Professional Documents
Culture Documents
CHAPTER
How a Breath Is Delivered
OUTLINE
Basic Model of Ventilation in the Lung During Inspiration Beginning of Inspiration: The Trigger Variable
Factors Controlled and Measured During Inspiration The Limit Variable During Inspiration
Pressure-Controlled Breathing Termination of the Inspiratory Phase: The Cycling Mechanism
Volume-Controlled Breathing (Cycle Variable)
Flow-Controlled Breathing Expiratory Phase: The Baseline Variable
Time-Controlled Breathing Types of Breaths
Overview of Inspiratory Waveform Control Summary
Phases of a Breath and Phase Variables
KEY TERMS
• Baseline variable • Negative end-expiratory pressure • Pressure triggering
• Continuous positive airway pressure • Patient triggering • Spontaneous breaths
• Control variable • Phase variable • Time cycled
• Cycle variable • Plateau pressure • Time triggering
• Flow cycling • Positive end-expiratory pressure • Trigger variable
• Flow limited • Pressure control • Volume cycled
• Flow triggering • Pressure cycling • Volume limiting
• Limit variable • Pressure limiting • Volume triggering
• Mandatory breath • Pressure support
LEARNING OBJECTIVES On completion of this chapter, the reader will be able to do the following:
1. Write the equation of motion, and define each term in the equation. 7. Recognize the effects of a critical leak (e.g., a patient disconnect)
2. Give two other names for pressure ventilation and volume on pressure readings and volume measurements.
ventilation. 8. Define the effects of inflation hold on inspiratory time.
3. Compare pressure, volume, and flow delivery in volume-controlled 9. Give an example of a current ventilator that provides negative
breaths and pressure-controlled breaths. pressure during part of the expiratory phase.
4. Name the two most commonly used patient-trigger variables. 10. Based on the description of a pressure–time curve, identify a
5. Identify the patient-trigger variable that requires the least work of clinical situation in which expiratory resistance is increased.
breathing for a patient receiving mechanical ventilation. 11. Describe two methods of applying continuous pressure to the
6. Explain the effect on the volume delivered and the inspiratory airways that can be used to improve oxygenation in patients with
time if a ventilator reaches the set maximum pressure limit during refractory hypoxemia.
volume ventilation.
S
electing the most effective mode of ventilation to use once it factors determine the mode of ventilation, and each of these con-
has been decided that a patient will require mechanical ven- cepts is reviewed in this chapter.
tilation requires an understanding of how a ventilator works.
Answers to several questions can help explain the method by which BASIC MODEL OF VENTILATION IN THE LUNG
a ventilator accomplishes delivery of a breath: (1) What is the DURING INSPIRATION
source of energy used to deliver the breath (i.e., is the energy pro-
vided by the ventilator or by the patient)? (2) What factor does the One approach that can be used to understand the mechanics of
ventilator control to deliver the breath (e.g., pressure, volume, flow, breathing during mechanical ventilation involves using a mathe-
or time)? (3) How are the phases of a breath accomplished (i.e., matical model that is based on the equation of motion. This equa-
what begins a breath, how is it delivered, and what ends the breath)? tion, which is shown in Box 3-1, describes the relationships among
(4) Is the breath mandatory, assisted, or spontaneous? All these pressure, volume, and flow during a spontaneous or mechanical
27
28 CHAPTER 3 How a Breath Is Delivered
Common Methods
BOX 3-1 Equation of Motion BOX 3-2 of Delivering Inspiration
Pmus + Pvent = PE + PR Pressure-Controlled Ventilation
The clinician sets a pressure for delivery to the patient. Pressure-
where controlled ventilation is also called
Muscle pressure + Ventilator pressure = Elastic recoil • Pressure-targeted ventilation
pressure + Flow resistance pressure • Pressure ventilation
If one considers that
Volume-Controlled Ventilation
Elastic recoil pressure = Elastance × Volume The clinician sets a volume for delivery to the patient. Volume-
= Volume/Compliance ( V/C), and controlled ventilation is also called
• Volume-targeted ventilation
Flow resistance pressure = Resistance × Flow = (R aw × V )
• Volume ventilation
Then the equation can be rewritten as follows:
Pmus + Pvent = V/C + (R aw × V )
Pmus is the pressure generated by the respiratory muscles that a number of combinations of Pmus and Pvent can be used to
(muscle pressure). If these muscles are inactive, Pmus = 0 cm H2O,
achieve the total force required during assisted ventilation.)
then the ventilator must provide the pressure required to
The right side of the equation in Box 3-1 represents the imped-
achieve an inspiration.
ance that must be overcome to deliver a breath and can be expressed
Pvent, or more specifically, PTR, is the pressure read on the
as the alveolar pressure (Palv) and the transairway pressure (PTA).
