Professional Documents
Culture Documents
14 Intended Use
The AC40 clinical audiometer is designed to be a device for
diagnosing hearing loss. Output and specificity of this type of device
are based on the test characteristics defined by the user, and may
vary depending on environmental and operating conditions. The
diagnosing of hearing loss using this kind of clinical audiometer
depends on the interaction with the patient. However, for patients not
responding well possibilities of various tests allow the tester of
having at least some evaluative result. Thus, a “normal hearing”
result should not allow for ignoring other contra indications in this
case. A full audiologic evaluation should be administered if concerns
about hearing sensitivity persist
15 Precautions
Notice - Be sure to use only stimulation intensities which will be
acceptable for the patient.
Notice - Never insert or in any way use the insert headset without a
new clean non defect test tip.
The “Common” setup in the AC40 must be set to “Print All” if patient
data are to appear automatically on printouts together with
audiometric data (press “Menu” (11)\Setup (F2)\Common (F3), Now
highlight the “Printer options”: “All” and turn it on with the “Change”
key (F1)).
No transfer to OtoAccess:
Patient data entered into the AC40 by use of keyboard will not
be transferred to the OtoAccess computer program.
17 Delete Function
The “Delete” button (41) carries 4 different functions:
18 Printout
The data from the AC40 can be printed in different ways:
External printer:
Select “Print for external printer”. If an external printer (HP-
GL2 language compatible) is connected to the AC40, then
printouts of speech and tone audiograms will be controlled by
the “Print” button. (If the printer does not respond, check if
“External” printer is selected in the printer setup: Menu
(11)\Setup (F2)\Common (F3)\Printer).
Computer:
If the AC40 is connected to a PC with OtoAccess program
installed, speech and tone audiograms from the AC40 can be
printed on the PC.
Talk back:
If the patient is equipped with a microphone connected to the
talk back input the operator may listen to the patient by
pressing the “TB” button (13). The volume in the patient’s
headset is adjusted on the volume control (N) placed just
above the “TB.” button (13).
Instructions to assistant:
There is always a direct connection through the goose neck
microphone to the assistant wearing a headset connected to
the “Assistant Monitor” output.
20 Monitoring
Monitoring of channel 1, 2 or both channels together is available by
selecting the “Monitor” button (14) once, twice or three times.
Selecting it a fourth time will switch off the monitoring function again.
In the pure tone speech testing screens the “Swap Ch.” function is
activated by pressing the F7 key. The Intensity Control (46) and
Tone Switch (47) on the left hand side of the audiometer are now
controlling channel 2, and the right hand Intensity Control (51) and
Tone Switch (50 are likewise operating channel 1. To return to
normal operation, select “Swap Ch” (F7) again. When “Swap Ch” is
selected it is active in all testing modes.
24 External Keyboard
Short keys:
F10 = Print:
Press F10 to print out.
3) Select test ear as desired and transducer channel 1 (32, 33, 34,
35, or F8)
4) Select frequency with the frequency buttons (48, 49)
5) Select desired intensity with channel 1 attenuator (46)
6) Store chosen threshold by selecting “Store” (36)
Extended range:
An additional 20dB is available as extended range by
selecting the “Ext. Range” (24) button.
5) Select the test ear by pressing Left and Right buttons (32 and
33).
6) Select the desired frequency with the frequency buttons (48 +
49).
7) Select Intensity with channel 1 Attenuator (46).
8) Present tone by touching channel 1 Tone switch (47).
9) Store threshold by selecting “Store” (36).
Multi-frequency:
Smaller steps between frequencies are available by selecting
the “Multi frequency” button (22).
Extended range:
An additional 20 dB is available as extended range by
selecting the “Ext. Range” (24) button.
The speech window will now appear on screen with default settings
as shown below:
Extended range:
An additional 20 dB is available as extended range by
selecting the “Ext. Range” (24) button.
Insert masking:
In cases of suspected over-masking, masking by insert phone
may avoid this and is recommended. This improves the cross
hearing of the masking sound from approximately 40dB with
traditional headphone, to approximately 70dB of the CIR22
insert phone. The procedure is as follows:
Synchronous masking:
For synchronous masking, lock attenuators together by
choosing “Synch” (button 23). Changing intensity in channel 1
(47) will now result in an equal change in the masking level of
channel 2.
30 SISI Test
The Short Increment Sensitivity Index (SISI) is designed to test
the ability to recognize one decibel increases in insensitivity
during a series of bursts of pure tones presented 20dB above
threshold.
