HANDS-ON MODULE 2
HANDS-ON MODULE 2
Setting up the FLOW-i (MX-0796)
Important: Refer to User´s Manual Software V3.0 Chapter 3 (System Overview) for more information
about components.
Locking
device
Insert
absorber
Locking
switch
1.2 MAINS
Plug in the FLOW-i to the mains outlet.
N 2O
3
4
AGS
Patient
tubing
SAMPLING LINE
1.7 VAPORIZER
Check that the vaporizer contains sufficient
agent. (If refilling is necessary, refer to
instructions in User’s Manual – Chapter 5)
Note: The agent level shall never fall below max max
Max. level
the minimum level or exceed the maximum
level. These minimum and maximum levels Level marker
are indicated on the vaporizer monitor tube.
min min
Min. level
1.9 RESUSCITATOR:
An independent means of ventilation, e.g. a
manual resuscitator, must be available
whenever the system is in use. Therefore,
check that there is one readily available and
working correctly.
SUCTION UNIT
AUXILIARY O2 FLOWMETER: