You are on page 1of 6

The British Journal of Radiology, 72 (1999), 781±786 E 1999 The British Institute of Radiology

Relative electron density calibration of CT scanners for


radiotherapy treatment planning
S J THOMAS, MA, MSc
Medical Physics Department, Addenbrooke's Hospital, Hills Road, Cambridge CB2 2QQ, UK

Abstract. Several authors have reported data on the variation of Houns®eld numbers with
electron density in CT scanners. The data can be ®tted with a double straight line approach. For
non-bone tissues (or phantom materials with similar atomic numbers) the data from all authors
can be ®tted to a single straight line. For bone-like materials the line varies between authors. The
method used to measure electron density has a greater effect than the differences between
scanners, or the kilovoltage used on a given scanner. The effect of variation of these slopes on the
accuracy of radiotherapy treatment planning is analysed. For typical radiotherapy beams, to
produce a 1% error in dosimetry would require errors of over 8% in bone electron density. Using
a single pair of calibration lines for all the scanners reported would give dosimetric errors of
under 0.8%. A formula is recommended as a default for use in planning systems in circumstances
where no data are available for a particular scanner.

CT is routinely used in radiotherapy treatment dosimetric errors that will result from uncertain-
planning to determine both clinical and physical ties in the HU calibration, to quantify the amount
information. The principal physical information of variation between scanners that is acceptable
comprises the size, shape and location of outlines for treatment planning purposes. This will be
and inhomogeneities and, via the Houns®eld illustrated by an analysis of published data for a
Units (HU), the relative electron density of number of scanners, together with data that have
tissues. been measured on four machines used to provide
HU are de®ned as: CT input to treatment planning at Addenbrooke's
Hospital.
k{k 
w
HU~1000 …1†
kw Method
where m is the attenuation coef®cient, and mw is In radiotherapy treatment planning, the relative
the attenuation coef®cient of water. Attenuation electron density at a point in the patient is used as
coef®cients depend on electron density, atomic the input into a calculation of the dose. Treatment
number (Z) and on the quality of the beam used planning systems use a large variety of algorithms
in the CT scanner. to correct for heterogeneities, ranging from one-
The relationship between relative electron dimensional methods such as the effective depth
density, re, and the CT number in HU has been (ratio of tissue maximum ratios (TMR)) algo-
measured by several authors [1±5]. For materials rithm and the power law algorithm, to methods
with an average atomic number similar to that of based on separation of primary and scatter, kernel
water, the HU lie on or near a straight line that convolution or Monte Carlo methods.
passes through HU521000 for air and HU50 The following analysis is based on the ratio of
for water. Bone substitute materials give HU TMR method. A water-equivalent effective depth
values that lie above this line, with different is calculated, based on the product of depth and
results being observed for different scanners. relative electron density. The ratio of (TMR at
These differences lead to the standard recommen- effective depth) to (TMR at true depth) is used to
dation that the relationship between re and HU calculate a correction factor CF, de®ned as the
be measured for each scanner used for treatment ratio of the (dose corrected for heterogeneities) to
planning. However, in many circumstances, large the (dose at same place if all material were water).
uncertainties in relative electron density give only As will be shown in the discussion, the error in
small uncertainties in dosimetry. dosimetry produced for a given error in relative
The aim of this work is to analyse the electron density is similar for more complex
methods [6, 7].
Received 1 December 1998 and in revised form 3 March Table 1 gives a typical HU, typical thick-
1999, accepted 8 April 1999. ness and typical relative electron density for

The British Journal of Radiology, August 1999 781


S J Thomas

Table 1. Data for typical inhomogeneities encountered in treatment planning. The ®nal column shows the change
in water equivalent thickness caused by the inhomogeneity

Structure Typical HU Typical Relative electron Product of t and


thickness (t) density (re) (re21)

