FB7.3.5-06 System Settings (SYS)


Ziehm Vision R V4.00

Change History:
Version Date Major changes / Remarks Author
Initial Version of this document, based on the document “MTS-
V2_Organ_V2.16”. This document is split into single documents,
one for each system.
Correction of typos and appearance in order to achieve a better
handing of the document.
New APRs and modifiers with according settings:
Bone Extremities
Bone Trunk
Vascular Extremities
Vascular Trunk
Bolus Chase
Low Dose
DSA/RSA/MSA Extremities
DSA/RSA/MSA Trunk
DSA/RSA/MSA Extremities CO2
DSA/RSA/MSA Trunk CO2
Auto LPK
Adapting several dose settings for US
1.00 16.03.2011 SW
New Gamma settings with Offset at 50 kV and 100 kV
Settings for Auto LPK and Auto Gain
Adapting US kV/mA-curves of system without boost caps in bone
APRs
Adapting Black Level in several APRs
Removing settings from the document which are now fixed
programmed.
New RSA parameters:
Noise Cut: 512
Trace Threshold: 1023
Trace Offset: 800
Adapting the RSA settings:
dose values
filter settings
Max Frms/sec
Grid off
Gamma 5
Introduction of Max Frms/sec and Start Frms/sec

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 1 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Based on the document “SYS_ZiehmVisionR_V1.00”.


Main Changes:
BoneE, BoneT, Heart, Abdomen, Endo, Soft, VascE, VascT,
Bolus Chase

1.01 06.07.2011 All DSA/RSA Modes JL

All changes w.r.t. to document “SYS_ZiehmVisionR_V1.00”


are marked in red.

Based on the document “SYS_ZiehmVisionR_V1.01”.


(Version 2.00 skipped t o make it consistent with vision
family)
Main Changes:
Add. Settings for IB and TW Versions: Bone, DSA, RSA
Low Dose VascE and VascT
Low Energy Settings: DSA_E, DSA_E CO2, DSA_T, DSA_T
CO2
3.00 31.05.2012 JL
kV/mA curve limit for all Fluoro modifier
kV/mA curve all SSHT modifier
20kW Settings for SSHT all APR
20kW Settings Heart Cine HQ

All changes w.r.t. to document “SYS_ZiehmVisionR_V1.01”


are marked in red.

Based on the document “SYS_ZiehmVisionR_V3.00”.


Main Changes:
Add. Settings for IB and TW Versions: Bone Low Dose
Change settings for IB and TW Versions: DSA HQ and
3.01 24.07.2012 DSA LPK according MTS 2.16 SW
Change settings for IB and TW Versions: RSA according
MTS 2.16
All changes w.r.t. document “SYS_ZiehmVisionR_V3.00”
are marked in blue.
Based on the document “SYS_ZiehmVisionR_V3.01”.
New Settings for all Low Dose modes
3.02 16.08.2012 SW
All changes w.r.t. document “SYS_ZiehmVisionR_V3.00”
are marked in blue.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 2 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Based on the document “SYS_ZiehmVisionR_V4.00


New kV/mA curves and dose rates for all fluoro modes,
20kW curves for Bone_T and Vac_T SSHT,
4.00 06.12.2012 JL
New kV/mA curves for Low Dose
All changes w.r.t. document “SYS_ZiehmVisionR_V3.00”
are marked in red

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 3 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Table of Content

1. Overview................................................................................ 5
2. General Requirements and Information ................................. 6
2.1 Dose and Dose Rate Measurements ..............................................6
3. Specification for Vision R ....................................................... 7
3.1 Overall Settings .........................................................................7
3.2 BONE Extremities .......................................................................8
3.3 BONE Trunk.............................................................................15
3.4 HEART ....................................................................................22
3.5 ABDOMEN ...............................................................................30
3.6 Soft ........................................................................................38
3.7 ENDO .....................................................................................45
3.8 U R O (MTS 2.16).....................................................................52
3.9 Vascular Trunk.........................................................................58
3.10 Vascular Extremities .................................................................65
3.11 Bolus Chase ............................................................................71
3.12 DSA Extremities .......................................................................75
3.13 RSA Extremities .......................................................................79
3.14 DSA Extremities (CO2) ..............................................................83
3.15 RSA Extremities (CO2) ..............................................................87
3.16 DSA Trunk...............................................................................91
3.17 RSA Trunk...............................................................................95
3.18 DSA Trunk (CO2)......................................................................99
3.19 RSA Trunk (CO2) .................................................................... 103
4. Risk Prevention .................................................................. 107
5. Installation Requirements.................................................. 108
6. Requirements from Homologations and Standards ............ 109
7. Production Requirements and Limitations.......................... 110

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 4 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1. Overview
This document describes all the parameters that have to be adjusted at the system
Ziehm Vision R. These values are set in the tab cards “Dose 1”, “Dose 2”, “Dose 3” in the
service settings; please see graphics below as an example. All values are defined and
may not be changed except those values, which are denoted with tba (to be adjusted).

