INDUSTRIAL
STERILIZATION
Process Optimization
and Modality Changes
I (c) 2020
e user license. Further copying, networking, and distribution prohibited.
Summer 2020
IN THIS ISSUE
3 Introduction 45 Radiation Sterilization:
Emily Craven, Andre Tuggles, Dose Is Dose
and Jami McLaren Joyce M. Hansen, Niki Fidopiastis,
AAMI President Editorial Board Industrial Sterilization: Process Editorial: Visit www.aami.org/BIT for Index, Archive: Biomedical
Robert D. Jensen Emily Craven Optimization and Modality Changes is complete author guidelines. Instrumentation & Technology is indexed,
Boston Scientific a supplement to Biomedical archived, or listed by the following:
Editor in Chief Advertising information: Visit
Instrumentation & Technology (ISSN CINAHL Info Systems, Crossref,
Gavin Stern Joyce M. Hansen www.aami.org/advertising.
Johnson & Johnson 0899-8205) and is published by the DeepDyve, Ebsco, Embase, Index
Managing Editor Association for the Advancement of Mention of any commercial product, Medicus, PubMed/Medline, Referativnyi
Melissa Coates Jami McLaren Medical Instrumentation (AAMI), process, or service by trade name, Zhurnal, and Thomson Reuters.
Boston Scientific 901 N. Glebe Rd., Suite 300, Arlington, trademark, manufacturer, or otherwise
Graphic Designers Photocopies: Authorization to photocopy
VA 22203. © 2020, Association for in Industrial Sterilization: Process
Ruth Moyer Jeffery Nelson items for internal or personal use, or the
the Advancement of Medical Optimization and Modality Changes
Marielyn Cobero Nelson Labs internal or personal use of specific
Instrumentation. Publishing services does not constitute or imply an
clients, is granted by Allen Press, Inc. for
Copy Editor Andre Tuggles provided by Allen Press. endorsement or recommendation by
libraries and other users registered with
Barbara Saxton Johnson & Johnson AAMI. The views and opinions of the
Membership: AAMI members receive the Copyright Clearance Center (CCC)
authors in Industrial Sterilization: Process
a complimentary subscription. Contact Transaction Reporting Service, 222
Optimization and Modality Changes do
AAMI membership at 800-332-2264, Rosewood Drive, Danvers, MA 01923.
not state or reflect the opinion of AAMI.
ext. 1214.
Visit AAMI on the web at www.aami.org.
SUPPORTING ORGANIZATIONS
BD is one of the largest global medical technology companies in the Nordion, a Sotera Health company, is a leading provider of medical
world and is advancing the world of health by improving medical gamma technologies used for the prevention and treatment of disease
discovery, diagnostics and the delivery of care. BD helps customers and infection and is committed to their mission, Safeguarding Global
enhance outcomes, lower costs, increase efficiencies, improve safety Health™ with every critical isotope they supply. www.nordion.com.
and expand access to health care. www.bd.com
The global Microbiological Quality & Sterility Assurance community WuXi AppTec Medical Device Testing is a comprehensive and
has used the forum of the Kilmer community for over 40 years to share integrated testing platform for medical devices, tailored to suit
innovations and collaborate across the healthcare industry including particular needs at every stage of product development. Our facilities
academia and regulatory agencies. Passion for Patients Driving in Atlanta, St. Paul and Suzhou are US FDA registered and ISO
Collaborative Innovation accredited. www.wuxiapptec.com
I (c) 20202 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
Introduction
“How do we simplify sterilization modality changes and During the Kilmer Conference in June 2019, we presented
process optimization?” This question was the catalyst for a the same four questions to the conference participants. Once
year-long collaboration that started with a small team of we gathered responses from the larger group, the team
sterility assurance professionals and eventually grew to worked to consolidate and analyze the input from the
include the entire Kilmer Sterility Assurance Community. original team meeting in Arlington and the input from the
As coleaders in answering this question, our first job was conference participants. The team identified several items
to bring together a team of individuals with the diversity of that would support changes needed in the industry; these
technical backgrounds necessary to seed our collaboration items include targeted publications, training/education
with great ideas and information. We recruited experts in opportunities, tools for information sharing, and better
established and novel sterilization modalities, reusable guidance on existing standards and regulations.
device processing, regulatory affairs, and microbiology. We are therefore very pleased to have worked with AAMI
These individuals represent medical device manufacturers, to present this publication of invited articles on topics that
contract sterilization service providers, contract labs, and can provide guidance and insight into process optimization
regulators. The team goal was not only to find answers to the and modality changes. This is the first step in many to help
questions we were asked, but also to find a way to make this answer these important questions.
information more widely available to the entire industry.
Our official “Collaboration Event” was hosted at the AAMI Emily Craven (Mevex), Andre Tuggles (Johnson & Johnson),
headquarters in Arlington, VA in May 2019. During this Jami McLaren (Boston Scientific)
two-day meeting, we worked together to define the scope of Kilmer Conference Collaboration Event Cochairs
what we wanted to accomplish through friendly debate and Special thanks to the collaboration team:
structured information gathering. The group came out with
four distinct challenge questions: Arlington event: Phil Cogdill (Medtronic),
1. What is the source or reason for resistance to changing Bart Croonenborghs (Sterigenics), Melissa Escobedo
modalities and/or optimizing sterilization processes? What (Johnson & Johnson), Tony Faucette (BD), Nupur Jain
tools, resources, etc. are needed to assist in overcoming the (Intuitive Surgical), Vu Le (Abbott), Brian McEvoy (STERIS),
resistance to change? George Ngatha (Food and Drug Administration),
2. What is the barrier to accept or adopt novel sterilization Neville Niessen (Baxter), Patrick Weixel (Food and Drug
methods? What tools or information might we gather to Administration). Thanks also to Martell Winters (Nelson Labs),
assist in a transition to a novel sterilization method? for his role as team scribe.
3. How do we efficiently use the capacity that is available for Expanded team participants: Alpa Patel (Nelson Labs),
gamma and ethylene oxide (EO) sterilization processing? John Williams (Medtronic)
4. What tools or resources are needed to assist with a transi-
tion from gamma or EO to other modalities?
Pictured left to right are Martell Winters, Brian McEvoy, Andre Tuggles, Pat Weixel, Emily Craven, Jami McLaren, Melissa Escobedo, Phil Cogdill, Tony
Faucette, Vu Le, and George Ngatha. Additional phone participants who are not pictured include Nupur Jain, Neville Neissen, and Bart Croonenborghs.
I (c) 20204 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
To share our passion for what we do and help explain why we are passionate about “collab-
orate to innovate” for current and future products, we decided to use the conference as a
means to create a tagline. This tagline provides a tool that everyone can use to promote a
unified passion for working together to innovate for the future. During the conference, the
community collaborated “real time” via crowdsourcing technology to identify themes for our
tagline, and to create an industry tagline that we all can use.
The Kilmer 2019 Conference Industry Advisory Board (IAB) is pleased to share that new
tagline for the community:
Please feel free to use this tagline to express the passion we share as an industry. Use it as
a tool to start a conversation on why we are collaborative: about the end-to-end support we
provide our supply chains, for new product development activities, and for connections with
our customers. Use it on your e-mails, presentations, and communications to share with
others our clear and compelling vision for what we do each and every day.
Collaboration among industry, academia, contract suppliers, regulatory authorities, and
professional associations is key to innovation. To support this means of collaboration across
the community, we are continuing to sponsor additional Kilmer Collaboration Events and
recently initiated teams to work on the two PAT topics identified. Together we have the ability
to support the development and manufacture of products that help improve the lives of the
patients, customers, and consumers.
Joyce M. Hansen
Chair, 2019 Kilmer Conference
I (c) 20206 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
therefore, thorough understanding of sensitivities to heat, • Manufacturing: E.g., reduction of bioburden to enable
moisture, ionizing radiation, certain chemicals, oxidation, lower radiation dose.
and pressure changes is critical when selecting a sterilization An alternate sterility assurance level may also be consid-
modality. Ultimately, the potential for a sterilization process ered, per AAMI/ANSI ST67,7 to reduce the impact of
to negatively impact the ability of the device to provide its sterilization on product functionality. Such consideration is
intended patient care needs to be completely understood. based on an assessment of the risk of harm due to a nonster-
Understanding how the product is manufactured (e.g., ile product compared to the benefit the product provides.
extruded, 3D printed, injection molded, chemical process-
ing) is also key to understanding potential sensitivities. For Examples of Sterilization Modality Selection
example, devices containing polymers manufactured with a Process
high degree of residual stress from manufacturing may be The following section provides six examples of product
more susceptible to damage from the effects of sterilization.3 evaluation for sterilization modality selection as described
Chlorine-containing chemicals used in manufacturing may above. The flowchart in Figure 1 provides a decision tree for
result in high levels of residual ethylene chlorohydrin for choosing commonly used sterilization modalities. While
ethylene oxide–sterilized products. For devices that are sterilization modality selection involves more complexity
reprocessed, impacts related to multiple cleaning and than shown in Figure 1, this flowchart presents a high-level
sterilization cycles must also be understood. thought process for sterilization modality selection. Figures 2
Ideally, the person responsible for selecting the steriliza- to 7 show examples of how the flowchart in Figure 1 may be
tion modality should begin to work closely with the product used to select a sterilization modality for various devices.
design engineers early in the design process to ensure The functionality aspects listed in these examples are not
proper evaluation of all potential product sensitivities. If the exhaustive and do not go into detail regarding potential
product design is conceived with sterilization in mind, this design changes that could enable successful sterilization, but
can minimize design failures and rework later during the are meant to provide examples of the connection between
product development process. The person responsible for patient care and device characteristics impacted by
selecting the sterilization modality should request a sample sterilization.
device or sample components/materials as they become
available in order to have hands-on interaction with the Example 1: Chemical Ice Pack
device and its packaging. This hands-on interaction allows The example shown in Figure 2 demonstrates the modality
for a better understanding of features of the device that may selection process for a chemical ice pack. The product is
pose a challenge to and/or be impacted by sterilization. The used for treatment of swelling due to injury, and only has
person responsible for selecting the sterilization modality contact with intact skin. Therefore, based on the intended
should also gain a clear understanding of whether there is use and mode of patient contact, sterilization for this device
opportunity to change the design of the product, if necessary, is not required.
to ensure that the device can be sterilized. Early involvement
by sterilization experts is key to avoiding time-consuming Example 2: Silicone Breast Implant
and costly design changes later in the development process, The example shown in Figure 3 demonstrates the modality
such as when a sterilization modality is determined to selection process for a silicone breast implant. This product
negatively impact the product’s intended use after the design is used for breast augmentation and/or reconstruction
elements have been selected. through implantation in a patient. Because of the mode of
Various design changes to improve compatibility and patient contact, sterilization is required. The device materials
potentially enable sterilization should be considered when consist of a silicone shell filled with silicone gel. The product
compatibility issues arise. This consideration may include can withstand temperatures of up to 250°C for up to 48
changes in: hours and can withstand pressures as low as seven pounds
• Design: E.g., packaging devices in a low-oxygen environ- per square inch (absolute). The critical functional aspects of
ment to reduce impacts of oxidative degradation during this product include joint integrity, breaking strength, and
sterilization, or packaging devices in a low-temperature elongation, which are negatively impacted by ionizing
environment (e.g., ice packs) to reduce thermal degrada- radiation. Therefore, moist or dry heat may be selected.
tion during sterilization. Gaseous sterilization may be an option, but pressure limita-
• Material: Consider removing, replacing, or altering tions will restrict the modality or cycle parameters.
materials impacted by sterilization (e.g., including addi-
tives such as antioxidants or stabilizers to enhance
radiation resistance of certain materials; this may be
accomplished by working with material suppliers).4–6
Start
Is the product
Stop here No intended to be
sterile?
Yes
Is product Is product
Dry heat
possible
No resistant to high Yes resistant to high
humidity? temperature?
Yes Or
Is product
Ionizing
resistant to
Moist heat ionizing
Yes radiaon
possible
possible radiaon?
Or
Can sterility be
Gaseous achieved through
sterilizaon Yes gaseous
possible sterilizaon
(ethylene oxide or
other)?
No
Is asepc
Use asepc processing
processing
Yes
feasible?
No
Redesign
product
I (c) 20208 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Start
Product descripon: Chemical
ice patch (flexible bag of
reagents that react
endothermically to cool)
Is the product
Stop here No intended to be Mode of paent contact:
sterile? Indirect contact to injured area
to relieve swelling
Figure 2. Example sterilization modality selection process for chemical ice pack.
Start
Sterilizaon required
Yes
Is product Is product
resistant to high Yes resistant to high
humidity? temperature?
Figure 3. Example sterilization modality selection process for silicone breast implant.
Example 3: Ureteral Stent include three polymers and a hydrophilic coating. Moist or
The example shown in Figure 4 demonstrates the modality dry heat sterilization is not possible, as polymers comprising
selection process for a ureteral stent and delivery system. the device experience softening at temperatures above
This product is delivered through the urethra and bladder approximately 50°C, and the coating functionality is nega-
and implanted in the ureter to maintain flow of urine tively impacted by high humidity. Radiation sterilization is
between the kidney and bladder. Key requirements of this not possible because of the risk that the polypropylene
device include flexibility and lubricity to enable navigation of component flexibility will be negatively impacted at the
the stent and delivery system through the relevant anatomy. radiation doses of 25 kGy to 50 kGy typically used in medical
The stent portion of this device is implanted within a device sterilization.3 Ethylene oxide sterilization is selected as
patient’s ureter, thus requiring sterilization. The stent a suitable method, with temperature and humidity condi-
delivery system also requires sterilization as it is used to tions confirmed not to impact product functionality.
place the stent within the ureter. The product materials
Start
No
Can sterility be
achieved through
Use gaseous gaseous
sterilizaon
Yes
sterilizaon
(ethylene oxide or
other)?
I (c) 202010 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Example 4: Prefilled Vaccine Syringe process because all the terminal sterilization modalities
The example flowchart for a prefilled vaccine syringe is (heat, ionizing radiation, gas) would negatively impact the
illustrated in Figure 5. Because the vaccine is intended for quality of the vaccine. Heat and ionizing radiation, for
parenteral use, product sterility of both the syringe and instance, could cause degradation of the drug substance.
contained vaccine is a strict requirement. While the materi- Gaseous sterilization would not be able to penetrate the
als of the syringe components could withstand various prefilled syringe. As a result, aseptic processing is the sole
sterilization modalities, the prefilled vaccine syringe in its viable option to reach a sterile product.
final configuration could not undergo a terminal sterilization
Start
Is the product
intended to be
sterile?
Yes
Product descripon:
Is product Vaccine in filled syringe is
resistant to high intended for parenteral
temperature? inoculaon, sterilizaon is
required
No Product sensivies:
Heat, ionizing radiaon, and
Is product sterilizing gas negavely impact
resistant to the safety, purity, and/or
ionizing potency of the vaccine;
radiaon? addionally, gas cannot contact
the drug
No
Asepc processing required
Can sterility be
achieved through
gaseous
sterilizaon
(ethylene oxide or
other)?
No
Is asepc
Use asepc
processing Yes processing
feasible?
Figure 5. Example sterilization modality selection process for prefilled vaccine syringe.
Example 5: Flexible Irrigation Bag to 250°C) encountered in a dry heat sterilization process
The example flowchart for a flexible irrigation bag is illus- would impair the functionality of the container-closure
trated in Figure 6. The saline irrigation solution is used to system. While dry heat temperatures are not suitable for this
exert a mechanical cleansing action for the irrigation of body product, the temperature (typically 110°C to 135°C), moisture,
cavities, tissues, or wounds and for washing, rinsing, or and pressure ranges encountered in a moist heat steriliza-
soaking surgical dressings, instruments, and laboratory tion process are confirmed not to affect product functionality.
specimens. Because the irrigation solution is used to clean The product configuration of the flexible irrigation bag
and irrigate open wounds, or to rinse other sterile medical includes an impermeable, nonbreathable primary packaging.
devices, the sterility of the product is strictly required. The Therefore, gaseous sterilization is not possible for this product
final product includes the solution on one side and the because the gas would not be able to reach the solution.
container-closure system, made of flexible polymeric materi- While the relatively high density of the flexible irrigation
als, on the other side. Because of the nature of the polymeric bag might make this product unsuitable for sterilization with
container material, the product is not suitable for dry heat all ionizing radiation sterilization processes (electronic beam
sterilization as the high temperature ranges (typically 150°C in particular might be a challenge), the dose ranges encoun-
Start
Is the product
intended to be
sterile?
