You are on page 1of 24

2011a SECTION V ARTICLE 8

• ARTICLE 8
MANDATORY APPENDICES

APPENDIX I - GLOSSARY OF TERMS APPENDIX II - EDDY CURRENT


FOR EDDY CURRENT EXAMINATION EXAMINATION OF
NONFERROMAGNETIC HEAT
1-810 SCOPE
EXCHANGER TUBING
This Mandatory Appendix is used for the purpose of
establishing standard terms and definitions of terms related 11-810 SCOPE
to eddy current examination, which appear in Article 8.
This Appendix provides the requirements for bobbin
coil, multifrequency, multiparameter, eddy current exami-
nation for installed nonferromagnetic heat exchanger tub-
1-820 GENERAL REQUIREMENTS ing, when this Appendix is specified by the referencing
Code Section.
(a) This standard terminology for nondestructive exami-
nation ASTM E 1316 has been adopted by the Committee
as SE-1316.

• (b) SE-1316, Section 6, Electromagnetic Testing, pro- 11-820 GENERAL


vides the definitions of terms listed in I-830(a).
This Appendix also provides the methodology for exam-
(c) For general terms, such as Interpretation, Flaw, Dis-
ining nonferromagnetic, heat exchanger tubing using the
continuity, Evaluation, etc., refer to Article 1, Mandatory
eddy current method and bobbin coil technique. By scan-
Appendix I.
ning the tubing from the boreside, information will be
(d) Paragraph I-830(b) provides a list of terms and defi-
obtained from which the condition of the tubing will be
nitions, which are in addition to SE-1316 and are Code
determined. Scanning is generally performed with a bobbin
specific.
coil attached to a flexible shaft pulled through tubing manu-
ally or by a motorized device. Results are obtained by
evaluating data acquired and recorded during scanning.
1-830 REQUIREMENTS


(a) The following SE-1316 terms are used in conjunc-
11-821 Written Procedure Requirements
tion with this Article: absolute coil, differential coils, eddy
current, eddy current testing, frequency, phase angle, 11-821.1 Requirements. Eddy current examinations
probe coil, reference standard, standard. shall be conducted in accordance with a written procedure
which shall contain, as a minimum, the requirements listed
(b) The following Code terms are used in conjunction
in Table 11-821. The written procedure shall establish a
with this Article.
single value, or range of values, for each requirement.
bobbin coil: for inspection of tubing, a bobbin coil is
defined as a circular inside diameter coil wound such that 11-821.2 Procedure Qualification. When procedure
the coil is concentric with a tube during examination. qualification is specified by the referencing Code Section,
text information: information stored on the recording a change of a requirement in Table 11-821 identified as an
media to support recorded eddy current data. Examples essential variable shall require requalification of the written
include tube and steam generator identification, operator's procedure by demonstration. A change of a requirement
name, date of examination, and results. identified as a nonessential variable does not require requal-

• unit of data storage: each discrete physical recording ification of the written procedure. All changes of essential
medium on which eddy current data and text information or nonessential variables from those specified within the
are stored. Examples include tape cartridge, floppy disk, written procedure shall require revision of, or an addendum
etc. to, the written procedure.

155
ARTICLE 8 2011a SECTION V

Tube material
TABLE 11-821
REQUIREMENTS FOR AN EDDY CURRENT EXAMINATION

Requirements as Applicable

Tube diameter and wall thickness


Essential
Variable

x
X
PROCEDURE
Nonessential
Variable

Mode of inspection - differential or absolute X
Probe type and size X
Length of probe cable and probe extension cables X
Probe manufacturer, part number, and description X
Examination frequencies, drive voltage, and gain settings X
Manufacturer and model of eddy current equipment X
Scanning direction during data recording, i.e., push or pull X


Scanning mode - manual, mechanized probe driver, remote X
controlled fixture
Fixture location verification X
Identity of calibration reference standard(s) X
Minimum digitization rate X
Maximum scanning speed during data recording X
Personnel requirements X
Data recording equipment manufacturer and model X
Scanning speed during insertion or retraction, no data recording X
Side of application - inlet or outlet X
Data analysis parameters X
Tube numbering X
Tube examination surface preparation X

11-822 Personnel Requirements

The user of this Appendix shall be responsible for


assigning qualified personnel to perform eddy current
examination in accordance with the requirements of this
Appendix and the referencing Code Section.
11-830.2 Analog Data Acquisition
11-830.2.1 Analog Eddy Current
System
Instrument
(a) The frequency response of the outputs from the eddy
current instrument shall be constant within 2% of full scale
from dc to Fmax, where Fmax (Hz) is equal to 10 (Hz-s/in.)

[0.4 (Hz-s/mm)] times maximum probe travel speed
(in.lsec) (mm/s).
11-830 EQUIPMENT (b) Eddy current signals shall be displayed as two-
11-830.1 Data Acquisition System dimensional patterns by use of an X -Y storage oscilloscope
11-830.1.1 Multifrequency-Multiparameter or equivalent.


Equipment. The eddy current instrument shall have the (c) The frequency response of the instrument output
capability of generating multiple frequencies simultane- shall be constant within 2% of the input value from dc
ously or multiplexed and be capable of multiparameter to Fmax, where Fmax (Hz) is equal to 10 (Hz-s/in.) [0.4
signal combination. In the selection of frequencies, consid- (Hz-s/mm)] times maximum probe travel speed.
eration shall be given to optimizing flaw detection and
11-830.2.2 Magnetic Tape Recorder
characterization.
(a) The magnetic tape recorder used with the analog
(a) The outputs from the eddy current instrument shall
equipment shall be capable of recording and playing back
provide phase and amplitude information.
eddy current signal data from all test frequencies and shall
(b) The eddy current instrument shall be capable of
have voice logging capability.
operating with bobbin coil probes in the differential mode
or the absolute mode, or both. (b) The frequency response of the magnetic tape
(c) The eddy current system shall be capable of real recorder outputs shall be constant within 10% of the input
time recording and playing back of examination data. value from dc to Fmax, where Fmax (Hz) is equal to 10


(d) The eddy current equipment shall be capable of (Hz-s/in.) [0.4 (Hz-s/mm)] times maximum probe travel
detecting and recording dimensional changes, metallurgi- speed.
cal changes and foreign material deposits, and responses (c) Signal reproducibility from input to output shall be
from imperfections originating on either tube wall surface. within 5%.

156
2011a SECTION V ARTICLE 8

• 11-830.2.3 Strip Chart Recorder


(a) Strip chart recorders used with analog equipment
shall have at least 2 channels.
(b) The frequency response of the strip chart recorder
shall be constant within 20% of full scale from dc to Fmax,
where Fmax (Hz) is equal to 10 (Hz-s/in.) [0.4 (Hz-s/mm)]
(b) The system shall have multiparameter mixing capa-
bility.
(c) The system shall be capable of maintaining the iden-
tification of each tube recorded.
(d) The system shall be capable of measuring phase
angles in increments of one degree or less.
times maximum probe travel speed. (e) The system shall be capable of measuring ampli-
11-830.3 Digital Data Acquisition System tudes to the nearest 0.1 volt.
11-830.3.1 Digital Eddy Current Instrument 11-830.6 Analog Data Analysis System
(a) At the scanning speed to be used, the sampling rate 11-830.6.1 Display. Eddy current signals shall be dis-
of the instrument shall result in a minimum digitizing rate played as Lissajous patterns by use of an X- Y storage


of 30 samples per in. (25 mm) of examined tubing, use display oscilloscope or equivalent. The frequency response
dr = srlss, where dr is the digitizing rate in samples per of the display device shall be constant within 2% of the
in., sr is the sampling rate in samples per sec or Hz, and input value from dc to Fmax, where Fmax (Hz) is equal to
ss is the scanning speed in in. per sec. 10 (Hz-s/in.) [0.4 (Hz-s/mm)] times maximum probe travel
(b) The digital eddy current instrument shall have a speed.
minimum resolution of 12 bits per data point.
11-830.6.2 Recording System
(c) The frequency response of the outputs of analog
(a) The magnetic tape recorder shall be capable of play-
portions of the eddy current instrument shall be constant
ing back the recorded data.
within 2% of the input value from dc to Fmax, where Fmax
(Hz) is equal to 10 (Hz-s/in.) [0.4 (Hz-s/mm)] times maxi- (b) The frequency response of the magnetic tape
mum probe travel speed. recorder outputs shall be constant within 10% of the input
(d) The display shall be selectable so that the examina-
value from dc to Fmax, where Fmax (Hz) is equal to 10
(Hz-s/in.) [0.4 (Hz-s/mm)] times maximum probe travel
tion frequency or mixed frequencies can be presented as


a Lissajous pattern. speed (in.ls) (mm/s).
(e) The Lissajous display shall have a minimum resolu- (c) Signal reproducibility input to output shall be
tion of 7 bits full scale. within 5%.
(f) The strip chart display shall be capable of displaying 11-830.7 Digital Data Analysis System
at least 2 traces. 11-830.7.1 Display
(g) The strip chart display shall be selectable so either
(a) The analysis display shall be capable of presenting
the X or Y component can be displayed. recorded eddy current signal data and test information.
(h) The strip chart display shall have a minimum resolu-
(b) The analysis system shall have a minimum resolu-
tion of 6 bits full scale.
tion of 12 bits per data point.
11-830.3.2 Digital Recording System (c) The Lissajous pattern display shall have a minimum
(a) The recording system shall be capable of recording resolution of 7 bits full scale.
and playing back all acquired eddy current signal data from (d) The strip chart display shall be selectable so either


all test frequencies. the X or Y component of any examination frequency or
(b) The recording system shall be capable of recording mixed frequencies can be displayed .
and playing back text information. (e) The strip chart display shall have a minimum resolu-
(c) The recording system shall have a minimum resolu- tion of 6 bits full scale.
tion of 12 bits per data point.
11-830.7.2 Recording System
11-830.4 Bobbin Coils (a) The recording system shall be capable of playing
11-830.4.1 General Requirements back all recorded eddy current signal data and test infor-
(a) Bobbin coils shall be able to detect artificial disconti~ mation.
nuities in the calibration reference standard. (b) The recording system shall have a minimum resolu-
(b) Bobbin coils shall have sufficient bandwidth for tion of 12 bits per data point.
operating frequencies selected for flaw detection and
11-830.8 Hybrid Data Analysis System
sizing.
(a) Individual elements of hybrid systems using both


11-830.5 Data Analysis System digital elements and some analog elements shall meet spe-
11-830.5.1 Basic System Requirements cific sections of 11-830, as applicable.
(a) The data analysis system shall be capable of dis- (b) When analog to digital or digital to analog convert-
playing eddy current signal data from all test frequencies. ers are used, the frequency response of the analog element

157
ARTICLE 8 2011a SECTION V

outputs shall be constant within 5% of the input value from


dc to Fmax, where Fmax (Hz) is equal to 10 (Hz-s/in.) [0.4
(Hz-s/mm)] times maximum probe travel speed.

