Professional Documents
Culture Documents
Release B.01
English
Notice
Proprietary Information
This document contains proprietary information, which is protected by
copyright.
Copyright
Copyright © 2011 Koninklijke Philips Electronics N.V.
Trademark Acknowledgements
SureSigns is a registered trademark of Koninklijke Philips Electronics N.V.
Other product names may be trademarks of their respective owners.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover, MA 01810
(978) 687-1501
Document Number
4535 642 11121
Warranty Disclaimer
The information contained in this document is subject to change without notice.
Philips Medical Systems makes no warranty of any kind with regard to this
material, including, but not limited to, the implied warranties or merchantability
and fitness for a particular purpose. Philips Medical Systems shall not be liable
for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
New editions of this document incorporate all material updated since the
previous edition. Update packages may be issued between editions and contain
replacement and additional pages to be merged by a revision date at the bottom
of the page. Pages that are rearranged due to changes on a previous page are not
considered revised.
The documentation printing date and part number indicate its current edition.
The printing date changes when a new edition is printed. (Minor corrections and
updates that are incorporated at reprint do not cause the date to change.) The
document part number changes when extensive technical changes are
incorporated.
Text Formatting
The following typographical conventions are used in this guide:
Decimal Points
Because the SureSigns monitor uses a period (.) to indicate a decimal point in
decimal numbers (for example, 10.0), all decimal numbers in this guide use a
period as a decimal point. Commas are not used as decimal points.
Explanation of Symbols
15%
limitation
K3$
PLQ
Declaration
The SureSigns VSi vital signs monitor is a Class IIb device and complies with
0123 the requirements of the Council Directive 93/42/EEC of 14 June 1993
concerning medical devices and carries CE-marking accordingly.
The radio device used in the SureSigns VSi vital signs monitor is in compliance
with the essential requirements and other relevant provisions of Directive
1999/5/EC (Radio Equipment and Telecommunications Terminal Equipment
Directive).
Authorized EU Representative
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen
Germany
Rx Only
Caution United States Federal Law restricts this device to sale by or on the order of
a physician.
2. Basic Operation
The Front Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
The Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Setting up the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
On/Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Deep Sleep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Changing the System Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Mounting the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Main Screen Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Vital Signs Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Patient Records Display Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Changing the Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Tool Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Changing System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Saving Patient Records to a USB Flash Drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Viewing Monitor Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Networked Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Verifying the Network Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Synchronizing Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Using the Monitor Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Contents-1
SureSigns VSi Instructions for Use
Entering Patient IDs by Scanning Individual Barcodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Entering Patient IDs with a Programmed Barcode Scanner . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Saving Measurements Without a Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Viewing Records in the Patient Records Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Selecting an Existing Patient ID. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Editing a Patient Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Editing a Patient Record Using the On-screen Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Editing a Patient Record by Scanning Individual Barcodes . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Editing a Patient Record Using a Programmed Barcode Scanner . . . . . . . . . . . . . . . . . . . . 3-22
Deleting Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Deleting Specific Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Deleting All Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
4. Alarms
Visual Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Flashing Numeric Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Alarms Off Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Audible Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Speaker Malfunction Notification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Latched and Non-Latched Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Adjusting the Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Silencing Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Audio Pause Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Audio Off Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Testing Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Nurse Call System Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Alarms Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
5. Monitoring SpO2
Selecting an SpO2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Connecting SpO2 Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
The SpO2 Numeric Pane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Changing SpO2 Response Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
SpO2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
6. Monitoring NBP
NBP Measurement Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Selecting an NBP Cuff. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Connecting the Cuff and Hose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Contents-ii
SureSigns VSi Instructions for Use
NBP Numeric Panes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Changing NBP Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Configuring the Initial Inflation Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Changing the NBP Units of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Recalculating the NBP Value if the Limb is not at Heart Level . . . . . . . . . . . . . . . . . . . . . . .6-6
Stopping an NBP Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
NBP Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
7. Monitoring Temperature
Connecting the Temperature Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
The Temperature Numeric Pane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Taking a Temperature Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Changing Temperature Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Changing the Temperature Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
Changing the Probe Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
Changing the Temperature Units of Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Verifying the Temperature Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Temperature Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
11. Specifications
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-1
Safety Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2
Contents-iii
SureSigns VSi Instructions for Use
Environmental Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
NBP Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Oscillometric NBP Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Temperature Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
SpO2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8
Interface Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-10
Radio Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
USA — Federal Communication Commission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
Industry Canada . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
European Union . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
12. Troubleshooting
Viewing System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Diagnosing a Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Troubleshooting Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2
Alarms and Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5
A. Alarm Specifications
Technical Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
B. Electromagnetic Compatibility
Instructions for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Reducing Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
Emissions and Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-3
Guidance and Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
Contents-iv
SureSigns VSi Instructions for Use
1
Overview
This guide describes how to operate the SureSigns® VSi vital signs monitor. For
information about setting up the monitor, see the SureSigns VS2+ and VSi
Installation and Configuration Guide, which describes how to install the battery,
power up the monitor, and configure some of the system settings before using
the monitor.
Warning Before each use, inspect the monitor and accessories for deterioration or
damage. Replace any damaged equipment or report it to your system
administrator.
Overview
SureSigns VSi Instructions for Use 1-1
Indications for Use
Note — This document includes information about functions and features that
are not available on all SureSigns VSi monitors or may only be available in
selected countries. For information specific to your country, please contact your
local Philips representative.
The SureSigns VSi is a vital signs monitor that can measure blood pressure,
pulse rate, oxygen saturation (SpO2), and temperature. Features include:
• NBP
• SpO2
• Temperature
Overview
1-2 SureSigns VSi Instructions for Use
Intended Use
Intended Use
The SureSigns VSi vital signs monitor is for measurement of multiple
physiological parameters of adults, pediatrics, and neonates in healthcare
environments. Additionally, the monitor is intended for use in transport
situations within a healthcare facility.
863275 ●
863276 ● ● o
863277 ● ● ● ●
Overview
SureSigns VSi Instructions for Use 1-3
SureSigns VSi Documentation
Overview
1-4 SureSigns VSi Instructions for Use
2
Basic Operation
This chapter describes how to begin using the SureSigns VSi monitor.
For information about setting up and configuring the monitor, see the SureSigns
VS2+ and VSi Installation and Configuration Guide.
Basic Operation
SureSigns VSi Instructions for Use 2-1
The Front Panel
Alarm Silence Press to pause an alarm for a specified period of time. For more
key information, see Chapter 4, “Alarms.”
Main Screen key Press to close a menu and return to the main screen. When the
main screen is displayed, press to switch between display modes.
On/Standby key Press once to power up the monitor. Press again to enter Standby
mode.
Charging LED Changes color based on the charging status of the battery. For
more information, see “Charging the Battery” on page 2-5.
AC Power LED When lit, indicates that the monitor is connected to an AC power
source.
Basic Operation
2-2 SureSigns VSi Instructions for Use
The Rear Panel
USB port
Ethernet port
Equipotential
Power connector Grounding Post
Connector Description
Basic Operation
SureSigns VSi Instructions for Use 2-3
Setting up the Monitor
Connector Description
T 1.0A 250V
Powering Up
The monitor operates on AC power or the internal battery.
Step
Basic Operation
2-4 SureSigns VSi Instructions for Use
Setting up the Monitor
Basic Operation
SureSigns VSi Instructions for Use 2-5
Setting up the Monitor
Note — To ensure that the battery is sufficiently charged, keep the monitor
plugged in to AC power when it is not in use.
Three indicators show the charging status of the battery: the Charging LED, the
Battery Status pane, and the large battery icon.
Charging LED
The Charging LED on the front panel provides the charging status of the battery.
Charging LED
The color of the LED indicates how much charge remains on the battery:
If the monitor is On (with battery or AC power), the Battery Status pane at the
bottom of the monitoring screen indicates battery status. The number and color
Battery Status of bars in the pane indicates how much charge remains on the battery:
Basic Operation
2-6 SureSigns VSi Instructions for Use
Setting up the Monitor
Your system administrator can configure the monitor so the large battery icon
does not appear on the monitor screen.
Note — If the battery level drops below 12%, the monitor enters Deep Sleep
mode. No measurements are taken and no alarms sound. For more information,
see “Deep Sleep Mode” on page 2-8.
On/Standby Mode
If you press the On/Standby key while the monitor is On, the monitor goes into
Standby mode and the following occurs:
Basic Operation
SureSigns VSi Instructions for Use 2-7
Setting up the Monitor
To resume monitoring, connect the monitor to an AC power source and press the
On/Standby key to turn the monitor back On.
