T51P Indicator
T51XW Indicator
5000 Series Indicators EN-1
TABLE OF CONTENTS
1. INTRODUCTION ..........................................................................................................................................EN-5
1.1 Safety Precautons .....................................................................................................................................EN-5
1.1.1 Relay Opton Safety Precautons ........................................................................................................EN-5
1.2 Overvew of Parts and Controls ...................................................................................................................EN-6
1.3 Control Functons .....................................................................................................................................EN-10
2. INSTALLATION .........................................................................................................................................EN-11
2.1 Unpackng ..............................................................................................................................................EN-11
2.2 External Connectons ................................................................................................................................EN-11
2.2.1 Scale Base wth Connector to T51P .................................................................................................EN-11
2.2.2 RS232 Interface Cable to T51P........................................................................................................EN-11
2.2.3 AC Power to T51P ........................................................................................................................EN-11
2.2.4 AC Power to T51XW .......................................................................................................................EN-11
2.2.5 Battery Power to T51P ....................................................................................................................EN-11
2.2.6 Mountng Bracket ...........................................................................................................................EN-12
2.3 Internal Connectons .................................................................................................................................EN-12
2.3.1 Openng the Housng .......................................................................................................................EN-12
2.3.2 Scale Base Wthout Connector to T51P or T51XW ............................................................................EN-12
2.3.3 RS232 Interface Cable to T51XW ......................................................................................................EN-13
2.3.4 Footswtch to T51P or T51XW ..........................................................................................................EN-13
2.4 T51P Rear Housng Orentaton .................................................................................................................EN-13
2.5 Mountng Bracket .....................................................................................................................................EN-13
3. SETTINGS ................................................................................................................................................EN-14
3.1 Menu Structure ........................................................................................................................................EN-14
3.2 Menu Navgaton .....................................................................................................................................EN-15
3.3 Calbraton Menu .....................................................................................................................................EN-15
3.3.1 Zero Calbraton ............................................................................................................................EN-16
3.3.2 Span Calbraton ...........................................................................................................................EN-16
3.3.3 Lnearty Calbraton ......................................................................................................................EN-17
3.3.4 Calbraton Test ............................................................................................................................EN-18
3.3.5 Geographcal Adjustment Factor .....................................................................................................EN-18
3.3.6 End Calbraton .............................................................................................................................EN-18
3.4 Setup Menu .............................................................................................................................................EN-20
3.4.1 Reset ...........................................................................................................................................EN-20
3.4.2 Range .........................................................................................................................................EN-20
3.4.3 Capacty ......................................................................................................................................EN-20
3.4.4 Graduaton ...................................................................................................................................EN-21
3.4.5 Power On Unt ..............................................................................................................................EN-21
3.4.6 Zero Range ..................................................................................................................................EN-21
3.4.7 Auto-Tare .....................................................................................................................................EN-22
3.4.8 Retan Weght Data .......................................................................................................................EN-22
3.4.9 Legal for Trade .............................................................................................................................EN-22
3.4.10 Beeper Volume .............................................................................................................................EN-23
EN- 5000 Series Indicators
7. MAINTENANCE .........................................................................................................................................EN-53
7.1 Model T51P Cleanng ...............................................................................................................................EN-53
7.2 Model T51XW Cleanng ............................................................................................................................EN-53
7.3 Troubleshootng .......................................................................................................................................EN-53
7.4 Servce Informaton ..................................................................................................................................EN-54
1. INTRODUCTION
Ths manual contans nstallaton, operaton and mantenance nstructons for the T51P and T51XW Indcators. Please read ths
manual completely before nstallaton and operaton.
For safe and dependable operaton of ths equpment, please comply wth the followng safety precautons:
• Verfy that the nput voltage range prnted on the data label matches the local AC power to be used.
• Make sure that the power cord does not pose a potental obstacle or trppng hazard.
• Use only approved accessores and perpherals.
• Operate the equpment only under ambent condtons specfed n these nstructons.
• Dsconnect the equpment from the power supply when cleanng.
• Do not operate the equpment n hazardous or unstable envronments.
• Do not mmerse the equpment n water or other lquds.
• Servce should only be performed by authorzed personnel.
• The T51XW s suppled wth a grounded power cable. Use only wth a compatble grounded power outlet.
