You are on page 1of 10

ORIGINAL ARTICLE INTERNATIONAL JOURNAL OF LA BO RATO RY HEMATOLOGY

Evaluation and performance characteristics of the


coagulation system: ACL TOP analyzer – HemosIL reagents
M. MILOS*, D. HERAK*, L. KURIC †, I. HORVAT †, R. ZADRO*

*
Clinical Institute of Laboratory SUMMARY
Diagnosis, Clinical Hospital Center
Zagreb University School of ACL TOP is a fully automated coagulation analyzer, designed for
Medicine, Zagreb, Croatia, simultaneous measurement of routine and special coagulation param-

Faculty of Pharmacy and
Biochemistry, University of Zagreb,
eters. We evaluated analytical and technical performance characteris-
Zagreb, Croatia tics of the coagulation system composed of the ACL TOP analyzer and
HemosIL reagent group for the determination of routine clotting (PT,
Correspondence: APTT, fibrinogen, FVII, and FVIII), chromogenic (protein C) and
Dr Renata Zadro, Clinical Institute
of Laboratory Diagnosis, Zagreb immunological assays (FXIII antigen). Within run and between run
Clinical Hospital Center, Kispaticeva CVs ranged from 0.9% to 7.7% and from 2.0% to 14.8% respectively.
12, 10000 Zagreb, Croatia. The obtained CVs for imprecision of calibration curves were <5% for PT
Tel.: +385 1 2388 247;
and <7% for fibrinogen. The method comparison study showed good
Fax: +385 1 2312 079;
E-mail: rzadro@mef.hr correlation between results obtained on the ACL TOP and BCS/BCT
analyzers, with correlation co-efficients ranging from 0.709 to 0.955,
doi:10.1111/j.1751-553X.2007.00999.x but with significantly different results for PT INR, APTT, fibrinogen and
Received 2 May 2007; accepted for
protein C, and wide dispersion of differences observed in difference
publication 2 October 2007 plots for most assays. Despite good correlation and agreement for FVIII,
problems in measuring FVIII<10% were encountered. The effective
Keywords throughput for the ACL TOP and BCS was 151 and 212 PT/APTT/
Coagulation analyzer, ACL TOP, fibrinogen tests per hour, respectively. Although the ACL TOP
HemosIL reagents, evaluation
is designed to run multiple assays on a large number of samples,
software limitations make the instrument suitable rather for mid-sized
laboratories.

designed for the screening of a large number of sam-


INTRODUCTION
ples, with capabilities for performing a wide range of
During the last 15 years, coagulation analyzers and different coagulation assays on a single or on multiple
reagents have changed significantly. The current gener- samples at the same time. Along with the technological
ation of coagulation analyzers is fully automated, and development of analyzers, major improvement has
combines several technologies in the same instrument been made in software support, offering users easier
(Walenga & Fareed, 1994). By using clotting, chromo- and more efficient processing of daily routine.
genic, and immunological methods for quantification The aim of the present study was to evaluate ana-
of basic and specific hemostatic parameters, they are lytical and technical performance of a coagulation
Ó 2007 The Authors
26 Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM 27

testing system composed of the ACL TOP coagulation for immunologic measurements. The control module
analyzer and HemosIL reagent group (Instrumentation consists of the software running under Windows 2000
Laboratory, Lexington, MA, USA) under routine labo- Operating System (version 2.1) (Microsoft Corpora-
ratory conditions. We tested prothrombin time (PT), tion, Redmond, WA, USA) and touch-screen. The
activated partial thromboplastin time (APTT), fibrino- analyzer has the software capabilities for automatic
gen (Fib), factor VII (FVII), factor VIII (FVIII) and pro- rerun, automatic reflex testing, STAT sample loading,
tein C (PC) activities, and FXIII antigen (FXIII : Ag). and factor parallelism testing. There is a large results
The evaluation addressed several issues, including database (20 000 samples with up to 30 assays per
available methodologies, validation of performance, sample) and a possibility of access to reaction curves.
throughput study, reagent and patient sample on- Daily maintenance is a semiautomatic procedure and
board capabilities and ease of operation. usually lasts approximately 7 min.

