Elcometer 456: Op - 456 - 3 - English - Book Page - 1 Tuesday, February 16, 2010 3:08 PM
Elcometer 456: Op - 456 - 3 - English - Book Page - 1 Tuesday, February 16, 2010 3:08 PM
English
3
Elcometer 456
Top Models
Operating Instructions
Op_456_3_English.book Page 0 Tuesday, February 16, 2010 3:08 PM
English
3
These instructions apply to the following Elcometer 456 models:
Ferrous (F), Non-Ferrous (NF) and Dual Ferrous/Non-Ferrous (FNF)
3
The following table associates the relevant 456 model number with its Japanese Mutal Recognition Agreement Law
identifiers.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Modifications not expressly approved by Elcometer Limited could void the user's authority to operate the equipment under
FCC rules.
This Class B digital apparatus complies with Canadian ICES-003.
1
Op_456_3_English.book Page 2 Tuesday, February 16, 2010 3:08 PM
CONTENTS
Section Page
1 About your gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 What this box contains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Conventions in these instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5 Quick-start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Fitting the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Battery condition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Fitting probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 The controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Switching the gauge on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.6 Switching the gauge off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.7 The screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.8 Selecting a language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 Taking a reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2
Op_456_3_English.book Page 3 Tuesday, February 16, 2010 3:08 PM
3
Op_456_3_English.book Page 4 Tuesday, February 16, 2010 3:08 PM
4
Op_456_3_English.book Page 5 Tuesday, February 16, 2010 3:08 PM
Thank you for your purchase of this Elcometer 4563 The gauge is available in
Coating Thickness Gauge. Welcome to Elcometer. three versions; Basic,
Elcometer are world leaders in the design, Standard and Top. This
manufacture and supply of coatings inspection manual describes the
equipment. Our products cover all aspects of operation of the
R
5
Op_456_3_English.book Page 6 Tuesday, February 16, 2010 3:08 PM
6
Op_456_3_English.book Page 7 Tuesday, February 16, 2010 3:08 PM
Rechargeable batteries can be used but they will 16% to 33%, replacement required.
only have 25% to 30% of the life of alkaline
batteries. <16%, gauge beeps every 10 seconds
and symbol flashes - immediate
replacement required.
5 loud beeps, gauge switches off
automatically.
8
Op_456_3_English.book Page 9 Tuesday, February 16, 2010 3:08 PM
9
Op_456_3_English.book Page 10 Tuesday, February 16, 2010 3:08 PM
probes:
F1
456 F
90/10 rule (see page 34). Also indicates when CAL DATA STAT S MENU
Softkeys
On/Off key Note: Before switching the gauge on for the first
time read “Selecting a language” on page 12.
Figure 3. Elcometer 4563 control keys
10
Op_456_3_English.book Page 11 Tuesday, February 16, 2010 3:08 PM
2.6 SWITCHING THE GAUGE OFF When the gauge is switched on a welcome
To switch off all gauge information screen may be displayed briefly (4).
types, press and hold 3 seconds
11
Op_456_3_English.book Page 12 Tuesday, February 16, 2010 3:08 PM
12
Op_456_3_English.book Page 13 Tuesday, February 16, 2010 3:08 PM
13
Op_456_3_English.book Page 14 Tuesday, February 16, 2010 3:08 PM
456
09 / 10 / 2000
F1
F
Automatic-switching
CAL DATA STATS MENU
probe symbol
R
(FNF probes only)
R
Battery state
Substrate
Note: If CAL softkey symbol is flashing the gauge 4.2 MAIN MENU
should be recalibrated. This is due to the calibration Gauge configuration and measurement functions
adjustment method having been changed or a are controlled using menus (12). The structure of
probe change - see “Calibration adjustment” on the menus is shown on page 19.
page 24. Batches cannot be created while the CAL
Menu title
softkey symbol is flashing.