Start the actual test by selecting the 1.0 dB increment (F6). After 20
increments at 1.0 dB have been presented the test stops
automatically and the score is shown on the display.
31 Weber Test
The Weber test distinguishes between conductive and sensorineural
hearing loss originally by means of a tuning fork. Strike it softly and
place the fork in the middle of the patients head. If the patient hears
the tone better in the poorer ear the hearing loss is conductive, and
the tone is heard better in the better ear the hearing loss is
sensorineural at the given frequency. The tuning fork method,
however, typically only allows for testing at one frequency, unless the
clinician has many tuning forks. Using an audiometer together with a
bone conductor to do the Weber is more reliable and flexible than
the tuning fork method and has therefore become widespread.
4) Select desired test ear (32, 33) and frequency (48, 49).
5) Use channel 1 attenuator (46) to set the intensity to 40dB
above the patient’s hearing threshold for the frequency being
tested.
6) Select increment intensity (F9) which is normally clearly
audible. Instruct the patient to respond via the patient
response button as long as they hear the beating character of
the tone. (Test of cooperation can be done by introducing the
0 dB increment (F1). This should make the patient cease to
respond).
Instruct the Patient that he will hear tones alternating Left and Right
ears, that he is expected to press the response switch corresponding
to the ear where the loudest tone is heard and that only loudness
and not character of the tone should be considered.
4) Set the tone intensity to the poorer ear at a level 5dB above
threshold. Select “Start” (F2).
5) Adjust the level of tone for the other ear so that loudness
levels perceived are matched. Store the result by pressing
“Store”
6) Increase the level to the poorer ear 20dB and repeat test.
Repeat test at increasingly higher intensities until discomfort
or limit of output is reached.
Menu function:
1) Press “Menu” once or more times to enter the “Main
Menu”
2) Press “Tests” (F3).
3) Press “MLD” (F5).
Test:
1) Instruct the patient to respond to tone only and that
when a pulsing tone is heard, the Patient Response
button should be activated.
2) Select the desired frequency with the frequency buttons
(48 + 49). 500Hz is usually preferred.
3) Now press “start” (F2).
Instruct the patient that he will hear two different tones alternating in
the same ear that the purpose of the test is to match the intensities
of the two tones, and that only loudness and not character of the
tone should be considered.
Instruct the patient that he will hear tones and should keep the
Patient response button activated whenever a tone is audible to him.
The test runs automatically. When the test is finished “Stop” above
F1 will be highlighted and the test score will be presented. Now the
other ear may be tested.
Instruct the patient that tones with different frequencies will be heard,
and that the patient response button should be activated whenever a
tone is audible.
4) Test
−Fixed Frequency Békésy is selected automatically
−Frequencies in addition to the standard frequencies are
included in the test, if they are selected in the setup (Menu
(11)/Setup (F2)/Tone (F1)/ Frequency selection).
For example: You want to delete a 2 kHz test result. Place the
cursor on 2 kHz by use of the frequency knobs (48 + 49).
Then press “Delete” (41).
Instruct the patient that he will hear tones with different frequencies
and that he is supposed to push the patient response button
whenever a tone is audible to him.
Trace / “Average”:
In order to calculate the audiogram from the HW tracing
simply press “Average” (F9) and the resulting audiogram will
be displayed. Return to the HW tracing by pressing “Trace”
(F9).
Printing:
A printout of the tracing or the calculated audiogram can be
made according to whether “Tracing” or “Average” is selected
with F9.
Deleting in a HW tracing:
Pressing “Delete” (41) will delete the test result of the
frequency indicated by the position of the cursor on the
screen.
For example: You want to delete a 2 kHz test result. Place the
cursor on 2 kHz by use of the frequency knobs (48 + 49).
Then press “Delete” (41).
Ask the patient to press the patient response button when he hears a
tone (do not mention which ear).
2) Select first the output: Left, Right, FF1 or FF2 by the buttons
(32), (33) or (35). The intensity – dB SPL is adjusted by means
of the attenuator in the bottom left-hand corner (46).
3) The indication from the patient (if the tone is not heard, if the
tone is comfortable or if the tone is too high etc.) is adjusted in
the form by means of the Frequency Decr/Incr (48) or (49) and
plotted by activating the Store Button (36).
