Technical specification Cherish your life, Cherish your health!

Ventilation mode Alarm and protection


PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, IPPV, A/C, PCV, PSV, AC power failure alarm Power failure or no connection
SPONT/CPAP, SIGH, MANUAL Internal backup battery low voltage alarm ≤ 11.3 ± 0.3 V
Ventilator parameter No tidal volume No tidal volume within 6 s
Tidal volume (Vt) 0 ~ 2000 mL High minute volume alarm 5 L/min ~ 99 L/min
Frequency (Freq) 1 min ~ 100 min Low minute volume alarm 1 L/min ~ 30 L/min

S1100 ICU Ventilator


Oxygen concentration 21 % ~ 100 % High airway pressure alarm 20 cmH2O ~ 100 cmH2O
I:E 4:1~1:8 Low airway pressure alarm 0 cmH2O ~ 20 cmH2O
PEEP 0cmH2O ~ 40 cmH2O High oxygen concentration alarm 19 % ~ 100 %
Pressure triggering sensitivity (Ptr) -20 cmH2O ~ 20 cmH2O (Based on PEEP) Low oxygen concentration alarm 18 % ~ 99 %
Flow trigger sensitivity (Ftr)
Pressure control (PC)
0.5 L/min ~ 30 L/min
5 cmH2O ~ 80 cmH2O
Continuous pressure alarm
Suffocation warning
(PEEP + 1.5 cmH2O) over 16 s
5 ~ 60 s ADULT · PEDIATRIC
Pressure support (PS) 0 cmH2O ~ 80 cmH2O Fan error Show on screen
SIGH
Apnea ventilation
0 (off) 1/100 ~ 5/100
OFF, 5 s ~ 60 s
Oxygen deficit
The maximum limited pressure
Show on screen
< 12.5 kPa
Friendly Powerful Reliable
Pressure limit 20 cmH2O ~ 100 cmH2O
Working condition
Monitoring parameter Gas source O2, Air
Frequency(Freq) 0 / min ~ 100 / min Pressure 280 kPa - 600 kPa
Tidal volume(Vt) 0 mL ~ 2500 mL Voltage 220 V ± 22 V
MV 0 L/min ~ 99 L/min Power frequency 50 Hz ± 1 Hz
Airway pressure 0 cmH2O ~ 100 cmH2O Input power 900 VA (With air compressor)
Dynamic lung compliance testing 1 mL/cmH2O ~ 1000 mL /cmH2O 250 VA (Without air compressor)
Oxygen concentration 15 % ~ 100 %
Oscillogram
P-T(Pressure-Time)
Packing size F-T(Flow-Time)
Main components: L 560 * W 560 * H 605 mm V-T(Volume-Time)
G.W. : 40 KG, N.W. : 17 KG P-V Loop(Pressure-Volume Loop)
Air compressor: L 670* W 700 * H 1160 mm F-V Loop(Flow-Volume Loop)
G.W. : 84 KG, N.W. : 46.2 KG P-F Loop(Pressure-Flow Loop)

Other models for your reference:

S1100B S1100 S1600

The picture is for reference only. For more information, please contact Superstar Medical sales representatives.

Nanjing Superstar Medical Equipment Co., Ltd.


Address: No.6 Building, No.9 Bofu Road, Yanjiang Industrial Development Zone, Liuhe District, 211505 Nanjing, P. R. China.
Tel: +86-25-69658833 Mobile: +86-15951815673 Fax: +86-25-69658833
Email: info@superstar-med.com Web: http://www.superstar-med.net
14 types of alarm and protection function

15’’ TFT LCD screen display


screen saver mode

3 of 5 oscillographs can be displayed


on the screen at the same time

S1100 ICU Ventilator 2 minutes fast oxygen supply

Application
The ventilator makes a good performance in operation room,ICU department and emergency treatment.It used to Parameters monitoring
Parameters setting
assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg. 25 years experience in
Market-oriented ventilator make us professional and reliable, satisfying all your needs in ventilation. Due to the flexible
configuration, good quality and competitive price, S1100A has soon become the superstar of market. Nebulization function

Features
easily.
15” TFT touch screen displays the ventilation parameters, alarm information, and oscillograms, make every operation more easil y.
Multiple ventilation modes,suitable for ICU,emergency department and operation room etc: ( IPPV, A/C, PCV, SIMV,PSV,
SPONT/CPAP,SIGH,MANUAL) Oxygen concentration sensor
5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time.
Humidifier can heat and wet breathing gas, makes it comfortable for patient to breathe.
Rapid oxygen supply,automatically offer high flow rate oxygen within two minutes
Nebulization make medicine into small liquid and particle,easier and quickier for patient to breathe in.
High temperature resistance breathing circuit is reusable and anti-pollution.

