You are on page 1of 8

Accred Qual Assur

DOI 10.1007/s00769-009-0630-8

REVIEW

Strategies to set global analytical quality specifications


in laboratory medicine: 10 years on from the Stockholm
consensus conference
Per Hyltoft Petersen • Callum G. Fraser

Received: 29 October 2009 / Accepted: 23 December 2009


 Springer-Verlag 2010

Abstract The setting of analytical quality specifications laboratory medicine for a very wide variety of purposes,
in laboratory medicine has attracted attention for many particularly in quality management. However, there is a
years. Over time, many strategies were advocated and all requirement for additional investigation of, inter alia,
had advantages and disadvantages. In the final decade of quality specifications for examinations done on measure-
the last millennium, considerable confusion existed on ments performed on ordinal and nominal scales, pre-
how to define analytical quality specifications correctly analytical factors and matrix effects, examinations done as
and how to apply them in everyday practice. This led to POCT, target values of control materials, and ways in
wide professional interest. In 1999, a consensus conference which analytical quality specifications can be used both to
sponsored by IUPAC, IFCC and WHO was held in set what is the optimum performance and as a tool for
Stockholm on ‘‘Strategies to Set Global Analytical Quality assessment of everyday practice.
Specifications in Laboratory Medicine’’. The consensus set
useful and well-documented strategies for the setting of Keywords Analytical bias and imprecision 
analytical quality specifications into a hierarchy with the Analytical goals  Analytical requirements 
best strategy at the highest level, namely, (1) Evaluation of Laboratory medicine  Matrix effects  Point-of-care-testing
the effect of analytical performance on clinical outcomes in
specific clinical situations, (2) Evaluation of the effect of
analytical performance on clinical decisions in general, (3) Introduction
Published professional recommendations, (4) Performance
goals set by regulatory bodies and EQAS organisers, and Over the last decade, since the landmark international
(5) Goals based on the current state of the art. Much consensus conference on ‘‘Strategies to Set Global Ana-
success has been achieved since the promulgation of the lytical Quality Specifications in Laboratory Medicine’’
statement with the approach being adopted by many in held in Stockholm [1], there has been wide introduction of
national accreditation in laboratory medicine, generally
based on ISO 15189 [2]. The basis for all such accredita-
tion is a robust quality management system that sets
P. H. Petersen (&) standards for each and every aspect of the organisation,
Norwegian Quality Improvement of Primary Care Laboratories, management and delivery of a quality laboratory medicine
NOKLUS, Division for General Practice, University of Bergen, service. However, before a decision can be made on
Box 6165, 5892 Bergen, Norway
whether or otherwise a quality service is being delivered, it
e-mail: Per.Petersen@isf.uib.no
is obligatory for the standards to be precisely defined.
C. G. Fraser While it is widely stated that current methodology and
Scottish Bowel Screening Centre Laboratory, Kings Cross, technology allow the delivery of laboratory data that are of
Dundee, Scotland, UK
more than sufficient quality to meet clinical needs, there is
P. H. Petersen considerable evidence that this is not the case [3]. It is
Flittig Lise Vej 20, 5250 Odense SV, Denmark therefore vital that objectively set numerical quality

