Operation Manual
1
APT8000 Operation Manual OPAPT8000.1.1.7ENG
Index
1. Summary - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1.1 Intruduction- - - - - - - - -- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1.2 Characteristic- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
2.1 Principle - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
2.3 Wiring - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
2.5 Display - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -- -- - - - - - - - - - - - - - - - - - - - - -- -- 5
3. Parameter setting - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6
2
APT8000 Operation Manual OPAPT8000.1.1.7ENG
1.Summary
1.1 Introduction
APT8000 series Smart pressure transmitter is a type of intelligent multi-functional digital instruament, it’s elaborate
designed on the basis of mature and reliable transmitter technology, combined with advanced Single-chip technology
and smart digital transmitter technology.
The centerpiece of the Smart pressure transmitter adopt 16 bit Single-chip whose powerful and high-speed
Calculating power guarantees the quality of the transmitter. The whole design focus on reliably, stability,
high-precision and intelligent.It also meets the requirements of industrial application.Hence,this software adopt digital
signal processing technology,has excellent anti-interference ability, zero point stability, Zero Sensitive Capability (ZSC)
and Temperature Sensitive Capability (TSC).
1.2 Characteristic
APT8000 not only has excellent anti-interference ability, zero point stability, Zero Sensitive Capability(ZSC)and
Temperature Sensitive Capability(TSC),but also has many other merits as follow
Optional HART modules, has HEAT communication capabilities.
4~20mA
T/T Current loop
CPU
HART
A/D
Display
‧Linear
Display
‧Range
Setting Keys
3
APT8000 Operation Manual OPAPT8000.1.1.7ENG
2.2 Dimensional Drawings
130mm
102mm 130mm
S Z
SET
50
0 PV 100%
0.000
KPA
160mm
2.3 Wiring
Z S 1250
1000
750
500
250
+ - + 0
Test
13 18 23 33 38
Power Supply (VDC)
+
mA
-
+ -
13~38VDC
4
APT8000 Operation Manual OPAPT8000.1.1.7ENG
2.4 The function of keypad
SET SET Enter in the Parameter setting, Save the Datas, Get to next step
MOVE
< Move the cursor to the right
1. Press and hold the button for about 2 seconds then release to enter the main menu settings.
If there is no response for 10 seconds in the window 0, it will return to the normal display automatically.
If there is no response for 2 minutes in the non-window 0, it will return to the normal display automatically.
2.Press and hold the SET
+ button for 5 seconds, release, then enter the menu setting window 6 for zero clearance.
Press botton and choose ”YES” to clear zero.
3. Press and hold + button for 5 seconds, release and enter the menu setting window 7 for zero shift and span shift.
Press and hold button for 5 seconds, release and then finish Zero Shift.
Press and hold button for 5 seconds, release and then finish span Shift.
2.5 Display
2.5.1 Normal Display
SET
50
0 PV 100%
0.000
KPA
SET
0.000
2
KPA
5
APT8000 Operation Manual OPAPT8000.1.1.7ENG
4. Parameter setting
4.1 Operate flow chart
50
0 PV 100%
0.000
KPA
SET
0 0 000
1
00002
1
SET
0.000
2 KPA
SET
0.0000
3 KPA
SET
120.00
4 KPA
SET
0.0000
5 S
SET
0.000
6 NO
SET
0.000
8 LIN
SET
0 0 000
0
SET
6
APT8000 Operation Manual OPAPT8000.1.1.7ENG
4.2 User operate description
50
0 PV 10 0%
Normal Display
0.000
KPA Press and hold for 2 seconds, then release to enter the main setting
0.000 rPress to enter unit selection, then press to activate the cursor and press
1 2 KPA again to select unit as following:
SET
InH2O,InHg,FtH2O,mmH2O,mmHg,PSI,Bar,mBar,g/cm2,Kg/cm2,Pa,Kpa,
Torr,ATM,Mpa,InH20@4˚C , MH20@4˚C,MH2O,mHg,M,Cm,mm,Special.
(The”Special” means it can input the unit which customers need.)
0.0000 SET
1 3 KPA Press to enter zero setting, and then press to activate the cursor .Press
SET
again to change the symbol to”-”. or press to move cursor and then press to
change the value. When the cursor move to the rightmost, continue to press toflash
all decimals, and press to change decimal position.
SET
Press or to confirm, and enter the next window Code4.
SET
again to change the symbol to”-”. Or press to move cursor and then press to
change the value.
SET
Press or to confirm, and enter the next window Code5.
00005
SET
again to change the symbol to”-”. Or press to move cursor and then press to
change the value.
SET
Press or to confirm, and enter the next window Code6.
00006
SET
0 PV
50
1 0 0% Normal Display
0.000 Press and hold + for 5 seconds, release and enter menu-setting window 7,
KPA the zero shift and full-span shift.
+
Zero Shift
mA
4.003 Press and hold for 5 seconds, and release to finish zero shift.
7 If press for 2 minutes without any responses, it will return to normal display automatically.
mA Span Shift
20.003 Press and hold for 5 seconds, and release to finish full-span shift.
7 If press for 2 minutes without any responses, it will return to normal display automatically.
0 PV
50
1 0 0% Normal Display
0.000
KPA
8
APT8000 Operation Manual OPAPT8000.1.1.7ENG
4.4 How to set Decimal point and minus
Example: Set zero to “- 15Kpa”
50
0 10 0 %
PV
Normal Display
0.000 Press and hold for 2 second, release and enter main menu setting.
KPA
9
APT8000 Operation Manual OPAPT8000.1.1.7ENG
4.5 Variable displayed setting
LCD can be set to display "Press valve", "mA valve", "%", displaying one(fixed) or two of these three kinds(alternate).
When the first variable and the second variabel displayed are the same, the display will be fixed.
When the first variable and the second variable displayed are different, the display will be alternate (the interval is 4 sec.).
In the normal display, to press and hold button can change the variable. (Lower-left display function code is 30)
50 50 50
0 PV 10 0% 0 PV 10 0% 0 PV 10 0%
mA
0.000 0.000 % 0.000
30 KPA 30 30
The different variable means alternate display: Example:” Press valve”,”mA valve” for alternate display.
(the first variable: “Press valve”, and the second variable:”mA valve”)
50 50
0 PV 10 0% 0 PV 10 0%
mA
0.000 0.000
KPA
The same variable is fixed display: Example: display “Press valve” fixed,
(the first and the second variable are both “Press valve”)
50 50
0 PV 10 0% 0 PV 10 0 %
0.000 0.000
KPA KPA
E.g1. When the variable “Press valve” ,“mA valve” display alternately, it should change to the fixed display“Press valve”.
So, the variable “mA valve” should be changed to “Press valve”.
It means the first variable and the second one are the same.
Operating Steps: when the screen displays “mA valve”, press and hold , waiting until the screen displays “Press valve”. This
step finishes the change “mA valve” to “Press valve”. At this time, the screen displays “Press valve” which is fixed.
E.g2. When the variable “Press valve” display fixed, it should change “Press valve” and “mA valve” to display alternately.
So, the variable “Press valve” should be changed to “mA valve”. It means the first variable and the second one are
different.
Operating Steps: when the screen displays “Press valve”, press and hold , waiting until the screen displays “mA valve”. This
step finishes the change “Press valve” to “mA valve”. At this time, the screen displays “Press valve” and “mA valve” which are alternate.
10