Gas steriliza
on not
Product descripon: Flexible
possible because of Yes
pressure constraints, gas bag system containing saline for
cannot contact solu
on wound irriga
on
Or
Can sterility be
achieved through
gaseous
sterilizaon
(ethylene oxide or
other)?
Figure 6. Example sterilization modality selection process for flexible irrigation bag.
I (c) 202012 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
tered in various ionizing sterilization processes are suitable Example 6: Reusable Flexible Endoscope
for the polyvinyl chloride container material and the saline The example shown in Figure 7 demonstrates the modality
solution. An ionizing radiation sterilization process such as selection process for a reusable flexible endoscope that is
gamma or X-ray could be selected for the sterilization of the used for therapeutic and diagnostic applications. Because
flexible irrigation solution bag. While both moist heat and these devices are reusable, the healthcare facility is responsi-
ionizing radiation modalities provide the same level of ble for processing them through cleaning, disinfection, and/
sterility assurance, sterilization by heat has a lower risk for or sterilization per validated methods provided by the
impacting the materials (e.g., moist heat does not introduce manufacturer. It should be noted that device compatibility
radiolysis impurities). For these reasons, moist heat is given with cleaning and disinfection methods must be considered
priority over ionizing radiation in the decision process. along with compatibility with sterilization modalities.
Start
Product descripon: Flexible
Materials: Mulple polymers,
endoscopes that are used for
polyurethane elastomer (outer
therapeuc and diagnosc
material), Teflon, rubber (distal
procedures
end), wire metal mesh/cables
(distal end) Is sterilizaon
an opon for Mode of paent contact:
this product? <6 hours use on the paent
Product materials are impacted
at high temperatures; coang
Sterilizaon required?
impacted by high humidity,
Yes No, however flexible
chemicals = moist/dry heat not
endoscopes are compable
acceptable
with ethylene oxide and
Is product hydrogen peroxide gas plasma
Product materials can be
resistant to high or hydrogen peroxide with
negavely impacted with
temperature? ozone sterilizaon methods
radiaon sterilizaon over
repeated uses = radiaon not
acceptable No
Selected modality: Novel
Is product
modality or low-temperature
resistant to
gas sterilizaon (ethylene oxide,
ionizing
vaporized hydrogen peroxide
radiaon?
(VHP), VHP + ozone ), with
careful assessment of product
(typically not
temperature in cycle and
available in
mulple uses over me
healthcare
facilies)
No
Can sterility be
Use low- achieved through
temperature gaseous
gaseous
Yes sterilizaon
sterilizaon (ethylene oxide or
other)?
Figure 7. Example sterilization modality selection process for reusable flexible endoscope.
I (c) 202014 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
I (c) 202016 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Waiting to explore an alternate sterilization modality at the time of need could result in supply chain issues that, in turn, could affect healthcare facility
access to medical devices. The time it takes to react to a disruption could come at the cost of patient care.
data presented at the 2019 International provide a buffer for the additional time
Meeting on Radiation Processing provided needed to deliver products to HDOs (i.e., to
an estimated average compound annual accommodate increased processing time). If
growth rate of 5% for Europe and Asia across this action is taken, the healthcare industry
EO, gamma, E-beam, and X-ray modalities.9 also must be aware of the burden that an
E-beam is a good complement to gamma influx of additional product would place on
but has inherent limitations with penetra- the available sterilization capacity. Therefore,
tion related to high-density products. X-ray the question becomes: “Is the supply chain
is known to have a penetration capability prepared for a disruption?”
comparable with gamma, and approximately Shifting all products from a primary to a
five facilities worldwide offer industrial X-ray backup sterilization site does not imply that
sterilization services. In addition, expansion product volume (cubic footage) can be
projects have been announced recently, processed in the same amount of time. A
with more X-ray sterilization facilities under change from one sterilization site to another
construction in North America, Europe, also affects regulatory agencies. Does the
and Asia. regulatory agency have the capacity to handle
Shortages of available EO sites and Co-60 the influx of submissions with a sudden
have placed challenges on the supply chain disruption in network capacity? This change
and, in turn, affected healthcare delivery is not instantaneous and involves an added
organizations (HDOs) and other users of layer of complexity within the supply chain
medical devices. Given this situation, the to manage product distribution as regulatory
manufacturing of additional products might approval is obtained in different markets.
I (c) 202018 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
I (c) 202020 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
backup, thereby allowing for just-in-time 4. Environmental Protection Agency. EPA Moves
sterilization, reduced turnaround time, and a Forward on Suite of Actions to Address Ethylene
decrease in product inventory. Oxide. www.epa.gov/newsreleases/epa-moves-
When selecting external contract services, forward-suite-actions-address-ethylene-oxide.
factors that should be considered include Accessed April 20, 2020.
vendor competency in managing complex 5. Haney JT Jr, Lange SS, Jenkins A, Myers JL.
equipment and speed to market (utilizing Ethylene Oxide Carcinogenic Dose Response
existing infrastructure). Assessment. www.tceq.texas.gov/assets/
public/implementation/tox/dsd/final/eto.pdf.
Conclusion Accessed June 28, 2020.
For product supply chain sustainability,
6. Illinois Environmental Protection Agency.
more than one sterilization modality should
Illinois EPA information on Ethylene Oxide.
be validated for products that are compatible
www2.illinois.gov/epa/topics/community-
with multiple modalities. The validation of
relations/sites/ethylene-oxide/Pages/default.aspx.
more than one sterilization modality will
Accessed April 20, 2020.
provide medical device companies with
flexibility when responding to high product 7. Food and Drug Administration. FDA Innovation
demand and sterilization supply chain Challenge 2: Reduce Ethylene Oxide Emissions.
interruptions. A BCP that includes validating www.fda.gov/medical-devices/general-hospital-
and maintaining validation of both steriliza- devices-and-supplies/fda-innovation-challenge-
tion modalities should be considered. In 2-reduce-ethylene-oxide-emissions. Accessed
addition, the BCP also should take into April 20, 2020.
account the potential to validate multiple 8. Food and Drug Administration. FDA Innovation
sterilization sites for supply chain flexibility. Challenge 1: Identify New Sterilization Methods
For products designed to allow for multiple and Technologies. www.fda.gov/medical-devices/
sterilization modalities, the BCP becomes general-hospital-devices-and-supplies/fda-
more robust and allows for constructive innovation-challenge-1-identify-new-sterilization-
conversations regarding turnaround time, methods-and-technologies. Accessed April 20,
processing flexibility, and resources. These 2020.
advantages are possible with a best-in-class
9. International Irradiation Association. IMRP
sterilization program that allows for steriliza-
Report: The Business and Science of Radiation
tion SME engagement in the early stages of
Processing in 2019. https://iiaglobal.com/news/
product development.
imrp-report-the-business-and-science-of-
radiation-processing-in-2019. Accessed April 20,
References
2020.
1. Gamma Industry Processing Alliance,
International Irradiation Association. A 10. International Atomic Energy Agency. IAEA
Comparison of Gamma, E-beam, X-ray and Ethylene Helps Remove Highly Radioactive Material from
Oxide Technologies for the Industrial Sterilization Five South American Countries. www.iaea.org/
of Medical Devices and Healthcare Products. http:// newscenter/pressreleases/iaea-helps-remove-
iiaglobal.com/wp-content/uploads/2018/01/ highly-radioactive-material-from-five-south-
White-Paper-Comparison-Gamma-Eb-Xray-and- american-countries. Accessed June 19, 2020.
EO-for-Sterilisation.pdf. Accessed June 28, 2020.
11. Tartal J. Design Controls. www.fda.gov/
2. Lambert BJ, Lam S, Hansen JM, Bryans media/116762/download. Accessed April 20, 2020.
TD (Eds.). Assurance of Sterility for Sensitive
12. ANSI/AAMI/ISO 11137-1:2006/(R)2015.
Combination Products and Materials: New
Sterilization of health care products—Radiation—
Paradigms to Bring Innovative Healthcare Products
Part 1: Requirements for development, validation,
to Patients. 1st ed. Cambridge, MA: Academic
and routine control of a sterilization process for
Press; 2019.
medical devices. Arlington, VA: Association for the
3. AAMI TIR17:2017. Compatibility of materials Advancement of Medical Instrumentation.
subject to sterilization. Arlington, VA: Association
for the Advancement of Medical Instrumentation.
Figure 1. Energy spectra of X-ray photons generated by bremsstrahlung on a tantalum target with electron of 5 and 7 MeV (right) in comparison with
gamma energy spectrum (left). Image on right reprinted from Radiation Physics and Chemistry, 57, Meissner J, Abs M, Cleland MR, Herer AS, Jongen Y,
Kuntz F, Strasser A. X-ray treatment at 5 MeV and above, 647–651, 2000, with permission from Elsevier.
I (c) 202024 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
similar maximum dose. This 11°C difference could be high-power irradiators or irradiators with a smaller irradia-
significant with regard to influence on materials properties. tion field, or lower for a low-power X-ray irradiator where
For example, if one considers the accelerated aging of sterile dose rate is typically a few kGy/h.
barrier systems for medical devices, the Arrhenius equation As shown in Figure 3, higher dose rates can also limit the
demonstrates a Q10 value of 2 (for every 10°C increase in oxidative degradation of polymers by minimizing the time
temperature, the rate of accelerated aging doubles) for for oxygen replenishment required for radical-oxygen
temperature.9 If Q10 is applied to the observed 11°C increase reactions. Consequently, as discussed in AAMI TIR17,10 a
in gamma vs. X-ray, it follows that the materials potentially material that is formerly qualified at a low dose rate (gamma)
age twice as fast in gamma vs. X-ray. will typically require minimal qualification to demonstrate
material compatibility at a higher dose rate, as the gamma
Dose Rate qualification might be considered as the worst case scenario.
The dose rate is the quantity of radiation absorbed per unit
time. In radiation processing, it is usually given in kGy/h or Material Compatibility
kGy/s and is related with the power of the irradiator or One of the known aspects of radiation processing common
source. Therefore, comparisons of X-ray and gamma dose to all three sources (Gamma, E-beam, and X-ray) is that the
rate must be assessed on equivalent throughput systems. An process may have a significant effect on the molecular
X-ray irradiator dose rate may vary significantly depending structure of the processed material, which may lead to a
on the irradiator design (e.g., beam current, converter modification of the medical device and/or its packaging
design, and distance to product). It can be much higher than integrity and properties. The influence of radiation-induced
a dose rate in a gamma irradiator (a few hundred kGy/h) for active chemical species on the properties and performance of
Figure 2. Photons intensity as a function of angle at 5, 7.5, and 10 MeV incident electron energies. Reprinted from Radiation Physics and Chemistry, 57,
Meissner J, Abs M, Cleland MR, Herer AS, Jongen Y, Kuntz F, Strasser A. X-ray treatment at 5 MeV and above, 647–651, 2000, with permission from
Elsevier.
a polymer is proportional to dose. When there is a distribu- rendering gamma sterilization unacceptable. As an example
tion of dose within a part or within a product load, the of the varying discoloration effect from differing radiation
resulting property changes can vary by location within the technologies, Figure 4 shows the discoloration effect of
part, or from part to part. Often the most significant mode of gamma and X-ray radiation (at same maximum dose) on
radiation-induced degradation is the embrittling chain PTFE. As demonstrated, significantly less discoloration is
scission reaction that results from interaction with oxygen.10 observed with PTFE processed with X-ray.
Comparisons made at the STERIS Däniken facility
between a gamma irradiator with a dose rate around 3 kGy/h Improved Dose Uniformity Ratio
and an X-ray irradiator (7 MeV, 560 kW) at a dose rate around The sterilization dose is the “minimum dose to achieve the
250 kGy/h on ultra-high-molecular-weight polyethylene specified requirements for sterility” and the maximum
(UHMWPE) at five different doses (50, 75, 100, 125, and 150 acceptable dose is the “highest dose that can be applied to a
kGy) tested in accordance with ASTM F2565,11 demonstrate specified product without compromising safety, quality, or
the impact on polymer properties, where X-ray was shown to performance.”13 In order for a radiation process to be feasible
be equivalent or better for some characteristics.12 Preliminary and suitable for the product, both the sterilization and
accelerated aging studies performed on common packaging maximum acceptable doses must be delivered.8
material at the same conditions revealed a reduced surface The dose uniformity ratio (DUR; defined as “ratio of the
oxidation index for X-ray–irradiated packaging compared maximum to the minimum absorbed dose within the
with gamma. radiation container”13) defines the available dose window
A well-known effect of the radiation process on material is between the sterilization dose and the maximum acceptable
color modification—where a number of polymers may dose. Therefore, the lower the DUR of the irradiation
discolor to yellow or brown following processing—as a process, the greater the opportunity for efficient processing
consequence of the maximum dose received and possibly the of products. A lower DUR for the same process load may
dose rate. The degree of coloration is also dependent on the result in improved product properties as the maximum dose
material and may potentially fade over time, and therefore delivered to the product will be reduced, therefore reducing
while undesirable in a medical device, may be acceptable. deleterious effects such as oxidation and discoloration.
Gamma sterilization is compatible with many materials. As a result of the X-ray photons having increased penetra-
However, materials such as polyvinyl chloride, acetal, and tion and optimized photon energy delivery, improvements
polytetrafluoroethylene (PTFE) can be severely affected, (over gamma) in DUR may be realized: As discussed in the
Figure 4. Polytetrafluoroethylene material coloration following processing with gamma and X-ray irradiation at the same maximum dose (circa. 250 kGy).
I (c) 202026 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Temperature section above, even with many lower-energy DUR is improved by 14% with X-ray. Such a reduction could
photons, bremsstrahlung photons from a 5.0-MeV E-beam be significant, especially with a polymer that might have
penetrate slightly more than the radiation from a large-area marginal compatibility with radiation processing. Specifically,
source of gamma rays.4 This greater penetration is partly due if one considers this example in conjunction with Figure 6,
to the higher energy photons in the bremsstrahlung spectrum four polymers that would not be compatible with gamma
and partly due to the angular distribution. X-ray systems have potentially become available for X-ray. This is very much in
a directive beam of photons concentrated in the direction of line with expectations as summarized by Grégoire et al.14
the product, optimizing the photon capture rate. In contrast, where “technical advantages of higher energies are better
the nearly isotropic radiation in an industrial gamma-ray power utilization, reduced treatment time and improved
facility has a wide angular distribution. Consequently, much dose uniformity” resulting in the benefits of throughput
of the gamma-ray emission is more divergent than a speed, reduced costs, and improved product quality.
high-energy bremsstrahlung beam and enters the products
at larger angles from the perpendicular direction. The Performance Qualification Comparison
outcome of this with regard to DUR performance is exempli- Improved DUR observed during OQ has been verified
fied in the following examples observed at STERIS Däniken. throughout performance qualification (PQ) studies on actual
product configurations. Case studies below show the differ-
Operational Qualification Comparison ence found on the same configuration when processing with
As shown in Figure 5, performance of operational qualifica- gamma and X-ray at Däniken. These case studies demon-
tion (OQ) of gamma and X-ray irradiators with identical strate how an improvement in DUR can yield additional
product density and process load show a significant gain in benefits to product processing, including more precise dose
DUR for X-ray. As an example of how to use this graphic, for a delivery, reduction in product temperature, and improved
pallet of medical device of density 0.25 g/cm3, and a minimal product throughput.
dose of 25 kGy, the maximum dose received by the process
load (pallet of 1.0 m × 1.2 m × 1.8 m = 2.2 m3) would be:
Figure 5. Dose uniformity ratio (DUR) as a function of density (g/cm3) for Däniken X-ray and gamma pallet irradiator.
Case study 1: Precise dose for product testing. Per Figure 7, Case study 2: Temperature sensitive product. In another
using a uniform configuration of UHMWPE, an 8% reduc- example, where a temperature-sensitive product was evalu-
tion in DUR was observed by moving from gamma to X-ray. ated, a DUR improvement from 1.47 to 1.22 was observed by
With improved DUR, the dose could be delivered to the moving process from gamma to X-ray. This lower DUR
product more precisely.
Figure 6. Relative radiation stability of medical polymer “families.” Adapted from AAMI TIR17.10
Figure 7. Gamma and X-ray DUR results from the treatment of UHMWPE at target doses of 50 to 150kGy.
Abbreviations used: DUR, dose uniformity ratio; UHMWPE, ultra-high-molecular-weight polyethylene.
