changes. Restart the system.
applying the changes. If the same error recurs after restarting the
system, reinstall the system.
20008 ClsAdv_Tomo 1% An error occurred in the interlocking Preform the operation again. If the same 1%: The spot of the
processing of image processing phenomenon recurred, restart DR-ID error and its cause
adjustment. 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-122
Event Event Source Description Causes Remedies Remarks
20008 [TS20008] Failed in Failed in switching to other QA function. Switch the QA function again.
switching to other QA
function.
20008 [ST20008] Patient ID that Patient ID that will be issued is duplicated. Change the patient ID.
will be issued is duplicated.
20009 CslAdv_Delivery Analyze XML Failure An exception occurred in analyzing XML. Restart. If the same phenomenon recurred, {0}: Path of the xml file
(path={0} Message={1} reinstall the system. {1}: Exception message
StackTrace={2}) {2}: Stack trace
20009 ClsAdv_Tomo Failed in updating display An error occurred in TomoViewCtrl.dll. Preform the operation again. If the same
patient information An error occurred when updating the phenomenon recurred, restart DR-ID
display of patient information. 300CL.
20009 CslAdv_Delivery Received an unexpected The topic of the message notified in Restart. If the same phenomenon recurred, {0}: Topic
Message (/{0}/{1}/{2}) Ansemble message receiving process was reinstall the system. {1}: Object
in error. {2}: Event
20009 [QA20009] Failed in editing Failed in editing study comments. Restart the system.
study comments.
20009 [RE20009] Hard disk space Hard disk space required for free layout Delete the study and try again.
required for free layout print print is insufficient.
is insufficient.
20009 [QC20009] Baseline values Baseline values cannot be generated. Select another image, and reexecute.
cannot be generated.
20009 [ST20009] The menu The menu cannot be changed because an Update only the patient information.
cannot be changed image has been input.
because an image has
been input.
20010 CslAdv_Delivery {0} is a delivery attribute The unsupported distribution attribute was Restart. If the same phenomenon recurred, {0}: Distribution attribute
not supported. (Coping notified. reinstall the system.
Combine DDOFile at
LifeSize)
20010 CslAdv_Delivery {0} is a delivery attribute The unsupported distribution attribute was Restart. If the same phenomenon recurred, {0}: Distribution attribute
not supported. (Creating notified. reinstall the system.
delivery host information)
20010 CslAdv_Delivery {0} is a delivery attribute The unsupported distribution destination Restart. If the same phenomenon recurred, {0}: Distribution attribute
not supported. (Making information was notified. reinstall the system.
delivery DDO-file)
20010 CslAdv_Delivery {0} is a modality not The unsupported abstract syntax was Restart. If the same phenomenon recurred, {0}: Abstract syntax
supported. (Registering notified. reinstall the system.
delivery host information in
file-attribute)
20010 CslAdv_Delivery {0} is not found. (Copying The DDO file of composition original image Restart. If the same phenomenon recurred, {0}: DDO file
combine ddo file) does not exist. reinstall the system.
20010 CslAdv_Delivery {0} is not found. (Copying The DDO file does not exist. Restart. If the same phenomenon recurred, {0}: DDO file
DDO file) reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-123
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery {0} is not found. (Copying The image data file does not exist. Restart. If the same phenomenon recurred, {0}: Image data file
image data file) reinstall the system.
20010 CslAdv_Delivery Catch some other An exception occurred in copying process. Restart. If the same phenomenon recurred, {0}: File name
Exception. [{0}]\r\n{1}\r\n{2} reinstall the system. {1}: Exception message
{2}: Stack trace
20010 CslAdv_Delivery Catch some other An exception has occurred. Restart. If the same phenomenon recurred, {0}: Exception message
Exception.\r\n{0}\r\n{1} reinstall the system. {1}: Stack trace
20010 CslAdv_Delivery DCOMObjExeption : {0}\r\ An exception occurred in DDOHelper. Restart. If the same phenomenon recurred, {0}: DDO tag
n{1}\r\n{2} reinstall the system. {1}: Exception message
{2}: Stack trace
20010 CslAdv_Delivery DDO tag {0} NULL value was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
DistributionDensity is null. DistributionDensity. reinstall the system.
(Creating delivery host
information of file attribute)
20010 CslAdv_Delivery DDO tag {0} NULL value was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
DistributionProc is null. DistributionProc. reinstall the system.
(Creating delivery host
information of file attribute)
20010 CslAdv_Delivery DDO tag {0} ID is illegal. NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
(Data creation of delivery ID value. reinstall the system.
beginning of multi frame
information processing)
20010 CslAdv_Delivery DDO tag {0} ID is illegal. NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
(Data creation of delivery ID value. reinstall the system.
beginning)
20010 CslAdv_Delivery DDO tag {0} ID is illegal. (Is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
excluded auto delivery in ID value. reinstall the system.
delivering processing)
20010 CslAdv_Delivery DDO tag {0} Failed in retrieving ImagingSystem. Restart. If the same phenomenon recurred, {0}: DDO tag
ImagingSystem is illegal. reinstall the system.
(Copying WS image data
file)
20010 CslAdv_Delivery DDO tag {0} is illegal. Failed in retrieving the image data file Restart. If the same phenomenon recurred, {0}: DDO tag
(Copying image data file) name. reinstall the system.
20010 CslAdv_Delivery DDO tag {0} PrintDensity is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO file
null. (Creating delivery host the output image density. reinstall the system.
information of print
attribute)
20010 CslAdv_Delivery DDO tag {0} PrintProc is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO file
null. (Creating delivery host the output image processing. reinstall the system.
information of print
attribute)
20010 CslAdv_Delivery DDO tag {0} ProcUID is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: DDO tag
null. (Checking multi frame) ProcUID. reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-124
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery Delivery message Failed in converting the data for Ansemble Restart. If the same phenomenon recurred, {0}: Topic
conversion failure. (Starting message distribution. reinstall the system. {1}: Object
PreDelivery)\r\nMessage=/ {2}: Event
{0}/{1}/{2}
20010 CslAdv_Delivery Delivery work DDO file Failed in generating the file name for Restart. If the same phenomenon recurred,
name is null. (Copying distribution process. reinstall the system.
DDO file)
20010 CslAdv_Delivery Delivery work image file Failed in generating the file name for Restart. If the same phenomenon recurred,
name is null. (Copying distribution process. reinstall the system.
combine ddo file)
20010 CslAdv_Delivery Delivery work image file Failed in generating the file name for Restart. If the same phenomenon recurred,
name is null. (Copying distribution process. reinstall the system.
image data file)
20010 CslAdv_Delivery FF32CSException :\r\n{0}\r\ An exception occurred in FF32CS Restart. If the same phenomenon recurred, {0}: Exception message
n{1} processing. reinstall the system. {1}: Stack trace
20010 CslAdv_Delivery FFNetException :\r\n{0}\r\ An exception occurred in FFNet Restart. If the same phenomenon recurred, {0}: Exception message
n{1} processing. reinstall the system. {1}: Stack trace
20010 CslAdv_Delivery ImagePosition is 0. (Multi NULL value or "0" was notified to Restart. If the same phenomenon recurred,
frame image delivery ImagePosition. reinstall the system.
judgement processing)
20010 CslAdv_Delivery ImgDBCExException :\r\ An exception occurred in mgDBCEx Restart. If the same phenomenon recurred, {0}: Exception message
n{0}\r\n{1} processing. reinstall the system. {1}: Stack trace
20010 CslAdv_Delivery IOException : [{0}]\r\n{1}\r\ An exception occurred during the file copy. Restart. If the same phenomenon recurred, {0}: File name
n{2} reinstall the system. {1}: Exception message
{2}: Stack trace
20010 CslAdv_Delivery new delivery host Failed in creating the distribution Restart. If the same phenomenon recurred,
informaiton if null. (Creating destination information. reinstall the system.
useurgent delivery host
information)
20010 CslAdv_Delivery new delivery host Failed in creating the distribution Restart. If the same phenomenon recurred,
informaiton if null. destination information. reinstall the system.
(Registering delivery host
information in file-attribute)
20010 CslAdv_Delivery Not a delivery message. Trigger type to be distributed was in error. Restart. If the same phenomenon recurred,
(Starting PreDelivery) reinstall the system.
20010 CslAdv_Delivery The acquisition information Number of acquired information was in Restart. If the same phenomenon recurred, {0}: DistributionHost
number is illegal. (Creating error. reinstall the system. {1}: DistributionDensity
delivery host information of {2}: DistributionProc
file attribute)\r\
nDistributionHost = {0}\r\
nDistributionDensity = {1}\r\
nDistributionProc = {2}
20010 CslAdv_Delivery The delivery attribute is not NULL value or blank letter was notified to Restart. If the same phenomenon recurred,
set. (Starting delivery the distribution attribute. reinstall the system.
perparation.)
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-125
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery The input file name is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Input file name
illegal. [{0}] (Creating the input file name. reinstall the system.
delivery work file name)
20010 CslAdv_Delivery The input parameter is Invalid value was notified to the input Restart. If the same phenomenon recurred,
illegal. (Creating new multi parameter. reinstall the system.
frame information)
20010 CslAdv_Delivery The input parameter is NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: HostName
illegal. (Registering delivery the parameter. reinstall the system. {1}: OutputDensity
host information in file- {2}: OutputProc
attribute)\r\nHostNmae =
{0}\r\nOutputDensity = {1}\r\
nOutputProc = {2}
20010 CslAdv_Delivery The input parameter is NULL value or blank letter was notified to Restart. If the same phenomenon recurred,
illegal. (Starting Delivery the parameter. reinstall the system.
host information
acquisition)
20010 CslAdv_Delivery The input parameter is NULL value or blank letter was notified to Restart. If the same phenomenon recurred,
illegal. (Starting delivery the parameter. reinstall the system.
judgment)
20010 CslAdv_Delivery The KeyType({0}) is illegal. Notified type was in error. Restart. If the same phenomenon recurred, {0}: Type
(PreDeluvery of manual reinstall the system.
delivery processing)
20010 CslAdv_Delivery The number of multi frame Number of registered multi-frame Restart. If the same phenomenon recurred,
information lists is 0. (Data information was in error. reinstall the system.
creation of delivery
beginning of multi frame
information processing)
20010 CslAdv_Delivery The parameter of the Notified type was in error. Restart. If the same phenomenon recurred,
message is illegal. reinstall the system.
(Creating SQL sentence of
auto delivery reading)
20010 CslAdv_Delivery The parameter of the Notified auto distribution setting type was Restart. If the same phenomenon recurred, {0}: Auto distribution
message is illegal. auto in error. reinstall the system. setting type
delivery type = [{0}]
(Creating SQL sentence of
auto delivery)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
ES Output Image Group is ES Output Image Group. reinstall the system.
illegal. (ES image delivery
judgement processing)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
ES Output Image Group is ES Output Image Group. reinstall the system.
null. (WS image delivery
judgement processing)
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-126
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
film output format is illegal. the film output format. reinstall the system.
(ES image delivery
judgement processing)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
film output format is null. the film output format. reinstall the system.
(Other image delivery
judgement processing)
20010 CslAdv_Delivery The value of DDO tag {0} Failed in acquiring the film UID. Restart. If the same phenomenon recurred, {0}: Tag
ImageUID is illegal. reinstall the system.
(Checking completed multi
frame information)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
MPMCode is illegal. (ES MPM code. reinstall the system.
image delivery judgement
processing)
20010 CslAdv_Delivery The value of DDO tag {0} Failed in acquiring the multi frame UID. Restart. If the same phenomenon recurred, {0}: Tag
MultiFrameUID is illegal. reinstall the system.
(Multi frame image delivery
judgement processing)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or "0" was notified to the Restart. If the same phenomenon recurred, {0}: Tag
number of sheets of image number of images in the film. reinstall the system.
in film is illegal. (Creating
new multi frame
information)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
Print Image Group is null. Print Image Group. reinstall the system.
(WS image delivery
judgement processing)
20010 CslAdv_Delivery The value of DDO tag {0} Failed in acquiring ProcUID. Restart. If the same phenomenon recurred, {0}: Tag
ProcUID is illegal. (Multi reinstall the system.
frame image delivery
judgement processing)
20010 CslAdv_Delivery The value of diagnosis Es NULL value or blank letter was notified to Restart. If the same phenomenon recurred,
Image Type is illegal. (ES the ESImageType for diagnosis images. reinstall the system.
image delivery judgement
processing)
20010 FFOutImgPrc {0} is a delivery attribute The unsupported distribution attribute was Restart. If the same phenomenon recurred, {0}: Distribution attribute
not supported. (Creating notified. reinstall the system.
SQL sentence of auto
delivery)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: Tag
MPMCode is illegal. (ES the MPM code. reinstall the system.
image delivery judgement
processing)
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-127
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery The value of diagnosis Es NULL value or blank letter was notified to Restart. If the same phenomenon recurred,
Image Type is illegal. (ES the ESImageType for diagnosis images. reinstall the system.
image delivery judgement
processing)
20010 CslAdv_Delivery %1 is not found. (Copying DDO file to be copied does not exist. Perform delivery again. If the error is %1: DDO file path
DDO file) caused by the system, reinstall the system.
20010 CslAdv_Delivery Delivery work DDO file Failed in creating a temporary file name for Perform delivery again. If the error is
name is null. (Copying delivery. caused by the system, reinstall the system.
DDO file)
20010 CslAdv_Delivery IOException : [%1] An IOException exception occurred. Perform delivery again. If the error is %1: DDO file path
%2 caused by the system, reinstall the system. %2: Exception
%3 message
%3: Stack trace
20010 CslAdv_Delivery Catch some other An exception occurred. Perform delivery again. If the error is %1: DDO file path
Exception. : [%1] caused by the system, reinstall the system. %2: Exception
%2 message
%3 %3: Stack trace
20010 CslAdv_Delivery The input file name is The temporary file name for delivery is Perform delivery again. If the error is %1: File path
illegal. [%1] (Creating abnormal. caused by the system, reinstall the system.
delivery work file name)
20010 CslAdv_Delivery DCOMObjExeption : (%1) An exception of DDO operation Perform delivery again. If the error is %1: DDO tag number
%2 components occurred. caused by the system, reinstall the system. %2: Exception
%3 message
%3: Stack trace
20010 CslAdv_Delivery ReconstDataTime or The ReconstDataTime or InfoFileURL of Perform delivery again. If the error is %1: Method name
InfoFileURL is illegal. tomosynthesis image is abnormal. caused by the system, reinstall the system.
(%1)
20010 CslAdv_Delivery ImgGrpDBCException : An exception of ImgGrpDBC components Perform delivery again. If the error is %1: Exception
%1 occurred. caused by the system, reinstall the system. message
%2 %2: Stack trace
20010 [ST20010] Left-right density Left-right density difference by %s% has Confirm the image.
difference by %s% has been detected in the image.
been detected in the image.
If the actual SID is %s cm,
that is equal to the value
set to the Console,
it may be oblique by %s
degree.
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart DR-ID 300CL. {0}: Tag
Modality is illegal. Modality. If the same error recurs after restarting,
(Registering delivery host reinstall DR-ID 300CL.
information in file-attribute)
20010 CslAdv_Delivery The value of DDO tag {0} NULL value or blank letter was notified to Restart DR-ID 300CL. {0}: Tag
MPMCode is illegal. MPM Code. If the same error recurs after restarting,
(Registering delivery host reinstall DR-ID 300CL.
information in file-attribute)
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-128
Event Event Source Description Causes Remedies Remarks
20010 CslAdv_Delivery FFNet An error occurs in the function of Restart DR-ID 300CL.
FFNetDBGetStorageClass FFNetDBGetStorageClass. If the same error recurs after restarting,
Failed. (Registering delivery reinstall DR-ID 300CL.
host information in file-
attribute)
20010 CslAdv_Delivery MammoMenu is a Modality not supported by Mammography Check the setting parameter. If the error is {0}: Modality
modality{0} not supported. Menu is set. caused by the system, reinstall the DR-ID
(Registering delivery host 300CL.
information in file-attribute)
20010 CslAdv_Delivery FFNet.GetEquip2 Failed. Acquisition of the delivery destination Check the setting parameter and then
information by GetEquip2 of FFNet has restart the DR-ID 300CL. If the same
failed. phenomenon recurred, reinstall the
system.
20010 CslAdv_Delivery FFNet.GetDevice2 Failed. Acquisition of the delivery destination Check the setting parameter and then
information by GetDevice2 of FFNet has restart the DR-ID 300CL. If the same
failed. phenomenon recurred, reinstall the
system.
20010 CslAdv_Delivery PrintMessageInfoFor12on1 Message data registered in the message Restart DR-ID 300CL. If the same
is illegal.(MessageInfo were data list for 12on1 print delivery are phenomenon recurred, reinstall DR-ID
registed in duplicate) duplicated. 300CL.
20010 CslAdv_Delivery PrintMessageInfoFor12on1 Progress to be registered in the message Restart DR-ID 300CL. If the same
is illegal.(progress were data list for 12on1 print delivery has phenomenon recurred, reinstall DR-ID
registed in duplicate) already been registered. 300CL.
20010 CslAdv_Delivery CancelOfImageStatusOut Cancellation of temporary control of Restart DR-ID 300CL. If the same
Processing failed. delivery status has failed. phenomenon recurred, reinstall DR-ID
300CL.
20010 CslAdv_Delivery RegistersPrintMessage 12on1 print delivery message processing Restart DR-ID 300CL. If the same
InfoFor12on1() Failed. has failed. phenomenon recurred, reinstall DR-ID
300CL.
20010 CslAdv_Delivery GetDeliveryHostInfoOf Acquisition of the delivery destination Check the setting parameter and then
FisrtPrintAttribute() Failed. information for the first printer has failed. restart the DR-ID 300CL. If the same
phenomenon recurred, reinstall the
system.
20010 CslAdv_Delivery Judgment12on1Print() 12on1 print delivery start judgment has Restart DR-ID 300CL. If the same
Failed. failed. phenomenon recurred, reinstall DR-ID
300CL.
20010 CslAdv_Delivery DeliveryPreparationOf 12on1 print delivery advance preparation Restart DR-ID 300CL. If the same
12on1Print() Failed. has failed. phenomenon recurred, reinstall DR-ID
300CL.
20010 CslAdv_Delivery CancelOfImageStatusOut Cancellation of temporary control of Restart DR-ID 300CL. If the same
Processing() Failed. delivery status has failed. phenomenon recurred, reinstall DR-ID
300CL.
20010 [QA20010] Failed in Failed in positioning the right-and-left Adjust the trimming position manually.
positioning the right and left mammography images.
mammography images.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-129
Event Event Source Description Causes Remedies Remarks
20011 CslAdv_Delivery Delivery message Failed in converting the data for Ansemble Restart. If the same phenomenon recurred, {0}: Topic
conversion failure. message distribution. reinstall the system. {1}: Object
(GetImageUIDList)\r\ {2}: Event
nMessage=/{0}/{1}/{2}
20011 CslAdv_Delivery ImageStatusOutProcessing An error of checking upper/lower limit for Restart. If the same phenomenon recurred, {0}: Image_UID
is illegal.imageID={0}({1}) output processing status. reinstall the system. {1}: Output processing
status
20011 CslAdv_Delivery The input parameter is NULL value was notified to Image_UID list. Restart. If the same phenomenon recurred,
illegal. (CalcImageStatus reinstall the system.
OutProcessing)
20011 CslAdv_Delivery The input parameter is NULL value was notified to Image_UID list. Restart. If the same phenomenon recurred,
illegal. (GetImageUIDList) reinstall the system.
20011 CslAdv_Delivery The input parameter is NULL value was notified to the parameter. Restart. If the same phenomenon recurred,
illegal. (StatusEndImage reinstall the system.
StatusOutProcessing)
20011 [QA20011] Failed in Failed in entering information. Enter information in the marked items
entering information. correctly.
20011 [RE20011] Failed in starting Failed in starting QA of the study. Check the connection to the other device.
QA of the study.
20011 CslAdv_Delivery ImgGrpDBCException : An exception of ImgGrpDBC components Perform delivery again. If the error is %1: Exception
%1 occurred. caused by the system, reinstall the system. message
%2 %2: Stack trace
20011 CslAdv_Delivery Catch some other An exception occurred. Perform delivery again. If the error is %1: Exception
Exception. caused by the system, reinstall the system. message
%1 %2: Stack trace
%2
20011 ClsAdv_Tomo 1% An error occurred while displaying job Preform the operation again. If the same 1%: The spot of the
setting dialog. phenomenon recurred, restart DR-ID error and its cause
300CL.
20011 [QC20011] Input values are Input values are incorrect. Input a correct value.
incorrect.
20011 [ST20011] Failed in the Failed in the portable alignment detection Click the button below to restart the
portable alignment processing. system.
detection processing.
20012 CslAdv_Delivery Invalid Finish Media OutputId for media output is not registered. Restart. If the same phenomenon recurred, {0}: OutputId
Storage OutputId reinstall the system.
(status={0})
20012 CslAdv_Delivery Invalid Finish Media The message status value received from Restart. If the same phenomenon recurred, {0}: MediaStorage_
Storage Result ({0}={1}) Linker was in error. reinstall the system. Result
{1}: Result
20012 CslAdv_Delivery Invalid Media Detail Status An error occurred in converting the error Restart. If the same phenomenon recurred, {0}: AE name
(ae_name={0}, media_ information type from the message. reinstall the system. {1}: Status
status={1}, detail_ {2}: detailStatus
status={2})
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-130
Event Event Source Description Causes Remedies Remarks
20012 CslAdv_Delivery Invalid Media Status (ae_ MediaStatErr execution function of Spooler Restart. If the same phenomenon recurred, {0}: AE name
name={0}, media_ returned an error. reinstall the system. {1}: Status
status={1})
20012 CslAdv_Delivery Invalid Media Status NULL value or blank letter was notified to Restart. If the same phenomenon recurred, {0}: AE name
Parameter (ae_name={0}) the AE name. reinstall the system.
20012 FFOutImgPrc Invalid image_id An error occurred in converting the image_ Restart. If the same phenomenon recurred, {0}: image_id
(status={0}) id type of ID list which has completed the reinstall the system.
preliminary preparation.
20012 [RE20012] The number of The number of studies that allow the free Select 100 studies or less and try again.
studies that allow the free layout print exceeds the upper limit.
layout print exceeds the
upper limit.
20012 [QC20012] Input values are Input values are incorrect. Input a correct value.
incorrect.
20012 [ST20012] No examinee ID No examinee ID that can be issued Confirm the range of issuance numbers.
that can be issued automatically exists.
automatically exists.
20012 [QA20012] Failed in Failed in cancelling the right-and-left Review the trimming position on the right-
cancelling the right-and-left mammography image positioning. and-left images.
mammography image
positioning.
20012 ClsAdv_Tomo 1% An error occurred when performing data Restart DR-ID 300CL. 1%: The spot of the
to synchronization. error and its cause
20015
20013 [RE20013] A study which A study which does not allow the free Select a study and try again.
does not allow the free layout print has been selected.
layout print has been
selected.
20013 [QC20013] Input values are Input values are incorrect. Input a correct value.
incorrect.
20013 [QA20013] Failed to open Failed to open the ROI measurement Restart the system. If the same error
the ROI measurement screen. recurs after restarting the system, reinstall
screen. the system.
20014 [RE20014] Failed in Failed in shutting down the study forcibly. Restart the system.
shutting down the study
forcibly.
20014 [QC20014] Input values are Input values are incorrect. Input a correct value.
incorrect.
20014 CslAdv_Delivery Catch some other Failed in creating the DDO for outputting Check the setting parameter and then %1: CreateReport()
Exception. SR. restart DR-ID 300CL. If the same return value
Creating a Structured phenomenon recurred, reinstall DR-ID %2: Stack trace
Report finished in Failure. 300CL.
(rc=%1)
%2
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-131
Event Event Source Description Causes Remedies Remarks
20014 CslAdv_Delivery Catch some other An exception occurred. Restart DR-ID 300CL. %1: Exception
Exception. If the same phenomenon recurred, reinstall message
%1 DR-ID 300CL. %2: Stack trace
%2
20014 [QA20014] Fail in the Fail in the process of QA reset. Click the button to continue other
process of QA reset. processing.
20015 [QC20015] Input values are Input values are incorrect. Input a correct value.
incorrect.
20015 [QA20015] Measurement Measurement window does not start. Restart the system.
window does not start. If the same error recurs after restarting the
system, reinstall the system.
20015 [ST20015] The hard disk The hard disk space is too full for Free After suspending the study in the middle of
space is too full for Free Layout Print. progress, delete the study on the study list
Layout Print. screen, and then perform Free Layout Print
again.
20016 [RE20016] A study which A study which cannot be output has been Select a study again.
cannot be output has been selected.
selected.
20016 [QC20016] Input values are Input values are incorrect. Input a correct value.
incorrect.
20016 [ST20016] Images that do Images that do not allow Free Layout Print Click the button that displayed on the
not allow Free Layout Print or unexposed menus are included. screen.
or unexposed menus are
included.
20017 [RE20017] An error An error occurred in displaying F-RIS Contact the F-RIS maintenance staff.
occurred in displaying client.
F-RIS client.
20017 [QC20017] File existed. File existed. Check the file name.
20017 [ST20017] There is no There is no already-exposed menu. Click the button that displayed on the
already-exposed menu. screen.
20018 [RE20018] The number of The number of studies that match the Narrow down the search conditions and try
studies that match the study conditions exceeds the limit value. again.
study conditions exceeds The search will be aborted.
the limit value. The search
will be aborted.
20019 [RE20019] The specified The specified study was not found. Restart the system.
study was not found. If the same error recurs after restarting the
system, reinstall the system.
20020 [RE20020] Failed in starting Failed in starting the study. Restart the system.
the study. If the same error recurs after restarting the
system, reinstall the system.
20021 [QA20021] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-132
Event Event Source Description Causes Remedies Remarks
20021 [RE20021] The study The study cannot be deleted. Unlock the study and try again.
cannot be deleted.
20023 [RE20023] Failed in auto- Failed in auto-start of the study. Select a study from the study list and start
start of the study. the study.
20024 [QA20024] The patient The patient orientation has not been Enter information correctly.
orientation has not been entered.
entered.
20024 [RE20024] The study The study cannot be started. Choose study from the list, and start.
cannot be started.
