You are on page 1of 24

CR-IR 391CL V1.0 / CR-IR 391V CL V1.

0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA Console
b. Specification and Function Guideline

1. Outline
This document explains the background and the concept that help understanding the specification and
the functions of “CR-IR 391CL” and “CR-IR 391V CL”. We hope this will be a help to understand them.
For accurate values or detailed operability, see such documents as the product specification or the
operation manual.

NOTE In this document, if not otherwise specified, "CR-IR 391CL" and "CR-IR 391V CL" are
called "PRIMA CL" as a whole.

Contents

1. Outline....................................................................................................................................... 1
2. Specification .............................................................................................................................. 2
2.1 Concept...................................................................................................................................2
2.1.1 Reader Unit ......................................................................................................................... 2
2.1.2 Image Processing................................................................................................................. 3
2.1.3 Difference from CR Console................................................................................................ 3
2.1.4 Difference from FCRView / FCR CAPSULA V VIEW....................................................... 3
2.1.5 Spec ...................................................................................................................................... 4
2.2 Serviceability..........................................................................................................................7
3. Architecture............................................................................................................................... 8
3.1 Hardware architecture ..........................................................................................................8
3.1.1 Restrictions.......................................................................................................................... 8
3.2 Composition of Software ........................................................................................................9
3.3 Structure of FCR PRIMA System .......................................................................................10
3.4 Software transition ..............................................................................................................11
3.5 Configuration items .............................................................................................................13
3.5.1 Standard items .................................................................................................................. 13
3.5.2 Optional items ................................................................................................................... 13
4. Function................................................................................................................................... 14
4.1 The list of functions .............................................................................................................14

1/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
2. Specification
2.1 Concept
“PRIMA CL” is all-in-one equipment that can perform study registration, image input, viewer function,
output to printer and storage to DVD.

2.1.1 Reader Unit


Following Reader Unit are available.

CR-IR 391CL CR-IR 391RU


CR-IR 391V CL CR-IR 391RU

Reading of 20 [pixel/mm] is not supported but 10 [pixel/mm] is supported.


HR-V type Imaging Plate is not available.
FR mode is not supported.

The following indicates times required for processing an image (14"x14in", “35x35cm”).

Time to display image Through put


CR-IR 391RU 46 [sec] 180 [sec] (20 [IPs/h])

Time to display image

Through put

2/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
2.1.2 Image Processing
CR-IR 391CL includes exposure menus for Human.
CR-IR 391V CL includes exposure menus for Veterinary.

“PRIMA CL” ’s menus and CR Console’s menus are not compatible. So “PRIMA CL” cannot import CR
Console’s menus.

Following image processing are available.

GP (Gradation Processing)
RP (Response Processing)
DRC (Dynamic Range Control Processing)
TAS (Tomographic Artifact Suppression Processing)
MFP* (Multi-Objective Frequency Processing)
GPR (Grid Pattern Removal processing)
FNC (Flexible Noise Control).
* MFP processing is option

About other function of Viewer screen such as window pattern and comparing past study, refer “5.1 The
list of functions”.

2.1.3 Difference from CR Console


The major difference from CR Console is that “PRIMA CL” equips viewer function, detailed QA,
measurement function and DVD storage as standard function.
DVD storage function can backup image data to media, and you can manage which image is stored in
which DVD. Viewer function can explain medical condition to a patient clearly by using such function as
image magnifying and comparison to past image.
Moreover, User control function and security function have been enhanced. For instance, it can prevent
an erroneous operation by an unauthorized user by setting user authority and password. In addition,
intended data falsification can be avoided by access record of a study and consistency check of a disk.

For detail comparison with CR Console’s function, see “5.1 The list of functions” of this document.

2.1.4 Difference from FCRView / FCR CAPSULA V VIEW


The major difference from FCRView / FCR CAPSULA V VIEW is that “PRIMA CL” does not equip
Client PC (option), DICOM US Storage (option), connecting high-resolution monitor, dual monitor
configuration, connectable models of RU and connecting with CR Console.

