Professional Documents
Culture Documents
CHEMISTRY ANALYZER
AU480
ISE USER GUIDE
© 2009 Beckman Coulter Biomedical K.K.. All Rights Reserved.
Beckman Coulter AU480 User Guide.
This User Guide as well as the system described in it may be used or copied only in accordance with
the terms of copyright. The content of this User Guide is intended for instructional use only and is
subject to change without notice.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form
or by any means, electronic, mechanical or otherwise, without the prior written permission of Beckman
Coulter Biomedical K.K..
Please remember that existing artwork or images from this guide that you may want to include in your
project may be protected under copyright law.
The unauthorized incorporation of such material into your new work could be a violation of the rights of
the copyright owner. Please be sure to obtain any permission required from the copyright owner.
Any references to any company names or peoples names in sample templates and windows are for
demonstration purposes only and are not intended to refer to any actual organization or names of
persons.
DCN-1
ISE User Guide
Version 1.0
Return to Main Table of Contents
DCN-2
ISE User Guide
Version 1.0
Return to Main Table of Contents
FORWARD
Thanks for using the optional ISE unit for Beckman Coulter Automated Chemistry Analyzers.
This ISE User Guide (Supplement) offers detail descriptions, limited troubleshooting to common
problems, and maintenance procedures to be used in conjunction with the conventional user guide.
Use also the conventional user guide.
Objective readers
This manual is intended for laboratory personnel who have completed the training course offered by
Beckman Coulter, or personnel who have been trained by those who have taken Beckman Coulter
training course.
ISE User Guide
Version 1.0
Return to Main Table of Contents
AU480 ISE
User Guide User Guide
This manual
ii
ISE User Guide
Version 1.0
Contents
ISE User Guide Contents
Version 1.0
Contents ISE User Guide
Version 1.0
Return to Main Table of Contents
Part I
ISE TROUBLESHOOTING
In order to keep the ISE unit of the AU480 in good condition and to
maintain it’s original performance, customers should perform periodic
maintenance.
This part of “ISE TROUBLESHOOTING” describes problems that can
occur and how to resolve them in flow chart form.
Also refer to “Part II MAINTENANCE FLOW”.
Contents
1 ISE WORKFLOWS AND PROBLEMS THAT CAN OCCUR...........................................I-2
2 OTHER ERRORS IN ISE..................................................................................................I-3
3 TROUBLESHOOTING.....................................................................................................I-4
I-1
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
Calibration
Analysis Operation
Start (2) ISE sample pot overflow
sensor error (3186)
(4) Concurrent
reproducibility error Refer to page I-8.
Maintenance
(7) Selectivity check error Refer to page I-11.
Power OFF
I-2
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
If any of the following errors are encountered, check the level of the solution
named in the Display column next to the corresponding Alarm No. Column.
These Errors indicate insufficient reagent liquid level
Alarm No. Display
3251 ISE Buffer Solution Short
3252 ISE MID Solution Short
3253 ISE REF Solution Short
3254 ISE Detergent Short
3255 ISE Serum Standard Solution H Short
3256 ISE Serum Standard Solution L Short
3257 ISE Urine Standard Solution H Short
3258 ISE Urine Standard Solution L Short
3259 ISE Electrode Check Liquid (Na) Short
3260 ISE Electrode Check Liquid (K) Short
3261 ISE CRS Standard Solution H Short
3262 ISE CRS Standard Solution M Short
3263 ISE CRS Standard Solution L Short
I-3
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
3 TROUBLESHOOTING
(1) Troubleshooting for ISE Sample Pot Liquid level Sensor Error
-1 (3186, 3187)
3186(1)
3187(1) ISE sample pot liquid level sensor error – 1
3187(2)
END
Details of operating
Check item Cause and action Service call
procedure
CALL
I-4
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
END
(2) b Troubleshooting for ISE Sample Pot Liquid level Sensor Error
-2 (3186)
3186(3) ISE sample pot liquid level sensor error – 2
END
Details of operating
Check item Cause and action Service call
procedure
CALL
I-5
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
NO
YES
Are air bubbles
gotten in the REF END
solution tubing?
END
I-6
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
END
END
END
Details of operating
Check item Cause and action Service call
procedure
CALL
I-7
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
END
The error results from The sample pot gets dirty, clogged or leaks.
the diluent pot. →Wash or replace the sample pot.
Part II 7.2, 9.2, 11.1
MThis error results from The tubing system is clogged or leaks.
MID solution dispense. →Check the roller tube.
Part II 9.1
The error results Air enters into the REF solution tubing.
from the tubing. →Replace the electrode set and O-ring.
Part II 8.1
The tubing get clogged, dirty, or leaks.
→Wash or replace the tubing.
Part II 5.2, 7.2, 9.1, 9.2, 9.3, 9.4, 11.2, 11.4
The error results from Air enters into the REF solution tubing.
the REF solution tubing. →Replace the REF electrode packing.
Part II 10.1
The amount of the REF electrode solution is insufficient.
→Add the REF electrode solution.
Part II 8.2
The indicated electrode/REF
The error results electrode has deteriorated.
from electrodes. →Replace the electrode.
Part II 8.1, 10.1
Details of operating
Check item Cause and action Service call
procedure
CALL
I-8
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
END
NO
YES Did the control sample
analyze next to the
highvalue sample?
The indicated electrode has deteriorated.
→Replace the electrode.
