Professional Documents
Culture Documents
RIF-9
USER MANUAL
Ver. 2.0
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
CONTENTS
RIF - 9
USER MANUAL 2
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Used abbreviations
RB – rechargeable battery;
SC – shortening coefficient of a cable;
PC – personal computer;
SW – software;
UM – user manual;
ETL – cable test van;
V/2 – propagation speed in a cable.
Protective Ground
Earth, overall
RIF - 9
3 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
1.1 Purpose
The digital impulse reflectometer RIF-9 is designed to analyse the
discontinuities of cable lines and can be used to conduct the following
measurements for symmetrical and asymmetrical cables:
determining the distance to characteristic impedance discontinuities
or faults;
determining a cable length;
determining the shortening coefficient of a line based on its given
length;
determining types of faults.
The reflectometer provides access to measurement results on a
colour digital display; saved reflectograms can be sent to a computer
via USB or stored on a USB Flash drive for future display and
processing. The reflectometer can interchange data with other
equipment of a cable test van via the copper interface RS-485.
RIF - 9
USER MANUAL 4
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
RIF - 9
5 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Distance
Fault
reflection
Probe pulse Time
Distance
L=(V×T)/2=(V/2)×T=(С/2×g)×T, (1)
RIF - 9
USER MANUAL 6
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
where:
L the distance to a fault;
T the time delay of reflected impulses relative to probing ones;
V the propagation speed of a pulse signal along a line;
С light speed in a vacuum (300 000 000 m/s);
g shortening coefficient.
It is commonly accepted that the propagation speed of pulse signals is
measured using the non-dimensional quantity of the shortening
coefficient g. The shortening coefficient is the speed signals travels
along a line relative to the speed of light in a vacuum; it is determined
according to the following formula:
g=C/V. (2)
Traditionally, CIS countries use the shortening coefficient g, while
European countries apply its reciprocal, propagation speed in a cable
V/2, which is determined according to the following formula:
V/2 = 150/g, m/μs. (3)
In practice, it is believed that the shortening coefficient (or V/2) also
depends on a cable structure. For instance, the "electrical length"
impulses go through when travelling along the cable conductors with a
large lay is bigger than the cable geometrical length measured along
its surface. The easiest way to factor in this variation is to adjust the
shortening coefficient (or V/2) so that the cable length measured with
the device ("electrical length") matches its surface measurement
(the cable actual length). It is exactly how it is done in practice.
Therefore, cables with a similar conductor diameter as well as
insulation type and thickness, but different in terms of the number of
conductors and their lay (pitch strand) can have different shortening
coefficients.
The value of the shortening coefficient (or V/2) depends on what a
cable sheath is made of and is unique for every cable. The value of
the shortening coefficient (or V/2) can be determined experimentally
based on a given cable length.
RIF - 9
7 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
16
4
5
3 1
18 2
19
13
9 14
10 15
6 11
7 12
17
1.5 Marking
On the device there is the following marking:
Marking reference
Description
designation
Protective grounding clamp (5019, IEC 60417-1)
Input (0351, ISO 7000)
Alternating current (5032, IEC 60417-1)
Direct current (5031, IEC 60417-1)
Stand-by mode (5009, IEC 60417-1)
RIF - 9
9 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
1.6 Packaging
The operation manual is wrapped with packaging film in a way
preventing dust and moisture from getting in and placed in a packing
crate.
Each component of the device is placed inside the crate in a way
ensuring that all empty space between it and the crate shell is densely
filled with shock absorbing material.
The crates are supplied with labels containing the device name as well
as shipping and storage instruction signs.
2 PROPER USE
ATTENTION:
ALL THE MEASUREMENTS ARE TO BE PERFORMED WHEN
THE LINE IS CUT OFF ON BOTH SIDES. TO PREVENT A
BREAKDOWN OF THE DEVICE, DISCHARGE THE LINE
BEFOREHAND BY CLOSING THE CONDUCTORS WITH EACH
OTHER AND THE GROUNDING DEVICE!
The mode "TDR" is used when the impedance value of a fault in a cable is not more
than 200 Ω. If the fault impedance is over 200 Ω, other (high-voltage) line analysis
modes are to be used ("ICE", "DECAY", "ARC").
