You are on page 1of 6

Science

Assessment of the Reliability of the Sysmex


XE-5000 Analyzer to Detect Platelet Clumps
Jennifer Hawkins, DO,1,2 Gene Gulati, PhD,2* Guldeep Uppal, MD,2 Jerald Gong, MD2
Laboratory Medicine 47:3:189-194

DOI: 10.1093/labmed/lmw016

ABSTRACT
Objective: Platelet counts generated by automated analyzers on size distribution [PAD] flag and platelet-clumps [CLP] flag) were
blood specimens that contain platelet clumps are often inaccurate and determined.
require verification by blood-smear review. In this study, we assessed
the reliability of the Sysmex XE-5000 instrument to detect platelet Results: The respective values for the sensitivity, specificity, efficiency,
clumps. and PPV were 42%, 83%, 63%, and 1% for the PAD flag and 57%, 99%,
78%, and 37% for the CLP flag. The NPV was 100%.
Method: We reviewed automated complete blood count (CBC)
results and the findings of the microscopic review of corresponding Conclusion: The overall reliability of the CLP flag is superior than that of
blood smears of 600 blood specimens specifically selected from the PAD flag but there is room for further improvement.
the routine laboratory workload. The sensitivity, specificity,
efficiency, positive predictive value (PPV), and negative predictive Keywords: Sysmex XE-5000, platelet clumps, assessment, automated
value (NPV) of its 2 platelet-associated flags (abnormal platelet- platelet counts, reliability, CBC

Clinical laboratories worldwide use the XE-5000, an result of the flagged parameter by other means. One such
automated hematology analyzer from Sysmex Corporation, parameter of clinical significance is the automated platelet
to perform complete blood counts (CBCs) and differential count, which is often unreliable and consequently
leukocyte counts (DIFFs) on ethylenediamineteraacetic acid unreportable if the blood specimen contains platelet clumps.
(EDTA)–anticoagulated blood specimens. The overall
reliability of the results generated by this analyzer has been We have assessed the reliability of the Sysmex XE-5000
assessed and found to be acceptable for clinical use.1 analyzer to detect platelet clumps by determining the
However, the CBC results generated on some of the blood sensitivity, specificity, efficiency, positive predictive value
specimens are flagged by the analyzer for verification of the (PPV), and negative predictive value (NPV) of its 2 relatively
frequently generated platelet-associated flags, namely, the
abnormal platelet size distribution (PAD) flag and the platelet
Abbreviations
clumps (CLP) flag. The PAD flag is generated by the analyzer
CBCs, complete blood counts; DIFFs, differential leukocyte counts; EDTA, based on automated analysis of the platelet histogram-
ethylenediamineteraacetic acid; PPV, positive predictive value; NPV,
negative predictive value; PAD, platelet size distribution flag; CLP, plate- associated parameters. The CLP flag is generated based on
let clumps; IMI, immature cell information; NRBC, nucleated red blood automated analysis of the specific area (upper right area of
cells; NA, not applicable, VPA, vortexed prior to analysis; PCIC, platelet the ghost population) in the DIFF, immature cell information
count inaccurate due to clumps; NPLT, platelet count not obtainable due
to clumps; NWC, normal with clumps; IWC, increased with clumps; DWC, (IMI), and nucleated red blood cell (NRBC) scattergrams.2
decreased with clumps; RBT, if clinically indicated, a reliable platelet These flags may be generated by the analyzer individually or
count may be obtained from a specimen collected in a blue-top (citrated) together, with or without accompanying white blood cell and/
tube
or red blood cell-associated flags.
1
Department of Pathology, Marshall University, 2Department of
Pathology, Anatomy, and Cell Biology, Sydney Kimmel Medical College, Our study focused on the selected platelet-associated flags
Thomas Jefferson University, Philadelphia, Pennsylvania
and their variant combinations. The latter category included 2
*To whom correspondence should be addressed. scenarios, one represented by PAD and/or CLP and the other
gene.gulati@jefferson.edu

