Program for calibration,

preventive maintenance and


repair for the equipment

 Record of schedule and updated certificate


of calibration and maintenance of
equipment
 Record of reports of preventive
maintenance and repair
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
BOECO GERMANY
ITEM MICROSCOPE
BM 300
SERIAL/LOT NUMBER 0000497
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
NIKON ECLIPSE MICROSCOPE
ITEM
E200
SERIAL/LOT NUMBER 762179
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
VIGISGSTEN LABORATORY
ITEM INSTRUMENT INC.
CENTRIFUGE DSC 200T
SERIAL/LOT NUMBER 1009496
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
8 PLACER CENTRIFUGE
ITEM
(CLINICAL)
SERIAL/LOT NUMBER 0709529
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
DIGISYSTEM CENTIFUGE
ITEM
(CLINICAL)DSC-200T
SERIAL/LOT NUMBER AR 1508
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
OLYMPUS MICROSCOPE
ITEM (BIOLOGICAL)
CX21LEDFS1
SERIAL/LOT NUMBER OE86577
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
DIGISYSTEM CENTRIFUGE 8
ITEM PLACER (CLINICAL)
DSC-158T
SERIAL/LOT NUMBER CT2-18110929
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
BIO-RD CENTRIFUGE
ITEM
ID 12 SII
SERIAL/LOT NUMBER 2004294
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

CERTIFICATE OF CALIBRATION
This is to certify that the equipment analyzer below has passed
the calibration standards of MEDTECH ASSOCIATION and it has
performed within the acceptable criteria of the National Laboratory
Standards.

THOMASIAN DOCTORS
LABORATORY/HOSPITAL
HOSPITAL
XL-180 (AUTOMATED
ITEM
BIOCHEMISTRY ANALYZER)
SERIAL/LOT NUMBER 160317
RESULTS OF ANALYSIS CALIBRATION PASSED
DATE CALLIBRATED NOVEMBER 2019

Certified by:

Bernard Brylle Anthony E. Lumberio


Regional Service Engineer Manager

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Vigisgsten Laboratory Instrument Inc. Centrifuge DSC 200T


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
A319786 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
200749 AOE758309 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the tub for damage  
Check the O-ring for damage  
Check the hinges for damage  
Check the fan filter for contamination  
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Digisystem centrifuge (Clinical) DSC-200T


PREVENTIVE MAINTENANCE REPORT

Case No: Institution: Address: Date:


B124089 Thomasian Doctors Hospital December 01. 2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
657021 ADF09876 3.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the tub for damage  
Check the O-ring for damage  
Check the hinges for damage  
Check the fan filter for contamination  
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Digisystem Centrifuge 8 Placer (Clinical) DSC-158T


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
C785099 Thomasian Doctors Hospital
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
830938 OOA67830 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the tub for damage  
Check the O-ring for damage  
Check the hinges for damage  
Check the fan filter for contamination  
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

BIO-RD Centrifuge ID 12 SII


PREVENTIVE MAINTENANCE REPORT

Case No: Institution: Address: Date:


D304934 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
700234 AAE002345 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the tub for damage  
Check the O-ring for damage  
Check the hinges for damage  
Check the fan filter for contamination  
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

8 Placer Centrifuge (Clinical)


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
E097342 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
472429 EIO872345 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the tub for damage  
Check the O-ring for damage  
Check the hinges for damage  
Check the fan filter for contamination  
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Boeco Germany Microscope BM 300


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
F458609 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
389208 EII048738 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the Lenses for damage  
Check the Eye piece for damage  
Check the hinges for damage  
 
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Nikon Eclipse Microscope E200


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
G093489 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
027634 CEO783458 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the Lenses for damage  
Check the Eye piece for damage  
Check the hinges for damage  
 
Check for functionality of Components Check Fix/Adjust/
REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Olympus Microscope (Biological) CX21LEDFS1


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
H954059 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
983438 AOI73495 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Lenses  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the Lenses for damage  
Check the Eye piece for damage  
Check the hinges for damage  

Check for functionality of Components Check Fix/Adjust/


REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

XL- 180 (Automated Biochemistry Analyzer)


PREVENTIVE MAINTENANCE REPORT
Case No: Institution: Address: Date:
099823 Thomasian Doctors Hospital December 01,2019
12 SIIS/N 12 SII PCN: Firmware Version:  Required as
673920 EEA037475 2.0 Part of PM
 If needed only
6 MONTH PM ANNUAL PM
Cleaning and disinfect of Components Clean Disinfect
REFERENCE:
Tub  
Components Cover  
Outer housing  
Perform Remove any liquid with a clean, dry, lint free cloth 
Visual Inspection of Components Check Fix/Replace
REFERENCE:
Components Visually check the seal for damage  
Check the Lenses for damage  
Check the Eye piece for damage  
Check the hinges for damage  

Check for functionality of Components Check Fix/Adjust/


REFERENCE: Replace
Components Keypad functionality  
LCD contrast  
Proper sound volume  
Lid Closing/Opening  
Verify on the screen if status is correctly identified  
Verify ID Cards Mobility  
Alignment and QC Verification Protocols Done Passed REFERENCE
Finals Checks Verify Target Speed (1030 rpm ±5  
rpm)  f 
Verify centrifugation time (10 mins  
±12 secs)  
Verify the speed at which a 10
unbalanced card will trigger the
vibration sensor (900±50rpm)
Verify AVR Output Voltage

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Vigisgsten Laboratory Instrument Inc. Centrifuge DSC 200T


PREVENTIVE MAINTENANCFE REPORT

The model Vigisgsten Laboratory Instrument Inc. Centrifuge DSC 200T have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Digisystem centrifuge (Clinical) DSC-200T


PREVENTIVE MAINTENANCFE REPORT

The model Digisystem centrifuge (Clinical) DSC-200T have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Digisystem Centrifuge 8 Placer (Clinical) DSC-158T


PREVENTIVE MAINTENANCFE REPORT

The model Digisystem Centrifuge 8 Placer (Clinical) DSC-158T have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

BIO-RD Centrifuge ID 12 SII


PREVENTIVE MAINTENANCFE REPORT

The model BIO-RD Centrifuge ID 12 SII have been serviced for


Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

8 Placer Centrifuge (Clinical)


PREVENTIVE MAINTENANCFE REPORT

The model 8 Placer Centrifuge (Clinical) have been serviced for


Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Boeco Germany Microscope BM 300


PREVENTIVE MAINTENANCFE REPORT

The model Boeco Germany Microscope BM 300 have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Nikon Eclipse Microscope E200


PREVENTIVE MAINTENANCFE REPORT

The model Nikon Eclipse Microscope E200 have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC.

Olympus Microscope (Biological) CX21LEDFS1


PREVENTIVE MAINTENANCFE REPORT

The model Olympus Microscope (Biological) CX21LEDFS1 have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi
SUNLIFE

SUNLIFE DIAGNOSTICS SUPPLIES INC. XL- 180 (Automated Biochemistry Analyzer)


PREVENTIVE MAINTENANCFE REPORT

The model XL- 180 (Automated Biochemistry Analyzer) have been serviced for
Prevention Maintenance and meet Bio-Rad Performance specifications.

Signature Date
Field Service Engineer December 01,2019
Customer December 01,2019

Direct +639 123-4567, +639 234-567 Fax. +639 345-6789, +639 888-3729, +639 9728-1172 Rawis, Legazpi