You are on page 1of 6

Efficiency and Safety of Mechanical Ventilation

with a Heat and Moisture Exchanger Changed


Only Once a Week
JEAN-DAMIEN RICARD, ERIC LE MIÈRE, PHILIPPE MARKOWICZ, SERGE LASRY, GEORGES SAUMON,
KAMEL DJEDAÏNI, FRANÇOIS COSTE, and DIDIER DREYFUSS
Service de Réanimation Médicale et de Bactériologie, Hôpital Louis Mourier (Assistance Publique-Hôpitaux de Paris), Colombes, France;
and Institut National de la Santé et de la Recherche Médicale, Unité 82, Faculté de Médecine Xavier Bichat, Paris, France

The cost of mechanical ventilation (MV) is high. Efforts to reduce this cost, as long as they are not
detrimental for the patients, are needed. MV with heat and moisture exchangers (HME) changed ev-
ery 48 h is safe, efficient, and cost-effective. Preliminary reports suggest that the life span of these fil-
ters may be prolonged. We determined prospectively whether a hygroscopic and hydrophobic HME
(Hygrobac-Dar; Mallinckrodt) provided safe and efficient heating and humidification of the inspired
gases when changed only once a week. Patients who were considered to require mechanical ventila-
tion for more than 48 h were included in the study. HMEs were initially set for 7 d. Efficient airway
heating and humidification were assessed by clinical parameters (number of tracheal suctionings and
instillations required, peak airway pressures) and hygrometric measurements performed by psy-
chrometry. Resistance was measured from Day 0 to Day 7. Bacterial colonization of circuits and HMEs
was studied. A total of 377 days of mechanical ventilation with 60 HMEs was studied. Clinical param-
eters and hygrometric measurements did not change between Day 0 and Day 7. Mean absolute hu-
midity was 30.3 6 1.3 mg H2O/L on Day 0 and 30.8 6 1.5 mg H2O/L on Day 7 (p 5 0.7). Endotracheal
tube occlusion never occurred. Three HMEs were replaced prematurely because of insufficient ab-
solute humidity. This rare event occurred only in patients with COPD and after the third day of
use. In addition, the absolute humidity delivered by the HMEs was significantly lower in patients with
COPD than in the rest of the population. Resistance did not change from Day 0 to Day 7 (2.4 6 0.3
versus 2.7 6 0.3 cm H2O/L/s; p 5 0.4). Bacterial samples of both circuits and ventilator sides of HMEs
were sterile in most cases. We conclude that mechanical ventilation can be safely conducted in non-
COPD patients using an HME changed only once a week, leading to substantial cost savings (about
$110,000 per year if these findings were applied to the university-affiliated hospitals in Paris). Ricard
J-D, Le Mière E, Markowicz P, Lasry S, Saumon G, Djedaïni K, Coste F, Dreyfuss D. Efficiency
and safety of mechanical ventilation with a heat and moisture exchanger changed only once
a week. AM J RESPIR CRIT CARE MED 2000;161:104–109.

Reducing costs is a major issue when treating critically ill pa- the circuit, circuit contamination, high cost, and increased
tients. Mechanical ventilation of these patients requires ade- work load (3–6). Disposable devices called heat and moisture
quate airway heating and humidification to counterbalance exchangers (HMEs) have been developed. Their clinical effi-
the bypassing of the upper respiratory tract. Without such cacy is comparable to that of the heated humidifier (6–8), de-
conditioning, the dry inspired gases may severely damage the spite their slightly lower humidity outputs. The types of HMEs
respiratory epithelium (1, 2). Heated humidifiers have been include (1) purely hydrophobic HMEs, which possess high an-
used for many years. While the hot water system used pro- timicrobial properties, but perform poorly in terms of humid-
vides adequate heat and humidity to the inspired gases, there ity output and have been responsible for endotracheal tube
are undesired side effects, such as excessive condensation in occlusions (9–12), (2) hygroscopic HMEs, which have better
humidifying qualities than the hydrophobic HMEs, but do not
possess antimicrobial filtration properties, and (3) hydropho-
(Received in original form in February 16, 1999 and in revised form June 10, 1999)
bic and hygroscopic HMEs, which have both satisfactory hu-
midity outputs and antimicrobial properties. The manufactur-
Disclosure of potential conflict of interest: We received no financial support from
and do not have any commitment to the brand of the device tested in this study. ers recommend that these disposable HMEs be changed every
Presented in part at the ATS meeting, April 24–29, 1998.
24 h, although this is not supported by objective data. It has
been previously shown that a hydrophobic and hygroscopic
Correspondence and requests for reprints should be addressed to Didier Drey-
fuss, Service de Réanimation Médicale, Hôpital Louis Mourier, 178, rue des Re- HME (Hygrobac-Dar; Mallinckrodt Medical, Mirandola, It-
nouillers, 92700 Colombes, France. E-mail: didier.dreyfuss@lmr.ap-hop-paris.fr aly) could be changed only every 48 h without any adverse ef-
Am J Respir Crit Care Med Vol 161. pp 104–109, 2000 fects for the patients (13). The rates of nosocomial pneumonia
Internet address: www.atsjournals.org were identical whether the HME were replaced every 24 h or
Ricard, Le Mière, Markowicz, et al.: Heat and Moisture Exchangers Changed Weekly 105