123
Accred Qual Assur

specifications for the reliability performance characteristics pre-analytical and within-subject biological variation as
of examinations in laboratory medicine are laid down so well: additionally, the probability with which clinical
that satisfactory methodology can be adopted and applied. decisions are made is not always based on the medians and
It is pointless undertaking quality control, assessment and at P \ 0.05 as assumed by all those adopting this strategy.
assurance unless the quality required has been defined Much information is available on the analytical ‘‘state of
a priori. In this paper, the background to the 1999 con- the art’’ attained in laboratory medicine. These data were
sensus conference, the results of the conference that did so used to set quality specifications in a variety of strategies
much to publicise objective approaches to the setting of [4, 5] but are clearly less than ideal since the state of the art
analytical quality specifications, and the achievements or represents what is achievable with the methodology and
otherwise are discussed and ways forward suggested. technology used to derive the data and thus may not be
related to the quality that is actually required to facilitate
optimum clinical decision-making.
Background
The concept that quality specifications could be derived
from numerical data on components of biological variation
The setting of numerical analytical quality specifications,
was elaborated some 40 years ago [8, 9]. This idea was
also termed analytical goals, quality goals and analytical
adopted by the College of American Pathologists who
performance goals, has been a subject of interest for a
documented that, for group screening, the analytical CV
considerable time. The concepts promulgated in the early
(CVA) should be less than half of the within-subject (CVI)
years has been discussed in detail previously [4, 5] and will
plus between subject (CVG) biological variation, that is,
be simply summarised here.
CVA \ 0.5 (CV2I ? CV2G)0.5 and, for diagnosis and moni-
The first attempt to delineate analytical quality specifi-
toring treatment, CVA \ 0.5 CVI. This approach has been
cations was that of Tonks [6] who suggested that the
considerably refined since this original publication as is
allowable limits of error (ALE), often taken as twice the
described below.
analytical co-efficient of variation (CV), although clearly
After the publication of these original strategies for the
more relevant to total laboratory error (imprecision plus
setting of analytical quality specifications, based on refer-
bias), could be calculated from the following formula:
ence values, the opinions of clinicians and the components
ALE ¼ 2CV ¼ 0:25ðul  ll)=l of biological variation and, as methodology and technology
advanced and significant interest in quality management
where l is the mean of the reference interval and ul is the
developed, there was further more detailed work in the
upper reference limit and ll is the lower reference limit of
setting of objective quality specifications for analytical
the 0.95 reference interval. By multiplying CV by 100, the
reliability performance characteristics.
coefficient of variation is transformed to percentage.
Gowans et al. [10] explored the setting of quality
However, it should be noted that this quality specifica-
specifications for bias and suggested that, in order to be
tion was advocated for analysis of performance in external
able to use common reference intervals across geography,
quality assessment schemes, not for judgement of indi-
the bias (B) had to be less than one-quarter of the group
vidual laboratory performance. Moreover, while similar
biological variation, that is,
ideas were expressed by others [4, 5] and these clearly were
useful at the time, it became obvious that, since the ref- 0:5
jBj\0:25 CV 2I þ CV 2G :
erence interval is influenced by imprecision and bias, the
specifications derived was not independent of existing In addition, a group from the Nordic countries undertook a
performance and, in consequence, the approach became series of detailed analyses of the effect of analytical
considered outmoded. performance (imprecision and bias) on clinical outcomes in
This was followed by the publication of ‘‘medically very specific clinical situations and showed that it was more
significant CV’’ by Barnett [7]. Analytical quality specifi- than possible to set quality specifications whose achievement
cations were said to be derived from the opinions of would ensure optimal delivery of patient care [11, 12]. This
clinicians and laboratory specialists. These were sub- approach has been followed by some others since and an
jective. Similar work was done by others [4, 5], and this excellent recent example is provided by Boyd and Bruns [13]
included analysis of responses to clinical vignettes where a who derived quality specifications for glucose meters based
change in results was used to derive a numerical quality on simulation modeling of errors in insulin dose.
specification based on the median difference reported. The idea that the opinions of users (both clinicians and
However, these studies were seriously flawed in that patients) of laboratory data can be used to derive objective
the changes documented were considered to be due to quality specifications through analysis of the numerical
analytical imprecision alone, whereas they are due to changes that stimulate action was further explored in the