Menu contents
Softkey functions
Figure 11. Reading screen in extended mode Some screens allow the status of a feature to be
and showing full set of statistical values. changed e.g. on to off or select or deselect, etc. A
tick box indicates this type of feature. A tick
If is shown flashing in the top right corner of against a menu item indicates the function is
your display, it indicates that your gauge and your activated or selected.
PC have established a Bluetooth® connection.
SEL softkey selects the option displayed and in
When your gauge and PC are connected by
some cases toggles the status of a tick box off/on.
Bluetooth®, you can transfer readings and batches
using ElcoMaster software - see “Transferring Up/Down softkeys Ï Ð move the cursor to the
readings to a computer” on page 44. menu item required. The menus scroll up/down and
a line across the screen indicates the start and end
of the menu.
15
Op_456_3_English.book Page 16 Tuesday, February 16, 2010 3:08 PM
BACK softkey returns the gauge to a previous 4.3 MAIN MENU - EXTENDED MENU OFF
screen. Holding this softkey down will rapidly exit
from any menu and return to the reading screen.
16
Op_456_3_English.book Page 17 Tuesday, February 16, 2010 3:08 PM
17
Op_456_3_English.book Page 18 Tuesday, February 16, 2010 3:08 PM
D
USA GAUGE : Resets gauge to USA default
settings e.g. MM/DD/YY date format and imperial
units.
18
Op_456_3_English.book Page 19 Tuesday, February 16, 2010 3:08 PM
19
Op_456_3_English.book Page 20 Tuesday, February 16, 2010 3:08 PM
20
Op_456_3_English.book Page 21 Tuesday, February 16, 2010 3:08 PM
ALL BATCHES
Figure 19. SINGLE BATCH screen Deletes all batches.
Use the Right/Left softkeys Í Î to locate the Gauge displays ARE YOU SURE? Press NO
batch to be deleted. At either end of the list of softkey to leave data unchanged or YES softkey to
batches the gauge beeps. Press OK when the delete all of the batches.
batch to be deleted is displayed. Gauge displays CAL METHOD
ARE YOU SURE? Press NO softkey to leave data
Allows selection of calibration method from list of
unchanged or YES softkey to delete the batch. If
options - see “Calibration method” on page 25 for
there is no data stored in the memory the gauge
more details.
SETUP
Used to select, change or activate gauge features:
STATISTICS: Activates statistics feature.
Display - see “Display” on page 36.
Set NDFT - see “Set NDFT” on page 34
21
Op_456_3_English.book Page 22 Tuesday, February 16, 2010 3:08 PM
Select stats - see “Select stats” on page 33. TAG - readings still appear in the batch (with a tag-
Softkey enabled - switches STATS softkey on/off. deleted symbol ), but they are not included in
DATA: Activates DATA softkey, SHOW DATE statistical calculations.
STAMP option and selects the DELETED DELETE - readings are permanently deleted.
READING menu (21). PROBE: Only for dual function probes (FNF and F1
2). Changes probe mode.
FNF probes - select from Automatic, F or N.
F1 2 probes - select from F1 or F2.
UNITS: Units are automatically set by the probe
Figure 21. DATA menu screen type, however the user can manually override the
automatic setting. Select from µm, mm, mil, thou or
SOFTKEY ENABLED allows DATA softkey to be
inch.
turned off when EXTENDED MENU is active.
CLOCK/ALARM: Allows setting of time, date and
SHOW DATE STAMP - toggles date stamp on or
alarm functions and setting of the date format
off in Review Batches - see “Review batches” on
DD/MM/YY or MM/DD/YY (23).
page 42.
DELETED READING screen (22) allows readings
to be tagged or deleted.
22
Op_456_3_English.book Page 23 Tuesday, February 16, 2010 3:08 PM
Time and date can be set to be displayed in large taken. See “Transferring readings to a computer”
digits when the gauge is switched off. on page 44 for further information.
