You are on page 1of 1

graphite furnace performance

barium
characteristic mass 3 sigma detection limit

6.7 pg 4.5 pg

linearity spectrometer
model Solaar 989QZ

 wavelength 553.6 nm
  bandpass half height 0.5 nm
 lamp current 100% maximum

  
 
 
  
 

background correction None


 high resolution selected
 
 furnace
 
 model GF90 Zeeman

atomisation Normal
 
cuvette ELC
      injection volume 10 µl


 matrix modification none

ash atomise plot signal





 
 





 

    
   

program notes
Sample solutions were prepared in 0.1%
nitric acid.
The characteristic mass with Zeeman
background correction is 9.0 pg, and the
detection limit is 9.9 pg.

issue 08/97