ventilator gauge (manometer) during inspiration with positive
pressure ventilation (i.e., the ventilator gauge pressure). Palv is produced by the interaction between lung and thoracic com-
V is the volume delivered, C is respiratory system compli- pliance and the pressure within the thorax. PTA is produced by
ance, V/C is the elastic recoil pressure, Raw is airway resistance, resistance to the flow of gases through the conductive airways
and V is the gas flow during inspiration (Raw × V = Flow resis- resistance = PTA/flow).
tance). It is important to recognize that various combinations
of Pmus + Pvent are used during assisted ventilation. FACTORS CONTROLLED AND MEASURED
Because Palv = V/C and PTA = Raw × V, substituting in the
DURING INSPIRATION
above equation results in
Delivery of an inspiratory volume is perhaps the single most
Pmus + PTR = Palv + PTA
important function a ventilator accomplishes. Two factors deter-
Where Palv is the alveolar pressure and PTA is the transairway mine the way the inspiratory volume is delivered: the structural
pressure (peak pressure minus plateau pressure [PIP − Pplateau]) design of the ventilator and the ventilator mode set by the clinician.
(see Chapter 1 for further explanation of abbreviations). The clinician sets the mode by selecting either a predetermined
pressure or volume as the target variable (Box 3-2).
The primary variable the ventilator adjusts to achieve inspira-
tion is therefore called the control variable (Key Point 3-1).6 As the
equation of motion shows, the ventilator can control four variables:
breath.1-4 The equation includes three terms, which were previously pressure, volume, flow, and time. It is important to recognize that
defined in Chapter 1, namely, PTR, or transrespiratory pressure; PE, the ventilator can control only one variable at a time. Therefore it
or elastic recoil pressure; and PR, or flow resistance pressure. Figure must operate as a pressure controller, a volume controller, a flow
3-1 provides a graphic representation of each of these pressures.5 controller, or a time controller (Box 3-3).
Notice that energy (i.e., pressure) required to produce motion
(i.e., flow) can be achieved by contraction of the respiratory muscles
(Pmus) during a spontaneous breath, or generated by the ventilator Key Point 3-1 The primary variable that the ventilator adjusts to
(Pvent) during a mechanical breath. In both cases, the total amount produce inspiration is the control variable. The most commonly used control
of pressure that must be generated to produce the flow of gas into variables are pressure and volume.
the lungs depends on the physical characteristics of the respiratory
system (i.e., elastance or, more specifically, compliance of the lungs
and chest wall, plus the amount of airway resistance [Raw] that must Pressure-Controlled Breathing
be overcome). When the ventilator maintains the pressure waveform in a specific
If the respiratory muscles are inactive, then the ventilator must pattern, the breathing is described as pressure-controlled. With
perform all of the work required to move air into the lungs. The pressure-controlled ventilation, the pressure waveform is unaf-
pressure generated by the ventilator therefore represents the trans fected by changes in lung characteristics. The volume and flow
respiratory pressure (PTR), that is, the pressure gradient between waveforms will vary with changes in the compliance and resistance
the airway opening and the body’s surface. For example, during characteristics of the patient’s respiratory system.
positive pressure ventilation, the pressure delivered at the upper
airway is positive and the pressure at the body surface is atmo- Volume-Controlled Breathing
spheric (ambient pressure, which is given a value of 0 cm H2O). When a ventilator maintains the volume waveform in a specific
Keep in mind that PTR represents the pressure gradient that must pattern, the delivered breath is volume-controlled. During volume-
be generated to achieve a given flow. (It is important to recognize controlled breathing, the volume and flow waveforms remain
How a Breath Is Delivered CHAPTER 3 29
Transairway pressure
Resistance =
Flow
Flow
Transairway
pressure
Transrespiratory
pressure
Volume Transthoracic
pressure
Fig. 3-1 Equation of motion model. The respiratory system can be visualized as a conductive tube connected to an elastic
compartment (balloon). Flow, volume, and pressure are variables and functions of time. Resistance and compliance are constants.
Transthoracic pressure is the pressure difference between the alveolar space (PA), or lung, and the body surface (Pbs). (See text for
further explanation.) (From Kacmarek RM, Stoller JK, Heuer AJ, editors: Egan’s fundamentals of respiratory care, ed 10, St Louis,
2013, Elsevier-Mosby.)
Yes
No
No
Fig. 3-2 Defining a breath based on how the ventilator maintains the inspiratory waveforms. (Modified from Chatburn RL:
Classification of mechanical ventilators, Respir Care 37:1009-1025, 1992.)
oscillators control time (both inspiratory and expiratory); however, Basic Points for Evaluating a Breath
distinguishing between inspiration and expiration during high- BOX 3-4 During Inspiration
frequency ventilation can be difficult. Time-controlled ventilation is
1. Inspiration is commonly described as pressure-controlled or
used less often than pressure- and volume-controlled ventilation.
volume-controlled. Although both flow- and time-controlled
ventilation have been defined, they are not typically used.
