Professional Documents
Culture Documents
Publications
2198945-100
Revision 0
do not duplicate
TABLE OF CONTENTS
REVISION HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
LIST OF EFFECTIVE PAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
5 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
SECTION 1 – CALIBRATION STEERING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
1–1 Vascular Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
1–2 Cardiac Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
1–3 Software Aperture Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
1–4 For Biplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
SECTION 2 – CALIBRATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
2–1 System Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
2–2 Technical Menu Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
Job Card CAL 0 A – SOFTWARE APERTURE CALIBRATION CURVE . . . . . . . . . . . . . . . . . 5–5
Job Card CAL 0 B – DEFAULT VASCULAR APERTURE SETTING . . . . . . . . . . . . . . . . . . . . 5–15
Job Card CAL 0 C – DEFAULT CARDIAC APERTURE SETTING . . . . . . . . . . . . . . . . . . . . . 5–27
Job Card CAL 0 D – DSA STEPPING APERTURE CALIBRATION . . . . . . . . . . . . . . . . . . . . 5–37
Job Card CAL 0 E – APERTURE CALIBRATION FOR MP11 . . . . . . . . . . . . . . . . . . . . . . . . . 5–49
Job Card CAL 0 F – BOLUS/SPIN APERTURE CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . 5–59
Job Card CAL 0 G – DSA STEPPING APERTURE CALIBRATION . . . . . . . . . . . . . . . . . . . . 5–69
Job Card CAL 0 H – DEFAULT PHOTOSPOT APERTURE SETTING . . . . . . . . . . . . . . . . . . 5–81
i
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ii
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
iii
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
iv
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
REVISION HISTORY
Title page 0
i thru vi 0
v
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
vi
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
CHAPTER 5 – CALIBRATION
SECTION 1
CALIBRATION STEERING
ADX–E/ADX
OPTIONS:
ADX–E
FOR THE OPTION DSA STEPPING
DSA MP E2 APERTURE SETTING MP E2 PERFORM JOB CARD CAL 0 D
ADX
FOR THE OPTION DSA STEPPING
DSA STEPPING MP12 PERFORM JOB CARD CAL 0 G
APERTURE SETTING MP 12
ADX
FOR THE OPTION DSA STEPPING
DSA MP 11 PERFORM JOB CARD CAL 0 E
APERTURE SETTING MP 11
ADX–E
FOR THE OPTION BOLUS/SPIN
BOLUS/SPIN APERTURE SETTING PERFORM JOB CARD CAL 0 F
ADX–E
FOR THE OPTION PHOTOSPOT
PHOTOSPOT APERTURE SETTING PERFORM JOB CARD CAL 0 H
ADX–E/ADX
DEFAULT CARDIAC
PERFORM JOB CARD CAL 0 C
APERTURE SETTING
The Software Aperture Calibration Curve procedure, CAL 0 A, only needs to be performed if
criteria are not met in the IMG 0 G procedure.
5–1
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Repeat for each plane as done for monoplane. Job card CAL 0D, CAL 0F and CAL 0G won’t
be performed on the lateral plane since bolus chasing, DSA stepping and SPIN are available
only on the frontal plane.
Check that the SOD of the lateral X ray tube is set at 70 cm (minimum SOD) when
CAUTION performing any JOB CARD on the lateral subsystem.
Note: For lateral calibration, set the x–ray beam in the same direction as the one described in the
JOB CARD for the frontal plane
Note: With the SOD set at 70 cm on the lateral plane, the configuration is similar to the frontal
configuration and JOB CARD are valid as well for calibrating or verifying the lateral
plane.
5–2
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 2
CALIBRATION PROCEDURES
SOD SID
ÉÉÉÉÉÉÉÉÉÉÉÉÉ
OBJECT TABLE PANEL
OR
30 cm (12”)
Cu 30 cm (12”)
ÉÉÉÉÉÉÉÉÉÉÉÉÉ ÉÉÉÉÉÉÉÉÉÉ
Alu
plexiglass
Table top at isocenter and X–Ray tube at minimum SID Table top at isocenter and X–Ray
Tube at minimum SID
Cu
Min SID
Click on Technical Menu in the DLX Others Menu to access the Tech Menu function.
5–3
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–4
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Note: THE DETAILED PROCEDURE BEGINS ON PAGE 3 OF THIS JOB CARD.
PURPOSE To build a look–up table repeating the function dose = f (aperture) for a constant
brightness on the PUT in progressive scan mode (512 and 1024).
PROCEDURE 1– Take repeated exposures, the adjust kVp, mA and exposure time (for final adjustment
should be ≥ 25ms) to obtain a mean value equal to 250 ± 3.
2– Click on ACC Table – Iris Aperture is set to 100.
3– Store the final value in the STORE field: DLX increments the aperture to 110.
4– Take an exposure with the same parameters.
5– Store the value for this aperture position: DLX increments the aperture to the next
position (increment plus 10).
6– Repeat steps 4 and 5 until DLX displays the complete aperture curve table.
7– Click on EXIT to quit.
RESULTS The entire ACC table is displayed with variations of less than 1% after
approximately 330.
ACTIONS This procedure is not mandatory. Refer to IMG 0 G for the ACC test.
5–5
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–6
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration procedure is to build a Progressive Scan Aperture Look–up table that
correctly represents the gamma correction function of the Advantx Pick–up Tube.
SECTION 1
DESCRIPTION
The Software Aperture Calibration Curve is used by the DLX Application software to automatically
choose the correct aperture setting after a trial exposure to ensure that proper Peak Video levels are
maintained during subsequent exposures.
The Peak Video (scaled at 0 to 255) versus delta Aperture Step table is used by the system software to
project a Mask Peak Video from a Trial Peak Video, and to calculate the recommended aperture change.
This table is only used in progressive–scan, low gain operations (DLX 512 or 1024 Vascular acquisitions).
The functions resulting from Peak Video to Aperture are nonlinear and provide correction for the gamma
characteristics of the Advantx VIC TV Pick–up Tube.
SECTION 2
SYSTEM USAGE
The DLX System Application Software database provides the operator with an initial aperture position
based on the operator selected Dose. (Refer to Job Card CALO B for setting this initial Aperture position in
the Vascular mode).
During a DLX 512 or 1024 Vascular acquisition,the Aperture position is recalculated and automatically
reset after a trial exposure. After each subsequent DLX 512 Vascular or 1024 Vascular Trial exposure, the
Aperture position is recalculated and automatically reset based on the Mean Value measured in the DLX
generated ROI (Region of Interest) window.
SECTION 3
WHEN TO PERFORM PROCEDURE
The Peak Video versus delta Aperture step is not a linear function and is unique to each Advantx/DLX
System. Therefore, we recommend that you generate a new Software Aperture Calibration Curve
whenever you replace the TV Camera Pick–up Tube, or when you have recalibrated the Advantx VIC and
haved not passed the IMG 0 G procedure.
SECTION 4
TOOLS REQUIRED
or, 2 mm copper
5–7
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 5
PROCEDURE
5.1 Boot Advantx Application and make these softkey selections on the Advantx Console
– PA = posterior anterior.
– Set collimator blades to fully open.
– Set table top at Isocenter.
– Place a 1/4 inch aluminum plate, and 1/32 inch copper plate, on top of the collimator (Vascular
Systems) or on the table top centered in FOV (R and F Systems). Place copper closest to the Image
Intensifier.
– Adjust the Image Intensifier to 12 inches (30 cm) above table top.
– Set the X–ray tube to the minimum SID (Vascular Systems).
5.3 Start the Application
1. Boot the DLX. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to
the DLX Operator’s Manual for instructions on booting the DLX, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on START EXAM.
5–8
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Access the Technical Menu. See the Chapter 6 Introduction for more information on the Technical
Menu.
2. Edit the Technical Menu as follows:
– 10 Bits,
– AEC = OFF.
100
Note: When you perform a prep, the Iris Ref. automatically goes to the default value.
5–9
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During the following steps you must fully release the PREP switch between exposures to continue with the
next exposure.
Due to the brightness response characteristics of the Advantx VIC TV Pick–up Tube, the Pick–up Tube
tends to “fold–over” the image if the image on the tube is too bright. You may notice this during the
following step when you increase the exposure time and the DLX System indicates a decrease in the mean
value. Normally an increase in exposure time results in an increase in the mean value displayed.
To avoid this misregistration, always approach the desired 98% mean value from a lower value (or dimmer
image). If the mean value is too high, reduce the exposure time and continue the step approaching the
correct mean value using a dimmer image.
5–10
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During the following procedure, start building the Aperture Calibration Curve for the DLX System at 98%
(250 ± 3) Mean Value. If you start building the curve at mean values outside of this range, the curve will not
properly correct Aperture positions during Application software acquisition sequences.
Be sure that the first mean value (Aperture position 100) that you store is 250 ± 3.
Establish an Advantx DLX System set–up that yields a 250 ± 3 mean value on the DLX Console Image
monitor Density Histogram display for the first exposure (Aperture position 100).
1. Disable the AEC mode in Technical Menu by clicking on AEC. The icon will be gray.
– Type: 40 <cr>.
– To increase the mean value during the following exposure, increase the exposure time.
– To decrease the mean value during the following exposure, decrease exposure time.
– Do not select an exposure time that is less than 25msec.
2. Continue taking repeated Mask exposures and adjusting the exposure time until the mean value is as
close as possible to 250.
Note: If required, make one or more of the following adjustments and continue taking Mask exposures
until mean value is 250.
Do not change the system set–up once you have established a mean value of 250 for the first exposure
(Aperture position 100). Since the DLX System interpolates between data points (aperture positions),
changing this reference will corrupt the accuracy for the remaining data points.
5–11
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
The DLX System automatically indicates that it has had enough aperture position mean values and has
completed the curve building. This occurs when the mean values change by less than 1% between aperture
positions.
– You can tell that the system has enough values when you click on STORE, and the new Aperture
Calibration Curve table is displayed on the DLX Console Image monitor.
– Click on END to close the APERTURE CALIBRATION CURVE window.
5.13 Verification
5–12
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Record the Aperture Calibration Curve table displayed on the DLX Console Image monitor in the SCAT
Report (Test Data Results).
5–13
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–14
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED PROCEDURE.
PURPOSE To calibrate the default Iris Aperture/NDF position used in the 512/1024 Vascular modes
for each dose selection.
5–15
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–16
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration is to determine and load initial aperture values for a particular Vascular
acquisition Dose selection (Dose A through D for both 512 and 1024 acquisitions). In addition, this
procedure can be used to determine the proper positioning of the Neutral Density Filter for optimum
aperture performance.
SECTION 1
DESCRIPTION
The default Vascular aperture values are used by DLX Application software to initially set the aperture
position to the value determined during this procedure for a given Dose and 512/1024 display selection.
Determining and loading these initial aperture values helps reduce the number of Trial exposures required
to determine an aperture position that will provide 80 ± 3 LSB as a mean value for a particular study.
SECTION 2
SYSTEM USAGE
The DLX System database provides the operator with an initial recommended aperture position based on
the selected Dose and 512/1024 display selection.
During a Vascular acquisition, the aperture position is recalculated and is automatically reset after a trial
exposure. After each subsequent Vascular Trial exposure, the aperture position is recalculated and reset
automatically based on the mean value measured in the DLX ROI (Region of Interest) window.
SECTION 3
WHEN TO PERFORM PROCEDURE
Peak Video versus delta aperture step is not a linear function and is unique to each Advantx/DLX System.
Therefore, we recommend that you generate a new Software Aperture Calibration Curve whenever you
replace the TV Camera Pick–up Tube or recalibrate the Advantx VIC.
SECTION 4
TOOLS REQUIRED
or, 2 mm copper
5–17
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 5
PROCEDURE
5.1 Boot the Advantx Application and make the following soft key selections on the Advantx Console
D TECHNIQUES
– Start with Dose A
– 70 kVp
– Depends on the adjustment (start with 100 mA)
– Focal spot depends upon the tube (small or medium)
D DLX VASCULAR MODE SELECTIONS
– NO SUB Image Display
– 9 inch Field of View (22 cm)
– PA = posterior anterior.
– Set collimator blades fully open (field size).
– Set table top at Isocenter.
– Place the CV phantom filtration on the table top or, if using the 2 mm of Cu, place the Cu on the
collimator.
– Adjust the Image Intensifier to 12 inches (30 cm) above table top.
– Set X–Ray Tube (Vascular Systems) to the minimum SID.
5–18
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
5–19
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
***
Note: ***Perform a prep; the Iris Ref. automatically goes to the default value. See the Iris Reference
Table in Section 5.7.
5–20
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
IRIS APERTURE and ND FILTER fields displayed on the Technical Menu Screen to determine Advantx
System Aperture position and Neutral Density Filter position.
Therefore, throughout this procedure you must enter values in the IRIS APERTURE field and select the
proper ND FILTER (ON or OFF) to start each vascular dose for 512 and 1024.
Please refer to the following table for the default values to be entered when beginning a calibration.
Note: The defaults values depend on the calibrated systems. The defaults values for a LCA are not exactly the
same as for a LC+ (LC/LP+) or a LCV+ (LCN+). Select in the following tables the one which
corresponds to the system being calibrated.
TABLE 1
Note: It is known that the iris values obtained after calibration can be very different from one system to one
another (because of actual II conversion factor or PUT sensitivity). Thus big differences between
default values and calibrated values can be observed on no broken systems and without miscalibration.
5–21
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter. If ND Filter is selected, the DLX System
adds a value of 450 to the Aperture position value.
Examples:
– If the DLX System requires the Advantx System VIC to set the Aperture at position 23 and bring
in the ND Filter for Vascular Dose C, 512 Display, the Iris Value field parameter field will display
a value of 680 (Aperture position 230 and 450 to reflect that the ND Filter is in the light field). To
properly start this procedure for Vascular Dose C, 512 Display enter the value 230 in the Iris
Aperture field and select ON in the Technical Menu ND Filter field.
– If the DLX System requires the Advantx System VIC to set the Aperture at position 36 and leave
out the ND Filter for Vascular Dose A, 512 Display, the Iris Value field parameter field will display
a value of 360 (Aperture position 360 and 0 to reflect that the ND Filter is not in the light field).
To properly start this procedure for Vascular Dose A, 512 Display enter the value 360 in the Iris
Aperture field and select OFF in the Technical Menu ND Filter field.
5–22
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Due to brightness response characteristics of the Advantx VIC Image TV Pick–up Tube, the Pick–up Tube
tends to “fold–over” the image if the image on the tube is too bright. You may notice this during the
following steps when you decrease the Aperture Value and the DLX System indicates a decrease in the Peak
Video display. Normally a decrease in Aperture value results in an increase in Peak Video display.
To avoid this misregistration, approach the desired mean value of 80 ± 3 LSB from a lower value (or
dimmer image). If the mean value is too high, increase the Aperture value and continue the procedure
approaching the correct mean value from a dimmer image.
We recommend that you keep Aperture values for all Doses, 512 or 1024 Display, at at least 8 Aperture
steps (80) from Aperture range extremes (0 to 44) to keep selected Apertures in the middle of the Aperture
operating range. Use the ND Filter to maintain this requirement, if necessary.
Examples:
– To get an Iris Aperture value Aperture position that is greater than 360 with the ND Filter set to
OFF (Iris Value of 360) with a mean value of 80 ± 3 LSB for either 512 or 1024 Display of Doses
A, B, C, or D then:
Select ON as the ND Filter selection (filter in) in the Technical Menu screen for the affected Dose
and 512 or 1024 Display, and repeat the step that determines the correct Aperture position for the
affected Dose and 512 or 1024 Display.
– To get an Iris Aperture value Aperture position that is less than 80 with the ND Filter set to ON
(Iris Value of 530) with a mean value of 80 ± 3 LSB for either 512 or 1024 Display of Doses A,B,C
or D then:
Select OFF as the ND Filter selection (filter out) on the Technical Menu screen for the affected
Dose and 512 or 1024 Display, and repeat the step that determines the correct Aperture position
for the affected Dose and 512 or 1024 Display.
Note: In the case of you cannot adjust the iris aperture neither with NDF ON nor with NDF OFF to obtain
the right mean value, prefer to select NDF ON and adjust iris aperture to reach this mean value (this
case could happen very rarely).
5–23
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a Dose A, 512 Display, Aperture Setting that yields a mean value of 80 ± 3 LSB during a
Vascular Dose A, 512 Display Mask exposure.
If the mean value in the LEVEL ADJUST screen is not 80 ± 3 LSB, STD DEV=
MEAN=
record the value and go on to step 5.12. PEAK=
If the mean value in the LEVEL ADJUST screen is 80 ± 3 LSB, skip to
step 5.13.
Note: It is essential to make the measurement on an image in steady state, i.e. fourth image. Never calibrate
on a single shot image.
