SCT-2200

Quick Setup Guide


To enter the setup menu, turn the instrument on and briefly press the mode key ( ) when the software version displays.
Press the print key ( ) in the setup menu to view/select the parameter choices. Key Function

1 DIV.DEC Press and to set decimals and minimum division. In setup: scroll forward through parameters
In numeric input: decreases the digit to be modified
2 CEL.CAP Press , and to set the total capacity of the load cells. ZERO

3 CEL.SEN Press , and to set the mV/V sensitivity of the load cell. In setup: scroll back through the functions
In numeric input: increases the digit to be modified
4 DEAD.LD Press , and to set the weight value present on the load cell. TARE
5 RANGE Press , and to set the scale capacity. In setup: quickly position the first step of a menu
In numeric input: selects the digit to be modified, from
6 ZERO Quick calibration of zero. MODE left to right
7 WEIGHT Press , and to set the weight value used for the calibration.
In setup: enter into a parameter or to confirm a setting
8 ADC.MVV Check the load cell signal in millivolts. PRINT In numeric input: confirms the entry made
Turn the instrument on and off
9 AN.OUT MODE Set the analog output on net or gross weight. C In setup: exit a step without confirming the setting
ON/OFF In numeric input: clears the present value
Optional Analog AO MAX Set the maximum value of the analog output.
Model Parameter AO ZER Set the analog output when the scale shows zero weight.
AO MIN Set the minimum value of the analog output.

10 INPUTS INP.01 Set the key to be emulated by the input.


INP.02

11 OUTPUT OUT.01 FUNC 0 NONE Disabled.


OUT.02 1 GROS Setpoint for gross weight.
2 NET Setpoint for net weight, even with no tare.
3 PCS Setpoint for pieces.
Default Factory Calibration
4 GRO.0 Gross weight at zero.
The instrument has a default calibration with the following 5 NET.0 Net weight at zero.
features: 6 MOTI Active on instability.
Capacity = 10,000 kg Active when is pressed.
23 K.PR
Load cell sensitivity = 2.000mV/V 25.K.MOD Active when is pressed.
Divisions = 1 26 K.C Active when the C key is pressed.
Theoretical Calibration 27.K.ZER Active when is pressed.
Use parameters 1,2,3,4 to perform a theoretical calibration. 28.K.TAR Active when is pressed.
29 ERR Error indication.
Calibration Using Known Weight
30.NET.T Setpoint for net weight when tare exists.
Use parameters 1,5,6,7 to perform a standard calibration
with a known weight. For parameters 6 & 7, press and NO/NC Set the output normal state.
wait for ERNOT, STORE?. Then press to accept. ONSTAT Set the output switching condition.
For parameter 7, apply test weight before pressing . RL.IST Enable or disable the hysteresis.
Setpoint Programming ENAB.TM Enable time to disable the output.
In the weigh mode, press and hold to access output’s DELAY Enable time to enable the output.
setpoint programming.

When settings are complete press until the indicator displays SAVE?. Press to save set up and return to the weigh mode.
Pressing any other key exits the setup and discards changes.

November 7, 2018 PN 184974 Rev A


Wiring Diagram
SCT-2200

J1

Analog Output CAL


IN 2
IN 1
V COMOUT

I OUT 1
OUT 2
RS-232

Load Cell RJ OUT

RJ IN

12/24 Vdc

• VE 12 / 24 Vdc POWER SUPPLY • LOAD CELL RECEPTORS


1. GND 0 Vdc (GND) CELL:
2. +Vdc +12 / 24 Vdc 15. SIG+ SIGNAL +
16. SIG- SIGNAL -
• CALIBRATION JUMPER
17. SEN+
18. SEN-
SENSE +
SENSE -
J1 = close to enable access to the metrological parameters 19. EXC+ EXCITATION +
20. EXC- EXCITATION -
• ANALOG OUTPUT (Analog Option Only)
On current: On voltage:
10. I+ +20 mA 8. V+ +10 V
11. I- 0 mA (GND) 9. V- 0 V (GND)
Note: the maximum resistance applicable on the output current is 350 Ÿ and the minimum resistance applicable on the output
voltage is 10 NŸ.

• INPUTS AND OUTPUTS


Optoisolated Inputs positive logic (12–24Vdc, 5–20mA max) Outputs (48Vac or 60Vdc, 150mA max):
6. IN1 input 1 3. OUT1 output 1
7. IN2 input 2 4. OUT2 output 2
Input common normally connected to ground. 5. COMOUT output common

• SERIAL PORT
RS 485 RS 232
RJ-IN 485 Line 12. TX Transmission
RJ-OUT 485 Line 13. RX Reception
Pin 4 = A(+) , Pin5 = B(-) 14. GND GND

© Rice Lake Weighing Systems Specifications subject to change without notice.


Rice Lake Weighing Systems is an ISO 9001 registered company.
230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319
November 7, 2018 www.ricelake.com PN 184974 Rev A