Professional Documents
Culture Documents
Meeting 2018
Workshops, presentations, business talk and more
No surprise - AZURA with the same performance as Ultimate 3000
No surprise
AZURA on par with UltiMate 3000 RS
AZURA vs. UltiMate
AZURA UHPLC System (998502200)
AZ U R A Au t o s a m p l e r AS 6 . 1 L 56.500* €
1000 bar 100 Hz
AZ U R A C o l u m n t h e r m o s t a t C T 2 . 1
AZ U R A D e t e c t o r D AD 6 . 1 L
AZ U R A P u m p P 6 . 1 L
1000 nm LPG
* W i t h o u t S o f t wa r e & P C
UltiMate™ 3000 RS LC System
W P S - 3 0 0 0 R S Au t o s a m p l e r 60.500* €
1000 bar 200 Hz
TCC-3000RS Column thermostat
D AD - 3 0 0 0 R S D e t e c t o r
LPG-3400RS Pump
800 nm LPG
S e m i a n a l yt i c a l f l o w c e l l 5 µ L
* W i t h o u t S o f t wa r e & P C
Technical comparison
DWELL VOLUME
550 µL 1100 µL
The difference in the dwell volumes results from the larger mixing chamber of the
thermal system, the 100 μL capillary on the autosampler valve
and the additional pre-column cooling.
DISPERSION VOLUME
14.5 µL 14.5 µL
15 µAU 17 µAU
< 0.1
mAU/h
CARRY OVER
Limit
0.01%
DMA
LOD
LOQ
Method parameter
1,8
Gradient Buffer ACN
mL/min
ESII 150 mm ID
5 µm
C18P length 4,6 mm
Injection
220 nm 23 °C 1 ppm
10 µL
Calibration
DMA mg/ mL
Level 1 0,010
Level 2 0,025
Level 3 0,030 R-Square or
0.9999 Correlation 0.9999
Level 4 0,050 Coeffecient
Level 5 0,100
Reproducibility
Peak
area
Peak
height Relative
Standard
< 1.0 % < 1.0 %
Deviation
Retention (RSD)
time
Chromatogram
1 ppm
0.001
mg/mL
LOD & LOQ
Signal Noise
Flow cell LOD [ppm] LOQ [ppm]
[mAU] [mAU]
UltiMate 7 mm, 5 µl 1.710 0.057 0.100 0.332
AZURA 10 mm, 10 µl 2.502 0.083 0.100 0.333
International Distributors
Meeting 2018
Workshops, presentations, business talk and more