BFM-800
FETAL MONITOR Ver 1.1
OPERATION MANUAL
BFM-800
FETAL MONITOR Ver 1.1
Table of Contents
Instruction
General 3
Warranty Period 4
Request for Services 5
How to Contact Us 6
3. Description of BFM-800
3.1 Characteristics and Principles of BFM-800 14
3.2 Composition of Equipment 14
3.3 Part Names 15
3.4 Description of Equipment Symbols 17
4. Installation of BFM-800
4.1 Cautions for Installation 18
4.2 Installation of Main Equipment 18
4.3 Connection of Cable 20
4.4 Inserting of Print Paper 20
7. Central Function(Option)
7.1 Installation of Central Function 42
7.2 DTE_TCP Setup 43
8. Power Supply
8.1 AC Power Source 44
8.2 Battery Power 44
8.3 Battery Specifications 44
9. Troubleshooting 46
10. Product Specifications 48
BFM-800
INTRODUCTION
General
Thank you for using BFM-800 (Fetal Monitor). Please read this User s Manual carefully before
operating this equipment for proper handling and maintenance. It is wise to become familiar with
all the functions and features of the equipment, to ensure safe handling and reliable performance.
Please, try to fully understand the features, operation and maintenance methods of this equipment
prior to the actual operation by thoroughly reading this manual. This is the only way to ensure safe
operation and reliable performance of the equipment for a long period of time. BIONICS Corp.
only provides reliable products to our customers.
All of assembling, extending, adjusting or repairing of this equipment shall be carried out only
by the service personnel authorized by either our company or BIONICS Corp. itself.
Electrical connection or installation location has to comply with relevant regulations. The
equipment has to be operated in accordance with the User s Manual.
Operation of this equipment must be as directed in this manual.
This equipment has to be operated under the supervision of licensed medical professionals.
This equipment is used to monitor the conditions of the fetus. To insure the safety of the fetus and
mother, use only the parts and accessories recommended by BIONICS Corp.
In the event that this equipment needs to be used in accordance with other equipment that is not
listed in this manual, be sure to notify us or our authorized dealers about this intent.
General 3
FETAL MONITOR
Warranty Period
Our products are produced through strict quality control and inspection processes. The
compensation standards in terms of repairing and exchange policies for our products complies
with the Compensation Regulations for Consumers issued by the Economic Planning Board
of the Korean Government.
The warranty period for this product is specified to be one year.
If the equipment malfunctions during the course of normal operation while under the warranty
period, it will be repaired without any cost to the customer at one of our Service Centers.
If the equipment malfunctions during the warranty period, please inform the service personnel
on the basic facts, such as model name, product No., purchase date, and a brief description of
the problem.
BIONICS or the local distributor will make available on request circuit diagrams, component
part lists, descriptions, calibration instructions or other information which will assist your
appropriately qualified technical personnel to repair those parts of equipment which are
designated by BIONICS repairable.
4 Warranty Period
BFM-800
Service on BIONICS products can be performed only by the personnel authorized by the
Customer Service Dept. of BIONICS Corp. or by BIONICS itself. In the event that the product is
serviced or tried to be serviced by an unauthorized third party during the warranty period, the
warranty shall be nullified. The Customer Service Dept. of BIONICS or the authorized dealers are
obliged to provide the services requested by the customers. When any abnormalities are found in
the equipment or when a danger to bodily harm exists, the equipment must be repaired
immediately and adequately by authorized personnel.
Immediately contact our Customer Service Dept. or local distributor. Please be prepared to give
the receiver the model name, product No., purchase date, and a brief description of malfunction
or enquire before making a call.
Our staff in the Customer Service Dept. will first try to analyze and solve the problem while
talking to the customer on the phone, if possible. And for the complicated problems which
cannot be dealt with over the telephone, we will provide on the site service as quickly as
possible.
How to Contact Us
Contact us at following phone numbers for product and order enquiries and service requests. Do
not hesitate to call us for any inquiries. We are here to help you.