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 5 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

2. General Requirements and Information

Software Version: ≥ 5.30.3 SP3 SV

For IB and TW Versions SYS_ZiehmVision_R_V3.02 stays valid!

An empty field means this field is not applicable (NA).

Rest of world ROW includes EU.

2.1 Dose and Dose Rate Measurements

Dose and dose rate measurements are done by 25 pulses/sec and with grid functionality
disabled.

For US:
Dose and dose rate measurements are done by 30 pulses/sec and with grid functionality
disabled.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 6 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3. Specification for Vision R


3.1 Overall Settings
Max Dose:
For all Fluoro and LPK modes: < 88mG/min (ROW)
For all Fluoro and LPK modes: < 10 R/min (US, non HLC)
For US HLC mode: < 20 R/min (US, HLC)

3.1.1 Overview Dose Settings


Dose and dose rate measurements are done by 25 pulses/sec (US: 30 pulses/sec) and
with grid functionality disabled.

3.1.2 Default Modifier RSA


The default modifier for RSA is all RSA modes “Fluoro”.

3.1.3 Table „Grid“


Grid 31 23 15 10 Res
Grid Regulation Parameters
Motion Level Min 4 4 4 4
Motion Level Max 100 100 100 100
Motion Level Step 2 2 2 2
Motion Count 7500 7500 7500 7500
Motion Hysteresis 500 500 500 500

Auto Dose Reduction Max Pulse 2 2 2 2 2 200 - 240V


pause
Auto Dose Reduction Max Pulse 2 2 2 2 2 USA
pause
Screen 15 15 15 15 0…15

kV reduction tba tba tba tba %


mA reduction / addition tba tba tba tba %

Cooling Water Pump


On Temp. 32 32 32 32 °C
Off Temp. 30 30 30 30 °C

Remark:
The kV reduction and the mA reduction have to be adjusted with every single system.
During a pulse down sequence the image has to keep the same impression regarding
brightness and contrast.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 7 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.1.4 Table „Collimator“


Collimator 12” II 9” II Res

Vignetting 50 30 0-99 %
HEAL 50 50 0-99 %

Arc Protection
Arc Protection on Off Off
Arc Protection kV na na 40-120
Time1 na na 0-3000 s
Step1 na na 1(00)-30(00) V
Time2 na na 0-3000 s
Step2 na na 1(00)-30(00) V

3.1.5 Table „Radiography“


Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap’s

kV mA kV mA kV mA kV mA
1 40 64 40 18 40 50 40 18
2 50 56 50 18 50 44 50 18
3 60 52 60 20 60 40 60 20
4 70 48 120 17 70 37 120 13,7
5 80 46 0 0 80 35 0 0
6 90 44 0 0 90 33 0 0
7 100 42 0 0 100 32 0 0
8 110 41 0 0 110 32 0 0
9 120 40 0 0 120 31 0 0

Short shot < 100ms


kV range 40 110 40 110 40 100 40 100

mA 65 19 60 15

3.2 BONE Extremities


3.2.1 Adjustments

3.2.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,43 0,43 0,52 0,52 0,52 µG/sec

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 8 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose rate US 0,8 1,1 1,6 1,1 1,6 2,0 µG/sec


Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 5 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 8 8 8 8 8 8 0-9
Offs. Lev. (50kV) On On On On On On On/Off
Offs. Lev. (50kV) 5 5 5 5 5 5 0-100%
Offs. Lev.(100kV) On On On On On On On/Off
Offs. Lev.(100kV) 25 25 25 5 5 5 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
2 5 10 2 5 10 2 5 10
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 1 1 2 1 1 2 1 1
Fluoro 3 1 1 3 1 1 3 1 1
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 1 1 2 1 1 2 1 1
Low Dose 3 1 1 3 1 1 3 1 1
REP 1 0 1 1 0 1 1 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5
2 45 2,0 45 2,0
3 60 2,7 60 2,7
4 70 3,5 70 3,5
5 80 5,1 80 5,1
6 90 8,5 90 8,5
7 100 9,5 100 9,5

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 9 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

8 120 9,5 120 9,5


9 0 0 0 0
Start (kV/mA) 60 60
Max Pulse Rate 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain Off Off
Threshold kV
Max Gain
Auto LPK US Off
Auto LPK Off Off
off kV
on kV

Grid
White Clipping
Offset 700 650 650 700 650 650 0…4095
Difference 1700 1600 1600 1700 1600 1600 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection On On On On On On On/Off
Auto Dose Red. On On On On On On On/Off
Motion Sensi. tba tba tba tba tba tba 0…4095

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 10 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.2.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Dose Gain US 0 0 0 0 0 0 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
120 45,0
120 45 46 48 50 52 54 56
120 62,5 120 62,5 58 60 62,
5

Start (kV/mA) 80 60 80 60

Max Frms/s 25 25 15 30
Pulse Length 20 10 16 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 11 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

3.2.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8
9
Start 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/s 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 12 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 100 60 100
4 70 140 70 140
5 75 160 75 160
6 80 160 80 160
7 90 155 90 155
8 100 145 100 155
9 120 140 120 145

Start kV 60 60

Mag Mode 0 ½ 0 1/2


Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

3.2.1.4 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Video
Black Level on on on on on on %
Black Level 5 5 5 5 5 5 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metall Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 13 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 14 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.3 BONE Trunk


3.3.1 Adjustments

3.3.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,52 0,60 0,95 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 5 5 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 3 7 7 7 7 7 %
White Level On On On On On On On/Off
White Level 100 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 7 7 7 7 7 7 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Level On On On On On On On/Off
Offs. lev.(100kV) 25 25 25 25 25 25 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
4 8 12 4 8 12 4 8 12
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 1 1 2 1 1 2 1 1
Fluoro 3 1 1 3 1 1 3 1 1
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 1 1 2 1 1 2 1 1
Low Dose 3 1 1 3 1 1 3 1 1
REP 1 0 1 1 0 1 1 0 1

Dose 2 200 - 240V 100 - 120V

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 15 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

kV mA kV mA
1 40 1,5 40 1,5
2 45 2,0 45 2,0
3 60 3,0 60 3,0
4 70 6,0 70 6,0
5 80 7,5 80 7,5
6 90 8,5 90 8,5
7 100 9,5 100 9,5
6 120 9,5 120 9,5
7
8
9
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain on On On/off
Threshold 110 110 kV
Max Gain 4 4
Auto LPK US Off
Auto LPK On On
Off 80 80
On 110 110

Grid
White Clipping
Offset 700 700 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection off off off off off off On/Off
Auto Dose Red. off off off off off off On/Off
Motion Sensi. 500 600 660 480 600 600 0…4095

3.3.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 16 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60 50,0


Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length 20 10 16 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 15 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pause Burst Pause Bursts Pause

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 17 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.3.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Iris Camera 100 100 100 100 100 100 %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 8 7 8 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8
9
Start 60 60
Max Frms/sec 25 30 10, 16, 20,
30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 18 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.3.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8
9
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 19 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 166 120 166
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0 0

3.3.1.5 Low Dose

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 5 5 5 5 5 5 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metall Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 20 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 21 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.4 HEART
3.4.1 Adjustments

3.4.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 25 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 11 11 14 11 11 11 %
White Level On On On On On On On/Off
White Level 90 90 90 90 90 90 0-200
Metal Black Level tba Tba Tba tba tba tba
Metall Peak tba tba Tba tba tba tba
Windowing
Window 90 90 90 90 90 90 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On
off off off
Offs. lev. (50kV) 2 2 2 2 2 2
Level Off Off Off Off Off Off
Offs. lev.(100kV)
Dynamic 10 10 10 10 10 10 5-100%

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
2 3 4 2 3 4 2 3 4 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 0 0 1 0 0 1 0 0 1
Fluoro 1 0 1 1 0 1 1 0 1
LPK 1 0 1 1 0 1 1 0 1
SSHT 0 0 1 0 0 1 0 0 1
Low Dose 3 1 1 3 1 1 3 1 1
REP 0 0 1 0 0 1 0 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 22 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1 40 1,7 40 1,7
2 50 3,0 50 3,0
3 70 6,0 70 6,0
4 85 10,0 85 10,0
5 100 10,0 100 10,0
6 120 9,5 120 9,5
7
8
9
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na Na Na na na na 0…4095