Part II 8.1
NO
YES Has the problem been corrected?
CALL Contact Beckman Coulter Technical Services.
NO
YES
END
END
I-9
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
END
Details of operating
Check item Cause and action Service call
procedure
CALL
I-10
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
END
Details of operating
Check item Cause and action Service call
procedure
CALL
I-11
ISE User Guide ISE TROUBLESHOOTING
Version 1.0
Return to Main Table of Contents
I-12
ISE TROUBLESHOOTING ISE User Guide
Version 1.0
Return to Main Table of Contents
Part II
ISE MAINTENANCE FLOW
In order to maintain the original performance of the AU480 ISE unit
and to keep it operating in good condition users should perform
periodic maintenance on the analyzer.
This part of “ISE MAINTENANCE FLOW” describes the daily,
periodic and occasional maintenance procedures to be performed in
case of trouble along with a series of operation flows, based on the
conventional ISE User Guide.
Also refer to “Part I ISE TROUBLESHOOTING”.
Contents
1 ISE MAINTENANCE CHECKLIST..................................................................................II-2
2 OVERVIEW OF ISE MAINTENANCE FLOW..................................................................II-4
3 LIST OF PARTS USED FOR ISE MAINTENANCE........................................................II-5
4 TUBING BLOCK DIAGRAM FOR ISE MAINTENANCE................................................II-7
5 DAILY MAINTENANCE...................................................................................................II-8
6 WEEKLY MAINTENANCE............................................................................................II-12
7 MAINTENANCE REQUIRED EVERY TWO WEEKS OR 3,000 SAMPLES.................II-16
8 MAINTENANCE REQUIRED EVERY TWO MONTHS OR 20,000 SAMPLES.............II-22
9 MAINTENANCE REQUIRED EVERY THREE MONTHS OR 20,000 SAMPLES.........II-26
10 MAINTENANCE REQUIRED EVERY TWO YEARS OR 150,000 SAMPLES..............II-38
11 MAINTENANCE ON AN AS-NEEDED BASIS..............................................................II-42
12 RESETTING THE SYSTEM WHEN SWITCHED TO THE STOP MODE
WHILE PERFORMING MAINTENANCE......................................................................II-64
When performing maintenance on the ISE unit be sure to wear protective gloves
to avoid infection and other injury and observe the “WARNING” and “CAUTION”
notations in this user guide.
CAUTION
The ISE maintenance flow, biweekly or longer periodic maintenance is managed on the basis
of either period of time or number of samples.
The AU480 ISE can process 200 samples per day.
II-1
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Date (below)
Maintenance
Month: Year:
Daily 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
5.1 Check the
Buffer Syringe for
Leaks
5.2 Auto-wash the
Sample Pot and
Electrode Line
Weekly 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
6.1 Check the
Selectivity of
the Na and K
Electrodes
6.2 Enhanced
Cleaning of
Electrode Line
Every two
weeks or 3000 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Samples
7.1 Manually
Wash the Mix Bar
and Liquid Level
Sensors
7.2 Manually
Wash the Sample
Pot and the
Sample Pot Tubes
Every two
months or 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
20000 Samples
8.1 Replace
the Na, K, or Cl
Electrode
8.2 Check
and Add REF
Electrode Solution
Every three
months or 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
20000 Samples
9.1 Replace the
Roller Tubes for
MID Solution
Dispense and
Mixture Aspiration
9.2 Replace the
Sample Pot Tubes
9.3 Replace the
REF Block Side
Drain Tube and
the Pinch Valve
Tubing
9.4 Replace the
Drain Tube and
Manually Wash
the Drain Tank
II-2
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Date (below)
Maintenance
Month: Year:
Every two
years or
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
150000
Samples
10.1 Replace the
REF Electrode
and Packing
On an as
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
needed basis
11.1 Replace the
Sample Pot
11.2 Clean the
Electrode Block
(Inlet Side)
11.3 Manually
Wash the K
Electrode
11.4 Manually
Wash and
Replace the REF
Electrode Block
11.5 Replace
Reagents
Replace Buffer
solution
Replace MID
solution
Replace REF
solution
11.6 Replace the
Buffer Syringe
11.7 Procedure
for Enhanced
Cleaning of the
Electrode Line
(Manual)
11.8 Replace the
Buffer Syringe
Case and Case
Head
II-3
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
DAILY MAINTENANCE
WEEKLY MAINTENANCE
7.1 Manually Wash the Mix Bar and Liquid Level Sensors .......II-16
7.2 Manually Wash the Sample Pot and the Sample Pot Tubes .......II-18
II-4
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Target
Maintenance
Maintenance Item Parts to Be Used Part Number Location
Interval
(in Diagram)
Daily 5.1 Check the Buffer - - (0)
Syringe for Leaks
5.2 Auto-wash the Sample ISE cleaning solution AUH1019 (1) to (11)
Pot and Electrode Line
Weekly 6.1 Check the Selectivity Selectivity check liquid AUH1018 Electrodes
of the Na and K
Electrodes
6.2 Enhanced Cleaning of ISE cleaning solution AUH1019 (1) to (11)
Electrode Line
Every two 7.1 Manually Wash the Ethyl alcohol (Ethanol) Commercial item (1) and (2)
weeks or 3,000 Mix Bar and Liquid Level
samples Sensors
7.2 Manually Wash the Wash solution OSR0001
Sample Pot and the
Sample Pot Tubes
Every two 8.1 Replace the Na, K, or Na electrode MU9194 Electrodes
months or Cl Electrode K electrode MU9195
20,000 samples
Cl electrode MU9196
O-ring MU9900