Service Request
BIONICS Corp. -Tel +82-2-3275-5422
When requesting a repair service, please have ready the product s model name, product
serial No., and a brief description of the problem when contacting our Customer Service
Dept.
6 How to Contact Us
BFM-800
This manual contains all the information needed to operate our Fetal Monitor, BFM-800,
which has been designed by giving the utmost consideration on the users convenience, and it
provides exact information on the conditions of the fetus including the measured values such
as fetal heart rate, and uterine contraction.
Because the chapters in this manual are constructed so that each chapter can be used
independently, some contents may appear in more than one chapter.
This manual is written to help users effectively operate BFM-800. Each feature of this
equipment could be used more effectively, if relevant medical books are referenced for a
clearer understanding of clinical meaning and pathology.
In case some problems occur during the operation of our products, please contact our
company or the A/S Dept. of our contracted A/S company.
Users have to carefully read this manual before operation of the equipment.
This manual consists of the various chapters :
Chapter 9 . Troubleshooting
Chapter 10 . Product Specifications
Symbols are used to emphasize the agreed details as follow. Users need to specifically follow all
the cautions and notes listed in this manual.
In the event that the product is damaged due to misuse or negligence by the user, the
manufacturer or the authorized dealers shall not be responsible for any damage or loss to the
product.
The title “Warning” is used to inform the users of possible causes that could inflict injury,
death, or property damage to the patient.
The title “Caution” is used to inform the users of possible causes that could inflict injury to
the patient although it might not be severe enough to cause death.
The title “Note” is used to inform the users of items that are of importance in terms of
installation, operation, or maintenance of the equipment although failure does not inflict
bodily harm to the patient.
Prior to operation of the equipment, make sure to check the following items:
The power supply is appropriate. (100 - 240Vac).
Connections (power line or selected equipment) are properly made to the equipment. For
proper connections, refer to Chapter 4.
The equipment is properly grounded. (If not, some noises could be generated.)
The proper measuring accessory for the desired parameter is connected to the equipment prior
to the turning on of the equipment.
The equipment should be placed far from a generator, X-ray equipment broadcasting
equipment or transmitting wires, so as to prevent electrical noises from being generated
during operation. When these devices are near the equipment, it can produce inaccurate
measurements. For BFM-800, both independent circuit and stable grounding are essential
requirements. In the event that the power source is shared with other electronic equipment,
inaccurate outputs may result.
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (e.g. IEC 950 for data processing equipment and
IEC 601-1 for medical equipment). Furthermore all configurations shall comply with the
system standard EN 60601-1-1:1993. Everybody who connects additional equipment to the
signal input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the system standard IEC 601-1-
1:1993. If in doubt, consult the technical service department or your local representative
In the hospital, doctors and patients are expose to dangerous, uncontrollable compensating
currents. These currents are due to the potential differences between connected equipment and
reachable conducting parts often in the medical rooms. The safety solution to the problem is
accomplished with consistent equipotential bonding. Medical equipment is connected with
connecting leads made up of angled sockets to equipotential bonding network in medical rooms.
BFM-800 and its accessories(including probes) can be cleaned using various methods. Please
follow the methods listed below to avoid unnecessary damage or contamination to the equipment.
In the event that harmful (unauthorized) materials are used for cleaning, the damaged or
contaminated equipment shall not be serviced without charges regardless of warranty period.
Please check carefully both frame and sensor after cleaning the equipment. Do not use
equipment that is worn or damaged.
At least once a month, clean and wipe off the frame by using a soft cloth after wetting it with
lukewarm water and alcohol. Do not use lacquer, thinner, ethylene or oxides, which could be
harmful to the equipment. Make sure both cables and accessories are free of dust or contaminants,
and wipe them off with a soft cloth wetted with lukewarm water(40 / 104 ), and at least once
a week, clean them by using clinical alcohol. Do not submerge the accessories in any liquid or
detergent. Also, make sure any liquid is not to get into the equipment or probe.
Place the system far from power generators, X-ray machines, broadcasting stations, and
transmission lines to avoid electrical noise during scanning. Close proximity to such items
may result in abnormal output. An independent circuit and a safety grounded outlet are
strongly recommended for the BPM-800. Poor or abnormal output may occur if the system
shares a power source with other electrical or electronic equipment.
