Professional Documents
Culture Documents
The size of a sample drawn from a ‘lot’ of food will almost invariably be considerably
larger than that required for analytical purposes. Processes such as ‘quartering’ that are
used to obtain a representative laboratory sample for chemical analysis (and even for some
mycological analyses; Jarvis, 1978), cannot usually be applied in microbiological analysis
because of contamination risks.
It is essential to ensure that test samples are drawn randomly from the primary samples
provided for examination such that they are truly representative of the material under test.
A detailed discussion of methods for taking and preparing laboratory samples for analysis
is given by ICMSF (1978). In general, solid food materials will be macerated during prepa-
ration of a suspension for subsequent dilution and examination.
The size of sample to be taken will depend upon the amount of primary sample avail-
able and its homogeneity, or otherwise. Ideally, the analytical sample should never be less
than 10 g and it should be weighed to the nearest 0.1 g into a sterile container on a top-pan
or other suitable balance. The precision (as reflected in the coefficient of variation) of the
weight of sample taken increases with increasing sample size (Table 6.1). Provided that the
TABLE 6.1
Effect of Sample Weight on the Coefficient of Variation (CV) of weighed samples
TABLE 6.2
Errors in Dispensed Nominal 9 ml volumes of Diluent
Diluent
*A: before autoclaving; B: after autoclaving; B1, B2: replicates stacked in top or bottom half of
basket, respectively.
It is widely recognized that a major potential source of error in counts is that caused by vari-
ations in the volume of diluent used. When dilution blanks are prepared by dispensing before
autoclaving, the potential errors are generally largest. The data in Table 6.2 illustrate this point.
Diluent volume errors will have a cumulative effect in any serial dilution series and there-
fore can have a major effect on the derived colony counts. Due attention is required not only
to minimize the variation between replicate volumes but also the mean volume dispensed. For
instance, whereas 5 10-fold serial dilutions give a dilution factor of 105 (1 in 100,000),
5 9.8 ml serial dilutions gives a dilution factor of 9.85 (approximately 1 in 90,000). This is
discussed in more detail below. Errors in the volumes of diluent used in the primary homog-
enization of samples will also affect the subsequent dilutions. It is therefore essential to ensure
that the volume of diluent added to a primary sample is measured as accurately as possible.
What is the effect of autoclaving on the actual dispensed volumes of a diluent to be used
in a serial dilution procedure?
A notional 9 ml of distilled water was dispensed into each of a number of tared uni-
versal bottles using an automatic dispensing pump or a graduated pipette. The weight of
water dispensed was determined before and after autoclaving, or after aseptic dispensing
of bulk sterilized water. The results are shown in Table 6.2
The coefficients of variation for hand-dispensed diluent increased from 0.8% to 1.5%
in samples weighed before autoclaving, to 2.2–2.6% in samples weighed after autoclav-
ing. Although calibrated to deliver 9 ml of diluent, the automatic dispensing pump deliv-
ered only 8.516 g 0.997 g/ml ( 8.49 ml); after autoclaving the mean volume was 8.01 ml
(8.037 g 0.997 g/ml).
In practice, some allowance can be made for evaporation of water during autoclaving,
but the volume needs to be checked carefully before use. Aseptic dispensing of diluents
after sterilization is the preferred method of use.
Over the years, many studies have been made of the variation in volumes of liquid dis-
pensed by bacteriological pipettes of various categories. Table 6.3 provides a summary of
some typical data.
The variation in the volume of liquid dispensed is dependent upon technical (e.g. operator)
errors in the use of pipettes, calibration errors and the extent to which organisms adhere to the
glass (or plastic) of the pipette; in addition, the multiple use of a single pipette to prepare a dilu-
tion series introduces covariance errors (Hedges, 1967, 2002, 2003; see below). It is worthy of
note that, manufacturers’ pipette calibration errors are of two kinds: manufacturing tolerance
errors (or inaccuracy) and imprecision (repeatability errors) generally cited as a % coefficient
of variation. Values for inaccuracy errors are not readily available from suppliers and reflect
the extent of the manufacturer’s acceptable production QC tolerance (Hedges, 2002). It is
highly desirable that all pipettes (including semi-automated pipette systems that use replaceable
pipette tips) should be recalibrated within the laboratory before use (Hedges, 2003).