AUTO SWITCH OFF: Changes delay before BAUD RATE can be set at values from 1200 to
gauge switches off when displaying Reading 19200. The default value is 9600 baud.
Screen. RS232 BIT IMAGE Toggle tick box to
• Minimum = 1 minute activate/deactivate. When activated, readings are
• Maximum = 10 minutes sent to the RS232 interface as they are taken. All
• Default = 1 minute images and characters are output as bit-maps. This
allows printing on the Elcometer Miniprinter (see
Auto switch off may be disabled by selecting ‘off’ (In
“Miniprinter” on page 55).
this case, switch off using On/Off key .)
RS232 PLAIN TEXT Toggle tick box to
OUTPUT: Selects Bluetooth® PC Reply option,
activate/deactivate. When activated, readings are
baud rate (24) and activates data output via the
sent to the RS232 interface as they are taken. The
interfaces - see “Interfaces” on page 13. gauge sends standard ASCII characters from the
Courier New font sete. This allows printing on
devices other than the Elcometer Miniprinter, e.g.
RS 232 printers or PC via Elcometer software
(page 44) or via HyperTerminal.
Figure 24. OUTPUT screen
23
Op_456_3_English.book Page 24 Tuesday, February 16, 2010 3:08 PM
BEEP VOLUME: Changes volume. When un-ticked, the softkey functions disappear
• 0 = off from the reading screen 5 seconds after the reading
• 5 = loudest screen has been displayed, or 5 seconds after
pressing any key. To view the functions again,
• Default = 3
press any one of the four softkeys.
LANGUAGES: Allows selection of language.
Softkey functions are always visible in menus.
LARGE FONTS: When ticked, displays all menus
OPENING SCREEN: Disables the opening
and screens using double height, single width fonts
(welcome) screens so that the gauge switches on to
(for improved legibility, if needed). display the reading screen.
If a personalised welcome screen has been
downloaded into the gauge (see page 49), Opening
Screen must be activated to display this screen.
24
Op_456_3_English.book Page 25 Tuesday, February 16, 2010 3:08 PM
5.2 PRESET CALIBRATION METHODS Note: When the calibration method is changed, e.g.
The gauge also contains four preset calibration from Smooth to Rough, the gauge will display a
methods which follow relevant standards. These message (28).
set the calibration method and the data collection
method (data collection method is only set when in
batching mode).
ISO: (ISO 19840) This uses the Zero Offset
calibration method with counted average data Figure 28. Recalibration required screen
collection set to 5 readings.
If the NO softkey is pressed the CAL softkey
SSPC: (SSPC-PA2 2004) This uses the 2-Point
symbol on the Readings Screen will flash to warn
calibration method with counted average set to 3
that calibration adjustment is still required. While
readings.
the CAL softkey symbol is flashing new batches
SWEDISH: (SS 18 41 60) This uses the 2-Point cannot be opened.
calibration method with counted average set to 5
If the YES softkey is pressed the calibration
readings.
adjustment procedure is activated - see “Calibration
AUSTRALIAN: (AS 3894) This uses the Zero adjustment procedure” on page 27.
Offset calibration method with a correction value of
1/3 the surface profile peak-to-valley height and a 5.3 CALIBRATION FOILS AND STANDARDS
counted average data collection set to 5 readings.
Calibration adjustment should be carried out with
the appropriate probe on the same type of metal,
the same curvature and similar finish to the item to
be measured. It is best to use an uncoated sample
of the items to be tested.
26
Op_456_3_English.book Page 27 Tuesday, February 16, 2010 3:08 PM
Calibration can be carried out using measured foils errors due to placing the foil labels between the
or coated standards. foils.
FOILS (SHIMS): These are coating thickness
standards which have been measured using Serial no. TK1009
Inspected by
techniques independent of the gauge. They are micron
502
mils/thou
19.75
ideal for calibration because they provide a known
mils/thou
6.93
Inspected by
measured. Calibration certificates for foils are
176.1
micron
available upon request.