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR
BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY
AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
Before makng connectons to the Relay termnals, remove power from the system. If the system contans an optonal
rechargeable battery system, be sure that the ON/ZERO Off button s used to fully turn off the system after removng the AC power
plug.
More detaled nstallaton nstructons are ncluded wth the Relay Opton Kt when purchased.
EN- 5000 Series Indicators
10 16 RS232 Connector
11 12 13 14 15 16
1
TABLE 1-. T51XW PARTS.
2 Item Description
1 Data Label
3 2 Front Housng
4 3 Control Panel
4 Adjustng Knob (2)
5 Mountng Bracket
6 Screw (4)
5 7 Rear housng
8 Data Label
9 Securty Screw
10 Stran relef for opton
11 Stran relef for RS232
12 Stran relef for opton
6 13 Stran relef for opton
10 11 12 13 14 15
/. /&&
1 2 3 4 7 5 6
1 2 3 4 56 78 9
27
26
25
24
23
22
21
20
19 18 17 16 15 14 13 12 11 10
Button
In Wegh mode,
temporarly dsplays 10x
expanded resoluton.
Secondary Functon Off Units Mode Menu
(Long Press) Turns the Indcator off. Changes the weghng Allows changng the Enter the User menu.
Unt. applcaton mode.
. INSTALLATION
.1 Unpacking
Unpack the followng tems:
• T51P or T51XW Indcator • Capacty Label Sheet
• AC Power Cord (T51P only) • LFT Sealng kt
• Mountng Bracket • Instructon Manual CD
• Knobs (2) • Warranty Card
For connectng bases wth a connector to a T51XW (whch does not have the external connector), a Load Cell Cable Adapter
Kt p/n 80500736 s avalable as an accessory. Ths kt connects to the termnal block nsde the T51XW and has an external
connector on the other end.
T51P
Remove the four Phllps head screws from the rear housng.
Remove the front housng beng careful not to dsturb the nternal connectons.
Once all connectons are made, reattach the front housng.
T51XW
Remove the four hex head screws from the rear housng.
Open the housng by carefully pullng the front housng forward.
Once all connectons are made, reattach the front housng.
The screws should be tghtened to 2.5 N m (20-25 n-lb) torque to ensure a watertght seal.
•
After wrng s completed and jumpers are n place, replace the ndcator housng screws. Make sure the lqud-tght connector s
properly tghtened.
5000 Series Indicators EN-1
Fgure 2-3. Wall Mount Confguraton. Fgure 2-4. Bench Top Confguraton.
. SETTINGS
.1 Menu Structure
Notes:
1) Hdden when LEGAL FOR TRADE s ON.
2) Locked at current settng when LEGAL FOR TRADE s ON.
3) Vsble only wth RS485/RS422 opton nstalled.
5000 Series Indicators EN-15
For menu tems wth numerc settngs such as Capacty, the current settng s dsplayed wth all dgts flashng.
Press the No button to begn edtng.
The frst dgt s dsplayed flashng.
Press the No button to ncrement the dgt or press the Yes button to accept the dgt and move to the next dgt.
Press the Yes button when the last dgt has been set.
The new settng s dsplayed wth all dgts flashng. Press the Yes button to accept the settng or press the No
button to resume edtng.
Ths method also apples to settng Checkwegh under and over targets.
For End menu tems, pressng the Yes button advances to the next menu, whle pressng the No button returns
to the top of the current menu.
The dsplay flashes 0 and the calbraton unt. Press the Yes button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.
When SPAN s dsplayed, press the Yes button to ntate Span Calbraton.
The dsplay flashes the span calbraton pont. Place the specfed weght on the scale and press the Yes
button.
To choose a dfferent span pont or calbraton unt, edt the settng as explaned n Secton 3.2 Menu
Navgaton. When the desred settng s dsplayed, place the specfed weght on the scale and press the Yes
button.
Wth no weght on the scale, press the Yes button to establsh the zero pont.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.
5000 Series Indicators EN-1
The dsplay flashes the full calbraton pont and calbraton unt. Place the specfed weght on the scale and
press the Yes button.
To choose a dfferent full pont or calbraton unt (kg or lb), edt the settng as explaned n Secton 3.2 Menu
Navgaton. When the desred settng s dsplayed, place the specfed weght on the scale and press the Yes
button.
