Professional Documents
Culture Documents
CX31 / CX41
REPAIR MANUAL
All rights reserved, Reproduction in whole or in part without written permission is prohibited. (1G)
CX31/CX41
INTRODUCTION
The CX31 is a biological microscope newly adopting the UIS optical system and is placed as the
successor to conventional CH30/40 microscopes, while the CX41 microscope is placed as the successor
to the CX40. The difference between these two microscopes is that the CX31 has a brightfield
configuration, while the CX41 is a system microscope that can be used with the CX40/BX2 systems.
This manual has been written on the premise that the service engineer already has experience repairing
the CH30/40 and CX40 microscopes. Therefore, the binocular tube section can be repaired by referring
to CH3-BI45 repair manual. The shape of the frame and the installation procedures are different from
that of CH30/40 or CX40 and so figures of the screw positions are provided at the beginning of section
C. Disassembly and Assembly Procedures.
The focusing guide unit and stage mechanisms are the same as the above conventional microscopes.
However, there are some differences in their installation position and shape such as the focusing guide
unit attached to the frame side, the hook position of wire, the shape of lower stage and X-guide. These
points will be explained in section D. Repair Procedure.
Voltage adjustments will be needed when replacing the rheostat ass’y (DZ290400) or circuit board
(DZ290300) respectively. These adjustments can be made by following the instructions given in this
manual.
This manual uses the following simplified method for explaining the disassembly and assembly
procedures.
- Disassemble them following the steps starting with (A).
- Assemble them following the disassembly procedure in the reverse order, while applying any
needed grease or adhesive, and while following any instructions in the “Remarks” column.
- The “Remarks” column provides important notes and supplemental information. Adjustments and
other procedures that cannot be adequately explained in the “Remarks” column are handled in
separate sections.
[Example]
Disassembly
No. Parts Name Screws Grease Adhesive Remarks
(A) RACK ABSK3X8SA Reference direction: Right from the
2 pcs. (*1) rear.
(B) BOTTOM CUK3X4SA Screw positions: CUKHWB screws
PLATE 3 pcs. (*2) shown in C-2
CUKHWB3X6SA
1 pc.
(C) BASE CUTB3X10SA The base cover ass’y and collector
COVER 3 pcs. (*3) ass’y come off together.
SECTION When assembling, fit the collector
ass’y to the base cover ass’y and
mount them to the frame so that the
protrusion at the front end of base
cover ass’y is inserted into the inner
side of frame (align the right and left
surfaces).
(Screw positions: See C-2.)
Assembly
Supplement Reference page
CX31/CX41
CONTENTS
A. OUTLINE OF PRODUCT
1. Name and Code in OLYMPUS.......................................................................................................... A-1
2. Outline............................................................................................................................................... A-1
3. Features ............................................................................................................................................. A-1
4. Using Conditions............................................................................................................................... A-2
5. Specifications .................................................................................................................................... A-3
5-1 CX31 ........................................................................................................................................ A-3
5-2 CX41 ........................................................................................................................................ A-5
6. Dimensions........................................................................................................................................ A-6
B. INSPECTION STANDARDS
1. Inspection Items and Methods........................................................................................................... B-1
D. REPAIR PROCEDURE
1. Stage Adjustment .............................................................................................................................. D-1
1-1 X-Guide Adjustment ................................................................................................................ D-1
1-2 How to Wind the X-Wire ......................................................................................................... D-2
1-3 X-Shaft Rotation Adjustment................................................................................................... D-2
1-4 X-Wire Tension Adjustment .................................................................................................... D-3
1-5 Y-Guide Adjustment ................................................................................................................ D-3
1-6 Y-Shaft Rotation Adjustment................................................................................................... D-4
1-7 How to Wind the Y-Wire ......................................................................................................... D-5
1-8 Y-Wire Tension Adjustment .................................................................................................... D-5
1-9 X/Y Knob Tension Adjustment................................................................................................ D-6
2. Focusing Guide Adjustment .............................................................................................................. D-7
3. Optical Adjustments .......................................................................................................................... D-9
3-1 Preparations.............................................................................................................................. D-9
3-2 Parfocality and Optical Axis Adjustments ............................................................................... D-9
4. Electrical Unit.................................................................................................................................. D-10
4-1 CX31/41 Wiring Diagram ...................................................................................................... D-10
4-2 Replacing Circuit Board/Rheostat Ass’y ............................................................................... D-10
4-3 Preparation of the voltage adjustments .................................................................................. D-11
4-4 Voltage Adjustments .............................................................................................................. D-11
A. OUTLINE OF PRODUCT
1. Name and Code in OLYMPUS
The following frames are available for the CX2 system.