Safety Humidifier
Heating and wetting, real-time
14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting. temperature display
Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration.
Easy to move with four casters, easy to stop with two brakes.
Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
Display continuous working
Compact long life internal battery can provide emergency power,avoid risk of patient.
hours for the air compressor
Self-check before operation,eliminate system mistake.

Friendly Powerful Reliable


S1100 ICU Ventilator --Nanjing Superstar Medical Equipment Co.,Ltd

Nanjing Superstar Medical Equipment Co.,Ltd


E-mail: info@superstar-med.com/diana001@superstar-med.com
Tel: +86-25-69658833
Mobile/Wechat/Whatsapp : +86-159-5181-5673
Add:No.6 Building, No.9 Bofu Road, Yanjiang Idustrial Development Zone, Liuhe District, 211505 Nanjing, P. R.
China
We are manufacturer, so we can give you more!
19.7 modo de Ventilación
——IPPV
——A/C P-A/C V-A/C
——V-SIMV
——P-SIMV
——PSV
——PCV
——SPONT/CPAP
——PRVC
——APRV
——DuoLevel
——SUSPIRO
——MANUAL
19.8 parámetro de Ventilación ajustamiento
 Frecuencia
—— Rango de ajuste 1 /min ~ 100 /min
(Bajo SIMV:1 /min~ 40 /min
Todo modo excepto SIMV:VTH 4 /min~ 40 /min
VTL 20/min~ 100 /min )
—— Error permitido ± 1 / min o ± (valor de ajuste del 10%), el que sea mayor
 Tiempo de inhalación(I:E)(Tinsp)
——Rango de ajuste 0s~ 12s (excepto SIMV modo, I:E 4:1~ 1:8)
——Error permitido ± 15 %
 Volumen corriente (VT)
——Rango de ajuste 0 mL ~ 1500 mL
Y :.
VTH 250 mL ~ 1500 mL
VTL 0. 20 mL~ 300 mL

——Rango de ajuste ± (10 mL + 10% valor de encuadre)


 Ventilación de minuto(MV)
——Max MV
VTH ≥18 L/min
VTL ≥10 L/min
 PEEP
——Rango de ajuste 0 cmH2O ~ 20 cmH2O
——Error permitido ± (2 cmH2O + 5% valor de encuadre)

 CPAP
——Ra 0 cmH2O ~ 20 cmH2O
——Error permitido ± (2 cmH2O + 5% valor de encuadre)

 Flujo continuo
——Rango de ajuste VTH 7 L/min ~ 60 L/min
VTL 2 L/min ~ 30 L/min
——Error permitido >3L/min, ±15%
≤3L/min, ±0.5 L/min
 Trigger inspiratorio
Trigger de presión(PTr)
——Rango de ajuste -20 cmH2O ~ 0 cmH2O(Bajo PEEP)
——Error permitido ±(1 cmH2O +10 % encuadre value)
Trigger de flujo
——Rango de ajuste 0.5L/min ~ 30 L/min
——Error permitido >3L/min, ±15%
≤3L/min, ±0.5 L/min
 control de Presión(Pc)
——Rango de ajuste 5 cmH2O ~ 60 cmH2O
——Error permitido ± (2 cmH2O + 5% valor de encuadre)
 Soporte de presión(Ps)
——Rango de ajuste 0 cmH2O ~ 60 cmH2O
——Error permitido ± (2 cmH2O + 5% valor de encuadre)
 Concentración de O2
——Rango de ajuste 21 %~100 %
——Error permitido ± 3% (v / v) o ± 10% del valor de configuración, el que sea mayor
—— Tiempo de respuesta del VENTILADOR de una fracción de volumen del 21% al 90%
Vt 500 mL 2s
Vt 150 mL 2.5 s
Vt 30 mL 3s
 Tiempo de mantenimiento (Plataforma de inhalación) (CONSEJO)
——Rango de ajuste 0 ~ 6s(0%~50% inhalación tiempo)
——Error permitido ± 0.1s o ± (valor de ajuste del 10%), el que sea mayor
 Ventilación de apnea
——Rango de ajuste FUERA,5 s ~ 60 s
 Caudal inspiratorio máximo
——Máximo Inspiratory Índice de Flujo:≥ 60 L/min
 Presión máxima de trabajo establecida
(rango límite de presión) 20 cmH2O ~ 100 cmH2O
 Presión máxima ≤125 cmH2O
 Producción MV bajo ventilación manual ≥25L/min
 Nebulizador presión presión de salida máxima≤0.2 MPa
flujo de salida máximo ≥8 L / min

explicación:

1)establezca el valor límite de presión para producir la presión máxima de trabajo (el límite superior de la alarma de alta
presión de la vía aérea).