TABLE 6.3
Errors in the Volumes of Diluent Delivered Using Different Pipettes
La Water 10 20.0 0.95 4.75
Ca Water 10 33.8 2.87 8.48 Spencer (1970)
Lb Water 20 19.0 0.63 3.33
Cb Water 20 33.3 1.49 4.47
‘Breed’ capillary C 1b Milk 21 10.1 0.30 2.97
C 2b Milk 12 5.6 0.77 13.75 Brew (1914)
C 3a Milk 5 9.0 0.70 7.78
Ca
Serological Water 50 1046.0 37.0 3.49
(1 ml disposable) Ca Water 50 983.0 28.7 2.92 Jarvis (1989)
Cb Water 25 1009.8 45.7 4.53
L laboratory made and calibrated; C commercially prepared and calibrated; C1, C2, C3 different makes of
pipette; a Variation within pipette; b variation between pipettes; c 0.9 ml dispensed; d 0.1 ml dispensed;
e
100 l disposable tips.
How significant are the errors associated with preparation of a dilution series? Is it better
to make six 10-fold dilutions or a 100-fold dilution followed by four 10-fold dilutions?
Assume firstly that the dilution series is to be prepared from a liquid sample. One
1 ml of the sample is transferred by pipette to a 99 ml volume of diluent and after mixing
a further pipette is used to prepare the four subsequent 10-fold serial dilutions to 106.
This gives us one 100-fold (n 1) and four 10-fold (m 4) sequential dilutions.
The equation for estimation of dilution error (based on Jennison and Wadsworth,
1940) is:
Assume that the SD of the volume of liquid delivered from each 1 ml disposable
pipette: a 0.04 ml (see Table 6.3); that the SD of the 9 ml diluent volumes: b 0.20 ml
(Table 6.2) and the SD of the 99 ml diluent volume is c 2.0 ml.
Then for: a 0.04, b 0.20, c 2.0, m 4, n 1, x 1.0, u 10.0 and v 100.0:
The % error of the 106 dilution is given by:
0 . 04 2 (4 1)2 4 2 (0 . 04 2 0 . 22 ) 12 (0 . 04 2 22 )
100 2
1 102 1002
0 . 0016(25) 16(0 . 0416) 1(4 . 0016)
100
1 100 10, 000
100 0 . 04 0 . 00656 0 . 00040016
100 0 . 04696 100(0 . 2167) 21 . 7 %
Assume now that the dilution series is produced using six 10-fold dilutions. For:
a 0.04, b 0.20, m 6, n 0, x 1.0 and u 10.0. Then the % error of 106 dilution
0 . 04 2 (6)2 62 (0 . 04 2 0 . 22 ) 02 (0 . 04 2 2 . 02 )
100 2
1 102 1002
0 . 0016(36) 36(4 . 0016)
100
1 102
100 0 . 0576 0 . 014976
100 0 . 07257 6 100(0 . 2694) 26 . 9 %
For the range of standard deviations and dilution volumes used in this calculation,
the overall percentage dilution error of the 106 dilution would be higher (26.9%) when
prepared using 6 10-fold dilutions than if 1 100-fold and 4 10-fold sequential dilu-
tion steps were used (21.7%).
These values are derived using high levels of error for the pipette and diluent volumes
in order to illustrate the importance of minimizing such errors.
Would the error of the 10 6 dilution be less if a separate pipette were used for each stage
in the dilution process?
Assume that a dilution series to 106 is prepared using six 10-fold sequential dilu-
tions. As before (Example 6.2) assume SD of the pipettes: a 0.04 ml, SD of diluent vol-
ume: b 0.20 ml, the number of dilution stages: m 6 and the volume of the inoculated
diluent: u 10; then: % error of 106 dilution prepared using a different randomly cho-
sen pipette for each stage is
a2 m m(a2 c2 )
100 2
x u2
2
0 . 04 (6) 6(0 . 04 2 0 . 22 )
100
12 102
100 0 . 0096 0 . 002496
100 0 . 012096 11 . 0 %
Hence the percentage dilution error for the 106 dilution prepared using 6 10 ml
serial dilutions with a different randomly drawn pipette for each dilution step is approxi-
mately half of that which would be obtained for an equivalent dilution series made using
a single pipette for all stages (21.7%; Example 6.2).
