You are on page 1of 8

DOI 10.

1515/cclm-2013-1090      Clin Chem Lab Med 2014; 52(7): 973–980

Koen Hens, Mario Berth, Dave Armbruster* and Sten Westgard

Sigma metrics used to assess analytical quality


of clinical chemistry assays: importance of the
allowable total error (TEa) target

Abstract Koen Hens and Mario Berth: Algemeen Medisch Laboratorium


(AML), Antwerpen, Belgium
Sten Westgard: Westgard QC, Madison, WI, USA
Background: Six Sigma metrics were used to assess the
analytical quality of automated clinical chemistry and
immunoassay tests in a large Belgian clinical laboratory
and to explore the importance of the source used for esti- Introduction
mation of the allowable total error. Clinical laboratories
are continually challenged to maintain analytical quality. Analytical quality is a prerequisite for the clinical labora-
However, it is difficult to measure assay quality objectively tory, but it can be difficult to assess it. Six Sigma metrics,
and quantitatively. hereafter referred to as Sigma metrics, have been used to
Methods: The Sigma metric is a single number that esti- measure quality in an objective and quantitative manner,
mates quality based on the traditional parameters used in combining the three traditional elements used to evaluate
the clinical laboratory: allowable total error (TEa), preci- assay performance: the allowable total error (TEa), bias
sion and bias. In this study, Sigma metrics were calculated and precision. Both IVD manufacturers and clinical labo-
for 41 clinical chemistry assays for serum and urine on five ratories have implemented the Sigma metrics approach,
ARCHITECT c16000 chemistry analyzers. Controls at two leading to reduced operational defects in the laboratory
analyte concentrations were tested and Sigma metrics and/or quantitation of test performance quality [1–4]. A
were calculated using three different TEa targets (Ricos higher Sigma metric value means fewer analytical errors
biological variability, CLIA, and RiliBÄK). and fewer questionable test results are accepted and
Results: Sigma metrics varied with analyte concentration, reported, and fewer acceptable test results are falsely
the TEa target, and between/among analyzers. Sigma rejected and not reported [5]. A Six Sigma assay is one for
values identified those assays that are analytically robust which 99.99966% of results are error free, corresponding
and require minimal quality control rules and those that to 3.4 defects per million opportunities, in this case, assay
exhibit more variability and require more complex rules. results. Sigma metrics can be translated into appropriate
The analyzer to analyzer variability was assessed on the and robust quality control rules thus improving overall
basis of Sigma metrics. laboratory quality. As a result, the laboratory can maxi-
Conclusions: Six Sigma is a more efficient way to control mize efficiency and reduce control costs through fewer
quality, but the lack of TEa targets for many analytes and re-runs and work-arounds.
the sometimes inconsistent TEa targets from different The objective of this study was to evaluate the perfor-
sources are important variables for the interpretation and mance of 41 chemistry assays on five Architect c16000®
the application of Sigma metrics in a routine clinical labo- instruments in terms of Sigma metrics.
ratory. Sigma metrics are a valuable means of comparing
the analytical quality of two or more analyzers to ensure
the comparability of patient test results. Materials and methods
Keywords: allowable total error (TEa); quality; Sigma Analyzers
metrics.
Five ARCHITECT c16000 clinical chemistry analyzers (Abbott Diag-
nostics, Chicago, IL, USA) were used. The c16000 is a general pur-
*Corresponding author: Dave Armbruster, Abbott Diagnostics pose, high volume instrument that uses spectrophotometry for a
Division, Department 09AA, Building CP1-1S, Abbott Park, IL 60064, variety of absorbance/colorimetric and immunoturbidimetric assays
USA, E-mail: David.Armbruster@abbott.com and ion-specific electrodes (ISE) for electrolytes (Na, K, Cl). Three

