You are on page 1of 7

MINERVA ANESTESIOL 1998;64:75-81

Comparison

A
of three different humidification systems

IC
during prolonged mechanical ventilation
M. LUCHETTI, A. STUANI*, G. CASTELLI*, G. MARRARO

T® D
H E
IG M
Background. An efficient humidification
system is expected to maintain fluid and
Azienda Ospedaliera
«Fatebenefratelli e Oftalmico» - Milano
easily drainable airway secretions. This study Servizio di Anestesia e Rianimazione
aims to compare the efficiency and safety of *Azienda Ospedaliera «Carlo Poma» - Mantova
three humidification systems during pro- Servizio di Anestesia e Rianimazione
R A
longed mechanical ventilation.
Design. Two-center, prospective, randomized
study.
Y V

Methods. 45 critically ill patients undergoing Conclusions. Passive humidification systems


mechanical ventilation were included in the provided low degrees of humidity and tem-
study and allocated to receive one of three perature and could not maintain good secre-
humidification techniques: 1) Bennett Cas- tions. Active systems appeared to satisfy the
P R

cade water-bath humidifier (Bennett group); recommended standards and to allow fluid
2) Fisher & Paykel servocontrolled humidi- and easily drainable secretions.
fier (F & P group); 3) HME Hygrobac DAR Key words: Respiration artificial, instrumenta-
O E

(HME group). Clinical and experimental tion - Intensive Care - Humidity - Ventilators
observations were conducted for 3 to 7 con- mechanical.
secutive days and included: body T°, room T°,
C IN

inspired gas T°, tracheal T°, relative and abso-


lute humidity, heat and water loss, airway
secretion score, need for endotracheal saline
instillation and incidence of ETT occlusion.
Results. The HME group showed a lower tem-
T he upper respiratory tract functions as a
heat and moisture exchanger. Under
normal physiologic conditions inhaled air
M

perature of inspired gases compared to the F reaches the lungs saturated with water
& P group (p<0.05); it also showed a lower vapor at a temperature corresponding to
absolute humidity compared to both Bennett the body temperature.
and F & P groups (p<0.05). A better airway The absolute humidity of inhaled air in
secretion score was obtained in Bennett and F
& P groups compared to the HME group the lungs, if completely saturated with
(p<0.01). water vapor at 37 °C, is 44 mg/l. In the
spontaneously breathing patient inspired
This work was presented in part as an abstract at the gases are usually warmed in the upper
APICE Congress held in Trieste, Italy, November 13-19,
1995. airway to 34 °C.1
Received September 11, 1997. During nasotracheal intubation and me-
Accepted for publication February 3, 1998. chanical ventilation a substantial part of the
upper respiratory tract is by-passed, elimi-
Address reprint requests to: M. Luchetti - Via Varese, 4 -
nating at least one-fifth of the air condi-
20121 Milan, Italy. tioning surface area. In this case adequate

Vol. 64, N. 3 MINERVA ANESTESIOLOGICA 75


LUCHETTI COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION

humidification and warming should be pro- intubated and artificially ventilated. The
vided by external devices. To such purpose patients were prospectively and randomly
active systems (humidifiers) or passive allocated to one of 3 different humidifica-
systems (heat and moisture exchangers, tion systems.
HMEs) are available. In 15 of them the humidification of venti-

A
The active humidifiers supply the patient lated gases was obtained by means of a
with heat and humidity by warming and Bennett Cascade II water-bath humidifier

IC
saturating the ventilated gas with water and a normal inspiratory line (Bennett
vapor. group). This type of humidifier does not
The HMEs achieve humidification similar allow precise temperature settings but only
to normal upper tract humidification. Du- approximate ones, on a numeric scale

T® D
ring expiration, the warm and wet gases ranging from 0 to 12. In this study it was set
pass through plastic fibers which retain heat at 8, as in our routine practice.
and water. During the following inspiration In 15 patients a Fisher & Paykel MR600