MATERIALS AND METHODS Evaluated samples

Patient samples referred to the laboratory for routine


Description of the ACL TOP analyzer
coagulation testing were used for evaluation proce-
The ACL TOP analyzer is a fully automated, random- dure. Platelet-poor plasma was prepared by centrifu-
access, multiparameter coagulation analyzer designed gation at 2000 g for 15 min at room temperature from
for simultaneous measurement of routine and special blood specimens collected into siliconized glass tubes
coagulation parameters. The system is composed of (Vacutainer Becton Dickinson, Meylan Cedex, France)
control and analytical modules. The analytical module containing 0.105 mol/l buffered sodium citrate. Plasma
consists of the functionality required to handle and samples were used for precision, correlation and
process samples, reagents and auxiliary materials, and throughput study. Commercially available lyophilized
can perform clotting, chromogenic and latex immuno- plasma samples from Instrumentation Laboratory: He-
assay methods. Sample and reagent areas have covers mosIL Normal Control, HemosIL Low Abnormal Con-
that provide increased operating safety and reduce the trol and Special Test Level 2 were used for quality
influence of external environment. The ACL TOP control and for precision study.
allows continuous loading of cuvettes, reagents and
samples. The cuvette loading area is on the left side of
Reagents and assay procedures
the instrument, allowing the maximum single load of
800 cuvettes. The sample area is at room temperature The reagents and the calibrators (brand name and
and can hold 12 racks, each capable of holding 10 manufacturer) used on all analyzers are listed in
samples either in primary sample tubes or in original Table 1, as well as test methodology. All reagents, cal-
sample cups. The reagent area is on the right side of ibrators, controls and diluents used on the ACL TOP
the analytical module and has 60 positions for were from HemosIL group of reagents. Assays were
reagents, with 44 positions cooled at 15 ± 3 °C, and performed according to the manufacturer’s specifica-
16 positions at room temperature dedicated for con- tions and the standard laboratory methods. Calibra-
trols, calibrators, and diluents. Samples and reagents tion plasma (HemosIL Calibration plasma) was used
are delivered to the instrument on sample and reagent and diluted automatically by the analyzer to create
racks, with barcode identification. The sample arm specific calibration curves.
consists of a heated probe and syringe used for aspi- PT was initiated by adding thromboplastin solu-
rating and dispensing samples. There are two reagent tion (RecombiPlasTin) to plasma sample, without
arms: the left arm for aspirating and dispensing incubation.
diluents and intermediate reagents and the right arm For APTT, two different reagents were evaluated
for aspirating and dispensing start reagents. Each on the ACL TOP: SynthASil (APTT1) and APTT-SP
probe has a level sensor. Optical reading is performed reagent (APTT2), both with colloidal silica as an
using two wavelengths: 405 or 671 nm for clotting activator. A mixture of APTT reagent and plasma sam-
assays, 405 nm for chromogenic, and 405 or 671 nm ple was incubated at 37 °C for 300 s, and calcium
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
28 M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM

Table 1. Reagents and calibrators used on the ACL TOP, BCS and BCT

ACL TOP reagent/calibrator BCS reagent/calibrator BCT reagent/calibrator


Test Methodology Instrumentation Laboratory Dade Behring Dade Behring

PT Coagulation RecombiPlasTin Innovin –


HemosIL Calibration plasma PT-Multi Calibrator
APTT1 Coagulation SynthASil Actin FS –
APTT2 Coagulation APTT-SP Actin FS –
Fib Coagulation FibrinogenC-XL Multifibren U –
HemosIL Calibration plasma Fibrinogen Standards 1–6
PC Chromogenic Protein C Berichrom Protein C –
HemosIL Calibration plasma Standard Human Plasma
FVII Coagulation FVII deficient plasma – FVII deficient plasma
RecombiPlasTin Innovin
HemosIL Calibration plasma Standard Human Plasma
FVIII Coagulation FVIII deficient plasma – FVIII deficient plasma
SynthASil Actin FS
HemosIL Calibration plasma Standard Human Plasma
FXIII : Ag Immunological FXIII Antigen – –
HemosIL Calibration plasma
FXIII : Act Chromogenic – Berichrom FXIII –
Standard Human Plasma