Auto Gain/Auto LPK


Auto Gain Off off
Threshold kV
Max Gain
Auto LPK US Off
Auto LPK on on
Off 80 80
On 110 110

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 23 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.4.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 15 15 15 15 15 15 %

Video
Black Level 11 11 11 11 11 11 %
White Level 95 95 95 95 95 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 25 25 15 30 1-25 / 30
Pulse Length in 20 10 16 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 15 30 1-25 / 30
Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 24 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 10 40 10
2 50 60 50 60
3 60 90 60 90
4 70 130 70 130
5 75 160 75 140
6 80 175 80 135
7 90 170 90 130
8 100 160 100 120
9 120 140 120 105
Start kV 60 60
Max Frames/s 25 25
Pulse Length in ms 10 7
Start Frames/s 8 8
Burstmode
Frames/s Bursts Pauses Bursts Pauses
1
2
4

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

3.4.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 25 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %
Video
Black Level on on on on on on %
Black Level 200 - 11 11 11 11 11 11
240V
Black Level US 15 15 15 15 15 15 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 75
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 25 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 3,0 50 2,5
4 70 6,0 60 5,0
5 85 15,0 70 9,0
6 100 14,0 80 9,0
7 120 13,0 120 11,2 -
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res


Auto Gain / Auto Dose
Auto Gain Off Off Off Off On/Off
Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.4.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 25 25 25 25 25 25 %

Video
Black Level 200 - 11 11 11 11 11 11 %
240V
Black Level US 17 17 17 17 17 17 %
White Level 95 95 95 95 95 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 26 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3 60 50,0 60 50,0 60 50,0 60 50,0


4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 69,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 27 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.4.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %

Video
Black Level 11 11 11 11 11 11 %
White Level 95 95 95 95 95 95 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 28 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 29 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.5 ABDOMEN
3.5.1 Adjustments

3.5.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,0 µGy/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Gamma USA 7 7 7 7 7 7 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Offs. lev. (50kV) 8 8 8 8 8 8 0-100%
USA
Level Off Off Off Off Off Off On/Off
Offs. lev.(100kV l 0-100%
Dynamic 10 10 10 10 10 10 5-100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 30 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 4 8 2 4 8 2 4 8 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 2 2 0 2 2 0 2
Fluoro 2 0 2 2 0 2 2 0 2
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 0 1 2 0 1 2 0 1
Low D. 2 0 8 2 0 8 2 0 8
REP 1 0 1 1 0 1 1 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
3 65 7,0 65 7,0
4 80 8,0 80 8,0
5 100 7,5 100 7,5
6 120 6,6 120 6,6
7
8
9
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 31 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. 400 400 400 400 400 400 0…4095
Auto Gain/Auto LPK
Auto Gain on on
Threshold 110 110 kV
Max Gain 3 3
Auto LPK US Off
Auto LPK on On
off 80 80
on 110 110

3.5.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µGy/S
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 32 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60 50,0

Max Frms/s 25 25 30 30 1-25 / 30


Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 33 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.5.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µGy/S
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on
Black Level 9 9 9 9 9 9
Black Level US 3 7 6 4 6 4 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 34 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.5.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 35 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.5.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 36 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 37 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.6 Soft
3.6.1 Adjustments

3.6.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 8 8 8 8 8 8 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 7 7 7 7 7 7 0-9
Level Off Off Off Off Off Off On/Off
Level 0 0 0 0 0 0 0-100%
Dynamic 10 10 10 10 10 10 5-100%
Filter MAG 0 MAG 1 MAG 2
Nr. 1 2 3 1 2 3 1 2 3
2 4 6 2 4 6 2 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 0 4 1 0 4 1 0 4
Fluoro 1 0 4 1 0 4 1 0 4
LPK 2 0 8 2 0 8 2 0 8
SSHT 1 0 4 1 0 4 1 0 4
Low D. 2 0 4 2 0 4 2 0 4
REP 1 0 1 1 0 1 1 0 1

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 38 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5
2 50 1,5 50 1,5
3 58 1,5 58 1,5
4 90 8 90 8
5 120 6,8 120 6,8
6
7
8
9
Start (kV/mA) 60 58
Max Frms/sec 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

Auto Gain/Auto LPK


Auto Gain on on
Threshold 100 100 kV
Max Gain 3 3
Auto LPK US Off
Auto LPK on on
off 80 80
on 100 100

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 39 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.6.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metall Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,5