Model Name Description
CX31RBSF Right handle stage brightfield set CX31 frame
CX31LBSF Left handle stage brightfield set CX31 frame
CX41RF Right handle stage CX41 frame
CX41LF Left handle stage CX41 frame
2. Outline
(1) Microscope for education and laboratory, which provides all of the necessary performance and
functions with cost-efficiency.
(2) The CX31 and CX41 are UIS optical system microscopes.
(3) The CX31 is a set-up microscope for brightfield, and the CX41 is a system microscope.
(4) These microscopes are applicable to the world-wide voltage.
3. Features
CX31
(1) Preparation for observation is easily made since the eyepiece, observation tube, objective and
condenser have been set to the frame.
CX41
(1) Wide range of observation methods possible including phase contrast, simple polarized light, and
darkfield.
- Simple phase contrast illumination possible by attaching CX-PH1/2/3 (ring slit for phase contrast
observation) to CH3-CD/CX-SLC/CH3-CDP (condenser for CX41) in combination with the UIS series
phase contrast objective.
- Darkfield observation is possible by attaching the CH2-FH (filter holder) and CH2-DS (darkfield ring) to
the CX41 condenser.
(2) A slider method is used for the simple polarized light analyzer.
(3) Observation at 2X without uneven illumination possible by attaching the low-magnification
adapter (CX-LA) to the CX41 condenser.
(4) A photo system and various TV cameras can be attached to the U-CTR30-2.
CX31/41
(1) Even better observation images are provided by adopting the UIS optical system.
(2) Adaptable to any power source around the world (world-wide voltage).
(3) 6V30W Halogen bulb is used to provide bright illumination.
(4) Specimens can be easily set and replaced by adopting the inward-facing revolving nosepiece.
(5) Fatigue-free design allows user to operate with hands put on the desktop.
(6) Electric safety and radio wave disturbance standards ensure safety and reliability.
(7) The built-in field diaphragm unit provides high contrast image.
(8) Easy to operate as there is not protrusion by the stage's horizontal (X-axis) directional guide.
(9) The stage and condenser knobs are located at opposite positions to make operations easier.
(10) Convenient hand grips at the front and back of frame make it easy to carry the microscope.
(11) The Plan C series objectives provide excellent image flatness.
(12) The anti-contamination structure of the observation tube along with anti-fungus agent prevents
contamination by harmful particles.
A-1
CX31/CX41 A. OUTLINE OF PRODUCT
4. Using Conditions
CX31
(1) The CX31 was exclusively designed for brightfield observation and so optional units will be
needed for conducting other methods of observations.
*The CX31 is a set-up microscope, but it can still be used in combination with CX-SLC and U-
RFA-2.
CX41
(1) A field diaphragm image cannot be focused by the 40X and 100X objectives when using the
phase contrast condenser (CX-PCD).
(2) Shading may occur when the phase contrast condenser is used with a 4X objective and at the
minimum AS diameter.
(3) The followings are the system assembly restrictions.
1) The field number is 18 when CX-RFA-2 is combined with U-CBI30/CBI30-2 or U-
CTR30/CTR30-2. (This is because shading may occur when the field number is 20 and
PlanC40X or PlanC100X is used.)
*In case of the above combination, the eyepiece adapter (R18: field restricting ring), which is
included with CX-RFA-2, should be attached to the eyepiece.
*The U-CBI30 and U-CBI30-2 (U-CTR30 and U-CTR30-2) were produced at different
locations.
2) When using U-TBI-3, the CWH10X (field number: 20) should be used as eyepiece.
*Vignetting may occur when using WH10X (filed number: 22) with a 4X objective.
(4) CH3-CD, CH3-CDP and CX-SLC are condensers that can be used in combination with CX-LA
(low magnification adapter) and CX-AL (attachment lens).
