Professional Documents
Culture Documents
0537
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices.
CAUTION: U.S. Federal law restricts this device to sale by or on the order
of a licensed medical practitioner.
Outside the USA, check local laws for any restriction that may apply.
August, 2000
CE Marking
Related documentation
Copyright
Instrumentarium Corp., 2000. All rights reserved.
The manual is protected by copyright. No part of this document may be photocopied, reproduced,
or translated to another language without the prior written consent of Datex-Ohmeda.
Table of Contents
TABLE OF CONTENTS
1 Introduction 1–1
Symbols and abbreviations ..............................................................................................1–1
Symbols.....................................................................................................................1–1
Abbreviations.............................................................................................................1–1
Safety precautions .....................................................................................................1–2
2 System description 2–1
Arrhythmia Workstation....................................................................................................2–1
Telemetry system .............................................................................................................2–2
Telemetry transmitter..................................................................................................2–4
Antenna plate ............................................................................................................2–5
Antenna system .........................................................................................................2–5
Parts installed in the computer....................................................................................2–7
Data transfer ..............................................................................................................2–8
3 Installation 3–1
Unpacking.......................................................................................................................3–1
Arrhythmia Workstation installation ..................................................................................3–1
Hardware setup..........................................................................................................3–1
Software installation...................................................................................................3–1
Arrhythmia Workstation functional check.....................................................................3–2
Cabling ......................................................................................................................3–2
Telemetry installation.......................................................................................................3–3
Pre-Installed workstation ............................................................................................3–3
Adding a Receiver Board.............................................................................................3–3
Antenna installation ...................................................................................................3–6
Functional check of the telemetry system....................................................................... 3–12
Checking the antenna system .................................................................................. 3–12
Checking the entire telemetry system ....................................................................... 3–12
4 Configuration 4–1
Steps of configuration......................................................................................................4–1
Editing Config.doc............................................................................................................4–1
User configuration ...........................................................................................................4–2
5 Products 5–1
Arrhythmia Workstation Computer, C-ARRWS and C-ARRWSTEL.........................................5–1
Patient telemetry transmitter, T-PE2..................................................................................5–2
Functional description ................................................................................................5–2
Specifications of the telemetry transmitter...................................................................5–4
Service and repair procedures.....................................................................................5–5
i
Datex-Ohmeda S/5 Arrhythmia Workstation
ii
List of Figures
LIST OF FIGURES
Figure 2-1 Arrhythmia monitoring system 2–1
Figure 2-2 Telemetry system 2–3
Figure 2-3 Transmitter 2–4
Figure 2-4 Dual system of the antenna plate 2–5
Figure 2-5 Maximal antenna systems 2–6
Figure 2-6 Converter board 2–7
Figure 2-7 Receiver board 2–7
Figure 3-1 Arrhythmia Workstation setup 3–1
Figure 3-2 Parts of the Pre-Installed Telemetry Workstation 3–3
Figure 3-3 Connecting the Receiver Board to the Converter Board 3–4
Figure 3-4 Receiver board dip switch settings 3–6
Figure 3-5 Antenna coverage 3–7
Figure 3-6 Maximum antenna system 3–8
Figure 3-7 Simple antenna system 3–9
Figure 3-8 Intensified antenna system 3–10
Figure 5-1 Block diagram of the telemetry transmitter 5–3
Figure 5-2 Changing batteries in the telemetry transmitter 5–5
Figure 5-3 Block diagram of the antenna plate 5–7
Figure 5-4 Block diagram of the converter board 5–9
Figure 5-5 Block diagram of the receiver board 5–11
LIST OF TABLES
Table 1 Dip switch settings for Receiver Boards ................................................................. 3–5
Table 2 Some factors affecting antenna coverage.............................................................. 3–7
iii
Introduction
1 INTRODUCTION
The arrhythmia monitoring system consists of the ECG monitors, Arrhythmia
Workstation, and optional telemetry system. This manual consists of following
parts:
- system description
- installation of the Arrhythmia Workstation
- installation of the telemetry system
- detailed information on separate products.
For technical information of the monitors, refer to the manuals of the monitor
used.
Symbols
Attention. Consult accompanying documents. In this device the
symbol means that the device must be supplied from an isolation
(separating) transformer when placed in the patient environment.