11.840 REQUIREMENTS
if) amplification for all channels of the eddy current
instrument shall be within 5% of the mean value, at all
sensitivity settings, at any single frequency
(g) the two output channels of the eddy current instru-
ment shall be orthogonal within 3 deg at the examination
frequency

11.840.1 Recording and Sensitivity Level 11-860.1.2 Digital Equipment. Analog elements of
(a) The eddy current signal data from all test frequencies digital equipment shall be calibrated in accordance with
shall be recorded on the recording media as the probe 11-860.1.1. Digital elements need not be calibrated.
traverses the tube.
(b) The sensitivity for the differential bobbin coil tech- 11-860.2 Calibration Reference Standards
nique shall be sufficient to produce a response from the 11-860.2.1 Calibration Reference Standard
through-wall hole(s) with a minimum vertical amplitude
of 50% of the full Lissajous display height.

11-840.2 Probe Traverse Speed. The traverse speed


shall not exceed that which provides adequate frequency
response and sensitivity to the applicable calibration dis-
continuities. Minimum digitization rates must be main-
Requirements. Calibration reference standards shall con-
form to the following:
(a) Calibration reference standards shall be manufac-
tured from tube(s) of the same material specification and
nominal size as that to be examined in the vessel.
(b) Tubing calibration reference standard materials heat

tained at all times. treated differently from the tubing to be examined may
be used when signal responses from the discontinuities
11-840.3 Fixture Location Verification described in 11-860.2.2 are demonstrated to the Inspector
(a) The ability of the fixture to locate specific tubes to be equivalent in both the calibration reference standard
shall be verified visually and recorded upon installation of and tubing of the same heat treatment as the tubing to be
the fixture and before relocating or removing the fixture. examined.


Independent position verification, e.g., specific landmark (c) As an alternative to 11-860.2.l(a) and (b), calibration
location, shall be performed and recorded at the beginning reference standards fabricated from UNS Alloy N06600
and end of each unit of data storage of the recording media. shall be manufactured from a length of tubing of the same
(b) When the performance of fixture location reveals material specification and same nominal size as that to be
that an error has occurred in the recording of probe verifi- examined in the vessel.
cation location, the tubes examined since the previous loca- (d) Artificial discontinuities in calibration reference
tion verification shall be reexamined. standards shall be spaced axially so they can be differenti-
11-840.4 Automated Data Screening System. When ated from each other and from the ends of the tube. The
automated eddy current data screening systems are used, as-built dimensions of the discontinuities and the applica-
each system shall be qualified in accordance with a written ble eddy current equipment response shall become part of
procedure. the permanent record of the calibration reference standard.
(e) Each calibration reference standard shall be perma-
nently identified with a serial number.
11-860 CALIBRATION
11-860.1 Equipment Calibration
11-860.1.1 Analog Equipment
The following shall be verified by annual calibration:
(a) the oscillator output frequency to the drive coil shall
be within 5% of its indicated frequency
11-860.2.2 Calibration Reference Standards for
Differential and Absolute Bobbin Coils
(a) Calibration reference standards shall contain the fol-
lowing artificial discontinuities:
(1) One or four through-wall holes as follows:
(a) A 0.052 in. (1.3 mm) diameter hole for tubing

(b) the vertical and horizontal linearity of the cathode with diameters of 0.750 in. (19 mm) and less, or a 0.067
ray tube (CRT) display shall be within 10% of the deflec- in. (1.70 mm) hole for tubing with diameters greater than
tion of the input voltage 0.750 in. (19 mm).
(c) the CRT vertical and horizontal trace alignment shall (b) Four holes spaced 90 deg apart in a single
be within 2 deg of parallel to the graticule lines plane around the tube circumference, 0.026 in. (0.65 mm)
(d) the ratio of the output voltage from the tape recorder diameter for tubing with diameters of 0.750 in. (19 mm)


shall be within 5% of the input voltage for each channel and less and 0.033 in. (0.83 mm) diameter for tubing with
of the tape recorder diameters greater than 0.750 in. (19 mm).
(e) the chart speed from the strip chart recorder shall (2) A flat-bottom hole 0.109 in. (2.7 mm) diameter,
be within 5% of the indicated value 60% through the tube wall from the outer surface.

158
20lla SECTION V ARTICLE 8

• FIG.II-860.3.1

50%
DIFFERENTIAL TECHNIQUE
RESPONSE FROM CALIBRATION REFERENCE
STANDARD
FIG.II-860.3.2 ABSOLUTE TECHNIQUE RESPONSE
FROM CALIBRATION REFERENCE STANDARD

100% through-wall
hole\'oon"
50%

25%
25%
20% flat bottom
/ hole response
•..
•..
.c
.c
Cl


Cl 'iii
J:
"iii
J: o c:
c:
Q)
~
0
I
1.0. groove response
\
Probe motion axis lJ)
Q)

~
u
u
lJ)

25%
25%

100% through-
wall hole response 50%
50% Screen Width
Screen Width

• (3) Four flat-bottom holes 0.187 in. (5 mm) diameter,


spaced 90 deg apart in a single plane around the tube
circumference, 20% through the tube wall from the outer
surface.
(b) The depth of the artificial discontinuities, at their
the instrument and shall be distinguishable from each other
as well as from probe motion signals.
11-860.3.2 Absolute Bobbin Coil Technique
(a) The sensitivity shall be adjusted to produce a mini-
center, shall be within 20% of the specified depth or mum origin-to-peak signal of 2 volts from the four 20%
0.003 in. (0.08 rom), whichever is less. All other dimen- flat-bottom holes or 3 volts from the four through-wall
sions shall be within 0.003 in. (0.08 rom). drilled holes.
(c) All artificial discontinuities shall be sufficiently sep- (b) Adjust the phase or rotation control so that the signal
arated to avoid interference between signals, except for response due to the through-wall hole forms up and to the
the holes specified in 11-860.2.2(a)(l)(b) and (a)(3). left as the probe is withdrawn from the calibration reference

• 11-860.3 Analog System Set-up and Adjustment


11-860.3.1 Differential Bobbin Coil Technique
(a) The sensitivity shall be adjusted to produce a mini-
mum peak-to-peak signal of 4 volts from the four 20%
flat-bottom holes or 6 volts from the four through-wall
drilled holes.
standard holding the signal response from the probe motion
horizontal. See Fig. 11-860.3.2.
(c) Withdraw the probe through the calibration refer-
ence standard at the nominal examination speed. Record the
responses of the applicable calibration reference standard
discontinuities. The responses shall be clearly indicated by
the instrument and shall be distinguishable from each other
(b) The phase or rotation control shall be adjusted so
as well as from probe motion signals.
the signal response due to the through-wall hole forms
down and to the right first as the probe is withdrawn from 11-860.4 Digital System Off-Line Calibration. The
the calibration reference standard holding the signal eddy current examination data is digitized and recorded
response from the probe motion horizontal. See Fig. 11- during scanning for off-line analysis and interpretation.
860.3.1. The system set-up of phase and amplitude settings shall be

• (c) Withdraw the probe through the calibration refer-


ence standard at the nominal examination speed. Record the
responses of the applicable calibration reference standard
discontinuities. The responses shall be clearly indicated by

159
performed off-line by the data analyst. Phase and amplitude
settings shall be such that the personnel acquiring the data
can clearly discern that the eddy current instrument is
working properly.
ARTICLE 8 20t1a SECTION V

100

90
FIG.II-880

I
I
I
FLAW DEPTH AS A FUNCTION OF PHASE ANGLE AT 400 kHz
[Ni-Cr-Fe 0.050 in. (1.24 mm) WALL TUBE]

,'~

I ~
80 I
I
(il I
til
0) 70
I ~
c I
-".~ I
I
"'"
.J:: I
I- 60 I ""
I
ctl I
"-


~ 50
I

I
~ I

•..
.J:: I ~
40
0-
0)
Cl
~
I
I
I
I

" "'-
ctl
30 I
u:: I

I
I
~
20 I '\
I

I
I
I\.
10 I
I
I
I ~
o
o 20 40 60 80 100 120 140 160 180
Phase Angle (deg From Left Horizontal Axis)

11.860.4.1 System Calibration Verification


(a) Calibration shall include the complete eddy current
examination system. Any change of probe, extension
cables, eddy current instrument, recording instruments, or
any other parts of the eddy current examination system
correlating the signal amplitude or phase with the indica-
tion depth shall be based on the basic calibration standard
or other representative standards that have been qualified.
This shall be accomplished by using curves, tables, or
software. Figure II-880 illustrates the relationship of phase

hardware shall require recalibration. angle versus flaw depth for a nonferromagnetic thin-walled
(b) System calibration verification shall be performed tube examined at a frequency selected to optimize flaw
and recorded at the beginning and end of each unit of data resolution.
storage of the recording media. 11.880.3 Frequencies Used for Data Evaluation. All
(c) Should the system be found to be out of calibration indications shall be evaluated. Indication types, which must
(as defined in II-860.3), the equipment shall be recalibrated. be reported, shall be characterized using the frequencies