Lung 2740 8 cm 0.26 25.92


Fat 250 4 cm 0.95 20.20
Liver +50 8 cm 1.05 0.40
Rib +600 1 cm 1.34 0.34
Humerus +700 2.5 cm 1.39 0.98
Cranium +900 1.5 cm 1.51 0.77

inhomogeneities found in the human body. The tissues, using a stochiometric calibration tech-
relative electron densities shown in this table are nique. Table 2 summarises the sources of data,
those published by Duck [8]. The ®nal column in listing in each case the scanner, tube potential
Table 1 shows the change in water equivalent (kV) and the bone substitute phantom material
thickness caused by an inhomogeneity of this used.
thickness and relative electron density.
Data from BJR Supplement 25 [9] were used to
calculate the error in the water equivalent depths Results
that would result in a 1% error in dose for a Table 3 shows the errors in relative electron
10 cm610 cm ®eld, and depths of 5 cm and density needed to give ¡1% error in dose.
10 cm, for a range of photon energies. These Greater accuracy is required for thick structures
results were combined with the ®nal column of with electron densities near to that of water (e.g.
Table 1, to give values of the error in relative liver) than for thinner structures of high density
electron density that would produce 1% error in (e.g. cranium). The higher the energy of X-rays
dose calculation at a clinically relevant depth used for radiotherapy, the lower the accuracy
(10 cm for all except cranium, where 5 cm has required in the values of relative electron density.
been used). The values for 6 MV from Table 3 have been
used in Figures 1 and 2.
Figure 1 shows the relationship between CT
Measurement of relationship between CT number and relative electron density for low Z
number and relative electron density materials (i.e. all tissues and phantom materials
other than bone or bone substitutes). All the data
With one exception, all measurements included fall on or near a single straight line:
in this analysis were obtained by imaging
phantoms containing various tissue substitute HU
materials, and tabulating the CT number obtained oe ~ z1:00 …2†
1000
against relative electron density.
For low Z materials (everything except bone This result is not unexpected; for materials of a
substitute phantoms), similar materials were used similar atomic number the attenuation coef®cient
by all authors. Air and water were universally is proportional to electron density, hence this
used. Other materials included plastics (PMMA, result will follow from Equation (1). The bars
polystyrene), wood, organic liquids and resin- shown on the graph at HU of 2740, 250 and +50
based tissue substitutes. As will be seen in show the range of relative electron densities
the results section, the choice of materials corresponding to ¡1% error in 6 MV dosimetry
did not lead to systematic differences between for lung, fat and liver. In all cases the points lie
results. within the range.
For bone substitute materials, the phantom Figure 2 shows the relationship between HU
materials fall into two distinct sets. Several and relative electron density for bone substitute
authors used aqueous solutions of CaCl2. materials. The ®lled symbols are all for measure-
Others used resin mix, to which CaCO3 or ments with aqueous solutions of CaCl2, the
CaHPO4?2H2O had been added. As will be seen hollow symbols are all for resin-based solid
in the results section, the choice of material did phantoms. The crosses are for a stochiometric
lead to systematic differences between authors calibration. In contrast to the low Z situation,
results. there is considerable difference between data sets
The results of Schneider et al [5] were calculated for different machines and for different measure-
from the known chemical composition of various ment techniques.

782 The British Journal of Radiology, August 1999


Electron density calibration of CT scanners for treatment planning

Table 2. Sources of the data used in Figures 1 and 2, giving in each case the scanner, kV, bone substitute phan-
tom material used and symbols used in Figures 1 and 2

Source of data Scanner kV Bone substitute Symbol

[1] EMI CT-5005 Aqueous CaCl2 X


[2] GE 9800 140 Aqueous CaCl2 {
[3] Siemens Somatom 2 125 CB4+CaHPO4 s
[4] Siemens DR3 125 CB2+CaCO3 or SB3 n
[5] GE9000 120 SR1, SR4 ,
[5] GE9000 120 Stochiometric calculations 6
J
N
This paper GE high-speed adv. 120 Aqueous CaCl2
This paper Siemens Som. Plus 120 Aqueous CaCl2
This paper Siemens Som. Plus 140 Aqueous CaCl2 m
This paper Nucletron Simulix 120 Aqueous CaCl2 .