I (c) 202028 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
(lower maximum dose = less energy absorbed by the product compatibility as lower maximum doses achieve the same
= lower temperature increase) coupled with a higher dose requirement of sterility assurance; and 4) improved product
rate resulted in reduced processing time and product outcomes with reduction in temperature during processing.
temperature. Furthermore, the levels of oxidation observed with X-ray
Case study 3: Increase of process load volume. As shown appear improved compared to those seen in gamma at the
in Figure 8, improved DUR results in more product being same dose levels.
processed on the processing unit: In this case, the volume of As highlighted by Plaček & Bartoníček,15 the oxidation of
product per pallet can be more than doubled. polymeric materials is “strongly influenced by the atmos-
phere in which they are irradiated,” such that the radiation
Processing Flexibility dose required for reaching a particular level of degradation
Many gamma facilities operate “shuffle and dwell” opera- changes with the dose rate. The authors observed that the
tions where the dwell period is a factor of the amount of rate of oxygen diffusion at higher dose rates is insufficient to
cobalt present and of the density of the products within the support the oxidation reactions, contrary to that observed at
irradiator. In such an operation, products of similar density lower dose rates.15 Impacts such as these require further
and dose requirements may be batched to achieve the in-depth investigation. Such investigative work will continue
desired result. Also, the transition effect from one density to to inform and impart knowledge and further develop the
another may need to be taken into account. As a consquence, understanding summarized in Table 1.
scheduling limitations may be experienced. As explained in As suppliers of X-ray irradiators bring innovations and
other sections of this article, X-ray processing, having the design concepts to the market, the variances in product flow,
benefit of being incremental and working over a reduced energy, power, and dose rate must be understood and
treatment area, results in no significant impact on the pallets evaluated. Regardless of such variances, the opportunity to
neighboring the processed pallet. Consequently, pallets of provide a viable, efficient, and economical alternative to
both different densities and target doses can be processed gamma radiation is a reality. This will help address the
simultaneously, offering improved flexibility over gamma. concerns regarding cobalt supply, transportation, and
security highlighted previously.3
Future Opportunities with X-ray
As described in this article, experience thus far with industri- Acknowledgments
al-scale X-ray radiation of healthcare products and materials The authors thank Emily Craven, cochair of AAMI ST/WG2
has yielded some very promising outcomes. For example, as on Radiation Sterilization for her valuable review and input
dose uniformity is improved, maximum dose is lowered, to this article. The authors also thank the Kilmer Collabora-
resulting in 1) improved economics as more individual tion Modalities group for the collaborative efforts considering
products are processed on the processing unit; 2) improve- and exploring how additional sterilization modalities may be
ment in processing efficiency and supply chain as more deployed and adopted by the healthcare products manufac-
products are processed more quickly; 3) greater material turing industry, for the benefit of patient care.
Figure 8. Example of gamma and X-ray pallet configuration. Because of improved dose uniformity ratio (DUR), a higher volume of product
per pallet can be processed with X-ray. Final pallet configurations in gamma and X-ray achieved similar DURs of 1.18 and 1.21, respectively.
Table 1. Summary of differences between radiation sources and applicability to sterilization of medical devices.
7. Meissner J, Abs M, Cleland MR, et al. X-ray treatment at 5 MeV and 15. Plaček V, Bartoníček B. The dose rate effect and the homogeneity
above. Radiation Physics and Chemistry. 2000;(57):647–651. of radio-oxidation of plastics. Nuclear Instruments and Methods in
Physics Research, Section B: Beam Interactions with Materials and
8. Lambert BJ, Mendelson TA, Craven MD. Radiation and ethylene Atoms. 2001;185(1–4):355–359.
oxide terminal sterilization experiences with drug eluting stent
products. AAPS PharmSciTech 2011;12(4):1116–1126.
I (c) 202030 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Figure 1. Image of an example gamma irradiator. © 2020 Nordion (Canada) Inc. All rights reserved.
in at least one of these areas. This article will randomly placed within the source rack as
discuss each area and approximate the the location directly impacts shape of the
impact to the gamma irradiator performance. radiation field. That is, the placement of the
Gamma irradiators employ two main Co-60 sources defines the locations of the
designs: minimum and the maximum activity and
1. Product overlap thus the locations of the minimum and
2. Source overlap maximum dose within the product.
Product overlap machines have product For example, consider a newly installed
stacks that are taller than the source rack. (less than 50% full of Co-60 sources) gamma
That is, the product stack begins below the irradiator of product overlap design. If 100
bottom of the source rack and ends above sources are installed in a configuration that
the top of the source rack. This design is the provides the gamma irradiator with its best
most efficient use of the radiation source as balance of throughput and dose uniformity
the source is almost entirely surrounded by within the product, we approximate a 7%
the products. increase in product throughput and a 4%
Source overlap machines use source racks improvement in dose uniformity within the
taller than the product stacks. That is, the product (using 0.1 g/cc product, based on
product stack begins above the bottom of the mathematical modeling*) vs. installation in a
source rack and ends below the top of the randomly distributed fashion (i.e., sources
source rack. The source overlap design has and activities evenly spaced out throughout
advantages, such as ease of scheduling and the rack).
changing of the product stack size to meet Now, let us consider a mature (over 50%
product uniformity requirements. However, full of Co-60 sources) source overlap gamma
cobalt efficiency is sacrificed. Figure 2 shows irradiator and perform the same calculation
a comparison of the two designs. as above. In this case, we see a 5% decrease
in throughput using the optimal source
Cobalt Efficiency arrangement vs. a random distribution. This
The Co-60 sources are the engine of the may seem counterintuitive; however; there is
gamma irradiator. In the gamma irradiation an 18% improvement in dose uniformity
process, the sources are contained in a within the product using the optimal source
simple mechanical device, called a source configuration (using 0.1 g/cc product, based
rack, that needs little adjustment while in on mathematical modeling).
operation. However, the location of the Co-60 The definition of “optimal” clearly depends
sources within the rack is critical. Co-60 on your objectives. As in the examples above,
decays at a rate of approximately 12% per the source overlap could rearrange the Co-60
year, requiring gamma irradiators to add new sources to achieve a higher throughput.
Co-60 sources on a regular basis, usually However, the resulting large negative impact
annually. These sources should not be this would have on dose uniformity to the
* In the context of this article, mathematical modeling is the creation of a virtual representa-
tion of the gamma irradiator using a computer program. This includes the product
container geometry and materials, the Co-60 activities and locations, and an approximation
of the product.
The geometry for the models used in this article was constructed based on Nordion
engineering drawings and the results were calculated using point-kernel methodology,
following the guidelines in the industry standard.2 Due to the limitation of the point-kernel
approach, all products and absorbers were approximated as rectangular cuboids, constructed
of homogenous material, and the product stack had density no greater than 0.4 g/cc.
For the cases modeled in this article, all models were validated by comparing model
results to data from existing Nordion-built gamma irradiators. The validation used data
from the operation qualification— which uses 0.01–0.4 g/cc homogenous product—and
followed ISO/ASTM 52303.3
I (c) 202032 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
product is undesirable and, in most cases, amount of cobalt in each given region
prevents most products from being irradi- (module) is defined. The activity in a given
ated. The optimal locations of the cobalt row or column is defined by the percent of
sources are either determined through exper- total activity. The amount of activity in a
iment (experience) or can be predicted given module is the percent activity in the
through mathematical modeling. row multiplied by the percent activity in a
To calculate the possible dose distributions column. To create the DoE, the amount of
within a given product, and thus throughput activity in each row is varied and the
and dose uniformity within the product, throughput and dose uniformity within the
mathematical modeling can be used. Using a product (dose uniformity ratio; DUR) is
“design of experiments” (DoE) methodology, calculated. The results can then be sorted to
many cases can be tested and sorted to determine the optimal activity distribution in
determine the optimal case for a given each module. Once the optimal distribution
gamma irradiator. is determined, this can be used as an input
The most common design to hold Co-60 to plan the Co-60 source installation.
sources is a planar source rack. This rack is Figure 3 shows a generalized example of a
made up of rows and columns. To create a DoE where each row, R, is varied (R-xy – R-xy),
DoE for this type of source design, the where x is the experiment number and y is
Figure 2. Source overlap (left) and product overlap (right). © 2020 Nordion (Canada) Inc. All rights reserved.
Figure 3. Generalized DoE for Co-60 activity for a given module. © 2020 Nordion (Canada) Inc. All rights reserved.
Abbreviation used: DUR, dose uniformity ratio.
the row number. For example, R-46 is the the product container guides the product,
amount of activity in row 6 in experiment 4. any modification to the container has an
This exercise can be repeated for each influence on the dose to the product.
column as well. For scale, there can be
thousands of experiments that need to be Source-to-Product Distance
run (i.e., there are thousands of possible The product container places the product
combinations of how to distribute the cobalt within the source pass (Figure 1). Modifying
within the rack). There are even more the position of the product container
possible combinations if we do not first limit requires substantial effort but has a large
to pragmatic distributions. influence on the dose rate and dose uniform-
ity within the product.
Product Container Efficiency In general, dose uniformity degrades as
The product container transports the you move closer to the source rack, as
product from the storage area into the source variation in the individual Co-60 sources and
pass. There are three common containers: product placement becomes more evident.
1. Tote—Usually an aluminum or cardboard Likewise, moving the product away from the
box that sits on a conveyor and is pushed sources smooths out any variations, allows a
into the source pass (Figure 4). more uniform radiation field, and improves
2. Carrier—An aluminum box, larger than a dose uniformity within the product.
tote, that hangs from an overhead rail and However, the opposite is true with the dose
has a keel on the bottom (Figure 5). This rate: As you move closer to the source, the
rail is used to guide the carrier into the radiation field becomes more intense. The
source pass, where the carrier bottom is higher dose rate allows the product to reach
secured into a keel guide. its minimum required dose faster, resulting
3. Pallet—Used for standard U.S. or European in more products moving through the
pallets where the product is stacked and gamma irradiator (i.e., higher throughput).
secured, often with shrink wrap. The pallet Figure 7 shows that as the product is moved
is placed onto a slave pallet, which is on a further away from the rack (to the right on
conveyor. The slave pallet guides the the x-axis), the throughput (blue) drops. It
product along the irradiation path (Figure 6). also shows that the uniformity (red); that is,
Each of the containers ensures that the the dose distribution within the product gets
product is placed into a reproducible posi- smaller, meaning more uniform.
tion relative to the source rack. Given that
Figure 4. Product tote. © 2020 Nordion Figure 5. Product carrier. © 2020 Nordion Figure 6. Slave pallet, pallet, and product stack.
(Canada) Inc. All rights reserved. (Canada) Inc. All rights reserved. © 2020 Nordion (Canada) Inc. All rights reserved.
I (c) 202034 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Product Overlap vs. Source Overlap container (typically a hanging carrier) must
The product overlap design uses Co-60 more be removed or modified to have two levels.
efficiently; however, in general, optimizing For example, a shelf could be added to the
the order of the product to be irradiated is carrier. Then, a product interchange would
more difficult with this design. The source need to be added to move the product
overlap design is more flexible regarding the between the top and bottom levels. A new
order in which products can be irradiated. installation qualification, operation qualifica-
Each of these designs can be seen in Figure 2. tion, and performance qualification (PQ)
A mathematical model was created for need to be completed after a modification of
each design type (product overlap vs. source this magnitude. This includes determining
overlap) with equal Co-60 activity; each was new product load configurations throughout
deemed to have an optimal Co-60 distribu- the PQ process. While large gains in perfor-
tion. For all common product densities used mance could be achieved, the cost and time
in gamma irradiators, the product overlap lost for the work would be significant.
design was calculated to be approximately
twice as efficient in terms of throughput and Mechanical Limitations
provide 3%–15% better uniformity, for Most gamma irradiators operate through a
densities 0.1–0.4 g/cc, respectively. “shuffle and dwell” design, meaning the
However, to achieve these gains, the product container dwells (is stationary) in
source pass equipment must be replaced. one position for a given time and then
This means that the single-level product shuffles to the next location. The cycle time
Figure 7. Size of throughput and uniformity change due to the source-to-product distance. © 2020 Nordion (Canada) Inc. All rights reserved.
Abbreviation used: DUR, dose uniformity ratio.
equals the total amount of time needed for a Additionally, electric drives require substan-
product container to shuffle and dwell. This tially less maintenance, which reduces the
means that the cycle time has a mechanical overall downtime of the gamma irradiator,
limitation (lower limit) for how long it takes thereby increasing overall efficiency.
for all of the shuffling to happen within the In some cases, product flow (the shuffle
source pass. time within the source pass) is the limiting
Mechanical limitations occur in two cases: factor of the cycle time. For example, gamma
1. The gamma irradiator must be run quickly irradiators that transport product on a
to meet product requirements. This can monorail (i.e., hanging systems) may use an
occur when products with a lower dose inefficient process flow that slows down to
specification need to be irradiated in a allow products to move between rows within
system with a higher dose rate. the source pass.
2. Incremental dose—Some gamma irradia- An example of removing a monorail and
tors cycle the product through the process replacing it with a cross-transfer is shown in
(through all dwell positions) multiple Figure 8. Here, the monorail was used to
times. This allows the operator to schedule move the product into and out of the cell.
low-dose products with high-dose prod- However, this is slow. This bottleneck can be
ucts and reduce the number of transition replaced with cross-transfers that allow a
products required. See Process Scheduling more efficient movement of the product
section below for more information. container and allow the overall process to
Examples of current approaches to move faster. In Figure 8, we can see that the
overcome mechanical limitations by modify- cross-transfer drops off to a different lane
ing the process include setting the target between product inlet/outlet. This allows
dose of the product higher than its required product to be more efficiently staged to enter
minimum and lowering one of the racks the gamma irradiator. Also, the cross-trans-
holding the sources to effectively lower the fer can move faster than the product
amount of activity in the gamma irradiator. container hanging on the monorail. In past
Both of these approaches solve the mechani- projects, we have seen up to a 30% reduction
cal limitation problem but reduce the in minimum cycle time by employing this
gamma irradiator’s efficiency. approach.
Mechanical limitations can also be over- Another design bottleneck can occur in the
come by modifying the equipment (e.g., by interim (maze) product flow. In some
replacing old pneumatic drives with newer designs, the incoming products and the
electric drives). Electric drives are more outgoing products share the same path. This
consistent than pneumatic drives, resulting requires the incoming product to wait for the
in shorter and more predictable cycles. outgoing product to exit before it can enter.
Figure 8. Source pass equipment change. Monorail (left); cross-transfer (right). © 2020 Nordion (Canada) Inc. All rights reserved.
I (c) 202036 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
For more than 50 years, Nordion has provided
Cobalt supply for the gamma processing industry. That means protecting patients and the
healthcare industry for another 50 years.
I (c) 202038 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
particularly “empties,” are normally of low or no value to the efficient scheduling is less versatile (cannot react easily to
gamma irradiator facility. quick turnaround products).
This transition product can have a large effect on the Table 1 presents an overview of the results of this article for
efficiency of a gamma irradiator. A typical gamma irradiator the potential increase in performance. However, it should be
loses 5%–15% of its throughput due to scheduling ineffi- cautioned that throughput is negatively correlated with dose
ciency. The best way to reduce the amount of transition time uniformity. That is, an improvement in throughput often
is to use a scheduler with many years of experience, who has results in degraded dose uniformity.
determined the optimum setup rules and who executes to To put some of the results into perspective, a mid-size
those rules perfectly every time. As the number of products gamma irradiator could have 3 MCi. A site that is able to
and dose ranges increase, optimal scheduling becomes improve its efficiency by 3% could process the same amount
nearly impossible for a person to accomplish. The only way of material while saving 90,000 Ci. Another way to see the
to guarantee optimal scheduling is to automate the schedul- gains would be to look at a 4-pass carrier design with 3 MCi,
ing process. An automated scheduling process would have which can process approximately 700,000 cubic feet of
access to all of the available products and rules, allowing it to material per year. An efficiency gain of 3% means that it can
optimally sort the products to reduce the required transition process an extra 21,000 cubic feet per year, which yields an
phases. extra 1,700 carriers per year.
A further optimization to the scheduling process is to
reduce the required amount of experiments to determine Acknowledgments
compatibility. This can be accomplished through a validated The author thanks the Nordion Engineering Team for
mathematical model that imports the product geometries, helping to identify examples to be included based on past
material, and dose requirements. The compatibility experi- work, and for reading this article to ensure accuracy and
ments can then be run virtually and used to create the rules consistency. The author also thanks the Nordion Design
that are fed back to the scheduler. Office for creating great figures on short notice.