20025 [RE20025] The batch The batch output of the selected studies Confirm the status of studies.
output of the selected cannot be performed.
studies cannot be
performed.
20026 [QA20026] Failed in Failed in displaying images from Synapse. Press the button below to continue other
displaying images from processing.
Synapse.
20026 [RE20026] The returned The returned record exceeds the limit. Set a higher maximum number of records.
record exceeds the limit. (The maximum value can exceed 10000.)
20027 [QA20027] Failed in Failed in outputting the mis-exposure log. Restart the system.
outputting the mis-exposure
log.
20028 [QA20028] Failed in Failed in outputting the mis-exposure log. Restart the system.
outputting the mis-exposure If the same error recurs after restarting the
log. system, reinstall the system.
20028 [RE20028] No record has No record has been selected. Choose study from the list.
been selected.
20029 [QA20029] Failed in Failed in changing the exposure menu. Restart the system.
changing the exposure
menu.
20029 [RE20029] The status of The status of the selected study is invalid. Choose study from the list.
the selected study is
invalid.
20030 [QA20030] The S value has The S value has exceeded the limit value. Check the S value.
exceeded the limit value.
20030 [RE20030] Study Study information could not be collected. Continue the processing with the collected
information could not be study information.
collected.
20030 [TS20030] The S value has The S value has exceeded the limit value. Check the S value.
exceeded the limit value.
20031 [QA20031] The L value has The L value has exceeded the limit value. Check the L value.
exceeded the limit value.
20031 [RE20031] Study could not Study could not be started. The study that couldn't be started has been
be started. registered for reservation. Start a study
from the study list.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-133
Event Event Source Description Causes Remedies Remarks
20031 [TS20031] The L value has The L value has exceeded the limit value. Check the L value.
exceeded the limit value.
20032 [QA20032] Failed in Failed in switching images. Click the button displayed on the screen. * This function cannot be
switching images. completed.
20032 [RE20032] None of the None of the selected studies contain Select a study and start it. After starting the
selected studies contain exposure menus. study, register an exposure menu.
exposure menus.
20033 [QA20033] Failed in Failed in switching images. Restart the system.
switching images.
20033 [RE20033] Study in which Study in which no exposure menu has Select studies in which their exposure
no exposure menu has been registered exists in the selected menu is registered and start the studies.
been registered exists in studies.
the selected studies.
20034 [QA20034] Failed in Failed in displaying the order information. Restart the system.
displaying the order
information.
20034 [RE20034] Failed in Failed in relocating the exposure menu. Confirm User Utility setting.
relocating the exposure
menu.
20035 [QA20035] Failed in Failed in outputting the mis-exposure log. Restart the system.
outputting the mis-exposure If the same error recurs after restarting the
log. system, reinstall the system.
20035 [RE20035] Failed in Failed in accessing the study. Update the list to reflect the latest status.
accessing the study. Press the button below to continue the
processing.
20036 [QA20036] Failed in Failed in outputting the mis-exposure log. Restart the system.
outputting the mis-exposure If the same error recurs after restarting the
log. system, reinstall the system.
20036 [RE20036] Auto-issuance Auto-issuance ID is not correct. Specify auto-issuance ID (%s-%s) within
ID is not correct. the range where MinID is lower than
MaxID.
20037 [QA20037] Failed in Failed in notifying the linking application. Restart the system.
notifying the linking
application.
20037 [QA20037] Failed in Failed in displaying or updating the option Restart the system.
displaying or updating the monitor.
option monitor.
20038 [RE20038] The range of The range of auto-issuance ID # is Re-input the range of auto-issuance ID
auto-issuance ID # is duplicated.
duplicated.
20039 [RE20039] The facility The facility identification code is incorrect. Set up a facility identification code using
identification code is single byte alphanumeric characters
incorrect. except for invalid characters such as *, @,
and space.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-134
Event Event Source Description Causes Remedies Remarks
20040 [QA20040] Failed in Failed in selecting exposure menus. Restart the system.
selecting exposure menus. If the same error recurs after restarting the
system, reinstall the system.
20041 [QA20041] Failed in Failed in changing the order of exposure Restart the system.
changing the order of menu list. If the same error recurs after restarting the
exposure menu list. system, reinstall the system.
20042 [QA20042] Failed in Failed in scrolling the exposure menu list. Restart the system.
scrolling the exposure If the same error recurs after restarting the
menu list. system, reinstall the system.
20043 [RE20043] It exceeded the It exceeded the maximum value of issuable Restore the issuance ID to the minimum
maximum value of issuable ID. value.
ID.
20044 [RE20044] The format of The format of the setting file is incorrect. Confirm the setting file.
the setting file is incorrect.
20045 [RE20045] A study which A study which cannot be started has been Select a study again.
cannot be started has been selected.
selected.
20047 [RE20047] The maximum The maximum duration that Patient ID is Specify the maximum duration that Patient
duration that Patient ID is stored is incorrect. ID is stored within the range of [1…365].
stored is incorrect.
20048 [RE20048] None of the None of the anatomical regions has been Select one or more anatomical regions.
anatomical regions has selected.
been selected.
20049 [RE20049] None of the None of the screening type has been Select one or more screening type.
screening type has been selected.
selected.
20050 [QA20050] The specified The specified position is invalid. The top image and the bottom image
position is invalid. cannot be specified.
Specify the center image and another
image.
20050 [RE20050] Input of Input of screening type has not been Input necessary information.
screening type has not completed yet.
been completed yet.
20051 [QA20051] The specified The specified position is invalid. Adjust the image position
position is invalid.
20051 [RE20051] A normal study A normal study and an order study that Select again.
and an order study that was started from RIS have been selected.
was started from RIS have
been selected.
20052 [QA20052] Rotation Rotation direction is invalid. Directions of images to be composed are
direction is invalid. not the same.
Set the images to the same direction.
20053 [QA20053] Failed in Failed in reading out patient information. Restart the system.
reading out patient If the same error recurs after restarting the
information. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-135
Event Event Source Description Causes Remedies Remarks
20053 [RE20053] An error An error occurred in converting a part of Click the button and continue the
occurred in converting a data. processing.
part of data.
20054 [RE20054] An error An error occurred in converting all data. Click the button and continue the
occurred in converting all processing.
data.
20055 [RE20055] All of the All of the selected studies are unable to Click the button displayed on the screen.
selected studies are unable execute.
to execute.
20056 [RE20056] Some exposure Some exposure menus cannot be Check the modality settings of Service
menus cannot be registered. Utility. Check the received exposure menu.
registered. There is an inconsistency between the
There is an inconsistency modality of Console and the exposure
between the modality of menus received.
Console and the exposure
menus received.
20056 [QA20056] Failed in image Failed in image stitching processing. Click the button that displayed on the
stitching processing. screen.
20057 [QA20057] Failed in Failed in resetting the stitching point. Restart the system.
resetting the stitching point. If the same error recurs after restarting the
system, reinstall the system.
20059 [RE20059] Output of some Output of some studies was disabled. Select studies which could not be outputted
studies was disabled. in the list.
20060 [RE20060] The selected The selected study is unable to execute Select studies again.
study is unable to execute the QC.
the QC.
20061 [RE20061] The studies of The studies of QC for general and Select studies again.
QC for general and mammography cannot be saved at the
mammography cannot be same time.
saved at the same time.
20063 [RE20063] A registered A study which can be suspended is Execute suspension again.
study is selected. selected in the list.
20064 [RE20064] A registered A registered study is selected. Select the correct study and execute again.
study is selected.
20065 [RE20065] A study with A study which can be suspended is Execute suspension again.
invalid patient information selected in the list.
is selected.
20066 [RE20066] A study with A study with invalid patient information is Select the correct study and execute again.
invalid patient information selected.
is selected.
20067 [RE20067] For studies For studies including tomosynthesis menu, Press the button to cancel the process.
including tomosynthesis patient ID auto-convert function cannot be
menu, patient ID auto- executed.
convert function cannot be
executed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-136
Event Event Source Description Causes Remedies Remarks
20235 CslAdv_Re Failed in accessing into the Failed in accessing the reception image Restart DR-ID 300CL.
recept image db db.
20400 CslAdv_Re Show status dialog failure Failed to show dialog at the time of starting Restart DR-ID 300CL.
studies/QA or restarting studies.
20401 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
to TomoImageDisplayMgrCtrl.dll. If the same phenomenon recurred, reinstall error and its cause
20403 Failed in displaying images. DR-ID 300CL.
20450 CslAdv_QA QACommon.dll File path has not been defined. Check the path of log files defined by
QAFileLog.ctor Directory.ini.
20453 CslAdv_Re Error when using GuiAdp to An error occurred in loading resources. Restart DR-ID 300CL.
load resource
20500 CslAdv_Re Update exposure menu Failed to update the exposure menu Restart DR-ID 300CL.
control failure control.
20600 CslAdv_Re Select changed of the Failed to handle the SelectChanged event Restart DR-ID 300CL.
lstStudyList control failure of the lstStudyList control.
20602 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when displaying output
range.
20603 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
TomoRecCondResOutputCtrl.dll. error and its cause
An error occurred when displaying output
range.
20607 CslAdv_Re Index out of range An error occurred while setting the body Restart DR-ID 300CL.
part of the exposure menu or setting focus If the same phenomenon recurred, reinstall
to the impedance of the study type. DR-ID 300CL.
20641 CslAdv_Re Read Failed to get result data at the end of FLP. Restart DR-ID 300CL.
ReleasePosssessionXXX.
xml failed
20642 CslAdv_Re Can not release possesion Failed to release possession at the end of Restart DR-ID 300CL.
of selected study. FLP.
20700 CslAdv_Re Abort synchronize thread Failed to synchronize data when the Restart DR-ID 300CL.
console_db was changed.
20701 [TS20701] Reconstruction Reconstruction processing cannot be Check the status of PPU and the
processing cannot be executed again. connection with PPU. Re-execute the
executed again. reconstruction processing.
20702 [TS20702] An error An error occurred while loading the Check the status of PPU and the
occurred while loading the reconstruction mode. connection with PPU.
reconstruction mode.
20703 [TS20703] DDO file is not DDO file is not authorized to be updated. Update the data again by the authorized
authorized to be updated. user.
20704 [TS20704] Reconstruction Reconstruction processing cannot be Click the button that displayed on the
processing cannot be executed again. screen.
executed again.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-137
Event Event Source Description Causes Remedies Remarks
20801 CslAdv_Re Initializes customize study Failed to open the customize study screen Restart DR-ID 300CL.
list form failure from the main form. If the same phenomenon recurred, reinstall
DR-ID 300CL.
20810 CslAdv_Re Initializes search host Failed to open the search host screen from Restart DR-ID 300CL.
screen failure the customize study form. If the same phenomenon recurred, reinstall
DR-ID 300CL.
20815 CslAdv_Re Initializes basic setting Failed to open the basic setting screen Restart DR-ID 300CL.
screen failure from the customize study form. If the same phenomenon recurred, reinstall
DR-ID 300CL.
20816 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in the process of turning Open the study and perform the operation
FNCCRFParamChgView ON VGP. again.
btnVGPOn_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20817 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in the process of turning Open the study and perform the operation
FNCCRFParamChgView OFF VGP. again.
btnVGPOff_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20818 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in changing the VGE Open the study and perform the operation
FNCCRFParamChgView parameter (--button). again.
btnVGEMinusMinus_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20819 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in changing the VGE Open the study and perform the operation
FNCCRFParamChgView parameter (-button). again.
btnVGEMinus_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20820 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in changing the VGE Open the study and perform the operation
FNCCRFParamChgView parameter (+button). again.
btnVGEPlus_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20821 CslAdv_QA QAFNCCRFParamChgCtrl.dll An error occurred in changing the VGE Open the study and perform the operation
FNCCRFParamChgView parameter (++button). again.
btnVGEPlusPlus_Click- If the same phenomenon recurred, restart DR-
ID 300CL.
20822 CslAdv_QA QAFNCCRFParamChgCtrl. Failed in changing the active status (enable/ Open the study and perform the operation
dll disable) of the VGP ON/OFF switching button. again.
FNCCRFParamChgView If the same phenomenon recurred, restart DR-
SetEnableButtonOnOffVGP- ID 300CL.
20823 CslAdv_QA QAFNCCRFParamChgCtrl.dll Failed in updating the VGE value. Open the study and perform the operation
FNCCRFParamChgView again.
RefreshVGE- If the same phenomenon recurred, restart DR-
ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-138
Event Event Source Description Causes Remedies Remarks
20900 CslAdv_Re Gereral worklist error An error occurred during the following Restart DR-ID 300CL.
operations.
• Handling the elapsed event of Refresh Timer.
• Displaying the status of data search dialog
from RIS.
• Handling SearchRisEvent from StatusDialog
to RIS.
• Registering the data returned from RIS.
• Updating the study date/time at starting
studies from the work list.
20901 [TS20901] The The reconstruction processing could not Check the status of PPU and the
reconstruction processing be canceled. connection with PPU.
could not be canceled.
20902 [TS20902] The priority of The priority of the specified reconstruction Check the status of PPU and the
the specified reconstruction processing could not be changed. connection with PPU.
processing could not be
changed.
20903 [TS20903] The The reconstruction job cannot be canceled. Click the button displayed on the screen.
reconstruction job cannot
be canceled.
21000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
21000 [DE21000] Failed in Failed in obtaining information from the Restart DR-ID 300CL.
obtaining information from database. If the same phenomenon recurred, reinstall
the database. DR-ID 300CL.
21000 ClsAdv_Tomo Value of '0' is not valid for An error occurred in TomoMenuDispCtrl. Restart DR-ID 300CL.
'emSize'. 'emSize' should dll. If the same phenomenon recurred, reinstall
be greater than 0 and less An error occurred when setting the font DR-ID 300CL.
than or equal to System. style in exposure menu information display
Single.MaxValue. processing.
Parameter name: emSize
21001 CslAdv_MdIf Exceeded the Limitation An error in exceeding upper/lower limit of Check the exposure conditions.
Value of XCON Exposure exposure conditions.
Data. [%1 = %2]
21001 CslAdv_Modality Corresponding Selector is Failed in acquiring the selector number. Finish the study, and perform the study %1: TechniqueCode
Nothing. InitialActiveButton again. If the same phenomenon recurred,
is Selected. DDO restart the application.
TECHCODE:%1
21001 [DE21001] Failed in Failed in canceling the output processing. *Output of selected images could not be
canceling the output canceled.
processing. Press the button below to continue other
processing.
21002 CslAdv_MdIf Timeout occurred while Timeout occurred while waiting exposure Check the X-ray device.
waiting exposure result. result.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-139
Event Event Source Description Causes Remedies Remarks
21002 CslAdv_Modality Exposure result is received An error occurs inside the program. Finish the study, and perform the study
but selected menu is again. If the same phenomenon recurred,
invalid. The selected menu restart the application.
may be image inputted
menu.
21002 [DE21002] Failed in Failed in sending the message. Restart DR-ID 300CL.
sending the message. If the same phenomenon recurred, reinstall
DR-ID 300CL.
21002 [MI21002] Failed in dose Failed in dose acquisition from dose area Confirm dose area product meter.
acquisition from dose area product meter.
product meter.
21003 [MI21003] An exposure An exposure condition that is not registered Reselect the selector to be used for
condition that is not in this system has been received from the exposure in DR-ID 300CL.
registered in this system exposure unit.
has been received from the The combination of X-ray tube, exposure
exposure unit. technique and DR panel type may not
The combination of X-ray have been specified correctly.
tube, exposure technique
and DR panel type may not
have been specified
correctly.
21004 [MI21004] Failed in Failed in connecting with the X-ray unit. Confirm the connected device.
connecting with the X-ray The image may have been lost.
unit.
The image may have been
lost.
21100 CslAdv_Re Failed conditions start The data of study being input from RIS is Check the following configuration items
study invalid. related to RIS connection.
• Item 15 “Examination No.Length” of “5.
CSL/IDT FUNCTION”
• Item 17 “Use Series Identification From
F-RIS” of “5. CSL/IDT FUNCTION”
21100 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
to TomoImgDispOperationCtrl.dll. If the same phenomenon recurred, reinstall error and its cause
21135 An error occurred in the processing of DR-ID 300CL.
displaying images.
21211 CslAdv_Re Error Occurs from BC:RisIf An error occurred from BC:Rislf. Restart DR-ID 300CL. Check that access
to the shared folder is possible with the
recept computer. Also check the following
configuration items.
• Item 97 “Recept Computer Info File Share
Folder” of “5. CSL/IDT FUNCTION”
• Item 307 “Simplified Ordering Report Share
Folder” of “5. CSL/IDT FUNCTION”
21212 CslAdv_Re Failed in writing information Failed in setting the character string into Restart DR-ID 300CL.
into registry registry. If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-140
Event Event Source Description Causes Remedies Remarks
21213 CslAdv_Re Error Occurs from BC:RisIf An error occurred in steps of processing Restart DR-ID 300CL. Check that access
the OnInitianlizeResult event from to the shared folder is possible with the
SimpleOrdering Controller. recept computer. Also check the following
configuration items.
• Item 97 “Recept Computer Info File Share
Folder” of “5. CSL/IDT FUNCTION”
• Item 307 “Simplified Ordering Report Share
Folder” of “5. CSL/IDT FUNCTION”
21214 CslAdv_Re Failed in accessing the An error occurred in steps of processing Restart DR-ID 300CL. Check that access
image database the OnErrorOccured event from to the shared folder is possible with the
SimpleOrdering Controller. recept computer. Also check the following
configuration items.
• Item 97 “Recept Computer Info File Share
Folder” of “5. CSL/IDT FUNCTION”
• Item 307 “Simplified Ordering Report Share
Folder” of “5. CSL/IDT FUNCTION”
21216 CslAdv_Re Failed in sending report to An error occurred in steps of processing Restart DR-ID 300CL. Check that access
Simple ordering the OnReceiveOrder event from to the shared folder is possible with the
SimpleOrdering Controller. recept computer. Also check the following
configuration items.
• Item 97 “Recept Computer Info File Share
Folder” of “5. CSL/IDT FUNCTION”
• Item 307 “Simplified Ordering Report Share
Folder” of “5. CSL/IDT FUNCTION”
21502 CslAdv_Re Out of maximum-setting An error occurred when the user clicked a Set a value within the allowable range.
exposure menu number menu and tried to add the menu from the
menu selection region to the menu list.
21603 CslAdv_Re SiteCode is not correct The user is not allowed to enter characters Enter the value again.
or an error occurred while converting the
character code to ANSI.
21607 CslAdv_Re Active target setting screen The user clicked the tab setting button in Restart DR-ID 300CL.
failed the MassOrder setting screen. If the same phenomenon recurred, reinstall
DR-ID 300CL.
21611 CslAdv_Re Gereral Massorder error An error occurred during the following Restart DR-ID 300CL.
operations.
• Processing the FormClosing event of the
SoftKeyBoardForm control.
• Getting the maximum value, the minimum
value, and the number of digits of the
automatically numbered patient ID.
• Setting the text of patient ID.
21621 CslAdv_Re Execute query delete An error occurred during deleting the Restart DR-ID 300CL.
record in IDHistory table history of IDHistory table when starting
failure DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-141
Event Event Source Description Causes Remedies Remarks
21851 CslAdv_Re Could not save DDO object A siteCode mapping error occurred while Restart DR-ID 300CL.
into DDO file starting a study or starting a next study,
and reading of DDO failed.
22000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
22000 ClsAdv_Tomo 1% An error occurred in TomoDataMgrCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
An error occurred in the image DataEntity error and its cause
control function.
22001 CslAdv_MdIf Exceeded the Limitation An error in exceeding upper/lower limit of Check the exposure conditions.
Value of XCON Exposure exposure conditions.
Data. [%1 = %2]
22001 ClsAdv_Tomo 1% An error occurred in TomoDataMgrCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
An error occurred in the image DataEntity error and its cause
control function.
22002 CslAdv_MdIf Timeout occurred while Timeout occurred while waiting exposure Check the X-ray device.
waiting exposure result. result.
22002 [MI22002] Failed in dose Failed in dose acquisition from dose area Dose area product values will not be
acquisition from dose area product meter. stored.Confirm dose area product meter.
product meter.
22002 CslAdv_Modality ExpSupportDev Failed in communication with the exposure Restart DR-ID 300CL. If the same %1: The spot of the
Interface.%1. ret = %2 supporting device. phenomenon recurred, reinstall DR-ID error
300CL. %2: The error code
22003 [MI22003] In the split In the X-ray cumulation expand mode Change the exposure menu with the [Menu
exposure menu, the menu, the function is not set to RT. change] button.
function is not set to RT.
22601 [TS22601] Reconstruction Reconstruction processing could not be Execute the reconstruction processing
processing could not be started. again.
started.
22601 ClsAdv_Tomo 1% An error occurred in TomoJobMgntCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
to An error occurred in the managing error and its cause
22605 processing of reconstruction queue.
22602 [TS22602] Reconstruction Reconstruction processing could not be Check the status of PPU and the
processing could not be started. connection with PPU.
started.
22603 [TS22603] The image The image information could not be Check the status of PPU and the
information could not be acquired from PPU. connection with PPU.
acquired from PPU.
22604 [TS22604] The parameter The parameter of reconstruction condition Check the parameters. The following items are
of reconstruction condition is invalid. changed. (Unit: mm)
is invalid. 1. Slice center height
2. Slice width
3. Slice interval
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-142
Event Event Source Description Causes Remedies Remarks
22605 [TS22605] The height of The height of the center slice for PPU Check the status of PPU and the
the center slice for PPU exceeded the maximum value (500mm). connection with PPU.
exceeded the maximum
value (500mm).
22606 [TS22606] An error An error occurred while creating a marker. Click the button displayed on the screen.
occurred while creating a
marker.
22607 [TS22607] A tomosynthesis A Tomosynthesis menu in process of Click the button displayed on the screen.
menu in process of reconstruction exists.
reconstruction exists.
22608 [TS22608] An image of An image of which the height of the slice is Check the height of the slice.
which the height of the slice 0 to 500mm can only be generated. Other
is 0 to 500mm can only be images cannot be generated.
generated. Other images
cannot be generated.
22609 ClsAdv_Tomo * String of the exception An error occurred while setting the position Preform the operation again. If the same
occurred is output. of the scroll bar. phenomenon recurred, restart DR-ID
300CL.
22609 [TS22609] The number of The number of exposure images could not Confirm the number of exposure images.
exposure images could not be obtained properly.
be obtained properly.
22610 ClsAdv_Tomo * String of the exception An error occurred while setting the position Preform the operation again. If the same
occurred is output. of the scroll bar. phenomenon recurred, restart DR-ID
300CL.
22611 ClsAdv_Tomo * String of the exception An error occurred while setting the indexes Preform the operation again. If the same
occurred is output. of the 1st column to be displayed. phenomenon recurred, restart DR-ID
300CL.
22612 ClsAdv_Tomo * String of the exception An error occurred when acquiring the Preform the operation again. If the same
occurred is output. values in the table. phenomenon recurred, restart DR-ID
300CL.
22613 ClsAdv_Tomo * String of the exception An error occurred while retrieving/setting Preform the operation again. If the same
occurred is output. values of the table. phenomenon recurred, restart DR-ID
300CL.
22614 ClsAdv_Tomo * String of the exception An error occurred while retrieving/setting a Preform the operation again. If the same
occurred is output. value of the scroll bar. phenomenon recurred, restart DR-ID
300CL.
22615 ClsAdv_Tomo * String of the exception An error occurred when generating the Restart DR-ID 300CL.
occurred is output. display data.
22802 [UU22802] The patient's ID The patient's ID # is too long, Enter information correctly.
# is too long,
22803 [UU22803] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in patient's ID#.
the patient's ID#.
22804 [UU22804] The patient's The patient's Name is too long. Enter information correctly.
Name is too long.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-143
Event Event Source Description Causes Remedies Remarks
22805 [UU22805] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in patient's name.
the patient's name.
22806 [UU22806] The Kanji The Kanji patient's name is too long. Enter information correctly.
patient's name is too long.
22807 [UU22807] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in Kanji Patient's Name.
the Kanji Patient's Name.
22808 [UU22808] The format used The format used for the sex is incorrect. Enter information correctly.
for the sex is incorrect.
22809 [UU22809] The format used The format used for the date of birth is Enter information correctly.
for the date of birth is incorrect.
incorrect.
22810 [UU22810] Patient Patient information has not been entered. Enter at least one of the following items:
information has not been patent ID #, patient name and sex.
entered.
23001 [MI23001] An exposure An exposure condition that is not registered Reselect the selector to be used for
condition that is not in this system has been received from the exposure in DR-ID 300CL.
registered in this system exposure unit.
has been received from the The combination of X-ray tube, exposure
exposure unit. technique and DR panel type may not
The combination of X-ray have been specified correctly.
tube, exposure technique
and DR panel type may not
have been specified
correctly.
23003 ClsAdv_Tomo 1% An error occurred in Restart DR-ID 300CL. 1%: The spot of the
to TomoImgProcReqQueueCtrl.dll. If the same phenomenon recurred, reinstall error and its cause
23007 An error occurred when initializing the DR-ID 300CL.
Tomosynthesis View.
23008 ClsAdv_Tomo Key [%1] is set default Failed to get the option setting values Check the settings corresponding to the * When an error occurs,
value = %2 when starting Tomosynthesis due to the key indicated by log message, and correct the default value is
following reasons. the error according to the causes described used automatically,
• There is no setting file. in the “Cause” column. and Tomosynthesis is
• A setting key is not assigned to the setting file. normally started.
• Set values of the setting file are incorrect
(type, range).
23101 ClsAdv_Tomo 1% An error occurred in TomoDBAccessor.dll. Restart DR-ID 300CL. 1%: The spot of the
to An database access error has occurred. If the same phenomenon recurred, reinstall error and its cause
23104 DR-ID 300CL.
23200 ClsAdv_Tomo + DCMObjExException Failed in reflecting the DDO tags. Perform the operation again. If the
+ General Exception operation cannot be performed, restart
DR-ID 300CL.
23202 ClsAdv_Tomo + Get Latitude Tag Value Failed in acquiring the value of Latitude tag Perform the operation again. If the
Failed as it does not exist. operation cannot be performed, restart
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-144
Event Event Source Description Causes Remedies Remarks
23203 ClsAdv_Tomo + DCMObjExException Failed in acquiring the value of Latitude tag Perform the operation again. If the
+ General exception as it does not exist. operation cannot be performed, restart
DR-ID 300CL.
