The best insurance for accurate volumes is simply to insure that your glassware is
clean. For burets and pipets, cleanliness of the glassware is indicated by the absence
of any 'water beads' on the inside surface of the glassware. When the item is clean, the
solution will exist in a thin, unbroken film on the inside of the glassware.
A buret can be easily cleaned by using a detergent solution and a 'buret brush', a long
handled brush that is designed to be able to reach the bottom of the buret. Rinse the
buret thoroughly with tap water, and follow with 3-4 portions of distilled or deionized
water.
Another check that you must do for the buret is a leak check. Fill the buret with water,
and check for leaks around the stopcock (the valve at the bottom of the buret that you
use to release the reagent. Allow the buret to stand for several minutes, and make sure
that the indicated volume remains constant. If you find a leak check with the
instructor for further instructions.
Calibration
Pipet
The proper procedure for delivering the indicated volume for a pipet is relatively
straightforward. First, using your pipet aid, draw enough liquid into the pipet such that
the level is well past the 'mark' on the neck of the pipet. BE SURE TO AVOID
DRAWING REAGENT INTO YOUR PIPET AID! Remove the pipet aid and quickly
cover the end of the pipet using your forefinger. At this point make sure there are no
bubbles in the bulk liquid, or foam at the reagent/air interface. Wipe any reagent off of
the exterior of the pipet at this time. This is easier if you tilt the pipet slightly while
wiping. Find a 'waste' beaker, and slowly drain the pipet to the calibration mark while
the tip of the buret is in contact with the wall of the waste beaker. Then, carefully
move the pipet tip to the receiving vessel, and allow the liquid to drain. When the
pipet appears to be drained, touch the tip of the pipet against the wall of the receiving
container for at least ten seconds. If this is done properly, there should be a small
residual volume in the pipet tip. THIS IS EXPECTED, and the pipet has been
calibrated for this!
To calibrate your 25 mL pipet, use a clean pipet and water for which the temperature
is known and constant ( temperature-equilibrated water). Using an analytical balance,
weigh a clean 125 mL Erlenmeyer flask with a rubber stopper in the neck and record
the mass. Deliver a known volume into the flask, stopper, and reweigh. Using the
following table, calculate the volume of water delivered from the pipet based on the
mass of water delivered and the volume corrected to 20OC. Perform this operation at
least 10 times, and calculate the mean and standard deviation for the volume
delivered. [NOTE: It is not necessary to scrupulously dry the flask each time, but it IS
necessary to obtain a new initial weight, as some of the water from the previous
calibration may adhere to the inside of the Erlenmeyer flask.]
Buret
Using temperature-equilibrated water, fill the buret past the 0 mL mark (be sure to
record the temperature), insuring that no air bubbles are trapped in the tip. Give the
water about a minute to drain, and then allow the water to drain to the 0.00 mL mark
in your waste beaker. Touch the side of the tip to the beaker to remove any water
sticking to the tip. Replace the waste beaker with the weighed 125 mL Erlenmeyer
flask, (analytical balance) and slowly transfer 10 mL of water to the Erlenmeyer flask
(this should take about 1 minute.) Reweigh the flask (again using the analytical
balance) and calculate the volume delivered as you did for the pipet. Repeat this
procedure at least 2 more times, or until your delivered volumes are ±0.02 mL. Once
you get an appropriate agreement between the 10 mL values, perform the same
procedure for 20, 30, 40, and 50 mL. Remember to allow the buret to drain at about
10 mL/min for all of these determinations!
Volumetric Flask