Basic Operation
2-8 SureSigns VSi Instructions for Use
Changing the System Date and Time
Step
Basic Operation
SureSigns VSi Instructions for Use 2-9
Mounting the Monitor
4 Press the Select key, and then press the Up/Down key until the
desired value appears.
7 Press the Up/Down key to highlight the Apply button, and then
press the Select key to save your changes and close the menu.
You can change the date format (mm/dd/yyyy or dd/mm/yyyy) and you can
hide the time display using options in the System Menu. For details, see
“Changing System Settings” on page 2-15.
Caution If your monitor is mounted on a roll stand, use the handle on the roll stand
to move the monitor. Do not use the monitor handle or the antenna to move
the monitor; doing so creates stress on the mounting bracket and could
cause the monitor to fall off the roll stand.
Note — The weight of objects placed in the basket of a roll stand must not
exceed 3.6 kg (8 lb).
For information about mounting the monitor, see the instructions for use
provided with the mounting hardware.
Basic Operation
2-10 SureSigns VSi Instructions for Use
Main Screen Display
Note — The illustrations in this guide show the screens on a fully configured
SureSigns VSi monitor.
Basic Operation
SureSigns VSi Instructions for Use 2-11
Main Screen Display
• The Date/Time pane displays the current time. You can hide the time, as
described in “Changing System Settings” on page 2-15.
• Patient Records buttons are used to view, save, edit, delete, or cancel
changes to patient records. For more information, see
Chapter 3, “Creating a New Patient Record.”
• The scroll bar is used to scroll through patient records when the display is
in Patient Records mode. For more information, see Chapter 3, “Creating
a New Patient Record.”
Basic Operation
2-12 SureSigns VSi Instructions for Use
Main Screen Display
Adult 0009875433
122 98
SYS mmHg SpO2%
DIA
80 80
bpm SpO2%
Numeric
panes
98 37.1
MAP ºC Oral
03:31:00
Basic Operation
SureSigns VSi Instructions for Use 2-13
Main Screen Display
If you highlight a record, the patient IDs for the highlighted record appear in a
separate pane at the bottom of the screen.
If you do not enter a patient ID, the text ID Unknown is displayed. You can also
edit or delete patient records. For more information about working with records
in the Patient Records table, see Chapter 3, “Creating a New Patient Record.”
03:31:00
Basic Operation
2-14 SureSigns VSi Instructions for Use
Changing System Settings
Tool Tips
When you press the Up/Down key to highlight a button in the main screen, a
description appears, as shown in the following illustration.
Step
Basic Operation
SureSigns VSi Instructions for Use 2-15
Changing System Settings
3 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main
Screen button, and then press the Select key.
Basic Operation
2-16 SureSigns VSi Instructions for Use
Changing System Settings
Warning Exported patient records contain patient IDs and patient data. Ensure that
the exported data is handled according to your facility’s electronic protected
health information (ePHI) guidelines.
Use the Save Patient Records button on the System Menu to export patient
data to a USB flash drive. The exported file is a space-delimited .csv file with
the name RecordLog_Monitor Name.csv.
Step
1 Insert a USB flash drive in the USB port on the back of the
monitor.
2 Open the System Menu and press the Up/Down key to highlight
the Save Patient Records button.
Basic Operation
SureSigns VSi Instructions for Use 2-17
Changing System Settings
• Serial Number — The monitor’s serial number, which also appears on the
back of the monitor. The serial number is configured in the factory.
• Hardware ID — The version for each of the following hardware
components:
<Main board> - 0 - <FPGA ID>
• Software Version — The software version installed on the monitor.
• LAN MAC Address — The unique MAC address assigned to the monitor.
The MAC address is configured in the factory.
• LAN IP Address — The IP address currently assigned to the monitor.
• WLAN MAC Address — The unique wireless MAC address assigned to
the monitor. The MAC address is configured in the factory.
• WLAN IP Address — The wireless IP address currently assigned to the
monitor.
• Language — The language currently configured on the monitor.
• Configuration — The monitor type and the parameters installed in the
monitor.
Basic Operation
2-18 SureSigns VSi Instructions for Use
Changing System Settings
Note — The IP addresses are 0.0.0.0 (all zeros) until a network connection is
established.
Step
1 Press the Up/Down key to highlight the System button, and then
press the Select key.
The System Menu appears. Current settings are displayed.
2 Press the Up/Down key to highlight the System Info button, and
then press the Select key.
The System Information window appears.
3 Press the Main Screen key on the front panel to close the
window.
Alternative: Press the Up/Down key to highlight the Return
button, and then press the Select key to close the window.
Basic Operation
SureSigns VSi Instructions for Use 2-19
Networked Monitors
Networked Monitors
You can connect the VSi monitor to a network by using a standard wired LAN
connection, a wireless LAN interface, or a serial interface adapter. A networked
monitor can export patient records from the monitor to an access point or router,
and then forward the data to an HL7 interface server or EHR system.
If your monitor is networked, the records in the Patient Records table change
from white to green after they have been exported successfully. Records that are
not exported successfully change from white to blue, and a message indicating
the reason the record was rejected may appear on the screen.
Caution If you are using the optional serial interface adapter to export data and you
disconnect the adapter to move the monitor to a different location, make
sure the black sheath completely covers the RS-232 connector after you
reconnect the cable.
If the monitor detects a LAN IP address and a WLAN IP address, the monitor
first selects LAN for network communication.
Basic Operation
2-20 SureSigns VSi Instructions for Use
Networked Monitors
If the clock on the monitor differs by more than five seconds from the clock on
the server, the monitor will automatically adjust the date and time, unless:
• A patient record is open; the time change will occur after the record is
closed.
• An NBP or temperature measurement is in progress; the time change will
occur when the measurement is complete.
If the time difference is greater than 30 seconds, the following occurs:
Note — If the system clock on the server or EHR system adjusts automatically
for daylight saving time, the monitor will synchronize with the server clock. If
the system clock on the server does not adjust automatically for daylight saving
time, you must manually change the time on the monitor.
Basic Operation
SureSigns VSi Instructions for Use 2-21
Using the Monitor Safely
SureSigns VSi vital signs monitors conform to CISPR 11. SureSigns VSi vital
signs monitors are suitable for use in all establishments except domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Ensure that the monitor is in working condition before clinical use. If the
accuracy of any measurement does not seem reasonable, first check the patient’s
vital signs by alternative means and then with the monitor to make sure it is
working properly. Always verify that the monitor’s settings match your intended
selections.
If you connect the monitor to any instrument, verify proper operation before
clinical use. Refer to the instrument’s Instructions for Use for full instructions.
Anyone who connects additional equipment to the signal input port or signal
output port configures a medical system and is therefore responsible to ensure
that the system complies with the requirements of system standard IEC Standard
60601-1-1. If in doubt, contact the Philips Customer Care Center or your local
Philips representative.
The monitor and its accessories must be tested by qualified service personnel at
regular intervals to ensure that performance has not been degraded by aging or
environmental conditions. Periodic performance verification tests must be
performed, as described in the SureSigns VS2+ and VSi Service Guide.
Basic Operation
2-22 SureSigns VSi Instructions for Use
Using the Monitor Safely
Warning This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and operated
with a minimum distance of 20 cm between the monitor and any person.
Do not open the monitor or attempt to change the battery. If you suspect a
problem with parts within the monitor, contact your biomedical engineer or
local Philips Representative.
Before you begin monitoring, make sure that the correct patient type is
selected. The monitor’s default initial cuff inflation pressure is based on the
selected patient type.
Do not place the monitor in any position that might cause it to fall on the
patient. Do not lift the monitor by the AC power cord or patient
connections.
Use only grounded power cords (three-wire power cords with grounded
plugs) and grounded electrical outlets. Never adapt a grounded plug to fit an
ungrounded outlet by removing the ground prong or ground clip.
Basic Operation
SureSigns VSi Instructions for Use 2-23
Using the Monitor Safely
Do not use the monitor or SpO2 sensors during magnetic resonance imaging
(MRI) scanning. Induced current could potentially cause burns. The
monitor may affect the MRI image, and the MRI unit may affect the
accuracy of the monitor’s measurements.
Do not connect this monitor to any equipment or device, other than those
specified in this manual.
Use only approved accessories with the SureSigns VSi monitor. The use of
unapproved accessories can diminish monitor performance or safety.