H E
this heat and humidity are released, thus servocontrolled humidifier with heater wire
warming and humidifying the cold and dry and a coated inspiratory line were emplo-
inspired gas.2 3
IG M yed (F & P group). In this case the temper-
Basically, an efficient humidification ature of the humidifier was set at 37°C, in
system is expected to maintain fluid and accordance with our routine practice.
easily drainable airway secretions. Exces- In other 15 patients (HME group) humid-
sive or poor humidification have to be ification was provided by passive systems,
R A
avoided, since both expose the patient to consisting of filters/HMEs (DAR Hygrobac)
related side-effects, such as impairment of and a normal inspiratory line.
the mucociliary function,4 5 damage to the The Hygrobac DAR consists of two water-
Y V

tracheal mucosa, atelectasis, hypothermia, repellent electrostatic pre-filters, a micro-


hyperthermia, water intoxication, bron- porous membrane performing mechanical
chopneumonia. filtration, and a layer of cellulose sponge
P R

There is no agreement as to which is the acting as the heat and moisture exchanging
best system for use during prolonged element, the four layers all housed in a
mechanical ventilation. Both passive and
O E

transparent plastic casing. The HME is


active systems have advantages and disad- claimed to allow an inspired relative
vantages and even among active systems the humidity of 95% at 32 °C and a moisture
C IN

differences in performance are important. output of at least 30 mg/L.6 The Hygrobac


The present study aims to assess which, DAR is employed in the majority of ICUs in
among three different humidification Italy and is reported to be one of the best
systems, two active and one passive, is HMEs available. This was the reason for
M

more efficient and safer during long-term selecting this particular brand.
mechanical ventilation, in terms of the heat The HME was used from the moment of
and humidity they are able to provide and intubation and was changed every 24
of their effect on airway secretions. hours, according to the manufacturer’s
advice.
Fluid balance was maintained according
Materials and methods to the individual needs of each patient. In
general, attention was paid that neither
After local ethical committee approval hyperhydration nor hypohydration oc-
and informed consent, 45 patients, 30 male curred.
and 15 female, aged between 18 and 84 All the measurements and recordings
years, were included in the study. They were performed every day for 7 days or
were affected by three different diseases: until the extubation of the patient if this
post-anoxic coma, complicated COPD and occurred earlier. Patients with fewer than 3
polytrauma and were all nasotracheally recordings were excluded from the study.

76 MINERVA ANESTESIOLOGICA Marzo 1998


COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION LUCHETTI

At each daily observation, ventilatory set- TABLE I.—Airway secretion score (AWSS).
tings, room and body temperatures, and Points 0 1 2
blood gases were also recorded.
Humidification and warming data were Quantity Poor Normal Abundant
obtained by using a chart recorder (Yoko- Quality Purulent Brown White

A
Viscosity Thick Fluid —
gawa MR 100, Japan) and 4 silica-titanium
fast-response thermal probes with 0.002% The final score is given by the sum of the scores of each
parameter and ranges from 0 to 5 (worst - best secretions).

IC
precision.
A rotatory circuit with unidirectional val-
ves, interposed between the Y piece or the
filter/HME and the ETT allowed separation used for the daily evaluation of airway secre-
of the inspired flow from the expired flow. tions in our intensive care unit. Due to the

T® D
The inspired gas temperature was meas- lack in the literature of any similar scale, it
ured by means of two probes inserted on was decided to employ the scale for the pur-
pose of the study. The sum of the 3 ratings

H E
the inspiratory flow, one of these naked
(dry probe, T1) and the other drapped in a gives a score (AWSS - Airway Secretion Score)
thin wet piece of cloth (wet probe, T2). The ranging between 0 (worst secretions) and 5
IG M
third probe (T 3 ), inserted into the ETT
down to its tip, measured the tracheal tem-
(best secretions). The AWSS is calculated
once a day at the same time (8 a.m,). For
each patient the average of all the determina-
perature. When active systems were used, a
fourth probe (T4) was positioned at the exit tions all along the study was obtained. Those
values were then averaged for each group.
R A
of the humidifier in order to measure gas
temperature at that level. The need for ETT instillation with saline
Continuous recordings were carried out was reported as well as the incidence of ETT
occlusion. This latter was defined as the
Y V

for at least 30 minutes, the time required for


the system to “warm up”. The sum of the nurse feeling of obstacle while suctioning
daily measurements in each patient was together with an increase of insufflation pres-
P R

averaged. The values obtained were then sure, and confirmed by observation of the
averaged for each group. inside of the ETT once it had been removed.
Relative humidity was obtained from dry Statistical analysis was performed by one
O E

and wet probe temperatures by means of a way ANOVA for comparison between
psychrometric method. Absolute humidity means, and by χ2 test for the other data. A
p value <0.05 was considered significant.
C IN

at saturation point (AHs) was derived


according to the following formula:
AHs=16.41563-0.731 T+0.03987 T2, Results
where: AHs=absolute humidity at 100% RH
M

