Purchaser’s Requirements Bidder’s Offer
Priori Conformity Remarks
ty Yes No
Parameters to be monitored:
a. ECG 3 Leads C
1. b. NIBP - ( Oscilometric) - 0 to 300mmHg
±4mmHg
c. SPO2 - 1 to100% ± 2 digits
Display Screen: C
a. Built in active TFT colour display, at 12
inch or more in size
b. Minimum resolution 800x600 pixel
2. c. Minimum No. of traces: 4
d. Simultaneously displaying 2x ECG & 1x
SPO2
e. Numerical information displayed in large
digits against respective waveforms
Pulse Oxymetry - Masimo or equivalent C
offering:
a. Motion tolerance
b. Ability to perform under low perfusion
3.
conditions
c. Washable & disinfectible probes
d. Connecting cable length should be min.2
meters
Must have features & facilities: C
a. Trend graphing & listing up to 24 hrs.
b. Defibrillator overload protection
c. Advisory & critical warning alarms in
4. continuously audible & visual form
d. Convenience user interface such as touch
screen
e. Possibility to programme NIBP in manual,
periodic & automatic modes
General Conditions
5. The unit shall operate on the power supply of C
230V±10% 50 Hz single phases.
6. Each unit shall be supplied with an instruction
manual and a service manual in English.
7. Full graphic illustrated original technical C
literature in English describing the equipment
offered and detailing the specifications shall be