Specification for Bed , Ward , Single Function

1 2 3 4
Purchaser’s Requirements Bidder’s Offer
Priori Conformity Remarks
ty Yes No
Technical Specifications:
1. Overall Size Should not be less than following C
parameters.
±10% variance accepted
Length: 2100mm, Width: 950mm, Height: 500mm
Required Functions: C
a. Head rest adjustment (0deg – 85 deg) shall
2. be operated by smooth manual crank and the
Manual crank should have an over
turning protection mechanism.
Required Features: C
a. Collapsible Alloy Side Railings should
be available on both sides of the bed.
b. Wall protection rubber bumpers should
be available at all four corners.
c. Baggage holding tray should be
available under the head section.
3. d. At least two provisions required for IV
pole on head end of the bed with one
IV pole.
e. ABS Molded Detachable Head and Foot
Boards required
f. At least 120mm diameter TPR, non-
rusting, Swiveling, castors with
diagonal locking
Material Specifications: C
a. Main framework made of 1.5” x 2.5”
GI Rectangular tubes of 1.6mm thick.
b. Legs should be made of 2”x2” GI Box tubes of
4. 1.6mm thick.
c. 18G Perforated CRC/MS Sheet for
mattress base in two sections
d. All steel materials should be pretreated
and finished with epoxy powder coating.
Mattress: C
Non Foldable Mattress of 25mm soft density layer and
5. 75mm high density bottom layer should be supplied
with the bed. It shall be covered with a medical grade
cover.

1 DIVISION OF BIOMEDICAL ENGINERRING SERVICES


[02.03.2020]
General Conditions
6. Full graphic illustrated original technical literature in C
English describing the equipment offered and detailing
the specifications shall be supplied with the
bid.
7. A detailed proforma invoice of the equipment C
describing the parts and accessories offered as
requested by the specification together with their make,
model, country of origin, unit price, quantity, total
price, ref. numbers of accessories and the period of
warranty offered etc. shall be compulsorily
provided with the bid.
8. A list of users if any, in Sri Lanka of the equipment
offered shall be provided together with the bid.
9. If the model is not available in Government Hospitals C
in Sri Lanka, a sample unit shall be submitted for
evaluation when requested.
10. The equipment shall be covered by a comprehensive C
“Parts & Labour” warranty for the period of not less
than 24 calendar months from the date of successful
Installation & commissioning.
11. Bidder shall furnish the documentary evidence to C
demonstrate that the good it offers meet the following
usage requirement.
a. Manufacturer shall have ISO
9000/GMP certification/equivalent.

Make & Model : ……………………………………………………


Country of Manufacture : ……………………………………………………
Relevant catalogue & technical information attached
Information given under “Bidder’s offer” is true & correct
Bidder’s Name, Signature & Date : ……………………………………………………
Company seal : ……………………………………………………

2 DIVISION OF BIOMEDICAL ENGINERRING SERVICES


[02.03.2020]
3 DIVISION OF BIOMEDICAL ENGINERRING SERVICES
[02.03.2020]