C American Society for Clinical Pathology, 2016. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com
V 189
Science

by PAD and CLP together. In our laboratory, which processes, fibrin strands by microscopic review, for the presence of
on average, 634 specimens for CBC per day (based on 1 PAD and CLP flags. We defined significant clumping and the
week of data), the number of specimens flagged in either or presence of fibrin strands as the degree of clumping and/or
both flag categories averages 48 (7.6%) per day. Among these the presence of fibrin strands that rendered the automated
flags, the PAD is the most prevalent, accounting for 5.7% of platelet count unreliable and consequently unreportable.
total specimens; the CLP flag accounts for 1.4%. The Similarly, for determination of specificity, we retrospectively
remaining 0.5% account for generation of the 2 flags (PAD and reviewed the CBC results of an equal number of blood
CLP) together by the analyzer. specimens (ie, 100), for which smears did not reveal any
platelet clumps and/or fibrin strands via microscopic review,
for the presence of PAD and CLP flags. The sensitivity was
defined as the percentage of specimens, among the
Materials and Methods morphologically positive cases, for which test results
revealed either or both of the analyzer-generated flags. The
In this study, we included automated CBC results generated specificity was defined as the percentage of specimens
by the analyzer and morphologic findings of microscopic revealing neither of the 2 flags among the morphologically
review of the corresponding blood smears of a total of 600 negative cases. The efficiency was defined as the
EDTA-anticoagulated blood specimens, specifically selected percentage of specimens among this group of 200
as described later herein, from the routine workload of our specimens that were correctly identified by the analyzer as
laboratory during a 5-month period. The analyzer-generated having true positive or true negative results.
acceptable platelet counts of these specimens spanned a
range of 5  103 per lL to 1102  103 per lL. The
categorical distribution of the platelet counts among the Positive Predictive Value (PPV)
entire group of specimens included 35.5% decreased (ie,
To determine the PPV, blood smears of 300 additional
<140  103/lL, of which 72.4% were <100  103/lL), 6.2%
specimens flagged by the analyzer for PAD (n ¼ 100) or CLP
increased (ie, >440  103/lL), and 58.3% normal (ie, within
(n ¼ 100), or both flags together (n ¼100) were reviewed for
the range of 140-440  103/lL).
the presence of platelet clumps and fibrin strands. PPV was
Our laboratory routinely performs CBCs on 2 Sysmex XE- defined as the percentage of flagged specimens that, via
5000 analyzers, which are calibrated and quality controlled microscopic review, revealed significant platelet clumping
per manufacturer instructions.2 The platelet count is routinely and/or the presence of fibrin strands.
performed via the impedance method by this analyzer. The
optical method for platelet count is also available on this
analyzer, but we did not consider it to be appropriate or Negative Predictive Value (NPV)
necessary for our study. Blood smears are routinely prepared
The NPV was defined as the percentage of specimens not
and stained with a Romanowsky stain (Wright stain) on the
revealing significant clumping or any fibrin strands by
Sysmex SP-1000 instrument. One of the coauthors (G.G.)
microscopic review. To determine the NPV, we reviewed
reviewed all smears microscopically for the presence of
blood smears of an additional 100 specimens, which had not
platelet clumps, fibrin strands, and large/giant platelets. The
been flagged by the analyzer for any of the platelet-associated
smear review process was unbiased, despite the fact that
flags, for the presence of platelet clumps and fibrin strands.
the reviewer was aware of the flagging status of each
specimen included in the determination of PPV and NPV by
virtue of the specimen-selection process.

Results
Sensitivity, Specificity, and Efficiency
Sensitivity
For determination of sensitivity, we retrospectively reviewed
the CBC results of 100 blood specimens, for which smears Among the specimens that tested morphologically positive,
revealed significant platelet clumping and/or the presence of the analyzer flagged 42 for the PAD and 57 for the CLP,

190 Lab Medicine 2016;47:3;189–194 www.labmedicine.com


190 DOI: 10.1093/labmed/lmw016
Science

Table 1. Sensitivity of Sysmex XE- 5000 in Table 2. Specificity of Sysmex XE- 5000 in
Detecting Platelet Clumpsa Detecting Platelet Clumpsa
Variable Specimens, No. Sensitivity Variable Specimens, No. Specificity