Do not dispose of used probes into any hazard place. Always consider environmental
contamination.
There is real time clock which contains an internal battery inside of the chip. When users
dispose of this chip, please dispose of waste in the proper place for environmental protection.
DS12C887 real time clock.
BFM-800 is the fetal monitor that measures the fetal heart rate and uterine contraction. BFM-800
irradiates ultrasound wave to the abdomen of a pregnant woman, and detects the doppler
frequency signal reflected from the heart of the fetus. BFM-800 analyzes this signal and displays
the waveform of the heart rate in the printer. Also, BFM-800 provides the sound from the heart of
fetus. During delivery, BFM-800 can measure the heart rate of fetus with the FECG electrode.
Also BFM-800 can measure the uterine contraction of a pregnant woman with pressure sensors
and display in waveform along with numerical values. This waveform informs the period and
magnitude of the uterine contraction. BFM-800 displays the various information(HR, UC) of a
fetus and mother in real time on LCD with numerical values. The user can set up or adjust the
setup values by using the keys in the front, such as setting alarm. BFM-800 can measure the
condition of twins and provide the user with the Central System through the use of an optional
Ethernet Card.
Standard Accessories
- Ultrasound Probe (DOP I, DOP II )
- UC Probe
- Event Mark Jack
- Printer Paper (2EA)
- Power Code (1EA)
- User s Manual (1EA)
- Ultrasound Gel
- Probe Belt (3EA)
- Accessory Case
- Probe Holder
- Wall Mounted Panel
- Ground Cable
14 Description of BFM-800
BFM-800
Options
- FECG Probe
- Central Cable/Ethernet Card/Interface Program
- Cart
Description of BFM-800 15
FETAL MONITOR
Right Side
In order to avoid electrical shock, do not open the cover. Disassembling of the equipment
must be done only by service personnel authorized by BIONICS Corp.
User must pay attention to the connection to LAN port. Always consider the summation of
leakage current and consult with your hospital biomedical engineer if in doubt.
16 Description of BFM-800
BFM-800
Type CF equipment.
Equipotentiality
Operation by AC Power
Operation by Battery
Volume
Alarm
UC Reference
Menu
Record
Description of BFM-800 17
FETAL MONITOR
18 Installation of BFM-800
BFM-800
Supporting Panel
Accessory Case
Installation of BFM-800 19
FETAL MONITOR
1) Power Connection
Plug in the power cable into power port of BFM-800.
2) Connection between the main Equipment and Probe
- Connect the desired probe to the appropriate connecting ports on the bottom of the
equipment.(Refer to below figure.)
- Malfunction may occur, if connection is not proper.
3) Connection of Ground Cable Connect the grounding port of the equipment with the earth
terminal at the site of installation.
4) Connection of LAN Cable(Option) Connect the LAN port of the equipment with the LAN port
at the site of installation.
- Push the top door of the equipment upward to insert the printer paper as in figure below.
- Pull the bottom door towards the front. The space for the printer and the printer paper is then
evident.
- Lift up the grip of the right side of printer to insert print paper into the black roller of printer.
- Insert the paper into roller of printer. The surface to print adjoins the
20 Installation of BFM-800
BFM-800
bottom and 240 ,the heart rate scaled paper, should be placed at the left of the printer.
- Draw the paper 4~5cm and lower the grip of printer to fix the paper.
- Attention! Lower the grip of printer with inserting paper into printer. Otherwise, the thermal
head of the printer may be damaged.
- Close the bottom door. The top door slides down automatically.
Please turn off the power switch to prevent malfunction when replacing paper.
Installation of BFM-800 21
FETAL MONITOR
Check the following items for the equipment and measuring module, beforeoperating the
equipment.
Check whether there exists any mechanical danger.
Inspect the lead that connects it to outside, the power plug, and accessories.
Check all the measuring instruments required to monitor the fetus.