chloride (CaCl2) solution was added (0.025 mol/l after adding of the 0.02 mol/l CaCl2 solution. The
CaCl2 for APTT-SP and 0.02 mol/l CaCl2 for analyzer automatically remeasured all samples with
SynthASil). FVIII activities <10%, using different dilution of
Fib was measured according to Clauss (1957) by plasma and factor diluent on the same calibration
adding plasma sample, diluted 1 : 10 in Factor Diluent curve.
solution, to thrombin solution (35 UNIH/ml) (Fibrino- For FXIII : Ag determination, plasma sample,
gen C-XL). Samples with Fib concentration >6.5 and diluted 1 : 10 in FXIII Diluent solution, was preincu-
<0.8 g/l were remeasured using Fib-high and Fib-low bated at 37 °C. After adding of FXIII Buffer, and incu-
assay procedures, respectively. bation, FXIII Latex reagent (latex polystyrene particles
PC activity was determined using a chromogenic coated with a rabbit polyclonal antibody against the
assay by incubating plasma sample, diluted 1 : 4 in A-subunit of FXIII) was added and the change in
Protein C Diluent solution, and PC activator solution OD671 nm was recorded.
(fraction from the venom of Agkistrodon contortrix con- Measured clotting times of coagulation methods
tortrix, 0.4 U/ml) for 270 s. After adding PC Chromo- (PT, Fib, FVII, and FVIII) or optical densities in chro-
genic substrate S2366 (1.5 mg/ml), the change in mogenic (PC) or immunological (FXIII : Ag) methods
OD405 nm was recorded. were automatically converted into final results of spe-
For FVII activity measurement, FVII deficient cific activities or antigen concentrations by using the
plasma was added to plasma sample, diluted 1 : 10 in appropriate calibration curve. International normal-
Factor Diluent solution. After incubation at 37 °C, ized ratio (INR) was calculated by the analyzer using
final measurement was performed after the addition international sensitivity index value (ISI = 0.800) sta-
of RecombiPlasTin. ted for the ACL TOP by the manufacturer. An auto-
For FVIII activity measurement, FVIII deficient matic rerun test with different sample dilution,
plasma was mixed with plasma sample, diluted 1 : 10 preprogrammed by the manufacturer, was used in
in Factor Diluent solution, and SynthASil. After incu- cases of levels above or below the measuring limits of
bation at 37 °C, final measurement was performed calibration curves.
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM 29

of rank correlation were used. The agreement


Evaluation procedure
between results obtained on different analyzers (ACL
Evaluation procedure included determination of TOP vs. BCS/BCT) was evaluated according to Bland
within run precision, between run precision, accuracy and Altman (1986).
and correlation study for all tested parameters (PT,
APTT1 and APTT2, Fib, PC, FVII, FVIII, and
RESULTS
FXIII : Ag), precision of calibration curves for PT and
Fib, and throughput study for global assays (PT,
Within run precision
APTT1, and Fib). Within run precision was evaluated
by measuring tested parameters in normal and abnor- The results of within run precision are shown in
mal plasma samples consecutively 20 times (PT, Table 2. CVs obtained for most assays were within the
APTT1, and APTT2, Fib) or 10 times (PC, FVII, FVIII, criteria for acceptance.
and FXIII : Ag), in the same run. Between run preci-
sion was determined by repeated testing of commer-
Between run precision and accuracy
cial normal (HemosIL Normal Control) and abnormal
control plasma samples (HemosIL Low Abnormal Con- As shown in Table 3, CVs for most assays were within
trol, Special Test Level 2) over a period of 10 days. the criteria for acceptance, except for PT in HemosIL
Accuracy was evaluated by calculating the biases from Normal Control (CV = 7.7%) and PC and FVIII in
the target values stated for commercial control plasma Special Test Level 2 (CV = 14.8% and 13.7%, respec-
samples for all assays. For calibration curve precision tively). In the accuracy study, the obtained biases
study, calibration curves were constructed on six sepa- from target values in the commercial normal and
rate occasions by using original applications for PT abnormal control samples ranged from 1.1% to 8.6%,
and Fib calibrations. Criteria for the acceptance of and from 2.3% to 18.2%, respectively.
obtained CVs were <5% for PT, APTT (NCCLS, 1996)
and PC (Woods, Kitchen & Preston, 1999; Meijer,
Precision of calibration curves
Haverkate & Kluft, 2006), <7% for Fib (NCCLS,
2001), and <10% for factor assays (NCCLS, 1997). For The results of precision of calibration curves are
the method comparison study, results obtained on the shown in Table 4. The obtained CVs for all dilution
ACL TOP were compared with those obtained on one
of Dade Behring analyzers: Behring Coagulation Sys-
tem (BCS) and Behring Coagulation Timer (BCT), by Table 2. Within run precision in normal and abnormal
analyzing plasma samples (n = 44–241) with different plasma samples
values over the whole concentration range. Results of Normal Abnormal
FXIII : Ag obtained on the ACL TOP were compared plasma sample plasma sample
with FXIII activities (FXIII : Act) from BCS. The
CV CV
throughput study was performed by using 100 routine
Test n Mean ± SD (%) Mean ± SD (%)
plasma samples with PT/APTT/Fib requests and pro-
cessed on the ACL TOP and BCS. PT
% 20 114.6 ± 2.4 2.1 61.8 ± 1.0 1.6
INR 20 0.94 ± 0.01 0.9 1.28 ± 0.01 1.0
Statistical analysis APTT1 20 31.6 ± 0.5 1.6 50.6 ± 1.0 2.0
(SynthASil), s
The MedCalc program, version 4.10 for Windows, was APTT2 20 28.6 ± 0.4 1.6 56.1 ± 1.3 2.4
used for statistical analysis. In the case of normal dis- (APTT-SP), s
tribution, data were expressed as mean values with Fib, g/l 20 2.9 ± 0.1 3.5 5.6 ± 0.4 7.7
SD, and correlation co-efficient and Student t-test PC, % 10 112.3 ± 1.9 1.7 55.5 ± 2.9 5.2
were performed. When data were not normally dis- FVII, % 10 108.2 ± 3.9 3.6 34.4 ± 1.1 3.1
FVIII, % 10 84.2 ± 4.6 5.5 33.6 ± 1.2 3.5
tributed, median values with ranges, Wilcoxon’s
FXIII : Ag, % 10 113.6 ± 5.2 4.6 68.1 ± 2.0 2.9
matched pairs t-test, and the Spearman’s co-efficient
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
30 M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM

Table 3. Between run precision and accuracy in commercial normal and abnormal plasma samples in different runs
over a period of 10 days

HemosIL Normal Control (lot no. HemosIL Low Abnormal Special Test Level 2
N0957778) Control (lot no. N0957718) (lot no. N0957718)

Bias from Bias from Bias from


target CV target CV target
Test Mean ± SD CV (%) value (%) Mean ± SD (%) value (%) Mean ± SD (%) value (%)

PT
% 104.9 ± 8.1 7.7 6.0 30.1 ± 0.01 3.3 3.3 – – –
INR 0.98 ± 0.03 3.0 3.0 2.01 ± 0.1 3.1 )2.9 – – –
APTT1 (SynthASil), s 28.0 ± 0.9 3.3 )1.1 47.4 ± 1.3 2.7 )2.3 – – –
APTT2 (APTT-SP), s 31.3 ± 1.2 3.9 3.0 52.3 ± 2.4 4.6 8.3 – – –
Fib, g/l 3.3 ± 0.1 3.4 5.1 2.6 ± 0.2 6.5 5.7 – – –
PC, % 92.9 ± 1.9 2.0 )5.2 – – – 26.0 ± 3.8 14.8 18.2
FVII, % 88.6 ± 4.8 5.4 )8.6 – – – 24.0 ± 1.9 7.9 4.3
FVIII, % 87.9 ± 5.2 5.9 )6.4 – – – 29.3 ± 4.0 13.7 8.5
FXIII : Ag, % 73.3 ± 3.2 4.3 )3.6 – – – 31.8 ± 1.7 5.3 2.6

The agreement between results obtained on differ-


Table 4. Precision of the calibration curves for PT and
Fib ent analyzers is demonstrated in difference-plots,
according to Bland and Altman, which show differ-
Calibration ences between results over the whole concentration
plasma (%) PT (%) Fib (g/l) Mean ± SD (s) CV (%) range (Figure 1). Despite the good correlation co-effi-
cient for PT% and low mean difference of 0.8%, there
PT 100 100.0 – 12.4 ± 0.2 1.3
50 50.0 – 19.6 ± 0.3 1.6 was an important dispersion of differences (±1.96
25 25.0 – 35.1 ± 0.8 2.3 SD = 24.9%) in PT% testing (Figure 1a). The overall
Fib 150 – 4.35 12.6 ± 0.6 4.8 mean difference for PT INR was 0.16. For higher PT
100 – 2.90 16.7 ± 0.8 5.0 INR values obtained on BCS, there was a trend
60 – 1.74 23.9 ± 1.2 5.0
towards lower PT INR values on the ACL TOP
40 – 1.16 33.0 ± 2.1 6.3
30 – 0.87 41.2 ± 2.9 7.0 (Figure 1b). In APTT testing (Figure 1c,d), similar
mean differences were found for APTT1 (0.10) and
APTT2 ()0.08). Mean difference for Fib was )0.88 g/l,
with wide dispersion of differences (±1.96SD = 1.5 g/
points of PT and Fib calibration curves were within l), and with 86% higher results on the ACL TOP than
the criteria for acceptance. Higher, but uniform CVs on BCS (Figure 1e). Mean differences for FVII
were obtained for the precision of Fib calibration and FVIII were )4.6% and 3.4%, respectively
curve, with increasing tendency in higher dilutions of (Figure 1g,h). For FVII, nonhomogeneous distribution
calibration plasma. of differences was shown, with higher negative devia-
tions for higher levels of FVII on the BCT.
The results obtained in comparison study according
Method comparison study
to reference intervals for the ACL TOP (reference
The results of correlation study between the ACL intervals proposed by the manufacturer, except for
TOP and BCS/BCT are shown in Table 5. The PT%) and for the BCS/BCT (laboratory specific refer-
obtained correlation co-efficients were between ence intervals) are presented in Table 6. The number
0.709 and 0.955. Statistically significant difference in of discrepant results within and outside the reference
results was obtained for PT INR, APTT1, APTT2, Fib, intervals on the BCS/BCT was from 9.4% to 56.8%
and PC. and from 5.4% to 50%, respectively. A total number
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM 31

