You are on page 1of 29

Ann Clin Biochem 1985; 22: 46D-488

Review Article

Electrodes In clinical chemistry*


A D HI R S T t and J F S T EVE N S :j:
From the i Department of ChemicaL PathoLogy, Kings College HospitaL, London SE5 and
i Department of ChemicaL PathoLogy, St Stephen's HospitaL, London SWlO, UK

It has been possible to detect salts and bases in one of its salts, some metal ions enter solution,
solution by electrochemical means for many leaving electrons behind on the metal strip (Fig.
years. This technique, which has proved to be 1). The result of this dissociation is a small
particularly useful for the analysis of simple negative charge on the metal (b-) and a similar
aqueous solutions, was found to have limita- small positive charge in the solution. This in
tions in the analysis of biological fluids. The turn causes metal ions to tend to return to the
most serious limitations were the lack of sensi- metal strip; eventually a dynamic equilibrium is
tivity and specificity of detectors, and a varying established.
relationship between calibrants and samples. The potential between the metal strip and the
Other problems associated with electrodes were solution, which is produced by the small charge
related to the non-linear response and the time differences is known as the electrode potential.
taken to respond to a change in concentration It is dependent on the nature of the metal and
of an analyte. the concentration of the salt. No current can be
Considerable progress has been made in the drawn from this system alone-it forms a half
improvement of specifity, or selectivity, of cell. If a similar half cell with a different
electrodes. Modern electronic techniques and concentration of salt is connected to the first,
microprocessors have reduced the problems of with an electrical conductor between the two
sensitivity and non-linear response. Such im- solutions (eg a 'salt bridge') and the two metals,
provements have led to the introduction of the difference in electrode potential between
analytical techniques which have certain advan- the two half cells can be measured with a
tages over conventional techniques. The prob- suitable voltmeter. The potential difference or
lem of calibration has not yet been solved, and voltage of the cell (two electrically connected
there is confusion caused by an apparent half cells), when a minimal current is flowing,
inaccuracy of results produced by these techni- relates to the difference in concentration of the
ques, when compared to results using conven- solutions. The relationship is described by the
tional photometric analysers. It is possible that Nernst equation:
electrochemical methods will gradually replace
conventional techniques because they offer the RT al
E=-log., -
advantage of speed and sensitivity, and because nF a2
they do not require optically-clear solutions.
The following is an account of the principles
of electrode action and the possible applications 8-
<,
of electrodes in clinical chemistry. Zn
8+
Electrochemical ceUs -,
ELECTRODE POTENTIALS Zn 2 +
--~
Jr
- - -
When a strip of metal is placed in a solution of
- - "-'- Zn 2 +
- - - +2e-
- SO 42- - - - - - -
"This review was prepared at the invitation of the - - - - - - -
Scientific Committee of the Association of Clinical - - - - - - -
Hi. chemists (Instrumentation and Data processing
sub-committee'Lbut does not necessarily reflect the
views of the Scientific Committee. FIG. 1. A zinc half cell.

460
Electrodes in clinical chemistry 461

where E is the potential difference of the cell H2 ~ 2H ~ 2e- + 2H+


for given chemical activities, at and a2, of ions hydrogen platinum surface solution
in the two solutions. The standardisation of one gas (negative) (positive)
of these electrodes would enable a practical use
of this system for estimating the chemical At equilibrium, the rate of ionisation of hy-
activity of unknown solutions. To this end, the drogen atoms is proportional to the hydrogen
hydrogen half cell has been adopted as the molecules adsorbed onto the platinum, and
primary reference electrode. therefore to the pressure of hydrogen gas in
contact with the electrode. Stability is main-
THE HYDROGEN HALF CELLI tained by bubbling hydrogen gas at constant
This electrode is, in principle, the same as the pressure around the electrode, replenishing the
zinc model discussed earlier, or any other metal hydrogen consumed by adsorption and ionisa-
electrode, with hydrogen replacing the zinc and tion.
hydrogen ions in solution replacing the zinc While hydrogen electrodes are said to be not
ions. Hydrogen, being a gas, cannot be used too difficult to construct.i they are not suitable
directly as an electrode. The electrode consists for routine measurements of pH, or the concen-
of platinum, which is inert as an electrode, tration of any other ion. The use of the
coated with finely divided platinum (platinum hydrogen electrode is reserved as a reference
black), which is in contact with hydrogen gas. standard for the measurement of relative elec-
In Figure 2, which shows two hydrogen half trode potentials of other half cells. A suitable
cells connected by a salt bridge, the left-hand replacement would be an electrode with a
side cell is' accepted as a reference, against known potential relative to the hydrogen elec-
which the electrode potentials of other cells are trode which is simple to use and stable over
measured u'nder all conditions. The reference long time periods. This type of half cell is called
electrode consists of platinum-platinum black a reference electrode, as opposed to the hy-
in contact with hydrogen gas (bubbled past) at a drogen electrode which is called 'the reference
pressure of one atmosphere, in a solution of electrode' .
1 M hydrogen ion activity. The right-hand cell
contains a 0·1 M H+ solution which is being REFERENCE ELECTRODES
compared with the reference. For electrodes to be used as sensors, two
In the solution, hydrogen molecules are connected electrodes (or half cells) are required
adsorbed onto the platinum black and disso- to enable electrochemical measurement to be
ciate into atoms which may then ionise. The made. The most practical arrangement is to
following equilibrium is established: have one half cell which has a constant potential

Pt

H2 _::
.. _ -

I atmosphere

00
o
o 0
:.0 II"
H+

FIG. 2. Two hydrogen half cells.


462 Hirst and Stevens

(a reference electrode), against which the solutions also contributes to the total potential
potential of a second, variable half cell (a difference of the cell. This potential, called 'the
measuring electrode), is compared. junction potential', will be discussed later.
The most common type of reference elec-
trodes in use are two chloride electrodes: THE NERNST EQUATION
Assuming that the junction potential is negli-
Ag;AgCIJ0·1 M HCI (silver-silver chloride), gible, the hydrogen cell may be used for the
and estimation of hydrogen ion activity (or pH) in
Hg;Hg2Cl2/3·5 M KCI (mercury-mercurous unknown solutions. The relationship between
chloride-calomel) . the potential differences obtained and activity
of hy.dro§en ions is defined by the Nernst
Both silver chloride and mercurous chloride are equation."
highly insoluble in aqueous media.
For the calomel electrode at 25°C, variation
EMF=E reference-E unknown
in the electrolyte concentration would, as ex-
RT (H+reference)
pected, produce a variation in potential. The = - log,
standard potentials, relative to the normal nF (H+unknown)
hydrogen electrode are.:'
R=the gas constant (8,315 J/molJdegree);
+0·3376 V for 0·1 M KCI, T=absolute temperature;
+0·2848 V for 1 M KCI and n=valency charge;
+0·2458 V for saturated KCI. F=Faraday (96 500 coulombs).

A CELL
If the reference and second half cell solutions
A cell consists of two half cells which are are 1 M and 0·1 M for hydrogen ions respec-
electrically connected to enable electrochemical tively, then at 25°C:
measurements to be made. Each half cell has an
electrode potential (but cannot produce a cur-
8·315xT (H+reference)
rent); it is the potential difference between the EMF- ·2·303·log lll -----:----
two, when connected together, which we are 1x96500 (H+unknown)
able to measure with a millivoltmeter.
For the measurement of a potential differ- =0·OOO1983·T·(logI01-logI01O- 1)
ence, electrical continuity is required within the =0·0591 (0+ 1)
cell. To satisfy the conditions of Ohm's law, the =59·1 mV
measured voltage, or potential of a cell is
determined as the product of the current and ie a lO-fold (decade) difference in activity for a
resistance of the circuit (E=IR). If there was a monovalent ion at 25°C will produce a potential
lack of electrical continuity, the resistance difference of 59·1 mV for a perfect electrode.
would be "infinite and the current would be This relationship applies at all concentrations at
zero. The resulting potential would be indeter- which the Nernst equation can be applied to the
minate. For a cell potential to be measured, a response of the electrode. The relationship
small, finite current must flow around the applies equally well to metal electrodes such as
circuit. In practice the current may be as low as the zinc model shown earlier (though changes
10- 10 amps. in activity rather than concentration will be
A salt bridge is usually chosen for electrical reflected by this relationship).
contact in the model shown in Figure 2. A The Nernst equation shown here is only
direct contact between the two solutions with applicable to an ideal cell, in which there are
flow restrained by a ceramic plug will also only two sources of potential. There are other
suffice. This type of contact, which will be potentials which must be incorporated into the
described later, is widely used. equation, one of which-the junction potential--
Under ideal conditions, the potential differ- is caused by the introduction of the salt bridge.
ence of the cell, which should be measured at
minimal current, should be equal to the differ- JUNCTION POTENTIALS
ence in electrode potentials of the two cells. In The junction potential arises at the junction of
practice, a third potential created at the june- two dissimilar solutions when the mobilities of
tiO:1S between the salt bridge and the two some of the ions at the interface differ, resulting
Electrodes in clinical chemistry 463
I
I En
M+
I
(. I
I
, M+
I Relative error=---, where n is the valency.
0·2568
I x-I-' x-
I I I There are three approaches used to reduce
M+ • I I M+
I M+ x- the error due to the junction potential.
I x-~ x- To change the nature of the electrode. One
I'. , M+
M+ I I
alternative to using the calomel electrode is to
I X-~ X- use a second measuring electrode as depicted in
I I Figure 4, immersed in a solution similar to the
-----
Charge separation,
giving rise to
one being analysed. An attribute of reference
electrodes such as calomel or silver-silver chlor-
junction potential (Ej) ide is their low internal resistance (impedance).
The impedance of an electrode is in parallel
FIG. 3. Representation of a liquid junction of a with the input impedance of the voltmeter used
solution of M+ X- with water. If M+ diffuses more
rapidly than X-, a potential will be established at the for detection. The two impedances should be
junction (Ej). matched," otherwise one will short the other (to
ground). Standard pH meters, used as millivolt-
meters for most ion-selective electrodes, have a
in a charge separation at the interface, as high-impedance input for a high-impedance
depicted in Figure 3. The potential of the cell glass electrode, and a low-impedance input for
described will now be: a reference electrode.
If a high-impedance (eg glass) electrode is
E=Em-Er-Ej used in place of a calomel or silver-silver
chloride reference electrode, a special volt-
where Em is the measuring electrode potential, meter with two high-impedancevinputs would
Er is the reference electrode potential and Ej is be required.P '
the junction potential. To change the physical shape of the electrode.
The junction potential may not only be large, The conventional reference electrode, which is
but may also fluctuate by several millivolts. a vertical calomel electrode with a ceramic plug
Since, for a monovalent ion, a change in at the bottom forming a junction, is far from
potential of 59·1 mV represents a decade (10- ideal. This electrode, which owes its popularity
fold) change in concentration, an error of 1 mV to hardiness and low KCl leakage (approxi-"
would produce an error in concentration terms mately 50 IJ.Uday), may be acceptable for the
of approximately 4%4 measurement of large changes of an ion, eg an

Like electrodes

-: FIG. 4. A system employing similar electrodes


for both sample and reference electrode.