OVERVIEW OF INSPIRATORY 2. Pressure-controlled inspiration maintains the same pattern
WAVEFORM CONTROL of pressure at the mouth regardless of changes in lung
condition.
Figure 3-2 provides an algorithm to identify the various types of 3. Volume-controlled inspiration maintains the same pattern
breaths that can be delivered by mechanical ventilators. Figure 3-3 of volume at the mouth regardless of changes in lung
shows the waveforms for pressure- and volume-controlled ventila- condition and also maintains the same flow waveform.
tion, and Box 3-4 lists some basic points that can help simplify 4. The pressure, volume, and flow waveforms produced at
evaluation of a breath during inspiration.6 the mouth usually take one of four shapes:
The airway pressure waveforms shown in Fig. 3-3 illustrate what a. Rectangular (also called square or constant)
the clinician would see on the ventilator graphic display as gas is b. Exponential (may be increasing [rising] or decreasing
delivered. The ventilator typically measures variables in one of [decaying])
three places: (1) at the upper, or proximal, airway, where the patient c. Sinusoidal (also called sine wave)
is connected to the ventilator; (2) internally, near the point where d. Ramp (available as ascending or descending
the main circuit lines connect to the ventilator; or (3) near the [decelerating] ramp)
exhalation valve.*
Microprocessor-controlled ventilators have the capability of
displaying these waveforms as scalar graphs (a variable graphed
relative to time) and loops on a screen.6 Most current generation PHASES OF A BREATH AND PHASE VARIABLES
ventilators, such as the Dräger Infinity V500 and the CareFusion
AVEA, have built-in screens. As discussed in Chapter 9, this The following section describes the phases of a breath and the vari-
graphic information is an important tool that can be used for the able that controls each portion of the breath (i.e., the phase vari-
management of the patient-ventilator interaction. able). As summarized in Box 3-5, the phase variable represents the
signal measured by the ventilator that is associated with a specific
*Newer ventilators often have a pressure-measuring device on both the aspect of the breath. The trigger variable begins inspiration. The
inspiratory and expiratory sides of a ventilator circuit. limit variable limits the value of pressure, volume, flow, or time
How a Breath Is Delivered CHAPTER 3 31
Time (s)
Pressure
(elastic)
Volume
Volume
Pelastic =
Compliance
(resistive)
Pressure
Presistive = Resistance × Flow
Flow
Fig. 3-3 Characteristic waveforms for pressure-controlled ventilation and volume-controlled ventilation. Note that the volume
waveform has the same shape as the transthoracic (lung pressure) waveform (i.e., pressure caused by the elastic recoil
[compliance] of the lung). The flow waveform has the same shape as the transairway pressure waveform (peak pressure minus
plateau pressure [PIP − Pplateau]) (shaded area of pressure–time waveform). The shaded areas represent pressures caused by
resistance, and the open areas represent pressure caused by elastic recoil. (From Kacmarek RM, Stoller JK, Heuer AJ, editors:
Egan’s fundamentals of respiratory care, ed 10, St Louis, 2013, Elsevier-Mosby.)
Peak pressure
3 seconds 3 seconds
Fig. 3-4 Controlled ventilation pressure curve. Patient Airway
pressure
effort does not trigger a mechanical breath; rather, Machine
inspiration occurs at equal, timed intervals. breath
No patient Time
inspiratory effort
Patient’s
inspiratory Time
effort
the ventilator is based on the length of the TCT. For example, if the
breathing rate is set at 20 breaths per minute, the ventilator triggers 40
inspiration after 3 seconds elapses (60 s/min divided by 20 breaths/ 35
min = 3 seconds).
30
In the past, time-triggered ventilation did not allow a patient to
initiate a breath (i.e., the ventilator was “insensitive” to the patient’s 25
(cm H2O)
Pressure
Patient Triggering machine is to the patient’s effort. For example, the ventilator is
In those cases where a patient attempts to breathe spontaneously more sensitive to patient effort at a setting of −0.5 cm H2O than at
during mechanical ventilation, a ventilator must be available to a setting of −1 cm H2O. Sensing devices usually are located inside
measure the patient’s effort to breathe. When the ventilator detects the ventilator near the output side of the system; however, in some
changes in pressure, flow, or volume, a patient-triggered breath systems, pressure or flow is measured at the proximal airway.
occurs. Pressure and flow are common patient-triggering mecha- The sensitivity level for pressure triggering usually is set at
nisms (e.g., inspiration begins if a negative airway opening pres- about −1 cm H2O. The clinician must set the sensitivity level to fit
sure or change in flow is detected). Figure 3-5 illustrates a breath the patient’s needs. If it is set incorrectly, the ventilator may not be
triggered by the patient making an inspiratory effort (i.e., the sensitive enough to the patient’s effort, and the patient will have to
patient’s inspiratory effort can be identified as the pressure deflec- work too hard to trigger the breath (Fig. 3-6). If the ventilator is
tion below baseline that occurs prior to initiation of the mechanical too sensitive, it can autotrigger (i.e., the machine triggers a breath
breath). To enable patient triggering, the clinician must specify without the patient making an effort) (Case Study 3-1).