Combining an automated scheduler and mathematical
modeling would allow nearly real-time addition of new References
products to the gamma irradiation process. 1. International Irradiation Association. Factsheet: Beneficial Uses
of Cobalt-60. http://iiaglobal.com/wp-content/uploads/2018/01/
Summary Factsheet_Cobalt_60.pdf. Accessed May 20, 2020.
This article has outlined many areas of possible improve-
2. ASTM E2232-16. Standard Guide for Selection and Use of
ment in the efficiency of gamma irradiators. Some of these Mathematical Methods for Calculating Absorbed Dose in Radiation
may be straightforward to implement, while others would be Processing Applications. West Conshohocken, PA: ASTM
a significant investment in time and cost. The performance International.
due to modifications in each of these areas is dependent on
the design of each irradiator and the product that is being 3. ISO/ASTM 52303:2015. Guide for Absorbed-Dose Mapping in
irradiated. Radiation Processing Facilities. West Conshohocken, PA: ASTM
All of the situations discussed present ideas to improve the International.
gamma irradiation process. However, it should be noted that
any change may have a tradeoff that should be considered.
For example, if a product container is made too thin, wear-
and-tear will become a problem; rearranging the cobalt
distribution may improve the throughput but will usually
negatively impact the product dose uniformity; and more
I (c) 202040 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
For example, if a minimum of 35 PCDs were data from routine production cycles, the EO
selected due to the usable chamber volume exposure time parameters can be deter-
and zero positives were observed following mined to more accurately define the EO
half-cycle exposure, then a SAL of approxi- exposure time required to deliver the desired
mately 10–1.5 is demonstrated using a SAL. Given that medical device companies
conservative D-value with the assumption of already have this data, a more accurate EO
one positive PCD. exposure cycle time can be a determined and
11135 Annex B, section B.1.2.b. provides documented. The documented analysis for
another qualification approach for calcula- the more accurate EO exposure time can be
tion of the EO exposure time for a routine used to support regulatory submission.
production cycle. This qualification method,
called the cycle calculation approach, uses Analysis
data collected over a series of sublethal cycles Before beginning the evaluation of cycle
to establish the PCD decimal reduction value data, the following prerequisites exist to
(D-value; i.e., a resistance value for steriliza- ensure proper analysis and associated
tion effectiveness). Sublethal cycles are conclusions:
exposure conditions where some of the 1. The cycles under evaluation must have
PCDs are expected to be positive for growth been validated in accordance with the
while others are expected to be negative for overkill half-cycle method per require-
growth. Using the PCD positives, an esti- ments in 11135.
mated PCD D-value can be determined. This 2. The relationship between the product
estimated PCD D-value is then utilized to bioburden and the PCD must be under-
determine the routine production cycle stood. Relative resistance and bioburden
parameters, providing the process parame- quantity must be less than that of the PCD
ters that will deliver the desired SAL. as previously defined.
The analysis presented in this article uses 3. The bioburden program for the product
both the data obtained from the overkill families associated with the sterilization
half-cycle approach and the foundations of cycle must be stable, with historical data to
the cycle calculation approach to determine demonstrate maintenance of bioburden
new EO exposure time parameters. Data quantity and resistance.
obtained from the overkill half-cycle qualifi- 4. The cycle stability during the gas injection
cation allows for a conservative estimate of and postexposure washes must be demon-
the PCD D-value. The D-value is considered strated. The analysis includes the time
conservative as one must assume one from the initiation of EO gas injection,
positive-growth when zero positives were through any inert gas injection (if used)
observed from the PCDs following exposure prior to the start of the EO dwell phase, the
to the overkill half-cycle. exposure time, and the postexposure
Following load conditioning (e.g., humidi- washes. It should be noted that variability
fication), cycle lethality occurs upon injection in the time and rate of injection and wash
of the sterilant into the chamber and contin- phases will have to be considered to
ues through the entire process. An equation establish worst-case conditions.
established by Mosley et al.3 can be used to
calculate the equivalent exposure time (U)
for an EO sterilization process, taking into The analysis presented in this article uses both the data obtained
consideration the lethality achieved during from the overkill half-cycle approach and the foundations of the cycle
gas injection through exposure and the gas calculation approach to determine new EO exposure time parameters.
evacuation phases. Data obtained in the
overkill half-cycle MPQ and calculation of
equivalent exposure time will allow a more
accurate understanding of the delivered cycle
lethality.
Using the equivalent exposure calculation
U from the overkill half-cycle MPQ data and
Evaluation of four different EO sterilization cycles from using the Stumbo-Murphy-Cochran Procedure4 with the
four different medical device companies was performed for following formula:
calculating equivalent exposure time. The approach used the
Mosely et al. equation for calculating equivalent exposure Ui
time. DT = ___________________
ni
log10N0 – log10�ln� __��
n ri
1 Ci
U = ∑ (anti log(log tT + __ (Ti – Tref)) ___ DT = D-value
z Cref
i =1
Ui = total equivalent exposure
Ui = equivalent exposure time for a given time interval
N0 = initial population
tT = time interval (t) at temperature (T)
ni = number of units tested
z = z-value (30°C)
ri = number of units sterile
Ti = temperature
Tref = reference temperature
Using the D-value (DT ) and the equivalent exposure time
(U) obtained from the previous calculations, a spore log
Ci = EO concentration reduction (SLR) was determined using the following formula:
Cref = reference EO concentration
SLR = U/DT
In Table 1, the reference EO concentration (Cref ) was
derived from the overkill half-cycle MPQ data. The EO Note: This formula is a derived from the SLR formula in
concentration calculations include total partial pressure of ANSI/AAMI/ISO 11138-7:2019.4
EO injected into the chamber and use the average chamber
temperature from the MPQ cycle documentation. The Of the three overkill half-cycle MPQ documentation, the
reference gas concentration was calculated using the follow- MPQ half cycle with the longest equivalent exposure time
ing formula: was used for the analysis in Table 1. This was selected as the
longer equivalent exposure time results in a longer estimated
Cref = n*(PEO)/R*(Tref + 273) D-value (i.e., minutes). A longer estimated D-value equates
to a conservative estimate of the PCD resistance, and
n = 44,000 mg/mole (molecular weight of EO) therefore provides a conservative estimate of the SAL based
PEO = partial pressure of EO in chamber after injection on exposure time.
Data from a typical routine production cycle was used in
R = 62.361 mmHg-L/gm-mole-Kelvin
the establishment of the routine process SAL. SLR value was
Tref = reference temperature (converted to Kelvin) used along with the microbiological challenge population
(N0 ) to calculate the production cycle SAL.
The reference temperature (Tref ) was also based on the
average chamber temperature from the MPQ documenta- SAL = 10�Log(N0 ) – SLR�
tion. The rationale for using these data as the reference
values for the formula is that the MPQ runs are conducted at Note: This formula is derived from the SAL formula in
lower temperatures and partial pressure of EO when com- 11138-7. MPQ SAL is derived from MPQ data for U, N0 ,
pared to the routine production cycles. Using the MPQ data and DT . Routine production SAL is derived from MPQ N0
to set the reference values adds to the conservative nature of and DT , and routine production cycle U.
the evaluation.
The EO concentration (Ci ) and temperature (Ti ) come Data in Table 1 represents the time from EO charge
from the run record data. Ci represents the calculated EO through the end of the process.
concentration for that time stamp entry. Ti represents the It should be noted that Cycle A has a longer equivalent
average chamber temperature for that same time stamp exposure time than the actual run time for the production
entry. Each time stamp entry on the run record correlates to cycle. This happens as the production cycle uses a higher
a new Ci and Ti used in the equation. partial pressure of EO than that used during the MPQ cycle.
D-values were calculated for each cycle using the MPQ Since the EO reference concentration is derived from the
data. As previously mentioned, all PCDs were negative for MPQ cycles, the higher EO concentration in the production
growth in the MPQ cycles, therefore a single positive PCD cycle equates a longer equivalent exposure time for each time
was assumed in order to calculate an estimated D-value stamp during EO dwell. When accumulated throughout the
I (c) 202042 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
cycle, a longer equivalent exposure time than Similarly, use of a microbiological challenge
the actual run time is achieved. For the other that is more consistent with the production
three cycles, only slight increases in EO partial bioburden (i.e., quantity and resistance)
pressure is seen during production cycles. would also allow for cycle development with
Based on this analysis and a required SAL decreased exposure time and/or EO concen-
of 10–6, Cycles A, C, and D have between 5.6 tration. This second method is referred to as
to 6.8 logarithms (logs) of additional inactiva- the bioburden/biological indicator method.
tion. Therefore, these three cycles are Qualification activities using the overkill
candidates for utilizing the proposed meth- half-cycle approach or calculation approach
odology to reduce exposure times while still can take existing cycles and convert them to
achieving the desired SAL. This information cycles using lower EO concentrations. In
could reduce cycle exposure times by up to some instances, this may result in longer
40%. Table 2 provides potential reduction in sterilization cycles but reduce EO utilization.
EO dwell time settings for each cycle. The analysis presented in this article uses
data that is already available and could allow
Conclusion for immediate exposure time reduction
There are multiple approaches to reducing while working through qualifications with
EO utilization for the sterilization of medical other methods.
devices. The 11135 standard provides other Using the proposed approach, in lieu of
approaches to qualify the sterilization cycle solely relying on the half-cycle overkill
in addition to the overkill half-cycle and cycle approach, to calculate new exposure times
calculation methods described earlier. One of can provide potentially significant overall
these approaches—referred to as the biobur- reduced routine production cycle time.
den approach—uses the product bioburden Reduction in exposure time could also
instead of a resistant microbiological reduce overall product absorption of EO,
challenge to demonstrate process lethality thereby reducing the amount of EO that
and would allow for use of shorter exposure needs to be removed from the product
times or reduced EO concentration. during the post EO dwell remainder of the
Table 1. Equivalent exposure time with Mosley et al.3 equation and sterility assurance level (SAL) impact. Abbreviation used: MPQ, microbiological
performance qualification.
Table 2. Potential dwell time reductions. Abbreviation used: SAL, sterility assurance level.
References
1. ANSI/AAMI/ISO 11135:2014, Sterilization
of health care products—Ethylene oxide—
Requirements for the development, validation and
routine control of a sterilization process for medical
devices. Arlington, VA: Association for the
Advancement of Medical Instrumentation.
AAMI eSubscription
The Easier Way to Access Standards
An eSubscription allows multiple individuals across your enterprise
to easily access standards. Whether you need a particular set of
standards—such as sterilization—or a wide range, you can choose a
plan that meets your organization’s needs.
I (c) 202044 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
sample size selected for BI testing was reduced from the aseptic transfer to a test tube containing 10 mL sterile water,
requirements for Method 2A. In addition, the number of and the paper carrier was homogenized using a sterile pestle.
incremental doses was reduced due to the known resistance The BI suspensions were serially diluted in sterile water to
of the B. pumilus spore. The stability of the BI over the time obtain concentrations of approximately 30 to 300 colony-
of the test period was demonstrated using population counts forming units (CFU) per milliliter. Pour plates were prepared
prior to and following testing. This species was selected using molten SCD agar and incubated at 30 to 35°C for three
based on its demonstrated, consistent response to radiation, to five days. Upon completion of incubation, the plates were
due to a higher radiation resistance than other Bacillus enumerated and the surviving populations documented.
species. It is important to note that the use of a spore
challenge model for validation of a product sterilization dose Population Controls
is not recommended (11137-1, sections 1.2.3 and A.1.2.3).3 The microbial populations present on the bandage material
The range of radiation dose rates and energies selected for (i.e., bioburden) and BI paper carriers were determined
this study are typical of those that might be utilized during preceding and subsequent to completion of radiation
verification testing or routine sterilization processing. The processing to demonstrate population stability of the test
radiation resistances of the test articles were assessed using articles throughout the test period.
gamma radiation (emits two rays with energies of 1.17 and The bioburden population present on nonirradiated
1.33 MeV) delivered at two separate dose rates (0.37 and bandages was determined using a validated bioburden
12.9 kGy/h) and E-beam radiation (10 MeV) delivered at two recovery procedure, where 10 samples from each of three
separate dose rates (3,100 and 36,000 kGy/h). batches of bandages were tested. Each bandage was
immersed in sterile diluent and placed on a shaker table.
Methods The diluent with test samples were shaken at approximately
The test articles consisting of nonwoven cellulosic bandages 450 rpm for 15 minutes. Serial dilutions were performed
(2 × 3.5 inches) from each of three batches were individually from the recovery fluid using sterile water to obtain concen-
packaged in pouches, and B. pumilus (ATCC 27142) BI paper trations of approximately 30 to 300 CFU/mL. Pour plates
carriers from each of three lots were individually packaged in were prepared using molten SCD agar and incubated at 20 to
pouches. All test articles were submitted for irradiation. 25°C for three days, followed by incubation at 30 to 35°C for
Test articles were irradiated together for each of the doses two days. Upon completion of incubation, the plates were
selected. Twenty bandages from each of three batches and enumerated and the initial bioburden populations docu-
five BIs from each of three lots were irradiated together. The mented for each batch of bandages (Table 1).
bandages were irradiated at nine incremental doses (2, 4, 6, The population counts on nonirradiated BIs were deter-
8, 10, 12, 14, 16, and 18 kGy), and the BIs were irradiated at mined from five BIs from each of the three lots of test
six incremental doses (2, 4, 6, 8, 10, and 12 kGy). Each of the articles. Each BI was aseptically transferred to a sterile
incremental doses was delivered using four dose rates: (1) blender jar containing 100 mL sterile water and homoge-
gamma radiation at a dose rate of 0.37 kGy/h, (2) gamma nized for one minute. The contents of each blender jar were
radiation at a dose rate of 12.9 kGy/h, (3) E-beam radiation at serially diluted using sterile water to obtain concentrations of
a dose rate of 3,100 kGy/h, and (4) E-beam radiation at a dose approximately 30 to 300 CFU/mL. The dilution tubes then
rate of 36,000 kGy/h. were exposed to a heat shock at 80°C for 10 minutes. Pour
For consistent measurement of the irradiation doses, a plates were prepared using molten SCD agar and incubated
common dosimetry system was used. The irradiation dose at 30 to 35°C for two to three days. Upon completion of
delivered was measured using FWT-60 radiochromic film incubation, the plates were enumerated and the populations
dosimeters (Far West Technology, Santa Ana, CA). During documented (Table 2).
several of the irradiation runs, alanine reference dosimeters
from the National Physical Laboratory (Middlesex, UK) were Test of Sterility Negative System Controls
placed side-by-side with the FWT-60 film dosimeters to verify Negative controls were conducted to demonstrate that
the traceability of the Far West Technology films. positive tests of sterility observed were attributed to the
After irradiation, all test articles were returned to the product tested and not due to testing and/or laboratory issues.
laboratory for testing. A test of sterility was conducted on Ten or 20 bandages irradiated at 50 kGy or greater were
each bandage by aseptic transfer into a container of sterile tested each day using the same methods applied for the test
soybean casein digest (SCD) broth. The SCD containers were articles to confirm the aseptic technique of the technicians
incubated at 28 to 32°C for 14 days per 11137-2.1 The containers performing the product tests of sterility for the Method 2A
were periodically examined and the results documented as incremental doses. A total of 200 samples were tested for the
the number of units positive for growth/total number of units two tests of sterility teams, and zero positive product tests of
tested. A population count was conducted on each BI by sterility were observed for the negative controls.
I (c) 202046 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Method Suitability contents of each tube were pour plated using molten SCD
Method suitability testing was performed to verify that the agar. The plates were incubated at 30 to 35°C for a maximum
bandage and BI paper carrier materials did not alter the of three days and enumerated (Table 4).
growth-promoting properties of the culture media to the
extent of preventing or inhibiting outgrowth of microorgan- Results and Discussion
isms, if present on the test articles.
Six bandages were irradiated at 50 kGy or greater and each Bandages
immersed into a tube of sterile SCD broth. Duplicate broth The product bioburden data demonstrate that the population
tubes were inoculated with not more than 100 CFU each of remained stable over the course of the study (Table 1), and
the following microorganisms: Bacillus subtilis ATCC 6633, therefore, no impact to the resistance analysis occurred as a
Candida albicans ATCC 10231, or Aspergillus niger ATCC result of die-off of bioburden organisms.