23204 ClsAdv_Tomo + General exception An exception occurred when acquiring the Perform the operation again. If the
DDO tags. operation cannot be performed, restart
DR-ID 300CL.
23205 ClsAdv_Tomo + DCMObjExException Failed in deleting the DDO tags. Perform the operation again. If the
operation cannot be performed, restart
DR-ID 300CL.
23205 FFQRP FFInternal2Output2 error The DDO file does not exist or is damaged. Search the study again.
23205 FFQRP ddo_GetUnicodeVR_ The DDO file does not exist or is damaged. Search the study again.
DICOMObject error.
23205 FFQRP All image of study is not An image which is not supported exists in Search the study again.
support: the study.
intNotSupportImage: {0}
- intImageTotalNumber: {1}
23205 FFQRP intNotSupportImage: {0} An image which is not supported exists in Search the study again.
- intImageTotalNumber: {1} the study.
- intDisplayedImage: {2} -
intImageError: {3}
23205 FFQRP Not support image(s): {0} An image which is not supported exists in Confirm whether the tomosynthesis image
- Total support image(s): {1} the study. or the image exposed with long view menu
- Totak image(s): {2} for outputting individually in the study
exists or not.
23305 FFQRP Can't get patient name The config file does not exist or is Reinstall Network Image Search function.
format type from damaged.
QRPConfig.XML.
23207 ClsAdv_Tomo + DCMObjExException An error occurred when copying the DDO Perform the operation again. If the
+ General exception tags data. operation cannot be performed, restart
+ EncoderFallbackException DR-ID 300CL.
23208 ClsAdv_Tomo + DCMObjExException An error occurred when reflecting the Perform the operation again. If the
+ General exception exposure order of DDO tags. operation cannot be performed, restart
DR-ID 300CL.
23209 ClsAdv_Tomo + ArgumentException An error occurred when convert from byte Perform the operation again. If the
+ FormatException to short. operation cannot be performed, restart
+ General exception DR-ID 300CL.
23210 ClsAdv_Tomo + DCMObjExException An error occurred when creating image Restart DR-ID 300CL.
+ General exception UID.
+ FF32CSException
23600 FFQRP Not found resource file : {0} The resource file does not exist or is Reinstall Network Image Search function.
damaged.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-145
Event Event Source Description Causes Remedies Remarks
24300 [UU24300] Patient Patient information has been registered Search by the key word.
information has been more than the number of displayable
registered more than the information.
number of displayable
information.
24301 [UU24301] Patient Patient information beyond the number of Change the key word and try again.
information beyond the displayable information was found.
number of displayable
information was found.
24302 [UU24302] Failed in Failed in connecting to the patient It will be connected to the data base of this
connecting to the patient information management host. system.
information management To connect to the management host again,
host. confirm the connection and input a
"-"(hyphen) to the patient ID# to search.
24308 [UU24308] Failed in • It may have been deleted by another system. Search again.
obtaining patient • Failed in obtaining patient information.
information.
24801 [UU24801] The kana The kana patient name is too long. Enter information correctly.
patient name is too long.
24802 [UU24802] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in kana patient's name.
the kana patient's name.
24803 [UU24803] The telephone The telephone number is too long. Enter information correctly.
number is too long.
24804 [UU24804] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in telephone number.
the telephone number.
24805 [UU24805] The format used The format used for the blood type (ABO) Enter information correctly.
for the blood type (ABO) is is incorrect.
incorrect.
24806 [UU24806] The format used The format used for the blood type (Rh) is Enter information correctly.
for the blood type (Rh) is incorrect.
incorrect.
24807 [UU24807] The physical The physical disorder information is too Enter information correctly.
disorder information is too long.
long.
24808 [UU24808] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in physical disorder information.
the physical disorder
information.
24809 [UU24809] The infection The infection information is too long. Enter information correctly.
information is too long.
24810 [UU24810] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in infection information.
the infection information.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-146
Event Event Source Description Causes Remedies Remarks
24811 [UU24811] The The contraindication information is too Enter information correctly.
contraindication information long.
is too long.
24812 [UU24812] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in contraindication information.
the contraindication
information.
24813 [UU24813] The patient The patient comment is too long. Enter information correctly.
comment is too long.
24814 [UU24814] Invalid Invalid character(s) are included in the Enter information correctly.
character(s) are included in patient comment.
the patient comment.
24815 [UU24815] Items to be Items to be displayed have not been set. Set more than one display items.
displayed have not been
set.
24816 [UU24816] The setting of The setting of warning time for decreased Enter information correctly.
warning time for decreased battery capacity exceeds a range of 10 to
battery capacity exceeds a 59.
range of 10 to 59.
24817 [UU24817] Input Input information is invalid. Enter correctly.
information is invalid. The
number of studies is from
one to 1000.
24818 [UU24818] Input Input information is invalid. Enter correctly.
information is invalid.
Select one or more hosts.
24819 [UU24819] The remaining The remaining battery level to cause Enter correctly.
battery level to cause system shutdown exceeds a range of 5 to
system shutdown exceeds 50%.
a range of 5 to 50%.
29999 CslAdv_Delivery Can Not Write CallLog.\ An exception occurred in outputting the Restart. If the same phenomenon recurred, {0}: Exception message
n{0}\n{1} CallLog. reinstall the system. {1}: Stack trace
29999 CslAdv_Delivery Can Not Write EnterLog.\ An exception occurred in outputting the Restart. If the same phenomenon recurred, {0}: Exception message
n{0}\n{1} EnterLog. reinstall the system. {1}: Stack trace
29999 CslAdv_Delivery Can Not Write ExitLog\n{0}\ An exception occurred in outputting the Restart. If the same phenomenon recurred, {0}: Exception message
n{1} ExitLog. reinstall the system. {1}: Stack trace
29999 CslAdv_Delivery Can Not Write ReturnLog.\ An exception occurred in outputting the Restart. If the same phenomenon recurred, {0}: Exception message
n{0}\n{1} ReturnLog. reinstall the system. {1}: Stack trace
30000 [TS30000] Do you wish to The markers is tried to be deleted. Click [No] and continue the operation
delete the markers and without deleting the markers.
continue the operation?
30000 CslAdv_IDAutoConvert FF32CS. Failed in obtaining the registry information. Restart DR-ID 300CL.
RegGetKeyDWORD If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-147
Event Event Source Description Causes Remedies Remarks
30001 CslAdv_Modality ImageStatus != MENU_ The menu which cannot expose has been Select the other menu.
CREATED - ImageStatus = selected. Add a menu.
%1
30001 CslAdv_Modality ImageStatus != MENU_ The menu which cannot expose has been Select the other menu.
CREATED/MENU_FIXED/ selected. Add a menu.
MENU_FIXED_FREE/
READYUP_FREE -
ImageStatus = %1
30001 CslAdv_ Application.Start Application launched. - Information to know
SystemManager that the application was
being started.
30001 CslAdv_Modality ImageStatus != IMAGING_ Menu status is invalid. Select the menu again. If the same %1 :The status of
COMPLETED/OUTPUT_ phenomenon recurred, restart the Console. selected menu
COMPLETED/READYUP_ %2 :Menu number
FREE/MENU_CREATED
- ImageStatus =
%1,ImageId = %2
30001 CslAdv_Modality ImageStatus != MENU_ Menu status is invalid. Select the menu again. If the same %1 :The status of
FIXED/MENU_FIXED_ phenomenon recurred, restart the Console. selected menu
FREE - ImageStatus = %2 :Menu number
%1,ImageId = %2
30001 CslAdv_Modality MenuOperation != Menu status is invalid. Select the menu again. If the same %1 :The status of
SELECTED_FREE/ phenomenon recurred, restart the Console. selected menu
SELECTED_LOCK - %2 :Menu number
MenuOperation =
%1,ImageId = %2
30001 CslAdv_Modality MenuOperation != Menu status is invalid. Select the menu again. If the same %1 :The status of
SELECTED_FREE - phenomenon recurred, restart the Console. selected menu
MenuOperation = %2 :Menu number
%1,ImageId = %2
30001 CslAdv_Modality ImageStatus != MENU_ Menu status is invalid. Select the menu again. If the same %1 :The status of
FIXED - ImageStatus = phenomenon recurred, restart the Console. selected menu
%1,ImageId = %2 %2 :Menu number
30001 [QC30001] Established Established criteria have been changed. -
criteria have been changed.
30001 [IC30001] Auto-conversion Auto-conversion of patient ID will be -
of patient ID will be executed.
executed.
30001 CslAdv_Modality High Sensitive mode has Timeout has occurred in Xfree high- Click the button and continue the
timed out sensitivity mode. operation.
30001 [MI30001] The function has The function has been changed. Reconfigure APR on the X-ray generator.
been changed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-148
Event Event Source Description Causes Remedies Remarks
30001 [OI30001] The number of The number of images in the ImageBox to After matching the image to the exposure
images in the ImageBox to be displayed on the image library window menu, open the image library again.
be displayed on the image exceeds the limit.
library window exceeds the
limit.
30002 CslAdv_ Application.Exit Application exited properly. - Information to know
SystemManager that the application was
being exited properly.
30002 [QC30002] The study result The study result cannot be stored. Input the study result.
cannot be stored.
30002 [OI30002] Input images. The image is to be input. Click the button and continue the Patient : %s
operation. Menu name : %s
Image seq.No : %s
Date of getting : %s
30003 CslAdv_ ConsoleEventIndicator. Indicator processing was started. - Information to know
SystemManager Start that the application was
being started.
30003 [QC30003] Some items Some items have not yet been input. Enter the user input items.
have not yet been input.
30003 CslAdv_Modality statusId = empty, Failed in reflecting the message notifying Preform the operation again. If the same %1: Data Source Type
DataSourceType = %1 , from the unit correctly. phenomenon recurred, restart DR-ID %2: Selector number
selectorNo = %2, 300CL. %3: Host name
hostName = %3
30004 CslAdv_ ConsoleEventIndicator.End Indicator processing was finished. - Information to know
SystemManager that the application was
being exited properly.
30004 [QC30004] Some items Some items have not yet been input. Execute the measurement processing.
have not yet been input.
30005 [OM30005] The screen The screen resolution has been changed. Restart the system.
resolution has been
changed.
30005 CslAdv_ MakeCabFile::Exception Failed in creating Cab file when outputting Restart DR-ID 300CL. One of the 9 types will
SystemManager MakeCabFile Retry log file. If the same phenomenon recurred, reinstall be set as an event
BackupLogCabMake DR-ID 300CL. content.
Exception::
MaintePCUpLoad
MakeCabFile NG
MaintePCUpLoad::
exception=Mainte
PCUpLoad::MainteFolder
Nothing
IIPBackupLogDataMake
false end
IIPBackupLogDataMake
LogFile Nothing =
IIPBackupLogDataMake
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-149
Event Event Source Description Causes Remedies Remarks
30005 [QC30005] Measurement Measurement processing will be -
processing will be implemented.
implemented.
30006 [MD30006] Invalid Invalid characters are included in IP#. Re-input correctly.
characters are included in
IP#.
30006 [QC30006] The study result The study result is not stored accordingly. Store the study result.
is not stored accordingly.
30006 [OM30006] The current The current password is not correct. Input a correct value.
password is not correct.
30007 [QC30007] Reconfirming. Reconfirming. Click [Yes] to save the data and exit QC.
Click [No] to exit QC without saving the
data.
30007 [OM30007] The new The new password and the new password Input a correct value.
password and the new for confirmation do not match.
password for confirmation
do not match.
30008 [ST30008] Please check Please check the order of images. Check the order of the images.
the order of images.
30008 [OM30008] The new The new password is identical to the Input a different password.
password is identical to the password currently used.
password currently used.
30009 [MD30009] The conditions The conditions that do not allow an Check the exposure conditions of the
that do not allow an exposure to start have been sent from an external device.
exposure to start have external device.
been sent from an external
device.
30009 [ST30009] Stitching Stitching processing cannot be executed Check the number of the inputted images.
processing cannot be because no more than one image has
executed because no more been inputted.
than one image has been
inputted.
30009 [OM30009] Your password Your password will expire in %s days. Change the password.
will expire in %s days.
30010 [ST30010] Stitching Stitching processing cannot be executed Check the direction of the inputted image.
processing cannot be since the image size or rotation information
executed since the image that has been inputted does not match one
size or rotation information another.
that has been inputted does
not match one another.
30010 [OM30010] Your password Your password has expired. Change the password.
has expired.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-150
Event Event Source Description Causes Remedies Remarks
30011 [ST30011] Please set the Exposure condition has not been set Set the exposure conditions.
exposure condition properly properly.
again when the technique
is changed.
30011 [MD30011] Image receiving Image receiving could not be completed. Receive again.
could not be completed.
30011 [OM30011] The length of The length of new password is insufficient. Make the password longer than the
new password is minimum string length.
insufficient.
30012 [QC30012] Baseline values Baseline values will be stored. -
will be stored.
30012 [ST30012] Stitching Stitching processing has been performed -
processing has been on already acquired images.
performed on already
acquired images.
30012 [OM30012] The new The new password contains an unusable Enter single-byte alphanumerics.
password contains an character.
unusable character.
30013 [MD30013] Exposure was Exposure was aborted. Check the images received.
aborted.
30013 [OM30013] Password is too Password is too long. Keep the password within 20 characters.
long.
30014 [MD30014] Failed in Failed in changing the panel. Initialize the panel.
changing the panel.
30014 [QC30014] Selected history Selected history information will be -
information will be deleted. deleted.
30014 [OM30014] You cannot use The password that has been used for the Set another password instead.
the passwords that have last %s times is entered.
been used for the last %s
times.
30015 [MD30015] Exposed The menu is tried to be deleted. When you select [Yes], the images
images will be lost and combining processing starts for only the
unfinished long-view menus retrieved images.
will be deleted if the menu When you select [No], images will be
is deleted. retrieved again without deleting them.
30015 [QC30015] Basic value has Basic value has been changed. Store the study result.
been changed.
30015 [OM30015] Enter a The password that is too similar to the Set a password at least %s different
password that has at least previous one is entered. characters than the previous one.
%s different characters
than the previous one.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-151
Event Event Source Description Causes Remedies Remarks
30016 [MD30016] The technique The technique cannot be changed because The exposure technique cannot be
cannot be changed the long-view menu has already been changed for the developed long-view
because the long-view relocated. menu.
menu has already been
relocated.
30016 [OM30016] Changing the Changing the password when in the period Change the password after the period is
password is still prohibited. that changing password is prohibited. over.
30017 [MD30017] The exposure The exposure technique cannot be The menu whose exposure technique
technique cannot be changed. cannot be changed is selected.
changed. Check the menu.
30017 [OM30017] Character types The password entered contains no capital Set a password that contains capital
a password must contain letter. letter(s).
are insufficient.
30018 [MD30018] The conditions The conditions to start exposure are not Perform the operation again.
to start exposure are not met. If the same phenomenon recurred, restart
met. DR-ID 300CL.
30018 [OM30018] Character types The password entered contains no small Set a password that contains small
a password must contain letter. letter(s).
are insufficient.
30019 [MD30019] There is a study There is a study under image stitching. Click the button that displayed on the
under image stitching. screen.
30019 [OM30019] Character types The password entered contains no special Set a password that contains special
a password must contain character. character(s).
are insufficient.
30020 [QA30020] Information has Information has been changed. Check the Apply button to confirm the
been changed. image processing adjustment results.
30020 [MD30020] Exposure was Exposure was aborted. Check the received images.
aborted.
30020 [OM30020] User account User account has been locked. Ask the administrator to unlock the user
was locked. account.
30021 [OM30021] Login is unable Intend to login while the user account is Ask the administrator to unlock the user
because user account has locked. account.
been locked.
30022 [TS30022] FNC-CRF image FNC-CRF image processing parameters Check the FNC-CRF image processing
processing parameters are are not configured properly. parameters.
not configured properly.
30022 [OM30022] Character types The password entered contains no Set a password that contains number(s).
a password must contain number.
are insufficient.
30023 [TS30023] The corrected The corrected information will be saved. -
information will be saved.
30023 [OM30023] The user is Intend to change to the user account while Change to the account that is not logged
about to be changed to a it is currently logged in. in.
user currently logging in.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-152
Event Event Source Description Causes Remedies Remarks
30024 [OM30024] Failed in login. Input an invalid user information. Input user information correctly.
30025 [OM30025] Your IC card is Failed in authenticating IC card. Or utilized Check if it is the registered IC card, and
not recognized. an unregistered IC card. read the IC card again.
30026 [OM30026] Your IC card Failed in reading IC card. Restart. If the same symptom recurs,
was not read. perform a reinstallation.
30027 [OM30027] Failed in login. Input an invalid user information. Input user information correctly.
30027 [RE30027] Failed in study Failed to start the study since the same Restart. If the same symptom recurs,
start. study as the study which is already perform a reinstallation.
controlled in DR-ID 300CL was received
from MWM.
The same studies are the studies which
the following tag values are the same.
(0020,000d): Study Instance UID
(0040,0009): Scheduled Procedure StepID
30028 [OM30028] The specified The specified user has already been -
user is logged in. logged in.
30029 [OM30029] This IC card is IC card is not registered. Register IC card. Also use the registered
not registered. IC card.
30030 [OM30030] Switch the Change the radiographer. -
name of a technologist
(Technologist name : %s)
30031 [OM30031] Cannot unlock Tempted to unlock it with a user other than Only the logged in user account can unlock
the registered one. it.
30031 [TS30031] The corrected The corrected information will be saved. -
information will be saved.
30035 [ST30035] Search could Search could not be performed from the Enter search criteria again for a search. If
not be performed from the ordering system. the error persists, make sure ordering
ordering system. equipment has the data.
30035 [RE30035] No results were No results were found through RIS search. Enter search criteria again for a search. If
found through RIS search. the error persists, make sure RIS has the
data.
30037 [RE30037] The history of The history of patient ID issuance will be
-
patient ID issuance will be deleted.
deleted.
30037 [ST30037] The study The study contains image(s) retrieved from Click [OK] to continue the operation.
contains image(s) retrieved a media.
from a media.
30038 [QA30038] Failed in Failed in calculating the S or L value. Specify two points for ROI again.
calculating the S or L value.
30038 [RE30038] A study that is A study that is unable to execute exists. Confirm the study status.
unable to execute exists.
30038 [ST30038] The patient The patient information has been modified. Confirm that all information for the study in
information has been execution has been modified correctly, and
modified. then complete the study.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-153
Event Event Source Description Causes Remedies Remarks
30039 [ST30039] The study The study contains Tomosynthesis Click [OK] to continue the operation.
contains Tomosynthesis menu(s).
menu(s).
30040 [QA30040] Invalid Invalid combination has been selected. Switching of general exposure will be
combination has been changed to the default value.
selected.
30040 [RE30040] All the patient ID All the patient ID of the facility identification -
of the facility identification code currently set up will be cleared.
code currently set up will
be cleared.
30040 [ST30040] Failed in Failed in relocating the study information. Check the study information registered in
relocating the study RIS.
information.
30041 [QA30041] This function This function cannot be executed. Click the button to proceed the process.
cannot be executed.
30041 [RE30041] Patient ID that Patient ID that will be issued is duplicated. Confirm the patient ID.
will be issued is duplicated.
30042 [QA30042] This function This function cannot be executed. Click the button to proceed the process.
cannot be executed.
30042 [RE30042] There is a study There is a study not in the Waiting QA list. Confirm the study status.
not in the Waiting QA list.
30042 [ST30042] This function This function cannot be executed. Press the button below to continue the
cannot be executed. processing.
30043 [QA30043] Markers exist in Markers exist in the ES paired image. If [No] is selected, the operation will be
the ES paired image. Do continued without deleting the markers.
you wish to delete the
markers and execute the
operation?
30044 [QA30044] Images cannot Images cannot be stitched because those Adjust the image position.
be stitched because those images are not linked with one another.
images are not linked with
one another.
30045 [QA30045] It exceeds the It exceeds the maximum number of Select ViewModifier not more than 4.
maximum number of ViewModifier.
ViewModifier.
30046 [QA30046] View modifier View modifier code sequence items cannot Select an image with view code data.
code sequence items be changed.
cannot be changed.
30046 [RE30046] Switch the Switch the name of a technologist. -
name of a technologist
(Technologist name: %s)
30047 [RE30047] The specified The specified user is logged in. -
user is logged in.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-154
Event Event Source Description Causes Remedies Remarks
30048 [QA30048] It may not have It may not have been properly stitched. Check the image.
been properly stitched.
30048 [RE30048] Your IC card is Failed in authenticating IC card. Or utilized Check if it is the registered IC card, and
not recognized. an unregistered IC card. read the IC card again.
30049 [QA30049] Image The right-and-left mammography image Click [OK] and the trimming position will be
positioning will be trimming position is changed. changed.
executed.
30049 [RE30049] Your IC card Failed in reading IC card. Restart. If the same symptom recurs,
was not read. perform a reinstallation.
30050 [RE30050] Failed in login. Input an invalid user information. Input user information correctly.
30051 [QA30051] It will be fixed Change to the long-view stitched image Click the button that displayed on the
as a long-view stitched (WS) and confirm it. screen.
image (WS).
30052 [QA30052] It will be fixed Change to the general image (RT) and Click the button that displayed on the
as a general image (RT). confirm it. screen.
30054 [QA30054] Trimming frame Trimming frame is set on the place that Set the trimming frame correctly.
is set to where “Cut” is trimming function cannot be implemented.
inexecutable.
30209 [MD30209] System error System error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
30216 [ST30216] There is no There is no settable trimming size in %s. Restart the system.
settable trimming size in If the same error recurs after restarting the
%s. system, reinstall the system.
30507 [ST30507] A communication A communication error with Ascent View Restart DR-ID 300CL. If the same
error with Ascent View occurred. phenomenon recurred, reinstall DR-ID
occurred. 300CL.
30600 [TS30600] The output The output range is incorrect. Check the output range. The output range
range is incorrect. returns to the value
before the setting.
30902 [TS30902] Do you want to The reconstruction processing is tried to Click [OK] to cancel the reconstruction
cancel the reconstruction be stopped. processing.
processing?
30903 [TS30903] Do you want to The reconstruction processing is tried to Click [OK] to cancel the reconstruction
cancel the reconstruction be stopped. processing.
processing even if it cannot
be executed again?
31000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
31003 [MD31003] Obtaining Obtaining images has been completed. Click the button and continue the
images has been operation.
completed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-155
Event Event Source Description Causes Remedies Remarks
31004 [MD31004] Obtaining Obtaining images has been canceled. Click the button and continue the
images has been canceled. operation.
31099 [MD31099] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
31108 CslAdv_Re Failed in switching current
Failed to change the current user when Restart DR-ID 300CL.
user starting studies from the work list.
31114 [ST31114] Failed in Failed in specifying the urgent processing. Restart the system.
specifying the urgent If the same error recurs after restarting the
processing. system, reinstall the system.
31249 [ST31249] EDR mode is EDR mode is illegal. Confirm the EDR Mode.
illegal.
31270 [ST31270] Input value is Input value is illegal. Re-input correctly.
illegal.
31305 CslAdv_Study Failed warning SL It fails when display warning SL value. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
31306 CslAdv_Study Failed confirming IRS It fails when finish dialog show the detail Restart DR-ID 300CL. If the same
confirming IRS analyze. phenomenon recurred, reinstall DR-ID
300CL.
31344 [ST31344] Image size Image size could not be identified. *The Image size sent from the reading
could not be identified. device has a problem.
Check the device.
31350 [ST31350] The parameter The parameter of reconstruction condition Check the reconstruction condition The following items are
of reconstruction condition is invalid. parameters. updated. (Unit: mm)
is invalid. 1. Slice center height
2. Slice range
3. Slice interval
31360 [ST31360] An image of An image of which the height of the slice is Check the height of the slice.
which the height of the slice 0 to 500mm can only be generated. Other
is 0 to 500mm can only be images cannot be generated.
generated. Other images
cannot be generated.
31398 [MD31398] Failed in Failed in terminating. Try the operation again.
terminating.
31407 [ST31407] Failed in setting Failed in setting the exposure size. Check the device.
the exposure size.
31408 [ST31408] Failed in setting Failed in setting the reference position for Confirm the connected device.
the reference position for exposure.
exposure.
31410 [MD31410] No response No response comes from the device that Check the device.
comes from the device that shares ID information.
shares ID information.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-156
Event Event Source Description Causes Remedies Remarks
31421 [MD31421] The exposure The exposure menu could not be Press the confirmation button at the bottom
menu could not be determined. of the window after reselecting the
determined. appropriate menus for the exposures.
31422 [MD31422] Not able to Not able to transit to High Sensitivity Mode. Select a correct type of DR exposure
transit to High Sensitivity panel.
Mode.
31423 [ST31423] Failed in Failed in stitching images. Restart the system.
stitching images. If the same error recurs after restarting the
system, reinstall the system.
31423 [MD31423] Not able to Not able to release High Sensitivity Mode. Reselect an exposure menu.
release High Sensitivity
Mode.
31429 [MD31429] Failed in setting Failed in setting patient information. Restart DR-ID 300CL.
patient information.
31465 [ST31465] It will be treated Failed in provisional linkage. Check the image and perform an exposure
as a regular image due to a again as needed.
failure of provisional
linkage.
31467 [ST31467] Failed in energy Failed in energy subtraction processing. Restart the system.
subtraction processing. If the same error recurs after restarting the
system, reinstall the system.
32000 CslAdv_MdIf Module=%1, Class=%2, Defective program (error). Restart. If the same phenomenon recurred, %1: File name
Method=%3, %4 reinstall the system. %2: Class name
%3: Method name
%4: Detail message
32000 ClsAdv_Tomo 1% An error occurred in TomoDataMgrCtrl.dll. Restart DR-ID 300CL. 1%: The spot of the
An error occurred in the image DataEntity error and its cause
control function.
32021 [ST32021] The study could The study could not be completed. Wait for a while and try again.
not be completed.
32023 [ST32023] Data for Data for exposure analysis is full. Back up and clear the data with the User If the operation is
exposure analysis is full. Utility. continued without
backing up and clearing
the data, saving the
analysis data will not be
guaranteed.