5.12 Aperture Value
1. Use the mouse to adjust iris aperture value in the Technical Menu.
Click on the IRIS APERTURE field and use the keyboard to enter a new Iris Aperture value. Larger
Aperture values result in lower mean values. Smaller Aperture values result in higher mean values.
2. Repeat step 7. in section 5.10 until the mean value for vascular the Dose A, 512 Display is 80 ± 3 LSB.
5–24
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Click on Store to store the final Aperture value for the Vascular Dose A, 512 display.
IRIS
The Iris Aperture value should be displayed in the Iris Aperture table field of the Technical Menu.
Aperture table.
Record final Aperture value and Neutral Density
Filter position for Vascular Dose A, 512 Display
in the “DATA RECORDS” section of the
Service Manual.
Determine a Dose A, 1024 Display Aperture Setting that yields a mean value of 80 ± 3 LSB during a
Vascular Dose A, 1024 Display Mask exposure.
Repeat these steps (above) for Doses B, C, and D for both 512 and 1024 Matrix.
5–25
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–26
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED PROCEDURE.
PURPOSE To calibrate the default Iris Aperture/NDF position used in CARDIAC mode for each
dose selection.
Note: Perform a prep; the Iris Ref. automatically goes to the default
value. Refer to the Iris Reference Table in Section 3.6 (i.e. the
first value for Dose D is 170 with the ND filter = ON).
PROCEDURE 1– On the Advantx Console select Dose A / DENS QUANT = OFF.
2– In the DLX Technical Menu enter the Iris Aperture for Dose A Cardiac default.
3– Take repeated 2–3 s runs, using the Exposure button, and adjust the focal spot (or
kVp) to obtain an exposure time between 3 ms and 9ms.
4– Adjust Iris Aperture value to obtain a mean value equal to 98 ± 2 LSB on the last
image, doing successive runs.
5– Store the final Aperture Value.
6– Repeat steps 1 to 5 successively selecting Dose B, C, D and Cine.
RESULTS Record final aperture values in the SCAT Report (Test Data Results).
5–27
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–28
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration procedure is to determine and load initial Aperture values for a particular Dose
selection (dose A through D, Cardiac and Cine) during subtracted DLX acquisitions.
Note: These Aperture values are used as a fixed Aperture for a particular Dose selection (Dose A through D,
Cardiac with Cine) during DLX Cardiac acquisitions in normal mode. In addition, this procedure can
be used to determine the proper positioning of the Neutral Density Filter for optimum Aperture
performance during any type of Cardiac acquisition.
SECTION 1
DLX CARDIAC WITHOUT DENS QUANT ACQUISITIONS
The Default Cardiac Aperture values are also used by DLX Application software without DENS QUANT
(ABC loop engaged). During these acquisition modes, selecting a particular Dose fixes the Aperture position
to the value determined for the selected Dose.
Because the Advantx DLX System does not perform an Aperture adjustment during any Trial sequence
without DENS QUANT (ABC loop engaged) selected, this Aperture position is not corrected for optimum
Peak Video.
SECTION 2
TOOLS REQUIRED
or 2 mm copper plate
5–29
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 3
PROCEDURE
3.1 Boot the Advantx Application Software and make the following Advantx Console soft key
selections
DLX CARDIAC
D TECHNIQUES
– 70 kVp
– Select PW (pulse width) at 10 ms
– Focal spot small or medium (depends on the tube)
– DLX CARDIAC MODE SELECTION
– DENS QUANT: OFF
– 6 inch Field of View
3.2 System Set–up
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to Job Card VF1
and the DLX Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
5–30
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Access the Technical menu and make the following menu selections:
– AEC ON
– Level Adjust = ON,
– Box Stat default: 50 x 50,
– NDF = ON.
***
Note: ***Perform a prep; the Iris Ref. automatically goes to the default value. See the Iris Reference Table
in Section 3.6 (i.e. the first value for Dose D is 170 with ND filter ON).
5–31
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During this procedure, the DLX System uses the Aperture and Neutral
Density Filter values entered in the IRIS APERTURE and
ND FILTER fields displayed on the Technical Menu Screen to
determine Advantx System Aperture position and Neutral
Density Filter position.
Therefore, throughout this procedure you must enter values in the IRIS APERTURE field and select the
proper ND FILTER (ON or OFF) to start each vascular dose for 512 and 1024.
Please refer to the Table 1 for default values already entered when beginning the calibration.
TABLE 1
IRIS APERTURE DEFAULT VALUES
5–32
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Examples:
– If the DLX System needs the Advantx System VIC to set the Aperture to position 14 and include the ND
Filter for cardiac Dose C, 512 Display, the Iris Value field parameter field will display a value of 590
(Aperture position 140 and 450 to reflect that the ND Filter is in the light field). To start this procedure
for cardiac Dose C, 512 Display enter 140 in the Iris Aperture field and select ON in the Technical Menu
ND Filter field.
– If the DLX System needs the Advantx System VIC to set the Aperture at position 33 and to omit the
ND Filter for cardiac Dose A, 512 Display, the Iris Value field parameter field will display a value of 330
(Aperture position 330 and 0 to reflect that the ND Filter is not in the light field). To start this procedure
for cardiac Dose A, 512 Display, enter 330 in the Iris Aperture field and select OFF in the Technical
Menu ND Filter field.
Note: In this procedure the maximum mean value is 98 ± 2 LSB.
3.7 Pick–Up Tube Image “fold–over” Misregistration
Due to the brightness response characteristics of the Advantx VIC TV Pick–up Tube, the Pick–up Tube tends
to “fold–over” the image if the image on the tube is too bright. You may notice this during the following steps
when you decrease the Aperture Value and the DLX System indicates a decrease in the mean value display.
Normally a decrease in Aperture value results in an increase in the displayed mean value.
To avoid this misregistration, approach the desired maximum mean value indication from a lower value (or
dimmer image). If the mean value is too high, increase the Aperture value and continue the procedure
approaching the correct mean value with a dimmer image.
We recommend that you keep Aperture values for all Doses (512 or 1024 Display) at least 8 Aperture Steps
(80) from Aperture range extremes (0 to 44) in order to keep all selected Apertures in the middle of the
operating range. Use the ND Filter to maintain this requirement, if necessary.
Examples:
– For an Iris Aperture position that is greater than 360 with the ND Filter OFF (Iris Value of 360) to obtain
a mean value = 98 ± 2 LSB for 512 Display only for Doses A,B,C, D or Cardiac and Cine then:
Select ON in the Technical Menu ND Filter field (filter in) for the affected Dose, and repeat the steps
that determine the correct Aperture position for this Dose.
– For an Iris Aperture position that is less than 80 with the ND Filter ON (Iris Value of 530) to obtain a
mean value = 98 ± 2 LSB for Doses A,B,C, D or Cardiac and Cine, then:
Select OFF in the ND Filter field (filter out) of the Technical Menu for the affected Dose and repeat the
steps that determine the correct Aperture position for this Dose.
Note: In the case of you cannot adjust the iris aperture neither with NDF ON nor with NDF OFF to obtain
the right mean value, prefer to select NDF ON and adjust iris aperture to reach this mean value (this
case could happen very rarely).
5–33
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine an Aperture Setting that yields a mean value of 98 ± 2 LSB during a Cardiac Dose A acquisition.
1. Position the cursor in the IRIS APERTURE field; on the keyboard type the desired Iris Aperture
value.
– Larger Aperture values result in lower mean values. Smaller Aperture values result in higher mean
values.
2. Repeat steps 3.9.6 through 3.11 until the mean value for Cardiac Dose A, 512 Display is within
specifications.
5–34
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Click on Store to store the final Aperture value for Cardiac Dose A.
IRIS
– Click on IRIS to view the Iris Aperture table values for Cardiac Dose A.
TABLE
The Iris Aperture value should be displayed in the Technical Menu IRIS APERTURE 512 field.
Record final the Aperture value and Neutral Density Filter position for Cardiac Dose A 512 in the SCAT
Report (Test Data Results).
5–35
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine Dose B, C, D Cardiac and Cine Aperture Settings that yield the specified mean values and store
these 4 values in DLX System data base by repeating the above steps for Doses B, C, D Cardiac and Cine.
Record the final Aperture values and Neutral Density Filter positions in the SCAT Report (Test Data Results).
Make a PREP and check that the value is automatically displayed in the field.
Click on IRIS to view the Iris Aperture table. Check that all information was correctly stored in the table.
TABLE
5–36
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Do not select (highlight) the TEST RUN button during the calibration
CAUTION process.
PURPOSE To calibrate the default Iris Aperture/NDF position used in 512/1024 DA/DSA
STEPPING modes for each A and B Dose selection.
SET–UP Positioner Grid IN / Basic / X–ray tube at minimum SID / PA
Advantx DSA: DOSE B 512 / No Sub / focal spot = depends on Dose / FOV 22
cm (9 inch) / 70 kVp / 200 mA
STEP: Table ready for stepping with collimator in Auto mode / focal
spot = depends on tube / FOV 22cm (9 inch) / 70kVp / 200 mA / 1
frame per sec / Infinite duration (symbol ”–”)
/ Masks after ON
DLX Be ready for ACQ on patient before entering TECHMENU
Techmenu / AEC = ON / Level Adjust = ON / 10 bits
Phantom From CV phantom: 2 1/4” plexiglass + 1/4 inch aluminum plate
filtration + 1/32 inch copper plate filtration on table top, or 2mm cop-
per, on the collimator
PROCEDURE 1– Select DSA and corresponding set–up parameters on Advantx Console.
2– Take repeated exposures and adjust SID to obtain a mAs value achievable in
STEPPING MODES (typically 5 mAs, with ± 0.1 mAs accuracy).
3– Select STEP 512 mode and corresponding set–up parameters on the Advantx Con-
sole. BE SURE THAT OPTIMIZ SUB BUTTON IS NOT SELECTED.
4– Select DOSE A. Dial in the exposure that will give the same mAs value as in step 2.
5– Adjust iris aperture value and set NDF position to obtain a mean value equal to
98 ± 3 LSB on the fourth image doing successive runs of 4 exposures.
6– Store the final Aperture Value.
7– Select dose B and dial 10 mAs.
8– Repeat steps 5 and 6.
9– Still in the STEP mode, repeat steps 4 through 8 selecting 1024 Matrix.
10– SELECT OPTIMIZ SUB BUTTON AND KEEP THIS STATUS UNTIL THE
END OF CALIBRATION. Repeat Steps 4 through 9 (starting with the STEP 512
mode) and adjust the iris to obtain a mean value equal to 80 ± 3 LSB.
RESULTS The Stepping mode is calibrated.
5–37
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–38
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration is to determine and load aperture values for a particular DSA stepping
acquisition Dose selection (Doses A and B, for both 512 and 1024 acquisitions).
SECTION 1
DESCRIPTION
The default Vascular DSA stepping Aperture values are used by DLX application software to set the
aperture position to the value determined during this procedure for a given Dose and 512/1024 display
selection.
SECTION 2
SYSTEM USAGE
The DLX System Database provides the operator with an Aperture position based on the selected Dose and
512/1024 display selection.
– This aperture is used in acquisition, and corresponds to a film sensibility. Because the DSA
STEPPING mode uses fixed time, the operator can select the Dose he wants (as with a manual
fluoro).
– The Dose value displayed on Advantx console (A or B) is only used during acquisition to inform
the DLX which Iris value to use. In this procedure, its modification has no effect.
– If a good image is obtained in DSA Stepping Dose A for a given technique, the same image should
be obtained in Dose B with twice the mAs value.
SECTION 3
WHEN TO PERFORM PROCEDURE
5–39
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
TOOLS REQUIRED
30 cm (12”) 30 cm (12”)
2 1/4” plexiglass from the CV phantom Cu
ÉÉÉÉÉÉÉÉÉ ÉÉÉÉÉÉÉÉ
OR
Alu
ÉÉÉÉÉÉÉÉÉ ÉÉÉÉÉÉÉÉ
+ 1/4” aluminum plate filtration plexiglass
Cu
Min SID
SECTION 5
PROCEDURE
1. Boot the Advantx Application Software, and make the following softkey selections on the Advantx
Console:
D TECHNIQUES
– VASCULAR 512 DSA
– 70 kVp
– 200 mA
– Large Focal Spot (depending on Dose)
Note: VIC – To be calibrated before performing this procedure.
– DLX VASCULAR MODE SELECTION
– NO SUB Image Display
– 9 inch Field of View (22cm)
– Dose B
5–40
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– PA = posterior anterior.
– Set collimator blades to fully open (field size).
– Position the filtration plates on the table top so that they are centered in the beam.
– Position table top at isocenter.
– Adjust II to 30 cm (12 inches) above isocenter.
– Set the X–ray tube to the minimum SID.
Boot the DLX Application Software and access the Technical Menu Screen on the DLX Console Monitor
as described below:
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
6. Access the Technical Menu.
7. Make the following Technical Menu selections:
– 10 bits,
– AEC = ON,
– Level Adjust = ON.
5–41
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: When you perform a prep the Iris Ref. automatically goes to the default value.
5–42
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Center and adjust the DLX ROI (Region of Interest), the area where the DLX System measures Peak Video,
size to approximately a 50 mm square (the default size) as displayed on DLX Console Image Monitor.
– Click on CAMERA CENTER to center, by default, the DLX ROI box on the monitor image.
Click again on CAMERA CENTER to cancel the function.
– Click on STATISTICS in the Technical Menu, and select BOX STATS.
– Click on RESIZE.
– Use the mouse to adjust the ROI box size to a 50 mm square on DLX Console Image monitor.
– When the box is adjusted, click on the CLOSE button to exit from the Box Statistics function.
Modify the Technical Menu Screen on the DLX Console Dialog monitor.
– AEC: ON
– LEVEL ADJUST: ON
Use the ARROW keys on DLX Console programmable keypad or the mouse to move cursor to desired
Technical Menu Screen fields. Use the DLX Console mouse and keyboard to modify entries in each field.
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
Technical Menu IRIS APERTURE and ND Filter fields to determine the Advantx System Aperture
position and Neutral Density Filter position.
Throughout this procedure, you must enter values in the Iris Apert and ND Filter fields to start each
Vascular Dose and 512 or 1024 Display Aperture Setting. To determine these initial values, decode the
values displayed in the ACQUISITION PARAMETERS field of the Technical Menu for the dose and
display to be calibrated.
The values displayed in the ACQUISITION PARAMETERS field for a particular Vascular Dose and 512
or 1024 Display represent the current DLX System database values.
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter. If ND filter is selected, the DLX System
adds a value of 450 to the Aperture position value.
5–43
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Prepare the Advantx console, the positioner console and the DLX console for an Acquisition in order to
calibrate the iris aperture.
– Check that on the Advantx console, the Mask After button is displayed.
MASK
AFTER
Note: If the Mask After button is not displayed, press the Mask Before button, then the Mask None button.
The Mask After button is now displayed (this avoids changes in the table top position).
5–44
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a Dose A, 512 Display Aperture Setting that will give a mean value of 98 ± 3 LSB during a Dose
A, 512 Display Mask exposure.
– Dose A
– Select the same kVp and mA as in step 5.7
– Select the same mAs as used in step 5.7
– Leave the Image Intensifier at the same setting (height) as was used in step 5.7
5–45
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Adjust the IRIS APERTURE value in the Technical Menu screen by positioning the cursor on the IRIS
APERTURE field. Use the DLX keyboard to type the desired value.
Note mean value displayed on image monitor. Adjust Iris Aperture and take several exposures until the
mean value = 98 ± 3.
Note: It is essential to make the measurement on an image in steady state, i.e. fourth image. Never calibrate
on a single shot image.
Store the final aperture value for DSA Stepping Dose A 512 Display as follows:
Record the final Aperture value and Neutral Density Filter position for the DSA Stepping Dose A, 512
Display in the SCAT Report (Test Data Results).
5–46
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a DSA Stepping Dose A, 1024 Display Aperture Setting that yields a 98 ± 3 LSB mean value
during a DSA Stepping Dose A, 1024 Display Mask exposure.
Determine a DSA Stepping Dose B, 512 Display Aperture Setting that yields a 98 ± 3 mean value during a
DSA Stepping Dose B, 512 Display Mask exposure.
Determine a DSA Stepping Dose B, 1024 Display Aperture Setting that yields a 98 ± 3 mean value during a
DSA Stepping Dose B, 1024 Display Mask exposure.
1. SELECT OPTIMIZ SUB BUTTON AND KEEP THIS STATUS UNTIL THE END OF THE
SECTION 5.14 CALIBRATION PROCEDURE.
2. Repeat steps 5.9 through 5.13 (starting with DSA–STEP 512 mode) and adjust the iris to obtain a mean
value equal to 80 lsb ± 3.
