MAGNETOM

MR
Update Instructions No. 0 2 6 / 9 8
Title: Release of NUMARIS/3 VB31D

Type of update: Safety Mandatory Mandatory

Recommended X As required

Urgency: Immediate X Next visit

Update materials required? X Yes, Part No.: 5751339 K2119 No

Materials free of charge? X Yes No N.A.

Spare Parts affected? Yes X No

Last order can be placed until: August 1999

Estimated completion time: approx. 1 hour

Systems/Products affected
Name Part No. Serial No. New component sta-
tus/ Rev. Level

MAGNETOM Open 16 42 086 K2140 all 57


MAGNETOM Open viva 47 62 469 K2161 8301-8337 05
8501-8612 05
MAGNETOM Impact 54 55 647 K2114 all 175
MAGNETOM Expert 37 61 108 K2114 all 57
MAGNETOM Expert Upgrade 37 61 116 K2114 9601-9918 48
MAGNETOM Vision 16 42 334 K2130 7001-7490 97
MAGNETOM Vision Upgrade 37 60 084 K2151 7601-7671 73

Components affected/to be modified


Name Part No. Serial No. New component status/
Rev. Level

see table above see table above see table above see table above
© Siem ens A G 1 998
ÄA No.: 88687 The reproduction, transmission or
use of this document or its contents
Name: Schülein is not permitted without express
written authority. Offenders will be
Dept.: TD PS 23 liable for damages. All rights,
including rights created by patent
Part No. of document(s) n.a. grant or registration of a utility
model _or_ design,_are_ reserved.

Print No.: MR-098.092.26.01.02 Doc. Gen. Date: 07/98


Replaces: n.a. Page 1 of 10
2 Update Instructions No. 026/98

Release of NUMARIS/3 VB31D 0

Purpose 0

To update all MAGNETOM systems to software NUMARIS/3 VB31D.


Effective July 3rd, 1998, this software version is released for all MAGNETOM Open, Open
viva, Impact, Expert and Vision systems including their plus-variants.
This update was developed to correct problems regarding DICOM transfers and
decreased image quality viewed via the MAXICAM VCR output. Trouble Reset has been
improved, too.
Since DICOM is not used for image transfers at all sites and also the MAXICAM VCR out-
put is not used very often the 3VB31D needs to be installed just as required!

Systems/Products affected 0

See front page!

Prerequisites 0

• The system is running with 3VB31C!


(Only this version can be updated to 3VB31D!)
• To be on the safe side:
- Have an actual dynamic data set available.
- Archive and subsequently delete all patient images.

Special tools / Documents required 0

• Actual password list (Speed Info)!

Update kit: part numbers 0

Update kit suitable for Part number


UMR.NUM. 3VB31D Open, Open viva, Impact, Expert, Vision and 5751339 K2119
their plus- variants

Contents of the update kit 0

The update kit 5751339 K2119 contains the following parts

Pos. Amount Part No. Name Note


0010 1 5751321 K2119 CD NUM.3 VB31D CD-ROM
0020 1 5751990 Etikett Num.3 VB31D sticker
0030 1 5565697 Versandverpackung -> packing

MAGNETOM Register 1 MR-098.092.26 Page 2 of 10 Siemens AG


Update Instructions 1998 Rev. 01 07/98 TD PS 23 Medical Engineering
Update Instructions No. 026/98 3

Order information 0

Order the above described update kit for every system to be updated.
Pease, state the serial number of the system you need the kit for.
Orders in the USA have to be handled via Iselin; orders from the districts will be canceled.

Parts to be returned 0

n.a.

Spare parts handling 0

n.a.

Procedure 0

Update from NUMARIS/3 VB31C to NUMAR/3 VB31D 0

• Insert the CD with VB31D into the CD-ROM drive.

Output Input Note


mrc_login: install↵ If the system consists of two
Password: ...↵ consoles (mrc + mrsc1) both
consoles must be booted.
The mrsc1 is at the login
prompt.

MENU : Installation of NUMARIS

Your Choice [E] : si↵

MENU : Software Installation

Your Choice [MU] : sucd↵ Preparation for SW update


from CD

***************************************************
** Info : Please insert the NUMARIS CD **
***************************************************
Press <Ret> to continue : ↵ Press <return> to continue.

Update NUMARIS from CDRom?


