You are on page 1of 3

COMMUNICATIONS 257

but triggering on the QRS complex should result in reliable INTRODUCTION


operation. Methods for the evaluation of human gait are still insuffi-
A number of modifications could be made to the unit to ciently objective. Estimating the rehabilitation degree of pa-
enhance its capability and performance. The instrument in its tients using orthotic or prosthetic devices is left to medical
present form makes two rate measurements for every four specialists, and the conclusions drawn are understandably not
QRS complexes. This could be changed so that a measure- absolutely objective. Different authors have designed and de-
ment is made for each QRS complex with a minimal change veloped measuring equipment which would provide the data
in control circuitry. Various pacemaker clinics may de- necessary for the quantitative evaluation of a gait. A variety
modulate the transmitted signal differently, consequently of methods -is known for measuring goniometric functions as
requiring other pacemaker pulse triggering circuitry. Other essential parameters of gait pattern.
methods of triggering, such as a frequency selective circuit or The most simple is the analog measurement of joint angles
slope differentiation, could be employed. Also, an automatic by potentiometers [1]-[41, [7], [9]. Advantages of this
gain control circuit could be incorporated in the analog method are simplicity, low cost and suitability for on-line
processor to eliminate the manual gain control and increase computer analysis. On the other hand, the equipment be-
its capability to trigger on very low amplitude signals. With comes quite heavy and obstructs walking if it is not carefully
recent advances in state-of-the-art MSI low power digital logic designed. Another method requires filming marked sections of
(CMOS or LPTTL), the circuit could be reconfigured to fit a leg by movie camera or by stroboscopic photography [5],
in a package the size of a pack of cigarettes. This would [8 ]. The method is inadequate for routine application since it
allow portable operation using self-contained rechargeable requires longer times for processing and is absolutely unsuit-
NiCad batteries. able for on-line computer evaluation. Far more attractive is
SUMMARY the application of television cameras, but this method requires
rather expensive instrumentation. At one time only five con-
The digital heart rate meter described is sufficiently accurate secutive steps of a single leg can be evaluated [6]. Therefore
to be used in most clinical situations with the advantages of adequate statistical analysis of a greater number of steps is dif-
low component cost and ease of operation. The device is ficult. It also does not allow measurement of the symmetry of
presently in use in a pacemaker clinic with a history of over left and right leg movement.
75 reliable operations. The instrument has the following The system for gait evaluation described in this paper is
features: (1) instantaneous digital heart rate indication; (2) based on the first-mentioned technique. The disadvantages of
choice of ventricular or pacemaker triggering; (3) system test this method were kept to the minimum.
capability; (4) computer interface capability. All of these
features combine to make an excellent general purpose instru-
ment with a variety of applications. DESCRIPTION OF MECHANICAL PARTS
Complete schematics and parts list are available on request The most crucial feature of the mechanical goniometric sys-
from the authors. tem is its simple, sufficiently firm, repeatable and fast attach-
ment to the experimental subject. Attachments have to be
ACKNOWLEDGMENT made by contacts at bony prominences rather than fleshy
The authors wish to thank Dr. Harold Knight and Mr. areas. The number of attachments must be as low as possible
Walter Wohlfert, Department of Cardiovascular Surgery, Saint since each of them disturbs the subject.
Francis Hospital, Hartford, Connecticut, for their assistance in Three attachments were chosen, namely about the belt, un-
the clinical trials of this instrument. der the knee (just above the fibula) and on the shoe heel
[Fig. 11. The customary separate attachments of goniometric
potentiometer legs to the limb were discarded: instead a spe-
cial external lightweight framework was developed, which
halved the necessary number of attachments. The attachment
to the femur was omitted, since of all major skeleton segments
of the lower limb, the femur is by far the most difficult to
attach a measuring device to [3]. The external framework
A Simple Electrogoniometric System and Its Testing eliminates all errors which would arise from the movements
between two adjoining potentiometer legs (e.g., the distal leg
AMADEJ TRNKOCZY AND TADEJ BAJD of the hip goniometer and the proximal leg of the knee
goniometer).
It was supposed that for gait evaluation it is sufficient to
Abstract-A simple goniometric system intended for quantitative eval- measure joint angles in the sagittal plane. Displacements in
uation of a patients' gait is described. The hip, knee and ankde angles other planes as well as displacements of the pelvis were ne-
of both legs in the sagittal plane and the contacts between ground and glected. Free displacements in the coronal plane (transverse
three parts of the sole were measured. By using a special knob to adjust displacements) of bars interconnecting the potentiometers
and fix the knee potentiometer axis position, a small number of selec- (steel 0 2.5 mm) were facilitated by a ball joint on one end
tively chosen attachments and very light construction, the gait was dis- [Fig. 1(3)], and a special hole in the potentiometer axis at the
turbed to the minimum extent. The system's accuracy was tested by other end of the bar [Fig. 1(2)]. The hole is oblong and runs
comparing the directly measured step length and the calculated one. along the axis. It allows free movement of the bars in the co-
The accuracy achieved was better than 5%. ronal plane and slipping in the distal-proximal direction. This
prevents bending stresses on the bars as well as tensile or com-
pression stresses caused by small changes in distances between
Manuscript received June 4, 1974. This work was supported by the goniometer axes. These changes are caused by the fact that
Slovene Research Community, the Boris Kidric Foundation, and the centers of rotation of joints during the gait are not fixed with
Department of Health, Education, and Welfare, Social and Rehabilita-
tion Service, Washington, D. C., under Grant 19-P-58415-F-O1. regard to segments of the extremities. They follow a certain
The authors are with the Institute "J. Stefan," the University of concave curve [1]. This phenomenon is most obvious at the
Ljubljana, Ljubljana, Yugoslavia. knee joint. The curved line cannot be traced easily with
258 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, MAY 1975