ATTENTION:
BEFORE CONNECTING TO A CABLE LINE, MAKE SURE THERE
IS NO VOLTAGE ON IT AND DISCHARGE THE LINE WITH THE
GROUNDING ROD BY CLOSING THE CONDUCTORS WITH
EACH OTHER AND THE GROUNDING DEVICE. ALL
MEASUREMENTS ARE TO BE PERFORMED WHEN THE LINE IS
CUT OFF ON BOTH SIDES.
THE FOLLOWING IS PROHIBITED:
PERFORMING MEASUREMENTS ON CABLE LINES THAT ARE
UNDER VOLTAGE!
RIF - 9
USER MANUAL 10
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Ground bus
Mains
~230 V
Cable fault
RIF - 9
11 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.4 The screen of the mode "TDR" allows setting the required
characteristic and observing reflectograms on the screen
(see Figure 5).
RIF - 9
USER MANUAL 12
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.6 This menu allows the user to select the device input to be connected
to a cable line. The device allows displaying reflectograms of one, two
or three cable lines, a reflectogram saved previously and stored in the
device memory as well as select the analysis mode of a cable line
(comparative or differential) (see Table 4).
RIF - 9
13 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
ATTENTION:
TO PROLONG THE SERVICE LIFE OF THE
IN-BUILT RELAYS OF THE DEVICE, DO NOT SWITCH ON
MODES NO. 2, 5 – 7 (WITH FREQUENT RELAY SWITCHING)
FOR A LONG TIME (OVER 10 MINUTES)!
2.1.7 The selected inputs of the device are displayed in the top left corner of
the screen, they are highlighted with the colour corresponding to
reflectograms.
RIF - 9
USER MANUAL 14
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.8 Then, select the required distance measurement range (see Figure 7).
2.1.9 Then, set the required duration of a probing impulse (see Figure 8).
RIF - 9
15 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Additionally, the V/2 value can be set using the cable table of the
device. To do it, escape from the "Main Menu", press the button
"Setup" and then select "Cable Type Selection" in the screen
"System Configuration" (see Figure 11).
RIF - 9
USER MANUAL 16
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
In the screen "Cable List" you can select a cable from the table
(see Figure 12). The value V/2 of the selected cable will be applied to
the set device operating mode.
Also, this screen allows entering new cables into the table (button
"Add New Cable") or removing them (button "Delete Cable").
2.1.12 The next adjustment step is matching the device with cable line
impedance. To do it, press the button "Impedance" and, using the
buttons on the display or the encoder knob, set the value of the device
output impedance (see Figure 13).
RIF - 9
17 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
If the device output impedance does not match the cable line impedance,
there will be extra over-reflection, which can be seen on the
reflectogram. Over-reflected signals are always located at equal
distances behind a cable end. When matching, you must bring the level
of spurious over-reflected signals to the possible minimum
(see Figures 13 − 14). If matching has been performed perfectly, the
reflectogram should not show any over-reflected signals.
The impedance of power cables usually falls within the range (20 – 30) Ω.
RIF - 9
USER MANUAL 18
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
At this stage, the initial device adjustment in the mode "TDR" is over.
2.1.13 The button "Y Scroll" enables vertically scrolling reflectograms up and
down using the encoder knob. Sometimes it is necessary to move
apart the reflectograms on the screen to see the differences more
clearly (see Figure 16).
RIF - 9
19 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.14 The button "File" allows saving a reflectogram on the screen in the
device memory (see Figure 17). For saving, the device offers a file name
with the current date and time. If necessary, the user can change the
name of the file to be saved using the keyboard.
2.1.15 Some types of reflectogram with examples of cable faults are shown in
Figure 18.
RIF - 9
USER MANUAL 20
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.16.2 The measured distance between the cursors is shown in the screen
"Distance".
2.1.16.3 To place the cursor more accurately, use scaling and gain.
RIF - 9
21 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.16.4 To place the cursor most accurately, use scaling (see Figure 20) by
pressing the buttons «Zoom In» (scale magnification) and «Zoom
Out» (scale reduction). It can help you achieve a higher resolution for
placing the cursor.