16404.4 The broth tubes were incubated at 28 to 32°C for a The first no positive (FNP) dose, first fraction positive (FFP)
maximum of seven days and examined for turbidity dose, difference between FNP and FFP doses, and DS kGy
(Table 3). (results and calculations from the Method 2A experiment)
Five B. pumilus BIs were irradiated at 50 kGy or greater, are provided in Table 5, and the d* kGy values (an initial
immersed into tubes of sterile water, and homogenized estimate of the dose required to achieve a sterility assurance
using sterile pestles. Each tube was inoculated with 10 to 100 level (SAL) of 10–2 for an individual product batch) for the
CFU of B. pumilus ATCC 27142 spores, and the entire four dose rates evaluated are provided in Table 6.
Table 1. Bioburden population on nonwoven cellulosic bandages at the beginning and end of testing. *Average bioburden = ∑ bioburden populations for
test articles 1-10/10. †Correction factor is calculated to be 1.8 in the bioburden recovery validation. ‡Corrected average bioburden = (average bioburden)
× (correction factor).
Table 2. Bacillus pumilus biological indicator population counts at the beginning and end of testing. *Δ = log of average population (end of testing) – log
of average population (beginning of testing). Abbreviations used: BI, biological indicator; CFU, colony-forming unit.
Table 3. Method suitability test results for nonwoven cellulosic bandages. Abbreviations used: CFU, colony-forming unit; NA, not applicable.
Table 4. Method suitability test results for Bacillus pumilus. Abbreviations used: BI, biological indicator; CFU, colony-forming unit.
I (c) 202048 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
COMING SOON!
AAMI
TIR76:2020
Sterilization of health care products—Radiation—
Substantiation of a selected sterilization dose at a
specified sterility assurance level: Method VDmaxSD-S
AAMI
TIR12:2020
Designing, testing and labeling reusable medical
devices for processing in health care facilities: A
guide for medical device manufacturers
The DS kGy is an estimate of the dose required to inacti- The D10 values (radiation dose required to reduce a micro-
vate 90% of the organisms surviving a 10–2 SAL dose (i.e., bial population by 90%) for the B. pumilus BIs irradiated
verification dose experiment). This provides an estimate of using two irradiation sources (gamma and E-beam) and four
the most resistant portion of the product bioburden and is an dose rates (0.37, 12.9, 3,100, and 36,000 kGy/h) were calcu-
estimate using a composite of the three batches tested for lated using linear regression and are reported in Table 7. The
each of the four dose rates. As the study design was not data were analyzed using an ANOVA statistical technique
intended to determine a sterilization dose, the verification (Figure 2). The analysis indicated that no significant differ-
dose experiment was not conducted. Therefore, the calcula- ence (95% confidence) in the D10 values existed for the
tion for DS kGy was conducted using the following B. pumilus BIs that could be detected over a 5-log difference
assumptions: D**kGy = DD*kGy; CD* = 2+/100; and FNP = in dose rates across the two radiation source types (gamma
DD*kGy. No difference in DS kGy values (Table 5) can be and E-beam).
observed over the 5-log difference in dose rates across the
two radiation source types (gamma and E-beam). Conclusion
A further evaluation was conducted by reviewing the d* Based on the analysis of the data, no significant differences
kGy data for each individual product batch (Table 6) using an could be detected in the rate of microbial lethality across the
analysis of variance (ANOVA) statistical technique (Figure 1). 5-log difference in dose rates evaluated for the natural
The analysis included a Tukey comparison (95% confidence) product bioburden or the BIs across the two radiation source
of values and this comparison indicates there is no signifi- types (gamma and E-beam).
cant difference in d* kGy values for the bandage material This data indicate that the radiation resistance of microor-
that could be detected over a 5-log difference in dose rates ganisms is not affected by any slight differences in energy
across the two radiation source types (gamma and E-beam). levels and dose rates of the radiation sources typically used
by sterilization facilities. Because of this, it can be concluded
BIs that the sterilization and verification doses can be safely
The BI population count data demonstrate stability over the transferred between modes of irradiation, as well as irradia-
course of the study (Table 2), and therefore, no impact to the tion facilities, without requiring proof of equivalent
resistance analysis occurred due to the stability of the BI microbial inactivation. As long as proof exists that the
population. specified dose is delivered, dose is dose.
Table 5. Nonwoven cellulosic bandages, Method 2A, first no positive (FNP), first fraction positive (FFP), FNP minus FFP, and DS kGy. *FNP is an estimate
of the dose at which only one sample of 100 irradiated samples is expected to be nonsterile. †FFP is an estimate of the dose at which only one sample of
20 irradiated samples will be nonsterile. ‡FNP minus FFP is a portion of the formula to determine DS, where (FNP – FFP) is less than 10 kGy, the formula
is DS = 2 + 0.2(FNP – FFP), and where (FNP – FFP) is greater than or equal to 10 kGy, the formula is DS = 0.4(FNP – FFP). §DS kGy is an estimate of the
dose required to inactivate 90% of the organisms surviving the verification dose (estimated 10–2 SAL). For the purposes of this study, the sterilization
dose was not established. Therefore, the calculation for DS kGy was conducted using the following assumptions: D**kGy = DD*kGy, CD* = 2+/100, and
FNP = DD*kGy, where D**kGy is the initial estimate of dose required achieve a SAL of 10–2, DD* is the actual dose delivered to the 100 samples for the
verification dose experiment, and CD* is the number of positive tests of sterility from samples exposed to the verification dose.
d* kGy
Dose rate (kGy/h) 0.37 12.9 3,100 36,000
Bandage batch no.
1 17.5 20.0 18.6 20.0
2 10.2 12.6 12.1 19.3
3 12.0 16.4 18.6 19.3
Table 6. Nonwoven cellulosic bandages, Method 2A, and d* kGy values (an initial estimate of the dose required to achieve a sterility assurance level of
1 × 10–2 for an individual product batch).
I (c) 202050 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Summary
Groups Count Sum Average Variance
0.37 3 39.7 13.23333 14.46333
12.9 3 49 16.33333 13.69333
3,100 3 49.3 16.43333 14.08333
36,000 3 58.6 19.53333 0.163333
ANOVA
Source of Variaon SS df MS F P F crit
Between groups 59.55 3 19.85 1.872494 0.212594 4.066181
Within groups 84.80667 8 10.60083
Total 144.3567 11
Figure 1. Statistical analysis for nonwoven cellulosic bandages using analysis of variance (ANOVA) to compare d*
kGy values between different dose rates. Analysis was performed using data from Table 6. *ANOVA for D value
(kGy) using adjusted sum of squares (SS) for tests. †S = 3.25589, R2 = 41.25%, R2 (adj) = 19.22%. ‡Grouping
information using Tukey method and 95.0% confidence. Means that do not share a letter are significantly
different. Abbreviations used: df, degrees of freedom; MS, mean squares.
Table 7. Bacillus pumilus biological indicator (BI) radiation resistance values calculated using linear regression. D10
values were calculated using the average starting population and average surviving population for each of four
incremental doses (0, 2, 4, 6, and 8 kGy) using the actual dose delivered to each sample set.
References
1. ANSI/AAMI/ISO 11137-2:2013/(R)2019.
Sterilization of health care products—Radiation—
Part 2: Establishing the sterilization dose. Arlington,
VA: Association for the Advancement of Medical
Instrumentation.
Summary
Groups Count Sum Average Variance
0.37 3 3.8 1.266667 0.003333
12.9 3 4.2 1.4 0.01
3,100 3 4.1 1.366667 0.003333
36,000 3 3.7 1.233333 0.003333
ANOVA
Source of Variaon SS df MS F P F crit
Between groups 0.056667 3 0.018889 3.777778 0.058939 4.066181
Within groups 0.04 8 0.005
Total 0.096667 11
Figure 2. Statistical analysis for biological indicators using analysis of variance (ANOVA) to compare D10 values
between different dose rates. Analysis was performed using data from Table 7. Abbreviations used: df, degrees of
freedom; MS, mean squares; SS, sum of squares.
I (c) 202052 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Thus, a calibrated dosimetry system provides ates as expected and the resulting measured
the best measurement of absorbed dose and, dose is within the predicted limits of the
therefore, values from dose measurements target dose, the process can then be consid-
should not be corrected by associated ered in control. The measured doses are
measurement uncertainty (see ISO 11137-3: used as a means for determining process
2017, section 4.1.3).4 acceptance and releasing the product. This is
Dosimetry systems have several compo- referred to as “dosimetric release.”
nents of uncertainty that may manifest
during routine radiation processing (i.e., Radiation Processing Measurements
uncertainties due to dose rate or tempera- The analysis of a routine monitoring
ture) and must be characterized for the dosimeter(s) can be used to indirectly
conditions of use. determine whether a product has received
the required sterilization dose without
Irradiator exceeding the maximum acceptable dose
The overall reproducibility of the irradiator is determined for that product. The purpose of
the final critical component for process the new ISO/TS 11137-45 standard is to
control. An irradiator delivers the dose to provide guidance on how to analyze and
products and has several critical components interpret this measurement.
(e.g., source, conveyance, irradiation path- A single, routine dosimeter measurement
ways, etc.) that also must be characterized to in isolation can be interpreted several ways.
determine the appropriate processing Repeated measurements provide some
parameters and conditions for processing a information on the range of doses that can
product. This process characterization, or be expected over time. Routine dose meas-
operational qualification, requires the urement made in the context of a desired
operator to understand the processing limits, target dose range provides information on
expected variability (common cause variabil- whether the process is in a state of control.
ity), and overall reproducibility of the In industrial radiation sterilization, the
radiation process. This characterization uses interpretation of these individual dose
a calibrated dosimetry system to measure the measurements in relation to product is used
process output; therefore, the user must be to set up target doses and establish and
careful not to confuse measurement system monitor an ongoing radiation process. There
uncertainty with radiation process variability is potential for confusion due to variations
and vice versa. A lack of understanding of that are observed and whether they are
the sources of variability and whether they related to what is going on in the product.
manifest during the process may lead to dou- ISO/TS 11137-4 provides information on
ble counting in the assessment of process the sources of variation that may contribute
variability. to the range of doses seen during a routine
process (Figure 1). These sources relate to
Radiation Process Control both the process itself as well as our ability to
Routine dosimetry is used to determine measure the process.
whether a radiation process is predictable,
stable, and consistently operating at the Process Variation
target level of performance. Dosimeters are Variation is caused both by things we actively
placed at defined locations within irradiation control and things that are beyond our
containers at predefined frequencies, and control, and can be anticipated (i.e., common
measured to evaluate whether the qualified cause) or unanticipated (i.e., special cause).
process delivered the predicted range of The factors that affect the output of a
absorbed dose for a predetermined loading radiation process will depend on whether it
configuration. A term that is typically used is gamma, electron beam, or X-ray, but
for a radiation process is target dose. This is ultimately, three factors are at play:
the dose that the radiation process parame- • The intensity of the radiation activity or
ters are set to deliver at a specified energy (the activity of the source or the
monitoring location. If the irradiator oper- power of the beam)
I (c) 202054 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
IT TAKES IMAGINATION
TO MAKE SCIENCE MORE
GROUNDBREAKING.
That’s the difference between making devices and
making progress.
An uncompromising pursuit of excellence. An unwavering focus on solutions
that will make the most difference for patients. Intently listening and collaborat-
ing side-by-side with clinicians to understand their most complex problems.
These are the principles for every product, solution and relationship we build.
It’s what we do every day to advance science for patients, for life.
To learn more about what it takes to advance science for life,
visit www.bostonscientific.com
© 2020 Boston Scientific Corporation and its affiliates. All rights reserved.
CORP-421303-AA JUL2020
I (c) 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
• The distribution of the radiation (the shape measurement of dose due to the equipment
of a scanned beam or placement of sources) used to measure it or the dosimeter itself,
• The path of exposure of the product to the and variations in the dosimeter placement or
source of radiation (conveyor speed or products surrounding it.
shuffle and dwell timing and positions) Additional components of measurement
Based on these factors, key parameters can uncertainty may or may not be apparent,
be identified that have a direct effect on dose including components relating to calibration,
to product if they are varied during the as well as influence quantities such as
radiation process. temperature, which may only be observed
ISO/TS 11137-4 Table 2 provides a list of seasonally.
process parameters that are critical to
radiation sterilization, the effect of the Establishing Radiation Process Parameters
variation of these parameters, and how they Minimum and maximum doses to product
are monitored as part of the process. The table for the process and their expected variation
provides a starting point for guidance on are determined through replicate direct dose
monitoring process parameters, in addition to measurements made during process qualifi-
dosimetry, to ascertain that product has been cation (e.g., product dose mapping). A
processed according to the specifications radiation process can be established based
stipulated in ISO 11137-1:2006/(R)2015, A.10.6.6 on these dose measurements. Each measure-
ment of dose made during dose mapping is
Measurement Uncertainty made with a calibrated dosimetry system and
There are components of measurement a known level of uncertainty.
uncertainty that will contribute to variability In order to determine appropriate process-
seen in a routine dosimetry measurement. ing parameters, the variations determined
These include variability inherent in the from the dose mapping data are used in
Figure 1. Components of process variability (σprocess ). ©ISO. This material is adapted from ISO/TS 11137-4:2020, with permission of the
American National Standards Institute (ANSI) on behalf of the International Organization for Standardization. All rights reserved.
I (c) 202056 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Therefore, when the uncertainty associated expected rather than a means to calculate an
with this indirect measurement is used to indirect measurement of maximum or
establish process parameters, which ensures minimum product dose.
that the minimum and maximum doses to the
product are achieved, the process can become Case Study: Product Dose
extremely conservative and restrictive. Measurement vs. Process Dose
There is nothing wrong with an approach Measurement
that is conservative and provides a high level The following is an example of data from a
of confidence; however, when you have tight performance qualification study (i.e., dose
process specifications or where there is a mapping) needed to establish routine
requirement to optimize a process, there is radiation processing parameters for a
an equally acceptable alternative. medical device. The data (Table 1) represents
the variation that is expected (i.e., common
Measurement of Process Dose cause or planned variation) during routine
Routine dose measurements can also be processing. The data was evaluated to
used as a process monitor where the determine minimum and maximum routine
expected variation of the routine dose dosimetry locations and to establish the base-
measurement can be utilized to determine line expectations for the radiation process
whether a process ran as expected, is in a (Figure 2).
state of control, and delivers the specified
product dose. When interpreting the repli- Product Dose Measurement
cate dose measurements made during the During routine processing, a reference dose
process qualification, rather than looking at location—which is neither the minimum
the ratios of the individual measurements of nor maximum dose location—will be
dose, a probability distribution function asso- monitored and measured. A calculation of
ciated with measurements of routine, the dose to product (indirect measurement)
maximum, and minimum doses can be will be made using reference dose ratios
developed to establish a predictable range of (including their uncertainty) and the refer-
expected dose measurements for a given set ence location dose measurement to determine
of process parameters. conformance to the product specifications.
The measured dose in routine processing The reference ratios, along with other
becomes a verification that the process ran as components of variability and uncertainty, are
I (c) 202058 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
then used to evaluate and establish process In this example, a coverage factor (k) of 2,
target doses that will ensure a process can representing a 95% confidence for a two-
routinely deliver the required minimum and sided distribution, was used. The expected
maximum doses to the product. range of maximum, minimum, and moni-
toring doses for this process are shown in
Minimum dose reference ratio (k=2) Figure 3.
= average - (2σ) = 0.86 - (2 × 0.023) The indirect measurement considers the
= 0.814 largest potential uncertainty that may have
occurred in the ratio and applies it in each
Maximum dose reference ratio (k=2) indirect measurement, whether it occurred
= average + (2σ) = 1.11 + (2 × 0.021) or not. The resultant calculations are used to
= 1.152
determine whether the processed product
meets the product specifications:
Figure 3. Expected probability distribution functions for the sample process when measuring product dose. Abbreviation used: Spec, specification.
Minimum product dose = routine monitoring dose × minimum dose reference ratio
Maximum product dose = routine monitoring dose × maximum dose reference ratio
= product is acceptable
These calculations are used to: a) set a process, a statistical analysis of various
process target dose, b) determine whether parameters from the process can provide a
your process is capable; and c) evaluate high degree of confidence the radiation
routine processing runs and determine process was executed as planned. For this
acceptance for release. In the example, the example, the process establishment utilized
target dose of 31.0 kGy yields an acceptable five replicate process runs, and the dose
product dose. This target dose can then be delivered to product (the absolute minimum
adjusted to determine the appropriate target and absolute maximum) and a routine
dose processing range. monitoring location were monitored. The
variability in the dose measurements was
Process Dose Measurement assessed and an expected range of doses
Using the same data set, another acceptable from future processing runs (under the
means of determining whether your product same process conditions) can be predicted
met its acceptance criteria is to verify the and establishes the baseline for determining
process ran as expected and was under a whether the process is reproducible and
state of control. In evaluating the output of a repeatable.