First, connect the probe to use with the equipment. Generally, connect the probe as follows :
- Antipartum Monitoring of only One Fetus : DOP I, UC, Mark Jack
- Antipartum Monitoring of Twins : DOP I, DOP II, UC, Mark Jack
- Intrapartum Monitoring : FECG
Step 1) Turn the power switch on. It will automatically start the self-diagnosis procedure for the
equipment (If an error message appears, refer to Ch9. Solutions to Simple Problems ).
After the self-diagnosis, the equipment is ready to be used.
Step 2) If you want to change the set values while in use, use the keys of the equipment
(refer to Ch.6)
Step 3) Put the probe to measure on the abdomen of a pregnant woman. Fix the probe by fastening
the belt. (refer to 5.5, 5.6, 5.7)
Step 4) Hear the sound of the heart and check the heart rate value displayed on the LCD.
Step 5) Print the value of measurement by record key. (Refer to related items in this manual about
the measuring and setup.)
5.4 Key
(1) (Volume Up) : Volume Up key of selected module of FHR I or FHR II. During the use
of the Menu for setup, this key is used to move between items of Menu or change the
setting value.
(2) (Volume Down) : Volume Down key of selected module of FHR I or II. During the use
of the Menu for setup, this key is used to move between items of Menu or change the
setting value.
Lamp
The LEDs on the lower right of the front panel show the kind of the power used.
(8) (AC Power) : Green LED is turned on when using AC Power.
(9) (Battery) : Orange LED is turned on when using Battery.
To measure FHR, put the ultrasonic gel on the surface of the probe.
If heart rate is beyond the heart rate limit of the alarm, the warning message (Tachycardia or
Bradycardia) is displayed and alarm rings.
During monitoring, if the line of probe breaks down or the probe comes out from the connector,
the message ( DOP I OFF or DOP II OFF ) is displayed and the information sound Ding
Dong is heard.
Measurement of FHR
(1) Touch the abdomen of a pregnant woman and search for the back of the fetus, then, put
Doppler probe on that region. In the case that a fetus lays itself down sideways, put down
Doppler probe as seen in the figure below.
In the case of fixing the probe on the front of a fetus, not the back, brings about an inaccurate
measurement of heart rate owing to unstable signal from the fetus.
When the fetus is heading down-ward or up-ward, you must pay attention to placing the probe.
Put the probe around the navel of the patient.
(2) Moving the probe around slowly, search for the position where you can hear the heart-beating
more clearly, then, fasten the belt after positioning the probe.
Do not push the volume key until you find a correct position with probe. When you position
the probe on the correct area and hear a relatively clearheart-beat and see a FHR rhythm symbol
blinking as per the heart-beat, you can adjust volume key as needed.
(3) Carefully fasten the belt so that probe doesn t slip.
As shown in the drawing, above you need to arrange the probe up-ward of the patient, to
minimize cable damage and use it more efficiently.
(4) Put the ultrasonic gel on Doppler probe to prevent the attenuation of the ultrasound signal and
remove the bubble between the abdomen and the probe.
(5) BFM-800 uses Doppler probe with the Button-Top form. Stick the button of Doppler probe in
the button-hole of belt to fix Doppler probe.
(6) 4~5 seconds is needed to calculate and display the heart rate. If the stable heart rate is
displayed, start printing.
5.6 UC Measurement
With the help of the belt, the UC probe is used in measuring external uterine contraction. UC
probe is affixed on the maternal abdominal wall and measures UC by the pressure made by
uterine contraction.
Connecting UC Probe
Connect UC probe with the connector(UC) at the bottom of the equipment.
Initial Screen according to connecting UC Probe
If you don t connect UC probe, two hyphens are displayed on the UC value area. If you begin
to measure UC, the value proportionate to the uterine contraction of patient is displayed. During
monitoring, if the line of probe breaks down or probe comes out from the connector, the
message UC OFF! is displayed and the information sound Ding Dong is heard.
UC Measurement
(1) Place UC probe at 10cm upper of navel (fundus) and fasten with belt. If UC probe are placed
at below navel, measurement can be influenced by maternal respiration.
(2) Fasten UC probe with belt. At this time, UC value must show 20 ~ 90.
