You are on page 1of 5

450014

and PlebaniJournal of Laboratory Automation


2012
JLAXXX10.1177/2211068212450014Lippi

Feature Story
Journal of Laboratory Automation
18(2) 184­–188
Continuous-Flow Automation and © 2012 Society for Laboratory
Automation and Screening

Hemolysis Index:  A Crucial Combination DOI: 10.1177/2211068212450014


jala.sagepub.com

Giuseppe Lippi1 and Mario Plebani2

Abstract
A paradigm shift has occurred in the role and organization of laboratory diagnostics over the past decades, wherein
consolidation or networking of small laboratories into larger factories and point-of-care testing have simultaneously
evolved and now seem to favorably coexist. There is now evidence, however, that the growing implementation of
continuous-flow automation, especially in closed systems, has not eased the identification of hemolyzed specimens since
the integration of preanalytical and analytical workstations would hide them from visual scrutiny, with an inherent risk
that unreliable test results may be released to the stakeholders. Along with other technical breakthroughs, the new
generation of laboratory instrumentation is increasingly equipped with systems that can systematically and automatically
be tested for a broad series of interferences, the so-called serum indices, which also include the hemolysis index.
The routine implementation of these technical tools in clinical laboratories equipped with continuous-flow automation
carries several advantages and some drawbacks that are discussed in this article.

Keywords
hemolysis, hemolyzed specimens, laboratory testing, interference

Errors in Laboratory Diagnostics that is, those comprised within the so-called preanalytical
and postanalytical phases.7,8 The manually intensive activi-
A paradigm shift has occurred in the role and organization ties of the preanalytical phase are indeed the most vulnera-
of laboratory diagnostics over the past decades, which has ble steps, being associated with roughly two-thirds of
been mainly driven by biological discoveries, incessant all errors that can be identified throughout the total testing
implementation of novel diagnostic tests within diagnostic process. As reliably mirrored by distribution patterns,
reasoning, remarkable technological advances, and eco- the vast majority of preanalytical errors occurs as a conse-
nomic pressures.1,2 All these forces have paved the way to quence of mistaken or mishandled procedures while col-
a revolution in the laboratory environment, wherein con- lecting and handling blood specimens, which finally results
solidation or networking of small laboratories into larger in the generation of samples of poor/unsuitable quality
factories and decentralization of some tests (i.e., point-of- for diagnostic testing.9 In terms of relative prevalence, spu-
care testing) at bedside, physician offices, pharmacies, rious hemolysis is by far the largely prevailing cause of
supermarkets, and other health care facilities have simulta- nonconformance (from 40%–70%), ahead of samples with
neously evolved and now seem to acceptably coexist.3,4 insufficient volume (from 10%–20%), those collected in
Contextually with these radical changes, the quality inappropriate containers (from 5%–15%) or showing undue
throughout the total testing process has improved in paral- clotting (from 5%–12%), and those containing labeling
lel, with a substantial reduction in vulnerability and error or identification problems (from ~2%) (Table 1).10,11
rates. Nevertheless, laboratory diagnostics is still not the Irrespective of the large burden, hemolyzed specimens are
safe enterprise it is supposed to be, and errors may occur at
an estimated frequency of ~0.3% of all tests performed.5 1
Clinical Chemistry and Hematology Laboratory, Academic Hospital of
Although most of them are not associated with adverse Parma, Parma, Italy
2
Department of Laboratory Medicine, University of Padova, Padova, Italy
health care outcomes, in ~20% of cases, they might gener-
ate further inappropriate investigations or increase health Received Apr 5, 2012.
care expenditure, whereas in up to 6% of cases, they might
Corresponding Author:
also jeopardize patient safety.6 Giuseppe Lippi, Professor, U.O. Diagnostica Ematochimica, Azienda
Several lines of evidence attest that the vast majority of Ospedaliero-Universitaria di Parma,Via Gramsci, 14, 43126–Parma, Italy.
mistakes occur in the extra-analytical activities of testing, Email: glippi@ao.pr.it, ulippi@tin.it

Downloaded from jla.sagepub.com by guest on February 16, 2016


Lippi and Plebani 185

Table 1. Leading Preanalytical Problems and Available Technology to Address Them.