/
Unknown
solution
Liquid junction
\
Constant solution
similar to unknown
464 Hirst and Stevens

H+ change of two decades, with a correspond- lyte in the reference electrode, the junction
ing change in EMF of approximately 120 mV. potential may be kept constant or reduced to a
The detection of small changes of, say, negligible voltage. The following three solu-
0·2 mmolJL of potassium at a concentration of tions to this problem have been proposed:
5 mmol/L, is not so easy. This is equivalent to (a) use of a double junction. eg
approximately 1 mV EMF compared with a Pt;H2!S!! NaCI(0'15 M) !!KCI (sat)! Hg 2Cl2;Hg
possible junction potential variability of several (S=sample,!! = junctions, ; = phase boundary);
millivolts. (b) addition of an inert electrolyte to the
The size and variation of the junction poten- sample as used in the Harned cell for the
tial may be reduced by having the KCI below determination of the pK of a weak acid. eg
the test solution, and by the introduction of Pt; H 2! S, Na2Cl04!! HCl!AgCI; Ag
cylindrical symmetry. 2 Several alternative (S = sample,!! = junctions,; = phase boundary);
liquid junctions have been tried on a commercial (c) use of salt M+X- where M+ and X-
basis, most of which have a more stable have similar mobilities, eg KCl.
potential than the ceramic plug. A few of these A titrimetric system has been described" in
are illustrated in Figure 5. The advantage of which both options (a) and (b) were used,
these junctions is that they reduce the effects, together with similar electrodes (fluoride) for
mentioned earlier, of gravity and symmetry. both sensor and reference,
To change the composition of the electrolyte. LaF3 !Fx, KN0 3 (0·1 M)!!KN0 3 (0·1 M) !! KN0 3
It is possible that, by correct choice of electro- (0·1 M), Fs!LaF

Calomel
electrode

Calomel
electrode

KCL
Porous
ceramic
plug Liquid
junction Cellophane
membrane

(a) (b 1

Ag/AgCL
reference
Liquid
junction
~_ISE
Buffer =..;....
....".--~~----.....,r------.=
~
-.=-- Sample

(c 1
Waste

FIG. S.. Ex.mples of alternative reference cell junctions: (a) a conventional dip electrodewithstaticsample and
KCI; (b) a ellophane membrane junction (Instrumentation Laboratories), with stopped flow sample and static
KCI; (c) a.liquid. junction (Technicon Ltd) with flowing sample and reference buffer.
Electrodes in clinical chemistry 465

(Fx and Fs are unknown and standard fluoride encountered, a suitable reference electrode
solutions). may be the thallium-amalgam thallous chloride
In this system, the left-hand side of the cell is electrode. It has been claimed that this elec-
kept constant, and the right-hand side is titrated trode does not exhibit a temperature hysteresis
with fluoride to produce a null reading. Of the effect. This electrode is commercially available,
three solutions, the most common by far in but it is not in common use in clinical measure-
clinical use is the third, in which the reference ment systems.
half cell consists of a calomel electrode To summarise, for the choice of reference
(Hg;Hg2Cl2 ) in contact with saturated or 3·5 M electrode in the detection of small changes in
KCI. Potassium chloride is chosen because K+ concentration of an ion, the following points
and Cl" have similar (but different) mobilities. should be considered for low, constant liquid-
The choice of concentration is influenced by the junction potentials:
finding that for a single junction cell, the (a) there should be a good liquid-liquid
junction potential decreases with increasing junction, constantly replenished with cylindri-
KCl concentration.? cal symmetry if possible;
A drawback associated with the calomel (b) the reference electrolyte should be satu-
electrode is that it exhibits a temperature ratedl3·5 M KCl if a calomel electrode is
hysteresis effect. While this effect is not very chosen with a static system or a solution similar
large within the temperature range encountered in composition to the test solution if a silver-
in clinical laboratories (2D-37°C) , the use of this silver chloride electrode is chosen with a
electrode is contra-indicated if wide variations continuous flow system;
in temperature are likely to be encountered. (c) the entire measurement system should be
This electrode is unsuitable for use in chlor- isothermal."
ide measurement systems. An alternative is It should be noted that if the choice of
Hg;HgS0 4lK2S04(saturated), though this elec- reference electrode is inappropriate, this elec-
trode has larger and more variable junction trode is as likely to be the cause of problems as
potentials." An alternative reference electrode is the measuring electrode.
in common use is silver-silver chloride in
contact with 0·1 M HC!. This electrode has a MEMBRANE POTENTIALS
stability similar to the calomel electrode, but The half cells described up to this point consist
also exhibits a temperature hysteresis effect. of simple electrodes in direct contact with a salt
Silver-silver chloride electrodes are usually solution. These electrodes may be useful for the
chosen as direct-contact electrodes in gas- analysis of pure aqueous solutions, but are of
diffusion electrodes, such as the CO 2 and NH 3 no practical value in dealing with the problems
electrodes, and in continuous flow systems (Fig. of analysis of biological fluids, or any multi-
5). Silver-silver chloride electrodes also have component solution. When placed in a multi-
photo-electric properties and should be component solution, the electrode will no
shielded from strong light. longer respond specifically to the ion under
If large variations in temperature are to be consideration, but will respond with differing

Gloss FIG. 6. An electrode with an ion-


electrode selective membrane (glass), used in
conjunction with a reference
electrode.

Unknown solution
466 Hirst and Stevens

characteristics to all ions of the appropriate surface will be changed if a current is drawn
charge in the solution. from the cell. It is for this reason that a high
The selectivity of response of an electrode to input impedance voltmeter is required.
an ion is improved by placing an ion-selective The membrane potential (Em) has a Nern-
membrane between the multi-component solu- stian type of response:
tion and the electrode. Figure 6 shows an
example of such an electrode. The characteris- RT al
tics of membranes, such as selectivity, will be Em=-log.,
described later. In this section, the potentials nF az
arising in different parts of the cell are under
discussion, and the membrane is described here where al and az are the activities of the salt on
because it is the source of another potential- either side of the membrane.
the membrane potential. This equation is an over-simplification of a
When a membrane electrode, permeable to complex situation, in which variables such as
metal ions M+ (Fig. 7) is dipped into a salt diffusion rates and effects of interfering ions are
solution M+ X- , there will be a momentary flux not taken into account (to be described later).
of M+ ions through the membrane from the The point of. this simple model is to show
stronger to the weaker salt solution. The that, by introducing a selective membrane, the
anions, X-, being unable to migrate through potential which will vary with the concentration
the membrane, will remain in situ, the two of the ion under consideration is no longer at
populations separated by the membrane. A the electrode surface, but across the selective
charge separation will occur between the membrane. Although the nature of this poten-
mobile M+ ions entering the membrane and the tial is not quite identical with the electrode
static X- ions remaining in solution. This will potential, and may vary with the nature of the
result in a potential in the same way that a membrane, the response may be described as
charge separation is responsible for junction Nernstian under some conditions. The variable
potentials. component of the potential of conducting mem-
Since electrical neutrality must be main- branes is the membrane--external solution in-
tained, migration of the M+ ions will quickly terfacial potential.
cease, opposed by the potential set up between
the membrane and solution (the membrane STREAMING POTENTIAL
potential). The potential will be related to the Liquids flowing through tubes, as encountered
differences in activity of M+ on each side of the in continuous flow analysers, may give rise to
membrane. There will be no further net move- streaming potentials." The potential produced
ment of M+ ions from one side of the mem- is dependent on several factors including the
brane to the other. flow rate, di-electric constant and length of the
The equilibrium established at the membrane tube, and is also inversely proportional to the

M+

I J.l.~t
-
t-
Hydrated layer
----------------
4~
M+

FIG. 7. An expanded representation


of a glass membrane.
Dry layer

Membrane-salut ian
interfacial
patent ial
EKternal salut ian
Electrodes In clinical chemistry 467

conductivity of the solution. Peristaltic pumps Test electrode Reference (calomel)


produce fluctuating streaming potentials which electrode
may be detected by the electrode. It is often
necessary to remove these potentials from
continuous flow systems by electrical isolation
and/or shorting from the liquid to the shielding
of electrode. ~ ,
SUMMARY
The EMF produced by a membrane electrode is
composed of internal electrode potentials, Et,
t- 1-
Er; a junction potential, Ej; and a membrane
potential, Em (Fig. 8). For a test solution and a d
cationic electrode, the EMF generated, Ex, ~
Er ~
may be described in simple terms by the
equation:
Et
Ex=Et-Er+Ej+Em.
Em Ej
RT
If Em=- loge ax, where ax is the activity
nF
of Ion X:
FIG. 8. A typical membrane electrode-reference
electrode, with an indication of the potentials in-
RT volved.
Ex=(Et-Er+Ej)+- log, ax
nF

RT they have a Nernst type of response over


Ex=E u + - log, ax or, for an anionic a limited range of concentration, typically
nF
10- 5_10- 1 M.
responsive electrode:
SELECTIVITY
RT Electrode response
Ex=Eo - - loge ax. All electrodes in current use respond in varying
nF
degrees to more than one ion. It is sometimes
possible to suppress the contribution of inter-
EO is a function of the internal potentials and ferents by modifying measurement conditions
junction potential, and is only constant if (eg pH). For clinical applications, it is often
temperature and junction potentials are con- difficult or inappropriate to change conditions,
stant. or remove interferents. It is important, there-
fore, to be aware of the response characteristics
Ion selective electrodes of an electrode.
The membrane potential of an electrode
Electrodes have been constructed which re- includes contributions from all the ions to which
spond in a relatively specific manner to a the membrane is responsive, each weighted by
chosen ion. By varying the composition of the the selectivity of the membrane to that ion. In a
membrane, a large variety of electrodes have simple form, the Pungor generalisation of the
been produced with a selectivity which gives Nicolsky-Eisenrnan" equation may be written
them a useful role in clinical areas. as:
There are three basic types of membrane: the
solid state, the liquid ion exchanger, and the RT
EMF=Eo+-loge(Ai+Sijl·Ajl+Sij2·Aj2 .... )
neutral carrier membranes. The mechanism of nF
action of these three types differ, particularly
with respect to selectivity and establishment of where Ai is the activity of the major ion being
membrane potentials. The end result, however, measured, Aj 1,Aj2, etc are activities of in-
which is important to clinical chemists, is that terfering ions of the same charge, Sijl , Sij2, etc
468 Hirst and Stevens

are the selectivity constants of the membrane ion exchange preference (Kij) for potassium
towards intefering ions. over sodium is approximately lOO-fold, but this
The contribution of each ion to the mem- is counteracted by a membrane mobility of
brane potential is the product of the activity of sodium ions which is 10 times greater than
the ion and the selectivity constant for that ion. potassium ions. The result is a lO-fold selectiv-
Selectivity constants are relative to the major ity for potassium over sodium.
ion (Ai), the selectivity constant of the major Liquid ion exchange membranes have a
ion being unity. rather more complex mechanism for the deter-
This relationship is an approximation, and mination of selectivity, while for neutral carrier
may vary with the type of membrane and membranes the selectivity constant is simply
composition of the electrode internal solution. related to the equilibrium state."