Your Choice ([N] / Y) : y↵
Previous Version is VB31C
Welcome to Numaris Upgrade Procedure for Expert
============================================

Siemens AG Register 1 MR-098.092.26 Page 3 of 10 MAGNETOM


Medical Engineering Rev. 01 07/98 TD PS 23 Update Instructions 1998
4 Update Instructions No. 026/98

Output Input Note

(1) Installation of Numaris/3 VB31D


(q) No action, quit
Your Choice : 1 Select option 1 for the
update to 3VB31D
-------------
sv_display will be copied to ......
........
........
........

Installation of Numaris/3 VB31D successfully Also the satellite console ( if


completed ! available) has been updated
Don’t forget to perform a new SaveSet via to 3VB31D!
SaveDynData!
Please reboot your system now !
***********************************************
** Info : Update of NUMARIS done **
***********************************************
Press <Ret> to continue : ↵ Press <return> to continue.

MENU : Installation of NUMARIS

Your Choice [EI] : ↵ Exit install

The sytem reboots automatically to activate the changes!

System check 0

mrc login mr↵ Log in to MR and perform a


system check e.g. a mea-
surement.
Afterwards perform
’Save Dynamic Data’.

Customer documents 0

To validate the customer documents also for NUMARIS/3 VB31D, put one of the stickers
right after the current software version on the cover page of each folder.

MAGNETOM Register 1 MR-098.092.26 Page 4 of 10 Siemens AG


Update Instructions 1998 Rev. 01 07/98 TD PS 23 Medical Engineering
Update Instructions No. 026/98 5
Deinstalling NUMARIS/3 VB31D 0

Since it is not absolutely clear that the 3VB31D (especially the DICOM - implementation)
will satisfy every customer also a deinstallation procedure is provided with the CD-ROM.
Furthermore, the 3VB31D needs to be deinstalled temporarily for performing host
upgrades.

To deinstall the 3VB31D insert the 3VB31D-CDROM in to the drive and proceed as fol-
lows:

Output Input Note


mrc_login: install↵ If the system consists of two
Password: ...↵ consoles (mrc + mrsc1) both
consoles must be booted.
The mrsc1 is at login
prompt.

MENU : Installation of NUMARIS

Your Choice [E] : si↵

MENU : Software Installation

Your Choice [MU] : sucd↵ Preparation for SW update


from CD

***************************************************
** Info : Please insert the NUMARIS CD **
***************************************************
Press <Ret> to continue : ↵

Update NUMARIS from CDRom?


Your Choice ([N] / Y) : y↵
Previous Version is VB31D
Welcome to Numaris Upgrade Procedure for Expert
=============================================

(1) Uninstall Numaris/3 VB31D


(q) No action, quit
Your Choice : 1 Select option 1 to deinstall
the 3VB31D!

Siemens AG Register 1 MR-098.092.26 Page 5 of 10 MAGNETOM


Medical Engineering Rev. 01 07/98 TD PS 23 Update Instructions 1998
6 Update Instructions No. 026/98

Output Input Note


-------------
.sv_display.3b31c will be restored to ......
........
........
........

Deinstallation of Numaris/3 VB31D successfully Also the satellite console ( if


completed ! available) is back to
If you want to use Numaris/3 VB31C perma- 3VB31C!
nently don’t forget to perform a new SaveSet
via SaveDynData !
Please reboot your system now !
***********************************************
** Info : Update of NUMARIS done **
***********************************************
Press <Ret> to continue : ↵

MENU : Installation of NUMARIS

Your Choice [EI] : ↵ Exit install

The sytem reboots automatically to activate the changes!

System check 0

mrc login mr↵ Log in to MR and perform a


system check e.g. a mea-
surement.
Afterwards perform the
’Save Dynamic Data’ if you
intend to use the 3VB31C
permanently.

MAGNETOM Register 1 MR-098.092.26 Page 6 of 10 Siemens AG


Update Instructions 1998 Rev. 01 07/98 TD PS 23 Medical Engineering
Update Instructions No. 026/98 7

Host upgrade (Turbosparc) 0

For host upgrades the 3VB31D needs to be deinstalled temporarily according "Deinstall-
ing NUMARIS/3 VB31D".
Perform the host upgrade according the appropriate hardware upgrade documents and
the chapter "Installing the 3VB31C software patch" of the software manual.
Install the 3VB31D again after checking the system for proper function under 3VB31C
with the new host.
Draw a new dynamical safe set.

