Professional Documents
Culture Documents
1-127-94
Hartmut Schmidt
1
© 1994 by Drägerwerk AG, Lübeck, Germany
All rights reserved. No part of this brochure may by reproduced or copied for public
use by any mechanical, photografic or electronic process.
2
Introduction
3
The Babylog 8000 has a different approach to pressure measure-
ment. Two pressure transducers, one at the inspiratory gas outlet and
one at the expiratory valve, measure the pressures. A mathematical
model of the tubing is used to calculate pressure at the wye. This type
of measurement has proved to be precise over the entire flow range.
It is compatible with all available humidifiers and tubing systems.
Fig. 1
4
Average (mean) airway pressure is calculated from the shaded area
under the pressure waveform.
Fig. 2
Pinsp and PEEP are the standard controls used in adjusting pressu-
res. However another parameter, the insipiratory flow rate, has an
often underestimated influence on airway pressure. Flow rate set-
tings effect the ability of the ventilator to attain desired levels of peak
inspiratory pressure, desired wave forms, I:E ratios, and, in some
cases, respiratory rate [1].
5
Setting Lower Flow Rates
Setting lower flow rates can effect tidal volumes and therefore CO2
removal. They can also effect mean airway pressure and therefore
oxygenation. Lower flow rates result in a slower increase in airway
pressure, depicted as a triangular (sloping) pressure wave form on
the Babylog display screen. This results in a sinusoidal gas flow into
the patient, better resembling a spontaneous breathing pattern. If
short inspiratory times are used, Ti may not be long enough to reach
the set peak pressure limit. In this case, tidal volumes will be impaired
which may result in hypercapnia. At the same Peak pressure setting,
a sloping wave form (as opposed to a square wave) will result in a
decreased area under the pressure wave form, and therefore a lower
mean airway pressure. This may result in a decrease in PaO2. It has
been proposed on theoretical grounds that less barotrauma may be
associated with ventilation using this curve[1 ].
Fig. 3
Fig. 4
Fig. 5
The measured PiP indicates the maximum pressure that has occured
during the last ventilatory cycle. It is normally below or equal to the set
maximum PIP. It may be higher than the set pressure limit, when a
patient has hick-ups, is coughing or crying .
PEEP / CPAP
8
As PEEP is so important, the PEEP effective at the wye piece should
equal the set PEEP value. High flows rates, and narrow tubing may
create an artificial PEEP, so called inadvertent PEEP. Unlike many
other ventilators, the PEEP of the Babylog 8000 is not flow depen-
dent. The Babylog 8000 controls the PEEP with a servo loop and
automatically compensates for any inadvertent PEEP, even at high
flow rates of 30 L/min. This feature is important because with PEEP
independent of flow, the flow need not always be set to the usual
8 L/min. Instead flow can be used freely to control the shape of the
pressure waveform.
Pressure-time Curves
In addition to measured pressure values, dynamic information is often
required. Of particular interest are the pressure rise time and the
length of the pressure limited plateau. At higher breath rates this
information can hardly be obtained from a standard pressure gauge.
Pressure time curves present this information very clearly.
9
Time Cycled Breaths
Without any trigger mechanism involved, an inspiration is started,
when the set expiratory time, Te has elapsed. These mandatory
strokes can be identified as strokes that have no negative deflection
before the start of inspiration.
Fig. 6
10
It takes longer for pressure to reach the pressure limit if a small flow
rate has to fill up a large lung compliance. A high flow rate will quickly
fill up a low lung compliance increasing pressure quickly, resulting in
a shorter pressure rise time.
Fig. 7
Fig. 8
Fig. 9
Fig. 10
12
Influence of the Humidifiers and Tubing Systems
The flow delivered by the ventilator ventilates both the patient and the
tubing system. If the compressible volume of the humidifier and of the
tubing is high, less flow is delivered to the patient. This situation is
indicated by a lower pressure rise during inspiration. Different humi-
difiers and tubing systems require different flow settings to achieve a
particular pressure waveform. Furthermore, increased compressible
volumes may occur as the water level falls in the humidification
chamber. Fig. 10 shows the resultant pressure waveforms produced
from using two different humidifiers at the same flow rate setting. Note
the differences in Mean Airway Pressure.
13
Spontaneous Breaths
Small fluctuations in pressure represent a patient’s spontaneous
breathing efforts. Negative pressure represents inspiration, while
positive pressure represents exhalation. Due to the continuous flow
system and the microprocessor controlled servo loop for the pressu-
re, these fluctuations are small with the Babylog 8000. They are often
hardly detectable on the graphics screen. Spontaneously breathing
patients with higher flow requirements may cause bigger fluctuations
if the flow rate is set too low (see Fig.13). In this case increase the flow
rate.
Fig. 13
Conclusion
This handbook shows some applications of the pressure time curves.
It is obvious, that the pressure real time graphics teaches us a great
deal about patient and ventilator interaction. Nevertheless, we have
to keep in mind, that pressure is only the driving force for flow and
volume. Detailed feedback on the patients pulmonary situation requi-
re assessment of flow waveforms and volume measurements [4].
14
References
[1] William W. Fox, Alan R. Spitzer, John G. Shutack. Positive
pressure Ventilation: Pressure and time cycled Ventilators. In Golds-
mith, Karotkin (eds): Assistet Ventilation of the neonate, Philadelphia,
W. Saunders Co. 1988.
[3] Perez-Fontan JJ, Heldt GP, Gregory GG. The effect of a gas leak
around the endotracheal tube on the mean tracheal pressure during
mechanical ventilation.
Am Rev Respir Dis 1985; 132:339-342.
15
Drägerwerk
Aktiengesellschaft
Germany
" Moislinger Allee 53 – 55
D-23542 Lübeck
§ (49) 4 51 - 8 82 36 86
$ 2 6 807 - 0
16