123
Accred Qual Assur

well designed and executed model studies of Sandberg and A real problem
colleagues [14, 15]. The group was convinced that quality
specifications derived from analysis of the effects of per- In spite of all the work done on approaches to setting
formance in specific and well-defined medical strategies analytical quality specifications for reliability performance
revealed the ideal basis for setting quality specifications. characteristics and the guidelines from international expert
Such conditions, however, are difficult to define and need groups, some very real problems remained near the end of
individual, and often complicated, models for evaluation of the last millennium.
the specifications; however, these may not be relevant for These included the following:
use of the same quantity in other clinical settings.
• there were, vide supra, many published recommenda-
Therefore, strategies based on formulae derived using
tions dating from 1963 onwards and it might be difficult
the components of biological variation were generally seen
for some professionals in laboratory medicine, unfa-
as the best for practical purposes and have been discussed
miliar with the details of these, to select the most
at many conferences, congresses and courses. In addition, a
appropriate,
number of professional groups including the EGE-Lab, the
• there continued to be further new recommendations
European Group for the Evaluation of Analytical Systems
published over time, again possibly causing difficulties
in Laboratory Medicine [16], and a number of the Working
in selection of the most appropriate strategy,
Groups of the European EQAS Organisers Groups [17, 18]
• data from examinations in laboratory medicine are used
supported the concepts.
in many different clinical situations, including diagnosis,
However, it was realised that there were some defi-
monitoring, screening, research, development, teaching
ciencies with the approaches, in particular that the
and training, and it might be that a single set of quality
biological variation for certain quantities was very small
specifications would be inappropriate for all of these,
and the quality specifications could not be met with
• some strongly advocated the hypothesis that neither
available methodology and technology; in addition, some
patients nor clinicians were harmed by current analyt-
quantities had large biological variation which led to
ical performance, so there was no reason to impose
quality specifications that were very easily attained. In
more stringent quality specifications, and
consequence, it was advocated [19] that a tiered approach
• it was not obvious that manufacturers of analytical
could be used to generate quality specifications for
systems and reagent kit sets used objective quality
imprecision, bias and total allowable error (TEa), that is,
specifications in development or marketing.
desirable quality specifications:
CV A \0:5 CV I :
0:5 A potential solution
jBj\0:25 CV 2I þ CV 2G
0:5
TEa\1:65  0:5 CV I : þ 0:25 CV 2I þ CV 2G In 1999, the Information for Authors of Clinical Chemistry
optimum quality specifications: was modified to state: ‘‘Analytical quality. Results obtained
for the performance characteristics should be compared
CV A \0:25 CV I :
0:5 objectively to well-documented quality specifications, e.g.,
jBj\0:125 CV 2I þ CV 2G published data on the state of the art, performance required
0:5 by regulatory bodies such as CLIA ‘88, or recommenda-
TEa\1:65  0:25 CV I : þ 0:125 CV 2I þ CV 2G
tions documented by expert professional groups’’.
minimum quality specifications: An Editorial [23] supporting this statement and giving
CV A \0:75 CV I : potential authors detailed advice on how to fulfil this
0:5 requirement was then published. This proposed that, at least
jBj\0:375 CV 2I þ CV 2G
0:5 for data reported on ratio (and difference) scales, an approach
TEa\1:65  0:75 CV I : þ 0:375 CV 2I þ CV 2G should be adopted that, like the types of evidence and grading
of recommendations used in clinical practice guidelines,
It should be noted that this approach is facilitated by
placed available and useful strategies in a hierarchy of
easy access to a data base on biological variation, regularly
objectivity, the best being first and the worst being last. The
updated by Ricos et al. [20]. Moreover, the applicability of
hierarchy suggested the following acceptable strategies:
such data is supported by the fact that, in chronic but stable
disease, the components of biological variation are not • assessment of the effect of analytical performance on
much different to those in health [21], which often can be clinical decision-making—quality specifications in spe-
considered as constant over time and geography [22]. cific clinical situations followed by