This symbol is located on the telemetry transmitters. It symbolizes the
Nurse Call button. When Nurse Call is pressed, respective message is
displayed on the ARRWS screen.
Abbreviations
ARRWS Datex-Ohmeda S/5 Arrhythmia Workstation
BPSK binary phase shift keying
CCM Datex-Ohmeda S/5 Critical Care Monitor
CRC cyclic redundancy check
ID identification
1–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Safety precautions
1–2
System description
2 SYSTEM DESCRIPTION
Arrhythmia Workstation
Antenna
system
HUB
Telemetry
monitoring
2–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Telemetry system
In the telemetry system, the ECG signal is measured with two leads, and sent
from the portable transmitter to antennas, and linked further to the Arrhythmia
Workstation Computer. In the computer, the converter board processes the
signal suitable for the receiver, and then the signal is analyzed for arrhythmias.
See Figure 2-2.
2–2
System description
ECG signal
A/D Transmitter
conversion
transmission
linking, Antenna
dual system System
converting Converter
signal CPU
analysis
arrhythmia analysis
Figure 2-2 Telemetry system
2–3
Datex-Ohmeda S/5 Arrhythmia Workstation
Telemetry transmitter
The transmitter transforms the measured analog signal to digital, and sends it to
antenna system using the black lead wire as antenna.
The carrier wave is either 210-230 MHz (VHF) or 430 -470 MHz (UHF), and it is
modulated, using the binary phase shift keying (BPSK).
Each transmitter’s signal’s bandwidth is 25 kHz. All the transmitters in a
telemetry system must be within 1 MHz range. This is adjusted by the
manufacturer, and cannot be changed in the field.
NOTE: The transmitter cannot be switched on or off. It is always on when
functioning batteries are inserted.
2–4
System description
Antenna plate
The telemetry antennas link the radiowaves from the transmitter to the
Arrhythmia Workstation Computer.
When a patient carrying the transmitter walks around in the ward area, the
signals will reflect from the walls, floors and ceilings causing fading.
On an antenna plate, there are two whip antennas, which are placed with a
quarter of the wavelength from each other, and each with a separate amplifier.
When the signal is weak or put-phased at one antenna, it will be at it’s
maximum at the other antenna. Because of this dual system, two signals are
sent from the antenna to the computer.
1/4
Antenna system
The antenna system consists of maximum of 12 antennas, arranged in 1 to 3
branches with maximum of 5 antennas in each, see Figure 2-5. Between the
branch and the computer an adder/divider is needed for each of the signals.
Each antenna amplifies the signal to the extent of the attenuation that takes
place in the cable. Thus, other cables but the specified ones must not be used.
Accordingly, the specified special terminators must be used in every input
connector not used in the antenna plates or adder/dividers.
2–5
Datex-Ohmeda S/5 Arrhythmia Workstation
2–6
System description
Converter
The Frequency Converter Board, B-RFC, has a common front end diversity UHF
and VHF converter which converts the carrier waves (operating within the
frequency bands officially recognized for telemetry transmission in each
country) to lower frequency for the receiver board.
Receiver
The Receiver Board, B-RC4, B-RC3 or B-RC2, receives signals from the
converter. The number in the board’s name tells for how many patients the
receiver is configured. Up to three receiver boards can be installed into the
computer.
Because of the dual antenna system all the signals are doubled. The receiver
receives the two signals (A and B), and selects the better one by comparing the
noise level. The received data packages are opened and possible transmission
errors are handled. Then the data is passed on to the computer CPU for
arrhythmia analysis.
2–7
Datex-Ohmeda S/5 Arrhythmia Workstation
Data transfer
Various steps have been taken to prevent errors during transmission.
In addition to the digital data, each transmitter sends continuously its own ID-
integrity code. If the ID code is lost, a system alarm will appear.
The information sent is packaged into a matrix:
Data1 ID CRC1
Data2 ID CRC2
Data3 ID CRC3
Data4 ID CRC4
Data = measured data
ID = transmitter ID
CRC = cyclic redundancy check
The signal carries the information: Data1 Data2 Data3 Data4 ID ID ID ID CRC1
CRC2 CRC3 CRC4 ... If any of the data is lost on the way, it can be calculated with
the help of CRC.