the sensitivity setting, also called the patient effort (or patient- Flow triggering occurs when the ventilator detects a drop in
triggering) control. This setting determines the pressure or flow flow through the patient circuit during exhalation. To enable flow
change that is required to trigger the ventilator. The less pressure triggering, the clinician must set an appropriate flow that must be
or flow change required to trigger a breath, the more sensitive the sensed by the ventilator to trigger the next breath. As an example,
How a Breath Is Delivered CHAPTER 3 33
a ventilator has a baseline flow of 6 L/min. This allows 6 L/min of has the option of breathing at a faster rate. Clinicians often refer
gas to flow through the patient circuit during the last part of exha- to this as the assist-control mode. (NOTE: The clinician must
lation. The sensors measure a flow of 6 L/min leaving the ventilator always make sure the ventilator is sensitive to the patient’s efforts
and 6 L/min returning to the ventilator. If the flow trigger is set at [Box 3-6].)
2 L/min, the ventilator will begin an assisted breath when it detects
a drop in flow of 2 L/min from the baseline (i.e., 4 L/min returning
to the ventilator [Fig. 3-7]). Case Study 3-1
When set properly, flow triggering has been shown to Patient Triggering
require less work of breathing than pressure triggering. Many
Problem 1: A patient is receiving volume-controlled ventila-
microprocessor-controlled ventilators (e.g., Servo-i, CareFusion
tion. Whenever the patient makes an inspiratory effort, the
AVEA, Hamilton G5, Covidien PB 840) offer flow triggering as an
pressure indicator shows a pressure of −5 cm H2O below
option.
baseline before the ventilator triggers into inspiration.
Volume triggering occurs when the ventilator detects a small
What does this indicate?
drop in volume in the patient circuit during exhalation. The
Problem 2: A patient appears to be in distress while
machine interprets this drop in volume as a patient effort and
receiving volume-controlled ventilation. The ventilator is
begins inspiration. The Dräger Babylog and the Cardiopulmonary
cycling rapidly from breath to breath. The actual rate is much
Venturi are volume-triggered ventilators.
faster than the set rate. No discernible deflection of the
As mentioned previously, manual triggering is also available.
pressure indicator occurs at the beginning of inspiration.
With manual triggering, the operator can initiate a ventilator
The ventilator panel indicates that every breath is an assisted,
breath by pressing a button or touch pad labeled “Manual” breath
or patient-triggered, breath. What does this indicate?
or “Start” breath. When this control is activated, the ventilator
delivers a breath according to the set variables.
It is important to recognize that patient triggering can be quite
effective when a patient begins to breathe spontaneously, but The Limit Variable During Inspiration
occasionally the patient may experience an apneic episode. For Inspiration is timed from the beginning of inspiratory flow to the
this reason, a respiratory rate is set with the rate control to guar- beginning of expiratory flow. As mentioned previously, the ventila-
antee a minimum number of breaths per minute (Fig. 3-8). Each tor can determine the waveform for pressure, volume, flow, or time
breath is either patient triggered or time triggered, depending on during inspiration. However, it also can limit these variables. For
which occurs first. Although the rate control determines the example, during volume-controlled ventilation of an apneic patient,
minimum number of mechanical breaths delivered, the patient the clinician sets a specific volume that the ventilator will deliver.
In
Flow measuring
devices Patient
connection
Out
Fig. 3-7 Schematic drawing of the essential features of flow triggering. Triggering occurs when the patient inspires from the
circuit and increases the difference between flow from the ventilator (inspiratory side, in) and flow back to the exhalation valve
(expiratory side, out). (From Dupuis Y: Ventilators: theory and clinical application, ed 2, St Louis, 1992, Mosby.)
1
In general, the volume delivered cannot exceed that amount; it may
be for some reason less than desired, but it cannot be more.
A limit variable is the maximum value that a variable (pressure,
volume, flow, or time) can attain. It is important to emphasize,
however, that reaching the set limit variable does not end inspira-
tion. As an example, a ventilator is set to deliver a maximum pres-
sure of 25 cm H2O, and the inspiratory time is set at 2 seconds. The
maximum pressure that can be attained during inspiration is
25 cm H2O, but inspiration will end only after 2 seconds has
2
passed. Such a breath therefore is described as a pressure-limited,
time-cycled breath (cycling ends inspiration [see Termination of
the Inspiratory Phase: The Cycling Mechanism section]). Fig. 3-9 Internal pneumatic circuit on a piston-driven ventilator. 1, Pressure release
valve; 2, heated humidifier. (Modified from Dupuis Y: Ventilators: theory and clinical
Pressure Limiting application, ed 2, St Louis, 1992, Mosby.)