RIF - 9
USER MANUAL 22
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.1.16.6 The distance between faults can be measured using two cursors as
shown in Figure 22.
RIF - 9
23 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
HIGH VOLTAGE UR
PULSE W
GENERATOR ROUT << W
RIF-9
INPUT "I"
CURRENT
CONNECTING
DEVICE
RIF - 9
USER MANUAL 24
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.2.1.2 To adjust the device, firstly set the required measurement range by
selecting the button "Range".
RIF - 9
25 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.2.1.3 Set the value of "V/2" relevant to the cable under test.
2.2.1.4 Set the device synchronization level minus 40% by pressing the
button "Sync".
2.2.1.5 Select the synchronization mode "Normal" by pressing the button
"Sync Mode".
2.2.1.6 Switch on the ETL pulse current generator.
2.2.1.7 In case of a breakdown in the cable, the device display will show a
reflectogram of a recurrent damped wave process.
2.2.1.8 Set the device characteristics (the measurement range "Range", the
gain of the path "Gain") so that the reflectogram shows 3 to 4
waveform cycles.
2.2.1.9 Press the button "Sync" to set the synchronization level equal to
approximately half the signal level in the reflectogram (minus 50 %).
2.2.1.10 Switch of the ETL pulse current generator and set the mode of a
single synchronization "Single" by pressing the buttons "Sync Mode".
2.2.1.11 Then, after a single activation of the ETL current pulse generator, the
device display will show a waveform reflectogram for a thorough
analysis.
2.2.1.12 Using two cursors, measure the distance on the reflectogram
(see Figures 25 and 26).
RIF - 9
USER MANUAL 26
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
ATTENTION:
IN THE MODE "ICE", THE DISTANCE IS MEASURED BETWEEN
TWO NEARBY POSITIVE SIGNAL PEAKS IN THE SECOND OR
THIRD CYCLE OF THE REFLECTOGRAM. IT IS WRONG TO SET
THE CURSORS IN THE FIRST CYCLE!
Note. In the right bottom corner of the reflectogram field there is a picture showing
proper cursor setting in the mode "ICE".
RIF - 9
27 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
RIF-9
INPUT "U"
VOLTAGE
CONNECTING
DEVICE
RIF - 9
USER MANUAL 28
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
One of the waves travels from the place of the breakdown to the
beginning of the cable, and the other one travels to the end of the
cable. Having reached the beginning of the cable, the first wave is
reflected from a high-resistance source and, without changing its
polarity, travels to the place of the fault. Then, everything is repeated:
a breakdown occurs at the place of the fault, which is followed with a
reflection with the opposite sign, and so on. While damping, the wave
process lasts until the wave energy is enough to cause a breakdown
at the place of the fault.
RIF - 9
29 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.3.1.4 To adjust the device, firstly set the required measurement range by
selecting the button "Range" (see Figure 29).
2.3.1.5 Set the value of "V/2" relevant to the cable under test. by pressing the
button V/2 and then rotating the encoder knob.
2.3.1.6 Set the device synchronization level minus 40% by pressing the
button "SYNC" and rotating the encoder knob.
2.3.1.7 Using the button "Sync Mode" select the synchronization mode
"Normal" (this mode corresponds the single shot mode in a
conventional oscillograph).
2.3.1.8 Switch on the ETL high-voltage unit and slowly increase high voltage
until a breakdown occurs. In case of a breakdown in the cable, the
device display will show a reflectogram of a recurrent wave process.
2.3.1.9 Decrease high voltage to 0 V.
2.3.1.10 Repeat steps 2.3.1.8 − 2.3.1.9 several times to adjust the values of
the device characteristics "Range", "Gain" so that the reflectogram
shows 3 to 4 waveform cycles. By increasing the synchronization level
(with the button "SYNC" and by rotating the encoder knob) make the
device show a sharp image of the reflectogram on the screen after a
breakdown in a cable line.
RIF - 9
USER MANUAL 30
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.3.1.11 Using the button "Sync Mode" change the synchronization mode to
"Single" (this mode corresponds the single shot mode in a
conventional oscillograph).
2.3.1.12 After steps 2.3.1.8 − 2.3.1.9, the device display will show a waveform
reflectogram for a thorough analysis.