Selectivity constant SOLID ION EXCHANGE MEMBRANES


The selectivity constant may be defined and Glass membranes'"
estimated in several ways, depending on the The most common electrode of this type is the
type of membrane. For a solid ion exchange glass electrode, used as a sensor for H+ or Na+
membrane (eg glass), the selectivity for the in biological solutions. Glass membranes con-
membrane between two ions 1+ and J+ is sist of a three-dimensional matrix of a varying
dependent on the mobility of both ions in the mixture of oxides of elements such as silicon
membrane, and on the equilibrium constant and aluminium (oxidation state III), with
(Kij) of the exchange of ions between the oxides of elements such as sodium or calcium
membrane and aqueous phase.? (oxidation state I or II). The most common type
of glass used is sodium aluminium silicate. The
V'
.. ] K"IJ
SIJ=-X membrane acts as a cation exchanger; monova-
Vi lent cations (eg H+, Na+) being the most
mobile charged species in the matrix.
where Sij is the selectivity constant, Vi and Uj
are the mobilities of ions in the membrane, and
The hydrogen electrode
Kij is the equilibrium constant of the reaction,
A thin glass membrane consisting of three
Kij layers separates two solutions of different hy-
J+ (membranej--I" (aqueous) drogen ion activity (Fig. 9). After a conditioning
:;:=:J+ (aqueousj--I" (membrane). period the two outer layers become hydrated,
while the middle layer remains dry. Hydrogen
In the case of the glass potassium electrode, the ions from the internal and external solutions

Hyd rated lcyers

'" '"
'"
/ \ I
I
" I
"
"" ""
....
I
I
....
..... I
Glass electrode ..... H+ H+ I
I
H+ H+
in standard .....
.... I
I
or unknown .... I I
solution
Glass membrane
(expanded)

~IG. 9. A glass electrode responsive to hydrogen ions, showing the differing potentials on either side of the dry
layer.
Electrodes in clinical chemistry 469

exchange with hydrogen ions in the two hy- membrane. To be suitable for use as a
drated layers, giving rise to a membrane poten- membrane.F the crystal should be:
tial. (a) chemically inert;
The membrane p, tential is related to the (b) highly insoluble;
difference in hydrogen ion activity on either (c) non-porous;
side of the membrane: (d) mechanically strong in thin sections.
The crystal should also be capable of watertight
sealing to an electrode body and have a rapid
Nernstian response.
The mode of action of these membranes is
not quite the same as the ion-exchange phe-
where a1 and a2 are the activities of H+ on nomenon of glass. Conduction through the
either side of the membrane. crystal occurs by the movement of mobile ions
If the hydrogen ion activity of the internal into adjacent defects in the crystal lattice. The
solution is constant, then size of crystal defects restricts the movement to
one particular species of ion. Ions which are the
RT wrong size or charge cannot move in this way,
E=const+- loge a1 and hence cannot contribute to the conduction.
F
Membranes constructed from crystals with a
very low solubility product tend to be highly
Although there may be a small, but finite, selective. The smaller of the favoured ions tend
current when used in conjunction with a refer- to be the most mobile. This has a bearing on the
ence electrode, hydrogen ions do not travel mechanism of action of the electrode which will
across the membrane. There is movement of be discussed in the section on halide electrodes.
ions through the hydrated layer, but the current Crystals of lanthanum fluoride and silver
in the dry layer is thought to be mediated by sulphide have the necessary properties for use
charge transference, eg from one atom to an as membranes. Silver halides alone, although
adjacent one, with a small amount of move- highly insoluble, are susceptible to photo-
ment. electric effects. Silver chloride membranes
Glass electrodes have a very high resistance have, additionally, a relatively high electrical
(600-900 Mohm), and require measuring de- resistance which may present voltage measure-
vices with a high input impedance. ment problems.
The sodium glass electrode The lanthanum fluoride electrode
At a low H+ concentration relative to sodium, This electrode.':' (Fig. 10) may be used as a
the sodium component of the membrane poten- sensor for lanthanum or fluoride, and is the
tial of a hydrogen ion responsive glass electrode
will become significant. 10 This effect is known
as the 'alkali error'. II The alkali error may be
exaggerated by modifying the composition of
the glass matrix to such an extent that under
non-acid conditions, the electrode is highly ~ Internal electrode
selective towards sodium.

Other glass electrodes PTFE


Several other glass electrodes have been pro-
duced, but low specificity, sensitivity or stability
limits their suitability for clinical use. Silver and Internal solt solution
lithium electrodes are claimed to be among the 0·1 M F-
better group. III 0') M CC

Solid state crystal membranes


It is possible to extend the range of ions which
can be estimated by substitution of the glass
membrane with a crystal lattice. A few crystals
1
LoF 3 crystal
·r
share the properties required for use as a FIG. 10. A crystal membrane electrudr .(LaF x)·
470 Hirst and Stevens

basis of a method for the estimation of F- in acts as an inert matrix for the silver halide
blood and urine.!" The membrane consists of a which is in equilibrium with the sample solution
thin crystal of LaF] , sometimes doped with at the membrane-liquid interface. The silver
traces of europium (Eu3+) to increase the ions, being smaller than halide ions, are the
conductivity of the crystal. mobile species within the membrane.
Crystal lattice membranes suffer interference At the membrane surface, the activity of
from ions of like size and charge to the one silver in solution (aAg ") is given by:
being measured. In the case of fluoride, the
greatest interference is from OH-. This in- K (AgX)
terference is suppressed by buffering the a Ag"
sample at a pH 5-5·5 using total ionic strength
adjusting buffer (TISAB).15
where aX is the activity of halide, and K (AgX)
is the solubility product. The EMF of a halide
Silver sulphide-silver halide membranes
electrode is:
Silver sulphide is well suited to the re~uire­
ments of a crystal lattice membrane. 2 In
RT
addition to suitable physical properties, silver E=EO+- loge a(Ag+)
sulphide is not readily oxidised or reduced, and F
is extremely insoluble. An electrode with a
RT K (AgX)
silver sulphide membrane is capable of detec- =El)+ _ log
tion of Ag+ or S2- down to a concentration of F e aX-
the order of lO-H M. l3
RT
The use of pure silver halides for the = E lI ' - - log aX
construction of halide selective membranes is F e

associated with problems, but when silver


halides are incorporated with a silver sulphide
ie the electrode responds to the halide activity.
matrix, a series of stable and relatively selective
There is a clinical use for the chloride electrode
halide electrodes may be produced, as shown in
in the measurement of plasma and sweat'" 17
Fig. 11. In these membranes, the silver sulphide
chloride. Little use has been made of this
electrode for plasma estimations, possibly due
to the effect of proteins. The membrane should
be protected from direct contact with protein
with a semi-permeable membrane, such as
cellophane, to prevent precipitation or adsorp-
tion of protein on the membrane surface.
Internal
electrode
Careful consideration should be given to the
choice of reference electrodes when estimating
chloride, since a saturated KCL salt bridge may
contaminate the sample. A double junction
Internal reference electrode is often used, with, eg, a
0+- f-
elec t ro Iyte potassium-nitrate bridge in contact with the
specimen.

Silver sulphide-metal sulphide membranes

. =f 0·1 mm
Replacement of the silver halide with a metal
.. sulphide may produce a membrane which is
responsive to the metal in question.V Elec-
0'5-1 cm
trodes have been produced using this principle
I which are responsive to copper, lead and
cadmium. These electrodes have a limited
clinical application because, even if they had
Compressed disc of a mixture the required sensitivity for blood concentra-
of Ag 2S and AgCL tions, the specimens would require pre-
FIG.11. A halide. electrode with a mixed Ag 2S/ treatment to release the metals from the proteins
AgCl membrane. to which they are bound, or from erythrocytes.
Electrodes In clinical chemistry 471

LIQUID ION EXCHANGE ELECTRODES


Liquid ion exchange membranes
A liquid ion-exchange mernbrane'" consists of a
water-immiscible solvent which contains an
ionisable species preferentially soluble in the
solvent phase, eg, a fatty acid. This membrane
separates the inner electrode solution from the PTFE body
outer sample solution. For physical stability,
the membrane is often supported by an inert
porous matrix such as cellulose acetate. This
~embrane .differs from the solid ion-exchanger
In that the IOn exchange sites are mobile within
the membrane.
If the membrane separates two aqueous
solutions of a salt M+ X-, as illustrated in
Figure 12, the cation exchanger, S-, will be
Exchangeable
capable of selectively binding and transporting
end-cap
the M+ ions across the hydrophobic membrane. Ag/AgCl
The X- ions on either side of the membrane --.Q<'"
internal Internal
electrode solution
will be excluded and will remain in situ. In a CaCl 2 (10- 3 hi)
~imilar wa~, the cation exchanger, ideally being
Insoluble In water, will remain trapped within
the membrane. Membrane
FIG. 13. A calcium electrode with a liquid ion
Selectivity exchanger membrane.
The behaviour of the membrane in determining
the selectivity of the electrode is complex and
The calcium selective membrane consists of the
varies with the solvent. A solvent with a high
calcium salt of an alkyl phosphate dissolved in
di-electric constant will contain a dissociated
di-n-octylphenyl phosphonate, supported in an
ion exchanger. In this case the selectivity of the
inert matrix such as PVc. 19
membrane is dependent only on the solvent. As
The calcium electrode has sufficient selectiv-
the di-electric constant of the solvent is de-
ity over magnesium, sodium and potassium to
creased, the association of ion and ion-
be useful for serum calcium estimation, particu-
exchanger increases, and the relative mobility
larly free (ionised) calcium. These electrodes
of ions and interfering ions (and binding sites)
have a high resistance (approximately 10
become more important, though not to the
Mohm), a detection limit of 10- 5 rnol/L Ca 2+
and a slope of 25-29 mV per decade change i~
same extent as in solid ion exchangers.
concentration. (Calcium is divalent.)
The calcium electrode
The electrode of this type in most common
NEUTRAL CARRIER ELECTRODES
clinical use is the calcium electrode (Fig. 13).
Neutral carrier membranes
Membrane Solution 2
Analysts have long envied the ability of living
Solution I
cell membranes to transport ions and small
compounds, often against opposing concentra-
tion gradients, with a very high selectivity.
Progress was made towards reproducing this
characteristic with the discovery that macro-
x- _ _ S----..J- x- cyclic antibiotics caused an increased ion per-
meability of cell membranes.r'' Membranes
1~ constructed of the antibiotic in phospholipid
bilayers or organic solvents exhibit the same
1 - - - MS - - - - . I
property. The antibiotic, valinomycin, and a
synthetic analogue, dibenzo-18-crown-6 (a
"!'
FIG. 12.. liquid ion exchanger membrane, showing 'crown compound'S")' (Fig. 14), are ring com-
permeability to cation M+. pounds containing several ring oxygen atoms.
472 Hirst and Stevens
solvents (usually nitroaromatic) supported in an
inert matrix such as PVC or silicone rubber
(Fig. 15). The potassium electrode has a
measuring range of 10- 5_10- 1 mmoUL and a
response of 57-59 mV/decade. The membrane
has a limited life because valinomycin is slowly
leached out into the aqueous phase. Substitu-
tion of the valinomycin by crown compounds
with hydrophobic substituents should eventu-
ally improve membrane life.
A summary of electrodes suitable for clinical
use is shown in Table 1.