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
974      Hens et al.: Sigma metrics

of the c16000s were used to test routine clinical chemistry analytes are based on replicate analysis performed by the manufacturer and/
and selected high volume-specific protein analytes. The other two or independent laboratories using Architect analyzers and reagents.
c16000s were used for low volume assays and special tests. The sys-
tems were maintained and used as per the manufacturer’s operations
manual. Testing was conducted in February, 2010.
Allowable total error
Tolerance limits, in the laboratory, are best expressed as an allow-
Assays able total error (TEa) specification. TEa is a model that combines both
imprecision and bias (trueness) of a method to calculate the impact
Only assays for which TEa goals from at least two independent
on a test result. The Sigma metrics were calculated using TEa goals
sources were available were included in this study. Thirty-three serum
from three sources in order to understand the effect of TEa on esti-
clinical chemistry tests were included: albumin BCG (bromocresol
mates of Sigma metrics: 1) CLIA (Clinical Laboratory Improvements
green method), alkaline phosphatase, alanine aminotransferase
Amendments) from the US; 2) RiliBÄK (German Medical Council for
(ALT, without pyridoxal-5-phosphate), amylase, aspartate ami-
the Quality Assessment of Quantitative Analyses in Medical Labora-
notransferase (AST, without pyridoxal-5-phosphate), bicarbonate,
tories, 2008 version; the inter-lab or “Ring Trials” values, in contrast
direct bilirubin, total bilirubin, calcium (Ca), chloride (Cl), choles-
to the intra-lab values); and 3) the Ricos biological variability data-
terol, creatinine (alkaline picrate Jaffé method), creatine kinase (CK),
base (desirable target values, in contrast to the minimal or optimal
CRP (Standard range and Vario 32), γ glutamyltransferase (GGT), glu-
target values) [7, 8]. These three sources are regularly updated and
cose, high density lipoprotein (direct HDL), immunoglobulin A (IgA),
can be freely accessed through http://www.westgard.com.
immunoglobulin G (IgG), immunoglobulin M (IgM), iron, lactic acid,
lactate dehydrogenase (LDH, IFCC reference method formulation),
lipase, magnesium (Mg), phosphorus, potassium (K), total protein,
sodium (Na), transferrin, triglycerides, urea and uric acid. Two thera-
peutic drugs in serum (carbamazepine and lithium) and six specific
Sigma metric calculation
urine assays were included: Ca, creatinine, phosphorus, protein, Sigma metrics were calculated using the standard equation:
urea, and uric acid. All reagents were obtained from Abbott and used
as per the manufacturer’s package inserts. Sigma metric = (TEa–bias)/precision [all values expressed as
percent (%)].

Three sets of Sigma metrics were calculated, each using different


Controls TEa targets (providing that more than one set of TEa targets were
Bio-Rad (Bio-Rad, Marnes-la-Coquette, France) controls used were: available for an analyte) and Sigma metrics were calculated for
1) Lyphochek Assayed Chemistry Control material (two levels, lyo- both of the two control concentrations. For the more common (high
philized); 2) Lyphochek Immunology Plus (two levels, lyophilized); volume) analytes, three Sigma metrics were calculated for each
and 3) Liquichek Urine Chemistry Control (ready to use, stored at control concentration, one for each of the three Architect c16000
2–8 °C, non-frozen). The manufacturer’s package inserts were fol- analyzers testing that analyte. For the less common (low volume)
lowed. Controls were mixed thoroughly prior to use. Open vial con- analytes, two Sigma metrics were calculated for each control con-
trols were handled as per the manufacturer’s directions and repeat centration, one for each of the two Architect c16000 analyzers test-
freeze/thaw cycles were avoided. Controls were tested within 30 min ing that analyte.
of being placed on the analyzers. Calibration curves were verified as
valid when concentrations of the two Bio-Rad Lyphochek Assayed
Chemistry Control levels were within the listed validity ranges. All
runs subsequent to the initial calibration included two control levels Results
to verify acceptability of the analysis. Data were not accepted if a con-
trol was out of the specified acceptance range.
Complete Sigma metrics for 41 assays are shown in
Table 1. Using RiliBÄK TEa targets, in general the Sigma
Precision values are very acceptable and indicate robust, high
quality results, ranging from about 5 or 6 up to a high
Precision, expressed as coefficient of variation (% CV), was deter-
value of about 30. The Sigma for Control 1 for Ca ranged
mined for all assays using a 20-day protocol performed according
to CLSI (formerly NCCLS) guideline EP5-A2 [6]. The controls for each only from 0 to 1 due to the high bias, about 10% for a
assay were tested twice a day, with a minimum of 2 h separating the target value of 2.02 mmol/L, but the Sigma values for
analytical events. One reagent lot was used for each assay. Control 2 ranged from 8 to 13. Lithium Sigma values
for both controls ranged from 1.7 to 3.1 due to very high
bias values. Control 2 creatinine Sigma metrics ranged
Bias from only 0.2 to 0.9 due to very high biases at a target
Bias was estimated based on the difference of the average of observed
value of 5.63 mg/dL. Potassium Control 2 (target value
results for each analyte from the target values provided in the Bio- 5.67  mmol/L) Sigma values were only 2.6–2.9. By and
Rad control package inserts. Target values for the Bio-Rad controls large, the Sigma metrics between two or among three