Consult the Instructions for Use that accompany the accessories.
Basic Operation
2-24 SureSigns VSi Instructions for Use
Using the Monitor Safely
• Set the main screen to Vital Signs mode to hide patient records. For
more information, see “Main Screen Display” on page 2-11.
Basic Operation
SureSigns VSi Instructions for Use 2-25
Using the Monitor Safely
Basic Operation
2-26 SureSigns VSi Instructions for Use
3
Creating a New Patient Record
This chapter describes how to save vital signs measurements to a patient record
and how to work with records in the Patient Records table.
03:31:00
Each entry contains a patient ID (or ID Unknown if you do not enter an ID), the
patient type, date and time, and up to four measurements. The Time column
shows the time at which the last measurement completed. A question mark (-?-)
in the record indicates an invalid measurement, and dashes (--) indicate that the
parameter was not measured.
Note — If your monitor is networked, the records in the Patient Records table
change from white to green after they have been exported successfully. Records
that are not exported successfully change from white to blue, and a message
indicating the reason the record was rejected may appear on the screen.
Button
Button Description
Name
Delete If the record is closed, select the Delete button to
delete the saved record.
Button
Button Description
Name
See the procedures later in this chapter for detailed information about using
these buttons.
Patient ID Overview
Your system administrator configures your SureSigns VSi monitor to display
any or all of the following patient ID input fields in the New Patient Menu:
The available patient ID input fields can only be changed by your system
administrator in the password-protected System Admin Menu. In this guide, the
term patient ID is used to refer to any of the patient ID types listed above.
When you scroll through the Patient Records table, the display splits into two
panes. The top pane displays the patient records stored on the monitor, and the
bottom pane displays all of the patient IDs for the highlighted record.
Warning The procedures in this section instruct you to enter patient IDs in the New
Patient Menu before taking the patient’s vital signs measurements. Follow
the steps in the order they are written. Taking measurements before entering
patient IDs may result in sending measurements for the wrong patient ID to
the EHR.
Caution Before you begin monitoring, make sure that the correct patient type is
selected. The monitor’s initial cuff inflation pressures are based on the
selected patient type.
Your system administrator configures the method for entering patient IDs in the
password-protected System Admin Menu. You can enter a patient ID by using
one of the following methods:
• The on-screen keyboard. See “Entering Patient IDs Using the On-screen
Keyboard” on page 3-5.
• A barcode scanner that reads each patient ID one at a time. The monitor
prompts you to scan the barcode for each patient ID field. See “Entering
Patient IDs by Scanning Individual Barcodes” on page 3-7.
• A programmed barcode scanner. When you scan the barcode, the monitor
automatically enters the scanned data in the patient ID fields. See
“Entering Patient IDs with a Programmed Barcode Scanner” on
page 3-10.
Step
11 2 3 4 5 6 7 8 9 0 - =
q1 w e r t y u i o p [ ]
a1 s d f g h j k l ; ‘ \
z1 x c v b n m , . / Back
Shift OK Cancel
4 To enter an ID:
• Press the Up/Down key to highlight a character, and then
press the Select key after each selection.
• If you enter an incorrect character, use the Back button to
erase the character or use the Cancel button to start over.
• Use the Shift button to access uppercase characters and
additional special characters.
When you are done entering the ID, press the Up/Down key to
highlight the OK button, and then press the Select key to return
to the New Patient Menu.
A new row appears at the top of the Patient Records table. The
new row contains the patient ID and the patient type. The text in
the new row is red.
Before you begin scanning, be sure that you are familiar with the barcodes used
at your facility. When the monitor prompts you to scan a patient ID, you need to
know which barcode corresponds with the highlighted ID field.
Note — If the barcode scanner cannot read the barcode or does not function
correctly, you can unplug the scanner and use the on-screen keyboard to enter
patient IDs. For more information, see “Entering Patient IDs Using the
On-screen Keyboard” on page 3-5.
Step
1 Hold the scanner over the barcode, pull the trigger, and center the
beam on the barcode.
Note — To get a proper read:
4 Hold the scanner over the barcode, pull the trigger, and center the
beam on the barcode.
The scanned ID appears in the corresponding patient ID field, and
the next enabled field is highlighted. A message at the bottom of
the screen prompts you to scan the next patient ID.
Before you begin scanning, make sure you are familiar with the barcodes used at
your facility.
Note — If the barcode scanner cannot read the barcode or does not function
correctly, you can unplug the scanner and use the on-screen keyboard to enter
patient IDs. For more information, see “Entering Patient IDs Using the
On-screen Keyboard” on page 3-5.
Step
1 Hold the scanner over the barcode, pull the trigger, and center the
beam on the barcode.
Note — To get a proper read:
3 If necessary, hold the scanner over the next barcode, pull the
trigger, and center the beam on the barcode.
The scanned identifiers appear in the corresponding patient ID
fields.
Caution Before you begin monitoring, make sure that the correct patient type is
selected. The monitor’s initial cuff inflation pressures are based on the
selected patient type.
This section describes how to take a set of vital signs measurements, and save
the measurements to the Patient Records table, without entering a patient ID. If
you save a record with an unknown ID and then decide to assign an ID, you can
edit the ID, as described in “Editing a Patient Record” on page 3-18.
Step
3 Press the Up/Down key to highlight the Patient Type field, and
then press the Select key. Press the Up/Down key to select a
patient type.
Press the Select key again to save the selected patient type.
4 Press the Up/Down key to highlight the OK button, and then press
the Select key to close the New Patient Menu.
Note — The OK button is unavailable until all required patient
ID fields are complete.
A new row appears at the top of the Patient Records table. The
text in the new row is red.
When you scroll through the Patient Records table, the display splits into two
panes. The top pane displays the patient records stored on the monitor, and the
bottom pane displays all of the patient identifiers for the highlighted patient
record.
The View Records button in the lower right corner of the Patient Records screen
toggles between two modes: View Patient and View All. Use View Patient mode
View Records to view all records for a specific patient. Use View All mode to view all records
button in the Patient Records table.
Note — When you highlight the View Records button, the tool tip displays the
opposite mode: if you select View Patient, the label changes to View All; if you
select View All, the label changes to View Patient.
Step
2 Press the Up/Down key to highlight the scroll bar to the right of
the Patient Records table, and then press the Select key.
The scroll bar changes from gray to red to indicate that it is
active.
Note — If you take a new measurement while scrolling mode is active, the
monitor automatically scrolls to the top and exits scrolling mode.
Step
1 Press the Up/Down key to highlight the scroll bar to the right of
the Patient Records table, and then press the Select key.
The scroll bar changes from gray to red to indicate that it is
active.
2 Press the Up/Down key to highlight the desired patient ID, and
then press the Select key.
Notes
• If you are using a barcode scanner to enter patient IDs, you can
select an existing patient ID from the patient ID list by unplugging
the scanner and using the on-screen keyboard.
• To save a new patient ID, you must take an associated set of vital
signs measurements and save them. If you enter a new ID but do
not save measurements for the new ID, the ID is not saved to the
database.
Step
3 To change the sort order of the list, press the Up/Down key to
highlight the List Sorted By button, and then press the Select
key.
4 Press the Up/Down key to highlight the list, and then press the
Select key.
5 Press the Up/Down key to select a patient ID from the list, and
then press the Select key.
Notes
• After you enter a patient ID and then exit the New Patient Menu,
you can no longer change the Patient Type.
• After a record has been exported successfully, it can no longer be
edited. Records that are exported successfully appear green in the
Patient Records table. Records that are not exported successfully
change from white to blue, and a message indicating the reason the
record was rejected may appear on the screen.
The method for editing patient IDs depends on the method used to enter patient
IDs. You can edit a patient record by using one of the following methods:
Step
1 Press the Up/Down key to highlight the scroll bar, and then
press the Select key.
11 2 3 4 5 6 7 8 9 0 - =
q1 w e r t y u i o p [ ]
a1 s d f g h j k l ; ‘ \
z1 x c v b n m , . / Back
Shift OK Cancel
6 Press the Up/Down key to highlight the Yes button, and then
press the Select key to save the changes.
Note — If the barcode scanner cannot read the barcode or does not function
correctly, you can unplug the scanner and use the on-screen keyboard to edit
patient IDs. For more information, see “Editing a Patient Record Using the
On-screen Keyboard” on page 3-19.
Step
1 Press the Up/Down key to highlight the scroll bar, and then
press the Select key.
