New Technologies of ARIETTA


Diagnostic Ultrasound System
Akifumi Otake1)      Akifumi Sakou2)

First Medical System Technology Headquarters, Hitachi Aloka Medical, Ltd.


1)

2)
Second Medical System Technology Headquarters, Hitachi Aloka Medical, Ltd.

We developed ARIETTA※1 V70 and ARIETTA V60 diagnostic ultrasound systems, the best among the systems in this
class, by inheriting the technologies and the performance of Hitachi Medical and Aloka, for realizing higher performance
and downsizing. Following is the report on the new imaging technologies towards high image quality centering on the
Symphonic Technology.

Key Words: Diagnostic Ultrasound, ARIETTA, Symphonic Technology

1. Introduction
Hitachi Aloka Medical, Ltd. (hereinafter referred to as
Hitachi Aloka) was established in April 2011 as a result of
the integration of Hitachi Group’s diagnostic ultrasound
business. Hitachi Aloka has now become Japan’s largest
ultrasound manufacturing company, marketing 12 different
diagnostic ultrasound systems and more than 250 types of
probes (transducers). However, to establish Hitachi Group’s
brand in the diagnostic ultrasound industry, it was necessary
to integrate the different operability of systems developed by
the two companies and harmonize overlapping functions and
probes as soon as possible.
The process was started by Hitachi Medical Corporation
and Aloka Co., Ltd. evaluating each other. Both had long but
different experiences of global introductions of diagnostic
ultrasound systems 1) 2), and also differed in the performance
of their systems in each discipline. In the venture to develop
the ARIETTA series*1, the goal was to integrate the basic
capabilities, functionalities, usabilities, and new technologies
of both companies to achieve a higher level of customer Figure 1. Diagnostic ultrasound ARIETTA series
satisfaction. Left: ARIETTA V70; Right: ARIETTA V60
The ARIETTA series comprises two models, ARIETTA
V70 and ARIETTA V60. The for mer integrates the
technical assets of the high-end systems of Hitachi Medical
Corporation and Aloka Co., Ltd., and the latter features
compactness and ease of use. Symphonic Technology
encompasses the separate technologies that ser ve as the
pillars of the ARIETTA series (Figure 1).

− 
1− MEDIX-E006 (VOL61, 27-30, 2014)
A. Otake et al.

2. Symphonic Technology the transducer is increased, and for probes with transducers
composed of multiple elements (Figure 4).
Symphonic Technology divides the process from In addition, the ARIETTA series comes with a selection
ultrasonic signal transmission to image display on the of probes made from single crystal piezoelectric material.
monitor into the following five modules: In engineering, a single cr ystal is defined as a material
- Probe in which atoms are regularly aligned with the axes of all
- Frontend cr ystals oriented in the same direction in the material. A
- Beam Former polycr ystal is defined as a random aggregation of single
- Backend crystals. The use of this single crystal piezoelectric material
- Monitor in the transducer allows the direction of the mechanical
It adopts and harmonizes the optimized technologies strains in the material to be aligned in the same intended
throughout the ultrasound system to achieve the highest direction to produce a high piezoelectric effect (Figure 5).
possible image quality (Figure 2).
The most important challenge in this quest is how to
secure efficiency and fidelity in energy transfer between
����
���������&�'� ��� ���� ������
(����'��������������&�'� ���

the system and the subject — in other words, how to )�"��'����������'����


communicate energy with high sensitivity over a broad ��������������
&������"�������
��������
frequency range. For probes and the frontend, in particular,
there are four factors to consider: ���������� ��
��&����������'����
������"�������������
1. Efficiency of electrical energy transfer between the main ����������
���������������
unit and the probe
2. Efficiency of electromechanical energy conversion within ������������� ������������� ������� ��������
������� �� ��
the probe
3. Efficiency of mechanical energy propagation between the ����������������
probe and the subject �� �!������������"�������
����������� ����������������� ������� ������������
4. Shaping and focusing of the ultrasonic beam in the subject
�� �!������������"������
������ ����������#������$� ��������%�
����������� ����������������������

Figure 3: Impedance matching between the probe and


the main unit
Poor matching results in an attenuation of the output waveform and
reduced sensitivity.

Figure 2: Symphonic Technology


Symphonic Technology, the imaging technology adopted in the
ARIETTA series, consists of five high-performance modules. �����*����������������+������������'���
����� ��������"��'����������������'���
(1) Probe
Probe design requires optimization of signal transmission !��������
'������������������
efficiency between the main unit of the diagnostic ultrasound ��4
system and the subject, and calls for an advanced architecture
)�������������

built on engineering technologies in the fields of electricity, �.4


machinery, acoustics, etc 3). We have developed a new group
of probes called Smart Transducer series for the ARIETTA
�/4
series. These probes use the newly developed Multi-layered
Crystal Technology as standard for the medium-frequency
�����*�������
,��-
��������������
�04
convex probes, to achieve yet greater image quality. .����/����0����1����2����3 �������
5������� �������
A transmitted ultrasonic signal reflected by tissues in
the body will be received back at the probe as echo signals. ���������������������6��������� ����������� ������������������

The energy of these echo signals are transferred through


the matching layer that allows ef ficient propagation of
ultrasonic waves, and the vibrations will be converted by the Figure 4: L
 amination using the Multi-layered Crystal
piezoelectric transducer to electrical energy. This electrical Technology
energy then passes through a cable to the preamplifier of the Reduced energy loss and improved sensitivity can be expected.
main unit of the diagnostic ultrasound system. The recent
trend has been for transducers to be composed of multiple
transducer elements to reduce artifact. However, the greater
the number of transducer elements, the smaller the reception
area per element. As a result, the output impedance of the
transducer becomes higher than the load impedance of
the cable and preamplifier, increasing energy loss due to
poor matching and the likelihood of reduced sensitivity of
the signal reception (Figure 3). The Multi-layered Crystal
Technology, where the transducer element is composed of
multiple layers, improves the impedance matching between Figure 5: D
 iagram showing the orientation of
the cable and the main unit. Sensitivity is improved especially mechanical strains in a piezoelectric material
for low-frequency probes where the electrical impedance of A single crystal allows energy to be aligned in the same direction.

− 
2− MEDIX-E006 (VOL61, 27-30, 2014)
A. Otake et al.

The electromechanical coupling factor (k33, a measure of the of individual scattered waves prevents calculation of the
magnitude of electromechanical energy conversion) of lead arithmetic sum for the envelope. As the value does not
zirconate titanate (PZT) is about 70-80%, whereas that of a directly correlate with the scattering intensity, the resulting
single crystal piezoelectric material exceeds 90%. The latter image takes on a fabric-like pattern called “speckle.” Since
material is responsive over a wider frequency range and is many organs are closely packed with scatterers, most of what
also higher in sensitivity. we see in a typical ultrasound image is speckle pattern. To
solve this problem, a variety of technologies have recently
(2) Frontend been proposed for reducing speckle using image processing
Featured in the frontend is the Compound Pulse Wave techniques such as filtering.
Generator, or CPWG+, which uses programmable waveform On the other hand, there are many clinical examinations —
transmission technology to enable high energy sound staging of cir rhosis, for example — in which visual
transmission. Programmable waveform transmission allows experience of speckle patterns and their correlation to
for fine control of the frequency, amplitude, and pulse form tissue properties has been accumulated and exploited as
of the transmitted sound waves to meet the transducer a diagnostic tool. In which case, unconditional removal of
characteristics. speckle may lead to loss of useful diagnostic information,
Previously, it has been necessary to keep the transmitted leaving no room to apply the established image reading
sound pressure low to alleviate the effect of increased probe expertise.
sur face temperature. Setting such a limit, however, can To address this challenge, we equipped ARIETTA with
compromise the full potential of the probe and may result in HI REZ*2 as an image filter to selectively reduce speckle that
poor image quality due to a lack of sensitivity. affects the visibility of tissue structures without sacrificing
The major cause of increased probe surface temperature the information in the speckle pattern that cor relates
is energy loss during electroacoustic conversion within the with tissue properties (Figure 6). Since HI REZ requires
transducer. To overcome this problem, a programmable sophisticated computation for simultaneous removal of
waveform transmitter which reduces energy loss during the speckle in multiple directions, it makes it difficult to secure
formation of transmission waveforms is a valid approach real-time processing by software, so we have developed a
to maximize the potential of the probe. The combination of dedicated UltraBackend hardware to realize the real-time
the above-mentioned Multi-layered Cr ystal technology or image processing in clinical fields that require high frame
the single crystal probe with the CPWG+ to overcome the rates, such as cardiology. HI REZ operates with linear,
problem of increased probe surface temperature has allowed convex, and sector probes, so it is applicable in all clinical
us to provide high quality images both in terms of resolution fields. HI REZ can be used in combination with spatial
and sensitivity. compound imaging to create genuine images with fewer
artifacts.
(3) Beam Former
During signal reception, the Pixel Focus technology used
in the Beam Former contributes to improved resolution. A
digital Beam Former is commonly used in many models. It
receives the echo signals reflected by different targets at a
time that is dependent on the distance of the target from the
individual transducer elements and performs digital delay
processing by which the difference in time taken by the
signal to reach each individual transducer elements is stored
in the memory for correction before the individual signals
are summed in the main unit of the ultrasound system. In a
full digital ultrasound system, received signals are summed
without the use of an analog delay line, resulting in a greater Figure 6: HI REZ
degree of freedom in delay processing and a dramatic Left: Speckle noise in an ultrasound image, Right: Clinical image of
improvement in the precision of received signal focusing. the gallbladder using HI REZ
In addition to conventional capabilities, the Pixel Focus
technology used in the ARIETTA series reduces the loss in (5) Monitor
focus precision by correcting refraction errors in the lens The monitor features the latest 21-inch IPS-Pro*3 liquid
and matching layer of each probe, and allows for improved cr ystal display, developed by reassessing the structural
accuracy with multiple beams by individual apodiziation members of the drive system, electrode structure, liquid
(weighting) in different beam directions during simultaneous cr ystal material, and oriented film material, to provide a
reception of multiple signals. The combined use with a wider field of vision, increased brightness, and greater
new phase-adjusting ASIC (Application Specific Integrated contrast performance. The resulting image is crisp and clear,
Circuit) for high-precision Focus technology allows for the even in a dark examination room.
formation of narrower reception beams.
Fur thermore, ARIETTA’s Beam Former achieves a In conclusion, the ARIETTA series of diagnostic
maximum four-fold increase in frame rate compared to that ultrasound systems are designed without compromise to
achieved with the conventional method of simultaneous achieve the highest possible levels of image quality by strong
multiple processing. As stated above, the use of Pixel performance in all modules from the probe through to the
Focus technology has enhanced spatial resolution, azimuth display monitor.
resolution, and temporal resolution in all modes, and realized
a dramatic improvement in image quality.