The recalibration shall be noted on the recording. All tubes or frequency mixes that were qualified.
examined since the last valid calibration shall be reex-
amined.
11.890 DOCUMENTATION
11.890.1 Reporting
11.870 EXAMINATION
11.890.1.1 Criteria. Indications reported in accor-
Data shall be recorded as the probe traverses the tube.
dance with the requirements of this Appendix shall be
described in terms of the following information, as a
minimum:
11.880 EVALUATION
(a) location along the length of the tube and with respect
11.880.1 Data Evaluation. Data shall be evaluated in to the support members
accordance with the requirements of this Appendix. (b) depth of the indication through the tube wall, when
11.880.2 Means of Determining Indication Depth.
For indication types that must be reported in terms of depth,
a means of correlating the indication depth with the signal
amplitude or phase shall be established. The means of

160
required by this Appendix
(c) signal amplitude
(d) frequency or frequency mix from which the indica-
tion was evaluated •
2011a SECTION V ARTICLE 8

• 11-890.1.2 Depth. The maximum evaluated depth of


flaws shall be reported in terms of percentage of tube wall
loss. When the loss of tube wall is determined by the
analyst to be less than 20%, the exact percentage of tube
wall loss need not be recorded, i.e., the indication may be
reported as being less than 20%.
eddy current examination system or part thereof .
(b) The report shall include a record indicating the tubes
examined (this may be marked on a tubesheet sketch or
drawing), any scanning limitations, the location and depth
of each reported flaw, and the identification and certifica-
tion level of the operators and data evaluators that con-
11-890.1.3 Non-Quantifiable Indications. A non- ducted each examination or part thereof.
quantifiable indication is a reportable indication that cannot (c) Tubes that are to be repaired or removed from ser-
be characterized. The indication shall be considered a flaw vice, based on eddy current examination data, shall be
until otherwise resolved. identified.
11-890.1.4 Support Members 11-890.2.4 Record Retention. Records shall be


11-890.1.4.1 Location of Support Members. The maintained in accordance with requirements of the refer-
location of support members used as reference points for encing Code Section.
the eddy current examination shall be verified by fabrica-
tion drawings or the use of a measurement technique.
11-890.2 Records APPENDIX III - EDDY CURRENT
11-890.2.1 Record Identification. The recording EXAMINATION ON COATED
media shall contain the following information within each FERRITIC MATERIALS
unit of data storage: ill-810 SCOPE
(a) Owner
(b) plant site and unit (a) This Appendix provides the eddy current examina-
(c) heat exchanger identification tion methodology and equipment requirements applicable
(d) data storage unit number for performing eddy current examination on coated ferritic
materials.


(e) date of examination
(b) Article 1, General Requirements, also applies when
(f) serial number of the calibration standard
eddy current examination of coated ferritic materials is
(g) operator's identification and certification level
required. Requirements for written procedures, as specified
(h) examination frequency or frequencies
in Article 8, shall apply, as indicated.
(i) mode of operation including instrument sample rate,
(c) SD-1186, Standard Test Methods for Nondestructive
drive voltage, and gain settings
Measurement of Dry Film Thickness of Nonmagnetic
(j) lengths of probe and probe extension cables
Coatings Applied to a Ferrous Base, may be used to develop
(k) size and type of probes
a procedure for measuring the thickness of nonmagnetic
(I) probe manufacturer's name and manufacturer's part
and conductive coatings.
number or probe description and serial number
(m) eddy current instrument serial number
(n) probe scan direction during data acquisition ill-820 GENERAL
(0) application side - inlet or outlet


111-821 Personnel Qualification
(p) slip ring serial number, as applicable
(q) procedure identification and revision The user of this Appendix shall be responsible for
11-890.2.2 Tube Identification assigning qualified personnel to perform eddy current
(a) Each tube examined shall be identified on the appli-
examination in accordance with requirements of this
cable unit of data storage and Appendix and the referencing Code Section.
(b) The method of recording the tube identification shall
correlate tube identification with corresponding recorded 111-822 Written Procedure Requirements
tube data.
The requirements of IV-823 shall apply. The type of
11-890.2.3 Reporting coating and maximum coating thickness also shall be essen-
(a) The Owner or his agent shall prepare a report of tial variables.
the examinations performed. The report shall be prepared,
filed, and maintained in accordance with the referencing


111-823 Procedure Demonstration
Code Section. Procedures and equipment used shall be
identified sufficiently to permit comparison of the examina- The procedure shall be demonstrated to the satisfaction
tion results with new examination results run at a later of the Inspector in accordance with requirements of the
date. This shall include initial calibration data for each referencing Code Section.

161
ARTICLE 8 2011a SECTION V

111-830 EQUIPMENT
The eddy current system shall include phase and ampli-
tude display.
the qualification flaws through the maximum coating thick-
ness regardless of flaw orientation (e.g., perpendicular,
parallel, or skewed to the scan direction). The signal ampli-
tude from each qualification flaw in the coated qualification
specimen shall be at least 50% of the signal amplitude
measured on the corresponding qualification flaw prior to

111.850 TECHNIQUE
coating.
The performance of examinations shall be preceded by
measurement of the coating thickness in the areas to be
examined. If the coating is nonconductive, an eddy current 111-870 EXAMINATION
technique may be used to measure the coating thickness.
If the coating is conductive, a magnetic coating thickness (a) Prior to the examination, all loose, blistered, flaking,


technique may be used in accordance with SD-ll86. Coat- or peeling coating shall be removed from the examina-
ing thickness measurement shall be used in accordance tion area.
with the equipment manufacturer's instructions. Coating (b) When conducting examinations, areas of suspected
thickness measurements shall be taken at the intersections flaw indications shall be confirmed by application of
of a 2 in. (50 mm) maximum grid pattern over the area to another surface or volumetric examination method. It may
be examined. The thickness shall be the mean of three be necessary to remove the surface coating prior to per-
separate readings within 0.250 in. (6 mm) of each inter- forming the other examination.
section.

111-890 DOCUMENTATION
111.860 CALm RATION 111.891 Examination Report
(a) A qualification specimen is required. The material
The report of examination shall contain the following


used for the specimen shall be the same specification and
information:
heat treatment as the coated ferromagnetic material to be
(a) procedure identification and revision
examined. If a conductive primer was used on the material
(b) examination personnel identity and, when required
to be examined, the primer thickness on the procedure
by the referencing Code Section, qualification level
qualification specimen shall be the maximum allowed on
(c) date of examination
the examination surfaces by the coating specification. Plas-
(d) results of examination and related sketches or maps
tic shim stock may be used to simulate nonconductive
of rejectable indications
coatings for procedure qualification. The thickness of the
coating or of the alternative plastic shim stock on the (e) identification of part or component examined
procedure qualification specimen shall be equal to or
greater than the maximum coating thickness measured on 111.893 Record Retention
the examination surface.
(b) The qualification specimen shall include at least one Records shall be maintained in accordance with require-


crack. The length of the crack open to the surface shall ments of the referencing Code Section.
not exceed the allowable length for surface flaws. The
maximum crack depth in the base metal shall be between
0.020 in. and 0.040 in. (0.5 mm and 1.0 mm). In addition,
if the area of interest includes weld metal, a 0.020 in. APPENDIX IV - EXTERNAL COIL
(0.5 mm) maximum depth crack is required in an as-welded EDDY CURRENT EXAMINATION OF
and coated surface typical of the welds to be examined. TUBULAR PRODUCTS
In lieu of a crack, a machined notch of 0.010 in. (0.25 mm)
IV-810 SCOPE
maximum width and 0.020 in. (0.5 mm) maximum depth
may be used in the as-welded surface. This Appendix describes the method to be used when
(c) Examine the qualification specimen first uncoated performing eddy current examinations of seamless copper,
and then after coating to the maximum thickness to be copper alloy, and other nonferromagnetic tubular products.
qualified. Record the signal amplitudes from the qualifica- The method conforms substantially with the following
tion flaws.
(d) Using the maximum scanning speed, the maximum
scan index, and the scan pattern specified by the procedure,
the procedure shall be demonstrated to consistently detect

162
Standard listed in Article 26 and reproduced in Subsection
B: SE-243, Electromagnetic (Eddy Current) Testing of
Seamless Copper and Copper-Alloy Heat Exchanger and
Condenser Tubes. •
2011a SECTION V ARTICLE 8

• TABLE IV-823
REQUIREMENTS OF AN EXTERNAL COIL EDDY
CURRENT EXAMINATION PROCEDURE

Requirements (As Applicable)


Essential
Variable
Non-
Essential
Variable
written procedure by demonstration. A change of a require-
ment identified as a nonessential variable does not require
requalification of the written procedure. All changes of
essential or nonessential variables from those specified
within the written procedure shall require revision of, or
an addendum to, the written procedure.
Frequency(ies) x
Mode (differential/absolute) X
Minimum fill factor X IV-830 EQUIPMENT
Probe type X
Maximum scanning speed during data X Equipment shall consist of electronic apparatus capable
recording
of energizing the test coil or probes with alternating cur-
Material being examined X


Material size/dimensions
rents of suitable frequencies and shall be capable of sensing
X
Reference standard X the changes in the electromagnetic properties of the mate-
Equipment manufacturer/model X rial. Output produced by this equipment may be processed
Data recording equipment X so as to actuate signaling devices and/or to record examina-
Cabling (type and length) X tion data.
Acquisition software X
Analysis software X
Scanning technique X IV-831 Test Coils and Probes
Scanning equipmenUfixtures X
Tube scanning surface preparation X Test coils or probes shall be capable of inducing alternat-
ing currents into the material and sensing changes in the
electromagnetic characteristics of the material. Test coils
should be selected to provide the highest practical fill
IV-820 GENERAL
factor.