7 When you are done scanning the IDs, press the Up/Down key
to select the OK button, and then press the Select key.
Note — The OK button is unavailable until all required patient
ID fields are complete.
8 Press the Up/Down key to highlight the Yes button, and then
press the Select key to close the Edit Patient Menu and save
the changes.
Note — If the barcode scanner cannot read the barcode or does not function
correctly, you can unplug the scanner and use the on-screen keyboard to edit
patient IDs. For more information, see “Editing a Patient Record Using the
On-screen Keyboard” on page 3-19.
Step
1 Press the Up/Down key to highlight the scroll bar, and then
press the Select key.
4 Hold the scanner over the barcode, pull the trigger, and center
the beam on the barcode for the highlighted patient ID.
The barcode will fill in the corresponding ID fields on the
monitor.
6 When you are done scanning the IDs, press the Up/Down key
to highlight the OK button, and then press the Select key.
Note — The OK button is unavailable until all required patient
ID fields are complete.
7 Press the Up/Down key to highlight the Yes button, and then
press the Select key to close the Edit Patient Menu and save the
changes.
Note — If you delete all records for a specific patient, the patient ID for the
deleted patient is also removed from the Patient Records database.
The options that appear in the Delete Records Menu depend on which View
mode is active.
• If View Patient mode is enabled, you can delete a specific patient record
or all records for the selected patient.
• If View All mode is enabled, you can delete a specific patient record, all
records for a specific patient, or all records in the Patient Records
database.
Note — When you highlight the View Records button, the tool tip displays the
opposite mode: if you select View Patient, the label changes to View All;
likewise, if you select View All, the label changes to View Patient.
Step
Main Screen
key
2 Press the Up/Down key to highlight the scroll bar, and then press
the Select key.
3 Press the Up/Down key to highlight the desired patient ID, and
then press the Select key.
4 Press the Up/Down key to highlight the View Patient button, and
then press the Select key to display all saved records for the
selected patient.
8 Press the Up/Down key to highlight the Yes button, and then
press the Select key.
The patient record(s) are deleted.
Step
Main Screen
key
6 Press the Up/Down key to highlight the Yes button, and then
press the Select key.
All patient records are deleted.
Alarms alert you to conditions that need immediate attention. The VSi monitor
supports only technical alarms; it does not support physiological alarms.
Alarms on the VSi monitor are divided into two severity levels:
Visual Alarms
The SureSigns VSi monitor uses flashing numeric values and alarm messages to
indicate an alarm.
Alarms
SureSigns VSi Instructions for Use 4-1
Audible Alarms
Alarm Messages
Alarm messages appear in the message pane in the bottom left side of the screen.
Alarm messages use the same colors as the numeric panes described above. For
a complete list of alarm messages and descriptions of each message, see
Appendix A, “Alarm Specifications.”
High-priority alarm messages preempt lower priority alarm messages. After the
high-priority alarm has been resolved, the lower priority alarm message appears.
If multiple alarms of the same priority occur at the same time, the alarm
messages rotate every 1.5 seconds.
Audible Alarms
The alarm sound and interval depend on the alarm priority. High priority alarms
beep at a faster rate and sound different than low priority alarms.
• Alarm volume — You can increase or decrease the alarm volume. See
“Adjusting the Alarm Volume” on page 4-5.
• Silence alarms — You can pause alarms or silence them indefinitely. See
“Silencing Alarms” on page 4-6.
• Alarm tones — The SureSigns VSi offers two sets of alarm tones. Only
authorized personnel can change the alarm tone in the password-protected
System Admin Menu.
Alarms
4-2 SureSigns VSi Instructions for Use
Audible Alarms
To verify that the speaker is working at any time, perform the procedure in
“Testing Alarms” on page 4-9.
Warning Never pause an audible alarm or decrease the alarm volume if it may
compromise patient safety.
Do not rely exclusively on the audible alarm system for patient monitoring.
The most reliable method of patient monitoring requires correct operation
of the monitor and close observation of the patient.
For information about manually testing the speaker, see “Testing Alarms” on
page 4-9.
AUDIO FAILED
message
AUDIO FAILED
Patient/Type Date Time Pulse NBP SpO2 Temp
0009875433 A 06/21/11 03:20:12 65 SpO2 -- / -- (--) 98 36.8
0007363627 A 06/21/11 03:16:44 64 SpO2 -- / -- (--) 98 39.1
0006352516 A 06/21/11 03:12:41 72 SpO2 -- / -- (--) 99 36.9
0008763842 A 06/21/11 03:08:03 60 SpO2 -- / -- (--) 98 37.1
0006655321 A 06/21/11 03:00:43 80 SpO2 -- / -- (--) 99 36.6
0008764332 A 06/21/11 02:55:11 56 NBP 111/60 (82) 97 37.2
0008893376 A 06/21/11 02:50:06 77 SpO2 -- / -- (--) 95 37.0
0008873221 A 06/21/11 02:44:32 78 SpO2 -- / -- (--) 98 36.8
Speaker Malfunc
message 0003323876 P 06/21/11 02:40:51 75 SpO2 -- / -- (--) 97 36.9
Alarms
SureSigns VSi Instructions for Use 4-3
Audible Alarms
Warning Audible alarms do not occur when the AUDIO FAILED message is
displayed. If this message appears, do not use the monitor and contact your
system administrator for repair.
A latched alarm continues even after the condition that caused the alarm has
resolved itself. For example, if an NBP Overpressure alarm occurs and the NBP
cuff returns to normal, the alarm will continue to sound to notify you of the
event.
Alarms
4-4 SureSigns VSi Instructions for Use
Adjusting the Alarm Volume
Note — Your system administrator can set a minimum alarm volume, which
prevents you from setting the volume below the specified level.
Step
3 Press the Select key, and then press the Up/Down key to increase
or decrease the volume.
5 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Return
button, and then press the Select key to return to the Alarm
Settings Menu.
Alarms
SureSigns VSi Instructions for Use 4-5
Silencing Alarms
Silencing Alarms
You can silence alarms by pressing the Alarm Silence key on the front panel of
the monitor.
When an alarm occurs and you press the Alarm Silence key once, the monitor
Alarm
Silence key responds in one of the following ways:
• For SpO2 technical alarms (such as SpO2 No Sensor) and a Low Batt
alarm, the audible alarm is silenced and visual alarm indicators continue
to flash until the error condition is corrected.
• For NBP technical alarms (such as NBP Loose Cuff) and temperature
technical alarms (such as Temp Probe Error), the audible alarm is
silenced and visual alarm indicators are cleared.
• For an Extreme Low Batt alarm, the audible alarm is silenced and visual
alarm indicators continue to flash. The audible alarm resumes after
60 seconds.
For a description of the technical alarm messages, see Appendix A, “Alarm
Specifications.”
Alarms
4-6 SureSigns VSi Instructions for Use
Silencing Alarms
During Audio Pause mode, a white box with the AUDIO PAUSED message
appears at the top of the main screen and a timer shows the amount of time
remaining until the Audio Pause mode ends.
Audio Paused
message
122 98
SYS mmHg SpO2%
DIA
80 80
bpm SpO2%
98 37.1
MAP Oral
ºC
03:31:00
Alarms
SureSigns VSi Instructions for Use 4-7
Silencing Alarms
Note — If your system administrator has changed the Allow Audio Off setting
in the password-protected System Admin Menu, Audio Off mode is disabled.
If you press and hold the Alarm Silence key for two seconds, the monitor enters
Audio Off mode. Audible alarms are silenced, and a red box with the AUDIO
OFF message appears at the top of the main screen.
Audio Off
message
122 98
85 80
DIA bpm SpO2%
98 37.1
MAP ºC Oral
03:31:00
Alarms
4-8 SureSigns VSi Instructions for Use
Testing Alarms
Testing Alarms
To verify that the audible alarm system is working:
Step
1 Connect the NBP hose to the NBP input connector, but do not
place the cuff on your arm.
NBP key
• Your system administrator can change the alarm priority level for the
nurse call signal. For example, if the priority level is set to high in the
password-protected System Admin Menu, only high-priority alarms will
sound on the nurse call system.
Alarms
SureSigns VSi Instructions for Use 4-9
Alarms Safety Information
Caution The alarm volume should be loud enough to be heard within a room or
through an open door. Set the volume based on the environment and
ambient noise levels.