– DSA Step No EDR Dose A, 512
– DSA Step No EDR Dose A, 1024
– DSA Step No EDR Dose B, 512
– DSA Step No EDR Dose B, 1024
3. Record values in the OPTIMIZE column of the DSA Stepping Aperture Setting SCAT report.
5–47
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–48
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 1
DESCRIPTION
The default Vascular stepping Aperture values are used by DLX application software to set the Aperture
position to the value determined during this procedure for a given Dose and 512/1024 display selection.
SECTION 2
SYSTEM USAGE
The DLX System Database provides the operator with an Aperture position based on the selected Dose and
512/1024 display selection.
– This aperture is used in acquisition, and corresponds to a film sensibility. (In a fixed time mode,
the operator can select the dose he wants, as with a manual fluoro.)
– The dose value displayed on Advantx console (A or B) is only used during acquisition to inform
the DLX which Iris value to use. In this procedure, its modification has no effect.
– If a good image is obtained in Dose A for a given Technique, the same image should be obtained
in Dose B with twice the mAs value.
SECTION 3
WHEN TO PERFORM PROCEDURE
5–49
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
TOOLS REQUIRED
2 1/4” plexiglass
or, 2 mm copper
SECTION 5
PROCEDURE
1. Boot the Advantx Application Software, and make the following softkey selections on the Advantx
Console:
D TECHNIQUES
DA
– 70 kVp
– 200 mA
– 0.6 Focal Spot
D VIC
Note: To be calibrated before performing this procedure.
D DLX VASCULAR MODE SELECTION
– NO SUB Image Display
– 9 inch Field of View (22cm)
– Dose B
5–50
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Boot the DLX Application Software and access the Technical Menu Screen on the DLX Console Monitor:
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
5.4 Edit DLX Technical Menu
D Use the DLX pop-up Main Menu to access the Technical Menu.
5–51
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– AEC (ON),
– 10 bits,
– Level Adjust: OFF,
– OFFSET INC: 0,
– ND FILTER: OFF,
– Iris Aperture: 330,
– Stat Mode: RAW.
Note: Enter the aperture value (0 to 440) in IRIS APERT and the ND filter value (ON or OFF) based on the
value displayed in the ACQ PRMS field for the dose and matrix, 512 or 1024, mode selected.
Use the ARROW keys on the DLX Console programmable keypad or the mouse to move cursor to desired
Technical Menu Screen fields. Use the DLX Console mouse and keyboard to modify entries in each field.
5–52
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Center and adjust the DLX ROI (Region of Interest) (the area where the DLX System measures Peak
Video) size to approximately a 50 mm square as displayed on DLX Console Image Monitor.
– Click on CAMERA CENTER to center, by default, the DLX ROI box on the monitor image.
Click again on CAMERA CENTER to cancel the function.
– Click on STATISTICS in the Technical Menu, and select BOX STATS.
– Click on RESIZE.
– Use the mouse to adjust the ROI box size to a 50 mm square on DLX Console Image monitor.
– When the box is adjusted, click on the CLOSE button to exit from the Box Statistics function.
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
Technical Menu IRIS APERTURE and ND Filter fields to determine the Advantx System Aperture
position and Neutral Density Filter position.
Throughout this procedure, you must enter values in the Iris Apert and ND Filter fields to start each
Vascular Dose and 512 or 1024 Display Aperture Setting. To determine these initial values, decode the
values displayed in the ACQUISITION PARAMETERS field of the Technical Menu for the dose and
display to be calibrated.
The values displayed in the ACQUISITION PARAMETERS field for a particular Vascular Dose and 512
or 1024 Display represent the current DLX System database values.
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter (if ND filter is selected, the DLX System
adds a value of 450 to the Aperture position value).
5–53
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a Dose B, 512 Display mAs value that yields as accurately as possible one of the three following
values: 6.4 mAs, 8 mAs, 10 mAs.
Take repeated exposures using the Advantx handswitch and adjust the kVp on the Advantx console until
the console displays an exposure time of 40 ms ± 10ms and a mAs value very close to one of the three values
6.4, 8 or 10.
Note: The mAs, mA and kVp values are displayed on Advantx console.
Prepare the Advantx console, the AFM console and the DLX console for an Acquisition.
DA
– On the Advantx console, select DA .
DSA
– Center AFM table laterally. Longitudinally, move the table at least 20 cm from its limits.
– Add the aluminium and copper filtrations.
SELECT/
– If is highlighted on PVC, press VALIDATE.
VALIDATE
CLEAR STORE
– Press for more than five seconds, flashes.
SEQUENCE POSITION
5–54
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
STORE
– Press again.
POSITION
TEST/ TEST/
– Press simultaneously
RETURN RETURN
TEST/
D On the positioner console (PVC), press
RETURN
until the table goes to the second position (the collimator automatically closes).
D On the Advantx Console, select the minimum kVp, 0.25 mAs and 1024.
D On the DLX Console, close the iris and ND filter by selecting: ND Filter: ON
IRIS Aperture: 440
D Take an exposure, then press BOX STATS.
D Record the mean value, that is displayed on image monitor, on the SCAT Report (Test Data Results).
D Take a series of three exposures, and adjust the Offset INC until the mean value displayed, when you
press BOX STATS, is close to 2.5+ 0.5 LSB.
D Record the newly displayed offset value on the SCAT Report (Test Data Results).
D Calculate the ”light bias correction (LBCV)” as follows:
LBCV = VAROFF x 0.153 + 0.50,
where VAROFF is the absolute value of the OFFSET INC variation.
VAROFF = OFFSET INC value.
D Take the integer part of the LBCV and record the number in the table.
TEST/
D On the Positioner Console (PVC), press
RETURN
until the table arrives at the first position (the collimator automatically opens).
5–55
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: The OFFSET INC value used in paragraph 5.9 for the determination of light bias correction in 1024
matrix is also used later, when calibrating the iris apertures in that mode.
Note: There is no correction in DA 512 since the FLURO vs. Offset is calculated by the DLX in a similar mode
(DLX fluoro).
5.10 Dose A 512 Display Aperture
Determine a Dose A, 512 Display Aperture Setting that yields a 100 ± 5 mean value during a Dose A, 512
Display Mask exposure.
Pos nbr: 1 2
D On the DLX Console, enter DA EDR Dose A in the ACQUISITION PARAMETERS field of
Technical Menu Screen.
D Reset the OFFSET INC value to zero.
5.11 Aperture Value Adjustments
Adjust the IRIS APERTURE value in the Technical Menu screen by positioning the cursor on the IRIS
APERTURE field. Use the DLX keyboard to type the desired value.
Note mean value displayed on image monitor. Adjust Iris Aperture and take several exposures until the
mean value = 100 ± 5.
5–56
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Store the final aperture value for DA EDR DOSE A 512 Display as follows:
Record the final Aperture value, and Neutral Density Filter position for EDR Dose A, 512 Display, on the
SCAT Report (Test Data Results).
Determine a DA EDR Dose A, 1024 Display Aperture Setting that yields a 100 ± 5 mean value during a DA
Dose A, 1024 Dispaly Mask exposure.
Determine a DA EDR Dose B, 512 Display Aperture Setting that yields a 100 ± 5 mean value during a Dose
B, 512 Display Mask exposure.
5–57
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a DA EDR Dose B, 1024 Display Aperture Setting that yields a 100 ± 5 mean value during a
Dose B, 1024 Display Mask exposure.
D Reset Advantx and boot the “BOOT and CONFIG” Advantx calibration floppy.
D Select the system configuration unit where Light Bias values are set (refer to the Advantx calibration
User’s Guide).
D If the new mean value recorded on the SCAT report (Test Data Results) sheet is greater than the original
value, then select the DLX VASC DA EDR1024 field and add the Light Bias Correction Value (LBCV)
to the current number.
D If the new mean value recorded on the SCAT report (Test Data Results) sheet is smaller than the
original value, select successively the following fields of acquisition, and add the Light Bias
Correction Value to the current value.
– CONVENTIONAL, PULSCAN FLUORO, DLX/DF FLUORO CINE/DF DYNAMIC LIVE,
DLX CARD DEF, DLX VASC DA EDR 512, DLX VASC DA EDR 1024, PHOTOSPOT, SPOT–
FILM, FILM CHANGER
Note: This step, enter new light bias, is done to enter the Light Bias Correction used to correct the black level
in DA EDR 1024 as well as to be certain that the Light Bias value entered in the configuration is never
less than the recommended value.
Note: Original and new mean values are the values found in Section 5.9, record these values on the SCAT
report (Test Data Results) sheet.
5–58
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED PROCEDURE.
Note: This procedure is to be performed for the Bolus/Spin No Mask Mode on systems that do not have
the DSA Stepping option.
PURPOSE To calibrate the default Iris Aperture/NDF position used in the 512/1024 Bolus/Spin No
Mask Mode for the only available dose B in the No Mask mode.
SET–UP Positioner Grid IN / PA / Table = ISO / II 12”(30cm) above table top /
collimator in Auto mode
Advantx If only Bolus OR both Bolus and spin options are installed, select Bolus
No Mask Mode. If only the Spin option is installed, select SPIN to per-
form calibration .
Perform 512 and 1024 / NOSUB /large focal spot/ FOV 22 cm (9 inch) /
rate = 1 i/s for 5 s.
Do NOT select Start/Stop on positioner if Bolus is selected.
Select Start/Stop if Spin is selected.
DLX Be ready for ACQ on patient before entering TECHMENU
Techmenu / AEC = ON / 10 bits / Level Adjust = ON
Iris ref = **see Note below
Box Stat default: 50 x 50
Phantom If Bolus is selected: CV phantom, 2 1/4” plexiglass + 1/4 inch thick alumi-
num plate filtration + 1/32 inch thick copper plate filtration.
If Spin is selected: install 2 mm copper plate on the X–Ray tube cover, and
carry out the necessary (Adhesive tape, etc...) so that the plate remains in
place.
Note: Perform a prep; the Iris Ref. automatically goes to the default values:
DA EDR dose A is displayed on technical menu whereas dose B is selected on
Advantx console. This is normal. Refer to the Iris Reference Table in Section 5.7.
PROCEDURE 1– Select Dose B/512 on Advantx Console.
2– Press the Prep button to update the iris aperture on the Tech Menu to match that of
the Advantx dose.
3– Take one sequence (4 images). Expose time should be between 5 and 20 ms.
4– Adjust the Iris Aperture value and set NDF position, if necessary, to obtain a Mean
Value equal to 98 ±3 LSB on the image, on the fourth image dpoing successive runs
of 4 exposures.
5– Store the final aperture Value (click on STORE IRIS).
6– Repeat steps 1 to 5 selecting VASC 1024 Dose B.
RESULTS Record final Aperture values and Neutral Density Filter positions in the SCAT Report (Test
Data Results).
5–59
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–60
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration is to determine and load initial aperture values for Bolus chasing and Spin
acquisitions in the No Mask mode. In addition, this procedure can be used to determine the proper
positioning of the Neutral Density Filter for optimum aperture performance.
SECTION 1
DESCRIPTION
There are three acquisition modes for Bolus/Spin: Mask Before, Mask After, and No Mask.
In the Mask Before and Mask After Modes, the acquisition is done in the linear mode (low gain path in the
VIC camera) to create a high quality image for subtracted images.
In the No Mask Mode, the acquisition is done in the EDR mode (high gain path in the VIC camera) to
guarantee a large dynamic range in unsubtracted images.
In the Mask Before and Mask After Modes, the iris aperture data used by the system is the same as that used
for the DSA mode. Therefore, Dose A and B iris aperture values are the same as Dose A and B DSA mode
values.
In the No Mask Mode, the Dose B iris aperture data used by the system is the same as for DSA stepping
Dose A.
D For the Mask Before Acquisition and Mask After Acquisition modes refer to the Job Card CAL 0 B
Default Vascular Aperture Setting.
D For the No Mask mode, on a system with the DSA Stepping option, refer to the procedure in Job Card
CAL 0 D “Stepping Iris Aperture Calibration”.
D If the system does not have the DSA Stepping option, for the No Mask mode, carry out the following
procedure.
SECTION 2
SYSTEM USAGE
The DLX System database provides the operator with an initial recommended aperture position based on
the selected Dose and 512/1024 display selection.
SECTION 3
WHEN TO PERFORM PROCEDURE
Peak Video versus delta aperture step is not a linear function and is unique to each Advantx/DLX System.
Therefore, we recommend that you generate a new Software Aperture Calibration Curve whenever you
replace the TV Camera Pick–up Tube or recalibrate the Advantx VIC.
5–61
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
TOOLS REQUIRED
If only Spin option is installed, install 2 mm copper plate on the X–Ray tube cover, and carry out the
necessary (Adhesive tape, etc...) so that the plate remains in place.
SECTION 5
PROCEDURE
5.1 Boot the Advantx Application and make the following soft key selections on the Advantx Console
DLX VASCULAR
5–62
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
5–63
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
***
Note: ***Perform a prep; the Iris Ref. automatically goes to the default value. Be aware that the iris aper-
ture Bolus/Spin No Mask, Dose B is the same as for DA Stepping EDR Dose A. See the Iris Refer-
ence Table in Section 5.7.
5–64
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
IRIS APERTURE and ND FILTER fields displayed on the Technical Menu Screen to determine Advantx
System Aperture position and Neutral Density Filter position.
Therefore, throughout this procedure you must enter values in the IRIS APERTURE field and select the
proper ND FILTER (ON or OFF) to start each vascular dose for 512 and 1024.
Please refer to the following table for the default values to be entered when beginning a calibration.
TABLE 1
DEFAULT VALUES
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter. If ND Filter is selected, the DLX System
adds a value of 450 to the Aperture position value.
5–65
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Due to brightness response characteristics of the Advantx VIC Image TV Pick–up Tube, the Pick–up Tube
tends to “fold–over” the image if the image on the tube is too bright. You may notice this during the
following steps when you decrease the Aperture Value and the DLX System indicates a decrease in the Peak
Video display. Normally a decrease in Aperture value results in an increase in Peak Video display.
To avoid this incorrect registration, approach the desired mean value of 98 ± 3 LSB from a lower value (or
dimmer image). If the mean value is too high, increase the Aperture value and continue the procedure
approaching the correct mean value from a dimmer image.
We recommend that you keep Aperture values for 512 and 1024 Display at at least 8 Aperture steps (80)
from Aperture range extremes (0 to 44) to keep selected Apertures in the middle of the Aperture operating
range. Use the ND Filter to maintain this requirement, if necessary.
Examples:
– To get an Iris Aperture value Aperture position that is greater than 360 with the ND Filter set to
OFF (Iris Value of 360) with a mean value of 98 ± 3 LSB for the 512 and 1024 Display of Dose
B then:
Select ON as the ND Filter selection (filter in) in the Technical Menu screen for 512 and 1024
Display, and repeat the step that determines the correct Aperture position for 512 and 1024 Display.
– To get an Iris Aperture value Aperture position that is less than 80 with the ND Filter set to ON
(Iris Value of 530) with a mean value of 98 ± 3 LSB for Dose B 512 and 1024 Display, then:
Select OFF as the ND Filter selection (filter out) on the Technical Menu screen for 512 and 1024
Display, and repeat the step that determines the correct Aperture position for 512 and 1024 Display.
5–66
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a Dose B 512 Display, Aperture Setting that will give a mean value of 98 ± 3 LSB in Bolus/Spin
Dose B, No Mask Mode, 512 Display.
If the mean value in the LEVEL ADJUST screen is not 98 ± 3 LSB, record
the value and go on to step 5.12.
If the mean value in the LEVEL ADJUST screen is 98 ± 3 LSB, skip to STD DEV=
step 5.13. MEAN=
PEAK=
Note: It is essential to make the measurement on an image in steady state,
i.e. fourth image. Never calibrate on a single shot image.
1. Use the mouse to adjust iris aperture value in the Technical Menu.
Click on the IRIS APERTURE field and use the keyboard to enter a new Iris Aperture value. Larger
Aperture values result in lower mean values. Smaller Aperture values result in higher mean values.
2. Repeat step 7. in section 5.10 until the mean value for the Dose B, 512 Display is 98 ± 3 LSB.
5–67
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Click on Store to store the final Aperture value for the Bolus/Spin Dose B (which is the same as DA EDR
IRIS
The Iris Aperture value should be displayed in the Iris Aperture table field of the Technical Menu.
Record final Aperture value and Neutral Density Filter position for Bolus/Spin Dose B, No Mask Mode,
512 Display on the SCAT Report (Test Data Results).
Determine a Dose B, 1024 Display Aperture Setting that yields a mean value of 98 ± 3 LSB during a
Bolus/Spin Dose B (which is the same as DA EDR Dose A), No Mask Mode, 1024 Display exposure.