(5) U-TV1X cannot be attached to the trinocular section due to mechanical interference.
(In this case, the new U-TV1X-2 should be used.)
(6) WHB10X and WHB10-H are the eyepieces that can be combined with U-CBI30-2.
(7) When the condenser for CX41 is combined with CX-AL, FS image can be focused at
magnifications between 10~40X.
(8) FS image cannot be focused when using the condenser for CX41 with CX-LA.
(9) When the condenser for CX41 is combined with CH2-DS, the objectives that can be used for
darkfield observation are N.A.0.65 and lower (40X and lower).
CX31/41
(1) FS image cannot be focused when using a 100X objective.
(2) An AS/FS ghost may occur when stopping down FS/AS diameter to the minimum.
(3) Be sure to turn OFF the main switch and unplug the power cord before replacing the bulb.
(4) When moving the stage, do not rotate the knob after coming into contact with the stopper.
(5) Do not move the stage by holding it or specimen holder without rotating the X/Y knob.
(6) Do not carry the microscope by holding the stage, X/Y knob or observation tube.
(7) When the light intensity adjustment knob is set near its maximum level and the main switch is
turned ON/OFF, there may be cases in which the protection circuit will operate and the lamp will
not light.
(8) Ambient Environment
Temperature: 0~40 degrees C (32~104 degrees F), Humidity: 30~90%, Altitude: within 2,000m
Pollution degree: 2 (based on IEC60664), Installation/Overvoltage Category: II (based on
IEC60664)
(9) When using immersion oil, there is the possibility of deteriorating the condenser lens.
(Use the specified immersion oil by OLYMPUS.)
A-2
CX31/CX41 A. OUTLINE OF PRODUCT
5. Specifications
5-1 CX31
Item Specifications Remarks
Optical System (1) UIS optical system (infinity corrected/
1
compensation-free)
2 Eyepiece (1) Inserted to the eyepiece sleeve Insertion diameter: φ23.2
(2) Field number: 20, Magnification: 10X
(3) The eyepiece is fixed to the sleeve of observation Screw (AC731800)
tube with a clamping screw.
3 Observation Tube (1) Circular dovetail type (mounted part) φ38
Fixed to microscope frame:
CX31RBSF: Attached using a clamping screw Screw (AC743900)
(Allen wrench included)
CX31LBSF: Attached using a clamping knob Knob (AA852600)
(2) With anti-fungus agent
(3) The observation tube can be rotated 360° when
the clamping screw or knob is loosened.
(4) Eyepiece sleeve diameter: φ23.2
(5) Siedentopf type, observation tube inclination: 30°
(6) Interpupillary distance adjustment: 48~75mm
(7) Diopter compensation: ±5 m-1 (left sleeve)
4 Revolving (1) Fixed quadruple nosepiece with inward tilt W20.32 P=36/inch
Nosepiece
5 Objective (1) Plan C series attached to the revolving nosepiece
6 Stage (1) X/Y coaxial vertical knob rotation Rubber grip equipped as
(2) X/Y movement: wire method standard
(3) Guide method: ball guide (CX-SHG)
(4) Operation range: Y direction = 50mm
X direction = 76mm
(5) Graduated scale in X and Y directions
X-scale: 0~50mm
Y-scale: 100~180mm
Minimum unit: 1mm for both X/Y
Minimum reading: 0.1mm by vernier
Graduation: Line width = 0.2mm, Height = 0.2mm
7 Specimen Holder (1) Attached to stage with screws
(2) Holding capacity: standard slide glass (76mm x
26mm)
Hold two slide glasses
Specimen thickness = 0.9mm~1.4mm
(3) Hold specimens with inclination angle of holder
8 Focusing (1) Coaxial coarse/fine adjustment knob
Mechanism (2) Stage height movement by roller guide (rack &
pinion)
(3) Full stroke range: 25mm
(4) Fine scale: 2.5 microns /1 graduation
(5) Coarse adjustment stroke per rotation: 39.6mm
(6) Fine adjustment stroke per rotation: 0.2mm
(7) Coarse adjustment knob
(with the tension adjustment mechanism)
(8) Simple stage upper limit stopper attached Don’t turn fine focus
adjustment knob after
coming into contact with
the stopper.
