Professional Documents
Culture Documents
REGULAR ARTICLE
Evaluation of the Automated
Coagulation Analyzer SYSMEX CA 6000
Peter Quehenberger, Stylianos Kapiotis, Sylvia Handler, Katharina Ruzicka and Wolfgang Speiser
From the Clinical Institute of Medical and Chemical
Laboratory Diagnostics, University of Vienna, A-1090 Vienna, Austria.
D
ue to increasing demands for coagulation
decreased factor VII). Interference studies with
testing, automated coagulation analyzers
lipemic, icteric, and hemolytic samples showed just
minor influences of these abnormal sample charac- have been developed for screening large
numbers of samples in coagulation laboratories.
Automated coagulation analyzers are expected to
Abbreviations: PT, prothrombin time; aPTT, activated partial be labor saving, quick, and safe, and should ensure
thromboplastin time; TT, thrombin time; AT, antithrombin; CV,
coefficient of variation; f.c., final concentration. accuracy and precision. In the last years new types
Corresponding author: W. Speiser, Clinical Institute of Medical of coagulation analyzers were developed that are
and Chemical Laboratory Diagnostics, General Hospital Vienna, capable of performing not only standard coagula-
Währinger Gürtel 18-20, A-1090 Vienna, Austria. Tel.: 143 (1)
40400 5369; Fax: 143 (1) 4036688; E-mail: ,wolfgang.speiser@ tion tests such as prothrombin time (PT), activated
univie.ac.at.. partial thromboplastin time (aPTT), or fibrinogen,
0049-3848/99 $–see front matter 1999 Elsevier Science Ltd. All rights reserved.
PII S0049-3848(99)00069-9
66 P. Quehenberger et al./Thrombosis Research 96 (1999) 65–71
but also more sophisticated analysis such as mea- agent holder table (158C) with 28 positions for
surement of single coagulation factors activities. reagents and two fixed positions for rinse solution.
Furthermore, these analyzers are usually equipped
with a photometric detection unit for chromogenic 1.2. Patients, Volunteers, and Blood Sampling
substrate assays.
In the present study we evaluated the new coagu- Citrated blood (Na-citrate, final concentration
lation analyzer SYSMEX CA 6000 (TOA Medical (f.c.) 0.013 M) was taken into siliconized glass tubes
Electronics Co., Kobe, Japan) [1]; the most striking (Vacutainer Becton Dickinson, Meylan Cedex,
features of the analyzer are cap-piercing technique, France). Platelet-poor plasma was prepared by
fully automated sampling, positive identification of centrifugation at 30003g for 20 minutes. Plasma
samples by bar-code detection, reagent cooling to was then assayed immediately or frozen at 2708C
158C, and the option to integrate the machine into until use. Determinations were performed on sam-
a fully automated “advanced coagulation analysis ples from patients whose specimens were sent to
system.” Coagulation tests are performed using a the routine coagulation laboratory of our hospital.
photo-optical clot detection unit that is also suit- For definition of normal ranges and estimation of
able for chromogenic substrate analysis [2]. It was analysis time, samples from 40 apparently healthy
our aim to evaluate the technical characteristics of volunteers (20 females, 20 males; 25–58 years of
SYSMEX CA 6000 to obtain new information (1) age, median 39 years) were used. The following
on the effect of cap piercing on the performance normal ranges (values within the 5th to 95th per-
of the analyzer; (2) on the comparability of values centile) were determined: PT, 11.8–13.5 seconds;
obtained by using photo-optical clot detection with aPTT, 30–41 seconds; thrombin time (TT), ,19 sec-
those measured by using mechanical clot detection, onds; fibrinogen, 2.0–3.5 g/L; antithrombin (AT),
in particular in samples with abnormal optical char- 80–100%; factor V, 74–136%; factor VII, 62–144%;
acteristics (lipemic, icteric, and hemolytic samples); factor VIII, 60–180%; factor IX, 71–150%.
and (3) on the precision of automated single factor
coagulation activity measurements. 1.3. Reagents and Laboratory Methods
100 mL of a 1:20 dilution of the sample; after 60 monitored because of alternating magnetic field.
seconds of incubation at 378C, 100 mL of the chro- Changes in the swing due to fibrin-clot formation
mogenic substrate solution were added. For cali- are taken as coagulation end point. For AT mea-
bration of AT, pooled normal plasma was diluted surement a chromogenic test (measurement at 405
1:20, 1:30, and 1:60 with Owren’s diluent buffer. nm) was used. Determinations of samples used for
The test configurations of PT, aPTT, TT, fibrino- the calibration of various tests were performed in
gen, and AT were similar on SYSMEX CA 6000 duplicate, whereas all the other measurements were
and on STA analyzer (Stago, Asnieres-Sur-Seine, done in single tests. Fibrin split product D-dimer
France). The activities of the coagulation factors was measured by ELISA (Asserachrom D-dimer,
V and VII were measured by mixing 50 mL of 1:10, Stago Diagnostics; normal range: ,400 mg/L).