The dashed line shown on Figure 2 is for a for treatment planning, this is not always
straight line with the equation of: practicable. Faced with a choice between using
this CT data, or planning using a non-CT method
HU
oe ~ z1:00 …3† (manual outlines and no inhomogeneity data), it
1950 is better to have a means of using the CT data.
The bars shown on the graph at HU of +700 and Even when a centre uses only a few CT
+900 show the range in relative electron densities scanners, and physics staff have ample access to
corresponding to ¡1% error in 6 MV dosimetry these scanners for the imaging of phantoms,
for humerus and cranium. Notwithstanding the uncertainties remain. The data in Figure 2 show
considerable differences between the data sets, greater variation by phantom material than by
they all lie within these bars. manufacturer or by kV. For relative electron
densities between 1.1 and 1.3, the highest CT
numbers occur for phantoms consisting of aque-
Discussion ous solutions of calcium chloride solution. Since
With the widespread introduction of CT, a these are solutions in water, they all tend back to
radiotherapy centre frequently ®nds itself using an HU of 0.0 at a relative electron density of 1.0,
CT data from a wide range of scanners, spread but then follow slopes with dHU/dre52370¡260.
over the geographical area from which referrals to The slopes (dHU/dre) for solid phantoms are
the centre are made. With the implementation of similar to those for CaCl2 solutions. Solid
cross-manufacturer standards such as DICOM, it phantoms tend to be based on a high density
is increasingly feasible to make use of this data for resin; the resin with no added calcium always lies
radiotherapy treatment planning. Whilst it is on or near the straight line of Figure 1, typically
clearly desirable for full electron density calibra- with an HU of 100 at a relative electron density of
tion curves to be available for every scanner used 1.1. Since the CaCl2 solutions typically have an

Table 3. Percentage change in tissue maximum ratio (TMR) per cm of depth and the change in depth required to
produce a 1% change in TMR. Data are taken from BJR Supplement 25 [9]. These are used to calculate the
range of relative electron densities that would give a dose correction factor within ¡1% of those calculated for
the typical inhomogeneities in Table 1

Cobalt 6 MV 10 MV 21 MV

% Change in TMR per cm of depth


5 cm 4.0 3.0 2.4 0.7
10 cm 5.4 3.6 3.0 2.3
Change in depth to produce 1% change in TMR
5 cm 2.5 mm 3.3 mm 4.2 mm 14.3 mm
10 cm 1.9 mm 2.8 mm 3.3 mm 4.3 mm
Range of relative electron density for ¡1% in dose
Lung 0.24±0.28 0.23±0.30 0.22±0.30 0.21±0.31
Fat 0.90±1.00 0.88±1.02 0.87±1.03 0.84±1.06
Liver 1.03±1.07 1.02±1.08 1.01±1.09 1.00±1.10
Humerus 1.31±1.47 1.28±1.50 1.26±1.52 1.22±1.56
Cranium 1.34±1.68 1.23±1.79 1.23±1.79 0.56±2.46

The British Journal of Radiology, August 1999 783


S J Thomas

Figure 1. The relationship between


Houns®eld Units (HU) and relative
electron density for material of low
atomic number. See Table 2 for the
key to the symbols. The horizontal
error bars show the uncertainty in
electron density required to give 1%
error in dosimetry for a 6 MV
beam passing through typical in-
homogeneities of lung, fat and
liver.

