ES series Firmwares update procedure
This is an instruction which helps you to update (1) DCPU firmware and (2) TS_COMM firmware. Before carrying out
the updates, please check that "TS_Loader","TSC_Loader", "DCPU firmware" and "TS_COMM firmware" are avaliable
and activated on your PC.
Note:
‐ In order to update correctly, ‘TS Loader’ and ‘TS_COMM Loader’MUST be installed before starting the following
procedure.
‐ .NET Framework 2.0 is required on your PC to activate TS_COMM Loader. If it is not avaliale on your PC, please obtain
the installer from the following web page and install it on your PC. :
http://www.microsoft.com/en‐us/download/details.aspx?id=1639
‐ Also, please check that your PC and Total Station are connceted via DOC210 interface cable. (Mostly, a USB cable
which connects DOC210 and your PC is required, too) In addition, please check the batteies are charged well.
‐ When a serial port is not mounted on the PC, a USB‐serial conversion cable is necessary to use USB port.
1. DCPU firmware update
1‐1. DCPU firmware update
[Total Sation]
1. Start it in boot‐mode.
Please hold down {ENT} & {ENT} keys, and press Power key on the instrument panel . The following screen will
appear on the instrument display. (The battery must be charged fully)
When pressing {ENT} key on the instrument panel, the following screen will appear on the instrument display.
(*This screen will keep appearing until DCPU firmware is updated)
* As next procedure, start TS Loader and then install DCPU program.
[PC]
1. Start TS Loader.
Please double‐click DCPU firmware file (Extension : *.sbc),then TS_Loader tool can be activated and the following
screen will appear.
* As for the installation of TS Loader tool, please refer to "TS Loader Installation Procedure"
*Please check if the proper port is activated with "WINDOWS Device Manager"
2. DCPUfirmware update
Please cleick [Load Now] button, then the update action will start.
After the firmware update is complete, the following dialog box will appear. (*It takes approx. Five (5) minutes)
Checksum
Note:
Pleaseensure that the checksums are matched. After that, please click [OK] button and close the TS_loader tool.
1‐2. Check
Please check the following screen on Total Station, so you could see if DCPU firmware can be updated properly.
Note:
* If "User Registeration screen" appears on the display, please activate Demo mode, so you could see the above
screen.
(The following procedure is required only to models into which TSshield is installed.)
2. TS_COMM firmware update
2‐1.TS_COMM firmware update
Note:
‐ Please check that your PC and Total Station are connceted via DOC 210 interface cable. (Mostly, a USB cable which
connects DOC210 and your PC is required, too) In addition, please check the batteies are charged well.
‐ "TsCommCE.dll" file should be stored in the folder of the file "TSC_Loader1006.exe".
[Total Station]
1. Start it in download‐mode.
Please hold down {F4} & {ENT} keys, and press Power key on the instrument panel . The following screen will
appear on the instrument display.
2. Please select [TS_COMM <‐‐‐> PC] button and then press {ENT} key on the panel.
After approx. fifteeen (15) seconds, the following screen will appear on the instrument display.
* As next procedure, start TS Loader and then install TS_COMM program.
Installation of TS_COMM program must be hold until the above standby state (black & white reversing
charactors) is displayed.
Note:
When update is completed, press [ESC] to go back to the menu screen.
[PC]
1. Start TS_COMM Loader.
Please double‐click "TSC_Loader1006.exe". Then, TS_COMM Loader tool ican be activated and the following
screen will appear.
Note:
.NET Framework 2.0 is required on your PC to activate TS_COMM Loader. If it is not avaliale on your PC, please
obtain the installer from the following web page and install it on your PC. :
http://www.microsoft.com/downloads/ja‐jp/details.aspx?familyid=5b2c0358‐915b‐4eb5‐9b1d‐10e506da9d0f
*Please check if the proper port is activated with "WINDOWS Device Manager"
2. TS_COMM firmware installation
Please click [INSTALL] button on the above screen. The installation will starts (It takes approx. one (1) minute)
After the installation is complete, the following dialog box will appear. The installtion is finished to to click 【OK】
button.
2‐2. Check
1. After the update is complete, please press {ESC} key on the instrument panel. Then, the following screen will
appear on the display.
2. Please select [Firmware version] button, then press {ENT} key on the panel. After approx. fifteen (15) seconds
passes, the following screen will appear, so you could see if TS_COMM firmware can be updated properly.
3. Power OFF