When using foils care must be taken to keep the
foils clean and free from dust and to avoid damage Figure 29. Stacking foils to increase thickness
by creasing particularly the thinner foils. Always
remove a foil from its storage wallet before use. COATED STANDARDS: Thickness standards using
typical substrate materials coated with hardwearing
When calibrating a High Temperature PINIP™
materials and measured using techniques
Probe use the special thickness standards supplied
independent of the gauge.
with the probe - see “Calibrating High Temperature
PINIP™ Probes” on page 32. Coated standards are most often used to confirm
that the gauge meets its specifications if it is not
Note: To calibrate 5 mm (200 mils) and 13 mm
possible to use foils (shims).
(500 mils) range gauges it will be necessary to
stack the foils (29). Care must be taken to avoid
5.4 CALIBRATION ADJUSTMENT PROCEDURE
Calibration adjustment can be carried out at any
time by pressing CAL softkey from the reading
screen. To prevent inadvertent calibration
27
Op_456_3_English.book Page 28 Tuesday, February 16, 2010 3:08 PM
28
Op_456_3_English.book Page 29 Tuesday, February 16, 2010 3:08 PM
these readings and the last reading. Repeat Note: - - - indicates over-range (32). Taking a
this action until a stable reading is obtained. reading within range clears this screen.
Average
29
Op_456_3_English.book Page 30 Tuesday, February 16, 2010 3:08 PM
reading. Repeat this action until a stable The gauge will display the option to test the
reading is obtained. calibration of the gauge.
30
Op_456_3_English.book Page 31 Tuesday, February 16, 2010 3:08 PM
31
Op_456_3_English.book Page 32 Tuesday, February 16, 2010 3:08 PM
32
Op_456_3_English.book Page 33 Tuesday, February 16, 2010 3:08 PM
• % between 90% and 100% of NDFT STATS are displayed and the CLOCK is not
• 90/10 pass or fail displayed.
For more information see: Combined
• “Select stats” on page 33. statistics
symbol
• “Statistics terminology” on page 51.
Use Up/Down softkeys to move cursor and SEL ABOUT PSPCI READINGS
softkey to select or deselect the statistical values. Your gauge is capable of displaying PSPC readings
Note: When PSPC readings is selected, no other according to the 90/10 rule as defined in the IMOj
values can be selected and the SET NDFT screen requirements.
will be displayed automatically. To pass the 90/10 rule (43):
See “Set NDFT” on page 34 for further information • At least 90% of all thickness measurements
about PSPC and NDFT. shall be greater than or equal to NDFT, and
• none of the remaining measurements shall be
6.5 SET NDFT
g h
below 0.9 x NDFT.
The factory default value for NDFT is 320 µm
(12.6 mils).
If you need to adjust this value, select ‘SET NDFT’,
use the Up/Down softkeys Ï Ð to adjust the value
and then press OK. If any readings have been
taken in immediate mode, a screen will be
displayed warning that the statistics will be cleared.
Press YES to continue (to change the NDFT value)
or NO to escape (and leave the value unchanged).
g. Default value from PSPC for Ballast Water Tanks i. Performance Standard for Protective Coatings
h. Nominal Dry Film Thickness j. International Maritime Organisation
34
Op_456_3_English.book Page 35 Tuesday, February 16, 2010 3:08 PM
36
Op_456_3_English.book Page 37 Tuesday, February 16, 2010 3:08 PM
37
Op_456_3_English.book Page 38 Tuesday, February 16, 2010 3:08 PM
Note: While the CAL softkey symbol is flashing, The initial OPEN NEW BATCH screen (48) stays
new batches cannot be opened. Calibrate the on while the gauge sets up the batch, as indicated
gauge before opening a new batch. FNF probes by the progress bar.
should be calibrated in F mode and in N mode.