123
Accred Qual Assur

• general quality specifications based on medical needs, 1. Evaluation of the effect of analytical performance on
including, strategies based on biological variation, clinical outcomes in specific clinical situations
• professional recommendations and guidelines from 2. Evaluation of the effect of analytical performance on
national or international expert groups followed by clinical decisions in general
guidelines from expert individuals or institutional
a. Data based on the components of biological
groups,
variation
• quality specifications laid down by regulation or by
b. Data based on analysis of clinicians’ opinions
EQAS organisers—quality specifications laid down by
regulation followed by quality specifications laid down 3. Published professional recommendations
by EQAS organisers, and finally,
a. From national and international expert bodies
• published data on the state of the art from external
b. From expert local groups or individuals
quality assessment and proficiency testing schemes and
then from published methodology. 4. Performance goals set by
Around the same time, a Working Group of ISO TC 212 a. Regulatory bodies
undertook the task of preparing a Standard or Guide on how b. Organisers of External Quality Assessment (EQA)
to define and apply quality specifications. Eventually, in schemes
2001, TC 212/SC/WG 3 published ISO/TR 15196, a Tech-
5. Goals based on the current state of the art
nical Report (type 2) entitled ‘‘Determination of analytical
performance goals for laboratory procedures based on a. As demonstrated by data from EQA or Proficiency
medical requirements’’. This was not an ISO International Testing Schemes
Standard and was distributed for review and comment: it b. As found in current publications on methodology.
was noted that it was subject to change without notice and
When available, and when appropriate for the intended
may not be referred to as an International Standard. It is
purpose, models higher in the hierarchy are to be pre-
much regretted that this work does not appear to have been
ferred to those at lower levels. The concept of such a
progressed, since it supported the concepts in the Editorial in
hierarchy is described in a recent Clinical Chemistry
Clinical Chemistry and, indeed, expanded these somewhat
Editorial in which the relative merits of the above models
and generally concurred with the consensus statement
are discussed [23]. This hierarchy has also been proposed
reached at the consensus conference next described.
by the ISO/TC 212/WG3 subgroup on ‘‘performance
Goals Based on Medical Needs’’ as the basis for the
Strategies to set global analytical quality specifications ongoing ISO/CD 15196. The following matters were also
in laboratory medicine discussed and agreed:
• The above hierarchy includes currently available
The question of, and difficulties in, setting analytical
models; however, new useful concepts will undoubtedly
quality specifications in laboratory medicine was attracting
evolve. Implementation of any of the models should use
very wide international interest and, as a result, the
well-defined and described procedures.
Stockholm Consensus Conference on Quality Specifica-
• To facilitate the future debate on the setting on
tions in Laboratory Medicine, was organised in 1999. This
analytical quality specifications, there is a need for
was supported by the Clinical Chemistry Section of the
agreement on concepts, definitions and terms.
International Union of Pure and Applied Chemistry (IU-
• There is a need for continuous improvement in the
PAC), the International Federation of Clinical Chemistry
exchange of information on quality issues between
and Laboratory Medicine (IFCC) and the World Health
clinical laboratory professionals and the diagnostics
Organization (WHO). Professionals who had published in
industry, and between clinical laboratory services and
the field were invited to present their views and others
the users of the laboratory service.
involved in this facet of laboratory medicine also partici-
pated. The papers and the consensus statement were
published in a Special issue of the Scandinavian Journal of
Clinical and Laboratory Medicine [1]. Successes
The majority of the consensus statement is reproduced
here. This approach was very well received by professionals in
The main outcome of the conference was agreement that laboratory medicine at the time and, since its promulgation,
the following hierarchy of models should be used to set has been cited on many occasions and very widely applied,
analytical quality specifications. particularly in method selection and assessment,