Platelet-Associated Flagged Not Flagged Platelet-Associated Flagged Not Flagged


Flag(s) by Analyzer by Analyzer Flag(s) by Analyzer by Analyzer
Platelet abnormal 42 58 42.0% Platelet abnormal 17 83 83.0%
distribution (PAD) distribution (PAD)
Platelet clump (CLP) 57 43 57.0% Platelet clump (CLP) 1 99 99.0%
PAD and/or CLP 73 27 73.0% PAD or CLP 18 82 82.0%
PAD and CLP 26 74 26.0% PAD and CLP 18 82 82.0%
a a
Data from specimens that tested positive for platelet clumps and/or fibrin strands Data from specimens that tested negative for platelet clumps and fibrin strands via
via blood smear (n ¼ 100). The Sysmex XE-5000 instrument is manufactured by blood smear (n ¼ 100). The Sysmex XE-5000 instrument is manufactured by
Sysmex Corporation. Sysmex Corporation.

yielding sensitivities of 42.0% and 57.0%, respectively


(Table 1). The sensitivity increased to 73.0% when a Table 3. Efficiency of the Sysmex XE 5000 in
specimen was considered flagged for platelet clumps if the
Detecting Platelet Clumpsa
PAD flag and/or the CLP flag was/were generated by the Platelet-Associated No. of Specimens Efficiency
Flag(s) Correctly Flagged By
analyzer. It decreased, however, to 26.0% when a specimen
Analyzer
was considered flagged for platelet clumps if the PAD and
Platelet abnormal distribution 125 62.5%
CLP flags were generated together by the analyzer. The
(PAD)
reported platelet estimates of this group of 100 specimens Platelet clump (CLP) 156 78.0%
ranged from 10 (10.0%) that were decreased with clumps to PAD and/or CLP 155 77.5%
76 (76.0%) that were normal with clumps to 14 (14.0%) that PAD and CLP 108 54.0%
a
were increased with clumps. n ¼ 200. The Sysmex XE-5000 instrument is manufactured by Sysmex Corporation.

Specificity 63.0% and 78.0% (Table 3). The efficiency decreased to 54%
when a specimen was considered flagged for platelet clumps
Among the specimens that tested morphologically negative, if both flags were generated together by the analyzer.
the analyzer flagged 17 for PAD and 1 for CLP, yielding However, this value remained relatively unchanged at 77.5%
specificities of 83.0% and 99.0%, respectively (Table 2). The when a specimen was considered flagged for platelet clumps
specificity dropped a bit, to 82.0%, when a specimen was if either of the 2 flags or both flags was/were generated by
considered flagged for platelet clumps, if either of the 2 flags the analyzer. The categorical distribution of the platelet
or both flags were generated by the analyzer. The analyzer- counts among this group of 200 specimens included the
generated platelet counts of this group of 100 specimens following: decreased, 46 (23.0%); increased, 23 (11.5%); and
ranged from 41  103 per lL to 1102  103 per lL. Based on normal, 131 (65.5%).
our laboratory’s reference range of 140 to 440  103 per lL,
the categorical distribution of the platelet counts among this
group included the following: decreased (ie, <140  103/lL), Positive Predictive Value
36 (36.0%; 28 of 36 [77.8%] were <100  103/lL); increased
(ie, >440  103/lL), 9 (9.0%); and normal (ie, within the range The respective PPVs for the PAD and CLP flags for detecting
of 140-440 103/lL), 55 (55.0%) . platelet clumps were 1.0% and 37.0% (Table 4). The PPV
increased minimally to 39.0% when a specimen was
considered flagged for platelet clumps if both flags were
Efficiency generated together by the analyzer. The analyzer-generated
platelet counts of this group of 300 specimens ranged from
The respective efficiencies for the PAD and CLP flags, as 5  103 per lL to 780  103 per lL. The categorical
determined from this data set from 200 specimens, were distribution of the platelet counts among this group included

www.labmedicine.com Lab Medicine 2016;47:3;189–194 191


DOI: 10.1093/labmed/lmw016
Science

The variant combination of PAD and CLP flags together


Table 4. Positive Predictive Value (PPV) for
minimally increased the PPV to 39.0%, at the expense of a
Platelet-Associated Flags Generated by Sysmex
XE- 5000 in Detecting Platelet Clumps reduction in sensitivity to 26.0% and efficiency to 54.0%, with
specificity remaining unchanged at 82.0%, compared with
Variable No. of Specimens, Result PPV
for Clumps and Fibrin the corresponding values for the CLP flag.