valinomycin Other electrodes


In addition to the basic ion-selective electrodes
described above, there are other types of other
electrode with a different mode of operation,
which may be direct, or, more often, indirect.
The three further categories are solid state, gas
dialysis and enzyme electrodes.

SOLID STATE ELECTRODES


The EMF generated by a cell consists of several
dibenzo-18-crown- 6 constant potentials and the membrane poten-
FIG. 14. Valinomycin and dicyclohexyl-18-crown- tial, which is variable. Since there is no require-
6. 4 ment for the internal reference electrode of the
measuring half cell to produce a potential (this
The oxygen atoms may act as a substitute for being constant), attempts have been made to
the hydration shell of some cations, eg, potas- produce direct contact between electrode and
sium, rubidium and caesium. These cations can membrane, omitting the internal aqueous
move from aqueous solution to the ring com- phase. A functional solid state sensor may be
pound and can be selectively transported produced if there is electron conduction
through a hydrophobic matrix containing the through the membrane and reversible electron
compound. exchange to replace reversible ion exchange at
the inner surface of the membrane. The
Such complexing agents (unlike the ion-
membrane--external solution interfacial poten-
exchange carriers) are electrically neutral, and
tial is the variable being measured. The most
carry the charge of the cation in the complexed
form.

Selectivity
Assuming that the different ion complexes of
the macrocyclic molecule have similar mobili- r--t;:::::=l---l- electrode
Ag/AgCL
ties, selectivity is largely determined by the
relative affinity of cations for the carrier, and is Internal
independent of solvent. The simplicity of this --+--+-- solution
function of selectivity should result in large 0·1 M K+
future expansion of the use of this type of
membrane.
Valinomycin
membrane
The potassium electrode
This is the most common example of a neutral
carrier type of electrodef in clinical use. The
most Common carrier is valinomycin rather than
crown compounds, dissolved in one of several FIG. 15. A potassium electrode.
Electrodes in clinical chemistry 473
TABLE 1. A summary of electrodes suitable for clinical use

Concentration Special
Electrode range (M) pH conditions Application

Ammonia 1~-Hf >11 adjust pH NH3 , serum, urine


(NH 3 ) NH3 , as a reaction
product, eg, urease
Carbon dioxide 1O-5_W--2 <pH 5 for total CO 2 PC0 2 , blood, serum
reaction product
Calcium 1O-5- Hf 6-8 Calcium, blood
serum, urine
Chloride 10-5_10° 2-11 Requires a double Chloride, sweat
junction ref., serum, urine
poisoned by protein
Fluoride 10-7_10° S-8 Sample requires Fluoride, serum, urine
buffering
Iodide 10-7 - 10" 3-12 1-, reaction product
Potassium 1(}-5_100 3-10 Requires a double Potassium, blood
junction ref. serum, urine
Sodium 1~-101l >pNa+3 Sodium, blood
serum, urine

Electrodes with limited clinical use Limitation

K+fNH 4+ 1O-5_Hi1 Low selectivity, interference from Na+, K+ and NH 4+


Lithium 1(}-5_100 Selectivity, interference from Na" (SN.=5·1O- 2 )
Other metals 1~-10" Most metals are protein-bound and would require sample
Pb, Cd pretreatment
Ag, Cu

stable readings are obtained if an electrode The two most common variations of the solid
membrane has dimensions (area, thickness) state electrode employ membrane contact
which produce a DC resistance of 1-100 Mohm. either via carbonlgra~hite, eg, the Pungor-
The input impedance of the measuring device Radelkis electrode/!' 2. (Fig. 16) or directly to
should be of the order of 1000 X the resistance metal. The physical form of direct metal contact
of the electrode." varies from electroplating of glass to coating a
metal wire with a membrane" (Fig. 17). Coated
wire electrodes have been produced for calcium
and potassium estimations, but are not used
routinely.

11-4-tt-- Ag
Interna I
t--H-- electrode
Glass

___ Graphite
Membrane
coating

- - - - Ag/AgCL
- Membrane
FIG. 16. A solid state electrode using a graphite FIG. 17. A coated wire electrode, consisting of
contact. anodised silver coated with a selective membrane.
474 Hirst and Stevens

ION SELECTIVE FIELD EFFECT present ISFETs have a limited active life (1G-20
TRANSISTORS (ISFET) days). Mass production of stable ISFETs would
Bipolar (normal) transistors are current-driven provide inexpensive assays for many analytes.
semiconductors, ie, they amplify a current.
They have a low input impedance, and are not GAS DIALYSIS ELECTRODES (NH3 AND COz)
suitable for the first (input) stage of an elec- In this type of electrode ,2M. 29 the gas to be
trode voltmeter. Field-effect transistors (FET) determined diffuses across a membrane, fol-
are voltage-driven semiconductors; amplifica- lowed by detection of the gas on the sensor side
tion is related to the voltage applied to the of the membrane. In two cases (carbon dioxide
device which produces an electrostatic field. and ammonia) the gas changes the pH of a
FETs do not consume current, operating only buffer in contact with a pH electrode (Fig. 19).
on voltage. They have, therefore, a high input Since the gas is neutral, movement across the
impedance, and are widely used for the first gas membrane is 'halted not for preservation of
stage of an electrode voltmeterv: 26 (Fig. 18). electrical neutrality, but upon achievement of
The ion selective component of this sensor is equilibrium of the buffer reaction. For the CO2
similar to membranes described earlier. electrode the reaction is:
The elimination of intermediate processes is
taken one step further than with the coated wire
electrode by connecting the first stage of the
millivoltmeter (a field-effect transistor) directly The production of carbonic acid (first step) is
with the ion-selective' membrane. The advan- the rate-limiting step, and this, rather than
tage of this arrangement is that the high diffusion through the membrane, is the reason
impedance of the electrode is transformed for the slow response of CO 2 electrodes. The
directly to the low impedance output of the rate of reaction may be increased by the
transistor, eliminating the need for heavy addition of carbonic anhydrase (eg, lysed red
shielding. cells), though this is not feasible for everyday
The attraction of this approach is that with use.
current micro-electronic technology, it should The membrane is a gas-permeable hydropho-
be possible to construct a cluster of electrodes bic material such as polypropylene, teflon, or,
on one chip, for possible in vivo applications.f" more recently, silicone rubber. The ammonia
ISFETs have been developed for Na+, K+, electrode incorporates nonactin, a macro cyclic
Ca2+, H+, and Nat and have been used in neutral carrier, in a polytetrafluoroethylene
vivo. Unfortunately development of this sensor membrane.'?
has been retarded by problems of electrical For clinical use, selectivity between CO 2 and
insulation of the chip from the solution. The NH 3 is based on the pH of the external
solution, the ammonia electrode requiring an

lon-selective membrane

Outer ----..; ~
n!2!:i:~~ni4---Gate body

pH electrode

Pt internal
electrode Ag/AgCL
l--- reference
I-
electrode
Buffer
solution -
~

Rubber a-ring ~ ~
FIG. 18. An ion-selective field-effect transistor. Two
areas of silicone with 'negative holes' (neg Si) are f
Gas dialysis
separauI by one with 'positive holes' (pos Si).
membrane
Application of a potential to the gate will induce a
current to flow through the pos Si area. FIG. 19. A carbon dioxide-ammonia electrode.
Electrodes in clinical chemistry 475

alkaline solution to convert NHt to the diffu- will produce a current at constant potential, as
sible NH 3 molecules. Both electrodes have a opposed to a potential at zero current. The
useful clinical function. The CO 2 electrode may response of a polarographic electrode tends to
be used for the estimation of PC02 and indi- be linear, rather than Nernstian.
rectly, total CO 2 in a large variety of fluida." The relationship between voltage and current
The ammonia electrode may be used for the which is seen in oxygen-free solutions at low
estimation of plasma NH 3 , 32 and for the detec- voltage becomes distorted in the presence of
tion of NH 3 as the product of a primary oxygen and other electroactive compounds such
reaction. The plasma or reaction solution must as paracetamol and catecholamines. Plateaux,
be made alkaline prior to ammonia estimation. such as those shown in Figure 20, are found.
The two gas electrodes described respond to Reduction reactions occur at negative vol-
the partial pressure of gas. The response over tages, and the current passing shows a plateau
small changes of concentration is almost Nern- at low voltage which corresponds to energy
stian in that the internal pH electrode has consumption by the following reactions:
Nernstian response to changes in the H+
-0·7 V
activity of the buffer, but deviates according to
the characteristics of the buffering system. 02+2H20+2e----+H 202+20H-
These gas dialysis electrodes are indirect, or
double membrane electrodes, in that the poten- +0·7 V
tial arises not at the gas-permeable membrane, H 202 ---+ 2H+ +02+2e-
but at the glass membrane of the pH electrode.
In this sense, they are comparable with 'enzyme It is possible to estimate dissolved oxygen in a
solution by measurement of the current at a
electrodes', (see Enzyme and other indirect
electrodes) . constant potential of -0·6 to -0·7 V. The
The oxygen electrode shares the characteris- current will be directly proportional to the
tics of the gas-permeable membrane, but has a oxygen consumed which is itself proportional to
different detection system (polarography). the P0 2 of the test solution. Peroxide may be
similarly estimated if a potential of +0·7 V is
applied. The earliest form of polarographic
POLAROGRAPHIC ELECTRODES
electrode, the dropping mercury electrode,
Polarographic, or amperometric, analysis in-
suffers from many disadvantages, such as pro-
volves the measurement of current produced in
tein poisoning and lack of specificity, and is not
electrochemical reactions. The electrodes dis-
suitable for routine laboratory use. Alternative
cussed up to this point have been voltametric;
electrodes which do not suffer from these
voltage is measured under equilibrium condi-
disadvantages are now widely used in clinical
tions at negligible current. Polarographic
laboratories.
systems are designed to measure a current
A large number of electroactive compounds,
produced by an electrochemical reaction at an
electrode surface. A polarographic electrode such as those shown in Figure 21/1 are detect-
able by this technique. Oxidation (positive
voltage) or reduction (negative voltage) condi-
tions, or a combination of the two, may be used.
Current (/Lamps) If 100% of the analyte is oxidised (or reduced),
the technique is referred to as 'coulometric
analysis'.