Table 5. Results of correlation study between the ACL TOP and BCS/BCT analyzers

Test n ACL TOP BCS/BCT P-value r

PT
% 241 80.0 (8.0–143.8) 87.0 (6.0–131.0) 0.247 0.913
INR 93 1.88 (0.94–5.41) 1.63 (0.91–6.65) 0.037* 0.899
APTT1 (SynthASil), ratio 208 0.95 (0.63–3.87) 0.98 (0.74–4.28) <0.001* 0.730
APTT2 (APTT-SP), ratio 109 1.14 (0.73–4.01) <0.001* 0.709
Fib, g/l 201 4.6 (0.4–11.2) 3.5 (0.4–10.6) <0.001* 0.923
PC, % 59 87.6 (18.8–186.9) 91.9 (15.0–173.8) 0.002* 0.955
FVII, % 44 87.8 ± 32.0 83.2 ± 29.8 0.150 0.775
FVIII, % 80 73.1 (3.1–253.4) 83.5 (1.0–303.0) 0.516 0.944
FXIII : Ag, % 66 89.5 ± 36.7 – 0.153 0.928
FXIII : Act, % 66 – 92.0 ± 34.8

In the case of normal distribution – data expressed as mean values with SD; P-values obtained using Student t-test for
paired samples; r, correlation co-efficient. In the case of non-normal distribution – data expressed as median values
with ranges; P-values obtained using Wilcoxon’s matched pairs T-test; r, Spearman’s co-efficient of rank correlation.
*Statistically significant.

of discrepant results ranged from 9.7% to 42.4%. put of samples and by reliability of coagulation assays
When analyzing APTT, a total of 33.7% of discrepant (Monagle et al., 2006). Currently, there are a number
results for APTT1 and 30.3% for APTT2 were of different automated coagulation analyzers that are
obtained, with a similar number of discrepant results commercially available and designed for testing a
within and outside the reference intervals. The high- large number of samples by combining different
est number of discrepant results was obtained for PC technologies.
and FVII (56.8% and 50%, respectively), although the In the present study, the coagulation testing system
mean differences were not high (4% and )4.6%, composed of the ACL TOP analyzer and HemosIL
respectively; Figure 1). FXIII : Ag results obtained on reagent group was evaluated. The evaluation included
the ACL TOP were correlated to FXIII : Act from the clotting (PT, APTT, Fib, FVII, and FVIII), chromogenic
BCS. Results and regression line (Y = 0.98X ) 0.34) (PC) and immunological (FXIII : Ag) assays. To our
are presented in Figure 2. knowledge, it is the first evaluation of the coagulation
testing system, which included the ACL TOP analyzer
and HemosIL reagents, calibrators and controls.
Throughput study
Concerning within- and between-run precision,
The results of throughput study are presented in the obtained CVs were within the criteria for accep-
Table 7. The total time needed to complete all analy- tance in most evaluated assays, and results are similar
ses on the ACL TOP and on the BCS was 3 h 20 min to already published evaluations of other fully-auto-
and 2 h 9 min, and the calculated number of tests mated coagulation analyzers with similar characteris-
performed per hour was 151 and 212, respectively. tics, such as the STAR (Flanders et al., 2002), and CA
7000 (Fischer et al., 2006; Dorn-Beineke et al., 2005),
as well as of the ACL TOP (Appert-Flory et al., 2007;
DISCUSSION
Eschwège, Catillon & Robert, 2006).
Nowadays coagulation laboratories are faced with an As there is little information available concerning
increasing number of requests for different coagula- accuracy of measurements within the field of hemo-
tion assays in daily laboratory practice. Automation in stasis, the obtained results are difficult to interpret.
coagulation testing has enabled laboratories to func- However, PT INR, APTT1, Fib, and PC fulfilled the
tion more cost-effectively by improving the through- desirable or minimum goals for biases given by Ricós
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
32 M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM

(a) 40 (b) 3.0


30 2.5
+1.96 SD

PT INR BCS - ACL TOP


PT (%) BCS - ACL TOP
20 24.1 2.0
10
1.5
Mean
0 +1.96 SD
- 0.8 1.0
–10 1.04
0.5
–20 Mean
–1.96 SD
–30 0.0 0.16
–25.7
–40 –0.5 –1.96 SD
–50 –1.0 –0.73
0 20 40 60 80 100 120 140 0 1 2 3 4 5 6 7
AVERAGE of PT (%) BCS and ACL TOP AVERAGE of PT INR BCS and ACL TOP

