You are on page 1of 2

Client:

Project:
Tag Number:
Service:
Model:

Please provide response/s under TQ #: 1

QUERY ITEM #1
TQ1: Query Date: Jul. 25, 2018
Please answer the following questions on the Response
1 row below:

1. Please confirm if you can provide Float Type


level instruments, except LAL/LSL-1172 and
LAL/LSL-1173.
NO

Please confirm if this model can handle a


density range of 932 – 1000 kg/m3. YES

2. Please confirm if you can provide Conductive


Limit Level Switch for LAL/LSL-1172 and
LAL/LSL-1173.

YES

3. Please confirm if this model can handle the


following liquids:

a. Deionized Water (LG) YES (DISTILLED)


b. Deionized Water (HG) YES (DISTILLED)
c. Dishwashing Liquid NO
Fabric Conditioner NO
d. Lotions YES (BODY LOTION)
e. Shampoo YES
f. Body Wash YES
g. Deodorant YES

4. Please confirm if Efector 160 LMT100 can


handle liquids with viscosity range of 1 –
10,000 cP.
YES
5. Please provide a model that can withstand a
temperature range of 120 and 164 degrees
Celsius for LAH/LSH-1402 and LAH/LSH-1182,
respectively. UP TO 100c ONLY AND THAT
LMT100
6. Can G ½ A be connected to DN 20, 40, and 50?
YES
7. Please provide us the pressure range this item
can withstand. We required 1 – 3.5 barg.
YES
8. Please confirm if the internal materials are
made of SS316L.YES

Response 1: Response Date:

OTHER INFORMATION
Prepared by:
Company:
Attachment #

You might also like