THE OXYGEN ELECTRODE


+ The Clarke oxygen electrode in common clinical
Valtage (V) use is an amperometric electrode based on
polarographic detection.'! Solid metal elec-
trodes are protected from protein poisoning by
a gas-permeable membrane (eg, polypropylene).
The dimensions of the Clarke electrode (Fig.
22) are critical. The diameter of the cathode tip
must be such that there is not excessive con-
FIG. 20. The relationship of current to voltage for sumption of oxygen. If oxygen consumption is
solutions with and without electroactive compounds. too high, diffusion nither than partial pressure
476 Hirst and Stevens

becomes rate-limiting. As gaseous oxygen dif- be difficult to measure. As a compromise, the


fuses more rapidly than dissolved oxygen, a gas platinum tip will be 1-100 urn in diameter, and
and a liquid with identical P0 2 will yield will yield a current of the order of 10 uamps for
differing currents. On the other hand, a very a normal blood Po 2 •
small cathode will yield a low current which will The oxygen electrode in this form is in
widespread use for P0 2 estimation of blood and
Reduction Oxidation other aqueous solutions, and is used as an
oxygen detector in indirect systems (see En-
zyme and other indirect electrodes).

-Olefins

Esters
Hydrocarbons
4

Azinesl
triazines

The selectivity of the electrode for oxygen is
derived from both the membrane and the
polarographic detection. Some foreign volatile
• compounds, eg, halothane.P' have been re-
Aminesl
Ketones amides ported as interferents, but normal blood consti-
tuents do not affect the estimation.
Aldehydes Phenathiazines
• OTHER POLAROGRAPHIC ELECTRODES
Olefinic Peroxide
esters Phenols Amperometric detection of peroxide is the
basis of an indirect enzyme electrode for glucose

-
Organometalics Aromatic OH estimation which is in widespread use. (See
Enzyme and other indirect electrodes).

-Diazo

Nitro
Quinoli nes

Imines
Electrochemical detectors for HPLC
Polarographic, or 'electrochemical', detection
has been successfully used in HPLC for com-
pounds which would otherwise re~uire deriva-
Halogens Halogens
... .........-..... tisation for optical detection.P: 33, 5 In the case
I
I
of catecholamines, the reaction at the electrode
,
I is:
-2 -I 0 2
HO~R +O.5V O~R
(Volts vs calomel electrode)
FIG. 21. Electroactivity of organic compounds. HoN ----- oN + 2H+ + 2e-

Gold, and gold amalgam electrodes have


been commonly used for (oxidative) polaro-
graphy; gold amalgam being particularly useful
in detecting thiols. There is considerable interest
in clinical chemistry in the development of a
+lfH--Ag anod e carbon electrode with a highly polished, hard
surface (glassy carbon). Typical working ranges
of the three electrodes are shown in Figure 23.
Electrolyte Voltage stabilisation is necessary because when
a current flows there is a potential drop
between the two electrodes, which will vary
with the composition and flow rate of the
solution. A constant voltage is maintained by
the introduction of a third 'counter' electrode
t4t-f5I-:-- Pt cat hod e
into the measurement cell, with a potential
Rubber O-ring which is controlled from the reference electrode
via a feedback circuit. An example of such a
Gas permeable circuit is shown in Figure 24. Continuous use of
Site ",f reduction membrane a polarographic cell may result in the accumula-
(inside 'hemembrline)
tion and adsorption of reaction products on to
FIG. 22. A C!lirkeoxygen electrode. the electrode surface. The electrode will then
Electrodes in clinical chemistry 477

become inactive. A further electronic refine- the HPLC solvent, and continuously purging it
ment to maintain the activity of the electrode is with nitrogen or helium.
to apply alternating pulses of a polarising
ENZYME AND OTHER INDIRECT
potential and a lower 'cleaning' potential at 36
E LECTROD ES
which the electrode reaction is reversed.
This section is concerned with electrodes with a
In practical terms there are two further
membrane in addition to that of the sensor. The
effects which should be considered for HPLC
gas-permeable membranes were described
applications. One is pulsation of the liquid flow,
separately because they allow gases to diffuse
producing streaming potentials which will be
unchanged through them. The membranes to
detected by the electrode. A pulse-free pump is
be described in this section allow the selective
essential for low background noise. The second
conversion of substances to products which are
effect is produced by the presence of dissolved
readily detectable by electrochemical means. A
oxygen. In the reduction mode, dissolved
prime requirement of this system is that either
oxygen will produce a high background current
the membrane or the detector must be selec-
(Figure 20). In the oxidation mode (positive
tive.
voltage), high noise signals are often found,
possibly due to the transient production of The urea electrode
electroactive species. Both the background One of the earliest enzyme electrodes, the urea
current and noise can be reduced by de-gassing electrode" consisted of an ammonium (NHt)
electrode enveloped by urease, as illustrated in
Figure 25. This electrode was unsuitable for
Glassy
car ban

u:;
c.
E
c
:L
....~ NHl glass
Illet radl

Cillophoni

0·1 0'8 1·6 Ureon


Valts
FIG. 23. Working ranges for three electrochemical
electrodes: response to thiols . FIG. 25. A urea electrode.

.....----.,-----{ A)-----,
Warking
electrade

Reference Pawer
supply FIG. 24. Circuit diagram for pro-
viding a constant potential for a
Caunter 3-electrode cell.
478 Hirst and Stevens

biological fluids, due to the poor selectivity of


the NHt electrode. More specific systems have
been described" using the more selective
ammonia electrode.
An instrument which is widely used (Beck-
man Instruments), for electrochemical urea N
determination is based upon the change in o
It
conductivity of a sample following the addition
of urease, due to the reaction:
NHz CONH z + 2H zD---+2NH 4+ -cos
The glucose electrode
The electrochemical detection of glucose is
based upon the action of glucose oxidase which 30 60
catalyses the reaction: Time (s)

glucose--Oj-e HzD---+gluconic acid+ HzO z FIG. 27. The change in Po; with time in a glucose
electrode exposed to air.
There are two types of glucose electrode in
common use, both of which use polarographic
The glucose-peroxide electrode. The
detectors. The first, and earliest, records the
electrode." illustrated in Figure 28 forms the
consumption of oxygen, or decrease in POz,
detector of a widely used glucose analyser
while the second records the production of
(Yellow Springs Instruments Ltd). The mem-
peroxide.
brane contains immobilised glucose oxidase at
The glucose-Paz electrode. The glucose
which site hydrogen peroxide is formed. The
electrode" as illustrated in Figure 26 may take
HzO z diffuses through the membrane to the
many forms. The glucose oxidase may be added
polarographic cell where it is detected at a
as a wet reagent, or may be immobilised eg on
potential of +0·7 V (see section on polaro-
to the stirrer. The drawback which applies to all
graphy).
these systems is that if the system is exposed to
This electrode does not suffer from the
~ir the oxygen consumed will be replenished,
problem of re-equilibration with atmospheric
ie, the POz will fall then rise until it is again in
oxygen, enabling the total reaction to be
equilibrium with air, as seen in Figure 27.
followed. The membrane is not, however,
Glucose is estimated by using relatively high
entirely selective for peroxide. The system may
glucose oxidase activity and detecting either the
respond to foreign substances such as
peak activity (point P), or the peak rate of
paracetamol. 3Y
reaction. Both approaches require a fast re-
Other enzymes. The use of the polarographic
sponding POz electrode, and may suffer poor
and the potentiometric type of enzyme elec-
precision.
trode. is ~ot restricted to glucose estimation. By
substituting the glucose oxidase with other

Reaction
chamber

Polarographic
......,;...q._ Reactian cell
chamber

Stirring bar P t - - - -....

~ -Membrane

/1 l~--- Oz detector
Ag-----iI

I----t
-
Double layer membrane
Pt cathode Annular silver with glucose oxidase
anode trapped between

FIe. 26. A glucose-Po, electrode. FIG. 28. A glucose-peroxide electrode.


Electrodes in clinical chemistry 479

enzymes, electrodes have been produced for rodes to yield results that may be significantly
urate, cholesterol and ethanol.r" different from those obtained by conventional
photometric methods. Some factors such as
The lactate electrode interfering ions and junction potentials have
This electrode'" 40 uses a yeast lactate dehy- already been discussed. Among these factors
drogenase which catalyses the reaction: are:

lactate + 2Fe( CN)~ - ~pyruvate + 2Fe( CN)~- + 2H + A. Effects on the activity detected by the
PI electrode
2Fe(CN)~--->2.Fe(CN)~- -ze (i) variations in ionic strength and activity
coefficient: both affect the activity of a given
The ferricyanide is detected by polarography
molar concentration of an ion.
with an 80 mV polarising voltage.
(ii) chemical interference.
(iii) space occupying effects of protein and
Other enzyme/redox systems lipid in the sample.
Although many other enzyme electrodes have
been described, such as deaminating enzymes!
B. Factors affecting the total potential of the
ammonia probes, few are suitable for routine
system
clinical use. Several amperometric techniques
(i) selectivity of the membrane.
for the NADH-NAD+ system have been des-
(ii) junction potentials.
cribed, which could have wide clinical applica-
(iii) rate of response.
tions. Unfortunately the reduction of NAD+ at
(iv) temperature.
the electrode surface is complicated by absorp-
tion and dimer formation. As a result, these
C. Calibration
electrodes are not suitable for continuous use,
(i) the nature of the cali brant.
and are at present of academic interest only to
(ii) the algorithm used for calibration.
clinical chemists.
Of these factors, the activity coefficient is the
Future developments most important and is influenced by ionic
(a) ISFETs. The range of ISFETs is expanding strength, particularly if direct measurements
and now includes enzyme and immunochernical are required.
sensors. 26
(b) lmmunochemical sensors. If an
ACTIVITY
antibody-antigen reaction causes a change in
Activity coefficient
charge density, immobilisation of an antibody
As discussed earlier, the EMF of a poten-
could be used to produce a potentiometric
tiometric electrode is related to the activity of
membrane. Attractive though this concept may
the analyte in solution which is related to the
sound, practical systems have not yet been very
concentration of analyte by the equation:
successful.
(c) Bipolar pulse conductance/" This techni- Ai=yi'Ci
que involves the excitation of an ion-selective
electrode using two equal and opposite 3 V where Ai is activity, yi is the activity coefficient
pulses lasting for 100 us. At the end of the and Ci is concentration. Activity coefficients in
second pulse the current is recorded, this being aqueous solutions or blood/plasma are not
related to the interfacial charge transfer and known exactly, though an approximate value
conductance. The advantages of this technique can be derived for dilute solutions using the
are speed (several readings may be taken over a Debye-Huckel equationY
short time) and elimination of a reference half
cell. A metal counter-electrode may be used, A·z 2 vT
eliminating problems of liquid-junction poten- logy=
l+BaYI
tials. A disadvantage is that, at present, the
technique cannot be used for enzyme elec- where A and B are constants, Z is the ion
trodes. charge, a is the ion size parameter (nrn), and I is
the total ionic strength (mmol/L). This equation
Factors affecting ISE measurements should only apply when ! is less than 0·1
There are several factors which cause elec- mrnol/L and therefore is no more than 'an
480 Hirst and Stevens
approximation for plasma. (Appendix A shows Interferents (chemical)
an example of an ionic strength calculation.) The effect of chemical interference is to remove
Plasma and urine present two separate prob- ions by binding so that they are not sensed by
lems in estimation of ionic strength. the electrode as 'active'.
The ionic strength of blood/plasma is gener- In vivo binding compounds. The classic case
ally assumed to be 166-168 mmoVL,43 and is is that of calcium in plasma, less than half of
affected principally by variations in plasma which is free, or 'ionised'. The remainder of the
sodium concentration. A change in ionic calcium is bound to proteins (mainly albumin)
strength of 15 mrnol/L, which is frequently or other organic anions. The binding of plasma
encountered, would result in an error of 2-3 calcium to protein is reversible and pH-
mrnol/L in a sodium estimation. dependent." the calcium activity falling as pH
The composition and ionic strength of urine increases. In the case of calcium, where estima-
varies so widely that direct analysis by electrode tion of the ionised species is of clinical interest,
should be avoided. Unlike plasma, a 'sodium blood specimens must be handled anaerobically
correction' may not provide adequate com- or re-adjusted by titration to the patient's pH or
pensation. An indirect approach is preferable, a standard pH.
in which the ionic strength of urine is adjusted Several other compounds in blood are bound
by dilution with a strong buffer. The rela- to proteins or cell membranes, including most
tionship of activity coefficient to ionic strength of the heavier metals-Cu, Zn, Pb, etc. In these
is depicted in Figure 29. Since the ionic strength cases, and for substances such as salicylate,
of plasma and urine is unknown and variable, barbiturates, etc, the total amount rather than
the activity coefficients of the constituents is the free ions is clinically important. Use of ISEs
similarly unknown and variable. As with ionic to estimate these ions and compounds, even if
strength, estimates of the activity coefficient are there were suitable ones, would require prior
widely used. liberation of the ions from protein.
There is a large variation in activity coeffi- Bicarbonate has been reported to bind
cients of different blood constituents, which are sodiurn.f' though there is some doubt over the
approximately 0·71 for potassium, 0·75 for nature of this effect.
sodium and 0·55 for calcium, though widely Foreign compounds. Calcium is subject to
differing values have been reported.P The further chemical interference, this time from
combination of the low activity coefficient for heparin.t" Two attractions of potentiometric
calcium, and the electrical charge results in a measurements is that they do not require
sensitivity of approximately 14 mV per decade optically clear solutions and should operate on
change in concentration. A reading error of whole blood. While blood is used for acid-base
1 mV will be equivalent to a 15% error for free studies, usually anticoagulated with lithium or
calcium, compared with 4% for potassium. sodium heparin, other popular anticoagulants
While there are several schools of thought on such as citrate and EDTA act by removing
calibration of electrodes for clinical use, com- calcium, and are unsuitable for calcium estima-
parison of the ionic strength of the calibrant with tion. Heparin has a different mode of action,
that of the test solution must be considered. but nevertheless binds calcium to a small but
significant degree. The two alternatives to using
serum are: (a) to control the amount of
Monovalent
heparin; and (b) to render the heparin non-
1'0 binding by presaturation with calcium.
~

.
co 0·75
Protein-lipid dilution effect
.
.."

a
u
0'5 Colorimetric and potentiometric analyses often
'"' differ in the way in which the sample is treated.
> 0·25 In colorimetric analysis the sample is usually
£Cl:
diluted. Electrode analysis is often performed
10- 7 10- 6 10- 5 10- 4 10- 3 10- 2 10- 1 10 directly on whole blood or plasma, since cells,
proteins, lipids, etc, should not affect the
Ionic strength (mmot/L)
reading of the electrode-this being related to
FIG. :l9. The variation- of activity coefficient with the concentration of analyte in the water
ionicstrength foro monovalent and a divalentcation. compartment of the plasma (Fig. 30). Since
Electrodes in clinical chemistry 481

Aqueous Normol serum Pathologicol


stondord serum

r r H2O

r
H2O
H2O II 140 mmol

1r
II
140 mmol No+
II
Abnormal protein

L
140 mmol

No+ L Na+ (ag myelomo) or


1-1Ipid(chylomicrons)

r
x x

\
Normal protein
\
Normal protein
and iipid and lipid
(7 % of serum volume)

FIG. 30. The effect of protein and lipid in plasma on the molal and molar concentration of sodium. In each of
the three cases, a direct ISE measurement will detect a constant molal concentration of sodium (140 mrnol/kg
H 20), whereas indirect methods will detect a variable molar concentration (mmoVtotal volume).

normal plasma contains approximately 7% phate and many other ions is much higher than
protein-lipid and 93% water by volume,42 one in serum. Potassium estimation may be affected
would expect results from direct potentiometric by caesium which is occasionally added to
analysis to be higher than techniques involving quality control materials. Measurements may
dilution by a similar degree. The differences require buffering to a fixed pH as H+ (sodium)
found are usually smaller than 7% (2-5%) or OH- (fluoride) may interfere.
because this effect is counteracted by activity
coefficient, junction potential effects and the Junction potentials
nature of the calibrant used. The effects of junction potentials were de-
The results obtained by indirect potentio- scribed in the section on Electrochemical
metry, when the sample is diluted prior to potentials. Aqueous calibrants may differ from
analysis, usually compare well with alternative biological fluids not only by the activity coeffi-
techniques. The results obtained by direct cients of their constituent ions, but also by the
potentiometry may be clinically more useful," junction potential at the reference electrode
in the cases of plasma sodium and calcium and interface. Although 3 mol/L or > 3·5 rnol/L
particularly when plasma proteins-lipids are KCI is usually chosen to minimise junction
grossly elevated. An alternative to using an potentials, variation of the potential still occurs
adjusting factor which has been recommended'" and tends to have the opposite effect of the
to achieve agreement, eg, with flame photo- plasma protein-lipid dilution effect. The varia-
metry, would be to make it plain that the tion of junction potentials between specimens is
estimation is not the same as 'plasma sodium'. sometimes referred to as the 'residual junction
potential' .
FACTORS AFFECTING THE TOTAL In whole blood, erythrocytes produce a
POTENTIAL OF THE SYSTEM further difference in junction potential with a
Selectivity 'sedimentation effect'-a form of streaming
Ionic interference has been described in the potential. It has been clairned'" that the
section on Electrochemical cells. The majority sedimentation effect is reduced by using a
of electrodes in clinical use have been chosen double junction reference electrode with a
for their suitable selectivity towards other ions sodium formate salt bridge.
in biological fluids, particularly blood/serum. For a situation where the calibrant and
Care should be exercised, however, in using sample solution have differing activity coeffi-
electrodes for urine analysis, in which the cients and junction potentials, the concentra-
concentration of potassium, ammonium, phos- tion of an analyte is given bj the equation.v'
482 Hirst and Stevens

Cs=Cc'explO ( AEC) yc
- - . -·explO (AEj)
--
slope ys slope

where Cs and Cc are concentrations of sample


and calibrant, ys and yc are activity coefficients 101
of sample and calibrant , AEc is the difference
in EMF between sample and calibrant, and AEj
u,
is the junction potential difference between :::;;;: Concentration
sample and calibrant. w change
The last two terms in the expression are error
terms. One approach'! to reduce the difference
in results produced by electrometric and other -10 1
methods is to design the reference electrode
such that the product of the error terms
(derived from the previous equation):

- AEj ) ys
explO ( - - - is unity. Time (s )
slope yc
FIG. 31. The variation in response rate to positive
and negative changes in concentration.
Response time
The response of an ISE to a change in analyte
concentration is slow compared with photo- rate or flow rate of a solution. Figure 32
metry, and may be variable. The rate of illustrates the change in response of an elec-
response of an electrode is affected by several trode with increasing stirring rate. Continuous
factors such as the thickness, dimension and flow systems may benefit in a similar way by
nature (ion exchange rates) of the membrane, using a faster flow rate, provided that this does
change in activity, temperature and stirring not produce a streaming potential.
rate. The relationship between EMF and time is
far from simple,49 being based on 3 or 4 time Temperature
constants, and can be represented by a series of Sensitivity. The EMF of an electrode is
exponentials only in an ideal electrode. For the temperature-dependent, the relationship being
commercially available electrodes in clinical shown in the Nernst equation:
use, the two most important variables affecting RT
response are change in concentration and E=Eo+- log, a
stirring rate. nF
Change in concentration. Electrodes respond where T is the absolute temperature. An
more rapidly to large changes in concentration increase in temperature of lOoC will produce an
than small ones, and more rapidly to increases
than to decreases, though this does not always (I)
apply. Theoretical response times are depicted
in Figure 31. An example of the effect of .L.:.---- (2)
concentration can be seen by comparing the
apparently instant response of a pH electrode
over a range of 2 pH units (2 decades change in u, '".-t:.-------- (3)
:::;;;:
concentration) with the sluggish response w
(5 min) of a fluoride electrode to changes in <J
urine fluoride concentration of 0·1 umol/L
(10-,-7 M). Since response time is so variable, it is
not good practice to take a reading at a fixed
time, unless the time interval is large compared
with the electrode response. The best practical Time
approach is to wait until the rate of change of
FIG. 32. The relationship of response rate to stirring
signal is negligible before taking a reading. rate. As the stirring rate increases from 1 to 3 the rate
Stirring (flow) rate. The second variable of response increases, but the final response de-
which can affect response time is the stirring creases.
Electrodes in clinical chemistry 483