3/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
2.1.5 Spec
“PRIMA CL” has 250GB size of internal HDD. 5GB of it is used for Windows system and 5GB for
margin. So 240GB is available for image storage area. The maximum number of image storage is
about 15,000 for 14"x17" size. The controllable number of study is 60,000 at the maximum. When the
volume of image files exceeds the capacity of the HDD, images are deleted with FIFO (First in First
out) method. Deleted images can be read and displayed from DVD.

Using “Free Layout Print” option, the size of image storage area will be reduced to 235GB.

HDD

Deleted old
images by
Database FIFO method.
Older studies
(No images on HDD)
60,000
Newer studies
studies (Images are on HDD)
About 15,000
images can
be stored in
the HDD

Those images that are delivered to the Storage disk can be retrieved and displayed on the monitor or
re-printed to the film after deleted by FIFO method.

After images are deleted by FIFO method, images delivered to the Storage disks can be
retrieved from disks. Images delivered to the Work disks can not be retrieved from disks.

HDD

Database Deleted old


images by
Older studies FIFO method.
(No images on HDD)
60,000
Newer studies
studies (Images are on HDD)
Storage disk

Work disk

4/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17

When the image size is 14"x14" for almost data, one DVD can stores 200 images (HQ &
Uncompressed), or 1300 images (ST & JPEG Lossless)
For detail see the product specification.
Accordingly, it can be said that:

About one DVD is necessary for a year under the following conditions.
5 shots per day
ST & JPEG Lossless storage method:

About eight DVDs are necessary for a year under the following conditions.
10 shots per day
HQ & JPEG Lossless storage method

In case of taking backup, the necessary number of DVD is doubled.

3 types of image storing are supported.

Type Function Media


Storage disk Storing images to the DVD. The destination of images DVD+RW, DVD-RW
( to the DVD ) are recorded in DB.
Work disk Storing images to the DVD. The destination of images DVD+RW, DVD-RW
(to the DVD) are not recorded in DB.
PDI Storing images to the CD or DVD with PDI viewer DVD-R, CD-R
software.

There is an operation manual of PDI Viewer for patient at “Appendix B PDI Disk Image Display” in
Reference Guide and this manual is designed to provide a copy to the patient with PDI CD-R or
DVD-R.

5/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
“Appendix B PDI Disk Image Display” in Reference Guide

6/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
2.2 Serviceability
On the service utility, Simplified setup wizard, which is used on installation, displays necessary settings
orderly according to options. This function will make the risk of erroneous operation lesser.

On the installer (activated from the install CD), an operation which has been done is displayed in green
and one which can be done next is displayed in red. The installer is used to change display language
and others.

7/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
3. Architecture
3.1 Hardware architecture
z Connection to one Reader Unit (1-to-1 connection to FCR PRIMA )
z Connection to two printers* (DRYPIX PRIMA and 7000 are supported.)
* Only one unit is available for automatic transmission of images.
z Connection to two storage devices*
* Only one unit is available for automatic transmission of images.
z 100Base-TX or more for network

DICOM CR STORAGE

FRUP DICOM PRINT or Free Layout Print


LAN

FCR PRIMA “FCR PRIMA DRYPIX DRYPIX Server / Viewer


Console” PRIMA PRIMA

FCR PRIMA and DRYPIX PRIMA have no LCD panel to display or edit configurations on them. It is
necessary to edit configurations (such as IP address) of FCR PRIMA and DRYPIX PRIMA by using
PC-TOOL or PC Utility working on the Service PC or FCR PRIMA Console. These service operations
are similar to that of XG-1 RU.

3.1.1 Restrictions
Only 1-to-1 connection is supported for connection to Reader Unit. Accordingly, in case of registration
or image reading, it is necessary to display the exposure menu (unlike N-to-N connection, image
reading is impossible with the background process, and the barcode reader cannot be used).
Accordingly, it is desirable that “PRIMA CL” should be installed beside Reader Unit. Direct connection
between FCR PRIMA and FCR PRIMA Console with cross LAN cable is not supported.