necting bars of different length and by using heel rings of


varying size.
Besides the goniometers tape-switches glued between two
thin leather pieces were used as foot-switches. They were
fastened into the shoe sole by plastic adhesive tape. Switches
on the heels and toes were made from three separate switches
connected in parallel and placed in a semicircular arrangement
while the switch in the middle of the sole was as wide as the
shoe. In this way a satisfactorily reliable switching was en-
sured even with patients (hemiplegics) who usually have in-
verted feet. Smaller switches, fixed centrally were far less
reliable. All of the equipment mounted on the subject
weighed less than 750 g.
Data collected were transmitted to the recording system via
a lightweight, 60 m long flexible 10-lead cable. The long walk-
(a) way permits statistical treatment of gait parameters. The
other possibility for transmission, telemetry, was abandoned
because of high cost, low reliability and the usual presence of
nonlinearity problems. The person conducting the experiment
accompanied the subject along the walkway (his presence was
anyway necessary for safety reasons when patients were in-
volved) carrying the cable and so removed its friction with the
ground.
DESCRIPTION OF THE ELECTRICAL PARTS
The potentiometers used were lightweight (14 g), miniature
(o 18 mm), precise (nonlinearity less than 1% and wire-wound
(1070 w/3500, finite resolution caused by individual wires
< 0,30) with ball bearings and no mechanical stops to prevent
mechanical damage due to inadvertently large motions. The
electronics consists of six resistor bridges. The bridge output
voltages are amplified by integrated circuit amplifiers. The
(b) output voltages are adjustable to all recorders and AID con-
verters (sensitivity from 10 V/1000 to 500 mV/1000). It is
possible to change signal polarity and to generate a calibration
voltage equivalent to 100. The magnitude of zero adjustment
and angle changes can be observed on an instrument. Switches
on the heel are electrically weighted so that each combination
of closed switches generates adequate output voltage. The
swing phase and different sections of stance phase can produce
a maximum of 8 voltage levels, although only 6 are physically
feasible.
ACCURACY TEST OF THE SYSTEM
The extent to which measured angles correspond to actual
joint angles was checked in the following manner. In each
heel-on time the angles were read out for all six joints. From
the measured distances between potentiometer axes, the foot
size and, from read-out angles, the step length were calculated
(c) using the trigonometric formula
Fig. 1. The attachment of (a) ankle, (b) knee and (c) hip goniometer.
(1) The knob used for fixing the potentiometer axis in the approxi- d = 11 sin as + r, sin 31 + 12 sin (a1 +C2) + r2 sin (13k - 132)
mate knee joint axis. (2) Oblong hole in potentiometer axis, allowing + 13 sin (a1 + 2 - 3) + r3 sin (11 -12 + 33)
free movement of the joining bar in the coronal plane and distal-
proximal direction. (3) Ball joint between knee goniometer and the
bar leading to the ankle.
+d[l cos(1
- -
2 +03)]+d2 [I cos(oil +2 -a3)]-
-

The definitions of lengths and angles are given in Fig. 2.