38
Op_456_3_English.book Page 39 Tuesday, February 16, 2010 3:08 PM
The second OPEN NEW BATCH screen (49) . The display toggles between two options,
shows the current batch settings. normal and counted average:
Lower and • NORMAL - Each reading is added to the
Data collection
upper limits
method number of readings and contributes to the
Number of statistical calculation.
Batch
readings in
number • COUNTED AVE: n - Readings are taken in
batch (0)
Limits Calibration preset groupsl, the default being 5 readings,
selected/ type n=5 (50). At the end of each group the gauge
deselected
symbol Calibration beeps and calculates the average for the group
method and stores this average value, which is used
Data collection
Offset for the statistical calculation. The individual
method softkey
readings in the group are not stored. The value
Figure 49. Second OPEN NEW BATCH screen -
Immediate mode
39
Op_456_3_English.book Page 40 Tuesday, February 16, 2010 3:08 PM
40
Op_456_3_English.book Page 41 Tuesday, February 16, 2010 3:08 PM
The second OPEN NEW BATCH screen (52) current batch will re-open when the gauge is
shows the new batch settings. switched back on.
Cloned batch Note: If the probe is changed to another probe of
symbol the same type the gauge will display a warning (53).
41
Op_456_3_English.book Page 42 Tuesday, February 16, 2010 3:08 PM
42
Op_456_3_English.book Page 43 Tuesday, February 16, 2010 3:08 PM
The gauge will display the current settings (56). 7.6 FREE MEMORY
Values shown are for illustration only. This option displays the amount of free memory
available for storage of readings (57).
Adjust the values with Up/Down softkeys. When Figure 57. FREE MEMORY screen
the correct value is displayed, press OK to enter.
With LIMITS ON activated the out-of-limits reading
will be indicated by a triple beep and a red light on
the keypad LED. A reading within limits is indicated
by a single beep and a green light. Out-of-limits
symbols and will be stored in the batch and can
be seen when the batch is reviewed.
43
Op_456_3_English.book Page 44 Tuesday, February 16, 2010 3:08 PM
44
Op_456_3_English.book Page 45 Tuesday, February 16, 2010 3:08 PM
45
Op_456_3_English.book Page 46 Tuesday, February 16, 2010 3:08 PM
If this reply is not received within half a second of This will ensure that the NDFT value is transferred
sending the reading, an error message will be from your gauge in addition to the readings data.
displayed (59).
9 PROBES
An extensive range of probes is available for the
Elcometer 4563 Coating Thickness Gauge. Probes
for ferrous (F), non-ferrous (N) and dual
Figure 59. Bluetooth PC Reply error message ferrous/non-ferrous (FNF) operation are available
as either integral (built-in) or separate options.
If the Bluetooth® connection is still established and
the reply is received after half a second then the Separate probes are fully interchangeable and are
message will clear automatically, (this indicates that available in standard, PINIP™ and miniature
the Bluetooth® connection is operating at the limit of formats.
its range). If the Bluetooth® connection has been PINIP™ format (Plug-In Integral Probe) is an
lost, then the error message will remain on the integral style probe which plugs in to a separate
screen until a key is pressed. gauge. This provides all the benefits of an integral
gauge and the flexibility of a separate gauge in a
8.3 TRANSFERRING PSPC READINGS DATA single unit.
TO ELCOMASTER Miniature separate probes permit measurements
When transferring PSPC readings to ElcoMaster, where space is restricted.
run ElcoMaster and then tick the ‘Use 456 Gauge
NDFT’ box in File>Preferences>Batch Info>NDFT 9.1 FERROUS (F) PROBES
Settings. F probes measure the thickness of non-magnetic
coatings on magnetic substrates. They can be used
46
Op_456_3_English.book Page 47 Tuesday, February 16, 2010 3:08 PM
FERROUS
FNF probes are dual function, F and N in one probe. F1 telescopic 9 8 9
FNF gauges will automatically detect the type of F2 telescopic 9 8 9
substrate and set the mode accordingly. F1 PINIP™ 9 8 9
Alternatively the mode can be set manually - see
F2 PINIP™ 9 8 9
“PROBE:” on page 22 and “Coatings on galvanised
F1 2 PINIP™ High Temp 9 8 9
or metallised steel” on page 48.