Comparative calculated errors for using a single pipette or different pipettes for vari-
ous serial dilution series are shown in Table 6.4. Note that the standard deviations used
for Examples 6.2 and 6.3 are different to those used in the table.
TABLE 6.4
Dilution Errors for Various Levels of Dilution* Prepared Either with
the Same Pipette Throughout or Using a Different, Randomly Selected
Pipette for Each Transfer
Other inter-related major sources of potential error occur in making sample dilutions. Solid
food samples require maceration and homogenization. Traditionally, laboratory macera-
tion was done using top-drive or bottom-drive homogenizers, although this has now been
largely superseded by techniques such as ‘stomaching’ and ‘pulsifying’. A potential source
of error with any type of traditional homogenizer is related to the inadequate homogeniza-
tion of the sample, such that a heterogeneous suspension results. Extensive maceration with
a bottom-drive homogenizer can result in a pronounced temperature rise that may affect
the viability of some organisms through sublethal damage and may adversely affect colony
counts. Studies by Barraud et al. (1967), Kitchell et al. (1973) and others have shown that
results varying by about 7% can be obtained with different makes of homogenizer and by
50% when a pestle, mortar and sand were used to grind meat samples. The second related
source of error is inadequate mixing of the inoculated diluent, such that the organisms are
not thoroughly distributed in the suspension, both at the primary and subsequent stages of
a serial dilution. The extent of maceration and mixing can affect the size and distribution
of cell aggregates within the suspension so that apparent differences are seen in the micro-
bial population in replicate plates both at a single dilution and between different dilution
levels.
The introduction of the ‘Stomacher™’ (Sharpe and Jackson, 1972) provided a totally
new approach to the preparation of primary food suspensions that avoids the need to main-
tain a large number of sterile macerators. In addition, the more gentle ‘massaging effect’ of
the stomacher separates organisms from the food such that there was less physical break-
down of the food into particles that might interfere in later stages of the test. Comparative
studies showed that colony counts on food suspensions prepared by ‘stomaching’ and by
homogenisation were generally comparable although, in some cases, counts after ‘stomach-
ing’ may be higher than counts after maceration (Tuttlebee, 1975; Emswilea et al., 1977). A
variation in the stomaching procedure uses ‘filter bags’ into which the food sample is placed
before immersion in the diluent so that the suspension contains the organisms without any
of the ‘stomached’ food sample.
Another alternative procedure uses the ‘Pulsifier®’, an instrument that combines a high-
speed shearing action with intense shock waves to liberate organisms with minimal dis-
ruption of the food sample matrix. Described originally by Fung et al. (1998), its use has
been described by, among others, Sharpe et al. (2000), Kang and Dougherty (2001) and Wu
et al. (2003). The lower level of both suspended and dissolved solids improves membrane
filtration rates and reduces interference in polymerase chain reaction (PCR) and similar
methods.
As pointed out by Mudge and Lawler (1928), among others, a further source of ‘dilution
error’ relates to variations in time between preparation of dilutions and plating. Major dif-
ferences in colony count can result from ignoring this effect; the time between preparation
of dilutions and plating should be consistent and as short as possible.