every 48 h. Finally, extending the use period to 48 h provided sodes of HME obstruction were identified by unexplained rises in air-
measurable savings per year (13). But clinical evaluation of way pressure and confirmed by visual inspection of the removed filter
the humidifying and heating performances of HMEs with pa- and the immediate normalization of airway pressures after replace-
rameters such as the number of tracheal suctionings and instil- ment of the HME.
lations required daily, and the peak airway pressure, is insuffi- Hygrometric Measurements
cient because it does not always detect less effective HMEs.
Absolute humidity, relative humidity, and tracheal temperature were
A preliminary study showed that the clinical parameter of
measured within the first 48 h and then daily from Day 3 until Day 7.
adequate airway humidification of a moderately performing Absolute humidity is the amount of water vapor contained in air (mg
HME was not different from that of efficient HMEs, whereas H2O/L). Absolute humidity at saturation (AHs) is the maximum
hygrometric measurements revealed much lower values for amount of water vapor that air can contain at a given temperature.
both absolute humidity and relative humidity than those ex- Relative humidity is the ratio of absolute humidity to absolute humid-
hibited by the efficient HMEs (14). These data clearly indicate ity at saturation, and is expressed as a percentage. These parameters
the need for an objective in vivo evaluation of the heating and may be measured by psychrometry, a technique widely used in clinical
humidifying performances of HMEs before extending the du- studies that evaluate HME performance (14–18). Briefly, inspiratory
ration of their use. The remarkable stability of the values for and expiratory gas flows were separated by a device containing two
one-way valves inserted between the endotracheal tube and the HME.
absolute humidity and relative humidity obtained with the Hy-
Two thermal probes—a dry one and a wet one—were placed in the in-
grobac-Dar after 48 h of use indicates that this hygroscopic spiratory part of the device. The temperatures recorded by the two
and hydrophobic HME may be appropriate for even longer probes were measured and displayed on a chart recorder (Yokogawa,
periods of mechanical ventilation without being changed. Al- Tokyo, Japan). Tracheal temperature was measured with another
though the economic impact of further extending the use of thermal probe inserted in the endotracheal tube and also displayed on
HMEs may be important in an era of diminished resources, the chart recorder. Mean temperatures were recorded from both
studies addressing this issue must certainly not, for obvious probes after a 30-min period to allow for optimal thermal equilibrium.
ethical reasons, place the patients at risk. Extending the use of The psychrometric method compares the temperatures obtained with
HMEs on clinical grounds only is insufficient. We therefore the two probes in the inspiratory part of the separating device. The
assessed the safety and efficacy of mechanical ventilation with dry probe is placed upstream and measures the actual gas tempera-
ture. The downstream probe is coated with sterile cotton wetted with
a hygroscopic and hydrophobic HME changed only once a
sterile water. Evaporation around the wet probe in the inspiratory
week, using both hygrometric measurements and bacterial col- part is proportional to the dryness of the gas. The temperature gradi-
onization surveillance. ent between the two probes increases as the inspired gas humidity de-
creases. There is no thermal gradient when the inspired gas is fully
METHODS saturated with water (100% relative humidity).
Relative humidity (RH) was calculated by reference to a nomo-
Study Design gram, taking into account the difference between temperatures mea-