(3) Push Reference key to make UC value 10 .
(4) When uterine contraction is not occurred, check the place of UC probe again.
After fastening UC probe with belt, you need to push Reference key one time before use.
Afterward, UC reference is automatically setted.
If you want to change UC probe position, you need to try above four steps again.
You must print by pushing Record key after Doppler probe and UC probe are placed at right
position.
Even if FECG probe is connected, but if you don t use FECG probe, no sound rings. And 0 is
displayed on the heart rate value area. If you begin to m o n i t o r, the measured heart rate is
displayed on that area and the heart rhythm symbol blinks in synchronism to the beating of the
heart. If only the outer ring of the heart rhythm symbol blinks, it represents the measured heart rate
is unstable. And you can hear the beep sound representing the period of the heart-beating. The
sound volume can be controlled by UP/DOWN key. If the heart rate is beyond the heart rate limit
of alarm, the warning message ( Tachycardia or Bradycardia) is displayed and alarm sound rings.
During monitoring, if the line of probe breaks down or the probe comes out of the connector, the
message FECG OFF is displayed and the information sound Ding Dong is heard.
FECG Measurement
(1) Connect FECG transducer with the equipment and fix the metal plate of FECG transducer to
the thigh of the mother.
(2) Connect intradermal electrode with the terminal of FECG transducer. (If necessary, refer to the
manual on packing of intradermal electrode.) Push the plug of FECG transducer terminal and
put each wire of the intradermal electrode in the hole of the terminal. The two pieces of wire
from the intradermal electrode may not be put in the same hole.
(3) Check the direction of the spiral of intradermal electrode and stick and affix the electrode into
the scalp of fetus head. Then, the Heart Rate of fetus is displayed on the FHR II area of LCD.
You can hear the beep sound and see the Heart Rhythm representing the period of heart-
beating.
(4) 4~5 seconds is needed to calculate and display the Heart Rate. If a stable Heart Rate is
displayed, start printing.
MARKER is used to record the fetal movement perceived by the mother. If a mother presses the
button of Marker, an arrow is printed on paper. MARK JACK is connected with the Mark
connector on the bottom of the equipment.
How to PRINT
(1) If you want to print, Press RECORD key. The symbol which informs the operation of printer
appears below the alarm symbol of the right of LCD. The monitoring measurements are
printed.
(2) During printing, if you want to stop printing, Press RECORD key again. The print symbol
disappears and the operation of the printer stops. At this time, the printer feeds paper 3 ~ 4cm
after stopping the recording.
(3) When you don t use the printer, if you want to feed paper through the printer, press
RECORD key for longer than 2 seconds. As long as you press RECORD key continuously,
the printer will continue to feed paper.
(4) Basically, the graph of FHR II area is printed dimmer than that of FHR II to distinguish FHR I
from FHR II.
(5) If a error message ( “Door Open!” or “Paper Out!” )is displayed on LCD, solve the problem
and print again.
Door Open ! : The grip of printer is lifted up.
Paper Out ! : paper is used up or caught in the roller of printer.
OFFSET of FHR II
(1) When you monitor FHR I, II at the same time, it is hard to distinguish FHR I from FHR II if
the measurements of FHR I are similar to those of FHR II. At this time, the OFFSET function
will help you. If you use OFFSET , FHR II measurement is printed by 20(bpm) added value.
(2) Before printing, set the Offset(+20 in Print setup Menu) as o n b y MENU/UP/DOWN/
RECORD key.
(3) This setting will not be changed even if the power is off. Therefore, if you don t want to use
OFFSET function afterwards, you must change the setup to off
ZOOM
(1) This function enables you to watch the variation of FHR closely.
(2) If you set rng as on in print setup menu, the printing range of FHR(50 ~ 240(bpm)) is
change to 105 ~ 200(bpm). You can see the FHR scale(105 ~ 200(bpm)) of a small size on
paper.
Recording Paper
If you begin to print, the various information regarding the present situation is printed first,
before printing the measurements. This information includes the date, time, Probe On/Off,
print setting, alarm limit, and the information about a mother.