8.2 Check and Add REF REF electrode solution AUH1017
Electrode Solution
Every three 9.1 Replace the Roller Roller Tube MU9623 (9)
months or Tubes for MID Solution
20,000 samples Dispense and Mixture
Aspiration
9.2 Replace the Sample Tube set MU5386 (4)
Pot Tubes
9.3 Replace the REF Tube set 2 MU8247 (7) and (8)
Block Side Drain Tube Pinch valve tubes ZM2970
and the Pinch Valve
Tubing
9.4 Replace the Drain Drain Tube 1 MU8302 (10) and (11)
Tube and Manually Wash Sodium hypochlorite Commercial item
the Drain Tank solution
Every two years 10.1 Replace the REF REF Electrode (with the MU9197 Electrodes
or 150,000 Electrode and Packing packing)
samples The packing for REF MU9202
Electrode
II-5
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Target
Maintenance
Maintenance Item Parts to Be Used Part Number Location
Interval
(in Diagram)
On an as 11.1 Replace the Sample Sample Pot MU9627 (3)
needed basis Pot
11.2 Clean the Electrode Mandolin string(f0.3) ZM0227 (5)
Block (Inlet Side)
11.3 Manually Wash the K - - Electrodes
Electrode
11.4 Manually Wash REF electrode block 2 MU8245 (6)
and Replace the REF
Electrode Block
II-6
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
4
Version 1.0
Level sensor connector
Mixer connector
MID solution
(2) Mix bar (3) Sample pot
Buffer solution
Roller pump
(for MID solution dispense)
(4) Tubes between the (0)
Return to Main Table of Contents
E
Description of Tubing Block Diagram Symbols
(5) Electrode block CI (0) to (11): The numbers of tubing, etc.,
(8) Pinch valve tubing
(inlet) for which maintenance is performed
A to H: The symbols of detaching locations, etc.,
Na Electrode that are shown in the detail description
F of maintenance routine
K
D
→Waste solution
(6) REF
electrode block H
(10) Drain tube
II-7
Return to Main Table of Contents
5 DAILY MAINTENANCE
Fixing screws
Case head
(0)
Syringe case Syringe
Buffer syringe
• If protective gloves, etc., get wet, thoroughly rinse them immediately with water.
• Verify that strong alkali such as AU wash solution alkali does not come in contact
CAUTION with the syringe cases. If the alkali wash solution comes in contact with the
syringe case cracks may occur. If there is contact with the syringe case rinse
thoroughly with water.
3. Using a dry cloth, wipe the bottom of the syringe case, the joint of the case head and syringe
case, and the vicinity of fixing screws to check for leaks.
If any of the liquid has leaked, wipe it off with a dry clean cloth.
4. Visually check the case head for cracks.
If any cracks are found on the case head, replace it with a new one. After replacing the case head
advance to step 9.
5. Check to see if the case head and the syringe are loose.
• If the case head and syringe case are loose, tighten them by turning the syringe case
clockwise toward the case head.
• If any leak was found, allow approximately five minutes after tightening the case head and
syringe case. Then, recheck the bottom of the syringe case for leaks.
II-8
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
6. Check to see if the two fixing screws on the case head, fixing nut, and piston fixing screw are
loose.
• If loosen, firmly tighten the fixing screws, fixing nut, and piston fixing screw.
• If a leak was found, allow approximately five minutes after tightening the case head and
syringe case. Then, recheck the bottom of the syringe case, etc. for leaks.
7. Visually check to see if any liquid is leaking inside the syringe case.
If any leak was found, go to section “11.6 Replace the Buffer Syringe” in this chapter.
8. From the AU480 “Home” window select Menu List>Maintenance>User Maintenance>ISE
Maintenance>Maintenance to display the “ISE Maintenance: Maintenance tab” window.
• Reagent may drip from the case head during syringe replacement. Wipe off the
reagent with a paper towel.
CAUTION • If leaks still occur even after tightening the loose parts, replace the syringe.
II-9
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
When handling wash solution, wear protective gloves to avoid contact with the wash
solution.
If hands or clothing come into contact with the wash solution, wash them
CAUTION immediately with water. Should wash solution get splashed accidentally in the eyes
or mouth, rinse it immediately with water and consult a physician.
If the sample pot and electrode line remain unwashed, contamination between samples will occur,
which may result in a calibration or data error. After completing daily analysis, be sure to auto-wash
the sample pot and electrode line. It takes approximately 4 minutes from start to auto-wash the sample
pot and electrode line.
If performing an analysis after auto-wash be sure to execute calibration before starting the analysis.
Avoid spilling wash solution when handling the reagent bottle. If the wash solution is
spilled, wipe it off immediately with a cloth or paper towel soaked in water.
CAUTION
II-10
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
II-11
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
6 WEEKLY MAINTENANCE
Do not leave the bottle of selectivity check liquid open. Evaporation may effect the
concentration of the selectivity check liquid or it may crystallize.
CAUTION
II-12
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
9. Select C h e c k S t a r t .
The “Selectivity Check” window appears.
10. Select O K .
The selectivity check procedure will start and display the check results.
11. Check the selectivity data.
For abnormal data, the background of the numeric value is displayed in yellow.
