Professional Documents
Culture Documents
Daryl W. Hanson
Process Consulting Services Inc., Houston, Texas
Norm P. Lieberman
Process Improvement Engineering, Metairie, LA
T
roubleshooting refinery vacuum unit Column internals generate pressure drop mmHg absolute, most digital gauges read a
revamps(Photo 1) begins with accu- if they are functioning properly, but so will few mmHg low. Special high precision
rate differential pressure measure- liquid level. Because pressure drop gener- gauges are required if accurate pressure
ments. Specific pressure measurement ele- ated by packed column internals is very readings are required. When measuring
vations and detail drawings of the column low, two gauges must be used with the pressure drops, this is not a problem.
internals are essential to properly interpret readings taken simultaneously. To ensure
the data. Without complete information there is no offset between the two instru- Rules of Thumb
wrong conclusions can be drawn from the ments, they should both be connected to Column internals generate pressure drop
differential pressure measurements. Three the same process point and each should and so does liquid head. Properly func-
case studies will show how accurate field have the same reading. By measuring dif- tioning stripping section trays will generate
measured differential pressure can be ferential pressure simultaneously the read- between 3.5-5 mmHg per tray. Liquid level
interpreted incorrectly. These cases present ings are not affected by normal pressure also produces P, therefore 10” of 0.75
only a few of the ways to misinterpret data. variations. specific gravity liquid generates approxi-
When measuring pressures less than 10 mately 13-14 mmHg. Since packing gener-
Measuring Low Pressure
Being able to accurately measure pressures
under deep vacuum requires proper equip-
ment. Acceptable measuring devices are
absolute pressure mercury manometers or
absolute pressure electronic devices.
Moreover, tubing from the instrument to
the process connection must not leak or
measured pressure will be higher than true
process pressure. To test for leaks the pro-
cess isolation valve should be first opened
so that the measuring device reads the
process pressure, then the valve must be
blocked in. If the measured pressure grad-
ually increases then the tubing is leaking, if
it remains constant there is no leak.
If the pressure does not fall to a steady
value within a few seconds, then the con-
nection is partially plugged. The steady
state reading will then read high because a
very minor air leak cannot be completely
off-set by a relatively larger opening into
the connection. Photo 1. Vacuum unit
Flash 25 Measured
760
zone
Measured
16 during survey
Figure 2. Pressure measurement at the upper-level tap on the slop Figure 3. Measured pressures above and below the stripping
wax collector, below the wash bed and in the vapor section
space above
on the pressure tap. Using this reading and a short bed, black HVGO product normal operation measured at 60% of
alone, will lead the troubleshooter to a pointed to coke. level. Vapor flowing through the risers
conclusion that the wash zone packing In addition, the pressure survey mea- entrained slop wax that contained vacuum
and overflash tray have high pressure sured 15 mmHg across the wash bed. High tower bottoms from flash zone entrain-
drop, but this is often not so when the P indicated that coke was likely causing ment. Also when tray level was lowered,
refiner opens the tower. The lesson is clear: the black HVGO. Figure 2 shows pressure HVGO product color became a translucent
a pressure tap immediately below the wash was measured at the upper level tap on the dark green. The packed bed was not coked
zone packing should be used to measure slop wax collector, below the wash bed and (Photo 4)!
critical wash zone differential pressure. in the vapor space above. Furthermore,
slop wax tray level measured 60% which Case #3:
Case #2: indicated the upper instrument tap was in Low HVGO Product Yield-
Black HVGO Product –“High” the vapor space. But, slop wax collector “Normal” Stripping Section
Wash Section Pressure Drop trays have no level glass to confirm the Pressure Drop
Black HVGO product contained 1.6 wt % instrument reading is a true level and these HVGO product yield had dropped by
microcarbon residue that increased FCC instruments often do not work. Therefore, 4,000 bpd following start-up even though
catalyst consumption by more that 4 pressure was measured in the flash zone as the same crude rates and blends were
tons/day. Black HVGO is a symptom of a check of slop wax tray liquid level. being processed. Problems with column
wash bed coking, yet the refiner had no Measured flash zone pressure was only boot level instrumentation were common
history of coking. Moreover, flash zone 21 mmHg versus the 31 mmHg measured during start-up and throughout normal
temperature was only 730°F and the wash on the slop wax tray. Level on the tray was operation. High liquid level is the most
bed packing depth was only 4 feet. reduced until the instrument reading was common cause of stripping section tray
Experience shows it is nearly impossible to 15%, at the same time the measured pres- damage. Yet in the most recent turnaround
form coke in short crude vacuum column sure decreased from 31 to 18 mmHg. Thus, no tray damage was found. Until the refin-
(versus visbreaker or hydrocracker) wash the initial pressure reading on the tray er switched to heavier crude, very little
zones. Because entrainment from the flash included 10 inches of liquid, which emphasis was placed on stripping section
zone and wash liquid from the top keep the increased the pressure by 13 mmHg. performance. But as crudes get heavier,
packing sufficiently wetted to avoid “dry- Further review showed that the top of the stripping section performance plays an
out” in short packed beds and operating collector tray risers were 10 inches higher increasing important role in meeting
temperature was only 730°F, coking was than the instrument top tap. Consequently, HVGO product yield targets. Properly
not likely. Yet in spite of low temperature slop wax was overflowing the riser during functioning stripping trays generate 3-5