Start (kV/mA) 60 60 60 60

Max Frms/sec 25 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Burs Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 40 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity na na na na

3.6.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 8 8 8 8 8 8 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20, 30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 41 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.6.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 42 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off


Auto Dose Reduction on Off Off Off Off
Motion Sensitivity 0 0 0 0

3.6.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on
Black Level 8 8 8 8 8 8 %
White Level on on on on on on
White Level 95 95 95 95 95 95 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 43 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 44 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.7 ENDO
3.7.1 Adjustments

3.7.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 15 0 15 15 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 5 %
White Level On On On On On On On/Off
White Level 95 85 75 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100
Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Level Off Off Off Off Off Off On/Off
Offs. lev.(100kV)
Dynamic 10 10 10 10 10 10 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 4 2 3 4 2 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 0 1 1 0 1 1 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 2 0 1 2 4 1 2 4 1
SSHT 1 0 1 1 0 1 1 0 1
Low D. 2 0 1 2 4 1 2 4 1
REP 1 0 1 1 4 1 1 4 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 45 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1 40 1,7 40 1,7
2 50 5,0 50 5,0
3 65 7,0 65 7,0
4 80 8,0 80 8,0
5 100 7,5 100 7,5
6 120 6,6 120 6,6
7
8
9
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. 0 0 0 0 0 0 0…4095

Auto Gain/Auto LPK


Auto Gain on on
Threshold 100 100 kV
Max Gain 3 3
Auto LPK US Off
Auto LPK on on
off 80 80
on 100 100

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 46 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.7.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 15 0 15 15 %
Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metall Black tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,7 40 1,7 40 5 40 5,0
2 50 5,0 50 5,0 50 20 50 20,0
3 60 60 60 50 60 50,0
4 75 38,0 75 38,0
5 80 80 80 40 44 46 50 51 54 58 80 67,0
80 80 60 62 65 67 71 73 75
80 80 75 75 75
6 90 90 90 37 40 42 46 47 50 53
90 90 55 57 60 62 65 68 70
90 90 72 75 75
7 100 100 100 34 37 39 42 44 46 49 100 58,0
100 100 51 53 56 58 60 63 65
100 100 67 70 73
8 120 28,0 120 28,0 120 30 32 34 36 38 40 42 120 50,0 < 10 R/min !!!
120 120 45 46 48 50 52 54 56
120 120 58 60 62,
5

Start (kV/mA) 75 75 80 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length 20 20 10 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 30 30 1-25 / 30
Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 47 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.7.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,5 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 80 80 %
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20, 30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 48 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.7.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,6 1,6 1,6 1,6 1,6 1,6 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 15 15 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/s 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 49 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.7.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 50 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 51 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.8 U R O (MTS 2.16)


3.8.1 Adjustments

3.8.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 4 %
Peak 95 95 95 95 95 95 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Level
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On On/Off
Level 15 15 15 15 15 15 0-
100%
Dynamic 10 10 10 10 10 10 5-
100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 52 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
5 8 16 5 8 16 5 8 16 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 2 0 1 2 0 1 2 0 1
SSHT 2 0 1 2 0 1 2 0 1
REP 1 0 1 1 4 1 1 4 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5
2 55 6,0 55 6,0
3 90 13,0 90 13,0
4 100 13,0 100 13,0
5 120 13,0 120 11,2 – < 10 R/min!
8,0
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 90 90
Pulse Length in ms 30 30 10, 16, 20,
30
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 53 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.8.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5 40 5,0
2 50 20,0 50 20,0 50 20 50 20,0
3 60 50,0 60 50,0 60 50 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,5

Start (kV/mA) 80 60 80 60
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Bilder/s 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 54 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

3.8.1.3 LPK

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 3 3 3 3 3 4 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba Tba tba tba
Level
Peak tba tba tba Tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
Start 80 80
Pulse Length in ms 30 30 10, 16, 20,
30
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 55 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.8.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap´s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Bilder/s 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 56 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 57 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.9 Vascular Trunk


3.9.1 Adjustments

3.9.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100
Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. On On On On On On
(100kV)
Offs. Lev. 25 25 25 25 25 25
(100kV)
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 6 2 3 6 2 3 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 0 1 2 0 1 2 0 1
Low Dose 3 1 1 3 1 1 3 1 1
REP 1 0 1 1 4 1 1 4 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 58 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1 40 1,7 40 1,7
2 50 5,0 50 5,0
3 65 7,0 65 7,0
4 80 8,0 80 8,0
5 100 7,5 100 7,5
6 120 6,6 120 6,6
7
8
9
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain on on
Threshold 100 100 kV
Max Gain 2 2
Auto LPK US Off
Auto LPK On on
off 80 80
on 100 100

