OPERATION MANUAL
UNICO® has been providing quality laboratory equipment for the medical labs for 12
years. We are constantly working to improve our products with the latest technological
innovations. UNICO® PowerSpin™ C8624 24 place centrifuge is released to meet the
needs and offer safety features needed in the fast growing medical and health care
industries.
Construction:
The enclosure of the PowerSpin C8624 24 place centrifuges is made of high impact
plastic. The internal rotor chamber is made of stainless steel. A safety device is installed to
prevent the lid from opening while the centrifuge is running. The motor is mounted on
rubber mounts to provide quiet running and less vibration. The Stainless Steel rotor
chamber can be easily cleaned and disinfected.
Accessories:
Each Power Spin C8624 centrifuge comes equipped with angled rotor with 24-place tube
shields already installed.
2
Unpacking and Installation:
The centrifuge is first packed in Styrofoam and then put in a carton box for shipping.
Please examine the unit upon receiving and retain all packing until the unit has been
tested. The warranty excludes any damage caused by shipping.
Inspect for any possible shipping damage. If you note any damage, please contact your
sales representative. Save the original packaging. If it becomes necessary to return the
instrument to us, we ask that it be returned in the original packaging.
Unpack the shipping carton carefully and verify that the shipment is complete. It should
include the following items:
No on-site installation is required. Follow simple procedures below to set the centrifuge
up:
Note: The rotor is protected by cushion foam in the rotor chamber.
Do not operate the centrifuge before the foam is removed!!
1. Take the centrifuge from the Styrofoam and place it on a sturdy flat surface.
Do not turn on the centrifuge!
2. Use a small screw driver (or a rod with Ø4.5mm or less in diameter) to open the
lid. Insert the tool into the emergency lid-opening hole on the up-right corner on
the left side (as showed below). Push the tool to release the safety lock and open
the lid.
Special packing cushion foam is packed inside the chamber of the centrifuge to
prevent the rotor and shields from moving during shipping.
Top lid
Emergency
Lid-opening Front
Left Side
Caution: Make sure the voltage of the power source matches the voltage input of
centrifuge.
3
Control Panel and Key Pad Controls
START
ddd
Spd: xxxx Time: xxxxx
OPEN
STOP
▲ ▼ ▲ ▼
Speed Timer
Timer setting: Use the up or down arrow to preset the running time. The timer range is
from 1 minutes to 60 minutes. The display will show counting down time.
Speed setting: Use the up or down arrow to preset the running speed. The speed range is
from 1,000rpm to 3,400rpm in 100rpm increments. The display will show
actual running speed and RCF (g) level.
Start Button: Press Start button to spin sample. The centrifuge will not spin if the lid is
not closed or not properly closed.
4
Stop Button: Stop the centrifuge at anytime.
Open Button: Open the centrifuge lid only when the centrifuge is not spinning.
Emergency Lid-opening:
The lid is designed to prevent it from being opened accidentally while the
centrifuge is spinning. There may be situation when the lid needs to be opened
manually (For example if you lose electrical power and samples are still in the
centrifuge).
To open the lid manually, use the lid-opening tool (or a small screw driver). Insert
the tool into the emergency lid-opening hole on the up-right corner on the left side
(as showed in Installation Section). Push the tool in to release the safety lock and
open the lid.
The centrifuge will count the total cycles of usage. To check the cycle usage, press
and hold Stop Button for 4 seconds, the total to date number of cycles will be
displayed.
Note: Cycles of usage is not total spinning time. It represents the total times the
centrifuge has spun.
Operation:
1. Plug the centrifuge into a power source outlet. Make sure the voltage is correct.
2. Open the lid by pressing the Open button on the control panel.
3. Place the specimen tubes into the tube shields of the horizontal rotor.
CAUTION!!
Make sure that the specimen tubes are balanced.
Always balance rotor with tubes of equal size and equal fill.
(See safety section below)
Note: When the centrifuge run is complete, the timer will be automatically reset to
the last timer setting and the centrifuge is ready for the next run.
5
Safety:
Your PowerSpin C8624 centrifuge is a carefully designed and built instrument. When
operated properly, it presents no safety hazards, however, please take note of these safety
precautions.
NEVER ATTEMPT TO OPEN THE LID UNTIL THE ROTOR HAS COME TO
COMPLETE STOP!
Load Balance:
Always make sure that the load is balanced before operating the instrument.
It is very important that tubes with equal loads are placed opposite each other in the rotor.
Also, when spinning a single specimen tube, another specimen tube of the same size filled
with water must be installed directly across from the specimen tube to maintain balance. If
the centrifuge is not properly balanced, damage could occur. If excessive noise or
vibration is noticed during a run, stop the centrifuge and recheck the balance.
Maintenance:
Calibration:
It is recommended that your centrifuge’s RPM be calibrated at least every 6 months. This
period is suggested by both the FDA and C.L.I.A. guidelines. The speed should fall within
±100rpm of the speed display. Since the speed displayed on Unico C8624 is from a real
time digital sensor mounted on the motor shaft, it reflects the real running speed of the
rotor. The timer is an electronic timer and no calibration is required.
Biohazard:
If a tube breaks, a biological hazard may exist. Unplug the instrument. Clean using
approved laboratory procedures. Always wear the proper protective clothing and gloves.
Use a suitable disinfectant and dispose of the sample properly. Refer to the guidelines set
forth by OSHA, CLIA, etc.