A-3
CX31/CX41 A. OUTLINE OF PRODUCT
A-4
CX31/CX41 A. OUTLINE OF PRODUCT
5-2 CX41
Item Specifications Remarks
1 Optical System (1) UIS optical system (infinity corrected
compensation-free)
2 Observation Tube (1) Circular dovetail type (mounted part) φ38
Fixed to microscope frame:
Attached using a clamping screw Screw (AC743900)
(2) Siedentopf type
(3) With anti-fungus agent
(4) The observation tube can be rotated 360° when
the clamping screw is loosened.
U-CBI30-2 U-CTR30-2
1) Filed Number 20 20
2) Observation 30° 30°
tube
inclination
3) Interpupillary 48~75mm 48~75mm
distance
adjustment
4) Diopter ±5 m-1 ±5 m-1 (left sleeve)
compensation
5) Light path None None U-CTR30-2: Fixed optical
selector path; binocular section
50%, TV-photo section
50%
3 Revolving (1) Fixed quintuple nosepiece with inward tilt W20.32 P=36/inch
Nosepiece (2) Slider inlet provided
4 Objective (1) Plan C series attached to the revolving nosepiece
5 Stage *Same as CX31.
6 Specimen Holder *Same as CX31.
7 Focusing (1) Stage upper limit stopped by simplified pre- When the stage upper limit
Mechanism focusing lever. is effective, the focusing is
*Otherwise, same as CX31. possible by turning the fine
focus adjustment knob.
8 Condenser Holder *Same as CX31.
9 Filter φ45 filter can be inserted to the collector lens section. φ45.2, depth 1.5mm
A daylight filter (32.5C-2) can be installed in the filter
holder or attachment lens and then attached under the
condenser.
10 Field Iris *Same as CX31.
Diaphragm
11 Transmitted Light *Same as CX31.
Illumination /
Electrical Unit
12 Dimensions Microscope and observation tube combined:
233(W) x 367.5(D) x 423(H) Unit: mm
Stage:
188(W) x 134(D) x 213(H)
*Refer to A-6 for details.
13 Weight Approx. 8.5kg (including the observation tube)
14 Safety Standards *Same as CX31.
*Refer to the Parts List (Binder No. 33-2) for the CX41 system diagram.
A-5
CX31/CX41 A. OUTLINE OF PRODUCT
6. Dimensions
1) CX31
OLYMPUS
CX31
*398.3
305.6
0
3 4
1 6
212.2
160 244.7
233
2) CX41
OLYMPUS
*419.3
CX41
326.6
3 0 4
1 6
212.2
244.7
160
233
Unit: mm
The length marked with an asterisk (*) in the above figures may vary according to the interpupillary
distance. The ∗ numerical figure shown above indicates the distance when the interpupillary distance
is set at 62mm.
A-6
CX31/CX41 B. INSPECTION STANDARDS
B. INSPECTION STANDARDS
1. Inspection Items and Methods
Inspection Item Standards Inspection Methods
Coarse/Fine Coarse focus *5.9~14.7N(600~1500gf) 1) Tie a string around the knob and
Focus adjustment knob *The above standard is then measure the tension when the
Adjustment tension based on the normal knob is rotated.
Knob function in coarse Tension gauge (30N): OT3223
Pre-focusing lever adjustment tension CX41:
(CX41) range 1) Lock the upper limit, lower the
Simple upper limit Adjustment range: stage and return the lock position:
stopper (CX31) 0.5~24.5N(50~2500gf) In the above state, confirm the
sound of the spring returning when
releasing.
2) Confirm that the coarse adjustment
knob does not rotate along with
the lock knob.
CX31:
1) Confirm that there is a complete
stop when coming into contact
with the upper limit stopper.
2) Confirm that the ring of stopper
(AD785000) rotates smoothly in
position adjustment range.
Fine focus 0.2~1.2N 1) Tie a string around the knob and
adjustment knob (20~120gf) then measure the tension when the
tension knob is rotated.
Tension gauge (3N): OT3225
Fine focus 1) Confirm that there is no rough
adjustment knob feeling.
rotation 2) Confirm that the knob does not
rotate naturally.