It has long been recognized that in seeking to assess the precision of a colony count it is nec-
essary to rely not just on the actual count of colonies but to take account also of the contri-
butions to variance of the preceding steps, especially those of the dilution series. Jennison
and Wadsworth (1940) derived a formula to estimate the magnitude of the dilution error
knowing the standard deviations of the pipette volume(s), diluent volumes and number of
dilutions prepared. For a dilution series consisting of both 1 in 100 and 1 in 10 dilutions,
the dilution error, as the standard deviation of the final dilution, is given by:
m2 (a2 b2 ) n 2 (a2 c 2 )
Percentage dilution error 100 a2 (m n)2
102 1002
This generalized equation of the percentage error, as standard deviation of the volume,
can be used for any series of dilutions with pipette and diluent volumes (a, b and c) for
any combination (m and n) of 10 and 100 ml dilution series. Since the absolute value of a2
should be small compared to the values of b2 and c2, and with careful control of the diluent
volumes, such that b 10a and c 100a, this simplifies to:
Jennison and Wadsworth (1940) provide a Table of dilution errors for various combinations
of pipette and diluent blank variances and various combinations of 1 in 10 and 1 in 100
dilutions to 108. However, in their original calculation, they assumed that only one pipette
would be used in the preparation of a dilution series; that is to say, that the same pipette
would be used throughout. Hedges (1967) showed that the basic equation is incorrect if a
fresh, randomly selected pipette is used for each stage in preparing a dilution series. Since
this latter method is the one adopted traditionally by most microbiologists, it is important
to recognize the difference in magnitude of the error that results from elimination of the
covariance terms the equations derived above. Thus, when a fresh randomly drawn pipette
is used at each stage of preparation of a dilution the equation given above for the percentage
dilution error simplifies to:
Percentage dilution errors for dilutions to 106 are illustrated in Table 6.4 for various com-
binations of 1 in 100 and 1 in 10 dilutions, prepared with a single pipette or with a fresh
pipette for each transfer. The magnitude of the dilution error can be limited to some extent
by reducing the number of dilutions prepared, that is in Table 6.4 the dilution error of the 106
dilution (using a separate pipette at each stage) is reduced from 13.4% to 8.5% by preparing
two 100-fold and two 10-fold dilutions, rather than six 10-fold dilutions. The calculations for
use of different pipettes at each stage of the dilution process show that the errors are lower than
in a series for which a single pipette is used throughout. For instance, six 10-fold dilutions made
with fresh randomly drawn pipettes at each stage have a dilution error of 5.5% (cf. 13.4%
when the same pipette is used throughout).
Hedges (2002) investigated the inherent inaccuracy of pipettes in terms of the pipette man-
ufacturers’ quoted calibration and repeatability errors (the latter being a measure of impre-
cision) and took account also of the impact of the Poisson distribution and the sampling
component of variance a consideration ignored by Jennison & Wadsworth (1940). As noted
previously, he observed that pipette manufacturers’ rarely provide information on manufac-
turing inaccuracy and generally cite only the imprecision value, usually as a coefficient of var-
iation (CV) 100 s/u, where s is the standard deviation and u is the nominal volume of
the pipette. In his treatment, Hedges (2002) assumed that a volume (u) of inoculum is pipet-
ted into a volume (v) of diluent to prepare a dilution u/(uv) and that the process is repeated
n times to give a final dilution, with D [u/(uv)]n as the nth step. He further supposed
two method scenarios: (a) the more traditional approach where separate randomly selected
pipettes are used for each of the u volumes and different randomly drawn pipettes are used
to dispense each of the v volumes of diluent at each stage; and (b) a more modern approach
where the same randomly selected dispenser pipette is used to deliver all u volumes (with
a fresh tip used at each step) and another randomly selected pipette is used to deliver all v
volumes.
He used Taylor’s series to derive equations to determine the dilution components of vari-
ance for each serial dilution stage of both methods. Let D p/q un/(uv)n, where p un
and q (uv)n and w (u v). The volumes u and v are uncorrelated, Var(ui) Var(uj) –
where i dilution i, and j dilution (i 1); similarly for v and w.
For method (b), the cumulative variances associated with the inoculum transfer volumes
(ui) and the dispensed diluent volumes (vi) are determined for each stage in the dilution
process:
TABLE 6.5
Dilution and Sampling Components of Variance for Serial Dilutions Made Using Separate Pipettes
at Each Step (Method (a)) and the Same Pipette at Each Step (Method (b))
Dilution component a
Source: Modified from Hedges (2002) and reproduced by permission of the author and Elsevier.
a
Values to be multiplied by (N*)2, where N* estimate of the colony count.
b
Values to be multiplied by N* and u2, where u volume delivered to the plate
However, for method (a), the variances of p and q are not affected by covariances, so the
equations for Var(p) and Var(q) are simplified by the exclusion of the covariance factors.