Place the specfed weght on the scale and press the Yes button.
Wth no weght on the scale, press the Yes button to establsh the zero pont.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.
EN-1 5000 Series Indicators
NOTE: Calbraton Test s always avalable (even when LFT s set to ON).
When tESt s dsplayed, press the Yes button to ntate Calbraton Test.
The dsplay flashes 0. Wth no weght on the scale, press the Yes button to record the current zero pont.
The dsplay flashes the span calbraton weght usng the value from the last calbraton. The example shows
test weght of 30 kg.
Place the specfed test weght on the scale and press the Yes button.
The dsplay shows --t-- whle the data s processed.
The dsplay flashes the actual dfference between the calbraton data and the test weght.
The example shows a 0.010 kg dfference. The result of the Calbraton Test s prnted.
After 5 seconds, Calbraton Test ends, the scale returns to the actve weghng mode and dsplays the current
weght.
NOTE: Only an authorzed manufacturer’s representatve or certfed verfcaton personnel may make these
•
changes. Changng the geographcal settng alters the calbraton values. •
•
..1 Reset
Reset the Setup menu to the factory defaults. (except Range, Capacty and Graduaton)
NO = not reset.
YES = reset.
NOTE: If the Legal for Trade menu tem s set to ON, the Range, Capacty, Graduaton, Zero Range, Auto Tare,
Retan Weght Data and Legal For Trade settngs are not reset.
.. Range
Set the number of weghng ranges.
SINGLE = one weghng range from zero to full capacty.
dUAL = two weghng ranges, where range 1 s from zero to half capacty and range 2 s from
half capacty to full capacty.
.. Capacity
Set the scale capacty as explaned n Secton 3.2 Menu Navgaton.
NOTE: If dUAL was selected n the rANGE menu tem, the Capacty settng defnes the Range 2 capacty. The
Range 1 capacty s automatcally defned as half of the Capacty settng. For example, f Capacty s set to
15, the Range 1 capacty becomes 7.5.
.. Graduation
Set the scale readablty.
0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000.
NOTE: Graduaton settngs are lmted to values from Capacty dvded by 1000 to Capacty dvded by
•
30000. Therefore, not all settngs are avalable for each capacty.
•
•
NOTE: If dUAL was selected n the rANGE menu tem, the Graduaton settng defnes the Range 1 graduaton.
The Range 2 graduaton s automatcally defned as one step greater than the Graduaton settng. For
example, f Graduaton s set to 0.001, the Range 2 graduaton becomes 0.002.
NOTE: Range 2 graduaton s retaned even under half capacty untl the scale returns to zero.
NOTE: Unts oz, lb:oz and C (custom) wll not be vald as Power On unts when Range s set to Dual. The
next avalable unt wll be dsplayed nstead
.. Auto-Tare
Set the Automatc Tare functonalty.
OFF = Automatc Tare s dsabled.
ON = the frst stable gross weght wll be tared.
ACCEPt = when the applcaton mode s CHECK, stable gross weght that s wthn the
Checkwegh accept lmts wll be tared.
NOTE: When Legal for Trade s set to ON, the Menu settngs are affected as follows:
• Calbraton functons are hdden except for Calbraton Test.
• Capacty s read-only.
• Range, Graduaton, Power On unt, Auto-Tare, Retan Zero, Gross Indcaton, Prnt Output, Unt and Mode
settngs are locked at ther current settngs.
• Zero Range s locked at 2%.
• Stable Range s locked at 1d.
• Auto-Zero Trackng s set to 0.5d.
• Contnuous Prnt s dsabled.
• IP and CP RS232 commands are dsabled.
NOTE: When Legal for Trade s set to ON, t s necessary to set the securty swtch to ON before extng the
menu. If the securty swtch s not set to ON, the message “NO.SW” s dsplayed and the ndcator returns to
the menu.
5000 Series Indicators EN-
.5.1 Reset
Set the Readout menu to factory default settngs.
NO = not reset
YES = reset
If the Legal for Trade menu tem s set to ON, the Stable Range, Averagng Level, Auto Zero Trackng, Auto Off
and Gross settngs are not reset.
NOTE: When LFT s set to ON, the settng s forced to 1 d. The settng s locked when the hardware lock
swtch s set to the ON poston.