Focusing Guide Guide movement 0.8~1.5N(80~150gf) 1) Confirm the working force.
Unit Tension gauge (3N): OT3225
Base Cover FS movement 1) Confirm the smooth rotation.
Ass’y FS shape / 1) Confirm that the long side / short
minimum diameter side ratio is 2 : 1 or smaller.
2) Minimum diameter: φ1.7 or
smaller
Collector lens 1) No apparent contamination when
viewing from the top.
Stage Squareness against 0.1/75mm 1) Set the SKN004 to the revolving
optical axis nosepiece and confirm the front-
back, left-right squareness of the
stage.
(Not go through 0.1mm thickness)
Thickness gauge: OT1949
Clearance between In whole area: 1) Confirm a clearance by a thickness
stage and specimen Can go through 0.05mm gauge (OT1949).
holder thickness, but not go
through 0.7mm thickness
X/Y movement 1) Confirm that there is no
interference between the specimen
holder and the vernier, and
between the specimen holder (or X
guide) and the guide cover.
B-1
CX31/CX41 B. INSPECTION STANDARDS
B-2
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Back Plate
Ass’y
Grounding Screw
*2
*1
Inside
from the back
C-1
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Bottom Plate
*1
C-2
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
3. Stage
3-1 X/Y Guide Unit
*4
(B)
*1
(K)
*10
*3 (A)
(M) To stage
(L)
a
(N)
Stage holder
*11 reference *2
(I) *(J) direction
(H)
(G) a
From power
*9 switch
(C)
(D) *8
(F)
* When the X-wire (J) from the holder (I), secure the X-wire with tape to keep the X-
*5 wire wound around pully one time.
*6 (E) This will make adjustment easier if the center shaft (See C-4) does not need to be
*7 removed.
(F) Y-GUIDE-L ABK3X8SA 4 pcs. (*8) OT2008 OT1378 Apply grease to the groove of Y-guide-L when the wire
OT1838 guides are replaced.
See D-4 to apply adhesive.
(G) BALL B3UO 12 pcs. OT2008 Apply grease to the balls.
(H) Y-CASING
(K) SPECIMEN CUK3X4SA 3 pcs. (*11) OT2008 OT1378 Apply grease to the groove of specimen holder mount
HOLDER when wire guides are replaced.
MOUNT
(L) X-GUIDE OT2008 OT1838 The X-guide and specimen holder mount come off
together.
Apply grease to the groove of X-guide when wire
guides are replaced.
Apply adhesive between parts, (K) and (L). (See D-1.)
(M) BALL B3UO 8 pcs. OT2008 Apply grease to the balls.
(N) X-CASING
C-3
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
(X-pulley)
X-shaft ass'y
Grease applying
positions.
Grease: OT2008
Bearing
(X)
(R)
*Apply grease to the inside
of Y-shaft. *2
Grease: OT2008
(W)
(V)
*4
(U)
(Q)
(P)
(T)
(S)
*3
*Apply enough grease to the
grease to fill the groove on the *1
top of X-knob.
(O)
Grease: OT2008
C-4
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
(A)
(D)
(E)
(B)
*3 (I)
*1 *5
(G) (F)
*2
C-5
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
a
*5
(G) (J)
(I)
c
(F)
a
(E)
*1 c
TO STAGE
b
(A) (H)
*6
(B)
Base cover ass’y *3
*4 *2
(C) (D)
b Collector ass’y
* Before disassembling the focusing guide unit, remove the back plate ass’y, disconnect the cables and remove
the stage holder. (See C-3.)
C-6
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Simple Stage
Upper Limit
Stopper
(only CX31)
(A) *1 *2
(F)
(D)
(G)
Coarse Adjustment
(B) Lock for Upper Limit
*7,8 (only CX41)
(E)
Dummy Mount
*3
(only CX31)
(K) (H)
(e)
(f) (L)
(M)
(I)
Stopper Holder
(N)
(only CX41)
(O)
(g) (i)
*4
(m )
(h) (d)
(r)
(a)
(p) (o)
(c)
(b)
(q) *5
(J) (n)
4 Apply grease (OT2008) to the shaft (j) of the fine focus adj. knob ass’y (C) and the sliding
surface:
Apply grease to the outer surface of the shaft. First, spread the grease by inserting it from the left
with rotating the fine adj. knob ass’y. Remove the fine focus adj. knob once. Again apply grease
to the shaft and then insert it to the right side. Use the lens tissue to wipe away any grease
exuding from the right and left.