Features:
1) Change attenuator steps to 1dB with (25)
2) Enter extended range levels with “Ext. Range” (24)
3) Add Warble to tone (21)
4) Enable smaller frequency steps with “Mult Freq.” (22)
5) Reverse function of “Tone Switch” by Man Rev” (37)
6) Compare Left audiogram to Right (F5)
Masking
In manual audiometry masking can be applied with channel 2
1) Select the output to carry the masking sound (42), (43),
(44), and (45). (Relevant type of noise is automatically
pre-selected as indicated by (29))
Speech testing
1) Enter Main Menu with “Menu” (11)
2) Enter manual audiometry with “Audio” (F1)
3) Select “Mic” (18) or “CD/Tape” (19) in channel 1
4) Select desired output (32), (33), (34), (35)
5) Select desired HL level (46)
6) Present word
7) Monitor patient reply with “TB” (13) and “N”
8) Enter score of each word by “Incor” (30) or “Correct” (31)
Features:
1) Monitor pre-recorded speech material with “Monitor” (14)
and volume “0”
2) Change attenuators steps to 1dB by (25)
3) Enter extended range levels with “Ext.Range” (24)
4) Create additional speech curves with (F3), (F4)
5) Test UCL level by “HL-UCL” (F1)
6) Compare headphone results to FF with (F5)
Re-programmed test
1) Enter Main menu with “Menu” (11)
2) Enter pre-programmed tests with “tests” (F3)
3) Select desired test with (F1-8)
4) If relevant select output (32), (33), (34), (35)
5) If relevant select frequency (48, (49)
6) If relevant select intensity (46)
7) If relevant control test “Start” (F2) – “Stop” (F1)
11) Menu key: This key will move you one step back in the
menu hierarchy until the “Main menu” is
reached.
12) Talk forward: Pushing this key will present your spoken
instructions in the patient’s headphones. When
this key is not pushed your instructions are
automatically presented in the assistant’s
headphones (Optional).
22) Multi-frequency:
Selects smaller steps in between the main
audiometric frequencies. The size of the steps
is set up in the “Tone Setup”.
25) 1 or 5 dB steps:
Selects between attenuator steps of 1 dB or 5
dB.
35) Free field 1/2: Selects free field speakers as transducer for
channel 1.
- First push: Selects FF speaker 1.
- Second push: Selects FF speaker 2.
39) Manual/Reverse:
Continuous or manual presentation has two
different settings for channel 2.
40) Simultaneous/Alternate:
Presentation in both channels offers the
possibility of locking presentation channel 1 and
channel 2 together, or to have the presentation
in the two channels alternating to each other.
41) Delete: Pushing this button will delete the data on the
screen. The data will be deleted in sequence,
so recently entered data may be deleted
without deleting other data. Simply hold down
the key until the desired amount of data has
been deleted.
C) Tone indication:
This indicator lights up when signal is
presented in channel 1.
G) VU.Meter channel 2:
Indicates the level of speech audiometry. Is
adjusted by “L” or “M” to reach “0”.
I) Display brightness:
Adjusts the display brightness.
J) Microphone calibration:
Adjust this so the channel 1 VU-meter deflects
to “0” for live voice in speech testing.
L) Microphone calibration:
Adjust this so the channel 2 VU-meter deflects to
“0” for live voice in speech testing.
Q) Gooseneck microphone:
May be used for live voice speech testing as well
as for talk forward and instructions to assistant.
Medical CE-mark:
The CE-mark indicates that Interacoustics A/S meets the
requirements of Annex II of the Medical Device Directive
93/42/EEC. TÜV Product Service, Identification No. 0123,
has approved the quality system.
Channels:
Two independent channels.
Channel 1: Input: Tone, Microphone 1&2, Tape/CD 1&2, NB,
SN, WN, and PN. Output: Left, Right, Bone L+R, Free Field
1+2, Insert-phones, HF-phones.
Channel 2: Input: Tone, Microphone 1+2,Tape/CD, 1&2, NB,
SN, WN, and PN. Output: Left, Right, Free Field 1+2, Insert-
phones, HF-phones, Insert masking.
Frequency Range:
125-20000 Hz divided in two ranges: 125-8000 Hz and 8000-
20000 Hz.
Synchronous Masking:
Locks Ch 2 attenuator to follow Ch 1 attenuator.
Attenuators:
Totally Click free, -10 to 120 dB HL in 1 or 5 dB Steps.
Talk forward:
0-110 dB SPL: Continuously adjustable on front panel, built-in
goose neck microphone.