The system will judge a value more than 160 mmol/L for Na-electrode and a value more than 6.0
mmol/L for K-electrode as abnormal data.
If the check results are abnormal, replace the relevant electrode.
For details on replacing the electrode, refer to section “8.1 Replace the Na, K, or Cl Electrode” in
this chapter.
II-13
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
12. Remove previously placed Na and K sample cups from STAT table.
13. The maintenance date will be automatically updated in the list.
When handling wash solution, wear protective gloves to avoid contact with the wash
solution.
If hands or clothing come into contact with the wash solution, wash them
CAUTION immediately with water. Should wash solution get splashed accidentally in the eyes
or mouth, rinse it immediately with water and consult a physician.
II-14
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
7. Select O K .
The enhanced cleaning operation starts. This procedure requires approximately 30 minutes to
complete.
Maintenance date is automatically updated in the list.
8. After completing enhanced cleaning operation, remove the sample cup filled with the wash
solution from the STAT table.
II-15
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
7.1 Manually Wash the Mix Bar and Liquid Level Sensors
If the mix bar and liquid level sensors are not washed over an extended period of time, they become
contaminated with serum, urine or a fur, etc.
Wash the mix bar and liquid level sensors with alcohol every two weeks or every 3,000 samples,
whichever comes first.
Maintenance locations: Nos. (1) and (2) in the ISE tubing block diagram
Level sensor connector
Mixing unit Mixing motor connector
Mixing unit clamp knob
Buffer solution and
MID solution connecting tubes
(1) Liquid level sensor
Nozzle
Procedure for Wash the Mix Bar and Liquid Level Sensors
Materials:
II-16
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Be careful not to bend or break the liquid level sensors when washing them.
CAUTION
11. Wipe the two liquid level sensors and mix bar with a cloth dampened with ethyl alcohol (ethanol).
Mix bar
Nozzle
Connecting tubes
Mixing unit
II-17
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
14. Reconnect the connectors of the liquid level sensor (714) and mixing motor (706) to the mixing
unit.
15. Press the TA B L E R OTAT I O N / D I A G button.
The MID solution is supplied.
16. Close the ISE cover.
17. Close the main cover.
Do not change the orientation and position of two nozzles attached to the mixing unit
and do not use force pulling on the connecting tubes of the mixing unit.
CAUTION
18. Select U p d a t e .
A confirmation window appears.
19. Select O K .
7.2 Manually Wash the Sample Pot and the Sample Pot Tubes
The sample pot tubes are the three tubes that run between the sample pot, electrode unit, and T-
connector. When the system processes samples that contain large amounts of fiblin and protein debris
may build up in the tubes over time.
Maintenance locations: Nos. (3) and (4) in the ISE tubing block diagram
(4)Tubes between
the sample pot and
electrode unit (tube set)
T-connector
Electrode unit
.
II-18
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Procedure for Wash the Sample Pot, Tubes between the Sample Pot and
Electrode Unit, and T-connector
Materials:
II-19
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
11. Remove the tube attached to the bottom of the sample pot.
12. Remove the tube from the electrode unit and the bypass tube.
13. Remove three tubes from the T-connector.
14. Supply washing water or the sample diluent of approximately 2% concentration of alkali wash
solution to the sonicator filled with deionized water.
To prevent the REF electrode block from deforming due to the ultrasonic cleaning,
follow these precautions.
• Do not perform ultrasonic cleaning for more than ten minutes.
CAUTION
• Use cleaning liquid at room temperature.
15. Wash the T-connector and the two tubes together with the sample pot in the sonicator filled with
deionized water for approximately 10 minutes.
16. Rinse the sample pot, T-connector, and the three tubes with deionized water or distilled water.
17. Wipe the sample pot, T-connector, and the three tubes with a wiper cloth.
18. Connect the three tubes to the T-connector.
Make sure to push the tubes all the way on the T connector to avoid leaks.
CAUTION
19. Connect the other end of the tube to the original position of the electrode unit and the bypass
tube.
20. Attach the tube to the bottom of the sample pot.
Tighten the tube knob firmly.
Tighten the knob firmly when attaching the tube to the bottom of the sample pot.
CAUTION
When remounting the mixing unit be careful not to get the tube caught in the
analyzer’s peripheral part.
CAUTION
II-20
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
24. Reconnect the connectors of the mixing unit and the mixing motor.
25. Press the TA B L E R OTAT I O N / D I A G two or three times.
The MID solution is supplied.
Confirm that liquid is properly aspired from the sample pot to the electrode unit. If liquid remains
in the sample pot, check the tubes connections.
26. Close the ISE cover.
27. Close the main cover.
28. Select U p d a t e .
A confirmation window appears.
29. Select O K .
II-21
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
E (O-ring)
CI Cl electrode cord (blue)
REF
Lock lever
• Na electrode : MU9194
• K electrode : MU9195
• Cl electrode : MU9196
• O-ring : MU9900
II-22
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
9. Move the lock lever to the back, if viewed from the front, then unlock the electrodes
to be replaced.
10. Remove the three electrodes.
11. Disconnect the cords from the Cl electrode, Na electrode, or K electrode.
Three O-rings are attached to one side of three new electrodes. Be careful not to
lose the O-rings when mounting the electrodes.
An O-ring is not attached to the right side of the Cl electrode. Use the separately
CAUTION packaged O-ring to insert between the right side of Cl electrode and the metal plate.