(mgH2O/L), and T=dry probe t° (°C). The groups were comparable with regard
Absolute humidity was then calculated to patients’ age, weight and body tempera-
from the following formula: ture, room temperature, ventilatory settings,
humidifier temperature and blood gases.
AH=(AHs×RH)/100 (in mgH2O/L). Different pathologies were homogeneously
The heat loss along the inspiratory line distributed in the groups (Table II).
was also calculated, as the difference bet- The other relevant results are summar-
ween T4 and T1. The difference between ized in Table III.
absolute humidity values derived from T4 The HME group showed a lower temper-
and T1 gave the water loss along the inspi- ature of inspired gases compared to the F &
ratory line. P group (p<0.05); it also showed a lower
The quantity, the quality and the viscosity absolute humidity compared to both Ben-
of airway secretions were evaluated accor- nett and F & P groups (p<0.05).
ding to the rating scale described in Table I. Relative humidity was found to be lower
This is a scale created by us and routinely in the F & P group (p<0.05).

Vol. 64, N. 3 MINERVA ANESTESIOLOGICA 77


LUCHETTI COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION

TABLE II.—Comparison of demographic data, underlying diseases, temperatures, ventilatory settings and blood
gases.
Bennet F&P HME

Patients (n°) 15 15 15

A
Age (years) 056.1±23.7 058.5±20.4 056.3±18.8
Weight (kg) 071.4±16.4 73.1±7.7 072.8±14.2
Underlying diseases

IC
Post-anoxic coma (n° pts) 4 3 5
Complicated COPD (n° pts) 6 7 6
Politrauma (n° pts) 5 5 4
Temparatures
Body (°C) 37.1±0.8 37.0±0.6 37.1±0.6

T® D
Room (°C) 24.8±1.5 23.4±1.7 24.2±1.8
Humidifier (T4) (°C) 37.5±0.4 37.3±0.3 —
Ventilatory settings
RR (breaths/min) 17.8±1.2 18.0±0.8 17.5±1.1

H E
TV (ml) 559.0±44.3 566.6±50.4 594.2±58.6
FiO2 00.42±0.05 00.45±0.05 00.46±0.06
PEEP (cmH2O)
IG M 08.2±0.3 08.8±0.3 08.7±0.4
Blood gases
PaO2 (kPa) 11.4±1.4 11.9±1.6 12.2±1.6
PaCO2 (kPa) 05.6±8.8 05.3±0.7 05.4±0.5
All values are expressed as mean ±SD except for “underlying diseases” (number of patients). No significant difference was
found; RR=respiratory rate; TV=tidal volume; FiO2=fraction of inspired oxygen; PEEP=positive end expiratory pressure; PaO2=arte-
R A
rial tension of oxygen; PaCO2=arterial tension of carbon dioxide.
Y V

TABLE III.—Comparison of temperature, humidity, No statistically significant difference bet-


heat and water loss, secretion, and incidence of ween groups was found with regard to tra-
saline instillation and ETT ostruction.
P R

cheal temperature, although the HME


Bennet F&P HME group showed a lower mean temperature.
In the F & P group lesser heat and water
O E

Patients (n°) 15 15 15
Measurements (n°) 48 50 59 loss was found compared to the Bennett
Insp gas dry T° (T1) group (p<0.05).
C IN

(°C) 32.8±1.4 34.9±1.3 30.5±1.6° A better airway secretions score was ob-
Insp gas wet T° (T2) tained in Bennett and F & P groups com-
(°C) 32.5±1.3 33.0±1.5 30.3±1.20
Tracheal T° (T3)
pared to HME group (p<0.01).
(°C) 35.6±1.4 36.0±1.2 33.4±2.40 Compared to the active humidification
Insp rel. hum groups, the HME group also showed a
M