(c) 3.0 (d) 1.5


APTT1 (ratio) BCS - ACL TOP

APTT2 (ratio) BCS - ACL TOP


2.5
1.0
2.0

0.5 +1.96 SD
1.5
0.43
1.0 Mean
0.0
+1.96 SD
0.63 –0.08
0.5
Mean 0.10 –0.5 –1.96 SD
0.0
–1.96 SD –0.58
–0.44
–0.5 –1.0
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5
AVERAGE of APTT1(ratio) BCS and ACL TOP AVERAGE of APTT2 (ratio) BCS and ACL TOP

(e) 1.5 (f) 25


+1.96 SD
1.0 20
+1.96 SD 20.9
Fib (g/L) BCS - ACL TOP

PC (%) BCS - ACL TOP

0.5 15
0.62
0.0 10
–0.5 5 Mean
Mean 4.0
–1.0 0
–0.88
–1.5 –5
–2.0 –10 –1.96 SD
–1.96 SD
–2.5 –15 –13.0
–2.39
–3.0 –20
0 2 4 6 8 10 12 0 20 40 60 80 100 120 140 160 180 200
AVERAGE of Fib (g/L) BCS and ACL TOP AVERAGE of PC (%) BCS and ACL TOP

(g) 100 (h) 100


80 +1.96 SD
FVII (%) BCS - ACL TOP

FVIII (%) BCS - ACL TOP

60 50
55.7
40 +1.96 SD
Mean
36.3 0
20 3.4
0 Mean –1.96 SD
–50
–20 –4.6 –48.9

–40 –1.96 SD –100


–60 –45.5

–80 –150
20 40 60 80 100 120 140 160 180 0 50 100 150 200 250 300
AVERAGE of FVII (%) BCS and ACL TOP AVERAGE of FVIII (%) BCS and ACL TOP

Figure 1. Comparison performed according to Bland and Altman of test results obtained on the ACL TOP and BCS/
BCT analyzers. (a) PT%, (b) PT INR, (c) APTT1, (d) APTT2, (e) Fib, (f) PC, (g) FVII and (h) FVIII.

Ó 2007 The Authors


Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM 33

Table 6. Discrepancy between test results obtained on the ACL TOP and BCS/BCT analyzers (according to reference
intervals)

No. discrepant test results (n, %) on ACL TOP

Results outside
Reference Reference Results within the the reference
intervals for intervals for reference interval interval on
Test ACL TOP* BCS/BCT† n on BCS/BCT BCS/BCT Total

PT
%† 70.0 70.0 241 17/147 (11.6) 11/94 (11.7) 28/241 (11.6)
INR‡ 2.00–3.50 2.00–3.50 93 5/19 (26.3) 4/74 (5.4) 9/93 (9.7)
APTT1 (SynthASil), ratio 0.89–1.11 0.84–1.16 208 45/134 (33.6) 25/74 (33.8) 70/208 (33.7)
APTT2 (APTT-SP), ratio 0.84–1.16 109 20/62 (32.3) 13/47 (27.7) 33/109 (30.3)
FIB, g/l 2.4–5.0 1.8–4.1 201 18/113 (15.9) 12/88 (13.6) 30/201 (14.9)
PC, % 93.9–158.7 70.0–140.0 59 21/37 (56.8) 4/22 (18.2) 25/59 (42.4)
FVII, % 50.0–129.0 70.0–120.0 44 3/32 (9.4) 6/12 (50) 9/44 (20.4)
FVIII, % 50.0–150.0 60.0–180.0 80 7/41 (17.1) 5/39 (12.8) 12/80 (15.0)

Results are expressed as the number (n) and percentage (%) of discrepant test results within the reference interval on
the BCS/BCT, results outside the reference interval on the BCS/BCT, and the total number of discrepant test results.
*Reference intervals proposed for the ACL TOP by the manufacturer.
†Laboratory specific reference intervals.
‡For PT INR, results were analyzed according to therapeutic range.