Expected range minimum dose (k=2) = average – (2σ) = 27.8 – (2 × 0.239) = 27.3 kGy
Expected range maximum dose (k=2) = average – (2σ) = 36.2 – (2 × 0.274) = 35.7 kGy
Expected range monitoring dose (k=2) = average – (2σ) = 32.5 – (2 × 0.770) = 31.0 kGy
I (c) 202060 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Figure 4. Expected probability distribution functions for the sample process when measuring process dose. Abbreviations used: PDF, probability
distribution function; Spec, specification.
Figure 5. Inputs and steps in establishing a process target dose. ©ISO. This material is adapted from ISO/TS 11137-4:2020, with
permission of the American National Standards Institute (ANSI) on behalf of the International Organization for Standardization.
All rights reserved.
I (c) 202062 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
• Options to choose the process control approach based on 3. ISO/ASTM 51707:2015. Guide for estimation of measurement
the amount of information known about the irradiator. uncertainty in dosimetry for radiation processing. Geneva: International
• The ability to accommodate innovative products that may Organization for Standardization.
require a tighter dose range. 4. ANSI/AAMI/ISO 11137-3:2017. Sterilization of health care products—
The use of process measurements provides an acceptable Radiation—Part 3: Guidance on dosimetric aspects of development,
alternative to traditional methods that require an indirect or validation and routine control. Arlington, VA: Association for the
direct measurement of minimum and maximum dose as a Advancement of Medical Instrumentation.
measure of conformance. Identifying sources of variability
provides an avenue to improve and optimize these processes. 5. ISO/TS 11137-4:2020. Sterilization of health care products—
Either method described in this article provides the accept- Radiation—Part 4: Guidance on process control. Geneva: International
able dose range for the product to meet its specification, but Organization for Standardization.
there are advantages of the process dose approach for the 6. ANSI/AAMI/ISO 11137-1:2006/(R)2015 & A1:2013 & A2:2019.
industry utilizing ISO/TS 11137-4. Overall, this new guidance Sterilization of health care products—Radiation—Part 1: Requirements
document can be used to broaden the application of radiation for development, validation, and routine control of a sterilization process
sterilization for products with challenging dose specifica- for medical devices. Arlington, VA: Association for the Advancement
tions and allow more efficient operation for irradiators. of Medical Instrumentation.
References
1. WebFinance Inc. Definition of “process control.” www.business
dictionary.com/definition/process-control.html. Accessed June 5,
2020.
AAMI Connect
Your one-stop shop for online
discussion groups!
• Connect with healthcare technology
professionals from around the globe.
• Have a question? Get it answered here!
• Share resources.
• Participate in discussions on key issues.
About the Authors Abstract out the revision history of the standards.
Trabue Bryans The validation of a radiation sterilization dose The terms implied throughout the radia-
is president at involves an initial sterilization dose determina- tion sterilization documents for the TOS
BryKor LLC in tion as well as maintenance of that sterilization outcomes of the verification experiment
Marietta, GA.
dose. The procedures for maintenance of the performed during a dose audit are “accept-
Email: trabue@
brykor.com
sterilization dose typically include the periodic able verification” and “unacceptable
use of two types of tests: bioburden and dose verification,” although one clause is titled
audits. The details for the procedures are “Failure of a sterilization dose audit.”2,3 In
Lisa Foster outlined in the ISO radiation sterilization the strictest sense, the term “failure” should
is principal at
standards. These documents also provide not apply to the initial positives in a TOS but
Adiuvo QS &
SA Consulting guidelines for recommended actions in response should only apply when those positives are
in Fairhope, AL. to the results of the two tests. The results for the determined to be true survivors of the
Email: lisafoster@ dose audit are based on the number of positive verification dose and exceed the acceptable
sterilization tests of sterility (TOS) for products that have limits described in the standards. However,
consulting.net
been irradiated at a verification or experimental the terms widely used in the healthcare
Kimberly dose. When the dose audit yields TOS positives, product industry are “pass” and “fail” of the
Patton is it is often thought that they indicate a steriliza- dose audit based on positive TOS results. For
WW sterilization tion failure and nonsterile product. The belief this discussion, and to be more in keeping
program that any TOS positive is a failure is an incorrect with industry understanding, the indus-
manager,
assumption because of the statistical basis used try-accepted terms of “pass” and “fail” will be
corporate global
sterility assurance for the determination of the sterilization dose. used here, as well as the terms “acceptable”
at BD in Sumter, SC. Email: This article will outline the truth of what dose and “unacceptable.”
kimberly.patton@bd.com audit TOS positives mean in terms of the When failure of a dose audit occurs, many
sterility assurance of product, as well as the companies typically assume the TOS posi-
Joyce M.
Hansen is vice
consequences of TOS positives. tives are indicative of a catastrophic failure.
president of The natural inclination is to assume the dose
microbiological The validation of terminal sterilization using audit is directly evaluating the routinely
quality & sterility radiation involves the establishment of an sterilized product and, consequently, to
assurance at initial sterilization dose as well as routine assume a failed dose audit means that the
Johnson &
maintenance of that sterilization dose. The fully processed product has also “failed” or is
Johnson in Raritan, NJ. Email:
jhanse2@its.jnj.com procedures for routine maintenance, nonsterile. The belief that dose audit posi-
typically carried out by performing biobur- tives or failures mean fully sterilized product
den and periodic dose audits, are outlined in is nonsterile is absolutely not true. This
the ANSI/AAMI/ISO 11137 series of stand- article will outline the truth of what a dose
ards1,2 as well as ANSI/AAMI/ISO TIR13004.3 audit failure means in terms of the sterility
The methods defined in these documents and sterility assurance of the product. In
have been successfully used for several addition, it will explore the consequences of
decades to determine and maintain steriliza- a dose audit failure on the sterilized product
tion doses. These documents provide associated with it.
guidelines for recommended actions based The dose audit TOS results are reported as
on the number of positive tests of sterility the number of positives in a set of samples,
(TOS) from a dose audit. Although some typically 100 samples for Method 1 and
wording and instructions have been modi- Method 2 and 10 samples for VDmax—the
fied slightly over the years to provide most widely used dose determination
additional clarification, the general actions to methods. The number of TOS positives
be taken have remained the same through- determines the action to be taken. In all
I (c) 202064 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
radiation sterilization dose audits, one and should not be referred to as such.
positive TOS for 10 samples and 2 positive Apart from this scenario of a repeat/
TOS for 100 samples are acceptable, and confirmatory test, a certain number of TOS
these results confirm the validity of the positives in the initial dose audit is consid-
sterilization dose for the specified sterility ered failing and will call for augmentation of
assurance level (SAL). Although a positive the sterilization dose (increase in steriliza-
TOS is acceptable and expected, sometimes tion dose to address the number of positives
any TOS positive will be wrongly viewed as observed) or cessation of sterilization for
unacceptable or in a negative context. This is products impacted, while corrective action is
tied to the misconception that the verifica- taken and the sterilization dose is reestab-
tion TOS positive is somehow directly lished. Each dose determination method
reflective of the sterility of fully sterilized specifies those criteria and the correspond-
product. It is also tied to the misconception ing actions. For this discussion, only those
that the verification test of sterility is the outcomes that indicate a confirmed dose
same as a lot release test for sterility. The audit failure will be addressed—not out-
verification dose is intended to provide a 10–1 comes prior to completion of a repeat/
to 10–2 SAL (sometimes referred to as confirmatory test, or outcomes of an accept-
“sublethal”), and therefore one or two able repeat/confirmatory test.
positives in the set of samples is considered
a statistically acceptable outcome of the
experiment and should never be viewed as ...positive TOS results that call for repeat/confirmatory tests
an indication of a problem with the product. are not “failures”...
When the number of TOS positives
exceeds the specified level for acceptability, It should be noted that the truths and the
different actions are to be taken, depending consequences to follow are not relevant for a
on the number of positives. For Methods 1 situation where there is a gross dose audit
and 2, a repeat test is called for, based on the failure due to an extremely high or out-of-
number of initial TOS positives. The initial control bioburden, or an excessive number
test in this case is not considered a dose of TOS positives (i.e., a number beyond that
audit failure and might only indicate that the which calls for augmentation). In these situa-
results fall slightly outside statistical accepta- tions, the principles presented here do not
bility, with the repeat test potentially apply and, based on the actual failure
confirming that fact. When using VDmax, the situation, there might be a significant impact
repeat test is called a “confirmatory test.” to the sterility of the product because of total
(Note that a VDmax document currently under loss of control of the manufacturing process.
development will not include the “confirma-
tory test” term). Again, the initial test in Truth
itself is not considered a dose audit failure,
and, as with Methods 1 and 2, might only Sterility and Safety
indicate that the results fall slightly outside The first truth about dose audit failures is
statistical acceptability. The outcome of the that there should not be an immediate
confirmatory test will determine whether the assumption that fully sterilized product is
dose audit should be considered passing or nonsterile or not safe for use simply because
failing. These two situations outlined for there are TOS positives after exposure to the
Method 1, Method 2 and VDmax will be verification dose. The incorrect assumption
referred to as “repeat/confirmatory” tests in by many companies is that a failing number
the remainder of this article. In the case of a of TOS positives means that product that has
repeat/confirmatory test, the additional been fully processed is nonsterile. This is
results will confirm whether the initial test incorrect, in that the TOS positives were
was a statistical expectation, and whether the based on irradiation at a verification dose,
sterilization dose continues to be valid. which is a much lower dose than that used
Hence, positive TOS results that call for for sterilization and realistically is expected
repeat/confirmatory tests are not “failures” to result in one or more TOS positives
(e.g., one positive out of ten tested demon- ing contamination is a significant factor in
strates a 10–1 SAL). The verification dose is the potential for a sterility test positive, and
set for the purpose of experimenting at a 10–1 the aspects that can contribute to this factor
or 10–2 SAL, not at the 10–6 or other SALs need to be recognized.
used for full sterilization. Therefore, verifica- The probability of lab contamination in a
tion dose positives arise from a 10–1 to 10–2 sterility test—resulting in false positives—is
SAL dose that statistically can result in a valid concern.6 In published information,
positives, and not from a full sterilization the false positive rate varies, but can be as
dose where the expectation of a positive is, high as 0.5%, depending on the level of
for example, one in one million (i.e., 10–6 environmental control and the competency
SAL). of the lab. Issues such as product design,
fatigue, the test environment, manipulation,
materials, technician error, and incubation
The probability of lab contamination in a sterility test—resulting in conditions all come into play in evaluating
false positives—is a valid concern. the likelihood of contamination. Labs that
are not set up for strict aseptic practices for
sterility testing, such as the sterility test
One cannot compare the safety of product isolator practices used in the pharmaceutical
that has been irradiated at a one in one industry, will likely have higher contamina-
million probability of a viable microorgan- tion rates. This expected laboratory
ism to that of product irradiated at a one in contamination rate is one reason sterility
ten or one in one hundred probability of a testing of terminally sterilized product is not
viable microorganism. The failed dose audit recommended.6 There is even a likelihood
might indicate that the calculated SAL of the that contamination will occur during the
fully processed product is not exactly at 10–6, incubation process, where container integrity
but, if calculations are performed based on may be an issue during incubation and
D-values, it is most often shown that the SAL sample examination.
is only slightly different than that claimed. Postprocessing contamination can include
Therefore, even if the SAL of the sterilized more than just lab contamination—it can be
product is slightly different than a claimed essentially anything that happens to the test
SAL of 10–6, the SAL might be only some- samples from the time the product exits the
thing such as 10–5.8, a probability of a viable irradiation process all the way through the
microorganisms of about one in six hundred final examination of containers in the testing
fifty thousand, which is certainly not a level process. Of utmost importance is the
that can be called nonsterile as far as patient package that maintains the sterility of the
safety is concerned.4 In ANSI/AAMI ST67, samples. A breach in the package integrity—
there are SALs other than 10–6 that are either a pre-existing issue or unknown/
recognized as “sterile,” such as 10–3, 10–4 and undetected damage—could also negate the
10–5, which are acceptable based on certain results of the TOS, especially considering the
criteria.5 Because of this, the issue is not handling and stresses involved in shipping
whether the fully processed product is (e.g., samples are frequently packaged in
sterile—it is sterile as far as patient safety is material other than what is validated for the
concerned. The primary concern relates to a finished product).
compliance perspective in demonstrating the
label claim, as explained further. Probability of Dose Audit Failures
Another truth about dose audit failures is the
TOS False Positive Rate statistical probability that one in approxi-
An important truth about a dose audit failure mately 11 or 12 dose audits will have results
is that there should not be an automatic that fall outside the acceptable number of
assumption that all TOS positives are true TOS positives, even if the bioburden has not
survivors of the verification dose. It is also changed.7 Dose determination is based on
possible that a TOS positive is the result of statistical probabilities, as previously dis-
postprocessing contamination. Postprocess- cussed, such as the likelihood of survival of
I (c) 202066 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
HELPING DELIVER BETTER OUTCOMES. A more efficient lab can lead to faster results, which in turn can improve patient
outcomes. Across the healthcare continuum, BD combines advanced automation with fully integrated products, expertise
from pre- to post-analysis, and sophisticated informatics to streamline lab workflows and accelerate turnaround times for
accurate results. This also helps measurably drive down costs across the entire lab for chemistry, microbiology, molecular
and clinical flow cytometry. From discovery to the delivery of care, this is the difference of an exceptionally broad portfolio of BD
solutions, insights and global partnerships. Discover the diagnostic difference one company can make. Discover the new BD.
BD and the BD Logo are trademarks of Becton, Dickinson and Company. © 2020 BD. All rights reserved. 5095 (0720)
I (c) 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
one in one million viable microorganisms. dose range will translate into SALs exceeding
Based on the probabilities of occurrence of the SAL for the sterilization dose. Therefore,
numbers of positives presented in the 11137 the vast majority of a given sterilization load
series dose determination methods, there is will possess SALs ranging from 10–6 to 10–9 or
around a 92% probability that there will be 10–10. This fact demonstrates the truth that
the acceptable number of positives, which the vast majority of a sterilization load will
leaves about an 8% probability that the have received the intended SAL of 10–6,
number of TOS positives will not be accept- because a slight change in SALs of 10–7, 10–8,
able, according to the interpretation criteria. or 10–9 will not approach that of 10–6. This is
This is the basis for the repeat or confirma- the reason that the majority of a sterilization
tory options in dose audits. load will achieve the designated SAL, leaving
only a portion of the load—in the minimum
dose zones—where the SAL might have
...the vast majority of a given sterilization load will possess SALs
been affected.8
ranging from 10–6 to 10–9 or 10–10.
Sterilization Dose vs. Minimum
In real life terms, for quarterly dose audits, Delivered Dose
this probability of unacceptable results can A final truth is that, in many cases, the
happen at any time during the course of minimum sterilization dose delivered to the
performing dose audits—not only after the product load is higher than the sterilization
11th or 12th dose audit. This is one reason dose that was determined (validated or
why the repeat/confirmatory option is substantiated). For instance, if a 25 kGy
specified—to demonstrate whether or not sterilization dose is substantiated, the
the failure was within the probability of routine sterilization process might show that
occurrence. If this truth is known and previous loads received minimum doses
understood, it should prevent a company higher than 25 kGy, such as 26 or 27 kGy. In
from jumping to the wrong conclusion about these cases, for a dose audit failure, the truth
the meaning of unacceptable results. For this is that there is potentially no impact to
truth, the assumed failure should be achieving the corresponding 10–6 SAL,
assessed in light of it potentially being a because the minimum dose of 25 kGy was
statistical probability, and therefore actions always exceeded, and therefore it could be
should be focused on determining this and demonstrated by calculations that the 10–6
demonstrating that the product is in a state SAL was always achieved.
of control, rather than only taking remedial Calculations for D-value and SAL based on
actions for a problem that might not exist. bioburden and radiation dose can show that,
for example, 25.5 kGy provides a 10–6.3 SAL,
Processed Loads in which case a slight change in that SAL
The next truth about a dose audit failure is due to a true dose audit failure still would
that, where there is a possibility of not mean the SAL of 10–6 was achieved. A series
achieving the specific SAL claimed, that of calculations can be performed for each
possibility only pertains to one small portion load to definitively show what the theoretical
of a sterilization load. For this discussion we SAL is, based on bioburden data coupled
will assume a 10–6 SAL. In all radiation with the minimum delivered dose for each
sterilization a dose range will be delivered, load. Depending on the circumstances, it is
which is typically the sterilization dose plus, often appropriate to specify a minimum dose
for example, 10–20 kGy, such as 20–32 kGy to the sterilization site that exceeds the
or 25–40 kGy. A dose range is specified validated minimum dose. If it is possible to
because, due to the physics of the irradiation specify the dose range for sterilization as a
process, the exact dose cannot be delivered minimum dose slightly above the dose that
throughout the product itself or the product was determined (e.g. 26–40 kGy versus
configuration. The selected SAL, such as 25–40 kGy), this could become the critical
10–6, corresponds to the sterilization (lowest) factor in defense of the nonimpact of a dose
dose, whereas all the higher doses in the audit failure.