Interlacing
Interlacing means that the transmission of one data block is spread over time.
This reduces the probability that all bits are damaged.
A brief disturbance affecting a number of consecutive bits will therefore have no
significant effect on the system’s possibility of reproducing an ECG signal,
because CRC is capable of correcting bit errors.
2–8
Installation
3 INSTALLATION
Unpacking
1. Ensure that the packing box is undamaged. If the box is damaged,
contact the shipper.
2. Open the top of the box and carefully unpack all components.
3. Ensure that all the components are undamaged. If any of the
components is damaged, contact the shipper.
4. Ensure that all the components are included. If any of the components is
missing, contact your Datex-Ohmeda distributor.
Hardware setup
Software installation
3–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Cabling
Cabling depends on the system you have, materials used etc. Refer to Datex-
Ohmeda S/5 Central and Network Technical Reference Manual.
3–2
Installation
Telemetry installation
NOTE: Before building a telemetry system you must have a license from local
authorities to use certain channel of VHF or UHF radiowaves. Inform your local
Datex-Ohmeda distributor of this channel, the telemetry transmitters are made
accordingly.
After hardware installation, perform the configuration and functional check.
Pre-Installed workstation
Telemetry Converter
board B-RFC
Smart key
When the telemetry option is purchased, the required boards (converter and
receiver boards) are pre-installed. However, it may be necessary to add more
receiver boards.
3–3
Datex-Ohmeda S/5 Arrhythmia Workstation
4. Ensure that the dip switch is set correctly (see Table 1), and that there is
only one board with each address (1, 2 or 3).
5. Press the board into the appropriate expansion slot.
6. Attach the board with a screw. Make sure that the bracket slot is well
against the screw, otherwise the bracket may interfere with an adjacent
bracket.
7. Connect the two cables from the receiver to the converter board
connector. Make sure they are in adjacent positions.
8. Terminate the empty connector positions with the converter terminators.
3–4
Installation
Dip switches
The dip switches must be set uniquely for each receiver board in the system to
address it right. Switches on the receiver board can be adjusted e.g. with a
small screwdriver through the three holes in the covering plate. The stickers
nearby show the correct switch positions, and a small round sticker shows the
card’s factory settings.
NOTE: Only Receiver boards 1, 2 or 3 can be used.
Table 1 Dip switch settings for Receiver Boards
Receiver 1, 304
Contact no. 1 2 3 4 5 6 7 8
Position Off Off On On On On On Off
Receiver 2, 308
Contact no. 1 2 3 4 5 6 7 8
Position Off On Off On On On On Off
Receiver 3, 30C
Contact no. 1 2 3 4 5 6 7 8
Position Off Off Off On On On On Off
3–5
Datex-Ohmeda S/5 Arrhythmia Workstation
Antenna installation
Antenna coverage
The thumb rule for planning the antenna system is that one antenna covers an
area of 30m x 30m. This can be used for approximating the need of antennas in
a system.
The final antenna density depends on the environment, e.g. factors listed in
Table 2. The simple antenna system, where the antenna functioning radius can
be up to 20 meters, can be used with good radio transmission conditions with
minimum disturbance. In difficult conditions intensified system is required, e.g.
one antenna coverage radius being 15 m.
3–6
Installation
30 m
20
m
30 m
Figure 3-5 Antenna coverage
3–7
Datex-Ohmeda S/5 Arrhythmia Workstation
120 m
90 m
30 m
CAUTION All cables are planned so that attenuation in them equals to the
amplifiers on the antenna plates. Do not use any other cables, or
shorten or extend the cables.
Consider the factors listed above in Table 2 whether simple or intensified
antenna system is needed.
3–8
Installation
Example:
Antenna system should cover an L-shaped department with 6 patient rooms
along the corridor, see Figure 3-7 and Figure 3-8.
Figure 3-7 shows an example of a simple system where radio transmission
conditions are good and disturbances minimal.
12m
ArrWs
72 m
Figure 3-7 Simple antenna system
Figure 3-8 shows a plan for place, where hinders (thick walls etc.) or
disturbances (MRI, electrosurgery etc.) force to use an intensified system.