Preanalytical Problem Prevalence, % Technology


Hemolyzed samples 40–70 Hemolysis index
Insufficient samples 10–20 Liquid-level sensors
Inadequate container 5–15 Optic detection of tube or color of the stopper
Undue clotting 5–12 Liquid-level sensors
Misidentification ~2 Barcode, radiofrequency identification, delta check on results
Lipemic or icteric samples ~1–2 Serum indices

also associated with relevant clinical, organizational, col- analysis develop in a continuous and logical sequence of
lective, and economical issues. The presence of cell-free activities (i.e., check-in, centrifugation, decapping, aliquot-
hemoglobin in the sample attributable to spurious break- ing, analysis and storage), has in fact partially—if not
down of red blood cells while drawing blood or handling totally—hidden the samples from visual scrutiny of medi-
the tube causes a kaleidoscope of biological, chemical, and cal technologists and laboratory technicians, with the inher-
spectrophotometric interferences that globally jeopardize ent risk that unsuitable samples are left unidentified and
the reliability of test results, especially hemostasis testing, thereby loaded and processed. Considering that phleboto-
potassium, aminotransferases, lactate dehydrogenase, and mists and nurses have a varying perception on whether a
cardiospecific troponins.12–14 When the concentration of sample is suitable for testing at the time of blood with-
cell-free hemoglobin in the sample exceeds manufacturer-, drawal (e.g., Stauss et al18 recently reported that it is more
instrument-, and analyte-specific thresholds, test results likely that a sample is unsuitable when the phlebotomist
should be suppressed, and a second—it is hoped suitable— thinks it is suitable and vice versa), innovative technical
sample should be recollected. This obviously results in tools should be put forward and integrated within labora-
increased costs due to the need for drawing a second blood tory automation to replace visual scrutiny before blood
tube, delay in reporting results to the referring clinicians, specimens are analyzed. A variety of technological advances
and frequent complaints from nurses and other health care have recently improved the automatic detection of the most
personnel.15 On the other hand, when left unidentified or common sources of nonconformity. One of the former solu-
unmanaged, hemolyzed samples are indeed a latent hazard tions has been the introduction of liquid-level sensors in the
that can seriously affect patient safety, so a standardized instruments’ needle to limit aspiration errors in samples
procedure for identification and management is needed, with low to insufficient volume or detect the presence of
indeed.16 microclots or bubbles. Another important tool is the imple-
mentation of barcode or even RFID (radiofrequency identi-
fication) technology, which now allows a bidirectional
Hemolyzed Samples and communication between instrumentation and laboratory
Laboratory Automation information system (LIS), to prevent identification and
As we move forward toward novel models of laboratory transcriptional errors (Table 1).
automation and integration, new challenges emerge. The Regardless of these breakthroughs, reliable detection
complexity and the radical changes that have occurred in and management of unsuitable specimens remain the major
laboratory organization have often culminated in the inte- tasks in laboratory diagnostics, and their identification,
gration and/or consolidation of analytical and preanalytical especially of those plagued by spurious hemolysis, remains
instrumentation. In particular, “closed systems” are gener- rather challenging in “closed” automation systems (Figure
ally referred to as packages of preanalytical automation 1). The recognition of hemolyzed samples has been tradi-
lines, analyzers, and postanalytical instruments that are tionally accomplished by visual inspection and by direct
more expensive than open systems, can hardly be modified comparison with standardized color-coded scales to deter-
after installation, and are characterized by poor flexibility. mine the degree of hemolysis and thereby roughly estimate
The unquestionable benefits of this type of automation the content of cell-free hemoglobin. Although largely per-
can be mostly found in complete replacement of manual, fectible due to the high arbitrariness of judgment,19,20 this
potentially dangerous, error-prone steps with automated technique has nevertheless represented the “gold standard”
activities that require negligible contribution from labora- for decades. Together with the technical innovations previ-
tory staff, increased throughput, and reduced turnaround ously discussed, the new generation of analytical and pre-
time (TAT). There is clear evidence, however, that the analytical workstations is increasingly equipped with
adoption of these solutions has not eased the identification systems that can systematically and automatically test for a
of some preanalytical problems.17 The implementation of broad series of interferences, the so-called serum indices,
continuous-flow automation, where sample processing and which also include the hemolysis index (HI).21

Downloaded from jla.sagepub.com by guest on February 16, 2016


186 Journal of Laboratory Automation 18(2)

Figure 1. Identification of hemolyzed specimens in “closed” continuous-flow automation.