.5. Filter
Set the amount of sgnal flterng.
LOW = less stablty, faster stablzaton tme (<1 sec.)
MEd = normal stablty, stablzaton tme (<2 sec.)
HI = greater stablty, slower stablzaton tme (<3 sec.)
NOTE: When the LFT menu tem s set to ON, the selectons are lmted to 0.5d, 1d and 3d. The settng s
locked when the hardware lock swtch s set to the ON poston.
5000 Series Indicators EN-5
.5.5 Backlight
Set the dsplay backlght functonalty.
OFF = always off.
ON = always on.
AUtO = turns on when a button s pressed or the dsplayed weght changes.
When Auto s selected, set Backlght shut off tme.
Settngs:
SEt 1 = backlght turns off after 1 mnute of no actvty.
SEt 2 = backlght turns off after 2 mnute of no actvty.
SEt 5 = backlght turns off after 5 mnute of no actvty.
..1 Reset
Set the Mode menu to the factory defaults.
NO = not reset.
YES = reset.
NOTE: If the Legal for trade menu tem s set ON, the settngs are not reset.
..1 Reset
Set the Unt menu to the factory defaults.
NO = not reset.
YES =reset
Note: If the Legal for Trade menu tem s set ON, the settngs are not reset.
Factor
Set the converson factor.
0.00001 to 9.99999
Refer to Secton 3.2 Menu Navgaton to enter settngs.
5000 Series Indicators EN-
Exponent
Set the factor multpler.
0 = 100 (Factor x 1)
1 = 101 (Factor x 10)
•
2 = 102 (Factor x 100)
•
3 = 103 (Factor x 1000)
•
-2 = 10-2 (Factor ÷ 100)
-1 = 10-1 (Factor ÷ 10)
..1 Reset
Set the GMP menu to factory defaults.
NO = not reset.
YES = reset.
12 hour format
12 A to 12 P = hour poston
00 to 59 = mnute poston
(current tme blnkng)
Refer to Secton 3.2 Menu Navgaton to enter settngs.
(Set hours 01 to 12 A or P)
.. User ID
Set the user dentfcaton.
000000 to 999999
.. Project ID
Set the Project dentfcaton.
000000 to 999999
Refer to Secton 3.2 Menu Navgaton to enter settngs.
.. Scale ID
Set the Scale dentfcaton.
000000 to 999999
Refer to Secton 3.2 Menu Navgaton to enter settngs.
Gross
Set the status.
OFF = Dsabled.
ON = the Gross weght s prnted.
Net
Set the status.
OFF = Dsabled.
ON = the Net weght s prnted.
Tare
Set the status.
OFF = Dsabled.
ON = the Tare weght s prnted.
Header
Set the status.
OFF = Dsabled.
ON = the Header s prnted.
User ID
Set the status.
OFF = Dsabled.
ON = the User ID s prnted.
EN- 5000 Series Indicators
Project ID
Set the status.
OFF = Dsabled.
ON = the Project ID s prnted.
Scale ID
Set the status.
OFF = Dsabled.
ON = the Scale ID s prnted.
Time
Set the status.
OFF = Dsabled.
ON = the Date and Tme s prnted.
Difference
Set the status.
OFF = Dsabled.
ON = the Calbraton Test dfference s prnted.
Reference Information
Set the status.
OFF = Dsabled.
ON = the Reference Informaton s prnted.
NOTE: The Reference Informaton s dependent on the actve mode (Wegh mode: None, Count mode: APW,
Percent mode: Reference Weght, Dynamc mode: Level, Check Wegh mode: Under and Over lmts).
Mode
Set the status.
OFF = Dsabled.
ON = the Mode s prnted.
Name
Set the status.
OFF = Dsabled.
ON = the Name lne s prnted.
5000 Series Indicators EN-5
Format
Set the prntng format.
MULtI = a mult-lne (sngle column style) prntout s generated. A CRLF s added after each
tem.
SINGLE = a sngle lne prntout s generated. (A TAB space s added between each tem and a
CLRF s used only after the very last tem.)
Line Feed
Set the paper feed.
LINE = move paper up one lne after prntng
4.LINE = move paper up four lnes after prntng
FOrM = a form feed s appended to the prntout
.. Output
Set the format of the seral output strng to a prnter or computer.