F3 PINIP™ 9 8 9
FM3 miniature straight 9 8 9
FM3 miniature 45° 9 8 9
o. Using an N probe (or an FNF probe manually set to
N1) on a ferrous substrate will give a reading, but the
FM3 miniature 90° 9 8 9
reading will be incorrect. FM5 miniature 90° 9 8 9
47
Op_456_3_English.book Page 48 Tuesday, February 16, 2010 3:08 PM
NON-FERROUS
appropriate protective clothing and take
N1 right angle 8 9 9 care to avoid bodily contact with the hot
N1A anodiser’s 8 9 9 surface during measurement.
N1 PINIP™ 8 9 9 Do not exceed the maximum measurement speed
NM3 miniature straight 8 9 9 when using these probes - see page 52.
NM3 miniature 45° 8 9 9 Note: Refer to special calibration procedure - see
NM3 miniature 90° 8 9 9 “Calibrating High Temperature PINIP™ Probes” on
FNF1 * 8 8 9 page 32.
DUAL
48
Op_456_3_English.book Page 49 Tuesday, February 16, 2010 3:08 PM
49
Op_456_3_English.book Page 50 Tuesday, February 16, 2010 3:08 PM
50
Op_456_3_English.book Page 51 Tuesday, February 16, 2010 3:08 PM
13 STATISTICS TERMINOLOGY
Term Meaning
Coefficient of Variation. The standard deviation divided by the mean for a group
COEF OF VARIAT’N
of readings, expressed as a percentage.
HIGHEST READING The value of the maximum thickness in a group of readings.
LOWEST READING The value of the minimum thickness in a group of readings.
The average of a group of readings; the sum of the individual readings divided
MEAN
by the number of readings.
Number of Readings. The running value for the number of readings taken in a
NO. OF READINGS group. In the case of the mode, the Number of Readings is the number of
values recorded, not the total number of readings taken.
Standard Deviation. A statistical measure of the spread of values in a group of
STD DEVIATION
readings.
PERCENT GREATER The percentage (%) of readings greater than or equal to the value of NDFT (see
OR EQUAL TO NDFT “Set NDFT” on page 34)
PERCENT BETWEEN The percentage (%) of readings less than the value of NDFT but greater than
90% AND 100% NDFT 90% of the value of NDFT.
90/10 test results: = pass, = failed on 90% test, = fail due to a reading
90:10 PASS / FAIL below 0.9xNDFT (if a number between 1 and 10 is displayed, insufficient
measurements have been taken to be statistically valid - take more measurements)
51
Op_456_3_English.book Page 52 Tuesday, February 16, 2010 3:08 PM
52
Op_456_3_English.book Page 53 Tuesday, February 16, 2010 3:08 PM
53
Op_456_3_English.book Page 54 Tuesday, February 16, 2010 3:08 PM
54
Op_456_3_English.book Page 55 Tuesday, February 16, 2010 3:08 PM
55
Op_456_3_English.book Page 56 Tuesday, February 16, 2010 3:08 PM
56
Op_456_3_English.book Page 57 Tuesday, February 16, 2010 3:08 PM
a. Whichever is the greater. Lower value achieved when calibrated close to the thickness to be measured.
