Dear Partner,
This troubleshooting guide aims at helping you find IPC errors quickly. Some of the following information contains
content that can also be found in PSDB (Problem Solution DataBase):
http://psdb2.health.ge.com/knowledgecentral (internal)
http://psdb2.gehealthcare.com/knowledgecentral (external)
Please help us improve this troubleshooting guide by notifying us about problems you may notice (write to
wendelsteinsupportL4@ge.com)
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
1
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
1. Boot-up problems
Problem persists
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
2
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
System boots several times (systems IPC does not complete Check connection and condition
with E or F image with external PC boot procedure? of 25 pole parallel cable, replace if
guard), for LEDs see chapter 1.1. necessary
External PC guard LEDs
25 pole cable OK
Power cable OK
BIOS OK
Progress bar on monitor appears, File system damaged? Use WADI tool to replace file
but does not complete (only for system
systems higher than 10k)
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
3
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
D1: power LED, must light bright, otherwise only parallel port is connected
2
D10 D9 D8 D7 D6 D5 D4 D3
X 0 0 0 0 0 0 0
X X 0 0 0 0 0 0
X X X 0 0 0 0 0
X X X X 0 0 0 B
X X X X X 0 0 B
X X X X X X 0 B
X X X X 0 0 0 0
0 = LED off, X= LED on, B = flashing
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
4
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
Meaning of LEDs:
D16: Hardware Reset, lights red when PC guard activates a hardware reset
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
5
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
2. Specific errors
Error #25 II Camera Error is Frame grabber OK? Check log file for frame grabber
displayed or system has init error. Unplug frame grabber
(intermittent) boot problems card, clean connector, replug. If
defective, replace
Frame grabber card OK
Cables OK
Fuse OK
Power supply OK
Replace B455.
Camera problem?
Replace camera
Camera OK
IPC problem?
Replace IPC
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
6
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
M2 Mode Error is displayed on the Frame grabber OK? Check log file for frame grabber
User Interface init error. Unplug frame grabber
card, clean connector, replug. If
defective, replace
Frame grabber card OK
Hostname OK
Connection problem?
Check cable between B380 (B766
Cables OK?
for systems with a S/N higher
than 10,000) and IPC. Connect
properly or, if defective, replace
Cable OK
Connections OK
-> only for systems with an
external PC guard: Check cables
and external PC guard. Connect
properly or, if defective, replace
COM port OK
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
7
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
3. IPC stuck
IPC stuck after battery replacement Display is dark after Follow the instructions of
Technical Bulletin 09TB003
booting?
The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF GENERAL
ELECTRIC COMPANY, ITS SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-employees or non-designees is
strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
8
GE OEC Medical Systems GmbH
Fluorostar_IPC_Troubleshooting_Guide.docx
4. Alarm sounds
-> Image E, F, G: unplug one after
the other to see if one of them is
IPC alarm sounds Graphic card defective? defective. If alarm goes out,
replace IPC.
Frame grabber OK
Replace IPC.
5. More help
If you have not found help in this Troubleshooting Guide, you may also want to look at:
- 10TB005_User_Interface_Pixel_Disturbances
- 09TB005_Lost_Settings_during_Software_Update