Some differences exist in the way the HI is assessed and Advantages and Limitations of the
reported, but most instruments can provide a qualitative or Hemolysis Index
quantitative measure of cell-free hemoglobin in the sample,
which can be further compared with specific thresholds that There are some obvious advantages in the widespread
would help determine if the sample is suitable or not for adoption of the HI in routine laboratory practice (Table 2).
testing. Although the assessment of this type of interference These include (a) overcoming visual inspection and arbi-
by using multiple wavelengths may be preferable,22 this trary judgment about sample quality; (b) transmission of
approach is unavailable on most automated analyzers, so the HI to the LIS, where the data are safely archived for
measurement of the HI is usually based on monitoring of both forensic reasons (i.e., complaints from the wards) and
serum or plasma absorbance at a bichromatic wavelength, practical reasons (i.e., potential inclusion in the laboratory
most commonly between 570 and 600 nm (the serum indi- report); and (c) assessment of sample quality in high-
ces for lipemia and icterus are instead typically monitored volume clinical laboratories that use continuous-flow auto-
at bichromatic wavelengths between 660–700 and 480–505 mation with integration of preanalytical and analytical
nm, respectively).23 By solving a set of predefined equa- workstations, where automatic/systematic monitoring of
tions, the HI is finally calculated and expressed as either a serum indices represents the only suitable approach to
quantitative value (e.g., cell-free hemoglobin in arbitrary detect unsuitable specimens in “real time,” before they are
units or g/L) or a qualitative index (e.g., from “0” to “+”). released from conveyor belts. Another important advantage
The medical technologist can also customize the cutoffs for of the HI is that it can be used as a surrogate to judge trans-
generating specific alerts. A recent evaluation of the HI on port time, transport temperature and centrifugation effi-
seven different automated clinical chemistry instruments ciency, and assess phlebotomists’ performance, wherein
has documented excellent performances in terms of impre- individual phlebotomists or phlebotomy centers encounter-
cision (i.e., intra-assay coefficient of variation between ing a high burden of spuriously hemolyzed specimens can
0.1% and 2.7%) and interlaboratory agreement.24 Specific be objectively identified and further subjected to technical
external quality assessment (EQA) schemes based on ship- (e.g., replacement of the material used for drawing blood)
ment of frozen materials can also be settled to harmonize or educational (e.g., specific training) interventions.26 The
and monitor interlaboratory performance.25 routine use of the HI, however, has some drawbacks. First,

Downloaded from jla.sagepub.com by guest on February 16, 2016


Lippi and Plebani 187

Table 2. Advantages and Potential Limitations of Routine automation and integrated preanalytical-analytical systems
Implementation of the Hemolysis Index in Clinical Laboratories because of the minor handling of specimens, prevention of
Designed around Continuous-Flow Automation. mislabeling,29 and possibility of relieving medical technol-
Advantages ogists and medical laboratory technicians of repetitive
• Overcoming visual inspection and arbitrary judgment about
activities such as transporting specimens and loading ana-
sample quality lyzers.30 Besides detailed analysis of workflow to make
•A  utomatic transmission of hemolysis degree to the laboratory automation as efficient as possible, the process
laboratory information system (LIS) also requires a straightforward knowledge of the potential
•A  ssessment of sample quality in high-volume clinical sources of poor-quality specimens as well as the best solu-
laboratories where preanalytical and analytical workstations tions to identify and manage this foremost preanalytical
are integrated problem. Continuous-flow automation indeed represents a
• S urrogate measure to judge phlebotomists’ performance remarkable improvement for laboratory organization, which
Limitations requires systematic monitoring of sample quality. Special
• Unclear impact on the turnaround time consideration hence should be given to routine implementa-
• Not validated for diagnostic purposes tion of serum indices in clinical laboratories equipped with
• Should not be used for adjusting test results
continuous-flow automation.