II-23
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
12. Connect the blue cord, yellow cord, and red cord to the Cl electrode, Na electrode, and K
electrode, respectively.
13. Mount the three new electrodes on the electrode block.
Mount the electrodes in order of Cl, Na, and K from the sample pot side to the front side.
14. Move the lock lever forward to clamp the electrodes.
15. Press the TA B L E R OTAT I O N / D I A G button.
The MID solution is supplied to the tube.
The two roller pumps turn on and move the solution in each roller tube.
Press and hold down the TA B L E R OTAT I O N / D I A G button until the roller pumps are
activated five times. Visually check the tube to see if the solution passing through the electrodes
includes air bubbles. Repeat this operation until there are no air bubbles in the tube.
16. Select U p d a t e .
A confirmation window appears.
17. Enter the serial number of each replacement electrode, and then select O K .
The “Electrode Serial No. Setup” window is closed.
18. Close the ISE cover.
19. Close the main cover.
20. Allows at least 5 minutes after closing the main cover, and then perform calibration analysis.
To obtain the best possible analysis data, perform two successive calibration
measurements to confirm the electrode stability.
If the difference in factor value between the first and second calibrations is within the
CAUTION following values, the electrodes are stable. The MID solution concentrations in the
following table have been set from “Specific Test Parameters: ISE” menus.
Na K CI
MID solution concentration 140 4.0 100
Difference between 1st and 2nd factors 0.020 0.045 0.025
• If the difference between those factor values is not within each value shown
above, of if slope 0 is output at the first calibration:
There is a possibility that air remains inside the electrode block. First, open the
main cover and ISE cover.
From the AU480 “Home” window select Menu List>Maintenance>User
Maintenance>ISE Maintenance>Maintenance to display the “ISE
Maintenance: Maintenance tab” window.
Select the “ISE Maintenance” check box. After selecting the M I D / R E F
P r i m e , press the TA B L E R OTAT I O N / D I A G button.
II-24
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
REF electrode
Reference line
1. Remove the cap on the REF electrode, then add the REF electrode solution until it is above the
reference line.
If air bubbles are visible in the electrode of the REF electrode block, turn the
electrode downward and tap it with finger. Refer to section “10.1 Replace the REF
CAUTION Electrode and Packing” in this chapter.
II-25
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
9.1 Replace the Roller Tubes for MID Solution Dispense and
Mixture Aspiration
The roller tubes will deteriorate due to friction with each roller pump and vibrations. If used for an
extended period of time these tubes will break. Replace the roller tubes with new ones every three
months or every 20,000 samples.
Roller pump
II-26
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
• Avoid spilling solution from the roller tube when replacing the tube.
• The roller tube consists of the MID solution roller tube and waste liquid roller
CAUTION tube. Replace these two roller tubes at the same time.
II-27
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
9. Connect the roller tube with the roller pump, matching connectors with the notches on the roller
pump.
Connect the MID solution roller tube and waste liquid roller tube with each roller pump in the
same manner as the roller tube pump.
Make the connection so the numbers on the tubes match those on the notches near the roller
pump.
II-28
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
(4)Tubes between
the sample pot and
electrode unit (tube set)
T-connector
Electrode unit
Replace the Tubes between Sample Pot and Electrode Unit, and T-connector
Materials:
Prepare the system for maintenance before beginning the procedure. If the operation
is not performed in Warm up or Standby, liquid may be dispensed from nozzles,
CAUTION tubes, etc. during maintenance.
II-29
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-30
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Make sure to push the tubes all the way on the T connector to avoid leaks. Tighten
the knob firmly when attaching the tube to the bottom of the sample pot.
CAUTION
When remounting the mixing unit in place, be careful not to splash the buffer solution
and MID solution.
CAUTION
9. Reconnect the connectors of the liquid level sensor (714) and mixing motor (706) of the mixing
unit in place.
10. Confirm that D ra i n - F l o w c e l l is selected on the “ISE Maintenance: Maintenance tab”
window.
11. Press the TA B L E R OTAT I O N / D I A G button two or three times.
Ensure that liquid is properly aspired from the sample pot to the electrode unit.
If liquid remains in the sample pot, check the tube connections.
12. Select U p d a t e .
A confirmation window appears.
13. Select O K .
14. Close the ISE cover.
15. Close the main cover.
Refer to “ (1) Troubleshooting for ISE Sample Pot Liquid level Sensor Error -1 (3186, 3187)” in section
“3. TROUBLESHOOTING” in Part I.
II-31
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
9.3 Replace the REF Block Side Drain Tube and the Pinch Valve
Tubing
The REF block-side drain tube is subject to accumulation of contaminants and may adversely affect
analysis data due to flow-rate decrease and noise interference if it is used over an extended period of
time. Replace the drain tube with a new one every three months or every 20,000 samples.
The pinch valve tubing is provided with pinch valves.
If the pinch valve tubing is used for an extended period of time, the section pinched by the valve
will deteriorate. Replace the pinch valve tubing with a new one every three months or every 20,000
samples. Concurrently replace the REF block-side drain tube and pinch valve tubing.