HU of 237 at the same electron density, it follows scanners give essentially the same line, showing
that, for all relative electron densities over 1.1, a that the manufacturers all meet the required
solid phantom will systematically give a lower HU standards; routine quality assurance (QA) for
than a water based phantom. diagnostic purposes should ensure that scanners
Results from stochiometric calibrations tend to meet the requirements for treatment planning.
lie in between these two sets. However, perform- In the region for HU.100, there is considerable
ing a full stochiometric calibration of a large variation both between machines and between
number of CT scanners is beyond the resources of calibration techniques. However, in this region
most radiotherapy departments. It is not the quite large errors in electron density calibration
purpose of this paper to pass an opinion on which are required before a signi®cant error in dosim-
phantom material gives the most clinically etry occurs. The higher the energy of radiation
applicable densities; this analysis presumes only used for therapy the greater the permissible
that the truth lies somewhere between the uncertainty in electron density. If the dashed
extremes of the measured data. line in Figure 2 were used for all machines, the
In the region for HU,100, the bars on greatest error in dosimetry that would occur at
Figure 1 show the accuracy required to achieve 6 MV would be 0.8%. At 21 MV this would
dosimetric accuracy ,1%. The requirements for reduce to 0.5%. Even at 60Co the error in
reproducibility of CT number are also important dosimetry would only be 1.3%.
for diagnostic use. For this reason, most radi- To illustrate that conclusions drawn from an
ology departments will have a system of quality analysis based on effective depth (TMR ratio) are
assurance in place to ensure at the very least that applicable to other methods of inhomogeneity
air gives 21000 HU, and water gives 0 HU. As calculation, the effect of changing density is
shown by Figure 1, there is no great difference calculated for a lung and a humerus by the
between manufacturers. In this region, all effective depth method, a modi®ed power-law

Figure 2. The relationship between


Houns®eld Units (HU) and relative
electron density for bone substitute
materials. See Table 2 for the key
to the symbols. The solid symbols
are all for aqueous solutions of
CaCl2, the hollow symbols for
resin-based solid phantoms. The
horizontal error bars show the
uncertainty in electron density
required to give 1% error in
dosimetry for a 6 MV beam
passing through typical bone
inhomogeneities.

784 The British Journal of Radiology, August 1999


Electron density calibration of CT scanners for treatment planning

Table 4. A comparison between the effective depth, modi®ed power law and ETAR methods of calculating the
correction factor for a 6 MV beam of X-rays passing through an inhomogeneity. Data were calculated using
beam data from a Varian 600C linear accelerator

rew CF value % change in CF from 10% change in re


Eff. depth Power law ETAR Eff. depth Power law ETAR

Lung
0.234 1.221 1.213 1.203 0.6% 0.7% 0.6%
0.260 1.214 1.205 1.196
0.286 1.207 1.197 1.189 0.6% 0.7% 0.6%
Humerus
1.251 0.977 0.980 0.982 1.2% 1.1% 1.0%
1.390 0.965 0.969 0.973
1.529 0.952 0.958 0.963 1.4% 1.1% 1.0%