(%) 98.5±0.5 99.0±0.4 98.9±0.20 greater need for endotracheal tube saline
Insp abs. hum.
(mgH2O/L) 34.9±1.5 35.6±1.6 31.0±2.5* instillation (p<0.05) and a higher incidence
Heat loss of ETT obstruction (p<0.05).
(°C) 04.7±1.7 002.4±1.3* — The three patients of the HME group who
Water loss
(°C) 07.8±1.6 003.6±1.5* — had their ETT occluded were then switched
AWSS over to active humidification.
(0-5) 03.5±0.8 03.8±1.0 01.9±1.3§
Saline instillation
(n° pts) 0 0 3*
ETT ostruction Discussion
(n° pts) 0 0 3*
All values are expressed as mean ±SD except for “saline Recently, a few works have addressed
instillation” and “ETT obstruction” (number of patients); the topic of humidification during long-
°p<0.05 compared to F&P group; *p>0.05 compared to the
other groups; § p<0.01 compared to the other groups; term mechanical ventilation with different
AWSS=Airway Secretions Score; ETT = endotracheal tube.
results.7-10

78 MINERVA ANESTESIOLOGICA Marzo 1998


COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION LUCHETTI

The first thing a humidification system is Compared to active humidifiers, heat and
asked is to maintain the airway secretions moisture exchangers are reported to have
and the mucociliary clearance as similar as also some advantages, such as ease of han-
possible to physiologic. The present study dling, low weight, small dead space, no risk
aimed to compare the efficiency and safety of water intoxication, bacterial filtering

A
of three different humidification systems, capability and reduced nursing require-
by assessing their capability of providing ments.6 HMEs’ performance and safety in

IC
given degrees of temperature and humidity recent years have greatly improved and
and, more important, their capability of their use is now widespread. Given the sim-
maintaining fluid and easily drainable plicity of the devices, many intensivists may
airway secretions, thus avoiding complica- find that the level of humidity they are able

T® D
tions such as obstruction of the ETT. to provide is acceptable.
The optimal humidity degree of the Undoubtedly, active humidifiers require
inspired gases during prolonged mechan- more nursing and do not offer a bacterial

H E
ical ventilation is not well established yet, filtering capability. When not properly
neither is the minimum accepted level. used, they can even increase the risk of
Actually, values published in the litera-
IG M bacterial contamination, and represent a
ture range from 17 to 44 mgH2O/L.11-14 potential for water-drowning. These are
In 1970 the British Standards Institution probably the major reasons for someone
recommended a tracheal temperature of avoiding their use.
35 °C and an absolute humidity of at least In the present study HMEs appeared not
R A
33 mgH2O/L to maintain good secretions to meet the aforementioned requirements,
and avoid complications during mechan- providing lower degrees of humidity and
ical ventilation.15 temperature than recommended.
Y V

In a recent paper Williams et al. 16 analyze The poor humidification appeared to


the evidence from the literature and suggest increase the need for endotracheal instilla-
that warming and humidifying gas to its nat- tion with saline prior to endotracheal suc-
P R

ural target of core temperature (37 °C) and tioning. This is a ritual that persists despite
100% relative humidity (44 mg/L) achieves a lack of benefit and a potential for infec-
O E

full gas conditioning, maintains the rhe- tion. In fact, saline instillation may dislodge
ology and normal volume of airway secre- viable bacteria from a colonized endotra-
tions, maximizes mucociliary clearance, cheal tube into the lower airway, over-
C IN

prevents inflammatory reactions, sustains whelming the defense mechanism of immu-


airway patency and preserves lung compli- nocompromised patients.20
ance. Any deviation of the inspired gas In three patients of the HME group the
from the core temperature and 100% rela- poor humidification led to endotracheal
M

tive humidity condition would result in tube occlusion, requiring the ETT change.
both thermal and water mass challenges to In these patients, the decision was taken to
the mucosa. The damage of the mucosa switch over to active humidification.
would be greater with increasing exposure Improved airway secretions were ob-
time. tained within a few hours of active humidifi-
Some previous works, which considered cation. Other A. have previously docu-
lower values of humidity as sufficient, mented the risk of endotracheal tube occlu-
reported apparently good performance for sion associated with the use of HMEs during
several kinds of HMEs,2 17-19 although with prolonged mechanical ventilation.21-23
many limitations, such as a relevant warm- Active humidification systems, in this
up time (45-60 minutes), a short activity study, behaved much better than passive
time (max 24 hours), a tidal volume-depen- systems. Both types of active humidifiers
dant efficiency (large tidal volumes cause provided mean tracheal temperature and
high water loss) and an inadequacy for use absolute humidity well above the minimum
in children (high dead space). recommended; this resulted in fluid and