Maintenance locations: Nos. (7) and (8) in the ISE tubing block diagram
B
Tubes between
the sample pot and
electrode unit (tube set)
Na
Electrodes
K
REF C
(6)REF electrode
block
(7)REF block-side
drain tube Pinch valve
(tube set 2)
A
Procedure for Replace the REF Block Side Drain Tube and Pinch Valve Tubing
Materials:
II-32
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Prepare the system for maintenance before beginning the procedure. If the operation
is not performed in Warm up or Standby, liquid may be dispensed from nozzles,
CAUTION tubes, etc. during maintenance.
II-33
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Do not mount each branch tube in a different groove on the pinch valve. If a tube is
mounted backwards, liquid will not be supplied and drained correctly.
Mount a shorter tube in the bottom groove of the pinch valve (between A and C in
CAUTION the tubing block diagram). Mount longer tube in the top groove of the pinch valve
(between B and C in the tubing block diagram).
11. Mount a new section of pinch valve tubing by connecting the branches at positions A, B, and C in
the tubing block diagram.
Connect the longer Y-branched tube at position B.
12. Insert the two branch tubes into the grooves on the pinch valve, so that each tube passes through
the middle of the groove.
Mount the tube in the bottom groove of the pinch valve while pressing the top end of the pinch
valve with finger, in the same way the tube was removed. Mount the tube in the top groove of the
pinch valve directly.
To position B
Groove
To position A
Pinch valve
II-34
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Refer to “ (1) Troubleshooting for ISE Sample Pot Liquid level Sensor Error -1 (3186, 3187)” in section
“3. TROUBLESHOOTING” in Part I.
9.4 Replace the Drain Tube and Manually Wash the Drain Tank
If the system uses samples that contain large amounts of fibrin and protein such as a dialysis sample,
fibrin, protein, etc. debris will accumulate at the drain tube outlet and drain tank, resulting in a
calibration or data error.
Replace the drain tube and manually wash the drain tank every three months or 20,000 samples,
whichever comes first.
Maintenance locations: Nos. (10) and (11) in the ISE tubing block diagram
D
Pinch valve tubing → Waste solution H
Roller pump
(for mixture aspiration)
G (Joint No.3)
(11) Drain tank
Procedure for Replace the Drain Tube and Manually Wash the Drain Tank
Materials:
II-35
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-36
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
When handling the solution hypochlorite solution, wear protective gloves to avoid
coming into contact with the sodium hypochlorite solution.
If hands or clothing come into contact with the wash solution, wash them
CAUTION immediately with water. Should wash solution get splashed accidentally in the eyes
or mouth, rinse it immediately with water and consult a physician.
When connecting the tubing, make sure to push the tube all the way to avoid leaking
joints.
CAUTION
Verify that the solution aspired reliably from the sample pot into the electrode unit
and drained from the drain tube, and verify that no air bubbles are present in the
drain tank.
CAUTION If solution remains in the sample pot, make sure the tubes are connected correctly.
8. Select U p d a t e .
A confirmation window appears.
9. Select O K .
10. Close the ISE cover.
11. Close the main cover.
II-37
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
CI
Na
REF electrode packing
K Cap screw
To avoid damage when replacing the REF electrode to not use force mounting or
dismounting the electrode. When dismounting or mounting the electrode, be careful
CAUTION not to break the electrode unit.
II-38
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
When loosening the set screw, do not press the connector of the thermistor cord by
hand, otherwise the cord could break.
CAUTION
12. Loosen the set screw, then gently remove the REF electrode along with the set screw.
13. Remove the packing of the REF electrode.
II-39
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
The REF electrode tip can be easily inserted into the packing if it is dampened with
deionized water.
TIP
5. After inserting the REF electrode tighten the coupling screw by rotating it 1/4 to 1/2 turn, then
secure the REF electrode.
6. Connect the REF electrode cord correctly.
7. Remount the block in place.
8. Clamp the lock lever.
9. Select M I D / R E F P r i m e .
10. Select O K .
11. Press the TA B L E R OTAT I O N / D I A G button.
The MID solution will be injected into the tube.
The two roller pumps are activated and solution flows into the individual roller tubes.
Press the TA B L E R OTAT I O N / D I A G button until the roller pumps are activated five
times or so. Visually check for air bubbles in the tubes.
Press the TA B L E R OTAT I O N / D I A G button again until no air bubbles are found in the
tube.
If air bubbles will not purge, push in the cap screw.
12. Select U p d a t e .
The “Electrode Serial No. Setup” window appears.
13. Enter the serial number of the new reference electrode, and then select O K .
14. Close the ISE cover.
15. Close the main cover.
II-40
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
16. Allows at least 5 minutes after closing the main cover, and then perform calibration analysis.
To obtain the best possible analysis data, perform two successive calibration
measurements to confirm the electrode stability.
If the difference in factor value between the first and second calibrations is within the
CAUTION following values, the electrodes are stable. The MID solution concentrations in the
following table are those set up in “Specific Test Parameters: ISE” menus.
Na K CI
MID solution concentration 140 4.0 100
Difference between 1st and 2nd factors 0.020 0.045 0.025
• If the difference between those factor values is not within each value shown
above, or if slope 0 is output at the first calibration:
There is a possibility that air remains inside the electrode block. First, open the
main cover and ISE cover.
From the AU480 “Home” window select Menu List>Maintenance>User
Maintenance>ISE Maintenance>Maintenance to display the “ISE
Maintenance: Maintenance tab” window.
After selecting the “ISE Maintenance” check box, select M I D / R E F
P r i m e , and then repeat steps 9 to 12 in the above maintenance procedure.