Repeat the procedures in sections 5.10 through 5.13 for the 1024 Display.
5–68
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Do not select (highlight) the TEST RUN button during the calibration
CAUTION process.
PURPOSE To calibrate the default Iris Aperture/NDF position used in 512/1024 DA/DSA
STEPPING modes for each A and B Dose selection.
SET–UP Positioner Grid IN / Basic / X–ray tube at minimum SID / PA
Advantx DSA: DOSE B 512 / No Sub / focal spot = depends on Dose / FOV 22
cm (9 inch) / 70 kVp / 200 mA
STEP: Table ready for stepping with collimator in Auto mode / focal
spot = depends on tube / FOV 22cm (9 inch) / 70kVp / 200 mA / 1
frame per sec / Infinite duration (symbol ”–”)
/ Masks after ON
DLX Be ready for ACQ on patient before entering TECHMENU
Techmenu / AEC = ON / Level Adjust = ON / 10 bits
Phantom From CV phantom: 2 1/4” plexiglass + 1/4 inch aluminum plate
filtration + 1/32 inch copper plate filtration, or 2mm copper, on the
collimator
PROCEDURE 1– Select DSA and corresponding set–up parameters on Advantx Console.
2– Take repeated exposures and adjust SID to obtain a mAs value achievable in
STEPPING MODES (typically 5 mAs, with ± 0.1 mAs accuracy).
3– Select STEP 512 mode and corresponding set–up parameters on the Advantx Con-
sole. BE SURE THAT OPTIMIZ SUB BUTTON IS NOT SELECTED.
4– Select DOSE A. Dial in the exposure that will give the same mAs value as in step 2.
5– Adjust iris aperture value and set NDF position to obtain a mean value equal to
98 ± 3 LSB on the fourth image doing successive runs of 4 exposures.
6– Store the final Aperture Value.
7– Select dose B and dial 10 mAs.
8– Repeat steps 5 and 6.
9– Still in the STEP mode, repeat steps 4 through 8 selecting 1024 Matrix.
10– SELECT OPTIMIZ SUB BUTTON AND KEEP THIS STATUS UNTIL THE
END OF CALIBRATION. Repeat Steps 4 through 9 (starting with the STEP 512
mode) and adjust the iris to obtain a mean value equal to 80 ± 3 LSB.
RESULTS The Stepping mode is calibrated.
5–69
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–70
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration is to determine and load aperture values for a particular DSA stepping
acquisition Dose selection (Doses A and B, for both 512 and 1024 acquisitions).
SECTION 1
DESCRIPTION
The default Vascular DSA stepping Aperture values are used by DLX application software to set the
aperture position to the value determined during this procedure for a given Dose and 512/1024 display
selection.
SECTION 2
SYSTEM USAGE
The DLX System Database provides the operator with an Aperture position based on the selected Dose and
512/1024 display selection.
– This aperture is used in acquisition, and corresponds to a film sensibility. Because the DSA
STEPPING mode uses fixed time, the operator can select the Dose he wants (as with a manual
fluoro).
– The Dose value displayed on Advantx console (A or B) is only used during acquisition to inform
the DLX which Iris value to use. In this procedure, its modification has no effect.
– If a good image is obtained in DSA Stepping Dose A for a given technique, the same image should
be obtained in Dose B with twice the mAs value.
SECTION 3
WHEN TO PERFORM PROCEDURE
5–71
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
TOOLS REQUIRED
30 cm (12”) 30 cm (12”)
2 1/4” plexiglass from the CV phantom Cu
ÉÉÉÉÉÉÉÉÉ ÉÉÉÉÉÉÉÉ
OR
Alu
ÉÉÉÉÉÉÉÉÉ ÉÉÉÉÉÉÉÉ
+ 1/4” aluminum plate filtration plexiglass
Cu
Min SID
SECTION 5
PROCEDURE
1. Boot the Advantx Application Software, and make the following softkey selections on the Advantx
Console:
D TECHNIQUES
– VASCULAR 512 DSA
– 70 kVp
– 200 mA
– Large Focal Spot (depending on Dose)
Note: VIC – To be calibrated before performing this procedure.
– DLX VASCULAR MODE SELECTION
– NO SUB Image Display
– 9 inch Field of View (22cm)
– Dose B
5–72
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– PA = posterior anterior.
– Set collimator blades to fully open (field size).
– Position the filtration plates on the table top so that they are centered in the beam.
– Position table top at isocenter.
– Adjust II to 30 cm (12 inches) above isocenter.
– Set the X–ray tube to the minimum SID.
Boot the DLX Application Software and access the Technical Menu Screen on the DLX Console Monitor
as described below:
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
6. Access the Technical Menu.
7. Make the following Technical Menu selections:
– 10 bits,
– AEC = ON,
– Level Adjust = ON.
5–73
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: When you perform a prep the Iris Ref. automatically goes to the default value.
5–74
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Center and adjust the DLX ROI (Region of Interest), the area where the DLX System measures Peak Video,
size to approximately a 50 mm square (the default size) as displayed on DLX Console Image Monitor.
– Click on CAMERA CENTER to center, by default, the DLX ROI box on the monitor image.
Click again on CAMERA CENTER to cancel the function.
– Click on STATISTICS in the Technical Menu, and select BOX STATS.
– Click on RESIZE.
– Use the mouse to adjust the ROI box size to a 50 mm square on DLX Console Image monitor.
– When the box is adjusted, click on the CLOSE button to exit from the Box Statistics function.
Modify the Technical Menu Screen on the DLX Console Dialog monitor.
– AEC: ON
– LEVEL ADJUST: ON
Use the ARROW keys on DLX Console programmable keypad or the mouse to move cursor to desired
Technical Menu Screen fields. Use the DLX Console mouse and keyboard to modify entries in each field.
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
Technical Menu IRIS APERTURE and ND Filter fields to determine the Advantx System Aperture
position and Neutral Density Filter position.
Throughout this procedure, you must enter values in the Iris Apert and ND Filter fields to start each
Vascular Dose and 512 or 1024 Display Aperture Setting. To determine these initial values, decode the
values displayed in the ACQUISITION PARAMETERS field of the Technical Menu for the dose and
display to be calibrated.
The values displayed in the ACQUISITION PARAMETERS field for a particular Vascular Dose and 512
or 1024 Display represent the current DLX System database values.
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter. If ND filter is selected, the DLX System
adds a value of 450 to the Aperture position value.
5–75
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Prepare the Advantx console, the positioner console and the DLX console for an Acquisition in order to
calibrate the iris aperture.
NO STEP
– On the Advantx console, select STEP.
STEP
SELECT/
– If VALIDATE is highlighted on PVC, press VALIDATE.
CLEAR
– On SEQUENCE press SEQUENCE for more than five seconds, and
STORE
POSITION will flash.
STORE
– Press POSITION button.
5–76
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
STORE
– Press POSITION
again.
TEST/ TEST/
– Press simultaneously RETURN RETURN
Determine a Dose A, 512 Display Aperture Setting that will give a mean value of 98 ± 3 LSB during a Dose A,
512 Display Mask exposure.
5–77
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Adjust the IRIS APERTURE value in the Technical Menu screen by positioning the cursor on the IRIS
APERTURE field. Use the DLX keyboard to type the desired value.
Note mean value displayed on image monitor. Adjust Iris Aperture and take several exposures until the
mean value = 98 ± 3.
Store the final aperture value for DSA Stepping Dose A 512 Display as follows:
Note: It is essential to make the measurement on an image in steady state, i.e. fourth image. Never calibrate
on a single shot image.
Record the final Aperture value and Neutral Density Filter position for the DSA Stepping Dose A, 512
Display in the SCAT Report (Test Data Results).
5–78
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a DSA Stepping Dose A, 1024 Display Aperture Setting that yields a 98 ± 3 LSB mean value
during a DSA Stepping Dose A, 1024 Display Mask exposure.
Determine a DSA Stepping Dose B, 512 Display Aperture Setting that yields a 98 ± 3 mean value during a
DSA Stepping Dose B, 512 Display Mask exposure.
Determine a DSA Stepping Dose B, 1024 Display Aperture Setting that yields a 98 ± 3 mean value during a
DSA Stepping Dose B, 1024 Display Mask exposure.
1. SELECT OPTIMIZ SUB BUTTON AND KEEP THIS STATUS UNTIL THE END OF THE
SECTION 5.14 CALIBRATION PROCEDURE.
2. Repeat steps 5.9 through 5.13 (starting with DSA–STEP 512 mode) and adjust the iris to obtain a mean
value equal to 80 lsb ± 3.
– DSA Step No EDR Dose A, 512
– DSA Step No EDR Dose A, 1024
– DSA Step No EDR Dose B, 512
– DSA Step No EDR Dose B, 1024
3. Record values in the OPTIMIZE column of the DSA Stepping Aperture Setting SCAT report.
5–79
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5–80
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
PROCEDURE SYNOPSIS
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED PROCEDURE.
PURPOSE To calibrate the default Iris Aperture/NDF position used in the 1024 Photospot mode for
each dose selection.
Note: Perform a prep; the Iris Ref. automatically goes to the default
values. Refer to the Iris Reference Table in Section 5.7 (i.e. the
first value for Dose D 1024 is 320 with NDF = ON).
PROCEDURE 1.Select Dose A on Advantx Console.
2.Press the Exposure button to update the Dose on the Tech Menu to match that of the
Advantx.
3.Take repeated exposures and adjust mA/Focal spot/ and kVp to obtain 40 ± 15 ms
exposure time.
4.Adjust the Iris Aperture value and set NDF position, if necessary, to obtain a Mean
Value equal to 42 ±2 LSB on the fourth image doing successive runs of 4 expo-
sures.
5.Store the final aperture Value (click on STORE IRIS).
6.Repeat steps 1 to 5 selecting successively dose B, C and D (see the Iris Reference
Table in Section 5.7 for values).
RESULTS Record final Aperture values and Neutral Density Filter positions in the SCAT Report
(Test Data Results).
5-81
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
5-82
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
The purpose of this calibration is to determine and load initial aperture values for a particular Photospot
acquisition Dose selection (Dose A through D for 1024 acquisitions). In addition, this procedure can be
used to determine the proper positioning of the Neutral Density Filter for optimum aperture performance.
SECTION 1
DESCRIPTION
The default Photospot aperture values are used by DLX Application software to initially set the aperture
position to the value determined during this procedure for a given Dose and 1024 display selection.
SECTION 2
SYSTEM USAGE
The DLX System database provides the operator with an initial recommended aperture position based on
the selected Dose and 1024 display selection.
SECTION 3
WHEN TO PERFORM PROCEDURE
We recommend that you generate the default Photospot aperture setting whenever you replace the TV
Camera Pick–up Tube or recalibrate the Advantx VIC.
SECTION 4
TOOLS REQUIRED
or 2 mm Copper.
5-83
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 5
PROCEDURE
5.1 Boot the Advantx Application and make the following soft key selections on the Advantx
Console
DLX PHOTOSPOT
D TECHNIQUES
– Start with Dose A
– 70 kVp
– Depends on the adjustment (start with 50 mA)
– Focal spot depends upon the tube (small or medium)
D DLX PHOTOSPOT MODE SELECTIONS
– NO SUB Image Display
– 9 inch Field of View (22 cm)
– PA = posterior anterior.
– Set collimator blades fully open (field size).
– Set table top at Isocenter.
– Place the 2 1/4” plexiglass, 1/4” aluminum and 1/32” copper filtration on the table top. Place the
filtration in the following order: copper on top, aluminum in the middle, plexiglass on the
bottom ; if using 2 mm Copper, place it on the collimator.
– Adjust the Image Intensifier to 12 inches (30 cm) above table top.
– Set X–Ray Tube (Vascular Systems) to the minimum SID.
5-84
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears. Refer to the DLX
Operator Manual for information on the DLX application, and selecting DLX screens.
2. Click on NEW PATIENT.
3. Create a new patient and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
5-85
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
***
Note: ***Perform a prep; the Iris Ref. automatically goes to the default value. See the Iris Reference Table
in Section 5.7.
5-86
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
During this procedure, the DLX System uses the Aperture and Neutral Density Filter values entered in the
IRIS APERTURE and ND FILTER fields displayed on the Technical Menu Screen to determine
Advantx System Aperture position and Neutral Density Filter position.
Therefore, throughout this procedure you must enter values in the IRIS APERTURE field and select the
proper ND FILTER (ON or OFF) to start each vascular dose for 1024.
Please refer to the following table for the default values to be entered when beginning a calibration.
TABLE 1
DEFAULT VALUES
These values reflect an Aperture position in tenths (from 0 to 440 corresponding to actual Advantx
Aperture positions from 0 to 44) and the position of the ND Filter. If ND Filter is selected, the DLX
System adds a value of 450 to the Aperture position value.
Example:
– If the DLX System requires the Advantx System VIC to set the Aperture at position 24 and bring
in the ND Filter for Photospot Dose C, 1024 Display, the Iris Value field parameter field will
display a value of 690 (Aperture position 240 and 450 to reflect that the ND Filter is in the light
field). To properly start this procedure for Photospot Dose C, 1024 Display enter the value 245
in the Iris Aperture field and select ON in the Technical Menu ND Filter field.
5-87
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Due to brightness response characteristics of the Advantx VIC Image TV Pick–up Tube, the Pick–up
Tube tends to “fold–over” the image if the image on the tube is too bright. You may notice this during the
following steps when you decrease the Aperture Value and the DLX System indicates a decrease in the
Peak Video display. Normally a decrease in Aperture value results in an increase in Peak Video display.
To avoid this misregistration, approach the desired mean value of 42 ± 2 LSB from a lower value (or
dimmer image). If the mean value is too high, increase the Aperture value and continue the procedure
approaching the correct mean value from a dimmer image.
We recommend that you keep Aperture values for all Doses, 1024 Display, at at least 8 Aperture steps
(80) from Aperture range extremes (0 to 44) to keep selected Apertures in the middle of the Aperture
operating range. Use the ND Filter to maintain this requirement, if necessary.
5-88
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Determine a Dose A, 1024 Display, Aperture Setting that yields a mean value of 42 ± 2 LSB during a
Photospot Dose A, 1024 Display exposure.
Note: The mean digital reading is important for the aperture setting. Since the contrast and brightness
defaults in technical menu are different than the defaults in applications, the image’s appearance is
different from tech menu to applications. Actually, the images in tech menu look darker than the
images in applications.
Note: It is essential to make the measurement on an image in steady state, i.e. fourth image. Never calibrate
on a single shot image.
1. Use the mouse to adjust iris aperture value in the Technical Menu.
Click on the IRIS APERTURE field and use the keyboard to enter a new Iris Aperture value. Larger
Aperture values result in lower mean values. Smaller Aperture values result in higher mean values.
2. Repeat step 7 in section 5.10 until the mean value for Photospot Dose A, 1024 Display is 42 ± 2 LSB.
5-89
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Click on Store to store the final Aperture value for the Photospot Dose A, 1024 display.
IRIS
The Iris Aperture value should be displayed in the Iris Aperture table field of the Technical Menu.
Photospot Dose A
Photospot Dose B
Photospot Dose C
Photospot Dose D
Repeat these steps (above) for Doses B, C, and D for 1024 Matrix.
5-90
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 1
DLX IMAGE QUALITY INTRODUCTION
This chapter provides instructions for the system’s initial Image Quality checks as well as
checks to correct Image Quality problems. Most of the Image Quality Job Cards are divided
into two parts: first, a short overview of the procedure, and secondly a detailed procedure.
Carry out the Calibration procedures before beginning the Image Quality checks. See the
Installation, Calibration, and Image Quality Steering Guides for more information. The
Image Quality Steering Guide is shown on page 6–3.
SECTION 2
COMMON PROCEDURES
For many of the procedures in this chapter, you will need to use the DLX TECHNICAL
MENU. To do this:
1. Log in as dlx.
2. When the system initialization has finished, the Patient Browser screen appears. Select,
start, or resume an exam before accessing the Technical Menu.
3. Position the cursor near the bottom of the screen and press the left mouse button.
The DLX Main Menu appears.
4. With the mouse button still pressed, select the DLX Others Menu.
5. Still pressing the mouse button, go to the TECHNICAL MENU and release the mouse
button.
2–2 Using the Box Statistics Function
6-1
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– Click on EXIT in the upper right hand corner of the Technical Menu screen.
2–4 Quitting DLXINSTAL
– Click on DLX Application in the pop-up DLX Main Menu to exit from
DLXINSTAL, and start the application.
2–5 Recording Image Quality Values
For each of the Image Quality procedures, you will need to record the results. IQ results
should always be recorded in the chapter Data Records, found in the back of this Service
Manual. Certain IQ values can also be recorded in the IQ Tool, for easy access.