IV-821 Performance

Tubes may be examined at the finish size, after the final IV-832 Scanners
anneal or heat treatment, or at the finish size, prior to the
final anneal or heat treatment, unless otherwise agreed upon Equipment used should be designed to maintain the
between the supplier and the purchaser. The procedure material concentric within the coil, or to keep the probe
shall be qualified by demonstrating detection of discontinu- centered within the tube and to minimize vibration during
ities of a size equal to or smaller than those in the reference scanning. Maximum scanning speeds shall be based on
specimen described in IV-833. Indications equal to or the equipment's data acquisition frequency response or
greater than those considered reportable by the procedure digitizing rate, as applicable.
shall be processed in accordance with IV -880.
IV.833 Reference Specimen (a)
IV.822 Personnel Qualification


The reference specimen material shall be processed in
The user of this Appendix shall be responsible for the same manner as the product being examined. It shall
assigning qualified personnel to perform eddy current be the same nominal size and material type (chemical
examination in accordance with requirements of this composition and product form) as the tube being examined.
Appendix and the referencing Code Section. Ideally, the specimen should be a part of the material
being examined. Unless specified in the referencing Code
Section, the reference discontinuities shall be transverse
IV -823 Written Procedure Requirements notches or drilled holes as described in Standard Practice
IV.823.1 Requirements. Eddy current examinations SE-243, Section 8, Reference Standards.
shall be performed in accordance with a written procedure,
which shall contain, as a minimum, the requirements listed
in Table IV-823. The written procedure shall establish a IV-850 TECHNIQUE
single value, or range of values, for each requirement.
Specific techniques may include special probe or coil

• IV -823.2
qualification
a change of
an essential
Procedure Qualification. When procedure
is specified by the referencing Code Section,
a requirement in Table IV -823 identified as
variable shall require requalification of the

163
designs, electronics, calibration standards, analytical algo-
rithms and/or display software. Techniques, such as chan-
nel mixes, may be used as necessary to suppress signals
produced at the ends of tubes. Such techniques shall be
ARTICLE 8 201la SECTION V

in accordance with requirements of the referencing Code


Section.

IV-860
IV-861
CALIBRATION
Performance Verification
(a) tube material specification, diameter, and wall thick-
ness condition
(b) coil or probe manufacturer, size and type
(c) mode of operation (absolute, differential, etc.)
(d) examination frequency or frequencies
(e) manufacturer, model, and serial number of eddy cur-

Performance of the examination equipment shall be veri- rent equipment
fied by the use of the reference specimen as follows: (j) scanning speed
(a) As specified in the written procedure (g) procedure identification and revision
(1) at the beginning of each production run of a given (h) calibration standard and serial number
diameter and thickness of a given material (i) identity of examination personnel, and, when
required by the referencing Code Section, qualification


(2) at the end of the production run
(3) at any time that malfunctioning is suspected level
(b) If, during calibration or verification, it is determined (j) date of examination
that the examination equipment is not functioning properly, (k) list of acceptable material
all of the product tested since the last calibration or verifi- (I) date of procedure qualification
cation shall be reexamined. (m) results of procedure requalification (as applicable)
(c) When requalification of the written procedure as
required in IV-823.2. IV-893 Record Retention
Records shall be maintained in accordance with require-
IV -862 Calibration of Equipment
ments of the referencing Code Section.
(a) Frequency of Calibration. Eddy current instrumen-
tation shall be calibrated at least once a year, or whenever
the equipment has been subjected to a major electronic


APPENDIX V - EDDY CURRENT
repair, periodic overhaul, or damage. If equipment has not
MEASUREMENT OF
been in use for a year or more, calibration shall be done
NONCONDUCTIVE-NONMAGNETIC
prior to use.
(b) Documentation. A tag or other form of documenta-
COATING THICKNESS ON A
tion shall be attached to the eddy current equipment with NONMAGNETIC METALLIC
dates of the calibration and calibration due date. MATERIAL
V-810 SCOPE

IV-870 EXAMINATION This Appendix provides requirements for absolute sur-


face probe measurement of nonconductive-nonmagnetic
Tubes are examined by passing through an encircling
coating thickness on a nonmagnetic metallic material.
coil, or past a probe coil with the apparatus set up in
accordance with the written procedure. Signals produced


by the examination are processed and evaluated. Data may V-820 GENERAL
be recorded for post-examination analysis or stored for
archival purposes in accordance with the procedure. Out- This Appendix provides a technique for measuring non-
puts resulting from the evaluation may be used to mark conductive-nonmagnetic coating thicknesses on a nonmag-
and/or separate tubes. netic metallic substrate. The measurements are made with
a surface probe with the lift -off' calibrated for thickness
from the surface of the test material. Various numbers of
IV-880 EVALUATION thickness measurements can be taken as the probe's spacing
Evaluation of examination results for acceptance shall from the surface is measured. Measurements can be made
be as specified in the written procedure and in accordance with various types of instruments.
with the referencing Code Section.
V-821 Written Procedure Requirements
V -821.1 Requirements. Eddy current examination


IV -890 DOCUMENTATION
shall be performed in accordance with a written procedure
IV -891 Examination Reports that shall, as a minimum, contain the requirements listed
A report of the examination shall contain the following in Table V-821. The written procedure shall establish a
information: single value, or range of values, for each requirement.

164
2011a SECTION V ARTICLE 8

• V.821.2 Procedure Qualificationffechnique Valida.


tion. When procedure qualification is specified by the refer-
encing Code Section, a change of a requirement in Table
V -821 identified as an essential variable shall require
requalification of the written procedure by demonstration.
A change of a requirement, identified as a nonessential
shall be adequate for the material and the coating thickness
range.

V.831 Probes
The eddy current absolute probe shall be capable of
inducing alternating currents into the material and sensing
variable, does not require requalification of the written
changes in the separation (lift-off) between the contact
procedure. All changes of essential or nonessential vari-
surface of the probe and the substrate material.
ables from those specified within the written procedure
shall require revision of, or an addendum to, the written
procedure. V.850 TECHNIQUE
A single frequency technique shall be used with a suit-


able calibration material such as nonconductive shim(s),
V.822 Personnel Qualification
paper, or other nonconductive nonmagnetic material. The
The user of this Appendix shall be responsible for shims or other material thicknesses shall be used to corre-
assigning qualified personnel to perform eddy current late a position on the impedance plane or meter reading
examination in accordance with requirements of this with the nonconductive material thicknesses and the no
Appendix and the referencing Code Section. thickness position or reading when the probe is against the
bare metal. If the thickness measurement is used only to
assure a minimum coating thickness, then only a specimen
V.823 Procedureffechnique Demonstration representing the minimum thickness need be used.
The procedure/technique shall be demonstrated to the
satisfaction of the Inspector in accordance with the require- V.860 CALmRATION
ments of the referencing Code Section. The probe frequency and gain settings shall be selected


to provide a suitable and repeatable examination. The probe
shall be nulled on the bare metal.
V.830 EQUIPMENT (a) Impedance Plane Displays. For instruments with
impedance plane displays, gains on the vertical and hori-
The eddy current instrument may have a storage type
zontal axes shall be the same value. The phase or rotation
display for phase and amplitude or it may contain an analog
control and the gain settings shall be adjusted so that the
or digital meter. The frequency range of the instrument
bare metal (null) and the air point are located at diagonally
opposite comers of the display. A typical coating thickness
calibration curve is illustrated in Fig. V-860.
TABLE V-821 (b) Meter Displays. For instruments with analog meter
REQUIREMENTS OF AN EDDY CURRENT displays, the phase and gain controls shall be used to pro-
EXAMINATION PROCEDURE FOR THE MEASUREMENT vide near full scale deflection between the bare metal and
OF NONCONDUCTIVE-NONMAGNETIC maximum coating thickness.


COATING THICKNESS ON A METALLIC MATERIAL
(c) All Instruments. For all instruments, the difference
in meter readings or thickness positions on the screen shall
be adequate to resolve a 10% change in the maximum
Requirement
thickness.
Examination frequency (d) Calibration Data. The screen positions or meter
Absolute mode readings and the shim thicknesses shall be recorded along
Size and probe type(s), manufacturer's with the bare metal position or meter reading.
name and description
Substrate material
(e) Verification of Calibration. Calibration readings
Equipment manufacturer/model shall be verified every two hours. If, during recalibration,
Cabling (type and length) a reading representing a coating thickness change greater
Nonconductive calibration material than :tl0% from the prior calibration is observed, examina-
(nonconductive shims) tions made after the prior calibration shall be repeated.
Personnel qualification requirements


unique to this technique
Reference to the procedure V.870 EXAMINATION
qualification records
Examination surface preparation Coating thickness measurements shall be taken at indi-
vidual points as indicated in the referencing Code Section.

165
ARTICLE 8 2011a SECTION V

FIG. V-8bO TYPICAL LIFT-OFF

\ Air Point
CALIBRATION
CALIBRATION
CURVE FOR COATING THICKNESS
POINTS ALONG THE CURVE
SHOWING THICKNESS


\
\
\
~

~,
5 ""'-

4
~
- 2
.....
I-+--
1
Bare
Metal
Point

If it is desired to measure the minimum coating thickness


or maximum coating thickness on a surface, a suitable grid
pattern shall be established and measurements shall be
taken at the intersections of the grid pattern. Measurements
shall be recorded.
APPENDIX VI - EDDY CURRENT
DETECTION AND MEASUREMENT OF
DEPTH OF SURFACE
DISCONTINUITIES IN NONMAGNETIC
METALS WITH SURFACE PROBES
VI.810 SCOPE

V-880 EVALUATION This Appendix provides the requirements for the detec-
Coating thicknesses shall be compared with the accept- tion and measurement of depth for surface discontinuities
ance standards of the referencing Code Section. in nonmagnetic-metallic materials using an absolute sur-
face probe eddy current technique.