DEF = use the default output format of the T51 ndcator (see Secton 5.2 Output Format).
C11 = use the output format of the Ohaus CD/CW-11 ndcators (see respectve
CD-11/CW-11 user manuals).
.10.1 Reset
Set the COM1 and COM2 menu to factory defaults.
NO = not reset.
YES = reset.
.10. Baud
Set the Baud rate.
300 = 300 bps
600 = 600 bps
1200 =1200 bps
2400 = 2400 bps
4800 = 4800 bps
9600 = 9600 bps
19200 = 19200 bps
.10. Parity
Set the data bts and party.
7 EVEN = 7 data bts, even party.
7 Odd = 7 data bts, odd party.
7 NONE = 7 data bts, no party.
8 NONE = 8 data bts, no party.
.10.5 Handshake
Set the flow control method.
NONE = no handshakng.
ON-OFF = XON/XOFF software handshakng.
HArd = hardware handshakng.
.10. Address
Set the communcaton address.
NOTE: Address s only dsplayed n the COM2 menu f the RS422/RS485 opton s nstalled.
OFF = no address.
01 to 99 = address 01 to 99 •
•
•
5000 Series Indicators EN-
Alternate Tare
Set the alternate command character for Tare.
A to Z.
Alternate Zero
Set the alternate command character for Zero.
A to Z.
.11.1 Reset
Set the I-O menu to factory defaults
NO = not reset.
YES = reset.
EN- 5000 Series Indicators
Type
Set the ntal state of the relay.
OPEN = the relay output s normally open.
CLOSEd = the relay output s normally closed.
CAUTION: The normally closed relay condton s only actve whle the Indcator s powered on.
When powered off or when power s removed, the relay condton returns to a normally open
condton. Restorng power to the Indcator wll restore the closed condton of the relays.
Output Sequence
Set how the relay outputs react as the weght readng changes from under / accept / over.
NOrM = the prevously enabled relay wll be dsabled as the next relay s enabled.
HOLd = the prevously enabled relay wll hold the same state as the next relay s enabled.
5000 Series Indicators EN-
Contact
Set the tmng of the relay contacts.
NOTE: A 100 ms delay or over-lap s used for the break-before-make and make-before-break tmng.
Stable
Set how the relay outputs react durng nstablty.
OFF = relay changes are mmedate.
ON = delays relay changes untl weght readng s stable.
.11.5 End I-O
Advance to the next menu.
Use ths menu to prevent unauthorzed changes to menu settngs. When the securty Reset No, Yes
swtch s set to ON, the locked menus can be vewed but not changed. Default settngs Lock Calbraton Menu Off, On
are bold. Lock Setup Menu Off, On
Lock Readout Menu Off, On
Lock Mode Menu Off, On
Lock Unt Menu Off, On
Lock Prnt1 Menu Off, On
Lock Prnt2 Menu Off, On
Lock Com1 Menu Off, On
Lock Com2 Menu Off, On
Lock GMP Menu Off, On
Lock I-O Menu Off, On
End Lock Menu
.1.1 Reset
Set the menu Lock menu to factory defaults.
NO = not reset.
YES = reset.
NOTE: Settngs for LFT controlled menu tems are not reset.
NOTE: When the Menu button s locked, the user may unlock ths button by holdng the Menu button for 10
seconds untl UNLOCK s dsplayed. The hardware Lock Swtch must be n the unlocked poston.
Open the housng as explaned n Secton 2.3.1. Set the poston of securty swtch SW2 to ON as shown n Fgure 1-3.
5000 Series Indicators EN-
. OPERATION
.1 Turning Indicator On/Off
To turn the Indcator on, press the ON/ZERO Off button. The Indcator performs a dsplay test followed by a
seres of nformatonal dsplays, and then enters the actve weghng mode.
To turn the Indcator off, press and hold the ON/ZERO Off button untl OFF s dsplayed.
Press the ON/ZERO Off button to zero the weght dsplay. The scale must be stable to accept zero operaton.
Place the empty contaner on the scale (example 0.5 kg) and press the TARE button. The dsplay wll show
the net weght.
To clear the Tare value, empty the scale and press the TARE button. The dsplay wll show the gross weght.
The dsplay wll show the Pre-set Tare as a negatve value, wth the PT Indcator on.
NOTES: 1. The PT value wll supersede any other Tare or PT value n memory.