Platelet-Associated Flag(s) Positive Negative These low levels of sensitivity and efficiency for the
a
Platelet abnormal distribution (PAD) 1 99 1.0% combination of PAD and CLP point to its inferiority to the
Platelet clump (CLP)a 37 63 37.0% CLP flag in detecting platelet clumps. Among the 3
PAD and CLPa 39 61 39.0%
categories of platelet counts (decreased, increased, and
a
n ¼ 100. The Sysmex XE-5000 instrument is manufactured by Sysmex Corporation.
normal), the flagging rate was fairly similar between the
decreased (47.3%) and normal categories (49.3%), with the
the following: decreased, 142 (47.3%; with 108 of the 142 remaining 3.3% representing the increased category.
[76.0%] having values of < 100  103/lL); increased, 10 However, among the flags, the PAD was the most prevalent
(3.3%); and normal, 148 (49.3%). (50.0%) in the decreased platelet count category, and the
CLP flag was the most prevalent in the increased (70.0%)
and normal (50.0%) platelet-count categories. The
Negative Predictive Value incidence of combined flagging (PAD and CLP together)
was also fairly similar between the decreased (36.6%) and
The NPV for the blood specimens not flagged by the normal (31.1%) platelet count categories and a bit lower
analyzer for any of the platelet-associated flags was 100%. (20.0%) in the increased platelet count category. From these
The analyzer-generated platelet counts of this group of 100 findings, one might infer that the analyzer generates
specimens ranged from 18  103 per lL to 771  103 per lL. platelet-associated flags at all levels of platelet counts but
The categorical distribution of the platelet counts among this the type of generated flag varies somewhat with the level of
group included the following: decreased, 25 (25.0%; with 14 the platelet count. The PAD flag, which has the lowest PPV
of the 25 [56.0%] having a value of less than 100  103/lL); of 1.0%, was more often associated with decreased platelet
increased, 4 (4.0%); and normal, 71 (71.0%). counts. In contrast, the CLP flag and the combination of the
2 flags (PAD and CLP together), which yield the relatively
A summary of all data, including the sensitivity, specificity, higher PPVs of 37.0% and 39.0%, respectively, were more
efficiency, PPV, and NPV, is presented in Table 5. often associated with the normal and decreased platelet
counts.

To our knowledge, there are no published study results on


Discussion the Sysmex XE-5000 with which we can compare our
findings. However, comparable study results on other
Among the 2 platelet-associated flags and their variant automated hematology analyzers have been published
combinations, which we assessed for their ability to detect during the past several years. The most comparable
platelet clumps, we found the CLP to be relatively more observations are from a study on the evaluation of
reliable, with sensitivity of 57.0%, specificity of 99.0%, PPV of automated platelet counts generated by the XE-2100, an
37.0%, and overall efficiency of 78.0%; the NPV was 100%. earlier model of automated hematology analyzer from
The PAD flag, compared with the CLP flag, had much lower Sysmex Corporation, the same manufacturer as the XE-
values: sensitivity, 42.0%; specificity, 83.0%; efficiency, 5000.3 For detection of platelet clumps, this study reported
62.5%; and PPV, 1.0%. When a specimen was considered a sensitivity of 1% for the PAD flag, 22.5% for the CLP flag,
correctly identified by the analyzer for the presence of and 35.4% for the CLP þ PAD flag when the blood
platelet clumps because it generated a PAD flag, a CLP flag, specimens collected in lavender-top tubes were used to
or both (ie, the PAD and/or CLP combination), the sensitivity perform the CBCs. The sensitivity for the CLP flag was
increased to 73.0%, but the specificity decreased to 82.0%, higher (57.1%) when blood specimens collected in
and efficiency remained relatively unchanged at 77.5%, Microtainer tubes (Becton, Dickinson and Company) were
compared with the corresponding values for the CLP flag. used to perform CBCs.3