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
Table 1 Sigma metrics using three different sources for TEa target values (RiliBÄK, biological variability, CLIA).

Assay   Control 1   Control 2   Control 1 Sigma   Control 2 Sigma   Control 1 Sigma   Control 2 Sigma   Control 1 Sigma CLIA   Control 2 Sigma CLIA
target value target value RiliBÄK RiliBÄK biological variability biological variability
a a a
Albumin   39 g/L   25 g/L   27/27.5/19.5   31.7/32/32   //   1.3/a/a   8.5/9.0/6.2   12.8/12/12
Alk. Phosph.   115 U/L   470 U/L   3.9/4.2/5.3   9.2/12/12.9   1.7/1.9/2.6   4.8/6.5/7.1   6.2/6.4/7.9   13.3/17.4/18.5
ALT   34.6 U/L   101 U/L   7.8/7.8/7.4   8.1/11.1/10.8   12.2/12.3/11.2   13.0/21.6/17.6   7.4/7.4/7.0   7.6/13.0/10.2
Amylase   66.1 U/L   411 U/L       5.0/6.8/5.5   9.1/9.8/7.9   11.2/15.3/11.8   19.5/21.3/21.2
AST   35.9 U/L   190 U/L   8.8/9.6/9.5   14.4/19.8/19.9   6.2/6.6/6.8   7.6/4.9/13.0   8.4/9.1/9.1   13.5/18.7/18.7
a a a a a a a a a
Bicarbonate   34.4 mmol/L   16.3 mmol/L       //   //   1.1/1.7/a   //
Bilirubin, direct   0.436 mg/dL   1.93 mg/dL       15.0/21.4/22.2   25.2/31.2/28.9    
Bilirubin, total   0.838 mg/dL   4.29 mg/dL   9/7.7/6.4   8.8/9.7/10.3   12.9/10.9/9.1   12.9/14.2/15.1   8.2/7.0/5.8   7.8/8.7/9.3
a a a
Ca   2.02 mmol/L   3.03 mmol/L   0.3/0/1.0   12.3/13.2/8.1   //   0.6/0.9/0    
Ca (urine)   0.86 mmol/L   2.75 mmol/L   6.3/6   9.1/14   17.8/16.8   18.8/28.5    
Carbamazepine   3.47 mg/L   15.1 mg/L   6.1/3.8   6.94/5.4       8.3/5.4   8.9/7.1
Cholesterol   249 mg/dL   97.1 mg/dL   20.5/17.9/17.1   19/16.9/15.8   12.4/10.5/10.2   11.1/10.5/9.7   15.1/12.9/12.5   13.7/12.7/11.7
a a a
Cl   103 mmol/L   83.1 mmol/L   9.3/9.6/8   7/7.1/5.7   //   1.1/0.8/0.4   5.0/5.0/3.9   5.9/4.2/3.2
a a a a a a
Creat.   2.25 mg/dL   5.63 mg/dL   8.9/6.1/9   0.7/0.9/0.2   3.3/2.4/3.6   //   6.6/4.6/6.7   //
Creat. (urine)   83.4 mg/dL   243 mg/dL   9.9/6.3   7.6/5.2   18.0/8.9   10.9/7.5    
CK   148 U/L   502 U/L   12.4/15.6/15.4   17.6/8.4/14.6   19.8/24.7/23.9   26.8/38.9/36.1   19.6/24.5/23.7   26.6/38.5/35.8
CRP   8.3 mg/L   49.8 mg/L   10.4/9.9/10.2   7.7/10.1/12.6   37.7/38.2/31.5   25.2/34.2/41.4    
GGT   65.1 U/L   161 U/L   17.8/22.9/11.3   18.5/12.6/10.5   18.8/24.2/12.6   19.6/13.4/11.1    
Glucose   93.5 mg/dL   297 mg/dL   14.5/13.1/12.4/   10.5/13.8/14.2/   6.4/5.6/5.7/3.3/8.6   4.5/5.3/5.4/3.9/7.0   9.5/8.4/8.2/5/13   6.8/8.5/8.8/8.9/10.6
7.5/20.2 9.8/12.5
HDL   64.4 mg/dL   32.2 mg/dL       5.7/9.3/6.4   3.3/2.9/3.1   18.6/25.2/21.3   10.5/11.2/8.9
IgA   1.16 g/L   3.07 g/L   12.7/19.5   18.5/25.3   7.1/10.4   11.4/16.0    
IgG   8.18 g/L   24.5 g/L   17.9/16.2   11.3/6   6.9/6.8   4.9/1.3   25.6/22.8   15.8/9.3

Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
Brought to you by | Universiteit Gent
IgM   0.536 g/L   1.69 g/L   9.7/9.1   14.3/33.5   5.2/4.3   7.1/18.1    
Iron   302 µg/dL   56.4 µg/dL       29.5/29.5/28.2   23.9/23.7/35.7   18.8/18.8/18.1   8.4/15.2/13.6
Lactic acid   445 mg/L   N/A   4.2/6.2   N/A   8.8/13.4   N/A    
LDH   181 U/L   362 U/L   5.2/6.5/5.7   14.9/19.3/22.5   2.9/3.9/3.1   8.6/10.5/13.5   5.8/7.3/6.5   16.9/21.9/25.2
Lipase   29.4 U/L   69.3 U/L   12.7/10.2/8.2   8.4/7.6/9.5   14.8/12.3/9.7   10.3/9.4/11.8    
Lithium   0.69 mmol/L   2 mmol/L   2.7/1.7   1.7/3.1       5.8/4.1   4.6/6.1
Mg   0.8 mmol/L   1.88 mmol/L   9.8/7.2/5.8   7.8/7.9/6.7   2.9/2.1/1.7   2.5/2.4/2.1   16.7/12.2/16.1   13.1/13.2/11.3
Phosphorus   1.05 mmol/L   2.28 mmol/L   7.9/11.3/9.8   19/24/17.3   4.6/6.8/5.9   11.6/15.2/10.5    
Phosphorus   12.5 mmol/L   23.7 mmol/L   10.4/9.9   7.7/16.8   16.9/15.7   12.3/26.4    
(urine)
Potassium, K   3.77 mmol/L   5.67 mmol/L   6.2/6.5/3.9   2.6/2.7/2.9   1.9/3.9/2.4   0.9/1.2/1.4    
Protein   68 g/L   42 g/L   16.9/16.7/14.1   15/7/9.3   5.6/5.5/4.4   4.8/1.6/3.1   17.0/16.7/14.1   15.0/7.0/9.3
Protein (urine)   221 mg/L   648 mg/L   5.8/3.5   7.6/13.9   12.3/6.6   13.8/25.3    
a a a
Sodium, Na   146 mmol/L   125 mmol/L   4.9/7.6/7.9   3.2/11.1/6.1   0.1/1.0/0.7   //    
a a a a a a
Transferrin   2.49 g/L   1.65 g/L   6/9.3/9.3   10.5/14.7/2.6   //   //    
Triglycerides   185 mg/dL   89.6 mg/dL   19/17/24   5.9/13.6/15.8   33.3/30.1/42.5   10.6/25.5/29   29.8/26.9/38.0   9.5/22.6/25.7
Hens et al.: Sigma metrics      975
976      Hens et al.: Sigma metrics

analyzers were comparable, although Control 2 Sigma


  Control 2 Sigma CLIA

4.5/4.9/5.4

24.5/21.6/18.9
values for Na showed ­considerable variability: 3.2, 6.1,
and 11.1. It is not surprising that the Sigma metric at one