V.890 DOCUMENTATION
VI-820 GENERAL
V.891 Examination Report
This Appendix provides a technique for the detection
The report of the examination shall contain the following and depth measurement of cracks and other surface discon-
information: tinuities in nonmagnetic metal components. An absolute
(a) procedure identification and revision surface probe containing a single excitation coil is scanned
(b) examination personnel identity, and, when required over the surface of the examination object. When a surface
by the referencing Code Section, qualification level discontinuity is encountered by the magnetic field of the
(c) date of examination probe, eddy currents generated in the material change their
(d) results of examination and related sketches or maps flow and provide a different magnetic field in opposition
of thickness measurements to the probe's magnetic field. Changes in the eddy current's
(e) identification of part or component examined magnetic field and the probe's magnetic field are sensed
by the instrument and are presented on the instrument's

V-893 Record Retention


Records shall be maintained in accordance with require-
ments of the referencing Code Section.

166
impedance plane display. These instruments generally have
capability for retaining the signal on the instrument's dis-
play where any discontinuity signal can be measured and
compared to the calibration data. •
20118 SECTION V ARTICLE 8

• VI-821 Written
VI-821.1 Requirements.
Procedure Requirements
Eddy current examination
shall be performed in accordance with a written procedure
that shall, as a minimum, contain the requirements listed
in Table VI-821. The written procedure shall establish a
single value, or range of values, for each requirement.
VI-830 EQUIPMENT
The eddy current instrument may have a storage type
display for phase and amplitude on an impedance plane.
The frequency range of the instrument shall be adequate
to provide for a suitable depth of penetration for the mate-
rial under examination.
VI-821.2 Procedure Qualification. When procedure
qualification is specified by the referencing Code Section, VI-831 Probes
a change of a requirement in Table VI-821 identified as
an essential variable shall require requalification of the The eddy current absolute probe shall be capable of
written procedure by demonstration. A change of a require- inducing alternating currents into the material and sensing


ment identified as a nonessential variable does not require changes in the depth of the notches in the reference speci-
requalification of the written procedure. All changes of men. The probe and instrument at the frequency to be used
essential or nonessential variables from those specified in the examination shall provide a signal amplitude for the
within the written procedure shall require revision of, or smallest reference notch of a minimum of 10% full screen
an addendum to, the written procedure. height (FSH). With the same gain setting for the smallest
notch, the signal amplitude on the largest notch shall be a
minimum of 50% FSH. If the amplitudes of the signals
VI-822 Personnel Qualification cannot be established as stated, other probe impedances or
The user of this Appendix shall be responsible for geometries (windings, diameters, etc.) shall be used.
assigning qualified personnel to perform eddy current
examination in accordance with requirements of this VI.832 Reference Specimen
Appendix and the referencing Code Section.
A reference specimen shall be constructed of the same


alloy as the material under examination. Minimum dimen-
VI-823 Procedureffechnique Demonstration sions of the reference specimen shall be 2 in. (50 mm) by
The procedure/technique shall be demonstrated to the 4 in. (100 mm) and shall contain a minimum of two notches.
satisfaction of the Inspector in accordance with the require- Notch length shall be a minimum of 0.25 in. (6 mm) and
ments of the referencing Code Section. notch depth shall be the minimum to be measured and the
maximum depth allowed. If smaller length notches are
required to be detected by the referencing Code Section,
the reference specimen shall contain a smaller length notch
TABLE VI-821 meeting the referencing Code requirements. The depth
REQUIREMENTS OF AN EDDY CURRENT shall have a tolerance of + 10% and -20% of the required
EXAMINATION PROCEDURE FOR THE
dimensions. A typical reference specimen for measuring
DETECTION AND MEASUREMENT OF DEPTH FOR
flaw depths in the range of 0.01 in. (0.25 mm) through
SURFACE DISCONTINUITIES IN NONMAGNETIC
METALLIC MATERIALS
0.04 in. (1 mm) is shown in Fig. VI-832.


When curvature of the examination object in the area of
Non- interest is not flat and affects the lift-off signal, a reference
Essential Essential
Requirement
specimen representing that particular geometry with the
Variable Variable
applicable notches shall be used.
Examination frequency x
Size and probe type(s), manufacturer's X
name and description
VI-850 TECHNIQUE
Material X
Equipment manufacturer/model X A single frequency technique shall be used. The fre-
quency shall be selected to result in an impedance plane
Cabling (type and length) X
presentation that will result in a 90 deg phase shift between
Reference specimen and notch depths X
Personnel qualification, when required by X the lift-off signal and the flaw signals. The resulting signals
the referencing Code Section will be displayed using an impedance plane presentation
Personnel qualification requirements with one axis representing the lift-off signal and the other


X
unique to this technique axis representing the reference notch and flaw signal
Reference to the procedure qualification X
responses. The gain control on each axis displaying the
records
Examination surface preparation X flaw signals shall be adjusted to present amplitude for the
flaw signal from the deepest notch to be at least 50% of

167
ARTICLE 8 2011a SECTION V

FIG. VI-832 REFERENCE SPECIMEN


n~ 1 in. (25 mm) ~ n~ 1 in. (25 mm) ~ n~ 1 in. (25 ~
U-Typica~Typica~Typical


Typical Notch Depths
0,010 in. 0.020 in. 0.040 in.
[ (0.25 mm) r (0.5 mm) ,{ (1mm)
t ..f

GENERAL NOTES:
(a) Typical notch dimensions are 0.25 in. (6 mm) length x 0.010 in. (0.25 mm) width.
(b) Tolerances on notch dimensions are :tlO% for length and width, and +10% and -20% for depth.

the vertical or horizontal display it is presented on. Typical VI-870 EXAMINATION


responses of the calibrated instrument are shown in Fig.
The area of interest shall be scanned with overlap on
VI-8S0. Note that the display may be rotated to show these
the next scan to include at least 10% of the probe diameter.


indications in accordance with the procedure. Typically,
If the direction of suspected discontinuities are known, the
the gain setting on the axis displaying the discontinuity
scan direction shall be perpendicular to the long axis of the
signal will have a gain setting higher than the axis dis-
discontinuity. The object shall be scanned in two directions,
playing lift-off. Discontinuity indications will be mostly
90 deg to each other. During the examination, the maxi-
vertical or horizontal (at 90 deg to lift-off). Any surface
mum scanning speed and lift-off distance shall not be
discontinuities in the examination specimen would provide
greater than those used for calibration.
similar indications.

VI-860 CALIBRATION VI-880 EVALUATION

The probe frequency and gain settings shall be selected The discontinuity shall be scanned perpendicular to its
to provide a suitable depth of penetration within the mate- long axis to determine its maximum depth location and
rial so that the depth of the deepest notch is distinguishable value. The maximum depth of any discontinuity detected


from the next smaller notch. The gain settings on the verti- shall be compared with the appropriate response of the
cal and horizontal axis shall be set so that there is a dB reference specimen as specified in the referencing Code
difference with the discontinuity depth gain being higher. Section.
The probe shall be nulled on the bare metal away from
the notches. The X-Y position of the null point shall be
placed on one corner of the screen. The phase or rotation VI-890 DOCUMENTATION
control shall be adjusted so that when the probe is lifted VI-891 Examination Report
off the metal surface, the display point travels at 90 deg
to the discontinuity depth. Increase the vertical or hori- The report of the examination shall contain the following
zontal gain, as applicable, if the smallest indication or the information:
largest indication from the notches do not make 10% or (a) procedure identification and revision
50% FSH, respectively. Maximum response from the (b) examination personnel identity, and, when required
by the referencing Code Section, qualification level


notches is achieved when the probe is scanned perpendicu-
lar to the notch and centered on the notch. Differences in (c) date of examination
the vertical and horizontal gain may have to be adjusted. (d) results of examination and related sketches or maps
The screen indication lengths from the baseline (lift-off of indications exceeding acceptance standard
line) for each of the notch depths shall be recorded. (e) identification of part or component examined

168
20118 SECTION V ARTICLE 8

• FIG. VI-850 IMPEDANCE PLANE REPRESENTATIONS OF INDICATIONS FROM FIG. VI-832

Greater
lift-off

~
i
Increasing
notch
Increasing depth


notch
depth
~ Greater lift-off
Larger Horizontal Gain Larger Vertical Gain

(j) identification of reference specimen TABLE VII-823


(g) calibration results, minimum and maximum discon- REQUIREMENTS OF AN EDDY CURRENT SURFACE
tinuity depth measured EXAMINATION PROCEDURE
Non-
Essential Essential
VI-893 Record Retention Requirements (As Applicable) Variable Variable

• x
Records shall be maintained in accordance with require- Frequencies
ments of the referencing Code Section. Mode (differential/absolute) X
Probe type X
Maximum scanning speed X
Material being examined X
Material surface condition X
(a) APPENDIX VII - EDDY CURRENT Reference specimen material and X
EXAMINATION OF MAGNETIC AND simulated flaws
ET instrument manufacturer/model X
NONMAGNETIC CONDUCTIVE
Data presentation - display X
METALS TO DETERMINE IF FLAWS Cabling (type and length) X
ARE SURFACE-CONNECTED Use of saturation X
Analysis method X
VII-810 SCOPE Scanning technique X
Surface preparation X


This Appendix provides the requirements for using an
eddy current examination (ET) procedure to determine if
flaws are surface-connected (Le., open to the surface being VII-822 Personnel Qualification
examined). With appropriate selection of parameters, the
method is applicable to both magnetic and nonmagnetic The user of this Appendix shall be responsible for
conductive metals. assigning qualified personnel to perform eddy current
examination in accordance with requirements of this
Appendix or the referencing Code Section.
VII-820 GENERAL
VII-823 Written Procedure Requirements
VII-821 Performance VII-823.1 Requirements. Eddy current examinations
This Appendix provides requirements for the evaluation shall be performed in accordance with a written procedure,
of flaws, detected by other nondestructive examinations, which shall contain, as a minimum, the requirements listed

• utilizing a surface probe operating at a suitable test fre- in Table VII-823. The written procedure shall establish a
quency or combination of frequencies. The resultant phase single value or range of values, for each requirement.
and amplitude responses are used to determine if flaws are VII-823.2 Procedure Qualification. When procedure
surface-connected. qualification is specified by the referencing Code Section,

169
ARTICLE 8 2011a SECTION V

FIG. VII-830.5 EDDY CURRENT REFERENCE SPECIMEN

T •
1.5 in.
(38mm)

1

+----1 in.~L1 in.----+-~1 in.~L1 in.~L1 in.~
(25 mm) I (25 mm) (25 mm)1 (25 mm)1 (25 mm)

5 in. (125 mm)

I 0.010 in. 0.020 in.

