You are on page 1of 5

Stryker Endoscopy

5900 Optical Court, San Jose, CA 95138

APPROVALS DATE TITLE


DRAWN BY:

D. Tea 5/30/07 Pneumo Sure High Flow Insufflator User Manual


ORIGINATOR

D. Tea 5/30/07
SIZE REV. DOCUMENT NO. SHEET OF

A A 1000-400-014 1 5
REVISIONS REVISIONS
LTR ECN NO. DATE APPROVED LTR ECN NO. DATE APPROVED
A ECO#105570 5/30/07 D. TEA
B
Printing Instructions
C
D 1. Black and White
2. 8.5” w × 11” h
E 3. Select binding and paper weight based on cost
F 4. Source files are available from Media Services
G
H Global Source
J
K
L
M
N
P
R
T
U
V
W
Y

STRYKER CONFIDENTIAL - THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF STRYKER ENDOSCOPY, SAN
JOSE. NEITHER THIS DOCUMENT NOR THE INFORMATION HEREIN MAY BE REPRODUCED, USED, OR DISCLOSED TO OR FOR THE BENEFIT OF
ANY THIRD PARTY WITHOUT THE PRIOR WRITTEN CONSENT OF STRYKER.
Insufflator for Laparoscopy and Vessel Harvesting

Quick Start Guide

2007/04 www.stryker.com 1000-400-014 Rev. A


Getting to know the control and connection elements

On/Off switch Touch screen Insufflation tube connection

How to use and read the touch screen display


(1) Real-Time Pressure Sensing display
(1) (2) (3) (4) (5) (2) Gas supply display
(3) Actual pressure display
(6)
(15) (4) Gas consumption display / function field for
mm Hg l/min
reset)

(14)
15 0 0.0
3
(7)
(5)
(6)
(7)
Gas heating connected/ready
Increasing nominal gas flow
Nominal gas flow display
(13) Li ter (8) (8) Decreasing nominal gas flow
Mode: High Flow (9) Actual gas flow display / menu function field
(10) START/STOP function field
Change Actual
Mode l/min (11) Status display/error & warning messages
(12) Insufflation mode display /selecting
insufflation mode
(12) (11) (10) (9) (13) Decreasing nominal pressure
Press the function fields (9) (Menu) or (14) Nominal pressure display
(12) (Change insufflation mode), depicted with frame and
(15) Increasing nominal pressure
hold for 2 seconds to trigger functions.

How to read the gas supply display

Bottle pressure >30 bar House gas supply pressure OK


(green) (green)

Bottle pressure 20-30 bar House gas supply pressure too


With pressure of < 30 bar ->message Low gas pressure (5 s) low
(yellow) (red)

Bottle pressure < 20 bar ->message Change gas bottle


Bottle pressure <15 bar -> message Check gas supply,
(red) Insufflation stops automatically

-2-
How to choose the insufflation mode
mm Hg l/min
Press the respective function field to select the desired mm Hg l/min

Pediatric High Flow mode (e.g. High Flow). Unavailable procedure modules Pediatric High Flow
are grayed out and cannot be selected.

Bariatric Bariatric

Available disposable tube sets


Heated tube set with Real-Time Pres- High Flow tube set with Real- High Flow II tube set
sure Sensing (RTP) Time Pressure Sensing (RTP)

Heating element Sensor line Integrated sterile filter

How to connect and remove the tube set


Connecting the tube set Removing the tube set
Press plug of tube in direction of Press plug of tube in direction
device to engage snap-in latch of device to release snap-in
and connect tube set latch and remove tube set

How to use the gas heating function


Connecting to use the gas heating Incorrect or defective heating
Thermometer symbol indicates Crossed out thermometer sym-
mm Hg l/min mm Hg l/min

gas heater is working ok


15 0 0.0
3 15 0 0.0
3
bol indicates gas heater is defect

Li ter Li ter

Gas heater OK Gas heater defective


Mode Actual Change Actual
High Flow l/min Mode l/min

How to use the Real-Time Pressure Sensing function


(Only with "Heated tube set with Real-Time Pressure Sensing" and "High Flow tube set with Real-Time Pressure Sensing")

Thin tube goes to a different trocar for pressure measurement

Thick tube goes to veress needle / first trocar for flow

-3-
Indicating Real-Time Pressure Sensing Defective or interrupted Real-Time Pressure Sensing

mm Hg l/min
Real-Time Pressure Sensing mm Hg l/min
Crossed out Real-Time Pressure
symbole indicates working
15 10 3 continuous pressure 15 0 3
Sensing symbole indicates de-
fective or interrupted conti-
0.0
Lit er measuring. 283
Liter
nuous pressure measuring
Veress insufflation
Mode Actual Change A ctual
High Flow l/min Mode l/min

How to set pressure and flow


mm Hg l/min
Press on up and down but- mm Hg l/min
Press on up and down buttons

15 0 3
tons to change nominal
pressure. 15 0 0.0
3
to change nominal flow.
0.0
Li ter Li ter

Mode: High Flow Mode: High Flow

Change Actual Change A ctual


Mode l/min Mode l/min

How to start and stop insufflation

mm Hg l/min Start insufflation: mm Hg l/min Stop insufflation:

15 0 0.0
3
Press on the START button
15 0 283
40
Press on STOP button.
Li ter Li ter

Mode: High Flow Insufflation: High Flow


Change Actual Mode A ctual
Mode l/min High Flow l/min

Configuration menu - Enter and navigate

mm Hg l/min
Depiction/functions in Configuration menu
15 0 0.0
3 Press the "arrow forward" function field to access the next
Li ter
menu on the same level.
Mode: High Flow
Change
Mode
Actual
l/min Press the "arrow backward" function field to access the
previous menu on the same level or to access the previous
When insufflation is stopped, hold down menu of the next higher menu level without saving settings.
on the Actual field for a second to access Press the (SAVE) function field to save settings. The display
the configuration menu. depicts (SAVED) for 2 seconds. After saving you return
automatically to the previous level.
Press the (EXIT) function field to exit the menu and return to
the work screen.

EC REP

Produced for Stryker European Rep-


Stryker Endoscopy RA/QA Manager
5900 Optical Court, San Jose ZAC Satolas Green Pusignan
CA 95138 ( USA) Av. De Satolas Green
(408) 754-2000 (800) 624-4422 69881 MEYZIEU Cedex France
www.stryker.com

2007/04 www.stryker.com 1200601/10000003898 00/ama


1000-400-014 Rev. A
-4-

You might also like