8/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
3.2 Composition of Software
Software consists of the server application and the client application and a number of utilities. Server
application receives / stores / transfers images and create web page of Client application and also
switches utilities. The client application runs on the web browser (Internet Explorer) and display GUI of
FCR PRIMA Console.

Indicator Client Application Client Utility FFUdfFmtTool

Routine Operation Disk Formatting

HTTP
SOAP

User Utility Backup Utility Disk Copy Tool


Server Application

Background Processing User Maintenance

Service Utility Application Installer

Service Operation

Applications of FCR PRIMA Console

9/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
3.3 Structure of FCR PRIMA System
Followings are software working on the FCR PRIMA Console PC and their relations.

FCR PRIMA Console


Shut down
Error code
Windows Vista

Internet Explorer 7.0 Internet Explorer 7.0


Configuration

Client Application PC Utility

DRYPIX PRIMA
ART PC

PC TOOL Magic packet


HTTP / SOAP
Logs
Wake on LAN

DICOM (TCP/IP)
IIS

Web Application Web Service


Magic packet
FTP Service

Configuration

Server Application

Server Application
FCR PRIMA

Win32 API, Jet, SQL, other Logs

FRUP (TCP/IP)

DICOM (TCP/IP)
System Master Image
Config DBs DB

Viewer / PACS

* ART PC is software installed by DRYPIX PRIMA Application Software CD.

* PC Utility is a web based maintenance tool of imager and it works by accessing web service on the DRYPIX PRIMA

* PC TOOL is a maintenance tool of reader unit and is installed by FCR PRIMA Application Software CD.

* Wake on LAN is software which sends magic packet to the network devices and start-up in synchronization.

10/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
3.4 Software transition
During routine operation, both the server application and the client application are activating on one
PC.
Although the server application does not have GUI except for the initial screen and icons displayed on
the taskbar, it serves as a core of various kinds of process of “PRIMA CL”. In addition, Service Utility
opens by clicking the upper left and upper right corners of the initial screen, and the startup menu for
User Utility, Backup Utility and Disk Copy Tool opens by right-clicking the server application icon on the
taskbar.
Server application and Client application start automatically (by the default). To start them manually,
click desktop icon. “FCR” is an icon to start Server application and “FCRWeb” is an icon to start Client
application.

11/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
Server application 2009.07.17
has already started
Server PC Start Wake on LAN

fieldengineer

Server application start


(Automatically)

Service Utility
Shut down
Client application start
Client Utility (Automatically)

Indicator
fieldengineer

Client application
Close

system

PC Lock
FFUdfFmtTool

Server PC shut down

Server application shut


down

Disk Copy Tool


Backup Utility
User Utility

12/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
3.5 Configuration items
3.5.1 Standard items
Standard items below are available.
Product Code Explanation
IR 391 CL A1-V1 E (for Human) Model: DELL OptiPlex 760 DT
or OS: Windows Vista(R) Business SP1 32bit
IR 391 CL VE A1-V1 E (for Veterinary) CPU: Pentium Core2 Duo E7400 (2.80GHz)
Memory: 2GB (1GB x 2)
HDD: Serial ATA 250GB
Drive: FDD and Super Multi Drive
Preinstalled Software:
“PRIMA CL” Application Software
“PRIMA CL” Application Key Software
Standard Kit EXP (Including FNC, GPR, HQ keys)
Standard Keys (Including FRUP, TAS, DRC keys)
CR-IR 391V Image Processing Software
etc.,
Manuals:
Operation Manual (paper)
Quick Guide (paper)
Software License Agreement (paper)
Reference Guide (CD-R)
etc.,
“CR-IR 391V Image Processing Software” is accessory of “CR-IR 391V CL”

3.5.2 Optional items


Optional items below are available.
Product Code Explanation
IR 391 LCD-17 E LCD (DELL E178FP)
1280x1024[pix], 17[inch] TFT color
IR 391 LCD-17H E LCD (EIZO FlexScan S1721-SHBK)
1280x1024[pix], 17[inch] TFT color
IR 391 MFP E Option key software of Multi-Objective Frequency Processing
IR 391 DICOM PRINT E Option key software of DICOM Print
IR 391 F-LAYOUT PRINT E Option key software of Free Layout Print including DICOM
Print key
IR 391 DICOM STORAGE E Option key software of DICOM Standard CR Storage

Only one monitor can be used as the specification.