_______
f 0.015 in.
(0.37 mm)
[0.004
__
in.
(0.1 mm)
I
(0.24 mm) I [ (0.5 mm)

___
I
~T ~T 4 s- l
11/4(6mm)
in.

DRAWING NOTES:
(a) Drawing not to scale.
(b) Typical notch length may vary from 1 in. (25 mm) to full block width. Full width notches will require welding at the ends or filling the notch
with epoxy to prevent block breakage.
(c) Maximum notch widths 0.010 in. (25 mm).
(d) Tolerance on notch bottoms +01-10% from the examination surface. •
(e) Block length, width, and thickness are as shown.
(f) Notch spacing and distance from ends of block are as shown.

a change of a requirement in Table VII-823 identified as VII-830.4 Instrumentation. The eddy current instru-
an essential variable shall require requalification of the ment shall be capable of driving the probes selected for
written procedure by demonstration. A change of a require- this examination with alternating current over a suitable
ment identified as a nonessential variable does not require range of frequencies. The eddy current instrument shall be
requalification of the written procedure. All changes of capable of sensing and displaying differences in phase and
essential or nonessential variables from those specified amplitude correlated to the depth of discontinuities. The


within the written procedure shall require revision of or instrument shall be capable of operating in either the abso-
an addendum to the written procedure. lute or differential mode. The persistence shall be adjusted
to display the phase and amplitude responses of the refer-
ence specimen notches and flaws in the material under
Vll.830 EQUIPMENT examination.
VII-830.1 System Description. The eddy current sys-
VII-830.5 Reference Specimen. The reference speci-
tem shall consist of an eddy current instrument, surface
probe, and cable connecting the instrument and the probe. men shall be constructed of the same alloy and product
form as the material being examined. The reference speci-
VII-830.2 Surface Probes. The eddy current probes men shall be as specified in Fig. VII-830.5. Calibration
shall be either differential or absolute type. They shall be references consist of two surface connected notches and
capable of inducing alternating currents in the material two bridged notches, representing both surface-connected
being examined and be capable of sensing changes in the and subsurface flaws.


resultant electromagnetic field. The specimen shall be a minimum of 5.0 in. (125 mm)
VII-830.3 Cables. Cables connecting the eddy current long, 1.5 in. (38 mm) wide, and ~ in. (6 mm) thick. Addi-
instrument and probes shall be designed and assembled to tional notches and bridged notches may be added and block
operate with these components. lengthened when additional information or higher precision

170
2011a SECTION V ARTICLE 8

• FIG. VII-862 IMPEDANCE PLANE RESPONSES FOR


STAINLESS STEEL (a) AND CARBON STEEL (b)
REFERENCE SPECIMENS

0.004 in.
Subsurface
notch
(0.1 mm) indications

0.015 in.
0.020 in.
(0.37 mm)
(0.5mm)


Lift-off
direction
0.010 in.
(0.25 mm)

Surface
connected
notch
indications

(a) Stainless Steel at Examination Frequency of 800 kHz

• 0.020 in.
(0.5mm)
Surface
connected
notch
indications

0.010 in.
(0.25 mm)

Lift-off


direction

0.004 in.
(0.1 mm)

0.015 in.
Subsurface
(0.37 mm)
notch
indications

(b) Carbon Steel at Examination Frequency of 800 kHz

• 171
ARTICLE 8 2011a SECTION V

is required. Surface conditions and finish of both the refer-


ence specimen and the material being examined shall be
similar.

VII-850 TECHNIQUE
response. The phase and amplitude of flaws and their loca-
tion will be recorded. During the examination the maxi-
mum scanning speed and lift-off distance shall not be
greater than those used for calibration. The surface finish
of areas scanned shall be comparable to the reference
specimen.

A single or multiple frequency technique may be used.
The frequency( s) shall be selected to result in an impedance
VII-880 EVALUATION
plane presentation of 90 deg to 180 deg phase shift between
the surface and subsurface notch indications. Discrimination of surface-connected flaw responses
from those of subsurface flaws shall be determined by
comparable phase and amplitude responses obtained from


VII-860 CALIBRATION similar surface-connected notches and subsurface, bridged
VII-861 General notches contained in the reference specimen.
The probe frequency(s) and gain settings shall be
selected to provide a suitable phase spread while providing VII-890 DOCUMENTATION
sufficient penetration to ensure that the shallowest subsur-
VII-891 Examination Report
face bridged notch indication is detected. Display gain of
the vertical and horizontal axis shall be set to provide equal The report of the examination shall contain the following
signal response. The ET instrument shall be adjusted to information:
rotate the phase for the lift-off response to be positioned at (a) procedure identification and revision
the 270 deg horizontal plane. Scanning shall be conducted (b) identification of examination personnel
perpendicular to the length of the notches. The gain shall (c) qualification of personnel, when required by the ref-
be set to display the 0.020 in. (0.5 mm) deep surface notch erencing Code Section


at 100% full screen height. At this gain setting, the 0.010 in. (d) date of examination
(0.24 mm) deep surface notch should be displayed at (e) identification of component or material examined
approximately 25% full screen height. The gain settings (j) scan plan including frequency(s) and gain
for these two reference notches may be accomplished on (g) flaw identity (e.g., surface-connected or not surface-
separate frequencies. Balancing the instrument will be con- connected)
ducted with the probe situated on the space between (h) identification and drawing of reference calibration
notches. Scanning speed shall be adjusted to allow the specimen
display to be formed for evaluation. The persistence of (i) calibration results (display) showing the indications
the screen shall be adjusted to allow a comparison of the of the bridged (subsurface) notches and surface notches
responses from each notch. The screen shall be cleared to detected
prevent the display to become overloaded. The presentation (j) ET equipment manufacturer, model, type, and serial
shall be balanced prior to making initial and final adjust- number
ments of phase and amplitude. Responses in terms of ampli- (k) probe manufacturer, model, type, and serial number


tude and phase angle resulting from scanning the surface (1) extension cable, if used, manufacturer, type, and
notches and notch bridges shall be recorded. length

VII-862 Calibration Response VII-892 Record Retention


Typical responses from carbon steel and stainless steel Records shall be maintained in accordance with require-
calibration specimens are shown in Fig. VIl-862. Note ments of the referencing Code Section.
that responses from magnetic materials and nonmagnetic
materials provide significantly different displays.
APPENDIX VIII - EDDY CURRENT (a)

VII-870 EXAMINATION EXAMINATION OF NONMAGNETIC


HEAT EXCHANGER TUBING
The flaw of interest shall be scanned with an overlap
on the adjacent scan to include approximately 50% of the VIII-810 SCOPE
probe diameter. Scanning shall be conducted perpendicular This Appendix provides the requirements for bobbin •
to the flaw length. The identity of the flaw will be deter- coil, multifrequency, multiparameter, eddy current exami-
mined from the phase and amplitude of the displayed nation for installed nonmagnetic heat exchanger tubing,

172
2011a SECTION V ARTICLE 8

• TABLE VIII-821
REQUIREMENTS FOR AN EDDY CURRENT EXAMINATION PROCEDURE

Requirements (As Applicable)

Tube material, size (outside diameter), wall thickness and


grade/temper
Essential
Variable

x
Nonessential
Variable

Mode of inspection - differential and/or absolute x


Probe type X
Probe manufacturer, part or serial number, and description X
Examination frequencies, drive voltage, and gain settings X
Manufacturer and model of eddy current equipment X
Maximum scanning speed X
Scanning mode - manual, mechanized probe driver, remote X


controlled fixture
Identity of calibration reference standard(s) including drawing X
Minimum digitization rate/samples per second X
Procedure qual ification X
Personnel qualifications X
Data recording equipment manufacturer and model X
Data analysis parameters X
Tube numbering X
Tube examination surface preparation X
Scanning equipment, extension cable, and fixtures X

excluding nuclear steam generator tubing, when this written procedure by demonstration. A change of a require-
Appendix is specified by the referencing Code Section. ment identified as a nonessential variable does not require


requalification of the written procedure. All changes of
essential or nonessential variables from those specified
VIII-820 GENERAL within the written procedure shall require revision of, or
an addendum to, the written procedure.
This Appendix also provides the methodology for exam-
ining nonferromagnetic heat exchanger tubing using the VIII-821.3 Personnel Requirements. The user of this
electromagnetic method known as near field eddy current Appendix shall be responsible for assigning qualified per-
testing (the coil that generates the magnetic field also senses sonnel to perform eddy current examination in accordance
changes in the magnetic field). The method may employ with requirements of the referencing Code Section.
one or more bobbin wound coils. By scanning the tubing
from the boreside, information will be obtained from which
the condition of the tubing will be determined. Scanning VIII-830 EQUIPMENT
is generally performed with the bobbin coil(s) attached to


a flexible shaft pulled through tubing manually or by a Vill-831 Data Acquisition System
motorized device. Results are obtained by evaluating data VIII-831.1 Multifrequency-Multiparameter Equip-
acquired and recorded during scanning. This Appendix ment. The eddy current instrument shall have the capability
does not address tubing with enhanced heat transfer sur- of generating multiple frequencies simultaneously or multi-
faces or saturation eddy current testing. plexed and be capable of multiparameter signal combina-
tion. In the selection of frequencies, consideration shall be
given to optimizing flaw detection and characterization.
VIII-821 Written Procednre Reqnirements
(a) The outputs from the eddy current instrument shall
VIII-821.1 Requirements. Eddy current examinations provide phase and amplitude information.
shall be conducted in accordance with a written procedure, (b) The eddy current instrument shall be capable of
which shall contain, as a minimum, the requirements listed operating with bobbin coil probes in the differential mode
in Table VIII-821. The written procedure shall establish a or the absolute mode, or both.
single value, or range of values, for each requirement.
(c) The eddy current system shall be capable of real