32025 [ST32025] failed in failed in relocating the exposure menu. Restart the system.
relocating the exposure If the same error recurs after restarting the
menu. system, reinstall the system.
32043 [ST32043] Failed in linking Failed in linking to the patient information Check the patient information for the study
to the patient information display terminal. that is currently in process and for the
display terminal. study that is displayed on the patient
information display terminal.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-157
Event Event Source Description Causes Remedies Remarks
32600 [TS32600] An error An error occurred when connecting to Check the status of PPU and the
occurred when connecting PPU. connection with PPU.
to PPU
32601 [TS32601] Failed in Failed in acquiring the image file from Check the status of PPU.
acquiring the image file PPU. Click the button displayed on the screen.
from PPU.
32602 [TS32602] Failed in Failed in acquiring the image file from Check the status of PPU.
acquiring the image file PPU. Click the button displayed on the screen.
from PPU.
32603 [TS32603] A time error A time error occurred while an image file Check the status of PPU and the
occurred while an image was being copied. connection with PPU.
file was being copied.
32768 CslAdv_QA FFGetPixelImageSizeFor Failed in reading Directory.ini when Restart DR-ID 300CL. If the same
DRImage outputting the image. phenomenon recurred, reinstall DR-ID
Directory.ini is NotFound. 300CL.
32768 CslAdv_QA FFGetPixelImageSizeFor Failed in reading the size definition file Restart DR-ID 300CL. If the same %1: The path of the
DRImage when outputting the image. phenomenon recurred, reinstall DR-ID size definition file.
File Open Error. File Path = 300CL.
%1
32768 CslAdv_QA FFGetPixelImageSizeFor Format of the loaded size definition file is Restart DR-ID 300CL. If the same %1: The path of the
DRImage abnormal while outputting image. phenomenon recurred, reinstall DR-ID size definition file.
File Format Error. File Path 300CL.
= %1
32768 CslAdv_QA FFGetPixelImageSizeFor An error occurred in file converting process Restart DR-ID 300CL. If the same
DRImage while outputting image. phenomenon recurred, reinstall DR-ID
Other Error. 300CL.
34019 [ST34019] A menu in which A menu in which reconstruction processing Check the target menu.
reconstruction processing has not been executed exists in the study. Specify mis-exposure for a Tomosynthesis
has not been executed menu which does not require
exists in the study. reconstruction processing and complete
the study.
34020 [ST34020] A menu which is A menu which is currently in process of Suspend the study.
currently in process of reconstruction exists in the study. The study is not automatically completed
reconstruction exists in the even after the reconstruction image
study. processing ends. After resuming the study,
check the image and complete the study.
34021 [ST34021] Please check The reconstructed image in the study Check the reconstructed image.
the reconstructed image in needs to be checked.
the study.
34022 [ST34022] A study that has A study that has been suspended due to Check the completion state of the study.
been suspended due to exposure conditions exists. If it contains a Tomosynthesis menu,
exposure conditions exists. resume the study to check the image and
then complete the study.
34038 [ST34038] The target menu The target menu is in process of Cancel the reconstruction processing for
is in process of reconstruction processing. the target menu and specify mis-exposure.
reconstruction processing.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-158
Event Event Source Description Causes Remedies Remarks
34048 [ST34048] An attempt was The number of menus exceeds the range Display the menu within the range which
made to display an image available for image display. can be displayed.
outside the specified range.
34049 [ST34049] A menu that A menu that exceeds the range available Select the menu within the range which
exceeds the range for image display was selected. can be displayed.
available for image display
was selected.
34050 [ST34050] The number of The number of registered menus exceeds Display the registration menu within the
registered menus exceeds the range available for image display. range which can be displayed.
the range available for
image display.
34057 [ST34057] Too many Too many menus are included in the study. Reduce the number of menus to %s or
menus are included in the less and perform exposures.
study.
34061 [ST34061] Failed in Failed in cancelling the right-and-left Review the trimming position on the right-
cancelling the right-and-left mammography image positioning. and-left images.
mammography image
positioning.
34063 [ST34063] Some of the Some of the set-menu images have not Input images in all the set-menus and then
set-menu images have not been received yet. operate again.
been received yet.
34064 [ST34064] A study that A study that includes this menu will be Select [OK] to delete them.
includes this menu will be deleted. Select [CANCEL] not to delete them.
deleted.
34065 [ST34065] Failed in Failed in positioning the right-and-left Check the image orientation and others,
positioning the right and left mammography images. and then re-perform image positioning.
mammography images.
34066 [ST34066] Image The currently set trimming position was Click [OK] and the trimming position will be
positioning will be changed. changed.
executed.
34067 CslAdv_Study Read QAExpansionSetting Failed in reading the setup file of image Restart DR-ID 300CL. If the same error
config fail : invalid area operation unit (invalid Argument). recurs after restarting, reinstall DR-ID
Argument 300CL.
34068 [ST34068] It may not have It may not have been properly stitched. In case of an error in the stitching
been properly stitched. processing, perform manual stitching
processing.
34068 CslAdv_Study Read QAExpansionSetting Failed in reading the setup file of image Restart DR-ID 300CL. If the same error
config fail : wrong format area operation unit (wrong format). recurs after restarting, reinstall DR-ID
300CL.
34069 [ST34069] S/L values are S/L values are outside the specified range. Check the exposure conditions.
outside the specified range.
34072 CslAdv_Study Read QAExpansionSetting Failed in reading the setup file of image Restart DR-ID 300CL. If the same error
config fail : area operation unit (OverflowException). recurs after restarting, reinstall DR-ID
OverflowException 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-159
Event Event Source Description Causes Remedies Remarks
34073 CslAdv_Study Read QAExpansionSetting Failed in reading the setup file of image Restart DR-ID 300CL. If the same error
config fail area operation unit (System error). recurs after restarting, reinstall DR-ID
300CL.
34074 CslAdv_Study SaveExpansionSetting error : Failed in saving the setup file of image Restart DR-ID 300CL. If the same error
ConfigurationErrorsException area operation unit recurs after restarting, reinstall DR-ID
(ConfigurationErrorsException). 300CL.
34075 [ST34075] Failed in Failed in positioning the right-and-left Adjust the trimming position manually.
positioning the right and left mammography images.
mammography images.
34075 CslAdv_Study SaveExpansionSetting Failed in saving the setup file of image Restart DR-ID 300CL. If the same error
error area operation unit (System error). recurs after restarting, reinstall DR-ID
300CL.
34080 [ST34080] Images cannot Images cannot be distributed because they Input images and stitch them appropriately.
be distributed because they are not stitched. Perform the necessary operations again.
are not stitched.
34107 [ST34107] Do you wish to You tried to delete markers while executing Select [Yes] to delete them.
delete the markers and the operation. Select [No] not to delete them.
execute the operation?
34108 [ST34108] Images Images generated from different image Press the button below to abort
generated from different generators cannot be replaced each other. processing.
image generators cannot
be replaced each other.
34109 [ST34109] Images were Images were input during the preview Try stitching operations again.
input during the preview display.
display.
34114 [ST34114] Conditions Conditions inhibiting the start of exposure Check the exposure conditions on the
inhibiting the start of had been sent from the external device. external device.
exposure had been sent
from the external device.
34118 [ST34118] Failed in menu Failed in menu selection. Select a menu again.
selection.
34300 [ST34300] The number of The number of patient information exceeds Search by the key word.
patient information exceeds the number that can be indicated.
the number that can be
indicated.
34301 [ST34301] Patient Patient information beyond the number Change the key word, and search again.
information beyond the which can be indicated was found.
number which can be
indicated was found.
34302 [ST34302] The QC study The QC study will be saved. Click [OK] and the QC study will be saved.
will be saved.
34303 [ST34303] Failed in Failed in initialization of Smart-QA. Restart DR-ID 300CL. If the same
initialization of Smart-QA. phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-160
Event Event Source Description Causes Remedies Remarks
34303 CslAdv_Study Failed initialze IRS It fails when initialize IRS Framwork. Restart DR-ID 300CL. If the same
phenomenon recurred, reinstall DR-ID
300CL.
34304 [ST34304] Failed in Failed in analyzing Smart-QA. Press the button below to continue other
analyzing Smart-QA. processing.
34304 CslAdv_Study Failed implement analyze It fails when Implement analyze IRS image. Restart DR-ID 300CL. If the same
IRS image phenomenon recurred, reinstall DR-ID
300CL.
34305 [ST34305] Failed in Failed in displaying Smart-QA analysis Press the button below to continue other
displaying Smart-QA result. processing.
analysis result.
34306 [ST34306] Failed in Failed in processing the image correction. Press the button below to continue other
processing the image processing.
correction.
34307 [ST34307] Failed in Failed in executing the exposure guidance Restart DR-ID 300CL. If the same
executing the exposure function. phenomenon recurred, reinstall DR-ID
guidance function. 300CL.
34313 [ST34313] Unfinished Unfinished exposure menu is selected. Stitch after selecting the menu for the
exposure menu is selected. density standard.
34317 [ST34317] Failed in Failed in matching the density of long-view Adjust the density manually.
matching the density of image stitching processing.
long-view image stitching
processing.
34318 [ST34318] Failed in setting Failed in setting of auto marker. Check the information of the automatic
of auto marker. marker.
34319 [ST34319] Failed in Failed in duplicating the exposure menu. Select an exposure menu registered with
duplicating the exposure the User Utility.
menu.
34320 [ST34320] The selected The selected studies contain a study which Check the condition of study.
studies contain a study cannot be started.
which cannot be started.
34321 [ST34321] Failed in Failed in initialization. Restart DR-ID 300CL.
initialization. If the same phenomenon recurred, reinstall
DR-ID 300CL.
34322 [ST34322] Failed in Failed in obtaining delivery information. Restart DR-ID 300CL.
obtaining delivery The target image will not be delivered. If the same phenomenon recurred, reinstall
information. The target DR-ID 300CL.
image will not be delivered.
34324 [ST34324] Cannot be Cannot be delivered because the Wait until the reconstruction is completed,
delivered because the reconstruction has not been completed. and then perform operations again.
reconstruction has not been
completed.
34325 [ST34325] Cannot be Cannot be delivered due to the mis- Release the mis-exposure setting, and
delivered due to the mis- exposure setting. then perform operations again.
exposure setting.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-161
Event Event Source Description Causes Remedies Remarks
34326 [ST34326] The already- The already-delivered image will be Click [OK] and the already-delivered image
delivered image will be delivered again. will be delivered again.
delivered again.
34327 [ST34327] A long-view A long-view menu exists with no IP Register the IP in a long-view menu.
menu exists with no IP registered.
registered.
34328 [ST34328] The exposure The exposure time exceeded the limit for Set the exposure time within the limit for
time exceeded the limit for Tomosynthesis. Tomosynthesis.
Tomosynthesis.
34329 [ST34329] Separate the Separating the stitched image and Click the button and continue the
stitched images to return returning to the status to before stitching. operation.
them to the state before
they were stitched.
34330 [ST34330] The stitched Due to being distributed, the stitched Stop the output queue of the connected
images cannot be image cannot be separated. device status from sending, or perform
separated because they again after sending.
are being distributed.
34331 [ST34331] Distributed Replacing the stitched image due to being Click the button and continue the
images will be replaced. distributed. operation.
34332 [ST34332] Images cannot Images cannot be replaced because they Stop the output queue of the connected
be replaced because they are being distributed. device status from sending, or perform
are being distributed. again after sending.
34333 [ST34333] The image The image library window cannot be Wait for a while and try again.
library window cannot be displayed because there are images in the
displayed because there process of being displayed in the
are images in the process ImageBox.
of being displayed in the
ImageBox.
34334 [ST34334] Failed in starting Failed in starting the image library window. Click the button to restart the system.
the image library window. If the same error recurs after restarting the
system, reinstall the system.
34335 [ST34335] The device you The device out of the target of memory Select the other selector.
selected does not support exposure mode is selected.
the memory exposure
mode.
34336 [ST34336] No image exists No image exists in the specified FPD's Click the button and continue the
in the specified FPD's ImageBox. operation.
ImageBox.
34337 [ST34337] As images are As images are not input in to the first Enter image in the first menu, and try
not input in to the first menu, menu, they cannot be stitched. again.
they cannnot be stitched.
34338 [ST34338] Menus that can Menus that can duplicate an image are Select a menu with unregistered IPs and
duplicate an image are not selected for the duplication destination. uninput images for the duplication
selected. destination.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-162
Event Event Source Description Causes Remedies Remarks
34339 [ST34339] Long-view, Long-view, energy subtraction, or For the duplication destination, select a
energy subtraction, and tomosynthesis menu are selected. menu other than long-view, energy
tomosynthesis menu subtraction, and tomosynthesis.
cannot duplicate an image.
34340 [ST34340] Duplicating an Duplicating an image has been failed. Perform operations again.
image has been failed. If the same error recurs, restart the system.
34341 [ST34341] Image effective The change for the image effective area Click the button displayed on the screen.
area has been changed. has been confirmed.
You cannot use a function
for changing image
effective area.
34342 [ST34342] Since Intend to change the image effective area Suspend the image distribution or change
distribution is being carried while image distribution. the image effective area after the
out, you cannot use a distribution is over.
function for changing image
effective area.
34343 [ST34343] Image effective The image effective area has not been Click [OK] to finish the study, or click
area has not been set. changed. There are images that exposure [Cancel] and set the image effective area
There is a deviation from region are smaller than 17”X17” in the for the image.
the automatic exposure study.
recognition result.
Automatic exposure
recognition result:
Within 17”X17”
Image effective area:
17”X17”X3
34344 [ST34344] Image effective The image effective area has not been Click [OK] to suspend the study, or click
area has not been set. changed. There are images that exposure [Cancel] and set the image effective area
There is a deviation from region are smaller than 17”X17” in the for the image.
the automatic exposure study.
recognition result.
Automatic exposure
recognition result:
Within 17”X17”
Image effective area:
17”X17”X3
40000 [RE40000] Movement Movement cannot be executed because Execute again after the image is input.
cannot be executed no image is input.
because no image is input.
40001 [RE40001] The source The source image cannot be moved Execute again after the output is
image cannot be moved because it is being output. completed.
because it is being output.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-163
Event Event Source Description Causes Remedies Remarks
40002 [RE40002] Movement Movement cannot be executed because Execute again after a while.
cannot be executed the destination study contains an image
because the destination waiting to be received.
study contains an image
waiting to be received.
40003 [RE40003] The image The image cannot be moved because the Execute again after the output is
cannot be moved because destination study is being output. completed.
the destination study is
being output.
40004 [RE40004] Images in a Images in a suspended study cannot be Move the image which is not in a
suspended study cannot be moved into a QA-completed study. suspended study.
moved into a QA-completed
study.
40005 [RE40005] Images in a Images in a QA-not-completed study Move the image after the QA is completed.
QA-not-completed study cannot be moved into a QA-completed
cannot be moved into a study.
QA-completed study.
40006 [RE40006] Images in a Images in a suspended study cannot be Move the image which is not in a
suspended study cannot be moved into a study waiting for QA. suspended study.
moved into a study waiting
for QA.
40007 [RE40007] Images in a Images in a study waiting for QA cannot be Move the image after the QA is completed.
study waiting for QA cannot moved into a suspended study.
be moved into a suspended
study.
40008 [RE40008] The image will The image will be moved. Click [OK] and the image will be moved.
be moved
>Are you sure?
40010 [RE40010] The source The source image cannot be moved Execute again after a while.
image cannot be moved because it is included in a study which is
because it is included in a receiving images.
study which is receiving
images.
40011 [RE40011] The source The source image cannot be moved To move the source image, cancel the
image cannot be moved because the right-and-left mammography positioning.
because the right-and-left image positioning has been performed on
mammography image it.
positioning has been
performed on it.
40012 [RE40012] The entry format The entry format of the specified period Input it correctly.
of the specified period (From) is invalid.
(From) is invalid.
40013 [RE40013] The entry format The entry format of the specified period Input it correctly.
of the specified period (To) (To) is invalid.
is invalid.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-164
Event Event Source Description Causes Remedies Remarks
40014 [RE40014] "To" has an To has an earlier date than “From” of the Input it correctly.
earlier date than "From" of specified period.
the specified period.
40015 [RE40015] The specified The specified image has been deleted. Update the display and execute the
image has been deleted. processing again.
40016 [RE40016] The study The study specified as the destination has Update the display and execute the
specified as the destination been deleted. processing again.
has been deleted.
40017 [RE40017] Moving images Moving images to the specified hierarchy Check the system settings.
to the specified hierarchy is is prohibited by the system settings.
prohibited by the system
settings.
40018 [RE40018] The source The source image cannot be moved Execute again after the study is completed.
image cannot be moved because it is being used for a study in
because it is being used for another system.
a study in another system.
40019 [RE40019] The image The image cannot be moved because the Execute again after the study is completed.
cannot be moved because destination study is being executed in
the destination study is another system.
being executed in another
system.
40020 [RE40020] Images of the Images of the study in modality: %s cannot Check the study modality.
study in modality: %s be moved into the study of modality:%s.
cannot be moved into the
study of modality:%s.
40021 [RE40021] You have no You have no authority to move images Check the authority.
authority to move images between studies.
between studies.
40022 [RE40022] The study The study information has been changed. Update the display and execute the
information has been processing again.
changed.
40023 [RE40023] Required hard Hard disk space is insufficient. Delete the study and execute again.
disk space is insufficient.
40024 [RE40024] The keyword The keyword contains an illegal character. Input it correctly.
contains an illegal
character.
40025 [RE40025] The date format The date format is invalid. Input it correctly.
is invalid.
40026 [RE40026] Filtering could Filtering could not be executed. Enter information in the marked items
not be executed. correctly.
40027 [RE40027] The source The source image cannot be moved Execute again after the reconstruction is
image cannot be moved because it is being reconstructed. completed.
because it is being
reconstructed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-165
Event Event Source Description Causes Remedies Remarks
40028 [RE40028] In In Tomosynthesis menu, images cannot be Check the study information.
Tomosynthesis menu, moved to another system.
images cannot be moved to
another system.
40029 [RE40029] Failed in moving Failed in moving the image. Restart the system.
the image. If the same error recurs after restarting the
system, reinstall the system.
40073 FW Can not get image data The bitmap cannot be loaded. Restart the system.
from file path: "{0}"
40073 FW Can not get image depth. The image depth cannot be acquired from Restart the system.
Path="{0}" DDO.
40073 FW Can not get image file path. The path of the bitmap cannot be acquired Restart the system.
Path="{0}" from DDO.
40073 FW Can not get image heigh. The image height cannot be acquired from Restart the system.
Path="{0}" DDO.
40073 FW Can not get image witdh. The image width cannot be acquired from Restart the system.
Path="{0}" DDO.
40073 FW Can not get the gamma LUT table cannot be created. Restart the system.
table
40073 FW Can not get the image No memory to assign for image loading. Restart the system.
buffer
40073 FW Can not get the XML AC cannot acquire the "create xml Check the AC parameter file again and
parameter parameter" from DDO file. restart the system.
50000 FFOutImgPrc Initialize Error. Failed in initializing COM. Restart DR-ID 300CL. %1: Error code
%1 : Com Initialize Error < If the same phenomenon recurred, reinstall %2: Method name
%2 > DR-ID 300CL.
50001 FFQCMsgBoxCtrl LoadImage An error occurred in reading DDO file (hq). Check the appropriateness of the data of
Open m_QCDataInfo({0}). the DDOFile (.hqFile) and restart DR-ID
strStdFilePath For Binary 300CL.
Access Read As #1
50001 FFQCMsgBoxCtrl LoadImage The data of DDO (INF) are not appropriate. Check the appropriateness of the data of
GetRotateParam({0}): the DDOFile (INF) and restart DR-ID
lRet={1} 300CL.
50002 FFOutImgPrc FFSysConfig::Initialize Failed in reading the set values from Confirm whether the system settings are %1: GetLastError()
Error < Run_Prc > system settings. set properly in UserUtility and return value
[LastError=%1] ServiceUtility. If the error is caused by the %2: Error code
[ErrorCode=%2] system, reinstall the system.
50003 FFOutImgPrc ACCMenuMasterCalc Failed in calling the method of menu Restart. If the same phenomenon recurred, %1: Call return value
MaskSize Failed nRet=(%1) master parameter accessor. reinstall the system. %2: Name of function
(%2) < ImgPrcExec > error occurred
50003 FFOutImgPrc ACCMenuMasterGet Failed in calling the method of menu Restart. If the same phenomenon recurred, %1: Call return value
MaskSize Failed nRet=(%1) master parameter accessor. reinstall the system. %2: Name of function
(%2) < ImgPrcExec > error occurred
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-166
Event Event Source Description Causes Remedies Remarks
50003 FFOutImgPrc ACCMenuMaster Failed in calling the method of menu Restart. If the same phenomenon recurred, %1: Call return value
LoadParams Failed master parameter accessor. reinstall the system. %2: Name of function
nRet=(%1) (%2) error occurred
< ImgPrcExec >
50003 FFOutImgPrc ConvertImageProc Failed in converting the image processing Restart. If the same phenomenon recurred, %1: Call return value
Parameter Failed lRet=(%1) parameters. reinstall the system. %2: Name of function
(%2) < ImgPrcExec > error occurred
50004 FFOutImgPrc ACCAppImgCreator Acquiring accessor of CSL image Restart. If the same phenomenon recurred, %1: Class and function
WSLifeSizeProcGetter generation (life size composition) image reinstall the system. name
Failed (%1) <ImgPrcExec> processing common parameters returned
an error.
50004 FFOutImgPrc ACCAppImgPreProc CSL preprocessing facade parameter Restart. If the same phenomenon recurred, %1: Name of caller
GammaTblGetter Failed accessor returned an error. reinstall the system. function
(%1) < ImgPrcExec >
50004 FFOutImgPrc ACCAppImgProc%1Getter CSL processing facade parameter Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec accessor returned an error. reinstall the system. function
>
50004 FFOutImgPrc ACCMenuMasterInitialize Initialization function of menu master Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec parameter accessor returned an error. reinstall the system. function
>
50004 FFOutImgPrc ACCMenuMasterSetMaster Perpetuation accessor of image processing Restart. If the same phenomenon recurred, %1: Name of caller
ParameterToDDO Failed common parameters returned an error. reinstall the system. function
(%1) < ImgPrcExec >
50004 FFOutImgPrc ACCModImgProcGetter FCR processing facade parameter Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec accessor returned an error. reinstall the system. function
>
50004 FFOutImgPrc BC::Facade Notification An error occurs inside the parameter Restart. If the same phenomenon recurred, %1: Error type
Type[%1] Code[%2] accessor. reinstall the system. %2: Error code
Time[%3] Message[%4] < %3: Error time and date
ImgPrcExec > %4: Internal error
information
50004 FFOutImgPrc CFcRawMake:: Output DDO file name acquiring function of Restart. If the same phenomenon recurred, %1: Name of caller
ExeImageProcess Failed creating RAW data returned an error. reinstall the system. function
(%1) < ImgPrcExec >
50004 FFOutImgPrc CFcRawMake:: Output DDO file name acquiring function of Restart. If the same phenomenon recurred, %1: Name of caller
GetOutDICOMFName creating RAW data returned an error. reinstall the system. function
Failed (%1) < ImgPrcExec
>
50004 FFOutImgPrc CFcRawMake:: RAW file name acquiring function of Restart. If the same phenomenon recurred, %1: Name of caller
GetRAWFName Failed creating RAW data returned an error. reinstall the system. function
(%1) < ImgPrcExec >
50004 FFOutImgPrc CFcRawMake:: Initialization setting common function of Restart. If the same phenomenon recurred, %1: Name of caller
ImgRawMakeInit Failed creating RAW data returned an error. reinstall the system. function
(%1) < ImgPrcExec >
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-167
Event Event Source Description Causes Remedies Remarks
50004 FFOutImgPrc CharToBstr Failed (%1) < Function of converting from C character Restart. If the same phenomenon recurred, %1: Name of caller
ImgPrcExec > strings to BSTR type character strings reinstall the system. function
returned an error.