II-41
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-42
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Tighten the knob firmly when attaching the tube to the bottom of the sample pot.
TIP
When remounting the mixing unit in place, avoid splashing the buffer solution and
MID solution.
CAUTION
10. Reconnect the connectors of the liquid level sensor (714) and mixing motor (706) of the mixing
unit in place.
11. Press the TA B L E R OTAT I O N / D I A G button two or three times.
Confirm that liquid is properly aspired from the sample pot to the electrode unit.
If liquid remains in the sample pot check, the tube connections.
12. Select U p d a t e .
A confirmation window appears.
13. Select O K .
II-43
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Refer to “ (1) Troubleshooting for ISE Sample Pot Liquid level Sensor Error -1 (3186, 3187)” and “ (4)
Troubleshooting for Concurrent Reproducibility Error” in section “3. TROUBLESHOOTING” in Part I.
Electrodes
REF
Mandolin: ZM0227
II-44
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
II-45
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Four O-rings are attached to each side of three electrodes and the metal part
(location E in the tubing block diagram) that comes in contact with the Cl electrode.
CAUTION Be careful not to loosen the O-rings when removing the electrodes.
5. Pass the mandolin strings through the flow path hole on the side of the electrode block.
If contamination disturbed, it may get stuck in the mixture flow path in the electrode block. Bind
the mandolin strings up to the maximum thickness that can pass through the flow path.
6. Remove the stuck contamination by turning the mandolin strings while holding both ends by
hand.
When contaminants stick on the mandolin strings, wipe them with a dry, clean cloth or paper
towel several times.
7. Connect the blue cord, yellow cord, and red cord to the Cl electrode, Na electrode, and K
electrode, respectively.
8. Mount the three electrodes on the electrode block.
Attach the electrodes in order of Cl, Na, and K from the sample pot side.
9. Move the lock lever to clamp the electrodes.
10. Mount the tube connecting from the sample pot to the electrode block.
11. Press the TA B L E R OTAT I O N / D I A G button.
The two roller pumps are activated and solution flows into individual roller tubes.
Press the TA B L E R OTAT I O N / D I A G button until the roller pumps are activated five
times or so. Visually check that there are no air bubbles in the tube.
Press the TA B L E R OTAT I O N / D I A G button again until no air bubbles are found in the
tube.
12. Select U p d a t e .
A confirmation window appears.
13. Select O K .
14. Close the ISE cover.
15. Close the main cover.
Refer to “ (3) Troubleshooting for Calibration Error – 1 of 2 (3264, 3265, 3266, 3267, 3268, 3269)”,
“ (4) Troubleshooting for Concurrent Reproducibility Error”, and “ (5) Troubleshooting for Daily
Variation” in section “3. TROUBLESHOOTING” in Part I.
II-46
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
E (O-ring)
II-47
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Three O-rings are attached to one side of three electrodes. Be careful not to lose the
O-rings when mounting the electrodes.
CAUTION
II-48
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
To obtain the best possible analysis data, perform two successive calibration
measurements to confirm the electrode stability.
If the difference in factor value between the first and second calibrations is within the
CAUTION following values, the electrodes are stable.
• Difference between 1st and 2nd factors Na: 0.020, K: 0.045, Cl: 0.025
• If the difference between those factor values is not within each value shown
above, or if slope 0 is output at the first time, there may reside air inside the
electrode block. First, open the main cover and ISE cover. Next, from AU480
“Home” window select Menu List>Maintenance>User Maintenance>ISE
Maintenance>Maintenance to display the “ISE Maintenance: Maintenance
tab” window. Then select the check box of “ISE Maintenance”, select M I D /
R E F P r i m e and O K , and press the TA B L E R OTAT I O N /
D I A G button.
• If slope 0 is output in both calibrations, the electrodes may not be set correctly.
Repeat steps 1 to 15 in the above maintenance procedure.
Refer to “ (3) Troubleshooting for Calibration Error – 1 of 2 (3264, 3265, 3266, 3267, 3268, 3269)” and
“ (4) Troubleshooting for Concurrent Reproducibility Error” in section “3. TROUBLESHOOTING” in Part
I.
CI
Na
REF electrode packing
K Cap screw
REF solution tube G Electrode unit
REF electrode cord (green)
REF
II-49
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-50
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
The REF electrode tip can easily be inserted into the packing if it is damped with
deionized water.
TIP
12. After inserting the REF electrode tighten the cap screw by rotating it 1/4 to 1/2 turn, then secure
the REF electrode.
13. Connect the roller tube for waste liquid with the roller pump.
14. Reconnect REF electrode cord in place.
15. Remount the REF electrode block in place.
16. Move the lock lever to the clamp position.
17. Confirm that D ra i n - F l o w c e l l has been selected from the “Single Operation” buttons on
the “ISE Maintenance: Maintenance tab” window.
18. Press the TA B L E R OTAT I O N / D I A G button.
The two roller pumps are activated and solution flows into the individual roller tubes.
Press the TA B L E R OTAT I O N / D I A G button until the roller pumps are activated five
times or so. Verify that there are no air bubbles in the waste liquid tube on the REF electrode
block side.
Press the TA B L E R OTAT I O N / D I A G button again until no air bubbles are found in the
tube.
II-51
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
19. Select U p d a t e .
A confirmation window appears.