192 Lab Medicine 2016;47:3;189–194 www.labmedicine.com


192 DOI: 10.1093/labmed/lmw016
Science

Table 5. Summary of Data on Sensitivity, Specificity, Efficiency, PPV, and NPV for Platelet-Associated
Flags Generated by the Sysmex XE-5000 in Detecting Platelet Clumpsa
Platelet-Associated Flags Sensitivityb Specificityb Efficiencyc PPVd NPVb
Platelet abnormal distribution flag (PAD) 42.0% 83.0% 62.5% 1.0% NA
Platelet clump flag (CLP) 57.0% 99.0% 78.0% 37.0% NA
PAD and/or CLP 73.0% 82.0% 77.5% NA NA
PAD and CLP 26.0% 82.0% 54.0% 39.0% NA
Neither PAD nor CLP NA NA NA NA 100%
NA, not applicable, PPV, positive predictive value, NPV, negative predictive value.
a
The Sysmex XE-5000 instrument is manufactured by Sysmex Corporation.
b
n ¼ 100.
c
n ¼ 200.
d
n ¼ 100/flag(s).

Table 6. Platelet-Associated Smear Review Findings and Suggested Course of Action


Platelet-Associated Smear Review Finding Suggested Course of Action
Significant platelet clumping, and automated platelet count Vortex the specimen to break up the clumps and rerun the specimen to determine
appears unreliable the CBC. If clumps break up, as judged by the smear review, report the platelet
count from the vortexed specimen and append to it the comment “VPA”
or
If clumps do not break on vortex or if vortex is not feasible, modify the platelet count
to “PCIC” (if already released) or “NPLT” (if not already released), report a platelet
estimate as “NWC,” “IWC,” or “DWC” from the smear, and attach a comment “RBT”.
Many fibrin strands Cancel the CBC (reason: unsuitable specimen) and ask for an appropriately
collected new specimen.
Rare platelet clumps and/or rare fibrin strands and automated Report automated platelet count and attach the comment “with rare
platelet count appears reliable clumps/fibrin strands”.
No platelet clumps and no fibrin strands; automated platelet count Report the automated platelet count and attach to it the comment
appears reliable “verified by smear review”.
Few giant platelets, and automated platelet count appears reliable Report the automated platelet count and attach to it the comment
“with few giant platelets”.
Many giant platelets, and automated platelet count appears falsely lowDo not report automated platelet count but report an estimated platelet
count from the smear; attach the comment “estimated”.
Platelet satellitosis, and automated platelet count appears unreliable Do not report automated platelet count but report an estimated platelet
count from the smear if feasible; attach the comment “platelet
satellitosis present”.
Platelet satellitosis, and automated platelet count appears reliable Report automated platelet count and attach to it the comment
“platelet satellitosis present, result may be affected”.
VPA, vortexed prior to analysis; PCIC, platelet count inaccurate due to clumps; NPLT, platelet count not obtainable due to clumps; NWC, normal with clumps; IWC, increased with
clumps; DWC, decreased with clumps; RBT, if clinically indicated, a reliable platelet count may be obtained from a specimen collected in a blue-top (citrated) tube; CBC, complete
blood count