Results are shown for three or five analyzers, at two different control levels. If no results are shown (blank field), no TEa is available. N/A Target value for control material not available.
control concentration may be very different from that at
another concentration because the precision and/or bias
can change considerably with analyte concentration.
Using biological variability (desirable) TEa targets,



most analytes again exhibited impressive Sigma metrics:


  Control 1 Sigma CLIA

0.9/1.3/1.5

17.4/13.6/13.9

5 or 6 and up to about 40. However, several analytes


produced low Sigma values: albumin, bicarbonate, Ca,
Mg, K, Na and transferrin (both controls); alkaline phos-
phatase, LDH, and HDL, total protein, and IgG (control 2).
It is notable that bicarbonate and the electrolytes exhib-



ited especially poor Sigma values, but the biological


biological variability
Control 2 Sigma

11.0/10.2/11.2
10.7/16.6
17.7/15.7/13.7
12.8/15.7

variability TEa goals are very demanding. Again, Sigma


metrics between two or among three analyzers were fairly
consistent.
Using CLIA TEa targets, once again Sigma metrics for
most assays were very impressive: 5 or 6 and up to about
35. Bicarbonate Sigma values ranged from negative to only




1.7, reflecting the bias for this difficult assay and the 8%
biological variability
Control 1 Sigma

5.1/4.7/6.5
18.3/19.0
12.3/9.5/9.8
7.5/10.1

TEa target. Surprisingly, urea Sigma metrics for control 1


ranged from 0.9 to 1.5 (target: 33.8 mg/dL) and for control
Bias exceeds the TEa target resulting in a negative value for the numerator in the Sigma metric equation.

2 from 4.5 to 5.4 (target: 97.9 mg/dL). The CLIA target for Cl
is 5% and Sigma values were 5, 5, and 3.9 (control 1) and
5.9, 4.2, and 3.2 (control 2). Creatinine Sigma values for




control 1 were respectable (6.6, 4.6, and 6.7) but negative


RiliBÄK
  Control 2 Sigma

15.3/13.5/15
11.2/12.6
  18.6/16.5/14.4
9.3/11.8

for control 2 because of the large bias at the higher con-


centration value. Sigma metrics between two or among
three analyzers were again consistent.
Although Table 1 is informative, displaying the same
data as normalized method decision charts, as pioneered


RiliBÄK

by Westgard, is more visually appealing and allows Sigma


Control 1 Sigma

7.8/6.9/9.7
13.7/14.5
12.9/10.1/10.3
5.7/7.8

metrics for a test at two analyte control concentrations


using three different TEa targets to be readily examined [9,
10]. Three graphs are included as examples, illustrating
especially the importance of the source of the TEa used
to estimate the Sigma metrics. Figure 1 displays Sigma




metrics for albumin. Using either the RiliBÄK or CLIA TEa


target value

97.9 mg/dL

9.14 mg/dL
24.8 mg/dL
760 mg/dL
Control 2

targets, assay quality is world class (Sigma metric   ≥  6) at


both low and high control concentrations, but using the
biological variability TEa, all Sigma metrics are unac-
ceptable. Figure 2 displays Sigma metrics for Cl. Using





target value

33.8 mg/dL

5.16 mg/dL
11.7 mg/dL
395 mg/dL

the RiliBÄK TEa, assay quality is excellent; using the


Control 1

CLIA TEa, assay quality ranges from marginal to good to


excellent; but using the biological variability TEa, assay
(Table 1 Continued)

quality is unacceptable. Figure 3 displays Sigma metrics





Uric acid (urine)  

for glucose. Using either the RiliBÄK or CLIA TEa targets,


Urea (urine)

assay quality is world class, and even using the biological


Uric acid

variability TEa, assay quality ranges from good to excel-


Assay

Urea

lent to world class.


a

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
Hens et al.: Sigma metrics      977

Figure 1 Normalized method decision chart for albumin comparing Figure 3 Normalized method decision chart for glucose comparing
Sigma metrics calculated using biological variability ( ), CLIA ( ) Sigma metrics calculated using biological variability ( ), CLIA ( )
and RiliBÄK ( ) TEa targets. and RiliBÄK ( ) TEa targets.
Data points represent the Sigma value for controls at two concentra- Data points represent the Sigma value for controls at two concentra-
tions and from three different analyzers. tions and from five different analyzers.