2. When usng Pre-Set Tare, make sure that Auto-Tare functon s set off n the Setup menu.
3. If the Tare entry ncludes dgts beyond the readablty of the Indcator, the tare value s rounded off
to the readablty of the Indcator.
To clear a Pre-set Tare value, empty the scale then press the TARE button. The dsplay wll show the Gross
weght.
.5 Auto-Tare
Auto-Tare automatcally tares the ntal weght (such as a contaner) placed on the empty scale, wthout
havng to press the TARE button. The tare value s cleared automatcally when the weght on the scale s fully
removed.
Durng Checkweghng operaton, f the On Accept settng s selected n the Setup menu, weght values that are
wthn the accept range wll be tared automatcally.
Press the PRINT Units button to send the dsplayed data to the communcaton port (the Auto-Prnt Mode n Secton 3.9 functon
must be Off).
..1 Weighing
Place the tem to be weghed on the scale. The llustraton ndcates a sample of 1.5 kg, Gross weght.
NOTE: Press the FUNCTION Mode button to temporarly dsplay the weght n 10x expanded resoluton.
NOTE: Press the FUNCTION Mode button to temporarly dsplay the APW value.
To establsh the APW, place the specfed quantty of samples on the scale and press the FUNCTION Mode
button to capture the weght.
APW s dsplayed shortly followed by the APW value wth the current unt of measure.
Begin Counting
Place the parts on the scale and read the count. If a contaner s used, be sure to tare the empty contaner
frst.
NOTE: Press the FUNCTION Mode button to temporarly dsplay the reference weght.
Establishing a New Reference Weight
The dsplay shows Put.rEF %.
To establsh the Ref Wt, place the sample on the scale and press the FUNCTION Mode button to capture the
weght. rEF.Wt s dsplayed shortly followed by the REF Wt value wth the current unt of measure.
contaner frst.
EN- 5000 Series Indicators
Checkweighing Limits
When the FUNCTION Mode button s released, CLr.rEF s dsplayed.
NOTE: Press the FUNCTION Mode button to temporarly dsplay the Under and Over Lmt values.
Settngs:
-999950 to 999950
Refer to Menu Navgaton Secton 3.2 to enter settngs.
to
NOTE: The mnus sgn s used together wth the frst dgt to show a negatve value.
Settngs:
-999950 to 999950
Refer to Menu Navgaton Secton 3.2 to enter settngs.
For loads greater than the Under Lmt and less than the Over lmt, the green Accept LED s lt.
For loads greater than the Over Lmt, the red Over LED s lt.
5000 Series Indicators EN-
If the manual mode s n use, press the FUNCTION Mode button to start measurement. If the sem-automatc
or automatc mode s n use, measurement s started automatcally.
•
NOTE: When usng manual mode, t s not necessary for the dsplay to be at zero gross or net. When usng
•
sem-automatc or automatc mode, the dsplay must be at zero gross or net before placng the object on
•
the scale. The example s for a settng of 5 seconds. Durng the averagng perod, the countdown tmer
decreases n one second ncrements.
NOTE: If SEt 0 was selected n the Dynamc menu tem, the countdown tmer s not dsplayed.
When the countdown has completed, the readngs are averaged and held on the dsplay.
If the manual or sem-automatc mode s n use, reset the countdown tmer by pressng the FUNCTION Mode
If the automatc mode s n use, the held readng s shown on the dsplay for 10 seconds after the object s
removed to wthn 5 dvsons of zero. Then the dsplay shows rEAdY.
5. SERIAL COMMUNICATION
The T51P and T51XW Indcators nclude an RS232 seral communcaton nterface.
The setup of RS232 operatng parameters are more fully explaned n Secton 3.10. The physcal hardware connecton s
explaned n Secton 2.6.
The nterface enables dsplay and GMP data to be sent to a computer or prnter. A computer can be used to control some
functons of the ndcator usng the commands lsted n Table 5-1.
NOTES:
1) Commands sent to the Indcator must be termnated wth a carrage return (CR) or carrage return-lne feed (CRLF).
2) Alternate command characters may be defned by the user (see Alternate Commands n Secton 3.10).
3) Data output by the Indcator s always termnated wth a carrage return-lne feed (CRLF).