As the example mentioned above illustrates, pressure limiting
allows pressure to rise to a certain value but not exceed it. Figure
3-9 shows an example of the internal pneumatic circuit of a piston
ventilator. The ventilator pushes a volume of gas into the ventilator during a specific interval. The clinician can set the volume of gas
circuit, which causes the pressure in the circuit to rise. To prevent that the ventilator delivers. With volume limiting, the ventilator
excessive pressure from entering the patient’s lungs, the clinician may include a bag, bellows, or piston cylinder that contains a fixed
sets a high pressure limit control. When the ventilator reaches the volume, which establishes the maximum volume of gas that can be
high pressure limit, excess pressure is vented through a spring- delivered. (NOTE: Reaching that volume does not necessarily end
loaded pressure release, or pop-off, valve (Fig. 3-9). The excess gas inspiration.) A piston-operated ventilator can be used to provide a
pressure is released into the room, just as steam is released by a simple example of volume limiting. Volume is limited to the
pressure cooker. In this example, reaching the high pressure limit amount of volume contained in the piston cylinder (see Fig. 3-9).
does not cycle the ventilator and end inspiration. The forward movement of the piston rod or arm controls the dura-
The pressure–time and volume–time waveforms shown in Fig. tion of inspiration (time-cycled breath).
3-10 illustrate how the set pressure and volume curves would Ventilators can have more than one limiting feature at a time.
appear for a patient with normal lung function and when the In the example just provided, the duration of inspiration could not
patient’s lungs are less compliant. Notice that a higher pressure is exceed the excursion time of the piston, and the volume delivered
required to inflate the stiff lungs and the pressure limit would be could not exceed the volume in the piston cylinder. Therefore a
reached before the end of the breath occurs. Consequently, the piston-driven ventilator can be simultaneously volume limited and
volume delivered would be less than desired. In other words, time limited. (NOTE: Current ventilators that are not piston driven
volume delivery is reduced because the pressure limit is reached at [e.g., Servo-i] provide a volume limit option. When special modes
Time A even though inspiration does not end until Time B (i.e., are selected, an actively breathing patient can receive more volume
the breath is time cycled). if inspiratory demand increases. The advantage of these ventilators
Infant ventilators often pressure limit the inspiratory phase is that the volume delivered to the patient during selected modes
but time cycle inspiration. Other examples of pressure-limiting is adjusted to meet the patient’s increased inspiratory needs.)
modes are pressure support and pressure-controlled ventilation.
Remember that when the clinician establishes a set value in Flow Limiting
pressure-targeted ventilation, the pressure the ventilator delivers to If gas flow from the ventilator to the patient reaches but does not
the patient is limited; however, reaching the pressure limit does not exceed a maximum value before the end of inspiration, the ventila-
end the breath. tor is flow limited; that is, only a certain amount of flow can be
provided. For example, the constant forward motion of a linear-
Volume Limiting drive piston provides a constant rate of gas delivery to the patient
A volume-limited breath is controlled by an electronically operated over a certain period. The duration of inspiration is determined by
valve that measures the flow passing through the ventilator circuit the time it takes the piston rod to move forward.
How a Breath Is Delivered CHAPTER 3 35
Pressure limit Pressure curve with a patient coughs or if there is an obstruction in the ventilator
set by operator decreased compliance tubing. Some ventilators allow inspiration to continue while excess
pressure is vented to the atmosphere through a pressure safety
valve. (In newer intensive care unit [ICU] ventilators, a “floating”
Pressure curve exhalation valve prevents pressures from abruptly rising as might
with normal occur when a patient coughs [Case Study 3-2]).
Pressure
During inspiration, positive pressure builds up in the patient With time-cycled, volume-controlled ventilation, an increase in
circuit, resulting in expansion of the patient circuit and compres- airway resistance or a decrease in compliance does not affect the
sion of some of the gas in the circuit (an application of Boyle’s law). flow pattern or volume delivery as long as the working pressures
The compressed gas never reaches the patient’s lungs. of the ventilator are adequate. Therefore volume delivery in a fixed
In most adult ventilator circuits, about 1 to 3 mL of gas is lost period remains the same, although the pressures vary. Appropriate
to tubing compressibility for every 1 cm H2O that is measured by alarms should be set to alert the clinician of any significant changes
the airway pressure sensor. As a result, a relatively large volume of in airway pressures.
gas may be compressed in the circuit and never reaches the patient’s With time-cycled, pressure-controlled ventilation, both volume
lungs when high pressures are required to ventilate a patient. Con- and flow vary. Volume (and flow) delivery depends on lung compli-
versely, a patient whose lung compliance is improving can be ven- ance and airway resistance, patient effort (if present), the inspira-
tilated at lower pressures; therefore less volume is lost to circuit tory time, and the set pressure. Time-cycled, pressure-controlled
compressibility. ventilation is commonly called pressure-controlled ventilation.