(1) If you press UP key, Doppler and FECG beep sound is louder than before. And if you press
DOWN key, Doppler and FECG beep sound is lower than before. You can see the change in
sound in the volume step symbol on LCD.
(3) Only the selected one of FHR I, FHR II is controlled by UP/DOWN key. If you want to
control unselected one, change the setting in sel o f VOLUME Menu. I means FHR I and
II means FHR II in sel
(1) Sound volume of selected FHR I and FHR II is controlled by UP/DOWN key.
(2) While printing, you can t use the Menu. Therefore, if you want to change the selection( FHR I
--> FHR II or FHR II --> FHR I), press the Menu key. Then, the selection is changed ( FHR I
--> FHR II or FHR II --> FHR I).
1) Alarm sound is ready to ring in the initial setting directly after turning on the Power. You can
see the bell symbol on the right of LCD.
2) If the measured Heart Rate is beyond the range of the alarm limit, the alarm rings. In this case,
you can stop the alarm to avoid the continuous ringing. If you press ALARM key, the bell
symbol with drawing X appears on the right of LCD and alarm stops ringing.
3) If you want to ring the alarm again, press ALARM k e y. Then, the bell symbol without X
appears and alarm is ready to ring.
5.12 SOUND
(1) DOPPLER : Doppler sound by the heart-beating of the fetus rings. Doppler sound is
controlled by UP/DOWN key.
(2) BEEP : When monitoring FECG, the beep sound representing the period of heart-
beating rings.
(3) ALARM : If the measured Heart Rate is beyond the range of the alarm limit, the
alarm sound( Bee-Bee-Bee ) rings.
(4) INFORMATION : Information sound( Ding-Dong ) rings when the probe comes off from
the connector during monitoring, if the battery is low, the grip of Printer is
lifted up or the paper is used up.
Step 1) If you press the Menu key on the initial screen, the Menu bar appears on the right of LCD
as the below figure. If you press the Menu key again, the screen returns to the initial
screen. In the Menu screen and setup screen, the Menu key functions as the hot key for the
screen to return to the initial screen.
Step 2) Move the triangular indicator point to the item in which you want to change a setting by
the UP/DOWN key.
Step 3) If you press Record key in the item to change a setting, the setup screen appears.
Step 4) If you press the Record key in the EACAPE item, the screen returns to the initial screen.
The below table illustrates the factory setup value which is set at the time of taking the
equipment out of the factory.
Step 1) Make the triangular indicator point to the Alarm Item on the Menu screen, and press the
Record key. Then, the alarm setup screen appears.
Step 2) Make the triangular indicator point the setup Item(HI, LO) on the setup screen with the
UP/DOWN key, and press the Record Key. Then, the set value is changed to the reverse image.
Step 3) Change the setting value with the UP/DOWN key.
Step 4) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
Step 5) To return to the Menu screen, select the ESCAPE item( ). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
Step 2) Make the triangular indicator point to the setup Item(SPD, CON, PRD, RNG, +20) on the
setup screen with the UP/DOWN key, and press the Record key. Then, the set value is
changed to the reverse image.
Step 3) Change the setting value with the UP/DOWN key.
Step 4) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
Step 5) To return to the Menu screen, select the ESCAPE item( ). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
A user can set the year, month, date, hour and minutes.
Item DESCRIPTION Range
YEA Year (1900~2099) 1900 ~ 2099
MON Month 01~12
DAT Date 01~31
HOU Hour 00~23
MIN Minute 00~59
Step 1) Make the triangular indicator point to the Time Item on the Menu screen, and press the
Record key. Then, the time setup screen appears.
Step 2) Make the triangular indicator point the setup Item(YEA, MON, DAT, HOU, MIN) on the
setup screen with the UP/DOWN key, and press the Record key. Then, the set value is
changed to the reverse image.
Step 3) Change the setting value with the UP/DOWN key.
Step 4) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
Step 5) To return to the Menu screen, select the ESCAPE item( .). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
A user can set the channel(FHR I, FHRII) to control the sound volume.