1. Click on the Exit icon in the lower right corner of the flat panel screen. The
6-2
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–1
DLX IMAGE QUALITY STEERING
NO
NO
CHECK CHECK
. VID CHAIN GND
. BEAM CURRENT . VAD BOARD
. CABLES (VID +GEBUS) . DARK CURRENT
. DLX HIGH GAIN
. LIGHT BIAS
. P.U.T ROTATION
NOISE
DLX ONLY
IMG0B
3 SPATIAL
OK 4 MTF
UNIFORMITY OK
(VASC)
(VASC)
IMG0E
IMG0D
NO
NO
CHECK CHECK
. II FOCUS
. II LENS FOCUS
. GRID ALIGNMENT
. P.UT. LENS FOCUS
. XRT TO II ALIGNMENT
. CAMERA ELECTROMAGNETIC FOCUS
. PICK UP TUBE BEAM ALGN.
. P.U.T. ROTATION
. PICK UP TUBE BEAM ALIGNMENT
STEPS 1, 2, 3, 4 AND 5
OK WERE SUCCESSFULLY OK IRIS APERTURE OK
TEMPORAL
5 UNIFORMITY COMPLETED WITH NO
CURVE
TEST
(VASC) ADJUSTMENTS DONE IMG 0G
OK OK
IMG0F
NO NO
IRIS APERTURE
NO CALIBRATION IRIS APERTURE
CHECK (CARD = VASC = DA) CURVE
CAL B–C–D–F–H CALIBRATION
. LIGHT BIAS ADJ. CAL 0A
. IRIS
. DOSE
. AEC
IF OPTION PRESENT
CV LEVELS
MULTI–FORMAT
(VASC + CARD)
MONITOR CAMERA
IMG0H
ADJUSTMENTS ADJUSTMENT END
CHECK IMG0I + IMG0P (HARD–COPY)
. GAINS IMG0J
. IRIS
. EDR
MANDATORY
6-3
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–2
BIPLANE MODE IMAGE QUALITY STEERING
CHECK :
. TIMING EXP–GATE
. VBLANK SIGNALS
. GEIM OUTPUTS
. II BLANKING SIGNALS
. II G2 STABILIZATION
MANDATORY PROCEDURE
Check that the SOD of the lateral X ray tube is set at 70 cm (minimum SOD) when
CAUTION performing any JOB CARD on the lateral subsystem.
Note: For lateral calibration, set the x–ray beam in the same direction as the one described in the
JOB CARD for the frontal plane
Note: With the SOD set at 70 cm on the lateral plane, the configuration is similar to the frontal
configuration and JOB CARD are valid as well for calibrating or verifying the lateral
plane.
6-4
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To test for system noise levels in the VIC and the DLX. Perform the test in:
– VASCULAR or DSA–NO STEP 512 / 1024
– VASCULAR or DA–STEPPING 1024
– DLX Fluoro Manual Mode
SET–UP Positioner Grid IN / PA / collimator blades closed / 1 mm lead ( or 2 mm copper
filtration) over the collimator to block residual radiation
Advantx DSA–NO STEP 512 and 1024 / NO SUB / FS = small / min kVp /
lowest mA for the selected focal spot / mA min / Dose A /
mask to run = 1s / segment #1 = 1 i/s for 5s
DA–STEPPING 1024 / NO SUB / FS = Small / min kVp / min mA
Manually blank VIC Camera S225 / VIC 1 / A11 ( S225 ON = up)
FOV 9 inch (22 cm)
DLX Fluoro / Manual / kVp min / mA min
DLX Be ready for ACQ on patient before entering TECHMENU
Techmenu / AEC = OFF / exposure time = 2 ms / 10 bits / Iris = 400 /
NDF = ON / Level Adjust = ON
PROCEDURE 1. Select DSA – NO STEP 512 and corresponding set-up parameters on Advantx
Console.
2. Adjust OFFSET INC to 127. (Note that once you have taken an exposure a new,
different value, and not 127, is displayed; this is normal.)
3. Take an exposure.
4. Click on BOX STATS and move the box around the image – look for the maximum
Standard Deviation reading (do not measure on areas representing structured noise).
5. Check that the S.D. is within specifications, and record the value on the SCAT (Test
Data Results) data sheet.
6. Repeat Steps 1 through 4 selecting DSA – NO STEP 1024 Mode.
7. Repeat Steps 1 through 4 selecting DA – STEPPING 1024 Mode.
8. Repeat Steps 1 through 4 making DLX Fluoro exposures.
9. Return to normal operating conditions: S225 (OFF = down) on VIC 1 / A11.
6-5
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
6-6
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
This performance procedure checks that the noise level through the Advantx TV Camera to the DLX
processor and display video signal chain is within specified SNR values.
SECTION 1
TOOLS REQUIRED
SECTION 2
PROCEDURE
2.1 Boot the Advantx Application Software and select the following on the Advantx Console:
D TECHNIQUES
– DLX VASCULAR
– kVp min
– mA min
D DLX VASCULAR MODE SELECTION
– 512 Line Display
– NO SUB Image Display
– 9 inch (22 cm) Field of View
– Dose A
– Small focal spot
– Auto Inject–OFF
– Mask to Run Delay–1.O Sec
– Segment One Rate–1.O Sec
– Segment One Duration–5.0 Sec
– Manual
– DLX Fluoro
Refer to Advantx Operation Manuals for instructions on booting the Advantx Application Software and
selecting DLX Vascular screens on the Advantx Console.
6-7
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Refer to the DLX Operator Manual for instructions on booting the DLX application software, and selecting
DLX screens.
Manually blank the Advantx VIC TV camera and reset the Advantx console to enable Vascular
acquisitions.
1. Manually blank Advantx VIC TV Camera by placing S225 (MAN TARGET BLANK) on VIC Sync
and set the Timing Board VIC1 A11 to ON (up).
Note: You do not need to reset the Advantx, as the system may not pass offsets.
2. Reselect DLX VASCULAR using the softkey on the Advantx console.
2.4 Set–up the System
6-8
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Enter the following parameters in the DLX Technical Menu screen as shown in Illustration
AUCUN LIEN .
– AEC = 2 ms,
– 10 bits,
– Iris = 400,
– NDF = ON,
– Level Adjust = ON.
ILLUSTRATION 6–3
THE TECHNICAL MENU SCREEN
4
2
400
6-9
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Click on CAMERA CENTERING in the Technical Menu. The DLX ROI (Region of Interest) box
size is adjusted to 50 mm square (the default size).
2. Click again on CAMERA CENTERING to exit from the CAMERA CENTERING function.
2.8 Establish the Offset Value
In the DLX Technical Menu screen, click on BOX STATS in the STATISTICS menu. Move the box until
you reach the maximum standard deviation value.
Check to see if the standard deviation meets the specifications; if so, record the standard deviation value on
the SCAT Report (Test Data Results), and continue with the procedure. See also ”Recommendations” in
Section 2.12.
2.10 If the 1024 Display option is installed, repeat the preceding procedure
Repeat the procedure with the 1024 line display parameter selected on the Advantx console.
1. Unblank Advantx VIC TV Camera by placing S225 (MAN TARGET BLANK) on VIC Sync and set
Timing Board VIC1 A11 to OFF (down).
2. Reset the Advantx System by turning the Advantx console power first OFF, then ON.
6-10
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
2.12 Recommendations
If any of the standard deviation values exceed the specifications, do the following:
6-11
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-12
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To test for noise level on the DLX only. Perform the test in: DSA–NO STEP 512 / 1024,
or Vascular 512 / 1024.
6-13
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-14
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
This performance procedure checks that the DLX processor noise level and display video signal chain are
within specified SNR values.
SECTION 1
TOOLS REQUIRED
None.
SECTION 2
PROCEDURE
2.1 Boot the Advantx Application Software and select the following on the Advantx Console:
D TECHNIQUES
– DLX VASCULAR
– Minimum kVp
– Minimum mA
D DLX VASCULAR MODE SELECTION
– 512 Line Display
– NO SUB Image Display
– 9 inch (22cm) Field of View
– Dose A
– Small focal spot
– Auto Inject–OFF
– Mask to Run Delay–1.O sec
– Segment One Rate–1.O sec
– Segment One Duration–5.0 sec
6-15
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Refer to the DLX ”Operator Manual” for instructions on booting the DLX application and selecting
screens.
1. Disconnect the video cable from the DLX1A4T1 video filter input on the DLX cabinet.
2. Click on NEW PATIENT in the PATIENT BROWSER FROM DISK screen .
3. Create a new patient (name: NOISE – in uppercase letters) and enter an identification number.
4. Click on NEW EXAM.
5. Click on START EXAM.
The SEQUENCE VIEWER screen appears.
Note: Refer to the ”Technical Menu Engineer’s Guide” in the Service Manual for more information on the
Technical Menu field selections.
6-16
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–4
EDIT THE TECHNICAL MENU
4
2
4
400
Click on CAMERA CENTERING. The DLX ROI (Region of Interest) box appears on the monitor screen
adjusted to a 50 mm square (the default size).
1. Edit the DLX Technical Menu Screen displayed on the DLX Console monitor to reflect the parameter
field entries shown in Illustration AUCUN LIEN .
2. Enter the value 127 in OFFSET INC.
3. Take an exposure.
Note: Once you have taken the exposure a new, different value, and not 127, is displayed in OFFSET INC;
this is normal – continue.
6-17
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Click on BOX STATS and move the box until you reach the maximum standard deviation value.
Check to see if the standard deviation meets the specifications; if so, record the standard deviation value on
the SCAT Report (Test Data Results), and continue with the procedure. See also ”Recommendations” in
Section 2.9.
Select 1024 Matrix on the Advantx console, and repeat steps 2.1 through 2.6.
Reconnect the video cable from VIC to the DLX cabinet video filter.
2.9 Recommendations
If any of the standard deviation values exceed the specifications, check the VDP rack board DLX1A1A31
VAD3.
6-18
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 2 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To check and measure the actual black level on a digital image acquired for each DLX
acquisition mode and verify DLX offset calculation is correct.
SET-UP Positioner Grid IN / Collimator blades fully closed / 1 mm lead (apron) or copper
plates on collimator (2 x 1/32 inch, or 2 x 1 mm copper plates)
Advantx DSA-NO STEP 512 NO SUB / Power Off/On the Advantx console to
force the sytem to calculate offsets
For Vascular mode – Dose A: small focal spot / 40 kVp / mA = mini-
mum / FOV = nominal
For DLX fluoro: manual / minimum kVp, mA, and FOV
DLX Be ready for ACQ on patient before entering TECHMENU
Techmenu / 10 bits / NDF filter ON / Iris Aperture = 400 /
AEC = OFF / Exposure time = 2 ms, level adjust ON
Phantom 1 mm lead (apron), or copper plates on collimator (2 x 1/32 inch copper
plates, or 2 x 1 mm copper plates)
PROCEDURE 1– Select DSA–NO STEP 512 and corresponding setup parameters on Advantx Console.
2– Adjust OFFSET INCREMENT to 0.
3– Take five exposures; verify that the displayed mean values do not vary by more than
4 LSB. Record the last displayed mean value.
4– If the mean value is < 0.1, increase the OFFSET INCREMENT by 4 and repeat step
3. Continue to increment by 4; repeat step 3 until the mean value is > 0.1.
5– Calculate the black level:
Offset Increment
Black Level = Mean –
4
6– Record the displayed offset value and the calculated black level value on the SCAT
Records (Test Data Results).
7– Repeat steps 1 to 6 selecting 1024 mode.
8– Repeat steps 1 to 6 selecting DLX Fluoro.
9– Repeat steps 1 to 7 selecting DA–STEPPING mode. running fluoro for 5 seconds.
10– Disconnect Video cable from DLX filter and reboot the Advantx for offset
calculation.
11– Reconnect the Video cable.
6-19
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ANALYSIS Typical OFFSET values for all doses is 62 ± 4 (without Video level on DLX).
The means of the five exposures should not differ by more than 4 LSB.
Typical black level value for all modes should be +/–2 LSB:
–2 LSB black level 2 LSB
ACTION SYSTEM CHECK: Light Bias instability
PUT rotation
VIC dark current and beam alignment
VIC/DLX video gains
DIGITAL CHECK: VAD board offset adjustment
DETAILED PROCEDURE
SECTION 1
DESCRIPTION
This procedure measures the DLX black level (in LSB). The black level of an image is an important
parameter that affects the visibility of the arteries in the image background.
The black level is the level at which, for a given point on the Image Intensifier, absolutely no X–rays (direct
or scattered) fall. The black level is easily measured (lowest technical parameters, collimator blades
closed) for each DLX acquisition mode.
This procedure verifies also if the DLX offset calculation is correct without Video signal at the VAD input.
SECTION 2
WHEN TO USE PROCEDURE
SECTION 3
TOOLS REQUIRED
6-20
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
PROCEDURE
4.1 SET–UP
1. Power OFF then power ON the Advantx console to force an offset calculation (during the boot).
2. Boot the Advantx Application Software.
3. Make the following selections on the Advantx Console.
D DLX VASCULAR MODE SELECTION
– Dose A
– DSA No Stepping 512, No Sub
– 40 kVp
– Minimum mA
– Small focal spot
– FOV nominal
D FLUORO SELECTION
– DLX Fluoro
– Manual mode
– Minimum kVp, minimum mA
– FOV minimum
D POSITIONER
– Set the collimator blades to fully closed.
4. Place a strong filtration in the path of the X-ray beam, for example a 1 mm lead (apron) or 2 mm copper
plates on the collimator, so that no X-rays reach the II entrance plane.
Lead or
copper
X–ray tube
6-21
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <Return>. The PATIENT BROWSER FROM DISK screen appears.
2. Click on NEW PATIENT. Enter a new patient (name:) and an identification number.
3. Select NEW EXAM.
4. Click on START EXAM.
5. Carry out the following steps to access the Technical Menu:
Edit the DLX Technical Menu screen displayed on the DLX console dialog monitor so that the parameter
field entries are set to the values shown in Illustration 6–5. Use the mouse to move the cursor to the desired
Technical Menu fields. Use the mouse and the keyboard to edit the parameter fields.
Edit the IRIS APERTURE and ND Filter DLX Technical Menu screen parameter fields as follows:
6-22
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–5
THE TECHNICAL MENU SCREEN
400
1. Take five exposures, using the console handswitch. Check that the mean values displayed after each
exposure do not differ from each other by more than 4 LSB.
2. Record the last displayed mean value.
3. If this mean value is less than 0.1, then increase the OFFSET INCREMENT by 4 and repeat step 3.
Continue to increment by 4 and repeat step 3 until the mean value is > 0.1. The offset increment of 4 is
chosen because a 4 LSB offset increment corresponds to a change of 1 LSB mean value on the DLX
monitor.
4. Calculate the black level: Black Level = Mean Value – (Offset Increment 4). This value is the black
level value on the monitor, which corresponds to a dynamic range of 0 to 255 LSB on the DLX monitor.
5. Record the offset value (displayed on the DLX console) and the calculated black level value, on the
data sheet in the chapter “Data Records”.
4.5 Measure the Black Level and Offset values in the 1024 mode
6-23
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
4.6 Measure the the Black Level and Offset values in DLX fluoro (using the foot switch)
6-24
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 5
ANALYSIS
1. If any of the offset values are not 62"4, with the video disconnected from the VIC, check the DLX
VAD board.
2. The observed means of the sets of five exposures should not differ by more than 4 LSB. This confirms
black level stability.
3. In all acquisition modes, the calculated black level values should be: " 2 LSB (–2 v black level v 2).
Although, in reality, black level values on the DLX monitor can only be positive or nul, this calculated
black level value gives a representative value of the video signal’s black level value before digitization.
It represents the video’s black level, which corresponds to a scale of 0 to 255 LSB on the DLX monitor.
If any of these values are different than expected in any of the acquisition modes, check:
– light bias levels,
– PUT rotation,
– VIC black level,
– PUT beam alignment,
– VIC gains.
Note: This procedure can be performed with inverted horizontal and/or vertical sweeps, in order to
check the black level stability with these sweeps.
6-25
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-26
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 2 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To check the image spatial uniformity for the largest field of view.
6-27
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 2 of 10
ACTION System check: Grid Alignment, XRT to II adjustment, PUT alignment, and beam
alignment.
DETAILED PROCEDURE
This Image Performance Check procedure determines whether the Advantx DLX System imaging chain
meets specified Brightness Level Uniformity criteria across an unsubtracted flat field image.