He introduced, also, an estimate of the sampling component due to the random distribu-
tion of organisms at the sequential stages of the dilution series. Table 6.5 summarizes the
calculated dilution and sampling components and the correlation coefficients for derivation
of covariances are shown in Table 6.6. The derivations of the %CVs of the colony count
for dilution methods (a) and (b) are shown in Table 6.7. The overall % error of a count
of 223 colonies at the 106 dilution (using the same pipette throughout) estimated by the
method of Hedges (2002) was 7.3%. The procedures to be followed to derive the dilution
components are illustrated in Example 6.4.
In a subsequent paper, Hedges (2003) assessed the impact on the precision of serial dilu-
tions and colony counts following recalibration of pipettes in the laboratory, a requirement
for accredited laboratories (ISO, 1999). He concluded that although recalibration improves
the precision of the methods, owing to the dominant effect of the final sampling variance,
the overall improvement in the precision of colony counts is small.
Augustin and Carlier (2006) applied the methods of Hedges (2002, 2003) in their eval-
uation of repeatability variance for data from laboratory proficiency testing. For ease of
calculation, they converted some of the equations into coefficients of variation (CV). For
instance, the CV for the volume (u) of inoculum transferred between dilutions is given by:
CV(2p) Var(p) / p2 Var(un )/ u2n, where Var variance, p un volume of inoculum
transferred at dilution n. They then applied the various equations to determine the overall CV
of the derived colony count, which they showed to be related more closely to the number of
colonies counted than to any other parameter examined. This again indicates the overwhelm-
ing importance of the statistical distribution of organisms both in the original foodstuff and in
the final countable test plates.
Suppose that you need to compare the overall precision of a 10-fold serial dilution series to
106, assuming (a) use of different randomly selected 1 and 10 ml pipettes for each dilution
stage and (b) use of the same 1 and 10 ml pipettes throughout, but with random selection of
pipette tips.
Although this procedure seems complex, it is relatively straightforward providing that
the calculations are done in the following sequence:
Whence, the variance of the dilution (D) the variance of (p/q) Var(D) Var(p/q)
D2{Var(p)/p2 Var(q)/q2 2Covar (p,q)/pq}.
Now Var(p/p2) the square of the coefficient of variation of p (i.e. (CVp)2); similarly
Var(q/q2) (CVq)2.
For both methods of dilution, the volumes u and v are uncorrelated, and the variance
of the volume of inoculum delivered u is given by Var(u) cal(u) rep(u); similarly the
variance of the volume of diluent (v) is Var(v) cal(v) rep(v), so the total pipetting var-
iance Var(w) is given by Var(w) Var(u) Var(v).
For our 10-fold dilution series we use 1 ml ( 1) pipettes that have a calibration inac-
curacy of 0.81% and an imprecision of 0.4% and diluent is dispensed in 9 ml (u 9) vol-
umes with pipettes having a calibration inaccuracy of 0.3% and an imprecision of 0.1%.
Then, using the equations from Hedges (2002), the calibration variance of the 1 ml
pipette cal(u) {(u inaccuracy)/(3 100)}2 {(1 0.81)/300}2 7.29 106; similarly,
the calibration variance of the 9 ml pipette cal(v) {(9 0.1)/100}2 8.1 105.
The repeatability variance of the 1 ml pipette rep(u) {(u imprecision)/100}2
{(1 0.4)/100}2 1.60 105; similarly, the repeatability variance of the 9 ml
pipette rep(v) {(9 0.1)/100}2 8.1 105.
The variance of the volume transferred at each step is given by: Var(u)
{7.29 106 1.60 105} 2.329 105; and the variance of the diluent volumes
is: Var(v) {8.1 105 8.1 105} 1.62 104; so the total pipetting error
Var(w) Var(u)Var(v) {2.329 105 1.62 104} 1.853 104.
However, for any dilution step a covariance factor (Covaru,w) may be necessary to
allow for compound errors in both the inoculum volumes (u) and the diluent volume (v).
To determine covariance, the relation: Covar(s, r) rs, r Var (s) Var (r) is used,
where rs,r the Pearson correlation coefficient, for which r2 represents the proportion of
the total variance shared linearly by the two covariates. The correlation coefficients for
the various possible covariates are given in Table 6.6.
TABLE 6.6
Correlation Coefficients for Covariates in a Serial Dilution Scheme
Source: Modified from Hedges, 2002 and reproduced by permission of the author and Elsevier.
a
Methods (a) and (b) are described in Example 6.4.