All of the patients hospitalized over a 9-mo period in the Service de sured by the two probes. Absolute humidity at saturation (100% RH;
Réanimation Médicale of the Louis-Mourier University Hospital (a AHs) was calculated with the following formula: AHs 5 16.451563 2
12-bed intensive care unit) who were considered likely to require con- 0.731T 1 0.03987T2 mg H2O/L, where T (8C) is the dry probe temper-
tinuous mechanical ventilation for more than 48 h were included. Ex- ature. Absolute humidity (AH) was obtained from the formula AH 5
clusion criteria included profound hypothermia (temperature , 338 C), (AHs 3 RH)/100 (in mg H2O/L). Room temperature was constant at
a bronchopleural fistula, or poisoning with breath-eliminated drugs 23.5–258 C.
(hydrocarbons). The HME used was the hygroscopic and hydropho- HMEs were replaced before the seventh day of use when the abso-
bic Hygrobac-Dar (Mallinckrodt Medical). The ventilators used were lute humidity was , 26 mg H2O/L or when there was partial obstruc-
Siemens Servo 900 D (Siemens-Elema, Solna, Sweden), Bird 8400 Sp tion of the HME by tracheal secretion.
(Bird Products, Palm Springs, CA), and César respirators (CFPO,
Paris, France). Each HME was initially installed for a period of 7 d Resistance of the HMEs to Air Flow
after which it was replaced. Premature replacement could occur un- HME resistance to air flow was measured daily in one of every three
der certain circumstances (see below). Patients ventilated for more patients, from Day 0 until Day 7, according to the International Stan-
than 7 d therefore provided several 7-d study periods. dards Organization ISO 9360 (19). Briefly, resistance was calculated
from the pressure drop from either side of the HME when a constant
Positioning of the HMEs 60-L/min flow was applied through the HME (Flowmeter; Fischer &
HMEs were placed between the endotracheal tube and the Y-piece of Porter, Warminster, PA). Pressure was measured with pressure sen-
the circuit. Particular care and attention were given to positioning sors (SCX 0.0045 DV, AST, Vanves, France) and the signal was ana-
each HME in relation to the endotracheal tube. The HME had to be lyzed and displayed on a two-track recorder (Windograf; Gould, Cleve-
placed vertically above the tracheal tube (by means of a flex tube) in land, OH).
order to reduce the risk of partial obstruction of the HME due to re-
fluxed secretions from the tracheal tube. Nurses and doctors repeatedly
checked the position of each HME. TABLE 1
NUMBER OF HMEs STUDIED ACCORDING
Clinical Evaluation of HME Safety and Efficacy TO THE NUMBER OF DAYS ON MECHANICAL
The same simple parameters as those used in our previous studies (7, VENTILATION AND THE REASON FOR VENTILATION
13) were recorded. The number of tracheal suctionings and tracheal Number of Days on
instillations was recorded daily (in our unit, tracheal aspirations are Mechanical Ventilation
performed every 4 h and whenever breathing sounds are heard, and
instillations are performed by the nurse only when the secretions are Reason for Ventilation 3 4 5 6 7
considered thick and difficult to suction). Peak airway pressures were
Chronic airway obstruction (n 5 10)* 2 2 4 3 15
recorded every 6 h and averaged (they are subsequently referred to as Other pulmonary disease (n 5 13) 0 1 3 1 16
mean peak airway pressures). These measurements were done after Other† (n 5 10) 0 1 2 1 9
patient suctioning. Tracheal tube occlusion was prospectively defined
Total HMEs studied 2 4 9 5 40
as an unexplained rise in the peak airway pressure without evidence
of filter obstruction and an inability to insert a suction catheter * Numbers in parentheses represent the number of patients studied.
through the previously patent tube, leading to its replacement. Epi- †
Postoperative mechanical ventilation, neurologic emergency, miscellaneous.
106 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 161 2000