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 59 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.9.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60

Max Frms/sec 25 25 30 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 60 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

3.9.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Dose rate USA 0,8 1,0 1,2 1,0 1,2 1,4 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Video
Black Level on on on on on on
Black Level 9 9 9 9 9 9
Black Level US 3 7 6 4 6 4 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain on on on on on On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 61 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.9.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 62 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 166 120 166
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 12 25
Pulse Length in ms 7 7
Start Frames/s 12 25 12 25

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.9.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose ROW 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Dose US 0,35 0,35 0,35 0,35 0,35 0,35
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 63 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 64 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.10 Vascular Extremities


3.10.1 Adjustments

3.10.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,52 0,60 0,95 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100
Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. 100kV) On On On On On On
Offs. Lev. 100kV) 25 25 25 25 25 25
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 5 2 3 5 2 3 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 2 2 0 2 2 0 2
Fluoro 2 0 2 2 0 2 2 0 2
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 0 1 2 0 1 2 0 1
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 65 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain off off
Threshold kV
Auto LPK US Off
Max Gain
Auto LPK off off
off
on

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 66 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.10.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30
Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 67 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.10.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 68 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

20 kW Generator
200 – 240 V 100 – 120 V
kV mA kV mA
1 40 1.5 40 1.5
2 50 40 50 40
3 60 120 60 120
4 70 200 70 200
5 75 190 75 190
6 80 185 80 185
7 90 170 90 170
8 100 160 100 160
9 120 145 120 145
Start kV 60 60
Mag Mode 0 1/2 0 1/2
Max Frames/s 12 25 15 30
Pulse Length in ms 7 7
Start Frames/s 12 25 15 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.10.1.4 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose ROW 0,27 0,30 0,35 0,27 0,30 0,35 µG/s
Dose US 0,35 0,35 0,35 0,35 0,35 0,35
Iris Camera tba tba tba tba tba tba %
CCD Fill 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) 20(tba) %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 69 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0.2 40 0.2
2 80 0.2 80 0.2
3 100 2.5 100 2.5
4 120 5.0 120 5.0
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 70 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.11 Bolus Chase


3.11.1 Adjustments

3.11.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,00 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100
Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. 100kV) On On On On On On
Offs. Lev. 100kV) 25 25 25 25 25 25
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 5 2 3 5 2 3 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 0 0 0 0 0 0 0 0 0
Fluoro 0 0 0 0 0 0 0 0 0
LPK 0 0 0 0 0 0 0 0 0
SSHT 0 0 0 0 0 0 0 0 0
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 71 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain off off
Threshold kV
Max Gain
Auto LPK US Off
Auto LPK off off
off
on

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 72 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.11.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,7 1,9 1,7 1,9 2,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 73 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

3.11.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,5 40 1,5 40 1,5 40 1,5
2 50 10,0 50 10,0 50 10,0 50 10,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 10 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 74 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.12 DSA Extremities


3.12.1 Adjustments

3.12.1.1 Fluoro

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,3 1,6 1,9 1,6 1,9 2,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 17 17 17 17 17 %
White Level 90 90 90 90 90 90 0-200
Metal Black Level 17 17 17 17 17 17 %
Peak 95 95 95 95 95 95 %

12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 45 45 45 45 45 45 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV On On On On On On On/Off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 75 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 3 0 1 3 0 1 3 0 1
Fluoro 3 0 1 3 0 1 3 0 1
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

Dose 2
7.5 kW Generator 20 kW Generator
kV mA kV mA
1 40 1.7 40 1.7
2 50 3.0 50 3.0
3 60 6.0 60 6.0
4 70 14.0 70 14.0
5 90 15.0 90 15.0
6 100 15.0 100 15.0
7 120 13.0 120 13.0
8 0 0 0 0
9 0 0 0 0
Start kV 60 60
Max Frames/s 25 30
Pulse Length in ms 20 20
Start Frames/s 25 30
Burstmode
Frames/s Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1 9 16
2 5 7
4 3 3