13/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
4. Function
4.1 The list of functions
This list aims to compare functions of CR Console V7.0, FCRView V5.0 and FCR PRIMA Console.

FCR PRIMA
CR Console FCRView
Console

INPUT Patient Information Manually ○ ○ ○


Patient DB ○ ○ ○

Import via .csv file ○ ○ ○

DICOM MWM PI ○ × ×

BCR ○ × ×

MCR ○ × ×

Modality Work List DICOM MWM Order ○ × ×

Image Reading 1:1 Connection ○ ○ ○

N:N Connection ○ × ×

Study Execution (1-image display) ○ ○ ○

Study Execution (6-image display) ○ × ×

ST & HQ ○ ○ ○

SH ○ × ×

<CR-VW 674> <CR-IR 391CL>

CR-IR 355RU CR-IR 391RU

CR-IR 356
FCR5000Plus
CR-IR 357
or later
Reader Unit CR-IR 359
(Except CR-IR
CR-IR 391RU
391RU)
<CR-IR 355V CL> <CR-IR 391V CL>

CR-IR 355RU CR-IR 391RU

CR-IR 356

FR (fast reading) ○ ○ ×

Energy Subtraction ○ × ×

Cassette for panoramic tomography ○ ○ ○

Long view image ○ × ×

Long view image via CR Console - ○ ×

HR-V type Imaging Plate ○ × ×

DICOM US Storage × ○ ○

General purpose image × ○ ○

X-CON ○ × ×

DAP ○ × ×

ID Online ○ × ×

PROCESSING EDR ○ ○ ○

S/L ○ ○ ○

14/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

GP/RP ○ ○ ○
DRC ○ ○ ○

MFP ○ ○ ○

PEM ○ × ×

FNC ○ ○ ○

GPR ○ ○ ○

Rotation / Inversion ○ ○ ○

Trimming ○ × ×

Image Alignment (Mammo) ○ × ×

Electronic Shutter ○ × ×

Stitching ○ × ×

LIST Query Create Query × ○ ○

Query By Patient × ○ ○

Query By Study × ○ ○

Query In Media ○ ○ ○

Filtering ○ ○ ○

VIEWER Comparing to past study × ○ ○

Types of image to be CR (ST) ○ ○ ○

CR (HQ) ○ × ×

US × ○ ×

○ ○

General purpose image × (images input via (images input via

General-purpose General-purpose

Study Window Pattern Study 1x1 ○ ○ ○

Study 2x1 (display 2 studies) × ○ ○

Image Window Pattern 1x1 ○ ○ ○

2x1 (display 2 images) ○ ○ ○

1x2 (display 2 images) × ○ ○

2x2 (display 4 images) × ○ ○

2x3 (display 6 images) × ○ ○

2x4 (display 8 images) × ○ ○

2x6 (display 12 images) × ○ ○

15/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

3x4 (display 12 images) × ○ ○

4x1 (display 4 images) × ○ ○

4x2 (display 8 images) ○ ○ ○

4x3 (display 12 images) × ○ ○

4x4 (display 16 images) × ○ ○

4x6 (display 24 images) × ○ ○

6x4 (display 24 images) × ○ ○

Measurement Length × ○ ○

Angle × × ×

CTR (Cardiothoracic Ratio) × ○ ○

Annotation Arrow × ○ ○

Ellipse × ○ ○

Rectangle × ○ ○

Polygon × ○ ○

Text ○ ○ ○

Marker Input manually ○ ○ ○

Input automatically ○ × ×

Display Magnify/Reduce ○ ○ ○

Auto fit to Window ○ ○ ○

Life size (100%) magnification × ○ ○

Pixel scale
○ ○ ○
(1 picture element represents 1 pixel of
Stack display × ○ ○

Applying modality LUT × ○ ○

Applying preset image processing parameter × ○ ○

Free Drawing × ○ ○

OUTPUT DICOM Print (FILM) Selectable Film Format ○ ○ ○

Special Print ○ × ×

Flee Layout Print × ○ ○

Output automatically ○ ○ ○

Output manually ○ ○ ○

16/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