SECTION 1
TOOLS REQUIRED
or; 2 mm copper
SECTION 2
PROCEDURE
2.1 Boot the Advantx Application Software and select the following on the Advantx Console:
D TECHNIQUES
– DLX VASCULAR
– Large focal spot
– 75 kVP
– 640 mA or 800 mA
D DLX VASCULAR MODE SELECTION
– 512 Line Display
– NO SUB Image Display
– Dose D
– Mask to Run Delay – 1.0 Sec
– Segment One Rate – 1.5 Sec (60Hz) – 1.3 Sec (50Hz)
– Segment One Duration – 2.2 Sec
– Segment One Delay – 0.0 Sec
– Segment two Duration – 0.0 Sec
– Segment Three Duration – 0.0 Sec
6-28
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 3 of 10
ILLUSTRATION 6–6
SEGMENT STATISTICS SCREEN ON MONITOR
L1
L2
D FOV SELECTIONS
– Select the nominal FOV:
- 9 inch Field of View (23 cm) for 9 inch (23 cm) Image Intensifiers, or
- 12 inch Field of View (30 cm) for 12 inch (30 cm) Image Intensifiers, or
- 16 inch Field of View (40 cm) for 16 inch (40 cm) Image Intensifiers.
6-29
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 4 of 10
1. To access the Technical Menu, click on Technical Menu in the DLX Others Menu:
ILLUSTRATION 6–7
THE TECHNICAL MENU
4
10
6-30
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 5 of 10
2.5 Adjust the DLX ROI box and Reset the ROI Box to Default Size
Note: The DLX ROI (Region of Interest) box is the area where DLX System measures mean level and standard
deviation.
1. Click on CAMERA CENTERING in the Technical Menu to enter the Camera Centering function.
The DLX ROI (Region of Interest) box is adjusted to 50mm square (the default size).
2. Click on CAMERA CENTERING a second time to exit the Camera Centering function.
1. Edit the Technical Menu on the DLX Console monitor so that the parameter field entries are the same
as those shown in Illustration 6–7.
2. Use the mouse to move the cursor to the desired Technical Menu fields. Use the mouse and keyboard
to modify the values in each field.
Find a Advantx Aperture position that yields a Mean Value (Level) of 100 3 indication during a Vascular
Dose D, 512 Display, with an exposure time of 15 ms or more:
1. Take a Vascular Dose D, 512 Display, exposure, using the Advantx Console Handswitch. Start with the
Exposure Time set to 10 ms.
2. Click on BOX STATS in the Technical Menu to enter Box Statistics function; check the Mean Value
and the Exposure Time.
3. If the Mean Value indicated on the DLX Console Image monitor screen is not 100 3:
Record the value, then adjust the EXPOSURE TIME value in the Technical Menu (use the mouse to
position the cursor on the EXPOSURE TIME field, and the keyboard to enter the desired value).
Repeat from step 1., using the new exposure time.
– Larger Exposure Time values result in larger mean value indications.
– Smaller Exposure Time values result in smaller mean value indications.
6-31
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 6 of 10
4. If the Mean Value indicated on the DLX Console Image monitor screen is 100 3:
Check that the exposure time is 15 ms or more. If it is not, select a lower mA value on the Advantx
console. Repeat from step 1., using the new mA value.
5. Repeat steps 1. through 4. until the mean value indicated for Vascular Dose D, 512 Display is 100 3,
and the Exposure Time is 15 ms or more.
Determine the DLX System Brightness Uniformity for the Mean Value of the 100 3 image determined in
step 2.6 (the % difference between the maximum and minimum level must be < 40%) using the Technical
Menu SEGMENT STATS function.
270 90
L1 LINE
starting point L1 LINE
ending point
180
3.2. Position the starting point in the left portion of the circle. Click with the mouse to confirm the
L1 Line starting point. See diagram on illustration 6–8.
3.3. The DLX monitor displays the L1 Line with a circled end–point. Place L1 Line end–point cross
hair in the right portion of the circle. See diagram above.
3.4. Check that L1 Line is 367 (for 60Hz DLX), or 400 for (50Hz DLX), pixels in length (as
displayed in lower left box on image) and that it is horizontal, near the horizontal line of the
camera centering function (approximately at 270° to 90°).
6-32
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 7 of 10
4. Draw a SEGMENT STATS vertical L2 continuous dotted line, across the center of the image
(approximately 0° to 180°). The line should be 367 pixels in length (60Hz DLX), or 400 pixels in
length (50Hz DLX).
4.1. Position the starting point in the upper portion of the circle. See Illustration 6–9.
Click to confirm the L2 starting point.
4.2. Position the L2 ending point. See Illustration 6–9.
The DLX Console Image Monitor displays the L2 continuous line with a circled end–point
cross hair.
ILLUSTRATION 6–9
0 - 360 L1 LINE
starting point
270 90
180 L1 LINE
ending point
4.3. Check that the vertical L2 continuous line is 367 (for 60Hz DLX, or 400 (for 50Hz DLX), pixels
in length (as displayed in lower left box on image), and is vertical, close to the vertical line of
the camera centering function (approximately at 0° to 180°).
5. Click on the left button of the mouse three times to display a full SEGMENT PROFILE DISPLAY
image on DLX Console Image monitor.
6-33
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 8 of 10
6. From the full SEGMENT PROFILE DISPLAY image, record the following absolute values:
ILLUSTRATION 6–10
_____________ ____________
2. Repeat the procedures 2.6 through 2.8 for 1024. The lines length is now 800 pixels (50Hz system) or
734 pixels (60Hz pixels).
6-34
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 9 of 10
Whatever the matrix size, the Brightness Uniformity value should be:
2.12 Recommendations
1. For each II and smallest measured FOV, if the Brightness Uniformity value is not correct, the video
camera is likely to be faulty. Repeat the measurements with Horizontal and Vertical sweeps reversed.
Check and/or adjust, in the corresponding 512 or 1024 mode:
– P. U. Tube alignments
– Beam alignments
2. If the Brightness Uniformity value is correct for the smallest FOV, but not for the largest FOV, the
problem is probably linked to one or more of the following:
Grid non-uniformity; XRT non-uniformity;
I. I. non-uniformity; Geometric misalignment between these components.
6-35
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
BRIGHTNESS UNIFORMITY
Job Card IMG 0 D 10 of 10
Blank page
6-36
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To measure the image system MTF in VASC 512 and 1024 at two spatial frequency
values which represent system use.
SET–UP Positioner SID 1m / Grid IN / collimator Auto mode
Advantx 512 or 1024 / 50kVp / 100 mA / 9 inch FOV (22 cm) / Small focal spot
/ No SUB / Dose D
DLX Be ready for ACQ on patient before entering TECHMENU
Technical Menu / 10 bits / AEC = OFF / start with 20 ms Exposure time
Phantom Resolution wedge, taped to the center of the grid with the lines running
vertically on the monitor, centered at:
0.6 lp/mm (50Hz) in 512 and 1024 (0.55 lp/mm in 60Hz), or
1.2 lp/mm (50Hz) in 1024 (1.1 lp/mm in 60Hz)
PROCEDURE 1– Select 512. The NDF and Iris aperture are automatically set when you do a prep.
2– Center the resolution wedge in fluoro as above. The measurement of white level,
black level and standard deviation should be made as in Illustration 6–11.
3– Take exposures, adjusting the exposure time to obtain a white level of 200 ± 3 lsb;
store the white level value (see Illustration 6–11 for the box position). Set LEVEL
ADJUST to ON.
4– Measure the white level with LEVEL ADJUST (see Illustration 6–11) and store it
(click on WHITE LEVEL).
5– Measure the black level with BOX STATS (see Illustration 6–11) and store it (click
on BLACK LEVEL).
6– Measure the standard deviation in the resolution wedge with BOX STATS (see
Illustration 6–11) and store it (click on STD DEV).
7– Record the MTF value from the Technical Menu on the SCAT Records (Test Data
Results).
8– Select 1024, Dose D. (The ND filter and Iris Aperture are automatically set.) Repeat
steps 2 through 7 for : 0.6 lp/mm and 1.2 lp/mm (50Hz) or,
0.55 lp/mm and 1.1 lp/mm (60 Hz).
9– Record the MTF value from the Technical Menu on the SCAT Records (Test Data
Results) data sheet.
ANALYSIS For 512 the MTF > 40% at 0.6 lp/mm
For 1024 the MTF > 40% at 0.6 lp/mm / MTF > 18% at 1.2 lp/mm
ACTION – Adjust the PUT/II lens focus (with iris open = maximum
standard deviation).
– Adjust the PUT/II focalisations.
6-37
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–11
BOX POSITIONS FOR BLACK/WHITE LEVELS
STEP
5
STEP
3 STE
P6
6-38
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
This measurement procedure gives the value of the modulation transfer function (MTF) for particular
spatial frequencies in the imaging system. The MTF is one of the most important parameters in the imaging
system.
SECTION 1
DESCRIPTION
The MTF is measured at a spatial frequency of 0.6 lp/mm for 512 Matrix, and at two spatial frequencies, 0.6
lp/mm and 1.2 lp/mm, for 1024 Matrix. These spatial frequency values were chosen because they are
representative of the detail size the user observes on the monitor, and because an imaging system
misadjustment can easily be detected for these values. Depending on the line frequency, the measurement
values chosen are 0.6 lp/mm and 1.2 lp/mm for 50HZ, or 0.55 lp/mm and 1.1 lp/mm for 60Hz.
The imaging system MTF is found by using a resolution wedge centered on the II entrance plane, and
measuring two densities, and one standard deviation, on the monitor image using the DLX Box Statistics
function.
SECTION 2
WHEN TO USE PROCEDURE
SECTION 3
TOOLS REQUIRED
6-39
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
PROCEDURE
4.1 SET–UP
Boot the Advantx Application Software. Select the following on the Advantx Console:
D TECHNIQUES
– DLX VASCULAR
– 50 kVp
– 100 mA
– Smallest Focal Spot for selected kVp and mA
D DLX VASCULAR MODE SELECTION
– 512 Line Display
– NO SUB Image Display
– Dose D
– 9 inch Field of View (22 cm)
– L and U positions to 0°
D POSITIONER
– Adjust SID to 1 meter (3 feet).
– Leave the Image Intensifier grid in place.
– Set the collimator blades to the Automatic mode.
6-40
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears.
2. Click on NEW PATIENT. Enter a new patient (name:) and an identification number.
3. Select NEW EXAM.
4. Click on START EXAM.
5. Access the Technical Menu.
4.3 Edit the DLX Technical Menu
Edit the DLX Technical Menu screen displayed on the DLX console dialogue monitor so that the
parameter field entries are set to the same values as those shown in Illustration 6–12. Use the mouse to
move the cursor to the desired Technical Menu fields. Use the mouse and the keyboard to edit the
parameter fields.
Start with a 20 ms exposure time, 10 bits, and AEC = OFF in the Technical Menu.
ILLUSTRATION 6–12
THE TECHNICAL MENU SCREEN
20
6-41
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– Tape the resolution wedge on the II grid and, using the DLX fluro, center it as shown in Illustration
6–13. The area to be measured, 0.6 lp/MM for 50HZ or 0.55 lp/mm for 60Hz, must be centered
on the camera centering function cross. The lower spatial frequencies should be towards the top
of the image, with the higher frequencies towards the bottom of the image.
– Click on CAMERA CENTERING a second time to exit from the Camera Centering function.
– Use the BOX STATS function to position the DLX ROI box over the
resolution wedge.
6-42
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–13
MTF MEASUREMENT SET–UP
PATIENT HEAD
RESOLUTION
CAMERA CENTERING WEDGE
CIRCLE
PATIENT FEET
POSITION RESOLUTION WEDGE SO THAT
MAIN AXIS OF PATTERN IS PERPENDICULAR
TO TV SCAN DIRECTION
6-43
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Find an Advantx Exposure Time that yields a mean value of 200 ± 3 during a vascular Dose D, 512 Display
Mask exposure.
1. Take a Vascular Dose D, 512 Display Mask exposure using the Advantx Console handswitch.
– Set LEVEL ADJUST to ON.
– If the Mean Value indicated on the DLX Console Image Monitor screen is not 200 3, go to step
2.
– If Mean Value indicated on DLX Console Image Monitor screen is 200 3, store the white level
value in the White Level area of the DLX Technical Menu, and go to step 4.7.
2. Adjust the exposure time (ms) value in the Technical Menu using the mouse to move the cursor to the
Exposure time (ms) parameter field. Use the keyboard to enter the desired value.
Note: Start with a low value (i.e. between 5ms to 20ms) to prevent fold–over misregistration. Larger Exposure
Time values result in larger mean value indications. Smaller Exposure Time values result in smaller
mean value indications.
3. Repeat steps 1. and 2. until the Mean Value indication for Vascular Dose D, 512 Display, is 200 3.
4.7 Determine MTF value for 512 Matrix
Determine DLX System MTF value for a Vascular Dose D, 512 MTF Value Display exposure using the
Exposure time determined in step 4.6.
1. Determine and load Black Level value for the DLX ROI box that is placed in the flat response gray area
of the resolution wedge, displayed on the image acquired in step 4.6.
– Click on BOX STATS to enter Box Statistics function. ROI BOX
– Position the ROI box to the left in the lead area of the
resolution wedge.
or
6-44
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
– Adjust the ROI box to the largest possible size without covering over any of the resolution wedge
markings.
– Store the black level value in the Technical Menu BLACK LEVEL field.
2. Determine the Standard Deviation value for a wide ROI box that is centered over the line pairs at the
0.55 lp/mm or 0.6 lp/mm resolution wedge mark in the image from step 4.6. Load this Standard
Deviation value.
ROI BOX
– Click on BOX STATS in the STATISTICS
menu to enter Box Statistics function.
– Adjust the ROI box height to a value approximately five times smaller than the box width. It may
be necessary to reduce the width to avoid touching the gray areas of the resolution wedge.
– Store the standard deviation value in the Technical Menu STANDARD DEVIATION field.
– The system automatically calculates the MTF value and displays it in the MTF field.
– Record the MTF value in the SCAT Records (Test Data Results).
6-45
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
If any of the measured values is lower than the given limit, in the appropriate mode, 512 or 1024, check:
This procedure describes the MTF measurement procedure for a 9 inch (22 cm) FOV. It is, however, also
possible to measure other FOVs.
Due to time/cost restraints, it is not necessary to repeat this procedure for other FOVs each time you check
the MTF. The 9 inch (22 cm) MTF is effective for adjusting the video camera, although it does not provide
information for other FOV focalizations, or Camera Lens adjustment.
Therefore, the two following procedures are strongly recommended for II or PUT focalization
optimization and for Camera lens adjustment.
6-46
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Measure the MTF value as above for the chosen FOV in 512 or 1024 mode, in the APPLICATION
mode.
2. In the SERVICE mode, record the corresponding value for the II or PUT focalization to be adjusted.
3. In the SERVICE mode, change the II or PUT focalization value.
4. Repeat steps 1. through 3. until the MTF value reaches its maximum.
The following procedure is also recommended for camera lens adjustments:
1. Open the VIC iris near the maximum setting (for example 10), and set the NDF Filter to ON in the
Technical Menu.
2. Adjust exposure time so that the mean value in the centered 0.6 lp/mm area is within 100 and 120 LSB
(that corresponds to about 200 LSB for the white area, but level accuracy is not needed at this time).
3. Read the standard deviation in the 0.6 lp/mm area.
4. Rotate the camera lens.
5. Repeat steps 2. and 4. until the standard deviation is maximum. The techniques must not change during
these operations.
Note: Because you are not performing a measurement, you do not have to adjust the white level to 200 LSB
during this lens adjustment. For the best results at this point, rotate the lens to obtain the highest
standard deviation value. When you have obtained this, the maximum difference between black and
white lines in the resolution wedge has been reached, and the MTF is at its maximum value.
6-47
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-48
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To test the system’s ability to give the same levels during an entire sequence.
SET–UP Positioner Grid IN / PA / Table = ISO / II = 30 cm (12”) above table top
Advantx Vascular 1024 / 80 kVp / 640 mA / Dose C / No Sub /
FOV = 9 inches (22cm) / Large Focal Spot
DLX Acquire / New Patient / Name
Phantom CV II phantom, centered on the image.
PROCEDURE 1– Measurement in FAST RUN, perform the step 3.4.
ACTION Check: light bias, iris aperture and Corrective Factor Setting T009, X–Ray Generator per-
formance, etc... See paragraph 3.6.
6-49
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 2 of 8
Blank page
6-50
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 3 of 8
DETAILED PROCEDURE
SECTION 1
DESCRIPTION
This Image Performance test procedure determines whether the Advantx DLX System provides good
stability of image brightness during an aquisition. This parameter is particularily important for substracted
images.
Note: This procedure performs the measurement in the same way it is done in Advantx–E Calibration User’s
Guide for LCA/LCV+/LC+, T009 Unit, in the verification part.