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 76 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 100 100 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.12.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,8 2,2 2,6 2,0 2,4 2,8 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 17 17 17 17 17 %
White Level 90 90 90 90 90 90 0-200
Metal Black Level 17 17 17 17 17 17 %
Peak 95 95 95 95 95 95 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 77 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,0 40 1,0
2 50 4,0 50 4,0
3 60 16,0 60 16,0
4 70 30,0 70 30,0
5 80 42,0 80 42,0
6 90 47,0 90 47,0
5 100 55,0 100 55,0
7 110 56,0 110 56,0
7 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length 20 20 10, 16, 20, 30
in ms
Start Frms/s 25 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 78 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.13 RSA Extremities


3.13.1 Adjustments

3.13.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,3 1,6 1,3 1,6 1,9 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 65 65 65 65 65 65 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 0 0 0 0 0 0 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 50 50 50 50 50 50 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 2 1 1 2 1 1 1 1
Fluoro 2 1 1 2 1 1 2 1 1
LPK 0 0 0 0 0 0 0 0 1
SSHT 0 0 0 0 0 0 0 0 1
Low D 0 0 0 0 0 0 0 0 1
REP 0 0 0 0 0 0 0 0 1

Dose 2 200 - 240V 100 - 120V

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 79 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

kV mA kV mA
1 40 1,7 40 1,7
2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 15,0 100 15,0
5 120 13,0 120 13,0

Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 20 20 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter U 622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U 622>K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

3.13.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 80 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose Gain tba tba tba tba tba tba 0-9


Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 12 12 15 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 12 12 15 30 1-25 / 30

Burstmode
Frms/sec Burs Pause Burst Pauses Burst Pauses Bursts Pauses
ts s s s

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 81 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain Off Off On/Off
Auto Dose Red. Off Off
Motion Sensitivity na na On/Off
RSA Parameter U622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U622> K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 82 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.14 DSA Extremities (CO2)


3.14.1 Adjustments

3.14.1.1 Fluoro

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,0 2,0 2,0 2,0 2,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 16 16 16 16 16 16 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 83 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

Dose 2
7.5 kW Generator 20 kW Generator
kV mA kV mA
1 40 1.7 40 1.7
2 50 3.0 50 3.0
3 60 6.0 60 6.0
4 70 14.0 70 14.0
5 90 15.0 90 15.0
6 100 15.0 100 15.0
7 120 13.0 120 13.0
8 0 0 0 0
9 0 0 0 0
Start kV 60 60
Max Frames/s 25 30
Pulse Length in ms 20 20
Start Frames/s 25 30
Burstmode
Frames/s Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1 9 16
2 5 7
4 3 3

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 84 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.14.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 4,5 4,5 4,5 4,5 4,5 4,5 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 85 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 2,0 40 2,0
2 50 8,0 50 8,0
3 60 16,0 60 16,0
4 70 30,0 70 30,0
5 80 42,0 80 42,0
6 90 47,0 90 47,0
5 100 55,0 100 55,0
7 110 56,0 110 56,0
7 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Pulse 25 30
Rate
Pulse Length 20 16 10, 16, 20, 30
in ms
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 86 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.15 RSA Extremities (CO2)


3.15.1 Adjustments

3.15.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 22 22 22 22 22 22 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 1 1 1 1 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 87 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 15,0 100 15,0
5 120 13,0 120 13,0
Start (kV/mA) 60 60
Max pulse rate 25 30 1-25 / 30
Pulse Length in ms 20 20 10, 16, 20,
30
Max Frms/sec 25 30 1-25 / 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter U622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U622> K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

3.15.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µGy/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 88 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/s 12 12 15 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 12 12 15 30 1-25 / 30

Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses
s s s s

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 89 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain Off Off On/Off
Auto Dose Red. Off Off
Motion Sensitivity na na On/Off
RSA Parameter U622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U622> K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 90 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.16 DSA Trunk


3.16.1 Adjustments

3.16.1.1 Fluoro

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,5 1,8 2,1 1,8 2,1 2,2 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 45 45 45 45 45 45 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 91 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 3 0 1 3 0 1 3 0 1
Fluoro 3 0 1 3 0 1 3 0 1
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

Dose 2
7.5 kW Generator 20 kW Generator
kV mA kV mA
1 40 1.7 40 1.7
2 50 3.0 50 3.0
3 60 6.0 60 6.0
4 70 14.0 70 14.0
5 90 15.0 90 15.0
6 100 15.0 100 15.0
7 120 13.0 120 13.0
8 0 0 0 0
9 0 0 0 0
Start kV 60 60
Max Frames/s 25 30
Pulse Length in ms 20 20
Start Frames/s 25 30
Burstmode
Frames/s Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 92 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.16.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,5 3,2 3,8 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba Tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 93 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 5,0 40 5,0
2 50 25,0 50 25,0
3 60 50,0 60 50,0
4 80 75,0 70 75,0
5 90 75,0 80 75,0
6 100 73,0 90 73,0
5 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length 20 16 10, 16, 20, 30
in ms
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 94 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.17 RSA Trunk