Declaration of Conflicting Interests


we know of no study that has established how the system- The authors declared no potential conflicts of interest with respect
atic measurement of HI on all samples would affect TAT. to the research, authorship, and/or publication of this article.
This may be relevant in terms of throughput and efficiency
since the HI is performed as an add-on test in some auto- Funding
mated instrumentations. Moreover, although the HI is often The authors received no financial support for the research, author-
provided by manufacturers also for immunoassays, it is ship, and/or publication of this article.
typically part of the chemistry (colorimetric) analyzer,
which would hence lead to inefficiencies when the degree References
of hemolysis in samples received only for immunoassay 1. Zaninotto, M.; Plebani, M. The “Hospital Central Labora-
has to be checked with a necessary transit of the specimens tory”: Automation, Integration and Clinical Usefulness. Clin.
through the chemistry module(s). It is then noteworthy that Chem. Lab. Med. 2010, 48, 911–917.
this measure has not been validated for diagnostic purposes 2. Brunetti, M.; Pregno, S.; Schünemann, H.; Plebani, M.; Trenti,
(e.g., for diagnosing and monitoring hemolytic anemia or T. Economic Evidence in Decision-Making Process in Labo-
for assessing blood substitutes), and it can only be used for ratory Medicine. Clin. Chem. Lab. Med. 2011, 49, 617–621.
obtaining information on sample quality. Although specific 3. Lippi, G.; Plebani, M.; Favaloro, E. J.; Trenti, T. Laboratory Test-
software might be available for automatic correction of test ing in Pharmacies. Clin. Chem. Lab. Med. 2010, 48, 943–953.
results carried out on hemolyzed specimens according to 4. Plebani, M.; Lippi, G. Is Laboratory Medicine a Dying Profes-
the degree of hemolysis, and some laboratories have actu- sion? Blessed Are Those Who Have Not Seen and Yet Have
ally incorporated the HI into their reporting algorithms, the Believed. Clin. Biochem. 2010, 43, 939–941.
HI must not be used for adjusting test results. We have 5. Carraro, P.; Plebani, M. Errors in a Stat Laboratory: Types
recently proven that the mechanical injury of erythrocytes and Frequencies 10 Years Later. Clin. Chem. 2007, 53, 1338–
and other blood cells that may arise during collection, han- 1342.
dling, transportation, or storage of blood specimens does 6. Plebani, M.; Carraro, P. Mistakes in a Stat Laboratory: Types
not occur homogeneously (i.e., it is largely dependent on and Frequency. Clin. Chem. 1997, 43, 1348–1351.
the individual erythrocyte fragility),27 nor is the intracellu- 7. Plebani, M.; Lippi, G. Closing the Brain-to-Brain Loop in
lar content of several molecules (especially ions, enzymes, Laboratory Testing. Clin. Chem. Lab. Med. 2011, 49, 1131–
hemoglobin, and other proteins) identical and hence suffi- 1133.
ciently predictable among different subjects, which would 8. Plebani, M.; Lippi, G. Improving the Post-Analytical Phase.
make the use of corrective formulas based on the HI of the Clin. Chem. Lab. Med. 2010, 48, 435–436.
sample unreliable and even misleading.28 Finally, manufac- 9. Lippi, G.; Chance, J. J.; Church, S.; Dazzi, P.; Fontana, R.;
turers often provide hemolysis limits for their assays and Giavarina, D.; Grankvist, K.; Huisman, W.; Kouri, T.; Palicka,
assume that the laboratory will take notice of the HI of that V.; et al. Preanalytical Quality Improvement: From Dream to
assay. It is therefore the laboratories’ duty to comply with Reality. Clin. Chem. Lab. Med. 2011, 49, 1113–1126.
that stated use. 10. Lippi, G.; Guidi, G. C.; Mattiuzzi, C.; Plebani, M. Preanalyti-
Most clinical laboratories, especially those with the larg- cal Variability: The Dark Side of the Moon in Laboratory Test-
est volume of samples, clearly profit from total laboratory ing. Clin. Chem. Lab. Med. 2006, 44, 358–365.