Vol. 64, N. 3 MINERVA ANESTESIOLOGICA 79


LUCHETTI COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION

easily drainable secretions with no inci- ricoverati in Terapia Intensiva e sottoposti a venti-
dence of endotracheal tube obstruction. lazione meccanica sono stati inclusi nello studio e
Servocontrolled humidifiers with heater suddivisi in maniera randomizzata in tre gruppi a
seconda della tecnica di umidificazione adottata:
wire inside a coated inspiratory limb ap-
1) umidificatore attivo a bagno d’acqua Bennett
peared to be more efficient and safer. The

A
Cascade (gruppo Bennett);
other active systems, although capable of 2) umidificatore attivo servocontrollato Fisher &
maintaining good secretions, showed a sub- Paykel (gruppo F & P);

IC
stantial heat and water loss along the inspir- 3) scambiatore di calore e umidità HME Hygro-
atory line. This caused abundant condensa- bac DAR (gruppo HME). Per 3-7 giorni consecutivi
tion in the circuit, which could possibly sono state condotte osservazioni cliniche e speri-
increase the risk of infection and water- mentali consistenti in: temperatura (T°) corporea, T°
ambiente, T° dei gas inspirati, T° tracheale, umidità
drowning. However, these latter were not

T® D
relativa e assoluta, perdita di calore e umidità,
demonstrated in this study, as we did not secrezioni delle vie aeree, necessità di lavaggi con
look into the incidence of either infectious soluzione fisiologica del tubo endotracheale e inci-

H E
complications or water-drowning. denza di occlusione del tubo endotracheale.
Risultati. Il gruppo HME ha mostrato una T° dei gas
inspirati più bassa rispetto al gruppo F & P (p<0,05),
IG M Conclusions
nonché un’umidità assoluta inferiore rispetto sia al
gruppo Bennett che F & P (p<0,05). Nei gruppi Ben-
nett e F & P è stato ottenuto un punteggio delle
Surely, more prospective, randomized, secrezioni superiore rispetto al gruppo HME (p<0,01).
controlled studies are needed on the topic Conclusioni. I sistemi di umidificazione passiva
hanno fornito bassi livelli di umidità e temperatura
R A
of humidification during prolonged e non sono stati in grado di mantenere fluide le
mechanical ventilation, but according to the secrezioni. I sistemi attivi sono sembrati soddisfare
data obtained in this study two conclusions gli standard raccomandati e hanno permesso di
Y V

can be drawn: mantenere fluide le secrezioni.


1) both types of active humidification Parole chiave: Respirazione artificiale - Terapia
Intensiva - Ventilatori meccanici - Umidificatori.
P R

systems tested in this study appear to be


able to maintain good airway secretions
even during prolonged mechanical ventila-
O E

tion by providing an adequate humidity References


degree, very close to core temperature; 1. Mc Fadden ER, Pichurko BM, Bowman FH, Ingenito
2) heat and moisture exchangers, at least E, Burns S, Dowling N et al. Thermal mapping of the
C IN

those which have been employed in this airways in humans. J Appl Physiol 1985;58:564.
2. Ogino M, Kopotic R, Mannino FL. Moisture con-
study, seem to be inadequate for humidifi- serving efficiency of condenser humidifier. Anaesthe-
cation during prolonged mechanical venti- sia 1985;40:990-5.
lation and, in our opinion, should be 3. Haslam KR, Nielsen CH. Do passive heat and mois-
ture exchangers keep the patient warm? Anesthesio-
M

employed only for short-time ventilation. logy 1986;64:379-81.