For visual alarms, the side-to-side viewing angle of the display is
approximately ±30 degrees relative to normal viewing.
The Alarms Off symbol in each measurement’s numeric pane indicates that
the SureSigns VSi monitor does not support physiological alarms.
Alarms
4-10 SureSigns VSi Instructions for Use
5
Monitoring SpO2
If an SpO2 value does not seem reasonable, use the perfusion indicator to assess
the signal quality. If the perfusion value is low (three bars or fewer) try adjusting
the sensor or using a different type of sensor.
Monitoring SpO2
SureSigns VSi Instructions for Use 5-1
Connecting SpO2 Cables
Caution Do not apply the blood pressure cuff to the same extremity as the one to
which an SpO2 sensor is attached because the cuff inflation disrupts SpO2
monitoring and leads to nuisance alarms.
Note — If the SpO2 measurement is delayed for more than 30 seconds (due to
an excessively noisy signal or because you are trying to measure NBP and SpO2
on the same limb), the SpO2 Extd Update alarm occurs and the SpO2 numeric
pane display alternates between the measurement value and a question mark
(-?-).
Monitoring SpO2
5-2 SureSigns VSi Instructions for Use
The SpO2 Numeric Pane
98
Perfusion
SpO2 %
indicator
SpO2 measurement
After 20 seconds of a valid SpO2 signal, the monitor saves the measurement to
the patient record, and the record closes. The measurement is cleared from the
numeric pane.
Step
Monitoring SpO2
SureSigns VSi Instructions for Use 5-3
SpO2 Safety Information
4 Press the Select key, and then press the Up/Down key to select one
of the following options:
• Slow — Use this setting when motion artifact is an issue. SpO2
changes are reported more slowly compared to the other
modes.
6 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main Screen
button, and then press the Select key.
Warning To minimize risk of damage to the monitor during defibrillation use only
approved supplies.
The SureSigns VSi monitor does not perform continuous SpO2 monitoring.
Monitoring SpO2
5-4 SureSigns VSi Instructions for Use
SpO2 Safety Information
Do not use disposable sensors on patients who have allergic reactions to the
adhesive.
Do not use the monitor or SpO2 sensors during magnetic resonance imaging
(MRI) scanning. Induced current could potentially cause burns. The
monitor may affect the MRI image, and the MRI unit may affect the
accuracy of the monitor’s measurements.
Caution Use only specified sensors and cables, otherwise patient injury can result.
Before using a sensor, verify that it is compatible with the monitor. For a
complete list of compatible accessories, see Chapter 10, “Accessories List.”
Do not leave the sensor on the patient once the measurement is complete;
remove the sensor from the patient.
Sensors are not sterile and should not be used in a sterile environment.
Do not apply the sensor too tightly as this results in venous pulsation. This
can severely obstruct circulation and lead to inaccurate measurements.
Follow the sensor’s instructions for use; adhere to all warnings and
cautions.
Check that the light emitter and the photo-detector are directly opposite
each other. All light from the emitter must pass through the patient’s tissue.
Remove colored nail polish from the application site.
Make sure that the sensor is the appropriate size. The sensor should not fall
off, nor should it be too tight.
When applying the M1193A or M1193T neonatal sensor, do not overtighten
the strap.
Monitoring SpO2
SureSigns VSi Instructions for Use 5-5
SpO2 Safety Information
When using the M1195A infant finger sensor, select a finger or toe with a
diameter of between 7 and 8 mm (0.27" and 0.31").
If a sensor is too loose, it might compromise the optical alignment or fall off.
If it is too tight — because the application site is too large or becomes too
large due to edema — the excessive pressure may cause venous congestion
distal, leading to interstitial edema, hypoxaemia and tissue malnutrition.
Do not use OxiCliq™ disposable sensors in a high humidity environment,
such as in neonatal incubators or in the presence of fluids, which can
contaminate sensor and electrical connections causing unreliable or
intermittent measurements.
For neonatal patients, place all sensor connectors and adapter cable
connectors outside the incubator. The humidity in the incubator can cause
inaccurate measurements.
Do not place the sensor on extremities with an arterial catheter or
intravascular venous infusion line.
Do not use more than one extension cable (M1941A). Do not use an
extension cable with Philips reusable sensors with part numbers ending in
-L (which indicates “Long” version). Do not use an extension cable with an
adapter cable.
To avoid electrical interference, position the sensor cable and connector
away from power cables.
To dispose of pulse oximeter equipment or components, follow local
regulations regarding disposal of hospital waste.
Monitoring SpO2
5-6 SureSigns VSi Instructions for Use
6
Monitoring NBP
The SureSigns VSi monitor measures Systolic, Diastolic, and Mean arterial
blood pressure by acquiring pressure pulses through a series of controlled
deflation steps of an inflated cuff.
Monitoring NBP
SureSigns VSi Instructions for Use 6-1
Selecting an NBP Cuff
A cuff that is too loose or too tight can cause inaccurate measurements. Also, if
the cuff is too loose, it may not deflate properly.
NBP input
connector
Monitoring NBP
6-2 SureSigns VSi Instructions for Use
NBP Numeric Panes
85
DIA
Diastolic value
98
MAP
Mean Arterial
Pressure (MAP) value
Monitoring NBP
SureSigns VSi Instructions for Use 6-3
Changing NBP Settings
Step
The factory default value for Initial Inflation Pressure is based on the patient
type:
Monitoring NBP
6-4 SureSigns VSi Instructions for Use
Changing NBP Settings
Step
1 Open the Blood Pressure Menu and press the Up/Down key to
highlight Initial Inflation Pressure.
2 Press the Select key, and then press the Up/Down key to select the
desired inflation pressure. The range of values is based on the
current patient type.
4 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main Screen
button, and then press the Select key.
If you start a new patient, change the patient type, or put the monitor in Standby
mode, the Initial Inflation Pressure setting reverts to the factory default value or
the default value set by your system administrator on the password-protected
System Admin Menu.
Monitoring NBP
SureSigns VSi Instructions for Use 6-5
Recalculating the NBP Value if the Limb is not at Heart Level
Step
1 Open the Blood Pressure Menu and press the Up/Down key to
highlight Blood Pressure Units.
2 Press the Select key, and then press the Up/Down key to select one
of the following options:
• mmHg
• kPa
4 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main Screen
button, and then press the Select key.
Higher than For each centimeter higher, add 0.75 mmHg (0.10 kPa)
heart level or
For each inch higher, add 1.9 mmHg (0.25 kPa)
Lower than For each centimeter lower, deduct 0.75 mmHg (0.10 kPa)
heart level or
For each inch lower, deduct 1.9 mmHg (0.25 kPa)
Monitoring NBP
6-6 SureSigns VSi Instructions for Use
Stopping an NBP Measurement
NBP key
Warning Continual NBP measurements can cause injury to the patient being
monitored. Weigh the advantages of frequent measurement against the risk
of injury.
In some cases, rapid, prolonged cycling of an NBP cuff has been associated
with any or all of the following: ischemia, purpura, or neuropathy. Apply
the cuff according to the directions and check the cuff site and cuffed
extremity regularly when blood pressure is measured at frequent intervals
or over extended periods of time.
Do not place the cuff on an extremity being used for intravenous infusion or
any area where circulation is compromised or has the potential to be
compromised.
Do not apply the blood pressure cuff to the same extremity as the one to
which an SpO2 sensor is attached because the cuff inflation disrupts SpO2
monitoring and leads to nuisance alarms.
Monitoring NBP
SureSigns VSi Instructions for Use 6-7
NBP Safety Information
Monitoring NBP
6-8 SureSigns VSi Instructions for Use
7
Monitoring Temperature
Monitoring Temperature
SureSigns VSi Instructions for Use 7-1
The Temperature Numeric Pane
37.1
oC Oral
Temperature
measurement
Monitoring Temperature
7-2 SureSigns VSi Instructions for Use
Taking a Temperature Measurement
Warning Use only approved, disposable probe covers. Probe covers are for single use
only. The use of unapproved probe covers or not using a probe cover can
result in the following:
• Patient cross-contamination
When you take rectal temperature measurements, insert the probe slowly
and carefully to avoid tissue damage.
Step
Monitoring Temperature
SureSigns VSi Instructions for Use 7-3
Changing Temperature Settings
Note — When the monitor switches from Standby mode to On mode, the
temperature module takes several seconds to reset. Do not remove the
temperature probe from the well while the module resets.