Downloaded from jla.sagepub.com by guest on February 16, 2016


188 Journal of Laboratory Automation 18(2)

11. Lippi, G.; Guidi, G. C. Risk Management in the Preanalytical 21. Plebani, M.; Lippi, G. Hemolysis Index: Quality Indicator or
Phase of Laboratory Testing. Clin. Chem. Lab. Med. 2007, 45, Criterion for Sample Rejection? Clin. Chem. Lab. Med. 2009,
720–727. 47, 899–902.
12. Lippi, G.; Blanckaert, N.; Bonini, P.; Green, S.; Kitchen, S.; 22. Harboe, M. A Method for Determination of Hemoglobin in
Palicka, V.; Vassault, A. J.; Plebani, M. Haemolysis: An Over- Plasma by Near-Ultraviolet Spectrophotometry. Scand. J.
view of the Leading Cause of Unsuitable Specimens in Clini- Clin. Lab. Invest. 1959, 11, 66–70.
cal Laboratories. Clin. Chem. Lab. Med. 2008, 46, 764–772. 23. Golf, S. W.; Schneider, S.; Friemann, E.; Temme, H.;
13. Lippi, G.; Fontana, R.; Avanzini, P.; Aloe, R.; Ippolito, L.; Roka, L. Correction of Catalytic Activities of Aspartate-
Sandei, F.; Favaloro, E. J. Influence of Mechanical Trauma Aminotransferase, Lactate-Dehydrogenase, Acid-Phosphatase
of Blood and Hemolysis on PFA-100 Testing. Blood Coagul. and Potassium Concentration in Hemolytic Plasma by Deter­
Fibrinolysis 2012, 23, 82–86. mination of Hemoglobin Concentration with Direct Spectro-
14. Lippi, G.; Avanzini, P.; Dipalo, M.; Aloe, R.; Cervellin, G. Influ- photometry. J. Clin. Chem. Clin. Biochem. 1985, 23, 585.
ence of Hemolysis on Troponin Testing: Studies on Beckman 24. Lippi, G.; Luca Salvagno, G.; Blanckaert, N.; Giavarina, D.;
Coulter UniCel Dxl 800 Accu-TnI and Overview of the Litera- Green, S.; Kitchen, S., Palicka, V.; Vassault, A. J.; Plebani, M.
ture. Clin. Chem. Lab. Med. 2011 Sep 6. [Epub ahead of print] Multicenter Evaluation of the Hemolysis Index in Automated
15. Lippi, G.; Plebani, M.; Di Somma, S.; Cervellin, G. Hemo- Clinical Chemistry Systems. Clin. Chem. Lab. Med. 2009, 47,
lyzed Specimens: A Major Challenge for Emergency Depart- 934–939.
ments and Clinical Laboratories. Crit. Rev. Clin. Lab. Sci. 25. Miller, W. G.; Erek, A.; Cunningham, T. D.; Oladipo, O.;
2011, 48, 143–153. Scott, M. G.; Johnson, R. E. Commutability Limitations Influ-
16. Simundic, A. M.; Topic, E.; Nikolac, N.; Lippi, G. Hemolysis ence Quality Control Results with Different Reagent Lots.
Detection and Management of Hemolysed Specimens. Bio- Clin. Chem. 2011, 57, 76–83.
chem. Med (Zagreb). 2010, 20, 154–159. 26. Lippi, G.; Plebani, M. Reduction of Unsuitable Specimens: A
17. Lam, C.W.; Jacob, E. Implementing a Laboratory Automation More Radical and Comprehensive Approach Is Needed. Clin.
System: Experience of a Large Clinical Laboratory. J. Lab. Chim. Acta 2011, 412, 400.
Autom. 2012, 17, 16–23. 27. Lippi, G.; Mercadanti, M.; Aloe, R.; Targher, G. Erythrocyte
18. Stauss, M.; Sherman, B.; Pugh, L.; Parone, D.; Looby-
Mechanical Fragility Is Increased in Patients with Type 2 Dia-
Rodriguez, K.; Bell, A.; Reed, C. R. Hemolysis of Coagula- betes. Eur. J. Intern. Med. 2012, 23, 150–153.
tion Specimens: A Comparative Study of Intravenous Draw 28. Lippi, G.; Avanzini, P.; Pavesi, F.; Bardi, M.; Ippolito, L.; Aloe,
Methods. J. Emerg. Nurs. 2012, 38, 15–21. R.; Favaloro, E. J. Studies on In Vitro Hemolysis and Utility
19. Glick, M. R.; Ryder, K. W.; Glick, S. J.; Woods, J. R. Unreli- of Corrective Formulas for Reporting Results on Hemolyzed
able Visual Estimation of the Incidence and Amount of Tur- Specimens. Biochem. Med (Zagreb). 2011, 21, 297–305.
bidity, Hemolysis, and Icterus in Serum from Hospitalized 29. Lippi, G.; Sonntag, O.; Plebani, M. Appropriate Labelling
Patients. Clin. Chem. 1989, 35, 837–839. of Blood Collection Tubes: A Step Ahead towards Patient’s
20. Simundic, A. M.; Nikolac, N.; Ivankovic, V.; Ferenec-Ruzic, Safety. Clin. Chem. Lab. Med. 2011 Sep 30. [Epub ahead of
D.; Magdic, B.; Kvaternik, M.; Topic, E. Comparison of Visual print]
vs. Automated Detection of Lipemic, Icteric and Hemolyzed 30. Plebani, M.; Lippi, G. Darwinian Evolution or Regression?
Specimens: Can We Rely on a Human Eye? Clin. Chem. Lab. The Fate of Laboratory Professionals. Clin. Chem. Lab. Med.
Med. 2009, 47, 1361–1365. 2010, 48, 1367–1368.

Downloaded from jla.sagepub.com by guest on February 16, 2016

You might also like