TABLE 2
VENTILATORY PARAMETERS AND PATIENT BODY TEMPERATURE:
COMPARISONS BETWEEN PATIENTS WITH AND WITHOUT COPD

COPD Others
(n 5 10) (n 5 23) p Value

Tidal volume, ml 484 6 73 495 6 95 NS


Respiratory rate, breaths per minute 18.6 6 2.4 20.6 6 2.8 NS
FIO2, % 55 6 10.2 49 6 13.7 NS
Time spent under pressure support ventilation (percentage
of total time under mechanical ventilation) 23.4 6 25.9 33.7 6 34.5 NS
Temperature, °C 37.4 6 0.2 37.3 6 0.6 NS

Colonization of HMEs and Circuits HME replacement (see below). Table 1 shows the number of
The bacteria-filtering properties of the HME were assessed by open- HMEs studied according to the number of days of mechanical
ing them aspetically after their seventh day of use (day of replace- ventilation and the reasons for it.
ment), and placing the ventilator side of the HME on CLED agar. Only two modes of mechanical ventilation were used: as-
Circuit bacterial colonization was assessed by plating a 1-cm2 swab- sist-control ventilation and inspiratory pressure support venti-
bing of the Y-piece on CLED agar and incubating the plates for 48 h. lation. The conditions of mechanical ventilation and body
Colonies were counted and identified. Bacterial counts are expressed temperature were identical in patients with COPD and in the
as colony-forming units (CFU) per HME or per square centimeter of
rest of the patient population (Table 2).
the Y-piece.
Clinical Evaluation of Humidifying Efficacy
Data Collection and Definitions
The parameters recorded for each patient included age, sex, indica-
The parameters used to evaluate the safety and efficacy of the
tion for ventilatory support according to the classification of Zwillich HMEs are shown in Table 3. No endotracheal tube occlusion
and colleagues (20), simplified acute physiology score on admission occurred. There were no differences between Days 0 and 7 in
(21), mean tidal volume, respiratory rate, FIO2 (fraction of inspired ox- terms of the number of tracheal suctionings, tracheal instilla-
ygen), body temperature over time, and the percentage of time spent tions, or the mean peak airway pressures. There were also no
on inspiratory pressure ventilation (defined as the ratio of time spent differences when these comparisons were performed from one
on inspiratory pressure ventilation to total duration of mechanical day to another (data not shown).
ventilation). Only 6 of the 60 HMEs used were changed prematurely (be-
This protocol was approved of by the Institutional Review Board fore Day 7) because of partial obstruction of the HME by tra-
for human studies of the French Intensive Care Society. Informed
cheal secretions, thus amounting to 0.015 HME changed per
consent was not requested since all procedures were considered to be
routine practice, and none was invasive. day of mechanical ventilation. This rare event was not corre-
lated with the duration of HME use (data not shown), was not
Statistical Analysis related to the performance of the HMEs, but was encountered
The parameters for the HMEs are given as means 6 SD, or as the in productive patients and had no adverse effect on the patient.
number of patients affected. Continuous data for the groups were as-
sessed by analysis of variance (ANOVA). When ANOVA indicated Hygrometric Measurements
differences between groups, they were compared using the protected The hygrometric parameters of the Hygrobac-Dar, measured
least significant difference. Hygrometric values over time were com- within the first 48 h of mechanical ventilation (referred to as
pared according to the indication for mechanical ventilation, using a t Day 0–2) and then daily (Day 3 to Day 7), were remarkably
test with Bonferroni’s correction. Categorical data were analyzed by constant throughout the 7-d study period (Table 4). The val-
contingency table analysis. A p value , 0.05 was considered significant.
ues for absolute humidity, relative humidity, and thracheal
temperature measured on Day 7 were identical to those mea-
RESULTS
sured on Day 0–2 (Table 4).
Study Population During the whole study period, only three HMEs were re-
Thirty-three patients were included in the study. The sex ratio placed prematurely because of insufficient absolute humidity.
(M/F) was 18/15. The mean age was 67.3 6 13.1 yr (range, 23– This represented 3 HMEs/377 d, or 0.7 time per 100 d of me-
86 yr). The marker of acute illness used was the Simplified chanical ventilation. All three HMEs replaced concerned pa-
Acute Physiologic Score 2 (21), with a mean of 49.2 6 15.7 tients with chronic obstructive pulmonary disease (COPD).
(range, 21–87). Reasons for mechanical ventilation were the One HME delivered less than 26 mg H2O/L of absolute hu-
following: there were 10 patients with chronic airway obstruc- midity on Day 5, and the other two HMEs were replaced, de-
tion, 13 patients with other pulmonary diseases, and 10 pa-
tients with nonrespiratory diseases (4 postoperative mechani-
cal ventilation, 1 neurologic emergency, and 5 miscellaneous). TABLE 3
Of the 60 HMEs studied, 40 were studied for 7 d. The 33 pa- CLINICAL ASSESSMENT OF HUMIDIFYING EFFICIENCY
tients enrolled provided a total of 377 d of mechanical ventila-
tion. Of these, 12 patients (6 chronic airway obstruction, 4 Day 0 Day 7 p Value
other pulmonary diseases, 1 neurologic emergency, 1 miscella- Tidal volume, ml 482 6 109 488 6 146 0.9
neous) provided several 7-d study periods: 7 periods in 1 pa- Respiratory rate, breaths per minute 19.6 6 3.5 21.1 6 3.9 0.3
tient, 4 periods in 2 patients, 3 periods in 2 patients, and 2 peri- Mean peak airway pressure,* cm H2O 29.3 6 6.7 33.8 6 6.5 0.2
ods in 7 patients; 5 patients provided one 7-d period. The No. of tracheal suctionings, per day 9.3 6 2.4 9.2 6 1.9 0.7
No. of tracheal instillations, per day 0.16 6 0.6 0.25 6 0.8 0.1
remaining patients (16) provided 20 periods of less than 7 d
because of mechanical ventilation cessation or premature * Peak airway pressures were recorded every 6 h and averaged.
Ricard, Le Mière, Markowicz, et al.: Heat and Moisture Exchangers Changed Weekly 107

TABLE 4
HYGROMETRIC PARAMETERS

Day

0–2 3 4 5 6 7 p Value

Absolute humidity,
mg H2O/L 30.3 6 1.3 30.5 6 1.5 30.4 6 1.4 30.3 6 1.6 31.0 6 1.8 30.8 6 1.5 0.7
Relative humidity, % 99.3 6 0.8 99.3 6 0.8 98.3 6 1.1 99.0 6 1.1 99.3 6 0.8 99.3 6 0.8 0.5
Tracheal temperature, 8C 33.1 6 0.8 33.2 6 0.8 33.3 6 0.7 33.0 6 0.9 33.4 6 0.9 33.2 6 0.8 0.9