Computed values were then compared to actual ones defined
simple technical approaches. A small error was introduced by the distance between two subsequent ink stamp pad marks
intentionally by placing the potentiometer axis into the ap- on paper bands which covered the walkway where the patient
proximate center of the form encircled by the curve. The walked.
movement of potentiometer axis away from preselected posi- The experiment was conducted for an average distance of 70
tion was restricted by a knob having a conical deepening in steps with five subjects of different age and height. Five mea-
which the goniometer axis was put into [Fig. 1(1)], and which surements on the same subject were performed to find out
had been attached by surgical adhesive strap. The attachments how precisely joint axes can be determined by simply observ-
of the hip and ankle goniometers with the aid of a belt and a ing the leg anatomy. Through step lengths calculation it was
metal heel ring are shown in Fig. 1. The adjustments required intended to reveal how accurately the angles can be measured
for different subjects were attained by exchanging intercon- depending on the use of a more or less subjective method for
COMMUNICATIONS 259

Fig. 2. Position of both legs at the moment when the values of angles,
needed for the calculation of step length, are read out.

determining the joint axes. The whole system was five times Department of Medicine and Surgery Veterans Administration,
mounted and dismounted and at each search for joint axes Washington, D. C., USA.
there was no information available on previous mountings, due [6] D. A. Winter, P. K. Greenlaw, and D. A. Hobson, "Television-
to long time intervals between the experiments which were computer analysis of kinematics of human gait," Computers and
Biomedical Research, 5, 1972, pp. 498 -504, Academic Press.
performed by three different experimenters. Inc., USA.
The mean systematic error and mean absolute error were for [7] L. A. Leavitt, E. N. Zuniga, J. C. Calvert, J. Canzoneri, and C. R.
all five mountings and for all five subjects measured smaller Peterson, "Gait analysis of normal subjects," Southern Medical
than 1.5% and 4.8%, respectively. The conclusion can be Journal, volume 64, September 1971, no. 9, pp. 1131-1138,
drawn that simple visual determination of potentiometer posi- Southern Medical Association, USA.
tion is sufficiently accurate. Statistical errors are most likely [8] B. Bresler and J. R. Frankel, "The forces and moments in the leg
to come from inexact readings of angles from graphical rec- during level walking," Transactions of the Asme, January, 1959,
ords, which were only a few centimeters wide. The resolution pp. 27-36.
was therefore limited. When outputs are connected to a com- [9] E. N. Zuniga, L. A. Leavitt, J. C. Calvert, J. Canzoneri, and C. R.
puter via an AID converter, it is expected that these errors are Peterson, "Gait patterns in above-knee amputees," Archives of
reduced substantially. Physical Medicine and Rehabilitation, volume 53, August 1972,
no. 8, Chicago, USA.
CONCLUSION [10] L. A. Leavitt, E. N. Zuniga, C. Calvert, J. Canzoneri, and C. R.
Peterson, "Gait analysis and tissue-socket interface pressures in
The described goniometric system of simple and expedient above-knee amputees," Southern Medical Journal, volume 65,
design gives sufficiently accurate goniometrical and basograph- October 1972, no. 10, Journal of the Southern Medical Associa-
ical data suitable for on-line computer processing. Because of tion, USA.
its lightnes and adequate attachments it only moderately inter-
feres with the subject's gait.
The accuracy attained for angles (since hip and knee angles
are rather small and sin 4 t is valid) is better than 5%. The
system can be connected to a processor computer for on-line
statistical analysis of stride and step durations, gait speed, step
length, symmetry and correlation study of goniograms, etc. A Simple Computer-Television Interface System
REFERENCES for Gait Analysis
[11 P. V. Karpovich, E. L. Herden, and M. M. Asa, "Electrogonio- IN-SHENG CHENG, S. H. KOOZEKANANI, MEMBER, IEEE,
metric study of joints," U. S. Armed Forces Medical Journal, AND M. T. FATEHI, STUDENT MEMBER, IEEE
volume 11, no. 4, April 1960, pp. 424-450, USA.
[2] C. M. Tipton and P. V. Karpovich, "Electrongoniometric records
of knee and ankle movements in pathologic gaits," Archives of Abstract-A new method for analysis and recording of gait paramn-
Physical Medicine & Rehabilitation, March 1965, pp. 267-272. eters is reported. This method consists of a television camera inter-
[3] L. W. Lamoreux, "Kinematic measurements in the study of Hu-
man walking," Bulletin of Prosthetic Research, 10-15, Spring, faced with a PDP-11/10 minicomputer. The TV camera picks up
1971, pp. 3-84, Department of Medicine and Surgery, Veterans anatomical points of interest such as knee joint, ankle joint, etc., to
Administration, Washington, D.C., USA. which small lights are attached and the computer calculates their co-
[4] R. W. Cunmings and L. W. Lamoreux, "The use of electrical ordinates and joint angles as a function of time.
transducers in the measurement of body motions," Bulletin of
Prosthetics Research, 10-18, Fall, 1972, pp. 46-59, Department
of Medicine and Surgery, Veterans Administration, Washington, Manuscript received June 14, 1974. This work was supported by the
D. C., USA. National Science Foundation under Grant GK-25292.
[51 E. Peizer, D. W. Wright, and C. Mason, "Human locomotion," The authors are with the Department of Electrical Engineering, the
Bulletin of Prosthetics Research, 10-12, Fall, 1969, pp. 48 -105, Ohio State University, Columbus, Ohio 43210.

You might also like