2.3.1.13 By setting two cursors on the screen, measure the distance to a fault.
The reflectogram corresponds to a recurrent wave process. The delay
between a positive half-wave and a negative one depends on the
cable length from the generator to the place of a breakdown.
Measurements should be performed with the use of two cursors as
shown in Figure 30.
ATTENTION:
IN THE MODE "DECAY", THE DISTANCE IS MEASURED
BETWEEN TWO NEARBY POSITIVE AND NEGATIVE SIGNAL
PEAKS IN THE SECOND OR THIRD CYCLE OF THE
REFLECTOGRAM. IT IS WRONG TO SET THE CURSORS IN THE
FIRST CYCLE!
In the right bottom corner of the reflectogram field there is a picture showing proper
cursor setting in the mode "DECAY".
RIF - 9
31 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
RIF - 9
USER MANUAL 32
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Arc W
maintenance
unit
RIF-9
High Matching
voltage divider
source
RIF - 9
33 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.4.2.4 The initial device adjustment in the mode "ARC" corresponds to the
adjustment in the mode "TDR" (see 2.1.8 – 2.1.12 and Figure 34).
2.4.2.5 After the initial adjustment, before switching on the ETL high-voltage
generator, press the button "STOP" (this mode is set by default).
Following this, the button will change its inscription to "START" and
become green. It means that the device mode has been changed to
the mode of waiting for an initiating impulse on the input "TRIG" from
the ETL. Also, the screen will display the reflectogram of the cable line
without high voltage feed.
RIF - 9
USER MANUAL 34
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
Figure 35 – Mode "ARC", usual measurement without high voltage feed, cable end (break)
is at distance of 1022.3 m
Figure 36 – Mode "ARC"; during high voltage feed short circuit occurred in cable
at distance of 316.9 m
2.4.2.7 The cursor for measuring the distance to a cable fault is set at the
place of the visible discrepancy between the two reflectograms
(see Figure 36).
RIF - 9
35 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.5.2 Then, measure the length of these cables using a reflectogram. With
this purpose, set the cursor "End Cursor" at the end (break) of the
cable, which corresponds to the beginning of a positive impulse
(see Figure 38). The cursor "End Cursor" should be set in the position
0.0 m.
RIF - 9
USER MANUAL 36
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.5.3 When the button "Save L1 end" is pressed, the value in the screen
"Distance", corresponding to the length of the measured additional
ETL cable connected to the device input «L1», will be saved in the
device memory.
2.5.4 The length of other additional ETL cables connected to the inputs
«L2» and «L3» is measured in the same way.
2.5.5 If, after that, in the menu "System Configuration" the option "Hide
lab cable" is selected (see Figure 39), the value of an additional ETL
cable will be automatically subtracted from the measured distance.
Following this, the reflectogram will be displayed beginning with the
end of the additional cable.
RIF - 9
37 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
RIF - 9
USER MANUAL 38
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
2.5.1.2 Having touched the required files in the menu "FLASH OPERATION",
to select them, you can do one of the following:
copy reflectograms from the device memory to an external USB
Flash drive by pressing the button "Copy to Flash";
remove reflectograms from the device memory to an external USB
Flash drive by pressing the button "Move to Flash";
copy reflectograms from an external USB Flash drive to the device
memory by pressing the button "Copy to Device" (see Figure 42).
RIF - 9
39 US E R M AN U AL
KharkovenergoPribor Ltd.
(+380) 57 393 1069
www.kep.ua | info@kep.ua
3 MAINTENANCE
ATTENTION:
TO PREVENT A PROBABLE DANGER OF THE BATTERY
SPONTANEOUS IGNITION, DO NOT LEAVE THE DEVICE
UNATTENDED DURING RECHARGING!
3.1.3 If RB or charging circuits are faulty, the button will be blinking with red
light:
infrequent blinking for 30 seconds means that the battery voltage is
lower than allowed and it can be faulty;
infrequent blinking for 50 seconds means that the battery voltage is
higher than allowed or there is a break in the charging circuits;
frequent blinking for 3 minutes means that during battery recharging
there was an emergency escape from the recharging mode.
The recharging mode will be on again in 25 minutes.
RIF - 9
USER MANUAL 40