SO it is not necessary to redo the procedure if T009 has been done successfully and if nothing has been
changed after. In that case, just copy the results of T009 Tables 2 and 3 into tables 1 and 2 of this
procedure.
SECTION 2
TOOLS REQUIRED CV II Phantom
CV II Phantom, including :
30 cm (12”)
– CV Phantom with :
aluminum
ÉÉÉÉÉÉÉ
ÄÄÄÄ
– 2 1/4” plexiglass from CV phantom plexiglass
SECTION 3
PROCEDURE
3.1 Boot the Advantx Application Software and select the following on the Advantx Console
. TECHNIQUES
. DLX VASCULAR
. Large focal spot
. 80 kVp
. 640 mA
. DLX VASCULAR MODE SELECTION
. 1024 Matrix
. NO SUB Image Display
. Dose C
. FOV SELECTION
6-51
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 4 of 8
2. Place the CV II Phantom on the table top, centered carefuuly in the beam.
4. Adjust the Image Intensifier so that it is 30cm (12”) from the table top.
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears.
6-52
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 5 of 8
6.3 (7.5) fps 0.0s 3.1 (3.8)fps 0.0s 1.6 (1.9) fps
0.5 s 3.0 s 5.0 s 5.0 s
2. Perform a TRIAL using default ROI, with no manual adjustment of SIZE, POSITION, or LEVEL.
5. Adjust size and position of a measuring box on the CV phantom LUNG Region, close to the center (see
figure).
Be sure NOT TO MOVE the box between this and following sequences.
ILLUSTRATION 6–1
6. On dilution curve, measure levels on images corresponding approximately to times of 3, 7 and 12s and the
level of which is representative of the steady state in the segment.
Note: Levels shall be strictly inferior to 250. If MEAN>250, calibration of iris values shall be checked.
7. Complete table 1
6-53
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 6 of 8
2.1 (2.5) fps 0.0s 1.6 (1.9) fps 0.0s 1.1 (1.3) fps
0.5 s 8.0 s 8.0 s 10.0 s
2. Perform a TRIAL using default ROI, with no manual adjustment of SIZE, POSITION, or LEVEL.
5. Adjust size and position of a measuring box on the CV phantom LUNG Region, close to the center (see
figure).
Be sure NOT TO MOVE the box between this and following sequences.
ILLUSTRATION 6–2
6. On dilution curve, measure levels on images corresponding approximately to times of 7, 14 and 24 s, and
the level of which is representative of the steady state in the segment.
Note: Levels shall be strictly inferior to 250. If MEAN>250, calibration of iris values shall be checked.
7. Complete table 2
6-54
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 7 of 8
3.6 Analysis
This procedure is designed to verify that the level stability during one acquisition is right, for the most
difficult types of acquisitions for the system.
If results in table 1 (Relative Difference in Fast Run) or in first part of table 2 (Relative Difference in Slow
Run) are not within the expected accuracy (± 4%), that means some component has not a good stability;
check:
. iris aperture
. light bias
. X–Ray generator performances (filament boost, kVp waveform, etc)
. AEC stability
. II high voltage, power supply
. PUT grids power supply
. XRT output ratio, tube spits
To help finding the origin of the problem, same procedure could be performed, changing the acquisition
conditions (i.e; acquisition matrix, FOV, etc) or mounting the camera head on the IET to separate the
camera from the system.
Table 2 compares same frame rate in presence of a fast segment or not. If results in second part of table 2
FAST to SLOW Relative Difference) are not within the expected accuracy, that means the iris corrective
factor is not well calibrated, so check:
. light bias
. iris aperture
. Iris corective factor setting (T009 for 1024 matrix)
6-55
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
TEMPORAL UNIFORMITY
Job Card IMG 0 F 8 of 8
TABLE 1
VERIFICATION (FAST RUN)
Approximative Time 3 7 12
(s)
Image Number
(1) (2) (3)
MEAN LEVEL
TABLE 2
VERIFICATION (SLOW RUN)
Approximative Time 7 14 24
(s)
Image Number
(4) (5) (6)
MEAN LEVEL
6-56
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To test two points on the look–up table, which represent the function dose = f aperture,
for a constant brightness on the PUT in the progressive scan mode (512 and 1024).
ANALYSIS For each dose A and D, you should get: m1 x 0.9 < m2 < m1 x 110
m2 – m1
RATIO = x 100 ≤ 10%
m1
ACTION If test is not successful, refer to the system adjustment Software Aperture Calibration
procedure in CAL 0 A.
6-57
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-58
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
SECTION 1
DESCRIPTION
This calibration test checks two points on a software aperture calibration curve which represent the gamma
correction function of the Advantx Pick–Up Tube.
The software aperture calibration curve is used by the DLX application software to automatically choose
the correct aperture setting after a trial exposure to ensure that proper Peak Video levels are maintained
during subsequent exposures.
A table of Peak Video levels (scaled from 0 to 255) versus delta Aperture Step is used by the system
software to project Mask Peak Video from Trial Peak Video and to calculate the recommended Aperture
change. This table is only used in progressive scan, low–gain operations (DLX 512 or 1024 Vascular
acquisitions). The resulting Peak Video to Aperture functions are nonlinear and provide correction for the
gamma characteristics of the Advantx VIC TV Pick–up Tube.
SECTION 2
SYSTEM USAGE
The DLX System Application Software Database provides the operator with an initial Aperture position
based on the operator selected Dose. (See Job Card CAL O B to set the initial Aperture position in the
Vascular mode).
During a DLX 512 or 1024 Vascular acquisition, the Aperture position is recalculated and automatically
reset after a Trial exposure.
After each subsequent DLX 512 Vascular or 1024 Vascular Trial exposure, the Aperture position is
recalculated and automatically reset based on the Peak Video measured in the DLX generated ROI (Region
of Interest) window.
Use the mouse to change the box size. Press the right button to change the size (repeat to stop the size
change). Press the left mouse button until the stats are calculated.
SECTION 3
WHEN TO PERFORM PROCEDURE
Peak Video versus delta Aperture step is not a linear function and is unique to each Advantx/DLX System.
Therefore, we recommend that you test this Software Aperture Calibration Curve whenever you replace
the TV Camera Pick–up Tube or recalibrate the Advantx VIC.
If this test fails, generate a new Software Aperture Calibration Curve using Job Card CAL O A.
6-59
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
TOOLS REQUIRED
SECTION 5
PROCEDURE
D TECHNIQUES
– DLX VASCULAR
– 60 kVp
– 160 mA (to be adjusted later to obtain an Exposure Time greater than 25ms)
– Smallest Focus
D DLX VASCULAR MODE SELECTION
– 512 Line Display
– NO SUB Image Display
– Segment One Rate – 1.0 i/Sec
– Segment One Duration – 5 Sec
– Dose A
– 9 inch Field of View (22 cm)
5.3 System Set–up
1. Place one1/32 inch copper plate (or 1mm copper) on top of the set–up collimator (Vascular systems) or
on a table top centered in FOV (R and F systems).
2. Set the Image Intensifier to 12 inches (30 cm) above table top.
3. Adjust the X-ray tube (Vascular systems) to minimum SID.
6-60
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Login: dlx <cr>. The PATIENT BROWSER FROM DISK screen appears.
2. Click on NEW PATIENT. Enter a new patient (name: CURVE TEST) and an identification number.
3. Click on NEW EXAM.
4. Click on START EXAM. The SEQUENCE VIEWER screen appears.
5. Take a Trial exposure and check that you have an Exposure Time of greater than 12.5ms for this Trial
(that means actual exposure will be greater than 25ms). If it is not, adjust the mA to obtain more than
12.5ms.
6. Perform the first acquisition and wait for its completeion.
5.5 Install the second copper plate and perform a second run
1. Place the second 1/32 inch copper plate (or 1mm copper) on the previously installed copper plate.
2. Take a Trial exposure and check that you have an Exposure Time of greater than 12.5ms for this trial
(that means actual exposure will be greater than 25ms). If not, adjust the mA to obtain more than
12.5ms.
3. Perform a second acquisition and wait until completion.
2. Click on TOOL KIT and then on REGION STATS to display the REGION STATISTICS menu.
1. Select the fourth image of the first run (image acquired with one copper plate).
2. Click on BOX STATISITCS and note the mean value displayed on the image monitor: MEAN 1.
Record this value in the ”DATA RECORDS”section of the Service Manual.
3. Select the fourth image of the second run (image acquired with two copper plates).
6-61
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
4. Click on BOX STATISITCS and note the mean value displayed on the image monitor: MEAN 2.
Record this value on the SCAT Report (Test Data Results) data sheet.
5. Calculate the mean ration and record it on the SCAT Report (Test Data Results) data sheet. The mean
ratio should be less than 10%.
MEAN 2 – MEAN 1
R = ____________________
M1
Note: MEAN 1 and MEAN 2 values should not differ by more than 10 %.
Repeat the complete procedure (5.2 to 5.7) with the parameters Dose D and 70 kVp selected on the Advantx
Console.
Repeat the complete procedure (5.2 to 5.8) with 1024 Matrix selected on the Advantx Console.
For each Dose A and D, and each 512 and 1024 Matrix, the R ratio should be less than 10%. If it is not, the
iris aperture curve is not correctly adjusted.
Perform the Software Aperture Calibration Procedure, CAL 0 A, then repeat the check procedures in this
Job Card.
6-62
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
6-63
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
B – VASCULAR
1– In vascular, the exposure time should not vary by more than 20 % compared to the
Ref. Exp. value on the SCAT data sheet.
2– The V values should not differ by more than 5% from the reference levels given in
the data sheet.
3– Check that the CV2 values fall within the ranges listed below:
Area: lung heart media abdomen ball
512 Dose C Range: 200/250 90/140 20/60 10/40 5/20
1024 Dose C Range: 200/250 90/140 20/60 10/40 5/20
6-64
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–3
BOX POSITIONS FOR CV LEVEL MEASUREMENTS
lead DISK
area HEART
One of 4
boxes
ABDOMEN
DETAILED PROCEDURE
SECTION 1
DESCRIPTION
This measurement procedure establishes a historical database of the video levels obtained through a CV
phantom. This database must begin when the entire system is completely and correctly calibrated, and
especially after installation. When it is first performed, reference levels are set, to which all subsequent
measurements over the life of system should be compared.
The proper transmission of image levels from the II entrance plane to the DLX monitor is one of the most
important functions in the system. This transmission can vary over time, due to changes in component
characteristics such as XRT, II, and PUT during their life time.
SECTION 2
WHEN TO PERFORM THIS PROCEDURE
6-65
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 3
TOOLS REQUIRED
SECTION 4
PROCEDURE
1. Boot the Advantx Application Software and select the following on the Advantx Console for DLX
Cardiac in the Application mode:
– 70 kVp
– Large focal spot
– FOV 6 inch (16 cm)
– Pulse width adjusted to maximum.
– Dose A
– 25 i/s (50 Hz), or 30 i/s (60 Hz)
– Grid in
2. Set collimator to the Auto mode.
3. Set the table top at isocenter.
4. Place CV1 phantom on the table top (without aluminum and copper).
5. Adjust the SID to 1 meter (40”).
Note: PA = posterior–anterior.
1. Login: dlx <Return>. The PATIENT BROWSER FROM DISK screen appears.
2. Click on NEW PATIENT. Enter a new patient (name:) and an identification number.
3. Select NEW EXAM.
4. Select START EXAM.
6-66
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. On the Advantx Console, select Cardiac, Autostabilize mode, 25 i/s (for 50 Hz) or 30 i/s (for 60 Hz).
2. Select Cardiac Dose A, set SID to 1m, and position CV1 as described below.
– Using DLX Technical Menu ; select Camera Centering Function..
– Take a DLX fluoro exposure and position the CV1 phantom center at the image center with HEART
at the top, and LUNG to the right of the image monitor. See Illustration 6–3 for the CV phantom
layout.
3. Take a cardiac run ; change X-ray tube focus as required to obtain an exposure time greater than 1.5 ms
and less than 9 ms.
4. Exit from DLX Technical Menu. In application, take a cardiac run (2 to 3 seconds).
5. In Application, click on Advanced Processing, then REGION STATS, then BOX STATS to access
the BOX STATS function. Use a 32 x 32 pixels box (about 1000 pixels) to measure the CV1 phantom
levels in the five areas (heart, abdomen, media, lung, and ball). The center of each box should be at the
middle of the each radius of the shutter circle. See Illustration 6–3.
6. Record this set of C1 values on the SCAT Report (Test Data Results) sheet.
7. Repeat steps 2. through 7 for Doses B, C, and D.
ILLUSTRATION 6–4
THE TECHNICAL MENU SCREEN
A – 207
6-67
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Position the CV2 phantom on the table top and set the SID to 1 meter (40”).
2. On the Advantx Console, select Vascular 512/NOSUB.
3. Make the following selections:
– 1 i/s for 5 sec
– Large focal spot
– FOV = 6 inches (16 cm)
– 80 kVp
– 640 mA
– Dose C
4. In the DLX Technical Menu, select AEC ON and Camera Centering Function. Verify the CV Phantom
center is at the image center using DLX fluoro.
5. The first time you perform this check procedure, take an exposure (AEC ON) and adjust the mA to
obtain an exposure time of 30 ±10 ms. Record this as the reference exposure time (ref. EXP).
6. Exit from DLX Technical Menu ; in Application take four images.
7. In the Application, click on Advanced Processing, then REGION STATS, then BOX STATS to
access the BOX STATS function. Use the BOX STATS function to measure the CV phantom levels in
the five areas (heart, abdomen, media, lungs, and ball) using a 32 x 32 pixels box (about 1000 pixels) in
512 matrix or a 64 x 64 pixels box (about 4000 pixels) in 1024 matrix. The center of each box should be
at the middle of each radius of the shutter circle. See Illustration 6–3.
8. Record this set of values (V) on the SCAT Records (Test Data Results) sheet.
9. Repeat this Vascular Measurement procedure for Vascular 1024.
6-68
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
C1 If any one of the C1 set values differs by more than 5% from the reference level obtained during SCAT
measurements, check:
– DLX Cardiac offsets and gains,
– kVp in secondary contacting,
– Cardiac II entrance doses,
– Beam current
– Iris apertures.
– Brightness uniformity of the video camera
Check that the values fall within the ranges listed below:
Area: lung heart media abdomen ball
Range: 140/190 70/120 20/50 10/30 5/20
C2 If one of the C2 set values is equal to 255 LSB, or if there is a visible saturation (fold–over) in the image,
check:
– EDR adjustments,
– Beam currents.
6-69
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
If the exposure time is 20% higher than the reference exposure time (ref. EXP), the AEC gives a higher exposure
time. Check:
– XRT technical parameters (kVp, mA),
– XRT output ratio,
– II conversion factor,
– II entrance Doses.
If any one of the V set values is more than 5% different than the reference level, check:
– DLX Vascular offsets and gains,
– kVp in primary contacting,
– Iris apertures,
– Brightness uniformity of the video camera
– Beam currents.
– Vascular upper threshold set up and setp optimal level parameters in DLX INSTAL software.
Check that the CV2 values fall within the ranges listed below:
Area: lung heart media abdomen ball
512 Dose C Range: 200/250 90/140 20/60 10/40 5/20
1024 Dose C Range: 200/250 90/140 20/60 10/40 5/20
6-70
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To adjust monitor contrast and brightness according to ambient light in the room, and to
optimize these adjustments.
TOOLS – Luminance–meter (recommended model: Minolta LS100) and stand
– Calibrated gray reference card, part no. 2110360, with known reflection ratio RG (RG is
often 18%)
– 2 1/4” plexiglass + 1/4 inch thick aluminum plate filtration + 1/32 inch thick copper
plate filtration from CV phantom, or 2mm copper
SET–UP Room monitors positioned as they are for the physician during the exam.
All light measurement is done from the area where the physician is positioned during an
exam (near the patient).
DLX in Technical Menu. ADX Console = DLX Fluoro / Manual mA at minimum setting.
PROCEDURE 1– Position the gray reference card on the monitor screen.
2– Measure the reflected light on the gray card: LG.
3– Calculate the reference luminance: LR = LG / RG
and the ambient illumination: IA = LR x π.
Record these on the SCAT (Test Data Results). IA must be between 30 and 50 lux.
4– Remove the gray card, power OFF the monitor (power plug disconnected) and mea-
sure the phosphore luminance: LP.
5– Calculate the monitor screen reflection ratio RM = LP/LR
RM must be between 7.5 % and 8.5 %.
6– Power ON the monitor, wait several minutes for the monitor to stabilize.
7– Do a short fluoro without X–Ray on the II (collimator closed, lead on the II) and
measure the luminance at zero dose (black level): LO.