4) These commands are only avalable when Prnt>Output s set to C11 (see Secton 3.9.6).
5000 Series Indicators EN-
Field: Weight Space* Unit Space* Stability Space* G/N Space* Term. Char(s)
Length: 1 5 1 1 1 N 1 **
5. Printouts
The followng sample prnt outs are generated by the Print button, “P” Command or alternate prnt command. The content of the
prntout s defned n the Prnt Content menu tem. A maxmum of 24 characters can be prnted on each lne.
NOTE: Shaded areas = ths date s prnted when set on n the Prnt Content menu.
Unshaded = typcal
Dynamic Mode Printout Check Weighing Mode Printout Calibration Test Printout
Ohaus Corporation Ohaus Corporation --------- Cal Test ---------
19A Chapin Road 19A Chapin Road Ohaus Corporation
P.O. Box 2033 P.O. Box 2033 19A Chapin Road
Pine Brook, NJ, 07058 USA Pine Brook, NJ, 07058 USA P.O. Box 2033
Tel: +1-973-377-9000 Tel: +1-973-377-9000 Pine Brook, NJ, 07058 USA
01/31/08 12:30 PM 01/31/08 12:30 PM Tel: +1-973-377-9000
Scale ID: 123456 Scale ID: 123456 01/31/08 12:30 PM
User ID: 123456 User ID: 123456 Scale ID: 123456
Project ID: 123456 Project ID: 123456 User ID: 123456
Name:___________________ Name:___________________ Project ID: 123456
Final Wt.: 0.200 kg NET Result: 10.00 kg NET OVER Name:------------------
12.34 kg G 11.00 kg G Mode: Test
11.11 kg NET 10.00 kg NET New Cal: 10.000 kg
1.22 kg T 1.00 kg T Old Cal: 10.000 kg
Level: 10 Under: 9.99 kg Diff: 0.000 kg
Mode: Dynamic Over: 10.01 kg Wt. ID:------------------
Mode: Checkweigh ------------ End -----------
5000 Series Indicators EN-51
.1 Settings
Before verfcaton and sealng, perform the followng steps:
1. Verfy that the menu settngs meet the local weghts and measures regulatons.
2. Perform a calbraton as explaned n Secton 3.
3. Set Legal for Trade to ON n the Setup menu.
4. Wthout extng the menu, turn the ndcator off.
5. Dsconnect power from the ndcator and open the housng as explaned n Secton 2.3.1.
6. Set the poston of the securty swtch SW2 to ON as shown n Fgure 1-3, tem 4.
7. Close the housng.
8. Reconnect power and turn the ndcator on.
NOTE: When Legal for Trade s set to ON and the securty swtch s set to ON, the followng menu settngs cannot be changed:
Zero Calbraton, Span Calbraton, Lnearty Calbraton, GEO, Range, Capacty, Graduaton, Power On Unt, Zero Range, Auto
Tare, Retan Weght, Legal for Trade, Stable Range, Auto Zero Trackng, Gross Indcator, Modes, Unts, Stable Only.
. Verification
The local weghts and measures offcal or authorzed servce agent must perform the verfcaton procedure.
. Sealing
The local weghts and measures offcal or authorzed servce agent must apply a securty seal to prevent tamperng wth the
settngs. Refer to the llustratons below for sealng methods.
Fgure 6-1. T51P Wre Seal Fgure 6-2. T51XW Wre Seal
EN-5 5000 Series Indicators
3%!,
Fgure 6-3. T51P Paper Seal Fgure 6-4. T51XW Paper Seal
When the scale base s attached to the ndcator usng a connector, t s necessary to seal the load cell cable to the ndcator n
some jursdctons. The load cell sealng collar P/N 80500737 (Fgure 6-5) s avalable as an accessory.
3%!,
3%!,
. MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
. Troubleshooting
Cannot zero the Scale, or wll not zero when Load on Scale exceeds allowable lmts. Remove load on Scale.
turned on.
Load on Scale s not stable. Wat for load to become stable.
Unable to calbrate. Lock Calbraton Menu set to On. Set Lock Calbraton Menu to Off.
Refer to Secton 3.12 Menu Lock.