220
Table 7. Results of the throughput study (PT/APTT/Fib,
200
n = 100)
180
FXIII:Ag (%) ACL TOP

160
ACL TOP BCS
140
120
Preanalytical time (h:min)* 1:06 0:35
100
Time to first result (h:min) 0:03 0:08
80
Time to last result (h:min) 1:59 1:25
60
Total time (h:min)† 3:20 2:09
40
No. samples performed per hour 50 71
20
20 40 60 80 100 120 140 160 180 200 No. tests performed per hour 151 212
FXIII: Act (%) BCS
*Comprise time to prepare the analyzer, (and reconstitu-
tion and stabilization of) reagents and control samples.
Figure 2. Correlation of the HemosIL FXIII Antigen on
†Comprise preanalytical time, analytical time and time
the ACL TOP with the Berichrom FXIII activity on
for maintenance and shutdown of the analyzers.
the BCS (Y = 0.98X ) 0.34, r = 0.928).

et al. (1999) and Meijer et al. (2006), while minimum dilution point of 25%. As fibrinogen concentration in
goals for PT%, APTT2, FVII, and FVIII could not be this dilution point was 0.725 g/l, and as low fibrino-
achieved in at least one control sample. For FXIII, gen concentration is known to be a limiting factor for
there is until now no published data concerning bio- PT measurement, it was not possible to extend the
logical variation (Ricós et al., 1999). calibration furthermore, making PT results below 25%
Although satisfactory precision of PT and Fib cali- questionable.
bration curves was obtained (Table 4), the disadvan- Despite good correlation between results obtained
tage of the proposed PT calibration was the last on the ACL TOP and BCS/BCT, statistically significant
Ó 2007 The Authors
Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
34 M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM

difference was established for PT INR, APTT1, APTT2, uncertain. Moreover, nine of 18 (50%) severe
Fib, and PC. Besides the correlation, which is a mea- hemophilia A patients would not be properly classi-
sure of relationship between the test results of two fied (data not shown), which is in concordance with
different methods, the testing of agreement, that Barrowcliffe (2004) who stressed that the lower
includes numerical identity between the test results of limit of a one-stage method (1%) may be difficult
two different methods, is also necessary to perform as to achieve with some reagents.
it provides additional information (Meijer, 2003). The Although FXIII : Act assay has been the method of
agreement between results analyzed using difference- choice, HemosIL FXIII : Ag, as an immunoturbidi-
plots (Figure 1) showed that mean differences could metric assay for measuring the subunit A of the FXIII
be of clinical relevance, i.e., in case of PT INR this dif- tetramer, offers new opportunities for FXIII measure-
ference may lead to inappropriate anticoagulant dos- ment. In our study, this assay showed good perfor-
age decisions. mances in precision, accuracy, as well as correlation
A high number of discrepant results obtained for with FXIII : Act. The investigation of agreement could
APTT, PC and FVII despite the low mean difference not be performed because of nonavailability of any
indicate that the reference intervals proposed by the other assay for FXIII : Ag measurement in our labora-
manufacturer for these assays on the ACL TOP seem tory.
to be inappropriate, and that new reference intervals During evaluation, we observed some potential
should be determined. advantages and disadvantages of the evaluated sys-
The Scientific Subcommittee on FVIII and FIX of tem. The analyzer enables the real time monitoring
the Scientific and Standardization Committee of the of each sample status on board together with the
International Society on Thrombosis and Haemosta- possibility of inspection of every single stored reac-
sis has recently recommended the classification of tion curve. Although the software offers the over-
severe, moderate and mild hemophilia A on the view of reagent volume on board, it is not possible
basis of biological FVIII levels of <1, 1–5, and >5%, to see how much reagent is needed to process the
respectively, rather than on the clinical severity of job list. As a consequence, in the case when there
affected individuals (Preston et al., 2004; White et al., is not enough volume of any component needed for
2001). According to that, coagulation laboratory has the reaction, the instrument gives the warning but
a vital role in the diagnosis and classification of proceeds with the procedure and consumes other
hemophilia patients, and results of determinations components not completing the analysis. The limita-
should be both precise and accurate (Chuansumrit, tion of the software is also that all relevant data for
McCraw & Preston, 2004). In our study, we reagents, controls and calibrators must be entered
encountered problems with FVIII determination on manually by the operator.
the ACL TOP in plasma samples with FVIII<10%. Although the ACL TOP was found to have a num-
We observed that the analyzer automatically remea- ber of performances similar to other modern auto-
sured all samples with FVIII activities below 10%, mated coagulation analyzers in regard to its ability to
using different dilution of plasma and factor diluent analyze a high number of samples, the obtained
on the same calibration curve, although the stated throughput that was by 29% lower than for the BCS,
linearity and test range for FVIII is 1–150% and together with inappropriate package volumes for some
0.8–150%, respectively. However, the analyzer gen- special tests makes the analyzer suitable rather for
erated messages with both results, making them mid-sized laboratories.