3–9
Datex-Ohmeda S/5 Arrhythmia Workstation
12m
ArrWs
72 m
Figure 3-8 Intensified antenna system
3–10
Installation
NOTE: Use only specified terminators. Other terminators may destroy the
system.
4. Attach the next antenna less than 30m from the first one towards the
workstation.
7. Connect the two cables from A and B output connectors of the last
antenna plate to an adder/divider.
3–11
Datex-Ohmeda S/5 Arrhythmia Workstation
3–12
Configuration
4 CONFIGURATION
Steps of configuration
1. Start the workstation computer.
2. Go to Normal Screen and make sure all menus are closed.
3. Quit the arrhythmia software by pressing Ctrl-F2 and giving the password
DADE, so that you are in operating system level.
4. Edit the Config.doc file.
5. Save the file, and take a back-up copy of it.
6. Write INIT and press Enter to initialize the system.
NOTE: Initialization clears all the patient data from the workstation’s
memory.
7. Reset the monitor.
Editing Config.doc
NOTE: Make sure you configure each monitor to one Arrhythmia Workstation
only.
Make sure you write the correct number of characters and spaces to the exactly
correct position. The #### string shows the place and maximum length of the
item. See the examples below (written in COURIER typeface).
1. Enter the name of your Server name in S/5 network (max 39 characters)
########################################
network server. Start at NAME Arrws ICU
column 6 and use max.
39 characters. This name
appears on the monitor
end when it
communicates with the
workstation.
2. Enter the printout Header string for printouts. Hospital/department name (max. 50
characters)
header. Start at column ##################################################
6 and use max. 50 HEAD The Royal Hospital of the Republic
characters.
3. Enter the name for each List of telemetry names (max 6 characters)
and transmitter ID codes (max 2 digits)
of your telemetry T##### ## (Telemetry names start with T)
transmitters (e.g. Tel1), CHAN Tel1 09
and the code of the CHAN Tel1 10
...
transmitter’s device CHAN Tel12 21
4–1
Datex-Ohmeda S/5 Arrhythmia Workstation
User configuration
For configuration of displayed parameters, different views of patient screen etc.,
please refer to the User’s Reference Manual.
4–2
Products
5 PRODUCTS
5–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Functional description
The electrode disconnection (‘Leads Off’) is monitored by feeding a bias current
through the electrodes.
The ECG amplifier ensures that the ECG signal is on an appropriate level for
further sampling and transmission. It also includes overload detection and
active recovery.
After amplification there is a pace maker pulse detection circuitry. It detects
pacer spikes by the shape of the signal.
Analog to Digital Converter takes samples 200 times/second and converts
every sample to a number with 10-bit resolution.
The signal is further processed and modulated for safe and fast transfer (see
chapter Data transfer).
High frequency transmitter is designed for low power consumption.
Transmission is modulated by the BPSK signals. The black lead wire works as an
antenna.
The power supply works with two 1.5 V batteries.
See the block diagram in Figure 2-1.
5–2
Products
ANT.
TRIPLER UHF * 2
HF OSC. *3 PA
MOD. AMP.
. AMP.
VHF AMP.
LEADS
400 kHz
PACER
DET.
LEADS OVER
OFF LOAD
NURSE CALL
PATIENT EVENT
ID
CODE
+3 V POWER BATTERY
2 * 1,5 V
-2 V SUPPLY
GND
5–3
Datex-Ohmeda S/5 Arrhythmia Workstation
5–4
Products
Changing batteries
Cleaning
Wipe with a moistened cloth. Do not let any liquid enter the transmitter.
5–5
Datex-Ohmeda S/5 Arrhythmia Workstation
5–6
Products
Functional description
An antenna plate consists of two whip antennas each having its own amplifier
on an aluminum plate.
Antenna whips are placed 1/4 wavelength from each other to enable that when
a wave is weak or put-phased at one whip, it will be at its maximum in another.
This is called true diversity, which also means that two signals (A+B) are passed
to the converter and receiver for each receiver channel.
The power for amplificators is supplied from the converter board in the computer
via the same cables as the signals but to the opposite direction.
The amplifying capacity is equal to attenuation taking place in the cables. Thus,
the cables must not be shortened or extended, and no other cables must be
used.
The connector(s) that are not used for cabling must be terminated with a
specified special terminator. The terminator is not a resistance only but has also
a condensator.