NEW hints: 0

C H A R M Entry ID : 269841 0

A MIP-series transfered to the Prominence will get the name of the original series MIP has
been performed with. An image containing graphics will not be transfered correctly. In the
resulting image multiple interfering images of the original MR-image are superimposed by
horizontal patterns.
At the Prominence some tranfered data are not displayed in general, respectively cannot
be found in any platform. Missing are among others:
Weight, Maiden Name, Ref. Phys., Adm. Diagn., Comments, Pat. Pos, View Dir., all data
from Patient Extension, Reg. Date.

C H A R M Entry ID : 269814 0

DICOM-Transfer 3VB31D->3VB31D
Study names are transfered incorrectly (e.g. with MIP, StuModify, Dyn.Anal.). The name of
the original study is used instead.
Study Type is not transfered correctly.
Ima Type is not transfered correctly.

C H A R M Entry ID : 269738 0

DICOM- Transfer 3VB1D->35VA11A


Studies with the same ’patient move time’ can be processed at the ’source system’ with-
out problems, that means the slice graphics is displayed without any warnings. At the des-
tination system , however, a warning regarding ’wrong patient move time’ comes up. The
warning needs to be acknowledged before the slice graphics is displayed.
This warning also comes up if the Scout is out of the selected study (e.g. display a Scout
and select the same Scout-Seq. for Stu/Dis).
An analog problem occurs when patient images, measured at a 35VA11A system, are
sent to a 3VB31D system and back to the 35VA11A system again.

Siemens AG Register 1 MR-098.092.26 Page 7 of 10 MAGNETOM


Medical Engineering Rev. 01 07/98 TD PS 23 Update Instructions 1998
8 Update Instructions No. 026/98
C H A R M Entry ID : 269715 0

With DICOM- transfers the time stamp entry in the data base differs by one hour.
This affects some functions how they display data, e. g. the browser (register-, image-
date), the viewer (register-, image-date) and study display (Warning: Image has a different
patient move time from localizer).
Possibly this applies also for other functions accessing the time stamp entry in the data
base.

C H A R M Entry ID : 270180 0

DICOM-Transfer
The function Patient/Modify does NOT create unique DICOM-UIDs. Some of the created
UIDs are identical to those of the former patient. A possible result of this problem could be
this: All patient images are sent from the source to a destination system; at the destination
system name, weight and comment are modified (e.g.) and the images sent back to the
source system again. This way a new patient is created at the source system.
If, however, just the comment of one image is modified at the destination system and all
images are sent back to the source system, there the images of all studies are duplicated.

C H A R M Entry ID : 270182 0

DICOM-Transfer 3B31D <--> Prominence


Study/Display does not work anymore, if patient images measured with 3VB31D are sent
to a Prominence and back to the source again.
Error message: ’unable to decode arguments’!

Completion 0

• Copy and fill out the attached "Update Completion Form", file it in Register 10 of the
System Manual.
• Enter the completed UI in the OPSIS UPDATE Handling system.
• Please fill out the Update Completion Form even if you completed the UI before it was
published.

Changes to previous version 0

NA

MAGNETOM Register 1 MR-098.092.26 Page 8 of 10 Siemens AG


Update Instructions 1998 Rev. 01 07/98 TD PS 23 Medical Engineering
Update Instructions No. 026/98 9

Update completion Form 0

Siemens AG
Medical Engineering
Technical support TD PS 23
Henkestr. 127
D-91052 Erlangen
Germany

Department: TD PS 23
Tel.: 49-(0)9131-84-6440
Fax: 49-(0)9131-84-3593

The Update Instruction UI-Nr. 026/98 has been performed.

System type: .........................................................................

Serialnumber: .........................................................................

Sitenumber: .........................................................................

Customer: .........................................................................

Name: .........................................................................

Department: .........................................................................

Date: .........................................................................

Signature: .........................................................................

NOTICE After an update has been performed, this page must be copied,
filled in and filed in register 10 / system folder or register 3 / log
book of the corresponding site.

Siemens AG Register 1 MR-098.092.26 Page 9 of 10 MAGNETOM


Medical Engineering Rev. 01 07/98 TD PS 23 Update Instructions 1998
10 Update Instructions No. 026/98
This page intentionally left blank.

MAGNETOM Register 1 MR-098.092.26 Page 10 of 10 Siemens AG


Update Instructions 1998 Rev. 01 07/98 TD PS 23 Medical Engineering