Talk Back:
Microphone input. Level adjusted on operation panel.
Monitor:
Built in speaker or external loudspeaker. Monitor output level
for Ch 1 and Ch 2 adjusted separately on operation panel.
Assistant monitor:
Output to external earphone.
Printer:
Output to external Laser Printer (HP-GL/2 language) via 25
pin D connector.
Computer Communication:
Built-in RS232C two way computer interface which allows the
computer to both monitor and control the AC40.
Tests:
Tone: Manual, continuous, single pulse, pulsing (variable).
Speech: Live voice through goose neck microphone or
external microphone, Tape or CD inputs. Score counter:
Calculates % of correct score for speech.
Auto Threshold: Patient controlled Hughson- Westlake Test
after ISO 8253-1. 3 out of 5 or 2 out of 3 as response criteria.
Reduced frequency range option for rapid testing.
Békésy Test: 125Hz to 16kHz. Fixed Frequency or Sweep
Frequency Békésy. Continuous or pulsed tone.
Difference Limen Intensity: 0.0dB - 5.0dB in 9 steps.
Difference Limen Frequency: 0.0% - 5.0% in 9 steps.
Loudness Balancing: 250Hz, 500Hz, 1kHz, 2kHz, 4kHz,
6kHz NB noise with direct comparison to standard curves.
Difference Masked Unmasked: Compares graphically the
threshold measurements with and without masking.
Weber: 250Hz to 8kHz with dedicated graphical presentation.
ABLB: Individually adjustable pulse speed and pulse length.
TT decay: Calculation according to Rosenberg (1958).
Masking Limen Difference: Noise out of phase and signal
out of phase. Automatic calculation.
Monaural Loudness Balancing: Programmable test setup.
SISI: 0, 0.2, 0.4, 0.6, 0.8, 1, 2, 3, 5 dB, 20 increments.
Automatic score counter which calculates in % the number of
responses to 1 dB increments.
Stenger: Pure tone or Speech can be used for Stenger test.
Lombard test.
Dofler Steward test.
Dimensions:
LxWxH: 50x47x20 cm/ 20x19x8 inches.
Weight: 13 kg/29 lbs.
Power Supply:
AC 50-60 Hz. 100-120 V, 220-240 V.
Consumption: Max. 180 VA.
Data Transfer:
Transmitting data from the AC40 is done from the OtoAccess.
Please refer to the OtoAccess manual for Details.
46 Contents of Shipment
As standard the AC40 is delivered with the following:
1 Instrument AC40
1 Audiometric Headset TDH39
1 Bone Conductor B71
1 Bone Conductor Bag
1 Patient Response Button APS3
1 Operation Manual
1 CE Manual
47 Reporting Imperfections
Inspect before connection:
Prior to connecting the AC40 to the mains it should once more
be inspected for damage. All of the cabinet and the
accessories should be checked visually for scratches and
missing parts.
Annual calibration:
The AC40 has been designed to provide many years of
reliable service but annual calibration is recommended due to
possible impact on transducers.
Check that the intensities are calibrated for correct level by (J)
(K), making the VU meter (A) deflect up to “0”.
Make sure that the correct signal input is selected (18) (19)
and that the correct microphone is selected in the “Common”
setup (Menu (11)/ Setup (F2)/ Common (F3)/ Microphone 1).
In Auto testing (Békésy or HW) the test stop after having tested
only one ear:
If there is any audiometric data in the memory for the other ear
the auto test is not automatically started on this ear as this
would delete the previously saved data without warning.
Please make sure no part of the audiogram is recorded for the
other ear before starting the auto test.
Frequency dB jump:
If you want to change intensity automatically when the
frequency is changed the number of dBs to increase or
decrease can be set up from here. Available jumps run from
+20dB to –80dB in steps of 5dB. -20dB is suggested.
Warble Modulation:
When warble (21) is selected in pure tone audiometry the
warble modulation (frequency changes) will be according to
this setup. Available modulations run from 0% to 25%. 5% is
advised.
Bone at 8 kHz:
Here you can decide if you want 8 kHz available for bone
conduction testing. The settings are “Yes” & “no”.
Frequency selection:
You can remove certain frequencies from the automatic test
routine by not selecting them in this setup.
Max Score:
Here you can enter the desired number of words in your word
lists. This will make the automatic speech score counter
calculate correctly, and when the full number of word scores
have been entered then the automatic speech score counter
will stop and display the final speech score. Available settings
run from 0 to 100 words on a list.