C-7
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
C-8
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
6. Arm Section
*4 *3
(F)
*2
(B)
(A)
(D) (a)
Apply grease (b)
(OT2008) on the five
click-ball sections:
20mm range each
(d)
*1 *
(C) (G)
(c)
Arm reference
Attaching rubber ring of direction
nosepiece
(CX31)
CX41
No. Parts Name Screws Grease Adhesive Remarks
(A) SLIDER CSK3X6SB
HOLDER ASS’Y 2 pcs. (*1)
(B) COVER PSTB2X4SA
2 pcs. (*2)
(C) SLIDER
HOLDER
(D) CLICK SPRING Be sure to have a click feeling, and the
dummy slider (AQ063400) can be easily
inserted and removed. (Adjust the feeling
by bending the click spring.)
(E) CAP
(F) ARM 41 AB5X65SA OT3189 Apply grease to the clamping screw (a).
4 pcs. (*3) Screw (*3) tightening force: 6N (60Kgf)
Reference direction: To the back and left
from the front. (See the above figure.)
(G) NOSEPIECE CUK2.6X16SA OT2008
4 pcs. (*4)
*CX31 revolving nosepiece cannot be removed from the arm separately, because nosepiece itself needs to be
disassembled to remove from the arm. Therefore, the entire CX31 arm ass’y (AW019700) will be handled as
assembled parts.
*Apply grease (0T3189) to the thread on the screw (c) for CX31RBSF, or knob (d) for CX31LBSF. Apply
grease (0T2012) to the thread of stopper (b). Wipe away any grease exuding from the thread.
C-9
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
7. Base Section
7-1 Overall Configuration
Peel off a paper of double side tape
*The base cover ass’y and hand rest are from the rubber foot, and then
removed only when these parts are replaced attach the rubber foot on the bottom
or the guide cover (E) on C-6 is taken off. of base.
See C-2 to check the bottom section screw Lead the socket cables through
positions. the outside of the heat shield
(Remove the right hand rest if guide cover plate.
needs to be removed. See C-6.) Bind the cables at two
locations.
(with strapping band)
*4
(E)
*5
(F)
*3
(D)
*1
(A)
a (B)
*2
b
(C)
b
C-10
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
*2
(C) (B)
Attach socket
mount (B) while
pushing it in
arrow direction.
(A)
(D) *1
C-11
CX31/CX41 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Spring (F):
grease and adhesive locations (A)
a
(D)
(G)
Collector Ass'y
(F)
c
(A) (E)
(B)
c d
*1
Screws (C)
(CUTB3X10SA)
*Take off the screws (CUTB3X10SA 3 pcs.), and then remove the base cover ass’y and the collector
ass’y. (Refer to C-2 for the base cover ass’y.)
[Collector Ass’y]
No. Parts Name Screws Grease Adhesive Remarks
(A) RING
(B) LENS OT1899 Referring to the above figure: Apply adhesive
to the “d” section (6 locations) except the ring
(A) and hook section. Place the cut section of
the ring (A) between “d” section.
(C) WIRE
(D) FROSTED GLASS
C-12
CX31/CX41 D. REPAIR PROCEDURE
D. REPAIR PROCEDURE
1. Stage Adjustment
This section explains assembly and adjustment procedures based on the premise that the stage is
overhauled. Refer to steps, (A) to (Y), on pages C-3 and C-4 to disassemble the stage. (This section
provides explanation for right-handle stage.)