8– Adjust the monitor brightness to get: LO = LP x K, with K = 1.5 for vascular, and K
= 2 for cardiac.
9– With the DLX gray level 15 (255 LSB), measure the maximum luminance: LMAX.
10– Adjust the monitor contrast to get: LMAX = LO x 150.
11– Repeat steps 7 to 10 until you have LO and LMAX equal to the calculated values.
Record LO and LMAX on the data sheet.
12– Do Fluoro in Auto mode (collimator open with the copper filtration) and measure the
luminance LN at nominal (regulated) dose.
LN should be between 30 and 40 cd/m2. Record this value on the SCAT (Test Data
Results) data sheet.
ACTION See page 2 of this Job Card.
6-71
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ACTION – There must be no direct light reflecting on the monitor screen; if there is, modify the
room lighting to obtain better conditions.
– If IA > 50 lux, decrease the room lighting.
– LMAX can vary between LO x 120 and LO x 180, but must not exceed 360 cd/m2,
otherwise readjust LO to a lower level to get LO between LP x 1.2 and LP x 1.5.
– If LN is not between 30 and 40 cd/m2 and the monitor has been correctly adjusted,
there may be a misadjustment of video gains.
6-72
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
This adjustment procedure adjusts the room monitor brightness and contrast according to ambient light and
optimizes the image gray scale according to the dynamic range of the video.
SECTION 1
DESCRIPTION
A clear view of all the of the radiological image details on a monitor depends mainly on the monitor
observation conditions and the operating conditions, versus the dynamic range of the video signal.
This procedure helps to measure the observation conditions and adjust the operating conditions.
SECTION 2
WHEN TO PERFORM THIS PROCEDURE
The monitor, the final component on the video chain is quite important to Image Quality.
We suggest that you perform this monitor check procedure each time the customer is not perfectly satisfied
with its image quality, or each time an adjustment is made on the monitor.
SECTION 3
TOOLS REQUIRED
Calibrated gray reference card (RG is often 18%) with known reflection ratio RG, Part number 2110360
From CV phantom:
2 1/4” plexiglass
or, 2 mm copper
6-73
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
PROCEDURE
1. Boot the Advantx Application Software and select the following on the Advantx Console:
– VASCULAR or CARDIAC,
– DLX Fluoro,
– 75 kVp,
– Manual mA at minimum.
2. Access the DLX Technical Menu.
3. In the Technical Menu, adjust the Manual mA to its minimum.
4. The positioner, table top, and room monitor should be positioned in the room as they are under working
conditions. Check:
– the monitor position,
– its height from the floor,
– its tilt,
– its direction towards the user.
5. The room lighting should be the same as under normal working conditions (check during fluoro if
there is an automatic switching of lights). Check:
– the artificial lighting (fluorescent, incandescent lighting),
– the adjustment of variable lighting,
– if there is natural lighting, and if it is the main lighting (curtains or door openings, meteorological
conditions).
6-74
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: In vascular or cardiac rooms, it is important to have a low level of lighting to get a good contrast in the
blacks, yet the physician will need sufficient light on area of the patient where he is to introduce the
catheter. Therefore, a directional spot light is recommended, which provides a lighted area with a
diameter of several decimeters (approximately 2 to 5).
6. Position the luminance–meter (if possible on its stand on the table top) to about the same level and
position as the physician eyes would be under working conditions.
Typically, the monitor screen is between 1.1 and 1.9 meters for a high line rate, and twice this for a low
line rate.
4.2 Luminance Procedure
1. Position the gray reference card in the center of the monitor screen with the gray card towards the
luminance meter.
2. Measure the reflected light on the gray card center: LG.
3. Calculate the reference luminance : LR = LG / RG.
Note: 1. RG is the reflection ratio of the gray card; this card should be clean with no fingerprints.
Note: 2. The measurement can be repeated on the whole card. The ratio between the higher and lower values
should not exceed 2. This could happen if a spot light illuminates the monitor screen. If so, modify the
monitor position or the room lighting conditions.
6-75
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Completely close the collimator blades and place a strong filtration over the X–Ray beam (i.e. lead
shirt the II entrance plane).
2. Do a short fluoro (1 to 2s) at manual, and minimum mA. Release the fluoro foot pedal.
Note: In the DLX Technical Menu, the LIH function stores the last fluoro image – it is not necessary to
continue fluoro during all of the measurement.
3. Measure the screen center luminance at zero dose, L0. L0 is the luminance at zero dose; it corresponds
to 50 mV on the monitor’s video input. It may be necessary to change this level.
4. Adjust the monitor brightness knob to get:
L0 = LP x K where:
K = 1.5 for vascular, or K = 2 for cardiac.
5. Select and display GRAY 15 (255 LSB) in the DLX Technical Menu.
6. Measure the maximum luminance (LMAX) on the screen center. LMAX is the maximum luminance at
DLX level 255 LSB; it corresponds to 650 mV on the monitor’s video input. It may be necessary to
change this level.
7. Adjust the monitor contrast knob to get: LMAX = L0 x 150.
8. Repeat steps 2. through 7. until L0 and LMAX are equal to the calculated values.
Note: LMAX can vary between L0 x 120 and L0 x 180, but must not exceed 360 cd/m2; if this occurs, readjust
L0 to a smaller value between LP x 1.2 and LP x 1.5.
9. Record L0 and LMAX on the SCAT (Test Data Results) data sheet.
10. Open the collimator blades and remove the filtration from the X–Ray beam.
6-76
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: If the lighting conditions are good and the monitor has been correctly adjusted, LN should be between
30 and 40 cd/m 2. If it is not, video gains may be incorrectly adjusted.
4. Record LN on the SCAT (Test Data Results) data sheet.
6-77
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-78
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DSA PATTERN
16 LEVELS
6-79
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–5
MULTIFORMAT CAMERA REFERENCE CURVES V3.1
LASER UP
6-80
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 1
DESCRIPTION
This procedure checks that both planes acquire cardiac or vascular independently. This means that the image
acquired on one plane is the result only of the X–Rays coming from the XRT of that same plane. Scatter or
even direct X–Rays from the other plane should not affect this image.
This procedure uses particular positioner set–up positions to lighten the II of one plane with the XRT of the
other plane. This lets you check that the biplane acquisition timings are correct.
SECTION 2
WHEN TO PERFORM PROCEDURE
SECTION 3
TOOLS REQUIRED
6–81
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
PROCEDURE
4.1 SET–UP
Boot the Advantx Application Software and select the following on the Advantx Console for both frontal and
lateral planes, depending upon whether the mode to be tested is Cardiac or Vascular.
6–82
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
2. Enter 100 as the new value for VASC SET–UP OPTIMUM LEVEL , in the DLXINSTAL tool, using the
procedure below.
6–83
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Click on NEW PATIENT. Enter a new patient (name:SCATTER) and an identification number
(ID:BIPLANE).
– Frontal positioner
. C–Arc = 0
. Horizontal pivot = 0
. Minimim SID
. Grid IN
– Lateral positioner
. Vertical pivot = 55
. Maximum SID
. C–Arc = 50
. Smallest SOD (distance from XRT to isocenter)
. Grid IN
1. Position the table top at isocenter.
2. Position CV phantom filtrations on the table top with the plexiglass on top of them.
3. Adjust the filter plates so that the lateral beam gets enough filtration.
– Check that the filtration is properly positioned: Select RECORD VIEW FRT and LAT (biplane
mode) and, taking repeated trials, move the filtration to obtain maximum dose ratio mAs LAT /
mAs FRT (usually 4) with an exposure time close to 50 ms (slightly modify the kVp / mA to adjust
the exposure time).
4.5 Frontal test acquisitions
1. Frontal Acquisition
6–84
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
2. Biplane Acquisition
– Frontal positioner
. C–Arc = 0
. Horizontal pivot = –50
. Maximum SID
. Grid IN
– Lateral positioner
. Vertical pivot = 0
. C–Arc = 0
. Minimim SID
. Maximum SOD (distance from XRT to isocenter)
. Grid IN
For CARDIAC mode selection:
3. If testing VASCULAR:
6–85
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
2. If testing CARDIAC:
. Do a 2 s acquisition run.
3. If testing VASCULAR:
. Do the complete sequence without taking a trial exposure (otherwise the iris aperture is
changed).
NOTE For the VASCULAR mode, be sure the exposure time displayed during the acquisition is close
to the maximum available (100 ms) on both planes.
Box Stats measurements are done with BOX STATS for the Frontal testing (Rep. Lateral testing).
2. Select the sequence acquired in step 1. of section 4.5 (for the lateral step 1. in 4.7).
6. Click on BOX STATS; record the mean value on the data sheet in ”DATA RECORDS” as frontal/single
(do the same for lateral/single).
8. Select the sequence acquired in step 2 of section 4.5 (for the lateral step 2 in 4.7).
12. Click on BOX STATS. Record the frontal/biplane mean value on the data sheet in ”DATA RECORDS”
(do the same for lateral/biplane).
Note: Use the BOX STATS default position (center) even though IIs do not project exactly on the image
center.
6–86
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Calculate the difference in LSB between frontal/single and frontal/biplane mean values (do the same for
lateral).
2. Determine the variation by dividing the above difference by the level recorded in FRONTAL RECORD
VIEW mode (frontal/single). Do the same for lateral.
4. Restore the original value for VASC SET-UP OPTIMUM LEVEL. Access DLXINSTAL as described in
section 4.2.
SECTION 5
ANALYSIS
If any resulting value for the frontal or lateral mode, in vascular or cardiac, does not meet the specified
values, check the following:
. II tube blanking signals (voltage on TP5/TP7 and BLK II PULSE on XRII board),
. II tube G2 stabilization,
. timing (EXPOSURE GATE / VBLANK SIGNALS).
Note: Typical values recorded in Vascular mode on a system where the Frontal II blanking was no longer
working:
frontal/single = 61 LSB
frontal/biplane = 182 LSB.
6–87
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6–88
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Note: SEE PAGE 3 OF THIS JOB CARD FOR THE DETAILED CHECK PROCEDURE.
PURPOSE To adjust monitor contrast and brightness according to ambient light in the room, and to
optimize these adjustments.
TOOLS – Luminance–meter (recommended model: Minolta LS100)
– Calibrated gray reference card, part no. 2110360, with known reflection ratio RG (RG is
often 18%)
SET–UP Console monitor positioned as it is for the physician during image processing.
All light measurement is done from the area where the physician is positioned during
image processing.
DLX in application mode.
PROCEDURE 1– Position the gray reference card on the monitor screen.
2– Measure the reflected light on the gray card: LG.
3– Calculate the reference luminance: LR = LG / RG
and the ambient illumination: IA = LR x π.
Record these in ”DATA RECORDS”. The IA should be between 10 and 30 lux, and
should not exceed 50 lux.
4– Remove the gray card, power OFF the monitor (power plug disconnected) and mea-
sure the phosphore luminance: LP.
5– Calculate the monitor screen reflection ratio RM = LP/LR. The RM must be less than
9 %.
6– Power ON the monitor, and wait several minutes for the monitor to stabilize.
7– On the SMPTE pattern, adjust the monitor brightnes so that you can just see the 5%
square inside the 0% square.
8– Display an image that doesn’t have too much contrast, and turn the DLX keypad
brightness/contrast buttons fully CCW to get a completely black image.
9– Measure the L0, and check that it matches the L0 value for the measured IA value
(see the L0 curve in Illustration 6–6). Readjust the monitor brightness, if necessary.
10– Turn the DLX keypad brightness knob fully CW to get a completely white image.
11– Measure LMAX, and adjust the monitor contrast so that the LMAX matches the LMAX
value which corresponds to the measured IA value.
12– Repeat steps 7 to 11 until the L0 and LMAX match the recommended values for IA.
13– Record the L0 and LMAX on the SCAT records (Test Data Results) sheet.
ACTION See page 2 of this Job Card.
6-89
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ACTION – There must be no direct light reflecting on the monitor screen; if there is, modify the
room lighting to obtain better conditions.
– If IA > 50 lux, decrease the room lighting.
– LO should fall within LP x 1.1 and LP x 1.3.
– The LO should not be too high, so as to not to “dazzle” the user’s eyes.
6-90
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
DETAILED PROCEDURE
This adjustment procedure helps the service engineer measure the observation conditions, and adjust the
operating conditions accordingly.
SECTION 1
DESCRIPTION
The accurate perception of radiological image details on a monitor depends mostly on the observation and
operating conditions, as opposed to the dynamic range of the video signal.
It is very important that the image on the monitor match the filmed image as much as possible, which means
that their dynamic ranges should be as similar as possible.
SECTION 2
WHEN TO PERFORM THIS PROCEDURE
The monitor is the final component in the video chain, and its contribution to the overal image quality is
often forgotten.
We recommend that you perform this procedure any time the customer is not completely satisfied with the
monitor image quality, when you believe a monitor adjustment has been changed, or any time the image on
the monitor does not correspond to the image on film (the hard-copy camera).
SECTION 3
TOOLS REQUIRED
Calibrated gray reference card (RG is often 18%) with known reflection ratio RG, p/n 2110360
6-91
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
SECTION 4
PROCEDURE
Note: In vascular or cardiac rooms, it is especially desirable to have a low level of lighting to get a good contrast
in the blacks and to allow a high dynamic range between blacks and highlights.
4. Position the luminnce meter at the same level as the user’s eyes, and at the luminance meter’s minimum
focalization distance.
6-92
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
4.2 Procedure
1. Position the gray reference card in the center of the monitor screen with the gray card towards the
luminance meter.
2. Measure the reflected light on the gray card center: LG.
3. Calculate the reference luminance: LR = LG / RG.
Note: 1. RG is the reflection ratio of the gray card; this card should be clean with no fingerprints.
Note: 2. The measurement can be repeated on the whole card. The ratio between the higher and lower values
should not exceed 2. This could happen if a spot light illuminates the monitor screen. If so, modify the
monitor position or the room lighting conditions.
4. Calculate the ambient illumination: IA : LR x π : IA is correct only if the gray card has a lambertian
response, therefore IA will be between 10 and 30 lux, and should not exceed 50 lux.
5. Record LR and IA on data sheet.
6. Remove the gray card from the monitor.
6-93
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
1. Display the SMPTE pattern on the monitor. Check that the DLX brightness and contrast window is
reset to the default value (depress the appropriate DLX keypad button).
2. Adjust the monitor brightness knob so that the 5% square (the smaller square inside the 0% square) is
just visible in the 0% square. Its level should be just above that of the 0% square.
3. In the DLX Browser screen, choose a patient, exam, and sequence image that does not have too much
contrast. There should be no completely black or white area.
4. Turn the brightness and contrast knobs, on the DLX keypad, fully counter-clockwise (CCW). You
should get a completely black image, with no white pixels. If you do not, return to step 3.
5. Measure the luninance, L0, of the black level with the luminance meter.
Illustration 6–6 shows the relation between the monitor luminance in cd/m2 and the ambient light in
lux. Adjust the L0 with the monitor brightness knob, to the recommended value, according to the
previously measured ambient light IA. For example, if IA = 30 lux, the L0 should be equal to 1.95
cd/m2.
6. Turn the brightness knob on the DLX keypad fully clockwise (CW). The image should be completely
white with no black pixels. If not, redo steps 3. and 4.
7. Measure the white level luminance (L0) with the luminance meter.
See Illustration 6–6 to find the recommended value for LMAX according to the previously measured
ambient light IA. Adjust the LMAX, using the monitor contrast knob, to this recommended value. For
example, if the IA = 30 lux, the LMAX should be equal to 190 cd/m2.
8. Repeat steps to until you obtain a L0 and LMAX equal to the recommended values. Record these on the
data sheet on the SCAT records (Test Data Results) sheet.
9. Record the L0 and LMAX on the SCAT records (Test Data Results) sheet.
Note: LO should be within LP x 1.1 and LP x 1.3 to maintain the maximum dynamic range.
Note: LMAX may be slightly different than the recommended value, but it should not be too much smaller to
avoid reducing the luminance dynamic range, and not too much higher so as not to “dazzle” the user’s
eyes.
Note: It is very important to carefully perform the adjustments. The image on the monitor should be as similar
as possible to the film image. While the dynamic range on the film is very big, optical densities are more
than 2.5 above base + fog, and this corresponds to a ratio LMAX /L0 of more than 300, it is very difficult
to achieve this goal without having the LMAX too high, causing the monitor to dazzle. The only way to
move towards this goal is to decrease the L0 , so as to decrease the ambient light to the minimum possible
level accepted by the user.
6-94
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
ILLUSTRATION 6–6
DISTAR 15” CONSOLE MONITOR
6-95
GE Medical Systems ADVANTX DLX CALIBRATION
REV 0 sm 2198945-100
Blank page
6-96