• VIII-821.2 Procedure Qualification. When procedure time recording.


qualification is specified by the referencing Code Section, (d) The eddy current equipment shall be capable of
a change of a requirement in Table VIII-821 identified as sensing and recording discontinuities, dimensional
an essential variable shall require requalification of the changes, resistivity/conductivity changes, conductive/

173
ARTICLE 8 20118 SECTION V

magnetic deposits, and responses from imperfections origi-


nating on either tube wall surface.

VIII-832 Analog Data Acquisition System


VIII-832.1 Analog Eddy Current Instrument
FIG. VIII-864.1 DIFFERENTIAL
RESPONSE FROM CALIBRATION

10% 10 groove 10%F


TECHNIQUE
REFERENCE

g'Oov.

(a) The frequency response of the outputs from the eddy
current instrument shall be constant within 2% of full scale
from dc to Fmax, where Fmax (Hz) is equal to 10 Hz-s/in.
(0.4 Hz-s/mm) times maximum probe travel speed [in.lsec
(mm/s)].
(b) Eddy current signals shall be displayed as


two-dimensional patterns by use of an X- Y storage oscillo-
scope or equivalent.
(e) The Lissajous display shall have a minimum resolu-
VIII-832.2 Magnetic Tape Recorder
tion of 7 bits full scale.
(a) The magnetic tape recorder used with the analog
equipment shall be capable of recording and playing back
if) The strip chart display shall be capable of displaying
at least 2 traces.
eddy current signal data from all test frequencies and shall
(g) The strip chart display shall be selectable so either
have voice logging capability.
the X or Y component can be displayed.
(b) The frequency response of the magnetic tape
(h) The strip chart display shall have a minimum resolu-
recorder outputs shall be constant within 10% of the input
tion of 6 bits full scale.
value from dc to F max, where F max (Hz) is equal to
10 Hz-s/in. (0.4 Hz-s/mm) times maximum probe travel VIII-833.2 Digital Recording System
speed [in.lsec (mm/s)]. (a) The recording system shall be capable of recording


(c) Signal reproducibility from input to output shall be and playing back all acquired eddy current signal data from
within 5%. all test frequencies.
(b) The recording system shall be capable of recording
VIII-832.3 Strip Chart Recorder
and playing back text information.
(a) Strip chart recorders used with analog equipment
(c) The recording system shall have a minimum resolu-
shall have at least 2 channels.
tion of 12 bits per data point.
(b) The frequency response of the strip chart recorder
shall be constant within 20% of full scale from dc to Fmax,
where Fmax (Hz) is equal to 10 Hz-s/in. (0.4 Hz-s/mm) VIII-834 Bobbin Coils
times maximum probe travel speed [in.lsec (mm/s)]. VIII-834.1 General Requirements
(a) Bobbin coils shall be able to detect artificial disconti-
nuities in the calibration reference standard.
VIII-833 Digital Data Acquisition System
(b) Bobbin coils shall have sufficient bandwidth for


VIII-833.1 Digital Eddy Current Instrument operating frequencies selected for flaw detection and
(a) At the scanning speed to be used, the sampling rate sizing.
of the instrument shall result in a minimum digitizing rate (c) Coils shall be mounted as close to the outside of the
of 30 samples per in. (1.2 samples per mm) of examined probe as practical while providing sufficient protection
tubing, use dr = sr/ss, where dr is the digitizing rate in against coil damage.
samples per in., sr is the sampling rate in samples per sec (d) The probe fill factor (probe diameter/tube inside
or Hz, and ss is the scanning speed [in.lsec (mm/sec)]. diameter) shall be a minimum of 0.80.
(b) The digital eddy current instrument shall have a
minimum resolution of 12 bits per data point.
(c) The frequency response of the outputs of analog VIII-850 TECHNIQUE
portions of the eddy current instrument shall be constant VIII-850.1 Probe Data Acquisition Speed. The tra-
within 2% of the input value from dc to Fmax, where Fmax verse speed shall not exceed that which provides adequate
(Hz) is equal to 10 Hz-s/in. (0.4 Hz-s/mm) times maximum frequency response and sensitivity to the applicable calibra-


probe travel speed [in.lsec (mm/s)]. tion discontinuities. Probe scanning speed is determined
(d) The display shall be selectable so that the examina- by test frequency and the length of the smallest defect to
tion frequency or mixed frequencies can be presented as be detected and sized. The maximum probe data acquisition
a Lissajous pattern as shown in Fig. VIII-864.I. speed shall be 36 in.lsec (90 cm/sec). For copper alloys

174
2011a SECTION V ARTICLE 8

• (materials with a resistivity less than 50 J.L ohm/cm), the


maximum probe data acquisition speed shall be 18 in.lsec
(45 cm/sec).
VIII-850.2 Automated Data Screening System. When
automated eddy current data screening systems are used,
each system shall be qualified in accordance with a written
VIII-862.2 Calibration Reference Standards for Dif-
ferential and Absolute Bobbin Coils. Calibration refer-
ence standards shall contain the following artificial
discontinuities that are located on either the inner or outer
surface of the tube specimen:
(a) One ~6 in. (1.6 mm) wide, 360 deg circumferential

procedure. groove 10% through from the tube inner surface.


(b) One Ys in. (3.2 mm) wide, 360 deg groove 10%
through from the tube outer surface.
VIII-860 CALffiRA TION (c) One hole drilled 100% through the tube wall ~6 in.
VIII-861 Equipment Calibration (1.6 mm) diameter.


VIII-861.1 Analog Equipment. The following shall be (d) A sufficient number of artificial defects, such as
verified by annual calibration: EDM manufactured inner surface originated cylindrical
(a) the oscillator output frequency to the drive coil shall rounded bottom pits, 180 deg outer surface originated wear
be within 5% of its indicated frequency scars, and inner surface originated 360 deg grooves to aid
(b) the vertical and horizontal linearity of the cathode in developing a calibration relationship for sizing tube
ray tube (CRT) display shall be within 10% of the deflec- degradation.
tion of the input voltage (1) There shall be a minimum of three sizing disconti-
(c) the ratio of the output voltage from the tape recorder nuities of each type having depths equal to 20%, 40%, and
shall be within 5% of the input voltage for each channel 60% of the nominal tube wall. As-built depths within 5%
of the tape recorder are sufficient.
(d) the chart speed from the strip chart recorder shall (2) The length of sizing defects shall be selected to
be within 5% of the indicated value simulate the expected degradation mechanism. When gen-
(e) amplification for all channels of the eddy current eral inside or outside originated wall loss is possible, the


instrument shall be within 5% of the mean value, at all minimum length shall be % in. (16 mm).
sensitivity settings, at any single frequency (3) Simulated internal pitting defects shall have a
VIII-861.2 Digital Equipment. Digital equipment diameter that simulates the size pitting which may be
shall be calibrated after repairs which may change the encountered. The ~6 in. (1.6 mm) diameter 100% through-
instrument's accuracy are made. wall hole specified in VIII-862.2(c) is intended to identify
the appropriate relationship between inside and outside
originated wall loss and should not be used to generate
VIII-862 Calibration Reference Standards
linear calibration curves for pit sizing.
VIII-862.1 Calibration Reference Standard
Requirements. Calibration reference standards shall con- VIII-863 Base Frequency
form to the following: The base frequency shall be between 190 and 2.1 x 190
(a) Calibration reference standards shall be manufac- as defined by the following equations:
tured from tube(s) of the same material specification, tem- (a) Minimum Base Frequency:


per, and nominal size as that to be examined in the vessel.
(b) The resistivity of the reference standard shall be p
within 2 J.L ohm/cm of the resistivity of the tubing to be
190 = 4.8 r:;;::.
!-Lr

examined to ensure good results. When the resistivity of


(b) Maximum Base Frequency:
the reference standard differs from the tubes being tested,
the resistivity of the reference standard shall be compared
to that of the tubes being tested with a button probe at a P
2.1X19o=l~
!-Lr
frequency of at least 5 times greater than 190.
(c) Artificial discontinuities in calibration reference where
standards shall be spaced axially so they can be individually
evaluated and their eddy current responses can be differen- 190 = the frequency which generates a 90 deg phase
tiated from each other and from the ends of the tube. The separation between a shallow inside originated
as-built dimensions of the discontinuities shall become defect and a shallow outside originated defect

• part of the permanent record of the calibration referenced p = tube material resistivity (1-1- cm) n.
specimen. t = tube wall thickness [in. or (mmJ25)]
(d) Each calibration reference standard shall be perma- J.Lr = relative magnetic permeability (I-I-r = 1.0 for non-
nently identified with a serial number. magnetic materials)