AMP. AMP.
OUT A
EXT A INPUT
AMP. AMP.
OUT B
EXT B INPUT
5–7
Datex-Ohmeda S/5 Arrhythmia Workstation
5–8
Products
Functional description
When a signal is received from the antenna, it will first be amplified in a LNA (low
noise amplifier).
After the amplifier has passed the signal through an image filter, it is mixed
down to 43-44 MHz. After that there is another amplifier with an image filter. In
the last part of the image filter there is a divider used to distribute the signals to
the receiver boards.
The two receivers (A and B) join the same injection signal. The signal is
generated by an oscillator, and a multiplicator multiplies the frequency by 3. A
buffer is placed after the multiplicator. In the VHF version this buffer is a normal
amplifier, while in the UHF version it multiplies the frequency by 2.
A1
A filte r / A2
LN A filte r M ixer filter AMP
BNC divid er A3
A4
B uffer
M CX
U H F x2 conn ector
OSC x3 V H F x1
Buffer
B1
B filte r / B2
LN A filte r M ixer filter AMP
BNC divid er B3
B4
5–9
Datex-Ohmeda S/5 Arrhythmia Workstation
5–10
Products
Functional description
The signal comes from the converter board output and goes into the two splitters
that divide the signal into 2 times 4 identical signals.
From the splitter, the signals go to diversity intermediate frequency receivers
that transmit a total of 4 differential BPSK signals. The BPSK signals are
converted into a digital bit flow in 4 separate demodulator circuits.
Then the signal goes to a microprocessor that unpacks data and corrects errors
that may have occurred during the transmission. After error correcting and
unpacking, the signals are transmitted to the central processor of the PC via the
PC bus.
RECEI- DEMODU-
VER 1 LATOR
RECEI- DEMODU-
VER 4 LATOR
5–11
Datex-Ohmeda S/5 Arrhythmia Workstation
5–12
Cleaning and maintenance
Cleaning
The Arrhythmia Workstation computer and transmitters can be wiped with soft
moisturized cloth.
Preventive maintenance
Perform following maintenance procedures at least once a year.
6–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Telemetry transmitter
1. Check the transmitters visually.
2. Check that all the transmitters in the system function as they should, see
Functional check of the telemetry system.
3. Check and clean the battery cabin: open the battery cabin, remove the
batteries and wipe the patient connectors with soft cloth.
6–2
Troubleshooting
7 TROUBLESHOOTING
Workstations with or without telemetry:
Workstation does not work, or • Check that smart key is properly inserted into the parallel
works unreliably (e.g. signals communication port of the computer.
misplaced). NOTE: The system must be initialized always after the
smart key is removed and inserted again.
• Initialize the system, see Configuration.
NOTE: Initialization clears all the patient data from the
workstation’s memory.
• Ensure that the configuration is done correctly.
• Check cabling connections.
• Check workstation computer; refer to computer’s
manuals.
‘Leads off’ message appears. Electrode impedance > 2.5 MOhm.
‘No ECG’ or ‘ID error’ message Configuration error: Channel and transmitter ID do not match.
appears.
• Check configuration.
7–1
Datex-Ohmeda S/5 Arrhythmia Workstation
7–2
Spare Parts
8 SPARE PARTS
Monitor/IC Net Cable 885 268 (1.5 m/4.9 ft.)
885 269 (3 m/9.8 ft.)
885 271 (6 m/19.6 ft.)
Telemetry Antenna package (incl. connecting cables) N-API
Telemetry Antenna Adder/Divider (incl. connecting cables N-ADD
and antenna branch terminators)
Datex-Ohmeda S/5 Frequency Converter Board B-RFC
Datex-Ohmeda S/5 Receiver Boards B-RC4, B-RC3, B-RC2
5-lead wire set 713702
Special terminator for antenna branches and 909951
adder/dividers
Converter board terminator 909950
Telemetry transmitter* (5-lead ECG cable included) T-PE2
Telemetry ECG cable 713702
*) Transmitters are prepared specially for a given wavelength, and thus, cannot
be sold directly from the stock. It is recommended to have a spare transmitter
for each telemetry system.
8–1
INDEX
8–1
Datex-Ohmeda S/5 Arrhythmia Workstation
Index-2