The actual volume delivered to the patient can be determined Pressure-controlled ventilation is sometimes used because the
by measuring the exhaled volume at the endotracheal tube or tra- inspiratory pressure can be limited, which protects the lungs from
cheostomy tube. If the volume is measured at the exhalation valve, injury caused by high pressures. However, the variability of tidal
it must be corrected for tubing compliance (i.e., the compressible volume delivery can be a concern. Alarm settings must be chosen
volume). To determine the delivered volume, the volume com- carefully so that the clinician is alerted to any significant changes
pressed in the ventilator circuit must be subtracted from the in the rate and volume.
volume measured at the exhalation valve. Most microprocessor-
controlled ICU ventilators (e.g., Covidien PB 840, Servo-i) measure Flow-Cycled Ventilation
and calculate the lost volume and automatically compensate for With flow-cycled ventilation, the ventilator cycles into the expira-
volume lost to tubing compressibility by increasing the actual tory phase once the flow has decreased to a predetermined value
volume delivered. For example, the Covidien PB 840 calculates the during inspiration. Volume, pressure, and time vary according to
circuit compliance/compressibility factor during the establishment changes in lung characteristics. Flow cycling is the most common
of ventilation for a new patient setup. The ventilator measures the cycling mechanism in the pressure support mode (Fig. 3-11). In
peak pressure of a breath delivered to the patient and calculates the the Covidien PB 840 ventilator, flow termination occurs when the
estimated volume loss caused by circuit compressibility. Then, for flow reaches a percentage of the peak inspiratory flow, which is
the next breath, it adds the volume calculated to the delivered set selected by the clinician. In some ventilators, the flow cycle per-
volume to correct for this loss. (Determination of the compressible centage can be adjusted from about 5% to 80%.
volume is discussed in more detail in Chapter 6.)
System leaks. The volume of gas delivered to the patient may be CRITICAL CARE CONCEPT 3-1
less than the set volume if a leak in the system occurs. The ventila-
tor may be unable to recognize or compensate for leaks, but the Early generation Bennett ventilators (Bennett PR-1 and
size of the leak can be determined by using an exhaled volume PR-2) relied on a Bennett valve to control gas flow to the
monitor. In cases where a leak exists, the peak inspiratory pressure patient. The principle of operation of these devices is the
will be lower than previous peak inspiratory pressures and a low- valve switches from the inspiratory phase to the expiratory
pressure alarm may be activated. The volume–time graph also can phase when flow to the patient drops to 1 to 3 L/min. This
provide information about leaks (see Chapter 9). lower flow results when the pressure gradient between the
alveoli and the ventilator is small and the pressures are
Time-Cycled Ventilation nearly equal. Because equal pressure is nearly achieved,
A breath is considered time cycled if the inspiratory phase ends along with the low gas flow, these machines sometimes are
when a predetermined time has elapsed. The interval is controlled called pressure-cycled ventilators. However, because the
by a timing mechanism in the ventilator, which is not affected by predetermined pressure is never actually reached, these
the patient’s respiratory system compliance or airway resistance. At ventilators were in reality examples of flow-cycled ventila-
the specified time, the exhalation valve opens (unless an inspira- tors. (NOTE: The rate control on these machines allowed
tory pause has been used) and exhaled air is vented through the them to function as time-cycled ventilators as long as flow
exhalation valve. If a constant gas flow is used and the interval is and/or pressure limits were not reached first.)
fixed, a tidal volume can be predicted:
Tidal volume = Flow (Volume/Time) × Inspiratory time
The Servo-i and Dräger Evita XL are examples of time-cycled ven- Pressure-Cycled Ventilation
tilators. These microprocessor-controlled machines can compare During pressure-cycled ventilation, inspiration ends when a set
the set volume with the set time and calculate the flow required to pressure threshold is reached at the mouth or upper airway. The
deliver that volume in that length of time. Consider the following exhalation valve opens, and expiration begins. The volume deliv-
example. A patient’s tidal volume (VT) is set at 1000 mL and ered to the patient depends on the flow delivered, the duration
the inspiratory time (TI) is set at 2 seconds. To accomplish this of inspiration, the patient’s lung characteristics, and the set
volume delivery in the time allotted, the ventilator would have pressure.
to deliver a constant flow waveform at a rate of 30 L/min (30 L/ A disadvantage of pressure-cycled ventilators (e.g., Bird Mark
60 s = 0.5 L/s), so that 0.5 L/s × 2 s would provide 1.0 L over the 7) is that these devices deliver variable and generally lower tidal
desired 2-second inspiratory time. volumes when reductions in compliance and increases in
How a Breath Is Delivered CHAPTER 3 37
plished in one of two ways. With the first method, negative pres-
50 sure was applied with a bellows or an entrainment (Venturi) device
positioned at the mouth or upper airway to draw air out of the
lungs. This technique was called negative end-expiratory pressure
(NEEP). Another method involved applying positive pressure to
25
the abdominal area, below the diaphragm. With this latter tech-
nique, it was thought that applying pressure below the diaphragm
would force the air out of the lungs by pushing the visceral organs
0 against the diaphragm (i.e., similar to the effects of performing a
Time Heimlich maneuver).