Discussion based on the inherent analytical quality of a test can be


formulated to ensure acceptable patient test results are
reported and false rejection of results is minimized. For
Calculating Sigma metrics offers the advantage of setting
example, a Six Sigma assay can be controlled using a
a “quality base line”. Using the Sigma values, readily
simple 1:3s or 1:3.5s control rule, i.e., test results may
calculated by, e.g., EZ Rules software (Westgard QC Inc.,
be accepted and reported if values for two controls fall
Madison, WI, USA), appropriate quality control (QC) rules
within 3–3.5 standard deviations (SD) of the mean value.
Assays with lesser Sigma values require more sophis-
ticated control rules and they can be chosen according
to the observed quality, being customized to match each
test. This is an improvement over the traditional but
simplistic “control ± 2 standard deviations” approach by
which all assays are assumed to be of equal quality and
the same rule applies to all analytes, despite the fact that
some assays are clearly more robust and some are typi-
cally more problematic. Gras et  al. reported calculating
Sigma metrics for 69 assays on the Roche Modular PPE
and Integra using TEa targets from biological variabil-
ity, bias assessed from peer group values, and precision
observed over 2 months [11]. A simple 4s 1:4s rule could be
used for assays with Sigma values  > 9 and a 1:3s rule for
assays with Sigma values  > 6. These authors found that
two thirds of assays could be controlled using control
rules based on biological variability and corresponding
Sigma metrics.
Figure 2 Normalized method decision chart for chloride comparing
However, as clearly illustrated in this study, it must be
Sigma metrics calculated using biological variability ( ), CLIA ( )
and RiliBÄK ( ) TEa targets.
recognized that distinctly different Sigma metrics can be
Data points represent the Sigma value for controls at two concentra- obtained depending on the TEa values selected and using
tions and from three different analyzers. the same bias and precision values in the Sigma equation.

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
978      Hens et al.: Sigma metrics