spite study protocol, on Day 3 and Day 4. Although their ab- HME can be extended (13). These findings may have impor-
solute humidity values were not less than 26 mg H2O/L, they tant medical and economical consequences.
were constantly declining and were close to the set threshold We believe that this study is the first to evaluate an HME for
(Table 5). Importantly, the gradual decrease in absolute hu- such a long period with daily hygrometric measurements. Indeed,
midity before HME replacement observed in these three in- more than 350 d of mechanical ventilation were monitored.
stances led to no adverse effect on the patients. As this rare Clinical parameters, such as the number of tracheal suc-
event occurred solely in patients with COPD, we analyzed the tionings and instillations required per day, and peak airway
results of this category of patients separately from those of the pressures, may help to assess adequate airway humidification,
remaining population. Table 6 shows the values for absolute but daily survey of these parameters is not sensitive enough to
humidity in patients with COPD (including those in whom the detect a gradual fall in the humidity delivered by a humidify-
HME was replaced prematurely) and the values for absolute ing device. We have previously measured wide differences in
humidity measured in the other patients. The HMEs used by the absolute humidity delivered by different HMEs, whereas
the patients with COPD had significantly lower values of ab- the clinical parameters used to assess airway humidification
solute humidity than did those of the other patients after 72 h indicated no such differences (14). We therefore consider it
of use. This difference did not reach significance on Day 6 and important to perform hygrometric measurements when inves-
Day 7 because of the smaller number of HMEs left for analysis. tigating HMEs, and most importantly, when exploring the ex-
tended duration of use of an HME. Hygrometric parameters
Resistance of HMEs to Air Flow were measured once a day after Day 2 in the present study, to
The resistance of the HME to air flow was measured daily in detect over time any fall in humidity performance that could
10 patients; the results of these measurements are shown in have been detrimental to the patient. Absolute humidity re-
Table 7, and indicate that the resistance did not change from mained remarkably stable over time in the non-COPD pa-
Day 0 to Day 7. tients. This constant humidity output reflects the integrity of
the epithelial and mucociliary function. The HMEs act as pas-
Bacteriological Study sive heating and humidifying devices: they store heat and
Thirty-three Y-pieces and ventilator sides of HMEs were sam- moisture from the expired gas and return them to the patient
pled. HMEs (ventilator side) were sterile in 20 cases, whereas via the inspired gas. As the capacity to heat and humidify fresh
13 grew negligible amounts of microorganisms (fewer than 10 gas coming into the respiratory tract depends on the integrity
CFU/side of coagulase-negative staphylococci in 11 cases, 1 of the epithelial and mucociliary function, a constant absolute
CFU/side of Candida albicans or of a Bacillus sp. in one in- humidity output over 7 d of mechanical ventilation indicates
stance each). The results were comparable for the Y-piece: no indirectly that the heating and humidifying functions of the
growth in 30 instances, and fewer than 10 CFU/cm2 of coagu- respiratory tract have been preserved during this time. How-
lase-negative staphylococci in 2 cases and an Enterococcus sp. ever, this was not true in all patients. Three HMEs were re-
in one case. placed prematurely. In one case, the absolute humidity was
below the safety threshold (26 mg H2O/L), and it was con-
DISCUSSION stantly decreasing in the other two cases, being significantly
below the mean absolute humidity in the other patients on the
This prospective clinical study clearly shows that a hygro-
same day. The protocol was violated in these two patients be-
scopic and hydrophobic HME (Hygrobac-Dar) can be used
cause the particularly low absolute humidity could have con-
safely for seven continuous days of mechanical ventilation in
tinued to decrease and place the patient at risk of endotra-
all ICU patients except patients with COPD. These important
findings are based on both clinical and hygrometric assess-
ment, and confirm that the duration of use of this particular
PATIENT 6
ABSOLUTE HUMIDITY VALUES FOR PATIENTS
TABLE 5 WITH COPD AND FOR ALL OTHER PATIENTS
ABSOLUTE HUMIDITY* DELIVERED BY THE THREE REPLACED HMEs Patients with
Day COPD Other Patients p Value
Day
0–2 29.9 6 1 (25)* 30.6 6 1.5 (30) 0.3
Patient 0–2 3 4 5
3 29.7 6 1.3 (26) 31.1 6 1.3 (33) 0.001
1 30.5 28.3 28.3 24.9† 4 29.6 6 1.2 (24) 31.0 6 1.4 (34) 0.002
2 28.5 27.9 26.5† 5 29.6 6 1.6 (20) 30.9 6 1.4 (28) 0.04
3 28.8 26.6† 6 30.0 6 1.1 (13) 31.5 6 1.9 (20) 0.07
7 30.3 6 0.9 (15) 31.0 6 1.7 (25) 0.7
* ln mg H2O/L.

The HME was replaced after this measurement. * Numbers in parentheses indicate the number of HMEs.
108 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 161 2000

TABLE 7
HME RESISTANCE TO AIR FLOW*

Day

0 1 2 3 4 5 6 7

2.4 6 0.3 2.6 6 0.5 2.7 6 0.5 2.7 6 0.5 2.7 6 0.5 2.4 6 0.3 2.5 6 0.3 2.7 6 0.3
* Resistance units, cm H2O/L/s; ANOVA for repeated measurements revealed a p value of 0.4 for the changes in resistance from Day 0
to Day 7.