Under normal circumstances, expiration during mechanical
Inspiration ends
ventilation occurs passively and depends on the passive recoil of
Fig. 3-11 Waveforms from a pressure support breath showing the pressure and flow the lung. High-frequency oscillation is an exception to this prin-
curves during inspiration. When flow drops to 25% of the peak flow value measured ciple (this type of mechanical ventilation is discussed later in this
during inspiration, the ventilator flow cycles out of inspiration. (From Dupuis Y: chapter and in Chapters 22 and 23.)
Ventilators: theory and clinical application, ed 2, St Louis, 1992, Mosby.)
Definition of Expiration
The expiratory phase encompasses the period from the end of inspi-
ration to the beginning of the next breath. During mechanical
ventilation, expiration begins when inspiration ends, the expira-
tory valve opens, and expiratory flow begins. As previously men-
resistance occur. An advantage of pressure-cycled ventilators is that tioned, opening of the expiratory valve may be delayed if an
they limit peak airway pressures, which may reduce the damage inflation hold is used to prolong inspiration.
that can occur when pressures are excessive. These ventilators are The expiratory phase has received increased attention during
most often used to deliver intermittent positive pressure breathing the past decade. Clinicians now recognize that air trapping can
(IPPB) treatments. These devices have also been used for short- occur if the expiratory time is too short. Remember that a quiet
term ventilation of patients with relatively stable lung function, exhalation normally is a passive event that depends on the elastic
such as postoperative patients. It is important that appropriate recoil of the lungs and thorax and the resistance to airflow offered
alarms are operational to ensure that the patient is being by the conducting airways. Changes in a patient’s respiratory
adequately ventilated. Ensuring that the humidification system is system compliance and airway resistance can alter time constants,
adequate is also important. (NOTE: As mentioned previously, which in turn can affect the inspiratory and expiratory times (I:E)
pressure cycling occurs in volume-controlled breaths when the required to achieve effective ventilation. If an adequate amount of
pressure exceeds the maximum safety high-pressure limit. A high- time is not provided for exhalation, air trapping and hyperinflation
pressure alarm sounds, and the set tidal volume is not delivered can occur, leading to a phenomenon called auto-PEEP or intrinsic
[see Case Study 3-2].) PEEP (see the section on Expiratory Hold later in this chapter).
38 CHAPTER 3 How a Breath Is Delivered
Plateau pressure
Airway
pressure
Time
Fig. 3-12 Positive pressure ventilation with an inflation hold, or end-inspiratory pause, leading to a pressure plateau (Pplateau).
Peak pressure
Airway
pressure
0
Time
Fig. 3-14 Positive pressure ventilation with expiratory retard (solid line) and passive expiration to zero baseline (dashed line).
Expiratory retard does not necessarily change expiratory time, which also depends on the patient’s spontaneous pattern.
However, it increases the amount of pressure in the airway during exhalation.
EPAP
IPAP
10
Airway
pressure
Time
Fig. 3-15 Simplified pressure–time waveform showing continuous positive airway pressure (CPAP). Breathing is spontaneous.
Inspiratory positive airway pressure (IPAP) and expiratory positive airway pressure (EPAP) are present. Pressures remain positive
and do not return to a zero baseline.
above and below the baseline (i.e., atmospheric pressure). During expiratory phase. Many of these patients can accomplish a pro-
exhalation, HFO actually creates a negative transrespiratory pres- longed expiration during spontaneous breathing by using a tech-
sure. HFO is most often used for ventilation of infant lungs, nique called pursed-lip breathing. Obviously, a patient cannot use
although it has also been used occasionally to treat adult patients pursed-lip breathing with an endotracheal tube in place. To mimic
with acute respiratory distress syndrome (see Chapters 22 and 23). pursed-lip breathing, earlier ventilators provided an expiratory
adjunct called expiratory retard, which added a degree of resistance
Expiratory Hold (End-Expiratory Pause) to exhalation (Fig. 3-14). Although theoretically expiratory retard
Expiratory hold, or end-expiratory pause, is a maneuver transiently should prevent early airway closure and improve ventilation, this
performed at the end of exhalation. It is accomplished by first technique is not commonly used in clinical practice. It is important
allowing the patient to perform a quiet exhalation. The ventilator to recognize that ventilator circuits, expiratory valves, and bacterial
then pauses before delivering the next machine breath. During this filters placed on the expiratory side of the patient circuit produce
time, both the expiratory and inspiratory valves are closed. Deliv- a certain amount of expiratory retard because they cause resistance
ery of the next inspiration is briefly delayed. The purpose of this to flow. This is especially true of expiratory filters, which can accu-
maneuver is to measure pressure associated with air trapped in the mulate moisture from the patient’s exhaled air. The clinician can
lungs at the end of the expiration (i.e., auto-PEEP). check for expiratory resistance by observing the pressure manom-
An accurate reading of end-expiratory pressure is impossible to eter and the ventilator pressure–time and flow–time graphics.