1:20, and 1:40 plasma dilutions (Owren’s diluent For estimation of practicability in a routine coag-
buffer) with 50 mL factor V (Behringwerke, Mar- ulation laboratory PT, aPTT, TT, fibrinogen, and
burg, Germany) or factor VII-deficient plasma AT measurements in 40 individuals with values
(Dade Diagnostics AG). The mixtures were then within the normal range were performed. In these
incubated for 180 seconds (on STA for 60 seconds) samples a throughput of 123 analysis per hour
at 378C and thereafter 100 mL PT reagent Dade was calculated.
Thromboplastin IS (Dade Diagnostics AG) was
added. The activities of the clotting factors VIII 1.4. Statistical Analysis
and IX were determined by mixing 50 mL of a 1:10,
1:20, and 1:40 plasma dilution (Owren’s diluent For calculating coefficients of variation and regres-
buffer) with 50 mL factor VIII- or factor IX-defi- sion analysis, after Passing/Bablok the statistic pro-
cient plasma (both from Immuno AG, Vienna, gram EVAPAK (Roche Austria) was used [3].
Austria). The mixtures were then incubated for 60
seconds at 378C and thereafter 50 mL aPTT reagent
Dade Actin-FS (Dade Diagnostics AG) were 2. Results
added; after a further incubation of 180 seconds at
378C, coagulation was started by the addition of 2.1. Intraassay Precision
50 mL of 0.025 M CaCl2 (on the STA analyzer,
patient plasma, deficient plasma, and aPTT reagent The intraassay precision of PT, aPTT, fibrino-
were incubated together for 240 seconds before gen, TT, and AT were determined in pooled sam-
the addition of CaCl2). For the calibration of single ples with values within the normal range and in
factor activity determination, dilutions of Coag Cal pathological samples (PT: liver disease, oral antico-
N [Dade Diagnostics AG; 1:1 (5100%), 1:2, 1:4, agulant treatment; aPTT: heparin treatment; TT:
1:8, 1:16, and 1:32 with Owren’s diluent buffer] heparin treatment; fibrinogen: hypo- and hyperfi-
with the respective factor-deficient plasma were brinogenemia; AT: liver disease). Values were ob-
used. All dilutions of samples and plasma for cali- tained from pooled plasma that was divided into
bration were done automatically by the analyzers. 20 separate closed vacutainer tubes. The procedure
All coagulation tests measured on SYSMEX CA was repeated on the following two days. The values
6000 were performed by using a percent detection given in Table 1 represent the mean values of the
method for coagulation end point determination results obtained on these 3 days. Intraassay precision
where the amount of light scattered immediately showed coefficients of variation (CVs) ,5% for
after addition of reagent to the sample is taken as all tests within the normal range. In pathological
0% and the amount of light scattered at comple- samples CVs were also ,5%, except for TT in
tion of the reaction is taken as 100%. Clotting time heparinized samples (5.4%), for fibrinogen in pa-
was determined at the 50% point. AT was de- tients with hypofibrinogenemia (6.36%), and with
termined photometrically by a synthetic substrate fibrinogen levels greater than 7.0 g/L (10.1.%), and
method, using serial measurements of changes for decreased AT due to liver insufficiency
in absorbance at 405-nm wavelength. The STA (9.26%).
analyzer uses mechanical clot detection for coagu- Plasma pools of 10 healthy persons and of factor-
lation assays. The swinging of a small steel ball is deficient patients were used to determine the in-
68 P. Quehenberger et al./Thrombosis Research 96 (1999) 65–71
traassay precision of coagulation factor measure- samples precision values were ,5% for all factors
ment. Each sample was divided into 20 separate except for factor VII (6.12%) (Table 4).