In this study, precision was based on a typical 20-day paragraph, it is unknown whether these controls are com-
study. An even more robust precision estimate can be mutable. The metrological traceability of the controls is
made from long-term QC data, e.g., over a 6-month or also uncertain. The Bio-Rad materials used are typical
1-year period, and it would be expected that this estimate “precision” controls, as opposed to “trueness” controls,
would result in a larger % CV, reflecting greater variability and are intended to assess assay performance on the basis
over a longer time period, but also % CVs that are more of precision, i.e., % CV, and not accuracy, i.e., bias or true-
typical of long-term analytical performance variation. ness. Trueness controls, like calibrators, must be manu-
Using a precision value calculated for a shorter period of factured following a strict traceability chain, anchored by
time could result in a more optimistic estimate and result established reference materials and/or reference method
in a higher Sigma metric. [13]. The target values of trueness controls are established
The Sigma metrics reported here reflect assay perfor- by direct linkage to “gold standard” reference materials/
mance at the time the data was collected and thus rep- methods and accompanied by uncertainty estimates. This
resent a “snapshot”. Naturally, performance can change is not the case with precision controls as is readily appar-
over time for a variety of reasons (e.g., reagent lot to lot ent by examining the mean values listed in control package
variation). Periodic calculation of Sigma metrics is appro- inserts, which typically vary for an analyte depending on
priate to determine if assay quality has been maintained, the assay manufacturer. Realistic estimates of assay bias/
has decreased, or has improved. Sigma metrics thus rep- trueness require proper metrological standardization of
resent another quality assurance tool to be monitored all field assays and analysis of trueness controls, of which
­periodically to assess changes in assay quality. there are few, which may not be prepared using appropriate
Bias is more difficult to realistically estimate. In this reference materials, which often are not readily available,
study, bias was estimated as the difference between the and which may be prohibitively expensive for typical clini-
target values (mean values) provided for the controls by cal laboratories. In addition, the commutability of trueness
the control manufacturer and the observed mean values. controls and reference materials must be established, and
The control target values were not assigned by reference commutability studies are challenging and rare.
method analysis and it’s uncertain how close they are to Another weakness of this study is that the bias esti-
“scientific truth” as determined by a gold standard method mates were made by comparing the observed results from
as opposed to a routine “field method”. Also, the controls the analyzers to the target values listed in the control
used in this study may not be commutable, that is, their manufacturer’s package insert. In retrospect, it would
analytical response with an assay may not be equivalent have been preferable to estimate the bias as the difference
to that of a fresh patient sample and observed bias may between the observed results and the Architect inter-lab-
be misleading, possibly being greater or smaller than oratory peer group means for each analyte. Bias estimates
that observed with commutable materials and patient based on the peer group means would have better reflected
specimens. Therefore the observed bias is only “relative” the typical bias for the assays as performed in the testing
instead of “absolute”. However, the reality is that labora- laboratory.
tories routinely assess their bias by comparison to assayed The choice of TEa is critical and has a major impact on
control target values and external quality assurance (EQA) the Sigma metric, as clearly illustrated by this study. Three
participant reports, which often use peer group or all par- common sources of TEa were chosen: biological variabil-
ticipant means instead of accuracy-based grading. Freid- ity, CLIA, and RiliBÄK. There are several other sources for
ecky and coworkers critically evaluated EQA acceptance TEa targets and a laboratory must decide which TEa goal
criteria and noted that EQA peer group evaluation was not is most appropriate for it. One has to take into account that
sufficient to determine analytical quality and may com- a TEa goal is not available for every analyte. Also, when
promise desired patient care, although satisfying partici- using biological variability as the basis of TEa, it must be
pant laboratories and manufacturers [12]. These authors noted that there are actually three possible TEa targets for
recommend that EQA results be based on comparison to analytes: minimal, desirable, and optimal.
reference method target values, requiring metrological Biological variability TEa values are often suggested
traceability and assessment of absolute trueness instead as the most stringent, which is generally true, and appro-
of relative bias (i.e., peer group comparison). The biases priate, but their appropriateness is debatable. Taking
used in our study reflect the “real world” conditions of albumin as an example, the TEa targets for RiliBÄK, bio-
regular laboratory operations, even if they are not optimal. logical variability, and CLIA were 20%, 3.9%, and 10%,
A weakness of this study is the use of standard com- respectively. The Sigma metrics under RiliBÄK ranged
mercially available controls. As noted in the previous from about 20 to about 32. With biological variability,

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
Hens et al.: Sigma metrics      979

using exactly the same bias and precision values, the which has the better precision and bias and for which the
Sigma values ranged from negative to 1.3. Using the CLIA bias and/or precision is less than desirable, allowing for
TEa target, Sigma metrics ranged from about 6 to 12. So, troubleshooting and correction of that analyzer.
depending on the TEa chosen, the same albumin assay
would be classified under CLIA as definitely of world class
quality (Sigma 6 or above) or even well above world class
using RiliBÄK, but unacceptable and not “fit for purpose”
Conclusions
using biological variability. But albumin is routinely
In conclusion, the classical Westgard rules need not
tested and apparently with acceptable clinical outcomes,
always be used for every assay. Sigma metrics can be a
suggesting typical albumin assays are “fit for purpose.”
more efficient way to control quality by matching the QC
This suggests the biological variability TEa is perhaps
rules to the analytical quality of each individual assay.
too demanding for analytical performance for some
However, the lack of TEa targets for many analytes and the
typical field assays. Bicarbonate and the electrolytes also
sometimes inconsistent TEa targets from different inde-
exhibited poor Sigma values, undoubtedly reflecting the
pendent sources are a major variable in the interpretation
demanding, and in our opinion unrealistic, TEa targets for
and the application of Sigma metrics. Estimation of Sigma
these analytes using biological variability.
metrics represents an alternative method of evaluating
Although it seems logical to use TEa targets consist-
comparability of analyzers.
ently from the same source, with experience a laboratory
may find it desirable to choose TEa values from different
Acknowledgments: The technical support by Ilse Van
sources, mixing and matching as seems appropriate for
Gysel and Patrick Rosiers was much appreciated. We also
individual assays. Some TEa targets are likely too liberal
thank Frank Heyvaert for his professional cooperation.
and give a falsely optimistic estimate of quality, while
others, in particular those established using biological
variability, are conversely too demanding and yield overly Conflict of interest statement
pessimistic estimates of quality. Currently it is usually up
to laboratory directors to use their best practical and pro- Authors’ conflict of interest disclosure: The authors
fessional judgment to choose appropriate TEa goals. stated that there are no conflicts of interest regarding
Whether a laboratory chooses to set QC on an analyte- the publication of this article. Employment and fees for
specific basis using Sigma metrics, estimation of Sigma lecturing played no role in the study design; in the collec-
metrics to compare analytical between two or among more tion, a
­ nalysis, and interpretation of data; in the writing
than two analyzers is a useful exercise. It is clear that the of the report; or in the decision to submit the report for
patient test results from two or more analyzers if used to publication.
analyze the same sample should be equivalent (within Research funding: None declared.
defined variability limits). Comparison of quantitative Employment or leadership: Mario Berth: has received fees
values for two analyzers can readily determine if they are from Abbott for lecturing. Dave Armbruster: is employed
exhibiting similar performance. However, if results are not by Abbott Diagnostics; receives salary from and holds
equivalent, the question is which analyzer is “right” and stock from Abbott. Sten Westgard: has received fees from
which is exhibiting aberrant performance. If Sigma metrics Abbott for lecturing and preparing educational materials.
are used for comparison analysis, the TEa value is the same Honorarium: None declared.
for both analyzers, and the Sigma metric difference is due
to either the precision or bias, or both. Comparison of the Received December 17, 2013; accepted February 17, 2014; previously
bias and precision of both analyzers will easily establish published online March 8, 2014