6
Service:
If you feel that your centrifuge needs factory service, please call us and speak to a
customer service representative to determine the nature of the problem. Simple problems
may be resolvable over the telephone. If the instrument needs to be returned to Unico, a
R.G.A. (return goods authorization) number will be issued. Please reference the R.G.A.
number on the paperwork packed with the instrument, and on the outside of the carton.
Problem Solving:
d) Check that Power switch in the back of unit is in the “on” position.
e) Check that the power indicator light (with the switch) is lit. A failure of the
power light to illuminate could indicate a need for a new fuse when the unit is
properly plugged in, switch is turned on and the lid is closed.
b) Fuse is blown.
4. Fuse Replacement:
b) Remove the power cord from the power inlet on the back of the centrifuge.
c) Locate the fuse holder. The fuse holder is a part of the power inlet.
7
|
Fuse Holder
d) Use a Philip type screwdriver to take the snap-in type holder cover off.
Remove the blown fuse and replace it with the same type and rating fuse
which is:
250V 4A
Specifications:
8
Warranty:
This instrument is warranted against defects in materials and workmanship for a period of
one (1) year from the date of shipment to the customer. This warranty is limited to the
repair and replacement of parts, which prove to be defective during the warranty period.
This warranty is not valid for parts damaged, lost, or which fail because of accident, fire,
theft, acts of nature, negligence, or the use of chemicals, which have a deleterious effect.
This warranty is conditioned upon the manufacturer retaining the option of replacing parts
up to and including the entire instrument.
Note: This warranty will not extend to any repairs or modifications made to the instrument
by some party other than the manufacturer or a party authorized by Unico. Also, this
warranty shall be in effect only upon the notice of the defect to the manufacturer or it’s
authorized distributor, or dealer, within five (5) days after the occurrence of said defect
No other warranty of any kind is made expresses or implied. The warranty described
above shall be the sole and exclusive remedy available to the purchaser. Correction of
defects, in the manner and for the period of time described above, shall constitute
complete fulfillment of all liabilities and responsibilities of the manufacturer to the
purchaser with respect to the product, and shall constitute full satisfaction of all claims,
whether based on contract, negligence, strict liability or otherwise. In no event shall the
manufacturer be liable, or in any way responsible for any damages or defects in the
product that were caused by repairs or attempted repairs performed by anyone other than
the manufacturer’s dealer or station, nor shall the manufacturer be liable or in any way
responsible for any incidental or consequential economic or property damages. Some
states do not allow the exclusion of incidental or consequential damage, so the above
exclusion may not apply to you.
9
PowerSpin MX C8606, C8624 and M24 Trouble Shooting
Problem Cause Solution
1 Error 1 Initialization failure: Main control board
related. 1. Check all connectors.
(Initialization failure) 1. Loose connection(s) 2. Replace TR1 transistor
2. Defective transistor (TR1) (p/n on main board
BATA06-600B) 3. Replace U6 chip on
3. Defective chip (U6) (p/n MOC3022) main board
4. Main board damaged 4. Replace the main
board
2 Error2 Lid unable to open
1. Loose connection from lock to 1. Secure the connection
(Lid unable to open) main board (J4 on board)
2. Lock-driven motor disengaged 2. Adjust or replace the
or defective motor
3. Lock pin on lid “too long” 3. Adjust the lid pin
(Use emergency opening
on left side)
3 Centrifuge is unable to The transportation protection locking Remove the two locking screws
sit on flat table surface screws not removed located underneath the
centrifuge base.
4 No power to 1. Power cord is not plugged in 1. Plug in the power cord
centrifuge 2. On/Off switch is at Off on the back.
3. Fuse blew 2. Turn on the switch
3. Replace the fuse
5 Touch button no 1. Loose connection or connection 1. Exercise the connector
response (no response reversed. (J2 on main).
sound) 2. Touch panel defective 2. Replace touch panel
6 Display show 1. Connector J6 loose 1. Secure the connector
“Running” but motor 2. Transistor(TR1) failed 2. Replace the transistor
is not running. (BTA06-600B) 3. Replace the chip
3. U6 chip failed (MOC3022) 4. Replace the relay
4. Relay (K3) failed (p/n G2R-2-
12VDC)
7 Running normal but 1. Loose connection (J1) between 1. Secure the connector
no RPM /RCF display main and speed sensor board 2. Replace the sensor
2. Speed sensor (opt-coupler) board
failed 3. Replace U1 chip on
3. Chip(U1 on main) failed (p/n main board
LM339)
8 Lid open but display 1. Loose connection: lock sensor 1. Secure the connection
show READY board to main board (J7) 2. Replace the opt-
2. Lock sensor opt-coupler coupler (older version)
failed(MOCH25A)
9 Sample separation 1. Running speed/RCF is not 1. Increase the RPM
failed enough 2. Increase total running
2. Running time is not enough time
3. Rotor is not secured to the 3. Engage the rotor to
motor shaft motor shaft.
10 Centrifuge shaking 1. Load is not balanced. 1. Balance the load.
2. Short tube adapters not evenly 2. Even the short tube
placed adapters.
3. Motor cushion failed 3. Replace motor
cushions
10
PowerSpin C8624 Centrifuge
Parts and Nomenclature
Safety
Lid lock
Rotor
On/off
Emergency switch
lid-opening (on back
(left side) side)
Control
panel
Power
indicator
Model C8624
11