Step 1) Make the triangular indicator point the Volume Item on the Menu screen, and press the
Record key. Then, the time setup screen appears.
Step 2) Make the triangular indicator point the setup Item(SEL, INFOR) on the setup screen with
the UP/DOWN key, and press the Record Key. Then, the set value is changed to the
reverse image.
Step 3) Change the setting value with the UP/DOWN key.
Step 4) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
VOLUME
sel : I
infor :
on
Step 5) To return to the Menu screen, select the ESCAPE item( ). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
While printing, you can t use the Menu. Therefore, if you want to change the selection( FHR
I --> FHR II or FHR II --> FHR I), press the Menu Key. Then, the selection is changed ( FHR
I --> FHR II or FHR II --> FHR I).
A user can select the probe of FHR II. The FHR II area displays the HR from Doppler 2 or FECG.
Therefore, a user must select the probe of FHR II to use. The FHR I area displays the HR from
only Doppler 1.
Step 1) Make the triangular indicator point to the channel Item on the Menu screen, and press the
Record key. Then, the channel setup screen appears.
Step 2) Make the triangular indicator point the setup Item(FHR II) on the setup screen with the
UP/DOWN key, and press the Record key. Then, the set value is changed to the reverse
image.
Step 3) Change the setting value with the UP/DOWN key.
Step 4) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
Step 5) To return to the Menu screen, select the ESCAPE item( ). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
The central monitoring function of BFM-800 is the user s option. If a user wants to use the
central function, the user must install the DTE_TCP Ethernet Card from our supplier in this
equipment and purchase the central monitoring S/W. At the time of purchasing a BFM-800, if a
user purchases a BFM-800 with the central function option, our company installs the DTE_TCP
Ethernet Card in the equipment.
Check whether the DTE_TCP Card for the central monitoring is installed. You can check it in the
Menu screen. If the DTE_TCP has not been installed, the central function symbol item is not
displayed on the Menu screen, as in the below figure.
If the DTE_TCP has been installed, you can see the central function symbol item on the Menu
screen as the figure below.
42 Central Function(Option)
BFM-800
A user must set the DTE_TCP to connect BFM-800 with the Central Monitor.
Step 1) Make the triangular indicator point to the central function Item on the Menu screen, and
press the Record key. Then, the central function setup screen appears.
Step 2) Make the triangular indicator point the setup Item(IP, GW, SM, IPR) on the setup screen
with the UP/DOWN key, and press the Record key. Then, the set value is changed to the
reverse image.
Step 3) Make the triangular indicator point to the number to change on the setup screen with the
UP/DOWN key, and press the Record key. Then, the set value is changed to the reverse image.
Step 5) Press Record key to store the setting value. (The setting value is remembered even if the
power is turned off.)
Step 6) To return to the Menu screen, select the ESCAPE item( ). If you don t need to change
other Menu items, press the Menu key to return to the initial screen.
Central Function(Option) 43
FETAL MONITOR
BFM-800 can be operated by both AC and DC power. In addition to the standard AC power
source, BFM-800 can be operated while on the move by getting the power from DC battery.
When AC power is supplied to the equipment, the green LED AC power light on the lower right
of the front panel is turned on. And when the equipment is not in use, it is automatically put into
auto recharge mode.
When the equipment is turned on while AC power is cut off, the power is supplied from the
battery and the orange LED battery power light on the lower right of the front panel is turned on.
- In case battery power is running low: When the battery capacity goes below 11.2V, the
information sound will ring and the low battery message will be displayed and the battery-
shaped icon will blink for a few minutes, and then the power will eventually turn of f . Therefore,
AC power has to be supplied soon enough in order to prevent the power shortage.
- Recharging period : Longer than 15 hours
- Continuous operation time per full recharge : 1 hours
BFM-800 uses a lead acid battery pack. The shelf lives shown on this table are the life spans when
they are not usedat all.
44 Power Supply
BFM-800
Battery Replacing
The worn-out battery has to be replaced with the same type of battery,
- Type : VT1202 ( Lead-acid battery)
- Manufacturer : VOLTA Industry Co., Ltd.