Hedges (2002) shows that for both methods the value of the correlation coefficient for
u,w 0.355 (a value supported by simulation studies) so, for the pipette volumes
referred to above, the covariance of u and w for method (a) is
Pipette variances
For method (a), where a different randomly drawn pipette is used at each step, the only
covariate is Covaru,w; then the total variance contribution to the 106 dilution for the
1 ml pipette volume is given by:
Similarly the total variance contribution to the 106 dilution for the 9 ml pipette volume is
The value of the correlation coefficient for u,w 0.355 (Hedges, 2002) so, for the pipette
volumes referred to above, the covariance of u and w for method (a) is
For method (b), covariance factors are required at all stages since single pipettes are
used to deliver each of the u and v volumes. Hence the variance of p is given by:
Next, we need to take account of the Poisson sampling variance (Var(P)) at each step
throughout the series up to and including the volume delivered to the culture plate.
Again, Hedges (2002) provides an approximation for this variance, where u represents
the volume delivered to the plate:
For an observed count of X colonies per volume (u) at step n the error is estimated by
(N*)2 Var(P), where N* is the estimate of the original count (N ): N* X 1/D 1/u.
If X 223, then at step n 6, N* 223 106 and the final pipetting error for:
There now remains the estimate of the total (Poisson) sampling error, which is the same
for both methods. There are two error components: the first is the delivery of the final
diluted inoculum to the plate and the second is the cumulative sampling error during
the preparation of the dilution series. The first is estimated by the number of colonies on
the plate (X 223), the second is derived as:
i2 n
u2 N * ∑ {1/ zi } u2 N * the sampling component (from Table 6.5).
in1
For this example, where n 6, the sampling component is 1.11111 107 and the total
sampling error is
method (a),
the combined variance of (X ) is 7.265 247.778 255.043;
method (b),
the combined variance of (X ) is 16.753 247.778 264.531.
The standard deviation and coefficient of variation of the colony count at the 106
dilution using the two different methods of preparing the dilution series, is:
method (a),
SD Var 255.043 15.97; %CV 100(15.97/223) 7.161%;
method (b),
SD 264.531 16.264; %CV 100(16.26/223) 7.293%.
It is worthy of note that for method (a), the pipetting error is only 2.85% of the combined
error of the colony count whilst for method (b) it is 6.33%.
These data are summarized in Table 6.7
TABLE 6.7
Examples of the Calculated Variance for Colony Counts Done by the Standard Plate Count
Method
Source: Modified from Hedges (2002) and reproduced by permission of the author and Elsevier.
a
from Table 6.5.
b
Final pipetting error (N*)2 Var(P) where Var(P) {u2 Var(D) (D2 Var(u)}2.
c
Total sampling error {u2 N* sampling component value (Table 6.5) X}.
d
CV (%) 100 Var(X)/X (e.g.) 100 255.046/223.
Gross errors in the volumes of inoculum and diluent affect the overall dilution level and
will therefore influence the apparent colony count. For instance, the sequential inoculation
of a 1 ml volume into, for example, 8.5 ml volumes will give a dilution level of 1 in 9.5, not
1 in 10 as intended. Hence the % error associated with the 106 dilution, assuming use of
different pipettes with variance 0.02 ml and diluent blanks with variance 0.1 ml (Table 6.4)
would be 5.5% for six 10-fold dilutions, but would be 5.7% for six 9.5-fold dilutions.
Possibly more important is the impact on the calculated number of organisms.
For instance, an average colony count of 100 colonies from the 106 dilution would be
recorded as 100 106 cfu/unit of sample. But if we take due note of the volume errors, the
derived count should be 100 9.56 (i.e. 100 735100 73.5 106 cfu/unit of sample). The
difference in these counts is highly significant since a ‘true’ count of 74 million would have
been recorded as 100 million. Such differences could affect the likelihood that a colony count
result on a sample would, or would not, comply with a microbiological criterion and justifies
the necessity for calibration of dispensing equipment and checking of dispensed volumes as
part of laboratory quality monitoring Anon (1999).