50004 FFOutImgPrc Create Instance Error (%1) Instance generation function of parameter Restart. If the same phenomenon recurred, %1: Class and function
<ImgPrcExec> accessor returned an error. reinstall the system. name
50004 FFOutImgPrc DOM GetAttributeValue Tag value configuration function of DOM Restart. If the same phenomenon recurred, %1: Absolute location
Failed TAG=%1 ATTR=%2 method returned an error. reinstall the system. path
(%3) < ImgPrcExec > %2: Tag name
%3: Name of caller
function
50004 FFOutImgPrc DOM GetSameTagNumber Function in acquiring the number of same Restart. If the same phenomenon recurred, %1: Absolute location
Failed TAG=%1 (%2) < tags for DOM method returned an error. reinstall the system. path
ImgPrcExec > %2: Name of caller
function
50004 FFOutImgPrc DOM GetTagValue Failed Tag value acquiring function of DOM Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 (%2) < method returned an error. reinstall the system. path
ImgPrcExec > %2: Name of caller
function
50004 FFOutImgPrc DOM LoadXmlString Failed XML file read function of DOM method Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > returned an error. reinstall the system. function
50004 FFOutImgPrc DOM RemoveAttribute Deleting attribution function of DOM Restart. If the same phenomenon recurred, %1: Absolute location
Failed TAG=%1 ATTR=%2 method returned an error. reinstall the system. path
(%3) < ImgPrcExec > %2: Attribution name
%3: Name of caller
function
50004 FFOutImgPrc DOM SaveXmlString Failed Saving XML file function of DOM method Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > returned an error. reinstall the system. function
50004 FFOutImgPrc DOM SetAttribute Failed Attribution value configuration function of Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 ATTR=%2 (%3) < DOM method returned an error. reinstall the system. path
ImgPrcExec > %2: Attribution name
%3: Name of caller
function
50004 FFOutImgPrc DOM SetTagValue Failed Tag value configuration function of DOM Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 (%2) < method returned an error. reinstall the system. path
ImgPrcExec > %2: Name of caller
function
50004 FFOutImgPrc DOM TAG Does Not Exist Checking tag existence function of DOM Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 (%2) < method returned an error. reinstall the system. path
ImgPrcExec > %2: Name of caller
function
50004 FFOutImgPrc FCDAppImgCreator Life size composition facade of long view Restart. If the same phenomenon recurred, %1: Class and function
WSLifeSize Failed (%1) image returned an error. reinstall the system. name
<ImgPrcExec>
50004 FFOutImgPrc FCDAppImgPreProc γ table generation processing facade Restart. If the same phenomenon recurred, %1: Name of caller
GammaTbl Failed (%1) < returned an error. reinstall the system. function
ImgPrcExec >
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-168
Event Event Source Description Causes Remedies Remarks
50004 FFOutImgPrc FCDAppImgProc%1 Failed CSL image processing facade returned an Restart. If the same phenomenon recurred, %1: Distribution
(%1) < ImgPrcExec > error. reinstall the system. destination
recognition
("Strg"/"Print")
%2: Image type
character string
50004 FFOutImgPrc FCDManualMarginProc Adding margin (manual arrangement) Restart. If the same phenomenon recurred,
%1: Class and function
Failed (%1) <ImgPrcExec> facade of an image returned an error. reinstall the system. name
50004 FFOutImgPrc FCDModImgProc Failed FCR processing control function returned Restart. If the same phenomenon recurred,
%1: Name of caller
(%1) < ImgPrcExec > an error. reinstall the system. function
50004 FFOutImgPrc FCDTrimmingProc Failed Cutting out process (trimming process) of Restart. If the same phenomenon recurred,
%1: Class and function
(%1) <ImgPrcExec> an image returned an error. reinstall the system. name
50004 FFOutImgPrc FCDWriteLosslessJPEG LosslessJPEG compression output facade Restart. If the same phenomenon recurred,
%1: Name of caller
Failed (%1) < ImgPrcExec of RAW data returned an error. reinstall the system. function
>
50004 FFOutImgPrc GetIndexOfImagePixel Not image attribution index value acquiring Restart. If the same phenomenon recurred, %1: Requesting IN/
Found TAG:image_pixel function of image processing common reinstall the system. OUT recognition
ATTR=%1 (%2) < parameters returned an error. %2: Name of caller
ImgPrcExec > function
50004 FFOutImgPrc GetValueFromXMLTag Tag value acquiring function of image Restart. If the same phenomenon recurred, %1: Absolute location
Failed TAG=%1:%2 (%3) < processing common parameters returned reinstall the system. path
ImgPrcExec > an error. %2: Tag name/
attribution name
%3: Name of caller
function
50004 FFOutImgPrc PushLUTDataToTag failed DDO tag setting processing function of γ Restart. If the same phenomenon recurred, %1: Name of caller
(%1) <ImgPrcExec> table returned an error. reinstall the system. function
50004 FFOutImgPrc RemoveXMLTag Failed Tag value deleting function of image Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 (%1) < processing common parameters returned reinstall the system. path
ImgPrcExec > an error. %2: Name of caller
function
50011 FFOutImgPrc DICOM ReplaceVR/Push Failed in acquiring the character string Restart. If the same phenomenon recurred,
Error < region. reinstall the system.
ImgConvReadyDicomDXSt
>
50011 FFOutImgPrc VR Area Get Error DDO tag replace function returned an Restart. If the same phenomenon recurred, %1: Function return
<ImgConvReadyDicomSt> error. reinstall the system. value
50011 FFOutImgPrc VR Area Get Error DDO tag character replace function Restart. If the same phenomenon recurred, %1: Function return
<ImgConvReadyDicomSt> returned an error. reinstall the system. value
50011 FFOutImgPrc VR Area Get Error DDO tag character replace function Restart. If the same phenomenon recurred, %1: Function return
<ImgConvReadyDicomSt> returned an error. reinstall the system. value
50012 FFOutImgPrc DDT Error ( DDO = [%1] DDO conversion function by SOP class Restart. If the same phenomenon recurred, %1: DDO file path
SOPUID = [%2] AEName = returned an error. reinstall the system. %2: SOP Class UID
[%3] ) < ImgConvDicomSt %3: DistributionHost
>
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-169
Event Event Source Description Causes Remedies Remarks
50012 FFOutImgPrc DICOM Data GetVR Error ( Acquired function of DDO tag returned an Restart. If the same phenomenon recurred, %1: GroupID
Gr : 0x%1 Er : 0x%2 ) < error. reinstall the system. %2: ElementID
DelTrimInfo >
50012 FFOutImgPrc DICOM Data GetVR Error ( Acquired function of DDO tag returned an Restart. If the same phenomenon recurred, %1: GroupID
Gr : 0x%1 Er : 0x%2 Ret : error. reinstall the system. %2: ElementID
0x%3 ) < SetImgInf > %3: Function return
value
50012 FFOutImgPrc DICOM Data GetVR Error Acquired function of DDO tag returned an Restart. If the same phenomenon recurred, %1: Group number of
TAG=(%1,%2) lRet=(%3) error. reinstall the system. tag
(%4) < ImgPrcExec > %2: Element number of
tag
%3: Call return value
%4: Name of function
error occurred
50012 FFOutImgPrc DICOM Data SetVR Error Configuration function of DDO tag returned Restart. If the same phenomenon recurred, %1: Group number of
TAG=(%1,%2) lRet=(%3) an error. reinstall the system. tag
(%4) < ImgPrcExec > %2: Element number of
tag
%3: Call return value
%4: Name of function
error occurred
50012 FFOutImgPrc DICOM ReplaceVR/Push DDO tag replace function returned an Restart. If the same phenomenon recurred, %1: Function return
Error ( error code = %2 ) < error. reinstall the system. value
%1 > %2: Name of caller
function
50012 FFOutImgPrc Parameter Object not The return value of pointer acquiring Restart. If the same phenomenon recurred, %1: Name of caller
found. < %1 > function of sequence data is NULL. reinstall the system. function
50012 FFOutImgPrc Parameter ObjectArray not NULL is notified to the converted result of Restart. If the same phenomenon recurred, %1: Name of caller
found. < %1 > SQ tag to the sequence data. reinstall the system. function
50012 FFOutImgPrc SaveFile Error lRet=(%1) Failed in saving the DDO. Restart. If the same phenomenon recurred, %1: Call return value
(%2) < ImgPrcExec > reinstall the system. %2: Name of function
error occurred
50012 FFOutImgPrc SaveFile Failed File=%1 Saving common function returned an error Restart. If the same phenomenon recurred, %1: DDO file path
(%2) < ImgPrcExec > to DDO file. reinstall the system. name
%2: Name of caller
function
50012 FFOutImgPrc DICOM ReplaceVR/Push Failed in replacing the DDO tag. Perform delivery again. If the error is
Error < caused by the system, reinstall the system.
ImgConvAfterDicomSt >
50012 FFOutImgPrc DICOM GetDICOMData Failed in acquiring DDO. Perform delivery again. If the error is
Error < caused by the system, reinstall the system.
ImgConvAfterDicomSt >
50013 FFOutImgPrc GetDICOMData No Data ( Acquired function of DDO tag returned an Restart. If the same phenomenon recurred, %1: GroupID
Gr : 0x%1 Er : 0xXX%2) < error. reinstall the system. %2: ElementID
ImgFilmCharMake >
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-170
Event Event Source Description Causes Remedies Remarks
50020 FFOutImgPrc Convert SizeCode Error : Value conversion function of image size Restart. If the same phenomenon recurred, %1: Class and function
(%1) [SizeCode=%2] < code before/after trimming to the size code reinstall the system. name
ExeOutMain > returned an error. %2: Size code or
trimming size code
50020 FFOutImgPrc Convert SizeCode Error : Character string conversion function of Restart. If the same phenomenon recurred, %1: Class and function
(%1) [SizeCode=%2] < trimming size code returned an error. reinstall the system. name
ExeOutMain > %2: Trimming size code
50020 FFOutImgPrc ConvertSizeCode Value conversion function of size code Restart. If the same phenomenon recurred, %1: Size code
CharToNum Error. returned an error. reinstall the system. %2: Function return
SizeCode=(%1), lRet=(%2) value
50020 FFOutImgPrc ConvertSizeCode Character string conversion function of Restart. If the same phenomenon recurred, %1: Size code
NumToChar Errror. size code returned an error. reinstall the system. %2: Function return
SizeCode=(%1), lRet=(%2) value
50020 FFOutImgPrc GetImageSize Error : Acquisition function of pixel/line returned Restart. If the same phenomenon recurred, %1: Size code
[SizeCode%1] < an error. reinstall the system.
ExeOutMain >
50020 FFOutImgPrc GetValueFromXMLTag Failed in acquiring a tag from XML. Restart. If the same phenomenon recurred, %1: Absolute path of
Failed TAG=%1:%2 reinstall the system. XML file
lRet=(%3) (%4) < %2: Tag name
ImgPrcExec > %3: Call return value
%4: Name of function
error occurred
50020 FFOutImgPrc Invalid FilmFormatString Energy subtraction processing image type Restart. If the same phenomenon recurred, %1: Image data index
(%1:%2) < characters were in error. reinstall the system. %2: Image type
ImgGetImageData > character string
50020 FFOutImgPrc Null or Invalid Notification parameter was in error. Restart. If the same phenomenon recurred, %1: Class and function
Parameter(%1) < reinstall the system. name
ExeOutMain >
50020 FFOutImgPrc S%1 = 0x%1 '%2' Error < Image processed flag value was out of Restart. If the same phenomenon recurred, %1: Image processed
ImgFilmCharMake > range. reinstall the system. flag position index
%2: Image processed
flag value
%3: Image processed
flag value
(character)
50020 FFOutImgPrc S%1 = 0x%2 Error < Image processed flag value was out of Restart. If the same phenomenon recurred, %1: Image processed
ImgFilmCharMake > range. reinstall the system. flag position index
%2: Image processed
flag value
50020 FFOutImgPrc Size Code Convert Error Value conversion function of size code Restart. If the same phenomenon recurred, %1: Size code
(SizeCode = %1) returned an error. reinstall the system.
50020 FFOutImgPrc CFcImgInf object is null < The object of image data class is not Perform delivery again. If the error is %s: Method name
%s > found. caused by the system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-171
Event Event Source Description Causes Remedies Remarks
50021 FFOutImgPrc OutputJob Impossible Specifying output processing being Restart. If the same phenomenon recurred, %1: GroupID
Returned Value acquired was in error. reinstall the system. %2: ElementID
(0x%1,0x%2) (%3) < %3: Function name
ImgPrcExec >
50021 FFOutImgPrc OutputJob Object Data Get Tag acquiring function of output JOB file Restart. If the same phenomenon recurred, %1: GroupID
(%1) Error < returned an error. reinstall the system.
ImgConvDicomSt >
50021 FFOutImgPrc OutputJob Object Data Get Tag acquiring function of output JOB file Restart. If the same phenomenon recurred, %1: GroupID
Error (0x%1,0x%2) (%3) < returned an error. reinstall the system. %2: ElementID
ImgPrcExec > %3: Name of caller
function
50021 FFOutImgPrc OutputJob Object Data Get Failed in acquiring information from the Check the setting parameter. If the error is %1: Output JOB tag
Error (%1,%2) (%3) delivery JOB file. caused by the system, reinstall the system. Function
lRet=(%4) < %2: Output JOB tag ID
ImgConvDicomSt > %3: Name of function
error occurred
%4: Function return
value
50030 FFOutImgPrc [MsxmlWrap] XML file read function of DOM method Restart. If the same phenomenon recurred, %1: Class and function
LoadXmlString Error (%1) returned an error. reinstall the system. name
<ImgPrcExec>
50030 FFOutImgPrc [MsxmlWrap] Saving XML file function of DOM method Restart. If the same phenomenon recurred, %1: Class and function
SaveXmlString Error (%1) returned an error. reinstall the system. name
<ImgPrcExec>
50030 FFOutImgPrc CalculateManualMargin Parameter acquiring function of margin Restart. If the same phenomenon recurred, %1: Name of caller
failed (%1) lRet=(%2) process returned an error. reinstall the system. function
<ImgPrcExec> %2: Function return
value
50030 FFOutImgPrc CalculateTrimming failed Magnification/reduction ratio calculation Restart. If the same phenomenon recurred, %1: Name of caller
(%1) lRet=(%2) function of trimming-related parameters reinstall the system. function
<ImgPrcExec> returned an error. %2: Function return
value
50030 FFOutImgPrc CalculateTrimmingFor1414Film Trimming region calculation function of "14 Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec > inch X 14 inch" single print returned an reinstall the system. function
error.
50030 FFOutImgPrc CalculateTrimmingPosition Calculation function of "14 inch X 14 inch" Restart. If the same phenomenon recurred, %1: Name of caller
For1414Film Failed (%1) film trimming process (cutting out direction reinstall the system. function
< ImgPrcExec > of edge of the image) returned an error.
50030 FFOutImgPrc Catch some other An exception occurred in the acquiring Restart. If the same phenomenon recurred, %1: Process
Exception (%1) < process of image invert type of the film reinstall the system. recognition
ImgFilmCharMake > character. character string
50030 FFOutImgPrc Catch some other An exception occurred in the image Restart. If the same phenomenon recurred, %1: Name of function
Exception (%1) < distribution process control function. reinstall the system. exception occurred
ImgPrcExec >
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-172
Event Event Source Description Causes Remedies Remarks
50030 FFOutImgPrc Catch some other An exception occurred during the process. Restart. If the same phenomenon recurred, %1: Value of exception
Exception LastError=(%1) reinstall the system. occurred
(%2) < ImgPrcExec > %2: Name of function
error occurred
50030 FFOutImgPrc CheckDiagnosisImage Determination common function of Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec diagnosis image returned an error. reinstall the system. function
>
50030 FFOutImgPrc CoCreateInstance Failed Generation method of instance returned an Restart. If the same phenomenon recurred, %1: ID
%1 (%2) < ImgPrcExec > error. reinstall the system. %2: Caller function
50030 FFOutImgPrc Convert CS/SS error. ( Modification function of CS tag and SS tag Restart. If the same phenomenon recurred, %1: Function return
error code = 0x%1 ) < %2 > returned an error. reinstall the system. value
%2: Name of caller
function
50030 FFOutImgPrc Convert CS/SS including Modification function of image parameters Restart. If the same phenomenon recurred, %1: Function return
image parameter error. ( returned an error. reinstall the system. value
error code = 0x%1 ) < %2 > %2: Name of caller
function
50030 FFOutImgPrc CopyInputImageArea Memory management copying function of Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec input image returned an error. reinstall the system. function
>
50030 FFOutImgPrc Create TrimmingProc Generation function of image processing Restart. If the same phenomenon recurred, %1: Class and function
Parameter Error (%1) common parameters returned an error. reinstall the system. name
<ImgPrcExec>
50030 FFOutImgPrc CreateMarginProc Generation function of image processing Restart. If the same phenomenon recurred, %1: Class and function
XmlPaaram Error (%1) common parameters returned an error. reinstall the system. name
<ImgPrcExec>
50030 FFOutImgPrc CreateXML_ Image processing common parameters of Restart. If the same phenomenon recurred, %1: Name of caller
ImageCompJPEG Failed LosslessJPEG compression returned an reinstall the system. function
(%1) < ImgPrcExec > error.
50030 FFOutImgPrc DDOPush Failed DDO writing common function returned an Restart. If the same phenomenon recurred, %1: GroupID
(0x%1,0x%2) (%3) < error. reinstall the system. %2: ElementID
ImgPrcExec > %3: Name of caller
function
50030 FFOutImgPrc DDOPush Failed Mask DDO writing common function of writing Restart. If the same phenomenon recurred, %1: GroupID
Processing Data the mask tag returned an error. reinstall the system. %2: ElementID
(0x%1,0x%2) (%3) < %3: Name of caller
ImgPrcExec > function
50030 FFOutImgPrc DOM RemoveTag Failed Tag deletion function of DOM method Restart. If the same phenomenon recurred, %1: Absolute location
TAG=%1 (%2) < returned an error. reinstall the system. path
ImgPrcExec > %2: Name of caller
function
50030 FFOutImgPrc Exception Error (%1) An exception occurred in the long-view Restart. If the same phenomenon recurred, %1: Class and function
<ImgPrcExec> composition process for life size. reinstall the system. name
50030 FFOutImgPrc ExecDecisionCSL Failed CSL processing (for storage) execution Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > determination function returned an error.
reinstall the system. function
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-173
Event Event Source Description Causes Remedies Remarks
50030 FFOutImgPrc ExecDecisionFCR Failed FCR processing execution determination Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > function returned an error. reinstall the system. function
50030 FFOutImgPrc ExecImageCSL%1 Failed CSL processing function returned an error. Restart. If the same phenomenon recurred, %1: Distribution
(%2) < ImgPrcExec > reinstall the system. destination
recognition
("Strg"/"Print")
%2: Image type
character string
50030 FFOutImgPrc ExecImageFCR Failed (%1) FCR processing function returned an error. Restart. If the same phenomenon recurred, %1: Name of caller
< ImgPrcExec > reinstall the system. function
50030 FFOutImgPrc ExecImagePreCSL Failed CSL preprocessing function returned an Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > error. reinstall the system. function
50030 FFOutImgPrc ExecMaskProceToTag Mask image and DDO tag configuration Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec function returned an error. reinstall the system. function
>
50030 FFOutImgPrc FCDMakeMaskImg Failed Mask image generation facade for overlay Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > returned an error. reinstall the system. function
50030 FFOutImgPrc Get Parameter Failed Acquisition function of image processing Restart. If the same phenomenon recurred, %1: absolute location
(Element = [%1] Attribute = common parameters returned an error. reinstall the system. path
[%2]) (%3) <ImgPrcExec> %2: Tag name/
attribution name
%3: Class and function
name
50030 FFOutImgPrc GetInputImageArea Failed Acquisition function of input image pointer
Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > returned an error. reinstall the system. function
50030 FFOutImgPrc GetInputImageData (%1) < Acquisition function of input image returned
Restart. If the same phenomenon recurred, %1: Name of caller
ImgPrcExec > an error. reinstall the system. function
50030 FFOutImgPrc GetOutputProcessing Acquisition function of specifying output
Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec process returned an error. reinstall the system. function
>
50030 FFOutImgPrc GetSrcImgArea Error (%1) Pointer acquiring function of input image Restart. If the same phenomenon recurred, %1: Class and function
<ImgPrcExec> region returned an error. reinstall the system. name
50030 FFOutImgPrc GetVR Error(Gr : 0x%04X Acquired function of DDO tag returned an Restart. If the same phenomenon recurred, %1: GroupID
Er : 0x%04X) (%1) < error. reinstall the system. %2: ElementID
ImgPrcExec > %3: Name of caller
function
50030 FFOutImgPrc GetXMLExecutableResult Function to acquire the result of whether or Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec not to perform in the image processing reinstall the system. function
> common parameters returned an error.
50030 FFOutImgPrc ImageCompJPEGtoRawFile LosslessJPEG compression control Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec > function of image data returned an error. reinstall the system. function
50030 FFOutImgPrc InitializeRawData Failed Initialization function of RAW data returned Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > an error. reinstall the system. function
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-174
Event Event Source Description Causes Remedies Remarks
50030 FFOutImgPrc MakeRotationInverseBit Creating function of rotation/inversion bit Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec character strings returned an error. reinstall the system. function
>
50030 FFOutImgPrc ManualMargin is invalid Calculated result of margin process was in Restart. If the same phenomenon recurred, %1: Name of caller
(%1) Top=(%2) error. reinstall the system. function
Bottom=(%3) Left=(%4) %2: Calculated result
Right=(%5) <ImgPrcExec> margin size (top)
%3: Calculated result
margin size
(bottom)
%4: Calculated result
margin size (left)
%5: Calculated result
margin size (right)
50030 FFOutImgPrc Memory Copy Error (%1) Image processed result copying function to Restart. If the same phenomenon recurred, %1: Class and function
<ImgPrcExec> the memory region returned an error. reinstall the system. name
50030 FFOutImgPrc PermanentImage DDO perpetuation function of image Restart. If the same phenomenon recurred, %1: DDO file path
ParmtoDDO Failed processing parameters returned an error. reinstall the system. name
PathName=%1 (%2) < %2: Name of caller
ImgPrcExec > function
50030 FFOutImgPrc PushNormalizedFlag Failed Modification function of standardized flag Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > (0009,XXF2) returned an error. reinstall the system. function
50030 FFOutImgPrc RemoveDICOMData Failed Deletion common function of DDO tag Restart. If the same phenomenon recurred, %1: GroupID
(0x%1,0x%2) (%3) < returned an error. reinstall the system. %2: ElementID
ImgPrcExec > %3: Name of caller
function
50030 FFOutImgPrc Set Parameter Failed (%1) Saving function of output results for life Restart. If the same phenomenon recurred, %1: Class and function
<ImgPrcExec> size composition returned an error. reinstall the system. name
50030 FFOutImgPrc Set Parameter Failed Configuration function of image processing Restart. If the same phenomenon recurred, %1: Absolute location
(Element = [%1] Attribute = common parameters returned an error. reinstall the system. path
[%2]) (%3) <ImgPrcExec> %2: Tag name/
attribution name
%3: Class and function
name
50030 FFOutImgPrc SetImagePixelofBitStored Configuration function of image attribution Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec parameters in image processing common reinstall the system. function
> parameters returned an error.
50030 FFOutImgPrc SetImagePixelofFileData Output tag deletion function of image Restart. If the same phenomenon recurred, %1: IN/OUT recognition
Failed (%1:%2) < attribution information returned an error. reinstall the system. %2: Name of caller
ImgPrcExec > function
50030 FFOutImgPrc SetImagePixelofRotInverse Rotation/inversion tag configuration Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec function of input file information in image reinstall the system. function
> attribution parameters returned an error.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-175
Event Event Source Description Causes Remedies Remarks
50030 FFOutImgPrc SetValueToXMLTag Failed XML value configuration function of image Restart. If the same phenomenon recurred, %1: Absolute path of
TAG:<%1><%2> (%3) < processing common parameters returned reinstall the system. the tag
ImgPrcExec > an error. %2: Tag name
%3: Name of caller
function
50030 FFOutImgPrc SetXMLParameterCSL XML configuration function of CSL Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec processing parameters returned an error. reinstall the system. function
>
50030 FFOutImgPrc SetXMLPrmPreCSL Failed Image common parameters configuration Restart. If the same phenomenon recurred, %1: Name of caller
(%1) < ImgPrcExec > function of CSL preprocessing returned an reinstall the system. function
error.
50030 FFOutImgPrc SetXMLProcedureCSL XML configuration function of CSL Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec processing execution condition returned an reinstall the system. function
> error.
50030 FFOutImgPrc SetXMLProcedureFCR XML configuration function of FNC Restart. If the same phenomenon recurred, %1: Name of caller
Failed (%1) < ImgPrcExec processing execution condition returned an reinstall the system. function
> error.
50030 FFOutImgPrc SetXMLProcedure Function to set whether or not to execute Restart. If the same phenomenon recurred, %1: Class and function
ScaleProc Failed (%1) the distance correction of the image reinstall the system. name
<ImgPrcExec> processing common parameters returned
an error.
50030 FFOutImgPrc DDT Error ( DDO = [%s] Failed in converting the tag. Perform delivery again. If the same %s: File name
SOPUID = [%s]) < phenomenon recurs, check if there is any %s: SOPClassUID
ImgConvPrivateLocal > incorrect part in the definition file created
by DicomTagEditTool.
50100 FFSpooler DiconDataObject Load The tag value of read density does not Restart. If the same phenomenon recurred, %1: Source file name
failure [GetReadDensity]\ exist. reinstall the system. %2: Source file line
nF=%1 L=%2 number
50100 FFSpooler FFGetMinimumOuterSize Acquisition function of DICOM notation Restart. If the same phenomenon recurred, %1: Source file name
Notation size returned an error. reinstall the system. %2: Source file line
ByPhysicalImageSize Error number
[PhysicalImageSizeTo
SizeCode]\nF=%1 L=%2
50100 FFSpooler FFGetSizeCode Error Function to convert from the DICOM Restart. If the same phenomenon recurred, %1: Source file name
[PhysicalImageSize notation size to the size code returned an reinstall the system. %2: Source file line
ToSizeCode]\nF=%1 L=%2 error. number
50100 FFSpooler MakeSizeCodeList Error List has not been created in the list Restart. If the same phenomenon recurred, %1: Source file name
[GetTrimBySizeCode]\ creation function of size code. reinstall the system. %2: Source file line
nF=%1 L=%2 number
50100 FFSpooler OutputJobFile Failed in acquiring the trimming processed Restart. If the same phenomenon recurred, %1: Source file name
GetItem(TrimSizeCode) image size code. reinstall the system. %2: Source file line
Failed [GetInfo]\nF=%1 number
L=%2 [OutputID=0x%3] %3: OutputID
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-176
Event Event Source Description Causes Remedies Remarks
50100 FFSpooler Trimming Size(0007F272) The tag value of cut out size does not Restart. If the same phenomenon recurred, %1: Source file name
Error exist. reinstall the system. %2: Source file line
[GetTrimBySizeCode]\ number
nF=%1 L=%2
50100 FFSpooler mbstowcs Error \nF=%s Failed to convert the character string. Perform delivery again. If the error is %s: File name
L=%d caused by the system, reinstall the system. %d: Number of lines
50100 FFSpooler VR Error \nF=%s L=%d Failed to reserve the VR data area. Perform delivery again. If the error is %s: File name
caused by the system, reinstall the system. %d: Number of lines
50100 FFSpooler Index Failed tJobInfo. An incorrect ID was given to the film data Perform delivery again. If the error is %s: File name
tpFilmInfo[nFilmNum] \ in the output job data. caused by the system, reinstall the system. %d: Number of lines
nF=%s L=%d %X: Output job ID
[OutputID=0x%X]
50100 FFSpooler Array Index Overflow An incorrect ID was given to the film data Perform delivery again. If the error is %s: File name
tJobInfo. in the output job data. caused by the system, reinstall the system. %d: Number of lines
tpFilmInfo[nFilmNum] \ %X: Output job ID
nF=%s L=%d
[OutputID=0x%X]
50103 FFSpooler MoveNext Failed. [%1] Failed in accessing to the database for Restart the Console and perform delivery %1: Name of function
ErrorCode=%2 output. again. If the error is caused by the system, error occurred
F=%3 L=%4 reinstall the system. %2: Error code
%3: Source file name
%4: Source file line
number
50103 FFSpooler GetOperationRS Failed. Failed in accessing to the database for Restart the Console and perform delivery %1: Name of function
[%1] output. again. If the error is caused by the system, error occurred
ErrorCode=%2 reinstall the system. %2: Error code
F=%3 L=%4 %3: Source file name
%4: Source file line
number
50104 FFSpooler Catch some other exception An exception occurred. Confirm the setting parameters. If the error %1: Error code
[SearchWorkFolder] is caused by the system, reinstall the %2: Source file name
ErrorCode=%1 \nF=%2 system. %3: Source file line
L=%3 number
50104 FFSpooler Catch some other exception An exception occurred. Confirm the setting parameters. If the error %1: Error code
[SelectNot is caused by the system, reinstall the %2: Source file name
ExistFileOnOutImage] system. %3: Source file line
ErrorCode=%1 \nF=%2 number
L=%3
50106 FFSpooler DistributionHost includes \\ Separator remained. Restart. If the same phenomenon recurred, %1: Source file name
is not allowed reinstall the system. %2: Source file line
[GetDDOFileInfo]\nF=%1 number
L=%2 [SpotCode=%2] %3: Spot code
50106 FFSpooler Catch exception Failed in retrieving the DDO file name for Restart DR-ID 300CL.
[CSpoolProcSR] outputting. If the same phenomenon recurred, reinstall
InfFileName is empty DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-177
Event Event Source Description Causes Remedies Remarks
50106 FFSpooler Catch exception Failed in acquiring AE name. Restart DR-ID 300CL.
[CSpoolProcSR] If the same phenomenon recurred, reinstall
AEName is empty DR-ID 300CL.