Step
Monitoring Temperature
7-4 SureSigns VSi Instructions for Use
Changing Temperature Settings
Caution Do not exceed the recommended measurement periods of three minutes for
oral and rectal measurements and five minutes for axillary measurements.
Monitoring Temperature
SureSigns VSi Instructions for Use 7-5
Changing Temperature Settings
Step
2 Press the Select key, and then press the Up/Down key to select
one of the following options:
• Monitored
• Predictive
4 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key until the Main Screen
button is selected and press the Select key.
Monitoring Temperature
7-6 SureSigns VSi Instructions for Use
Changing Temperature Settings
Note — The currently selected probe site appears in the Temperature numeric
pane when you remove the probe from the probe well.
Step
2 Press the Select key, and then press the Up/Down key to select a
measurement site. The choices are:
• Blue Probe Site: Oral or Axillary
• Red Probe Site: Rectal
4 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main
Screen button, and then press the Select key.
Monitoring Temperature
SureSigns VSi Instructions for Use 7-7
Verifying the Temperature Accuracy
Step
2 Press the Select key, and then press the Up/Down key to select
one of the following options:
• oC (Celsius)
• oF
(Fahrenheit)
4 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main
Screen button, and then press the Select key.
Monitoring Temperature
7-8 SureSigns VSi Instructions for Use
Verifying the Temperature Accuracy
Step
3 Press the Select key and then press the Up/Down key to select
Monitored.
4 Press the Select key to save the selected option, and then press
the Main Screen key
The Monitored Mode icon appears in the
Temperature numeric pane when you remove the
Monitored Mode probe from the well.
icon
5 Remove the temperature probe from the probe well and firmly
push the probe into a probe cover.
Monitoring Temperature
SureSigns VSi Instructions for Use 7-9
Temperature Safety Information
Warning To minimize risk of damage to the monitor during defibrillation, use only
approved supplies.
Do not use the thermometer if you see any signs of damage to the probe.
Caution When you replace the temperature probe after a measurement is complete,
verify that the probe is firmly seated in the well, but do not forcefully insert
the probe into the well.
Biting the probe may cause damage to the probe.
Do not take an axillary temperature through a patient’s clothing. Direct
probe cover-to-skin contact is required.
Monitoring Temperature
7-10 SureSigns VSi Instructions for Use
8
Monitoring Pulse Rate
The SureSigns VSi monitor calculates and displays a pulse rate value, which can
be derived from either an SpO2 or NBP measurement.
Warning The SureSigns VSi monitor does not perform continuous Pulse
measurements.
Units of measurement
85
bpm SpO2 Pulse Rate Source
(SpO2 in this example)
Note — If the pulse rate source is SpO2, the pulse rate measurement is cleared
from the numeric pane when the patient record is closed.
If you select NBP as the pulse rate source, note the following:
Step
1 Press the Up/Down key to highlight the Pulse Rate numeric pane.
4 Press the Select key, and then press the Up/Down key to select
one of the following options:
• SpO2 — The monitor derives the pulse rate value from
SpO2.
• NBP — The monitor derives the pulse rate value from NBP.
• Auto — The monitor searches for an available pulse rate
source: SpO2 first, then NBP.
6 Press the Main Screen key on the front panel to close the menu.
Alternative: Press the Up/Down key to highlight the Main
Screen button, and then press the Select key.
Notes
• If the pulse rate source is NBP, after a measurement completes, the
value is cleared from the monitor screen as soon as the record is
closed or if the value is older than three minutes, whichever comes
first.
• If your patient has a very low pulse rate or strong arrhythmia, pulse
rate readings derived from SpO2 may cause nuisance alarms. Use
the perfusion indicator to assess signal quality. If necessary, use a
different method to measure the patient’s pulse rate.
To clean or disinfect your SureSigns VSi monitor, use only the approved
cleaning agents listed in this chapter.
For information about how to clean accessories, see the instructions for use
provided with the accessory.
General Guidelines
Keep the monitor, cables, and accessories free of dust and dirt. After cleaning
and disinfecting, check the equipment carefully. Do not use the equipment if you
see signs of deterioration or damage.
If you need to return any equipment to Philips, clean and disinfect it first.
Caution If you spill liquid on the exterior of the monitor, use a clean cloth to dry it.
If you believe the liquid may have entered the monitor, contact your
biomedical engineer, who can verify the performance and safety of the
equipment.
Step
2 Wring any excess moisture from the cloth and gently clean the
monitor.
Step
2 Wring any excess moisture from the cloth and wipe the monitor to
disinfect it.
Note — If your monitor has an attached SureSigns VSi and VS2+ Tabletop
Base, use the same solutions to clean the base and the monitor.
Caution Do not use alcohol to clean the cables. Alcohol can cause the cables to
become brittle.
Step
2 Wring any excess moisture from the cloth and gently clean the
cables.
3 Clean the cables again with a damp cloth moistened with water only.
Step
2 Wring any excess moisture from the cloth and gently clean the
cables.
3 Clean the cables again with a damp cloth moistened with water
only.
Caution When cleaning the temperature module, follow these general precautions:
• Do not use steam, heat, or gas sterilization on the probe or probe well.
• Dampen a soft cloth with mild soap and warm water and wipe the probe
and cord.
To disinfect the probe and cord:
• Dampen a soft cloth with any one of the following solutions and wipe the
probe and cord:
– Isopropyl alcohol (70% solution in water)
– Sodium hypochlorite (chlorine bleach, 10% solution in water)
– A nonstaining disinfectant
Step
3 Dampen a soft cloth with mild soap and warm water and wipe the
inner and outer surfaces.
5 Thoroughly dry all of the surfaces before replacing the well in the
monitor.
This chapter lists accessories that are compatible with the SureSigns VSi vital
signs monitor.
Note — The accessory list is subject to change. For the latest information about
supported accessories, contact your authorized Philips representative or refer to
www.medical.philips.com.
Accessories List
SureSigns VSi Instructions for Use 10-1
SpO2 Accessories
SpO2 Accessories
Accessories List
10-2 SureSigns VSi Instructions for Use
SpO2 Accessories
Adult/ Finger sensor, for patient size > 44 lb (20 kg) M1131A Adapter Cable
Pediatric M1943A (1.1 m)
or
Infant Digit sensor for patient size 7 lb – 22 lb M1132A M1943AL (3 m)
(3 kg – 10 kg)
Accessories List
SureSigns VSi Instructions for Use 10-3
SpO2 Accessories
Note — The Nellcor disposable sensors listed below are only available from
Philips in Europe and Japan.
Neonate/ Sensor for neonatal foot or adult digit, for M1901B Adapter Cable
Adult patient size < 7 lb or > 88 lb (< 3 kg or > 40 kg) M1943A (1.1 m)
or
Infant Digit sensor for patient size 7 lb – 44 lb M1902B M1943AL (3 m)
(3 kg – 20 kg)
Adult Digit sensor for patient size > 66 lb (30 kg) M1904B
Accessories List
10-4 SureSigns VSi Instructions for Use
NBP Accessories
Cable
Description Part Number
Length
Masimo LNOP Adapter Cable (adapts LNOP sensors to Philips 3.6 m M1020-61100
monitors)
Masimo LNCS Adapter Cable (adapts LNCS sensors to Philips 3.0 m 989803148221
monitors)
NBP Accessories
Accessories List
SureSigns VSi Instructions for Use 10-5
NBP Accessories
Infant 10 cm – 19 cm 6 cm 40401A
Accessories List
10-6 SureSigns VSi Instructions for Use
NBP Accessories
Accessories List
SureSigns VSi Instructions for Use 10-7
NBP Accessories
1. The luer connector cuffs and air hoses are not available in EEA (European Economic Area) countries.
2. The safety connector cuffs and air hoses may not be available in all countries. Check with your local
sales organization.