References
Anon. (1999) General Requirements for the Competence of Testing and Calibration Laboratories, 1st
edition. International Organisation for Standardisation, Geneva. ISO/IEC 17025
Anon. (2006) Official Methods of Analysis, 18th edition. Association of Official Analytical Chemists
Inc, Washington DC.
Augustin, J-C and Carlier, V (2006) Lessons from the organization of a proficiency testing program in
food microbiology by interlaboratory comparison: Analytical methods in use, impact of methods
on bacterial counts and measurement uncertainty of bacterial counts. Food Microbiol., 23, 1–38.
Barraud, C, Kitchell, AG, Labots, H, Reuter, G, and Simonsen, B (1967) Standardisation of the total
aerobic count of bacteria in meat and meat products. Fleischwirtschaft, 12, 1313–1318.
Brew, JD (1914) A comparison of the microscopical method and the plate method of counting bacte-
ria in milk. NY Agric. Expt. Sta., Bull, 373, 1–38.
Emswilea, BS, Pierson, LJ, and Kotula, AW (1977) Stomaching versus blending. Food Technol., pp,
40–42. October 1977
Fung, DYC, Sharpe, AN, Hart, BC, and Liu, Y (1998) The Pulsifier® a new instrument for preparing
food suspensions for microbiological analysis. J. Rapid Meth. Autom. Microbiol., 6, 43–49.
Harrigan, WF (1998) Laboratory Methods in Food Microbiology. Academic Press, London.
Hedges, AJ (1967) On the dilution errors involved in estimating bacterial numbers by the plating
method. Biometrics, 23, 158–159.
Hedges, AJ (2002) Estimating the precision of serial dilutions and viable bacterial counts. Int. J. Food
Microbiol., 76, 207–214.
Hedges, AJ (2003) Estimating the precision of serial dilutions and colony counts: Contribution of
laboratory re-calibration of pipettes. Int. J. Food Microbiol., 87, 181–185.
ICMSF (1978) Microorganisms in foods. 1 Their significance and methods of enumeration, 2nd edi-
tion. University of Toronto Press, Toronto.
Jarvis, B (1978) Methods for detecting fungi in foods and beverages. In Beuchat, L (ed.), Food &
Beverage Mycology. AVI Publ. Co .
Jarvis, B (1989) Statistical aspects of the microbiological analysis of foods. In Progress in Industrial
Microbiology 21. Elsevier, Amsterdam.
Kang, DH and Dougherty, RH (2001) Comparison of Pulsifier® and Stomacher™ to detach microor-
ganisms from lean meat tissues. J. Rapid Meth. Autom. Microbiol., 9, 27–32.
Kitchell, AG, Ingram, GC, and Hudson, WR (1973) Microbiological sampling in abattoirs. In Board,
RG and Lovelock, DW (eds.) Sampling – Microbiological Monitoring of Environments. Academic
Press, London, pp. 43–61. SAB Technical Series No. 7
Jennison, MW and Wadsworth, GP (1940) Evaluation of the errors involved in estimating bacterial
numbers by the plating method. J. Bacteriol., 39, 389–397.
Mudge, CS and Lawler, BM (1928) Is the statistical method applicable to the bacterial plate count?
J. Bacteriol., 15, 207–221.
Sharpe, AN and Jackson, AK (1972) Stomaching: a new concept in bacteriological sample prepara-
tion. Appl. Microbiol., 24, 175–178.
Sharpe, AN, Hearn, EM, and Kovacs-Nolan, J (2000) Comparison of membrane filtration rates and
hydrophobic grid membrane filter coliform and Escherichia coli counts in food suspensions using
paddle – Type and Pulsifier sample preparation procedures. J. Food. Prot., 62, 126–130.
Snyder, TL (1947) The relative errors of bacteriological plate counting methods. J. Bacteriol., 54,
641–654.
Spencer, R (1970) Variability in colony counts of food poisoning clostridia. Research Report No. 151.
Leatherhead Food Research Association.
Tuttlebee, JW (1975) The Stomacher – Its use for homogenisation in food microbiology. J. Food
Technol., 10, 113–123.
Wu, VCH, Jitareerat, P, and Fung, DYC (2003) Comparison of the Pulsifier® and the Stomacher® for
recovering microorganisms in vegetables. J. Rapid Meth. Autom. Microbiol., 11, 145–152.