increase in sensitivity of approximately 3·5% and the original unknown concentration:


for a univalent ion and half that for a divalent
ion. While there is little to be gained by u=~C(1O~EJS-1)
moderate increases in temperature, there is a
possibility of inaccuracy if the temperature of This technique can be used for complex and
calibrants and samples is not matched. variable solutions, eg, urine, but it is not
Response rate. The response rate of some practical for high throughput serum analyses.
electrodes is significantly improved with moder-
ate increases in temperature. Calibration curves
An electrode with a Nernstian response should
show a limited linear relationship between log
Calibration (activity) and EMF. Figure 33 shows a typical
electrode response curve and the relationship
CALCULATION between EMF and log activity is mainly linear.
There are several techniques for converting the At the lower end the relationship becomes
millivolts produced by an electrode to concen- unlinear as the limit of detection is approached
tration, most of which are based on the Nernst (or passed, depending on the definition of
equation. 'limit'). At the high end, the relationship with
activity remains linear, but there is a divergence
Direct use of the Nernst equation of the relationship with concentration.
The Nernst equation required is in the form: An electrode may be calibrated by plotting
the EMF readings from two or more calibrant
RT
Ecell=Eind-- loge Ax-Eref+Ej solutions on semi-log graph paper, and assum-
nF ing a linear relationship. The concentration of
yxu=Ax 1 analyte is simply read off the graph. The slope
=antilog lO Ecell+Eind-Eref+Ej of the electrode response is relatively stable.
O·1984·T Most electrodes may be recalibrated with a
single calibrant, and the slope checked at
Eind is a composite of internal potentials of the infrequent (eg, 6 h) intervals. The slope of
indicator electrode and includes potentials from many electrodes is lower than the theoretical
the inner reference electrode and the inner 59·1 mY/decade and often slowly decreases as
phase boundary. The uncertainty of Eind, Ej the electrode ages to less than 55 mV/decade.
and the activity coefficient make direct use of , 'Calibration of electrodes by graph plotting is
the Nernst equation impractical. suitable for clinical applications. This method
has had wide use and is still used in the form of
Standard addition or subtraction pH meters with logarithmic 'concentration'
The effect of adding a known amount of analyte scales. Graph plotting has now been largely
(or removal by chelation), on the EMF of a cell superseded by modern direct readout electro-
can be used to estimate the original concentra- meters.
tion in the sample. For the unknown solution,
Activity
RT
E =Eo+- log u·yx+EJ·
1 nF e

After the addition of an amount ~C of analyte,


LJ..
::!: Concentration
RT w-
E:z=Eo+- loge (u+~C).y~+Ej ~
nF I

if the addition does not significantly alter the


ionic strength and volume of the solution, the
activity coefficient and junction potential
should be unaltered. Then:

~E (ex+~C) RT Activity/concentration (rrvnol/L)


--=loglO where slope=2·303·-
slope ex nF FIG. 33. A typical electrode response curve,
484 Hirst and Stevens

Electronic signal linearisation The electronic approach is not, then, a simple


It is possible to linearise a logarithmic signal solution. Software on a microprocessor may
with an antilog amplifier. Unfortunately an offer a better means of linearising an electrode
antilog amplifier used alone will take the response.
antilog of the whole cell potential, and will not
necessarily produce an output which is linear. Computer algorithms
For the circuit A number of instruments now available use a
dedicated computer or microprocessor to com-
Ex Eout pute the concentration of an analyte, eg,
plasma sodium or potassium. The algorithm
used by the computer will depend on the
composition of the manufacturer's calibrant
solutions. Information on either the calibrant
composition or algorithm is not always freely
available, though this information should be
RT
Ex=Eo +- log, Ax; requested by prospective purchasers. Examples
nF of two commercially-used algorithms are given
in Appendix 2.
RT
put a=- and E=antilogcEO, and: CALIBRATION
Nf An electrode is calibrated by converting the
potential (mV) generated by a calibrant solution
Ex = 10g.,E +10g.,(Ax) to a practical quantity such as activity or molar
= 10g.,(E. Ax") concentration. The conversion is made by the
application of a variety of algorithms of greater
After passing through the antilog amplifier: or lesser complexity (Appendix 2). The choice
of calibrant and definition of its value presents
Eout=E·(Ax)a certain difficulties. To calibrate an electrode
with a primary aqueous calibrant of defined
The output has an exponential, rather than a activity is not possible at present. It would be
linear, relationship to activity. necessary first to find a way of defining and
It is possible to linearise the relationship by calculating activity as has been evolved over
introducing a pre-amplifier with gain b into the many years for the pH scale.
circuit: A more practical solution to the problem of
calibration is to use a standard calibrant which
Eout
. .. is comparable to the solution being analysed.l"
The EMF obtained could then be converted to
concentration units (eg, rnmol/L) and related to
By the same argument, the input to the antilog the activity of analyte in the calibrant. Serum or
amplifier is: blood-based materials used for quality control
in blood gas measurements have been found to
b·log.,(E·Ax a ) be unstable" and would be unsuitable for use
as calibrants. An alternative to serum is an
and the output will be: aqueous calibrant containing physiological
amounts of the major stable ions and buffers,
Eout= antilog.,(b·log.,E·Ax") but omitting the unstable components such as
= antilog c(1og.,E. Ax ab) proteins, in particular haemoglobin. Such a
=E'(Ax)ab calibrant may contain Na+, K+, Ca 2+, Mg2+,
Cl", HC03" and inorganic phosphate, with a
If the gain of the pre-amplifier is adjusted so pH adjusted to 7·40.
that ab=l, ie, b=l/a, then Eout=E·Ax 1= Most manufacturers have adopted a calibrant
(antilog.,Eo)·Ax. ie, the output is directly pro- similar in composition to the one outlined, with
portional to activity, but has a slope dependent a few modifications eg, substituting Tris buffer
on £0 which-may vary. The slope variable may for the inorganic phosphate and adjusting the
be rdroved·with the. addition of a further total ionic strength to 160-166 mmoVL. The
amplifier before Wlift¢r the antilog amplifier. rationale for the adjustment of the ionic
Electrodes in clinical chemistry 485

strength of calibrants is that the ionic strength the calculated activity/concentration and true
of normal serum is around 166 mmol/L. If the activity/concentration. The discrepancy will in-
calibrant has a normal composition (eg, sodium crease with the deviation of the serum sodium
concentration = 140 mmol/L) the activity and concentration from normality. The agreement
jun,ction potentials for calibrant and normal between instruments calibrated in this way and
sera should be similar. In practice, at least two conventional techniques (eg, flame emission)
calibrants are required, because the slope of the will only be close at normal sodium concentra-
electrode must be established before an al- tions. Results from flame emission and ion
gorithm derived from the Nernst equation can selective electrodes have been shown to cor-
be used (Appendix 2). The slope is derived relate well over a wide range of sodium
from the EMF difference of two calibrants. concentration.V In this system the ionic
Many manufacturers maintain a constant total strength of all the calibrants was not adjusted to
ionic strength for all calibrants (eg, 160 mmol/L) 160 mmol/L,
by addition of a non-interfering salt. The effect of
this adjustment is that the activity coefficients and SUMMARY
junction potentials for a 'low sodium' calibrant On theoretical grounds it would appear prefer-
will differ from those found for a serum sample able to use calibrants which are not adjusted to
with a similar molar concentration of sodium." a constant ionic strength, whether concentra-
(Figure 34). tion or activity is being estimated directly. For
Since calibrants with an adjusted ionic indirect methods, when sample and calibrant
strength do not resemble sera with a similar are diluted in an ionic strength adjusting buffer,
molar concentration of sodium (ie, they do not there is probably little difference between
have a physiological ionic strength), the slope adjusted and unadjusted calibrants. In this
(S2) derived from these calibrants will be case a simple calculation of concentration is
incorrect. The effect of using an incorrect slope possible.
in the calibration will be a discrepancy between A simple algorithm may be used for activity

Molar concentration response


{unilinear due to change in yl

Activity slope /
= S2 adjusted "
ionic strength "

Calibratian
paints
\
L; "
/
"
"
...
",,:,y
LL Activity
:::!:
"
•.
w slope = SI

""
<l physiolog ical
I ionic strength
• .....J"
."" .: I
" I
/" I
" I
I
I

Sodium concentration (mmol/Ll

FIG. 34. Relationship of EMF to sodium concentrations at physiological ionic strength (••••.) anddl~;~sted
ionic strength (- - - -).
i'
486 Hirst and Stevens