3.17.1 Adjustments

3.17.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,9 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 70 70 70 70 70 70 0-100
Level 50 50 50 50 50 50 0-100
Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 0 0 0 0 0 0 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 50 50 50 50 50 50 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 6 3 4 6 3 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 2 1 1 2 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT 0 0 0 0 0 0 0 0 1
Low D 0 0 0 0 0 0 0 0 1
REP 0 0 0 0 0 0 0 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 95 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

1 40 1,7 40 3,0
2 60 10,0 45 8,0
3 80 15,0 50 8,0
4 100 18,0 70 15,0
5 120 17,2 75 16,0
6 120 13,0
Start (kV/mA) 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%
Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter U 622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U 622> K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

3.17.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,5 3,0 3,5 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 96 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 80 60
Max Frms/s 12 12 15 15 1-25 / 30
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Start Frms/s 12 12 15 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 97 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain Off Off On/Off
Auto Dose Red. Off Off
Motion Sensitivity na na On/Off
RSA Parameter U622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U622> K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 98 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.18 DSA Trunk (CO2)


3.18.1 Adjustments

3.18.1.1 Fluoro

Dose 1 12” II 9” II Res


Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,0 2,0 2,0 2,0 2,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 16 16 16 16 16 16 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 99 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

Dose 2
7.5 kW Generator 20 kW Generator
kV mA kV mA
1 40 1.7 40 1.7
2 50 3.0 50 3.0
3 60 6.0 60 6.0
4 70 14.0 70 14.0
5 90 15.0 90 15.0
6 100 15.0 100 15.0
7 120 13.0 120 13.0
Start kV 60 60
Max Frames/s 25 30
Pulse Length in ms 20 20
Start Frames/s 25 30
Burstmode
Frames/s Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 100 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%
Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.18.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 4,5 4,5 4,5 4,5 4,5 4,5 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 5,0 40 5,0
2 50 20,0 50 20,0
3 60 50,0 60 50,0
4 80 75,0 70 75,0
5 90 75,0 80 75,0
6 100 73,0 90 73,0
7 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Frms/s 25 30

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 101 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Pulse Length 20 16 10, 16, 20, 30


in ms
Start Frms/s 12,5 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 102 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.19 RSA Trunk (CO2)


3.19.1 Adjustments

3.19.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 22 22 22 22 22 22 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 1 1 1 1 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na
Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 3,0

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 103 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

2 60 10,0 45 8,0
3 80 15,0 50 8,0
4 100 18,0 70 15,0
5 120 17,2 75 16,0
6 120 13,0
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30
Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 15 15 0…100%
Video low 0 0 0…100%
high 80 80 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%
Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter U 622≤ K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U 622> K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 104 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

3.19.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/s 12 12 15 15 1-25 / 30
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Start Frms/s 12 12 15 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain Off Off On/Off

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 105 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

Auto Dose Red. Off Off


Motion Sensitivity na na On/Off
RSA Parameter U622 K15
Noise Cut 1023 1023 1023 1023 1023 1023 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023
RSA Parameter U622 ≥ K15
Noise Cut 600 600 600 600 600 600 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 1000 1000 1000 1000 1000 1000 0…1023

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 106 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

4. Risk Prevention
The definition of the dose rate is done with respect to the DIN EN 60601-1 in every APR
including the modifiers (HQ, LPK, SSHT). The dose value must be adjusted for each
single APR, magnifications and modifiers separately.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 107 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

5. Installation Requirements
For installation use the latest valid working instruction for system calibration.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 108 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

6. Requirements from Homologations and


Standards
NA.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 109 von 110


Eigentum der Ziehm Imaging GmbH
Kopien / Ausdrucke des Dokuments unterliegen nicht der Dokumentenlenkung. Die jeweils gültige Version ist auf dem Server.
Dokumententyp: Formblatt
FB7.3.5-06 System Settings (SYS)
Ziehm Vision R V4.00

7. Production Requirements and Limitations


NA.

SYS_ZiehmVisionR_V4.00_Service.doc | Version: A Seite 110 von 110