2)No use presión negativa en la exhalación.


3)la presión límite máxima es la presión de la válvula de seguridad
19.9 Parámetro de ventilación Rango de monitoreo
 Los siguientes son parámetros de monitoreo bajo el ambiente de temperatura corporal y presión saturada.
 Fig display:

Parámetro Gama Resolviendo poder Exactitud

± 1 / min o ± (10% de lectura


Frecuencia(Freq) 0 /min ~ 100 /min 1 /min
real), lo que sea mayor

> 50 ml, ± (4.0 ml + 15%


Tidal Volumen
0 mL~2000mL 10 mL de lectura real)
(VT)
≤50 ml, ± 20 ml
> 3L / min, ± ( 15% de
Ventilación de
0 L/min~99 L/min 0.1 L/min lectura real)
minuto(MV)
≤3L / min, ± 0.5 L / min
± (2% escala llena +
(Pcumbre) 0 cmH2O ~100 cmH2O 1 cmH2O
4% lectura real)
± [2.5% (v / v) +
O2 concentración 15 % ~ 100 % 0.1 % 2.5% ( nivel de
concentración)]
Conformidad 1~1000mL/ cmH2O 1mL/ cmH2O

Forma de onda en pantalla:


—— Presión de tiempo-vía aérea (en todos los modos)
——Tiempo—Flujo(en todos los modos)
—— bucle de volumen de presión(todos los modos)

Observaciones: la máquina no puede registrar todos los valores de ajuste y monitoreo.


 El tipo、de posición de sensor、de propósito y método de muestreo de medida、de control y dispositivo de exhibición
Posición de
El propósito Tipo Método de muestreo
sensor
Presión de la vía Final de Elija valor promedio
Presión- voltaje、dato Simulado
aérea exhalación de muestreo múltiple
Final de Elija valor promedio
PEEP Presión- voltaje、dato Simulado
exhalación de muestreo múltiple
Final de Elija valor promedio
Presión continua Presión- voltaje、dato Simulado
exhalación de muestreo múltiple
Elija valor promedio
freq Construido-en Tiempo、dato simulado
de muestreo múltiple
Tiempo de Elija valor promedio
Construido-en Tiempo、dato simulado
inhalación de muestreo múltiple
Aguantando Elija valor promedio
Construido-en Tiempo、dato simulado
tiempo de muestreo múltiple
Flujo- voltaje、 Elija valor promedio
tidal Volumen Exhalando fin
Dato simulado de muestreo múltiple
Exhalando fin、 Flujo-voltaje、 Elija valor promedio
Flujo
Inhalando fin Dato simulado de muestreo múltiple
O2 concentración-voltaje、 Elija valor promedio
O2 concentración Inhalando fin
Dato simulado de muestreo múltiple
Packing List
Name:ICU Ventilator Type: S1100
NO
Name Picture Unit NO. Remark
.

1 Main body 1 PC.

2 Power cord 1 PC.

Accessory

3 Air compressor 1 PC.

4 Filter component 1 PC.


High pressure hose for
5 1 PC.
air compressor

6 High pressure O2 hose 1 PC.

7 CO2 sensor 1 PC.

8 Power cord 1 PC.

9 Fuse 4 PCS.
10 Fuse 2 PCS.

11 Fuse 2 PCS.

12 Fuse 2 PCS.

13 Screw 6 PCS.
14 Serrated saddle 6 PCS.

15 Hanger(support arm) 1 PC.

16 Humidifier (adult) 1 PC.


17 Breathing tube (C-Ⅰ-1) 2 PCS.

18 Breathing tube(C-Ⅰ-1) 3 PCS.

19 Water trap 2 PCS.

20 Y connector 1 PC.

21 Y connector 1 PC.

22 L connector 2 PCS.
23 Test lung 1 PC.

24 Elbow adapter 1 PC.

25 Operation manual 1 PC.

26 Certificate of Quality 1 PC.

27 Packing list 1 PC.

28 Adult mask 1 PC.