Accessories List
10-8 SureSigns VSi Instructions for Use
Temperature Accessories
Temperature Accessories
Patient
Description Part Number
Category
Accessories List
SureSigns VSi Instructions for Use 10-9
Miscellaneous Accessories
Miscellaneous Accessories
1D Barcode Scanner (includes mounting arm for use with roll stand) 989803167691
2D Barcode Scanner (includes mounting arm for use with roll stand) 989803147821
Accessories List
10-10 SureSigns VSi Instructions for Use
11
Specifications
General Specifications
Parameter Specification
Size
Display
Refresh Frequency 60 Hz
Specifications
SureSigns VSi Instructions for Use 11-1
Safety Standards
Safety Standards
Parameter Specification
Protection Against Hazards Equipment is not suitable for use in the presence
of Ignition of Flammable of a flammable anaesthetic mixture with air or
Anaesthetic Mixtures oxygen or nitrous oxide, per IEC 60601-1
Electrical Specifications
Parameter Specification
Power Sources
Battery Operating Time > 4.5 hours, taking 20 NBP and SpO2
(new, fully-charged measurements every 4 hours
battery)
Specifications
11-2 SureSigns VSi Instructions for Use
Environmental Specifications
Parameter Specification
Frequency 50/60 Hz
Environmental Specifications
Caution The monitor might not meet performance specifications if stored or used
outside the specified temperature and humidity ranges.
Parameter Specification
Specifications
SureSigns VSi Instructions for Use 11-3
Environmental Specifications
Parameter Specification
Thermal
Humidity
Specifications
11-4 SureSigns VSi Instructions for Use
NBP Specifications
NBP Specifications
In neonatal mode, the blood pressure measurements determined with this device
comply with the American National Standard for Electronic or Automated
Sphygmomanometers (ANSI/AAMI SP10-1992) in relation to mean error and
standard deviation, when compared to intra-arterial or auscultatory
measurements in a representative patient population. The NBP measurement is
suitable for use in the presence of electrosurgery and during the discharge of a
cardiac defibrillator according to IEC 601-2-30:1999/EN 60601-2-30:2000.
Specifications
SureSigns VSi Instructions for Use 11-5
NBP Specifications
The following table lists the specifications for the Philips NBP module:
Parameter Specification
Specifications
11-6 SureSigns VSi Instructions for Use
Temperature Specifications
Parameter Specification
Temperature Specifications
Parameter Specification
Specifications
SureSigns VSi Instructions for Use 11-7
SpO2 Specifications
SpO2 Specifications
The update rate for the SpO2 value and pulse rate is typically 1 second. Data
averaging and other signal processing on the displayed and transmitted data
values of SpO2 and pulse rate is controllable by the user-selectable SpO2
Response Mode: Slow (20 seconds), Normal (10 seconds), and Fast (5 seconds).
Depending on the magnitude of difference between the alarm limit and the
displayed value, the alarm signal generation delay may be from 1 second to the
value of the response time (5, 10, or 20 seconds).
Parameter Specification
Specifications
11-8 SureSigns VSi Instructions for Use
SpO2 Specifications
Parameter Specification
Specifications
SureSigns VSi Instructions for Use 11-9
Interface Specifications
Interface Specifications
Parameter Specification
Isolation 1.5 kV
Specifications
11-10 SureSigns VSi Instructions for Use
Radio Regulatory Compliance
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one of the following
measures:
Caution FCC Caution: Any changes or modifications not expressly approved by the
party responsible for compliance could void the user’s authority to operate
this equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference
that may cause undesired operation.
Specifications
SureSigns VSi Instructions for Use 11-11
Radio Regulatory Compliance
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and operated with
minimum distance 20 cm between the monitor and any person.
Industry Canada
To prevent radio interference to the licensed service, this device is intended to
be operated indoors and away from windows to provide maximum shielding.
Equipment (or its transmit antenna) that is installed outdoors is subject to
licensing.
For the specifications for antennas that are used with the monitor, see “Interface
Specifications” on page 11-10.
European Union
This device complies with the essential requirements of the R&TTE
Directive 1999/5/EC. The following test methods have been applied in order to
prove presumption of conformity with the essential requirements of the R&TTE
Directive 1999/5/EC:
• EN60950-1:2001 A11:2004
Safety of Information Technology Equipment
• EN 300 328 V1.7.1: (2006-10)
Electromagnetic compatibility and Radio spectrum Matters (ERM);
Wideband Transmission systems; Data transmission equipment operating
in the 2,4 GHz ISM band and using spread spectrum modulation
techniques; Harmonized EN covering essential requirements under article
3.2 of the R&TTE Directive
• EN 301 489-1 V1.6.1: (2005-09)
Electromagnetic compatibility and Radio Spectrum Matters (ERM);
ElectroMagnetic Compatibility (EMC) standard for radio equipment and
services; Part 1: Common technical requirements
Specifications
11-12 SureSigns VSi Instructions for Use
Radio Regulatory Compliance
In Italy the end-user should apply for a license at the national spectrum
authorities in order to obtain authorization to use the device for setting up
outdoor radio links and/or for supplying public access to telecommunications
and/or network services.
This device may not be used for setting up outdoor radio links in France and in
some areas the RF output power may be limited to 10 mW EIRP in the
frequency range of 2454 MHz to 2483.5 MHz. For detailed information the
end-user should contact the national spectrum authority in France.
Hereby, Philips declares that this monitor is in compliance with the essential
requirements and other relevent provisions of Directive 1999/5/EC.
Specifications
SureSigns VSi Instructions for Use 11-13
Radio Regulatory Compliance
Specifications
11-14 SureSigns VSi Instructions for Use
12
Troubleshooting
Use the information in this chapter to diagnose and correct problems with your
monitor. If you cannot resolve a problem after using the information in this
chapter, contact your system administrator or biomedical engineer.
Diagnosing a Problem
Before you begin to troubleshoot a problem, check the following basics:
Troubleshooting
SureSigns VSi Instructions for Use 12-1
Troubleshooting Table
Note — It may take several seconds for the AC Power LED to light/turn off
after the power cord is connected/disconnected.
When the monitor has power and a functioning display, use the information in
this chapter to diagnose other monitor problems.
Troubleshooting Table
Use the following table to diagnose and fix monitor problems. The table
describes a monitor problem by symptom, lists possible causes and suggested
actions. If you cannot resolve a problem by using the following table, or if the
table does not list the problem you are experiencing, see your system
administrator.
Troubleshooting
12-2 SureSigns VSi Instructions for Use
Troubleshooting Table
Troubleshooting
SureSigns VSi Instructions for Use 12-3
Troubleshooting Table
Troubleshooting
12-4 SureSigns VSi Instructions for Use
Alarms and Error Messages
Message Action
NBP Equip Malfunc - Not Calibrated See your system administrator.
SpO2 Error. The front end will now reset Allow the monitor to reset. If the problem persists,
itself. see your system administrator.
Temp Error. The front end will now reset Allow the monitor to reset.
itself.
Verify that the Temperature module is positioned
correctly and secured to the side of the monitor.
If the problem persists, see your system
administrator.
System Error See your system administrator.
Troubleshooting
SureSigns VSi Instructions for Use 12-5
Alarms and Error Messages
Troubleshooting
12-6 SureSigns VSi Instructions for Use
A
Alarm Specifications
Technical Alarms
The following table contains an alphabetical listing of technical alarm messages.
Alarm Specifications
SureSigns VSi Instructions for Use A-1
Technical Alarms
Alarm Specifications
A-2 SureSigns VSi Instructions for Use
Technical Alarms
Alarm Specifications
SureSigns VSi Instructions for Use A-3
Technical Alarms
Alarm Specifications
A-4 SureSigns VSi Instructions for Use
B
Electromagnetic Compatibility
This appendix lists the tests and compliance levels that make the SureSigns VSi
vital signs monitor suitable for use in the specified electromagnetic environment
according to IEC 60601-1-2:2001.
Radio frequency (RF) interference from nearby transmitting devices can degrade
performance of the product. Electromagnetic compatibility with surrounding
devices should be assessed prior to using the product.
Fixed, portable, and mobile radio frequency communications equipment can also
affect the performance of medical equipment. See your service provider for
assistance with the minimum recommended separation distance between RF
communications equipment and the product.
Electromagnetic Compatibility
SureSigns VSi Instructions for Use B-1
Reducing Electromagnetic Interference
Warning Use of accessories, transducers, and cables other than those specified may
result in increased emissions and/or decreased immunity of the SureSigns
VSi vital signs monitor.
The SureSigns VSi vital signs monitor should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is necessary, the
monitor should be observed to verify normal operation in the configuration
in which it is used.