method [6] and the equivalent tissue±air ratio not clinically signi®cant, as the highest density for
method (ETAR) [7]. Although the correction low Z tissues lies under 100 HU (e.g. skin at
factors produced by the three methods are not HU578) whilst the lowest density bone is well
identical, the dosimetric changes produced by above this value (e.g. spongiosa at HU5260).
small changes in density are of the same order. These equations should not be used for materials
Calculations are given for the lung and with Z values signi®cantly different from those
humerus situations considered above, for 6 MV occurring naturally in the human body. For
X-rays. In each case, the correction factor (dose example, they should not be applied to scans
with inhomogeneity/dose in water) was calculated through metal hip prostheses.
10 cm deep for the density given in Table 1, and In the region where Equation (2) holds, the
for densities 10% higher and lower. For the lung equation agrees with experimental data within
case the region was assumed to extend from a measurement uncertainties for all CT scanners
depth of 1 cm to a depth of 9 cm. For the reported in the literature.
humerus it was assumed to extend from a depth In the region where Equation (3) is recom-
of 4 cm to a depth of 6.5 cm. In each case a mended, its use will not cause dosimetric errors in
10 cm610 cm beam at 100 cm source-to-surface excess of 1%. Using an accurate conversion curve
distance was assumed. for an individual scanner will further reduce these
Table 4 shows the correction factors calculated uncertainties, but care should be taken in using
in each case, both as raw numbers and as measurements with simple bone substitute phan-
percentage changes in dose caused by a 10% toms; the difference from Equation (3) is more a
change in electron density. Although different function of the phantom material chosen than it is
algorithms give different CF values, the change in of the scanner.
CF value caused by a 10% change in electron
density is very similar. This shows that the
analysis in this paper, while based on effective Acknowledgments
depth, is equally applicable to other methods of
calculating the dose. The only signi®cant differ- I am grateful to Tim Fryer, Stuart Yates and
ence is seen when going to a very high density on Tony Geater for performing relative electron
humerus, where the effective depth algorithm density calibration measurements on the
overestimates by 0.4% the effect of increasing Somatom, Simulix and GE scanners.
density by 10%. This strengthens the arguments of
the paper that large errors in electron density are
required before measurable errors in dosimetry References
are observed. 1. Parker RP, Hobday PA, Cassell KJ. The direct use
of CT numbers in radiotherapy dosage calculations
for inhomogeneous media. Phys Med Biol
Conclusions 1979;24:802±9.
2. McCullough EC, Holmes TW. Acceptance testing
The same conversion between CT numbers and computerized radiation therapy treatment planning
relative electron density can be used for all systems: direct utilization of CT data. Med Phys
scanners reported in the literature. For 1985;12:237±42.
3. KnoÈoÈs T, Nilsson M, Ahlgren L. A method of
HU,100, Equation (2) should be used; for conversion of Houns®eld numbers to electron density
HU¢100, Equation (3) should be used. and prediction of macroscopic pair production cross
This gives a discontinuity at HU5100. This is sections. Radiother Oncol 1986;5:337±45.

The British Journal of Radiology, August 1999 785


S J Thomas

4. Constantinou C, Harrington JC, DeWerd LA. An 7. Sontag MR, Cunningham JR. The equivalent tissue±
electron density calibration phantom for CT-based air ratio method for making absorbed dose calcula-
treatment planning computers. Med Phys 1992; tions in a heterogeneous medium. Radiology 1978;
19:325±8. 129:787±94.
5. Schneider U, Pedroni E, Lomax A. The calibration 8. Duck FA. Physical properties of tissue. London:
of CT Houns®eld units for radiotherapy treatment Academic Press, 1990.
planning. Phys Med Biol 1996;41:111±24. 9. BIR/IPEMB Working Party. Central axis depth dose
6. Thomas SJ. A modi®ed power-law formula for data for use in radiotherapy: 1996, BJR Supplement
inhomogeneity corrections in beams of high-energy 25. London: British Institute of Radiology, 1996.
X rays. Med Phys 1991;18:719±23.

Book review
Computed Body Tomography with MRI viewing. The images are annotated and clearly
Correlation (3rd edn and CD-ROM). Ed. by J marked.
K T Lee, S S Sagel, R J Stanley and J P Heiken, In the text, the references are numbered and
1998 (Lippincott, Philadelphia), £165.00 can be accessed immediately by clicking on the
ISBN 0781716691 numbers. In addition, there are facilities to print/
This CD-ROM is the 3rd edition of the well- copy sections of the text, tables and images, which
known textbook of CT and MRI by Lee et al. should prove valuable as a teaching tool.
The CD-ROM requires 8 MB of RAM and 11 MB I have perused through several sections of this
of free hard disk space and will operate on
CD-ROM. The search facility is particularly
Windows systems as well as Apple Macintosh.
helpful, especially when looking up a dif®cult
The CD-ROM contains the text, images and
references of the two volume CT/MRI textbook. diagnostic problem. All aspects of CT/MRI of the
The whole body is covered systematically in body are covered in detail making this a useful
chapters. The text is comprehensive and is reference of cross-sectional imaging CD-ROM for
accompanied by referenced CT/MRI illustrations personal and department libraries.
down the side which can be magni®ed when A K BANERJEE

786 The British Journal of Radiology, August 1999

You might also like