123
Accred Qual Assur

international, national and regional external quality are usually based on parametric statistics with estimated
assessment scheme performance evaluation and other fac- Gaussian or log-Gaussian distributions as described by
ets of quality management. Ricos et al. [20]. These assumptions and models can easily
Moreover, it is pleasing to note that the majority of the be extrapolated to quantitative quality specifications for
international and national professional bodies that develop acceptable bias or acceptable imprecision. However, many
evidence-based guidelines, including IFCC and CLSI quantities in other matrices, including urine, are measured
(Clinical and Laboratory Standards Institute), now gener- on an ordinal scale where the result has an uncertainty: not
ally use the approaches based upon biological variation to as a quantitative distribution, but simply as a dichotomy of
develop analytical quality specifications when these are positive or negative results which can have large concen-
including in such works, the level 3 standards now being tration intervals in which both positive and negative results
developed using the level 2 strategy. can be correct. Here, the uncertainty is in a form where it
Interestingly, since 1999, there have been many pub- can be acceptable with, for example, 20% positive and 80%
lications on the generation of data on biological variation negative results for a certain control material. How can
for an ever expanding range of quantities of interest in reliable quality specifications be outlined for this type of
laboratory medicine [20]. Almost without exception, examination?
these describe the setting of analytical quality specifica- Measurements performed on ordinal scales were dis-
tions using the data generated as well as other cussed by Kouri et al. during the Stockholm Conference
applications. [26]. They generally investigated dichotomously report-
The rationale behind the need for analytical quality able examinations (positive/negative, present/absent,
specifications, the advantages and disadvantages of the detected/not detected) and introduced quality specifica-
strategies in the hierarchy and ways in which these can be tions in the form of detection limit, defined as the
applied in everyday practice have been detailed in a concentration for which the percentage of positive results
monograph [22] and will not be addressed further. should be below a certain value, and confirmation limit,
However, two outstanding projects with publications defined as the concentration for which the percentage of
must attract particular attention. First a project on estab- negative results should be below a certain value: the ratio
lishing common reference intervals for 25 common clinical between these values should be 5. These proposals have
chemical quantities in a homogeneous Nordic population now been published in European Urinalysis Guidelines
across country borders where all elements of quality were [27]. Here, the group also investigated the more detailed
considered, including traceability of calibration standards possible ordinal scale approach (negative, 1?, 2?, etc.),
and external control of measurements according to quality but did not distinguish this from what are usually
specifications for bias based on specifications for common described as semi-quantitative examinations, which actu-
reference intervals, beside painstaking description of the ally have an underlying ratio scale which makes it
reference individuals [24]. Second, a report from the possible to standardise and control the examination. Later,
National Institute of Standards and Technology (NIST) on improved models described the characteristics of dichot-
consequences of poor calibration and thereby analytical omous tests [28] and semi-quantitative tests [29] which
bias on serum calcium, related to the increase in number of made it relatively easy to describe performance variables
extra laboratory examinations and other investigations for, for example, different reagent kit sets whereby con-
performed on patients or extra costs in USD for increasing trol of the different tests according to defined analytical
bias [25]. quality specifications could be derived. But the problem
However, although these outcomes are very positive, remains as to how specifications for allowable deviations
this does not mean that the consensus cannot be improved of the results of examinations made on ordinal scales
upon with the passage of time and a number of issues should be defined?
remain that require exploration by professionals in labo-
ratory medicine.
Pre-analytical factors and matrix effects

Examinations reported on ordinal and nominal scales Guder described the many pre-analytical factors which can
increase the variability of measurement results or change
The vast majority of investigations on analytical quality the concentration or activity of a quantity, ranging from the
specifications have been concerned with quantitative ‘‘pre-analytical time’’, which is important when fast
measurements on components measured in blood, serum or delivery of results is needed to ‘‘influence of interfering
plasma using ratio or difference scales and use those factors’’, which can be divided into in vivo and in vitro
derived from data on biological variation. These models influences [30]. Both can be reduced by painstaking