4. Forbes AR. Humidification and mucus flow in the
intubated trachea. Br J Anaesth 1973;45:874-8.
5. Dick W. Aspects of humidification: requirements and
Riassunto techniques. Int Anesthesiol Clin 1974;4:217-39.
6. Department of Health, Medical Devices Directorate,
London, UK. Heat and moisture exchanger (HME):
Tre differenti sistemi di umidificazione a confronto DAR Hygrobac/Hygroboy. Evaluation nº 103, Feb-
durante ventilazione meccanica prolungata. ruary 1992.
7. Martin C, Papazian L, Perrin G, Saux P, Gouin F. Pres-
Obiettivo. Un efficiente sistema di umidificazione ervation of humidity and heat of respiratory gases in
dovrebbe mantenere le secrezioni delle vie aeree patients with a minute ventilation greater than
fluide e facilmente drenabili. Lo scopo di questo 10L/min. Crit Care Med 1994;22:1871-5.
studio è quello di confrontare l’efficacia e la sicu- 8. Misset B, Escudier B, Rivara D, Leclercq B, Nitemberg
rezza di tre differenti sistemi di umidificazione G. Heat and moisture exchangers vs heated humidi-
durante ventilazione meccanica prolungata fier during long-term mechanical ventilation. Chest
1991;100:160-3.
Disegno sperimentale. Studio comparativo, pro- 9. Sottiaux T. Use of a heat and moisture exchanger
spettico, randomizzato. during long-term mechanical ventilation. Chest
Pazienti e metodi. Quarantacinque pazienti 1992;102:979.

80 MINERVA ANESTESIOLOGICA Marzo 1998


COMPARISON OF THREE DIFFERENT HUMIDIFICATION SYSTEMS DURING PROLONGED MECHANICAL VENTILATION LUCHETTI

10. Tenaillon A, Cholley G, Boiteau R, Perrin Gachadoat 18. Eckerbom B, Lindholm CE. Performance evaluation
D, Burdin M. Heat and moisture exchanging bacterial of six heat and moisture exchangers according to the
filter versus heated humidifier in long-term mechani- draft international standard (ISO/DIS 9360). Acta
cal ventilation. Care Crit Ill 1991;7:56-61. Anaesthesiol Scand 1990;31:404-9.
11. Weeks DB. A laboratory evaluation of recently avail- 19. Martin C, Papazian L, Perrin G, Bantz P, Gouin F. Per-
able heat and moisture exchangers. Anesthesiol Rev formance evaluation of three vaporizing humidifiers

A
1986;13:33-6. and two heat and moisture exchangers in patients
12. Lowe GD. Efficiency of heat and moisture exchan- with minute ventilation >10 L/min. Chest 1992;
gers: a reply. Anaesthesia 1987;42:1118. 102:1347-50.
13. Epstein RA. Humidification during positive pressure 20. Hagler DA, Traver GA. Endotracheal saline and suc-

IC
ventilation of infants. Anesthesiology 1971;35:532-6. tion catheters: sources of lower airway contamina-
14. Cigada M, Elena A, Solca M, Damia G. The efficiency tion. Am J Crit Care 1994;3:444-7.
of 12 heat and moisture exchangers: an in vitro eval- 21. Cohen IL, Weinberg PF, Fein IA, Rowinsky GS. Endo-
uation. Intens Care World 1990;7:98-101. tracheal tube occlusion associated with the use of
15. British Standards Institution. Specifications for heat and moisture exchangers in the Intensive Care
humidifiers for use with breathing machines. BS Unit. Crit Care Med 1988;16:277-9.

T® D
4494. London, 1970. 22. Roustan JP, Kienlen J, Aubas S, DuCailar J. Compar-
16. Williams R, Rankin N, Smith T, Galler D, Seakins P. ison of hydrophobic heat and moisture exchangers
Relationship between the humidity and temperature with heated humidifier during prolonged mechanical

H E
of inspired gas and the function of the airway ventilation. Intens Care Med 1992;18:97-100.
mucosa. Crit Care Med 1996;24:1920-9. 23. Martin C, Perrin G, Gevaudan MJ, Saux P, Gouin F.
17. Shelly M, Bethune DW, Latimer RD. A comparison of Heat and moisture exchangers and vaporizing humidi-
five HME’s. Anaesthesia 1986;41:527-32.
IG M fiers in the Intensive Care Unit. Chest 1990;97:144-9.
R A
Y V
P R
O E
C IN
M

Vol. 64, N. 3 MINERVA ANESTESIOLOGICA 81

You might also like