D-1
CX31/CX41 D. REPAIR PROCEDURE
D-2
CX31/CX41 D. REPAIR PROCEDURE
b Standard 0.3~0.4N(30~40gf)
(2N fan-shaped tension gauge:
(B) OT3232)
D-3
CX31/CX41 D. REPAIR PROCEDURE
D-4
CX31/CX41 D. REPAIR PROCEDURE
Screws:
CUK3X6SA (*1) 2 pcs.(clamping screws)
(f)
CSK3X6SA (*2) 1 pc. (adjustment screw)
Standard 2.5N~3.0N(250~300gf)
(5N tension gauge: OT3224)
D-5
CX31/CX41 D. REPAIR PROCEDURE
After repairing the stage, as well as the focusing guide unit that is explained later, be sure to check the
right/left, forward/backward squareness of the stage (Jig: SKN0004, not go through 0.1mm thickness
gauge).
4 If the setting of stage holder is correctly made in 2), check the assembled condition of focusing
guide and stage guide units, and then readjust the defective part.
D-6
CX31/CX41 D. REPAIR PROCEDURE
Grease: OT2008
Guide
groove 4) Place the CASING (D) in about the
section (D) middle position of the WIRE GUIDES
(E), and set the ROLLERS (C)
(C) alternately so that the direction of top
surface changes each other as shown in
the left figure.
Roller: 8 pcs.
Lower side
5) In the same manner as described above,
place the casing on the upper groove
section on the inner guide and set the
rollers.
D-7
CX31/CX41 D. REPAIR PROCEDURE
Guide working
0.8~1.5N (80~150gf)
force:
(3N tension gauge: OT3225)
D-8
CX31/CX41 D. REPAIR PROCEDURE
3. Optical Adjustments
For the mechanism of binocular section, refer to the CH3-BI45 repair manual.
3-1 Preparations
Tools and Jigs:
(a) KN0048 (standard eyepiece; with adapters-1 and 2) + KC2040 (3mm ring adapter)
(b) KN0041 (standard objective)
(c) BXKN001 (UIS Standard straight tube; only straight tube part is used.)
(d) FT36
D-9
CX31/CX41 D. REPAIR PROCEDURE
4. Electrical Unit
4-1 CX31/41 Wiring Diagram
U003
4 Voltage Adjustment
The circuit board ass’y (AQ802400) consists of the circuit board (DZ290300) and rheostat ass’y
(DZ290400). In case where the above parts are replaced as AQ802400, the voltage adjustments for the
circuit board ass’y are not necessary. However, voltage adjustments are necessary when individually
replacing either the circuit board (DZ290300) or rheostat ass’y (DZ290400). The followings explain
procedures for replacing, preparation and adjustment of the minimum and maximum voltages.
*3
(A ) 2) Disconnect the cables from CIRCUIT BOARD (B)
connectors.
CN1: Output side connector
*2 *4 CN2: Lamp connector
(B ) CN3: Rheostat ass’y connector
C N1
a
(C ) 3) Disconnect the inlet CABLES (C) from the power
switch terminals, and then take off the BACK
C N3 PLATE ASS’Y (A). (No need to disconnect the “a”
a
connection.)
C N2
D-10
CX31/CX41 D. REPAIR PROCEDURE
a
4 Use a new Halogen bulb in the voltage
b(up) adjustments.
(down)
CN2
b
(white)
a
(black)
Halogen bulb
Standard: DC1.10~1.15V
VR23 (adjustment target: 1.123V)
CN3
∗ Do not turn the trimmer VR22 mounted on circuit board 2) Rotate the circuit board’s trimmer VR23 to
because it is adjusted to the prescribed current value. adjust so that the lamp output voltage
(VR22 is set for overcurrent protection) between the CN2’s 1 and 2 pins is within
the standard shown below using a digital
∗ Do not touch the AC inlet terminals with power turned
on because they could cause an electric shock. multimeter.
Be sure to turn off power during disassembly and
reassembly. Standard: DC 5.65~5.75V
(adjustment target: 5.70V)
D-11
CX31/CX41 E. JIGS AND TOOLS
E-1
CX31/CX41 F. LUBRICANTS AND CHEMICALS
2. List of Adhesives
<Adhesives>
No. Name Page
OT1026 Solvent-based adhesive (red) D-9
OT1378 Solvent-based adhesive (transparent) C-3, 8, D-1, 3, 4, 5
OT1837 Anaerobic adhesive (transparent) C-6, D-8
OT1838 Cyanoacrylate adhesive C-3, 12, D-1, 4
OT1899 Elastic adhesive C-12
F-1