estimation. The algorithm for direct concentra- 6 Covington AK, ed. Ion Selective Electrode Metho-
tion estimation is more complex. If concentra- dology. Florida: CRC Press, 1979.
tion, rather than activity, is required an indirect 7 Freiser H, ed. Ion Selective Electrodes in Analy-
tical Chemistry. New York: Plenum Press, 1980.
method must be used. If activity is required a 8 Fricke GH. Ion selective electrodes. Anal Chem
direct method must be used. 1980; 52: 259--75.
The quantitative effects of variations in 9 Pinkerton TC, Lawson BL. Analytical problems
calibration procedure may be smaller than facing the development of electrochemical trans-
variations in junction potential. ducers for in vivo drug monitoring: an overview.
Clin Chem 1982; 28: 1946-55.
Advantages and disadvantages of electrodes
References
ADVANTAGES
(a) Analysis is rapid. 1 Clark WM. The determination of hydrogen ions.
(b) Instrumentation is often simple, inexpen- London: Bailliere, Tindall and Cox, 1928: 221-9.
2 Covington AK. Reference electrodes. In: Durst
sive and portable. RA, ed. Ion Selective Electrodes, NBS special
(c) Optically clear solutions are not required publication 314. Washington DC: US Govern-
(whole blood may be used). ment Printing Office, 1969: 107-41.
(d) Analysis is simple. 3 Worth HGI. A comparison of the measurement
(e) Analysis is often non-destructive. of sodium and potassium by flame photometry
and ion-selective electrode. Ann Clin Biochem
1985; 22: 343--50.
DISADVANTAGES 4 Light TS. Industrial analysis and control with ion
(a) Primary calibration is not possible. selective electrodes. In: Durst RA, ed. Ion
(b) Junction potentials may vary between Selective Electrodes, NBS special publication 314.
samples in direct systems. Washington DC: US Government Printing Office,
(c) Analysis may be affected by ionic or 1969: 355-6.
5 Brand MID, Rechnitz GA. Differential poten-
chemical interference. tiometry with ion selective electrodes. Anal Chem
(d) Response is usually non-linear (except 1970; 42: 616-22.
for amperornetric methods). 6 Durst RA. Fluoride microanalysis by null point
(e) Response is sometimes slow, particularly potentiometry. Anal Chem 1968; 40: 931-5.
at low concentrations, limiting the use of 7 Simpson RI. Practical techniques for ion-selective
electrodes in automated analysis. electrodes. In: Covington AK, ed. Ion Selective
Electrode Methodology. Florida: CRC Press,
(f) Some electrodes have a limited life, or 1979; 59.
are easily poisoned. 8 Mertens I, van den Winkle P, Massart DL.
Kinetic study of the fluoride electrode in fast flow
Acknowledgements and automatic systems. Anal Chem 1976; 48:
272-7.
We are indebted to all the instrument manufac- 9 Eisenman G. Chapter 1. In: Durst RA, ed. Ion
turers who supplied information on calibrant Selective Electrodes, NBS special publication 314.
composition and instrument design. We also Washington DC: US Government Printing Office,
thank Professor Smith for his help and en- 1969: I-57.
couragement, Mrs Margaret Coles for typing 10 Moody GI, Thomas lDR. Selective Ion Sensitive
the manuscript and the members of the Instru- Electrodes. Watford: Merrow, 1971: 33-42.
11 Covington AK, Ferra MIA. Buffer solutions for
mentation and Data-Processing Sub- testing glass electrode performance in aqueous
Committee, especially Mr R Beetham, for their solutions over the pH range 0-14 at 25"C. Anal
help and advice. Chem !')77; 49: 1363--8.
12 Moody Gl, Thomas lDR. In: Selective Ion Sensi-
Bibliography tive Electrodes. Watford: Merrow, 1971: 57-88.
13 Frant MS, Ross lW. Electrode for sensing
lIves DIG, Janz GI, eds. Reference electrodes. fluoride ion activity in solution. Science 1966; 154:
New York: Academic Press, 1961. 1553--5.
2 Durst RA, ed. Ion Selective electrodes. NBS 14 Cowell DC. Automated fluoride ion determina-
special publication 314, Washington DC: US tion: Determination of urine/serum fluoride ion
Government Printing Office, 1969. levels. Ann Clin Biochem 1977; 14: 269--78.
3 Eisenman G, ed. Glass electrodes for Hydrogen 15 Moody GI, Thomas lOR. In: Selective Ion Sensi-
and other Cations. New York: Dekker, 1967. tive Electrodes. Watford: Merrow, 1971;70.
4 Moody 01, Thomas IDR. Selective ion sensitive 16 Szabo L, Kenny MA, Lee W. Direct measure-
electrodes. Watford: Merrow, 1971. ment of chloride in sweat with an ion-selective
5 Buck RP. Ion selective electrodes. Anal Chem electrode in situ. Clin Chem 1973; 19: 727-30.
19:'6; 50: 17-29. 17 Benson RG, Isherwood OM, McEvoy FA, Wood
Electrodes in clinical chemistry 487

A. Effect of temperature on response of the sweat inhibiting its effects. Clin Chern Acta 1979; 99:
chloride ion selective electrode. Clin Chern 1982; 229-34.
28: 2447-8. 35 Stoll M, Smith HR. HPLC in clinical chemistry.
18 Ross lW. Solid-state and liquid membrane ion- ACB Technical bulletin (in preparation).
selective electrodes. In: Durst RA, ed. Ion 36 Vadgama P. Enzyme electrodes as practical
Selective Electrodes. NBS special publication 314. biosensors. J Med Eng Tech 1981; 5: 293-8.
Washington DC: US Government Printing Office, 37 Montalvo 1, Guilbaut GG. Sensitised cation
1969: 57-88. selective Electrode. Anal Chern 1969; 41: 1897-9.
19 Moody GK, Oke RB, Thomas IDR. A calcium 38 Thevenot DR, Steinberg R, Coulet PR, et al.
selective electrode based on a liquid ion exchan- Enzyme collagen membrane for electrochemical
ger in a poly (vinyl chloride) matrix. Analyst 1970; determination of glucose. Anal Chern 1979; 51:
910-8. 96-100.
20 lain MK, Wagner RC. Introduction to biological 39 YSI glucose analyser handbook. Colorado: Yel-
membranes. Chichester: 1 Wiley, 1980; 192-231. low Springs Instrument Co.
21 Covington AK, Davison P. Liquid ion exchanger 40 Durliat H, Com tat M, Baudras A. Spectropho-
types. In: Covington AK, ed. Ion Selective tometric and electrochemical determinations of
Electrode Methodology. Florida: CRC Press, L( + }-lactate in blood by use of lactate dehy-
1979; 94. drogenase from yeast. Clin Chern 1976; 22: 1802.
22 Ladenson JH. Direct potentiometric measure- 41 Powley CR, Geiger RF, Nieman RA. Bipolar
ment of sodium and potassium in whole blood. pulse conductance measurements with a calcium
Clin Chern 1977; 23: 1912-6. ion-selective electrode. Anal Chern 1980; 52:
23 Moody GK, Thomas IDR. In: Selective Ion 705-9.
Sensitive Electrodes. Watford: Merrow, 1971; 88. 42 Robinson RA, Stokes RH. Electrolyte solutions.
24 Cattrall RW, Freiser H. Coated wire ion selective 2nd edn. London: Butterworth and Co, 1970.
electrodes. Anal Chern 1971; 43: 1905-6. 43 Czaban JD, Cormier AD, Legg KD. Establishing
25 lanata 1, Moss SD. Chemically sensitive field- the direct-potentiometric 'Normal' range for
effect transistors. Biomed Eng 1976; 11: 241-5. Na/K: Residual liquid junction potential and
26 Pinkerton TC, Lawson BL. Analytical problems activity coefficient effects. Clin Chern 1982; 28:
facing the development of electrochemical trans- 1936-45.
ducers for in vivo drug monitoring. Clin Chern 44 Robertson WG. Measurement of ionised calcium
1982; 28: 1946-55. in body fluids. Ann Clin Biochem 1976; 13: 540-8.
27 Sibbald A, Covington AK, Carter RF. Simul- 45 Butler IN, Huston R. Activity coefficients and ion
taneous on-line measurement of blood K + , Ca2+ , pairs in the systems sodium chloride-sodium
Na" and pH with a four function ChemFET bicarbonate-water and sodium chloride-sodium
integrated circuit sensor. Clin Chern 1984; 30: carbonate-water. J Phys Chern 1970; 74: 2976-83.
135-7. 46 Baron DN. Whole plasma or plasma water: The
28 Severinghaus lW, Bradley AF. Electrodes for problem of the reference base. Clin Chern 1981;
blood Po, and Pea; determinations. J Appl 27: 642.
Physiol 1958; 13: 515-20. 47 IFCC News 1982; 2:5.
29 Ross lW, Riseman IH, Kruger lA. Potentio- 48 KNA1 Handbook. Copenhagen: Radiometer Ltd.
metric gas sensing electrodes. Pure Appl Chern 49 Fleet B, Ryan TH, Brand MID. Investigation of
1973; 36: 473-87. . the factors affecting the response time of a
30 Meyerhoff ME, Fraticelli YM, Greenberg lA, calcium selective liquid membrane electrode.
Rosen 1, Parks Sl, Opdyke WN. Polymer mem- Anal Chern 1974; 46: 12-15.
brane electrode-based potentiometric sensing of 50 Covington AK, Robinson RA. Reference stan-
ammonia and carbon dioxide in physiological dards for the electrometric determination with
fluids. cu« Chern 1982; 28: 1973-8. ion-selective electrodes of potassium and calcium
31 Beetham R. A review of blood pH and blood gas in blood. Anal Chim Acta 1975; 78: 219-24.
analysis. Ann cu« Biochem 1982; 19: 198-213. 51 Kaufman RA, Tietz NW. Ion effects in measure-
32 Attili AF, Autizi D, Capocaccia L. Rapid deter- ment of ionised calcium with a calcium-selective
mination of plasma ammonia using an ion specific electrode. Clin Chern 1980; 26: 640-4.
electrode. Biochem Med 1975; 14: 109-16. 52 Durst RA. Automated analyzer for the deter-
33 Ryan TH, Wilson PS. Electrochemical detection mination of potassium and sodium in whole
in high performance liquid chromatography. Lab blood. Clin Chim Acta 1977; 80: 225-34.
Practice 1979; 501-506.
34 Norden AGW, Flynn FV. Halothane interference
with Po, measurements and a method of Accepted for publication 12 December 1984
488 Hirst and Stevens

Appendix 1
The ionic strength of a solution is the sum of where Ci is the molar concentration of Ion i
contributions from all the ionised species in with charge Zi. For a simple solution of com-
solution and is defined as: pletely dissociated salts, the ionic strength is not
I=ljl~Ci'Zi2 difficult to calculate:

Molar Concentration
Salt concentration Ions (mmol/L)

NaHCO] 25 Na+ 110+25 135


NaCI 110
KCI 5 cr 110+5+2 117
CaCI 1
K+ 5 1 5
Ca2+ 1 4 4
HCO; 25 1 25
S=286

Appendix 2 This modification of the measured sodium


The 'KNA1' Sodium-Potassium analyser activity is designed to compensate for the
(Radiometer Ltd) calculates results by software protein-lipid dilution effect on flame photo-
from measured electrode potentials using the meters. The system is designed to produce the
following algorithm for sodium:" same results as flame photometry within the
normal range, as recommended by the IFCC. 47
The 'KNA1', being a direct measurement sys-
Erne••- E I52 ) tem, produces results which are based on
(
CNa=152·1Q S·61·54 activity, rather than concentration, and which
will not agree closely with flame photometry at
high or low sodium activity.
where S The Coming 902 Sodium-Potassium analyser
61·54 (log 152-log 40) (Coming Instruments Ltd) uses the following
algorithm for the calculation of sodium activity/
concentration.
152 rnmol/L and 40 mmoUL calibrants are used;
E l52 and E 40 are the potentials of the two
calibrants; Erne as is the potential of the un- ( Erne••- E c• 1 )
known solution; 61·54 is the theoretical (Ner- CNa=CcaJ.10 Slope
stian) slope in mV at 37°c; and S is the slope
correction factor. This algorithm is in essence identical to the
KNA1 algorithm. Coming also have an addi-
The final result, which is displayed, is calcu- tional (optional) algorithm which is designed to
lated as: adjust the reading to a flame emission equiva-
lent, by subtraction of 2% of the calculated
CNare.dout=0·92981·CNameas+5·385. activity.

You might also like