cheal tube occlusion. This rare decrease in HME objective sistent with these results. The bacterial colonization of the
performance occurred only in patients with COPD. The val- Y-piece and the ventilator side of the HME was low after 7 d
ues for absolute humidity were significantly lower in the of use. In one of these previous studies addressing the issue of
COPD patient group than in the rest of the study population. nosocomial pneumonia (13), we showed that changing the Hy-
This difference remained even when the three HMEs replaced grobac-Dar every 48 h did not affect the rate of nosocomial
prematurely were excluded from the analysis. These surpris- pneumonia.
ingly low values for absolute humidity in patients with COPD There is now considerable evidence that HMEs may be
had not been encountered in our previous study, in which 371 used for longer than the 24 h recommended by the manufac-
measurements were performed, one-third on patients with turers (13, 14, 27, 28). However, and most importantly, this is
COPD (14). The HMEs were changed every 2 d in this previ- not true of all HMEs. Our previous study of three different
ous study. Thus, it is possible that this is the longest period hydrophobic and hygroscopic HMEs found that one of them
that some patients with COPD can be on an HME before its was clearly not suited for use beyond 24 h (14). These results
replacement; however, this remains to be confirmed. The rea- were based on daily measurements of humidity outputs. Such
sons for this decline in absolute humidity are not clear. As measurements are essential since the clinical parameters used
mentioned above, an HME acts passively and cannot restitute to evaluate the humidification performances did not detect
during the inspiratory phase more humidity than it has re- the lower performances of this particular HME, and obviously
ceived during the previous expiratory phase (22). The anatom- not all available HMEs are appropriate for extended use (14).
ical and functional alterations in the bronchial epithelium of We therefore believe that hygrometric performance must be
patients with COPD may have contributed to the gradual de- assessed in any clinical study investigating the extended use of
crease in absolute humidity in these three particular patients. HMEs. Tracheal tube occlusion is the most dangerous risk of
Thus, a decrease in HME performance after 48 h of use may insufficient humidification. Although it appears to be rare, fa-
result in a marked decrease in humidity in those patients with tal cases have been reported (11). Reducing the cost of me-
an altered respiratory epithelium. However, all patients with chanical ventilation is a laudable goal but it must certainly not
COPD did not behave in the same way. Indeed, one patient be done at the expense of a patient’s life.
with COPD was ventilated for 47 d and two for 21 d with their This study demonstrates that a particular HME (Hygrobac-
HME changed every week without any problem. Dar) can be safely used in non-COPD patients for 1 wk. Given
Insufficient humidification may lead to respiratory tract the fact that most patients are ventilated for shorter periods
desiccation and damage, and to life-threatening endotracheal (29), this implies that this HME can be used with total security
tube occlusion (9–12, 23). Tracheal tube occlusion never oc- in patients for more than 2 d, without change. This is the first
curred during the 377 d of mechanical ventilation we have objective demonstration of this possibility. One study (27)
studied. Furthermore, HMEs have been changed every week compared the use of a hygroscopic HME (without bacteria-fil-
in our unit (except for patients with COPD) since this study tering properties) over 7 d with a classic heated humidifier.
was completed (1 yr ago), and no tracheal tube occlusion has The authors concluded that this hygroscopic HME may be used
occurred in 246 ventilated patients, 176 of whom were venti- during the first 7 d of mechanical ventilation, reducing staff la-
lated for more than 3 d outside the research setting. bor and the cost of mechanical ventilation. But the study had
Placing an HME in the breathing circuit generates resis- some drawbacks: first, the HMEs were systematically replaced
tance to gas flow. Thus a fear of a major increase in this resis- by heated humidifiers at the end of the first week of mechani-
tance over time is a legitimate concern. This is why we mea- cal ventilation, which precluded assessment of longer periods
sured the resistance of 10 HMEs every day from Day 0 to Day of use; second, several HMEs were replaced by a heated hu-
7. Resistance did not change significantly over the 7-d use of midifier before the seventh day of use because of abundant se-
the HME. The additional resistance due to an HME is no cretions. Tracheal tube occlusion might have occurred if the
greater than that of a heated humidifier (24). Additional dead HME had been continued in these patients, and hygrometric
space may be a potential drawback to the use of HMEs. The measurements (which were not performed) might have de-
use of HMEs leads to greater PaCO2 and minute ventilation tected insufficient humidification in these patients; third, pre-
than with heated humidifiers, during weaning from mechani- maturely replaced HMEs were excluded from the analysis;
cal ventilation (25). Pelosi and coworkers showed that the use last and most important, the mean duration of mechanical
of HMEs increased the work of breathing, although this in- ventilation was 4.2 6 5.1 d, thus only 21 of 147 HMEs were
crease can be easily overcome by increasing pressure support used for at least 7 d. Our study included 40 HMEs used for
(26). The clinical impact of these observations on the course, more than 7 d.
duration, and outcome of weaning remains unknown. In our It is important to note that the economic savings obtained
daily practice, patients are kept on HMEs during weaning by prolonging the use of HMEs may be considerably reduced
from mechanical ventilation. if they are replaced prematurely owing to partial obstruction
Several studies have shown that the bacteria-filtering prop- by secretions. For instance, in the study by Kollef and co-
erties of the HME used in this study (Hygrobac-Dar) remain workers (27), the number of prematurely replaced HMEs was
stable after 48 h of use (13, 14). Our present findings are con- 0.036 per day of mechanical ventilation (27 HMEs/750 d of
Ricard, Le Mière, Markowicz, et al.: Heat and Moisture Exchangers Changed Weekly 109