123
Accred Qual Assur

CoaguChek S, lot 568


to the time needed to get phlebotomy, transport the sample
Difference (%) Difference plot
to the laboratory, perform the examination, and report back
CVMatrix = 7.9 % 30
the result. Although it has been cogently argued that the
CVA = 3.1 % analytical quality specifications required in POCT situa-
CV Total2 = 0 tions should be as for examinations done in other clinical
CV Matrix2 + CV A2 situations and these should best be based on data on bio-
CV Total = 8.5 % logical variation [31], the gain from speediness is obvious.
-30
Bias = – 15 % 0,5 1,5 2,5 3,5 4,5 5,5 The question remains: can the requirement for low turn-
Reference method (INR units) around time speed modify the analytical quality
specifications for a quantity?
Fig. 1 Difference plot illustrating the percentage difference between
a POCT INR method and a reference method as function of the
reference method. There is a mean bias of -15% and the dispersion
of points is CV = 0.085 (8.5%), which compared to the imprecision
Target values of control materials
of CV = 0.31, demonstrates a variation due to matrix-effects of
CV = 0.79. (Personal communication: Esther Jensen Department of According to International Vocabulary of Metrology, VIM
Clinical Biochemistry, Hillerød Hospital, Hillerød, Denmark) [32], measurements performed with analytical methods,
which in clinical chemistry usually means reagent kit sets,
should be traceable to a ‘true’ concentration value, defined
standardisation of preparation of the patient for sampling by method or international reference preparation, and for
and proper handling of samples from the patient, but also each kit an uncertainty should be stated. This is an ideal,
the stability of the quantity, choice of anticoagulant, con- which should result in examinations of the same control
tamination (for example. from haemolysis, icterus or lipid) sample within a narrow range around the ‘true’ value.
may cause additional variation in the results of examina- This, however, is not achieved in clinical chemistry,
tions. How are these pre-analytical factors related to the where kits from different manufacturers reveal different
analytical quality specifications? results for the same control material, which is specific for
Differences in methods and reagents used may cause the individual kit. Moreover, these method biases may be
individual but constant differences due to matrix effects, persistent for years. For example, the control results for
which are reproducible, in contrast to imprecision. In these serum TSH show permanent individual bias for the five
cases, the differences in examinations done by the different dominating kits on the European market over a 5- year
methods or with different reagents show a larger variance period from 2000 to 2005, as extracted from the results
than calculated from imprecision alone. This increase in from the external quality assessment scheme performed by
variation is thus related to the matrixes of samples and the ‘‘Deutsche Gesellschaft für Klinische Chemie’’ where
reagents, but it is not disclosed by the usual variables of the biases are constant and cover a range of 30% [33]. In
bias and imprecision, estimated in traditional analytical many EQA schemes, so-called peer group target values
control. An obvious example is POCT instruments for are calculated as the mean of control results for a single
measurements of INR, where samples examined with dif- kit, and this may simply lead to confusion, especially
ferent kits with different methods demonstrate a variation when the same EQAS organiser uses two models for the
which can be 2–3 times the coefficient of variation as analysis of results in a single scheme. The acceptable bias
estimated as imprecision alone, as illustrated in the per- should in principle be estimated from a ‘true’ concentra-
centage difference-plot for two INR-methods shown in tion but, in laboratory practice, reference intervals and
Fig. 1. How should the allowable size of these matrix decision limits are often based on the biased standardi-
effects be described and/or defined? sation and, consequently, the allowable bias should be
validated according to the local use. This is not an ideal
solution but, as long as kits for clinical chemistry reveal
Point-of-care-testing permanent bias in their standardisation, it is important to
distinguish between ‘‘true’’ target values and peer group
For point-of-care-testing (POCT), the analytical results can means in validation in EQAS where a laboratory can have
be obtained in a short time after sampling and, when good performance but a permanent bias due to a poor kit.
examination is performed by the patient, the result is A proposal for separating kit bias from performer bias is
available almost instantly. This saves time, whereby a to present the results in an x–y plot with performer
change in the concentration of the quantity can be imme- deviation as function of kit bias, e.g. as illustrated in
diately apparent and treatment initiated promptly, for Fig. 2. How can the problem be solved until kit producers
example a fall in plasma glucose concentration, compared deliver error free methods?

123
Accred Qual Assur

8 and developing countries where the requirements could


Acceptable
Deviation from peer-group mean
7 evolve at different tempo in a two-tier coordination. Is it
6 bias
reasonable to distinguish between goals and requirements?
5
In all cases, performance standards should be decided
4
regarding internal control as well as external assessment or
(µmol/L)