FILM Format Single ○ ○ ○

Right and Left Image Processing ○ × ×

2on1 ○ ○ ○

4on1 ○ ○ ○

12on1 ○ × ×

General-purpose image Output automatically × ○ ×

Output manually ○ ○ ○

JPEG ○ ○ ○

BMP × ○ ○

DCR ○ ○ ○

DVD Storage Output automatically × ○ ○

Output manually × ○ ○

Output image with annotation × ○ ○

Output image with marker × ○ ○

Unprocessed CR Image Storage × ○ ○

Processed CR Image Storage × × ×

CR Image Storage (HQ/ST) × ○ ○

CR Image Storage (SH) × × ×

MG Image Storage × × ×

US image (8bit or 24bit JPEG Lossless × ○ ○

General-purpose image (24bit JPEG × ○ ○

DVD Storage Output automatically ○ × ×

Output manually ○ ○ ○

Output image with annotation ○ ○ ○

Output image with marker ○ ○ ○

Unprocessed CR Image Storage ○ ○ ○

Processed CR Image Storage ○ ○ ○

CR Image Storage (HQ/ST) ○ ○ ○

CR Image Storage (SH) ○ × ×

MG Image Storage ○ × ×

17/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

US image (8bit or 24bit JPEG Lossless × ○ ○

General-purpose image (24bit JPEG × ○ ○

PDI CD-R ○ ○ ○

DVD-R × ○ ○

Output image with annotation ○ ○ ○

Output image with marker ○ ○ ○

Paper Print Output automatically × × ×

Output manually × × ×

DICOM Storage CR Storage (10bit) ○ ○ ○

CR Storage (12bit) ○ × ×

Private CR Storage ○ × ×

Private Unstandardized CR Storage ○ × ×

LUT Processed ○ ○ ○

TagReplace ○ ○ ○

TagLookup ○ × ×

Storage Commitment Push ○ × ×

MG Storage ○ × ×

FINP ○ × ×

Local Print ○ × ×

LINKAGE Clustering ○ × ×

Server / Client × ○ ×

IP ID Sharing ○ × ×

DICOM MPPS ○ × ×

Synapse Web Query ○ × ×

DICOM Q/R ○ × ×

Pocket id Console ○ × ×

UTILITY Reconstructing Image DB by DVD ○ ○ ○

DVD Consistency Check × ○ ○

DVD Copy Tool × ○ ○

Retake Analysis ○ × ×

18/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

Reader QC ○ × ×

Exposure Log ○ × ×

Access Log Login / logout history × ○ ○

Display / editing study history × ○ ○

Transmission history of study × ○ ○

Backup / Restore history × ○ ○

Supervise User ○ ○ ○

Supervise User's Authority ○ ○ ○

Login account for field engineer × ○ ○

Service Utility login password ○ ○ ○

Config Backup/ Restore ○ ○ ○

Backup notification × ○ ○

Maintenance log acquisition ○ ○ ○

Deleting Personal Info. from Log ○ ○ ○

Create and edit Exposure/Study menus ○ ○ ○

Parameter bat updating ○ × ×

Auto Log Off ○ ○ ○

OTHER Touch Panel Screen ○ × ×

Wake on LAN ○ ○ ○

Veterinary use × ○ ○

Screen Resolution 1024x768 Landscape (color) ○ × ×

1024x768 Portrait (color) × × ×

1280x1024 Landscape (color) ○ ○ ○

1280x1024 Portrait (color) × ○ ○

1600x1200 Landscape (monochrome) ○ × ×

1600x1200 Portrait (monochrome) ○ × ×

1600x1200 Landscape (color) × ○ ×

1600x1200 Portrait (color) × ○ ×

2048x1536 Landscape (monochrome) ○ × ×

2048x1536 Portrait (monochrome) × × ×

19/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
FCR PRIMA
CR Console FCRView
Console

