Rosemount Pressure Transmitter

Total Probable Error


Model Family: 3051T
Tag No: PT-200C
Model Number: 3051TG2A2B21AWR3J1M5DOC1Q4
Measurement Type: Gage Pressure

Calibrated Span: 1 to 3 psig


Ambient Temperature: 20 to 35 C
Static Line Pressure: 1 to 3 psig

Reference Accuracy: 0.011 psig


Line Pressure Effect: 0.000 psig
Ambient Temperature Effect: 0.030 psig

Total Probable Error: 0.032 psig


1.588 % of Span

Stability for One Year: 0.037 psig


1.875 % of Span
This report is provided according to the terms and conditions of the Instrument Toolkit(TM) End-Use Customer License Agreement.

Version: 3.0 (Build217D) Printed: 8/4/2020