50106 FFSpooler Catch exception Failed in issuing the spool request Restart DR-ID 300CL.
[CSpoolProcSR] message. If the same phenomenon recurred, reinstall
PostThreadMessage failed. DR-ID 300CL.
50106 FFSpooler Catch badalloc exception Failed in generating the spool request Restart DR-ID 300CL.
[CSpoolProcSR] method (for StructuredReport) parameter. If the same phenomenon recurred, reinstall
Memory allocation error. DR-ID 300CL.
50106 FFSpooler Catch exception An exception occurred. Restart DR-ID 300CL. %1: Exception
[CSpoolProcSR] If the same phenomenon recurred, reinstall message
%1 DR-ID 300CL.
50106 FFSpooler Catch logic_error Failed in retrieving the DDO file information Restart DR-ID 300CL. %1: Error code
exception[tSpoolProcSR] except for the device information. If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 DR-ID 300CL.
SpotCode=%2
DDOInfo Load Failed
[tSpoolProcSR]
50106 FFSpooler Catch some other exception An exception occurred. Restart DR-ID 300CL. %1: Error code
[tSpoolProcSR] If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 DR-ID 300CL.
SpotCode=%1
50106 FFSpooler Catch logic_error exception When adding a new record for Restart DR-ID 300CL. %1: Error code
[tSpoolProcSR] StructuredReport to the OutputDB, failed in If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 retrieving a pointer to the structure for DR-ID 300CL.
SpotCode=%2 DDO object.
DDOInfo is null
50106 FFSpooler Catch logic_error exception When adding a new record for Restart DR-ID 300CL. %1: Error code
[tSpoolProcSR] StructuredReport to the OutputDB, failed in If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 adding a new record to the Request table. DR-ID 300CL.
SpotCode=%2
AddNewRequestRecord
failed.
50106 FFSpooler Catch logic_error exception When adding a new record for Restart DR-ID 300CL. %1: Error code
[tSpoolProcSR] StructuredReport to the OutputDB, failed in If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 adding a new record to the Operation DR-ID 300CL.
SpotCode=%2 table.
AddNewOperationRecord
failed.
50106 FFSpooler Catch logic_error exception When adding a new record for Restart DR-ID 300CL. %1: Error code
[tSpoolProcSR] StructuredReport to the OutputDB, failed in If the same phenomenon recurred, reinstall %2: Spot code
ErrorCode=%1 adding a new record to the OutImage DR-ID 300CL.
SpotCode=%2 table.
AddNewOutImageRecord
failed.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-178
Event Event Source Description Causes Remedies Remarks
50106 FFSpooler Catch runtime exception. When retrieving the name of the setting file Restart DR-ID 300CL. %1: Spot code
[SetReportFileNameToJob] for Report and setting it to a JOB, NULL If the same phenomenon recurred, reinstall
Parameter(s) is null. was set to any one of the parameters (AE DR-ID 300CL.
title, Transfer syntax, Que management
object, or Dispatch pointer of the JOB).
50106 FFSpooler Catch runtime exception. Failed in retrieving the leading element of Restart DR-ID 300CL. %1: reportFileName
[SetReportFileNameToJob] the Report setting file name storage array. If the same phenomenon recurred, reinstall (Name of the
SafeArrayGetElement DR-ID 300CL. Report setting file)
failed. hr=%1
50106 FFSpooler DDO Load failed. Failed in reading the DDO file. Restart DR-ID 300CL. %1: DDO file name
[ExistsOutputHost] If the same phenomenon recurred, reinstall
%1 DR-ID 300CL.
50106 FFSpooler Catch runtime exception Failed in retrieving the Output ID from a Restart DR-ID 300CL. %1: Error code
[AddNewOperationRecord] newly added or already existing record. If the same phenomenon recurred, reinstall %2: Spot code
GetOperationRecordSet DR-ID 300CL.
failed. ErrorCode=%1
SpotCode=%2
50110 FFSpooler BackupOrRestore Failed to copy the DDO file for backup or Restart DR-ID 300CL. %s: File name
RetryDDO Failed [CopyFile] restore. If the same error recurs after restarting, %d: Number of lines
\nF=%s L=%d reinstall DR-ID 300CL. %X: Output job ID
[OutputID=0x%X]
50113 FFSpooler Catch some other exception An exception occurred in ImageStatus Restart. If the same phenomenon recurred, %1: Error code
[tNotifyUpdateImage updating result notification process. reinstall the system. %2: Source file name
StatusResult]\n %3: Source file line
ErrorCode=%1 \nF=%2 number
L=%3]
50113 FFSpooler Catch some other exception An exception occurred in completion Restart. If the same phenomenon recurred, %1: Error code
[tSetCommitProc processing of commitment process. reinstall the system. %2: Source file name
CompleteAfter]\ %3: Source file line
nErrorCode=0x%1 \nF=%2 number
L=%3
50113 FFSpooler Catch some other exception An exception occurred in completion Restart. If the same phenomenon recurred, %1: Error code
[tSetOutputProc processing of output process. reinstall the system. %2: Source file name
CompleteAfter]\ %3: Source file line
nErrorCode=0x%1 \nF=%2 number
L=%3
50113 FFSpooler Not Found Operation No record is found in Operation recordset. Restart. If the same phenomenon recurred, %1: Source file name
Record [tSetOutputProc reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [RequestID=%3] %3: OutputID
50113 FFSpooler Not Found OutImage No record is found in recordset. Restart. If the same phenomenon recurred, %1: Source file name
Record [tSetOutputProc reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [RequestID=%3] %3: RequestID
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-179
Event Event Source Description Causes Remedies Remarks
50113 FFSpooler Not Found Queue Failed in searching the DISPATCH pointer Restart. If the same phenomenon recurred, %1: Source file name
[tSetOutputProc of the output JOB object. reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=%3] %3: OutputID
50113 FFSpooler Not Found Queue Failed in deleting an OutputJob file for Restart. If the same phenomenon recurred, %1: Source file name
[tSetOutputProc output. reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=0x%3] %3: OutputID
50113 FFSpooler Operation Recordset Not No record is found in Operation recordset. Restart. If the same phenomenon recurred, %1: Source file name
Found [tSetOutputProc reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=0x%3] %3: OutputID
50113 FFSpooler TDDOEditData Copy Failed Failed in copying DDO tags. Confirm the setting parameters. If the error %1: Source file name
[CopyDDOFortData] is caused by the system, reinstall the %2: Source file line
F=%1 L=%2 [Group=%3, system. number
Element=%4] %3: DDO tag Group
%4: DDO tag Element
50113 FFSpooler SetMaxFreeTrimming Failed in setting the trimming region to the Confirm that the trimming region is added %1: Source file name
SizeToDDO Error. maximum size in composition original properly to the composition original image. %2: Source file line
[SetMaxFreeTrimming images. If the error is caused by the system, number
SizeToDDO] reinstall the system. %3: Function return
F=%1 L=%2 [lRet=%3] value
50113 FFSpooler LoadFile() Failed. Failed in reading the DDO files of Perform composition and delivery again. If %1: Path of
[GetMaxFreeTrimmingSize] composition original images. the error is caused by the system, reinstall composition original
[DDOFilePath=%1] the system. image DDO
50113 FFSpooler SetFree Failed in setting the trimming region in Perform delivery again. If the error is %1: Source file name
TrimmingSizeToDDO Error. DDO tags. caused by the system, reinstall the system. %2: Source file line
[SetFreeTrimmingSize number
ToDDO] %3: Function return
F=%1 L=%2 [lRet=%3] value
50113 FFSpooler Setting Failed in setting the maximum rate tag Perform delivery again. If the error is
MagRate(0029,xx34) is (0029,xx34). caused by the system, reinstall the system.
failed.
[SetLifesizeMagRate]
50114 FFOutImgPrc ImgFilmCharMake--Invalid Acquisition function of DDO sequence tag Restart. If the same phenomenon recurred, %1: Acquired size
Element Number ( num = size returned an error. reinstall the system.
%2 ) < ImgFilmCharMake >
50115 FFSpooler Not Found Queue Failed in searching the DISPATCH pointer Restart. If the same phenomenon recurred, %1: Source file name
[tSetCommitProc of the output JOB object. reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=%3] %3: OutputID
50115 FFSpooler Not Found Queue Failed in deleting the applicable DISPATCH Restart. If the same phenomenon recurred, %1: Source file name
[tSetCommitProc pointer from the queue list. reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=0x%3] %3: OutputID
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-180
Event Event Source Description Causes Remedies Remarks
50115 FFSpooler Operation Recordset Not No record is found in recordset. Restart. If the same phenomenon recurred, %1: Source file name
Found [tSetCommitProc reinstall the system. %2: Source file line
CompleteAfter]\nF=%1 number
L=%2 [OutputID=0x%3] %3: OutputID
50116 FFSpooler Invalid Parameter OutputID value was in error. Restart. If the same phenomenon recurred, %1: Source file name
[CNotifyUpdateImage reinstall the system. %2: Source file line
StatusResult]\nF=%1 L=%2 number
[OutputID=%2] %3: OutputID
50116 FFSpooler No Memory NULL value was notified to ImageStatus Restart. If the same phenomenon recurred, %1: Source file name
[CNotifyUpdateImage updating result notification message. reinstall the system. %2: Source file line
StatusResult]\nF=%1 L=%2 number
50116 FFSpooler PostThreadMessage Failed PostThreadMessage function returned an Restart. If the same phenomenon recurred, %1: Source file name
[CNotifyUpdateImage error. reinstall the system. %2: Source file line
StatusResult]\nF=%1 L=%2 number
50117 FFOutImgPrc DICOM Data GetVR Error ( Failed in acquiring the DDO tag. Check the setting parameter. If the error is %1: DDO tag Group
Gr : %1 Er : %2 ) < caused by the system, reinstall the system. %2: DDO tag Element
LUTProc >
50117 FFOutImgPrc Area Get Error < LUTProc Failed in securing a memory region for Perform the delivery again. If the error is
> LUT. caused by the system, reinstall the system.
50117 FFOutImgPrc Catch some other An exception has occurred. Perform the delivery again. If the error is %1: Error code
Exception ErrCode = %1 caused by the system, reinstall the system.
<LUTProc>
50117 FFOutImgPrc ModLUTGamma Error Failed in securing a memory region for Perform the delivery again. If the error is
<LUTProc> tone tables. caused by the system, reinstall the system.
50123 FFSpooler Catch some other exception An exception occurred in output completion Restart. If the same phenomenon recurred, %1: Error code
[tMainErrQueClearAfter]\ waiting queue execution process. reinstall the system. %2: Source file name
nErrorCode=%1 \nF=%2 %3: Source file line
L=%3 number
50124 FFSpooler SetNetDevInfo Failed FFNetDBGetMaskProcessing function Restart. If the same phenomenon recurred, %1: Source file name
[FFNetDBGetMask returned an error. reinstall the system. %2: Source file line
Processing]\n number
ErrorCode=0x%1 \nF=%2
L=%3
50300 FFDicomSC CoInitialize Error Failed in CoInitialize. Restart DR-ID 300CL. %s: File name
[tThreadProc] \nF=%s If the same error recurs after restarting, %d: Number of lines
L=%d reinstall DR-ID 300CL.
50301 FFDicomSC \nF=%s L=%d LParam is The parameters have not been Check the setting parameter. If the error is %s: File name
null. determined. caused by the system, reinstall the DR-ID %d: Number of lines
300CL.
50302 FFDicomSC \nF=%s L=%d LParam is The parameters have not been Check the setting parameter. If the error is %s: File name
null. determined. caused by the system, reinstall the DR-ID %d: Number of lines
300CL.
50303 FFDicomSC \nF=%s L=%d LParam is The parameters have not been Check the setting parameter. If the error is %s: File name
null. determined. caused by the system, reinstall the DR-ID %d: Number of lines
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-181
Event Event Source Description Causes Remedies Remarks
50600 [ME50600] Failed in Failed in initialization of media control. Restart the system.
initialization of media If the same error recurs after restarting the
control. system, reinstall the system.
50601 [ME50601] Failed in media Failed in media mounting process. Restart the system.
mounting process. If the same error recurs after restarting the
system, reinstall the system.
50602 [ME50602] Failed in Failed in notifying the conditions of a Restart the system.
notifying the conditions of a medium. If the same error recurs after restarting the
medium. system, reinstall the system.
50603 [ME50603] Failed in Failed in requesting to store images in a Restart the system.
requesting to store images medium. If the same error recurs after restarting the
in a medium. system, reinstall the system.
50610 [ME50610] Failed in Failed in creating a folder. Restart the system.
creating a folder. If the same error recurs after restarting the
system, reinstall the system.
50611 [ME50611] Failed in The specified drive could not be accessed. Select a connection destination for media
changing the connection again.
destination for media as the
specified drive could not be
accessed.
50612 [ME50612] Failed in Failed in acquiring a study from a medium. Restart the system.
acquiring a study from a If the same error recurs after restarting the
medium. system, reinstall the system.
50613 [ME50613] Failed in Failed in initialization of the media list. Restart the system.
initialization of the media If the same error recurs after restarting the
list. system, reinstall the system.
50614 [ME50614] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
50615 [ME50615] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
50616 [ME50616] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
50617 [ME50617] Failed in Failed in accessing the database. Restart the system.
accessing the database. If the same error recurs after restarting the
system, reinstall the system.
50618 [ME50618] Failed in Failed in checking images. Restart the system.
checking images. If the same error recurs after restarting the
system, reinstall the system.
50619 [ME50619] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-182
Event Event Source Description Causes Remedies Remarks
50620 [ME50620] Failed in Failed in updating the study conditions/ Restart the system.
updating the study image conditions. If the same error recurs after restarting the
conditions/image system, reinstall the system.
conditions.
50621 [ME50621] Failed in Failed in customizing a media list. Restart the system.
customizing a media list. If the same error recurs after restarting the
system, reinstall the system.
50622 [ME50622] Failed in setting Failed in setting a distribution code Restart the system.
a distribution code If the same error recurs after restarting the
system, reinstall the system.
50623 [ME50623] Failed in media Failed in media mounting process. Change the connection destination for
mounting process. The media and try again.
mounting process will be
performed in the originally
connected destination.
50624 [ME50624] Failed in media Failed in media mounting process. Change the connection destination for
mounting process. media and try again.
50625 [ME50625] Failed in Failed in changing the connection Restart the system.
changing the connection destination for media. If the same error recurs after restarting the
destination for media. system, reinstall the system.
50626 [ME50626] Failed in media Failed in media unmounting process. Restart the system.
unmounting process. If the same error recurs after restarting the
system, reinstall the system.
50627 [ME50627] Failed in Failed in information acquisition processing Restart the system.
information acquisition for mounting a medium. If the same error recurs after restarting the
processing for mounting a system, reinstall the system.
medium.
50628 [ME50628] Failed in Failed in acquiring a study from a medium. Restart the system.
acquiring a study from a If the same error recurs after restarting the
medium. system, reinstall the system.
50629 [ME50629] Failed in Failed in delivery processing. Wait for a while and try again.
delivery processing.
50630 [ME50630] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
50631 [ME50631] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
50632 [ME50632] Failed in Failed in acquiring a host name. Restart the system.
acquiring a host name. If the same error recurs after restarting the
system, reinstall the system.
50633 [ME50633] Failed in Failed in acquiring device information. Restart the system.
acquiring device If the same error recurs after restarting the
information. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-183
Event Event Source Description Causes Remedies Remarks
50634 [ME50634] Failed in Failed in acquiring screen information. Restart the system.
acquiring screen If the same error recurs after restarting the
information. system, reinstall the system.
50635 [ME50635] Failed in page Failed in page forward processing. Restart the system.
forward processing. If the same error recurs after restarting the
system, reinstall the system.
50636 [ME50636] Failed in page Failed in page backward processing. Restart the system.
backward processing. If the same error recurs after restarting the
system, reinstall the system.
50637 [ME50637] Failed in Failed in accessing the data base. Restart the system.
accessing the data base. If the same error recurs after restarting the
system, reinstall the system.
50638 [ME50638] Failed in Failed in obtaining a list of distribution Restart the system.
obtaining a list of codes. If the same error recurs after restarting the
distribution codes. system, reinstall the system.
50639 [ME50639] Failed in Failed in mounting media. Restart the system.
mounting media. If the same error recurs after restarting the
system, reinstall the system.
50640 [ME50640] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
50641 [ME50641] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
50642 [ME50642] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
50643 [ME50643] Failed in Failed in acquiring free space information. Restart the system.
acquiring free space If the same error recurs after restarting the
information. system, reinstall the system.
50644 [ME50644] Failed in Failed in acquiring drive information. Restart the system.
acquiring drive information. If the same error recurs after restarting the
system, reinstall the system.
50645 [ME50645] Failed in Failed in displaying process of a storage Restart the system.
displaying process of a folder selecting dialog. If the same error recurs after restarting the
storage folder selecting system, reinstall the system.
dialog.
50646 [ME50646] Failed in Failed in displaying a media list customize Restart the system.
displaying a media list dialog. If the same error recurs after restarting the
customize dialog. system, reinstall the system.
50647 [ME50647] Failed in Failed in displaying a progress dialog. Restart the system.
displaying a progress If the same error recurs after restarting the
dialog. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-184
Event Event Source Description Causes Remedies Remarks
50648 [ME50648] Failed in Failed in displaying a distribution code Restart the system.
displaying a distribution selecting dialog. If the same error recurs after restarting the
code selecting dialog. system, reinstall the system.
50649 [ME50649] Failed in Failed in displaying a medium selecting Restart the system.
displaying a medium dialog. If the same error recurs after restarting the
selecting dialog. system, reinstall the system.
50650 [ME50650] Failed in Failed in checking the available characters. Restart the system.
checking the available If the same error recurs after restarting the
characters. system, reinstall the system.
50651 [ME50651] Failed in device Failed in device auto recognition Restart the system.
auto recognition processing. If the same error recurs after restarting the
processing. system, reinstall the system.
50652 [ME50652] Failed in writing Failed in writing registry information. Restart the system.
registry information. If the same error recurs after restarting the
system, reinstall the system.
50653 [ME50653] Failed in Failed in changing the display position. Restart the system.
changing the display If the same error recurs after restarting the
position. system, reinstall the system.
50654 [ME50654] Failed in writing The mounted media was removed, and the Remove the media after unmount it.
on a medium. process to write into the media failed. Reconnect the media and try the study
output again.
50670 [ME50670] Failed in Failed in checking images. Restart the system.
checking images. If the same error recurs after restarting the
system, reinstall the system.
50671 [ME50671] Failed in Failed in acquiring image information. Restart the system.
acquiring image If the same error recurs after restarting the
information. system, reinstall the system.
50672 [ME50672] Image Image information does not exist. Restart the system.
information does not exist. If the same error recurs after restarting the
system, reinstall the system.
50673 [ME50673] There is no There is no study that can be created. Change the target study and try again.
study that can be created.
50674 [ME50674] Failed in Failed in acquiring screen information. Restart the system.
acquiring screen If the same error recurs after restarting the
information. system, reinstall the system.
50675 [ME50675] There is no data There is no data to write. Change the target study and try again.
to write.
50676 [ME50676] An error An error occurred while verifying. Restart the system.
occurred while verifying. If the same error recurs after restarting the
system, reinstall the system.
50677 [ME50677] An error An error occurred while comparing. Restart the system.
occurred while comparing. If the same error recurs after restarting the
system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-185
Event Event Source Description Causes Remedies Remarks
50678 [ME50678] An IO error An IO error occurred against the drive. Restart the system.
occurred against the drive. If the same error recurs after restarting the
system, reinstall the system.
50679 [ME50679] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
50681 [ME50681] Failed in Failed in checking free space. Restart the system.
checking free space. If the same error recurs after restarting the
system, reinstall the system.
50682 [ME50682] Failed in Failed in selecting a storage folder. Restart the system.
selecting a storage folder. If the same error recurs after restarting the
system, reinstall the system.
50684 [ME50684] Failed in writing Failed in writing a medium. Restart the system.
a medium. If the same error recurs after restarting the
system, reinstall the system.
50685 [ME50685] Failed in image Failed in image processing. Restart the system.
processing. If the same error recurs after restarting the
system, reinstall the system.
50686 [ME50686] Failed in Failed in generating a DICOM file. Restart the system.
generating a DICOM file. If the same error recurs after restarting the
system, reinstall the system.
50687 [ME50687] Failed in Failed in generating a media image. Restart the system.
generating a media image. If the same error recurs after restarting the
system, reinstall the system.
50688 [ME50688] A program error A program error has been detected. Restart the system.
has been detected. If the same error recurs after restarting the
system, reinstall the system.
50689 [ME50689] Failed in Failed in acquiring DICOM tag information. Restart the system.
acquiring DICOM tag If the same error recurs after restarting the
information. system, reinstall the system.
50690 [ME50690] Failed in Failed in checking available characters. Restart the system.
checking available If the same error recurs after restarting the
characters. system, reinstall the system.
50691 [ME50691] Failed in Failed in acquiring a host name. Restart the system.
acquiring a host name. If the same error recurs after restarting the
system, reinstall the system.
50692 [ME50692] Failed in Failed in accessing a definition file. Restart the system.
accessing a definition file. If the same error recurs after restarting the
system, reinstall the system.
50693 [ME50693] Failed in Failed in acquiring device information. Restart the system.
acquiring device If the same error recurs after restarting the
information. system, reinstall the system.
51300 [MQ51300] Measurement Measurement processing cannot be Check that the images are entered.
processing cannot be implemented.
implemented.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-186
Event Event Source Description Causes Remedies Remarks
51301 [MQ51301] The hard disk is The hard disk is full. Increase the free hard disk space and then
full. retry operation.
51303 [MQ51303] Measurement Measurement processing cannot be Use only the images from the same reader
processing cannot be implemented.
implemented.
51304 [MQ51304] Measurement Measurement processing cannot be Check that three images have been
processing cannot be implemented. entered.
implemented.
51305 [MQ51305] Measurement Measurement processing cannot be Check the size of the 1shot Phantom
processing cannot be implemented. image.
implemented.
51306 [MQ51306] Input values Input values are illegal. Input a correct value.
are illegal.
51307 [MQ51307] Input values Input values are illegal. Input a correct value.
are illegal.
51308 [MQ51308] Input values Input values are illegal. Input a correct value.
are illegal.
51309 [MQ51309] Input values Input values are illegal. Input a correct value.
are illegal.
51310 [MQ51310] Image data Image data storage processing cannot be When save processing is performed, check
storage processing cannot implemented. that images have been entered.
be implemented.
51311 [MQ51311] Measurement Measurement processing cannot be Reset the phantom position to a correct
processing cannot be implemented. position and expose again.
implemented.
51312 [MQ51312] Failed in image Failed in image output. Select an image that can be output.
output.
51313 [MQ51313] Input values Input values are illegal. Input a correct value.
are illegal.
51314 [MQ51314] Input values Input values are illegal. Input a correct value.
are illegal.
51315 [MQ51315] Input values Input values are illegal. Input a correct value.
are illegal.
51316 [MQ51316] Input values Input values are illegal. Input a correct value.
are illegal.
51317 [MQ51317] Input values Input values are illegal. Input a correct value.
are illegal.
51318 [MQ51318] Input values Input values are illegal. Input a correct value.
are illegal.
51319 [MQ51319] Input values Input values are illegal. Input a correct value.
are illegal.
51320 [MQ51320] Failed in Failed in accessing image files. Click the close button on the QC execution
accessing image files. window.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-187
Event Event Source Description Causes Remedies Remarks
51321 [MQ51321] Failed in Failed in accessing the database. Click the close button on the QC execution
accessing the database. [ReaderQC] window.
[ReaderQC]
51322 [MQ51322] Failed in data Failed in data storage processing. Click the close button on the QC execution
storage processing. window.
51323 [MQ51323] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall DR-ID
300CL.
51324 [MQ51324] Measurement Measurement processing cannot be Use only images with a same reading
processing cannot be implemented. density.
implemented.
51325 [MQ51325] The image is The image is illegal. Reset the phantom position to a correct
illegal. position and expose again.
51326 [MQ51326] Measurement Measurement processing cannot be Confirm if the specified reader has been
processing cannot be implemented. registered.
implemented.
51327 [MQ51327] Image with Image with unsupported reading density is Select an image with the designated
unsupported reading included. reading density.
density is included.
51328 [MQ51328] Images from Images from more than one Image Reader Acquire images from the same Image
more than one Image exist. Reader.
Reader exist.
51329 [MQ51329] Exposure Exposure submenu not for QC tests is Select the specified exposure menu for
submenu not for QC tests included. QC.
is included.
51330 [MQ51330] Image with the Image with the unsupported size for QC Expose with the specified image size.
unsupported size for QC test is included.
test is included.
51331 [MQ51331] Cannot obtain Cannot obtain data on the exposed Images are read in for all displayed menus.
data on the exposed images.
images.