(4) Backend
When a scanned object is densely populated with multiple
scatterers, the interference between carrier frequencies

− 
3− MEDIX-E006 (VOL61, 27-30, 2014)
A. Otake et al.

3. Ergonomics and environmental fatigue to the user, are easy to use, compact, and are capable
considerations of storing the entire peripheral equipment required for the
examination.
In recent years, increasing attention has been paid to
musculoskeletal disorders amongst sonographers, mainly in 4. Summary
Europe and the United States 4). In the pursuit of ergonomic
design of the system housing, however, it was difficult to Our latest diagnostic ultrasound systems ARIETTA V70
quantify the operator’s muscle fatigue. To address this and ARIETTA V60 feature the highest performance in their
challenge, we carried out a research study in 2010 to measure class, achieved through enhanced efficiency and downsizing
muscle load using electromyography and determined the without compromising the technologies and capabilities
ideal specifications for a diagnostic ultrasound system. In inherited from Hitachi Medical Corporation and Aloka
particular, the ARIETTA models are designed so that the Co., Ltd. We continue our efforts to meet challenges posed
monitor and the console panel can achieve positions and by advanced examination techniques in all disciplines of
heights such that the operator can work with less muscle diagnostic ultrasound and create standards that will lead the
load in a seated position, and the increased adjustable range increasingly competitive diagnostic ultrasound field.
of the monitor arm allows the operator to view the screen
at an optimal downward angle of 15 degrees (Figure 7). *1 A RIETTA is a registered trademark of Hitachi Aloka
Looking slightly downward reduces the strain on the neck Medical, Ltd.
and shoulders and is expected to contribute to the prevention *2 HI REZ, *4 HI VISION Avius, and Avius are registered
of VDT (Visual Display Terminal) syndrome. trademarks of Hitachi Medical Corporation.
ARIETTA’s improved environmental performance includes *3 I PS and IPS-Pro are registered trademarks of Japan
the coexistence of compactness with high performance. Display Inc.
We have worked to meet this challenge from the initial
phase of development by considering ever y possibility Reference
that reduces the number of parts and boards used in the
system, in order to downsize the system housing while 1) Akifumi Otake: Technology for enhanced image quality
fulfilling the demanding requirements for image quality, in diagnostic ultrasound systems ProSound Alpha 10
functionality, operability, recording, and so on. Fortunately,
and Alpha 7. Japanese Journal of Medical Ultrasound
the last few years have seen advancement in mobile
devices and terminals that have led to the development of Technology, Vol. 33, 44-49, 2008.
programmable high-performance devices (e.g., integrated 2) Naohiro Yoshida, et al.: Development of digital ultrasound
circuits) that can be used in diagnostic ultrasound systems, diagnostic scanner “HI VISION Preirus”. Hitachi Review,
allowing for downsizing and improved functionality through March 2011, 34-39.
optimal programming without compromising the scalability 3) A kifumi Sakou: High-per formance technologies of
of the circuit. As a result of the enhanced efficiency and ultrasound probe. Journal of Breast and Thyroid Sonology,
optimization of these individual parts or their integration Vol.3, No.1, January 2014, 29-33.
by the use of integrated circuits, the ARIETTA series — 4) Katsumi Kondo: Recent IPS Alpha technologies and future
especially ARIETTA V60 — has successfully reduced possibilities. Hitachi Review, October 2014, 60-65.
the number of boards and the volume of the housing to
Research and Development Team, the Japan Society of
approximately 50% compared to our previous conventional
Ultrasonics in Medicine: Recommendations to ensure
model, the HI VISION Avius*4, and decreased its weight to
89 kg. This downsizing has also led to approximately 40% the safety, comfort, and health of ultrasonographers —
reduction in electric power consumption, which is another instruments and working environment to prevent work-
great improvement in its environmental performance. The related musculoskeletal disorder and eye disorder. The
ARIETTA series of diagnostic ultrasound systems cause less Japan Society of Ultrasonics in Medicine, 2012.

Figure 7: Ergonomic design


When seated at the ARIETTA models (left and centre) the hands can be placed on the console panel in a lower position than when seated at a
conventional model (right: Alpha7), and the monitor can be viewed from a slightly downward direction to the recommended maximum angle of 15
degrees.

− 