Supplied parts
1 2 3
1 Spring
2 Snap ring
3 Vial detector
Required tools
- Allen key 3 mm
- Small slotted screw driver
- Torx T10 screw driver
- Torx T20 screw driver
6. Push the vial detector in and hold it in that position, see (1) in fig. 3.
7. Remove the snap ring with a small slotted screwdriver and pull the vial detector out of its guiding, see (2) in
fig. 3.
1 2
8. Check the slide bearings. If the bearings moved a bit out of the drilling, cut off the ends with a utility knife.
The two slide bearings have to be flush with the surface.
NOTICE
• During reassembling, especially when fixing the screws, move the vial detector in
and out to prevent blockage.
• Do not apply any kind of lubricant.