51332 [MQ51332] Failed to Failed to initialize the system. Restart DR-ID 300CL. If the same
initialize the system. phenomenon recurred, reinstall DR-ID
300CL.
51333 [MQ51333] Failed to Failed to access the definition file. Restart DR-ID 300CL. If the same
access the definition file. phenomenon recurred, reinstall DR-ID
300CL.
51334 [MQ51334] Failed to Failed to access the registry information. Restart DR-ID 300CL. If the same
access the registry phenomenon recurred, reinstall DR-ID
information. 300CL.
51335 [MQ51335] Failed to Failed to access the database.[%s] Restart DR-ID 300CL. If the same
access the database.[%s] phenomenon recurred, reinstall DR-ID
300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-188
Event Event Source Description Causes Remedies Remarks
51337 [MQ51337] Failed to obtain Failed to obtain the information on the Restart DR-ID 300CL. If the same
the information on the Image Reader. phenomenon recurred, reinstall DR-ID
Image Reader. 300CL.
51338 [MQ51338] A required A required image is missing. Expose the images if there are not
image is missing. exposed images.
51339 [MQ51339] An error An error occurred when calling the QC Restart DR-ID 300CL. If the same
occurred when calling the function. phenomenon recurred, reinstall DR-ID
QC function. 300CL.
51340 [MQ51340] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall DR-ID
300CL.
51341 [MQ51341] Multiple Multiple exposure menus are included. Select only 1 specified study menu.
exposure menus are
included.
51342 [MQ51342] An exposure An exposure menu not for QC test is Select only 1 specified study menu.
menu not for QC test is included.
included.
51343 [MQ51343] The exposure The exposure menu includes an improper Register the specified exposure menu with
menu includes an improper submenu. the study menu.
submenu.
51344 [MQ51344] Proper Proper exposure submenus are not listed Register the specified exposure menu with
exposure submenus are (too many). the study menu.
not listed (too many).
51345 [MQ51345] Proper Proper exposure submenus are not listed Register the specified exposure menu with
exposure submenus are (too few). the study menu.
not listed (too few).
51346 [MQ51346] No exposure No exposure submenu information is Select the specified study menu and
submenu information is included. perform QC.
included.
51347 [MQ51347] Same IP is Same IP is used multiple times. Use multiple different IP.
used multiple times.
51348 [MQ51348] Incorrect Incorrect exposure menu is selected. Select the same menu or end the study
exposure menu is selected. and start again from the beginning.
51349 [MQ51349] QC test cannot QC test cannot be performed. Restart DR-ID 300CL. If the same
be performed. phenomenon recurred, reinstall DR-ID
300CL.
51350 [MQ51350] Failed to Failed to access the database. Click the close button on the QC execution
access the database. [MammoQC] window.
[MammoQC]
51351 [MQ51351] Failed to Failed to access the database. [QCItems] Click the close button on the QC execution
access the database. window.
[QCItems]
51352 [MQ51352] Input value is Input value is incorrect. Input a correct value.
incorrect.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-189
Event Event Source Description Causes Remedies Remarks
51353 [MQ51353] Input value is Input value is incorrect. Check the exposure conditions and expose
incorrect. again/calculate again.
51354 [MQ51354] Input value is Input value is incorrect. Check the phantom position and expose
incorrect. again/calculate again.
51355 [MQ51355] Some items Some items have not yet been input. Check and input the exposure conditions
have not yet been input. (%s).
51356 [MQ51356] Some items Some items have not yet been input. Enter %s of the %s tab.
have not yet been input.
51357 [MQ51357] Cannot perform Cannot perform measurement. Check the exposure conditions and expose
measurement. again/calculate again.
51358 [MQ51358] Image is Image is incorrect. Check the phantom position and the
incorrect. exposure conditions and expose again/
calculate again.
51359 [MQ51359] Failed in Failed in obtaining data from the image Restart DR-ID 300CL. If the same
obtaining data from the information file. phenomenon recurred, reinstall DR-ID
image information file. 300CL.
51360 [MQ51360] Failed to Failed to display the image. Restart DR-ID 300CL. If the same
display the image. phenomenon recurred, reinstall DR-ID
300CL.
51361 [MQ51361] Max. (min.) Max. (min.) criteria value for %s is Check the baseline value or the criteria
criteria value for %s is incorrect. and set again.
incorrect.
51362 [MQ51362] The input value The value set to the initial setting is out of Set the value to the initial setting in range.
is invalid. range.
51363 [MQ51363] Fail/Pass is not Fail/Pass is not specified. Specify the Fail/Pass.
specified.
51364 [MQ51364] Input value is Input value is incorrect. Input a correct value.
incorrect.
51365 [MQ51365] Input value is Input value is incorrect. Input a correct value.
incorrect.
51366 [MQ51366] Input value is Input value is incorrect. Input a correct value.
incorrect.
51367 [MQ51367] Input value is Input value is incorrect. Input a correct value.
incorrect.
51368 [MQ51368] Some items Some items have not yet been input. Enter the user input items.
have not yet been input.
51369 [MQ51369] Some items Some items have not yet been input. Enter the %s.
have not yet been input.
51370 [MQ51370] Input value is Input value is incorrect. Input a correct value.
incorrect.
51371 [MQ51371] Cannot perform Cannot perform measurement. Check the image size of the input image.
measurement.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-190
Event Event Source Description Causes Remedies Remarks
51372 [MQ51372] Input value is Input value is incorrect. Input a correct value.
incorrect.
51373 [MQ51373] Input value is Input value is incorrect. Input a correct value.
incorrect.
51374 [MQ51374] Input value is Input value is incorrect. Input a correct value.
incorrect.
51375 [MQ51375] Input value is Input value is incorrect. Input a correct value.
incorrect.
51376 [MQ51376] Failed to obtain Failed to obtain a baseline value. Restart DR-ID 300CL. If the same
a baseline value. phenomenon recurred, reinstall DR-ID
300CL.
51377 [MQ51377] Cannot save Cannot save the test result. Check the user input contents and perform
the test result. automatic calculation.
51378 [MQ51378] Cannot save Cannot save the test result. Check the user input contents.
the test result.
51379 [MQ51379] Cannot save Cannot save the test result. Perform automatic calculation.
the test result.
51380 [MQ51380] Max. criteria Max. criteria value for %s is incorrect. Check the baseline value or the criteria
value for %s is incorrect. and set again.
51381 [MQ51381] Min. criteria Min. criteria value for %s is incorrect. Check the baseline value or the criteria
value for %s is incorrect. and set again.
51382 [MQ51382] Cannot perform Cannot perform measurement. Check the input number.
measurement.
51383 [MQ51383] Cannot perform Cannot perform measurement. Set the CNR baseline value correctly.
measurement.
51384 [MQ51384] Cannot obtain Cannot obtain data of the images captured Images are read in for all displayed menus.
data of the images captured by exposure.
by exposure.
51385 [MQ51385] Input value is Input value is incorrect. Input a correct value.
incorrect.
51400 [MQ51400] Failed to Failed to access the database. [%s] Click the close button on the QC execution
access the database. [%s] window.
51401 [MQ51401] IP with the IP with the same number is used multiple Reread the image so there is no duplicate
same number is used times. IP number.
multiple times.
51402 [MQ51402] Value of the Value of the registry information is Restart DR-ID 300CL. If the same
registry information is abnormal. phenomenon recurred, reinstall DR-ID
abnormal. 300CL.
51403 [MQ51403] Failed in Failed in deletion processing. Delete again.
deletion processing.
51404 [MQ51404] X-ray device X-ray device name is invalid. Change the X-ray device name in the
name is invalid. system settings.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-191
Event Event Source Description Causes Remedies Remarks
51405 [MQ51405] Conditions for Conditions for QC implementation are not Check the QC implementation conditions.
QC implementation are not satisfied.
satisfied.
51406 [MQ51406] If the number of If the number of menus exceeds 32, Check the QC menu number.
menus exceeds 32, Mammo QC cannot be executed.
Mammo QC cannot be
executed.
51600 [ME51600] Failed in Failed in mounting the media. Ensure that %s media has been set and
mounting the media. connected to a network.
51601 [ME51601] The free space The free space of the media is running out. Replace the media or specify another
of the media is running out. folder.
51610 [ME51610] The specified The specified folder does not exist. Select a folder again.
folder does not exist.
51615 [ME51615] Insert a disk to The disk is not inserted to the drive. Insert a disk to the drive. The mounting
the drive. The mounting process will be performed in the originally
process will be performed connected destination.
in the originally connected
destination.
51616 [ME51616] A delivery A delivery destination does not exist in the Select a distribution code again.
destination does not exist specified distribution code.
in the specified distribution
code.
51617 [ME51617] The maximum The maximum number of input characters Up to 16 characters can be entered for a
number of input characters has been exceeded for the volume label. volume label.
has been exceeded for the
volume label.
51618 [ME51618] Input of volume Input of volume label is invalid. Enter correctly.
label is invalid.
51672 [ME51672] No media is No media is inserted. Insert media.
inserted.
51673 [ME51673] The inserted The inserted medium is not blank. Insert a blank medium or specify an
medium is not blank. unused folder.
51674 [ME51674] Disk space is Disk space is insufficient. Replace the disk with high-capacity media.
insufficient.
51675 [ME51675] The maximum The maximum number of input characters Up to 4 characters can be entered for a file
number of input characters has been exceeded for a file name. name.
has been exceeded for a
file name.
51676 [ME51676] Input of file Input of file name is invalid. Enter correctly.
name is invalid.
51678 [ME51678] The destination The destination folder for storage is not Select an empty folder.
folder for storage is not empty.
empty.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-192
Event Event Source Description Causes Remedies Remarks
52100 FFSpooler FFGetSizeCode Failed. Conversion function to the size code Restart. If the same phenomenon recurred, %1: Source file name
[MakePermitted returned an error. reinstall the system. %2: Source file line
SizeCodeList]\nF=%1 number
L=%2 Gr=%3, El=%4 %3: Size code
%4: DDU group ID
%5: DDU element ID
52100 FFSpooler Operation Table Invalid Dat No record is found in recordset. Restart. If the same phenomenon recurred, %1: Source file name
a[tNotifyUpdateImageStatu reinstall the system. %2: Source file line
sResult]\nF=%1 L=%2 number
RequestID=%2 %3: RequestID
52103 FFQCMsgBoxCtrl SetCaption An error has occurred at the time of Check that the file “Fujifilm\IIP\Param\
GetStrDBString2: {0} database reading/writing. MammoQC.mdb” exists at the database
and restart DR-ID 300CL.
52105 [MQ52105] Failed to Failed to access the definition file. Restart the system.
access the definition file (ID If the same error recurs after restarting the
change required). system, reinstall the system.
52106 FFSpooler remove DDOfile Failed Failed in deleting the DDO file while Restart DR-ID 300CL. %1: ImageID
[tSpoolProcSR] canceling the SR output request due to the If the same phenomenon recurred, reinstall
[ImageID=%1] failure of the device information retrieval. DR-ID 300CL.
52106 [MQ52106] Failed to Failed to access the database. Restart DR-ID 300CL.
access the database. (ID If the same phenomenon recurred, reinstall
change required) DR-ID 300CL.
[MammoQC]
52107 [MQ52107] Image with Image with unsupported reading density is Select a suitable menu and expose.
unsupported reading included.
density is included (ID
change required).
52110 FFQCMsgBoxCtrl cmmOK_Click An error has occurred at the time of Check that the file “Fujifilm\IIP\Param\
Rollback {0} : {1} database reading/writing. MammoQC.mdb” exists at the database
and restart DR-ID 300CL.
52110 FFQCMsgBoxCtrl InitValue An error has occurred at the time of Check that the file “Fujifilm\IIP\Param\
GetXRayDevInfo({0}, m_ database reading/writing. MammoQC.mdb” exists at the database
tXRayUnit): lRet = {1} and restart DR-ID 300CL.
52113 FFSpooler Array Index Overflow An incorrect ID was given to the film data Perform delivery again. If the error is %s: File name
tJobInfo. in the output job data. caused by the system, reinstall the system. %d: Number of lines
tpFilmInfo[nFilmNum] \
nF=%s L=%d
53600 [ME53600] Required hard Required hard disk space is insufficient. Check the free space of the hard disk.
disk space is insufficient.
53601 [ME53601] Disk space is Disk space is insufficient. Check the free space of the disk.
insufficient.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-193
Event Event Source Description Causes Remedies Remarks
53611 [ME53611] Failed in Failed in obtaining a study from the media. Check the image processing parameters. If
obtaining a study from the The study contains an image with invalid the error is caused by the system, reinstall
media. image processing conditions. the DR-ID 300CL.
The study contains an
image with invalid image
processing conditions.
54207 [MQ54207] The period The period exceeds the allowable range. Input the date again.
exceeds the allowable
range.
54208 [MQ54208] Out of disk Out of disk space. Change the save destination and save
space. again.
54209 [MQ54209] Cannot display Cannot display the detailed graph for the Use full display mode.
the detailed graph for the selected item.
selected item.
54210 [MQ54210] Cannot display Cannot display the exposure conditions. Check the existence of exposure
the exposure conditions. conditions.
54215 [MQ54215] Failed in writing Failed in writing a file. Check the output media status.
a file.
54308 [MQ54308] Some items Some items have not yet been input. Check the input item.
have not yet been input.
54309 [MQ54309] The study result The study result cannot be stored. Input the study result.
cannot be stored.
54310 [MQ54310] Some items Some items have not yet been input. Enter the user input items.
have not yet been input.
54312 [MQ54312] Image Image sharpness is illegal. The phantom is inclined against IP. Set the
sharpness is illegal. phantom correctly and expose again.
54327 [MQ54327] Some items Some items have not yet been input. Enter the %s.
have not yet been input.
54332 [MQ54332] Value outside Value outside the specified range is input. Enter a value greater than %s.
the specified range is input.
54333 [MQ54333] Max. criteria Max. criteria value for %s is incorrect. Check the baseline value or the criteria
value for %s is incorrect. and set again.
54334 [MQ54334] Min criteria Min criteria value for %s is incorrect. Check the baseline value or the criteria
value for %s is incorrect. and set again.
54335 [MQ54335] The latest test The latest test result is not saved. Store the study result.
result is not saved.
54337 [MQ54337] No baseline No baseline value is input. Create a new baseline values.
value is input.
54338 [MQ54338] Delivery setting Delivery setting was not made on the Perform output again after implementing
was not made on the Console. the delivery setting with DR-ID 300CL.
Console.
54339 [MQ54339] The boxes for The boxes for targets and filters are not Apply at least one check mark.
targets and filters are not checked.
checked.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-194
Event Event Source Description Causes Remedies Remarks
54403 [MQ54403] Failed to Failed to access the database. Restart DR-ID 300CL. If the same
access the database. [%s] phenomenon recurred, reinstall DR-ID
300CL.
54613 [ME54613] The media The media management file is for read Confirm contents of the media.
management file is for read only.
only.
54616 [ME54616] This media is This media is not allowed by this system to Confirm contents of the system setup
not allowed by this system write data. database.
to write data.
54617 [ME54617] This is not an This is not an operating media. Confirm contents of the media.
operating media.
54625 [ME54625] Power Power interruption was detected. Shut down the system immediately.
interruption was detected.
54626 [ME54626] No media No media management files exist. Check the contents of the media.
management files exist.
54627 [ME54627] The media The media management file is for read Confirm contents of the media.
management file is for read only or write-protected.
only or write-protected.
54628 [ME54628] The file cannot The file cannot be handled because it is Confirm contents of the media.
be handled because it is unformatted or illegally formatted.
unformatted or illegally
formatted.
54640 SKBLauncher Not Found Bmp folder There is no BMP folder. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Not Found SKBOffU.bmp There is no BMP file. Restart DR-ID 300CL.
or SKBOffD.bmp If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Not Found SKBOnU.bmp There is no BMP file. Restart DR-ID 300CL.
SKBOnD.bmp If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Not Found SysConfig.mdb There is no SysConfig.mdb. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Faild DB Connect Failed in connecting to database. Restart the PC. If the same phenomenon
recurred, reinstall DR-ID 300CL.
54640 SKBLauncher Not Found KEY There is no received KEY string in Restart DR-ID 300CL.
SysConfig. If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Not Found USB Drive Cannot find USB drive. Confirm whether the USB is available for
DR-ID 300CL and whether the settings for
MediaStorage is correct.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-195
Event Event Source Description Causes Remedies Remarks
54640 SKBLauncher BootEXE() An exception occurred. Restart DR-ID 300CL.
Exception error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher IsUnLockUSB() An exception occurred. Restart DR-ID 300CL.
Exception error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher DeleteAllXMLFile() An exception occurred. Restart DR-ID 300CL.
Exception error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher ChangeDrive() An exception occurred. Restart DR-ID 300CL.
Exception error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Not Found NetConfig.mdb There is no NetConfig.mdb. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Faild DB Connect Failed in connecting to database. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Faild FFGetHostname Failed in FFGetHostname. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Abnormal Termination Get Failed in acquiring the registry “HKEY_ Restart DR-ID 300CL.
media registory LOCAL_MACHINE\SOFTWARE\FujiFilm\ If the same phenomenon recurred, reinstall
IIP\Media\Drive”. DR-ID 300CL.
54640 SKBLauncher Failed GetRecordCount Failed in acquiring records from MediaDB. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher Failed Connect MediaDB Failed in connecting to MediaDB. Restart the PC. If the same phenomenon
recurred, reinstall DR-ID 300CL.
54640 SKBLauncher Not Found Bitmap Path There is no BMP folder. Restart DR-ID 300CL.
If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher SKBOnButton Bitmap Set There is no BMP file. Restart DR-ID 300CL.
Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher SKBOffButton Bitmap Set There is no BMP file. Restart DR-ID 300CL.
Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher USBPwdButton Bitmap Set There is no BMP file. Restart DR-ID 300CL.
Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher SKBOpenButton Bitmap There is no BMP file. Restart DR-ID 300CL.
Set Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-196
Event Event Source Description Causes Remedies Remarks
54640 SKBLauncher SKBRisUSBPwd Bitmap There is no BMP file. Restart DR-ID 300CL.
Set Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher SKBHotPlug Bitmap Set There is no BMP file. Restart DR-ID 300CL.
Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
54640 SKBLauncher SKBHotPlug Bitmap Set There is no BMP file. Restart DR-ID 300CL.
Error If the same phenomenon recurred, reinstall
DR-ID 300CL.
59198 FFQCMsgBoxCtrl DispQC An error has occurred at the time of DDO Check that the DDO file exists and restart
m_clsQCDicomCommon. files reading/writing. DR-ID 300CL.
UpdateFolderPath(inputPat
h,nDicomArray)
59198 FFQCMsgBoxCtrl JudgeImgInfoObj An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file and restart DR-ID 300CL.
GetImagePixelSpacing(m_
QCDataInfo({0}).
nDicomObj,
strImageDensity): lRet={1}
59198 FFQCMsgBoxCtrl JudgeImgInfoObj An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file and restart DR-ID 300CL.
GetImagePixel(m_
QCDataInfo({0}).
nDicomObj, nImagePixel):
lRet={1}
59198 FFQCMsgBoxCtrl JudgeImgInfoObj An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file and restart DR-ID 300CL.
GetImageFileUrl(m_
QCDataInfo({0}).
nDicomObj,
strImageFileUrl): lRet={1}
59198 FFQCMsgBoxCtrl JudgeImgInfoObj An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file and restart DR-ID 300CL.
GetThmImageFileUrl(m_
QCDataInfo({0}).
nDicomObj,
strImageFileUrl): lRet=={1}
59198 FFQCMsgBoxCtrl JudgeImgInfoObj An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file and restart DR-ID 300CL.
GetStationName(m_
QCDataInfo({0}).
nDicomObj,
strStationName): lRet={1}
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-197
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl JudgeMenuName An error has occurred at the time of DDO Check the appropriateness of the data of
Correspondence files reading. the DDO file and restart DR-ID 300CL.
m_clsQCDicom.GetExam
MenuNameSbcs(m_
QCDataInfo({0}).
nDicomObj,
strExamination({1})):
lRet={2}
59198 FFQCMsgBoxCtrl JudgeMenuName An error has occurred at the time of DDO Check the appropriateness of the data of
Correspondence files reading. the DDO file and restart DR-ID 300CL.
m_clsQCDicom.
GetExposureMenuName
Sbcs(m_QCDataInfo({0}).
nDicomObj,
strExposure({1})): lRet={2}
59198 FFQCMsgBoxCtrl JudgeMenuName An error has occurred at the time of DDO Check the appropriateness of the data of
Correspondence files reading. the DDO file and restart DR-ID 300CL.
m_clsQCDicom.
GetPlateID(m_
QCDataInfo({0}).
nDicomObj, strPlateID({1})):
lRet={2}
59198 FFQCMsgBoxCtrl InitQcComponent An error has occurred at the time of DB Check that the file “Fujifilm\IIP\Param\
GetXRayDevInfo({0}, m_ reading/writing. MammoQC.mdb” exists at the database
tXrayUnit): lRet={1} and restart DR-ID 300CL.
59198 FFQCMsgBoxCtrl InitQcComponent An initialization error has occurred for Check that the file “Fujifilm\IIP\Param\
InitQcAutoCalc MammoQCAutoCalcCom.dll. MammoQCAutoCalcCom.dll” exists.
After check the appropriateness of the
data, restart DR-ID 300CL.
59198 FFQCMsgBoxCtrl InitQcComponent An initialization error has occurred for Check that the file “Fujifilm\IIP\System\
InitQcResultJudge MammoQCResultJudgeCom.dll. MammoQCResultJudgeCom.dll” exists.
After check the appropriateness of the
data, restart DR-ID 300CL.
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageId(m_ 300CL.
QCDataInfo({0}).
nDicomObj, m_
lImageID({1})) : lRet={2})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetStdImageFileUrl(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-198
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetStdImagePixel(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetStdImageLine(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetStdImageDepth(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetExposureMenuName 300CL.
Sbcs(m_QCDataInfo({0}”).
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetExamMenuName 300CL.
Sbcs(m_QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetMenuNameDbcs(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetPlateID(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetDetectorID(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-199
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetSValue(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetExposureResultKVP 300CL.
(m_QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetXconExposure(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetExposureResultMas 300CL.
(m_QCDataInfo({0}").
nDicomObj, lbuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetAnodeTargetMaterial 300CL.
(m_QCDataInfo({0}").
nDicomObj,
strTargetMaterial): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetFilterMaterial(m_ 300CL.
QCDataInfo({0}").
nDicomObj,
strFilterMaterial): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetBodyPartThickness(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetCompressionForce(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-200
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageFileUrl(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicomGetImagePi files reading. the DDO file (INF) and restart DR-ID
xelSpacing(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageDepth(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetMachineCode(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImagePixel(m_ 300CL.
QCDataInfo({0}").
nDicomObj, nBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageHighBit(m_ 300CL.
QCDataInfo({0}").
nDicomObj, lBuf): lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetScanDirection(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetRotationInverse(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-201
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageInfFileUrl(m_ 300CL.
QCDataInfo({0}").
nDicomObj, strBuf):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData LUT table could not be created. Check the appropriateness of the data of
MakeLUTTable(strDDOFile the DDO file (INF) and restart DR-ID
Path, {0}) : lRet={1}) 300CL.
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDiCom. files reading. the DDO file (INF) and restart DR-ID
GetDeviceName(m_ 300CL.
QCDataInfo(0).nDicomObj,
strBuf) : lRet={1}
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDiCom. files reading. the DDO file (INF) and restart DR-ID
GetStationName(m_ 300CL.
QCDataInfo({0}).
nDicomObj, strReader):
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetParamedicCode(m_ 300CL.
QCDataInfo(0).nDicomObj,
m_strPhysician) : lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetParamedicName(m_ 300CL.
QCDataInfo(0).nDicomObj,
strAnk, strKanji, strdata) :
lRet={1})
59198 FFQCMsgBoxCtrl CreateData An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDiCom. files reading. the DDO file (INF) and restart DR-ID
GetStudyDate(m_ 300CL.
QCDataInfo(0),m_
strStudyDate) : lRet={1})
59198 FFQCMsgBoxCtrl MakeLUTTable Failed in accessing the dll. Check the appropriateness of the data of
CCAppImgPreProcGamma the file “Fujifilm\IIP\System\
TblGetter({0}, {1}): lRet={2} ACCAppImgPreProc.dll” and restart DR-ID
300CL.
59198 FFQCMsgBoxCtrl MakeLUTTable Failed in accessing the dll. Check the appropriateness of the data of
FCDAppImgPreProc the file “Fujifilm\IIP\System\
GammaTbl({0}, {1}): ACCAppImgPreProc.dll” and restart DR-ID
lRet={2} 300CL.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-202
Event Event Source Description Causes Remedies Remarks
59198 FFQCMsgBoxCtrl JudgeReExposure An error has occurred at the time of DDO Check the appropriateness of the data of
m_clsQCDicom. files reading. the DDO file (INF) and restart DR-ID
GetImageId(qcData 300CL.
Info({0}).nDicomObj,
nNowImageData({1}).lID) :
lRet={2}
59198 FFQCMsgBoxCtrl DispResultQC An error has occurred at the time of Check that the file “Fujifilm\IIP\Param\
initComXRay database reading/writing. MammoQC.mdb” exists in the database
and restart DR-ID 300CL.
59198 FFQCMsgBoxCtrl Initialize An error has occurred at the time of Check that the file “Fujifilm\IIP\Param\
initComXRay database reading/writing. MammoQC.mdb” exists in the database
and restart DR-ID 300CL.
60000 [UU60000] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall the
system.
60002 [UU60002] User Utility will User Utility will be terminated. Restart the system.
be terminated. If the same error recurs after restarting the
system, reinstall the system.
60050 [UU60050] Failed in Failed in accessing the character string Restart the system.
accessing the character database. If the same error recurs after restarting the
string database. system, reinstall the system.
60051 [UU60051] Failed in Failed in accessing the system setting Restart the system.
accessing the system database. If the same error recurs after restarting the
setting database. system, reinstall the system.
60052 [UU60052] Failed in Failed in obtaining the registry information. Restart the system.
obtaining the registry If the same error recurs after restarting the
information. system, reinstall the system.
60053 [UU60053] Value of registry Value of registry information is abnormal. Restart the system.
information is abnormal. If the same error recurs after restarting the
system, reinstall the system.