I (c) 202068 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
much, or whether sterilization must be mum dose of 15.7 kGy for all loads under
halted and a new dose reestablished. The review. For loads that indicate an augmented
radiation guidelines for how to proceed in dose is required, it may be determined upon
this case are straightforward. For augmenta- review that the augmented dose was already
tion, the calculations are outlined based on achieved in routine processing, because the
the number of TOS positives, and the calculated augmented dose was actually
augmented dose is to be continued until achieved for all batches processed at the
either the underlying issue is resolved or the minimum delivered dose. Where the
dose is reestablished. For results that augmented dose was not achieved for certain
indicate cessation of sterilization, dose loads, a calculation of the theoretical SAL
reestablishment must be pursued immedi- might be appropriate to determine how
ately for sterilization to resume. different the calculated SAL is from the
designated SAL.
Once a theoretical SAL range has been
In essence, this means considering the validity of the sterilization
determined for each batch of product using
dose for the product that was sterilized since the previous passing
minimum dose delivered and maximum
dose audit.
dose delivered, then the percentage of
product at the minimum dose, intermediate
Dose in Retrospect doses, and maximum dose can be evaluated
In addition to the sterilization dose going through the product dose mapping per-
forward, another consequence is the sterili- formed during performance qualification.
zation of batches prior to the dose audit Armed with this information, a company can
failure. The 11137-2 guidelines specify that then assess if any product is at risk and how
one must consider the validity of the sterili- significant that risk might be, based on the
zation dose in retrospect. In essence, this calculated SAL for minimum dose locations,
means considering the validity of the as well as the percent of product that might
sterilization dose for the product that was not meet the SAL claim, if applicable. In
sterilized since the previous passing dose these calculations, if the bioburden has
audit. The guidelines state that “...the effect changed, the D-value of the population and
of processing product at the sterilization the calculation of SAL as indicated above
dose that has failed sterilization dose audit might not be appropriate. However, a change
on the achievement of the specified SAL for in bioburden would naturally be assessed,
previously processed batches of product shall initially, as a critical factor in the overall
be considered and a risk assessment under- investigation.
taken on their suitability for use.”2 Depending There will be cases where sterilized
on the data available, such as minimum product might be augmented, as specified in
delivered doses, bioburden determinations, the 11137 series and TIR13004. A company
and dose mapping, the consideration and the must assess whether this is a possibility
risk assessment might be simple. based on several factors, such as time
In performing a risk assessment, one elapsed since sterilization and whether the
would initially document the product additional dose will result in exceeding the
batches under review—as well as the maximum dose for the product. Based on
minimum dose delivered to these batches the findings laid out above and product use,
during sterilization—and compare these a company should assess the risk to patient
data with the calculated augmentation dose, and determine what, if any, action is war-
where appropriate. Many of the batches may ranted. The considerations discussed should
have already met the augmented dose, be undertaken with technical experts either
depending on the target dose determined by within the company or contracted when
the sterilizer and the actual delivered dose determining the path forward.
reported. For example, a product might have
a sterilization dose of 15 kGy, with a mini- Impact to Product Family
mum specified dose of 15.5 kGy for A final consequence that must be considered
irradiation, and an actual delivered mini- for a true dose audit failure is the impact to
I (c) 202070 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
other product in the family. The ISO 11137-2 dose audit failure. The same level of scrutiny
standard states that in the event of a dose and assessment must be applied to all
audit failure “…all members of that family members of the product family. Additionally,
shall be considered to be affected.”2,3 There- if a root cause is determined that could be
fore, actions taken for the dose audit systemic, assessment across product families
failure—augmentation, dose reestablish- might be warranted.
ment, or cessation of sterilization—will have
to apply to all members of the family. Summary
Considering the many variables that could Table 1 is a summary of the preceding
apply to a family, it would be impossible to sections concerning truths and conse-
cover all potential approaches here. Suffice it quences of dose audit failures.
to say that there are several options for
separating out certain family members or Conclusion
sub-groups and confirming the validity of a In conclusion, for a confirmed dose audit
sterilization dose for those members or failure there are several truths and several
sub-groups apart from the product in the consequences. The truths are that: a) product
Verification
Dose audit TOS Positives experiment Truth Action Consequence
Method 1 or 2 (10 ) –2
0, 1, or 2 Valid Acceptable No further action No
Pass
Method 1 or 2 (10–2) > 2 Invalid Unacceptable Repeat verification test No
Not a failure
Method 1 or 2 (10 ) –2
3 or 4 Valid Unacceptable Augment dose No
Not a failure Repeat verification test
Inconclusive
Method 1 or 2 (10–2) 5–15 Valid Unacceptable Augment dose Yes
Dose audit failure Reestablish dose
Assess previous batches
Method 1 or 2 (10 ) –2
> 15 Valid Unacceptable Stop sterilization Yes
Dose audit failure Reestablish dose
Assess previous batches
VDmax (10–1) 0, or 1 Valid Acceptable No further action No
Pass
VDmax (10 )–1
≥ 2 Invalid Unacceptable Repeat verification No
experiment
Not a failure
VDmax (10 )–1
2 Valid Unacceptable Perform confirmatory test No
Not a failure
Inconclusive
VDmax (10 )–1
3–6 Valid Unacceptable Augment dose Yes
Dose audit failure Reestablish dose
Assess previous batches
VDmax (10 )–1
> 6 Valid Unacceptable Stop sterilization Yes
Dose audit failure Reestablish dose
Assess previous batches
Table 1. Summary of dose audit truths and consequences. Abbreviation used: TOS, tests of sterility.
dose audit failure are that a) the sterilization 8. Logar JR. Exploring the probability of a device
dose going forward must be considered, being nonsterile following irradiation. Biomed
augmented if required, and guidelines for Instrum Technol. 2019;53(suppl. 2)43–48.
reestablishment followed; b) the sterilization
9. Hansen J, et al. Investigating AAMI Radiation
dose in retrospect must be considered and a
Audit Results. MD&DI. May 1994.
risk analysis performed; and c) the impact to
and subsequent action for other product in a
family must be considered.
The overarching fact is that having
positives in a dose audit test of sterility does
not automatically mean product that has
received the full sterilization dose is nonster-
ile or unsafe for use. In many cases, those
sterilized products possess an acceptable
SAL where patient safety is concerned.
References
1. ANSI/AAMI/ISO 11137-1:2006/(R)2015 & A1:2013
& A2:2019 (Consolidated Text), Sterilization
of health care products—Radiation—Part 2:
Establishing the sterilization dose. Arlington, VA:
Association for the Advancement of Medical
Instrumentation.
2. ANSI/AAMI/ISO 11137-2:2013/(R)2019,
Sterilization of health care products—Radiation—
Part 1: Requirements for development, validation
and routine control of a sterilization process for
medical devices. Arlington, VA: Association for the
Advancement of Medical Instrumentation.
I (c) 202072 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
NAVIGATING CHANGES—
TOGETHER.
From MDR to ISO 10993 and beyond, the regulatory landscape is constantly evolving.
Learn how we bring a robust approach, active leadership and strategic partnership to
every medical device product we test, with our exceptional line of services in:
• Consulting
• Testing
• Manufacturing
About the Authors Abstract the radiation source (cobalt, in the case of
Hydrogen peroxide has a multitude of uses and gamma irradiation) and material compatibil-
Brian McEvoy,
BSc, MBA, the vapor form was first identified as a sterilant ity, described in AAMI TIR17.2 Therefore, a
is senior in late 1970s. Following a number of develop- continual need for extension of the available
director global ments, vaporized hydrogen peroxide (VHP) terminal sterilization technologies remains
technologies became widely adopted in early 90s as a and was highlighted in 2019 with the launch
at STERIS in
substitute for ethylene oxide (EO) in device and of the Food and Drug Administration (FDA)
Tullamore,
Ireland. Email: brian_j_mcevoy@
instrument processing and reprocessing in Innovation Challenge to identify new
steris.com healthcare facilities. Often VHP was hailed as sterilization methods and technologies.
the replacement technology for EO. Because of Given its extensive history in healthcare
Randal key limitations such as scale, penetration, and processing and being deemed a Category B
Eveland, PhD, is
compatibility with packaging materials, sterilization method by FDA, vaporized
senior manager,
research and adoption to terminal sterilization of single-use hydrogen peroxide (VHP) is now being
development devices has not commenced to any significant considered as such a candidate technology
at STERIS in level. However, recent developments in steriliza- for inclusion.
Mentor, OH. tion chamber design and process development
Email: randal_eveland@steris.com
provide new opportunity for consideration. For Vaporized Hydrogen Peroxide
future products, such as those that require “end With hydrogen peroxide established as a
of production line sterilization,” such limita- biocidal agent since the 1800s and the first
tions may be reconsidered and overcome. This uses of the vapor form in sterilization
article describes those challenges and how they emanating from the late 1970s, VHP is a
have been addressed, with practical examples. well-established sterilization method of
The development of global consensus standards choice in the healthcare setting for the
and leveraging the well-established knowledge of processing of reusable devices.3 Further-
VHP sterilization with regard to microorganism more, VHP is used in many other
inactivation and material compatibility will applications, such as room and facility
help facilitate wider consideration of VHP decontamination.4 The disinfection and
technology as a true alternative to EO in certain sterilization efficacy of hydrogen peroxide in
product applications. both aqueous and gas form are well docu-
mented,4,5 with the latter form considerably
Terminal Sterilization of outperforming the liquid system.4,6 It has
Medical Devices been suggested that VHP can penetrate the
Terminal sterilization of single-use medical three-dimensional protein structure and
devices may be broadly subdivided into two cause breaking of bonds between subunits
main technologies, namely radiation and more easily than liquid hydrogen peroxide.7
sterilization by ethylene oxide (EO) gas.1 The The International Agency for Research on
continued reliance on EO is a consequence Cancer (IARC) has determined that hydro-
of its wide-ranging material compatibility gen peroxide’s carcinogenicity to humans is
and availability for scale-up. The challenges not classifiable.8 The advantages of using
of using EO relate to the hazardous nature of gaseous hydrogen peroxide are described by
the gas coupled with the prolonged treat- Hultman et al.6: 1) it will have uniform
ment times. Radiation, on the other hand, is contact with all exposed surfaces, including
a relatively quick process with no toxic those with complex topographies; 2) it may
residues, but is limited by the availability of be safely maintained in a chamber environ-
I (c) 202074 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
ment; and 3) it may be efficiently and quickly Limitations of VHP and the
removed from a chamber. Hydrogen peroxide Role for Process Design
is a strong oxidizing agent, having multiple As VHP offers a number of distinct
targets within a cell as well as in almost advantages, including material compatibility
every biomolecule: it can react strongly with (as highlighted in AAMI TIR17 and Table 1
thiol groups in enzymes and proteins, DNA, of this article), efficacious microbial
and the bacterial cell membrane.9 inactivation, and lack of toxic residues, it
Such a strong oxidizing agent can cause does present some limitations that must be
the formation of radicals such as ferryl considered and addressed, namely:
radical, which is formed from DNA-associ- 1) material compatibility; 2) packaging;
ated iron and has an important role in DNA 3) equipment scale and throughput
oxidation.9 As highlighted by the work of capability; and 4) surface penetration.
researchers such as Young and Setlow,10 Strategies to address such limitations may be
Fichet et al.,11 and Setlow,12 the site of summarized as follows.
microbial inactivation by VHP appears to
reside at the inner membrane. Material Compatibility
VHP has a wide range of material compati-
Industrial-Scale VHP bility but is limited whereby a small subset
Industrial VHP sterilization chambers of materials must be excluded due to
(Figure 1) are typically one-half to four pallets absorption and decomposition of the vapor.
(35–280 ft3). This is in contrast with large EO Liquids, powders, and cellulose materials are
chambers, often ranging from 1,000 ft3 to not compatible with the process.5
2,200 ft3. A distinct advantage with VHP
sterilizers, however, is the minimal require- Packaging
ment for supplies and services: water, Packaging for VHP sterilization needs to be
compressed air, electricity, and sterilant. This compatible with the sterilant and allow for
allows equipment to be supplied on a single diffusion of the sterilant to the medical
skid and installed with ease, which may be device. Packaging materials that prevent
advantageous for ‘in-line’ applications. VHP from reaching the devices (e.g., foil
Equipment may be commissioned and pouches) should not be used; however, the
validated in a manner similar to a steam or external surfaces of the material such as
EO sterilizer. glass (e.g., a liquid-containing ampoule) will
be sterilized. Nonwoven polyethylene and tion of H2O2 in the vapor phase, causing the
polypropylene materials have been proposed cycle to abort. Another potential approach
by the Sterile Barrier Association and might be lacquering or coating of absorbent
currently are used in healthcare applications. materials like cardboard or paper. Meszaros
Such materials provide a sterile barrier for et al.14 have examined the impact of surface
the product while allowing sterilant gases material on the process lethality, and did not
and steam to penetrate and escape quickly. observe any deleterious effects with beech-
Corveleyn et al.13 have shown that H2O2 wood, although it is a cellulose-based
penetration across Tyvek package was material. The authors have speculated the
considerably greater (87.7% of reference reason might be the processed (lacquered)
concentration as measured inside the nature of the material. Also, as cardboard
package) compared to that of medical paper cartons must be excluded, operators may
(30%). Therefore, it is essential that devices incur additional handling of materials.
are packed in a suitable sterile barrier. Possible mitigating strategies may involve
Cellulose is associated with product the presentation of product in high-density
packaging (cardboard) and instructions for polyethylene tote box systems: Such totes
use (paper), rather than with the medical have been used in the past in EO processing
device itself. Cellulose can absorb H2O2 to to improve product residual outcomes by
such an extent that it reduces the concentra- removing cardboard cartons from the process.
Plastics* Metals*
Delrin (polyoxymethylene, POM)
1 †
Aluminium
EVA (ethylene vinyl acetate) Brass
KRATON2 polymers (styrenic block copolymer, SBC) Cobalt chrome alloy
Neoprene (polychloroprene) †
Copper**
Noryl3 (polyphenylene ether and polystyrene) Gold
Nylon (polyamide)
1 †
Nitinol
PPMA (polymethyl methacrylate) Platinum
PEEK (polyether ether ketone) Silver
Polycarbonate, polyethylene, polypropylene, polystyrene, polyurethane Stainless steel‡
PVC (polyvynil chloride) Titanium‡
Radel4 (polyphenylsulfone)†
Santoprene (thermoplastic vulcanizates, TPVs)
Silicone
Teflon1 (polytetrafluoroethylene, PTFE)
Ultem3 polymers (polyetherimide, PEI)
Ceramics and others* Coatings*
Alumina (Al2O3) Aluminum titanium nitride (AlTiN)
Diamond, ruby, sapphire Aluminum titanium nitride chromium Nitride (AlTiN CrN)
Glass Diamond-like carbon (DLC)
Silicone nitride (Si3N4) Titanium nitride (TiN)
Zirconium nitride (ZrN) Titanium nitride titanium carbonitride (TiN TiCN)
Zirconia (ZrO2 with or without Y2O3) Tungsten carbide (WC)
Table 1. Materials compatible with Vaprox HC sterilant. Source: V-PRO maX 2 Low Temperature Sterilization System operator manual; 10085896 rev A;
Steris. ¹Delrin, Nylon, and Teflon are registered trademarks of the DuPont Corporation. 2KRATON Polymers is a trademark of KRATON Polymers U.S.L.L.C.
3
Ultem and Noryl are registered trademarks of SABIC Innovative Plastics IP BV. 4Radel is a registered trademark of Solvay Advanced Polymers, L.L.C.
*Consult device manufacturer prior to processing. †May have limited life after repeated sterilization. ‡Non-lumen and fast non-lumen cycles should not
be used to sterilize mated surface configurations other than stainless steel and titanium. **When used in power and electrical conditions.