57
Op_456_3_English.book Page 58 Tuesday, February 16, 2010 3:08 PM
19 PROBE CAPABILITIES
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
58
Op_456_3_English.book Page 59 Tuesday, February 16, 2010 3:08 PM
59
Op_456_3_English.book Page 60 Tuesday, February 16, 2010 3:08 PM
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
60
Op_456_3_English.book Page 61 Tuesday, February 16, 2010 3:08 PM
61
Op_456_3_English.book Page 62 Tuesday, February 16, 2010 3:08 PM
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
62
Op_456_3_English.book Page 63 Tuesday, February 16, 2010 3:08 PM
63
Op_456_3_English.book Page 64 Tuesday, February 16, 2010 3:08 PM
64
Op_456_3_English.book Page 65 Tuesday, February 16, 2010 3:08 PM
20 ERROR MESSAGES
Under certain conditions the gauge will display error messages (60). These messages are normally cleared
by pressing any one of the softkeys. The cause of the error will be indicated by the message and should be
corrected before proceeding.
Error message Causes Action to take
#1 - PROBE Probe-to-gauge communication failure. Integral gauge - return to Elcometer*.
Separate gauge - remove probe and refit.
If error persists, return to Elcometer*.
#2 - PROBE Corrupt data output from probe. Integral gauge - return to Elcometer*.
Separate gauge - try new probe.
If error persists, return to Elcometer*.
#3 - PROBE Internal error. Return to Elcometer*.
PROBE a) Probe is dual FNF, but gauge is ferrous a) Gauge model F and gauge model N
UNSUPPORTED only, or non-ferrous only. See page 46. cannot be used with a dual FNF probe.
b) Old gauge software does not support b) Return to Elcometer* for software
new probe. upgrade.
PROBE Probe has been changed. Calibration adjustment required.
CHANGED
UNSTABLE a) External electro-magnetic interference. a) Isolate from electro-magnetic
READING b) Gauge faulty. interference.
b) Return to Elcometer*.
65
Op_456_3_English.book Page 66 Tuesday, February 16, 2010 3:08 PM
66
Op_456_3_English.book Page 67 Tuesday, February 16, 2010 3:08 PM
21 INDEX
Numerics Precautions 49 Shims 27
2 Point calibration method 25 Rechargeable Calibration Adjustment 27
90/10 rule 34 Charging 53 Procedure, 2Point 31
Life of 8 Procedure, Rough 31
Specification 52 Procedure, Smooth 28
A
Baud rate 23 Procedure, Special substrate 31
Accessories 53 Beep Calibration certificates 54
AS 2331.1.4 6 Changing volume 24 Calibration methods
AS 3894 26 Switching off 24 2 POINT 25
AS 3894.3-B 6 Bluetooth 44 AUSTRALIAN 26
AS/NZS 1580.108.1 6 Icon in display 15 ISO 26
ASTM B 499 6 Transferring readings 44 ROUGH 25
ASTM D 1186-B 6 Bluetooth module 55 SET OFFSET 25
ASTM D 1400 6 BS 3900 (C5) 6 SPECIAL SUB 25
ASTM D 7091 6 BS 3900(C5) 6 SSPC 26
ASTM E 376 6 BS 5411 (11) 6 SWEDISH 26
ASTM G 12 6 BS 5411 (3) 6 ZERO OFFSET 25
Australian calibration method 26 BS 5599 6 Character size 11
Clock and alarm 22
B C Cloning a batch 37
Batch mode 36 Cable Coated thickness standards 27
Batch numbering 41 PC to gauge 55 Spares 54
Batches CAL softkey Coatings
Copying (Cloning) 37 Flashing 26 On galvanised steel 48
Reviewing 42 Calibration On metallised steel 48
Batching 36 Adjustment 26 Coefficient of Variation 51
Creating new batch 37 Coated standards 27 Computer
Data collection method 39 Coated thickness standards 27 Connection cable 55
Exiting 37 FNF probes 24 Computer, transferring data to 44
Opening existing batch 41 Foils 27 Copying a batch 37
Batteries Locking 16
Fitting 8 Methods 25 D
Life of 53
67
Op_456_3_English.book Page 68 Tuesday, February 16, 2010 3:08 PM
68
Op_456_3_English.book Page 69 Tuesday, February 16, 2010 3:08 PM
69
Op_456_3_English.book Page 70 Tuesday, February 16, 2010 3:08 PM
70