References
1. Lasky FD, Boser RB. Designing in quality through design control: 3. Gras JM, Philippe M. Application of the Six Sigma concept in
a manufacturer′s perspective. Clin Chem 1997;43:866–72. clinical laboratories: a review. Clin Chem Lab Med 2007;45:
2. Stankovic AK, Romeo P. The role of in vitro diagnostics 789–96.
companies in reducing laboratory error. Clin Chem Lab Med 4. Llopis MA, Trujillo G, Llovet MI, Tarrés E, Ibarz M, Biosca C,
2007;45:781–8. et al. Quality indicators and specifications for key analytical-

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM
980      Hens et al.: Sigma metrics

extranalytical processes in the clinical laboratory. Five years’ 9. Westgard JO. Internal quality control: planning and
experience using the Six Sigma concept. Clin Chem Lab Med implementation strategies. Ann Clin Biochem 2003;40:593–611.
2011;49:463–70. 10. Westgard JO, Darcy T. The truth about quality: medical
5. Kinns H, Pitkin S, Housley D, Freedman DB. Internal quality usefulness and analytical reliability of laboratory tests. Clin
control: best practices. J Clin Pathol 2013;66:1027–32. Chim Acta 2004;346:3–11.
6. CLSI EP5-A2. Evaluation of precision performance of quantitative 11. Gras JM, Goffinet P, Bormans F. Practical internal QC protocol
measurement methods; approved guideline – 2nd ed. Wayne, PA: based on Six Sigma and biological variation in a routine clinical
CLSI, 2004. chemistry laboratory. Clin Chem 2009;55(Suppl):A32.
7. Fraser CG, Petersen PH, Ricos C, Haeckel R. Proposed quality 12. Friedecky B, Kratochiva J, Budina M. Why do different EQA
specifications for the imprecision and inaccuracy of analytical schemes have apparently different limits of acceptability? Clin
systems for clinical chemistry. Eur J Clin Chem Clin Biochem Chem Lab Med 2011;49:743–5.
1992;30:311–7. 13. Armbruster DA. Accuracy controls: assessing trueness (bias).
8. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, In: Westgard JO, Westgard S, editors. Clinics in laboratory
Jiménez CV, et al. Current databases on biological variation: pros, medicine, quality control in the age of risk management, 1st ed.
cons and progress. Scand J Clin Lab Invest 1999;59:491–500. Vol 33. Philadelphia, PA: Elsevier, 2013:125–37.

Brought to you by | Universiteit Gent


Authenticated | 157.193.96.165
Download Date | 5/27/14 2:27 PM

You might also like