51 Common Setup
1) Press “Menu” once or more times to enter the “Main Menu”
2) Press “Setup” (F2)
3) Select “Common” (F3) to enter the Setup window:
L/R settings:
Automatic ear selection makes it possible to change ear
selection in Channel 2 when you change test ear in channel 1.
To achieve this select “L/R CH1=>R/L CH2”.
Bone symbols:
Two different types of bone conduction signs are available.
Select the type used in your area.
Available settings are: R = “<” – L=”>” and R=”>” – L=”<”
International settings are R = “<” – L=”>”.
Background:
On the left hand side of the display, the desired graphics may
be selected or inverted to give a different visual presentation.
Monitor loudspeaker:
The internal monitor loudspeaker may be turned on to monitor
the AC40 and to listen to patient talk back. Possible settings
are “On” and “Off”. On is recommended.
Printer:
To use an external printer connected to the AC40 is must be
selected in the setup by slecting “External”.
Move on Store
On/of
Microphone 1:
Select the desired microphone for speech testing; either
“internal” goose neck or “external” connected to “Mic 1” socket
on the rear plate.
Type of Noise:
Selection of noise as stimulus can be done from the front
control board by selecting Noise (29) as stimulus for channel
2.
- When “NB” (29) is chosen in channel 2, narrow band noise
according to the frequency setting will be presented.
- When “N” (29) is chosen in channel 2 the noise will be
speech noise when channel 1 is set to Mic (18) or Tape/CD
(19). If however, channel 1 is set to Tone (17), noise in “N”
setting will be white or pink noise depending on the setup
here. Settings are “White noise” and “Pink noise”. White is
suggested.
Headset:
Selection of TDH39 headphone or EAR-Tone insert phones
shall be selected as default when you turn on the AC40.
Settings are “Phones” and “Insert”.
Default intensity:
When pure tone audiometry is selected the attenuator will
automatically place itself on the value set up here. If “Off” is
selected the default intensity is set to 30 dB. Available settings
are: -10 dB to 6 0dB in 5 dB steps and “Off”.
External Keyboard:
If an external keyboard is connected to enter patient data the
type of keyboard must be selected in the Setup. The settings
are: “IBM AT”, ANK10 and “Off”.
Baud rate:
Baud rate is the communication speed with connected
computer equipment. The baud rate of the transmitting
instrument (AC40) must match the baud rate of the receiving
instrument or computer in order for communication to be
possible. If you use OtoAccess software in your computer you
must choose identical baud rates in the OtoAccess and in
AC40. Baud rate Settings are 9600, 19200 and 38400 and
USB Converter.
Printer Information:
Available settings are: “None”, “Clinic”, “Subject” or “Both”.
54 Test Setup
1) Press “menu” once or more times to enter the “Main Menu”.
2) Press “Setup” (F2)
3) Select “Tests” (F6)
ABLB:
1) Press “ABLB” (F1) to enter the “ABLB Test Setup”
window:
MLD:
1) Press “MLD” (F2) to enter the “MLD Test Setup” window
MLB:
1) Press “MLB” (F3) to enter the “MLB Test Setup”
Auto Test:
1) Press “Auto” (F4) to enter the “Auto test setup”
Curve to average:
In fixed frequency Békésy the curve for continuous tone or the
curve for pulsed tone may be used for averaging. Select which
curve to use for averaging. Suggested setting: Continuous.
Threshold Method:
For speed and accuracy. Select if 2 out of 3 or 3 out of 5
correct responses to a given intensity shall determine the
threshold. Suggested setting: For speed: 2/3. For accuracy:
3/5.
Loudness Scaling
1) Press “LOUD” (F5) to enter the “Loudness Scaling Setup”
Baud rate:
Select baud rate by “Change”. *Press “Change” (F1) to
choose between baud rates of 9600, 19200 or 39400. Select
baud rate to match that of the connected computer.
Transmission:
Select Verification by “Change”. * “Verification” or “No
verification” may be selected. The OtoAccess program will
need “Verification” in order to operate, whereas other
programs may not need “No Verification” in order to operate
correctly.
Transmit to:
Select the instrument to transmit data to by “Change”. “PC” or
“MS40/MS25” may be selected.
Transmit:
Select the way in which to transmit data by “Change”.
“Normal” and “Special” may be selected.
- For maximum electrical safety, turn off the power from a mains
powered instrument when it is left unused.