Electromagnetic Compatibility
B-2 SureSigns VSi Instructions for Use
Emissions and Immunity
Electromagnetic Compatibility
SureSigns VSi Instructions for Use B-3
Guidance and Manufacturer’s Declaration
RF emissions Class A
CISPR 11
Harmonic emissions Class A The SureSigns VSi vital signs monitor is suitable for
IEC 61000-3-2 use in all establishments, except domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Table B-2. Electromagnetic Immunity (ESD, EFT, Surge, Dips and Magnetic Field)
Electromagnetic Compatibility
B-4 SureSigns VSi Instructions for Use
Guidance and Manufacturer’s Declaration
Table B-2. Electromagnetic Immunity (ESD, EFT, Surge, Dips and Magnetic Field)
70% UT
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
< 5% UT
(> 95% dip in UT)
for 5 sec
Electromagnetic Compatibility
SureSigns VSi Instructions for Use B-5
Guidance and Manufacturer’s Declaration
Electromagnetic Compatibility
B-6 SureSigns VSi Instructions for Use
Guidance and Manufacturer’s Declaration
7
d = --- P ; 800 MHz – 2500 MHz
3
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the SureSigns VSi vital signs monitor
is used exceeds the applicable RF compliance level above, the SureSigns VSi vital signs monitor should be
observed to verify normal operation. If abnormal performance is observed, additional measures are necessary,
such as re-orienting or relocating the SureSigns VSi vital signs monitor.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Respiration
measurement may be subject to interference at 900 kHz to 1100 kHz and 70 MHz to 80 MHz at less than 3 V/M
field strength.
Electromagnetic Compatibility
SureSigns VSi Instructions for Use B-7
Guidance and Manufacturer’s Declaration
Frequency of 150 kHz – 80 MHz 80 MHz – 800 MHz 800 MHz – 2.4 GHz
Transmitter
For transmitters rated at a maximum output power not listed above, the separation distance d
can be estimated, in meters, using the equation in the corresponding column, where P is the
maximum output power rating of the transmitter in watts according to the transmitter’s
manufacturer.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Electromagnetic Compatibility
B-8 SureSigns VSi Instructions for Use
Index
A Blood Pressure Menu, 6-4
AC power LED, 2-5
AC power, troubleshooting, 12-2 C
accessories charging the battery
miscellaneous, 10-10 Charging LED, 2-6
NBP, 10-5 charging status, 2-6
SpO2, 10-2 overview, 2-5
temperature, 10-9 cleaning
alarms approved cleaning agents, 9-2
and nurse call system, 4-9 cables, 9-4
audible, 4-2 guidelines, 9-1
audio failure, 4-3 monitor, 9-2
colors, 4-1 tabletop base, 9-3
latched and non-latched, 4-4 temperature probe and cord, 9-5
pausing, 4-6 temperature probe well, 9-5
silencing, 4-6 configuration, viewing, 2-18
speaker malfunction, 4-3 connector
technical, A-1 NBP, 6-2
testing, 4-9 SpO2, 5-2
troubleshooting, 12-2 temperature input, 7-1
visual, 4-1 connectors, rear panel, 2-3
volume, adjusting, 4-5 cuff, NBP, selecting, 6-2
Alarms Off symbol, 4-2
audible alarms, 4-2 D
settings, 4-2
troubleshooting, 12-2 Date / Time menu, 2-9
Audio Off mode, 4-8 date format, changing, 2-16
Audio Pause mode, 4-7 date, changing, 2-9
Auto Suspend setting, 2-8 Deep Sleep mode, 2-8
deleting patient records
all, 3-25
B for selected patient, 3-24
barcodes overview, 3-23
scanning one at a time, 3-7 diagnosing problems, 12-1
using programmed scanner, 3-10 disinfecting
battery cables, 9-4
disposal, 2-7 monitor, 9-2
status icon, Standby mode, 2-7 display mode
status indicators, 2-6 changing, 2-14
status pane, 2-6 Patient Records, 2-14
Index-1
SureSigns VSi Instructions for Use
Vital Signs, 2-12 display, Patient Records, 2-14
display, Vital Signs, 2-12
E Monitored temperature, 7-5
On/Standby, 2-7
editing patient records, 3-18
Predictive temperature, 7-5
scanning individual barcodes, 3-21
SpO2 response, 5-3
using on-screen keyboard, 3-19
monitor
using programmed barcode scanner, 3-22
configurations, 1-3
electrical specifications, 11-2
mounting, 2-10
electromagnetic compatibility, B-1 name, 2-16
environmental specifications, 11-3
overview, 1-1
error messages, 12-5
system information, 2-18
Monitored mode, 7-5
F mounting options, 2-10
front panel, 2-1
N
H NBP
hiding the system time, 2-16 accessories, 10-5
changing units of measurement, 6-6
I configuring inflation pressure, 6-4
measurement limitations, 6-1
initial inflation pressure, 6-4 oscillometric measurement, 11-5
Intended Use statement, 1-3 panes, 6-3
recalculating if limb not at heart level, 6-6
K safety information, 6-7
keyboard, on-screen, 3-5 settings, 6-4
keys, front panel, 2-1 specifications, 11-5
stopping a measurement, 6-7
L troubleshooting, 12-3
NBP cuff
latched alarms, 4-4 connecting, 6-2
LED reasons for deflation, 6-5
AC power, 2-5 selecting, 6-2
network connection status, 2-20
M networked monitors
main screen display, 2-11 connection status, 2-20
measurement limitations, NBP, 6-1 synchronizing date and time, 2-21
mode non-latched alarms, 4-4
Audio Off, 4-8 nurse call system
Audio Pause, 4-7 alarms, 4-9
Deep Sleep, 2-8 troubleshooting, 12-4
display, changing, 2-14
Index-2
SureSigns VSi Instructions for Use
O setting up the monitor, 2-4
silencing alarms, 4-6
On/Standby mode, 2-7
speaker malfunction, 4-3
on-screen keyboard, 3-5
specifications
electrical, 11-2
P environmental, 11-3
patient ID general, 11-1
scanning barcodes one at a time, 3-7 interface, 11-10
selecting existing, 3-16 NBP, 11-5
using on-screen keyboard, 3-5 safety, 11-2
using programmed barcode scanner, 3-10 SpO2, 11-8
patient records temperature, 11-3, 11-7
deleting all, 3-25 SpO2
deleting for selected patient, 3-24 accessories, 10-2
editing, 3-18 cables, connecting, 5-2
saving to USB flash drive, 2-17 pane, 5-3
using the View Records button, 3-14 readings, troubleshooting, 12-4
viewing, 3-14 response mode, 5-3
Patient Records display mode, 2-14 safety information, 5-4
Patient Records table, 2-14, 3-1 specifications, 11-8
patient type SpO2 Menu, 5-3
changing, 3-5, 3-8, 3-11, 3-12 SpO2 sensors
setting default, 2-16 selecting, 5-1
power, troubleshooting, 12-2 types of, 5-1
powering up the monitor, 2-4 synchronizing date and time, 2-21
Predictive mode, 7-5 system date, changing, 2-9
pulse rate system errors, 12-5
changing the source, 8-2 System Information window, 2-18
pane, 8-1 System Menu, 2-15
settings, 8-1 system settings, changing, 2-15
Pulse Rate source, 8-2 system time
changing, 2-9
R hiding, 2-16
radio regulatory compliance, 11-11
rear panel, connectors, 2-3 T
records, See patient records tabletop base, cleaning, 9-3
technical alarms, A-1
S temperature
accessories, 10-9
safety standards, 11-2
changing the probe site, 7-7
saving measurements without patient ID, 3-12
changing units of measurement, 7-8
saving patient records to USB flash drive, 2-17
measuring, 7-3
selecting existing patient ID, 3-16
pane, 7-2
Index-3
SureSigns VSi Instructions for Use
safety information, 7-10
settings, 7-4
specifications, 11-7
verifying accuracy, 7-8
temperature modes
Monitored mode, 7-5
Predictive mode, 7-5
temperature probe and cord, cleaning, 9-5
temperature probe well, cleaning, 9-5
temperature probe, connecting, 7-1
temperature readings, troubleshooting, 12-4
testing alarms, 4-9
time synchronization, 2-21
time, system
changing, 2-9
hiding, 2-16
tool tips, 2-15
troubleshooting
AC power, 12-2
audible alarms, 12-2
diagnosing problems, 12-1
NBP, 12-3
nurse call system, 12-4
SpO2 readings, 12-4
temperature readings, 12-4
U
units of measurement
NBP, 6-6
temperature, 7-8
V
visual alarms, 4-1
Vital Signs display mode, 2-12
volume, adjusting alarm, 4-5
Index-4
SureSigns VSi Instructions for Use