3
2
proficiency testing in order to cope with analytical quality
1 specifications and the level of power for obtaining the
Acceptable
0
performance
required quality should be defined. The quality managing
-1 process should be performed at all levels of control in
-2 planning, processing, control, assessment and improvement
-3 [34] and with choice of the sigma level [35].
-4
-4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 10 11 12
Peer-group bias (µmol/L)
Conclusions
Fig. 2 Illustration of the two aspects of bias (validation of method
and validation of participant). Example on S-Transferrin. The mean An objective assessment 10 years on from the Consensus
peer-group bias is shown on the abscissa with confidence interval and
the laboratory deviation from the peer group mean is shown on the Conference on Strategies to Set Global Analytical Quality
ordinate together with the dispersion of results from the same peer- Specifications in Laboratory Medicine, clearly demon-
group. The example illustrates a good performer (result within strates that much success has been achieved with the
acceptable performance) using a method with bias (peer-group mean hierarchical approach being adopted by many in laboratory
outside the acceptable bias)
medicine for a wide variety of purposes in laboratory
medicine, particularly in quality management. However,
A ‘two-tier’ approach to analytical quality
the conference was not all encompassing, and there is a
specifications
need for further consideration of, and work on, inter alia,
quality specifications for examinations that generate
A further important question to be solved is related to the
reports on ordinal and nominal scales, pre-analytical fac-
difference between desirable quality, as estimated from the
tors and matrix effects, examinations done as POCT, target
clinical or biological investigations from the highest levels
values of control materials and ways in which analytical
of the hierarchy agreed at the consensus conference and the
quality specifications can be used both to set not only what
real practical world, where even the best methods and
is the optimum performance but also that will facilitate the
performers sometimes are unable to attain this quality. This
best patient care and as a tool for assessment of everyday
question was addressed in the EGE-Lab publication on
practice. We hope that professionals in laboratory medicine
quality specifications, where the idea was to set the spec-
will rise to these challenges and that this facet of the
ifications according to the 20% best performers until the
discipline of laboratory medicine will continue to push
goal was obtained for all [16].
boundaries.
This idea can be elaborated if we distinguish between
the ideal—which can be called analytical performance
goals for acceptable bias, imprecision and matrix effects References
based on the effect of variation of the clinical use of the
quantity or data on biological variation—and the analytical 1. Hyltoft Petersen P, Fraser CG, Kallner A, Kenny D (eds) (1999)
performance requirements needed to attain either a ‘‘pass’’ Strategies to set global analytical quality specifications in labo-
in proficiency testing (PT) or ‘‘satisfactory’’ performance in ratory medicine. Scan J Clin Lab Invest 59:475–585
2. International Organization for Standardization (2007) ISO
EQAS. 15189:2007. Medical laboratories—particular requirements for
Such analytical performance requirements could evolve quality and competence
in a stepwise manner with time as methodology and tech- 3. Westgard JO (2004) Nothing but the truth. Westgard QC,
nology advanced and the analytical performance goals Madison
4. Fraser CG (1981) Analytical goals in clinical biochemistry. Prog
became attainable, when the two-tier approach could col- Clin Pathol 8:101–122
lapse to a single set of analytical quality specifications 5. Fraser CG (1983) Desirable performance standards for clinical
based on a strategy high in the hierarchy, an idea advanced chemistry tests. Adv Clin Chem 23:299–339
some years ago [19]. This approach could also make it 6. Tonks DB (1963) A study of the accuracy and precision of
clinical chemistry determinations in 170 Canadian laboratories.
possible to retain the same analytical quality specifications Clin Chem 9:217–223
which document the optimum analytical quality’ and at the 7. Barnett RN (1968) Medical significance of laboratory results. Am
same time differentiate the requirements for rich nations J Clin Pathol 50:671–676