60054 [UU60054] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information character string database. If the same error recurs after restarting the
from the character string system, reinstall the system.
database.
60055 [UU60055] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information system setting database. If the same error recurs after restarting the
from the system setting system, reinstall the system.
database.
60056 [UU60056] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information system setting file. If the same error recurs after restarting the
from the system setting file. system, reinstall the system.
60057 [UU60057] Failed in setting Failed in setting information to the system Restart the system.
information to the system setting file. If the same error recurs after restarting the
setting file. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-203
Event Event Source Description Causes Remedies Remarks
60080 [UU60080] Failed in Failed in obtaining the display group Restart the system.
obtaining the display group information. If the same error recurs after restarting the
information. system, reinstall the system.
60081 [UU60081] Failed in Failed in obtaining the menu information. Restart the system.
obtaining the menu If the same error recurs after restarting the
information. system, reinstall the system.
60100 [UU60100] Failed in Failed in obtaining the MPM code. Restart the system.
obtaining the MPM code. If the same error recurs after restarting the
system, reinstall the system.
60114 [UU60114] Failed in Failed in obtaining the XCON parameter. Restart the system.
obtaining the XCON If the same error recurs after restarting the
parameter. system, reinstall the system.
60183 [UU60183] Failed in Failed in accessing the technologist Restart the system.
accessing the technologist information database. If the same error recurs after restarting the
information database. system, reinstall the system.
60184 [UU60184] Failed in Failed in accessing the technologist Restart the system.
accessing the technologist information database. If the same error recurs after restarting the
information database. system, reinstall the system.
60185 [UU60185] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information system setting database. If the same error recurs after restarting the
from the system setting system, reinstall the system.
database.
60203 [UU60203] Failed in Failed in accessing the requesting Restart the system.
accessing the requesting department information database. If the same error recurs after restarting the
department information system, reinstall the system.
database.
60204 [UU60204] Failed in Failed in accessing the requesting Restart the system.
accessing the requesting department information database. If the same error recurs after restarting the
department information system, reinstall the system.
database.
60205 [UU60205] Failed in Failed in obtaining the information from the Restart the system.
obtaining the information system setting database. If the same error recurs after restarting the
from the system setting system, reinstall the system.
database.
60215 [UU60215] Failed in Failed in accessing the film annotation Restart the system.
accessing the film database. If the same error recurs after restarting the
annotation database. system, reinstall the system.
60216 [UU60216] Failed in Failed in accessing the film annotation Restart the system.
accessing the film database. If the same error recurs after restarting the
annotation database. system, reinstall the system.
60251 [UU60251] The data for The data for image processing parameter Restart the system.
image processing is invalid. If the same error recurs after restarting the
parameter is invalid. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-204
Event Event Source Description Causes Remedies Remarks
60260 [UU60260] System has not System has not been rebooted. Restart the system.
been rebooted. If the same error recurs after restarting the
system, reinstall the system.
60300 [UU60300] An option and a An option and a database aren't Restart the system.
database aren't conformable. If the same error recurs after restarting the
conformable. system, reinstall the system.
60301 [UU60301] Failed in saving Failed in saving the information. Re-try it.
the information.
60302 [UU60302] Failed in Failed in deleting. Restart the system.
deleting. If the same error recurs after restarting the
system, reinstall the system.
60303 [UU60303] User Utility The necessary option has not been Restart the system.
cannot be started because installed. If the same error recurs after restarting the
the necessary option has system, reinstall the system.
not been installed.
60304 [UU60304] System error System error occurred. Confirm the connected device.
occurred.
60305 [UU60305] System error System error occurred. Confirm the connected device.
occurred.
60306 [UU60306] A A communication error occurred with the Perform the operation again after checking
communication error X-ray generator. the X-ray generator.
occurred with the X-ray
generator.
61060 [UU61060] Wrong Wrong password has been input. Re-input correctly.
password has been input.
61082 [UU61082] No menu has No menu has been selected. Select item prior to selection.
been selected.
61083 [UU61083] Already Already registered for displaying and Release the display registration for
registered for displaying cannot be deleted. deleting.
and cannot be deleted.
61084 [UU61084] Already Already registered to the study menu and Delete from the study menu for deleting.
registered to the study cannot be deleted.
menu and cannot be
deleted.
61085 [UU61085] Display group Display group has no menu. Register menu or delete/display group.
has no menu.
61090 [UU61090] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the display group name.
display group name.
61094 [UU61094] Display group Display group name is illegal. Either display group is too long (up to 64)
name is illegal. or invalid character has been input.
Re-input correctly.
61103 [UU61103] Same extension Same extension menu code already exists. Input other extension menu code.
menu code already exists.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-205
Event Event Source Description Causes Remedies Remarks
61106 [UU61106] Same ANK Same ANK exposure menu name already Input other ANK exposure menu.
exposure menu name exists.
already exists.
61109 [UU61109] Same kanji- Same kanji-exposure menu name already Input other kanji-exposure menu name.
exposure menu name exists.
already exists.
61110 [UU61110] Extension menu Extension menu code is illegal. Either extension code is too long (up to 16)
code is illegal. or invalid character has been input.
Re-input correctly.
61111 [UU61111] ANK exposure ANK exposure menu name is illegal. Either ANK exposure menu name is too
menu name is illegal. long (up to 64) or invalid character has
been input.
Re-input correctly.
61112 [UU61112] Kanji-exposure Kanji-exposure menu name is illegal. Either kanji-exposure menu name is too
menu name is illegal. long (up to 28) or invalid character has
been input.
Re-input correctly.
61122 [UU61122] MPM code is MPM code is illegal. Either MPM code length is invalid (fix; 4) or
illegal. invalid character has been input.
Re-input correctly.
61125 [UU61125] Same film Same film annotation already exists. Input other film annotation.
annotation already exists.
61130 [UU61130] Manufacturer/ Manufacturer/Equipment Code is illegal. Re-input correctly.
Equipment Code is illegal.
61131 [UU61131] Film-related Film-related code is illegal. Re-input correctly.
code is illegal.
61132 [UU61132] AEC setting is AEC setting is incorrect. Input correctly.
incorrect.
61142 [UU61142] Same study Same study menu code already exists. Input other study menu code.
menu code already exists.
61145 [UU61145] Same ANK Same ANK study menu name already Input other ANK study menu name.
study menu name already exists.
exists.
61148 [UU61148] Same kanji- Same kanji-study menu name already Input other kanji-study menu name (up to
study menu name already exists. 28).
exists.
61149 [UU61149] No exposure No exposure menu has been selected. Select item prior to selection.
menu has been selected.
61151 [UU61151] Exposure menu Exposure menu cannot be registered. Register the exposure menu to another
cannot be registered. study, or delete unnecessary exposure
menus.
61153 [UU61153] Study menu Study menu code is illegal. Either study menu code is too long (up to
code is illegal. 16) or invalid character has been input.
Re-input correctly.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-206
Event Event Source Description Causes Remedies Remarks
61154 [UU61154] ANK study ANK study menu code is illegal. Either ANK study menu name is too long
menu code is illegal. (up to 64) or invalid character has been
input.
Re-input correctly.
61155 [UU61155] Kanji-study Kanji-study menu code is illegal. Kanji-study menu name is too long (up to
menu code is illegal. 28) or invalid character has been input.
Re-input correctly.
61171 [UU61171] No technologist No technologist code has been input. Input the technologist code.
code has been input.
61172 [UU61172] technologist technologist code is illegal. Technologist code should be %s characters
code is illegal. (FIX).
Re-input correctly.
61173 [UU61173] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the technologist code.
technologist code.
61174 [UU61174] Same Same technologist code already exists. Input other technologist code.
technologist code already
exists.
61175 [UU61175] No ANK No ANK technologist name has been input. Input ANK technologist name.
technologist name has
been input.
61176 [UU61176] ANK ANK technologist name is too long. ANK technologist name can be input up to
technologist name is too 64.
long.
61177 [UU61177] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the ANK ANK technologist name.
technologist name.
61178 [UU61178] Same ANK Same ANK technologist name already Input other ANK technologist name.
technologist name already exists.
exists.
61179 [UU61179] No kanji- No kanji-technologist name has been Input kanji-technologist name.
technologist name has input.
been input.
61180 [UU61180] Kanji- Kanji-technologist name is too long. Kanji-technologist name can be input up to
technologist name is too 28.
long.
61181 [UU61181] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the kanji- kanji-technologist name.
technologist name.
61182 [UU61182] Same kanji- Same kanji-technologist name already Input other technologist name.
technologist name already exists.
exists.
61191 [UU61191] No requesting No requesting department code has been Input the requesting department code.
department code has been input.
input.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-207
Event Event Source Description Causes Remedies Remarks
61192 [UU61192] Requesting Requesting department code is too long. Requesting department code can be input
department code is too up to 16.
long.
61193 [UU61193] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the requesting department code.
requesting department
code.
61194 [UU61194] Same Same requesting department code already Input other requesting department code.
requesting department exists.
code already exists.
61195 [UU61195] No ANK No ANK requesting department name has Input ANK requesting department name.
requesting department been input.
name has been input.
61196 [UU61196] ANK requesting ANK requesting department name is too ANK requesting department name can be
department name is too long. input up to 64.
long.
61197 [UU61197] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the ANK ANK requesting department name.
requesting department
name.
61198 [UU61198] Same ANK Same ANK requesting department name Input other ANK requesting department
requesting department already exists. name.
name already exists.
61199 [UU61199] No kanji- No kanji-requesting department name has Input kanji-requesting department name.
requesting department been input.
name has been input.
61200 [UU61200] Kanji-requesting Kanji-requesting department name is too Kanji-requesting department name can be
department name is too long. input up to 8.
long.
61201 [UU61201] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the Kanji- Kanji-requesting department name.
requesting department
name.
61202 [UU61202] Same kanji- Same kanji-requesting department name Input other kanji-requesting department
requesting department already exists. name.
name already exists.
61203 [UU61203] The number of The number of split exposure cannot be Set the amount of exposure division to 1.
split exposure cannot be set more than two except when the
set more than two except function is RT.
when the function is RT.
61211 [UU61211] No film No film annotation characters have been Input the film annotation characters.
annotation characters have input.
been input.
61212 [UU61212] Film annotation Film annotation characters are too long. Film annotation characters can be input up
characters are too long. to 10.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-208
Event Event Source Description Causes Remedies Remarks
61213 [UU61213] Invalid character Invalid character has been input in the film Re-input correctly.
has been input in the film annotation characters.
annotation characters.
61214 [UU61214] Same film Same film annotation already exists. Input other film annotation.
annotation already exists.
61224 [UU61224] The drive is not The drive is not ready yet. Confirm the drive.
ready yet.
61234 [UU61234] Drive is not Drive is not ready. Confirm the drive.
ready.
61235 [UU61235] The wrong disk The wrong disk is inserted. Confirm the drive.
is inserted.
61236 [UU61236] The wrong disk The wrong disk is inserted. Insert %s the backup disk.
is inserted.
61260 [UU61260] No default No default menu has been selected. Select item.
menu has been selected.
61280 [UU61280] No Coding No Coding Scheme Designator has been Input Coding Scheme Designator.
Scheme Designator has input.
been input.
61281 [UU61281] Coding Scheme Coding Scheme Designator is too long. Coding Scheme Designator can be input
Designator is too long. up to 16.
61282 [UU61282] Invalid character Invalid character has been input in Coding Re-input correctly.
has been input in Coding Scheme Designator.
Scheme Designator.
61290 [UU61290] The patient's The patient's information maintenance Restart the system.
information maintenance cannot be executed. If the same error recurs after restarting the
cannot be executed. system, reinstall the system.
61291 [UU61291] The patient's The patient's information database cannot Restart the system.
information database be backed up. If the same error recurs after restarting the
cannot be backed up. system, reinstall the system.
61292 [UU61292] The patient's The patient's information database cannot Restart the system.
information database be restored. If the same error recurs after restarting the
cannot be restored. system, reinstall the system.
61300 [UU61300] Failed in backup Failed in backup process. Re-try it.
process.
61301 [UU61301] Failed in restore Failed in restore process. Re-try it.
process.
61302 [UU61302] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
61310 [UU61310] No direction of No direction of columns has been input. Input direction of columns.
columns has been input.
61311 [UU61311] Direction of Direction of columns is too long. Direction of columns characters can be
columns is too long. input up to 3.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-209
Event Event Source Description Causes Remedies Remarks
61312 [UU61312] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the direction of columns.
direction of columns.
61313 [UU61313] No direction of No direction of rows has been input. Input direction of rows.
rows has been input.
61314 [UU61314] Direction of Direction of rows is too long. Direction of rows characters can be input
rows is too long. up to 3.
61315 [UU61315] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the direction of rows.
direction of rows.
61316 [UU61316] No code value No code value has been input. Input code value.
has been input.
61317 [UU61317] Code value is Code value is too long. Code value characters can be input up to
too long. 16.
61318 [UU61318] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the code code value.
value.
61319 [UU61319] No coding No coding scheme designator has been Input coding scheme designator.
scheme designator has input.
been input.
61320 [UU61320] Coding scheme Coding scheme designator is too long. Coding scheme designator characters can
designator is too long. be input up to 16.
61321 [UU61321] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the coding scheme designator.
coding scheme designator.
61322 [UU61322] No coding No coding scheme version has been input. Input coding scheme version.
scheme version has been
input.
61323 [UU61323] Coding scheme Coding scheme version is too long. Coding scheme version characters can be
version is too long. input up to 16.
61324 [UU61324] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the coding scheme version.
coding scheme version.
61325 [UU61325] No code No code meaning has been input. Input code meaning.
meaning has been input.
61326 [UU61326] Code meaning Code meaning version is too long. Code meaning characters can be input up
version is too long. to 64.
61327 [UU61327] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the code code meaning version.
meaning version.
61329 [UU61329] It registers for It registers for the DX parameter of the Unregister it from exposure menu to
the DX parameter of the exposure menu. It can't be deleted. delete.
exposure menu. It can't be
deleted.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-210
Event Event Source Description Causes Remedies Remarks
61331 [UU61331] Same code Same code value, coding scheme Input other information.
value, coding scheme designator and coding scheme version
designator and coding already exists.
scheme version already
exists.
61332 [UU61332] No user name No user name has been input. Input user name.
has been input.
61333 [UU61333] User name User name version is too long User name characters can be input up to
version is too long 20.
61334 [UU61334] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the User User Name.
Name.
61335 [UU61335] An input user An input user name can't be used. Input other user name.
name can't be used.
61336 [UU61336] Same user Same user name already exists. Input other user name.
name already exists.
61337 [UU61337] No password No password has been input. Input password.
has been input.
61338 [UU61338] Password is too Password is too long. Password can be input up to 20.
long.
61339 [UU61339] Invalid character Invalid character has been input in the Re-input correctly.
has been input in the password.
password.
61341 [UU61341] This file is not This file is not for the contrast medium data Select contrast medium data base.
for the contrast medium base.
data base.
61342 [UU61342] A backup file A backup file does not exist. Confirm the file path.
does not exist.
61343 [UU61343] Direction of Direction of rows includes the same Re-input correctly.
rows includes the same character.
character.
61344 [UU61344] Direction of Direction of columns includes the same Re-input correctly.
columns includes the same character.
character.
61345 [UU61345] No media No media management files exist. Folder that is not subject to reconfiguration
management files exist. may be selected.
61500 [UU61500] All the images Deleting all images within Image Box. Click the button and continue the
in ImageBox will be operation.
deleted.
64311 [UU64311] Data transfer Data transfer service cannot be started at Check the connection and then re-try it.
service cannot be started at the destination device.
the destination device.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-211
Event Event Source Description Causes Remedies Remarks
64312 [UU64312] Cannot connect Unable to connect to the destination Check the connection and then re-try it.
to the destination device. device.
64313 [UU64313] Cannot obtain Cannot obtain the system configuration at Restart the system.
the system configuration at the destination device. If the same error recurs after restarting the
the destination device. system, reinstall the system.
64314 [UU64314] Cannot obtain Cannot obtain the system configuration. Restart the system.
the system configuration. If the same error recurs after restarting the
system, reinstall the system.
64315 [UU64315] An error occurs. An error occurs. This processing cannot be Retry.
This processing cannot be continued.
continued.
64457 [UU64457] The function ES The function ES cannot be specified by the Select other functions.
cannot be specified by the exposure menu that registered in study
exposure menu that menu.
registered in study menu.
64459 [UU64459] An exposure An exposure menu that has been Specify another function.
menu that has been registered to a multiframe-specified study
registered to a multiframe- menu cannot specify the ES function.
specified study menu
cannot specify the ES
function.
64460 [UU64460] MPM code MPM code cannot be extended anymore Assign any existing MPM code.
cannot be extended because it has reached the limit number.
anymore because it has
reached the limit number.
64461 [UU64461] DX parameters DX parameters have not been determined. Set image laterality and patient orientation.
have not been determined.
64509 [UU64509] A query server A query server isn't set up. Choose a query server when you do
isn't set up. automatic retrieve.
Don't check automatic retrieve check when
you can't choose it.
64560 [UU64560] No user group No user group has been entered. Enter a user group.
has been entered.
64563 [UU64563] No display No display items have been assigned. Assign display items appropriately.
items have been assigned.
64566 [UU64566] Failed in data Failed in data transmission processing. Retry.
transmission processing.
64569 [UU64569] No display No display items have been selected. Select display items accordingly.
items have been selected.
64580 [UU64580] The minimum The minimum allowable S value exceeds Input it correctly.
allowable S value exceeds the maximum value.
the maximum value.
64581 [UU64581] The minimum The minimum allowable L value exceeds Input it correctly.
allowable L value exceeds the maximum value.
the maximum value.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-212
Event Event Source Description Causes Remedies Remarks
64584 [UU64584] Failed in Failed in calculating the count how many Restart the system.
calculating the count how times the IP was used. If the same error recurs after restarting the
many times the IP was system, reinstall the system.
used.
64585 [UU64585] Failed in Failed in printout of the file list. Restart the system.
printout of the file list. If the same error recurs after restarting the
system, reinstall the system.
64586 [UU64586] Failed in Failed in resetting the count how many Restart the system.
resetting the count how times the IP was used. If the same error recurs after restarting the
many times the IP was system, reinstall the system.
used.
64587 [UU64587] Failed in Failed in launching the RetakeAnalysis Restart the system.
launching the tool. If the same error recurs after restarting the
RetakeAnalysis tool. system, reinstall the system.
64590 [UU64590] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
64592 [UU64592] Failed in local Failed in local image retrieval and Restart the system.
image retrieval and launching the image saving tool. If the same error recurs after restarting the
launching the image saving system, reinstall the system.
tool.
64595 [UU64595] Failed in Failed in activation of Ascent View setting Restart the system.
activation of Ascent View window. If the same error recurs after restarting the
setting screen. system, reinstall the system.
65535 [MD00001] Failed in Failed in initialization. Check whether the serial port settings of
initialization. OS are in accordance with those of DR-ID
300CL.
68000 [UU68000] Failed in Failed in initialization. (Error ID:%s) Restart the system.
initialization. (Error ID:%s) If the same error recurs after restarting the
system, reinstall the system.
68001 [UU68001] Failed in Failed in retrieval. (Error ID:%s) Restart the system.
retrieval. (Error ID:%s) If the same error recurs after restarting the
system, reinstall the system.
68002 [UU68002] Failed in saving Failed in saving the image. (Error ID:%s) Restart the system.
the image. (Error ID:%s) If the same error recurs after restarting the
system, reinstall the system.
68012 [UU68012] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
68015 [UU68015] The time setting The time setting of QA auto confirmation Enter information correctly.
of QA auto confirmation exceeds a range of 5 to 720 minutes.
exceeds a range of 5 to
720 minutes.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-213
Event Event Source Description Causes Remedies Remarks
68016 [UU68016] Failed in Failed in exporting the patient information Restart the system.
exporting the patient database. If the same error recurs after restarting the
information database. system, reinstall the system.
68017 [UU68017] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall the
system.
68018 [UU68018] An error An error occurred. Restart DR-ID 300CL. If the same
occurred. phenomenon recurred, reinstall the
system.
68020 [UU68020] Writing data Writing data onto media was disabled. Check if the media is write-protected.
onto media was disabled.
68023 [UU68023] The parameter The parameter of reconstruction condition Check the reconstruction condition The following items are
of reconstruction condition is invalid. parameters. changed. (Unit: mm)
is invalid. 1. Slice center height
2. Slice width
3. Slice interval
68024 [UU68024] Failed in starting Failed in starting up the QC tool of the Restart DR-ID 300CL.
up the QC tool of the reader. If the same error recurs after restarting,
reader. reinstall DR-ID 300CL.
68025 [UU68025] No media No media management files exist. Check the folder to be initialized, and then
management files exist. select the folder again.
68026 [UU68026] The setting The setting value of reconstruction Input a correct value.
value of reconstruction condition is invalid.
condition is invalid.
68027 [UU68027] An image of An image of which the height of the slice is Check the height of the slice.
which the height of the slice 0 to 500mm can only be generated. Other
is 0 to 500mm can only be images cannot be generated.
generated. Other images
cannot be generated.
68028 [UU68028] The number of The number of menus for selected MPM Select a menu for other MPM codes and
menus for selected MPM code exceeded the maximum value. then retry.
code exceeded the
maximum value.
68029 [UU68029] Extension menu The extension menu code (1st half) is not Enter the correct value.
code (1st half) is invalid. composed of 16 characters or invalid
character has been input.
68030 [UU68030] Extension menu The extension menu code (2nd half) is not Enter the correct value.
code (2nd half) is invalid. composed of 16 characters or invalid
character has been input.
68031 [UU68031] Failed in Failed in obtaining JJ1017V3 extension Restart DR-ID 300CL.
obtaining JJ1017V3 menu code. If the same phenomenon recurred, reinstall
extension menu code. DR-ID 300CL.
68032 [UU68032] Same extension Same extension menu code already exists. Input other extension menu code.
menu code already exists.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-214
Event Event Source Description Causes Remedies Remarks
69000 [UU69000] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
70000 [UU70000] Failed in Failed in conversion of mis-exposure *Collected log files may include illegal
conversion of mis-exposure image log information. information. Confirm the log file.
image log information.
70001 [UU70001] Failed in Failed in deletion of log files. Restart the system.
deletion of log files. If the same error recurs after restarting the
system, reinstall the system.
70002 [UU70002] Failed in Failed in accessing the definition file. Restart the system.
accessing the definition file. If the same error recurs after restarting the
system, reinstall the system.
70003 [UU70003] Connection to Connection to the log sharing database Restart the system.
the log sharing database resulted unsuccessful. If the same error recurs after restarting the
resulted unsuccessful. system, reinstall the system.
70004 [UU70004] Failed in Failed in acquiring information from the log Restart the system.
acquiring information from sharing database. If the same error recurs after restarting the
the log sharing database. system, reinstall the system.
70005 [UU70005] Connection to Connection to the system setting database Restart the system.
the system setting resulted unsuccessful. If the same error recurs after restarting the
database resulted system, reinstall the system.
unsuccessful.
70006 [UU70006] Failed in Failed in acquiring information from the Restart the system.
acquiring information from system setting database. If the same error recurs after restarting the
the system setting system, reinstall the system.
database.
70007 [UU70007] Connection to Connection to the character string Restart the system.
the character string database resulted unsuccessful. If the same error recurs after restarting the
database resulted system, reinstall the system.
unsuccessful.
70008 [UU70008] Failed in Failed in acquiring information from the Restart the system.
acquiring information from character string database. If the same error recurs after restarting the
the character string system, reinstall the system.
database.
70009 [UU70009] Failed in Failed in acquiring registry information. Restart the system.
acquiring registry If the same error recurs after restarting the
information. system, reinstall the system.
70010 [UU70010] Failed in writing Failed in writing registry information. Restart the system.
registry information. If the same error recurs after restarting the
system, reinstall the system.
70011 [UU70011] Failed in Failed in acquiring technologist Restart the system.
acquiring technologist information. If the same error recurs after restarting the
information. system, reinstall the system.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-215
Event Event Source Description Causes Remedies Remarks
70012 [UU70012] Failed in Failed in acquiring classifications of mis- Restart the system.
acquiring classifications of exposure images. If the same error recurs after restarting the
mis-exposure images. system, reinstall the system.
70013 [UU70013] Failed in Failed in acquiring comments on mis- Restart the system.
acquiring comments on exposure images. If the same error recurs after restarting the
mis-exposure images. system, reinstall the system.
70014 [UU70014] Connection to Connection to the log information database Restart the system.
the log information resulted unsuccessful. If the same error recurs after restarting the
database resulted system, reinstall the system.
unsuccessful.
70015 [UU70015] Failed in Failed in acquiring information for setup of Restart the system.
acquiring information for mis-exposure image analysis. If the same error recurs after restarting the
setup of mis-exposure system, reinstall the system.
image analysis.
70016 [UU70016] Failed in Failed in reading image files. Restart the system.
reading image files. If the same error recurs after restarting the
system, reinstall the system.
70017 [UU70017] Failed in Failed in displaying images. Restart the system.
displaying images. If the same error recurs after restarting the
system, reinstall the system.
70018 [UU70018] Failed in Failed in acquiring the button image. Restart the system.
acquiring the button image. If the same error recurs after restarting the
system, reinstall the system.
70019 [UU70019] Failed in Failed in conversion of date format. Restart the system.
conversion of date format. If the same error recurs after restarting the
system, reinstall the system.
70020 [UU70020] Failed in Failed in storage of the mis-exposure Restart the system.
storage of the mis-exposure image log list. If the same error recurs after restarting the
image log list. system, reinstall the system.
70021 [UU70021] Failed in Failed in storage of the calculation result. Restart the system.
storage of the calculation If the same error recurs after restarting the
result. system, reinstall the system.
70022 [UU70022] An error An error occurred while an image is being Restart the system.
occurred while an image is displayed. If the same error recurs after restarting the
being displayed. system, reinstall the system.
70023 [UU70023] An error An error occurred. Restart the system.
occurred. If the same error recurs after restarting the
system, reinstall the system.
70024 [UU70024] Connection to Connection to the database of regions to Restart the system.
the database of regions to be exposed resulted unsuccessful. If the same error recurs after restarting the
be exposed resulted system, reinstall the system.
unsuccessful.
010-305-19E
06.29.2018 FM9522
DR-ID 300CL Service Manual Appx MT B-216