1 DIV.DEC Press and to set decimals and minimum division. In setup: scroll forward through parameters
In numeric input: decreases the digit to be modified
2 CEL.CAP Press , and to set the total capacity of the load cells. ZERO
3 CEL.SEN Press , and to set the mV/V sensitivity of the load cell. In setup: scroll back through the functions
In numeric input: increases the digit to be modified
4 DEAD.LD Press , and to set the weight value present on the load cell. TARE
5 RANGE Press , and to set the scale capacity. In setup: quickly position the first step of a menu
In numeric input: selects the digit to be modified, from
6 ZERO Quick calibration of zero. MODE left to right
7 WEIGHT Press , and to set the weight value used for the calibration.
In setup: enter into a parameter or to confirm a setting
8 ADC.MVV Check the load cell signal in millivolts. PRINT In numeric input: confirms the entry made
Turn the instrument on and off
9 AN.OUT MODE Set the analog output on net or gross weight. C In setup: exit a step without confirming the setting
ON/OFF In numeric input: clears the present value
Optional Analog AO MAX Set the maximum value of the analog output.
Model Parameter AO ZER Set the analog output when the scale shows zero weight.
AO MIN Set the minimum value of the analog output.
When settings are complete press until the indicator displays SAVE?. Press to save set up and return to the weigh mode.
Pressing any other key exits the setup and discards changes.
J1
I OUT 1
OUT 2
RS-232
RJ IN
12/24 Vdc
• SERIAL PORT
RS 485 RS 232
RJ-IN 485 Line 12. TX Transmission
RJ-OUT 485 Line 13. RX Reception
Pin 4 = A(+) , Pin5 = B(-) 14. GND GND