I (c) 202076 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Figure 2. Three-phase pattern for vaporized hydrogen peroxide (VHP) sterilization. Courtesy of STERIS.
I (c) 202078 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
nation of knowledge and experience, the possibilities VHP 10. Young SB, Setlow P. Mechanisms of killing of Bacillus subtilis spores
adds to the technology portfolio available to manufacturers by Decon and Oxone, two general decontaminants for biological
can be realized. As projects are completed with a diverse agents. Journal of Applied Microbiology. 2004;96(2):289–301.
range of products, experience and knowledge are gained that
11. Fichet G, Antloga K, Comoy E, Deslys JP, McDonnell G. Prion
may be shared in future publications. inactivation using a new gaseous hydrogen peroxide sterilisation
process. J Hosp Infect. 2007;67:278–86.
Acknowledgments
The authors thank and acknowledge the Kilmer Collabora- 12. Setlow P. Summer meeting 2013—when the sleepers wake:
tion Modalities group for the cooperative efforts considering The germination of spores of Bacillus species. J Appl Microbiol.
and exploring how additional sterilization modalities may be 2013;115(6):1251–68.
deployed and adopted by the healthcare products manufac- 13. Corveleyn S, Vandenbossche GMR, Remon JP. Near-infrared (NIR)
turing industry, for the benefit of patient care. monitoring of H2O2 vapor concentration during vapor hydrogen
peroxide (VHP) sterilisation. Pharm Res. 1997;14(3):294–98.
References
1. Gamma Irradiation Processing Alliance, International Irradiation 14. Meszaros JE, Antloga K, Justi C, Plesnicher C, McDonnell G. Area
Association. A Comparison of Gamma, E-beam, X-ray and Ethylene Fumigation with Hydrogen Peroxide Vapor. Applied Biosafety.
Oxide Technologies for the Industrial Sterilization of Medical 2005;10(2):91–100.
Devices and Healthcare Products. 2017. http://iiaglobal.com/wp- 15. McDonnell G, Sheard D. A Practical Guide to Decontamination in
content/uploads/2018/01/White-Paper-Comparison-Gamma-Eb- Healthcare. West Sussex: Wiley Blackwell Publishers, John Wiley &
Xray-and-EO-for-Sterilisation.pdf. Accesed June 2, 2020. Sons, 2012.
2. AAMI TIR17:2017. Compatibility of materials subject to sterilization. 16. ANSI/AAMI/ISO 14937:2009/(R)2013. Sterilization of health care
Arlington, VA: Association for Advancement of Medical products—General requirements for characterization of a sterilizing
Instrumentation. agent and the development, validation and routine control of a
3. McEvoy B, Rowan NJ. Terminal sterilization of medical devices sterilization process for medical devices. Arlington, VA: Association for
using vaporized hydrogen peroxide: a review of current methods and the Advancement of Medical Instrumentation.
emerging opportunities. J Appl Microbiol. 2019;127(5):1403–20. 17. Food and Drug Administration. Biological Indicator (BI) Premarket
4. McDonnell G. The Use of Hydrogen Peroxide for Disinfection and Notification [510(k)] Submissions. October 4, 2007. https://www.fda.
Sterilization Applications. In: Rappoport Z, Ed. PATAI’S Chemistry gov/media/71134/download. Accessed March 30, 2020.
of Functional Groups. John Wiley and Sons; 2014. 18. ANSI/AAMI/ISO 11135:2014. Sterilization of health care products—
5. Block SS. Peroxygen Compounds. In: Block SS, Ed. Disinfection, Ethylene oxide—Requirements for development, validation and routine
Sterilization and Preservation. Philadelphia: Lea & Febiger; 1991:167– control of a sterilization process for medical devices. Arlington, VA:
81. Association for the Advancement of Medical Instrumentation.
6. Hultman C, Hill A, McDonnell G. The physical chemistry of 19. McDonnell G. Chemical Sterilization. In: McDonnell G, Ed.
decontamination with gaseous hydrogen peroxide. Pharmaceutical Antisepsis, Disinfection and Sterilization: Types, Action and Resistance.
Engineering. 2007;27(1):22–32. Washington: ASM Press; 2007:191–216.
Flexible Endoscopes:
Terminal Sterilization and
Impact to Patient Safety
Alpa Patel and Nupur Jain
I (c) 202080 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
endoscopes. AERs are commonly used to The high HAIs associated with duodeno-
reprocess flexible endoscopes and require scopes, bronchoscopes, and colonoscopes
Food and Drug Administration (FDA) relate to the challenges that these devices
approval before they are marketed. If the add to the decontamination process at a
flexible endoscope is manually cleaned, it is healthcare facility. These devices have
then disinfected, rinsed with critical/treated difficult-to-clean areas and their complex
water, and dried in drying cabinets. design does not allow for visualization of
Few healthcare facilities will then sterilize these areas during decontamination. Fur-
flexible endoscopes using liquid chemical thermore, these difficult-to-clean areas are
sterilization. While liquid chemicals tend to not always highlighted in the IFU as loca-
be more compatible with the materials of the tions that require attention. As the devices
endoscope, the device cannot be packaged are repeatedly used and reprocessed, there
prior to sterilization. With no sterile barrier, could be an impact to their service life: Wear
the flexible endoscopes need to be used and tear (e.g., scratches) on the devices make
immediately or reprocessed again before the device make more difficult to clean. Also,
use. Not only is reprocessing an unused these flexible endoscopes are sometimes
medical device wasteful for a healthcare serviced at third-party vendors, who may add
facility, it also degrades the materials of a new materials or parts that make these
flexible endoscope. Other sterilization devices more difficult to clean and bring the
methods available for flexible endoscopes cleaning validation into question.
include ethylene oxide (EO; either at the
healthcare facility or an industrial sterilizer),
vaporized hydrogen peroxide, or hydrogen The high HAIs associated with duodenoscopes, bronchoscopes, and
peroxide gas plasma. These methods allow colonoscopes relate to the challenges that these devices add to the
for packaging the device prior to steriliza- decontamination process at a healthcare facility.
tion. Therefore, these terminally sterilized
flexible endoscopes can be stored without
requiring reprocessing again before use. The inherent design of flexible endoscopes
is not the only reason they are difficult to
Analysis reprocess. Other contributing factors are the
Healthcare-associated infections (HAIs) healthcare facility’s environment and the
related to contaminated flexible endoscopes processing department: There is a lack of
are an increasing concern in recent years, knowledge and sufficient training needed for
not only because of the high volume of HAIs those responsible for reprocessing endo-
but also because of the death rate associated scopes.3 These devices require multiple
with antibiotic-resistant infections. A recent decontamination steps that can be difficult to
study from John Hopkins University follow. Some of the processing departments
reviewed more than 2.3 million patients in do not have the appropriate equipment (e.g.,
six states and reported that the infection risk brushes, water, light, AERs, connectors,
is as follows1: inspection tools, containers, etc.) to perform
• Colonoscopy—about one patient per 1,000 processing. Malfunctioning AERs are an
surgeries added challenge to the decontamination
• Upper gastrointestinal endoscopy—about process of flexible endoscopes.4 In the
three patients per 1,000 surgeries working group of ANSI/AAMI ST91, Flexible
• Cystoscopy—about four patients per 1,000 and semi-rigid endoscope processing in health
surgeries care facilities, contaminated equipment was
• Bronchoscopy—about 15.6 patients per one of the items that was identified as a
1,000 surgeries challenge to decontaminating flexible
Furthermore, a review of the FDA’s endoscopes (e.g., contaminated drying
Medical Device Reports shows that there cabinets can contribute additional infection
were 79 deaths from January 2015 to July risk to the flexible endoscopes during
2019 resulting from the use of contaminated storage). Damaged and compromised
duodenoscopes.2 flexible endoscopes can harbor microorgan-
I (c) 202082 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
Hamed M, Shamseya M, Naeem Dafa Alah 4. Kenters N, Huijskens EGW, Meier C, Voss A.
I, Sawaf G. Estimation of average bioburden Infectious diseases linked to cross-contamination
values of flexible gastrointestinal endoscopes of flexible endoscopes. Endosc Int Open.
after clinical use and cleaning: Assessment 2015;3(4):E259–E265.
of the efficiency of cleaning process. Alexan- 5. Molloy-Simard V, Lemyre JL, Martel K,
dria Journal of Medicine. 2015;51(2):95–103. Catalone BJ, Elevating the standard of
endoscope processing: Terminal sterilization of
International Association of Healthcare duodenoscopes using a hydrogen peroxide-ozone
Central Service Materiel Management. sterilizer. Am J Infect Control. 2019;47(3):243–50.
Endoscope Reprocessing Manual, 1st Edition.
Chicago: IAHCSMM, 2017. 6. Alfa MJ, Degagne P, Olson N. Worst-case soiling
levels for patient-used flexible endoscopes
before and after cleaning. Am J Infect Control.
Ofstead CL, Heymann OL, Quick MR, et al.
1999;27(5):392–401.
The effectiveness of sterilization for flexible
ureteroscopes: A real-world study. Am J 7. Rutala WA, Weber DJ. Gastrointestinal
Infect Control. 2017;45(8):888–895. endoscopes: A need to shift from disinfection to
sterilization? JAMA 2014;312(14):1405–6.
Ofstead CL, Heymann OL, Quick MR, et al.
8. Food and Drug Administration. Infections
Residual moisture and waterborne patho-
Associated with Reprocessed Duodenoscopes.
gens inside flexible endoscopes: Evidence
www.fda.gov/medical-devices/reprocessing-
from a multisite study of endoscope drying.
reusable-medical-devices/infections-associated-
Am J Infect Control. 2018;46(6):689–96.
reprocessed-duodenoscopes. Accessed May 12,
2020.
SGNA Practice Committee. Standards
of infection prevention in the gastroenterol-
ogy setting. Chicago: Society of
Gastroenterology Nurses and Associates,
2015. www.sgna.org/Portals/0/Education/
PDF/Standards-Guidelines/Standard%20
of%20Infection%20Prevention_FINAL.pdf.
Accessed May 26, 2020.
*
The European Medical Devices Regulation 2017/745 (MDR) came into force on May 25, 2017 to replace
the two Directives 93/42/EEC (MDD) and 90/385 EEC (AIMDD) by May 26, 2020.
I (c) 202084 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
FEATURE
potentially increasing EO fugitive emissions that will further increase the paper content Carlos Taira is
to the environment and negatively impacting per sterilization load. program
worker health and safety. Because of the increased absorption manager, package
characteristics of cellulosic material, this engineering at
Medtronic in
Case Study change in pallet configuration is directly
Mansfield, MA.
We will review the impact of this change on related to the increase in cellulosic material Email:
four different medical devices to help better needed for IFUs to meet the new the carlos.a.taira@medtronic.com
understand what impact this change will language requirement. If a sterilization
have on the package and sterilization load. process uses EO gas to sterilize a load with Suzanne Butler
is EMEA senior
As we can see from Table 1, the impact the current IFU, and now the load has 185 kg
microbiology
may be very small (only 2.3% change in more weight in cellulosic materials, the & sterilization
pallet weight), or it may be very large amount of EO gas used to sterilize that load manager at
(a 71.1% change in pallet weight). is estimated to be more than double the Medtronic in
Figure 1 illustrates the effect of the previous amount of EO gas needed. The ster- Galway, Ireland.
Email: suzanne.butler@medtronic.
additional paper. An IFU accompanies the ilization cycle would need to be revalidated
com
product in each individual product package. for the substantial change in the configura-
There are 170 individual unit boxes per tion of the sterilization load. This will add
distribution case and there are 16 distribu- significant time to the validation process and
tion cases per pallet. This means each pallet strain the already limited EO capacity, and
contains 2,720 IFUs. Considering a threefold will also require regulatory approval—which
increase in IFU size, the new total IFU adds more time to any updates required. In
weight will be 277 kg per pallet. This is a addition, the EO residuals for the product
185-kg increase in IFU paper when com- that are in that load will require revalidation
pared to the current state of a sterilizer pallet to ensure compliance with the ISO 10993-7
and requires a larger pallet volume. requirements.7
c
a b d c / d x 100
axb
Pallet weight
Pallet weight Current pallet percent
Weight Total # IFU per increase due to weight with increase due to
Description of change increase per IFU sterilizer pallet increased IFU existing IFU increased IFU
One-page map fold pamphlet to 29 gm 270 7.8 kg 220 kg 3.6%
two-page map fold pamphlet
Two-page pamphlet to 32-page 55 gm 96 5.3 kg 234 kg 2.3%
book
Booklet increase from 44 pages 103 gm 320 32.9 kg 85 kg 38.7%
to 145 pages; increase from
10 languages to 33 languages
Booklet increase from 20 pages 68 gm 2,720 184.9 kg 260 kg 71.1%
to 44 pages; increase from eight
languages to 20 languages
Table 1. Examples of instructions for use (IFU) size increase in medical devices.
OUTCOMES,
every device, therapy, and
service we offer has two
priorities in mind:
BETTER VALUE.
improving patient
outcomes and providing
more overall value to the
IT’S ACHIEVABLE.
healthcare system. This
means devices that reduce
complications, minimally
invasive therapies that
shorten or eliminate
hospital stays, services
Because we all do better when that improve how care is
delivered, and more.
the patient and the healthcare LET’S PUT THE VALUE
system do better. INTO VALUE-BASED
HEALTHCARE —
TOGETHER.
medtronic.com/
transforminghealthcare
Depending on the number of sterilization being placed in our waste streams, as well as
loads processed each year, this would add a promote compliance without extensive time-
significant amount of EO into the environ- consuming and capacity-limiting validation.
ment with no added benefit to the sterilization Digital instructions for medical devices
or safety of the product to the patient. make sense for several reasons:
• They provide the timeliest delivery of the
Discussion latest version of an IFU, assuring that the
In accordance with regulations (e.g., EU customer always has the most updated
Regulation 207/20128 and FDA Blue Book), information.
medical device instructions can be provided • Ease of accessibility makes viewing an IFU
in digital format (eIFU). However, some a capability from a variety of electronics: a
professional use devices are excluded from phone, tablet, or a laptop, which makes it a
the regulation, including non-implantables, convenient alternative for healthcare
accessories that can be used for multiple providers and field personnel who often
purposes beyond implantation, movable may not be in an office setting with access The digital IFU also allows
capital equipment, patient materials, ablation to a paper copy. ongoing enhancement
devices, instruments not used for program- • They are a more effective communication of instructions in the
ming, temporary leads, and devices with a tool, leading to improved customer future without extensive
high risk of off-label use. This remains one satisfaction. and time-consuming
of the primary challenges with overcoming • When digital instructions replace printed validation that strains
the EO sterilization load burden increase instructions, they allow manufacturers to the market and timeline
that the device industry faces with new MDR reduce the waste stream for EO gas usage for compliance.
compliance requirements. While the EU and the amount of paper used.
MDR doesn’t require eIFU, it does make it a • The digital IFU also allows ongoing
permissible form of distribution to all enhancement of instructions in the future
member states; therefore, one could argue without extensive and time-consuming
that it should be a requirement for them to validation that strains the market and
accept eIFU. timeline for compliance.
Instead of allowing member states to
require physical IFUs, the European Com- Conclusion
mission should expand the existing eIFU The unintended consequences of the
Regulation 207/2012 to include all medical increased languages for paper IFUs are
devices and allow organizations the opportu- significant and can impact the radiation and
nity to leverage eIFUs as a distribution EO sterilization processes, add paper into the
method, if desired. This green alternative waste stream, and increase the potential for
would help drive down the amount of EO gas exposure to EO gas for people handling the
used and help reduce the amount of paper product post sterilization.
Figure 1. Impact on pallet volume of new European Union Medical Device Regulation for instructions for use.
I (c) 202088 Industrial Sterilization: Process Optimization and Modality Changes Summer 2020
e user license. Further copying, networking, and distribution prohibited.
AAMI
TIR97:2019
Principles for medical device security—
Postmarket risk management for
device manufacturers
This technical information report provides guidance on methods to
perform postmarket security risk management for a medical device
in the context of the Safety Risk Management process required by
ISO 14971. This TIR is intended to be used in conjunction with AAMI
TIR57:2016.
Industrial Sterilization
Explore the latest sterilization technologies, methods, standards, FDA requirements,
designs, and product release decisions, as well as how to apply sterilization meet
requirements during product design.
October 6 - 9 | Virtual
Radiation Sterilization
for Medical Devices
Understand the FDA's expectations regarding successful submissions and inspections,
recalls, problem solving, and risk avoidance, as well as the principles of radiation
sterilization and 11137 series of radiation sterilization standards.