20. Select O K .
21. Close the ISE cover.
22. Close the main cover.
To prevent the REF electrode block from becoming deformed due to ultrasonic
cleaning, follow these precautions. If the REF electrode block has been deformed or
cracked, replace it.
CAUTION
• Do not perform ultrasonic cleaning for more than ten minutes.
Refer to section “10.1 Replace the REF Electrode and Packing” in section
“10. MAINTENANCE REQUIRED EVERY TWO YEARS OR 150,000 SAMPLES” in this chapter.
II-52
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
• When the aspiration tube is removed from the tank, the solution in the tube may
drip in the system. If this occurs wipe it off with a dry cloth or paper towel wearing
protective gloves.
CAUTION
• Do not handle the aspiration tube with bare hands. Reagents will become
contaminated, resulting in defective data.
4. Place the aspiration tube in a new bottle, tighten the cap, and place it on the instrument.
Dispose of the old solution as liquid waste is handled.
II-53
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
• Push the reagent tank all the way until it stops. If the reagent tank is positioned forward,
the liquid level sensor will not detect the reagent level. This may display an alarm message
indicating there is insufficient reagent.
CAUTION
• If the roller tube for the MID solution is removed from the roller pump, the solution in the
tubing flows backwards into each reagent tank. Do not remove the roller tube for the MID
solution from the roller pump when replacing either of the reagent tanks.
7. Select one of following maintenance item from the list of maintenance items.
II-54
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Fixing screws
Syringe case
Mounting groove
Piston fixing screw
R Syringe: ZM0112
When removing the fixing screws from the syringe case do not use excessive force,
or the syringe case may be damaged.
CAUTION
6. While holding the case head and syringe case, turn the syringe case counterclockwise to remove
it.
II-55
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Syringe case
If the O-ring remains in the case head, carefully remove the O-ring with tweezers. Be careful not
to scratch the case head.
• Set the R syringe in place. The S syringe and R syringe are identified with the piston shaft
diameter.
Piston shaft diameter: D
CAUTION
D = 5mm: R syringe
D = 2mm: S syringe
• Do not pull out the piston from a new syringe. Once the piston has been pulled out, the
syringe cannot maintain the dispensing accuracy even if the piston is inserted in the
original position.
Case head
Syringe case
Piston fixing screw
II-56
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
12. From the AU480 “Home” window select Menu List>Maintenance>User Maintenance>ISE
Maintenance>Maintenance to display the “ISE Maintenance: Maintenance tab” window.
If air bubbles get in the syringe, the bubbles will affect the dispense amount of buffer solution
and cause an error in data. If air bubbles have gotten in the syringe, perform B u f f e r
P r i m e until the air bubbles are fully eliminated.
CAUTION If air bubbles are not eliminated after the B u f f e r P r i m e is repeatedly performed, re-
exercise steps 8 to 16 in the above procedure.
Avoid getting AU Alkali wash solutions on the syringe case, be sure to wipe any excess off the
syringe case with a damp cloth.
CAUTION
II-57
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Materials:
• Pipette
When handling wash solution, wear protective gloves to avoid contact with the wash
solution.
If hands or clothing come into contact with the wash solution, wash them
CAUTION immediately with water. Should wash solution get splashed accidentally in the eyes
or mouth, rinse it immediately with water and consult a physician.
II-58
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Pipette
11. Select U p d a t e .
12. Select C a n c e l .
13. Repeat the procedure in steps 6 to 10.
II-59
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-60
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
Fixing screws
Syringe case
Mounting groove
Piston fixing screw
Procedure for replace the syringe case and the case head
Materials:
Case head
Fixing screws
Syringe case
4. Loosen the piston fixing screw located at the bottom of the syringe.
5. Loosen the lock nut located at the top of the syringe.
6. Pull the syringe case to the front to remove the syringe from the mounting grooves.
To avoid damaging the syringe case, do not apply excessive force to the fixing
screws when removing them.
CAUTION
7. While holding the case head and syringe case, turn the syringe case counterclockwise to remove
it.
II-61
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
Case head
O-ring
Syringe
ZM0112
Syringe case
If the O-ring remains in the case head, carefully take out the O-ring with tweezers be careful not
to scratch the case head.
• Set the syringe case and the case head for ISE.
• The syringe case and the case head are identified with their figures and color.
CAUTION
F
R
O
N
T
9. Insert the new R syringe for ISE into the new case head.
Avoid getting AU Alkali wash solutions on the syringe case and syringe head. The
strong alkali adhering to the syringe case and case head will cause cracks on them.
If the strong alkali has adhered to, remove the syringe case and case head and
CAUTION wash them with water.
II-62
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0
Return to Main Table of Contents
11. Insert the syringe into the mounting grooves by holding the case head. Tighten the lock nut while
holding the case head by hand.
Mounting groove
Lock nut
Fixing screws
Case head
Syringe case
Piston fixing screw
If air bubbles get in the syringe, it will affect the dispense amount of buffer solution
and cause a data error. If air bubbles have gotten in the syringe, perform B u f f e r
P r i m e until the air bubbles are fully eliminated.
CAUTION If air bubbles are not eliminated after the B u f f e r S o l u t i o n P r i m e is
repeatedly performed, repeat step 9 to 18 in the above procedure.
II-63
ISE User Guide ISE MAINTENANCE FLOW
Version 1.0
Return to Main Table of Contents
II-64
ISE MAINTENANCE FLOW ISE User Guide
Version 1.0