2048x1536 Landscape (color) ○ × ×

2048x1536Portrait (color) × × ×

Dual monitor × ○ ×

Language Japanese ○ ○ ○

English ○ ○ ○

German ○ ○ ○

French ○ ○ ○

Italian ○ ○ ○

Swedish ○ ○ ○

Spanish ○ ○ ○

Finnish ○ ○ ○

Danish ○ ○ ○

Norwegian ○ ○ ○

Portuguese ○ ○ ○

Czech ○ × ×

Polish ○ ○ ○

Hungarian ○ × ×

Korean ○ ○ ○

Chinese Simplified ○ ○ ○

Chinese Traditional ○ ○ ○

Russian ○ ○ ○

Slovenia × × ×

20/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
4.2 Free Layout Print
4.2.1 Outline
This function is to print out multiple images with an arbitrary positioning/display size. It is also possible to lay
out images of different patients on one film.
Available film format: 1x1 to 10x10

4.2.2 Operation procedure


It can be printed out from the study list screen or the viewer screen.
On study list screen: Select a study and click “Free Layout Print”

On viewer screen: Click the “Free Layout Print” icon

21/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17

(3) Select patients and images (2) Select film format. (1) Select Printer and Film size
by clicking buttons. and film orientation. Sizes set at
Printer setting of Service Utility
will be displayed here.

(4) Drag and drop images to the cells.

(5) Select or create / edit film annotation pattern.


Film annotation patter can be
added and deleted.

Select film annotation area


from left sided boxes and
select information to be

displayed and click .

22/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17

(6) Footer can be displayed and strings can be edited.

(8) Vertical border can be adjusted


by dragging the edge of images.

(7) Horizontal border can be


adjusted by dragging each edge of
image.

(10) Exit Free Layout Print function.


(9) Print.

When the images of different patients are printed on one film, Patient ID must be displayed on the annotation
area of each image.
For the reference refer the reference guide.

4.2.3 Installation procedure


Install the following option key software, and then install the Application Key Software.

23/24
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
b. Specification and Function Guideline
2009.07.17
CR-VW 674 : CR-IR 391 Free Layout Print Key Software
This options include DICOM Print key and Free Layout Print key. So there is no need to purchase
another DICOM Print key software to use Free Layout Print function.

A film is printed out to the printer which has been registered in [Service Utility] – [Connected Equipment] –
[Film Printer]. The setting values of the registered printer is used as the printer’s parameters. When a sorter(s)
is mounted to the printer, a film is output to the default bin for free layout print.

4.2.4 Notice on introduction


When the free layout print option is installed, the preparation area for image processing on the HDD becomes
10GB. (if not installed, 5GB). When the HDD capacity is 250GB, the area available for image storage is as
follows:
Free layout print option is not installed: 240GB (excluding 5GB for the system area and 5GB for the
preparation area for image processing)
Free layout print option is installed: 235GB (excluding 5GB for the system area and 10GB for the
preparation area for image processing)

When the amount of image data exceed the capacity above, the data will be deleted with FIFO method.

Before installing the option key, confirm that there is no study which is not verified yet. Otherwise such study
could be deleted due to the expansion of the preparation area for image processing.

After the free layout print is performed, the “Storage/Print” status of the study remains as “Un-Delivered” and
never changes to “Delivered”. Make sure to use DICOM Media Storage, DICOM Print or DICOM CR Storage.
Otherwise Automatic image data deletion function does not work when the HDD becomes full.

24/24