Our findings regarding the XE-5000, based on the CBCs previously reported efficiency in the range of 33.0% to
performed on specimens collected primarily (98.0%) in 67.0% for platelet-clump flagging by 3 different
lavender-top tubes, show an improvement in the reliability hematology analyzers (Beckman Coulter LH750 [Beckman
of its automated platelet flagging system compared with Coulter, Inc.], Sysmex XE-2100 [Sysmex Corporation], and
that of the XE-2100, but the degree of improvement does Siemens ADVIA 120 [Siemens Healthcare GmbH]). 4 A
not approach the desirable level of at or close to 100% for study on evaluation of the automated flagging system of
each analyzed statistical parameter. The overall efficiency an older hematology analyzer (GenS) from a different
of 78.0%, which we observed for the CLP flag generated manufacturer (Beckman Coulter, Inc) reported a sensitivity
by the XE-5000, also compares favorably with the of 12.0% to 44.0%, and specificity, efficiency, NPV, and

www.labmedicine.com Lab Medicine 2016;47:3;189–194 193


DOI: 10.1093/labmed/lmw016
Science

PPV of more than 95.0% each for the CLP flag.5 Our recent past count. We chose these criteria to ensure patient
findings on the XE-5000 reveal better sensitivity and safety in a cost-effective way, with the premise that the
similar or slightly lower NPV, PPV, and specificity likelihood of the patient being subjected to unnecessary
compared with that of the GenS. platelet transfusion, additional diagnostic work-up, and/or
other undue measures is low in such circumstances. Our
Another relatively recently published study reported laboratory’s policy on reporting the results of the platelet
sensitivity in the range of 40.4% to 82.8% for the detection of count based on the findings of the blood smear review, as
platelet clumps/fibrin strands in blood smears by the outlined in Table 6, may be used as a guide by others for
CellaVision DM96 (Siemens Healthcare GmbH), which is an potential implementation in their laboratories.LM
automated image analysis system.6 These 2 systems may
not be considered completely comparable because one (the
CellaVision DM96) is semiautomated and uses stained blood Acknowledgments
smears for performing DIFFs and platelet scans, and the
other (the Sysmex XE-5000) is fully automated and uses We thank Nandita Patel, B.Sc., and Vikas Patel, B.Sc., for
whole blood to perform the same analyses. their help in obtaining the slides and the data from the ana-
lyzer that we needed for the study.
All of these previously published reports and our findings
point towards the continued need for improvement in the
reliability of the automated flagging system to detect platelet
clumps. To maximize the sensitivity while maintaining the References
specificity, it is highly desirable that users of the XE-5000 1. Arneth BM, Menschikowki M. Technology and new fluorescence flow
analyzer include 1 or more additional criteria along with the cytometry parameters in hematological analyzers. J Clin Lab Anal 2015;
29: 175–183.
CLP flag to verify the automated platelet count by smear
2. Sysmex XE-5000 Automated Hematology Analyzer: Instructions for use
review and that the manufacturer of the analyzer continue its (North American Edition) Kobe, Japan: Sysmex Corporation; 2006-2008.
efforts to enhance the automatic detectability of platelet 3. Gulati G, Cote S, Behling E, Kocher W. Optimization of criteria for verifica-
clumps to the desirable level of at or close to 100%. The tion of automated platelet counts generated by the Sysmex XE-2100
hematology analyzer. Lab Medicine 2009; 40(9): 535–539.
manufacturer of the analyzer recommends verification of the
4. Sandhaus LM, Osei ES, Agrawal NN, Dillman CA, et al. Platelet counting by
flagged results by other means before reporting; the the Coulter LH 750, Sysmex XE 2100, and Advia 120: A comparative ana-
lysis using the RBC/platelet ratio reference method. Am J Clin Pathol 2002;
verification method generally used by the clinical laboratories
118 (2): 235–241.
is blood-smear review. The clinical laboratory at our 5. Gulati GL, Kocher W, Schwarting R, Hyland LJ, et al. An assessment of the
institution routinely verifies automated platelet counts by Coulter Gen•S automated flagging system. Lab Medicine 2001; 32(6):
310–317.
smear review if they are accompanied by the CLP flag, less
6. Gulati G, Uppal G, Dulau-Florea A, Gong J. Detection of platelet clumps on
than 100,000 per lL on initial encounter, and/or reveal delta peripheral blood smears by CellaVision DM96 system and microscopic re-
failure with a 50% or greater drop compared with the most view. Lab Medicine 2014; 45(4): 368–371.

194 Lab Medicine 2016;47:3;189–194 www.labmedicine.com


194 DOI: 10.1093/labmed/lmw016

You might also like