. TECHNICAL DATA
.1 Specifications
Materials
T51XW Housng: stanless-steel
T51P Housng: ABS plastc
Dsplay wndow: polycarbonate
Keypad: polyester
Feet: Rubber
Ambient conditions
The techncal data s vald under the followng ambent condtons:
Ambent temperature: -10°C to 40°C / 14°F to 104°F
Relatve humdty: Maxmum relatve humdty 80% for temperatures up to 31°C decreasng lnearly to 50%
relatve humdty at 40°C.
Heght above sea level: up to 2000m
Operablty s assured at ambent temperatures between -10°C and 40°C.
The Rechargeable Battery Kt, RS232 Kt, RS422/485 Kt, AC Relay Kt, DC Relay kt and Foot swtch must be
nstalled by a qualfed techncan.
5000 Series Indicators EN-5
. Compliance
Complance to the followng standards s ndcated by the correspondng markng on the product.
Marking Standard
Ths product conforms to the EMC Drectve 2004/108/EC, the Low Voltage Drectve 2006/95/EC and the
Non-automatc Weghng Instruments Drectve 2009/23/EC. The complete Declaraton of Conformty s
avalable from Ohaus Corporaton.
UL60950-1: 2003
AS/NZS4251.1, AS/NZS4252.1
EU Emissions Note
Ths devce comples wth EN55011 / CISPR 11 Class A Group 1.
FCC Note
Ths equpment has been tested and found to comply wth the lmts for a Class A dgtal devce, pursuant to Part 15 of the FCC
Rules. These lmts are desgned to provde reasonable protecton aganst harmful nterference when the equpment s operated
n a commercal envronment. Ths equpment generates, uses, and can radate rado frequency energy and, f not nstalled
and used n accordance wth the nstructon manual, may cause harmful nterference to rado communcatons. Operaton of
ths equpment n a resdental area s lkely to cause harmful nterference n whch case the user wll be requred to correct the
nterference at hs own expense.
Weghng Instruments verfed at the place of manufacture bear one of the precedng marks on the pack-
ng label and the green ‘M’ (metrology) stcker on the descrptve plate. They may be put nto servce
mmedately.
Weghng Instruments to be verfed n two stages have no green ‘M’ (metrology) on the descrptve plate
and bear one of the precedng dentfcaton mark on the packng label. The second stage of the ntal
verfcaton must be carred out by the approved servce organzaton of the authorzed representatve
wthn the EC or by the natonal weghts & measures (W+M) authortes.
The frst stage of the ntal verfcaton has been carred out at the manufacturer’s work. It comprses all tests accordng to the
adopted European standard EN 45501:1992, paragraph 8.2.2.
If natonal regulatons lmt the valdty perod of the verfcaton, the user of the weghng nstrument must strctly observe the
re-verfcaton perod and nform the respectve W+M authortes.
Disposal
In conformance wth the European Drectve 2002/96/EC on Waste Electrcal and Electronc Equpment
(WEEE) ths devce may not be dsposed of n domestc waste. Ths also apples to countres outsde
the EU, per ther specfc requrements.
The Batteres Drectve 2006/66/EC ntroduces new requrements from September 2008 on removablty
of batteres from waste equpment n EU Member States. To comply wth ths Drectve, ths devce has
been desgned for safe removal of the batteres at end-of-lfe by a waste treatment faclty.
Please dspose of ths product n accordance wth local regulatons at the collectng pont specfed for
electrcal and electronc equpment.
If you have any questons, please contact the responsble authorty or the dstrbutor from whch you
purchased ths devce.
Should ths devce be passed on to other partes (for prvate or professonal use), the content of ths
regulaton must also be related.
Ohaus products are warranted aganst defects n materals and workmanshp from the date of delvery through the duraton of
the warranty perod. Durng the warranty perod Ohaus wll repar, or, at ts opton, replace any component(s) that proves to be
defectve at No charge, provded that the product s returned, freght prepad, to Ohaus.
Ths warranty does Not apply f the product has been damaged by accdent or msuse, exposed to radoactve or corrosve
materals, has foregn materal penetratng to the nsde of the product, or as a result of servce or modfcaton by other than
Ohaus. In leu of a properly returned warranty regstraton card, the warranty perod shall begn on the date of shpment to the
authorzed dealer. No other express or mpled warranty s gven by Ohaus Corporaton. Ohaus Corporaton shall Not be lable
for any consequental damages.
As warranty legslaton dffers from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further detals.