Replace damaged or leaking batteries immediately. Do not use a damaged battery on the
equipment. When AC power is supplied to the equipment, the battery is recharged automatically.
The battery cannot be recharged while being separated from the equipment.
Refer to authorized local dealer or BIONICS for the periodical checking or replacement of
internal battery.
Power Supply 45
FETAL MONITOR
CHAPTER 9 Troubleshooting
This chapter explains the solutions to simple problems. One might feel lost sometimes while
operating the equipment. You don t need to call the service Dept. every time you have a problem.
It might be as simple as an operational problem. Before you call for help, please check the
following points.
System
Problem Diagnosis/Solution
Problem Diagnosis/Solution
46 Troubleshooting
BFM-800
During Printing
Problem Diagnosis/Solution
When “Door Open!” message Check whether the grip of printer is lowered. If the
appears on massage area. grip has been lifted, lower the grip.
Problem Diagnosis/Solution
When “FAIL” message appears for - contact our Customer Service Dept. or A/S
various initial check. Receptions.
Troubleshooting 47
FETAL MONITOR
Monitor Specifications
Power Specifications
Power Requirements : 100 - 240VAC, 50/60Hz, about 80VA
Power Fail Protection
Battery Specifications : Sealed Rechargeable Battery (12V 2Ah)
Operation Time : about 1 hours
Charge Time (after connection to AC power) : about 15 hours
48 Product Specifications
BFM-800
UC
External Type
Frequency Response : DC ~ 0.5 Hz
Reference(Zero) Control
Measurement Range : 0 ~ 99 units
FECG
FHR range : 50 ~ 240 bpm
Input Impedance : >10 M ohm
Input Circuit Current : Less than 50uA
Baseline Recovery Time : Less than 5 seconds
Input Signal Range : 5mVpeak(DC offset : 300mVpeak)
Bandwidth : 0.5 to 35Hz
Printer
Thermal Array Type
Print Speed : 1, 2, 3 cm/min
Print Contrast : 1, 2 ,3, 4 steps
Auto Print Period : 10, 20, 30, 40, 50, 60, off min
FHR Print Area Zoom Function
FHR II Offset Function
Paper Feeding Function
Display
Graphic LCD
Resolution : 128 X 64 pixels
Viewing Area : 83 X 18.6 mm
LED Backlight
3 Channels (HR I, HR II, UC)
Indicators
Heart Rhythm
Signal Quality
Alarm on/off State
Printing State
Operation Channel
Volume State
Low Battery State
Clock Display
Battery On (green LED)
Product Specifications 49
FETAL MONITOR
Sound
Doppler Sound with Volume Control
FECG beep sound with Volume Control
Information Sound
Alarms Sound
External Link
Ethernet (Central) : Option
Standard Accessories
- Ultrasound Doppler Probe 2 ea
- UC Probe 1 ea
- Event Marker Jack 1 ea
- Print Paper 2 ea
- Power cord 1 ea
- Fuse(1A, 250V) 2 ea
- Ultrasound Gel 1 ea
- Probe Belt 3 ea
- Accessory Case 1 ea
- Probe Holder 1 ea
- Wall Mounted Panel 1 ea
- Ground Cable 1 ea
- Operation Manual 1 ea
Optional Accessories
- FECG Probe
- Cart
- LAN CABLE
50 Product Specifications
BFM-800
Product Warranty
Product Specifications 51
FETAL MONITOR
EC Declaration of Conformity
0470
Manufacturers Name : BIONICS Corp.
Manufacturers Address : 214-3, Sangoan-Ri, Hongchun-Eub,
Hongchun-Gun, Kangwon Do
Korea
93/42/EEC
“Council Directive of 14 June 1993 on the approximation of the laws of the Member States concerning
medical devices.”
The technical documetation and full compliance with the standard(s) or other normative documents
listed below proves the conformity with the essential requirements of the above-mentioned EC Directive
:
Supplementary Information
The devices are Class IIb.
The devices were tested in a typical configuration of a BIONICS BFM-800 Fetal monitor.
52 Product Specifications