tubes. The mean values of the results calculated To determine the amount of carryover of hepa-
from three dilutions (1:10, 1:20, 1:40) were used rin, we analyzed the aPTT of pooled normal plasma
for calculating intraassay precision; CVs were ,5% run after a sample of pooled normal plasma spiked
for all tests in the normal and in the pathological with three units of unfractionated heparin per mL.
range (Table 2). In 10 alternating determinations of these two sam-
ples no prolongation of the aPTT of normal plasma
2.2. Interassay Precision and Carryover could be detected.
The interassay precision for PT, aPTT, TT, fibrino- 2.3. Correlation between
gen, and AT was tested over a period of 10 days Photo-Optical and Mechanical Clot Detection:
using aliquots of pooled and once-frozen plasma Basal Coagulation Tests, Single Factors,
samples. As shown in Table 3 the interassay preci- and Optical Abnormal Plasma Samples
sion within the normal range was below 5% for all
tests except for fibrinogen (5.62%). Precision for Coefficients of correlation between values ob-
abnormal plasmas was higher than 5% for aPTT tained with SYSMEX CA 6000 and with STA ana-
(11.6%), fibrinogen (11.7%), and AT (10.7%). lyzer were calculated for PT, aPTT, TT, fibrinogen,
For calculating the interassay precision of single and AT in optical normal samples (r values be-
factor activity determinations, values obtained on tween 0.96 and 0.992; n5100) (Table 5). Fibrinogen
10 days from frozen aliquots of patients plasma levels measured with SYSMEX CA 6000 photo-
samples were used. In normal and in pathological optical clot detection in 15 samples with markedly
The determination of clotting single factor activi- the exact interpretation of the measured values;
ties are usually performed by laborious techniques discrepancies between the results of several dilu-
on semiautomated analyzers. Due to several manu- tions may point to the existence of an inhibitor (e.g.,
ally performed pipetting and dilution steps these a lupus anticoagulant). Therefore, the technician can
determinations are of limited precision. In contrast, choose between the possibility of automatic averag-
single factor determinations on SYSMEX CA 6000 ing or obtaining the values for each of the dilutions
were found to be of satisfying precision. The possi- separately shown. Mechanical clot detection (STA)
bility of automatically performing determinations and photo-optical clot detection (SYSMEX CA
on several dilutions of a plasma sample, as it is 6000) showed good correlation concerning coagula-
executed on SYSMEX CA 6000, is important for tion single factor activity determinations.
Photo-optical and mechanical clot detection also for coagulation laboratories, not only with respect
showed satisfactory correlation for the basal coagu- to basal coagulation testing but also concerning the
lation tests prothrombin time, activated partial determination of single factor clotting activities.
thromboplastin time, fibrinogen, and antithrom- The comparison between values obtained with
bin, not only in samples with normal optical quality photo-optical and with mechanical clot detection
but also in optically abnormal specimens (e.g., in showed that determinations on SYSMEX CA 6000
lipemia, hyperbilirubinemia, and hemolysis). We are rather insensitive to interferents. The cap-pierc-
could also show that photo-optical and mechanical ing technique, a new technique in coagulation test-
clot detection do not show discrepancies in fibrino- ing, is of adequate technical quality.
gen measurement using the method described by
Clauss [4], even in samples with markedly elevated
fibrin split products. The only test that showed a References
weak correlation between photo-optical and me-
chanical clot detection was TT. The effect of the 1. Kurosaki T. Overview of the Sysmex CA-6000
interferents on TT measurement on SYSMEX CA automated coagulation analyzer. Sysmex J Int
6000 may be due to the fact that plasma dilution 1996;6:51–5.
within this assay is less pronounced compared to 2. Kasai T, Motoi S. Principles of analysis by the
the other coagulation tests. We found a throughput automated coagulation analyzer CA-6000. Sys-
of 123 analyses per hour in normal samples using mex J Int 1997;1:43–7.
SYSMEX CA 6000. This relatively low value can 3. Passing H, Bablok W. A new biometrical proce-
be increased if the analyzer is integrated into a fully dure for testing the equality of measurement
automated “advanced coagulation analysis system” from two different analytical methods. J Clin
offered by SYSMEX, consisting of an automated Chem Clin Biochem 1983;21:709–20.
centrifugation unit and several CA 6000 units. 4. Clauss A. Gerinnungsphysiologische Schnell-
Our data show that the automated coagulation methode zur Bestimmung des Fibrinogens. Acta
analyzer CA 6000 is a precise and reliable analyzer Haemat 1957;17:237–46.