Ó 2007 The Authors


Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35
M. MILOS ET AL. EVALUATION OF ACL TOP – HEMOSIL REAGENT SYSTEM 35

REFERENCES CA 7000. Thrombosis Research 117, NCCLS Document H30-A2 (ISBN 1-


721–729. 56238-439-2). NCCLS, Wayne, PA.
Appert-Flory A., Fischer F., Jambou D. & Flanders M.M., Crist R., Safapour S. & Preston F.E., Kitchen S., Jennings I.,
Toulon P. (2007) Evaluation and perfor- Rodgers G.M. (2002) Evaluation and Woods T.A.L. & Makris M. (2004) SSC/
mance characteristics of the automated performance characteristics of the STA- ISTH classification of hemophilia A: can
coagulation analyzer ACL TOP. Throm- R coagulation analyzer. Clinical Chem- hemophilia center laboratories achieve
bosis Research 120, 733–743. istry 48, 1622–1624. the new criteria? Journal of Thrombosis
Barrowcliffe T.W. (2004) Monitoring hae- Meijer P. (2003) Method comparison: cor- and Haemostasis 2, 271–274.
mophilia severity and treatment: new relation is not enough to decide about Ricós C., Alvarez V., Cava F., Garcia-Lario
or old laboratory tests? Haemophilia agreement and clinical accuracy. J.V., Hernandez A., Jimenez C.V., Min-
10(Suppl. 4), 109–114. Thrombosis and Haemostasis 90, 555– chinela J., Perich C. & Simon M. (1999)
Bland J.M. & Altman D.J. (1986) Statisti- 556. Current database on biological varia-
cal methods for assessing agreement Meijer P., Haverkate F. & Kluft C. (2006) tion: pros, cons and progress. Scandina-
between two methods of clinical mea- Performance goals for the laboratory vian Journal of Clinical and Laboratory
surement. Lancet 1, 307–310. testing of antithrombin, protein C and Investigation 59, 491–500.
Chuansumrit A., McCraw A. & Preston protein S. Thrombosis and Haemostasis Walenga J.M. & Fareed J. (1994) Auto-
E.F. (2004) Essential issues of labora- 96, 584–589. mation and quality control in the coag-
tory investigation for patients with hae- Monagle P., Barnes C., Ignjatovic V., ulation laboratory. Clinics in Laboratory
mophilia and bleeding disorders. Furmedge J., Newall F., Chan A., De Medicine 14, 709–728.
Haemophilia 10, 105–108. Rosa L., Hamilton S., Ragg P., Robinson White G.C., Rosendaal F., Aledort L.M.,
Clauss A. (1957) Gerinnungsphysiologi- S., Auldist A., Crock C., Roy N. & Row- Lusher J.M., Rothschild C. & Ingerslev
sche Schnellmethode zur Bestimmung lands S. (2006) Developmental haemo- J. (2001) Definitions in Hemophilia:
des Fibrinogens. Acta Haematologica 17, stasis. Thrombosis and Haemostasis 95, Recommendations of the Scientific Sub-
237–246. 362–372. committee on FVIII and FIX of the SSC
Dorn-Beineke A., Dempfle C.E., Bertsch NCCLS (1996) One-Stage Prothrombin of ISTH. Thrombosis and Haemostasis
T. & Wisser H. (2005) Evaluation of the Time (PT) Test and Activated Partial 85, 560.
automated coagulation analyzer Sysmex Thromboplastin Time (APTT) Test; Woods T.A.L., Kitchen S. & Preston F.E.
CA-7000. Thrombosis Research 116, Approved Guideline. NCCLS Document (1999) Quality assessment of haemo-
171–179. H47-A (ISBN 1-56238-301-9). NCCLS, static assays and external quality assess-
Eschwège V., Catillon C. & Robert A. Wayne, PA. ment schemes. In: Laboratory
(2006) Evaluation of the automated NCCLS (1997) Determination of Factor Techniques in Thrombosis – A Manual
coagulation analyser ACL TOP (Instru- Coagulant Activities; Approved Guide- (eds J. Jespersen, R.M. Bertina & F.
mentation Laboratory). Annales de Bio- line. NCCLS Document H48-A (ISBN 1- Haverkate), 2nd revised edition of
logie Clinique 64, 259–264. 56238-321-3). NCCLS, Wayne, PA. ECAT assay procedures, pp. 29–36. Klu-
Fischer F., Appert-Flory A., Jambou D. & NCCLS (2001) Procedure for the Determi- wer Academic Publishers, Dordrecht,
Toulon P. (2006) Evaluation of the nation of Fibrinogen in Plasma; The Netherlands.
automated coagulation analyzer Sysmex Approved Guideline-Second Edition.

Ó 2007 The Authors


Journal compilation Ó 2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 26–35

You might also like