175
ARTICLE 8 20l1a SECTION V

VIII-864 Set-up and Adjustment


VIII-864.1 Differential Bobbin Coil Technique
(a) The sensitivity shall be adjusted to produce Lissa-
jous and/or strip chart signals which clearly display the
smallest defect expected to be measured by the differential
FIG. VIll-864.2
CALIBRATION

.
ABSOLUTE TECHNIQUE
REFERENCE

...........
10% 10 .
STANDARD
FROM

Through wall
hole
.... ;..

signal.
(b) The phase rotation shall be adjusted so the signal
response due to the 10% inside originated groove is within
5 deg of the horizontal axis (max rate). The response due
;5~egi1\tJ'jC,'?o~~~
to the through-wall hole forms either up and to the left or
down and to the right first as the probe is withdrawn from
the calibration reference standard.
(c) Withdraw the probe through the calibration refer-
ence standard at the nominal examination speed. Record the
responses of the applicable calibration reference standard
discontinuities. The responses shall be clearly indicated by
the instrument and shall be distinguishable from each other
as well as from probe motion signals.
any other parts (essential variables) of the eddy current
examination system hardware shall require recalibration.
(b) System calibration verification shall be performed
and recorded at the beginning and end of each unit of data

(d) The 190 frequency should be verified by a 90 deg storage of the recording media and every 2 hr.
phase separation between the inside and outside originated (c) Should the system be found to be out of calibration
10% deep grooves. See example in Fig. VIII-864.1. (as defined in VIII-864.2), the equipment shall be recali-
brated. The recalibration shall be noted on the recording.
VIII-864.2 Absolute Bobbin Coil Technique
All tubes examined since the last valid calibration shall be
(a) The sensitivity shall be adjusted to produce Lissa-
reexamined.
jous and strip chart signals that clearly display the smallest
defect to be measured with the absolute signal.
(b) The phase rotation control shall be adjusted so the
signal response due to the 10% inside originated groove
is within 5 deg (peak-to-peak) of the horizontal axis. The
signal response due to the through-wall hole can be formed
up and to the left or down and to the right as the probe is
VIII-870 EXAMINATION
The maximum probe travel speed used for examination
shall not exceed that used for calibration. Data shall be
recorded as the probe traverses the tube.

withdrawn from the calibration reference standard.
(c) Withdraw the probe through the calibration refer-
VIII-880 EVALUATION
ence standard at the nominal examination speed. Record the VIII-880.1 Data Evaluation. Data shall be evaluated
responses of the applicable calibration reference standard in accordance with the requirements of this Appendix.
discontinuities. The responses shall be clearly indicated by VIII-880.2 Means of Determining Indication Depth.
the instrument and shall be distinguishable from each other For indication types that must be reported in terms of depth,


as well as from probe motion signals. a means of correlating the indication depth with the signal
(d) The 190 frequency should be verified by a 90 deg amplitude or phase shall be established. The means of
phase separation between the inside and outside originated correlating the signal amplitude or phase with the indica-
10% deep grooves. See example in Fig. VIII-864.2. tion depth shall be based on the basic calibration standard
or other representative standards that have been qualified.
VIII-864.3 Digital System Off-Line Calibration. The
This shall be accomplished by using curves, tables, or
eddy current examination data is digitized and recorded
equations and aided by software.
during scanning for off-line analysis and interpretation.
The system set-up of phase and amplitude settings shall be VIII-880.2 Frequencies Used for Data Evaluation.
performed off-line by the data analyst. Phase and amplitude All indications shall be evaluated. Indication types, which
settings shall be such that the personnel acquiring the data must be reported, shall be characterized using the frequen-
can clearly discern that the eddy current instrument is cies or frequency mixes that were qualified.
working properly.


VIII-864.4 System Calibration Verification VIII-890 DOCUMENTATION
(a) Calibration shall include the complete eddy current VIII-890.1 Reporting
examination system. Changes of any probe, extension VIII-890.1.1 Criteria. Indications reported in accor-
cables, eddy current instrument, recording instruments, or dance with the requirements of this Appendix shall be

176
20113 SECTION V ARTICLE 8

• described in terms of the following information, as a


minimum:
(a) location along the length of the tube and with respect
to the support members, when the indication identification
is relevant to a specific location (i.e., fretting @ baffle 2)
(b) depth of the indication through the tube wall
(j) lengths of probe and probe extension cables
(k) size and type of probes
(1) probe manufacturer's name and manufacturer's part
number or probe description and serial number
(m) eddy current instrument model and serial number
(n) probe scan direction during data acquisition
(c) frequency or frequency mix from which the indica- (0) application side - inlet or outlet
tion was evaluated (p) slip ring serial number, as applicable
VIII-890.1.2 Depth. The maximum evaluated depth (q) procedure identification and revision
of flaws shall be reported in terms of percentage of tube
VIII-890.3.2 Tube Identification
wall loss. When the loss of tube wall is determined by the
(a) Each tube examined shall be identified on the appli-
analyst to be less than 20%, the exact percentage of tube


cable unit of data storage and should be consistent with the
wall loss need not be recorded, i.e., the indication may be
manufacturer's as-built drawings and previous inspection.
reported as being less than 20%.
(b) The method of recording the tube identification shall
VIII-890.1.3 Non-Quantifiable Indications. A non- correlate tube identification with corresponding recorded
quantifiable indication is a reportable indication that cannot tube data.
be characterized. The indication shall be considered a flaw
until otherwise resolved. Vill -890.3.3 Reporting
(a) The Owner or his agent shall prepare a report of
VIII-890.2 Support Members the examinations performed. The report shall be prepared,
Vill-890.2.1 Location of Support. The location of filed, and maintained in accordance with the referencing
support members used as reference points for the eddy Code Section. Procedures and equipment used shall be
current examination shall be verified by fabrication draw- identified sufficiently to permit comparison of the examina-
ings or the use of a measurement technique. tion results with new examination results run at a later


VIII-890.3 Records date. This shall include initial calibration data for each
VIII-890.3.1 Record Identification. The recording eddy current examination system or part thereof .
media shall contain the following information within each (b) The report shall include a record indicating the tubes
unit of data storage: examined (this may be marked on a tubesheet sketch or
(a) Owner drawing), any scanning limitations, the location and depth
(b) plant site and unit of each reported flaw, and the identification and certifica-
(c) heat exchanger identification
tion level of the operators and data evaluators that con-
(d) data storage unit number
ducted each examination or part thereof.
(e) date of examination (c) Tubes that are to be repaired or removed from ser-
(/) serial number of the calibration standard vice, based on eddy current examination data, shall be
(g) operator's identification and certification level
identified.
(h) examination frequency or frequencies VIII-890.3.4 Record Retention. Records shall be
maintained in accordance with requirements of the refer-


(i) mode of operation including instrument sample rate,
drive voltage, and gain settings encing Code Section.

• 177
ARTICLE 9 2011a SECTION V

ARTICLE 9

VISUAL EXAMINATION

T-910 SCOPE TABLE T-921


REQUIREMENTS OF A VISUAL EXAMINATION


(a) This Article contains methods and requirements for PROCEDURE
visual examination applicable when specified by a referenc-
ing Code Section. Specific visual examination procedures
required for every type of examination are not included in Requirement (As Applicable)
this Article, because there are many applications where
visual examinations are required. Some examples of these Change in technique used
Direct to or from translucent
applications include nondestructive examinations, leak
Direct to remote
testing, in-service examinations and fabrication proce- Remote visual aids
dures. Personnel performance requirements,
(b) The requirements of Article 1, General Require- when required
ments, apply when visual examination, in accordance with Lighting intensity (decrease only)
Configurations to be examined and base
Article 9, is required by a referencing Code Section.
material product forms (pipe, plate,
(c) Definitions of terms for visual examination appear


forgings, etc.>
in Article 1, Appendix I - Glossary of Terms in Nonde- Lighting equipment
structive Examination, and Article 9, Appendix I. Methods or tools used for surface
preparation
Equipment or devices used for a direct
technique
T -920 GENERAL Sequence of examination
T-921 Written Procedure Requirements Personnel qualifications

T-921.1 Requirements. Visual examinations shall be


performed in accordance with a written procedure, which
condition or artificial imperfection should be in the least
shall, as a minimum, contain the requirements listed in
discernable location on the area surface to be examined to
Table T-921. The written procedure shall establish a single
validate the procedure.
value, or range of values, for each requirement.
T -921.2 Procedure Qualification. When procedure T -922 Personnel Requirements
qualification is specified by the referencing Code Section,
a change of a requirement in Table T-921 identified as an
essential variable shall require requalification of the written
procedure by demonstration. A change of a requirement
identified as a nonessential variable does not require requal-
ification of the written procedure. All changes of essential
The user of this Article shall be responsible for assigning
qualified personnel to perform visual examinations to the
requirements of this Article. At the option of the manufac-
turer, he may maintain one certification for each product,
or several separate signed records based on the area or
type of work, or both combined. Where impractical to use

or nonessential variables from those specified within the specialized visual examination personnel, knowledgeable
written procedure shall require revision of, or an addendum and trained personnel, having limited qualifications, may
to, the written procedure. be used to perform specific examinations, and to sign the
T -921.3 Demonstration. The procedure shall contain report forms. Personnel performing examinations shall be
or reference a report of what was used to demonstrate that qualified in accordance with requirements of the referenc-
the examination procedure was adequate. In general, a ing Code Section.


fine line J.32 in. (0.8 mm) or less in width, an artificial
imperfection or a simulated condition, located on the sur- T -923 Physical Requirements
face or a similar surface to that to be examined, may be Personnel shall have an annual vision test to assure
considered as a method for procedure demonstration. The natural or corrected near distance acuity such that they are

178

You might also like