123
Accred Qual Assur

8. Cotlove E, Harris EK, Williams GZ (1970) Biological and ana- 22. Fraser CG (2001) Biological variation: from principles to prac-
lytic components of variation in long-term studies of serum tice. AACC Press, Washington
constituents in normal subjects. III: physiological and medical 23. Fraser CG, Hyltoft Petersen P (1999) Analytical performance
implications. Clin Chem 16:1028–1032 characteristics should be judged against objective quality speci-
9. Harris EK (1979) Statistical principles underlying analytic goal- fications. Clin Chem 45:321–323
setting in clinical chemistry. Am J Clin Pathol 72:374–382 24. Rustad P, Felding P (eds) (2004) Transnational biological refer-
10. Gowans EMS, Hyltoft Petersen P, Blaabjerg O et al (1988) ence intervals. Scand J Clin Lab Invest 64:263–441
Analytical goals for the acceptance of common reference inter- 25. Gallaher MP, Mobley LR, Klee GG, Schryver P (2004) The
vals for laboratories throughout a geographical area. Scand J Clin impact of calibration error in medical decision error. NIST
Lab Invest 48:757–764 26. Kuri T, Laippala P, Kutter D, Gant V, Hallander H, Guder WG
11. Hørder M (ed) (1980) Assessing quality requirements in clinical (1999) Quality specifications for ordinal scale measurements with
chemistry. Scand J Clin Lab Invest 40 suppl 155:1–144 multiproperty (multiple) urine test strips. Scand J Clin Lab Invest
12. deVerdier C-H, Groth T, Hyltoft P (eds) (1993) Medical need for 59:523–526
quality specifications in clinical laboratories. Upsala J Med Sci 27. Kouri T, Fogazzi G, Gant V, Hallander H, Hofmann W, Guder
98:187–491 WG (eds) (2000) European urinalysis guidelines. Scan J Clin Lab
13. Boyd JC, Bruns DE (2001) Quality specifications for glucose Invest 60 Suppl. 231
meters: assessment by simulation modeling of errors in insulin 28. Hyltoft Petersen P, Gade Christensen N, Sandberg S, Nordin G,
dose. Clin Chem 47:209–214 Pedersen M, Nordic Control Organisations (2008) How to deal
14. Skeie S, Thue G, Sandberg S (2001) Patient-derived quality with dichotomous tests? Application of a rankit-ordinal scale
specifications for instruments used in self-monitoring of blood model with examples from the Nordic ordinal scale project on
glucose. Clin Chem 47:67–73 screening tests. Scand J Clin Lab Invest 68:298–311
15. Skeie S, Thue G, Sandberg S (2001) Interpretation of hemoglobin 29. Hyltoft Petersen P, Gade Christensen N, Sandberg S, Nordin G,
a1c (Hba1c) values among diabetic patients: implications for Pedersen M, Nordic Control Organisations (2009) How to deal
quality specifications for hba1c. Clin Chem 47:1212–1217 with semi-quantitative tests? Application of an ordinal scale
16. Fraser CG, Hyltoft Petersen P, Ricos C, Haeckel R (1992) model to measurements of urine glucose. Scand J Clin Lab Invest
Quality specifications for the acceptability of analytical systems 69 (in press)
for clinical chemistry. Eur J Clin Chem Clin Biochem 30:311– 30. Guder WG (1999) Preanalytical factors and their influence on
317 analytical quality specifications. Scand J Clin Lab Invest 59:545–
17. Stockl D, Baadenhuisjen H, Fraser CG et al (1995) Desirable 550
routine analytical goals for quantities assayed in serum. Eur J 31. Fraser CG (2003) Analytical quality requirements for point-of-
Clin Chem Clin Biochem 33:157–169 care testing. In: Price CP, St John A, Hicks J (eds) Point-of-care
18. Libeer JC, Baadenhuisjen H, Fraser CG et al (1996) Character- testing, 2nd edn. AACC Press, Washington, DC
ization and classification of external quality assessment schemes 32. International Vocabulary of Metrology (2008) Basic and general
(EQA) according to objectives such as evaluation of method and concepts and associated terms, 3rd edn. JCGM (Joint Committee
participant bias and standard deviation. Eur J Clin Chem Clin for Guides in Metrology) 200
Biochem 34:665–678 33. Hyltoft Petersen P, Blaabjerg O, Jensen E, Hegedüs L (2008) Clin
19. Fraser CG, Hyltoft Petersen P, Libeer JC, Ricos C (1997) Pro- Chem Lab Med 46 Special Suppl:S148
posals for setting generally applicable quality goals solely based 34. Westgard JO (2004) Nothing but the truth about quality: essays
on biology. Ann Clin Biochem 34:8–12 on quality management in the healthcare laboratory. Westgard
20. Ricos C, Alvarez V, Cava F et al (2009) Desirable specifications QC Inc., Madison
for total error, imprecision, and bias, derived from intra- and 35. Westgard JO (2001) Six sigma quality design and control:
inter-individual biologic variation. http://www.westgard.com/biodata desirable precision and requisite QC for laboratory measurement
base1.htm. Accessed 1 October 2009 processes. Westgard QC Inc., Madison
21. Ricos C, Iglesias N, Garcia-Lario J-V et al (2007) Within-subject
biological variation in disease: collated data and clinical conse-
quences. Ann Clin Biochem 44:343–352

123

You might also like