mechanical ventilation) whereas this number is 0.015 (6 HMEs/ 9. Roustan, J. P., J. Kielen, P. Aubas, and J. du Cailar. 1992. Comparison of
377 d of mechanical ventilation) in our study. This difference hydrophobic heat and moisture exchangers with heated humidifiers dur-
ing prolonged mechanical ventilation. Intensive Care Med. 18:97–100.
is probably due to the attention paid by the nurses and doctors
10. Villafane, M. C., G. Cinella, F. Lofaso, D. Isabey, A. Harf, F. Lemaire,
in our unit to keeping the HMEs above the endotracheal tube: and L. Brochard. 1996. Gradual reduction of endotracheal tube diam-
they are kept vertically above the endotracheal tube by means eter during mechanical ventilation via different humidification de-
of a flex-tube in order to limit refluxes of secretion. vices. Anesthesiology 85:1341–1349.
The balance between cost savings and quality improvement 11. Martin, C., G. Perrin, M. J. Gevaudan, P. Saux, and F. Gouin. 1990. Heat
in clinical research in the ICU is a major issue. In the present and moisture exchangers and vaporizing humidifiers in the intensive
unit. Chest 97:144–149.
study, most of the patients with COPD were ventilated with-
12. Cohen, I. L., P. F. Weinberg, I. A. Fein, and G. S. Rowinski. 1988. En-
out any problem when their HME was changed only once a dotracheal tube occlusions associated with the use of heat and mois-
week. If hygrometric measurements had not been performed, ture exchangers in the intensive care unit. Crit. Care Med. 16:277–279.
the conclusion of this study (based on the results of the clinical 13. Djedaïni, K., M. Billiard, L. Mier, G. Le Bourdelles, P. Brun, P. Marko-
parameters only) would probably have been that mechanical wicz, P. Estagnasie, F. Coste, Y. Boussougant, and D. Dreyfuss. 1995.
ventilation with an HME changed only once a week was safe Changing heat and moisture exchangers every 48 hours rather 24
hours does not affect their efficacy and the incidence of nosocomial
and efficient in all patients, COPD included. However, because
pneumonia. Am. J. Respir. Crit. Care Med. 152:1562–1569.
measurements were performed, three patients with COPD in 14. Markowicz, P., J.-D. Ricard, D. Dreyfuss, L. Mier, P. Brun, F. Coste, Y.
whom HMEs delivered an absolute humidity significantly Boussougant, and K. Djedaïni. 1999. Safety, efficacy and cost effec-
lower than the values measured in all of the other patients tiveness of mechanical ventilation with humidifying filters changed ev-
were detected. This event occurred only 0.7 time per 100 d of ery 48 hours: a prospective, randomized study. Crit. Care Med. (In
mechanical ventilation. However, we prefer to stay on the safe press)
15. Jackson. C., and A. Webb. 1992. An evaluation of the heat and moisture
side, and recommend a 48-h change of HMEs in patients with
performance of four ventilator circuit filters. Intensive Care Med. 18:264–
COPD. We decided to adopt a conservative approach in order 268.
that patient safety issues not be sacrificed for economic rea- 16. Martin, C., L. Papazian, G. Perrin, P. Bantz, and F. Gouin. 1992. Perfor-
sons. In addition, less frequent HME changes reduce the num- mance evaluation of three vaporizing humidifiers and two heat and
ber of septic procedures, thus improving quality of care. moisture exchangers in patients with minute ventilation . 10 L/min.
In conclusion, our study shows that mechanical ventilation Chest 102:1347–1350.
17. Martin, C., L. Thomachot, B. Quinio, X. Viviand, and J. Albanese. 1995.
with an HME changed every week is safe, efficient, and cost-
Comparing two heat and moisture exchangers with one vaporizing hu-
effective for non-COPD patients. Patients with COPD should midifier in patients with minute ventilation greater than 10 L/min.
have their HMEs changed every 48 h (13). Expanding the life Chest 107:1411–1415.
span of this HME may result in considerable cost savings— 18. Sottiaux, T., G. Mignolet, P. Damas, and M. Lamy. 1993. Comparative
more than $110,000 if these findings are extended to the 800 evaluation of three heat and moisture exchangers during short-term
intensive care beds (medical and surgical ICUs) of the 20 postoperative mechanical ventilation. Chest 104:220–224.
19. International Organization for Standardization. 1988. Humidifiers for
acute-care teaching hospitals of the Administration Générale
medical use: heat and moisture exchangers. Draft International Stan-