obtain if a patient is breathing spontaneously. However, measure- (Increased resistance is present if pressure and flow return to base-
ment of the exact amount of auto-PEEP present is not always line very slowly during exhalation [see Chapter 9].)
necessary; simply detecting its presence may be sufficient. Auto-
PEEP can be detected in the flow curve on a ventilator that pro- Continuous Positive Airway Pressure (CPAP) and
vides a graphic display of gas flow; it is present if flow does not Positive End-Expiratory Pressure (PEEP)
return to zero when a new mandatory ventilator breath begins (see Two methods of applying continuous pressure to the airways
Chapter 9). (NOTE: A respirometer also can be used if a graphic have been developed to improve oxygenation in patients with
display is not available. The respirometer is placed in line between refractory hypoxemia: continuous positive airway pressure
the ventilator’s Y-connector and the patient’s endotracheal tube (CPAP) and PEEP.
connector. If the respirometer’s needle continues to rotate when the CPAP involves the application of pressures above ambient pres-
next breath begins, air trapping is present [i.e., the patient is still sure throughout inspiration and expiration to improve oxygen-
exhaling when the next mandatory breath occurs].) ation in a spontaneously breathing patient (Fig. 3-15). It can be
applied through a freestanding CPAP system or a ventilator. CPAP
Expiratory Retard has been used for the treatment of a variety of disorders, including
Spontaneously breathing individuals with a disease that leads to postoperative atelectasis and obstructive sleep apnea (see Chapter
early airway closure (e.g., emphysema) require a prolonged 13 for more details on the use of CPAP).
40 CHAPTER 3 How a Breath Is Delivered
20
Fig. 3-16 Positive end-expiratory pressure (PEEP) during
controlled ventilation. No spontaneous breaths are taken
Airway
between mandatory breaths, and there are no negative 10
pressure
deflections of the baseline, which is maintained above zero.
0
Time
IPAP
Time
Fig. 3-19 Criteria for determining phase variables during delivery of a breath with mechanical ventilation. (From Kacmarek RM,
Stoller JK, Heuer AJ, editors: Egan’s fundamentals of respiratory care, ed 10, St Louis, 2013, Elsevier-Mosby.)
Control Variables
• The equation of motion provides a mathematical model for
Control variables are the main variables the ventilator adjusts
to produce inspiration. The two primary control variables are describing the relationships among pressure, volume, flow, and
pressure and volume. time during a spontaneous or mechanical breath.
• The work of breathing can be accomplished by contraction of
Phase Variables
the respiratory muscles during spontaneous breathing or by the
Phase variables control the four phases of a breath (i.e., begin-
ventilator during a mechanical ventilatory breath.
ning inspiration, inspiration, end inspiration, and expiration).
Types of phase variables include • Two factors determine the way the inspiratory volume is deliv-
• Trigger variable (begins inspiration) ered during mechanical ventilation: the structural design of the
• Limit variable (restricts the magnitude of a variable during ventilator and the ventilator mode set by the clinician.
inspiration) • The primary variable that the ventilator adjusts to produce
• Cycle variable (ends inspiration) inspiration is the control variable. Although ventilators can be
• Baseline variable (the parameter controlled during volume, pressure, flow, and time controlled, the two most com-
exhalation) monly used control variables are pressure and volume.
Types of Breaths • Determining which variable is controlled can be determined
• Spontaneous breaths: Breaths are started by the patient by using graphical analysis. The control variable will remain
(patient triggered), and tidal volume delivery is constant regardless of changes in the patient’s respiratory
determined by the patient (patient cycled). characteristics.
• Mandatory breaths: The ventilator determines the start • Pressure and flow waveforms delivered by a ventilator are often
time for breaths (time triggered) or the tidal volume identified by clinicians as rectangular, exponential, sine wave,
(volume cycled). and ramp.
• Phase variables are used to describe those variables that (1)
begin inspiration, (2) terminate inspiration and cycle the ven-
start time (time triggering) or tidal volume (or both). In other tilator from inspiration to expiration, (3) can be limited during
words, the ventilator triggers and cycles the breath. inspiration, and (4) describe characteristics of the expiratory
Box 3-8 summarizes the main points of control variables, phase phase.
variables, and breath types. Figure 3-19 summarizes the criteria for • CPAP and PEEP are two methods of applying continuous pres-
determining the phase variables that are active during the delivery sure to the airways to improve oxygenation in patients with
of a breath.4 refractory hypoxemia.
42 CHAPTER 3 How a Breath Is Delivered