de l’Assistance Publique de Paris (to which the Hôpital Louis dards (ISO/DIS 9360, 1988). International Organization for Standard-
Mourier belongs). Nevertheless, focusing only on cost reduc- ization. Technical Committee 121:1–13.
tion, on the assumption that all available HMEs are safe and 20. Zwillich, C., D. Pierson, C. Creagh, F. Sutton, E. Schatz, and T. Petty.
suitable for prolonged use, may lead to unacceptable adverse 1974. Complications of assisted ventilation: a prospective study of 354
effects (30). consecutives episodes. Am. J. Med. 57:161–170.
21. Legall, J., P. Loirat, A. Alperovitch, P. Glaser, C. Granthil, D. Mathieu,
P. Mercier, R. Thomas, and D. Villers. 1984. A simplified acute physi-
References ologic score for ICU patients. Crit. Care Med. 12:975–977.
1. Chalon, J., C. Patel, M. Ali, S. Ramanathan, L. Capan, C. Tang, and H. 22. Wilkes, A. R. 1998. Heat and moisture exchangers: structure and func-
Turndorf. 1979. Humidity and the anesthetized patient. Anesthesiol- tion. Respir. Care Clin. N. Am. 4:261–279.
ogy 50:195–198. 23. Misset, B., B. Escudier, D. Rivara, B. Leclerq, and G. Nitenberg. 1991.
2. Forbes, A. R. 1974. Temperature, humidity and mucus flow in the intu- Heat and moisture exchangers vs heated humidifier during mechani-
bated trachea. Br. J. Anaesth. 46:29–34. cal ventilation. Chest 100:160–163.
3. Craven, D., M. Connolly, D. Lichtenberg, P. Primeau, and W. McCabe. 24. Oh, T. E., E. S. Lin, and S. Bhatt. 1991. Resistance of humidifiers, and
1982. Contamination of mechanical ventilators with tubing changes inspiratory work imposed by a ventilator-humidifier circuit. Br. J. An-
every 24 or 48 hours. N. Engl. J. Med. 306:1505–1509. aesth. 66:258–263.
4. Craven, D., T. Goularte, and B. Make. 1984. Contaminated condensate 25. Le Bourdellès, G., L. Mier, B. Fiquet, K. Djedaïni, G. Saumon, F. Coste,
in mechanical ventilator circuits: a risk for nosocomial pneumonia. and D. Dreyfuss. 1996. Comparison of the effects of heat and moisture
Am. Rev. Respir. Dis. 129:625–628. exchangers and heated humidifiers on ventilation and gas exchange dur-
5. Dreyfuss, D., K. Djedaïni, P. Weber, P. Brun, J. Lanore, J. Rahmani, Y. ing weaning trials from mechanical ventilation. Chest 110:1294–1298.
Boussougant, and F. Coste. 1991. Prospective study of nosocomial 26. Pelosi, P., M. Solca, I. Ravagnan, D. Tubiolo, L. Ferrario, and L. Gatti-
pneumonia and of patient and circuit colonization during mechanical noni. 1996. Effects of heat and moisture exchangers on minute ventila-
ventilation with circuit changes every 48 hours versus no change. Am. tion, ventilatory drive, and work of breathing during pressure-support
Rev. Respir. Dis. 143:738–743. ventilation in acute respiratory failure. Crit. Care Med. 24:1184–1188.
6. Tenaillon, A., G. Cholley, R. Boiteau, D. Perrin-Gachadoat, and M. 27. Kollef, M., S. Shapiro, V. Boyd, P. Silver, B. Von Hartz, E. Trovillion,
Burdin. 1989. Filtre échangeurs de chaleur et d’humidité versus hu- and D. Prentice. 1998. A randomized clinical trial comparing an ex-
midificateurs chauffant en ventilation mécanique. Réan Soins Intens. tended-use hygroscopic condenser humidifier with heated-water hu-
Méd. Urg. 5:5–10. midification in mechanically ventilated patients. Chest 113:759–767.
7. Dreyfuss, D., K. Djedaini, I. Gros, L. Mier, G. Le Bourdellés, Y. Cohen, 28. Thomachot, L., R. Vialet, J. M. Viguier, B. Sidier, P. Roulier, and C.
P. Estagnasié, F. Coste, and Y. Boussougant. 1995. Mechanical venti- Martin. 1998. Efficacy of heat and moisture exchangers after changing
lation with heated humidifiers or heat and moisture exchangers: ef- every 48 hours rather than 24 hours. Crit. Care Med. 26:477–481.
fects on patient colonization and incidence of nosocomial pneumonia. 29. Esteban, A., A. Anzueto, I. Alia, M. Tobin, S. Benito, L. Brochard, and
Am. J. Respir. Crit. Care Med. 151:986–992. T. Stewart. 1999. Clinical characteristics of patients receiving mechan-
8. Hurni, J. M., F. Feihl, R. Lazor, P. Leuenberger, and C. Perret. 1997. ical ventilation (abstract). Am. J. Respir. Crit. Care Med. 159:A47.
Safety of combined heat and moisture exchangers in long-term me- 30. Durbin, C. G. J. 1998. Extended use of disposables: economically sound
chanical ventilation. Chest 111:686–691. or asking for trouble? [Editorial]. Crit. Care Med. 26:426–427.

You might also like